YOKOGAWA SR10001, SR10006, SR10002, SR10003, SR10004 User Manual

User’s Manual
Models SR10001/SR10002/SR10003/SR10004/
SR10006
SR10000 Recorder
Yo kogawa Electric Corporation
IM 04P03B01-01E 2nd Edition

Foreword

Thank you for purchasing the YOKOGAWA SR10000 Recorder. This user’s manual explains how to use the SR10000 recorder excluding the communication functions. To ensure correct use, please read this manual thoroughly before operation. The following three manuals including this manual are available for the SR10000 Recorder.
• Paper Manual
Manual Title Manual No. Description
SR10000 Recorder IM 04P03B01-02E Explains concisely the operations of the Operation Guide SR10000 Recorder. It is also provided in the
CD-ROM.
• Electronic Manuals Provided on the Accompanying CD-ROM
Manual Title Manual No. Description
SR10000 Recorder IM 04P03B01-02E Explains concisely the operations of the Operation Guide SR10000 Recorder. This is the electronic
version of the paper manual.
SR10000 Recorder IM 04P03B01-01E This manual. User’s Manual
SR10000 IM 04P03B01-17E Explains the communication functions of the Communication Interface SR10000 Recorder using Ethernet interface and User’s Manual the RS-422A/485 communication interface.
Notes
Trademarks
• The contents of this manual are subject to change without prior notice as a result of continuing improvements to the instrument’s performance and functions.
• Every effort has been made in the preparation of this manual to ensure the accuracy of its contents. However, should you have any questions or find any errors, please contact your nearest YOKOGAWA dealer as listed on the back cover of this manual.
• Copying or reproducing all or any part of the contents of this manual without the permission of Yokogawa Electric Corporation is strictly prohibited.
• The TCP/IP software of this product and the document concerning the TCP/IP software have been developed/created by YOKOGAWA based on the BSD Networking Software, Release 1 that has been licensed from the University of California.
• All the brands or names of Yokogawa Electric’s products used in this manual are either trademarks or registered trademarks of Yokogawa Electric Corporation.
• Microsoft, MS-DOS, Windows, Windows NT, and Windows XP are either registered trademarks or trademarks of Microsoft Corporation in the United States and/or other countries.
• Adobe, Acrobat, and PostScript are trademarks of Adobe Systems incorporated.
• For purposes of this manual, the TM and ® symbols do not accompany their respective trademark names or registered trademark names.
• Company and product names that appear in this manual are trademarks or registered trademarks of their respective holders.
Revisions
1st Edition February 2006 2nd Edition October 2006
2nd Edition: October 2006 (YK)
All Rights Reserved, Copyright © 2006 Yokogawa Electric Corporation
IM 04P03B01-01E

How to Use This Manual

Structure of the Manual
Read the Operation Guide first to familiarize yourself with the basic operation, and then read this manual. For a description of the communication function, see the
SR10000 Communication Interface User's Manual (IM 04P03B01-17E)
This user's manual consists of the following sections.
Chapter Title and Description
1 Functional Explanation and Setup Guide
Describes the functions of the SR10000 Recorder and provides a function setup guide. Refer to this chapter when you are unsure of the details of the function that you are operating.
2 Frequently Used Setup Operations (Setting Mode)
Describes how to change the input range, alarms, chart speed, etc.
3 Setup Operations for Convenient Functions (Setting Mode)
Describes the setup operations for convenient functions such as how to assign tags to channels and how to set message strings that are to be printed.
4 Setup Operations for Changing/Adding Functions (Basic Setting Mode)
Describes the setup operations for changing or adding functions such as setting the recorder to detect sensor burnouts and changing the contents that are printed on the chart paper.
5 Troubleshooting
Describes error message and troubleshooting measures of the SR10000 Recorder.
6 Maintenance
Describes periodic inspection, calibration, and pen adjustment/printer carriage adjustment.
7 Specifications
Gives the specifications of the SR10000 Recorder.
Appendix Describes the printout contents.
Index
Note
• This user’s manual covers information regarding the recorders with English as the printout
font (suffix code “2”).
Recorder’s Version and Functions Described in This Manual
The contents of this manual corresponds to the recorder with version 1.31.
SR10000 Versions and Functions
Version
1.21 or earlier
1.31
Suffix Code
– /BT1
• Checking the Version Number Press the Hold down the
FUNC
key, key, or key to select VER ( ), then press the key.
FUNC
Software (Sold Separately)
The table below shows the relationship between the RXA10 Configuration Software revisions and the SR10000 recorder versions.
RXA10 Configuration Software revision
Yes: Compatible
Added or Modified Functions
– (Added) Header printout
key to return to Operation mode.
Recorder version
1.21 or earlier
R3.01
1.31 YesYes
Reference
– Section 1.3
ii IM 04P03B01-01E
How to Use This Manual
Conventions Used in This Manual
Unit
K ........ Denotes 1024. Example: 768 KB (file size)
k ........ Denotes 1000.
Safety Markings
The following markings are used in this manual.
WARNING
CAUTION
Improper handling or use can lead to injury to the user or damage to the instrument.
instrument to indicate that the user must refer to the user’s
manual for special instructions. The same symbol appears in
the corresponding place in the user’s manual to identify those
instructions. In the manual, the symbol is used in conjunction
with the word “WARNING” or “CAUTION.”
Calls attention to actions or conditions that could cause serious
or fatal injury to the user, and precautions that can be taken to
prevent such occurrences.
Calls attentions to actions or conditions that could cause light
injury to the user or damage to the instrument or user’s data,
and precautions that can be taken to prevent such occurrences.
This symbol appears on the
1
2
3
4
5
6
7
Note
Subheadings
On pages that describe the operating procedures in Chapter 2 through 4 and 6, the following symbols are used to distinguish the procedures from their explanations.
Procedure
Explanation
Calls attention to information that is important for proper
operation of the instrument.
Follow the numbered steps. All procedures are written with
inexperienced users in mind; depending on the operation, not
all steps need to be taken.
This subsection describes the setting parameters and the
limitations on the procedures. It does not give a detailed
explanation of the function. For details on the function, see
chapter 1.
App
Index
IM 04P03B01-01E
iii

