Performance
Test conditions: Temperature and humidity = 23 ± 5°C, 80% RH or less; Accuracy = ± (% rdg + dgt).
Note: A response time is the time required for achieving the accuracy specified for the corresponding range.
TY700 General Specifications
AC Current Measurement [RMS] (~A) RMS detection crest factor: 3
Range Resolution
Accuracy (Upper: TY710; Lower: TY720; the display of “—” is not specified)
Voltage Drop
Maximum Input
Current
500μA 0.01μA
<0.11mV/μA
5000μA 0.1μA 1.5+20 1+20
—
50mA 0.001mA 1+20 0.75+20 1+30
<4mV/mA
500mA
*30.01mA
5A 0.0001A 1.5+20 1+20
—
<0.1V/A
10A
10A 0.001A 1.5+20 1+20 2+30 fuse-protected
Shown above is the accuracy at 5 to 100% of range (10 to 100% for 10A range). Response time: 1 second or less
*3: Maximum testing current at 500mA of range is 440mA.
10 - 20Hz 20Hz - 1kHz 1k - 5kHz
440mA
fuse-protected
AC Voltage Measurement [MEAN] (~A)
Mean-value detection and RMS-value calibration
(sinusoidal wave)
Range Resolution
Accuracy TY720
Voltage Drop
Maximum Input
Current
500μA 0.01μA
<0.11mV/μA
5000μA 0.1μA
2+20 1.5+20 2+30
50mA 0.001mA
<4mV/mA
500mA
*30.01mA
5A 0.0001A
3+20 2+20 4+30 <0.1V/A
10A
10A 0.001A
fuse-protected
Shown above is the accuracy at 5 to 100% of range (10 to 100% for 10A range). Response time: Approx. second or less
*3: Maximum testing current at 500mA of range is 440mA.
10 - 20Hz
20 - 500Hz 500Hz - 1kHz
440mA
fuse-protected
15+40*
2
1000V DC
1000V rms AC
Range Resolution
Accuracy
Input Resistance Maximum Input Voltage
TY710,TY720
50mV 0.001mV 0.05+10
500mV 0.01mV
0.02+2
Approx. 100MΩ
2400mV 0.1mV
5V 0.0001V 0.025+5
50V 0.001V
10MΩ
500V 0.01V 0.03+2
1000V 0.1V
NMRR: 80dB or greater for 50/60Hz ± 0.1%At 50mV of range, 70dB or greater for 50/60Hz ± 0.1%
CMRR: 100dB or greater for 50/60Hz(Rs=1kΩ) Response time: 0.3 seconds or less
DC Voltage Measurement( V)
Range Resolution
Accuracy (Upper: TY710; Lower: TY720; the display of “—” is not specified)
Voltage Drop
Maximum Input
Current
500μA 0.01μA
<0.11mV/μA
5000μA 0.1μA 2+10 1.5+10
—
50mA 0.001mA 1.5+10 1+10 1.5+10
<4mV/mA
500mA
*30.01mA
5A 0.0001A 2+10 1.5+10
—
<0.1V/A
10A
10A 0.001A 2+10 1.5+10 3+10
fuse-protected
Shown above is the accuracy at 5 to 100% of range (10 to 100% for 10A range). Response time: Approx. 2 seconds *3: Maximum testing current for 500mA of range is 440mA.
