INSTALLATION AND OPERATING PRECAUTIONS FOR TIIS FLAMEPROOF
EQUIPMENT
Revision Information
iv
IM 01F06A00-01EN
<1. INTRODUCTION>
1. INTRODUCTION
1-1
Thank you for purchasing the digitalYEWFLO
vortex fl owmeter.
To ensure correct use of the instrument, please
read this manual thoroughly and fully understand
how to operate the instrument before operating it.
■ Regarding This Manual
• This manual should be provided to the end
user.
• The contents of this manual may be changed
without prior notice.
• All rights are reserved. No part of this manual
may be reproduced in any form without
Yokogawa’s written permission.
• Yokogawa makes no warranty of any kind with
regard to this material, including, but not limited
to, implied warranties of merchantability and
suitability for a particular purpose.
• All reasonable effort has been made to ensure
the accuracy of the contents of this manual.
However, if any errors or omissions are found,
please inform Yokogawa.
• The specifi cations covered by this manual are
limited to those for the standard type under the
specifi ed model number break-down and do not
cover custom-made instruments.
• Please note that this manual may not be
revised for any specifi cation changes,
construction changes or operating part changes
that are not considered to affect function or
performance.
• Yokogawa assumes no responsibilities for this
product except as stated in the warranty.
• If the customer or any third party is harmed by
the use of this product, Yokogawa assumes
no responsibility for any such harm owing to
any defects in the product which were not
predictable, or for any indirect damages.
■ Safety and Modifi cation Precautions
• The following general safety precautions must
be observed during all phases of operation,
service, and repair of this instrument. Failure
to comply with these precautions or with
specifi c WARNINGS given elsewhere in
this manual violates safety standards of
design, manufacture, and intended use of the
instrument. Yokogawa assumes no liability for
the customer’s failure to comply with these
requirements. If this instrument is used in
a manner not specifi ed in this manual, the
protection provided by this instrument may be
impaired.
• Yokogawa will not be liable for malfunctions or
damage resulting from any modifi cation made
to this instrument by the customer.
• The following safety symbol marks are used in
this manual and instrument.
WARNING
A WARNING sign denotes a hazard. It calls
attention to procedure, practice, condition or the
like, which, if not correctly performed or adhered
to, could result in injury or death of personnel.
CAUTION
A CAUTION sign denotes a hazard. It calls
attention to procedure, practice, condition or the
like, which, if not correctly performed or adhered
to, could result in damage to or destruction of the
product.
IMPORTANT
An IMPORTANT sign denotes that attention is
required to avoid damage to the instrument or
system failure.
NOTE
A NOTE sign denotes information necessary
for essential understanding of operation and
features.
IM 01F06A00-01EN
<1. INTRODUCTION>
1-2
1.1 Using This Instrument Safety
(1) Installation
WARNING
• Installation of the vortex fl owmeter must
be performed by expert engineer or skilled
personnel. No operator shall be permitted to
perform procedures relating to installation.
• The vortex fl owmeter must be installed within
the specifi cation conditions.
• The vortex fl owmeter is a heavy instrument.
Be careful that no damage is caused to
personnel through accidentally dropping
it, or by exerting excessive force on the
vortex fl owmeter. When moving the vortex fl owmeter, always use a trolley and have at
least two people carry it.
• When the vortex fl owmeter is processing
hot fl uids, the instrument itself may become
extremely hot. Take suffi cient care not to get
burnt.
• Where the fl uid being processed is a toxic
substance, avoid contact with the fl uid and
avoid inhaling any residual gas, even after
the instrument has been taken off the piping
line for maintenance and so forth.
• Do not apply excessive weight, for example,
a person stepping on the vortex fl owmeter.
• Do not open the cover in wet weather or
humid environment. When the cover is open,
stated enclosure protection is not applicable.
• When opening the cover, wait for more than
2 minutes after turning off the power.
• All procedures relating to installation must
comply with the electrical code of the country
where it is used.
(2) Wiring
WARNING
• The wiring of the vortex fl owmeter must
be performed by expert engineer or skilled
personnel. No operator shall be permitted to
perform procedures relating to wiring.
• When connecting the wiring, check that the
supply voltage is within the range of the
voltage specifi ed for this instrument before
connecting the power cable. In addition,
check that no voltage is applied to the power
cable before connecting the wiring.
(3) Operation
WARNING
• Do not open the cover in wet weather or
humid environment. When the cover is open,
stated enclosure protection is not applicable.
• When opening the cover, wait for more than
2 minutes after turning off the power.
(4) Maintenance
WARNING
• Maintenance of the vortex fl owmeter should
be performed by the trained personnel
having knowledge of safety standard. No
operator shall be permitted to perform any
operations relating to maintenance.
• Do not open the cover in wet weather or
humid environment. When the cover is open,
stated enclosure protection is not applicable.
• When opening the cover, wait for more than
2 minutes after turning off the power.
• Always conform to maintenance procedures
outlined in this manual. If necessary, contact
Yokogawa.
IM 01F06A00-01EN
<1. INTRODUCTION>
1-3
(5) Explosion Protected Type Instrument
WARNING
• The instruments are products which have
been certifi ed as explosion proof type
instruments. Strict limitations are applied
to the structures, installation locations,
external wiring work, maintenance and
repairs, etc. of these instruments. Suffi cient
care must be taken, as any violation of the
limitations may cause dangerous situations.
Be sure to read Chapter 14 “EXPLOSION
PROTECTED TYPE INSTRUMENT”
before handling the instruments. For TIIS
fl ameproof type instruments, be sure to
read “INSTALLATION AND OPERATING
PRECAUTIONS FOR TIIS FLAMEPROOF
EQUIPMENT” at the end of this manual.
• Only trained persons use this instrument in
the industrial location.
• Take care not to generate mechanical
spark when access to the instrument and
peripheral devices in hazardous locations.
(6) European Pressure Equipment Directive
(PED)
WARNING
• When using the instrument in compliance
with PED, be sure to read Chapter 15 “PED
(PRESSURE EQUIPMENT DIRECTIVE)”
before use.
1.2 Warranty
• The terms of this instrument that are
guaranteed are described in the quotation.
We will make any repairs that may become
necessary during the guaranteed term free of
charge.
• Please contact our sales offi ce if this instrument
requires repair.
• If the instrument is faulty, contact us with
concrete details about the problem and the
length of time it has been faulty, and state the
model and serial number. We would appreciate
the inclusion of drawings or additional
information.
• The results of our examination will determine
whether the meter will be repaired free of
charge or on an at-cost basis.
■ The guarantee will not apply in the
following cases:
• Damage due to negligence or insuffi cient
maintenance on the part of the customer.
• Problems or damage resulting from handling,
operation or storage that violates the intended
use and specifi cations.
• Problems that result from using or performing
maintenance on the instrument in a location
that does not comply with the installation
location specifi ed by Yokogawa.