Contents

Foreword ......................................................................................................................................... i
How to Use This Manual ................................................................................................................. ii
Chapter 1 Functional Explanation and Setup Guide
1.1 Measuring Input Section .................................................................................................. 1-1
1.2 Alarms .............................................................................................................................. 1-6
1.3 Recording ......................................................................................................................... 1-8
1.4 Remote Control Function (/R1 Option) .......................................................................... 1-19
1.5 Other Functions ............................................................................................................. 1-21
1.6 Function Setup Guide .................................................................................................... 1-23
Chapter 2 Frequently Used Setup Operations (Setting Mode)
2.1 Setting the Input Range ................................................................................................... 2-1
2.2 Setting the Alarm.............................................................................................................. 2-9
2.3 Setting the Unit on Linearly Scaled Channels ................................................................ 2-11
2.4 Changing the Chart Speed ............................................................................................ 2-12
2.5 Setting the Date/Time .................................................................................................... 2-13
Chapter 3 Setup Operations for Convenient Functions (Setting Mode)
3.1 Setting the Trend Recording Interval (Dot Model)............................................................ 3-1
3.2 Setting the Filter (Pen Model) .......................................................................................... 3-2
3.3 Setting the Moving Average (Dot Model) ......................................................................... 3-3
3.4 Setting Recording Zones for Each Channel (Zone Recording) ........................................ 3-4
3.5 Setting the Partial Expanded Recording .......................................................................... 3-5
3.6 Turning Trend Recording (Dot Model) and Periodic Printout ON/OFF for Each Channel 3-6
3.7 Setting Tags on Channels ................................................................................................ 3-7
3.8 Setting the Message String .............................................................................................. 3-8
3.9 Setting the Secondary Chart Speed (Remote Control Function, /R1) ............................. 3-9
3.10 Applying a Bias on the Measuring Input Signal ............................................................. 3-10
3.11 Performing Calibration Correction (/CC1 Option) .......................................................... 3-11
3.12 Setting Up Start Printout and End printout (/BT1 Option) .............................................. 3-13
3.13 Regarding the Message Format (/BT1 Option) .............................................................. 3-18
3.14 Setting the Date/Time for Switching between Standard Time and DST ......................... 3-20
Chapter 4 Setup Operations for Changing/Adding Functions
(Basic Setting Mode)
4.1 Changing the Auxiliary Alarm Function ............................................................................ 4-1
4.2 Changing the Integration Time of the A/D Converter ....................................................... 4-3
4.3 Setting the Burnout Detection Function ........................................................................... 4-4
4.4 Setting the RJC Function on TC Input Channels ............................................................. 4-5
4.5 Changing the Channel Recording Color (Dot Model) ...................................................... 4-6
4.6 Recording by Compensating for the Pen Offset along the Time Axis (Pen Model) .......... 4-7
4.7 Turning Printouts ON/OFF (Selecting the Channel/Tag Printout and Turning ON/OFF the Channel, Alarm, Recording Start, New Chart Speed, Scale, and Pen Color
Printouts) .............................................................................................................. 4-8
4.8 Turning Periodic Printout ON and OFF and Setting the Interval .................................... 4-10
4.9 Setting the Key Lock ...................................................................................................... 4-12
4.10 Enabling the Moving Average Function (Dot Model)...................................................... 4-15
4.11 Enabling the Filter Function (Pen Model) ....................................................................... 4-16
iv IM 04P03B01-01E
Contents
4.12 Enabling the Partial Expanded Recording Function ...................................................... 4-17
4.13 Changing the Printout Font ............................................................................................ 4-18
4.14 Changing the Print/Display Format of the Date ............................................................. 4-19
4.15 Enabling the Bias, Low-Cut, and Calibration Correction (/CC1 Option) Functions ........ 4-20
4.16 Changing the Time Printout Format ............................................................................... 4-22
4.17 Initializing the Settings ................................................................................................... 4-24
4.18 Assigning Functions to the Remote Control Input Terminals (/R1 Option) ..................... 4-25
4.19 Selecting to Show/Hide the FUNC Key Menus .............................................................. 4-27
4.20 Selecting to Show/Hide Setting Mode Menus ................................................................ 4-29
4.21 Enabling/Disabling the Customized Menu ..................................................................... 4-31
4.22 Setting the Calibration Correction Function (/CC1 Option) ............................................ 4-34
4.23 Enabling Start Printout, End printout, and Message Format (/BT1 Option) ................... 4-35
4.24 Changing the Temperature Unit ..................................................................................... 4-37
Chapter 5 Troubleshooting
5.1 A List of Error Messages .................................................................................................. 5-1
5.2 Troubleshooting Flow Charts ........................................................................................... 5-4
Chapter 6 Maintenance
6.1 Periodic Inspection ........................................................................................................... 6-1
6.2 Cleaning the Recorder ..................................................................................................... 6-2
6.3 Calibrating the Recorder .................................................................................................. 6-3
6.4 Adjusting the Pen Position (Pen Model) .......................................................................... 6-5
6.5 Adjusting the Dot Printing Position (Dot Model) ............................................................... 6-7
1
2
3
4
5
6
7
Chapter 7 Specifications
7.1 Input Specifications .......................................................................................................... 7-1
7.2 Alarm Function Specifications .......................................................................................... 7-3
7.3 Recording Function Specifications ................................................................................... 7-4
7.4 Display Function Specifications ....................................................................................... 7-7
7.5 Specifications of Optional Functions ................................................................................ 7-9
7.6 General Specifications ................................................................................................... 7-13
Appendix
Appendix 1 Periodic Printout Interval ................................................................................ App-1
Index
App
Index
IM 04P03B01-01E
v