DC,10 - 20Hz DC,20Hz - 1kHz DC,1k - 5kHz
440mA
fuse-protected
DCA + ACA ( +~) Maximum effective display: 50,000, crest factor: 3
Range Resolution
Accuracy
Voltage Drop
Maximum Input
TY710,TY720 Current
500μA 0.01μA
<0.11mV/μA
5000μA 0.1μA 0.2+2
50mA 0.001mA
<4mV/mA
500mA
*30.01mA
5A 0.0001A
0.6+2 <0.1V/A
10A
10A 0.001A fuse-protected
Response time: 0.3 seconds or less *3: Maximum testing current at 500mA of range is 440mA
440mA
fuse-protected
DC Current Measurement ( A)
Temperature Measurement (TEMP)
Range Resolution
AccuracyTY710,TY720
Input Protection Voltage
-200 - 1372°C 0.1°C 1+1.5°C 1000V rms
Temperature probe: Type K thermocouple sensor (optional)
Frequency Measurement (Hz)
AC coupling, Maximum effective display: 9999
Range (auto-ranging) Resolution Accuracy TY710,TY720
2.000 - 9.999Hz 0.001Hz
9.00 - 99.99Hz 0.01Hz
90.0 - 999.9Hz 0.1Hz
0.900 - 9.999kHz 0.001kHz
9.00 - 99.99kHz 0.01kHz
*
2
*1: At 10 to 100% of input voltage or current range
*2: At 40 to 100% of input voltage or current range
0.02+1*
1
Duty Cycle (%)
Range Resolution
Accuracy TY710,TY720
10 - 90% 1% ± 1%*
1
*1: For input of a square wave with a frequency within 10.00 to
500.0Hz At 40 to 100% of input voltage or current range
Peak Hold Function (PH) TY720 only
Maximum effective display: 5000
Range Accuracy TY720 Response Time
DCV, DCA
±
100 digit >250μs
Capacitance ( ) Maximum effective display:5000
Range Resolution
AccuracyTY710,TY720
Input Protection Voltage
5nF 0.001nF
50nF 0.01nF
500nF 0.1nF
5μF 0.001μF
50μF 0.01μF
500μF 0.1μF 2+5
5mF 0.001mF
3+5
50mF 0.01mF
*1: Accuracy after zero calibration
1+5*
1
1000V rms
Diode Test ( )
Range Resolution Accuracy TY710,TY720 Testing Current (Vf = 0.6 V) Open-circuit Voltage Input Protection Voltage
2.4V 0.0001V 1 + 2 Approx. 0.5mA <5V 1000V rms
Model and Specification Code
Optional Accessories
Name Model
TY710
Digital Multimeter
TY720
Name Model Specification
DMM communication package 92015 USB communication adapter + USB
communication cable + Application software
Communication package for printer
97016 Printer adapter + Printer cable
Printer 97010 Thermal printer Paper width 112mm
AC adapter for printer 94005 AC 100V ± 10%
94006 For Printer, Europe
Thermal printing paper 97080 10 rolls (1 set)
Test leads 98015 1000V CAT#, 600V CAT$ Red/black (1 set)
Name Model Specification
Fuse 99015 440 mA/1000V (1 piece/1 unit)
99016 10 A/1000 V (1 piece/1 unit)
TC-K temperature probe 90050 -50 to 600°C (For liquids)
90051 -50 to 600°C (For liquids)
90055 -20 to 250°C (For surfaces)
90056 -20 to 500°C (For surfaces)
Current clamp probe 96001 For 400A, AC Output: 10mV/A, AC
Carrying case 93029 Hard type (Houses the DMM, the test leads and communication cable)
Measurement Functions
:DC voltage, AC voltage, DCV+ACV, DC current, AC current, DCA+ACA, resistance, frequency, temperature, capacitance,
duty cycle, decibel calculation, continuity check, diode test, low-power resistance (TY720 only)
For AC voltage/current, RMS/MEAN detection can be switched (TY720 only).
For AC voltage/current, the low-pass filter can be turned on/off (TY720 only).
Additional Functions :
Data hold/auto hold/peak hold (TY720 only), range hold, maximum/minimum/average values
resistance, capacitance zero, relative and percentage value calculation, manual-mode memory, logging-mode memory,
auto power off, backlight (white LED)
Display :5-digit LCD: 7-segment
Digital display: Main display; [50,000] counts
Sub-display ; [50,000] counts
Bar graph display: 51-segment
Polarity indicator: “-” appears automatically when the polarity is negative
Overrange indicator: “OL”
Low-battery indicator: “ ” appears at or below the minimum operating voltage.
