■ Inrush Current Measurement Method
Figure 1 Small motor Figure 2
Inrush AC current
Displays RMS value of 100ms period.
Explanation
A function to measure inrush current is available.
This function allows you to easily measure inrush current
that occurs when starting a motor, relay, or the like.
Just clamp the CW10 to the motor and set INRUSH
before starting the motor to automatically measure
and hold the inrush current that occured in 100 ms
period (Figure 2).
What is inrush current?
Starting current or inrush current is the
instantaneous electrical current that flows when
the power is initially turned on.
Application
Specifications
Inrush current
Accuracy
23± 5°C, 80%RH or less
Accuracy: ±(% of reading + digits)
Voltage
Rms-value detection
Function Range
Resolution
(Maximum reading)
Accuracy*
DCV
100 V 99.99 V
0.7% + 2
1000 V 999.9 V
ACV
100 V 99.99 V
1.0% + 5
50 ~ 500Hz
1000 V 999.9 V
LPF
ACV
100 V 99.99 V
50 ≤ f ≤ 60Hz: 1.0% + 5
60 < f ≤ 400Hz: 5.0% + 5
1000 V 999.9 V
* DCV<1000digits: add 6digits to accuracy
ACV<1000digits: add 3digits to accuracy
Maximum input voltage: 1000 Vrms, 1414.2 Vpk
Input impedance: approx. 3.5MΩ, <100pF
AC+DC Vrms accuracy=ACV accuracy + DCV accuracy
Crest factor effects
1.4 < CF ≤ 2.0: add 1.0% of reading to accuracy
2.0 < CF ≤ 2.5: add 2.5% of reading to accuracy
2.5 < CF ≤ 3.0: add 4.0% of reading to accuracy
Maximum input voltage: 690 Vrms CF=2 460 Vrms CF=3
Current
Rms-value detection
Function Range
Resolution
(Maximum reading)
Accuracy*
DCA
100 A 99.99 A 1.5% + 20
600 A 600.0 A*** 1.5% + 5*
ACA
100 A** 99.99 A
50 ≤ f ≤ 60Hz : 1.5% + 5*
60 < f ≤ 400Hz : 2.0% + 5*
600 A 600.0 A***
LPF
ACA
100 A** 99.99 A
50 ≤ f ≤ 60Hz : 1.5% + 5
60 < f ≤ 400Hz: 5.0% + 5
600 A 600.0 A***
*The measured value <1000digits: add 5digits to accuracy
**Input current ≥ 0.10A at 100 A range of ACA and LPF ACA
***600 A : Guaranteed accuracy (not maximum reading)
Maximum input current: 600 Arms, 848.5Apk
Conductor position effects: ±1.0% of reading
AC+DC Arms accuracy=ACA accuracy + DCA accuracy
Crest factor effects
1.4 < CF ≤ 2.0: add 1.0% of reading to accuracy
2.0 < CF ≤ 2.5: add 2.5% of reading to accuracy
2.5 < CF ≤ 3.0: add 4.0% of reading to accuracy
Maximum input current: 420 Arms CF=2 280 Arms CF=3
Peak Hold (AC mode only)
Function Range
Resolution
(Maximum reading)
Accuracy
ACV
100 V 140.0 V
3.0% + 15
1000 V 1400 V
ACA
100 A 140.0 A
3.0% + 15
600 A 850 A
PEAK MAX: polarity+, polarityMaximum input voltage and current: 1000 Vrms, 600 Arms
Sine wave, ACV ≥ 5 Vrms, ACA ≥ 5 Arms, 50 to 400 Hz continuous wave
Frequency (Hz)
Function Resolution (Measuring range) Accuracy
100 Hz 20.00 to 99.99 Hz
0.5% + 31000 Hz 20.0 to 999.9 Hz
10 kHz 0.020 to 9.999 kHz
Maximum input voltage and current: 1000 Vrms, 600 Arms
Input condition; 100 V range: 10 to 100 Vrms
(Sine wave) 1000 V range: 100 to 1000 Vrms
100 A range: 10 to 100 Arms (<400Hz)
600 A range: 100 to 600 Arms (<400Hz)
The measured value < approx. 10 Hz: 0.00Hz
Harmonic Measurement
Individual Harmonic
Harmonic order Resolution (Maximum reading) Accuracy
1st to 12th (h01- h12)
99.9 %
5% + 10
13th to 25th (h01- h12) 10% + 10
Maximum input voltage and current: approx. 1000 Vrms, 600 Arms
The “rdy” is displayed at ACV < 10 Vrms, ACA < 10 Arms
The “OutF” is displayed at f < 45, 65 < f (f: fundamental frequency)
Inrush Current
Function Range
Resolution
(Maximum reading)
Accuracy
ACA
100 A 99.99 A 2.5% + 20
600 A 600.0 A* 2.5% + 5
Maximum input current: approx. 600 Arms
*600 A : Guaranteed accuracy (not maximum reading)
100A range: ACA1 ≥ 10 Arms (Sine wave, 50Hz/60Hz)
600A range: ACA ≥ 100 Arms (Sine wave, 50Hz/60Hz)
Measurement time: approx. 100ms
Active Power
Function Range
Resolution
(Maximum reading)
Accuracy
ACW
DCW
10 kW 9.999 kW*
ACW: 2.5% + 11**
DCW: 2.2% + 22**
100 kW 99.99 kW
600 kW 600.0 kW**
* The measured value < 1.000kW: add 10digits to the accuracy.
