4320(W) Overview
2
3
4
1
4320
4320W
OPERATING MANUAL

1 Overview

1.1 Structure and function

Structure
Automatic sensor
recognition
1 Shaft
2 Connecting head with active sensor electronics
3 IDS plug (4320W)
4 Connection cable (4320)
The sensor electronics with the stored sensor data is in the connecting head. The data include, among other things, the sensor type and series number. With each calibration, the calibration data is written in the sensor and the calibration history is recorded. The data is recalled by the meter when the sensor is connected and is used for measurement and for measured value documentation. Storing the calibration data in the sensor ensures that the correct cell constant is automatically used if the sensor is operated with several meters.
The digital transmission technique guarantees the failure-free communi­cation with the meter even with long connection cables. If the sensor firmware is enhanced by YSI, it can be updated with the meter.

1.2 Recommended fields of application

Measurements in ultrapure water.
ba76140e02 10/2017
3
Measurement / Operation 4320(W)

2 Measurement / Operation

2.1 Commissioning

Scope of delivery Conductivity measuring cell4320(W)
Operating manual
Commissioning Connect the electrode to the meter.
4320 – via the sensor cable to a free IDS connec-
tor on the meter
4320W – via a connecting cable (accessory) to a
free IDS connector on the meter
or – wireless via an IDS WA-S adapter (acces-
sory) to a WA-capable meter
Accessories for the connection of the 4320W sensor to the meter: See chapter 6 W
PARTS AND ACCESSORIES.
EAR
Opening and closing the IDS plug connection, see section 2.2 O IDS
PLUG CONNECTION (4320W).
PENING AND CLOSING THE
4
ba76140e02 10/2017
4320(W) Measurement / Operation
2
4
3
1
Plug connection closed and locked
Guidance
Groove

2.2 Opening and closing the IDS plug connection (4320W)

This section only applies to the IDS plug variant, 4320W.
Opening the plug
connection
Closing the plug
connection
If necessary, clean the plug connection.
Open the locking device (step 1).
Use your thumb and index finger to press the clips of the connector togeth-
er, and pull the connector out of the plug (step 2).
Make sure that the plug connection is completely dry and clean.
Align the guidance of the connector with the groove in the plug and insert
the connector in the unlocked plug until it catches (step 3).
Close the locking device (step 4).
ba76140e02 10/2017
5
Cleaning 4320(W)

3 Cleaning

Note
To clean the sensor, disconnect it from the instrument.
Exterior cleaning We recommend to clean the sensor thoroughly, especially before measuring
low conductivity values.
Contamination Cleaning procedure
Lime sediments Immerse in acetic acid for 5 minutes (volume
share = 10 %)
Fat/oil Clean with warm water that contains wash-
ing-up liquid
Aging of the con­ductivity measur-
ing cell
Disposal We recommend to dispose of the measuring cell as electronic waste.
After cleaning, thoroughly rinse with deionized water and recalibrate if neces­sary.
Normally, the conductivity measuring cell does not age. Special measuring media (e.g. strong acids and bases, organic solvents) or temperatures that are too high may considerably reduce its lifetime or lead to damage. The war­ranty does not cover failure caused by measuring conditions and mechanical damage.
6
ba76140e02 10/2017
4320(W) What to do if ...

4 What to do if ...

Error symptom Cause Remedy
No temperature or con­ductivity display
Measurement delivers implausible conductivity values
Incorrect temperature display
– No connection between meter and
conductivity measuring cell
– Cable defective
– Measuring range exceeded
– Contamination in the area of the
electrodes
– Electrodes damaged
– The temperature sensor is not
immersed deep enough in the measuring solution
– Temperature sensor defective
– Establish connection between
meter and conductivity measuring cell
– Make sure the correct sensor is
being used for the application
– Clean the conductivity measuring
cell (see section 3).
– Return the sensor
– Observe the minimum immersion
depth
– Return the conductivity measuring
cell
ba76140e02 10/2017
7
Technical data 4320(W)
12,0
12,0
120,0
120,0
176,0
222,8
21,7
21,7
4320
4320W

