Thank you for purchasing the DL350 ScopeCorder. This User’s Manual explains how to use the
DL350. To ensure correct use, please read this manual thoroughly before operation.
After reading this manual, keep it in a safe place.
List of Manuals
The following manuals, including this one, are provided as manuals for the DL350. Please read all
manuals.
Manual TitleManual No.Description
DL350 ScopeCorder Features GuideIM DL350-01EN The supplied CD contains the PDF file of this manual.
This manual explains all the instrument’s features other
than the communication interface features.
DL350 ScopeCorder User’s ManualIM DL350-02EN This manual. The supplied CD contains the PDF file of
this manual. The manual explains how to operate this
instrument.
DL350 ScopeCorder Getting Started
Guide
DL350 ScopeCorder Communication
Interface User’s Manual
Precautions Concerning the ModulesIM 701250-04EThe manual explains the precautions concerning
Battery Pack Handling PrecautionsIM 739883-01EN This manual is included in models with the /EB option
DL350 ScopeCorderIM DL350-92Z1 Document for China
739883 Battery PackIM 739883-92Z1 Document for China
720923 Battery Pack CoverIM 720923-92Z1 Document for China
The “EN”, “E”, and “Z1” in the manual numbers are the language codes.
IM DL350-03EN This guide explains the handling precautions and basic
operations of this instrument.
IM DL350-17EN The supplied CD contains the PDF file of this manual.
The manual explains the instrument’s communication
interface features and instructions on how to use them.
the modules. This manual is included if you ordered
modules.
(battery pack + battery pack cover). It explains the
handling precautions of the battery pack.
This manual is included in models with the /EB option
(battery pack + battery pack cover).
This manual is included in models with the /EB option
(battery pack + battery pack cover).
Contact information of Yokogawa offices worldwide is provided on the following sheet.
Document No.Description
PIM 113-01Z2List of worldwide contacts
Notes
• The contents of this manual are subject to change without prior notice as a result of continuing
improvements to the instrument’s performance and functions. The figures given in this manual may
differ from those that actually appear on your screen.
• Every effort has been made in the preparation of this manual to ensure the accuracy of its
contents. However, should you have any questions or find any errors, please contact your nearest
YOKOGAWA dealer.
• Copying or reproducing all or any part of the contents of this manual without the permission of
YOKOGAWA is strictly prohibited.
• The TCP/IP software of this product and the documents concerning it have been developed/created
by YOKOGAWA based on the BSD Networking Software, Release 1 that has been licensed from
the Regents of the University of California.
• Microsoft, Internet Explorer, Windows, Windows 7, Windows 8, Windows 8.1, and Windows 10 are
registered trademarks or trademarks of Microsoft Corporation in the United States and/or other
countries.
• Adobe and Acrobat are either registered trademarks or trademarks of Adobe Systems Incorporated.
• PIEZOTRON is a registered trademark of Kistler Instrumente AG.
• ICP is a registered trademark of PCB Piezotronics Incorporated.
• Isotron is a registered trademark of Meggitt Group, PLC.
• VJE is a registered trademark of Yahoo Japan Corporation.
• MATLAB is a registered trademark of The MathWorks, Inc. in the United States.
• ScopeCorder and GIGAZoom ENGINE are registered trademarks of Yokogawa Electric Corporation.
• In this manual, the ® and TM symbols do not accompany their respective registered trademark or
trademark names.
• Other company and product names are trademarks or registered trademarks of their respective
holders.
Revisions
• 1st Edition: July 2017
• 2nd Edition: December 2017
• 3rd Edition: April 2018
ii
IM DL350-02EN
Page 4
Conventions Used in This Manual
Notes and Cautions
The notes and cautions in this manual are categorized using the following symbols.
Improper handling or use can lead to injury to the user or damage to the
instrument. This symbol appears on the instrument to indicate that the user must
refer to the user’s manual for special instructions. The same symbol appears
in the corresponding place in the user’s manual to identify those instructions.
In the manual, the symbol is used in conjunction with the word “WARNING” or
“CAUTION.”
WARNING
CAUTION
Calls attention to actions or conditions that could cause serious or fatal injury to
the user, and precautions that can be taken to prevent such occurrences.
Calls attention to actions or conditions that could cause light injury to the user or
damage to the instrument or user’s data, and precautions that can be taken to
prevent such occurrences.
French
AVERTISSEMENT
ATTENTION
Calls attention to information that is important for the proper operation of the
Note
Attire l’attention sur des gestes ou des conditions susceptibles de
Attire l’attention sur des gestes ou des conditions susceptibles
de provoquer des blessures graves (voire mortelles), et sur les
précautions de sécurité pouvant prévenir de tels accidents.
provoquer des blessures légères ou d’endommager l’instrument ou les
données de l’utilisateur, et sur les précautions de sécurité susceptibles
de prévenir de tels accidents.
This manual explains procedures using English menu screens. In procedural explanations, applicable
menus and items are indicated with bold characters (e.g., Setup).
For instructions on how to set the menu language, see section 20.3.
Touch Panel Operations
The basic touch panel operations are described below.
Tap
Tap refers to the act of gently hitting the screen with your finger.
Tapping is used on the instrument screen to select areas with a mark, close a setup menu, and so on.
Drag, Swipe, and Slide
Press your finger against the screen and move your finger across the screen.
Drag refers to the act of selecting and moving items, such as a trigger level icon.
Swipe refers to the act of moving a relatively wide display range, such as scrolling the setting screen.
Slide is also a term sometimes used depending on the movement operation.
Pinch Out and Pinch In
Pinch out refers to the act of pressing two fingers against the screen and spreading them apart. Pinch
in refers to the act of pressing two fingers against the screen and drawing them together.
On a screen displaying waveforms, you can pinch out to zoom in and pinch in to zoom out.
iv
IM DL350-02EN
Page 6
Entering Values
• Holding down + or - changes the value consecutively.
Tap to enter an exponent.
When a control for entering a value is displayed (such as the one shown below), tapping the value
area opens an input box (numeric keypad). After entering a value with the numeric keypad, tap Enter
to confirm the value.
Example of a Control for Entering a Value
• The least significant digit increases or decreases. Carrying over and
borrowing occur automatically.
Example of an Input Box (Numeric keypad)
Operation Screen and Common Operations
Tap the number area.
Tap + or - to adjust.
Tap here to set the default value and close the input box
(numeric keypad).
(You do not need to tap Enter.)
IM DL350-02EN
v
Page 7
Character insertion position
Moves the character insertion position
Switches between uppercase and lowercase
Operation Screen and Common Operations
Entering Character Strings
Use the keyboard that appears on the screen to enter character strings such as file names and
comments.
The keyboard used in the next example has the message language (see section 20.3) set to English.
How to Operate the Keyboard
With the keyboard displayed, tap the character you want to enter.
1.
Repeat step 1 to enter all of the characters in the string.
2.
Tap ENTER. The character string is confirmed, and the keyboard disappears.
3.
Deletes the previous character
Note
• @ cannot be entered consecutively.
• File names are not case-sensitive. Comments are case-sensitive. The following file names cannot be
AUX, CON, PRN, NUL, CLOCK, COM1 to COM9, and LPT1 to LPT9
Confirms the characters that you have
entered
Enter symbols.
used due to MS-DOS limitations:
vi
IM DL350-02EN
Page 8
of functions
Top menu
This is the entrance to the setup menu of each function.
of functions
Top menu
This is the entrance to the setup menu of each function.
Operation Screen and Common Operations
Menu Shown in the Footer Area
A menu is shown in the footer area of the waveform screen. You can set various items by using the
menu, list (options), or input box that appears when you tap the menu commands.
For Scope Mode
When the Time Base1 Is Internal
Help
Displays explanations
When the Time Base1 Is External
Help
Displays explanations
1 For setting the time base in scope mode, see section 3.1.
2 The sample rate varies depending on the time scale and record length settings.
Time scale ► section 3.1
(Can be adjusted in steps)
Record length
► section 3.1
This operation is not possible.
Displays time
base External
Record length
► section 3.1
Sample rate
Pulse/rotation
► section 3.1
2
Trigger
► chapter 4
Trigger
► chapter 4
Custom menu
► section 20.5
All channel settings
► section 2.1
Custom menu
► section 20.5
All channel settings
► section 2.1
IM DL350-02EN
vii
Page 9
Top menu
This is the entrance to the setup menu of each function.
Displays explanations of functions
Top menu
This is the entrance to the setup menu of each function.
Displays explanations of functions
Top menu
This is the entrance to the setup menu of each function.
Displays explanations of functions
Top menu
This is the entrance to the setup menu of each function.
Displays explanations of functions
Operation Screen and Common Operations
For Recorder Mode
When the Acquisition Method* Is Set to Memory, Memory + Save on Stop, or
Memory + SD Numeric Recording
• When the Time Base* Is Internal
• When the Time Base* Is External
When the Acquisition Method* Is Set to SD Recording
This section explains how to select scope mode or recorder mode.
• Navigation
• Scope Mode and Recorder Mode
► Features Guide: “Navigation”
Navigation Screen
When you start the instrument, the navigation screen appears.
Tap the measurement mode and then Start. The waveform screen for the selected mode appears.
Navigation Screen
Tap Scope Mode (Scope Mode).
If you select this check box, the waveform screen will appear the next time you start
the instrument. Tapping MENU > Navigation on the waveform screen displays this
navigation screen.
Tap Recorder Mode (Recorder Mode).
IM DL350-02EN
1-1
Page 14
1.2 Loading Setup Files
This section explains how to use the navigation feature to load a setup file according to your
application.
• Navigation
• Setup file
• Executing the load
Navigation Screen
When you start the instrument, the navigation screen appears.
