The cable is six wire multicore and covered
with a thermoplastic TPE. The wires are also
covered with thermoplastic TPE individually
and coloured. The cable connections
are supplied with 2 mm terminations for
connection to the transmitter, or connection
box: these pins quarantee a correct and
simple connection to the terminals.
11 Temperature
12 Temperature
13 Outer electrode
14 Outer electrode
15 Inner electrode
16 inner electrode
Figure 2-0. Sensor Wiring
At the instrument end the cable termination
tags are numbered to correspond with their
respective terminals of the transmitters.
NOTE:
If the cable is used with a connection box,
type BA10 then special purpose cable, type
WF10 must be used between connection
box and transmitter connected to the same
terminals as the cell cable. A maximum of
50 metres total cable length is permissable.
2-2. Mounting
The mounting is by a compression gland or
sanitary adapter, giving a simple effective
method of direct insertion in process
pipework. For mounting the retractable
sensor refer to the instruction manual of the
PR4A retractable holder. Before mounting
a cell in a process plant environment the
following points should be considered:
ssecorp fo wolf ,evitatneserper ,doog a liquid through the cell should exist. Hence,
the cell must
be mounted in the process
IM 12D7J4-E-E
in such a way that the flow through it
represents the true composition of the
liquid. The flow through the cell should be
uninterrupted and the cell should not be
mounted at a “dead” angle.
ssecorp eht ni desremmi eb tsum llec a liquid to a level above the outlet to ensure
an uninterrupted liquid path between the
electrodes.
to allow easy removal for
maintenance. It is recommended that
the cell is mounted in a ‘by-pass’ directly
behind a drain valve.
2-3. Safety warning
For SC4A..AD to avoid the sensor
shooting out, never install in application
where pressure can exceed specified
maximum. Never loosen mounting nut while
system is pressurised.
2-4. Commisioning
The sensor cable is marked with a label
specifying complete MS mode, Serialnumber
and calibrated cell constant*. This cell
constant must be entered in the EXA
transmitter to achive an accurate
measurement.
* Always specify complete MS code
and serialnumber in communication with
Yokogawa organisation with regards to
performance issues or warranty matters.
C
A
E
B
screen
D
overall
shield
Figure 2-1. WF10 Cable
yaw a hcus ni detnuom eb dluohs llec eht -
3. OPERATION AND MAINTENANCE
7
3-1. Checking the cell
If a fault occurs, first check the cell for
visible damage. If damage is not apparent, check the cell resistances measuring
between the cable connections.
a. 11 and 12 (or connector pins 1 and 2)
b. 11 and 13 (or connector pins 1 and 3)
c. 12 and 13 (or connector pins 2 and 3)
d. 12 and 15 (or connector pins 4 and 5)
e. 15 and 16 (or connector pins 5 and 6)
f. 13 and 14 (or connector pins 3 and 4)
g. 13 and 15 (or connector pins 3 and 5)
Celltype a, b, c, d, g e, f
SC4A Pt1000 value > 100MΩ< 2Ω
If the cell has become fouled resistance will
be developed across the electrodes. These
resistances can be measured by making
contact directly with the electrode and the
relevant pin of the connector.
3-2. Calibration of the cell
The cell constant is determined under
laboratory conditions. It is possible that
slight variations in the cell constant may
occur dependant on the installation.
This possible error can be corrected by
recalibrating with a solution of known
conductivity value. The procedure is
descibed in ASTM D1125 or OIML
recommendation 56. Sensors with
long
cables may exhibit an offset in the
temperature readings. This can be
adjusted in the transmitter or converter.
A calibrated thermometer should be used
as reference measurement.
3-3. Cleaning of the cell
If the cell becomes fouled an insulating
layer may be formed on the electrodes and
consequently, an apparent increase in cell
constant may accur, giving a measuring
error. This error is:
2 •x 100 %
(Rv)
(Rcel)
Where:
Rv = the resistance of the fouling layer
Rcel = the cell resistance.
Cleaning methods
htiw retaw toh snoitacilppa lamron roF .1
domestic washing-up liquid added will
be effective.
%01...5 a .cte ,sedixordyh ,emil roF .2
solution of hydrochloric acid is
recommended.
dna dica cirolhcordyh esu reveN *
domestic bleach simultaneously.
The liberation of the very poisonous
gas Chlorine will result.
eb nac ).cte ,staf ,slio( sgniluof cinagrO .3
NOTES:
The specific conductivity value of the
solution of known value must be near the
value of the liquid to be measured. Specific
conductivity is highly temperature
dependent, therefore in the above
calibration procedure the temperature of
the cell and the liquid should be allowed
to equalise and should be accurately
measured with a calibrated thermometer.
IM 12D7J4-E-E
8
3-4. SPARE PARTS AND ACESSOIRES
3-4-1- Spare parts and acessories for SC4A-S- AD and SC4A-T-AD
K1542DF Compression fitting in Stainless Steel ( /PS)
K1500DY Ferrule set for K1542DF
K1542CW Compression fitting in PVDF (/PF)
K1598AC Flowfitting with 3.1 material certificate
3.4.2. Spare parts and acessories for SC4A-E-SA
K1542FA Straight welding socket and mounting nut Stainless Steel
K1500BJ O-rings in EPDM (5), FDA migration tested
3.4.3. Spare parts and acessories for SC4A-E-SB
K1500HN Sealing gasket in EPDM for K1542FC (3), FDA migration tested
(welding socket, sealing gasket and clamp ring) (/SB1)
(welding socket, sealing gasket and clamp ring) (/SB2)
K1500BN Sealing gasket in EPDM for K1542FF (3), FDA migration tested
3.4.3. Spare parts and acessories for SC4A-E-SC
(welding socket, sealing gasket and clamp ring) (/SC1)
K1500BP Sealing gasket in EPDM for K1542FE (3), FDA
3.4.4. Spare parts and acessories for SC4A-T-PR and SC4A-S-PR
K1500BE O-rings in Viton (10)
K1500ED O-rings in Kalrez (2)
pmalcirt ”1 rof tes gnitnuom pmalC-irT CF2451K
pmalcirt ”5,1 rof tes gnitnuom pmalC-irT FF2451K
pmalcirt ”2 rof tes gnitnuom pmalC-irT EF2451K
migration tested
IM 12D7J4-E-E
9
IM 12D7J4-E-E
YOKOGAWA
YOKOGAWA EUROPE B.V.
Euroweg 23825 HD AMERSFOORT
The NetherlandsTel. +31-88-4641 000Fax +31-88-4641 111www.yokogawa.com/eu
Yokogawa has an extensive sales and
distribution network.
Please refer to the European website
(www.yokogawa.com/eu) to contact your
nearest representative.
Printed in The Netherlands
07-901 (A) I
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.