YOKOGAWA CM6G User Manual

User ’s Manual
Model CM6G Gas Calorimeter
IM 11R02A01-02E
IM 11R02A01-02E
6th Edition

u Introduction

Thank you for purchasing Model CM6G Gas Calorimeter.
In order to have the product deliver its full capabilities, read this instruction manual thoroughly before you use it.
On how to use the Gas density meter, read the instruction manual which comes with the product.
WARNING
Since the ignition starts until it completes, or right after the ame is extinguished, sample gas mixed with the air will be released from the top of the equipment into the installation space. Even though the sample gas will be diluted su󰀩ciently by the air, great attention should be paid to the
following points if the sample gas contains harmful gases such as CO.
• Conrm that the air is supplied before introducing the sample gas.
• Use the equipment in a well-ventilated environment equipped with a ventilation system.
• Do not expose your face above the top of the calorie detector.
<INTRODUCTION>
i
n About This Manual
• This manual should be passed on to the end user so that the user can refer to the manual whenever they need.
• If the product is not used in a manner specied in this manual, the safety of this product may
be impaired.
• The contents of this manual are subject to change without prior notice.
• The contents of this manual shall not be reproduced or copied, in part or in whole,without permission.
• This manual explains the functions contained in this product, but does not warrant that they
are suitable for the particular purpose of the user.
• Every e󰀨ort has been made to ensure accuracy in the preparation of this manual. However, when you realize mistaken expressions or omissions, please contact the nearest Yokogawa Electric representative or sales o󰀩ce.
n Drawing Conventions
Some drawings may be partially emphasized, simplied, or omitted, for the convenience of
description.
n Inspection on delivery
After the delivery, open the product package carefully and inspect for damages caused during the transport. Should there be any damage or breakage on the product, please contact YOKOGAWA
immediately. Retain all the packing materials, containers and boxes that came with the product. After unpacking, make sure the delivered products include all of the component equipment you ordered. When conrming the specications, refer to the model and su󰀩x codes indicated on the
nameplate on the product. The models are indicated on the name plates. Check also those of accessories. For further information, please read subsection 1.2 of this document.
Media No.IM 11R02A01-02E 6thEdition : May 10, 2019 (YK) All Rights Reserved Copyright © 2016, Yokogawa Electric Corporation

u Safety Precautions

n Safety,Protection,ModicationoftheProduct
• In order to protect the system controlled by the product and the product itself and ensure safe operation, observe the safety precautions described in this user’s manual. We assume no liability for safety if users fail to observe these instructions when operating the product.
• If this instrument is used in a manner not specied in this user’s manual, the protection
provided by this instrument may be impaired.
• If any protection or safety circuit is required for the system controlled by the product or for
the product itself, prepare it separately.
• Be sure to use the spare parts approved by Yokogawa Electric Corporation (hereafter simply referred to as YOKOGAWA) when replacing parts or consumables.
• Modication of the product is strictly prohibited.
• The following safety symbols are used on the product as well as in this manual.
WARNING
This symbol indicates that an operator must follow the instructions laid out in this manual in order to avoid the risks for the human body and health including risk of injury, electric shock, or fatalities. or the damages to instruments. The manual describes what special care the operator must take to avoid such risks.
<INTRODUCTION>
ii
CAUTION
This symbol indicates that the operator must refer to the instructions in this manual in order to prevent the operator or instrument from being injured, or damaged.
The following are signal words to be found only in our instruction manuals.
CAUTION
This symbol gives essential information to avoid damages of soft or hardware, or system failure.
NOTE
This symbol indicates information that complements the present topic.
n Warning and Disclaimer
• The product is provided on an “as is” basis. YOKOGAWA shall have neither liability nor responsibility to any person or entity with respect to any direct or indirect loss or damage arising from using the product or any defect of the product that YOKOGAWA can not predict in advance.
n Trademark policy
All company or product and product names mentioned in this document are trade names, trademarks or registered trademarks of their representative companies. In this document, trademarks or trade names are not indicated with TM or ® as trademark symbols.
u After-sales Warranty
n Do not modify the product.
n Yokogawa warrants the product for the period stated in the pre-purchase quotation Yokogawa
shall conduct dened warranty service based on its standard. When the customer site is located
outside of the service area, a fee for dispatching the maintenance engineer will be charged to the customer.
n During the warranty period, for repair under warranty, carry or send the product to the local
sales representative or service o󰀩ce. Yokogawa will replace or repair any damaged parts and
return the product to you.
• Before returning a product for repair under warranty, provide us with the model name
and serial number and a description of the problem. Any diagrams or data explaining the
problem would also be appreciated.
• If we replace the product with a new one, we won’t provide you with a repair report.
n In the following cases, customer will be charged repair fee regardless of warranty period.
• Failure of components which are out of scope of warranty stated in instruction manual.
• Failure caused by usage of software, hardware or auxiliary equipment, which Yokogawa
Electric did not supply.
• Failure due to improper or insu󰀩cient maintenance by user.
• Failure due to modication, misuse or outside-of-specications operation which Yokogawa
does not authorize.
• Failure or damage caused by relocation of the instrument.
• Failure due to power supply (voltage, frequency) being outside specications or abnormal.
• Failure caused by any usage at location or its maintenance that Yokogawa does not authorize.
• Any damage from re, earthquake, storms and oods, lightning, disturbances, riots, warfare,
radiation and other natural changes.
n Yokogawa does not warrant conformance with the specic application at the user site.
Yokogawa will not bear direct / indirect responsibility for damage due to a specic application.
<INTRODUCTION>
iii
n Yokogawa Electric will not bear direct / indirect responsibility for any damage an end user might
incur, when the user congures the product into systems or resells the product.
n Maintenance service and supplying repair parts will be covered for ve years after the
production ends. For repair for the product, please contact the nearest sales o󰀩ce described in
this instruction manual.
Blank Page
Model CM6G Gas Calorimeter
IM 11R02A01-02E 6th Edition
CONTENTS
u Introduction ....................................................................................................i
u Safety Precautions .......................................................................................ii
1. Outline ....................................................................................................... 1-1
1.1 StandardSpecications ................................................................................... 1-1
1.2 ModelandSu󰀩xCodes ...................................................................................1-3
1.3 ..............................................................Standard Systems for Each Application
1-4
1.4 ExternalDimensions ........................................................................................ 1-6
1.5 ....................................................................................Principle of Measurements
1-10
Toc-1
2. Installation, Piping, Wiring ...................................................................... 2-1
2.1 Installations .......................................................................................................2-1
2.2 Externalpiping .................................................................................................. 2-1
2.3 ExternalWirings ................................................................................................ 2-4
3. Construction and Function ..................................................................... 3-1
3.1 Air Pressure Regulating System ..................................................................... 3-1
3.2 Gas Pressure Control Section ......................................................................... 3-1
3.3 Di󰀨erentialPressureDetectionPart ............................................................... 3-4
3.4 Calorie Detector ................................................................................................ 3-5
3.4.1 Burner Unit ......................................................................................... 3-8
3.4.2 High/Low Alarm Action ....................................................................... 3-9
3.4.3 Ignition ............................................................................................... 3-9
3.5 Computing Station .......................................................................................... 3-10
3.5.1 Display ............................................................................................. 3-10
3.5.2 Contents of the Data Display ........................................................... 3-13
3.5.3 Correcting Computation ................................................................... 3-14
3.6 Density Meter ................................................................................................... 3-16
4. Preparation of Operation ......................................................................... 4-1
4.1 Sampling Section (Outside panel) .................................................................. 4-1
4.2 Status of Valves ................................................................................................. 4-1
4.3 Water Supply (For Steel Mill Use) .................................................................... 4-1
4.4 Supply of the Air ................................................................................................ 4-2
4.5 Supply of Power ................................................................................................ 4-2
4.6 TheZeroAdjustmentoftheDi󰀨erentialPressureTransmitter .................... 4-2
4.7 AirDi󰀨erentialPressureAdjustment .............................................................. 4-3
IM 11R02A01-02E
6th Edition :May 10, 2019-00
Toc-2
4.8 Pressure Adjustment of the Gas Line .............................................................4-3
5. Operation ................................................................................................... 5-1
5.1 Start Operation .................................................................................................. 5-1
5.2 Stopping Operations ........................................................................................ 5-2
6. Calibration ................................................................................................. 6-1
6.1 Supply of the Calibration Gas .......................................................................... 6-1
6.1.1 For Town Gas Use .............................................................................6-1
6.1.2 For Steel Mill Use ............................................................................... 6-2
6.2 Span Adjustment of the Calorie Detector.......................................................6-2
6.3 Calibration of the Computing Station ............................................................. 6-4
6.4 Calibration of the Density Meter ...................................................................... 6-4
7. Maintenance .............................................................................................. 7-1
7.1 Daily Check ........................................................................................................ 7-1
7.1.1 Air, Gas Di󰀨erential Pressure Adjustment ......................................... 7-1
7.1.2 Take Out the Water Out of the Drain Pot (Pump for Steel Mill Use) .. 7-1
7.1.3 Take Out the Drain from the Air Set ................................................... 7-1
7.2 Regular Check ................................................................................................... 7-2
7.2.1 Cleaning of the Orice Plate and Replacement of O-ring .................7-2
7.2.2 Fulo Filter (For Steel Mill Use)..........................................................7-3
7.2.3 Line Filter (For Town Gas Use) .......................................................... 7-3
7.2.4 Washing Bubbler or Pressure Regulating Pot ................................... 7-4
7.2.5 Dehumidier (For Steel Mill Use) ....................................................... 7-4
7.2.6 Density Meter ..................................................................................... 7-4
7.3 Check at the Regular Service .......................................................................... 7-5
7.3.1 Check the Burner Flame .................................................................... 7-5
7.3.2 Zero Adjustment of the Di󰀨erential Pressure Transmitter .................. 7-6
7.3.3 Others ................................................................................................ 7-6
8. Troubleshooting ....................................................................................... 8-1
8.1 Gas Sampling Pressure Regulating Section ................................................. 8-1
8.2 Air Pressure Adjustment Section ....................................................................8-2
8.3 Di󰀨erentialPressureTransmitterSection ...................................................... 8-2
8.4 Signal Section ................................................................................................... 8-3
8.5 Computing Station ............................................................................................ 8-5
8.6 Other Troubleshooting ..................................................................................... 8-5
Revision Information ...............................................................................................i
IM 11R02A01-02E 6th Edition :May 10, 2019-00
<1. Outline>