Tap Load File and then Start. A file list appears.
Navigation Screen
► Features Guide: “Navigation”
If you select this check box, the waveform screen will appear the next time you start
the instrument. Tapping MENU > Navigation on the waveform screen displays this
navigation screen.
Tap Load File (Load File).
1-2
IM DL350-02EN
Page 15
Selecting a Setup File
For instruction on how to use
each item, see section 6.10.
On the file list, tap the setup file you want to load.
1.
Tap Load. A confirmation message appears.
2.
Storage media (drive)
List sorting
Display format
Type of file to list
Displays file information
Create folders (directories).
1.2 Loading Setup Files
Tap the setup file to load.
Executing the Load
Tap OK. The setup file is loaded.
3.
File utility
File list
Slide to display files and folders
that do not fit in the screen.
Then, tap Load.
IM DL350-02EN
1-3
Page 16
1.3 Configuring the Recorder Mode Using the Easy
Setup
This section explains how to use the wizard of the navigation feature to configure the recorder mode
settings.
• Navigation
• Turning recording channels on and off
• Recording time
• Sample interval
• All channels setup menu
► Features Guide: “Navigation”
Navigation Screen
When you start the instrument, the navigation screen appears.
Tap Easy Setup and then Start. A Select Channel screen appears.
Navigation Screen
If you select this check box, the waveform screen will appear the next time you start
the instrument. Tapping MENU > Navigation on the waveform screen displays this
navigation screen.
• Select the check boxes of the channels you want to record. Clear the check boxes otherwise.
• For the detail settings of the instrument’s internal logic, see section 2.9. For the detail settings of the
GPS, see chapter 18.
Tap Next. A Recording time screen appears.
2.
Tapping Cancel displays the waveform screen.
Channel Selection Screen
1.3 Configuring the Recorder Mode Using the Easy Setup
4-CH module (720254 (4CH 1M16))
IM DL350-02EN
1-5
Page 18
Number of channels that was turned on in the previous
1.3 Configuring the Recorder Mode Using the Easy Setup
Setting the Record Time
Tap the record time value. An input box appears.
1.
Use the input box to set the record time.
2.
• Set in the range of 10 s to 50 days.
If you enter the value and unit and then tap Enter, the input record time is applied.
If you enter only the value and tap Enter, the input value is assumed to be in seconds (s).
If you do not enter the value or unit and tap Enter, the default value is applied.
• The following combinations of time units can be used. Other combinations are not allowed.
Minutes and seconds, hours and minutes, days and hours
Tap Next. A Sample interval screen appears.
3.
• Tapping < Back displays the Select Channel screen.
• Tapping Cancel displays the waveform screen.
Recording Time Screen
channel selection screen
Recording time value
Recording times and the
selectable sample intervals
Input Box That Appears When You Tap the Record Time Value
1-6
IM DL350-02EN
Page 19
The content of the
sample interval list
varies depending
on the number of
channels and
record time.
Number of channels that was turned on in the channel
opens the waveform screen.
1.3 Configuring the Recorder Mode Using the Easy Setup
Setting the Sample Interval
Tap the sample interval value. On the list of available options, tap the sample interval you
1.
want to use.
• The content of the list varies depending on the number of recording channels and record time.
• The sample rate and recording data size vary according to the specified sample interval.
Tap Finish. An All Channel Setup screen appears.
2.
• Tapping < Back displays the Recording time screen.
• Tapping Cancel displays the waveform screen.
Sample Interval Screen
List box that
appears when you
tap the sample
interval value
Example of the All Channels Setup Screen
For the setting procedure on the All Channels Setup screen, see section 2.1.
selection screen (two screens earlier)
Record time set in the previous record time screen
Sample interval value
Clearing this check box and tapping Finish
IM DL350-02EN
1-7
Page 20
1.4 Initializing the Settings
This section explains how to initialize the settings to their factory default values.
► Features Guide: “Initialization of Settings (Initialize)”
On the waveform screen, tap MENU > Preparation. The following menu appears.
1.
Tap Initialize. A confirmation message appears.
2.
Initialization
Executing Initialization
Tap OK. Initialization is executed.
3.
Settings That Cannot Be Reset to Their Factory Default Values
• Date and time settings
• Communication settings
• Language setting (English or Japanese)
• System mode
• Network settings
1-8
IM DL350-02EN
Page 21
1.5 Performing Auto Setup
This section explains how to perform auto setup, which automatically sets the instrument’s settings to
the values that are most suitable for the input signals.
On the waveform screen, tap MENU > Preparation. The following menu appears.
1.
Tap Auto Setup. A confirmation message appears.
2.
Auto setup
► Features Guide: “Auto Setup (Auto Setup)”
Executing Auto Setup
Tap OK. Auto setup is executed.
3.
IM DL350-02EN
1-9
Page 22
1.6 Calibrating the Instrument
Executes calibrationTurns auto calibration on and off
This section explains how to calibrate the instrument. Execute calibration when you want to make
accurate measurements.
On the waveform screen, tap MENU > Preparation. The following menu appears.
1.
Tap Calibration. A calibration menu appears.
2.
Calibration
► Features Guide: “Calibration (Calibration)”
Calibration
Executing Calibration
Tap Execute. Calibration is executed.
Turning Auto Calibration On and Off
Tap ON or OFF under Auto CAL.
To enable auto calibration, select ON. Otherwise, select OFF.
1-10
IM DL350-02EN
Page 23
Probe attenuation and
Channels
To collectively turn the waveform display of all channels on and off, set Disp in the All row.
Chapter 2 Vertical Axis
2.1 Configuring on the All Channels Setup Screen
This section explains the following settings for configuring all channels.
• Input settings
• Linear scaling
• Copying channels
• Balancing and offset canceling
(Strain balancing (strain module) and DC offset cancellation)
► Features Guide: “All Channel Settings (All CH Setup)”
On the waveform screen, tap MENU > Channel > All CH Setup. An All Channels Setup screen
appears.
You can also display the All Channels Setup screen by tapping All Channel on the bottom menu bar of the
waveform screen.
All Channel Setup Screen
Input Settings (Setup)
Tap the Setup tab. An input setting screen appears.
1.
The items on the input setting screen vary depending on the installed modules and measurement items.
For details on the settings, see the explanation of each measurement item in section 2.2 and later.
Tap the items of each channel. Use the displayed list (options) or input box to set the items.
2.
Example of an Input Setting Screen in Scope Mode
Turns the waveform display on and off
Display label
Input coupling
Vertical scale
Bandwidth
limit
Slide to display channels that
do not fit in the screen.
Channels 5 and 6 are built-in logic signal measurement channels of
the instrument. Tap + to set each bit.
Vertical scale
Vertical position
Set the zoom magnification.
current-to-voltage
conversion ratio
IM DL350-02EN
2-1
Page 24
To collectively turn the waveform display of all channels on and off, set Disp in the All row.
Probe attenuation and
Channels
Number of decimal places
Linear scaling mode
Channels
2.1 Configuring on the All Channels Setup Screen
Example of an Input Setting Screen in Recorder Mode
Turns the waveform display on and off
Display label
Input coupling
Channels 5 and 6 are built-in logic signal measurement channels of
the instrument. Tap + to set each bit.
Upper and lower limits
of the display range
Measurement range
Bandwidth limit
Slide to display channels that
do not fit in the screen.
current-to-voltage
conversion ratio
Linear Scaling (Linear Scale)
Tap the Linear Scale tab. The Linear Scale screen appears.
1.
For details on the settings, see the explanation of linear scaling in section 2.2.
Tap the items of each channel. Use the displayed list (options) or input box to set the items.
2.
Scaling coefficient A or
measured value at point P1
Offset value B or
scaled value of point P1
Measurement value
of point P2
Slide to display channels that
do not fit in the screen.
Scaled value
of point P2
Unit
Display mode
Unit prefix
2-2
Linear scaling is not possible on channels 5 and 6 because they are logic signals.
IM DL350-02EN
Page 25
Copying Channels (Channel Copy)
Set the channels to perform balancing
Tap the Channel Copy tab. The Channel Copy screen appears.
1.
Tap each item to set options from the displayed list or execute commands.
2.
Balancing and Offset Canceling (Balance/Offset Cancel)
Tap the Balance/Offset Cancel tab. The Balance/Offset Cancel screen appears.
1.
For instructions on how to set balancing and DC offset cancellation, see the explanation of DC offset
cancellation in section 2.2 and strain balance in section 2.6.
Tap each item to set options and execute commands.
2.
2.1 Configuring on the All Channels Setup Screen
Select the copy source channel.
Set the copy destination channels to ON.
Starts copying
and offset canceling to ON.
Executes balancing and offset canceling
IM DL350-02EN
2-3
Page 26
2.2 Configuring Voltage Measurements
(Can be adjusted in steps)
current-to-voltage conversion ratio
This section explains the following settings for the vertical axis of voltage measurements.
• Display settings (vertical scale (zoom method), vertical position, vertical zoom (zooming by
setting the magnification), zooming by setting the high limit and low limit of the display range,
waveform inverted display on/off, display groups)
• Detail settings (linear scaling, channel copying, gain adjustment, DC offset cancellation on/off)
• Display settings (high limit and low limit of the display range, waveform inverted display on/off,
display groups)
• Detail settings (linear scaling, channel copying, gain adjustment, DC offset cancellation on/off)
► Features Guide: “Voltage Measurement”
Channel Setting Menu
On the waveform screen, tap MENU > Channel > any channel from CH1 to CH4. A channel
1.
setting menu appears.
• You can also tap or double-tap any channel from CH1 to CH4 in the channel information area to display
the channel setting menu. (If the channel is selected, tap; otherwise, double-tap.)
