Yokogawa SC25V User Manual

Instruction
YOKOGAWA
Manual
Model SC25V Combined 12mm sensor; pH, Ref, LE and Temperature
IM12B6J1-40E-E
2nd edition
2
Table 1
Production year according to DIN. IEC 60062
Year Yearcode
2010 A 2011 B 2012 C 2013 D 2014 E 2015 F 2016 G 2017 H 2018 J 2019 K 2020 L 2021 M 2022 N
IM12B6J1-40E-E
Year Yearcode
2023 P 2024 Q 2025 R 2026 S 2027 T 2028 U 2029 V 2030 W 2031 X 2032 A 2033 B 2034 C 2035 D
Table 2
Production month according to DIN. IEC 60062
Month Monthcode
January 1 February 2 March 3 April 4 May 5 June 6 July 7 August 8 September 9 October O November N December D
TABLE OF CONTENTS
1. Preface 4
1.1 Introduction 4
1.2 ATEX/IECEx compliance 4
1.3 Unpacking and checking 4
1.4 Warranty and service 4
2. General Specifications 5
2.1 Measuring elements 5
2.2 Construction materials 5
2.3 Functional specifications 5
2.4 Operating range 5
2.5 Dynamic specification 5
2.6 Sensor dimensions 6
2.7 Equipment and protective systems 6
3. Installation 7
3.1 Typical installation 7
3.2 Preparing the sensor for use 7
3.3 Mounting the sensor 7
4. Wiring to the pH analyzer 9
4.1 Cable marking and pin allocation 9
4.2 Instrument connection of sensor 9
4.3 Instrument connection setting for PH402G and PH202G(S) 9
4.4 Instrument connection setting for PH150, PH450 and FLXA21 9
4.5 Sensor VP connection 10
3
5. General calibration and maintenance procedure 10
5.1 Calibration for pH measurement 10
5.2 Maintenance 10
5.3 Chemical cleaning HCN 11
6. Model and suffix codes 12
6.1 SC25V sensor 12
6.2 Cable for Sensor with Variopin Connector 12
7. Spare parts 12
8. Appendix 13
Chemical Compatibility Chart 13 EC DECLARATION OF CONFORMITY 14 Application datasheet SC25V 15
IM12B6J1-40E-E
4
1. PREFACE
1.1 Introduction
This instruction manual provides information for the installation and use of the SC25V, combined 12mm glass sensor; pH, Ref, LE and Temperature. The SC25V is the best choice for the majority of typical process­and waste water applications. The sensor is designed with titanium liquid earth, increased electrolyte reservoir and large surface area PTFE diaphragm combined with a long pathway inner junction resulting in a longtime stability and prolonged lifetime.
1.2 ATEX/IECEx compliance
The model SC25V is intended for use in potentially explosive atmospheres. A number of restrictions apply on the use of this sensor. All varieties of SC25V are ATEX certified: II1G Ex ia IIC T3...T6 Ga (see chapter 2.7).
EC-type Examination Certificate Number:
DEKRA 11ATEX0014 X
IECEx-type Examination Certificate Number:
IECEx DEK 11.0064x
Special conditions for safe use:
For ambient temperature see chapter 2.7. Electrostatic charges on the enclosure and cable shall be avoided.
1.3 Unpacking & Checking
Upon delivery, unpack the sensor carefully and inspect it to ensure it was not damaged during shipment. If damage is found, retain the original packing materials (including outer box) and immediately notify the carrier and the relevant Yokogawa sales office. Make sure the Model Code on the sensor is the same as on the packing list. The Model Code and Serial Number are found on the VP-connector of the sensor. Check that option(s) that were ordered are also included.
1.4 Warranty & Service
Yokogawa products and parts are guaranteed free from defects in workmanship and materials under normal use and service for a period of twelve months from the date of shipment from the manufacturer. Individual sales organizations can deviate from the typical warranty period, and the conditions of sales relating to the original purchase order should be consulted. Damage caused by wear and tear, inadequate maintenance, corrosion, or by the effects of chemical processes are excluded from this warranty coverage. In the event of warranty claim, the defective goods should be sent (freight paid) to the Service Department of the relevant sales Organization for repair or replacement (at Yokogawa’s discretion).
The following information must be included in the letter accompanying the returned goods:
• Model Code and Serial Number.