Measuring Rate :6 times/sec (Frequency: 1 time/sec, Capacitance: max. 0.03 times/sec (50mF), Resistance: 4 times/sec)
Bar graph display: 15 times/sec
Operating Temp. and Humidity
:-20 to 55°C; 80% RH or less (no condensation) 40 to 55°C: 70% RH or less
Storage Temp. and Humidity
:-40 to 70°C; 70% RH or less (no condensation)
Temperature Coefficient
:Add the accuracy 0.05/°C to the basic accuracy at a temperature within -20 to 18°C and 28 to 55°C.
Power Supply : Four AA (R6) dry cells
Battery Life :Approx. 120 hours (for continuous DC voltage measurement with alkaline cells)
Withstanding Voltage :6.88kV for 5 seconds (between input terminals and casing)
Dimensions :Approx. 90(W) x 192(H) x 49(D) mm
Weight :Approx. 560g (including batteries)
Compliance with Standards
:Safety EN61010-1, EN61010-031, 1000V CAT#, 600V CAT$, pollution level 2, indoor, 2000m max. above sea level
EMC: EN61326-1 Class B, EN55011 Class B Group 1
Standard Accessories :AA (R6) dry cells: 4, Test lead set (98015): 1, Fuse (installed) 440mA/1000V and 10A/1000V, Instruction manual: 1
+
--
AC Voltage Measurement [RMS] (~V)
AC coupling, RMS detection, crest factor for 1000V of range: 1.5 ; crest factor for ranges other than 1000V: 3
Input Maximum
Range Resolution
Impedance Input Voltage
50mV 0.001mV
———— —
2+80*
2
0.4+40*25+40*25.5+40*
2
11MΩ<50pF
500mV 0.01mV
5V 0.0001V 1.5+30*
1
2+50*2——
50V 0.001V 1+30*
1
1+40*12+70*25+200*
2
500V 0.01V
10MΩ<50pF
1000V 0.1V
*
2
*
2 3+30*
2
—
*
2
*
2 3+30*
2
—
*1: At 5 to 100% of range *2: At 10 to100% of range CMRR: 80dB or greater for DC to 60Hz(Rs= 1kΩ) Response time: 1 second or less
20Hz 1kHz
10 20Hz
1k -
10kHz
10k 20kHz
20k 50kHz
50k 100kHz
1000V rms AC
1000V DC
0.7+30*
1
0.4+30*
1
Accuracy (Upper: TY710; Lower: TY720; the display of “—” is not specified)
Input Maximum
Range Resolution
Impedance Input Voltage
5V 0.0001V 11MΩ<50pF
50V 0.001V
500V 0.01V
1000V 0.1V
*
2
*2
—
10MΩ<50pF
*
2
*
2
—
*1: At 5 to 100% of range *2: At 10 to 100% of range CMRR: 80dB or greater for DC to 60Hz (Rs = 1kΩ) Response time: Approx. 2 seconds
DC,20Hz
-
1kHz
DC,10
-
20Hz
DC,1k
-
10kHz
DC,10k
-
20kHz
DC,20k
-
50kHz
DC,50k
-
100kHz
1000V rms AC
1000V DC
DCV + ACV ( +~)
AC coupling, RMS detection crest factor for 1000V of range: 1.5 ; crest factor for ranges other than 1000 V: 3
Accuracy (Upper: TY710; Lower: TY720; the display of “—” is not specified)
1.5+10*
1
1+10*
1
2+10*2——
1.5+10*10.5+10*
1
1+10*12+10*25+20*
2
Resistance Measurement (Ω)
Range Resolution
Accuracy
Maximum Testing Open-circuit Input Protection
Current Voltage Voltage
500Ω 0.01Ω <1mA
5kΩ 0.0001kΩ <0.25mA
50kΩ 0.001kΩ <25μA
500kΩ 0.01kΩ <2.5μA
5MΩ 0.0001MΩ 0.5+2 <1.5μA