1. Outline

Model CM6G Gas calorimeter measures and controls a caloric value of gases, Wobbe-Index (hereinafter referred to as WI), a theoretical air requirement, and a heat input for various kind of
gas burning furnaces.
It detects a temperature rise of a sample gas, whose pressure is normally controlled, by burning it at a burner through the medium of air.
It picks up the ow rate of the sample gas and the air as the di󰀨erential pressure signal and outputs a signal of WI after compensating calculation of the indication di󰀨erence caused by the ow rate variation.
It also detects density of the sample gas by a density meter, and add to WI signal density
compensation, then, outputs a caloric signal.
1.1 StandardSpecications
EMC : EMC Regulatory Arrangement in Australia and New Zealand Korea Electromagnetic Conformity Standard Class A
A급 기기 (업무용 방송통신기자재)  이 기기는 업무용(A급) 전자파적합기기로서 판매자 또는 사용자는 이 점을 주의하시기 바라며, 가정외의 지역에서 사용하는 것을 목적으로 합니다.
한국 전자파적합성 기준
1-1
n Town Gas Application
Measurement Object : Measurement and control of WI or the caloric value of fuel gas
Measuring Range : 3 to 62 MJ/Nm
Sample Conditions :
Dust ; 5 mg/Nm Temperature ; 50°C or less Humidity ; dew point of 0°C or less Pressure ; (1) 10 to 20 kPa
Range : Select scale range (Span) :
General Gas ; 30 to 50% of maximum value of the span Butane or Butene + Air ; 20 to 30% of maximum value of the span Propane or Propylene + Air ;25 to 40% of maximum value of the span
Output : 1 to 5 V DC, 4 to 20 mA DC (simultaneously), non-isolated,
load resistance; 750 Ω or less
Alarm Contact Output :
Flame o󰀨 alarm; 100 V AC, 5 A, closed when alarm occurs (resistance
Orice Temperature alarm; 100 V AC, 3 A, closed when alarm occurs
Contact Input : Remote ignition (Custom order); 24 V DC, 0.1A or more
for town gas.
3
3
or less
(2) 10 kPa or under: with pump (3) 100 to 600 kPa: with pressure reducing valve
load) when contact is opened, the leakage current is 2 mA or less (100 V AC)
(resistance load)
<1. Outline>
Repeatability
Measurement Measuring range (Note 1) Repeatability
WI
Caloric value
3
MJ/Nm
Note 1: High caloric value means 6.3 MJ/Nm3 or more. Low caloric value means below 6.3 MJ/Nm3.
High caloric value
Low caloric value
High caloric value
Low caloric value
± 0.5% of measured value ± 1.0% of measured value
± 1.0% of measured value ± 1.5% of measured value
Sample Gas Flow Rate : Approx. 10 l/min
Response Time (Note 2):
Max.WImeasured Dead time Response time (63.2%)
50 or more 32 or more, less than 50 13 or more, less than 32 Less than 13
Note 2: Response time varies depending on the WI of a sample gas. This is due to the di󰀨erent sample gas ow rate of the calorimeter. The ow rate is preset depending on the WI of the sample gas to prevent the caloric value at the detector burner from exceeding the upper limit.
Approx. 30 sec or less Approx. 27 sec or less Approx. 23 sec or less Approx. 21 sec or less
Approx. 60 sec or less Approx. 53 sec or less Approx. 47 sec or less Approx. 41 sec or less
Utility :
Instrument Air ; Approx. 50 Nl/min, pressure 300 to 700 kPa, dew point of 0°C or less Power Supply ; 100 V AC ± 10%, single phase, 50/60 Hz (Note 3), 860 VA max.
Note 3: When you perform a low caloric value measurement, the frequency variation should be within ±0.4%. If the frequency variation
exceeds ±0.4% (especially for exporting overseas), consult with Yokogawa.
1-2
n Steel Mill Application
Measurement Object : Measurement and control of WI or the caloric value of fuel gas for a
Measuring range : 3 to 62 MJ/Nm
Sample Conditions:
Dust; 100 mg/Nm Temperature; 50°C or less
Pressure; (1) 8 to 15 kPa (2) 8 kPa or under: with pump
Range : Select scale range (Span):
General Gas; 30 to 50% of maximum value of the span Butane or Butene + Air; 20 to 30% of maximum value of the span Propane or Propylene + Air; 25 to 40% of maximum value of the span
Output : 1 to 5 V DC, 4 to 20 mA DC (simultaneously), non-isolated, load
resistance 750Ω or less
Alarm Contact Output :
Flame o󰀨 alarm;
Temperature alarm; 100 V AC, 3 A, closed when alarm occurs (resistance load)
Contact Input: Remote ignition (Custom order); 24 V DC, 0.1A or more
Repeatability :
Measurement Measuring range (Note 1) Repeatability
WI
Caloric value
3
MJ/Nm
steel mill.
High caloric value
Low caloric value
High caloric value
Low caloric value
3
3
or less
100 V AC, 5 A, closed when alarm occurs (resistance load) when contact is opened, the leakage current is 2 mA or less (100V AC)
0.5% of measured value
1.0% of measured value
1.0% of measured value
1.5% of measured value
Note 1: High caloric value means 6.3 MJ/Nm3 or more. Low caloric value means below 6.3 MJ/Nm3.
<1. Outline>
Sample Gas Flow Rate :Approx. 10 l/min.
Response Time (Note 2):
Max.WImeasured Dead time Response time (63.2%)
50 or more 32 or more, less than 50 13 or more, less than 32 Less than 13
Note 2: Response time varies depending on the WI of a sample gas. This is due to the di󰀨erent sample gas ow rate of the calorimeter. The ow rate is preset depending on the WI of the sample gas to prevent the caloric value at the detector burner from exceeding the upper unit.
42 sec or less 39 sec or less 36 sec or less 30 sec or less
70 sec or less 60 sec or less 50 sec or less 45 sec or less
Utility :
Water ; Approx. 0.2 l/min, pressure 200 to 600 kPa Instrument Air ; Approx. 50 Nl/min, pressure 300 to 700 kPa, dew point of 0°C or less Power Supply ; 100 V AC ± 10%, single phase, 50/60 Hz (Note 3), 1100 VA max.
Note 3: In case of low caloric value measurement, frequency variation should be within ± 0.4%. If frequency variation exceeds ± 0.4%,
consult with Yokogawa.
Panel:
Construction : For indoor installation, rack panel
Paint Color: Munsell 3.2PB7.4/1.2 (inside and outside)
Ambient Temperature:
0 to 40°C (little temperature variation,particularly no
rapid change )
1-3
1.2 ModelandSu󰀩xCodes
n CM6G Gas Calorimeter
Model Su󰀩xCode Option Code Description
CM6G
Gas
Pressure
Measurement 00
Power supply -5
Range R - - - - - - - - - - Measuring range
Style *C - - - - - - - - - - Style C
Note: Measuring range and unit must be specied.
- - - - - - - - - - - - - - - - -
-S6 - - - - - - - - - - Always - S6
1 2 3 4 5 6 7 8
10
-6
- - - - - - - - - - Gas calorimeter
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
Gas pressure 10 to 20 kPa for town gas, quake-proof Gas pressure 10 to 20 kPa for town gas Gas pressure 10 kPa or under for town gas Gas pressure 100 ro 600 kPa for town gas Gas pressure 8 to 15 kPa for steel mill, without preheating Gas pressure 8 to 15 kPa for steel mill, with preheating Gas pressure 8 kPa or under for steel mill, without preheating Gas pressure 8 kPa or under for steel mill, with preheating
WI measurement
Caloric value measurement (GD400G should be purchased
separately)
100 V AC 50 Hz 100 V AC 60 Hz
l Standard Accessories
Followings are the standard accessories supplied.
• Calorie Detector
Name Q’ty Remarks
Mirror Fuse
For burner ame inspection
1 2
3.15 A (A1113EF)
<1. Outline>
• Orice Assembly
Name Q’ty Remarks
O-Ring O-Ring
Hexagon Wrench Hexagon Wrench
1
P16 (Viton) (Y9114XB)
3
P20 (Silicon) (L9817MT)
1
Nominal size 1.5 mm
1
Nominal size 2.5 mm
n Gas Density Meter
Gas Density Meter compensates a density value in caloric value measurement. It is not required for WI measurement.
Converter: GD400G-N-10-N-/PA
Detector: GD300S-J-/KU
Measuring range and unit (specic gravity or density) should be specied. See GS 11T3E1-01E
for further information.
n Option
You must purchase options separately.
Name Part no. Description
Probe H7800HA Insertion length 650 mm
Probe H7800HB Insertion length 1150 mm
Probe H7800HC Insertion length 1650 mm
Fulo lter G7043XJ Element material: Polypropylene
Pressure reducing valve G7008XF Primary pressure: 15 MPa max.
Pore size: 50 μm
Body: SUS 316 Connection: Rc 1/2
Secondary pressure: 0 to 200 kPa
Material: Brass
1-4

1.3 Standard Systems for Each Application

Application Measurement Systemspecication Su󰀩xcode*
WI
Town Gas
Caloric value
3
MJ/Nm
WI
Steel Mill
Caloric value
3
MJ/Nm
* The code is basic for gas pressure and measurement.
Note: A wet sample gas in the town gas application is outside the scope of the standard specications. Consult
with Yokogawa.
Without density meter
With density meter
Without density meter
With density meter
Gas pressure 10 to 20 kPa: Standard Gas pressure 10 kPa or under: With pump Gas pressure 100 to 600 kPa: With pressure
reducing value
Gas pressure 10 to 20 kPa: Quake-proof Gas pressure 10 to 20 kPa: Standard Gas pressure 10 kPa or under: With pump Gas pressure 100 to 600 kPa: With pressure
reducing value
Gas pressure 8 to 15 kPa: Without preheating Gas pressure 8 to 15 kPa: With preheating Gas pressure 8 kPa or under: Without preheating Gas pressure 8 kPa or under: With preheating
Gas pressure 8 to 15 kPa: Without preheating Gas pressure 8 to 15 kPa: With preheating Gas pressure 8 kPa or under: Without preheating Gas pressure 8 kPa or under: With preheating
-S6200
-S6300
-S6400
-S6110
-S6210
-S6310
-S6410
-S6500
-S6600
-S6700
-S6800
-S6510
-S6610
-S6710
-S6810
<1. Outline>
n Instructions for System Selection
The quake-proof type gas calorimeter is always equipped with the density meter.
The CM6G Gas Calorimeter controls the ow rate under a constant di󰀨erential pressure. In
the caloric value measurement, if the density of a sample gas changes, a ow rate error proportional to the reciprocal of the square root of the density of the sample gas, will be generated, which directly a󰀨ects the caloric value. Therefore, density compensation is required using a density meter.
For the WI measurement, a density meter is not required since the WI is a value
proportional to
ρ
.
g
1/
1-5
ρ
g
1/
,
<1. Outline>
1.4 ExternalDimensions
n For Town Gas Applications
CM6G-S6200, S6210, S6300, S6310, S6400, S6410
1-6
Unit: mm
Mark Name Connection
A SAMPLE GAS IN Rc1/4
SAMPLE GAS OUT
B C INST.AIR IN  Rc1/4 D STD.GAS IN  Rc1/4 E STD.GAS IN  Rc1/4
B D E
A
C
250
View X
130
900
700 550 400
300
0
OR VENT Rc1/2
30
740 800
Density meter converter
Computing station
Approx.70
30
*
1800100
Approx.1970
100
4-Ø14 holes
X
* CM6G-S610 (with density meter) only
Density meter detector
700 900
*
Calorie detector
Flowmeter for density meter
100
*
Maintenance Space
100
800800 500
900
Back
Front
500
Weight: Approx. 300kg
Dim-6210.ai
<1. Outline>
n For Town Gas Application (Quake-proof Type)
CM6G-S6110
1-7
30 640 30
700
Ø120 hole
Density meter converter
1500100
Approx. 1670
X
D
400
Computing station
FAIL
---CM6G---
ALM
TREND 3
90.00
18.00
Y2Y1X3X1500.0
35.00
C
1000.0
37.67
100.00
23.02
A
M
▼▼▼
PF
0.00
14.65
0.0
23.02
▼ C
O
SHIFT
<
<<<
YOKOGAWA◇
1000
4-Ø15 holes
100600
Calorie detector
Flowmeter for density meter
100800100
300700
90
A
B
C2
View X
C1
90
Mark Name Connection
A SAMPLE GAS IN  Rc1/4 B INST.AIR IN  Rc1/4 C1 STD.GAS IN (ZERO)  Rc1/4 C2 STD.GAS IN (SPAN)  Rc1/4 D SAMPLE GAS OUT OR VENT  Rc1/2
Wiring to switch box should be made through the bottom.
Unit: mm
1000 850
0
Maintenance Space
1000 100500
Back
Front
500
700
800
Weight: Approx. 350kg
Dim-6110.ai
<1. Outline>
n For Steel Mill Application
CM6G-S6500, S6510, S6600, S6610, S6700, S6710, S6800, S6810
1-8
*
Flowmeter for density meter
Calorie detector
Approx.1970
100
Unit: mm
*
X
Mark Name Connection A SAMPLE GAS IN Rc1/2 B SAMPLE GAS OUT Rc1/2 C SAMPLE GAS OUT Rc1/2 D STD.GAS IN Rc1/4 E WATER IN Rc1/2 F DRAIN OUT Rc1/2 G INST.AIR IN Rc1/4 H SAMPLE GAS VENT Rc1/4
500
600
700
800
80
1900
1650 1500
150
750
View X
60
100
300 200 100 0
G
D
Fulflo filter
Dehumidifier
E
F
740 30(30)
800
0
350
Washing bubbler
Drain pot
Pressure regulating pot
Density meter converter
Approx. 70
H B
C A
**
Pump
*
Computing station
4-Φ14 holes
80 60
40
Approx.1160
(100)
* CM6G-S610 (with density meter and flowmeter)
** CM6G-S670, CM6G-S680 (with pump)
Density meter detector
700
900
100
500
800
800
Maintenance Space
540
900260
Back
Front
500
Weight: Approx. 350kg
Dim-6810.ai
<1. Outline>
n Dimensions of Options
l Probe
Unit: mm
Part number L H7800HA Approx. 650 H7800HB Approx. 1150 H7800HC Approx. 1650
)( 350Insertion length L
18
Φ34
1-9
JIS 10K-80-FF Equiv. SUS304
l FuloFilter(Partno.:G7043XJ)
108
Rc1/2Rc1/2
Approx. 350
Unit: mm
l PressureReducingValve(Partno.:G7008XF)
Secondary pressure gauge
0.3MPaG
Rc1
Primary pressure gauge 25MPaG
H7800HA.ai
Unit: mm
Approx. 119
Approx. 38
Handle for pressure control
Approx. 62
Approx. 120
Outlet
Handle for stop valve
Stop valve
Rc1/4
Outlet
H
C
I
A
H
K
O
A
T
Safety valve
Approx. 166
Daiflon packing
W22 Left-hand thread
Inlet
Nut
Approx. 81
F2-8.ai
<1. Outline>