• Problems or damage resulting from repairs or
modifi cations not performed by Yokogawa or
someone authorized by Yokogawa.
• Problems or damage resulting from
inappropriate reinstallation after delivery.
• Problems or damage resulting from disasters
such as fi res, earthquakes, storms, fl oods, or
lightning strikes and external causes.
■ Trademarks:
• ‘digitalYEWFLO’, ‘DY’, ‘DYA’, ‘DYC’, and
‘BRAIN TERMINAL’ are registered trademarks
of Yokogawa Electric Corporation. Company
names and product names used in this material
are registered trademarks or trademarks of their
respective owners.
• In this manual, trademarks or registered
trademarks are not marked with ™ or ®.
IM 01F06A00-01EN
<1. INTRODUCTION>
1.3 ATEX Documentation
This is only applicable to the countries in European Union.
1-4
GB
DK
E
NL
SK
CZ
I
LT
LV
EST
PL
SF
P
F
D
S
SLO
H
BG
RO
M
GR
IM 01F06A00-01EN
<2. HANDLING PRECAUTIONS>
2. HANDLING PRECAUTIONS
2-1
The Model DY Vortex Flowmeter and Model DYA
Vortex Flow Converter are thoroughly tested at the
factory before shipment. When these instruments
are delivered, perform a visual check to ascertain
that no damage occurred during shipment.
This section describes important cautions in
handling these instruments. Read carefully before
using them.
If you have any problems or questions, contact
your nearest YOKOGAWA service center or sales
representative.
2.1 Checking Model and
Specifi cations
The model and important specifi cations are
indicated on the name plate attached to the case.
Verify that they are the same as those specifi ed in
the original order, refer to Chapter 13 “GENERAL
SPECIFICATIONS .” In any correspondence,
always give model (MODEL) and serial number
(NO.) from the name plate.
3UA
MPa at 38°C
TAG NO.
4 ~ 20mA DC / PULSE
*1)
2.2 Transportation and Storage
Precautions
If the instrument is to be stored for a long period of
time after delivery, observe the following points.
(1) The instrument should be stored in its original
packing condition in the storage location.
(2) Select a storage location that fulfi ls the following
conditions:
• A place where it will not be exposed to rain or
water
• A place subject to minimal vibrations or shocks
• Temperature and humidity levels should be as
follows:
Temperature:-40 to +80°C
Humidity:5 to 100% RH (no condensation)
The preferred ambient temperature and
humidity levels are +25°C and approximately
65% RH.
(3) If the digitalYEWFLO vortex fl owmeter is
transferred to the installation site and stored
without being installed, its performance may
be impaired due to the infi ltration of rainwater
and so forth. Be sure to install and wire the
digitalYEWFLO vortex fl owmeter as soon as
possible after transferring it to the installation
location.
(4) The vortex fl owmeter is a heavy instrument. Be
careful that no damage is caused to personnel
through accidentally dropping it, or by exerting
excessive force on the vortex fl owmeter. When
moving the vortex fl owmeter, always use a
trolley and have at least two people carry it.
10.5 ~ 42V DC
*2)
F0201.ai
Figure 2.1(a) Example of Name Plate for Integral Type
TAG NO.
TAG NO.
F0202.ai
10.5 ~ 42V DC
MPa at 38°C
*1)
*2)
4 ~ 20mA DC / PULSE
*1)
*2)
Figure 2.1(b) Example of Name Plate for Remote Type
*1): K factor at + 15°C
*2): The product - producing country.
3UA
3WA
3YA
IM 01F06A00-01EN
<3. INSTALLATION>
3. INSTALLATION
3-1
WARNING
This instrument must be installed by expert
engineer or skilled personnel. The procedures
described in this chapter are not permitted for
operators.
3.1 Installation Precautions
(1) Ambient Temperature
Avoid an area which has wide temperature
variations. When the installation area is
subjected to heat radiation from process plant,
ensure adequate heat prevention or ventilation.
(2) Atmospheric Conditions
Avoid installing the vortex fl owmeter in a
corrosive atmosphere. When the vortex
fl owmeter must be installed in a corrosive
atmosphere, adequate ventilation must be
provided
(4) Precautions Regarding Piping
(a) Ensure that the process connector bolts are
tightened fi rmly.
(b) Ensure that no leak exists in the process
connection pipeline.
(c) Do not apply a pressure higher than the
specifi ed maximum working pressure.
(d) Do not loosen or tighten the fl ange mounting
bolts when the assembly is pressurized.
(e) Handle the vortex fl owmeter carefully when
measuring dangerous liquids, so that the liquids
do not splash into eyes or on face. When using
dangerous gases, be careful not to inhale them.
(5) Other Considerations
• Choose a location where is suffi cient clearance
around digitalYEWFLO exist to allow such work
as routine inspections.
• Choose a location that ensures easy wiring and
piping.
3.2 Piping Precautions
(3) Mechanical Shock or Vibration
The vortex fl owmeter is of sturdy construction,
but select an area subject to minimize
mechanical vibration or impact shock. If
the fl owmeter is subject to vibrations, it is
recommended that pipeline supports to be
provided as shown in Figure 3.1.
digitalYEWFLO
Vortex Flowmeter
Pipeline
Pipeline Support
Figure 3.1 Example of Pipeline Support
F0301.ai
Straight Pipe Length and
Recommendations
Refer to Table 3.1 about Valve Position and
Straight Pipe Length and so on.
● Piping support
Typical vibration immunity level is 1G for normal
piping condition.Piping support should be fi xed in
case of over 1G vibration level.
● Installation direction
If a pipe is always fi lled with liquids, the pipe can be
installed vertically or at inclined angle.
● Adjacent pipes
The process pipline inner diameter should be larger
than the digitalYEWFLO inner diameter.
Use the following adjacent pipe.
Model CodeAdjacent Pipe
DY015 up to DY050
DY025/R1 up to DY080/R1
DY040/R2 up to DY100/R2
DY080 up to DY400
DY100/R1 up to DY200/R1
DY150/R2 up to DY200/R2
Sch40
or larger inner
diameter than
Sch40
Sch80
or larger inner
diameter than
Sch80
IM 01F06A00-01EN
<3. INSTALLATION>
Table 3.1 (a) Straight pipe length and recommendations (1)
D: Nominal diameter (mm)
DescriptionFigure
Reducer pipe:
Ensure the upstream straight pipe length to be 5D or more, and the
downstream straight pipe length to be 5D or more for per reducer
pipe.
Reducer
Flow
5D or more
digitalYEWFLO
5D or more
3-2
Expander pipe:
Ensure the upstream straight pipe length to be 10D or more, and
the downstream straight pipe length to be 5D or more for per
expander pipe.
Bent pipe and straight pipe length:
1. Single bent pipe
2. Double bent pipe; coplanar
3. Double bent pipe; non coplanar
Valve position and straight pipe length:
Install the valve on the downstream side of the flowmeter.
