YSI ODO200 User Manual [en, es, fr]

User Manual
EcoSense® ODO200
Dissolved oxygen & temperature instrument
USER MANUAL
English, Français, Español, Português
Item #606335 Rev B, September 2014 For the latest version of this manual, visit ysi.com
©2014 YSI Incorporated.
The YSI logo and EcoSense are registered trademarks of YSI Incorporated.
CONTENTS
Warranty ................................................................................................i
Introduction .........................................................................................1
Getting Started ....................................................................................1
Initial Inspection .....................................................................1
Precautions .............................................................................1
The Probe .........................................................................1
Initial Setup ............................................................................2
Preparing the Probe ........................................................2
Battery Installation ...........................................................3
Key Pad ....................................................................................4
Main Display ..........................................................................5
Operational Procedures - Taking Measurements .............................6
Measurement Modes - % or mg/L ........................................7
Sampling modes - Continuous or Manual ...........................7
Saving, Viewing and Deleting Data ......................................7
Calibration ..............................................................................8
Temperature .....................................................................8
Dissolved Oxygen ...........................................................8
Maintenance and Troubleshooting ...................................................9
Sensor Maintenance - Dissolved Oxygen ............................9
Cleaning the Sensor Cap ................................................9
Sensor Cap Replacement ............................................10
Entering Sensor Cap Coefficients into the EcoSense
ODO200 .........................................................................11
Rehydrating the Sensor Cap ......................................12
Sensor Maintenance - Temperature ...................................12
Sensor Storage .....................................................................12
Short-term Storage ......................................................12
Long-term Storage ........................................................13
Troubleshooting ...................................................................13
Specifications .....................................................................................14
Accessories / Part Numbers .............................................................15
Contact Information and Service .....................................................16
Ordering and Technical Support ........................................16
Service Information ..............................................................16
Recycling ...............................................................................16
Appendix A-DO% Calibration Values ..............................................17
WARRANTY
The EcoSense® ODO200 Instrument, probe, cable and sensor cap are warranted for one year from date of purchase by the end user against defects in materials and workmanship, exclusive of batteries and any damage caused by defective batteries. Within the warranty period, YSI will repair or replace, at its sole discretion, free of charge, any product that YSI determines to be covered by this warranty.
To exercise this warranty, write or call your local YSI representative, or contact YSI Customer Service in Yellow Springs, Ohio at +1 937 767-7241, 800-765-4974 or visit YSI.com for a Product Return Form. Send the product and proof of purchase, transportation prepaid, to the Authorized Service Center selected by YSI. Repair or replacement will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days from date of repair or replacement.
Limitation of Warranty
This Warranty does not apply to any YSI product damage or failure caused by: (i) failure to install, operate or use the product in accordance with YSI’s written instructions; (ii) abuse or misuse of the product; (iii) failure to maintain the product in accordance with YSI’s written instructions or standard industry procedure; (iv) any improper repairs to the product; (v) use by you of defective or improper components or parts in servicing or repairing the product; or (vi) modification of the product in any way not expressly authorized by YSI.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY WARRANT Y OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. YSI’s LIABILITY UNDER THIS WARRANTY IS LIMITED TO REPAIR OR REPLACEMENT OF THE PRODUCT, AND THIS SHALL BE YOUR SOLE AND EXCLUSIVE REMEDY FOR ANY DEFECTIVE PRODUCT COVERED BY THIS WARRANTY. IN NO EVENT SHALL YSI BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES RESULTING FROM ANY DEFECTIVE PRODUCT COVERED BY THIS WARRANTY.
i
THIS PAGE LEFT INTENTIONALLY BLANK
ii
INTRODUCTION
Thank you for purchasing the EcoSense ODO200, a precise handheld instrument that measures temperature and dissolved oxygen in % and mg/L. The EcoSense ODO200 measures dissolved oxygen in water using optical, lifetime luminescence technology and uses a digital signal to send information between the instrument and probe. Key advantages of the EcoSense ODO200 include the elimination of sensor flow dependence (stirring) and sensor warm-up time, greater stability, less frequent calibrations, and the elimination of frequent membrane/ electrolyte changes. This instrument uses one 9V battery.
For additional product specification information, please visit ysi.com or contact Technical Support at 800-897-4151 (+1 937 767-7241) or environmental@ysi.com.