• Original Purchase Order and Date.
• Length of time in service and description of the process.
• Description of the fault and circumstances of the failure.
• Process/environmental conditions that may be related to the failure of the sensor.
• Statement as to whether warranty or nonwarranty service is requested.
• Complete shipping and billing instructions for return of material, plus the name and phone number of a contact person that can be reached for further information.
Returned goods that have been in contact with process fluids must be decontaminated and disinfected prior to shipment. Goods should carry a certificate to this effect, for the health and safety of our employees. Material Safety Data sheets must be included for all components of the process to which the sensors/fittings have been exposed.
IM12B6J1-40E-E
2. GENERAL SPECIFICATIONS
5
2.1 Measuring elements
pH glass electrode Silver/Silver Chloride reference Titanium liquid earth Pt1000 temperature sensor.
2.2 Construction materials Wetted materials:
O-ring : Viton Earth Ring : Titanium Reference Junction : Porous PTFE Measuring Sensor : Glass Body Sensor : Glass, PEEK
2.3 Functional specifications
Isothermal point : pH 7 Glass impedance G-Glass : 225 MΩ Glass impedance L-Glass : 600 MΩ Reference system : Double junction, Ag/
AgCl with saturated KCI, including Ag+ ion trap
Junction resistance : 0.2 to 10 kΩ Temp. element : Pt1000 to IEC 751 Asymmetry potential : 8 ±15 mV Slope : > 96 %
(of theoretical value)
Sensor length : 120 mm and 225 mm
2.4 Operating range
pH : 0 to 14 Temperature
Type G-glass : -10 to 80°C
(14 to 176°F)
Type L-glass : +15 to 130°C
(59 to 266°F) Pressure : 0 to 10 bar (0 to 145 psi)* Conductivity : > 10 μS/cm Storage temp. : -10 to 50°C (14 to 122°F)
* Damaging the screw thread might influence
the max process pressure.
Note: Specifications should not be considered
in isolation. For example the pH range can be 2-12pH, where the measurement is at elevated temperatures. For advice about specific applications please contact your local sales office.
2.5 Dynamic specifications
Response time pH step (7 to 4) : < 5 sec for 90% Response time temp step (25°C) : < 90 sec for 90% Stabilization time (0.02 pH unit/10 s) : < 30 seconds
Note: The temperature sensor included in the
SC25V is intended to provide indication and cell compensation. The construction has not been tested to the pressure vessel standards required for plant temperature control.
IM12B6J1-40E-E
6
2.6 Sensor Dimensions
Dimension according to DIN 19263:2007-05
Criteria : 120 mm version L (below plug head) : 120 ± 2 mm Ø shaft : 11.9 ± 0.1 mm Criteria : 225 mm version L (below plug head) : 225 ± 2 mm Ø shaft : 11.9 ± 0.1 mm
Concentricity : < 0,25 mm (centerline plug
head - 25 mm from tip)
Perpendicularity : < 0,5 mm (plug head
- 25 mm from tip)
Cilindricity : < 0,5 mm (centerline
sensor)
2.7 Equipment and protective systems
intended for use in potentially explosive atmospheres
ATEX II1G Ex ia IIC T3...T6 Ga. EC-type Examination Certificate Number:
DEKRA 11ATEX0014 X
IECEx-type Examination Certificate Number:
IECEx DEK 11.0064x
Description
pH sensor series SC25V for connection to a certified associated pH/ORP transmitter are used for measurement of pH, redox(ORP) and temperature.
Criteria : All types All measuring elements within 25mm from tip.
Wrench 19
PG13.5
YOKOGAWA
Made in the netherlands NL-3825 HD-2
173 / 278
II 1 G EX ia IIC T3/T4/T5/T6 Ga
DEKRA 11ATEX0014X
Ø12
0344
0344
S
-
A
C
5
2
V
X
6
1
-
1
0
0
120 / 225
< Ø11.8
25.0
Ambient temperature range:
-40°C to +40°C for temp. class T6,
-40°C to +55°C for temp. class T4 and T5,
-40°C to +105°C for temp. class T3,
Electrical data
Sensor input circuits (connector): In type of protection intrinsic safety Ex ia IIC, only for connection to a certified intrinsically safe circuit, with the following maximum values: Ui = 18V; Ii =170 mA; Pi = 0,4 W; Ci = 0 nF (connector) or 3,6 nF (permanently connected cable); Li = 0 mH;
Or for connection to the certified intrinsically safe YOKOGAWA pH/ORP Transmitter Model FLXA21 series or Model PH202S series.