1.5 Principle of Measurements

The instrument detects the temperature di󰀨erence, using a thermocouple, between the exhaust
combustion gas made after sample gas is burnt in the burner, and the feed air at the inlet of
the burner. Then the instrument amplies the value of temperature di󰀨erence. It adds the compensating calculation to the output signal and measures the caloric value of WI.
1-10
DIFFERENTIAL
PRESSURE
DETECTION
GAS
AIR
Figure 1.1 Measurement principle diagram
BURNING
PRIMARY
MIXING
SECONDARY
TERTIARY
AIR DIVIDING
BLOW
COMPENSATING
CALCULATION
OUTPUT
SIGNAL
When the sample gas is completely burnt by the air, the formula of the increased temperature is as follows:
∆θ = (1.1)
K • Fg
Cps • Fs
where K: Caloric value of the sample gas
Fg: Flow rate of the sample gas
Fs: Air-diluted combustion exhaust gas ow rate
Cps: Constant pressure heat ratio of air-diluted combustion exhaust gas
Air ow rate Fa is big enough compared with the sample gas ow rate Fg (Fa:Fg=50-200:1)
and Cps Cpa, Fs Fa+Fg=Fa ( 1 + g ), therefore, the formula (1.1) is as follows:
∆θ = (1.2)
K • Fg
Cpa • Fa (1 + g)
where Fa: Air ow rate
Cpa: Air constant pressure heat ratio
g: Fg/Fa
When using orice and take out Fa, Fg as a di󰀨erential pressure of before and after orice, Fa
and Fg are represented by the following formula:
Fa = Ka (1.3)
Pa
ρ
a
Fg = Kg
Pg
ρ
g
where: ∆Pa, ∆Pg: Air, gas di󰀨erential pressure between before and after orice
ρ
a, ρg: Density of the air and the gas
ka, kg: Orice constant gure of the air and the gas (Orice coe󰀩cient  orice sectional area)
<1. Outline>
If insert formula (1.3) into formula (1.2), ∆θ is represented by the following formula:
1-11
∆θ = C.K. (1.4)
1
ρ
g
(C =
Pg
Pa
1
Cpa (1+g)
Kg
)
Ka
ρ
a
According to formula (1.4), if ∆Pa, ∆Pg are constant, temperature di󰀨erence ∆θ is in proportion
ρ
g
to WI (
K/
) or caloric value (K). Thus we can measure continuously WI after ∆θ
measurement, and the caloric value after the measurement and calculation of the density.
Blank Page
<2. Installation, Piping, Wiring>

2. Installation, Piping, Wiring

2.1 Installations

Observe the following conditions when you install instruments.
(1) Adequate space for maintenance should be provided around the gas calorimeter.
* See “External Dimention” on 1.4.
(2) The base should be horizontal..
(3) Ambient temperature is 0 to 40°C and no rapid change in ambient temperature is allowed.
*Rapid change here means a change of approximately 10°C within 30 minutes.
(4) Install the instrument in the place where it is not directly exposed to the current of a
conditioned air.
(5) Minimal vibration( If much vibration is unavoidable, take an appropriate measure to absorb
shock, e.g, use of vibration-proof robber).
(6) A ventilation system should be provided.
(7) Corrosive gases and dust are present in small quantities and humidity is low.
(8) The water of the sampling system and the drain line do not freeze up.
2-1
WARNING
From the moment the ignition starts until it completes, or right after the ame extinction, sample gas mixed with the air will be released from the top of the equipment into the installation space of the equipment. Use the equipment in a well ventilated environment equipped with a ventilation
system.
2.2 Externalpiping
Refer to the drawing Figure 1.4. on the connections on the panel
The principal points to be taken care are as follows:
• Bent the outlet of the bent piping as U shape, and prevent it from the rain penetration. Set the outlet at higher position as much as possible where there is little fear of physical danger.
• Drain piping shall be conducted below the level of drain outlet on the panel and conrm no
drain is accumulated on the panel bed.
• It is necessary for the sample line of the instrument for steel mill use to provide a slope of more than 1/3, so that there is no blockade to the gas line by the drain at the bending part of
the piping. Make piping as short as possible. Equip the sample line with thermal insulation
so as to prevent the drain in the pipe line from freezing.
• Locate the standard gas cylinders at the place where they are comparatively cool and not
exposed to direct sunshine.
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<2. Installation, Piping, Wiring>
Probe Sampling Point
Steam for thermal insulation
Water In
Vent to the
atmosphere
Drain
Safely vent to the atmosphere
Filter
Filter
Tetoron Braid Hose
Pipe Φ6/Φ4 Cu
Pipe Φ6/Φ4 Cu
Zero Gas
Span Gas
Valves for maintenance should be installed.
Pipe SUS 15A
Sample Gas Inlet
Std. Gas
Inlet
Pressure Reducing Valve
Wash water inlet
Pipe SUS 15A
Pressure:0.1-0.2MPa
Sample Gas Out
CM6G
Gas
Calorimeter
Pipe SUS 15A or greater
Sample Gas Vent
Sample Gas Out
Pipe SUS 15A
Drain Outlet
Pipe SUS 15A
Steam for blowing back
Pipe SUS 15A
Pressure Reducing Valve
∆H
∆H: Water sealing greater than pressure at sampling point is required.
Seal pot
Note: denotes that piping should be installed at an angle that allows drain to flow downstream and smoothly. F4-1.ai
House Requirements
Temperature: 0 to 40˚C
Not rapid temperature change allowed. Adequate ventilation and lighting provided.
Power Supply
100 V AC±10% 50/60Hz
Max. approx.1100 VA
Output Signal
Alarm Contact
Output Signal
Instrument Air Pressure: 0.3-0.7MPa
Flow rate: Approx. 50Nl/min
Figure 2.1 Recommended Sampling for Steel Mill Use
2-2
Drain
• Sampling point shall be made at the location above or side of the transfer pipe. When the
instrument is mounted on wall, x it with a slope that the top end of the probe is facing
downward.
• When xing the probe, use a ange JIS 10K 80A.
• At the sample gas outlet of the probe, it is recommended to provide a gate valve of 1/2 inch.
Probe
Gate valve (1/2 inch)
Flange
100 or more
Figure 2.2 Mounting of the Probe
Transfer pipe
F4-2.ai
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<2. Installation, Piping, Wiring>
• Fix lter vertically with their drain outlets facing downward. Mount it rmly to wall or to pillar with U bolt-like bolts. Provide 1/2 inch gate valves at the sample gas inlet and outlet. Take an ample space under lters so that checking and replacing of elements can be held easily. For drain exhaust pipes, use exible pipes.
Gate valve (1/2 inch)
2-3
Rc1/2
Out
Fulflo filter
Bush
Flexible tube
Drain
Rc1/4
Drain
Figure2.3ExampleFuloFilterswiththePiping
In
Union (1/2 inch)
U-bolt
F4-3.ai
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<2. Installation, Piping, Wiring>
2.3 ExternalWirings
External wires shall be connected from the terminal block of the switch box inside the
panel. Use M4 terminal screws. Use appropriate crimp terminals at the wire ends.
MAIN PANEL
DENSITY METER CONVERTER
COMPUTING STATION
SWITCH BOX
350
CALORIE DETECTOR
2-4
200
Figure 2.4 SwitchBoxandEarth
SWITCH
NFB
POWER TERMINAL
150
EARTH TERMINAL
SIGNAL TERMINAL
810
Figure 2.5 InsideaSwitchBox
FUSE
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<2. Installation, Piping, Wiring>
n Notes on Wiring
• Turn o󰀨 all power supplies while connecting cables and wires.
• An electrical wiring duct must not be used for the simultaneous connection of a large capacity converter, motor, or power supply.
• When cables are connected in a place with high or low ambient temperatures, use cables suitable for the place.
• The material of cables must be capable of withstanding harmful gases, lizids, oil , or solvents when cables are used in the atmosphere where these harmful substances are present.
• Use crimp terminals with insulated sleeve (M4 screws) for the wire ends.
n Wiring to Peripheral Equipment
l Power Supply
Use a 600 V insulated vinyl cabtyre cable (JIS C3312) with a cross-section area of 2 mm2 or
more, or a wire or cable that is the equivalent or better.
l Grounding
Connect a grounding wire to the grounding terminal inside the panel. A grounding wire must be
connected so that the grounding resistance becomes 100 Ω or less (equivalent to JIS Class D).
2-5
l Analog Output (4 to 20 mA DC)
Use a shielded twisted pair cable with a cross-section area of 0.5 mm2 or more, or a cable that is
the equivalent or better, and install it separately from the power supply and alarm output cables
and sources of electromagnetic interference. A shielded cable must be connected to the frame ground (FG) beside each of the output terminals. The load resistance from the perspective of this
equipment must be 750 Ω or less.
l Contact Output
Use a 600 V insulated vinyl cabtyre cable (JIS C3312) with a cross-section area of 2 mm2 or
more, or a wire or cable that is the equivalent or better.
The ame extinction alarm contact and orice chamber temperature drop alarm contact must be
non-voltage dry contacts and the contact rating must be as follows.(when contact is opened, the leakage current is 2 mA or less (100V AC))
Contact Rating
Flame distinction alarm contact 100 V AC 5 A
Orice chamber temperature drop alarm contact 100 V AC 3 A
l Contact Input (Remote Ignition: Made to order)
Input a non-voltage contact as a contact input signal. The contact rating is 24 V DC 1 A. The open or closed state of the input is determined by the resistance value from the perspective of
this equipment. The resistance value also includes the wiring resistance.
Contact closed: 200 Ω or less. Contact open: 100 kΩ or more.
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<2. Installation, Piping, Wiring>
2-6
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<3.Construction and Function>
F00.ai

3. Construction and Function

TypicalSystemConguration
Gas Calorimeter
Calorie Detector
Solenoid Valve
Computing Station
---CM6G---
TREND 3
18.00
Y2Y1X3X1500.0
35.00
1000.0
100.00
0.00
0.0
▼ C
<<<
YOKOGAWA◇
Differential Pressure Transmitter
Orifice Assembly
FAIL ALM
90.00
C
37.67
23.02
A
M
▼▼▼
PF
14.65
23.02
O
SHIFT
<
Output 4-20mA DC
1 to 5V DC
Density Meter
Gas Pressure Regulating Unit
Air Pressure Regulating Unit
P
Calorific Value WI (Wobbe Index)
Sample Gas
Instrument Air
3-1
Item Function / Description
Calorie detector
Detects WI or caloric value. Generates an alarm and takes protective
actions when the burner ame goes out or abnormal combustion occurs.
Computing station (digital) Calculates WI or caloric value. Displays selected parameters, e.g., each
di󰀨erential pressure and caloric value. Adjusts zero / span and others.
Density meter Measures density used for calculation of caloric value. Not required for
WI measurement.
Di󰀨erential pressure transmitter Detects di󰀨erential pressure of gas and air before and after orice, and
converts it to an electrical signal.
Orice assembly Gas and air orices housed in the constant temperature chamber.
Solenoid valve Serves as a safety valve to shut o󰀨 the sample gas ow.
Figure 3.1 Components and Functions of Model CM6G Gas Calorimeter

3.1 Air Pressure Regulating System

Refer to Figure 3.2, 3.3 and 3.4. The instrument air pressure (300 to 700 kPa) is reduced by air set (2-1) to about 200 kPa, further reduced to about 20 kPa by the pressure reducing valve (2-2). Then the pressure is set by needle valve (V-16) to a di󰀨erential pressure of 500 Pa.
The air is controlled at 40°C in the preheater, which helps the temperature control by the orice.
When the temperature rises above 60°C, thermostat in the preheater operates and intercepts the power supply of the heater. This air pressure regulating section is common to all systems.