Chapter 1 Functional Explanation and Setup Guide

1.1 Measuring Input Section

Input Section
Number of Measurement Channels and Scan Interval
The recorder samples the input signals on the measurement channels at the scan interval to obtain the measured values.
Model Number of Channels Scan Interval
1-pen model 1 125 ms 2-pen model 2 125 ms 3-pen model 3 125 ms 4-pen model 4 125 ms Dot model 6 1 s
Input Type, Measurable Range, and Computation
The recorder can measure the following types of inputs.
Input Type Measurable Range
DC voltage DC voltage in the range of ±20 mV to ± 50 V 1-5V See “1-5V” below. Thermocouple Temperature range corresponding to each type: R, S, B, K, E, J, T, N, W, L, U,
and WRe
RTD Temperature range corresponding to each type: Pt100 and JPt100 ON/OFF input Contact input: Open contact is OFF (0). Closed contact is ON (1).
Voltage input: Less than 2.4 V is OFF (0). Greater than or equal to 2.4 V is ON (1). Within ±6 V.
1
Functional Explanation and Setup Guide
(However, the scan interval is 2.5 s when the integration time of the A/D converter is 100 ms.)
• 1-5V
1-5V is scaled to values in the appropriate unit to be used as measured values. Also, the low-cut function (input less than 0% is fixed to 0% (scale left value)) can be used.
• Current Input
A shunt resistor is attached to the input terminal. The current signal is converted to a voltage signal and measured. The measurable range is the range equivalent to the “DC voltage” range indicated above after converting the current to the voltage signal.
Note
Three types of shunt resistors (250 , 100 , and 10 ) are available for current input (See
“Optional Accessories (Sold Separately)” in the
shunt resistor is used to convert the signal to the range of 1 to 5 V for 4 to 20 mA input.
• Range Type, Measurable Range, and Recording Span
Various “range type” are available for the different types of inputs (for example thermocouple type R). Each range type has a preset measurable range (0.0 to
1760.0°C for thermocouple type R). Measurement can be made by specifying an arbitrary range within the measurable range as the in the input range are recorded on the chart paper. The range of measured values that are recorded is called the
Measurable range (Thermocouple Type R example)
1760.0°C
recording span
Input range or recording span
Operation Guide
input range
). For example, a 250-
. The measured values
1500.0°C (rightmost value of span)
IM 04P03B01-01E
300.0°C (leftmost value of span)
0.0°C
<Related Topics> Setting the input range: Section 5.1
For the procedure to set the functions, see section 1.6, “Function Setup Guide.”
1-1
1.1 Measuring Input Section
• Delta Computation
The value obtained by subtracting the measured value of another channel (called the
reference channel
) from the input value of the channel set to delta computation is used as the measured value of that channel. The reference channel must be assigned to a channel whose channel number is less than that of the channel on which delta computation is specified. The channel on which delta computation is specified is automatically set to the same range type as the reference channel.
Channel set to delta computation
Input value
Measured value on the reference channel
Measured value
Note
A channel whose input type is set to DC voltage, TC, or RTD can be designated as a
reference channel. However, channels set to scaling or square root computation cannot be
designated.
• Scaling
The input values are scaled to values in the appropriate unit to be used as measured values.
Measured valueInput value
10 V
0 V 100.0°C
• Square Root Computation
When the input type is DC voltage, the square root of the input value is calculated, the result is scaled to a value in the appropriate unit, and used as the measured value of the channel. Also, the low-cut function (input less than a given measured value is fixed to 0% (scale left value)) can be used.
300.0°C
1-2
Channel set to square root computation
Input value
Scaling
Measured value
Low-cut value
Measured value
Result of square root computation
Input value
Note
The square root computation on the recorder uses the following formula.
min
V - V
F = ( F - F )
x
max
min min
where V
min
(leftmost value of span) < V
F
min
(leftmost value of scale after scaling) < F
scaling)
Vx is the input voltage and Fx is the scaled value
x
max min
V - V
<Related Topics> Setting the input range: Section 5.1
For the procedure to set the functions, see section 1.6, “Function Setup Guide.”
+ F
max
(rightmost value of span)
max
(rightmost value of scale after
IM 04P03B01-01E
1.1 Measuring Input Section
• Bias
A given value (bias value) is added to the input value and used as the measured value of that channel.
Biased channel
Input value
+
Bias value
Measured value
<Related Topics> Setting the bias: Sections 4.15 and 3.10
• Calibration Correction (/CC1 Option)
Corrects the measured value of each channel using segment linearizer approximation and makes the resultant value the measured value of the channel. You can set arbitrary correction values for 2 to 16 points of arbitrary measured values. Linear approximation is used between two segment points. Correction values can be assigned using revise values or absolute values.
Scale value
Measured value
B3
B2
B1
A2
A1
A4
A3
B4
B5
A5
Correction value
Calibration point
Correction using revise values Measured value = A Correction value = B – A
A1 to A5: Measured value (measured value before correction)
B1 to B5: Correction value (measured value after correction)
Correction using absolute values Measured value = A Correction value = B
<Related Topics> Setting the calibration correction function: Sections 4.15, 4.22, and
3.11
1
Functional Explanation and Setup Guide
Burnout Detection of Thermocouples
This function makes the recording go off the scale to the right or left when the thermocouple burns out while measuring temperature with a thermocouple. This function can also be used on 1-5V. The burnout detection function can be set for each channel. By default, this function is disabled.
Note
For 1-5V, a burnout occurs when the input value is less than or equal to 0.2 V.
<Related Topics> Setting the burnout detection function: Section 4.3
Reference Junction Compensation of Thermocouple Input
When measuring the temperature using a thermocouple, the reference junction compensation on the recorder can be used. When using external reference junction compensation, you can set the reference voltage. The reference junction compensation can be set for each channel. By default, the recorder is configured to use the internal reference junction compensation function.
IM 04P03B01-01E
For the procedure to set the functions, see section 1.6, “Function Setup Guide.”
1-3
1.1 Measuring Input Section
Note
When using external reference junction compensation, set an appropriate reference junction
compensation voltage. For example, if the reference junction temperature of the external
reference compensation is T0 °C, set the reference compensation junction voltage to the
thermoelectromotive force of the 0°C reference of T0 °C.
Example when using external reference junction compensation
Recorder
<Related Topics> Setting the reference junction compensation function: Section 4.4
Noise Elimination from Input Signals
Filter and Moving Average
This function used to suppress the effects of noise that is riding on the signal. The pen model and dot model are equipped with a filter function and a moving average function, respectively. The function can be set for each measurement channel. However, it does not operate on channels set to ON/OFF input.
• Filter (Pen Model)
The filter is a low-pass filter. The time constant can be set to 2 s, 5 s, or 10 s.
Filter result (output for a step input)
Copper wire
External reference junction compensation (Hold the contact point of the thermocouple and copper wire at T
0
°C)
Thermocouple
Input signal
2, 5, 10 s (time constant, the time it takes to reach 63.2% of the output value)
63.2% of the output value
Output response curve
(when using the filter)
• Moving Average (Dot Model)
The average value of the m most recent values acquired at the scan interval is used as the measured value of the channel. The number of moving-averaged data points (m) can be set in the range 2 to 16. The figure below shows an example indicating the operation of the buffer for the moving average computation when the number of moving averaged data points is set to 5.
Buffer data for the
n+2th sampling time
10.0 mV
15.0 mV
10.0 mV
5.0 mV
0.0 mV
Deleted
8.0 mV
Moving average
Buffer data for the
nth sampling time
10.0 mV
1
5.0 mV
2
0.0 mV
3
–5.0 mV
4
–10.0 mV
5
0.0 mV
Buffer data for the
n+1th sampling time
Most recent data Most recent data
15.0 mV
10.0 mV
5.0 mV
0.0 mV
–5.0 mV
Deleted
5.0 mV
1-4
<Related Topics> Setting the filter: Sections 4.11 and 3.2
Setting the moving average: Sections 4.10 and 3.3
For the procedure to set the functions, see section 1.6, “Function Setup Guide.”
IM 04P03B01-01E
1.1 Measuring Input Section
Integration Time of the A/D Converter
The recorder uses an A/D converter to convert the sampled analog signal to a digital signal. By setting the integration time of the A/D converter to match the time period corresponding to one cycle of the power supply or an integer multiple of one cycle, the power supply frequency noise can be effectively suppressed. The integration time of the A/D converter is selected according to the model from the table below.
Model Integration Time of the A/D Converter
Pen model Select 16.7 ms (60 Hz), 20 ms (50 Hz), or Auto
Dot model Select 16.7 ms (60 Hz), 20 ms (50 Hz), 100 ms or Auto
• If Auto is selected, the recorder detects the power supply frequency and automatically selects 16.7 ms or 20 ms.
• If Auto is specified when using the 24-VDC power supply on a recorder with the 24­VDC/AC power supply (/P1 option), the integration time is fixed to 20 ms (50 Hz).
• Because 100 ms is an integer multiple of 16.7 ms and 20 ms, this setting can be used to suppress the power frequency noise for either frequency, 50 Hz or 60 Hz.
• The scan interval on the dot model is 1 s when the integration time is set to 16.7 ms or 20 ms and 2.5 s when the integration time is set to 100 ms.
<Related Topics> Setting the A/D integration time: Section 4.2
1
Functional Explanation and Setup Guide
IM 04P03B01-01E
For the procedure to set the functions, see section 1.6, “Function Setup Guide.”
1-5