Special conditions for safe use:
For ambient temperature see above. Electrostatic charges on the enclosure and cable shall be avoided. pH sensor series SC25V is suitable for use with the PH202S, FLXA21-EA conform to ATEX.
Note: When the sensor has been connected
to none intrinsically safe equipment or equipment which exceeds the restrictions regarding the sensor input circuit, the sensor is not suitable for intrinsically safe use anymore.
Figure 1
IM12B6J1-40E-E
3. INSTALLATION
3.1 Typical installation
The SC25V sensor can be installed in-line, in a bypass loop or in an immersion assembly. For best results the SC25V should be mounted with the process fluid flowing towards the sensor and positioned at least 15° above the horizontal plane to eliminate air bubbles in the glass bulb. See fig. 6 section 3.3
7
SC25V
O-RING VITON 11x3
PR10
3.2 Preparing the sensor for use
Remove the sensor from its shipping box and slide off the “wet pocket” (the flask including the rubber ring). This wet pocket is filled with a weak acid and saline solution to prevent the sensor from drying out during storage and making sure it is ready for immediate use.
Before mounting the sensor, it should be validated with buffer solutions and calibrated when necessary. The specific calibration procedure is described in the Instruction Manual of the pH transmitter, but a general procedure is also provided in section 5 of this manual.
3.3 Mounting the sensor
The SC25V can be mounted using:
• PR10 retractable fitting (see fig 2).
• FF20 flow fitting / FS20 subassembly / FD20 immersion fitting (see fig 4)
• For mounting in Yokogawa fittings, a PG13.5 to M25 adapter is available in different materials (see fig 3) more information in chapter 7.
• PD20 immersion fitting / PF20 flow fitting / PS20 flow fitting (see fig 5)
• FF40 fitting with PG13,5 adapter (K1523JA) or the small flow fitting K1598AC with PG13,5 adapter (K1523JB), (see fig 7).
Figure 2
Figure 3
Figure 4
SC25V
SLIDE RING RYTON
O-RING 11 x 3
SC25V
K1520JN
2 x O-RING 11x3
FD20 / FF20 / FS20
SC25V
Figure 5
SLIDE RING RYTON
O-RING 11 x 3
PD20 / PF20 / PS20
IM12B6J1-40E-E
8
Position
23
30
Position
15°
.
Minimum
Figure 6 Figure 7
ø
PG13.5
PG13,5 adapter
10
5
ø30 ø36
20
K1523JA
1
Flow fitting FF40 with K1523JA Adapter to fit sensors with a PG13,5 process connection in FF40/FS40 and FD40 fittings. Material: Polypropylene.
Conductivity
Flow Fitting
FF40
K1523JA
32
Ø 36
Ø 54
PG13,5 adapter K1523JB
153 (6.0”)
PG 13,5
60
1/4” NPT
3/4” NPT
SC4A
option
Ø30
/FF
100 (3.94”)
1/4” NPT
30 (1.18”)
K1598AC
Flow fitting option /FF K1598AC (incl. 3.1 certificate) with Adapter K1523JB to fit sensors with PG13,5 process connection.