GETTING STARTED
INITIAL INSPECTION
Carefully unpack the instrument and accessories and inspect for damage. Compare received parts with items on the packing list. If any parts are damaged or missing, contact YSI Customer Service at 800­897-4151 (+1 937 767-7241) or the authorized YSI distributor where the instrument was purchased.
PRECAUTIONS
THE PROBE
Sensor caps last longer if stored in a moist environment and kept clean and free of scratches in the paint layer. Erratic readings can result from damaged or dirty caps. If unstable readings occur from large scratches in the paint or dye layer, replace the sensor cap. See the Maintenance and Troubleshooting section of this manual for information on how to clean or replace the sensor cap. The average sensor cap replacement interval is 12 to 18 months, although they may last longer.
Avoid substances that may damage probe materials such as concentrated acid, caustics and strong solvents. Probe materials include urethane (cable), Xenoy (probe body), acrylic (sensor cap) and acetal and stainless steel (probe guard).
1
To prevent the sensor cap from drying out, store the probe in the gray calibration/storage sleeve with a moistened, clean sponge. In the event the sensor cap dries out, rehydrate it following the instructions in the Maintenance and Troubleshooting section of this manual.
INITIAL SETUP
Throughout the manual, the term “probe” refers to the end of the cable where the sensor is located, the term “sensor” refers to the Optical Dissolved Oxygen sensing portion of the cable/probe assembly, and the term “sensor cap” refers to the removable sensing cap that is replaced about once per year (Figure 1).
Sensor
Probe
Figure 1
PREPARING THE PROBE
Sensor Cap
Each EcoSense ODO200 sensor cap has unique calibration coefficients associated with it. These coefficients are labeled on the instruction sheet included with each EcoSense ODO200 cable/probe assembly and replacement ODO200 sensor cap.
Remove the probe/cable assembly from the shipping container and locate the sensor cap instruction sheet. Be sure to save the instruction sheet in case you need to reload these calibration coefficients. New probe/cable assemblies have the sensor cap installed and coefficients loaded into the probe at the factory. When a cable is installed, the coefficients are automatically sent to the instrument.
Preparing the probe for the first time:
1. Remove the probe guard from the probe by turning it counterclockwise.
2. Remove the red storage cap which contains a moist sponge from the end of the probe by pulling it straight off the sensor. Save this to use later for long term storage.
2
3. Reinstall the probe guard by sliding it carefully over the sensor and
then threading it onto the cable/probe assembly with a clockwise rotation (Figure 2).
4. Locate the gray calibration/storage sleeve that was shipped with
your probe/cable assembly. Moisten the sponge in the gray calibration/storage with a small amount of clean water.
5. Slide the calibration/storage sleeve over the probe guard to keep
the probe in a moist environment for storage and calibration (Figure 3). It is important to always keep your sensor in a moist environment so the sensor cap does not dry out.
6. Connect the cable to the instrument by aligning the connectors,
pushing them together and then screwing down the retainer clockwise.
Figure 2 Figure 3
BATTERY INSTALLATION
To install (or replace) the battery, remove the two battery cover screws and battery cover. Replace the 9V battery. Replace the battery cover and o-ring (be sure to align the o-ring correctly to prevent a bad seal) and fasten the two battery cover screws.
An initial display of “LOW BAT” on the LCD indicates approximately one hour of
Figure 4
battery life for unit operation within specifications. Replace battery when “LOW BAT” appears on the LCD.
Battery Disposal
This instrument is powered by a 9 volt battery, which the user must remove and dispose of when the battery no longer powers the instrument. Disposal requirements vary by country and region, and users are expected to understand and follow the battery disposal requirements for their specific locale.
3
KEY PAD
1
2
3
4
Figure 5
Number Key Description
1
Power key. Turns the unit on or off.
5
6
7
8
2, 6
3
4
Up and down arrow key. Used to adjust inputs for salinity correction, calibration and sensor cap coefficients
Calibration key. Press to calibrate the dissolved oxygen sensor.
Cap key. Press to enter new sensor cap coefficients.
4
5
Mode key. Adjusts between measurement modes (%, mg/L), recalling saved data (rCL) and deleting saved data (dEL).
