The EcoSense® pH100 Instrument is warranted for one year from date of purchase by the end user
against defects in materials and workmanship. pH100 probes and cables are warranted for six months
from date of purchase by the end user against defects in material and workmanship. Within the
warranty period, YSI will repair or replace, at its sole discretion, free of charge, any product that YSI
determines to be covered by this warranty.
To exercise this warranty, write or call your local YSI representative, or contact YSI Customer Service
in Yellow Springs, Ohio. Send the product and proof of purchase, transportation prepaid, to the
Authorized Service Center selected by YSI. Repair or replacement will be made and the product
returned, transportation prepaid. Repaired or replaced products are warranted for the balance of the
original warranty period, or at least 90 days from date of repair or replacement.
Limitation of Warranty
This Warranty does not apply to any YSI product damage or failure caused by: (i) failure to install,
operate or use the product in accordance with YSI's written instructions; (ii) abuse or misuse of the
product; (iii) failure to maintain the product in accordance with YSI's written instructions or standard
industry procedure; (iv) any improper repairs to the product; (v) use by you of defective or improper
components or parts in servicing or repairing the product; or (vi) modification of the product in any way
not expressly authorized by YSI.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED,
INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE. YSI's LIABILITY UNDER THIS WARRANTY IS LIMITED TO REPAIR OR
REPLACEMENT OF THE PRODUCT, AND THIS SHALL BE YOUR SOLE AND EXCLUSIVE
REMEDY FOR ANY DEFECTIVE PRODUCT COVERED BY THIS WARRANTY. IN NO EVENT
SHALL YSI BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL
DAMAGES RESULTING FROM ANY DEFECTIVE PRODUCT COVERED BY THIS
WARRANTY.
CONTACT INFORMATION
YSI Inc. • 1725 Brannum Lane • Yellow Springs OH, 45387
800-897-4151 • 937-767-7241 • Fax: 937-767-1058
Email: ecosense@ysi.com
RECOMMENDED SPARE PARTS LIST ........................................................ 7
2
GENERAL INTRODUCTION
The model pH100 is one of three instruments in the EcoSense product line from YSI. The
pH100 is a precision tool that measures pH, mV and temperature. A built-in microprocessor
stores, calculates and compensates for all parameters related to pH determinations including pH
electrode temperature characteristics, electrode slope deviations and buffer solutions.
This unit has a splash-proof IP65 case. The mechanical touch keys are highly reliable with
tactile and audio feedback. This meter uses one 9V battery. Re-calibration is not required when
power is restored.
The front of the meter has a large LCD that displays pH or mV and temperature simultaneously
along with user prompts and mode indicators. The unit prompts the user through calibration and
measurement procedures.
An AUTOLOCK feature for both pH and mV measurements enables the unit to automatically
sense the end point and "lock" the display to indicate the end point value of a measurement. The
pH100 can also be used in non-AUTOLOCK mode. AUTOLOCK and user prompts help
eliminate most errors in determining pH and mV values, resulting in precise, repeatable, errorfree measurements.
The model pH100 is available with pH, mV, ORP and ATC (Automatic Temperature
Compensation) probes. Other features include electrode offset recognition, electrode slope
recognition, electrode efficiency display, built-in buffer coefficients, automatic or manual
temperature compensation, long battery life, and 50/60 Hz AC noise rejection. This meter is
universal, user-friendly, for field, industrial and laboratory applications.
INITIAL INSPECTION
Carefully unpack the unit and accessories, and inspect for shipping damages. Compare
received parts with materials listed on the packing list. Notify YSI immediately of any damage or
missing parts. Save all packing materials until satisfactory operation is confirmed.
SPLASH RESISTANCE
Though the pH100 meter is housed in a watertight case, DO NOT use it underwater. The splashresistant case prevents permanent damage to the unit if accidentally dropped into non-corrosive
solutions. Follow these steps immediately if the unit is immersed in any solution:
1. Rinse unit carefully with distilled water. After rinsing and drying, inspect and clean connectors to
remove all contaminants that may affect probe connections.