• For a 4-CH module (720254 (4CH 1M16)), select the sub channel (if the display label is set to default,
the channel number followed by an underscore and a number as in CH3_1).
Basic Setup (Basic)
Tap the Basic tab.
2.
Tap each item. Use the displayed list (options) or input box to set the items.
3.
For Scope Mode
Displays the color of the channel being configured
Turns the waveform display on and off
To the previous channel
Vertical scale
Channel being configured
To the next channel
Input coupling
Bandwidth limit
Display label
Probe attenuation and
2-4
IM DL350-02EN
Page 27
For Recorder Mode
Turns the waveform display on and off
Attenuation
If AC is selectedIf DC is selected
To the previous channel
2.2 Configuring Voltage Measurements
Channel being configured
To the next channel
Measurement range
(Can be adjusted in steps)
Input Coupling (Coupling)
The following figure shows the frequency characteristics when the input coupling is set to AC or DC.
Note that if the input coupling is set to AC, the instrument does not acquire low-frequency signals or
signal components as indicated in the figure below.
–3dB
Low frequency –3 dB point when AC coupled
* This value differs depending on the input module.
For details, section 6.13, “Module Specifications,” in the Getting Started Guide (IM DL350-03EN).
CAUTION
0dB
Input coupling
Input frequency
Bandwidth limit
–3 dB point*
Display label
Probe attenuation and
current-to-voltage conversion ratio
Attenuation
0dB
–3dB
Input frequency
–3 dB point*
IM DL350-02EN
If the input coupling is AC, in accordance with the frequency response, the input signal is
attenuated more in lower frequencies. As a result, even when a high voltage signal is actually
applied, it may not be measured as a high voltage signal. Furthermore, the over-range
indicator may not be displayed on the screen. As necessary, switch the input coupling to DC
to check the input signal voltage.
Applying an input signal whose voltage exceeds the maximum input voltage of the input
module may damage the input section.
French
ATTENTION
Si le courant du couplage d’entrée est alternatif (CA), conforme à la réponse en fréquence, le
signal d’entrée est davantage atténué aux fréquences plus basses. Par conséquent, même si
vous appliqué un signal de tension élevée, ce dernier risque de ne pas être mesuré comme
tel. De plus, le voyant de dépassement de plage risque de ne pas s’afficher à l’écran. Le cas
échéant, basculez le couplage d’entrée sur CC (courant continu) afin de vérifier la tension du
signal d’entrée.
Si la tension du signal d’entrée dépasse la tension d’entrée maximale du module d’entrée, la
section d’entrée risque d’être endommagée.
2-5
Page 28
Turning the inverted waveform
► See “Setting the Waveform Vertical and Horizontal Positions” (explained later).
Can be set as follows within the setting range
display on and off
Can be set as follows within the setting range
display on and off
2.2 Configuring Voltage Measurements
Probe Attenuation and Current-to-Voltage Conversion Ratio (Probe)
The current-to-voltage conversion ratio of the current probe is displayed.
Note
If the probe attenuation or current-to-voltage conversion ratio is not set correctly, the voltage and scale
values of the input signals will not be displayed correctly. For example, if you use a 10:1 voltage probe but
set the probe type to 1:1, the automatically measured amplitude of the waveform will be 1/10 the real value.
Configuring the Display (Display)
Tap the Display tab.
2.
Tap each item. Use the displayed list (options) or input box to set the items.
3.
For Scope Mode
When V Scale (vertical scale) Is Set to DIV
The vertical position can be moved also by tapping the up and
down arrows that appear when you slide the waveform screen.
When V Scale (vertical scale) Is Set to SPAN
For Recorder Mode
Vertical position
(Tap + or – to adjust.)
Upper and lower limits of the display range
(Tap + or – to adjust.)
Vertical zoom
Display group
display on and off
• Retain the upper and lower limit interval
• Not retain the upper and lower limit interval
Display group
Turning the inverted waveform
• Retain the upper and lower limit interval
• Not retain the upper and lower limit interval
Upper and lower limits of the display range
(Tap + or – to adjust.)
2-6
Display group
Turning the inverted waveform
IM DL350-02EN
Page 29
Detail Setting (Detail)
on and off
* Set these when measuring voltage with a voltage module or strain with a strain module.
Tap the Detail tab.
2.
Tap each item. Use the displayed list (options) or input box to set the items.
3.
Items Common to Scope Mode and Recorder Mode
Linear scaling
Note
• When DC offset cancellation is set to ON, you cannot adjust the gain.
• The gain adjustment is reset to 1.0000 when you execute DC offset cancellation or set it to on.
Linear Scaling (Linear Scale)
Tap Linear Scale. The following screen appears.
Copies channels
2.2 Configuring Voltage Measurements
Adjusts the gain
(Tap + or – to adjust.)
Acquires the DC
offset value
Turns DC offset canceling
• When the Scaling Mode Is Set to AX+B
Display mode*
Scaling coefficient
Offset
Unit
When using exponent
When using floating
Number of decimal places
IM DL350-02EN
Unit prefix
2-7
Page 30
Retrieves the current measured value
* Set these when measuring voltage with a voltage module or strain with a strain module.
2.2 Configuring Voltage Measurements
• When the Scaling Mode Is Set to P1–P2
Scale values
Measured values (tap Get Value or
set using the input box)
Display mode*
• When the Scaling Mode Is Set to Shunt
Shunt mode is available for the strain module (701271 (STRAIN_DSUB)). This module has a
built-in relay circuit for shunt calibration.
Unit
When using exponent
When using floating
Number of decimal places
Unit prefix
Current measured value when the relay circuit is off
(tap Shunt Cal Exec or set using the input box)
Executes shunt calibration
2-8
Value for when the relay circuit is off (normally 0).
Current measured value when the relay circuit is on
(tap Shunt Cal Exec or set using the input box)
Strain value corresponding to the shunt resistance
when the relay circuit is on
Unit
When using exponent
When using floating
Number of decimal places
Unit prefix
IM DL350-02EN
Page 31
Copying Channels (Copy to)
• Tap the arrows.
• Slide the waveform
Tap the screen to display.
• Tap the arrows.
• Slide the waveform
Tap the screen to display.
Tap Copy to. The following screen appears.
2.2 Configuring Voltage Measurements
Select the copy source channel.
Set the copy destination channels to ON.
Starts copying
Setting the Waveform Vertical and Horizontal Positions (Sliding
the waveform screen)
For Scope Mode
When you slide your finger across the waveform screen, up and down arrows appear. You can move
the vertical position by performing the following operation while the arrows are displayed.
Waveform zoom ► chapter 14
XY display ► chapter 11
FFT ► chapter 10
screen in the direction
of the arrow.
On the display setting screen of the channel setting menu described earlier, you can set the vertical
position by entering a value.
Up and down arrows
IM DL350-02EN
For Recorder Mode
When you slide your finger across the waveform screen, up, down, left, and right arrows appear. You
can move the vertical or horizontal position by performing the following operation while the arrows are
displayed.
screen in the direction
of the arrow.
When you slide vertically
Up and down arrows
When you slide horizontally
Left and right arrows
Tap here to expand
the time range of the
waveform display.
Displays the waveform at the
left edge of the acquisition
time or record time bar
When waveform acquisition
is in progress, Hold or
Resume is displayed here.
Tap here to reduce the
time range of the waveform
display.
Displays the waveform at the
right edge of the acquisition
time or record time bar
2-9
Page 32
2.3 Configuring Voltage Measurements (For 16-CH
All sub channel settings
voltage input modules)
This section explains the following settings for the 16-CH voltage input module:
• Display settings (vertical scale (zoom method), vertical position, vertical zoom (zooming by
setting the magnification), zooming by setting the high limit and low limit of the display range,
waveform inverted display on/off, display groups)
► Features Guide: “Voltage Measurement (For the 16-CH Voltage Input Module)”
Sub Channel Setting Menu
On the waveform screen, tap MENU > Channel > CH1 or CH3. A sub channel setting menu
1.
appears.
You can also tap or double-tap a sub channel (if the display label is set to default, the channel number
followed by an underscore and a number as in CH3_1) in the channel information area to display the sub
channel setting menu.
(If the channel is selected, tap; otherwise, double-tap.)
Setting All Sub Channels (All Sub Channel Setup)
Tap the Common tab.
2.
Tap All Sub Channel Setup. The All Sub Channel Setup screen appears.
3.
Configuring Input Settings (Setup)
Tap the Setup tab. An input setting screen appears.
4.
Tap the items of each sub channel. Use the displayed list (options) or input box to set the items.
5.
If you do not want to measure sub channels individually, set the input coupling to OFF.
2-10
IM DL350-02EN
Page 33
To set all the sub channels to the same setting, change the settings in the All row.
Lower, respectively.
To set all the sub channels to the same setting, change the settings in the All row.
2.3 Configuring Voltage Measurements (For 16-CH voltage input modules)
Example of an Input Setting Screen in Scope Mode
Display label
Input coupling
Vertical scale
Slide to display channels that
do not fit in the screen.
Vertical scale
Bandwidth limit
When the vertical scale DIV is set to SPAN,
Position and V Zoom change to Upper and
Vertical position
Set the zoom magnification.
Example of an Input Setting Screen in Recorder Mode
Input coupling
Display label
Measurement range
Bandwidth limit
Slide to display channels that
do not fit in the screen.
Upper and lower limits
of the display range
IM DL350-02EN
2-11
Page 34
To set all the sub channels to the same setting, change the settings in the All row.
Starts copying
2.3 Configuring Voltage Measurements (For 16-CH voltage input modules)
Linear Scaling (Linear Scale)
Tap the Linear Scale tab. The Linear Scale screen appears.
4.
For details on the settings, see the explanation of linear scaling in section 2.2.