IM12B6J1-40E-E
4. WIRING TO THE PH ANALYZER
4.1 Cable marking and pin allocation
Signal description VP connector (Using WU10-V-S)
pH A Transparent (core coax) pH-shield C Black Reference B Brown (shield coax) Liquid earth D Yellow (cable shield) Temp -1 E Red Temp -2 F Blue
4.2 Instrument connection of sensor
Instrument Terminal for Measurement pH pH&ORP ORP only Signal description pH&rH
pH 15 - 13 pH-shield 16 - 17 Reference 13 13 ­Liquid earth / (ORP) 14 15 15 Temp -1 11 11 11 Temp -2 12 12 12 Jumper - 13-14 14-15
The settings for the pH instruments should follow the instruction Manual of the instrument The following tables summarize these settings:
4.3 Instrument connection settings for PH402G and PH202G(S)
Function pH pH compensated Setting (default)
code 01 0(pH) 0(pH) 0(pH) 1(ORP) 1(ORP) code 02 0(off) 1(ORP) 2(rH) 0(off) 0(off) code 03 1.1.1 1.1.1 1.1.1 0.0.1 0.0.1 code 04 0.0.1 0.0.1 0.0.1 0.0.1 1.1.1 code 31 0 2 2 Jumper input 1 no no no yes yes Jumper input 2 yes yes yes yes no
pH&ORP pH&rH ORP
Cable Wire Colour
pH-compensated
ORP
ORP
9
4.4 Instrument connection settings for PH150, PH450 and FLXA21
Function pH pH compensated Setting (default)
sensor setup pH pH + ORP pH + ORP ORP ORP measurement setup pH + ORP pH+rH Impedance settings input 1 High High High Low Low input 2 Low Low Low Low High jumper input 1 no no no yes yes jumper input 2 yes yes yes yes no
pH&ORP pH&rH ORP
ORP
IM12B6J1-40E-E
10
4.5 Sensor VP connection
VP-Connector
(Temp 2) F
(Temp 1) E
(LE) D
(pH) A
(Ref.) B
Figure 8
5. GENERAL CALIBRATION AND MAINTENANCE PROCEDURE
5.1 Calibration for pH measurement
To calibrate a pH sensor, at least two buffer solutions with known pH values are required. It is recommended to use one buffer solution with a value of pH=7. The second buffer solution should be either acidic (pH<7) or alkaline (pH>7), depending on the process value to be measured. Example buffer solutions are the IEC buffers with pH=4.01, pH=6.87 and pH=9.18.
The following is a general 2-point calibration procedure: 1 Clean the sensor (deposits may be removed
using a 5% HCl-solution).
2 Rinse sensor thoroughly with clean water.
Carefully slap out residual fluid drops.
3 Immerse the sensor in the first buffer (pH=6.87
buffer is recommended).
4 Refer to appropriate Instrument Instruction
Manual for detailed instrument calibration input settings (Auto; Manual; Sample, etc.),
5 Rinse sensor thoroughly with clean water.
Carefully slap out residual fluid drops.
6 Immerse the sensor in the second buffer
(pH=4.01 buffer or pH=9.18 buffer recommended).
7 Repeat Step 3.
Most instruments automatically correct the measured pH value at process temperature to a pH-value at a standard temperature of 25°C. After the calibration is complete, replace or re-install the sensor into the process.
5.2 Maintenance
Cleaning the electrodes
When the sensitivity of the electrode has decreased or the response has slowed down, the electrode should be cleaned. If cleaning with hot water is not sufficient, more aggressive water based agents should be used.
Examples:
• Deposits of lime, hydroxides or carbonates can be removed by immersing the sensor in or a dilute solution of hydrochloric acid (5% is recommended). Afterwards rinse with water.
• Deposits of oils and fats can be removed with hot water in conjunction with a detergent.
When the results are unsatisfactory, a mild (carbonate based) abrasive can be used.
IM12B6J1-40E-E
5.3 Chemical cleaning HCN
The chemical cleaning system is effective in processes where deposits can be removed by a suitable solvent. Good cleaning effects can be obtained from periodically spraying of an electrode. EXA PH450G has a built-in wash timer with programmable wash time, interval time and relaxation time for automatic cleaning of the electrodes, preventing pollution of the pH sensitive parts. After washing, it is possible to check the response time of the electrodes. A built-in (no return) nozzle in the spray unit prevents penetration of the process liquid in the cleaning system. The standardized dimensions allow mounting in all flow-, insertion- and immersion fittings of Yokogawa as well as back-end mounting on the 4-in-one pH/Redox electrodes.
Specifications
Materials Nozzle : Hastelloy O-rings : EPDM rubber Mounting set : Stainless steel Tubing : 1/4” (OD Ø) Nylon tubing Process cond. : Max. 1 MPa (10 bar) at 100 ºC
Part no.