7
Enter key. Used to confirm calibration, to save data from run screen, to select data mode (rCL/dEL), and confirm coefficient entries.
8
Sample/Mode key. Press and hold for 3 seconds to switch between continuous and manual measurement modes. Press and release to restart the measurement when in manual mode. Continuous mode: the instrument will continuously update the dissolved oxygen measurement. Manual mode: the instrument will update the DO reading until reaching a stable measurement. The stable measurement will be ‘locked’ on the display until the user initiates a new measurement reading by pressing the Sample/Mode key.
MAIN DISPLAY
1
2
3
10
9
8
7
6
4 5
5
Number Description
1
2
BAT: Low battery indicator. Continuous or Manual sampling mode indicator.
When in Manual mode, manual will blink until the measurement is stable.
3
4
5
6
CAL: Calibration mode indicator. Data record indicator. Temperature reading in Celsius. mbar: Displays during calibration to prompt user for
true barometric pressure.
7
Main display for dissolved oxygen measurements in either % or mg/L.
8
9
Save, Full, Recall, Each, All, Delete: Instrument’s data storage indicators.
Figure 6
SAL ppt: Displays during calibration when user is prompted for the approximate salinity (in ppt) of the sample to be tested.
10
CAP, K1, K2, K3, K4, K5, KC: Sensor cap coefficient indicators when entering new sensor cap coefficients.
OPERATIONAL PROCEDURES ­TAKING MEASUREMENTS
Press to turn the unit on or off. When turned on, the instrument will perform a self-diagnostic test during which an error message may display. This is normal. After the self-diagnostic test completes, the temperature displays in the lower right and the dissolved oxygen measurement displays in the center of the display (figure 6). The unit is ready for operation. If necessary, perform a 1-point dissolved oxygen calibration. The digital instrument and probe can hold its calibration for many months but should be verified regularly.
Immerse the probe into the sample solution to take a measurement.
When the unit is not in use, turn it off to save battery life. It is not necessary to recalibrate the instrument after turning it off and back on. The instrument has a 30 minute auto shut off feature when not in use.
6
MEASUREMENT MODES - % OR MG/L
This unit provides three distinct measurements: Temperature and Dissolved Oxygen in either % or mg/L.
The dissolved oxygen measurement unit is indicated on the right side of the display. Press Mode to switch between % and mg/L.
SAMPLING MODES - CONTINUOUS OR MANUAL
The ODO200’s sampling mode can be set to Manual or Continuous.
Press and hold the Sample/Mode key for 3 seconds to switch between Continuous and Manual mode. When in Manual mode, press and release Sample/Mode to initiate a new measurement.
Continuous mode: the instrument will continuously update the dissolved oxygen measurement.
Manual mode: the instrument will update the DO reading until reaching a stable measurement. The stable measurement will ‘lock’ on the display until the user initiates a new measurement by pressing the Sample/Mode key. Manual mode conserves battery power.
SAVING, VIEWING AND DELETING DATA
The ODO200 can save 50 data records. When in measurement mode,
press Enter to save a record. The instrument will confirm saving the data by displaying SAVE and the data record number for one second. “Full” is displayed when trying to save data and memory is full. To view saved data, press Mode until Recall (rCL) is displayed and then press Enter. Use the Up or Down arrow keys to review different saved records. Press Mode to escape back to measurement mode.
To delete data records, press Mode while in measurement mode until Delete (dEL) is displayed. Press Enter. “All” will be displayed and blinking. Press the Up or Down arrow key to switch between delete ‘All’ or ‘Each’ options. Select either ‘All’ or ‘Each’ by pressing Enter while that option is displayed.
If ‘All’ is selected, all records will be deleted from memory and ‘None’ will be displayed. Press Mode twice to return to the measurement mode.
7
If ‘Each’ is selected, the Up and Down arrow keys will allow you to scroll through the saved data records. Press Enter to delete the selected record. All records after the deleted record will shift up to keep the records in sequential order. For example, if record 3 is deleted, record 4 will become record 3 and record 5 will become record 4. Press Mode twice to return to the measurement mode.
CALIBRATION
TEMPERATURE
All EcoSense ODO200 cables have a built-in thermistor. Temperature calibration is not available or required. To verify the temperature sensor, compare it to a NIST traceable thermistor by touching the thermistors together and observing the measurements.
