Yokogawa CW500 Specifications

A Clear Path to Electrical Power Excellence!
CW500
Power Quality Analyzer
Bulletin CW500-01EN
Electrical power measurement is a key for innovation.
The Yokogawa CW500 is a portable power meter that utilizes a current clamp sensor for use in the eld enabling consumption and power quality measurement of the power line.
By the navigation screen of the unit, the connection setting and detail setting of the unit is simple and direct by following the screens.
Keys which directly turn on desired measuring screens allow comfortance at the use of eld.
Stored measured data are easily accessed by a click on the le with included PC software.
Simplied power measuring – Several direct
keys keep you simple and direct access to Voltage/ Current/Power/Power factor/ Phase Angle/Frequency/ Integrated power/Demand by values or by a trend graph.
Firmly capture line power quality issue –
Captures short period power malfunction conforming IEC 6100-4-30 Class S.
User support – Function of
Quick Start Guide to support secure wiring and setting of units.
Features
Power Measure and Logging
Power Quality Measuring
User Support
Analysis of Data and Report Generation
Power Measuring and Logging
Simultaneously measures 3 CH Voltage input, 4 CH current
clamp-on probe input, 2 CH DCV input.
Displays value list or trend graph screen of Instantaneous/
Average/Maximum/Minimum of Voltage/Current/Power/Power factor/Phase Angle/Phase Advanced Capacitance Calculation and DCV input.
Integration Value of Active/
Reactive/Apparent Energy is each displayed by consumption and generation.
Demand value can be
monitored by screens of present power consumption compared to aimed demand power value.
User Support
Quick Start Guide Function:
Start Guide Function supports secure wiring and setting
before measuring.
Automatically recognize the type of current clamp-on probe.
Vector Display:
Indicates Voltage and Current phase difference and values
between input channels of voltage and current.
Checks whether the wiring is appropriate or not.
Power Quality Measuring
Analysis of Data and Report Generation
Measure Temporary Malfunction of Power Line
Captures temporary malfunction phenomena of power
line which causes malfunction or destruction of devices by types (Voltage swell, Voltage dip, Voltage interruption, Transient overvoltage, inrush current) as an event by high sampling rate of 24 µs and RMS calculation.
Event data contains the type of malfunction, occurred
time or occurrence nish time, measured value and waveform of voltage and current of all channels for approx. 200 ms period.
Measurement method conforms to IEC standard 61000-
4-30 Class S
Measure Continuous Malfunction of Power Line
Harmonics
Measure and display graphs and list of up to the 50
Harmonic components of voltage, current and power for each phase and in total.
Waveform
Displays with up to 10 or 12 waveforms of voltage and
current for each CH.
Flicker
Measures, 1 minute icker (Pst, 1 min),
short icker (Pst) and long icker (Plt).
Unbalance rate
Displays voltage and current unbalance
rate on 3 phase wiring.
th
CW500Viewer (Included PC software)
Automatically generates graph and report by simple
clicking on a le data displayed on screen.
Uniform management of main unit settings
Realtime measuring by USB communication.
Types of data
Power data, Power quality event data, Main Unit Setting
data, Screen capture data.
Memory card and interface
SD Memory card, USB communication, Bluetooth
communication (Available for USA, Canada and Japan only)
Function
CW500Function
Power Measuring
Power line and Input Channel: 3 CH Voltage, 4 CH for Current Clamp Probe.
1P2W (up to 4 system), 1P3W (up to 2 system), 3P3W2current (up to 2 system), 3P3W3current, 3P4W
Power measuring item:
Instantaneous, Average, Maximum and Minimum values:
Voltage/Current/Power (Active, Reactive, Apparent)/Power factor/ Phase Angle/Frequency/Calculated Phase Shift Condensor/DC voltage value 2 CH.
Integration Value
Energy (Active, Reactive, Apparent) each by
consumption and generation
Demand (Occurrence of max. demand time, current
demand, estimated demand value)
Recording Interval Period:
1/2/5/10/15/20/30 second, 1/2/5/10/15/20/30 minute, 1 hour/2 hours
Recording method
Manual, time setting, period setting
List Display
Hems of measuring values can be selected with position.