The upstream straight pipe length dependent on the element
located on the upstream such as reducer/expander, bent and etc.,
refer to description as above. Keep 5D or more for downstream
straight pipe length.
In case the valve has to be installed on the upstream of the
fl owmeter, ensure the upstream straight pipe length to be 20D or
more, and the downstream straight pipe length be 5D or more.
Expander
1.
2.
3.
Refer to each element above for
straight pipe run.
Flow
digitalYEWFLO
Flow
10D or more
5D or more
digitalYEWFLO
Flow
10D or more5D or more
digitalYEWFLO
Flow
10D or more5D or more
digitalYEWFLO
Flow
20D or more5D or more
digitalYEWFLO
Valve
digitalYEWFLO
20D or more
5D or more
IM 01F06A00-01EN
<3. INSTALLATION>
Table 3.1 (b) Straight pipe length and recommendations (2)
D: Nominal diameter (mm)
DescriptionFigure
Fluid vibration:
For a gas line which uses a position-type or roots-type blower
compressor or a high-pressure liquid line (about 1MPa or more)
which uses piston-type or plunger-type pump, fluid vibrations may
be produced.
In these case, install valve on the upstream side of digitalYEWFLO.
For inevitable fluid vibration, put a vibration damping device such
as throttling plate or expansion section in the upstream side of
digitalYEWFLO.
Piston-type or plunger pump:
Install the accumulator on the upstream side of digitalYEWFLO to
reduce fl uid vibrations.
Valve positon (T-type piping exist):
When pulsation causes by a T-type piping exist, install the valve on
the upstream of the fl owmeter.
Example: As shown in the figure, when the valve V1 is turned off,
the fluid flow throught B as to meter A the flow is zero. But due to
the pulsating pressure is detected, the meter is zero point become
fl uctuating. To avoid this, change the valve V1 location to V1'.
Note: In case of the Reduced Bore Type, moisture may be
remained upstream of the fl owmeter. Drain it appropriately.
Flow
B
Relocating
V1’V1
digitalYEWFLO
digitalYEWFLO
digitalYEWFLO
digitalYEWFLO
A
3-3
Valve (Off)
Pressure and Temperature Taps:
Pressure tap outlet: install this tap between 2D and 7D on the
downstream side of a fl owmeter.
Temperature tap outlet: install this on the downstream side 1D to
2D away from a pressure tap.
Mounting Gasket:
Avoid mounting gaskets which protrude into the pipe line. This may
cause inaccurate readings.
Use the gaskets with bolt holes, even if digitalYEWFLO is the wafer
type.
When using a spiral gasket (without bolt holes), confirm the size
with the gasket -manufacturer, as standard items may not be used
for certain fl ange ratings.
Flow
Upstream
No good
Pressure tap
digitalYEWFLO
digitalYEWFLO
Temperature tap
downstream
2 to 7D1 to 2D
Pipeline Flange
Pipeline
IM 01F06A00-01EN
<3. INSTALLATION>
Table 3.1 (c) Straight pipe length and recommendations (3)
DescriptionFigure
Heat-Insulation:
When an integral-type fl owmeter or a remote type detector is
installed and the pipe carrying higt-temperature fl uids is heat-
insulated, do not wrap adiabatic materials around the installation
the bracket (DY015 to DY100) or the nozzle (DY150 to DY400) of
the converter.
digitalYEWFLOdigitalYEWFLO
Bracket
Heat-InsulatorHeat-Insulator
3-4
Nozzle
Note: Refer to Section 3.4 "Cryogenic and High Process
Temperature Version Insulation" and install it rightly.
Flushing of the pipe line:
Flush and clean scale, incrustation and sludge on the inside of
pipe for newly installed pipe line and repaired pipe line before the
operation. For fl ushing, the fl ow should fl ow through bypass-piping
to avoid damaging the fl owmeter. If there is no bypass-piping,
install short pipe instead of the fl owmeter.
Mounting Precautions
WARNING
In case of high process temperature, care
should be taken not to burn yourself because
the surface of body and case reach a high
temperature.
[DY015 to DY100][DY150 to DY400]
digitalYEWFLO
Short pipe
(2) Liquid Measurement Precautions
To insure accurate measurement, the
digitalYEWFLO must always have a full pipe.
• Piping Requirements for Proper Operation
Allow the fl ow to fl ow against gravity. When
the fl ow is moving with gravity, lift the down-
stream pipe length above the digitalYEWFLO
installation level to maintain full pipeline.
(1) Gas or Steam Measuring Precautions
• Piping to Prevent Standing Liquid
Mount digitalYEWFLO in a vertical pipeline
to avoid liquid traps. When digitalYEWFLO
is installed horizontally, raise that part of
the pipeline in which the digitalYEWFLO is
installed.
(Good)
(Good)
Flow
Flow
(No Good)
Flow
F0302.ai
Flow
(No Good)(No Good)
(Good)
h h>0
Flow
Flow
(Good)
h
h>0
Flow
F0303.ai
IM 01F06A00-01EN
<3. INSTALLATION>
3-5
• Piping for Avoiding Bubbles
Flows containing both gas and liquid cause
problems. Avoid gas bubbles in a liquid fl ow.
Piping should be carried out to avoid bubble
generation.
Install the valve on the downstream side of the
fl owmeter because pressure drop across the
control valve may cause gas to come out of the
solution.
(Good)
Control
Value
Flow
(Good)
Flow
Flow
(No Good)
(No Good)
(3) Multi-Phase Flow
digitalYEWFLO can measure gas, liquid
and steam when there is no change in state.
However, accurate measurement of mixed
fl ows (e.g. gas and liquid) is not possible.
(No Good)
F0304.ai
(4) Pipeline Diameter and digitalYEWFLO
The process pipeline inner diameter should be
slightly larger than the vortex fl owmeter inner
diameter, schedule 40 or lower pipe should be
used for 1/2 to 2 inch fl owmeters and schedule
80 or lower pipes for 3 to 16 inch fl owmeters.
(No Good)(Good)
D
D
1
2
<
D
D
1
2
D1D
D
2
D
1
2
F0306.ai
(5) Waterproof Construction
The vortex fl owmeter is of IP67, Type 4X, JIS C
0920 watertight protection. However, it cannot
be used under water.
3.3 Maintenance of Piping
(1) Pipe cleaning
• Flushing of pipe line (Cleaning)
Flush and clean scale, incrustation and sludge
on the inside of pipe wall for newly installed pipe
line and repaired pipe line before the operation.
• Fluid Carrying Solids
Do not measure fl uids that carry solids
(e.g. sand and pebbles). Make sure users
periodically remove solids adhering to the
vortex shedder.
• Obstruction of fl ow fl uids may cause to make
a chemical reaction and the fl uid will be
crystallized and hardened, and be deposited on
the pipe wall and shedder bar.
In those cases, clean shedder bar.