2. Wait for the unit and probe to dry completely before resuming operation.
3. If the unit does not function correctly after steps 1 and 2, call YSI for possible repair or
replacement (see Warranty).
BATTERY INSTALLATION
An initial display of “BAT” on the LCD indicates approximately one hour
of battery life for unit operation within specifications. Replace battery
when “BAT” appears on the LCD. (See Figure 1.)
To replace battery, remove the two battery cover screws and battery
cover and o-ring. Replace the 9V battery. Replace battery cover and
o-ring (align the o-ring properly to insure a good seal) and fasten the
two battery cover screws for the splash-resistant feature.
3
Figure 1.
Battery Installation
KEY FUNCTIONS OF THE MODEL pH100
1. : Turns the unit ON or OFF. The pH calibration values are not erased when the unit is
turned off. The unit powers up in the same status as when it was turned off. When the unit is
not in use, turn it off to save battery life. For long term storage, remove the batteries.
2. MODE: Selects display mode. Press MODE to sequentially display pH-AUTOLOCK , mVAUTOLOCK , pH, and mV. Calibration values are not affected by changing display mode.
3. STAND and SLOPE Keys: Used for dual-point pH calibration of the unit. Press and hold
STAND while turning on the power to change buffer sets.
4. Δ and ∇ Keys: Press to enter temperature values in manual (MAN) mode. These keys are
inoperative when operating in ATC mode.
5. MEA./EFF.: Press to release the unit from AUTOLOCK status when operating in pHAUTOLOCK or mV-AUTOLOCK mode. Press and hold for 5 seconds to display the
electrode efficiency.
6. ESC: Press to clear the unit when an error signal appears; it clears all calibration values
stored in internal memory. To prevent accidental clearing during normal use, the key does
not respond unless pressed and held for 2 seconds. The key responds only when an error
message displays.
When ESC is pressed, all LCD elements light. After about 2 seconds, the unit enters pH-
AUTOLOCK mode. “AUTOLOCK” displays and “STAND” begins to flash indicating the need
for calibration. Press ESC only when errors occur that require unit re-calibration.
THE LCD DISPLAY
1. WAIT: Displays while unit waits for a stable
reading or end point sensing.
2. BAT: Low battery indicator.
3. ATC/MAN: “ATC” displays if an ATC probe
is connected. Otherwise, “MAN” displays.
4. STAND/SLOPE: “STAND” or “SLOPE”
remains steady if the parameter has been
calibrated. If either one has not been
calibrated, it flashes.
5. AUTO: Autolock mode indicator.
6. HOLD: Indicates a reading is frozen during
Autolock mode.
7. EFF%: Displays when the user views electrode efficiency. It is recommended to replace the
electrode when efficiency is less than 75%.
8. pH/mV : Unit and mode indicators.
9. Main display for pH, mV and probe efficiency values.
10. °C: Temperature display.
1
WAITHOLD
2
3
4
5
T
A
B
STAND
SLOPE
AUTO
ATC
MAN
EFF%
mV
pH
C
Figure 2. LCD Display
6
7
8
9
10
OPERATIONAL PROCEDURES
Buffer Set Selection
The pH100 has two buffer sets: 7.00, 4.01, 10.01 pH and 6.86, 4.00, 9.18 pH. The factory
default is buffer set 7.00, 4.01, and 10.01. To change the buffer set, turn off the unit and place
the sensor in buffer 7.0. Next, turn the unit on while pressing and holding the STAND key.
Continue pressing the STAND key until the unit beeps. If the unit is uncalibrated and in pH
mode, it displays “7.00” if the first set is active, and “6.86” if the second set is active.
4 5
pH Calibration
The pH100 uses a 2-point calibration. The first point must be a 6.86/7.00 buffer, and the second
either a 4.00/4.01 or 9.18/10.01. These buffers can be purchased from a YSI representative.
1. Turn the unit on. Connect the pH electrode to the BNC connector and the ATC/Temp probe
to the ATC/Temp connector of the unit; “ATC” displays. Press MODE until “pH” displays.