DISSOLVED OXYGEN
The ODO200 sensor is an optical luminescent sensor which has greater stability and is less susceptible to calibration drift than traditional electrochemical sensors. This increased stability means the instrument can hold its calibration for many months; however, calibration should be verified regularly for highest data accuracy.
To verify the instrument’s calibration, place the sensor in its calibration environment and check to see that the DO% is reading its calibration value based on the barometric pressure. Refer to Appendix A for the DO% calibration values based on barometric pressure. Calibration Requirements:
The approximate true barometric pressure in millbars (mBar) of the location at the time of calibration.
True Barometric Pressure (mmHg) =
[Corrected BP mmHg] – [2.5 * (Local Altitude in ft. above sea level/100)]
mBar = mmHg *1.333
The approximate salinity of the water to be analyzed. Fresh water’s salinity value is approximate zero parts per thousand (ppt). Seawater has an approximate salinity of 35 ppt.
1. Moisten the sponge in the gray storage sleeve with a small amount of clean water. The moistened sponge creates a 100% water-saturated air environment for the probe, which is ideal
8
for calibration, transport, and storage of the EcoSense ODO200 probe.
2. Make sure there are no water droplets on the sensor cap or
temperature sensor and then install the storage sleeve over the guarded probe. Make sure the DO and temperature sensors are not touching the sponge or immersed in water. Turn the instrument on and wait approximately 5 to 10 minutes for the storage sleeve to become completely saturated and to allow the temperature and dissolved oxygen sensors to stabilize.
3. Press CAL.
4. The display prompts for the local, true barometric pressure in
mBar (millibars). Use the Up and Down Arrow keys to increase or decrease the barometric pressure value, then press Enter.
5. The display will show the current dissolved oxygen measurement
on the main display and the calibration value in the lower right corner. Wait for the DO measurement to stabilize, then press Enter.
6. The display prompts for the approximate salinity of the water
to be analyzed. Use the Up and Down Arrow keys to increase or decrease the salinity compensation value to the salinity value of your sample (between 0 to 40 parts per thousand [ppt]). When the correct salinity displays, press Enter.
7. The unit holds calibration for several months, even when powered
off and on.
MAINTENANCE AND TROUBLESHOOTING
SENSOR MAINTENANCE - DISSOLVED OXYGEN
CLEANING THE SENSOR CAP
The Sensor Cap should be kept clean since some types of fouling can consume or produce oxygen or block the water sample from reaching the sensor which could affect the dissolved oxygen measurements.
To clean the Sensor Cap, gently wipe away any fouling with a lens cleaning tissue that has been moistened with water. Do not use organic solvents to clean the Sensor Cap. Using an organic solvent to clean the Sensor Cap may cause permanent damage to the cap. For example, alcohol will dissolve the outer paint layer and other organic solvents will likely dissolve the dye in the cap. Dissolving or scratching the paint or dye layer will cause erratic dissolved oxygen measurements.
To disinfect the probe, soak the probe in 3% hydrogen peroxide for 15 to 30 minutes and then rinse it with clean water.
9
SENSOR CAP REPLACEMENT
The sensor cap should be replaced about once per year but may last longer. It should also be replaced if it is cracked or damaged (see Troubleshooting section for instructions on how to inspect the cap for damage). The instruction sheet shipped with the replacement ODO sensor cap includes the calibration coefficients specific to your sensor cap.
The replacement ODO sensor cap is shipped in a humidified container. The package should not be opened until ready for use. Once the sensor cap has been installed on the sensor as described below, it is important to keep the sensor in a 100% humid environment. Therefore, the sensor should be stored in either the gray calibration/storage sleeve with a moistened sponge or immersed in water, see Sensor Storage for more information. If the sensor dries out, refer to the Rehydration procedure in this manual.
Refer to Figure 7 below when following the instructions for replacing the cap.
Figure 7
Caution: Avoid touching the sensing end of the sensor cap during the following maintenance procedures.
1. Remove the old sensor cap assembly from the sensor by grasping the probe body with one hand and then rotating the sensor cap counterclockwise until it is completely free. Do not use any tools for this procedure.