1.2 Alarms

This function generates an alarm when the measured data meets a certain condition. The alarm status is displayed on the screen while recording the alarm occurrence/ release on the chart paper. Also, alarm output relays can be used to output contact signals when alarms occur (/A1, /A2, and /A3 options).
Alarm Types
Number of Alarm Point Marks
Up to four alarms can be set for each channel.
Alarm Conditions
The following four conditions are available: The alphanumeric character or symbol inside the parentheses is used on the recorder to denote each alarm.
• High Limit Alarm (H/
• Low Limit Alarm (L/
)
An alarm occurs when the input value exceeds the alarm value.
)
An alarm occurs when the input value falls below the alarm value.
High limit alarm
Alarm occurrence
Measured value
Alarm value
Alarm release
Low limit alarm
Measured value
Alarm occurrence
Alarm release
Alarm value
• Difference High Limit Alarm (h/ )*
An alarm occurs when the difference in the input values of two channels is greater than or equal to the specified value.
• Difference Low Limit Alarm (l/
)*
An alarm occurs when the difference in the input values of two channels is less than or equal to the specified value.
* Can be specified on channels set to delta computation.
Alarm Hysteresis
Hysteresis can be specified to the values for activating and releasing the alarm. The hysteresis applies only to high limit alarm (H) and low limit alarm (L). The hysteresis width can be set in the range of 0.0% (OFF) to 1.0% of the recording span in 0.1 steps. The setting applies to all high limit alarms and low limit alarms. By default, the hysteresis width is set to 0.5%.
High limit alarm
Alarm occurrence
Measured value
Alarm value
Alarm release
Low limit alarm
Measured value
Alarm occurrence
Alarm release
Hysteresis (1% or less)
Alarm value
1-6
<Related Topics> Setting alarms: Section 2.2
Setting the alarm hysteresis: Section 4.1
For the procedure to set the functions, see section 1.6, “Function Setup Guide.”
IM 04P03B01-01E
1.2 Alarms
Alarm Indication
When an alarm occurs, the ALM indicator in the status display section illuminates, and
nd
the 2
digit of the LED shows the alarm status for each channel. When the alarm clears,
the indicator and the LED turn OFF.
Alarm Recording
The alarm occurrence/release can be recorded on the chart paper. See section 1.3.
Alarm Output Relay (/A1, /A2, and /A3 Options)
Contact signals can be generated from alarm output relays when alarms occur. The number of output relays is 2 (/A1), 4 (/A2), or 6 (/A3). The alarm output relays are denoted as I01 to I06 on the recorder. The following functions can be assigned to the alarm output relay.
Diagnosis Output
The diagnosis output can be assigned to alarm output relay I01. The relay is activated when there is an error in the plotter operation on the pen model, when a burnout is detected, or when there is an error in the A/D converter. Output relay I01 is normally energized and de-energizes when an error is detected (de-energized operation).
Note
If diagnosis output is enabled, I01 becomes a relay dedicated to diagnosis output.
1
Functional Explanation and Setup Guide
<Related Topics> Setting the diagnosis output: Section 4.1
Energized/De-energized Operation of Alarm Output Relays
You can select whether the alarm output relay is energized or de-energized when an alarm occurs. If de-energized is selected, the status of the alarm output relay when an alarm occurs is the same as the status that results when the recorder is turned OFF (including power failures). The setting applies to all alarm output relays. The default setting is energized.
NO
Energize
De-energize
NO : Normally Opened, C : Common, NC : Normally Closed
CNC NOCNC NO C NC
NC
When power is turned OFF
NO C NC NO C NCNO C
When an alarm is not occurring (During normal operation)
When an alarm is occurring (During a malfunction)
Note
If diagnosis output is enabled, I01 is fixed to de-energized operation.
IM 04P03B01-01E
<Related Topics> Setting the energized/de-energized operation of alarm output relays:
Section 4.1
Alarm Output Relay Operation
When the output destination of multiple alarms is assigned to a single alarm output relay, the relay is activated when any of the assigned alarms is occurring (OR operation).
For the procedure to set the functions, see section 1.6, “Function Setup Guide.”
1-7