Autolock may be on or off as desired.
2. Place the pH and ATC/temp probes into the first buffer solution (either 7.00 or 6.86). Allow
temperature readings to stabilize, then press and hold“STAND” for 3 secondsto calibrate.
If AUTOLOCK is off, the first point has been calibrated. If AUTOLOCK is on, “WAIT” flashes
until the unit detects a stable reading. Once the unit calibrates the first point, “SLOPE”
flashes.
NOTE: If no temperature probe is connected, adjust the temperature reading to that of the
first buffer using the Δ or ∇ keys (0.0 to 60ºC) BEFORE pressing “STAND”.
3. Rinse the pH and ATC/temp probes in distilled water, then place into the second buffer
solution (either 4.01/4.00 or 10.01/9.18). Allow temperature readings to stabilize, then press
“SLOPE” to calibrate. If AUTOLOCK is off, the second point has been calibrated. If
AUTOLOCK is on, “WAIT” flashes until the unit detects a stable reading. Once the unit
calibrates the second point, the unit beeps twice andboth “STAND” and “SLOPE” display
steadily.
NOTE: If no temperature probe is connected, adjust the temperature reading to that of the
first buffer using the Δ or ∇ keys (0.0 to 60ºC) BEFORE pressing “SLOPE”.
4. The unit calculates and compensates for the pH electrode slope deviation corresponding to
the values of the two calibration buffers. The unit is now dual-point calibrated and ready for
measurements. After calibration, press and hold MEA. /EFF. for about 5 seconds to display
the new electrode efficiency.
Using the model pH160 Electrode Simulator
The model pH160 Electrode Simulator can be used to confirm proper instrument calibration. To
use the simulator:
1. Install the 9V battery provided.
2. Attach the pH160 to the pH100. Turn both units on. The pH160 has a small switch to the
right of the pH buttons.
3. In pH measurement mode, press one of the pH buffer simulator buttons on the pH160. The
corresponding pH value should appear on both screens.
Note: Calibration with the pH simulator calibrates only the instrument - NOT the instrument and
probe. For best accuracy, calibrate the pH instrument and probe together using buffer solutions.
pH Measurements
To take pH measurements, “STAND” and “SLOPE” must display steadily, indicating the unit is
dual-point calibrated and ready for measurements. If “STAND” and “SLOPE” are blinking,
perform a pH calibration before taking measurements.
1. Press MODE to enter pH mode with AUTOLOCK on or off as desired. For inherently
unstable samples, the unit will not AUTOLOCK. Turn AUTOLOCK off in this case.
2. Rinse the pH electrode and/or ATC/temp probe with distilled water and immerse in the
sample to be measured. Remove any air bubbles trapped around the probe by shaking or
stirring the probe. Allow the pH and/or temperature to stabilize. If no ATC/temp probe is
connected, “MAN” displays, indicating manual temperature compensation. Set unit to display the
sample temperature by pressing the Δ and ∇ keys (-10.0 to 120°C). If an ATC/temp probe is
connected “ATC” displays along with the sample temperature.
3. If AUTOLOCK is off, the pH value of the sample displays on the screen. If both pH and
temperature readings are stable, take a reading. If AUTOLOCK is on, press MEA./EFF.
“WAIT” flashes until the unit determines a stable pH reading.
Temperature Measurements
The model pH100 can measure temperature independently with the ATC/Temp probe without
using the pH electrode. Place the ATC/Temp probe in the media to be measured. The measured
temperature displays.
mV Measurements
1. Connect the optional combination mV electrode to the unit. Press MODE to enter mV mode
with AUTOLOCK on or off as desired. For inherently unstable samples, the unit will not
AUTOLOCK. Use mV mode with AUTOLOCK off in this case.
2. Rinse electrode with distilled water and immerse it in sample to be measured. If AUTOLOCK is
off, the mV value of the sample will be displayed on the screen. If AUTOLOCK is on, press
MEA./EFF. “WAIT” flashes until the unit determines a stable mV reading.