2. Inspect the o-ring on the probe for damage. If there is any indication of damage, carefully remove the o-ring and replace it with the new o-ring included with the replacement sensor cap. Do not use any tools to remove the o-ring.
10
3. Ensure that the o-ring installed on the probe is clean. If necessary,
wipe clean with a lint free cloth or replace the o-ring as described in the previous step.
4. Locate the o-ring lubricant included with the new sensor cap.
Apply a thin coat of o-ring lubricant to the installed o-ring. Remove any excess o-ring lubricant from the o-ring and/or probe with a lens cleaning tissue.
5. Remove the new sensor cap from its hydrated container and dry
the inside cavity of the sensor cap with lens cleaning tissue. Next, clean the clear surface of the sensor on the end of the probe with lens cleaning tissue.
6. Using clockwise motion, thread the new sensor cap onto the probe
assembly until it is finger-tight. The o-ring should be compressed between the sensor cap and probe. Do not over-tighten the sensor cap and do not use any tools for the installation process.
7. After installing the new sensor cap, store the sensor in the gray
calibration/storage sleeve with a moistened sponge.
8. Follow the procedures below for entering in the new sensor cap’s
coefficients into the EcoSense ODO200 instrument.
ENTERING SENSOR CAP COEFFICIENTS INTO THE ECOSENSE ODO200
After installing a new sensor cap, connect the probe/cable assembly to the instrument and turn the instrument on. Locate the Calibration Code Label at the top of the instruction sheet and note the six numbers which are listed as K1 through K5 and KC. These six numbers contain the calibration code for this particular sensor cap. Follow these instructions to enter the new calibration coefficients into your instrument:
1. Press and hold the CAP key for 3 seconds. The CAP key is located
in the lower left corner of the keypad.
2. The instrument will prompt for K1 and show 8 alpha/numeric
values with the first value blinking.
3. Use the up and down arrow keys to adjust the first value to match
the first value of the new cap’s K1. Press the Enter key to accept.
4. Continue using the up and down arrow keys to adjust all 8 alpha/
numeric values for K1 pressing the Enter key to confirm each entry.
5. After confirming the last value for K1, the instrument will prompt
you for K2. Follow the instructions in step 3 and 4 to enter and confirm the value for K2. Continue with K3, K4 and K5.
6. After confirming K5, the EcoSense ODO200 will prompt you for KC.
KC has two values. Follow the instructions in step 3 and 4 to enter the coefficients. After confirming KC, the instrument will return to the measurement screen.
11
7. After entering the Sensor Cap Coefficients, perform a 1-point calibration as described in the Calibration Section.
8. Press and hold the CAP key for 3 seconds at any time to exit and cancel the cap coefficient update.
If errors are made in entering the Sensor Cap Coefficients, the instrument will block the update and an error message will appear on the display and the instrument will return to the K1 entry screen. If you see this error message, re-enter the coefficients and check them carefully for correct transcription from the Calibration Coefficient Label.
The coefficient values are hexadecimal. Due to the EcoSense ODO200’s display, 6 and b look very similar. Hexadecimal values on the ODO200 will be displayed as: A, b, C, d, E, F, 0, 1, 2, 3, 4, 5, 6, 7, 8, and 9.
If you continue to get an error message after several entry attempts, contact YSI Technical Support for assistance.
REHYDRATING THE SENSOR CAP
The Sensor Cap must remain in a moist environment; see Sensor Storage for storage recommendations. If the sensor cap is allowed to dry out, it is likely to drift slightly at the beginning of your next study unless it is rehydrated. If the cap dries out, you can rehydrate it by soaking the probe tip with the sensor cap installed in room temperature tap water for 24 hours. After rehydration is complete, recalibrate and be sure to store the probe in a moist environment.
SENSOR MAINTENANCE - TEMPERATURE
You must keep the temperature portion of the sensor free of build up. Other than that, the sensor requires no maintenance. A soft bristle brush, like a tooth brush can be used to scrub the temperature sensor if needed.
SENSOR STORAGE
SHORT-TERM STORAGE
When the ODO sensor is not in use, store it in a moist environment.
For short-term storage (less than 30 days), moisten the sponge in the calibration/storage sleeve with a small amount of clean water and place this over the probe with the sensor cap and sensor guard installed. This will provide a 100% saturated air environment.