3.2 Gas Pressure Control Section

Refer to Figure 3.2, 3.3 and 3.4.The gas pressure regulating section has two di󰀨erent types; town
gas use and steel mill use.
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<3.Construction and Function>
l Town Gas Use
The sample gas, introduced through line lter (3-1), increases its pressure by pump, or
decreases by pressure reducing valves, according to the pressure at the sampling point. The
pressure gauge (3-2) indicates 8 to 18 kPa and the owmeter (3-3) approx. 10 l/min, respectively. The sample gas, then, is set its di󰀨erential pressure to 500 Pa by pressure reducing valve (3-4) and (3-6), and at this time the pressure gauge (3-5) indicates approx. 3 kPa. When the density compensation system is equipped, the sample gas is introduced to the density meter with the ow rate of 1 l/min, through the owmeter (5-3).
The standard gas is reduced its pressure to 8 to 18 kPa by the pressure reducing valve (4-1), and supplied, same as the sample gas, with the ow rate approx. 10 l/min.
l Steel Mill Use
The pressure of the sample gas is increased, by the pump, according to that of the sampling point.
The pressure gauge (3-1) indicates approx. 6 kPa. The sample gas then ows through the washing bubbler (3-2) and the fulo lter (3-3), and then secure a constant pressure in the pressure regulating pot (3-4) through the water sealed pipe from the dehumidier (3-5), then set the di󰀨erential pressure by the pressure reducing valve (3-6) to 500 Pa.
To increase the pressure by the pump, a drain pot (3-9) is added. When the density
compensation system is equipped, the sample gas is supplied to the density meter with its ow rate of 1 l/min, through the owmeter (5-3). The pressure of the standard gas is reduced by the pressure reducing valve (4-1) to about 6 kPa and is supplied by the owmeter (4-2) with the ow
rate about 10 l/min.
3-2
Standard Flow Sheet
1. Town Gas Application (Standard Type)
CM6G-S6200, S6210, S6300, S6310, S6400, S6410
No. Item 1-1 Calorie detector 1-2 Solenoid valve 1-3 Orifice assembly 1-4 Differential pressure transmitter (air) 1-5 Differential pressure transmitter (gas) 1-6 Computing station 2-1 Air set 2-2 Pressure reducing valve 2-3 Pressure gauge 2-4 Preheating chamber 2-5 One touch coupler
V-1
V-2
SAMPLE GAS IN
SAMPLE GAS OUT OR VENT
Maximum back pressure is 1.5 kPa and no pressure fluctuation is allowed.
A
B
No. Item 3-1 Line filter 3-2 Pressure gauge 3-3 Flowmeter 3-4 Pressure reducing valve 3-5 Pressure gauge 3-6 Pressure reducing valve 3-7 Diaphragm pump (when specified) 3-8 Pressure reducing valve (when specified) 4-1 Pressure reducing valve for standard gas
*3
3-8
V-4
3-1
*1
*2
3-7
V-5
V-3
P
DT
5-1
5-2
*1: CM6G-S610 (with density meter) *2: CM6G-S630 (with diaphragm pump) *3: CM6G-S640 (with pressure reducing valve)
V-11
V-6
V-13
5-3
V-7
3-2
3-3
3-4
3-6
3-5
V-12
GAS DP.T
1-1
1-2
No. Item
5-1 Density meter detector (when specified) 5-2 Density meter converter (when specified) 5-3 Flowmeter for density meter (when specified) 6-1 Zero gas (supplied by customer) 6-2 Span gas (supplied by customer) 7-1 Pressure reducing valve for cylinder (optional) 7-2 Pressure reducing valve for cylinder (optional)
V1...16 Ball valve, needle valve
1-6
1-3
AIR DP.T
1-5
1-4
V-16
4-1
2-4
2-3
V-15
2-1
V-14
2-2
2-5
V-9
V-8
V-10
FLOW-6410.ai
INST.AIR IN
0.3~0.7MPa
C
50Nl/min
STD GAS IN (ZERO)
7-1
7-2
D
E
6-2
STD GAS IN (SPAN)
6-1
Figure 3.2 Flow Sheet (for Town Gas)
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<3.Construction and Function>
2. Town Gas Application (Quake-proof Type)
CM6G-S6110
No. Item 1-1 Calorie detector 1-2 Solenoid valve 1-3 Orifice assembly 1-4 Differential pressure transmitter (air) 1-5 Differential pressure transmitter (gas) 1-6 Computing station 2-1 Air set 2-2 Pressure reducing valve 2-3 Pressure gauge 2-4 Preheating chamber 2-5 One touch coupler
No. Item 3-1 Line filter 3-2 Pressure gauge 3-3 Flowmeter 3-4 Pressure reducing valve 3-5 Pressure gauge 3-6 Pressure reducing valve 4-1 Pressure reducing valve 5-1 Density meter detector 5-2 Density meter converter 5-3 Flowmeter for density meter
No. Item 6-1 Zero gas (supplied by customer) 6-2 Span gas (supplied by customer) 7-1 Pressure reducing valve for cylinder (optional) 7-2 Pressure reducing valve for cylinder (optional)
V1...16 Ball valve, needle valve
3-3
V-4
V-1
V-2
SAMPLE GAS IN
(10 to 20kPa)
SAMPLE GAS OUT
OR VENT Maximum back pressure is 1.5 kPa and no pressure fluctuation is allowed.
A
D
V-3
3-1
5-1
5-2
Figure 3.3 Flow Sheet (for Town Gas)
1-5
1-3
AIR DP.T
1-4
1-6
V-16
4-1
2-4
2-2
2-3
V-15
2-5
V-8
2-1
V-14
V-9
V-10
FLOW-6110.ai
INST.AIR IN
0.3 to 0.7MPa
B
50Nl/min
STD GAS IN (ZERO)
7-1
7-2
C1
C2
STD GAS IN (SPAN)
6-1
6-2
1-1
3-3
3-4
3-6
3-5
V-12
GAS DP.T
1-2
3-2
V-11
V-5
V-6
V-7
V-13
P
5-3
DT
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<3.Construction and Function>
3. Steel Mill Application
CM6G-S6500, S6510, S6600, S6610, S6700, S6710, S6800, S6810
No. Item 1-1 Calorie detector 1-2 Solenoid valve 1-3 Orifice assembly 1-4 Differential pressure transmitter (air) 1-5 Differential pressure transmitter (gas) 1-6 Computing station 2-1 Air set 2-2 Pressure reducing valve 2-3 Pressure gauge 2-4 Preheating chamber 2-5 One touch coupler
No. Item 3-1 Pressure gauge 3-2 Washing bubbler 3-3 Fulflo filter 3-4 Pressure regulating pot 3-5 Dehumidifier 3-6 Pressure reducing valve 3-7 Line filter 3-8 Diaphragm pump (when specified) 3-9 Drain pot (when specified) 4-1 Pressure reducing valve for cylinder 4-2 Flowmeter
No. Item 5-1 Density meter detector (when specified) 5-2 Density meter converter (when specified) 5-3 Flowmeter for density meter (when specified) 6-1 Water Flowmeter 7-1 Probe (optional) 7-2 Fulflo filter (optional) 9-1 Pressure reducing valve for cylinder (optional) 9-2 Pressure reducing valve for cylinder (optional) 10-1 Standard gas cylinder (supplied by customer) 10-2 Standard gas cylinder (supplied by customer) 11-1 Drain pot (supplied by customer)
V1...16 Ball valve, needle valve
3-4
WATER IN
DRAIN
9-1
9-2
7-1
10-2
SAMPLE GAS IN
11-1
STD.GAS IN
(SPAN)
STD.GAS IN
10-1
7-2
WATER IN
DRAIN OUT
(ZERO)
SAMPLE GAS OUT
2-5
B
*1
*2
4-1
V-9
D
V-10
D
A
V-1
3-9
E
F
V-8
3-1 V-3
3-2
V-13
V-4
P
V-6
V-5
6-2
3-8
V-2
V-7
*1: CM6G-S610 (with density meter) *2: CM6G-S670, CM6G-S680 (with diaphragm pump)
SAMPLE GAS VENT
SAMPLE GAS OUT
H
C
4-2
V-11
V-12
3-3
*1
5-3
3-6
3-7
3-5
3-4
1-2
5-1
5-2
1-1
1-5
1-3
V-16
1-4
2-4
2-2
P
1-6
V-15
2-3
2-1
V-14
INST.AIR IN
G
FLOW-6810.ai
Figure 3.4 Flow Sheet (Steel Mill Use)
3.3 Di󰀨erentialPressureDetectionPart
In order to calculate the di󰀨erential pressure correction value, the ow rate is acquired from the orice as a di󰀨erential pressure and converted to an electrical signal using the di󰀨erential pressure transmitter. The orice is housed in a constant temperature chamber (orice assembly), the temperature of which is maintained at approximately 50°C by the temperature controller, to prevent the temperature drift of the actual ow rate.
When the temperature exceeds 90°C, the safety thermostat is activated to shut o󰀨 the heater power supply. When the temperature falls after the power supply is shut o󰀨, the alarm thermostat
is activated to provide an alarm (closed contact) output.
CAUTION
Power is supplied to the di󰀨erential pressure transmitter and the temperature controller from the
calorie detector. To operate the transmitter and the controller, turn on the calorie detector switch
of the switch box and the POWER switch on the front panel of the calorie detector.
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<3.Construction and Function>
MAIN PANEL
ORIFICE ASSEMBLY
DIFFERENTIAL PRESSURE
TEMPERATURE CONTROLLER
TRANSMITTER
Figure3.5Di󰀨erentialPressureDetectionEquipment
3-5
3
9
4
19
16
15
8
10
18
No. Name
Air inlet
1
Gas inlet
2
Air outlet
3
Gas outlet
4
Air differential pressure take-out
5
Gas differential pressure take-out
6
Orifice section
7
No. Name
8
9 10 11 12 13 14
Figure3.6OriceAssembly

3.4 Calorie Detector

The calorie detector consists of a burner unit, which detects the temperature di󰀨erence before and after the sample gas is burned, a detected signal amplication and ignition-and safety sequence circuit, and distributor circuits of the transmitter for air and the transmitter for gas.
The detected signal is converted from approximately 0 - 20 mV to 1 - 5 V DC and input to the computing station. If burner ame extinction or excessive combustion occurs, an alarm (closed
contact) output is produced.
7
12
11
Gas orifice section Crotchet joint Fitting screw Heat conversion block Heater plate Safety thermostat (up) Alarm thermostat (down)
13·14
5
6
2
17
No.
Thermistor (check)
15
Thermistor (control)
16
Temperature check terminal
17
Neon lamp
18
Terminal
19
1
Name
F0506.ai
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<3.Construction and Function>
PRINTED CIRCUIT BOARD ASSEMBLY
BURNER UNIT
IGNITION TRANS. ASSEMBLY
Figure 3.7 Calorie Detector
3-6
POWER Lamp
IGNIT Lamp
0
Figure 3.8 Front Panel Assembly
SAFE EV Switch
LOW AL Volume
MEAS Lamp
ALARM Lamp
START Switch
POWER Switch
METER Switch
SPAN Volume
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<3.Construction and Function>
Table 3.1 Function and operation procedure for each switch of the calorie detector
Name Function and operation procedures
“POWER” switch • Power supply switch.
• When the switch turns to “ON”, the power is supplied to the calorie detector and at the
same time the lamps, “POWER” and “ALARM” are lit on.
“START” switch • One push on “START” switch introduces gas to the burner unit and operates the ignition.
• At the same time, the lamp “IGNIT” is on while the lamp “ALARM” is lit o󰀨.
• If ignition occurs within the preset time, the lamp “IGNIT” is lit o󰀨 but the lamp “MEAS”
turns on.
• If ignition does not occur within the preset time, the lamp “IGNIT” is lit o󰀨 and the lamp
“ALARM” turns on again.
“SAFE EV” switch • Normally leave the switch at “SET”.
• When you adjust the di󰀨erential pressure, turn it to “RELEASE”. Then the lamp “ALARM” is let o󰀨, alarm is released and solenoid valve of the gas line is opened.
• Switching over from “SET” to “RELEASE” during the operation does not a󰀨ect anything.
However, if you leave the switch at “RELEASE”, no alarm signal is given even when the
burner ame is o󰀨.
• When the switch shows “RELEASE”, “START” switch does not start the operation.
“METER” switch •The indication of the indicator and the X5 display(*1) of the computing station switches
between “P.MEAS“ and “LOW AL“
• The switching does not a󰀨ect the nal output and sequence action.
“SPAN” volume • With this span volume you can adjust the output of the calorie detector to 1 - 5 V DC.
• If you turn this to the right, the span point becomes bigger.
“LOW AL” volume • The level for the extinction alarm (lower alarm) is set by this volume.
• If you turn this to the right, the alarm level becomes higher.
“POWER” lamp • When the lamp is lit, power is being supplied to the calorie detector, di󰀨erential pressure
transmitter, and temperature controller.
“IGNIT” lamp • While the lamp is lit, the ignition proceeds.
“MEAS” lamp • While the lamp is lit, the system is under the measurement condition.
“ALARM” lamp • While the lamp is lit, alarm is being released.
(*1) The display range of X5 is 0 to 30.0 mV. “P.MEAS“ and “LOW AL“ indicate the following values.
3-7
Signal Name Display Value
“P.MEAS“ Thermocouple electromotive force 0 - 20.0 mV “LOW AL“ Flame extinction alarm threshold value 0 - 10.0 mV (= 0 - 100%)
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<3.Construction and Function>