Tap the items of each sub channel. Use the displayed list (options) or input box to set the items.
5.
Linear scaling mode
Scaling coefficient A or measured value at point P1
Offset value B or scaled value of point P1
Measurement value of point P2
Scaled value
of point P2
Unit
Display mode
Number of decimal places
Unit prefix
Slide to display channels that
do not fit in the screen.
Copying Channels (Channel Copy)
Tap the Channel Copy tab. The Channel Copy screen appears.
4.
Tap each item to set options from the displayed list or execute commands.
5.
Select the copy sub source channel.
Set the copy destination
sub channels to ON.
2-12
Sets all sub channel copy to OFF.
Sets all sub channel copy to ON.
Copying channels can also be set using Copy to on the Detail tab.
IM DL350-02EN
Page 35
Basic Setup (Basic)
To the previous sub channel
Turns waveform display on and off (all sub channels on and off)
To the previous sub channel
Tap the Basic tab.
2.
Tap each item. Use the displayed list (options) or input box to set the items.
3.
For Scope Mode
Displays the color of the sub channel being configured
Waveform display on/off (all sub channels on/off)
To the previous main channel
2.3 Configuring Voltage Measurements (For 16-CH voltage input modules)
• Display settings (temperature unit, high limit and low limit of the display range, display groups)
• Detail settings (copying channels)
► Features Guide: “Temperature Measurement”
Channel Setting Menu
On the waveform screen, tap MENU > Channel > any channel from CH1 to CH4. A channel
1.
setting menu appears.
You can also tap or double-tap any channel from CH1 to CH4 in the channel information area to display the
channel setting menu.
(If the channel is selected, tap; otherwise, double-tap.)
Basic Setup (Basic)
Tap the Basic tab.
2.
Tap each item. Use the displayed list (options) or input box to set the items.
3.
Displays the color of the channel being configured
Turns the waveform display on and off
To the previous channel
Thermocouple type
Input Coupling (Coupling)
To measure temperature, set the input coupling to TC.
Tap each item. Use the displayed list (options) or input box to set the items.
3.
• Retain the upper and lower limit interval
• Not retain the upper and lower limit interval
Upper and lower limits
of the display range
Display groupTemperature unit
Detail Setting (Detail)
Tap the Detail tab.
2.
Copying Channels (Copy to)
►section2.2
Setting the Waveform Vertical and Horizontal Positions (Sliding
the waveform screen)
►section2.2
2-16
IM DL350-02EN
Page 39
2.5 Configuring Temperature Measurements (For
All sub channel settingsData update period
16-CH temperature/voltage input modules)
This section explains the following temperature measurement settings for the 16-CH temperature/
voltage input module. For voltage measurement settings, see section 2.3.
• Display settings (temperature unit, high limit and low limit of the display range, waveform inverted
display on/off, display groups)
• Detail settings (channel copying)
► Features Guide: “Temperature Measurement (For the 16-CH
Temperature/Voltage Input Module)”
Sub Channel Setting Menu
On the waveform screen, tap MENU > Channel > CH1 or CH3. A sub channel setting menu
1.
appears.
You can also tap or double-tap a sub channel (if the display label is set to default, the channel number
followed by an underscore and a number as in CH3_1) in the channel information area to display the sub
channel setting menu.
(If the channel is selected, tap; otherwise, double-tap.)
Setting the Data Update Period (Data Update Period) and Setting
All Sub Channels (All Sub Channel Setup)
Tap the Common tab.
2.
Tap Data Update Period. Use the displayed list (options) to set the data update period.
3.
Tap All Sub Channel Setup. The All Sub Channel Setup screen appears.
4.
IM DL350-02EN
2-17
Page 40
Turns RJC (reference junction
To set all the sub channels to the same setting, change the settings in the All row.
2.5 Configuring Temperature Measurements (For 16-CH temperature/voltage input modules)
Configuring Input Settings (Setup)
Tap the Setup tab. An input setting screen appears.
5.
Tap the items of each sub channel. Use the displayed list (options) or input box to set the items.
6.
Example of an Input Setting Screen in Scope Mode
• When the Input Coupling Is Set to TC
Thermocouple type
Input coupling
Display label
Temperature unit
Upper and lower limits
of the display range
Slide to display channels that
do not fit in the screen.
compensation) on and off
Turns burnout
detection on and off
• When the Input Coupling Is Set to DC, GND, or OFF
►section2.3
For the 16-CH temperature/voltage input module, there is no bandwidth limit setting.
2-18
IM DL350-02EN
Page 41
Turns RJC (reference junction
To set all the sub channels to the same setting, change the settings in the All row.
2.5 Configuring Temperature Measurements (For 16-CH temperature/voltage input modules)
Example of an Input Setting Screen in Recorder Mode
• When the Input Coupling Is Set to TC
Input coupling
Display label
Thermocouple type
Temperature unit
Upper and lower limits
of the display range
Slide to display channels that
do not fit in the screen.
compensation) on and off
Turns burnout
detection on and off
• When the Input Coupling Is Set to DC, GND, or OFF
►section2.3
For the 16-CH temperature/voltage input module, there is no bandwidth limit setting.
Linear Scaling (Linear Scale)
You can set linear scaling on sub channels with the input coupling set to DC, GND, or OFF.
Tap the Linear Scale tab. The Linear Scale screen appears.
5.
For details on the settings, see the explanation of linear scaling in section 2.2.
Tap the items of each sub channel. Use the displayed list (options) or input box to set the items.
6.
Linear Scaling Screen
►section2.3
Copying Channels (Channel Copy)
Tap the Channel Copy tab. The Channel Copy screen appears.
5.
Tap each item to set options from the displayed list or execute commands.
6.
Channel Copy Screen
►section2.3
Copying channels can also be set using Copy to on the Detail tab.
IM DL350-02EN
2-19
Page 42
Turns burnout detection
To the previous sub channel
2.5 Configuring Temperature Measurements (For 16-CH temperature/voltage input modules)
Basic Setup (Basic)
Tap the Basic tab.
2.
Tap each item. Use the displayed list (options) or input box to set the items.
3.
When the Input Coupling Is Set to TC
Displays the color of the sub channel being configured
Turns waveform display on and off (all sub channels on and off)
To the previous main channel
Sub channel being configured
To the next main channel
Thermocouple type
To the next sub channel
Sub number of the sub channel
being configured
Set input coupling to TC.
Display label
on and off
Turns RJC (reference junction
compensation) on and off
When the Input Coupling Is Set to DC, GND, or OFF
►section2.3
For the 16-CH temperature/voltage input module, there is no bandwidth limit setting.
2-20
IM DL350-02EN
Page 43
2.5 Configuring Temperature Measurements (For 16-CH temperature/voltage input modules)
Configuring the Display (Display)
When the Input Coupling Is Set to TC
►section2.4
For the sub channel settings (Sub Ch.), see “Basic Setup (Basic)” on the previous page.
When the Input Coupling Is Set to DC, GND, or OFF
►section2.2
For the sub channel settings (Sub Ch.), see “Basic Setup (Basic)” on the previous page.
Detail Setting (Detail)
►section2.2
For the 16-CH temperature/voltage input module, there is no gain adjustment or DC offset cancellation
setting.
For the sub channel settings, see “Basic Setup (Basic)” on the previous page.
Linear Scaling (Linear Scale)
You can set linear scaling on sub channels with the input coupling set to DC, GND, or OFF.
►section2.2
Copying Channels (Copy to)
►section2.3
Copying channels can also be set using Channel Copy tab on the All Sub Channel Setup screen.
Setting the Waveform Vertical and Horizontal Positions (Sliding
the waveform screen)
►section2.2
IM DL350-02EN
2-21
Page 44
2.6 Configuring Strain Measurements
(Can be adjusted in steps)
This section explains the following settings for strain measurements:
On the waveform screen, tap MENU > Channel > any channel from CH1 to CH4. A channel
1.
setting menu appears.
You can also tap or double-tap any channel from CH1 to CH4 in the channel information area to display the
channel setting menu.
(If the channel is selected, tap; otherwise, double-tap.)
Basic Setup (Basic)
Tap the Basic tab.
2.
Tap each item. Use the displayed list (options) or input box to set the items.
3.
Displays the color of the channel being configured
Turns the waveform display on and off
To the previous channel
Measurement range
Channel being configured
To the next channel
Bandwidth limit
Set the sensor.
Display labelExecutes strain balancing
2-22
IM DL350-02EN
Page 45
Configuring the Sensor (Sensor Setup)
Can be set as follows within the setting range
display on and off
Tap Sensor Setup. The following screen appears.
4.
2.6 Configuring Strain Measurements
Configuring the Display (Display)
Tap the Display tab.
2.
Tap each item. Use the displayed list (options) or input box to set the items.
3.
Upper and lower limits
of the display range
(Tap + or – to adjust.)
Range Unit (Range Unit)
• μSTR:Unitsofstrain(×10−6 strain)
• mV/V: Units of strain-gauge-transducer output
The mV/V range is calculated from the following equation.
mV/V=0.5×(µSTR/1000)
Bridge voltage
Gauge factor
(Tap + or – to adjust.)
• Retain the upper and lower limit interval
• Not retain the upper and lower limit interval
Display groupRange unit
Turning the inverted waveform
Detail Setting (Detail)
Setting the Waveform Vertical and Horizontal Positions (Sliding
the waveform screen)
IM DL350-02EN
►section2.2
For the strain module, there is no gain adjustment or DC offset cancellation setting.