Zoom Display
Zoom display can be selected by 4 or 8 division.
Trend Graph
Trend graph of voltage/ current/power/power factor/frequency/ advanced phase condenser/DC voltage
4
Estimated Recording Length with 2 GB SD card.
Interval Power recording +Harmonics
1 sec. 13 days 3 days 1 min. Over 1 year 3 months
30 min. Over 10 years Over 7 years
The recording length shortens according to the number
of power quality events.
Only included SD memory card or dedicated SD
memory card is guaranteed.
Various Measuring Screen
List display, Zoom display by 4 or 8 division,
Trend graph display
Integration Value Display for Energy
Demand List display of value,
Demand Graph display for change of period, Demand Graph display for whole recording
Energy Display
List of Active, Apparent, Reactive by consumption or generation
5
Analysis and Report of recorded data
Demand measuring
Demand is the average power value of a specic period (usually 30 minutes). Contract with the power supplier conditionally concerns the maximum average power value between period for the consumption fee. This function supports how to maintain within the target consumption by monitoring the estimated demand value to the setting rate, with the maximum demand value.
List of Demand Value
Graph display of change on specic period
When the estimated demand value line is displayed inside this area, it mean that the demand value will be over the target value.
Analysis and report is enabled by simply clicking on the desired data on software screen. Analysis is capable by trend graph, average, maximum and minimum value of voltage, current, power, power factor. Report generating could be set by day time, night time, working day, off day or monthly period. Additionally energy data to quantity of crude oil barrels, or to CO
can be converted.
2
Trend analysis graph
Sample of report
Estimated demand value
Graph display of whole demand trend
Max. demand value
Aimed target demand value
CW500Function
Power Quality Function
Short period power line malfunctions such as voltage swells/dips/interruptions/transient overvoltage or inrush currents, or long period such as harmonic distortion, icker may damage or reset your devices. The CW500 helps to identify each of those short period malfunctions, by recording occurrence time, occurrence nish time and waveforms. Additionally there is a digital output on occurrence. Long period malfunctions can be analyzed by harmonics, icker, waveform or checking on the unbalance rate calculation for 3 phase measurement. All data can be nalized to a report format with included software. The CW500 conforms to the IEC standard of 61000-4-30 Class S.
Power line malfucntion phenemona
Voltage Swell
Voltage Dip
Voltage interruption
Example of cause and inuence Example of waveform Notes
Lightnings or heavy load switching on power line may cause momentary swell on voltage.
Generating moments on motor load may cause an inrush current and cause dips on voltage.
Occurred time and
Momentary or long interruption to power distribution by lightning or open breakers
waveform is recorded. Digital output is set on during the occurred period.
6
Transient overvoltage (impulse)
Inrush current
Flicker
Harmonics
Unbalance rate
Lightning or heavy load switching on power line may cause momentary change on voltage.
Generating moments on motor load may cause an inrush current.
Increase and decrease on certain phases could cause ickering distortion on voltage and currents.
Inverter and Thyristor circuits (phase control cirtuits) which are used for the control circuit of general devices could affect currents and cause harmonic distortion.
Heavy loading on a specic phase, could inuence motor operation and could cause harmonic distortion.
Pst (1 min), Pst or Plt is measured.
Up to 50th harmonic contents are measured.
Voltage and Current unbalance rate measured on vector screen for 3 phase
7
Power Quality Measuring Function
Capture temporary power line issues
Various kind of power issues are captured by high sampling rate of 24 µs and overlapping RMS measuring by half cycles. They are recorded as an event by occurred time, occurrence nish time and waveforms are recorded.
Main unit displays list of occurred issue by types.
Recorded data are easily analyzed and report is
generated by PC software.
Voltage swell
Voltage dip
Voltage Interrupt
Transient overvoltage
Inrush Current
Power line malfunction event
Display event, timing and measurement
Capture continuous power line issues
Harmonic Measuring
Displays components of up to the 50
contents by individual channel or overall by graph or list
Displays the maximum occurred point on graph.
Graph display List display
Flicker
Displays list or graph of 1minute icker (Pst, 1 min), short term icker (Pst) or long term icker (Plt).
th
Harmonics
Waveforms and measured data are stored on a SD memory card.