3.4.1 Burner Unit

The sample gas burns inside the burner unit and a thermocouple detects the burning temperature increase. The air is introduced from the air inlet and divided to the primary, secondary and tertiary air. The primary and
secondary ones are for burning the sample gas and the tertiary is for diluting and stirring the exhaust gas. The sample gas is mixed with the primary air (In case of the low calorie gas, the primary air is throttled), and burnt
completely by the secondary air. Then, the combustion temperature generates, and burnt gas is promptly
diluted and stirred by the tertiary air. Finally the gas is exhausted out from the top of the calorie detector. The increased temperature is measured with the di󰀨erence of the electromotive force between the cold junction point (located at the air inlet) and the hot junction point (inside the mixed diluted exhaust gas). The heating
wire wound the burner tip is used for both ignition and preheating (when lower calorie gas is being used.)
Heater wire
Burner
tip
Lead wire
3-8
Figure 3.9 Burner Tip Assembly
1
14
13
12 11 10
2
3 4
6
5
9
7
8
Figure 3.10 Burner Unit
F0509.ai
No. Name
1 Hot junction detect point 2 Cold junction detect point 3 Air divider 4 Secondary air throttle screw 5 Primary air throttle screw 6 Air inlet 7 Gas inlet 8 Connector 9 Burner tip assembly 10 Combustion pipe 11 Stream contact pipe 12 Measuring pipe 13 Reverse flow pipe 14 External pipe
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<3.Construction and Function>
Ignition action starts

3.4.2 High/Low Alarm Action

Higher and lower alarm limits are set for the amplied thermocouple output, respectively, to execute a sequence. The higher alarm limit is the point to alarm the excessive combustion of the burner, which is set to approximately 120% of the span. The lower alarm limit is the point to alarm the ame extinction
of the burner. You can change the limit by “LOW AL”volume on the front panel of the calorie
detector. However, the lower alarm limit is set to an appropriate value in the nal adjustment test
at the factory before shipment. If the higher or lower limit alarm occurs, the electromagnetic valve (EV) is closed to stop the supply of gas

3.4.3 Ignition

Press START on the front panel of the calorie detector to start ignition. One cycle of an
ignition constitutes (T1) and (T2). (T1) is when the ignition voltage is applied to the heater
(T1). (T2) is when the voltage is not applied (when using low caloric value, the time when the preheating voltage is applied). The cycle repeats ve times normally. (T1) time and (T2) time are set independently within the range from approximately 2 to 20 seconds depending on the
measurement range and gas composition.
If the amplied thermocouple output exceeds the lower alarm limit, the burner is determined to
be ignited and the ignition action is stopped even if it is not completed. If the burner is not ignited
even after the ignition voltage is applied ve times, the ignition action is stopped after the ignition voltage is applied for the fth time. In that case, press START to start the ignition again.
If the burner ame is extinguished in the event of an alarm, after removing the cause of the failure, press START to start the ignition again. If the alarm occurs due to excessive combustion,
you cannot resume the ignition until the thermocouple output falls below the lower alarm limit (*1).
Indication lamps turn ON or OFF in accordance with each stage of the sequence. Figure 3.11 shows the ignition sequence for which the lower alarm limit is set to 50%.
3-9
Thermocouple output
(Higher alarm limit)
(Lower alarm limit)
Ignition voltage
EV
ALARM lamp
IGNIT lamp
MEAS lamp
Figure 3.11 Ignition Sequence
120%
50%
Hi Low Open Close
ON OFF ON OFF ON OFF
Ignition action starts
T1T2
Ignition action
Alarm status
Not ignition
Measurement status
Ignitied
Ignition action
Alarm status
Excessive combustion
Time
*1
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<3.Construction and Function>

3.5 Computing Station

The computing station generates WI signal by calculating calorie detector signal with each
di󰀨erential pressure signal. It compensates the density with the density signal and generates the caloric signal.
Each input signal is digitally computed after A to D conversion then D to A conversion. It generates DC 4 to 20 mA (DC 1 to 5 V) output.

3.5.1 Display

n Display indication
FAIL lamp
TREND 3
Y1: Calorific value (green)
Y2: WI value (light blue)
X3: MAX range of gas differential pressure(pink)
X1: MAX range of NON­ CORR. WI value(yellow)
X1: MIN range of NON­ CORR. WI value(yellow)
X3: MIN range of gas differential pressure(pink)
Unused
---CM6G---
TREND 3 Y1 Y2
C
<
18.00
X1
35.00
X3
1000.0
100.00
0.00
0.0
SHIFT
YOKOGAWA
90.00
500.0
37.67
23.02
14.65
23.02
FAIL
ALM
C
A
M
PF
O
<
F0516.ai
ALARM lamp
X1: NON-CORR.WI% (yellow)
X3: Gas differential pressure (Pa) (pink)
Y2: MAX range of WI (light blue)
Y1: MAX range of Calorific value (green)
Trend data (indicator 1 to 4)
Y1(green) : Calorific value Y2(light blue): WI X1(yellow) : NON-CORR. WI X3(pink) : differential gas pressure Pa
Y1: MIN range of Calorific value (green)
Y2: MIN range of WI (light blue)
The key for switching screens.
Press this key to go to ALARM screen. Push this again to return.
*1
3-10
Figure 3.12 Measuring Display
---CM6G---
ALARM
SYSTEM
PROCESS
* * * *
* *
STC EVENT
<
<
YOKOGAWA
X1
X2
X3
X4
Y1
CALC
SHIFT
FAIL
ALM
C
ALM
A
CLR
Left is ALARM screen you see when input or output is error. Do as *1 of Fiigure 3.12 to go to ALARM.
M
EVT
ON
EVT
PF
CLR
<
<
Figure 3.13 Alarm Display
ALARM items, such as X1 to X4,Y1 and Y2, are displayed.
F0517.ai
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<3.Construction and Function>
n Switching to alarm display
3-11
---CM6G---
TREND 3
Y1
18.00
Y2
35.00
1000.0
100.00
X3X1500.0
FAIL
ALM
90.00
C
37.67
23.02
ALARM
PROCESS SYSTEM
A
M
Push key
Push key
<
C
<
<
0.00
0.0
SHIFT
YOKOGAWA
PF
14.65
23.02
O
<
<
(1) Measuring display (2) Alarm display
Figure 3.14 Switching operation to alarm display
n ConrmationofInput/outputdata
---CM6G---
TREND 3 Y1 Y2
C
FAIL
1000.0
100.00
ALM
90.00
X1
500.0
X3
C
37.67
23.02
A
Push key
SHIFT
with key.
M
18.00
35.00
---CM6G---
TUNING MENU
PID 1
PID 2
STC 1
STC 2
P&T REG
PF
14.65
0.00
23.02
0.0
O
I/O DATA
FAIL
ALM
PF
C
A
M
---CM6G---
X1
*
X2
*
X3
*
X4
*
*
Y1
*
CALC
EVENTSTC
SHIFT
YOKOGAWA
Push key
FAIL
ALM
C
ALM
A
CLR
M
EVT
ON
EVT
PF
CLR
<
F0518.ai
PF
---CM6G---
I/O DATA
X1 X2 X3 X4 X5 X6 X7 X8
Y1 Y2 Y3 Y4 Y5 Y6
DI01
0
DI02
0
DI03
0
DI04
0
DI05
0
DI06
0
DI07
0
DI08
0
DI09
0
DI10
0
FAIL
ALM
90.00
500.0
C
500.0
0.500
18.00
-25.0
-25.0
A
-25.0
18.00
35.00
-6.3
0.0
M
0.0
0.0
DO01
0
DO02
0
DO03
0
DO04
0
DO05
0
DO06
0
DO07
0
DO08
0
DO09
0
DO10
0
DO11
0 0
DO12 DO13
0
PF
DO14
0
DO15
0
DO16
0
<
<
SHIFT
YOKOGAWA
<
<
<
<
SHIFT
YOKOGAWA
<
<
(1) Measuring display
Figure 3.15 Switching operation to Input/output status display
SHIFT
<
<
YOKOGAWA
<
<
F0519.ai
(3) Input/Output status display(2) Tuning display
This screen is for confirming Input/Output status and for setting X1,X2 and X3. Displayed characters become bigger by pushing key.
A
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<3.Construction and Function>
n Parameter setting
3-12
---CM6G---
TUNING MENU
PID 1
PID 2
STC 1
STC 2
P&T REG
I/O DATA
<
<
YOKOGAWA
FAIL
ALM
C
A
Push key
M
PF
SHIFT
<
<
---CM6G--- ---CM6G---
P&T REG 1/3
P01 P02 P03 P04 P05 P06 P07 P08 P09 P10 P11 P12 P13 P14 P15 P16 P17 P18 P19 P21 P22 P23 P24 P25 P26 P27 P28 P29 P30
<
<
YOKOGAWA
FAIL
ALM
60.00
2.500
C
0.00
100.0
7.00
100.0
7.00
A
5.00
0.0
0.0
0.0
0.0
0.0
M
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
PF
SHIFT
<
<
A
Push key
A
(Press key to enlarge characters of parameter to set.)
P&T REG 1/3
P01 P02 P03 P04 P05 P06 P07 P08 P09 P10 P11 P12 P13 P14 P15 P16 P17 P18 P19 P21 P22 P23 P24 P25 P26 P27 P28 P29 P30
<
<
FAIL
ALM
60.00
2.500
C
0.00
100.0
7.00
100.0
7.00
A
5.00
0.0
0.0
0.0
0.0
0.0
M
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
PF
SHIFT
<
<
YOKOGAWA
(1) Tuning display (2) Parameter setting display (3) Parameter setting display
Push key.
---CM6G---
TREND 3
18.00
Y2Y1X3X1500.0
35.00
C
<
YOKOGAWA
FAIL
ALM
90.00
C
1000.0
37.67
100.00
23.02
A
M
PF
0.00
14.65
0.0
23.02
O
SHIFT
<
(5) Measuring display
Push key
SHIFT
with key twice.
The screen is for setting various parameters such as ZERO and SPAN.
In setting other parameters
Parameter setting is finished. Return to the measuring display.
(4) Input/Output status display
This screen is for confirming Input/Output status.
Select a parameter to confirm. Press key to enlarge the character to display.
---CM6G---
I/O DATA
X1 X2 X3 X4 X5 X6 X7 X8
Y1 Y2 Y3 Y4 Y5 Y6
DI01 DI02 DI03 DI04 DI05 DI06 DI07 DI08 DI09 DI10
<
<
By using and key, change the parameter value which becomes larger characters.
After returning to Tuning display (1) by pushing key, push key.
90.00
500.0
500.0
0.500
18.00
-25.0
-25.0
-25.0
18.00
35.00
-6.3
0.0
0.0
0.0
0
0
DO01
0
0
DO02
0
0
DO03
0
0
DO04
0
0
DO05
0
0
DO06
0
0
DO07
0
0
DO08
0
0
DO09
0
0
DO10
0
DO11
0
DO12
0
DO13
0
DO14
0
DO15
0
DO16
SHIFT
YOKOGAWA
PF
FAIL
ALM
C
A
M
PF
<
<
F0520.ai
A
Figure3.16FlowChartofVariousParameterSettingandConrmation
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<3.Construction and Function>