**Conditions of accuracy (combination of Voltage and Current
range)
10 kW range: 100 V and 100 A
100 kW range: 100 V and 600 A or 1000 V and 100 A
600 kW range: 1000V and 600 A
Other combinations:
Accuracy: (Current accuracy×Voltage reading) + (Voltage
accuracy×Current reading)
**600 kW : Guaranteed accuracy (not maximum reading)
Maximum input voltage and current: 1000 Vrms, 600 Arms
ACW: ACV ≥ 10 Vrms and ACA ≥ 5 Arms (Sine wave, 50 ≤ f
≤
60Hz, PF=1.00)
DCW: at DCV ≥ 10 V and DCA ≥ 5 A
Power Factor
Function Resolution (Measuring range) Accuracy
Power factor -1.00~0.00~1.00 ±(3°+2digits)
Maximum input voltage and current: 1000 Vrms, 600 Arms
PF: ACV ≥ 10 Vrms and ACA ≥ 5 Arms (Sine wave, 50 ≤ f ≤ 60Hz)
Resistance/Continuity check
Function Range
Resolution
(Maximum reading)
Accuracy
Resistance
Ω
1000
Ω
999.9
Ω
1.0% + 5
10 k
Ω
9.999 k
Ω
1.0% + 3
100 k
Ω
99.99 k
Ω
Continuity check
1000
Ω
999.9
Ω
1.0% + 5
The buzzer turns on for resistances lower than approx. 30Ω. (Response time:
approx. 100msec)
Maximum input voltage: 1000 Vrms
Maximum test current: approx. 0.5mA
Open circuit voltage: approx. 3V
Diode Test
Function Resolution (Measuring range) Accuracy
Diode Test 0.40~0.80 V ±0.1 V
Maximum test current: approx. 0.5mA
Open circuit voltage: approx. 1.8V
General Specifi cations
Display count
:
9999 / 6000
Measuring rate
:
3 times / sec.
Over range indicator: “OL” or “-OL”
Auto Power Off
:
Approx. 15 minute.
Low-battery indicator
:
(four steps)
Power supply
:
9V alkaline battery (6LR61)
Battery life
:
When using alkaline battery, backlight off
Approx. 20 hours
Operating temperature and humidity: 0 ~ 50 °C (with no condensation)
≤
80% RH (0 ~ 30 °C)
≤
75% RH (30 ~ 40 °C)
≤
45% RH (40 ~ 50 °C)
Temperature coeffi cient: At 0 to 18 °C and 28 to 50 °C
Add 23±5 °C accuracy x 0.2 / °C
Storage temperature: -10 to 50 °C, 80% RH or less (remove the battery)
Withstand voltage: AC 6880 Vrms 5 sec. (between the core and the case)
AC 4300 Vrms 5 sec.
(between the core and the voltage input terminals)
AC 6880 Vrms 5 sec.
(between the voltage input terminals and the case)
Insulation resistance: 100MΩ or greater at 1000 VDC
(between the core and the case, the core and the voltage input
terminals and the voltage input terminals and the case)
Compliant standards:
Safety standards: EN 61010-1, EN 61010-2-032
1000V CAT.III, 600V CAT.IV
EN 61010-031 (the test leads)
Pollution degree 2, Indoor use, Altitude 2000m or less
EMC standards: EN 61326-1, EN 61326-2-1, EN 61326-2-2,
EN 55011
Dimensions: Approx 87.5 mm(W) x 242 mm(L) x 51 mm(D)
Diameter of measurable conductors
: ϕ
37mm (Maximum)
Weight
:
Approx. 435g (including the battery)
Accessories
:
Test leads 1set (Red and Black)
Carrying case
9V alkaline battery (6LR61)
User’s Manual
Accessories (Sold Separtely): Lead with Alligator Clip Model code 99014