Main operation
Analyze and report is created by PC
List display
Waveform
Displays voltage and current waveform by all or individually. It can be zoomed in and out by vertical or horizontal with rate selection below.
Vertical rate: 0.1, 0.5, 1, 2, 5, 10 Horizontal rate: 1, 2, 5, 10
CW500Function
User Support Function
User Friendly
Quick start guide and Automatic type of clamp identication
A quick start guide will show how to wire and how to set the range before measuring which ensures the settings are correct. The sensor identication will detect the type of clamp-on probe and set the highest range of the type.
Start of guidance
PUSH “START/STOP”key and select “Quick start guide”. Recording item selection will be displayed.
Wiring
Select the wiring and the appropriate connection diagram will be displayed. Connect the voltage probe and clamp-on probe accordingly.
Wiring check/self diag./self id.
Wiring check, self check and type of clamp-on probe identication will be executed and the results will be displayed. If NG is displayed, detail could be conrmed by ENTER key.
Recording setting
Set recording interval time and recording time or period.
Power supply from measuring line
Power (under 240 VAC) can be supplied by using the “Power supply adapter” (sold separately).
Example diagram
of power supply
Note: This is not available for EU region.
Power line
Leakage current measuring
External magnetic eld effect is 0.002 A or less, at 400 A/m. Yokogawa’s proprietary technology has achieved a magnetic eld impact amount of 30 ppm even in adjacent power lines (at 100 A), Below is an example to measure neutral line of 3 phase 4 line.
1R
2S
3T
N
Source
Clamp-on probe
96060
N U1 U2 U3 CH1 CH2 CH3 CH4
3 Phase 4 wireCW500 side
Multiple line measuring
4 system load measuring
Maximum of 4 systems is capable for 1 phase 2 line Maximum of 2 systems is capable for 1 phase 3 line or 3 phase 3 line.
8
Load
9
USB communication
Bluetooth communication*
SD Memory Card
Software for Analysis and Setting (Free) CW500Viewer
Data analysis , making reports of data, making setting le and doing a real time measuring can be easily done with the CW500Viewer. The data can be transferred by SD memory card, USB communication or Bluetooth communication (for USA, Canada, Japan only).
Graph and lists are created by a click on the data le.
Graph and lists of the recorded data le are created by
clicking on the desired data le. They can be displayed and copied on a clip board so it can be used on other software such as Word or Excel spread sheet.
The time axis and measure axis can be zoomed in or out
easily. With this capability, micro and macro changes, can be displayed desirably.
Waveforms of power quality events (Voltage swell,
voltage dips, voltage interrupt, transient overvoltage and inrush current) which are recorded by 200 ms period can be displayed and printed.
The integrated data can be added on one graph
which enables the whole energy integration data to be displayed.
Integrated energy data can be scaled into CO
oil value.
or crude
2
Settings management
Setting data can be read out from and to the main unit
via SD memory card, USB communication cable or Bluetooth communication*.
Settings data can be easily edited saved and managed.
Real time measurement
Real time measurement can be achieved up to 2 units
simultaneously via USB communication or Bluetooth communication*
System requirement
OS Windows 8/7/Vista (32 bit/64 bit)
CPU Pentium 4 or Pentium Processor over 2 GHz
Display 1024 × 768 dots, 65536 colors or more
HDD (Hard-disk space required)
Others .NET Framework (3.5)
Note: Windows is a registered trademark of Microsoft in the United States.
Pentium is a registered trademark of Intel in the United States. Bluetooth is a registered trademark of Bluetooth SIG. Other company names or names of merchandise are trademarks of their company.
1 GByte or more (including Framework)
Example of multiple graph display
Example of setting display
*Only available for US, Canada and Japan.
Application
CW500Application
Power Measuring + Power Quality Examination
Examination of main power line quality for factory system control and assets
Conrm to check quality of power line for factory system
control and assets.
Simultaneously measure the consumption of energy
trend and consider solutions for energy saving.
CW500 not only continuously measures harmonics but captures malfunction of power line and records occurrance time with waveforms. Simultaneously it can measure power and leakage current of neutral phase. Additionally users can conrm the condition of assets when signals are connected to the DCV input channels.