1.3 Recording

The recorder is capable of recording the measured values with pens or dots (trend recording) as well as various other types of information.
Trend Recording
The measured values are printed within a width of 100 mm.
Recording Method (Pen Model)
• The measured value is updated every scan interval and continuously recorded.
• The recording colors in order from channel 1 are red, green, blue, and violet.
Recording Method (Dot Model)
• The most recent measured value is recorded with a dot every dot printing interval. The dot printing interval is in the range of 10 s to 90 s. There are two recording methods from which you can select. One method automatically adjusts the dot printing interval according to the chart speed so that the dots do not overlap. The other method records at the fastest dot printing interval at all times.
• The recording colors in order from channel 1 are purple, red, green, blue, brown, and black. The recording color of each channel can be changed among these six colors.
• For each channel, trend recording can be enabled or disabled.
<Related Topics> Setting the trend recording interval: Section 3.1
Changing the recording color: Section 4.5 Enabling/Disabling trend recording for each channel: Section 3.6
Chart Speed
On the pen model, the chart speed can be selected from 40 settings in the range of 10 to 12000 mm/h. On the dot model, the chart speed can be selected from 28 settings in the range of 10 to 1500 mm/h. The default setting is 20 mm/h.
<Related Topics> Setting the chart speed: Section 2.4
Zone Recording
A recording zone is assigned to each channel. This function is useful such when the recording results overlap making them difficult to be viewed.
1-8
Zone 1 Zone 2 Zone 3 Zone 4
<Related Topics> Setting the zone recording: Section 3.4
For the procedure to set the functions, see section 1.6, “Function Setup Guide.”
IM 04P03B01-01E
1.3 Recording
Partial Expanded Recording
This function expands a section of the recording range. By default, partial expanded recording is disabled.
Compressed Expanded
<Related Topics> Setting the partial expanded recording: Sections 4.12 and 3.5
Pen Offset Compensation (Pen Model)
This function compensates for the pen offset (phase difference) along the time axis. On 2-pen, 3-pen, and 4-pen recorders, there are offsets along the time axis (phase difference) between the pens. This offset is corrected when pen offset compensation is used.
1
Functional Explanation and Setup Guide
Same time
Below is an explanation for the 2-pen model. The recording of these two pens are offset by an amount of phase P. If pen offset compensation is enabled, the measured values of pen 1 are stored in the memory, and recorded when the chart paper is fed by an amount corresponding to P.
Reference pen (pen 2)
Pen 1
Recorder front panel
Chart paper
P
Chart feeding direction
By default, this function is disabled.
<Related Topics> Setting the pen offset compensation: Section 4.6
IM 04P03B01-01E
For the procedure to set the functions, see section 1.6, “Function Setup Guide.”
1-9
1.3 Recording
Printout
The figure below is used to explain the printout contents. The actual printout and font are different from those illustrated in the figure. The printout positions are also slightly different.
Printout Example on the Pen Model
Manual printout
Jan.31.05 15:00 1 223.5mg/cm
3
2 437.2µS/cm
3 H 591.6˚C 4d 0.222V
New chart speed printout
50mm/h 14:55
Periodic printout
Time tick cancel mark
Jan.31.05! 13:50* 1 218.7mg/cm 2 390.6µS/cm
Pen offset compensation mark
3
Scale
3 H 598.4˚C 4 d 0.222V
0.0 500.0 1CH mg/cm RED 50mm/h_
Recording color
Alarm Delta computation
Time tick
Buffer overflow mark
3
Alarm printout
1H3*10:09 1H3 10:05
Message printout
09:52*START#205 ABCDEF
Recording start printout
08:00 25mm/h
Time tick
The time ticks are marks that indicate the positions of the date/time on the chart paper.
Time tick cancel mark
An exclamation point (!) is printed when the periodic printout time tick was not printed at the
correct position.
Channel number or tag printout
Channel numbers or tags can be printed.
1-10
<Related Topics>
Switching between channel number printout and tag printout: Section 4.7 Setting the periodic printout Turning printout ON/OFF
printout for periodic printout, and recording color printout for periodic printout)
(interval, reference time, and periodic printout ON/OFF): Section 4.8
(alarm printout, recording start printout, new chart speed printout, scale
: Section 4.7
Setting the time format (alarm printout, message printout, recording start printout, and new chart
speed printout): Section 4.16
Turning periodic printout ON/OFF for each channel: Section 3.6 Setting the message string: Section 3.8 Executing manual printouts, executing message printouts, clearing the alarm printout
buffer, and printing out settings: see the
Operation Guide
)
For the procedure to set the functions, see section 1.6, “Function Setup Guide.”
IM 04P03B01-01E
1.3 Recording
Printout Example on the Dot Model
Manual printout
Jan.31.05 16:00 1 223.5mg/cm
3
2 437.2µS/cm 3 H 591.6˚C 4d 0.222V 5 0.665V 6 L −0.448V
New chart speed printout
Periodic printout
_50mm/h 14:55
Jan.31.05 13:50 1 218.7mg/cm
3
Time tick
2 390.6µS/cm 3 H 598.4˚C 4 d 0.222V 5 0.995V 6 L 0.448V
0.0 500.0 1CH mg/cm 50mm/h_
Delta computation
Alarm
Time tick
Buffer overflow mark
09:52*START#205 ABCDEF
1
Functional Explanation and Setup Guide
Scale
3
Alarm printout
1H3*10:09 1H3 10:05
Message printout
Recording start printout
Channel printout
Channel Printout (Dot Model Only)
Prints the channel No. or tag by the trend recording. The channel No. or tag is printed every
approximately 25 mm on the chart paper. The channel printout can be enabled or disabled. By
default, the channel printout is enabled.
Time tick
_08:00 25mm/h
<Related Topics>
Switching between channel number printout and tag printout: Section 4.7 Setting the periodic printout (interval, reference time, and periodic printout ON/OFF): Section 4.8 Turning printout ON/OFF
speed printout, and scale printout for periodic printout): Section 4.7
(channel printout, alarm printout, recording start printout, new chart
Setting the time format (alarm printout, message printout, recording start printout, and new chart
speed printout): Section 4.16
Turning recording and printout ON/OFF for each channel (trend recording and periodic
printout): Section 3.6
Setting the message string: Sections 3.8 Executing manual printouts, executing message printouts, clearing the alarm printout
buffer, and printing out settings: see the
Operation Guide
)
IM 04P03B01-01E
For the procedure to set the functions, see section 1.6, “Function Setup Guide.”
1-11
1.3 Recording
Channel data
Recording color
Chart speed
Date
Time
Scale
Periodic Printout
Values such as the measured values are printed at a determined interval. The contents of the printout vary between the pen model and dot model. Printout cannot be performed at the following chart speeds.
Pen model: 1800 mm/h or higher; Dot model: 120 mm/h or higher
Channel No. or tag
Measured value
Unit
Time tick cancel mark
Offset compensation mark
Mar.31.2006! 15:50* TAG-001 1.000V 2 -1.000V 3 H-2.1050UNIT03 4 d 2.000V
0.000 2.000 TAG-001 V RED 50mm/h_
Time tick
The font used in the figure differs from that of the actual printout. The printout positions may
also differ from those of the actual printout.
Prints the letter “d” for channels set to delta computation.
Alarm
• Printout Contents
• Date/Time: The date/time when the periodic printout was executed.
• Time ticks: Marks that indicate the first pen position of the date/time on the chart paper. If the time tick is not printed in the correct position, the pen model prints a time tick cancel mark (!), and the dot model does not print the time tick.
• Offset compensation mark: When pen offset is being executed on the pen model,
asterisks () are printed.
• Channel data: Prints the channel numbers or tags, measured values (instantaneous values), and units.
• Alarm status: Prints the alarm that is occurring. Prints the letter “d” on channels set to delta computation. If multiple alarms are occurring, the alarm with the highest precedence is printed. Alarm printout precedence: (Higher) H, L, h, and l (lower)
• Scale: Prints the leftmost and rightmost values of the recording span and the channel number or tag for channels that have scale printout specified. The scale is printed for one channel at each periodic printout. The channel whose scale is printed changes in ascending order. The scale of channels that are being zone recorded is printed within the recording range of the zone for 40 mm or greater.
• Recording color (pen model): Prints the recording colors of channels that have
scale printout and recording color printout specified.
• Chart speed: The chart speed can be printed.
• For the measured values and alarm status, you can select whether to print them for
each channel. For the scale and recording color (pen model), you can select whether to print them.
1-12
• Interval (for details, see appendix 1)
The printout interval can be set by specifying the value or set automatically in sync with the chart speed.
• Turning ON/OFF the Periodic Printout
Periodic printout can be turn ON/OFF. By default, periodic printout is enabled with the interval synchronized to the chart speed.
For the procedure to set the functions, see section 1.6, “Function Setup Guide.”
IM 04P03B01-01E
1.3 Recording
Alarm Printout
Alarm information is printed when an alarm occurs or releases. Printout cannot be performed at the following chart speeds.
Pen model: 1800 mm/h or higher; Dot model: 120 mm/h or higher
Time of alarm occurrence/release
Indicates that there are alarms that are not printed because the alarm printout buffer is full.
Level number