12
LONG-TERM STORAGE
For long-term storage (>30 days), remove the battery from the instrument. Moisten the sponge in the calibration/storage sleeve with a small amount of clean water and place this over the probe with the sensor cap and sensor guard installed. Inspect the sponge every 30 days to make sure it is still moist.
Alternatively, you can place the probe with sensor cap directly in a beaker or other container of water, making sure that the water does not evaporate over time.
TROUBLESHOOTING
Erroneous dissolved oxygen readings typically indicate a need to clean the sensor cap, replace the sensor cap, and/or recalibrate the instrument. First, clean the sensor cap following the instructions in the Maintenance section and then perform a calibration. If erroneous readings persist, inspect the sensor cap for damage (cracks, large scratches in paint or dye layer), replace if necessary, enter new coefficients and then attempt to recalibrate the instrument.
If the sensor cap is not damaged, try to rehydrate the sensor cap then recalibrate.
If you are still getting erroneous dissolved oxygen readings, try replacing the sensor cap, enter the new coefficients and then recalibrate. If the erroneous readings continue, contact YSI Technical Support to help determine the next step.
Symptom Possible Solution
Instrument will not turn on, the BAT symbol displays on the screen.
Temperature values display Over or Undr on Run screen.
1. Low battery voltage, replace batteries.
2. Batteries installed incorrectly, check battery polarity.
3. Return system for service.
1. Sample temperature is less than 0° C or more than +50°C. Increase or decrease the sample temperature to bring within the allowable range.
2. Contact YSI Tech Support. Possible temperature sensor failure.
13
Symptom Possible Solution
Instrument will not calibrate dissolved oxygen.
DO readings are inaccurate.
Dissolved Oxygen values display Over or Undr on Run screen.
1. Verify calibration routine - correct barometric pressure, salinity input, and calibration environment.
2. Allow sufficient stabilization time for dissolved oxygen and temperature.
3. Check sensor cap. Clean, rehydrate or replace as necessary.
4. Contact YSI Tech Support.
1. Verify temperature readings are accurate.
2. Sample temperature should be between 0 and 45 ºC, the temperature compensation range for DO mg/L.
3. DO sensor not properly calibrated, recalibrate the sensor.
4. Check sensor cap. Clean, rehydrate or replace as necessary. Recalibrate.
5. Contact YSI Tech Support.
1. Sample dissolved oxygen concentration is more than 20 mg/L or 200%, or less than 0.0 mg/L or 0.0%.
2. Verify temperature readings are accurate.
3. DO sensor not properly calibrated, recalibrate the sensor.
4. Check sensor cap. Clean, rehydrate or replace as necessary. Recalibrate.
5. Contact YSI Tech Support.
SPECIFICATIONS
These specifications represent typical performance and are subject to change without notice. For the latest product specification information, please visit YSI’s website at ysi.com or contact YSI Tech Support.
Parameter Range Resolution Accuracy
Temperature 0 to 50°C* 0.1°C ± 0.3°C
Dissolved
Oxygen
Barometer 500 to 1125
0.0 to 200% air saturation
0.00 to
20.0 mg/L
mBar
1% or 0.1%, user selectable
0.1 or 0.01 mg/L ± 1.5% of reading
1.0 mBar manual entry
14
± 1.5% of reading or ± 1.5% air saturation, whichever is greater
or ± 1.5 mg/L, whichever is greater
Parameter Range Resolution Accuracy
Temperature
Operating
Range
Certifications
Memory 50 data sets
Salinity 0 to 40 ppt, manual entry Battery 9 volt alkaline, 100 hour battery life
0 to 50ºC
RoHs, WEEE, CE, IP-67
in manual sampling mode
* Automatic dissolved oxygen temperature compensation range is 0 to 45°C
ACCESSORIES / PART NUMBERS
Part Number Description
606326 EcoSense ODO Sensor Cap Kit
606327 1 meter (approximately 3.3 feet)
probe and cable assembly*
606328 4 meter (approximately 13 feet)
probe and cable assembly*
606304 10 meter (approximately 33 feet)
probe and cable assembly*
606329
605139
EcoSense ODO200 instrument
Carrying case, small, soft-sided (holds instrument and up to a 4 meter cable/probe assembly
606330 ODO200 hard-sided carrying case with form
fitted foam inserts
*All cables include optical-based dissolved oxygen sensor and temperature sensor.