Power Line Measuring Examination
LED lighting introduction for checking power saving effect
Introduce the effect of power saving to use LED lightning compared to the non LED lightning by measuring before and after.
10
Asset operation check
Checking power consumption of driving pumps of a manufacturing line
Inverters were applied to many driving pumps in a liquid manufacturing line. Power consumption checking was needed to check the effect before and after. Many settings to measure many pumps is easily and efciently arranged by PC software. Additionally the power line quality is checked to be safely operated with the new inverters.
Voltage/Current/ Power/Power factor/ Demand/ Analog input 2 CHLeakage current
11
[Before]
[After]
and malfunction rate of printing machine decreased.
Improving power line of a printing factory by measuring harmonics
(Printing factory)
Purpose: Investigation of the cause of periodic
malfunction for printing machine.
Harmonics distortion on line could be the reason?
Measuring: CW500 merit
Compact size and easy to carry.
Up to 50
Long term recording
Result: Conrming high level harmonic contents
on 5
Found harmonics were generated by internal
assets.
Especially 5th harmonics damages the direct reactor
of condensor for improving power factor
Found cause
Harmonics on current input Harmonic contents on current input
th
Harmonics measuring.
th
and 7th.
Others
Improvement of Power factor for power efciency
CW500 can calculate the appropriate condensor value by setting the aimed power factor value. By setting the appropriate advanced condensor and improving the power factor, users will benet for power saving with less load current and improve the capacity of the whole power system.
Harmonics on current input Harmonic contents on current input
Effect of correspondence: Harmonic contents of 5th and after dropped
Conrmation
of inow
harmonics
500kV transformers
of factory “A”.
External Appearance
17
CW500External Appearance
12
18
1
2
3
4
5
9 10 11
6
7
8
19
13
14
15
16
12
1
START/STOP SWITCH
2
Screen switch of W and Wh item
3
Vector and wiring check switch
4
Power button
5
Waveform screen switch
6
Harmonic screen switch
7
Power quality screen switch
8
Setting screen switch
9
Print screen switch
10
Data Hold/Key lock
(long press)
11
LCD Key/Contrast
(long press)
Turn ON/OFF the display and change brightness and contrast
by long press
12
Cursor and Enter Key
13
USB terminal
Mini B pin for PC communication
14
Digital output terminal
Trigger signal output on event
15
Analog input terminal 2 CH
For 100 mV, 1 V, 10 V, DC input
16
SD card interface
17
Voltage input terminal
18
AC power supply input
19
Clamp-on probe connection
Accessories
Accessories
13
Clamp-on Probe (sold separately)
For Leakage For Load current For load current (exible type)
Model code
Photo
96060 96061 96062 96063 96064 96065 96066
Measurable
diameter
Input Range
Output
voltage
50 Hz/
60 Hz
40 Hz to
1 kHz
Level
1 kHz to
3.5 kHz
Accuracy
Accuracy
Degree
Max Circuit
voltage
Dimension
Weight
40 mm 18 mm 24 mm 30 mm 40 mm 110 mm 150 mm
AC 2 A AC 50 A AC 100 A AC 200 A AC 500 A AC 1000 A
AC 50 mV (25 mV/A)
±1.0% rdg
±0.05 mV
±2.0% rdg
±0.1 mV
±3.0% rdg
±0.2 mV
AC 300 Vrms AC 300 Vrms AC 300 Vrms AC 600 Vrms AC 600 Vrms AC 600 Vrms AC 600 Vrms
approx.
70(W) × 120(H) ×
25(D) mm
approx. 230 g approx. 170 g approx. 160 g approx. 240 g approx. 260 g approx. 170 g approx. 950 g
AC 500 mV AC 500 mV AC 500 mV AC 500 mV AC 500 mV
±0.5% rdg
±0.1 mV
±0.8% rdg
±0.2 mV
±1.0% rdg
±0.4 mV
Less than ±2.0°
(0.5 to 50 A,
40 Hz to 3.5 kHz)
approx.