Mist flow
(No Good)
Stratified flow
(No Good)
Bubble flow
Liquid
Flow
Gas Flow
F0305.ai
(2) Bypass piping
Bypass piping is convenient for the
maintenance of digitalYEWFLO (vortex
shedder cleaning, etc.).
Bypass shut-off valve
digitalYEWFLO
Flow
Upstream shut-off valveDownstream shut-off valve
F0307.ai
IM 01F06A00-01EN
<3. INSTALLATION>
3-6
3.4 Cryogenic and High Process
Temperature Version
Insulation
When you are using Cryogenic and High Process
Temperature version of digitalYEWFLO Vortex
Flowmeter (Option code: /HT, /LT), refer to following
contents.
Installing Cryogenic Version
For cryogenic applications, use stainless steel
mounting bolts and nuts to install the fl owmeter.
These can be ordered separately from
YOKOGAWA. Cover the fl owmeter body with
heat insulating material so that the fl owmeter
can be maintained at ultra-low temperatures.
Maintenance for Cryogenic Applications
DY/LT uses special materials that produce
vortex fl owmeter for cryogenic applications.
When you are replacing a shedder bar, specify
Cryogenic Version shedder bar. To avoid
condensing in the terminal box, ensure that the
wire connecting port is well sealed.
Bracket
Cold insulating material
CAUTION
Keep the upper limit of heat insulating material to
prevent overheating of the terminal box.
Seal the Heat-Insulator to avoid hot-air leakage.
50mm min.50mm min.
UPPER LIMIT OF
Heat-Insulator
Nominal Size: 100mm or under
UPPER LIMIT OF
Heat-Insulator
Nominal Size: 150mm or over
F0309.ai
Maintenance for High Process
Temperature Applications
DY/HT uses special materials that produce
vortex fl owmeter for High Process Temperature
applications When you are replacing a
shedder bar or a gasket, specify High Process
Temperature Version.
3.5 Mounting Procedures
F0308.ai
Installing High Process Temperature
Version
Installation of the fl owmeter is the same as the
standard type. Cover the fl owmeter body with
heat insulating material following instruction of
“CAUTION”.
WARNING
The Vortex Flowmeter is a heavy instrument.
Please be careful to prevent persons from
injuring whin it is handled.
Before installing the instrument verify the following.
The direction of fl ow should match to the arrow
mark on the instrument body. When changing the
orientation of the terminal box, refer to Chapter 11
“MAINTENANCE.”
1. Installation of Vortex fl owmeter of the wafer and fl ange type is shown in Table 3.3.
When installing the wafer type vortex fl owmeter,
it is important to align the instrument bore with
the inner diameter of the adjacent piping.
To establish alignment, use the four collars
supplied with the instrument.
IM 01F06A00-01EN
<3. INSTALLATION>
3-7
• Four collars are supplied for 1/2 inch (15mm) to
1- 1/2inch (40mm), 2 inch of JIS 10K or ANSI
class 150, and 3 inch of ANSI class 150. Install
the instrument as illustrated in Table 3.3.
• If the adjacent fl anges have eight bolt holes,
insert the stud bolts in the holes on the
instrument shoulder.
• Stainless steel stud bolts and nuts are available
on order. When they are to be supplied by the
user, refer to Table 3.2 for stud bolt length.
Gaskets must be supplied by the user.
Pipeline Flange
Pipeline
F0311.ai
2. Avoid mounting gaskets which protrude into the
pipeline. This may cause inaccurate readings.
Use gaskets with bolt holes, even if
digitalYEWFLO is of the wafer type.
When using a spiral gasket (without bolt holes),
confi rm the size with the gasket-manufacturer,
as standard items may not be used for certain
fl ange ratings.
Table 3.2 Flange Rating
Size
mm
(inch)
15mm
(1/2B)
25mm
(1B)
40mm
(1-1/2B)
50mm
(2B)
80mm
(3B)
100mm
(4B)
d
Flange Rating
JIS 10K, 20K/DIN 10,
16,25,40
JIS 40K
ANSI 150, 300, 600
JIS 10K, 20K, 40K
ANSI 150
ANSI 300, 600
DIN 10, 16, 25, 40
JIS 10K, 20K/DIN 10,
16, 25, 40
JIS 40K
ANSI 150
ANSI 300, 600
JIS 10K, 20K, 40K/ DIN
10, 16, 25, 40 ANSI
150, 300, 600
JIS 10K/DIN 10, 16,
25, 40
JIS 20K, 40K
ANSI 150
ANSI 300, 600
JIS 10K/DIN 10, 16
JIS 20K/DIN 25, 40
JIS 40K
ANSI 150
ANSI 300
ANSI 600
Major Diameter of
External Threed
of Stud Bolt d
(mm)
Length "
Collar
12
16
12.7
16
12.7
15.9
12
16
20
12.7
19.1
16
15.9
16
20
15.9
19.1
16
20
22
15.9
19.1
22.2
Length
(mm)
160
160
155
160
155
160
160
160
170
155
170
200
200
220
240
240
240
220
240
270
240
240
270
Stud Bolt
F0310.ai
IM 01F06A00-01EN
<3. INSTALLATION>
Table 3.3 (a) Installation of Wafer Type Vortex Flowmeter
3-8
Wafer type
When Installation Collar are required, the
installation vortex flowmeters applied to the
following line sizes and flange ratings.
Size mm (inch)
15 to 40
(1/2 to 1-1/2)
50(2)
80(3)
Flange Rating
All ratings
JIS 10K, ANSI class 150,
DIN PN10 to PN40
ANSI class 150
WARNING
The inside diameter of the gasket must
be larger than the pipe inner diameter
so that it will not disturb the flow in the
pipeline.
WARNING
When installing the Flowmeter vertically in the
open air, change the electrical connection port
direction to the ground. If the electrical
connection port is installed upwards, rain
water might leak in.
Horizontal Installation
Flow
Direction
Nut
Gasket
Vertical Installation
Gasket
Collar (4 pcs.)
Gasket
Description
Electrical
Connection
Flange
Collar (4 pcs.)
Nut
Gasket
Flange
Flange
Electrical
Connection
Flange
Bolt (4 pcs.)
Nut
Bolt (4 pcs.)
Nut
WARNING
In case of vertical installation, two collars in
the upper part might move after the
installation. But it doesn't influence the
performance, please use the flowmeter under
such condition.
When Installation Collars are not required,the
installation vortex flowmeters applied to the
following line sizes and flanges.
Size mm (inch)Flange Rating
50(2)
80(3)
100(4)
JIS 20K, 40K
ANSI class 300,600
JIS 10K, 20K, 40K
ANSI class 300, 600
JIS 10K, 20, 40K
ANSI class 150, 300, 600
Flow Direction
(1) Insert two collars on each two bolts of bottom side of the flowmeter.
(2) Fit the flowmeter body to the collars. And tighten the four bolts and nuts
uniformly.
(3) Check for leakage from the flange connections.