3.5.2 Contents of the Data Display

Contents of the data display are as follows:
Table 6.2 Contents of the data display
Kind Data abbreviated Data contents Data range
Input
X
N
Output Y
N
Variable data
P
N
X1 NON-CORR. WI [%] WI value before correction 0.0 to 100.0
X2 A-PRESS [Pa] Air di󰀨erential pressure 0.0 to 1000
X3 G-PRESS [Pa] Sample gas di󰀨erential pressure 0.0 to 1000
X4 SQT. DENSITY A square root of the sample gas
density
X5 TC or LOW ALARM *2[mV] Thermocouple electromotive force or
ame extinction alarm threshold value
X6~X8 Unused
Y1
cal [MJ/
Nm
3
Caloric value *1
]
Y2 WI Wobbe index *1
Y3 Option Option *1
Y4 BIAS CHECK Preheating check output 0.0 to 100.0 Y5, Y6
Unused
P01 ZERO Zero adjustment 0.0 to 100.0
P02 SPAN Span adjustment *1
P03 BIAS Preheating adjustment 0.0 to 100.0
P04 A-CORR. RATE [%] Air di󰀨erential pressure signal
computing correction rate
P05 A-TIME [sec] Time constant of air di󰀨erential
pressure signal delay time
P06 G-CORR. RATE [%] Sample gas di󰀨erential pressure
signal computing correction rate
P07 G-TIME [sec] Time constant of sample gas
di󰀨erential pressure signal delay time
P08 PRESS ALARM SET [%] Di󰀨erential pressure warning setting 0.0 to 100.0
P09 to P30 Unused
3-13
*1
0 to 30.0
0.0 to 200.0
0.0 to 100.0
0.0 to 200.0
0.0 to 100.0
*1: Di󰀨ers depending on each specication. *2: Press “METER” to switch.
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<3.Construction and Function>

3.5.3 Correcting Computation

CM6G sets the di󰀨erential pressure of both sample gas (∆Pg) and air (∆Pa) to 500 Pa. In order to correct the indication error due to the variation of the di󰀨erential pressure (Flow rate), the
correcting computation is practiced.
The detection signal is obtained through the measurement with the standard di󰀨erential pressure of 500 Pa, but if each di󰀨erential pressure change to ∆Pg and ∆Pa (≠500 Pa), the detection
signal shall be changed from Eo to E’o.
3-14
E’o = Eo
Therefore, if you multiply the detection signal E’o by
Pg
Pa
/
, you can correct to the value
PgPa
(3.1)
at the standard di󰀨erential pressure.
In the computing program, each di󰀨erential pressure is root-extracted and provides a di󰀨erential
pressure correcting computation. The signal after the correction is range suppressed (ZERO)
and to further extended to SPAN. When it has a preheat circuit, deduct the amount of preheated value and extend to SPAN, then being range suppressed and shall be extended again to SPAN. Further as to generate the output of the caloric signal, the following density correction is
necessary:
ρ
K = C(WI)
g
(3.2)
Also it is necessary for each signal to operate with a timing matched with the signal of the caloric
detector and for this reason, dynamic characteristic function is provided. (A-CORR. RATE, A-TIME, G-CORR. RATE, G-TIME)
Remarks: In case of pre-heating circuit exists, the formula is as follows:
∆θ = C1
ρ
K
ρ
g
H : Pre-heat calorie
1, C2 : Constant
C
g : Density of the gas
Pg
Pa
+ C
H
2
Pa
(3.3)
If the correction computation of the di󰀨erential pressure is provided to ∆θ in the formula (6.3), it is
shown as per the following formula:
∆θ (3.4)
= C1
Pa
Pg
K
• ρ
g
+ C
2
H
Pg
The second clause of the right part in the formula (3.4) represent the preheat calorie, which, as you can see in the formula (3.4), the matters concerning the preheating is varied according to the
change of di󰀨erential pressure (∆Pg).
In this computing program, if there is preheating circuit, multiply the constant by the di󰀨erential pressure signal, and after correcting the di󰀨erential pressure to the preheating, the deduction of
the preheat calorie is computed.
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<3.Construction and Function>
3-15
(G-CORR. RATE)
(G-TIME)
Gas differential pressure signal
DYNAMIC CHAR.
FUNCTION
CORRECTION
Air differential
pressure signal
DYNAMIC CHAR.
FUNCTION
CORRECTION
(A-CORR. RATE) (A-TIME)
Extraction Extraction
Calorie detector signal
(ZERO)
(SPAN)
Correction of density
density
signal
Extraction
Remarks: ( ) is variable constant.
WI signal
Calorific signal
Figure3.17ComputingBlockDiagram(PreheatingCircuitDoesNotExist)
F0521.ai
Gas differential
pressure signal
(G-CORR.RATE)
(G-TIME)
DYNAMIC CHAR.
FUNCTION
CORRECTION
(BIAS)
Remarks: ( ) is variable constant.
Root Extraction
(SPAN)
(ZERO)
Air differential
pressure signal
DYNAMIC CHAR.
FUNCTION
CORRECTION
Root Extraction
(A-CORR.RATE) (A-TIME)
Calorie detector signal
Correction of density
Density
signal Root Extraction
Pre-heat temperature
check signal
Calorie signalWI signal
Figure3.18ComputingBlockDiagram(PreheatingCircuitExists)
F0522.ai
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<3.Construction and Function>

3.6 Density Meter

The GD400G is used for gas density meter.
Regarding GD400G Gas density meter, refer to attached Instruction Manual IM 11T3B1-01E.
3-16
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<4. Preparation of Operation>

4. Preparation of Operation

CAUTION
Before starting operation, check outside wirings or pipings and conrm if no gas leakage is found.
See Figure 3.2, 3.3 and 3.4.

4.1 Sampling Section (Outside panel)

If the drain pot is located outside the panel, supply water fully until the water overows from
the drain outlet.
Open the gate valve when it is in the sample line.
If two fulo lters are located in parallel on the sample line, one of them is not used.

4.2 Status of Valves

4-1
See the following table, Table 4.1 and Table 4.2 to open/close valves.
Table 4.1 For town gas use
Valve No. Open or Close Valve No. Open or Close
V-1 V-2 V-3 V-4 V-5 V-6 V-7 V-8
Remarks: Valves with the mark * are the ow adjustment use and preliminary adjusted. They don’t need to be adjusted.
Table 4.2 For steel mill use
Valve No. Open or Close Valve No. Open or Close
V-1 V-2 V-3 V-4 V-5 V-6 V-7 V-8
Close Open Open Close * Open Close Close
Close * * Close Close Open Close Close
V-9 V-10 V-11 V-12 V-13 V-14 V-15 V-15
V-9 V-10 V-11 V-12 V-13 V-14 V-15 V-15
Close Close Open Open * Close Open *
Close Close * * Close Close Open *
Remarks: Valves with the mark * are the ow adjustment use and preliminary adjusted. They don’t need to be adjusted.

4.3 Water Supply (For Steel Mill Use)

Open valve V-13 and supply water until it overows from the drain exhaust outlet of the
washing bubbler (3-2) and the pressure regulating pot (3-4). (Refer to Figure 3.4).
The owmeter (6-1) indicates the ow rate of approx. 0.2 l/min.
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<4. Preparation of Operation>

4.4 Supply of the Air

Open valve V-14 and supply the air. The normal pressure is as under mentioned:
Table 4.3
Pressure Gauge Normal pressure
Pressure gauge of air set Pressure gauge (2-3)
When the pressure is not normal, adjust it by the following procedures.
1) Adjust the pressure gauge of the air set to approx. 200 kPa, using the air set valve (2-1).
Turn the valve to the right to increase the pressure.
2) Adjust the pressure gauge (2-3) to about 20 kPa, using the pressure reducing valve (2-2)
and V-16. Turn the pressure reducing valve to the right to increase the pressure.
approx. 200 kPa approx. 20 kPa

4.5 Supply of Power

1) Set the positions of each switch on the front panel of the calorie detector to the followings:
Table.4.4
4-2
Switch Position
POWER
SAFE EV METER
2) Turn on the switches of the following components in the switch box.
• Calorie Detector
• Orice Assembly (NFB)
• Pre-heater (NFB)
• Computing station
• Density meter (With density compensation)
• Dehumidier (For steel mill use )
3) When you turn on the “POWER” switch on the front of the calorie detector, both POWER”
and “ALARM” lamps are lit to show an alarm status.
4) When the power is supplied to the computing station, the alarm lamp on the front (yellow) lit
on, but this does NOT indicate “out of order.”
5) Turn on “POWER” in gas density meter (for the one with a gas density meter attached)
6) Approx. 60 minutes later, the lamp of the orice assembly starts ashing and the
temperature of the orice constant chamber becomes stable.
OFF SET
P.MEAS
4.6 TheZeroAdjustmentoftheDi󰀨erential
Pressure Transmitter
1) Remove the impulse lines from transmitters.
2) Turn on the main breaker of the switch box and the calorie detector switch.
3 Turn on the POWER switch on the front panel of the calorie detector.
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<4. Preparation of Operation>
4) Measure each of the output terminals on the terminal block with a voltmeter. Use the “zero
adjustment screw” on the di󰀨erential pressure transmitter to adjust the voltage to 1 V.DC.
The terminal number assigned for gas is 15 and 16, and one for air is 17 and 18.
5) Use a at-head screwdriver to make the adjustment. Turn the screw clockwise to increase
the output, and counterclockwise to decrease. The adjustment amount of the zero point
varies depending on how quickly the zero adjustment screw is turned. To make a ne adjustment, turn the screw slowly, and to make a rough adjustment, turn the screw quickly.
6 When the adjustment is completed, connect the impulse lines while making sure there is no
leakage.
CAUTION
Be sure to make this adjustment every time the installation location of the panel is changed.
● Power is supplied to the di󰀨erential pressure transmitter from the calorie detector. To
operate the di󰀨erential pressure transmitter, turn on the calorie detector switch of the switch box and the POWER switch on the front panel of the calorie detector.
● After the zero-point adjustment is completed, do not turn o󰀨 the transmitter immediately. If
the power is turned o󰀨 within 30 seconds after the completion of the adjustment, the value
will return to the last one before the adjustment takes place.
4-3
Screw
Figure4.1Di󰀨erentialPressureTransmitter
F0401.aiZero Adjustment
4.7 AirDi󰀨erentialPressureAdjustment
1) Switch the computing station to the Input/Output status display (refer to Figure 3.15) and
check X2 (A-PRESS).
2) Adjust V-16 so that the indicated value becomes “500.”
3) If the indicated value of the pressure gauge (2-3) has deviated from 20 kPa, set the value to 20 kPa using the pressure reducing valve (2-2) and make the adjustment described in (2)
again.