52(W) × 106(H) ×
25(D) mm
(excluding pointed part)
±0.5% rdg
±0.1 mV
±1.0% rdg
±0.2 mV
Less than ±2.0°
(1 to 100 A,
45 Hz to 65 Hz)
approx.
60(W) × 100(H) ×
26(D) mm
±0.5% rdg
±0.1 mV
±0.8% rdg
±0.2 mV
±1.0% rdg
±0.4 mV
Less than ±1.0°
(2 to 200 A,
40Hz to 3.5kHz)
approx.
73(W) × 130(H) ×
30(D) mm
±0.5% rdg
±0.1 mV
±1.0% rdg
±0.2 mV
Less than ±1.0°
(5 to 500 A,
45 Hz to 65 Hz)
approx.
81(W) × 128(H) ×
36(D) mm
±0.8% rdg
±0.2 mV
±1.5% rdg
±0.4 mV
Less than ±2.0° (45 Hz to 65 Hz) Less than ±3.0° (40 Hz to 1 kHz)
73(W) × 130(H) ×
30(D) mm
*45 to 65 Hz (measuring at the center of sensor)
approx.
AC 300 A AC 1000 A AC 3000 A
For each range
AC 500 mV
±1.0% rdg*
Less than ±1.0° (for each range/
45 to 65 Hz)
approx.
61(W) × 111(H) ×
43(D)mm
Other accessory (sold separetely)
Extension code
*1 Extension code 98082 corresponds to below current probe. 96060, 96061, 96062 *2 Non-CE product. Not available for CE marking necessary region. *3 The following clamp-on probe are able to be connected. 96030, 96033, 96036
98082
*1
Power supply adapter
98031
*2
Portable case (for CW500)
93047
Conversion Cable (Banana-DIN)
99073
*2, 3
Specications
CW500Specications
14
Model code CW500-B0 CW500-B1
Bluetooth function No Bluetooth Function With Bluetooth
Wiring connections 1P2W (max. 4 systems), 1P3W (max. 2 systems), 3P3W (max. 2 systems), 3P3W3current, 3P4W
Measurements items Voltage, Current, Frequency, Active power, Reactive power, Apparent power, Active energy, Reactive energy,
Other function Digital output, Analog DCV input function
Voltage (RMS)
Current (RMS)
Active power
Frequency meter range 40 to 70 Hz
Power supply (AC Line) AC100 to 240 V/50 to 60 Hz/7 VA max.
Power supply (DC Battery) Alkaline size AA battery LR6 or Ni-Mh (HR15-51) × 6 pcs
Internal memory Flash memory (4 MB)
External memory card SD Card (2 GB)
PC communication USB Ver. 2.0 USB Ver. 2.0/Bluetooth Ver. 2.1 + EDR Class2
Display 320 × 240 (RGB) Pixel, 3.5 inch color TFT
Display update period 1 s
Display Language English, French, Spanish, Polish, Korean, Chinese, Japanese
Temperature and humidity range 23±5°C, less than 85% RH (without condensation)
Operating temperature and himidity range 0 to 45°C, less than 85% RH (without condensation)
Storage temperature and humidity range −20 to 60°C, less than 85% RH (without condensation)
Dimensions 120 (W) × 175 (H) × 68 (D) mm
Weight Approx. 900 g (with battery)
Included accessories (attached) 98078 Voltage Probe,
Optional accessories (sold separately) 96060, 96061, 96062, 96063, 96064 (Clamp-on Probe)
*1 Bluetooth model is available only for USA, Canada and Japan *2 98031 Power supply adapter is not available for EU region.