Vertical InstallationHorizontal Installation
Flow
Direction
Nut
Stud Bolt (8 pcs.)
Flange
Gasket
(1) Insert two stud bolts in the bolt holes
on the flowmeter shoulder to align
the instrument body with the inner
diameter of the adjacent piping.
(2) Tighten all bolts uniformly and check
that there is no leakage between the
instrument and the flanges.
Electrical Connection
Bolt Hole
Flange
Gasket
Nut
Electrical
Connection
Flow Direction
F0312.ai
IM 01F06A00-01EN
<3. INSTALLATION>
Table 3.3 (b) Installation of Flange Type Vortex Flowmeter
Flange typeDescription
3-9
Use the stud bolts and nuts supplied with the
flowmeter of the user.
The gaskets should be supplied by the user.
Flow Direction
CAUTION
The inside diameter of the gasket must be
larger than the pipe inner diameter so that it
will not disturb the flow in the pipeline.
Flow Direction
Vertical Installation
Table 3.3 (c) Installation of Remote Type Converter
Remote type converterDescription
Nut
Stud Bolt
Horizontal Installation
Flange
Flange
Gasket
Nut
Gasket
F0313.ai
CAUTION
A signal cable (DYC) is used between the
remote type flowmeter and the converter.
The maximum signal cable length is 97.5ft
(30m).
The converter is mounted on a 2-inch (60.5mm outer dia.) stanchion or horizontal
pipe.
Do not mount the converter on a vertical pipe. It makes wiring and maintenance
difficult.
The converter mounting orientation can be changed as illustrated below.
Horizontal Pipe MountingStanchion Mounting
Nut
Bracket
2-inch Pipe
U-Bolt
F0314.ai
IM 01F06A00-01EN
<4. WIRING>
4. WIRING
4-1
WARNING
The wiring of the vortex fl owmeter must
be performed by expert engineer or skilled
personnel. No operator shall be permitted to
perform procedures relating to wiring.
CAUTION
Once all wiring is complete, check the
connections before applying power to the
instrument. Improper arrangements or wiring
may cause a unit malfunction or damage.
4.1 Load Resistance of Output
Condition
Be sure to observe the following precautions when
wiring:
CAUTION
• In cases where the ambient temperature
exceeds +50°C (+122°F), use external heatresistant wiring with a maximum allowable
temperature of +70°C (+158°F) or above.
• Do not connect cables outdoors in wet
weather in order to prevent damage from
condensation and to protect the insulation.
• Do not splice the cable between the fl owtube
terminal and the converter if it is too short.
Replace the short cable with a cable that is
the appropriate length.
• All the cable ends must be provided with
round crimp-on terminals and be securely
wired.
• Be sure to turn power off before opening the
cover.
• Before turning the power on, tighten the
cover securely.
• Explosion protected types must be wired in
accordance with specifi c requirement (and,
in certain countries, legal regulations) in
order to preserve the effectiveness of their
explosion protected features.
• The terminal box cover is locked by the
Locking Screw. In case of opening the
terminal box cover, use the hexagonal
wrench attached.
• Be sure to lock the cover by the Locking
Screw using the hexagonal wrench attached
after installing the cover.
Table 4.1 shows the connection method of several
output conditions.
(1) Analog Output (4 to 20 mA DC)
This converter uses the same two wires for
both, the signal and power supply. A DC power
supply is required in a transmission loop.
The total leadwire resistance including the
instrument load and power distributor (supplied
by the user) must conform to a value in the
permissible load resistance range. Refer to
Figure 4.1.
IM 01F06A00-01EN
<4. WIRING>
4-2
600
E–10.5
R=
0.0236
250
Load resistance R ()
10.516.424.7
Power Supply Voltage E (V)
Figure 4.1 Relationship between Power Supply
Voltage and Load Resistance (4 to 20
mA DC Output)
Communication
Applicable range
BRAIN and HART
3042
F0401.ai
(2) Pulse output and Alarm, Status Output
This version uses three wires between the
converter and the power supply. A DC power
and load resistance are required, and pulse
output is connected to a totalizer or an electric
counter. Low level of the pulse output is 0
to 2V. No communication is possible over
a transmission line. Communication via the
amplifi er board is always possible irrespective
of the wiring condition.
4.2 Selection of Wires
The following should be taken into consideration
when selecting cables for use between the
converter and distributor.
(1) Use 600V PVC insulated wire or equivalent
standard wire or cable.
(2) Use shielded wire in areas susceptible to
electrical noise (both analog and pulse output
versions).
(3) In areas with high or low ambient temperatures,
use wires or cables suitable for such
temperatures.
(4) In atmospheres where oils or solvents,
corrosive gases or liquids may be present, use
suitable wires or cables.
(5) Use cable which withstand temperature up to
+60°C and more, when ambient temperature is
more than +60°C.
IMPORTANT
For the remote type, use DYC signal cable
to connect the converter and remote type
fl owmeter(DY-N).
4.3 Connection
(3) Simultaneous Analog-Pulse Output
When using digitalYEWFLO in the
simultaneous analog -pulse output mode, the
communicable distance of the transmission
line is restricted on the wiring method. Table
4.1 shows the examples of connection for this
output mode. Communication via the amplifi er
board is always possible irrespective of the
wiring condition.
IMPORTANT
For pulse output and the simultaneous analogpulse output ,use the load resistance. Refer to
Table 4.1.
Table 4.1 shows the connection sample of
connection for power supply and load resistance.
The terminal position of each connection is shown
in Figure 4.2.
Integral type
Supply
+
–
Pulse
+
Figure 4.2 Terminal Position
Remote type
4 to 20 mA DC Output Power Supply
and Output Signal Terminals
Pulse Output Terminal
Grounding Terminal
F0402.ai
IM 01F06A00-01EN
<4. WIRING>
Table 4.1 (a) The wiring example for the analog and pulse and status, alarm output.
ConnectionDescription
Analog Output
In this case, Communication
is possible (up to a distance
of 2km when a CEV cable is
used.)
digitalYEWFLO Electrical Terminal
+
SUPPLY
–
PULSE
+
+
–
Distributor
24V DC
250Ω
4-3
Pulse Output
In this case, No
communication is possible.
Status Output
Alarm Output
In this case,
No communication is
possible.
digitalYEWFLO Electrical Terminal
+
SUPPLY
Shielded Cable
–
PULSE
+
digitalYEWFLO Electrical Terminal
Shielded Cable
+
SUPPLY
–
PULSE
+
E
*2
R
Electric counter
External Power supply
30V DC, 120mA max
(Contact Rating)
Use the Three-wire shielded cable.
This supply voltage requires
a power sourse with a
maximum output current of
no less than E/R+25mA.
*1
Use the Three-wire shielded cable.
E
*1: To avoid the infl uence of external noise, use an electric counter which fi ts to the pulse frequency.
*2: Resistor is not necessary in case of an electric counter which can receive contact pulse signal directly.