Linear Scaling (Linear Scale)
►section2.2
Copying Channels (Copy to)
►section2.2
►section2.2
2-23
Page 46
2.7 Configuring Acceleration Measurements
Gain
(Can be adjusted in steps)
acceleration sensor on and off
This section explains the following settings for acceleration measurements:
For Scope Mode
• Basic settings (waveform display on/off, gain, setting the input coupling to ACCEL, bandwidth
limit, bias current supply to the acceleration sensor on and off, display labels)
• Display settings (vertical scale (zoom method), vertical position, vertical zoom (zooming by
setting the magnification), zooming by setting the high limit and low limit of the display range, unit
of acceleration, display groups)
• Basic settings (waveform display on/off, gain, setting the input coupling to ACCEL, bandwidth
limit, bias current supply to the acceleration sensor on and off, display labels)
• Display settings (high limit and low limit of the display range, unit of acceleration, display groups)
On the waveform screen, tap MENU > Channel > any channel from CH1 to CH4. A channel
1.
setting menu appears.
You can also tap or double-tap any channel from CH1 to CH4 in the channel information area to display the
channel setting menu.
(If the channel is selected, tap; otherwise, double-tap.)
Basic Setup (Basic)
Tap the Basic tab.
2.
Tap each item. Use the displayed list (options) or input box to set the items.
3.
Items Common to Scope Mode and Recorder Mode
Displays the color of the channel being configured
Turns the waveform display on and off
To the previous channel
Channel being configured
To the next channel
2-24
Bandwidth limit
Set Coupling to ACCEL.
Display label
Bias current supply to the
IM DL350-02EN
Page 47
► section 2.2
Can be set as follows within the setting range
Acceleration unit
Can be set as follows within the setting range
Acceleration unit
2.7 Configuring Acceleration Measurements
Input Coupling (Coupling)
To measure acceleration, set the input coupling to ACCEL.
To measure voltage, set the input coupling to an appropriate voltage measurement setting.
►section2.2
Configuring the Display (Display)
Tap the Display tab.
2.
Tap each item. Use the displayed list (options) or input box to set the items.
3.
For Scope Mode
When V Scale (vertical scale) Is Set to DIV
The vertical position can be moved also by tapping the up and
down arrows that appear when you slide the waveform screen.
When V Scale (vertical scale) Is Set to SPAN
For Recorder Mode
Vertical position
(Tap + or – to adjust.)
Upper and lower limits of the display range
(Tap + or – to adjust.)
Vertical zoomDisplay group
Acceleration unit
• Retain the upper and lower limit interval
• Not retain the upper and lower limit interval
Display group
• Retain the upper and lower limit interval
• Not retain the upper and lower limit interval
IM DL350-02EN
Upper and lower limits of the display range
(Tap + or – to adjust.)
Display group
2-25
Page 48
Copies channels
2.7 Configuring Acceleration Measurements
Detail Setting (Detail)
Tap the Detail tab.
2.
Tap each item. Use the displayed list (options) or input box to set the items.
3.
Items Common to Scope Mode and Recorder Mode
Acceleration sensor sensitivity
Copying Channels (Copy to)
►section2.2
Setting the Waveform Vertical and Horizontal Positions (Sliding
the waveform screen)
►section2.2
(Tap + or – to adjust.)
2-26
IM DL350-02EN
Page 49
2.8
(Can be adjusted in steps)
Configuring Frequency, Revolution, Period, Duty
Cycle, Power Supply Frequency, Pulse Width,
Pulse Integration, and Velocity Measurements
This section explains the following settings for frequency, revolution, period, duty cycle, power supply
frequency, pulse width, pulse integration, and velocity measurements:
• Display settings (vertical scale (zoom method), vertical position, vertical zoom (zooming by
setting the magnification), zooming by setting the high limit and low limit of the display range,
display groups)
On the waveform screen, tap MENU > Channel > any channel from CH1 to CH4. A channel
1.
setting menu appears.
You can also tap or double-tap any channel from CH1 to CH4 in the channel information area to display the
channel setting menu.
(If the channel is selected, tap; otherwise, double-tap.)
► Features Guide: “Frequency Measurement”
Basic Setup (Basic)
Tap the Basic tab.
2.
Tap each item. Use the displayed list (options) or input box to set the items.
3.
For Scope Mode
Displays the color of the channel being configured
Turns the waveform display on and off
To the previous channel
Vertical scale
Channel being configured
To the next channel
FV setting
Display label
Input condition
IM DL350-02EN
2-27
Page 50
Turns the waveform display on and off
(Can be adjusted in steps)
2.8 Configuring Frequency, Revolution, Period, Duty Cycle, Power Supply Frequency, Pulse Width, Pulse Integration, and Velocity Measurements
For Recorder Mode
To the previous channel
Channel being configured
To the next channel
Measurement range
FV Setup (FV Setup)
Tap FV Setup. The following screen appears.
4.
Tap each item. Use the displayed list (options) or input box to set the items.
5.
• When the measurement item is Frequency
FV setting
Display label
Set to Frequency.
Offset (Tap + or – to adjust.)
(This item is not available in recorder mode.)
Filter (Tap + or – to adjust.)
• Smoothing
• Pulse average
Turns decelerating prediction on and off
Input condition
• When the measurement item is Revolution (rpm)
2-28
Stop prediction
Set to Revolution (rpm).
Offset (Tap + or – to adjust.)
(This item is not available in recorder mode.)
Pulse/rotate (Tap + or – to adjust.)
(Number of pulses per rotation)
Filter (Tap + or – to adjust.)
• Smoothing
• Pulse average
Turns decelerating prediction on and off
Stop prediction
IM DL350-02EN
Page 51
2.8 Configuring Frequency, Revolution, Period, Duty Cycle, Power Supply Frequency, Pulse Width, Pulse Integration, and Velocity Measurements
• When the measurement item is Revolution (rps)
Set to Revolution (rps).
Offset (Tap + or – to adjust.)
(This item is not available in recorder mode.)
Pulse/rotate (Tap + or – to adjust.)
(Number of pulses per rotation)
Filter (Tap + or – to adjust.)
• Smoothing
• Pulse average
Turns decelerating prediction on and off
Stop prediction
• When the measurement item is Period
Set to Period.
Offset (Tap + or – to adjust.)
(This item is not available in recorder mode.)
Filter (Tap + or – to adjust.)
• Smoothing
• Pulse average
Turns decelerating prediction on and off
Stop prediction
IM DL350-02EN
2-29
Page 52
2.8 Configuring Frequency, Revolution, Period, Duty Cycle, Power Supply Frequency, Pulse Width, Pulse Integration, and Velocity Measurements
• When the measurement item is Duty
Set to Duty.
Offset (Tap + or – to adjust.)
(This item is not available in recorder mode.)
Measurement pulse selection
Filter (Tap + or – to adjust.)
• Smoothing
Timeout period (Tap + or – to adjust.)
• When the measurement item is Power Freq
• When the measurement item is Pulse Width
Set Function to Power Freq.
Center frequency
Filter (Tap + or – to adjust.)
• Smoothing
• Pulse average
Set to Pulse Width.
Offset (Tap + or – to adjust.)
(This item is not available in recorder mode.)
Measurement pulse selection
2-30
Filter (Tap + or – to adjust.)
• Smoothing
IM DL350-02EN
Page 53
the pulse count
2.8 Configuring Frequency, Revolution, Period, Duty Cycle, Power Supply Frequency, Pulse Width, Pulse Integration, and Velocity Measurements
• When the measurement item is Pulse Integ
Set to Pulse Integ.
Offset (Tap + or – to adjust.)
(This item is not available in recorder mode.)
Unit/pulse (Physical amount per pulse)
• When the measurement item is Velocity
Pulse integration unit
Filter (Tap + or – to adjust.)
• Smoothing
• Pulse average
Turns auto reset at pulse count
overflow on and off
Executes the manual reset of
Set to Velocity.
Offset (Tap + or – to adjust.)
(This item is not available in recorder mode.)
Distance/pulse (Distance per pulse)
Time unit
Velocity unit
Filter (Tap + or – to adjust.)
• Smoothing
• Pulse average
IM DL350-02EN
Turns decelerating prediction on and off
Stop prediction
2-31
Page 54
2.8 Configuring Frequency, Revolution, Period, Duty Cycle, Power Supply Frequency, Pulse Width, Pulse Integration, and Velocity Measurements
Input Setup (Input Setup)
Tap Input. The following screen appears.
4.
Tap each item. Use the displayed list (options) or input box to set the items.
5.
Preset item
Voltage range
Input coupling
Probe type
Bandwidth limit
When Preset is set to Pull-up 5V
Setting Preset Items
You can set the preset to one of the following 10 options: Logic 5V, Logic 3V, Logic 12V, Logic
24V, Pull-up 5V, ZeroCross, AC100V, AC200V, EM Pickup, or User (user-defined).
The settable input items differ depending on the preset that you set. You can only turn pull-up on
and off when the preset is set to Pull-up 5V.
Threshold level (Tap + or – to adjust.)
Hysteresis ( , , )
Slope ( , )
Chatter elimination (Tap + or – to adjust.)
Turns pull-up on and off
2-32
IM DL350-02EN
Page 55
2.8 Configuring Frequency, Revolution, Period, Duty Cycle, Power Supply Frequency, Pulse Width, Pulse Integration, and Velocity Measurements
Configuring the Display (Display)
For Scope Mode
►section2.2
• There are no inverted waveform display items.
• If the FV setup measurement item is Power Freq, V Scale is fixed to DIV, so no options will appear.
• If the FV setup measurement item is Pulse Integ or Velocity, the setting for retaining the upper and
lower limit interval for when V Scale is set to SPAN is not available.
For Recorder Mode
►section2.2
• There are no inverted waveform display items.
• If the FV setup measurement item is Power Freq, the same vertical position and vertical zoom
menu as scope mode is displayed.