50MΩ 0.001MΩ 1+2 <0.13μA
*1: Accuracy after zero calibration Response time: 1 second or less for 500Ω to 500kΩ, 5 seconds or less for 5MΩ to 50MΩ
TY710 TY720
<2.5V 1000V rms
0.1+2*
1
0.05+2*
1
Low-power Resistance Measurement (LP-Ω)
Maximum effective display: 5000
Range Resolution
Accuracy
Maximum Testing Open-circuit Input Protection
Current Voltage Voltage
5kΩ 0.001kΩ <10μA
50kΩ 0.01kΩ 0.2+3 <1.0μA
500kΩ 0.1kΩ <0.6μA
5MΩ 0.001MΩ 1+3 <0.05μA
TY720
<0.7V 1000V rms
Range Resolution Continuity Beeper TY710, TY720 Testing Current Open-circuit Voltage
Input Protection Voltage
500Ω 0.1Ω Buzzer sounds at 100 ± 50Ω or less. Approx. 0.5mA <5V 1000V rms
Continuity Check ( )
Maximum effective display: 5000
1000V rms AC
1000V DC
AC Voltage Measurement [MEAN] (~V)
AC coupling, Mean-value detection and RMS-value calibration (sinusoidal wave)
Range Resolution
Accuracy
Input Maximum
TY720
Impedance Input Voltage
50mV 0.001mV 4+80*
2
1.5+30*
2
5+30*
2
500mV 0.01mV 11MΩ<50pF
5V 0.0001V
50V 0.001V
10MΩ<50pF
500V 0.01V
1000V 0.1V
*
2
*
2
*
2
*1: At 5 to 100% of range *2: At 10 to 100% of range CMRR: 80dB or greater for DC to 60Hz (Rs= 1kΩ) Response time: 1 second or less
10 - 20Hz
2+30*
1
20 - 500Hz 500 - 1kHz
1+30*
1
3+30*
1
YOKOGAWA CORPORATION OF AMERICA (U.S.A.) Phone: +1-770-253-7000 Facsimile: +1-770-251-2088
YOKOGAWA EUROPE B. V. (THE NETHERLANDS) Phone: +31-334-64-1611 Facsimile: +31-334-64-1610
YOKOGAWA ENGINEERING ASIA PTE. LTD. (SINGAPORE) Phone: +65-6241-9933 Facsimile: +65-6241-2606
YOKOGAWA AMERICA DO SUL LTDA (BRAZIL) Phone: +55-11-5681-2400 Facsimile: +55-11-5681-1274/4434
YOKOGAWA MEASURING INSTRUMENTS KOREA CORPORATION (KOREA) Phone: +82-2-551-0660 Facsimile: +82-2-551-0665
YOKOGAWA AUSTRALIA PTY. LTD. (AUSTRALIA) Phone: +61-2-8870-1100 Facsimile: +61-2-8870-1111
YOKOGAWA INDIA LTD. (INDIA) Phone: +91-80-4158-6000 Facsimile: +91-80-2852-1441
YOKOGAWA SHANGHAI TRADING CO., LTD. (CHINA) Phone: +86-21-6880-8107 Facsimile: +86-21-6880-4987
YOKOGAWA MIDDLE EAST E.C. (BAHRAIN) Phone: +973-358100 Facsimile: +973-336100
YOKOGAWA ELECTRIC CIS LTD. (RUSSIAN FEDERATION) Phone: +7-095-737-7868 Facsimile: +7-095-737-7869
Represented by:
World Wide Web site at
http://www.yokogawa.com/MCC
Before using the product, read the instruction manual carefully to ensure proper and safe operation.
NOTICE
Tachihi Bld. No.2, 6-1-3 Sakaecho, Tachikawa-shi, Tokyo, 190-8586 Japan
Phone: +81-42-534-1413 Facsimile: +81-42-534-1426
YOKOGAWA METERS & INSTRUMENTS CORPORATION
International Sales
Subject to change without notice. Printed in Japan: Aug 2008(C)/3000(GT)[Ed: 08]
All Rights Reserved. Copyright © 2008, Yokogawa Meters & Instruments Corporation.