15
CONTACT INFORMATION AND SERVICE
ORDERING AND TECHNICAL SUPPORT
Telephone: 800 897 4151 (US) +1 937 767 7241 (Globally) Monday through Friday, 8:00 AM to 5:00 ET Fax: +1 937 767 9353 (orders) +1 937 767 1058 (technical support) Email: environmental@ysi.com Mail: YSI Incorporated 1725 Brannum Lane Yellow Springs, OH 45387 USA Web address: ysi.com When placing an order please have the following available:
1.) YSI account number (if available)
2.) Name and phone number
3.) Purchase Order or Credit Card
4.) Model Number or brief description
5.) Billing and shipping addresses
6.) Quantity
SERVICE INFORMATION
YSI has authorized service centers throughout the United States and Internationally. For the nearest service center information, please visit ysi.com and click ‘Support’ or contact YSI Technical Support directly at 800-897-4151.
When returning a product for service, include the Product Return form with cleaning certification. The form must be completely filled out for a YSI Service Center to accept the instrument for service. The form may be downloaded from ysi.com by clicking on the ‘Support” tab, then the Product Return Form button.
RECYCLING
YSI is committed to reducing the environmental footprint in the course of doing business. Even though materials reduction is the ultimate goal, we know there must be a concerted effort to responsibly deal with materials after they’ve served a long, productive life-cycle. YSI’s recycling program ensures that old equipment is processed in an
16
environmentally friendly way, reducing the amount of materials going to landfills.
Printed Circuit Boards are sent to facilities that process and reclaim as much material for recycling as possible.
Plastics enter a material recycling process and are not incinerated or sent to landfills.
Batteries are removed and sent to battery recyclers for dedicated metals.
When the time comes for you to recycle, follow the easy steps outlined at ysi.com.
APPENDIX A-DO% CALIBRATION VALUES
Calibration Value
D.O. % in Hg mmHg kPa mbar 101% 30.22 767.6 102.34 1023.38 100% 29.92 760.0 101.33 1013.25 99% 29.62 752.4 100.31 1003.12 98% 29.32 744.8 99.30 992.99 97% 29.02 737.2 98.29 982.85 96% 28.72 729.6 97.27 972.72 95% 28.43 722.0 96.26 962.59 94% 28.13 714.4 95.25 952.46 93% 27.83 706.8 94.23 942.32 92% 27.53 699.2 93.22 932.19 91% 27.23 691.6 92.21 922.06 90% 26.93 684.0 91.19 911.93 89% 26.63 676.4 90.18 901.79 88% 26.33 668.8 89.17 891.66 87% 26.03 661.2 88.15 881.53 86% 25.73 653.6 87.14 871.40 85% 25.43 646.0 86.13 861.26 84% 25.13 638.4 85.11 851.13 83% 24.83 630.8 84.10 841.00
Pressure
17
82% 24.54 623.2 83.09 830.87 81% 24.24 615.6 82.07 820.73 80% 23.94 608.0 81.06 810.60 79% 23.64 600.4 80.05 800.47 78% 23.34 592.8 79.03 790.34 77% 23.04 585.2 78.02 780.20 76% 22.74 577.6 77.01 770.07 75% 22.44 570.0 75.99 759.94 74% 22.14 562.4 74.98 749.81 73% 21.84 554.8 73.97 739.67 72% 21.54 547.2 72.95 729.54
18
Guide d’Utilisation
DOCUMENT nº 606335
EcoSense® ODO200
Appareil de mesure de l’oxygène dissous et de la température
GUIDE D’UTILISATION
English, Français, Español, Português
Article nº 606335 Rév. B, septembre 2014 Pour obtenir la dernière version de ce manuel, visitez ysi.com
©2014 YSI Incorporated.
Le logo YSI et EcoSense sont des marques déposées d’YSI Incorporated.