4.8 Pressure Adjustment of the Gas Line

The pressure adjustment of the gas line uses air.
n For town gas use
1) Close V-6 and connect the one touch coupler (2-5).
2) Open V-7 and V-8.
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<4. Preparation of Operation>
3) Turn the switch “SAFE EV” of the calorie detector to “RELEASE”. The solenoid valve of the gas line opens and the alarm is released.
4) Adjust the owmeter (3-3) to approx. 10L/min by using the pressure reducing valve (4-1) and V-13. The pressure gauge (3-2) indicates 8 to 18 kPa.
5) Adjust the pressure gauge (3-5) to approx. 3 kPa. by using the pressure reducing valve (3-
4).
6)
When a density meter is attached, adjust the throttle valve of the owmeter (5-3) to 0.5 to 1 l/min.
7)
Set the indication of the computing station to Input/Output status display and check X3 (G-PRESS).
8) Adjust by pressure reducing valve (3-6) so that “X3” indicates “500”. When the value on the pressure gauge (3-5) deviates from about 3 kPa, repeat the procedure (5), then conduct the
adjustment.
9) When the adjustment is completed, close V-8, V-7, separate the one touch coupler (2-5), and open V-6.
10) Set the “SAFE EV” switch of the calorie detector to “SET”.
n For steel mill use
1) Close V-4, connect the one touch coupler (2-5) and open V-5 and V-8.
2) Set the switch “SAFE EV” of the calorie detector to “RELEASE”. The solenoid valve of the gas line opens and the alarm is released.
4-4
3) Adjust the ow rate of the owmeter (4-2) to about 10 l/min by using the pressure reducing valve (4-1). The pressure gauge (3-1) indicates approx. 6 kPa.
4) When the density meter is attached, adjust the owmeter (5-3) to 0.5 to 1 l/min by using
V-11.
5) Adjust, by using V-12, the number of bubbles that come out of pipe A of the pressure regulating pot (3-4). The number of bubbles has to be 3 to 6 pcs/sec. However, if no bubbles come out, when low calorie gas is being used, open V-12 fully,
6) Set the indication of the computing station to Input/Output status display and check X3 (G-PRESS).
7) Adjust the pressure to “500” by using the pressure reducing valve. “X3” the pressure reducing valve (3-6), adjust the indication “X3” to become “500”. When the number of bubbles from pipe A uctuates, adjust the number by using V-12, then carry out the
adjustment.
8) When completing the adjustment, close V-8 and V-5. Separate the one touch coupler (2-5), and open V-4.
9) Turn the switch “SAFE EV” of calorie detector to “SET”.
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<5. Operation>

5. Operation

5.1 Start Operation

WARNING
Conrm the air is supplied before starting operation. Conrm that the ventilation system is running before introducing the gas.
n Introduction of the Sample Gas
l For town gas use
1) Open V-1 fully.
2) When the pump is attached, turn ON the power supply before proceeding to (3).
5-1
3) When the pressure is increased by the pump, adjust the pressure by V-5 so that the
pressure gauge (3-2) indicates 8 to 18 kPa. When the pressure is reduced by the pressure
reducing valve, adjust the pressure by the pressure reducing valve (3-8) so that the
pressure gauge (3-2) indicates 8 to 18 kPa, .
l For steel mill use.
1) When you use the standard pressure, open V-4. When the pressure is increased by the pump, open V-1, V-4 respectively.
2) When the pump is attached, turn on the power supply switch of the pump at this time.
3) Use V-3 for standard pressure or use V-2 and V-3 for the pressure raised by pump to make
the pressure gauge indicate approx. 6kPa.
n Start Operation of the Calorie Detector
Conrm the switch “SAFE EV” is set at “SET” then push the switch “START”.
n Di󰀨erentialPressureReadjustment
When the burner is ignited and the output becomes stable (after 20 to 30 min.), adjust again the
di󰀨erential pressure.
l For town gas use
1) When the pressure gauge (3-5) does not indicate approx. 3 kPa, adjust the pressure by the
pressure reducing valve (3-4).
2) When the density meter is attached and the owmeter (5-3) indicates out of 0.5 to 1 l/min,
adjust its indication again.
3) Set the computing station to Input / Output status display. If the indication of “X2 (A-PRESS)” and “X3 (G-PRESS)” is not ”500”, adjust each with the pressure reducing valve (3-6) and
V-16.
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<5. Operation>
l For steel mill use
1) When the density meter is attached, adjust again the indication of the owmeter (5- 3) by
V-11.
2) When the number of bubbles coming out from the pipe A is not constant, adjust it by V-12.
3) Set the computing station to Input / Output status display. If either the indication of “X2 (A-PRESS)” or “X3 (G-PRESS)” is not ”500”, adjust each with the pressure reducing valve
(3-6) and V-16.

5.2 Stopping Operations

n Long Time Stopping
1) Close V-1. To stop standard pressure gas for the steel mill use, close V-4. But when a pump
is attached, turn o󰀨 the power of pump in advance.
2) When the gas supply is suspended and the burner is extinguished, the lamp “MEAS” of the calorie detector is lit o󰀨 and the lamp “ALARM” on.
3) Switch the calorie detector from “SAFE EV” to “RELEASE”.
4) To stop the ow of a town gas , close V-6 and open V-7, V-8. To stop the ow of steel mill gas, close V-4 and open V-5, V-8. The sample gas in the gas line should be blown o󰀨 by the
air then the subtle sample gas will be emitted
5-2
5) After continuing the blow for 3 to 5 min., turn o󰀨 the “POWER” switch of the calorie detector.
If the density meter is attached, cut the power supply inside the density meter.
6) Turn o󰀨 all switches of the switch box.
7) Lastly, close V-14 and stop the air supply.
n Short Time Stopping
Same as mentioned above, the mode suspends the supply of sample gas only.
If this state is kept on hold, you can skip the warm-up time before another operation starts.
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<6. Calibration>

6. Calibration

For calibration, zero and span gases are necessary. Use the gas with the specications as near
to each measuring range for both lower and upper limit as possible.

6.1 Supply of the Calibration Gas

6.1.1 For Town Gas Use

Refer to Figure 3.2 and 3.3.
1) Close V-6 and open V-7. At this time, the ame of burner is extinguished. But if the pump is attached, turn o󰀨 the power supply of the pump rst, then close V-6 and open V-7.
2) Adjust the secondary pressure of the pressure reducing valve for calibration gas cylinder, to
approx. 200 kPa.
3) Open V-9 or V-10 and introduce the calibration gas.
4) Adjust the indication of the owmeter (3-3), by the pressure reducing valve (4-1), to become about 10 l/min. In case of only indication check, the adjustment to approximately 2 to 3 l/min
is permissible.
5) Push the “START” switch of the calorie detector to ignite.
6-1
6) Set the computing station to Input/Output display screen. If the indication of “X2 (A-PRESS)” and “X3 (G-PRESS)” is not ”500”, adjust each with the pressure reducing valve (3-6) and
V-16.
7) In the case of the density meter being attached, conrm and readjust (if necessary) the indication of the owmeter (5-3).
8) When the calibration is completed, close V-9, V-10, V-7, and open V-6. When a pump is attached, turn on the power of pump now.
9) Push the “START” switch of the calorie detector to ignite.
10) Conrm and readjust (if necessary) the gas di󰀨erential pressure and the ow rate to the
density meter.
11) Do not forget to close the root valve of the calibration gas cylinder.
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<6. Calibration>

6.1.2 For Steel Mill Use

Refer to Figure 3.4 Flow Sheet (Steel Mill Use).
1) Close V-4 and open V-5. The burner is extinguished. However, if the pump being attached, turn o󰀨 the power supply of it in advance.
2) Set the secondary pressure of the pressure reducing valve for calibration gas cylinder to
approx. 200 kPa.
3) Open V-9 or V-10 and introduce the calibration gas.
4) Using the pressure reducing valve (4-1), adjust the indication of the owmeter (4-2) to become about 10 l/min. In case of only indication check, the adjustment to approximately 2
to 3 l/min is permissible.
5) Push the “START” switch of the calorie detector to ignite.
6) Set the computing station to Input/Output status display. The indication of “X3 (G-PRESS)”
is adjusted to ”500” with the pressure reducing valve (3-6). In the case of the density meter
being attached, conrm and readjust the ow rate.
7) Conrm “X2 (A-PRESS)” in Input/Output status display and adjust “X2” to “500” by using
V-16.
8) When the calibration is completed, close V-9, V-10, V-5 and open V-4. When the pump is attached, turn on the power of the pump.
6-2
9) Push the “START” switch of the calorie detector to ignite.
10)
Readjust both the gas di󰀨erential pressure and the ow rate to the density meter.
11)
Do not forget to close the root valve of the calibration gas cylinder.

6.2 Span Adjustment of the Calorie Detector

This adjustment is not necessary for the usual calibration. This adjustment is held when replacement such as the thermocouple of the burner unit takes place. This adjustment is
di󰀨erent depending upon if the preheating circuit exists or not. This adjustment is to be held after regulating each di󰀨erential pressure to “500”.
n WhennoPreheatingCircuitExist
1) Set the computing station to Input/Output status display with X1 (NON-CORR. WI%).
2) Introduce the span gas and after ignition, wait until the indication “X1” becomes stable.
3) Adjust the indication, using the “SPAN” potentiometer on the front side of the calorie detector, to become the specied value (x%). WI is used to nd out the x value. Example: WI value of the span gas = 59.00 WI WI value of the upper range = 60.00 WI
59.00
x = 100 = 98.3%
60.00
n WhenPreheatingCircuitExists
1) Suspend the supply of the sample gas.
2) Same as the gas line pressure adjustment explained in paragraph 4.8), introduce the air to the gas line.
3) Set the “SAFE-EV” switch of the calorie detector to “RELEASE”, when the preheating voltage is applied to the ignition heater of the burner unit, by which the increased
temperature with an e󰀨ect of the preheating is detected by the thermocouple.
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<6. Calibration>
4) Switch the “METER” switch of the calorie detector to “P.MEAS” and measure X5 of the
computing station. Wait until the indicated value is stabilized and then record the value. (E0 mV)
5) Switch the “SAFE EV” switch of the calorie detector to “SET.”
6) Stop the air supply to the gas line, supply the span gas, and start the ignition action.
7) Measure X5 of the computing station. Wait until the indicated value is stabilized and then
record the value. (Ec mV)
8) Set the computing station to Input/Output status display and adjust the “SPAN” potentiometer of the calorie detector so that “X1” will show the specied value (x%). WI is to be used to nd out the value of x.
For example:
WI value of the upper range = 5.00 WI WI value of the span gad = 4.90 WI Eo = 4 mV, Ec = 19 mV
x = 100 = 98.4 %
(19-4) x
19
5.00
4.9
100
x
+ 4
Ec
6-3
NON-CORR. WI [%]
WI value of the span gas: Wc WI value of the upper range: Ws
x =
(Ec – Eo) + Eo
Figure6.1HowtoFindOutxWhenPreheatingCircuitExists
Ec
Ws Wc
100
output (mV)
Thermocouple
Eo
0
WI value of the span gas
WI value of the upper range
WOBBE INDEX [WI]
F0703.ai
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<6. Calibration>

6.3 Calibration of the Computing Station

This chapter explains an ordinary calibration of the computing station. The calibration is
conducted as follows according to the type of the output.
• WI output .............. Calibrate with WI value.
• WI calorie output .. Calibrate with WI value and conrm the caloric value.
• Calorie output ....... Calibrate with caloric value.
In case of existing the preheating, the bias adjustment is carried out rstly.
n Bias Adjustment
This adjustment is carried out only when the preheating circuit exists.
1) Suspend the supply of the sample gas, and introduce the air to the gas line.
2) Set the “SAFE EV” switch of the calorie detector to “RELEASE”.
3) Set the computing station to Input/Output status display. After the indication Y4 (BIAS
CHECK) becomes stable (20 to 30 minutes), adjust the (Y4) value of the P03(BIAS) will
become [0.0] by the data setting key.
Note) When you perform a ne adjustment on a potentiometer, select alternatively between (Y4) and (P03) and display
4) When the adjustment is completed, turn the switch “SAFE EV” of the calorie detector to “SET”, and suspend the air supply to the gas line.
6-4
n Zero and Span Adjustment
1) Introduce the zero gas and ignite.
2) Set the computing station to Input/Output status display. Adjust, with the “P01 (ZERO)” of
the operator, to be the “Y2 (WI)” or “Y1 [MJ/Nm calibration gas).
3) Change from zero gas to span gas, when, if let the span gas start owing before the zero gas valve is not completely closed, the burner is not extinguished. Same in the case of the other way. If burner is extinguished, ignite.
4) Set the indication of the computing station to Input/Output status display. After the indication
“Y2 (WI)” or “Y1 (cal)” becomes stable, adjust with the “P02 (SPAN)” of the operator to become WI value or the caloric value of the calibration gas. (Display P02, adjust by the data setting key). (Conrm the result of the adjustment after displaying it as WI or cal)
5) Repeat 2 or 3 times the above adjustments.
6) When complete the adjustment, suspend the supply of the calibration gas.
n Other Adjustment
The adjustment of P04 (A-CORR, RATE), P06 (G-CORR, RATE) and P05 (A-TIME), P07
(G-TIME) is usually unnecessary. (Reserve the record of the data in the initial stage).
But when replacing the thermocouple, the regular adjustment is required. For the adjustment in
detail, consult our service personnel.
3
]” (WI value or the caloric value of the

6.4 Calibration of the Density Meter

Regarding GD400G Gas density meter, refer to an attached Instruction Manual IM 11T3B1-01E.
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<7. Maintenance>
F11-1.ai

7. Maintenance

7.1 Daily Check

7.1.1 Air,GasDi󰀨erentialPressureAdjustment
Conrm the air or gas di󰀨erential pressure or adjust the indication of X2 (A-PRESS), X3 (G-PRESS) to “500”. Refer to the paragraph 4.7 or 4.8 then follow the instruction on the
adjustment.