K1547PA : /HCN3, 3-hole flow-, insertion-,
immersion fitting
K1547PB : /HCN4, 4-hole flow-, insertion-,
immersion fitting
Chemical cleaning
Tubing
O-rings (11x3)
11
Figure 9
K1547PP (spare tubing, EPDM)
IM12B6J1-40E-E
12
6. MODEL AND SUFFIX CODES
6.1 SC25V sensor
Model Suffix Code Option code Description
SC25V Combined 12mm sensor: pH, Ref, LE, Temp. Equiped with Variopin connector Sensor type -AGP25 General purpose
-ALP25 High temp: Chemical resistant. Sensor length -120 120 mm
-225 225 mm
6.2 Cable for Sensor with Variopin Connector
Cable Model Suffix code Description
WU10-V-S Cable-Variopin-Single COAX Cable length -02 2 meters
-05 5 meters
-10 10 meters
-15 15 meters
-20 20 meters
7. SPARE PARTS
Sealings
Prod. No. Description
K1500BV O-RINGS EPDM 11X3 (6 PCS.) K1500BZ O-RINGS VITON 11X3 (6 PCS.) K1500GR O-RINGS SILICON 11X3 (8PCS) K1524AA O-ring 10.77X2.62 SILICON 70°sh FP20-R12 Blind plug set for 1-hole FP20-R12M Blind plug set for 1-hole (x12)
Adapters
Prod. No. Description
K1523JA Adapter for mounting Pg13,5-sensors in F*40 K1520JN Adapter M25x1.5 - PG13.5 PVC K1520JP Adapter M25x1.5 - PG13.5 RVS K1500DV Adapter M25x1.5 - PG13.5 PVDF K1523JB Adapter PG13.5 to ¾"NPT K1598AC Flow fitting (3.1), for SC4A (In combination with K1523JB) K9148NA Adapter for mounting Pg13,5-sensors in HA405-120-S3 K9148NB Adapter for mounting Pg13,5-sensors in HA405-120-PP K9148NC Adapter for mounting Pg13,5-sensors in HA405-120-PV
IM12B6J1-40E-E
8. APPENDIX
CHEMICAL COMPATIBILITY CHART
Material
Viton EPDM Ti PTFE PEEK Glass (teflon)
13
Conc.% Sulfiric acid 10 o o o o o o - - - o o o o o o o o o 50 o o o o x - - - - o o o o o o o o o 95 o o o x - - - - - o o o - - - o o o fuming o o o - - - - - - o o o - - - o o o Hydrochloric acid 10 o o o o o o - - - o o o o o x o o o sat. o o o x x x - - - o o o o o x o o o Nitric acid 25 o o x o x - o o o o o o o o o o o o 50 - - - - - - o o o o o o x x x o o o 95 - - - - - - o o o o o o - - - o o o fuming - - - - - - o o o - - - o o o Phosphoric acid 25 o o o o o o x x - o o o o o o o o o 50 o o o o o o x - - o o o o o o o o o 95 x x - o o o x - - o o o o o o o o o Hydrofluoric acid 40 o o o - - - - - - o o o - - - x x x 75 o o x - - - - - - o o o - - - - - ­Acetic acid 10 - - - o o o o o o o o o o o o o o o glacial - - - x x x o o o o o o o o x o o o Formic acid 80 - - - o o x x x - o o o x x x o o o Citric acid 50 o o o o o o x x x o o o o o o o o o Calcium hydroxide sat. o o o o o o o o o o o o o o o o o o Potassium hydroxide 50 o o o o x - o x - o o o o o o o o x Sodium hydroxide 40 x x x o x - x x - o o o o o o o o x Ammonia in water 30 x x x o o o x x - o o o o o o o o x Ammonium chloride sat. o o o o o o o o o o o o o o o o o o Zinc chloride 50 o o o o o o o o o o o o o o o o o o Iron(III) chloride 50 o o o o o o o o o o o o o Sodium sulfite sat. - - - o o o o o o o o o o o o o o o Sodium carbonate sat. o o o o o o o o o o o o o o o o o o Potassium chloride sat. o o o o o o o o x o o o o o o o o o Sodium sulfate sat. o o o o o o o o o o o o o o o o o o Calcium chloride sat. o o o o o o o o o o o o o o o o o o Sodium chloride sat. o o o o o o o o o o o o o o o o o o Sodium nitrate 50 o o o o o o o o o o o o o o o o o o Aluminium chloride sat. o o o o o o o o x o o o o o o o o o Hydrogen peroxide 30 o o o o o x o o o o o o o o o o o o Sodium Hypochloride 50 o o x o o o x - - o o o o o o o o o Potassium dichromate sat. o o o o o o o o o o o o o o o o o o Chlorinated lime o o o x x x o o o Ethanol 80 x - - o o o o o o o o o o o o o o o Cyclohexane o o o - - - o o o o o o o o o o o o Toluene - - - - - - o o o o o o o o o o o o Trichloroethane x x x - - - o o o o o o o o o o o o
Organic Oxidizing Neutral Basic Acid Alkali Organic Introganic
acid agent salt salt salt acid acid
Water o o o o o o o o o o o x o o o o o o
o = can be used / x = shortens useful life / - = cannot be used / Blank = no data currently available Note: Information in this list is based on our general experience and literature data and given in good
faith. However Yokogawa is unable to accept responsebility for claims related to this information.