• If the FV setup measurement item is Pulse Integ or Velocity, the setting for retaining the upper and
lower limit interval is not available.
Detail Settings (Detail)
Items Common to Scope Mode and Recorder Mode
►section2.2
• There are no gain adjustment and DC offset items.
• The linear scaling setting menu does not have the display mode item.
Setting the Waveform Vertical and Horizontal Positions (Sliding
the waveform screen)
►section2.2
IM DL350-02EN
2-33
Page 56
2.9 Configuring Logic Signal Measurements
Set bits.
Display label
Copies channels
This section explains the following settings for logic measurements:
• Display settings (vertical position, vertical zoom (zooming by setting the magnification), bit
mapping, display groups)
► Features Guide: “Logic Measurement”
Channel Setting Menu
On the waveform screen, tap MENU > Channel > any channel from CH1 to CH6. A channel
1.
setting menu appears.
You can also tap or double-tap any channel from CH1 to CH6 in the channel information area to display the
channel setting menu.
(If the channel is selected, tap; otherwise, double-tap.)
Basic Setup (Basic)
Tap the Basic tab.
2.
Tap each item. Use the displayed list (options) or input box to set the items.
3.
Displays the color of the channel being configured
Turns the waveform display on and off
To the previous channel
Channel being configured
To the next channel
2-34
IM DL350-02EN
Page 57
Bit Settings (Bit Setup)
Turns the display ON for all bits
Chattering elimination times
(Tap + or – to adjust.)
Display group
Tap Bit Setup. The following screen appears.
4.
Tap each item. Use the displayed list (options) or input box to set the items.
5.
Display on/off state and label of each bit
2.9 Configuring Logic Signal Measurements
Turns the display OFF for all bits
Copying Channels (Copy to)
►section2.2
Configuring the Display (Display)
Tap the Display tab.
2.
Tap each item. Use the displayed list (options) or input box to set the items.
3.
Vertical position
Vertical zoom
Bit mapping
Setting the Waveform Vertical and Horizontal Positions (Sliding
the waveform screen)
►section2.2
IM DL350-02EN
2-35
Page 58
2.10 Configuring the Monitoring of CAN and CAN FD
Bus Signals (/VE option)
This section explains the following settings for monitoring CAN and CAN FD bus signals:
Items Common to Scope Mode and Recorder Mode
• CAN port configuration (waveform display on/off, port and all sub channel settings, loading CAN/
CAN FD data definition files, scaling all sub channels)
• Display settings (display range of each sub channel, scaling each sub channel, vertical position
of each sub channel, vertical zoom of each sub channel (zooming by setting the magnification),
display groups)
• One shot output
► Features Guide: “CAN and CAN FD Bus Signal Monitoring”
Channel Setting Menu
On the waveform screen, tap MENU > Channel > any channel from CH1 to CH4. A channel
1.
setting menu appears.
You can also tap or double-tap any channel from CH1 to CH4 in the channel information area to display the
channel setting menu.
(If the channel is selected, tap; otherwise, double-tap.)
CAN Port Configuration (CAN Port Config)
Tap the CAN Port Config. tab.
2.
Tap each item to set options and execute commands.
3.
Turns the waveform display on and off
To the previous main channel
Sub channel being configured
To the next main channel
Configure ports and
all sub channels.
Load a CAN/CAN FD
data definition file.
Executes auto scale
Scaling of all sub channels
Executes default scale
2-36
Port and All Sub Channel Settings (Port & All SubChannel Setup)
Tap Port & All SubChannel Setup. A Port & All SubChannel Setup screen appears.
4.
IM DL350-02EN
Page 59
Turns the sub channel monitoring on and off
CAN or CAN FD data conversion conditions
* You cannot set Factor, Offset, or Unit when the data type (Value Type) is set to Logic.
2.10 Configuring the Monitoring of CAN and CAN FD Bus Signals (/VE option)
All Sub Channel Setup (All SubChannel Setup
Conditions
Tap the All SubChannel Setup tab. The following screen appears.
5.
Tap each item. Use the displayed list (options) or input box to set the items.
6.
- CAN or CAN FD Data Extraction
)
To set all the sub channels to the same setting, change the settings in the All row.
CAN or CAN FD data extraction conditions
Slide to display channels that
do not fit in the screen.
All Sub Channel Factor/Offset (All SubChannel Factor/Offset
Conversion Condition
Tap the All SubChannel Factor/Offset tab. The following screen appears.
5.
Tap each item. Use the displayed input box to set the items.
6.
- CAN or CAN FD Data
)
***
Slide to display channels that
do not fit in the screen.
IM DL350-02EN
2-37
Page 60
2.10 Configuring the Monitoring of CAN and CAN FD Bus Signals (/VE option)
Port Settings (Port Setup)
Tap the Port Setup tab. The following screen appears.
5.
Tap each item. Use the displayed list (options) or input box to set the items.
6.
• CAN Bus Signal Monitoring (for the 720240 (CAN MONITOR), 720241 (CAN & LIN) modules)
Bit rate
Sample point
Resynchronization jump width (Tap + or – to adjust.)
Sample count
• CAN and CAN FD Bus Signal Monitoring (for the 720242 (CAN/CAN FD) module)
Turns listen-only on and off
Turns the terminator on and off
Bit rate
Sample point
CAN FD standard
Data bit rate
Data sample point
2-38
Turns listen-only on and off
Turns the terminator on and off
IM DL350-02EN
Page 61
Display group
Scaling of the specified sub channel
(Tap + or – to change.)
2.10 Configuring the Monitoring of CAN and CAN FD Bus Signals (/VE option)
Loading a CAN/CAN FD Data Definition File (Symbol File Load)
Tap Symbol File Load. A file list appears.
4.
On the file list, tap a symbol file (SBL file).
5.
For details on file list operations, see section 6.10.
Select the load destination channel.
6.
►section6.9
Tap Load. A confirmation message appears.
7.
Tap OK. The symbol file is loaded.
8.
Sub Channel Display Settings (Display)
Tap the Display tab.
2.
Tap each item. Use the displayed list (options) or input box to set the items.
3.
When the Data Type (Value Type) Is Unsigned, Signed, or Float
Upper and lower limits
of the display range
Specify the sub channel.
(Tap + or – to change.)
Executes auto scale Executes default scale
When the Data Type (Value Type) is Logic
Specify the sub channel.
Vertical position
(Tap + or – to adjust.)
Display groupVertical zoom
IM DL350-02EN
2-39
Page 62
2.10 Configuring the Monitoring of CAN and CAN FD Bus Signals (/VE option)
Executes one shot outputConfigure the one shot output settings.
* These are the items you set when the frame type is set to data (Data).
* These are the items you set when the frame type is set to data (Data).
Frame type
(only when Message
Type is set to CAN)
This item is not displayed
when Message Type is
set to CAN FD because
the frame type is fixed at
Data.
One Shot Output (One shot)
Tap the One Shot tab.
2.
Tap each item to set options and execute commands.
3.
One Shot Output Settings (One Shot Out Setup)
Tap One Shot Out Setup. A One Shot Out Setup screen appears.
4.
CAN Frame One-Shot Output
CAN or CAN FD Frame One-Shot Output
Message type
Message format
(for the 720240 (CAN MONITOR), 720241 (CAN & LIN) modules)
Message format
Message ID (Tap + or – to adjust.)
Frame type
Size, in bytes, of the data area
(Tap + or – to adjust.)*
Data*
(for the 720242 (CAN/CAN FD) module)
Message ID
(Tap + or – to adjust.)
Size, in bytes,
of the data area
(Tap + or – to adjust.)*
2-40
Data*
The number of transmission data bytes
that can be set varies depending on
the message type and DLC settings.
► See the next page.
IM DL350-02EN
Page 63
2.10 Configuring the Monitoring of CAN and CAN FD Bus Signals (/VE option)
DLC value and the number of transmission data bytes
When DLC = 0 to 8
DLC
000111222
333444555
666777888
Number of Data Bytes
CANCAN FDCANCAN FDCANCAN FD
DLC
Number of Data Bytes
DLC
Number of Data Bytes
WhenDLC≥9
• When the message type is CAN
DLCNumber of Data Bytes
9 to 158
• When the message type is CAN FD
DLCNumber of Data Bytes DLCNumber of Data Bytes DLCNumber of Data Bytes
91210161120
122413321448
1564
IM DL350-02EN
2-41
Page 64
2.11 Configuring the Monitoring of LIN Bus Signals
Scaling of all sub channels
(/VE option)
This section explains the following settings for monitoring LIN bus signals:
Items Common to Scope Mode and Recorder Mode
• LIN port configuration (waveform display on/off, frame and all sub channel settings, loading LIN
data definition files, scaling all sub channels)
• Display settings (display range of each sub channel, scaling each sub channel, vertical position
of each sub channel, vertical zoom of each sub channel (zooming by setting the magnification),
display groups)
► Features Guide: “LIN Bus Signal Monitoring”
Channel Setting Menu
On the waveform screen, tap MENU > Channel > CH2 or CH4. A channel setting menu
1.
appears.
You can also tap or double-tap CH2 or CH4 in the channel information area to display the channel setting
menu.
(If the channel is selected, tap; otherwise, double-tap.)
LIN Port Configuration (LIN Port Config)
Tap the LIN Port Config. tab.
2.
Tap each item to set options and execute commands.
3.
Turns the waveform display on and off
Frame and All Sub Channel Settings (Frame & All SubChannel
Setup)
Tap Frame & All SubChannel Setup. A Frame & All SubChannel Setup screen appears.
4.
2-42
To the previous main channel
Sub channel being configured
To the next main channel
Configure the Frame
and all sub channels.
Load a LIN data
definition file.