Alarm type (H: high limit, L: low limit, h: difference high limit, and l: difference low limit)
Channel No. or tag
: Alarm occurrence, : Alarm release
• The print condition can be set to (1) print when alarms occur and release, (2) print only when alarms occur, or (3) do not print.
• Alarms that occur while an alarm printout is in progress are temporarily saved to the buffer memory in a printout-wait condition. Alarms are cleared from the buffer memory when they are printed.
• The number alarms that can be stored in the buffer is 8 and 12 on the pen model and dot model, respectively. Alarms that occur while the buffer is full are not printed. A buffer overflow mark is printed when there are alarms that cannot be printed because the buffer is full.
• The time printout format can be selected.
1
Functional Explanation and Setup Guide
Manual Printout
Measured values and alarm status can be printed manually using the keys. When manual printout is executed, trend recording stops and restarts when manual printout is complete. <For the operation procedure, see the
Operation Guide
.>
Message Printout
Printout cannot be performed at the following chart speeds.
Pen model: 1800 mm/h or higher; Dot model: 120 mm/h or higher
Preset messages can be printed on the chart paper using the keys. Five messages, each within 16 characters, can be registered in advance.
• If message printout is executed while another message is being printed, the most recent message is temporarily stored to the buffer memory in a printout-wait condition. Messages are cleared from the buffer memory when they are printed.
• The number of messages that can be stored in the buffer is 5. If message printout is executed when the buffer is full, the message is not printed. A buffer overflow mark is printed when there are messages that cannot be printed because the buffer is full.
• The time printout format can be selected.
New Chart Speed Printout
Printout cannot be performed at the following chart speeds.
Pen model: 1800 mm/h or higher; Dot model: 120 mm/h or higher
• When the chart speed is changed, the time tick (dot model), the date/time of change, and the new chart speed are printed. An asterisk () shows there are messages that cannot be printed.
• The time printout format can be selected.
IM 04P03B01-01E
For the procedure to set the functions, see section 1.6, “Function Setup Guide.”
1-13
1.3 Recording
Recording Start Printout
Printout cannot be performed at the following chart speeds.
Pen model: 1800 mm/h or higher; Dot model: 120 mm/h or higher
When recording is started, the time tick (dot model), the time, and the chart speed can be printed. An asterisk () shows there are messages that cannot be printed.
• The recording start printout can be enabled or disabled. By default, the recording start printout is disabled.
• The time printout format can be selected.
Printout/Display Format of the Date.
The printout/display format of the date can be selected from the list below.
Selectable Type Printout Format Display Format Notes Settings Example Example
Year/Month/Day 2006/03/31 06 03 31 Default value Month/Day/Year 03/31/2006 03 31 06 Day/Month/Year 31/03/2006 31 03 06 Day.Month.Year 31.03.2006 31 03 06 Month.Day.Year Mar.31.2006 03 31 06
*1 These do not apply for the date printout format for message printouts that include measured
values (/BT1 option). Specify that in the message format.
Printout Format of the Time
The printout format of the time can be selected from the list below.
Selectable Type Printout Format Notes Settings Example
Hour:Minute 10:00 Default value Hour:Minute:Second 10:00:00 Month/Day Hour:Minute 03/31 10:00 Month/Day Hour:Minute:Second 03/31 10:00:00 Year/Month/Day Hour:Minute:Second 2006/03/31 10:00:00
*1: The year/month/day format varies depending on the printout/display format of the date. *2: Can be set to the alarm printout, message printout, recording start printout, and new chart
speed printout.
*3 These do not apply for the date printout format for message printouts that include measured
values (/BT1 option). Specify that in the message format.
1-14
<Related Topics> Setting the printout/display format of the date: Section 4.14
Setting the printout format of the time: Section 4.16
For the procedure to set the functions, see section 1.6, “Function Setup Guide.”
IM 04P03B01-01E
1.3 Recording
Setting Printout
List or setup list can be printed. When setting printout is executed, trend recording stops and restarts when the printout is complete. List printout contains Setting Mode settings such as the input range and alarm for each channel. Setup List contains Basic Setting Mode settings such as the alarm output relay operation and printout method.
• Printout Example of List on the Pen Model
1
Functional Explanation and Setup Guide
IM 04P03B01-01E
The printout examples may appear differently from the actual printout as a result of functional
improvements made on the recorder after this manual was written.
For the procedure to set the functions, see section 1.6, “Function Setup Guide.”
1-15
1.3 Recording
• Printout Example of List on the Dot Model
1-16
The printout examples may appear differently from the actual printout as a result of functional
improvements made on the recorder after this manual was written.
For the procedure to set the functions, see section 1.6, “Function Setup Guide.”
IM 04P03B01-01E
1.3 Recording
Header Printout (/BT1 Option)
When recording is started, the Start printout is performed, and recording starts. During trend recording, you can print out messages (up to 5) that include measured values. When recording is stopped, End printout is performed.
Operation
(Operation
instruction)
• Example Printout (Dot Model)
1. Chart paper feed
2. Start printout
3. Recording start (Record start)
Message printout 1 including measured values (Message printout 1)
End printout
Comment Batch name (batch number and lot number) date/time, and chart speed
Message printout 2 including measured values (Message printout 2)
Message printout 2 including measured values
1
Functional Explanation and Setup Guide
1. Recording stop
2. End printout
3. Chart paper feed (Recording stop)
Message printout 1 including measured values
Start printout
Comment Batch name (batch number and lot number) date/time, and chart speed
The printout examples may appear differently from the actual printout as a result of functional
improvements made on the recorder after this manual was written.
Start Printout and End printout
You can set “Start printout” and “Start printout 2” for the printout when recording starts. Also, you can set “End printout” and “End printout 2” for the printout when recording stops.
Printout/Operation Description Notes
Comment Prints 32 characters x 5 lines or less.
Batch name
Batch number Prints up to 26 characters.
Lot number Prints a number from 4-digits or 6-digits. You can automatically increment
Date/time The date format prints out according to the date Date and time cannot be turned
printout/display format. On/Off independently.
Chart speed Prints the current chart paper feed speed.
Chart paper feed Feeds the chart paper 50 mm or less before Start printout. Steps of 1 mm
Feeds the chart paper 50 mm or less after End printout. Steps of 1 mm
Ejection of pen offset You can record the portion of the data that remains after When pen offset compensating compensating data recording stops. Also, when recording the remaining is On (pen model).
portion of the data, you can change the chart speed to 450 mm/h (fixed).
by 1 when recording stops.
IM 04P03B01-01E
You can select whether to print out the batch name, date/time, and chart speed. By default, the printout is enabled.
For the procedure to set the functions, see section 1.6, “Function Setup Guide.”
1-17
1.3 Recording
• Switching between Start Printout and Start printout 2, and between End printout and End printout 2
With the remote control function (/R1 option), you can change the items that are printed out. For example, when a process ends successfully, End printout is performed and the lot number is updated. If the process fails, you can have End printout 2 be carried out and the lot number remain not updated.
Depending on the status of the “batch comment switching signal,” the following switches occur when the “record start/stop signal” switches:
Batch Comment Switching Signal Status Record start/Stop Signal Status Level: 0 (Open) Level: 1 (Closed) Upon start Edge (rising) Start printout Start printout 2 Upon stop Edge (falling) End printout End printout 2
Remote Control Signal
Record start/stop signal
Start Stop
Batch comment switching signal
Start printout End printout
Concluded successfully
Start Stop
Start printout End printout 2
Concluded unsuccessfully
Message Printout Including Measured Values
Following the specified message format, the date/time, message strings of the standard function (5 strings of up to 16 characters), and measured instantaneous values are printed out together.
• Up to 5 messages of 35 characters can be entered.
• Messages are printed out in the order in which they are set.
• The specified number of characters specified for standard function messages is used, then if a subsequent character string has been set, it is used next. Also, it can only be used once for the message format.
Message Example
06/30 10:10 Process-1 134.8°C
Measured value on CH1 (no units) Character string set for message 1
of the standard function Date/time
<Related Topics> Setting start printout and stop printout: Sections 4.23 and 3.12
Assigning functions to the remote control input terminals: Section 4.18 Setting the message string: Section 3.8 Enabling the message format: Section 4.23 Regarding the message format: Section 3.13
1-18
For the procedure to set the functions, see section 1.6, “Function Setup Guide.”
IM 04P03B01-01E