Temp. °C2060
1002060
1002060
1002060
1002060
1002060
IM12B6J1-40E-E
100
14
EC DECLARATION OF CONFORMITY
IM12B6J1-40E-E
APPLICATION DATA SHEET SC25V COMBINED PH SENSOR
Yokogawa has an extensive sales and distribution network. Please refer to the European website
(www.yokogawa.com/eu)
to contact your nearest representative.
A5 formaat
YOKOGAWA ELECTRIC CORPORATION World Headquarters
9-32, Nakacho 2-chome, Musashino-shi Tokyo 180-8750 Japan www.yokogawa.com
YOKOGAWA CORPORATION OF AMERICA
2 Dart Road Newnan GA 30265 USA www.yokogawa.com/us
YOKOGAWA ELECTRIC ASIA Pte. LTD.
5 Bedok South Road Singapore 469270 Singapore www.yokogawa.com/sg
YOKOGAWA CHINA CO. LTD.
3F Tower D Cartelo Crocodile Building No.568 West Tianshan Road Changing District Shanghai, China www.yokogawa.com/cn
YOKOGAWA MIDDLE EAST B.S.C.(c)
P.O. Box 10070, Manama Building 577, Road 2516, Busaiteen 225 Muharraq, Bahrain www.yokogawa.com/bh
Customer Please complete a separate form for each Address process stream to be analyzed and return, marked for the attention of: Contact Telephone Fax
Serial no. Date received Date returned
1. Process data
Description
15
Application
Operation
Type of solution
pH Value Nor Min Max Temperature ( ºC, ºF) Nor Min Max Pressure (psig, bar, kg/cm²) Nor Min Max Flow Rate (m/s : ft/s) Nor Min Max Conductivity (mS/cm) Nor Min Max
Concentration of salts Concentration of organics Type(s) Type(s)
Problems with Present measuring procedure
Description
Yokogawa Confidential
Pulp & Paper Power Pharmaceutical
Batch
Fouling Other
Continuous Monitor
Poisoning
Chemical Other
Control
Coating
YOKOGAWA
Euroweg 2 3825 HD Amersfoort The Netherlands www.yokogawa.com/eu
IM12B6J1-40E-E
A5 formaat
YOKOGAWA ELECTRIC CORPORATION World Headquarters
9-32, Nakacho 2-chome, Musashino-shi Tokyo 180-8750 Japan www.yokogawa.com
YOKOGAWA CORPORATION OF AMERICA
2 Dart Road Newnan GA 30265 USA www.yokogawa.com/us
Euroweg 2 3825 HD Amersfoort The Netherlands www.yokogawa.com/eu
YOKOGAWA ELECTRIC ASIA Pte. LTD.
5 Bedok South Road Singapore 469270 Singapore www.yokogawa.com/sg
YOKOGAWA CHINA CO. LTD.
3F Tower D Cartelo Crocodile Building No.568 West Tianshan Road Changing District Shanghai, China www.yokogawa.com/cn
YOKOGAWA MIDDLE EAST B.S.C.(c)
P.O. Box 10070, Manama Building 577, Road 2516, Busaiteen 225 Muharraq, Bahrain www.yokogawa.com/bh
Yokogawa has an extensive sales and distribution network. Please refer to the European website
(www.yokogawa.com/eu)
to contact your nearest representative.
YOKOGAWA
IM12B6J1-40E-E 02-1106 (G) I Subject to change without notice Printed in The Netherlands Copyright ©
Loading...