CONTENTS
Garantie ................................................................................................. i
INTRODUCTION ..................................................................................1
PREMIERS PAS .....................................................................................1
Inspection initiale ...................................................................1
Précautions .............................................................................1
Sonde ...............................................................................1
Configuration initiale ............................................................2
Préparation de la sonde ..................................................2
Installation de la pile .......................................................4
Clavier .....................................................................................4
Affichage principal ...............................................................6
PROCEDURES D’UTILISATION – PRISES DE MESURES ...................7
Modes de mesure –% ou mg/l .............................................7
Modes d’échantillonnage – Continu ou Manuel .................7
Enregistrement, affichage et suppression des données ....8
Étalonnage ..............................................................................8
Température ..................................................................... 8
Oxygène dissous .............................................................8
ENTRETIEN ET DÉPANNAGE ...........................................................10
Entretien des capteurs – Oxygène dissous .......................10
Nettoyage de l’embout du capteur .............................10
Remplacement de l’embout du capteur ...................10
Saisir les coefficients de l’emboutdecapteurdans
l’EcoSenseODO200 .....................................................12
Réhydratation de l’embout du capteur ....................13
Entretien des capteurs – Température ...............................13
Stockage du capteur............................................................13
Stockage à court terme ..............................................13
Stockage à long terme ..................................................13
Dépannage ...........................................................................13
CARACTÉRISTIQUES .........................................................................15
ACCESSOIRES / NUMÉROS DEPIECES ............. 16
COORDONNÉES ET SERVICE ..........................................................16
Commande et assistance technique ..................................16
Informations sur le service ..................................................17
Recyclage ..............................................................................17
ANNEXE A – VALEURS DE CALIBRAGE DU POURCENTAGE
D’OD ...................................................................................................18
GARANTIE
L’appareil EcoSense® ODO200, la sonde, le câble et l’embout de capteur sont garantis pour une période de un an à compter de la date d’achat par l’utilisateur final contre les défauts matériels et de fabrication, à l’exception des piles et des dommages causés par des piles défectueuses. Pendant la période de garantie, YSI s’engage à réparer ou à remplacer, gratuitement et à sa discrétion, tout produit qu’YSI peut établir comme étant couvert par la garantie.
Pour faire valoir cette garantie, écrivez à votre représentant YSI ou appelez­le, ou contactez le service clientèle d’YSI à Yellow Springs, Ohio, États-Unis, aux numéros +1937 767-7241, 800-765-4974, ou visitez YSI.com pour obtenir un formulaire de retour de produit. Envoyez le produit et son justificatif d’achat, en port payé, au Centre de service homologué sélectionné par YSI. La réparation ou le remplacement seront effectués et le produit vous sera retourné en port payé. Les produits réparés ou remplacés sont garantis jusqu’à expiration de la période de garantie initiale ou pour une période d’au moins 90 jours, à compter de la date de réparation ou de remplacement.
Limitation de garantie
Cette garantie ne s’applique pas aux produits YSI endommagés ou présentant des dysfonctionnements pour les raisons suivantes : (i) installation, exploitation ou utilisation du produit d’une façon non conforme aux instructions écrites d’YSI; (ii) abus ou mésusage du produit; (iii) manquement à l’entretien du produit conformément aux instructions écrites d’YSI ou aux procédures industrielles normales; (iv) réparation non conforme du produit; (v) utilisation par vous de pièces ou de composants défectueux ou non conformes lors de l’entretien ou de la réparation du produit, ou; (vi) modification du produit d’une façon non expressément autorisée par YSI.
CETTE GARANTIE REMPLACE TOUTES LES AUTRES GARANTIES, EXPRESSES OU INDUITES, Y COMPRIS LES GARANTIES DE QUALITÉ MARCHANDE OU D’ADAPTATION À UN USAGE PARTICULIER. LA RESPONS ABILITÉ D’ YSI SELON LES TERME S DE CETT E GARANT IE SE LIMITE À LA RÉPARATION OU AU REMPLACEMENT DU PRODUIT, CONSTITUANT VOTRE SEUL ET UNIQUE RECOURS POUR TOUT PRODUIT DÉFECTUEUX COUVERT PAR CETTE GARANTIE. YSI NE POURRA EN AUCUN CAS ÊTRE TENU RESPONSABLE DE DOMMAGES SPÉCIAUX, INDIRECTS, ACCIDENTELS OU CONSÉCUTIFS RÉSULTANT DE L’UTILISATION DE TOUT PRODUIT DÉFECTUEUX COUVERT PAR CETTE GARANTIE.
i
Loading...
+ 67 hidden pages