Executes auto scale
Executes default scale
IM DL350-02EN
Page 65
Frame Settings (Frame Setup)
To set frames 0 to 59 to the same setting at once, change the settings in the All row.
Turns the sub channel monitoring on and off
Tap the Frame Setup tab. The following screen appears.
5.
Tap each item. Use the displayed list (options) or input box to set the items.
6.
You cannot change the settings for frames 60 or above at once.
Data length
2.11 Configuring the Monitoring of LIN Bus Signals (/VE option)
Checksum typeBit rate
Slide to display zoom frames
that do not fit in the screen.
Note
All IDs are displayed. Only the settings for the frames that have IDs that data will be read for will be enabled.
The settings for frames that have other IDs will be ignored.
All Sub Channel Setup (All SubChannel Setup
Tap the All SubChannel Setup tab. The following screen appears.
5.
Tap each item. Use the displayed list (options) or input box to set the items.
6.
To set all the sub channels to the same setting, change the settings in the All row.
LIN data extraction conditions
Slide to display channels that
do not fit in the screen.
- LIN Data Extraction Conditions
)
IM DL350-02EN
2-43
Page 66
LIN data conversion conditions
* You cannot set Factor, Offset, or Unit when the data type (Value Type) is set to Logic.
2.11 Configuring the Monitoring of LIN Bus Signals (/VE option)
All Sub Channel Factor/Offset (All SubChannel Factor/Offset
Condition
Tap the All SubChannel Factor/Offset tab. The following screen appears.
5.
Tap each item. Use the displayed input box to set the items.
6.
- LIN Data Conversion
)
***
Slide to display channels that
do not fit in the screen.
Loading a LIN Data Definition File (Symbol File Load)
Tap Symbol File Load. A file list appears.
4.
On the file list, tap a symbol file (SBL file).
5.
For details on file list operations, see section 6.10.
Select the load destination channel.
6.
►section6.9
Tap Load. A confirmation message appears.
7.
Tap OK. The symbol file is loaded.
8.
2-44
IM DL350-02EN
Page 67
Specify the sub channel.
(Tap + or – to change.)
Display group
Specify the sub channel.
(Tap + or – to change.)
2.11 Configuring the Monitoring of LIN Bus Signals (/VE option)
Sub Channel Display Settings (Display)
Tap the Display tab.
2.
Tap each item. Use the displayed list (options) or input box to set the items.
3.
When the Data Type (Value Type) is Unsigned or Signed
Upper and lower limits
of the display range
Executes auto scale Executes default scale
Scaling of the specified sub channel
When the Data Type (Value Type) is Logic
Vertical position
(Tap + or – to adjust.)
Display groupVertical zoom
IM DL350-02EN
2-45
Page 68
2.12 Configuring the Monitoring of SENT Signals
resetting of the error count
(/VE option)
This section explains the following settings for monitoring SENT signals:
Items Common to Scope Mode and Recorder Mode
• SENT port configuration (waveform display on/off, SENT format, error channel, input settings, all
sub channel settings, error count reset)
• Display settings (display range of each sub channel, scaling each sub channel, vertical position
of each sub channel, vertical zoom of each sub channel (zooming by setting the magnification),
display groups)
► Features Guide: “SENT Signal Monitoring”
Channel Setting Menu
On the waveform screen, tap MENU > Channel > any channel from CH1 to CH4. A channel
1.
setting menu appears.
You can also tap or double-tap any channel from CH1 to CH4 in the channel information area to display the
channel setting menu.
(If the channel is selected, tap; otherwise, double-tap.)
SENT Port Configuration (SENT Port Config)
Tap the SENT Port Config. tab.
2.
Tap each item to set options and execute commands.
3.
Turns the waveform display on and off
Port settings
Port Settings (Port Setup)
Tap Port Setup. A Port Setup screen appears.
4.
2-46
To the previous main channel
Sub channel being configured
To the next main channel
All sub channel settingsExecutes the manual
IM DL350-02EN
Page 69
2.12 Configuring the Monitoring of SENT Signals (/VE option)
Turns error count integration
Turns error detection on and off (Successive Calibration Pulses (Option2) only)
to adjust.)
SENT Format (SENT Format)
Tap the SENT Format tab. The following screen appears.
5.
Tap each item. Use the displayed list (options) or input box to set the items.
6.
Clock tick (Tap + or – to adjust.)
Number of data nibbles of fast channel
messages (Tap + or – to adjust.)
Sets whether to include pause pulses in Fast CH messages (ON or OFF)
Sets whether to handle fast channel multiplexing (ON or OFF)*
CRC type
Slow channel message format
* You can set fast channel multiplexing when the SENT monitor module (720243(SENT)) version
is 0x07 or later.
Error Channel Settings (Error Channel Setup)
Tap the Error Channel Setup tab. The following screen appears.
5.
Tap each item to set options and execute commands.
6.
Error detection (Detect) is always on for Fast Channel CRC, Slow Channel CRC,
Nibble Value, and Pulse Number.
Turns error trigger display on and off
Turns error count reset on start on and off
Input Settings (Input Setup)
Tap the Input Setup tab. The following screen appears.
5.
Tap each item. Use the displayed list (options) or input box to set the items.
6.
on and off
Executes the manual
resetting of the error count
Probe attenuation
Threshold level (fixed)
IM DL350-02EN
Timeout period (Tap + or –
2-47
Page 70
Turns the sub channel monitoring on and off
If you tap a sub channel whose data type (Data Type) is set to FastCH, the
Switch the data type.
2.12 Configuring the Monitoring of SENT Signals (/VE option)
All Sub Channel Setup (All Sub Channel Setup)
Tap All Sub Channel Setup. An All Sub Channel Setup screen appears.
4.
All Sub Channel Settings (All SubChannels Setup
Tap the All SubChannel Setup tab. The following screen appears.
5.
Tap each item. Use the displayed list (options) or input box to set the items.
6.
Data type
- SENT Data Extraction Conditions
SENT data extraction conditions*
Slide to display channels that
do not fit in the screen.
)
When Fast Channel Multiplexing is set to ON
ID column changes to an FC column, and you can set FC (Frame Control).
If you tap a sub channel whose data type is set to SlowCH, you can set ID.
2-48
IM DL350-02EN
Page 71
SENT data conversion conditions
2.12 Configuring the Monitoring of SENT Signals (/VE option)
Sub Channel and Data Type
The types of data acquired in sub channels are as follows.
Sub ChannelData Type
1
FastCH2
3
4S&C (Stauts & Communication)
5
6
7
8
9
10Error Trigger
11Error Count
SlowCH, FastCH*
* You can change this to FastCH only when
Fast Channel Multiplexing is set to ON.
All Sub Channel Factor/Offset (All SubChannel Factor/Offset
Condition
Tap the All SubChannel Factor/Offset tab. The following screen appears.
5.
Tap each item. Use the displayed input box to set the items.
6.
- SENT Data Conversion
)
Slide to display channels that
do not fit in the screen.
IM DL350-02EN
2-49
Page 72
Specify the sub channel.
(Tap + or – to change.)
Display group
Specify the sub channel.
(Tap + or – to change.)
2.12 Configuring the Monitoring of SENT Signals (/VE option)
Sub Channel Display Settings (Display)
Tap the Display tab.
2.
Tap each item. Use the displayed list (options) or input box to set the items.
3.
When the Data Type Is Fast CH, SlowCH, or Error Count
Upper and lower limits
of the display range
Executes auto scaleExecutes default scale
Scaling of the specified sub channel
When the Data Type Is S&C (Status & Communication) and Error Trigger
Vertical position
(Tap + or – to adjust.)
Display groupVertical zoom
2-50
IM DL350-02EN
Page 73
2.13 Configuring the GPS Position Information
All items setup
Monitor
This section explains the following settings for monitoring position information with the GPS (Global
Positioning System).
To acquire GPS position information, a GPS unit (an accessory sold separately) must be connected to
the instrument and the position information acquisition setting must be turned on (see section 18.1).
Items Common to Scope Mode and Recorder Mode
• Common settings (All Items Setup)
• Basic settings (position information, display labels, range, 3D positioning status, GPS time sync
status)
Set position information acquisition to ON. For the procedure, see section 18.1.
1.
Channel Setting Menu
On the waveform screen, tap MENU > Channel > GPS. A channel setting menu appears.
2.
You can also tap or double-tap GPS in the channel information area to display the channel setting menu.
(If the channel is selected, tap; otherwise, double-tap.)
Common Settings (Common)
Tap the Common tab.
3.
Tap each item to set options.
4.
Displays the position information of the channel being configured
Turns the waveform display on and off
To the previous main channel
Channel being configured
IM DL350-02EN
2-51
Page 74
Display on/off state of each position information entry
When the position information is status (Status),
Latitude, Longitude, or Direction.
2.13 Configuring the GPS Position Information Monitor
All Items Setup (All Items Setup)
Tap the All Items Setup tab. The following screen appears.
5.
Tap each item. Use the displayed list (options) or input box to set the items.
6.
Basic Setup (Basic)
Tap the Basic tab.
3.
Tap Item. Use the displayed list (options) to set the position information.
4.
Tap each item. Use the displayed list (options) or input box to set the items.
5.
Display label
Measurement range
Upper and lower limits of the display range
set the vertical position and vertical zoom.
Latitude (Latitude), Longitude (Longitude), Direction (Direction)
Displays the position information of the channel being configured
Display on/off state of each position information entry
Set the position information to
Display label
2-52
IM DL350-02EN
Page 75
Set position information to Altitude.
Set position information to Velocity.
Display label of each bit
Set position information to Status.
Longitude, Direction, Altitude, or Velocity.
Set position information to Status.