Range 600.0/1000 V
Accuracy ±0.2% rdg ±0.2% rng. (sine wave, 40 to 70 Hz)
Allowable input 1 to 120% (rms) of each range, 200% for peak of each range
Display range 0.15 to 130% of each range
Crest factor 3 or less
Sampling speed of voltage transient 24 µs
Range 96060 (2 A type): 2000 mA
Accuracy ±0.2% rdg ±0.2% rng. + accuracy of clamp-on probe (sine wave, 40 to 70 Hz)
Allowable input 1 to 110% (rms) of each range, 200% for peak of each range
Display range 0.15 to 130% of each range
Crest factor 3 or less
Accuracy ±0.3% rdg ±0.2% rng. + accuracy of clamp-on probe (Power factor 1, sine wave, 40 to 70 Hz)
Effect of Power Factor ±1.0% rdg (40 to 70 Hz, reading at power factor 0.5 against 1.0)
Apparent energy, Power factor, Phase Advancing Condensor, Neutral current, Demand, Harmonics, Power Quality (Swell/Dip/Interrupt/Transient overvoltage, Inrush current, Unbalance rate. IEC icker
96061 (50 A type): 5000 mA/50 A/AUTO 96062 (100 A type): 10/100 A/AUTO 96063 (200 A type): 20/200 A/AUTO 96064 (500 A type): 50/500 A/AUTO 96065 (1000 A type): 100/1000 A/AUTO 96066 (3000 A type): 300/1000/3000 A
Battery life approx. 3 hours (LR6 Backlight OFF)
93046 Carrying case 97060 SD Memory Card 2 GB USB cable, Power cord, Quick manual, Alkaline size AA battery LR6 × 6pcs, Input terminal plate × 6 pcs, PC software (CD-ROM)
96065, 96066 (Clamp-on probe, exible type) 98031 (Power supply adapter) 93047 (Portable case with magnet)
*2
Function
*1
*1
15
Applicable Standard
Safety Standard EN 61010-1 CAT IV 300 V, CAT III 600 V, CAT II 1000 V Pollution level 2
*
EMC
Power Quality IEC 61000-4-30 Ed. 2 Class S, IEC 61000-4-15, IEC 61000-4-7
Wireless FCC approval, IC approval, Radio Electric technology engineering Radio technology standard
* This is a Class A instrument designed for an industrial environment. Operation of this equipment in a residential area may cause radio interference, in which case users will be
responsible for any interference which they cause.
EN 61010-2-030 EN 61010-2-033 EN 61010-031
EN 61326-1 Class A Table 2 EN 55011 Class A Group1
Power list of choosing clamp-on probe
For 1P2W (multiply 2 for 1P3W and 3P3W, multiply 3 for 3P4W)
Clamp-on Probe Model code (rate)
Voltage
range
600.0 V 3000 W 30.00 kW 6000 W 60.00 kW 12.00 kW 120.0 kW 30.00 kW 300.0 kW 60.00 kW 600.0 kW 180.0 kW 1800 kW
1000 V 5000 W 50.00 kW 10.00 kW 100.0 kW 20.00 kW 200.0 kW 50.00 kW 500.0 kW 100.0 kW 1000 kW 300.0 kW 3000 kW
96060 is dedicated for leakage only and is uncapable for power measuring
96061 (50 A) 96062 (100 A) 96063 (200 A) 96064 (500 A) 96065 (1000 A) 96066 (3000 A)
5000 mA 50.00 A 10.00 A 100.0 A 20.00 A 200.0 A 50.00 A 500.0 A 100.0 A 1000 A 300.0 A 3000 A
Current range
2
Model and sufx
NOTICE
Model code Sufx code Notes
CW500 Power Quality Analyzer
*Available for USA, Canada and Japan only
-B0 No Bluetooth Function
-B1 With Bluetooth Function*
-D AC code (UL/CSA)
-F AC code (VDE)
-H AC code (GB)
-N AC code (NBR)
-P AC code (KC)
-R AC code (SAA)
-S AC code (BS)
Accessories (included with CW500)
Model code Model name Notes
98078 Voltage Probe
93046 Carrying Case
97060
SD Memory Card (2 GB)
1 set 4 pcs Red Black White Blue
4 mm Approx. 3 m
CW500 and Clamp-on probe can be contained
2 GB SD Memory Card
Relevant products
CW120 series Clamp on Power Meter
CW120/CW121
Dim. 117 (W) × 161 (H) × 51 (D) mm
Approximately 0.6 kg
WT300E series Power Meter (Direct current input type)
WT332E/WT333E
Dim. 213 (W) × 132 (H) × 350 (D) mm
Approximately 5 kg
WT310E/WT310EH
Dim. 213 (W) × 88 (H) × 350 (D) mm
Approximately 3 kg
Simple and low cost model
3P3W model 2 CH input (CW120), 3P4W model
3CH input (CW121)
Frequency range: 45 Hz to 65 Hz, Basic power
accuracy: ±(0.8% rdg + 0.4% rng)
Multiple system (CW120: 3 system 1P2W,
CW121: 4 system for 1P2W)
Recording from 1 sec.