1.4 Remote Control Function (/R1 Option)

Specified operations can be carried out by applying remote signals (contact or open collector signals) to the remote control input terminals. There are five remote control input terminals. An action can be assigned to each terminal.
Recorder
Contact
Open collector
Assignable Functions
• Recording start/stop
• Remote input signal: Rising edge signifies start; falling edge signifies stop
• Starts/stops recording.
• Applying a rising edge signal when recording is already in progress produces no effect. Applying a falling edge signal when recording is stopped produces no effect.
• Chart Speed Switch
• Remote input signal: Level
• The chart paper is fed at the secondary chart speed while a level signal is applied to the terminal. The secondary chart speed is set in advance.
1
Functional Explanation and Setup Guide
• Internal Clock Adjustment
• Remote input signal: Trigger
• The internal clock of the recorder is adjusted to the nearest hour depending on the time when the remote signal is applied.
Time When Signal Is Input Adjustment
00 min 00 s to 01 min 59 s Truncates the minutes and seconds.
02 min 00 s to 57 min 59 s The time is not changed.
58 min 00 s to 59 min 59 s Rounds up the minutes and seconds.
• Message 1 Printout to Message 5 Printout
• Remote input signal: Trigger
• Manual Printout
• Remote input signal: Trigger
• Priority to Remote Recording (/BT1 Option)
• Remote input signal: Edge (rising/start or falling/stop)
• Starts/stops recording.
• When started with a remote signal (on a remote signal rise), stop per key operation or communication is disabled.
Example: 10 hours 01 min 50 s becomes 10 hours 00 min 00 s.
Example: 10 hours 59 min 50 s becomes 11 hours 00 min 00 s.
IM 04P03B01-01E
For the procedure to set the functions, see section 1.6, “Function Setup Guide.”
1-19
1.4 Remote Control Function (/R1 Option)
• Switching Batch Comment (/BT1 Option)
• Remote input signal: Level
• Switches between Start printout and Start printout 2, and between End printout and End printout 2 depending on the status of the batch comment switching signal when recording is started/stopped remotely.
• When starting/stopping by key operation, performs Start printout and End printout.
Remote Signal (Edge, Trigger, and Level)
The above actions are carried out on the rising or falling edge of the remote signal (edge), the ON signal lasting at least 250 ms (trigger), or the ON/OFF signal (level).
Rising/Falling edge Trigger Level
Rising
Falling
For contact inputs, the remote signal rises when the contact switches from open to closed and falls when the contact switches from closed to open. For open collector signals, the remote signal rises when the collector signal (voltage level of the remote control terminal) goes from high to low and falls when the collector signal goes low to high.
250 ms or more
250 ms or more
Operates at the secondary chart speed
<Related Topics> Assigning functions to the remote control input terminals: Section 4.18
Setting the secondary chart speed: Section 3.9
1-20
For the procedure to set the functions, see section 1.6, “Function Setup Guide.”
IM 04P03B01-01E