TROUBLESHOOTING
DISPLAY
OvEr/Undr
MAIN
Er 1
Er 2
Er 3
POSSIBLE CAUSE CORRECTIVE ACTION
• pH electrode offset is greater/less than
+/-1.5 pH.
• STAND was pressed before the
electrode and ATC/Temp probe settle to
within ±1.5 pH of the buffer value.
• pH electrode is faulty
• pH electrode slope is off by more than
30% of ideal slope.
• SLOPE was pressed before the
electrode and ATC/Temp probe settled
to within 30% of the buffer value.
• Buffer 4.00, 4.01, 9.18 or 10.01 is not
correct.
• Temperature is out of the 0.0 to 60.0 °C
range.
• Measured pH is out of the 16.00/-2.00
pH range.
• Measured mV is out of the 1250/-2000
mV range.
• Measured temperature is out of the 10/120 °C range.
• Replace the buffer and/or the
pH electrode. Press ESC.
• Allow sufficient time for the
electrode and ATC/Temp probe
to stabilize.
• Return for service.
• Check that the correct buffer is
used and that the electrode
slope is not off by more than
30% from the theoretical slope.
• Allow sufficient time for the
electrode and ATC/Temp probe
to stabilize.
• Replace the buffer and/or the
pH electrode. Press ESC.
For the latest version of this manual, visit www.ysiecosense.com
Revision E • March 2008
7
GARANTIE
L'appareil EcoSense® pH100 est garanti pour une période d'un an, à compter de la date d'achat par
l'utilisateur final, contre tout défaut matériel et de fabrication. Les sondes et les câbles de l' pH100
sont garantis pour une période de six mois, à compter de la date d'achat par l'utilisateur final, contre
tout défaut matériel et de fabrication. Pendant la période de garantie, YSI s'engage à réparer ou à
remplacer, gratuitement et à sa discrétion, tout produit qu'YSI peut établir comme étant couvert par la
garantie.
Pour faire valoir cette garantie, écrivez ou appelez votre représentant YSI ou contactez le Service
clientèle d'YSI à Yellow Springs, Ohio, États-Unis. Envoyez le produit et son justificatif d'achat en port
payé au Centre de service homologué sélectionné par YSI. La réparation ou le remplacement seront
effectués et le produit vous sera retourné en port payé. Les produits réparés ou remplacés sont
garantis jusqu'à expiration de la période de garantie originale ou pour au moins 90 jours, à compter
de la date de réparation ou de remplacement.
Limitation de garantie
Cette garantie ne s'applique pas aux produits YSI endommagés ou présentant des
dysfonctionnements pour les raisons suivantes : (i) installation, exploitation ou utilisation du produit
d'une façon non conforme aux instructions écrites d'YSI ; (ii) abus ou mésusage du produit ; (iii)
manquement à l'entretien du produit conformément aux instructions écrites d'YSI ou aux procédures
industrielles normales ; (iv) réparation non conforme du produit ; (v) utilisation par vous de pièces ou
de composants défectueux ou non conformes lors de l'entretien ou de la réparation du produit, ou ;
(vi) modification du produit d'une façon non expressément autorisée par YSI.
CETTE GARANTIE REMPLACE TOUTES LES AUTRES GARANTIES, EXPRESSES OU
INDUITES, Y COMPRIS LES GARANTIES DE COMMERCIABILITÉ OU D'ADAPTATION À UN
USAGE PARTICULIER. LA RESPONSABILITÉ D'YSI SELON LES TERMES DE CETTE
GARANTIE SE LIMITE À LA RÉPARATION OU AU REMPLACEMENT DU PRODUIT,
CONSTITUANT VOTRE SEUL ET UNIQUE RECOURS POUR TOUT PRODUIT DÉFECTUEUX
COUVERT PAR CETTE GARANTIE. YSI NE POURRA EN AUCUN CAS ÊTRE TENU
RESPONSABLE DE DOMMAGES SPÉCIAUX, INDIRECTS, ACCIDENTELS OU
CONSÉCUTIFS RÉSULTANT DE L'UTILISATION DE TOUT PRODUIT DÉFECTUEUX
COUVERT PAR CETTE GARANTIE.