Relay
AC power supply
Magnetic
valve
IM 01F06A00-01EN
<4. WIRING>
4-4
Table 4.1 (b) The wiring example for the simultaneous analog and pulse output, the calculation formula of the
range of load registance R for the pulse output.
ConnectionDescription
Simultaneous
Analog
-Pulse Output
Example 1
In this case, Communication
is possible(up to a distance
of 2km when a CEV cable is
used).
Example 2
In this case, Communication
is possible (up to a distance
of 200m when a CEV cable
is used) and R = 1k).
Example 3
In this case, No communi
-cation is possible (when
shielded cable is not used).
When analog and pulse output are used, the length of communication line is subjected to wiring conditions. Refer to
example 1 to 3. If the communication carries out from amplifier, no need to consider wiring conditions.
*3
Shield
Shielded Cable
+
SUPPLY
digitalYEWFLO Electrical Terminal
SUPPLY
PULSE
digitalYEWFLO Electrical Terminal
SUPPLY
PULSE
–
+
Shield
Shielded Cable
+
–
Outer Jacket
+
+
–
PULSE
+
digitalYEWFLO Electrical Terminal
Outer Jacket
Distributor (or communication medium)
For the shielded cables in this
example of flowmeter installation,
use two-wire separately
shielded cables. *4
This supply voltage requires a
power sourse with a maximum
output current of no less than
E/R.
(or communication medium)
*2
(R)
E(10.5 to 30V DC)
Counting input
Common
Electric counter *1
*2
(R)
E(16.4 to 30V DC)
*2
(R)
Counting input
Common
Electric counter *1
(or communication medium)
24V DC
250
Recorder or
other instrument
E(16.4 to 30V DC)
Counting input
Common
Electric counter *1
(or communication medium)
Recorder or
other instrument
For the shielded cables in this
example of flowmeter installation,
use two-wire separately shielded
cables. *4
This supply voltage requires a
power sourse with a maximum
output current of no less than
E/R+25mA.
The supply voltage requires output
impedance no more than 1/1000
of R (load resistance).
This supply voltage requires
a power sourse with a
maximum output current of
no less than E/R+25mA.
The range of load
The load resistance should be selected by calculation as shown below.
resistance R for the
pulse output.
E (V)
120
P (mW) =
R (k
E
R (k
)
2
(V)
)
0.1
μF ) × f ( kHz )
C (
Where
E = Supply voltage (V)
f = Frequency of pulse output (kHz)
R = Value of load resistance (k
Example of CEV cable capacitance
0.1μF/km
)
C = Cable capacitance (μF)
P = Power ratio of the load resistance
(mW)
*1: To avoid the infl uence of external noise, use an electric counter which fi ts to the pulse frequency.
*2: Resistor is not necessary in case of an electric counter which can receive contact pulse signal directly.
*3: When using analog and pulse output simultaneously, the HART communication may be infl uenced by noise comparing analog output
only.
*4: Signal Cable for ADMAG AXF, AXFC-0 (No Termination) is available.
IM 01F06A00-01EN
<4. WIRING>
Unit : mm
C
(Black)B(White)A(Red)
A
(Red)B(White)C(Black)
(Blue)
Specified
Length (L)
30m (max.)
DYC
Flowmeter
Converter
80
70
70
80
60
25
95
60
20
50
50
T
(Yellow)
T
(Yellow)
T: Only for / MV
4-5
4.4 Connection of DYC Signal
Cable
The remote type signal cable is shown in Figure 4.3
and Figure 4.4, and the terminal is shown in Figure
4.5.
The maximum cable length is 30 m (97.5 feet).
Remove terminal box cover and wiring connection
dust-cap before wiring.
For remote type the converter has two electrical
connections (cable inlets). Use the left connection
as viewed from the terminal box for the DYC signal
cable and the right connection for the transmission
cable.
If a signal cable kit is supplied by YOKOGAWA,
both ends of the cable must be fi nished in
accordance with the following instructions. Refer to
Section 4.5 “End Processing Method of DYC Signal
Cable“.
CAUTION
TT
C
B
A
A
B
C
T: Only for /MV
Input Terminal from built-
T
in temperature sensor
Input Terminals from
A
vortex detector
B
Common Terminal
C
F0405.ai
A
B
C
Flowmeter(DY-N)Converter(DYA)
DYC
T
Detector (DY-N.../E1)
Figure 4.5 Terminal of Detector and Converter
Signal Cable(DYC)
Vortex Flow Converter
(Yellow) T
(White) B
(Black) C
After completing the signal cable connections,
install the shielded cover to signal cable terminal
as shown in Figure 4.6.
Figure 4.3 DYC Signal Cable
Outer shield
To Flowmeter
(Red) A
Inner shield
T: Only for / MV
Figure 4.4 Construction of DYC Signal Cable
To Converter
T
(Yellow)
A (Red)
B (White)
C (Black)
(Bule)
F0404.ai
Shield Cover
Power Cable
F0406.ai
Figure 4.6 Shielded Cover
IM 01F06A00-01EN
<4. WIRING>
4.5 End Processing Method of DYC Signal Cable
4.5.1 For Vortex Flowmeter (DY-N)
DescriptionFigure
Strip off the outer polyethylene jacket, outer braided
1
shield and inner jacket, and inner braided shield as
per the dimensions below.
Strip off the black conductive layer convering two
2
wires completely, as per the dimensions below.
Twist each of the conductor and drain wires so that
there are no free strands.
Do not short-circuit the conductive layer and the
3
terminals (A, B, C and T
Strip off about 5 mm (0.2 in.) of insulation for each
4
of wires A, B, and T*
1
*
).
1
, and twist the strands of each
wire. Twist the inner and outer drain wires together.
Slide FEP (fluorinated ethylene propylene) tubing
5
over the twisted inner and outer drain wires C until
the tubing cannot be slid any further, and then cut
off the tubing leaving 5 mm (0.2 in.) of the stranded
drain wires exposed.
Slide heat shrinkable tubing over the cable end so
6
that the tubing covers the braided shield and
overlaps both the polyethylene jacket and loose
wires A, B, C, and T*
Slide a short piece of heat shrinkable tubing over
7
each of wires A, B, C, and T*
1
.
1
. Install a crimp-on
terminal lug at the tip of each wire. Crimp and
solder each lug.
Slide each short piece of heat shrinkable tubing
8
over the crimp sleeve. Heat all pieces of heat
shrinkable tubing with a heat blower or dryer.
Attach an identification label to the end of the cable.