5 Technical data

5.1 General data

General features Measuring principle 2-electrodes measurement
-1
Cell constant 0.100 cm
Temperature sensor Integrated NTC 30 (30 k at 25 °C / 77 °F)
Dimensions
(in mm)
± 2 %
Weight 4320 Approx. 90 g (without cable)
4320W Approx. 95 g
Materials Shaft Stainless steel 1.4571
Connection head 4320: POM
4320W: POM
Conductivity electrodes Stainless steel 1.4571
Thermistor enclosure Stainless steel 1.4571
8
ba76140e02 10/2017
4320(W) Technical data
Connection cable Lengths 4320: 1,5 m
4320W: 1,5 / 3 m
Diameter 4.3 mm
IDS plug
(4320W)
Pressure resis-
tance
Smallest allowed bend radius
Fixed installation: 20 mm Flexible use: 60 mm
Plug type Socket, 4 pins
Type of connection 4-pole, watertight plug connection with lock,
reverse polarity protected
Materials  Synthetic materials: Glass fiber reinforced
Noryl, TPU, TPC-ET, POM, PEEK, PBT
O-ring: FPM
Contacts gold-plated
5
Sensor with connection
IP 68 (2 x 10
Pa or 2 bar)
cable
Cable plug IP 67 (when plugged in)
The 4320(W) meets the requirements according to article 3(3) of the direc­tive, 97/23/EC ("pressure equipment directive").
Measurement con-
ditions
Conductivity measuring range
Temperature range -5 ... 70 °C (100 °C) 23 ... 158 °F (212 °F)
Max. admissible overpres­sure
Minimum depth of immer­sion
Maximum depth of immer­sion (at temperature)
Operating position Any
Storage conditions Recommended storing
method
Storage temperature 0 ... 50 °C (32 ... 122 °F)
μS/cm ... 200 μS/cm
0.01
5
2 x 10
Pa (2 bar)
30 mm
Whole sensor + cable up to 70 °C (158 °F) Sensor shaft only (=120 mm) up to 100 °C (212 °F)
In air
ba76140e02 10/2017
9
Technical data 4320(W)
Characteristics when delivered
Measuring ranges,
resolution
Temperature responding behavior
Accuracy of the temperature
t99 (99 % of the final value display after) <20s
± 0.2 K
sensor

5.2 Measuring ranges and resolution

Measured parameter Measuring range Resolution
ӕ [µS/cm] 0.01 ... 19.99
0.0 ... 199.9
ρ (resistivity) [kOhm*cm] 5,00 ... 19,99
20.0 ... 199.9 200 ... 1999
ρ (resistivity) [MOhm*cm] 2.00 ... 19.99
20.0 ... 199.9
T [°C] 5,0 ... + 100,0 0.1
0.01
0.1
0.01
0.1 1
0.01
0.1

5.3 Accuracy of the IDS measuring technique

Measured parameter Accuracy (± 1 digit)
ӕ, ρ ± 0.5 % of measured value
T [°C] ± 0.1
10
ba76140e02 10/2017
4320(W) Wear parts and accessories

6 Wear parts and accessories

Connection cable
4320W
- meter
Radio connection
4320W - meter
Description Model Order no.
IDS connection cable, 1.5 m IDS-CABLE-
903 850Y
1.5
IDS connection cable, 3 m IDS-CABLE-3 903 851Y
Description Model Order no.
WA capable IDS meter
see Internet
+ radio module for IDS meter
Radio module for plug head sensor IDS WA-S 108 141Y
ba76140e02 10/2017
11
Wear parts and accessories 4320(W)
12
ba76140e02 10/2017
4320 Contact Information

7 Contact Information

7.1 Ordering & Technical Support

Telephone
Fax
: (937) 767-1058
Email
Mail: YSI Incorporated
Internet
When placing an order please have the following information available:
YSI account number (if available) Name and Phone Number Model number or brief description Billing and shipping address Quantity Purchase Order or Credit Card
: (800) 897-4151
(937) 767-7241 Monday through Friday, 8:00 AM to 5:00 PM ET
: environmental@ysi.com
1725 Brannum Lane Yellow Springs, OH 45387 USA
: www.ysi.com
ba76140e02 10/2017

7.2 Service Information

YSI has authorized service centers throughout the United States and Internationally. For the nearest service center information, please visit
www.ysi.com
directly at 800-897-4151.
When returning a product for service, include the Product Return form with cleaning certification. The form must be completely filled out for an YSI Service Center to accept the instrument for service. The Product Return form may be downloaded at www.ysi.com ‘Support‘ tab.
and click ‘Support’ or contact YSI Technical Support
and clicking on the
13
Contact Information 4320
14
ba76140e02 10/2017
1) The tissue in plants that brings water upward from the roots;
2) a leading global water technology company.
We're 12,500 people unified in a common purpose: creating innovative solutions to meet our world's water needs. Developing new technologies that will improve the way water is used, conserved, and re-used in the future is central to our work. We move, treat, analyze, and return water to the environment, and we help people use water efficiently, in their homes, buildings, factories and farms. In more than 150 countries, we have strong, long-standing relationships with customers who know us for our powerful combination of leading product brands and applications expertise, backed by a legacy of innovation.
For more information on how Xylem can help you, go to www.xyleminc.com
Xylem |' m|zīlə
YSI 1725 Brannum Lane Yellow Springs, OH 45387 Tel: +1 937-767-7241; 800-765-4974 Fax: +1 937-767-1058 Email: environmental@ysi.com Web: www.ysi.com
©Xylem Inc
Loading...