COMMENT NOUS CONTACTER
YSI Inc. • 1725 Brannum Lane • Yellow Springs OH, 45387, États-Unis
800-897-4151 • 937-767-7241 • Télécopie : 937-767-1058
E-mail : ecosense@ysi.com
LISTE DES PIÈCES DÉTACHÉES RECOMMANDÉES................................ 8
2
INTRODUCTION GÉNÉRALE
Le modèle pH100 est un des trois instruments de la ligne de produits EcoSense de YSI. Le
modèle pH100 est un instrument de précision mesurant le pH, les mV et la température. Un
microprocesseur incorporé stocke, calcule et compense tous les paramètres relatifs aux
déterminations liées au pH, y compris les caractéristiques de température de l'électrode de
détection du pH, les écarts de pente de l'électrode et les solutions tampons.
L'appareil est doté d'un boîtier de protection contre les éclaboussures conforme à la norme
IP65. Les touches mécaniques sont très fiables et produisent une réaction tactile et sonore. Cet
appareil de mesure utilise une pile de 9 V. Aucun réétalonnage n'est nécessaire lorsque
l'alimentation électrique est rétablie.
L'avant de l'appareil dispose d'un écran à cristaux liquides de grande taille affichant
simultanément les mesures de pH, de millivoltage et de température, ainsi que les invites
destinées à l'utilisateur et les indicateurs de mode. L'appareil émet des invites destinées à
l'utilisateur lors des procédures d'étalonnage et de mesure.
Une fonctionnalité de verrouillage automatique AUTOLOCK, applicable aux mesures de pH et
de millivoltage, permet à l'appareil de détecter automatiquement le point de virage et de
verrouiller l'affichage pour indiquer la valeur du point de virage de la mesure. Le pH100 peut
également être utilisé alors que le mode de verrouillage AUTOLOCK est désactivé. La
fonctionnalité de verrouillage automatique et les invites permettent d'éliminer la plupart des
erreurs de détermination des valeurs de pH et de millivoltage et d'obtenir des mesures précises,
fiables et pouvant être répétées.
Le modèle pH100 est disponible avec des sondes à pH, millivoltage, potentiel redox et
compensation automatique de la température (CAT). Parmi les autres fonctionnalités, on peut
compter la reconnaissance du décalage de l'électrode, la reconnaissance de la pente de
l'électrode, l'affichage de l'efficacité de l'électrode, des coefficients de tampon incorporés, la
compensation automatique ou manuelle de la température, une pile longue durée et un dispositif
d'atténuation du bruit C.A. à 50 ou 60 Hz. Cet appareil est convivial et particulièrement souple
dans les applications sur le terrain, industrielles et en laboratoire.
INSPECTION INITIALE
Déballez soigneusement l'appareil et les accessoires et vérifiez qu'ils n'ont pas été
endommagés lors de l'expédition. Comparez les pièces reçues aux matériaux répertoriés dans
le bordereau d'emballage. Notifiez immédiatement YSI s'il s'avère que des pièces sont
endommagées ou manquantes. Mettez de côté les matériaux d'emballage jusqu'à ce que le
fonctionnement correct de l'appareil soit confirmé.
PROTECTION CONTRE LES ÉCLABOUSSURES
Bien que le modèle pH100 soit enfermé dans un boîtier étanche, NE l'utilisez PAS sous l'eau. Le
boîtier résistant aux éclaboussures prévient les dommages permanents si l'appareil est
accidentellement immergé dans des solutions non corrosives. Suivez immédiatement ces étapes
si l'appareil est immergé dans une solution quelconque :
1. Rincez soigneusement l'appareil avec de l'eau distillée. Après le rinçage et le séchage,
inspectez et nettoyez les connecteurs en vue d'éliminer tout contaminant pouvant affecter les
connexions de la sonde.
2. Attendez que l'appareil et la sonde soient parfaitement secs avant de reprendre les
opérations.