2.13 Configuring the GPS Position Information Monitor
Altitude (Altitude)
Display labelAltitude range
Velocity (Velocity)
Display labelVelocity range
Status (Status)
Configuring the Display (Display)
Tap the Display tab.
3.
Tap Item. Use the displayed list (options) to set the position information.
4.
Tap each item. Use the displayed list (options) or input box to set the items.
5.
Latitude (Latitude), Longitude (Longitude), Direction (Direction),
Altitude (Altitude), Velocity (Velocity)
Set the position information to Latitude,
Status (Status)
Display range
The display range setup menu varies
depending on the specified position information.
Display label
Display group
IM DL350-02EN
Vertical position
(Tap + or – to adjust.)
Vertical zoom
Display group
2-53
Page 76
(Tap + or – to adjust.)
(Can be adjusted in steps)
Chapter 3 Waveform Acquisition
3.1 Setting Conditions for Waveform Acquisition
(Scope mode)
This section explains the following settings for acquiring waveforms.
On the waveform screen, tap MENU > Acquire. The Acquire menu appears.
1.
Tap the Basic tab.
2.
Tap each item. Use the displayed list (options) or input box to set the items.
3.
Time scale
Acquisition Settings (Acquisition Setup)
Tap Acquisition Setup. Tap Acquisition Mode. Use the displayed list (options) to select the
4.
acquisition mode.
Acquisition Mode
Normal: Displays waveforms without processing the sampled data. Set the waveform acquisition
Envelope: Displays waveforms in envelope mode. Set the waveform acquisition count.
Average: Displays averaged waveforms. Set the average count or attenuation constant.
• Normal Mode (Normal), Envelope Mode (Envelope)
Acquisition settingsRecord length
count.
Set to Normal or Envelope.
Waveform acquisition count
(Tap + or – to adjust.)
IM DL350-02EN
3-1
Page 77
(Tap + or – to adjust.)
(Tap + or – to adjust.)
(Tap + or – to adjust.)
(Tap + or – to adjust.)
3.1 Setting Conditions for Waveform Acquisition (Scope mode)
• Averaging Mode (Average)
Set to Average.
Waveform Acquisition Detail Setting Menu
Tap the Detail tab.
2.
Tap Time Base. Use the displayed list (options) to set the time base.
3.
Tap each item. Use the displayed input box to set the items.
4.
When the time base is the internal clock signal
Set to Internal.
When the time base is an external clock signal
Trigger position
Average count
Slide to display counts that do not fit in the screen.
Attenuation constant
Set this when the average count is set to Infinite.
Trigger delay
Set to External.
Pulse/rotation
Trigger position
3-2
IM DL350-02EN
Page 78
Set to Action.
3.1 Setting Conditions for Waveform Acquisition (Scope mode)
Waveform Acquisition Action/SD Recording
Tap the Action/SD Recording tab.
2.
Action (Action)
When the acquisition mode is set to Normal (Normal) or Envelope (Envelop), you can set an action.
Tap Action/SD Recording, and select the action (Action).
3.
Tap each item. Use the displayed list (options) or input box to set the items.
4.
Action Settings (Action Setup)
Tap Action Setup. The following screen appears.
5.
Turns the saving
of waveform
data on and off
Turns the screen
capture save
feature on and off
Save the
screen image*
Save the waveform data*
Action mode
Set the actions you want to execute to ON.
* Items other than beep can be turned on and off on the action setting screen.
Email transmission*
Beep
(Alarm)
Set the actions.
Data format
► “Waveform Save
Conditions” in section 6.3
Waveform data save
destination
► “Setting the File Name”
in section 6.3
Screen image save
destination
► “Setting the File Name”
in section 6.3
IM DL350-02EN
Turns email
transmission on
and off
Email transmission count (Tap + or – to adjust.)
3-3
Page 79
Auto namingFile name
Set to SD Recording.
Comment
3.1 Setting Conditions for Waveform Acquisition (Scope mode)
SD Recording (SD Recording)
When the acquisition mode is set to Normal (Normal) or Envelope (Envelop), you can set SD
recording.
Tap Action/SD Recording, and select SD Recording (SD Recording).
3.
Tap each item. Use the displayed list (options) or input box to set the items.
4.
Detail Settings (Detail Setup)
Tap Detail Setup. The following screen appears.
5.
Detail settings
Turns file division on and off
When file division is set to ON
Number of file divisions
3-4
IM DL350-02EN
Page 80
3.2 Setting Conditions for Waveform Acquisition
The display changes depending
Sample interval
(Recorder mode)
This section explains the following settings for acquiring waveforms.
Tap each item. Use the displayed list (options) or input box to set the items.
5.
Set to Memory + SD Numeric Recording.
Auto naming
Numeric recording interval
SD Recording (SD Recording)
Tap each item. Use the displayed list (options) or input box to set the items.
5.
Comment
Decimal point type
File name
Turns time information on and off
Set to SD Recording.
Turns file division
on and off
File name
Comment
When file division is set to ON
Number of file divisions
IM DL350-02EN
3-9
Page 85
3.3 Starting and Stopping Waveform Acquisition
This section explains how to start and stop waveform acquisition.
Items Common to Scope Mode and Recorder Mode
START/STOP key
Waveform Acquisition (START/STOP)
Press START/STOP to start or stop waveform acquisition.
The key is illuminated while the instrument is acquiring waveforms.
► Features Guide: “Waveform Acquisition (START/STOP)”
3-10
IM DL350-02EN
Page 86
Trigger mode
Chapter 4 Triggering
4.1 Setting the Trigger Mode
This section explains the following setting for updating the displayed waveform.
Applicable to Scope Mode
Trigger mode
This setting is not available in recorder mode.
Waveform Acquisition Basic Setting Menu
On the waveform screen, tap MENU > Acquire. The Acquire menu appears.
1.
Tap the Basic tab.
2.
Tap Trigger Mode. Use the displayed list (options) to select the trigger mode.
3.
► Features Guide: “Trigger mode (Trigger Mode)”
Trigger mode (Trigger Mode)
Auto
If the trigger conditions are met within a 50-ms timeout period, the instrument updates the displayed
waveforms on each trigger occurrence. If not, the instrument automatically updates the displayed
waveforms. If simple trigger is in use and the trigger source is set to Time, the instrument operates
in Normal mode even when Auto mode is specified. If the time axis is set to a value that would
cause the display to switch to roll mode, roll mode display will be enabled.
Normal
The instrument updates the waveform display only when the trigger conditions are met. If no
triggers occur, the display is not updated. If you want to view waveforms that the instrument cannot
trigger on, or if you want to check the ground level, use Auto mode.
Single
When the trigger conditions are met, the instrument updates the displayed waveform once and
stops signal acquisition. If the time axis is set to a value that would cause the display to switch to
roll mode, roll mode display will be enabled. When the instrument triggers, it begins recording data.
When data has been acquired up to the amount specified by the set record length, the waveform
display stops.
On Start
Regardless of the trigger settings, when you press the START key, the instrument updates the
displayed waveforms once and stops signal acquisition. If the time axis is set to a value that would
cause the display to switch to roll mode, roll mode display will be enabled. When data has been
acquired up to the amount specified by the set record length, the waveform display stops.
IM DL350-02EN
4-1
Page 87
4.2 Setting the Trigger Position and Trigger Delay
(Tap + or – to adjust.)
(Tap + or – to adjust.)
This section explains the following setting for updating the displayed waveform.
Applicable to Scope Mode
Trigger position and trigger delay
These settings are not available in recorder mode.
► Features Guide: “Trigger Position (Position)”
“Trigger Delay (Trigger Delay)”
Waveform Acquisition Detail Setting Menu
On the waveform screen, tap MENU > Acquire. The Acquire menu appears.
1.
Tap the Detail tab > Time Base, and select Internal.
2.
Tap each item. Use the displayed input box to set the items.
3.
Trigger position
Trigger delay
4-2
IM DL350-02EN
Page 88
4.3 Triggering on an Edge Trigger
Set to Simple.
Source bit
Source channel
(CH1 to CH4,
16CH VOLT,
16CH
CAN, CAN FD,
LIN, SENT)
Sub channel
This section explains the following settings for triggering on the edges of an analog-signal trigger
source:
For Scope Mode
Setting the trigger to Simple, trigger source, trigger slope, trigger level, trigger hysteresis
For Recorder Mode
Setting the trigger type to Edge, trigger source, trigger slope, trigger level, trigger hysteresis
► Features Guide: “Trigger Settings (Setting),” “Trigger (Trigger)”
On the waveform screen, tap MENU > Acquire. The Acquire menu appears.
1.
For Scope Mode
Tap the Trigger tab > Setting, and select Simple.
2.
Tap each item. Use the displayed list (options) or input box to set the items.
3.
Trigger Source (Source)
Tap Source. Select the trigger source from the list (a context-sensitive list that varies depending
4.
on the installed module).
The displayed options vary depending on the installed module, waveform display on/off state, and
waveform label settings. For CAN, CAN FD, LIN, and SENT, sub channel whose input (Input) is set to OFF
cannot be selected.
TEMP/VOLT,
Logic signal*
► section 4.6
* When the GPS feature is on (see section 18.1), GPS is displayed
as an option in place of CH6. CH6 cannot be selected.
Trigger slopeTrigger sourceTrigger hysteresis
16CH VOLT,
16CH TEMP/VOLT,
CAN, CAN FD, LIN
Slide to display sources
that do not fit in the screen.
Trigger level
(Tap + or – to adjust.)
SENT
Error item
IM DL350-02EN
4-3
Page 89
4.3 Triggering on an Edge Trigger
For Recorder Mode
Tap the Detail tab > Type in the Trigger area, and select Edge.
2.
Tap each item in the Trigger area. Use the displayed list (options) or input box to set the items.