9
T*
B (White)
5 (0.2)
C (Black)
Lug tip
1
(Yellow)
A (Red)
50 (2.0)
60 (2.4)
Drain wires
C (Black)
A (Red)
5 (0.2)
C(Black)
Crimp and Solder Here
10
Heat Shrinkable Tubing
10 (0.4)
5 (0.2)
5 (0.2)
90 (3.5)
Conductive Layer (Black)
T*1 (Yellow)
40 (1.6)
5 (0.2)
3 (0.1)
or less
Conductive Layer (Black)
B (White)
1
(Yellow)
T*
5 (0.2)
5 (0.2)
FEP Insulation Tubing
(Black)
T*1 (Yellow)
A (Red)
B (White)
Heat Shrinkable Tubing
T*1 (Yellow)
A (Red)
B (White)
Heat Shrinkable Tubing
(approx. inches)
4-6
Unit : mm
NOTE
Check that the insulation resistance between each
wire including the inner shield is 10M or greater at
500V DC. Ensure that both ends of the wires are
disconnected (open-circuited) during the check.
(*1): Only for /MV
NOTE
In case that the cable end fi nish
)
Black
White
Yellow
80 (3.15
(C)
(B)
(A)Red
*1
(T)
60 (2.36
70 (2.76
50 (1.97
)
20 (0.79
)
)
)
Unit : mm
(approx. inches)
F0408.ai
Figure 4.7 End Processing Method of DYC Signal Cable for Detector
parts assembly is necessary
after delivery, contact your
nearest Yokogawa sales offi ce
or the sales representative
from which you purchased the
product.
F0407.ai
CAUTION
Do not touch the '' conductive layer"
(black area covering the signal cables A
and B) to the converter case, terminal,
and other leadwires. If it is touched,
operation of the converter may be
incorrect. When the cable is terminated,
remove the conductive layer properly.
IM 01F06A00-01EN
<4. WIRING>
(A)Red
(B)White
Black (C)
(G)Blue
Unit : mm
(approx. inches)
F0410.ai
70 (2.76)
95 (3.74)
60 (2.36)
80 (3.15)
25 (0.98)
MAX
3
Yellow*1(T) 50 (1.97)
4.5.2 For Vortex Flow Converter (DYA)
DescriptionFigure
Strip off the outer polyethylene jacket, outer braided
1
shield and inner jacket, and inner braided shield as
per the dimensions as shown.
Cut of the black conductive layers(convering the
2
two wires) completely, as per the dimensions below.
Twist each of the conductor and drain wires so that
there are no free strands.
Do not short-circuit the conductive layer and the
3
terminals (A, B, C, G, and T*
Strip off about 5 mm (0.2 in.) of insulation for each
4
of wires A, B, and T*
1
).
1
, and twist the strands of each
wire.
Slide black FEP (fluorinated ethylene propylene)
5
tubing over the inner shield drain wire C and blue
FEP tubing over outer shield drain wire G until
the tubing cannot be slid any further, and then cut
off the tubing leaving 5 mm (0.2 in.) of the drain
wires exposed.
Slide heat shrinkable tubing over the cable end so
6
that the tubing covers the braided shield and
overlaps both the polyethylene jacket and loose
wires A, B, C, G, and T*
Slide a short piece of heat shrinkable tubing over
7
each of wires A, B, C, G, and T*
1
.
1
. Install a crimp-on
terminal lug at the tip of each wire. Crimp and
solder each lug.
Slide each short piece of heat
8
shrinkable tubing over the crimp sleeve. Heat all
pieces of heat shrinkable tubing with a heat blower
or dryer.
Attach an identification label to the end of the cable.
9
B (White)
A (Red)
*1
(Yellow)
T
C (Black)
C (Black)
Lug-Tips
Conductive
Layer (Black)
G
C
G
G
B (White)
(0.4)
15 (0.6) 10 (0.4)
95
5 (0.2)
(3.7)
*1
T
3 (0.1) or less
40 (1.6)
50 (2.0)
60 (2.4)
5 (0.2)
5 (0.2)
5 (0.2)
5 (0.2)
A
10
Drain wires
5 (0.2)
FEP Insulation Tubing (Black)
T
(
)
Red
Crimp and Solder
Heat Shrinkable Tubing
(approx. inches)
(Yellow)
Conductive
Layer (Black)
5 (0.2)
T*1 (Yellow)
A (Red)
B (White)
FEP Insulation Tubing (Blue)
T*1 (Yellow)
A (Red)
B (White)
25 (1.0)15 (0.6)
Heat Shrinkable Tubing
Heat-shrinkable tubing
4-7
Unit : mm
NOTE
Check that the insulation resistance between each
wire including the inner shield is 10M or greater at
500V DC. Ensure that both ends of the wires are
disconnected (open-circuited) during the check.
(*1): Only for /MV
NOTE
In case that the cable end
fi nish parts assembly is
necessary after delivery,
contact your nearest
Yokogawa sales offi ce or the
sales representative from
which you purchased the
product.
Figure 4.8 End Processing Method of DYC Signal Cable for Converter
F0409.ai
CAUTION
Do not touch the '' conductive layer"
(black area covering the signal cables
A and B) to the converter case,
terminal, and other leadwires. If it is
touched, operation of the converter
may be incorrect. When the cable is
terminated, remove the conductive
layer properly.
IM 01F06A00-01EN
<4. WIRING>
F0412.ai
Terminal Box
Steel Conduit for
Flameproof
Converter
Te e
Drain Fitting
Flexible Metal Conduit
4.6 Wiring Procedures and
Precautions
NOTE
Once all wiring is complete, check the
connections before applying power to the
instrument. Improper arrangements or wiring
may cause a unit malfunction or damage.
(1) Lay wiring as far as possible from electrical
noise sources such as large capacity
transformers, motors, and power supplies.
(2) Remove the terminal cover and dustproof plug
of an electrical connection before wiring. When
you open the cover of explosion proof type(*),
turn the Locking Screw to the right, and unlock.
When you close a cover after wiring, be sure to
turn the Locking Screw to the left and lock.
and a duct for waterproofi ng or external
protection of an electric wire. Refer to Figure
4.9 and Figure 4.10.
(4) The fl ameproof packing adapter (option
code: /G11 or /G12) should be used for the
external wiring of TIIS Flameproof. Refer
to “INSTALLATION AND OPERATING
PRECAUTIONS FOR TIIS FLAMEPROOF
EQUIPMENT.”
Terminal Box
Figure 4.10 Example of Wiring (Remote Type
Converter)
B. Coupling
Clamp Nut
Clamp Ring
Packing Case
Adapter Body
Apply a nonhardnening
sealant to the threads
for watweproofing
Packing Gland
O-Ring
Cable
Lock Nut
Wrench
Union Nut
Figure 4.11 Cable Wiring
4-8
Washer
Packing
Wrench
Lock Nut
F0413.ai
Steel Conduit
Flameproof
Converter
Te e
Drain Fitting
Figure 4.9 Example of Wiring (Integral Type and
Remote Type Detector)
Flexible Metal Conduit
F0411.ai
NOTE
Be sure to use the fl ameproof packing adapter
(option code:/G11 or /G12) for TIIS fl ameproof
type at the time of cable wiring work. Refer to
Table 4.2.