Measuring voltage up to 400 V
Communication protocol: Modbus, PC link
Compact three-phase model with optional harmonic measurement function
Three-phase model (three-phase, three-wire: two
input elements; three-phase, four-wire: three input elements)
Power measurement frequency range:
DC and 0.1 Hz to 100 kHz
Basic power accuracy: 0.1% of reading.
Max. 3 channels simultaneous Harmonic
measurement (needs /G5 option)
A variety of other features, including line lter,
maximum hold, and integration function with categorization of positive and negative polarity, and average active power function
Low-cost model providing mobility for standalone measurement of standby consumed power and rated power
Single-phase model
Power measurement frequency range: DC and
0.1 Hz to 100 kHz (WT310EH: up to 20 kHz)
Basic power accuracy: 0.1% of reading.
Wide current input range (5 mA to 20 A)
(WT310EH 1 A to 40 A)
A variety of other features, including line lter,
maximum hold, and integration function with categorization of positive and negative polarity, and average active power function
Accessories sold separately
Model code Model name Notes
96060 Clamp-on probe
96061 Clamp-on probe
96062 Clamp-on probe
96063 Clamp-on probe
96064 Clamp-on probe
96065 Clamp-on probe
LOCK
SD
GB
96066 Clamp-on probe
98082 Extension cable
98031*
93047 Portable case Case with magnet
99073*
*Non-CE product. Not available for CE marking necessary region.
Power supply adapter
Conversion Cable (Banana-DIN)
Outline drawing
Before operating the product, read the user's manual thoroughly for proper and
safe operation.
40 mm AC 2 A, Leakage
current measurement
18 mm AC 50 A, Load current
measurement
24 mm AC 100 A, Load current
measurement
30 mm AC 200 A, Load current
measurement
40 mm AC 500 A, Load current
measurement
max. approx. 110 mm AC 1000 A exible type load current measurement
max. approx. 150 mm AC 3000 A, 3 CH Load current measurement
Extension cable for Clamp-on Probe
Power supply from measure line (100 to 240 V)
for 96030, 96033, 96036
120
175
(Unit: mm)
68
YOKOGAWA METERS & INSTRUMENTS CORPORATION
Global Sales Dept. /Phone: +81-422-52-6237 Facsimile: +81-422-52-6462
Copyright © 2015
E-mail: tm@cs.jp.yokogawa.com
YOKOGAWA CORPORATION OF AMERICA Phone: +1-770-253-7000 Facsimile: +1-770-254-0928 YOKOGAWA EUROPE B.V. Phone: +31-88-4641000 Facsimile: +31-88-4641111 YOKOGAWA SHANGHAI TRADING CO., LTD. Phone: +86-21-6239-6363 Facsimile: +86-21-6880-4987 YOKOGAWA ELECTRIC KOREA CO., LTD. Phone: +82-2-2628-3810 Facsimile: +82-2-2628-3899 YOKOGAWA ENGINEERING ASIA PTE. LTD. Phone: +65-6241-9933 Facsimile: +65-6241-2606 YOKOGAWA INDIA LTD. Phone: +91-80-4158-6000 Facsimile: +91-80-2852-8656 YOKOGAWA ELECTRIC CIS LTD. Phone: +7-495-737-7868 Facsimile: +7-495-737-7869 YOKOGAWA AMERICA DO SUL LTDA. Phone: +55-11-5681-2400 Facsimile: +55-11-5681-4434 YOKOGAWA AUSTRALIA PTY. LTD. Phone: +61-2-8870-1100 Facsimile: +61-2-8870-1111 YOKOGAWA MIDDLE EAST & AFRICA B.S.C(c) Phone: +973-17-358100 Facsimile: +973-17-336100
YMI-KS-MI-SE01
, Yokogawa Meters & Instruments Corporation
Subject to Change without notice.
Printed in Japan, 512(KP)