3. Si l'appareil ne fonctionne pas correctement après les étapes 1 et 2, appelez YSI en vue d'une
réparation ou d'un remplacement éventuels (voir la Garantie).
3
INSTALLATION DE LA PILE
Lorsque l'écran à cristaux liquides affiche pour la première fois « BAT », il
reste environ une heure de fonctionnement sur pile selon les
spécifications. Remplacez la pile lorsque l'indication « BAT » s'affiche sur
l'écran. (Voir Figure 1.)
Pour remplacer la pile, enlevez les deux vis du compartiment ainsi que le
couvercle et le joint torique. Remplacez la pile de 9 V. Replacez le
couvercle et le joint torique (alignez le joint correctement afin d'assurer
une bonne étanchéité) et revissez les deux vis du compartiment
pour conserver une bonne résistance aux éclaboussures.
Figure 1.
Installation de la pile
FONCTIONNALITÉS CLÉS DU MODÈLE pH100
1. : Met l'appareil hors ou sous tension. Les valeurs d'étalonnage du pH ne sont pas
effacées lorsque l'appareil est mis hors tension. L'appareil démarre dans le même état que
lorsqu'il a été mis hors tension. Lorsque l'appareil n'est pas utilisé, mettez-le hors tension
pour économiser la pile. Enlevez les piles pour un entreposage prolongé.
2. MODE : Permet de sélectionner le mode d'affichage. Appuyez plusieurs fois sur MODE
pour afficher successivement pH-AUTOLOCK (verrouillage automatique de la mesure du
pH), Mv-AUTOLOCK (verrouillage automatique de la mesure du millivoltage), pH et mV. Les
valeurs d'étalonnage ne sont pas affectées par le changement de mode d'affichage.
3. Touches STAND et SLOPE : Utilisées pour l'étalonnage en deux points du pH de l'appareil.
Maintenez la touche STAND enfoncée tout en mettant l'appareil sous tension pour changer
les jeux de tampons.
4. TouchesΔet ∇ : Appuyez sur ces touches pour entrer les valeurs de température en mode
manuel (MAN). Ces touches sont inopérantes en mode CAT.
5. MEA./EFF. : Appuyez sur cette touche pour désactiver le verrouillage automatique lorsque
l'appareil est en mode pH-AUTOLOCK ou mV-AUTOLOCK. Maintenez la touche enfoncée
pendant 5 secondes pour afficher l'efficacité de l'électrode.
6. ESC : Appuyez sur cette touche pour effacer la mémoire de l'appareil lorsqu'un message
d'erreur apparaît. Toutes les valeurs d'étalonnage stockées en mémoire interne sont
effacées. Afin d'éviter un effacement accidentel lors d'un fonctionnement normal, la touche
doit être maintenue enfoncée pendant 2 secondes pour être active. La touche n'est active
que lorsqu'un message d'erreur est affiché.
Lorsque la touche ESC est enfoncée, tous les éléments de l'écran à cristaux liquide
s'allument. Au bout d'environ 2 secondes, l'appareil se met en mode pH-AUTOLOCK. La
mention « AUTOLOCK » s'affiche et la mention « STAND » commence à clignoter,
indiquant qu'un étalonnage est nécessaire. Appuyez sur la touche ESC uniquement lorsque
surviennent des erreurs nécessitant un réétalonnage.
ÉCRAN À CRISTAUX LIQUIDES
1. WAIT : S'affiche lorsque l'appareil
attend une lecture stable ou la détection
d'un point de virage.
2. BAT : Indicateur de pile déchargée.
3. ATC/MAN : La mention « ATC »
s'affiche si une sonde à compensation
automatique de la température est
connectée. Sinon, la mention « MAN »
est affichée.
4. STAND/SLOPE : Les mentions
« STAND » et « SLOPE » restent
1
WAITHOLD
2
3
4
5
T
A
B
STAND
SLOPE
AUTO
ATC
MAN
Figure 4. Écran à cristaux liquides
4 5
EFF%
mV
pH
6
7
8
9
C
10