1.5 Other Functions

Key Lock
Key lock is a function that prohibits key operations. When key lock is enabled, pressing keys produces no effect. To release the key lock, a password is entered.
Key Lock Items
Each of the following keys can be included or excluded from the key lock function.
In the case of the FUNC key, each function of the FUNC key can be included or excluded from the key lock function. FUNC key functions: Manual printout, list printout, setup list printout, message printout,
<Related Topics> Setting the key lock function: Section 4.9
Using the key lock function: see the
Customize Menu
The menu can be customized to display only the menus that you use.
• Display only the items that you use on the FUNC key menu.
• Display only the items that you use on the Setting mode menu.
• Lock Basic Setting mode (use a password to enter the mode). The pen position adjustment (pen model) and dot printing position adjustment (dot model) can be set so that they can be used without the password.
Keys that can be locked
printout buffer clear, pen exchange (pen model), and ribbon cassette exchange (dot model)
Operation Guide
)
1
Functional Explanation and Setup Guide
Font
Operation mode
Displays only the FUNC key menu items to be used
<Related Topics> Setting the FUNC key menu: Section 4.19
The characters used in the printout can be set to English, Japanese, German, or French. The characters that are available vary depending on the selected font. English: Alphabet, numbers, and symbols Japanese: Alphabet, numbers, Katakana, and symbols German: Alphabet (German), numbers, and symbols French: Alphabet (French), numbers, and symbols
<Related Topics> Changing the language: Section 4.13
Hold down MENU for 3 s
Setting the Setting mode menu: Section 4.20 Enabling/Disabling the customized menu: Section 4.21
Setting mode
Displays only the setup menu items to be used
Hold down + for 3 s
Password
No password
Basic Setting mode
Pen and dot printing position adjustment
IM 04P03B01-01E
For the procedure to set the functions, see section 1.6, “Function Setup Guide.”
1-21
1.5 Other Functions
DST
Temperature Unit
If the recorder is used in a region that has DST, time can be switched automatically between DST and standard time by setting the date/time when switching from the standard time to DST and the date/time when switching back from DST to standard time. When switching from standard time to DST, the clock is set ahead by 1 hour. When switching back from DST to standard time, the clock is set back by 1 hour.
<Related Topics> Using the DST: Section 3.14
The temperature unit can be set to Celsius or Fahrenheit. The setting applies to all channels.
<Related Topics> Changing the temperature unit: Section 4.24
1-22
For the procedure to set the functions, see section 1.6, “Function Setup Guide.”
IM 04P03B01-01E

1.6 Function Setup Guide

This section explains the settings necessary to use various functions of the recorder. Read the section corresponding to the function you wish to use.
Note
This section contains all the settings related to each item. If the desired setting is the same as
the default value, you do not have to set it.
1
Functional Explanation and Setup Guide
Item Description Reference
Date/Time setting Use CLOCK in Setting mode 2.5
DST Sets the date/time for switching between DST and standard time using 3.14
AUX > DST in Setting mode.
Setting initialization Use INIT in Basic Setting mode to initialize the settings of Setting mode 4.17
and Basic Setting mode to their default values.
Measuring input functions
Item Description Reference
Range and span of the TC, RTD, or DC voltage
Use RANGE in Setting mode. 2.1
1-5V • Range, span, and scale 2.1
Use RANGE in Setting mode.
• Unit 2.3
Use UNIT in setting mode to set the unit after scaling.
• Low-cut 4.15
Use PERS. > 1-5V low-cut in Basic Setting mode and select Use or Not. If Use is selected, turn ON/OFF the low-cut function using RANGE in Setting 2.1 mode. If Not is selected, the Low-cut item does not appear in the RANGE setting.
Scaling • Range, span, and scale 2.1
Use RANGE in Setting mode.
• Unit 2.3
Use UNIT in setting mode to set the unit after scaling.
Square Root Computation • Range, span, and scale 2.1
Use RANGE in Setting mode.
• Unit 2.3
Use UNIT in setting mode to set the unit after scaling.
• Low-cut 4.15
Use PERS. > SQRT low-cut in Basic Setting mode and select Use or Not. If Use is selected, set the low-cut value using RANGE in Setting mode. 2.1 If Not is selected, the Low-cut item does not appear in the RANGE setting.
Unused channels Use RANGE > SKIP in Setting mode to disabling the trend recording (dot model) 2.1
and periodic printout of the target channel.
Bias Use PERS. > BIAS in Basic Setting mode and select Use or Not. 4.15
If Use is selected, set the bias value that is added to the input using BIAS in 3.10 Setting mode. If Not is selected, the BIAS item does not appear.
Calibration Correction (/CC1 option)
Use PERS. > CALIB in Basic Setting mode and select Use or Not 4.15 If Use is selected,
• Use CALIB in Basic Setting mode to set the correction mode and the number 4.22
of calibration points.
• Use CALIB in Setting mode to set the measured value and correction value 3.11
for each channel.
If Not is selected, the CALIB item does not appear.
Burnout detection function (TC input and 1-5V input) 4.3
Use B_OUT in Basic Setting mode to set the burnout detection function for each channel.
Section
Section
IM 04P03B01-01E
1-23
1.6 Function Setup Guide
Item Description Reference
RJC of TC input Use RJC in Basic Setting mode to select whether to use the internal 4.4
RJC function or an external RJC function.
Filter (pen model) Use FILTR in Basic Setting mode and select Use or Not. 4.11
If Use is selected, set the filter time constant using AUX > FILTR in Setting mode. 3.2 If Not is selected, the AUX > FILTR item does not appear.
Moving average (dot model)
Use M_AVE in Basic Setting mode and select Use or Not. 4.10 If Use is selected, set the number of samples of moving average using AUX > 3.3 M_AVE in Setting mode. If Not is selected, the AUX > M_AVE item does not appear.
Integration time of the A/D converter 4.2
Use INTG in Basic Setting mode to set the integration time of the A/D converter.
Temperature Unit Select the temperature unit using TEMP in Basic Setting mode. 4.24
Alarm functions
Item Description Reference
Alarms for each channel Use ALARM in Setting mode. 2.2
Set a hysteresis on the alarm occurrence/release value of high limit alarm and low limit alarm
Use ALARM > HYS in Basic Setting mode to set the hysteresis to be applied 4.1 to the high limit alarm and low limit alarm of measurement channels.
Diagnosis output Use ALARM > DIAG in Basic Setting mode to set the function. 4.1
Change the alarm output relay operation
If ALARM > RELAY in Basic Setting mode set to “DE_EN”, alarm output relay is 4.1 energized during normal operation and de-energized when an alarm occurs.
Section
Section
Recording functions
Item Description Reference
Section
Recording interval (dot model)
Change the recording color of measurement channels (dot model).
Turn trend recording ON/OFF (dot model)
Chart speed Use CHART in Setting mode to set the chart speed. 2.4
Record by setting recording zone
Partial expanded recording Use PART in Basic Setting mode and select Use or Not. 4.12
Record by compensating for the pen offset along the time axis (pen model)
Date format Use DATE in Basic Setting mode to set the printout/display format 4.14
Time format Use T_PRN in Basic Setting mode to set the printout format of the time. 4.16
Use AUX > TREND in Setting mode to set the recording interval to AUTO or FIX. 3.1
Use COLOR in Basic Setting mode to set the recording color of measurement 4.5 channels.
Use AUX > PRINT in Setting mode to turn ON/OFF the trend recording for each 3.6 measurement channel.
Use AUX > ZONE in Setting mode to set the recording zone for each channel. 3.4
If Use is selected, set the display mode using AUX > PART in Setting mode. 3.5 If Not is selected, the AUX > PART item does not appear.
Use POC in Setting mode to turn ON/OFF offset compensation. 4.6
of the date.
1-24
IM 04P03B01-01E