Table 4.2 Flameproof packing adaptor
Option Code
G11
G12
Diameter for
screw
G11
G12
Cable outer
diameter
mm (inch)
ø8.0 to ø10.0
(ø0.31 to ø0.39)
ø10.0 to ø12.0
(ø0.39 to ø0.47)
Identifi cation
mark
16 8-10
16 10-12
Parts NO.
G9601AM
IM 01F06A00-01EN
<4. WIRING>
4-9
(5) Perform attachment of fl ameproof packing
adaptor in the following ways. Refer to Figure
4.11.
(a) Loosen the locking screw and remove the
terminal box cover.
(b) Measure the cable outer diameter in two
directions to within 0.1 mm.
(c) Calculate the average of the two diameters,
and use packing with an internal diameter
nearest to this value. Refer to Table 4.2.
(d) Screw the fl ameproof packing adapter into
the terminal box until the O-Ring touches
the wiring port (at least 6 full turns), and
fi rmly tighten the lock nut.
(e) Insert the cable through the union nut, the
B. coupling, the clamp nut, the clamp ring,
the packing gland, the washer, the packing,
and the packing case, in that order.
(f) Insert the end of the cable into the terminal
box.
(g) Tighten the union cover to grip the cable.
When tightening the union cover, tighten
approximately one turn past the point where
the cable will no longer move up and down.
Proper tightening is important. If it is too
tight, a circuit break in the cable may
occur; if not tight enough, the fl ameproof
effectiveness will be compromised.
(h) Fasten the cable by tightening the clamp
nut.
(i) Tighten the lock nut on the union nut.
(j) Connect the cable wires to each terminal.
(6)
(a) Do not connect cables outdoors in wet
weather in order to prevent damage from
condensation and to protect the insulation.
(b) Do not splice the cable between the
fl owtube terminal and the converter if it is
too short. Replace the short cable with a
cable that is the appropriate length.
(c) The signal cables must be routed in
separate steel conduit tubes 16 (JIS C
8305) or fl exible conduit tubes 15 (JIS C
8309).
(d) Always route the power and output signal
cables in separate steel conduit tubes,
except when the power supply voltage is 24
V and four-core cables are used for wiring.
Keep conduits or fl exible tubes watertight
using sealing tape.
(7) For the TIIS fl ameproof type with wiring using
a fl ameproof packing adapter, wire cables
through the packing adapters approved by
Yokogawa (option code: /G11 or /G12).
T1
16.5
18
L
F
G
*Packing
(Choose from the table
below depend on cable
outside diameter)
T1T2CDLFG
G 1/2 G 1/2
(1.38)
Size
T2
Cable outer
diameter
ø8.0 to ø10.0
35
39
(1.54)
94.5
(3.72)
(ø0.31 to ø0.39)
ø10.0 to ø12.0
(ø0.39 to ø0.47)
Figure 4.12 Flameproof Packing Adapter (option
code: /G11, /G12)
Adapter body (M. Screw)
O-Ring
Packing case
Hexagon socket set screw
Packing *
Hexagon socket set screw
O-Ring
O-Ring
Washer
Union nut
Packing gland
Clamp ring
Clamp nut
O-Ring
B.coupling
Cable (user’s scope)
Packing
dimensions
ø10.0
(ø0.39)
(ø0.79)
ø12.0
(ø0.47)
ø20.0
C
Identifi cation
mark
16 8-10
16 10-12
Unit : mm
(approx. inch)
D
F0414.ai
Weight
kg (lb)
0.26
(0.57)
4.7 Grounding
IMPORTANT
When a lightning protector (option code: /A) is
selected, use a grounding resistance of 10 or
less.
(1) The grounding terminals
inside and outside of the terminal area. Either
terminal may be used.
(2) For pulse output version, ground the fl owmeter.
Also ground the shielded cable between the
converter and the pulse receiver.
(3) Grounding should satisfy Class D requirements
(ground resistance 100 or less).
(4) Use 600V PVC insulated wire for grounding.
Example: Integral Type
Figure 4.13 Grounding Terminal
are located on the
Grounding
Terminals
F0415.ai
IM 01F06A00-01EN
<5. BASIC OPERATING PROCEDURES>
5. BASIC OPERATING PROCEDURES
Data setting can be performed with the three keys
on the front panel (SET,SHIFT and INC) or using a
handheld BRAIN TERMINAL (BT200) and HART
communicator.
5.1 Display Confi guration
Figure 5.1 shows the confi guration of the
digitalYEWFLO display panel (if equipped).
Data Display
1
(Upper)
Unit Display
4
Alarm Display
3
Data Display
2
(Lower)
4
Unit Display
INCSHIFT
SET
5-1
Setting Keys
5
Figure 5.1 Display Confi guration
1
Data Display(Upper) : fl owrate data, setting data,
total data
temperature data (/MV)
2
Data Display(Lower) : total data, alarm data
temperature data (/MV)
3
Alarm Display : alarm of a fl ow error and a
vibration error
4
Unit Display : fl owrate unit
5
Setting Keys : These keys are used to
change fl ow rate data
displays and type of
setting data
F0501.ai
IM 01F06A00-01EN
<5. BASIC OPERATING PROCEDURES>
5.2 Display Contents
The display content items are classifi ed in the following three items.
Table 5.1 Mode Name List
Mode (status) NameDisplay Contents
Flow rate display modeA mode in which instantaneous fl ow rates or totalized values are displayed.
Display content is usually selected either in display content selection mode or by setting
parameters via BRAIN communication.
Setting modeIn this mode, parameter contents are confi rmed or data is updated using the setting section. The
mode is changed to this mode when “SET” key is pressed in normal mode.
Alarm number display modeThis mode is overlapped when an alarm is occurring in display mode. The alarm number
presentation to indicate alarm contents (about 2 sec) and the normal data display (about 4 sec ) are
repeated alternatively.
Mode represents that the system is in a state where the relevant setting or display is possible.
● Display Example
SET
SHIFT + SET
Flowrate Display ModeSetting ModeError Mode
5-2
UPPER
Flow rate
UPPER
Flow rate (%)
This mode display can be selected
below.
Upper display : Flow rate
Lower display : Total rate or Blank
LOWER
Total rate
LOWER
Blank
Switching
of setting
number
SHIFT
This mode is used to check parameter
content and rewrite data. This mode can
be called up from the flowrate display
mode by pressing the “SET” key.
Setting item and setting number can be
changed when pressing “SHFT” key.
This mode can be called up by pressing
“SET” key while pressing “SHIFT” key
when setting mode is displayed.
NORMAL
INDICATION
(4 sec)
ERROR INDICATION
(2 sec)
When an alarm situation occurs,
this mode will replace the current
mode (flow rate or setting mode)
to show what type of alarm has
occurred.
Refer to Section 6.5 “Self-
Diagnostic (Error Code List)”
about the error descriptions and
error number.
Alternately
F0502.ai
IM 01F06A00-01EN
Loading...
+ 143 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.