7.1.2 Take Out the Water Out of the Drain Pot (Pump for Steel Mill Use)

If the drain is accumulated, open V-7 and take it out. After the drain is taken out, be sure to close the valve.

7.1.3 Take Out the Drain from the Air Set

Turn the knob at the bottom of the air set and take out the drain. After the drain is taken out, be sure to turn the knob to close tightly.
ADJUSTMENT KNOB
LOCK NUT
7-1
GAUGE FOR SECONDARY PRESSURE
DRAINAGE KNOB
Figure 7.1 Air Set
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<7. Maintenance>

7.2 Regular Check

Refer to Figure 3.2 , 3.3 and 3.4. The ow sheet depends on the specication.
7.2.1 CleaningoftheOricePlateandReplacementof
O-ring
When something like dust adhere to the orice holes, the output declines. Follow the instruction as below to clean.
CROTCHET JOINT
7-2
HOLDING SCREW
ORIFICE ASSEMBLY
O-RING (P16)
O-RING (P20)
ORIFICE PLATE
COVER PLATE
HEX. HOLE SCREW
F10-2.ai
Figure7.2OriceSystem
1) Suspend the supply of the sample gas, then blow the gas line with the air.
2) Remove the crotchet joint.
3) Turn the holding screw and take out.
4) Pull out slowly to the direction as shown on Figure 10.2.
5) Loosen the set screw by using the hexagonal wrench attached to take out the orice plate together with the cover plate. The following is part number of the orice plate.
Oricehole(Ø) Part number Oricehole(Ø) Part number
0.35 G7025XJ 0.80 G7034XJ
0.40 G7026XJ 0.85 G7035XJ
0.45 G7027XJ 0.90 G7036XJ
0.50 G7028XJ 1.0 G7037XJ
0.55 G7029XJ 1.1 G7038XJ
0.60 G7030XJ 1.2 G7039XJ
0.65 G7031XJ 1.3 G7040XJ
0.70 G7032XJ 1.4 G7041XJ
0.75 G7033XJ 1.5 G7042XJ
6) For the cleaning use something like a supersonic cleaner and never insert into the orice
holes anything like a stick or rod.
7) When the O-rings are worn out, replace them with the spare parts. Part number of O-rings are Y9114XB (P16) and L9817MT (P20).
8) After the cleaning and replacement are completed, assemble the parts in the reverse order of the disassembly as mentioned above.
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<7. Maintenance>
NOTE
The orice system is temperature-controlled by a plate heater, so better nish the cleaning and
replacement as swiftly as possible.
7.2.2 FuloFilter(ForSteelMillUse)
The material of the element is polypropylene. The cleaning and the replacement of the element
shall be carried out as follows. Same process applies to the option fulo lter whose element (50µm) is G7057XJ in the part number.
1) Turn the nut and remove the cover.
2) Pull out the element and clean or replace.
3) Assembly is carried out in contrary order to above.
4) When the packing or gasket is worn out, replace it. We provide a packet which contains 4
pieces of packing and a gasket. (G7086XL)
PACKING GASKET ELEMENT PACKING
7-3
NUT
COVER
Figure7.3FuloFilter
TENTION BOLT
CASE
F0703.ai

7.2.3 Line Filter (For Town Gas Use)

Disassemble 3 screws, and open the cover, there inside exist an element (G7005XJ.) If it is stained, replace by a new one. Also if the O-ring (Y9116XB) is fatigued, replace it with new one.
SCREW
O-RING
GAS INLET
FILTER
ELEMENT
ELEMENT
GUIDE
BODY
Figure 7.4 Line Filter
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<7. Maintenance>

7.2.4 Washing Bubbler or Pressure Regulating Pot

If inside of case becomes dirty by the weeds grown in the water, pull the plug at the bottom of the
bubbler to drain and rell it with clean water. If inside the bubbler is stained, remove the cover and
clean inside of the bubbler. Be careful not to scatter the water around the panel when you drain the water.
When the water level becomes higher than the drain outlet position, there would be a possible
contamination in the exhaust pipe line, so the cleaning is necessary.
7-4
1
4
Figure 7.5 Washing Bubbler and Pressure Regulating Pot
7.2.5 Dehumidier(ForSteelMillUse)
The washing bubbler cannot eliminate all of the naphthalene if it is contained too much in a
sample gas. The naphthalene may be crystallized and clog the piping inside of dehumidier. In such a case, stop the sample gas ow, remove the 4 pipes at the top of the dehumidier and
introduce hot water of 70°C to 80°C. The naphthalene is dissolved into the hot water and is eliminated.
3
2
F12-6.ai
PRESSURE REDUCING VALVE
FULFLO FILTER
POWER
COOL
TERMINAL BLOCK
CIRCUIT BREAKER
POWER LAMP
COOL LAMP
DRAIN (TO PRESSURE REGULATING POT)
(ORIFICE ASSEMBLY)
DENSITY METER
Figure7.6Dehumidier

7.2.6 Density Meter

It is recommended to replace the O-ring of the GD300S detector every 2 or 3 years and wash the sensor if necessary. Consult to our service for washing the sensor.
F11-7.ai
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<7. Maintenance>

7.3 Check at the Regular Service

7.3.1 Check the Burner Flame

Whether the sample gas is completely burnt or not shall be judged by observing the shape and
the color of the burner ame.
• The shape of the ame for the complete combustion is as shown in Figure 7.7 - (A) of a shape sharply outlined, and the height of the inner ame is about a half size of the outer ame.
• When the primary air is too much, the ame becomes, as shown in Figure 7.7 - (B), ickering and when the secondary air is not su󰀩cient, the ame outline becomes like (B) as
well.
• When the primary air is not su󰀩cient, the color of the ame is clear yellow (C) and
sometimes soots come out.
• When there is any blockade or leakage is the air line, the ame becomes like the shape as
shown in Figure 7.7 - (D).
OUTER FLAME
INNER FLAME
7-5
A B C D
F11-8.ai
Figure 7.7 Burner Flames
Check the ames of the burner according to the following procedures. As to the adjustment of the squeezing screw of the primary air and the secondary air and the cleaning of the air distribution
part, ask to our service personnel.
1) Remove the screw (4 pcs) xing the external pipe, then remove the connectors to the
terminal of the hot junction detecting point.
2) Disassemble the external pipe and take out the jet plate by pulling upward. (The direction to
pull for removal is shown in Figure 7.8 by the sign of arrow).
3) Insert the attached mirror herewith into the ventilator.
4) When you ignite, the ame reected on the above mirror can be observed from slanting
upside.
5) Assemble the burner by following the procedures above in the reverse order.
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<7. Maintenance>
Connector
Detector for
Hot Junction
Exhaust Pipe
Terninal Board
7-6
Detector for
Cold Junction
Burner Unit
Ventilator
Secondary Air
Throttle Screw
Primary Air
Throttle Screw
Figure 7.8 Flame Checking
Screw(4pcs)
Mirror
Jet Plate
Combustion Pipe
Ignition Trans Assembly
Relay ( A1574MR)
External Pipe
Constant-Voltage Power Supply Transformer
F0708.ai
7.3.2 ZeroAdjustmentoftheDi󰀨erentialPressure
Transmitter
Adjust according to the procedure described in paragraph 4.6.

7.3.3 Others

If necessary, conduct each regular check mentioned in paragraph 7.2.
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<8. Troubleshooting>
8-1

8. Troubleshooting

Two major factors are considered to make large di󰀨erence between current measuring result and normally expected ones. One factor is that the process condition is changed, which leads to the
change of gas component . The other one is that the measuring system has some errors. If the measuring system has some errors, carry out the following troubleshooting in this chapter.

8.1 Gas Sampling Pressure Regulating Section

n For Town Gas Use
1) See Figure 3.2 and 3.3 Flow Sheet. Check if any blockade or leakage exist in the sampling
line up to the panel.
→If any blockade is found, blow the sampling line by air.
2) Check the lter element of the line lter. →According to the paragraph 7.2.3, clean or replace.
3) If conditions above mentioned are normal, the pressure gauge (3-2) indicates 8 to 18 kPa. →When the pump is attached and if the pressure does not indicate normal value even
after opening V-5 fully, the decrease of the suction ability of the pump might be a cause of the trouble.
→If the pressure reducing valve is attached, and the indication of the pressure gauge
(3-2) does not change even after the operation of the pressure reducing valve (3-8), the pressure reducing valve may have a defect.
4)
When introducing the check gas, conrm the indication of the owmeter (3-3) is about 10 l/min. If you need to check only an indication, adjust the owmeter to approximately 2 to 3 l/min. →It is due to a defect of the pressure regulator that the ow rate cannot be successfully
set by the pressure regulator (4-1).
5) Conrm the pressure gauge (3-5) indicate approx. 3 kPa. →If the pressure cannot be satisfactorily set by the pressure reducing valve (3-4), it may
have a defect.
6)
When the density meter is attached, conrm the indication of the owmeter (5-3) is 0.5 to 1 l/min. →It may be due to the clogging of the gas pipe inside the density meter that no indication appears on the owmeter.
7) Conrm the indication of the computing station is about “500”, after setting the indication to “X3 (G-PRESS)”. →When the di󰀨erential pressure setting cannot be executed with the pressure reducing
valve (3-6), it is due to a defect of the pressure reducing valve.
8) For all gas lines, check any leakage at each joint connection.
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<8. Troubleshooting>
n For Steel Mill Use
See Figure 3.4 Flow Sheet. The sampling system here (outside the panel) refers to the one we recommend.
1) Check if the water level of the drain pot (11-1) equipped outside the panel is higher than 15 kPa.
2) Check for any clogging, accumulation of dust or leakage in the pipings between the probe
(7-1) and the fulo lter (7-2) equipped outside. →When there is any clogging, blow by the air. →When the piping has any U bending, tilt the piping so that the water ows smoothly.
3) Check the element of the outside fulo lter (7-2). →When you nd any clogging, clean or replace as according to the paragraph 7.2.2.
4) Check the indication of the pressure gauge (3-1) is approx. 6 kPa. → If there is no water accumulated in the washing bubbler, no indication appears on the
pressure gauge. Supply water to the washing bubbler.
→ If the pump is attached, and its suction ability decreases,the pressure gauge may not
indicate the normal value after the adjustment by only V-2 and V-3.
5) When introducing the calibration gas, check the indication of the owmeter (4-2) is approx. 10 l/min.
→ When the pressure valve (4-1) cannot adjust it, there is a high chance that the pressure
reducing valve has a defect.
8-2
6) Check the element of the fulo lter (3-3). →If it is stained, clean or replace as according to the paragraph 7.2.2.
7) Check the water level of the pressure regulating pot.
→If necessary, supply water or clean.
8) Check if the bubbles are coming out 3 to 6 pcs/sec from the pipe A of the pressure regulating pot.
→ When V-11, V-12 fail to adjust, clean the piping inside the dehumidier as according to
the paragraph 7.2.5. But no bubbles are coming out, when you use low calorie gas.
9)
When the density meter is attached, check if the indication of the owmeter (5-3) is 0.5 to 1 l/min.
10) Check if the indication of the computing station is “500” when setting the indication of “X3 (G-PRESS)”. → When the adjustment is impossible even by using the pressure reducing valve (3-6), there
is a high chance that pressure reducing valve has a defect.
11) For all gas lines, check the leakage at each joint connection.

8.2 Air Pressure Adjustment Section

1) Check the pressure gauge of the air set (2-1) is about 200 kPa → When the primary pressure of the instrument air is normal, but the adjustment of the air
set is impossible, there is a high chance that the air set has a defect.
2) Check the pressure gauge (2-3) indicates approx. 20 kPa. → If you cannot adjust it when even using the pressure reducing valve (2-2), there is a high
chance that the pressure reducing valve has a defect.
3) If the above are found normal, we can adjust by using V-16 the air di󰀨erential pressure to
“500”.
8.3 Di󰀨erentialPressureTransmitterSection
1) Check the orice section of the orice assembly. →When the orice plate is contaminated, clean it as according to the paragraph 7.2.1.
IM 11R02A01-02E 5th Edition :Dec. 28, 2016-00
<8. Troubleshooting>
2) Check any leakage in the pressure transmission pipe of each di󰀨erential pressure
transmitter.
3) Check the temperature (thermistor resistance value) setting of orice assembly. →If the resistance value is not 3.6 to 4.2 kΩ, the temperature controller may have a failure.

8.4 Signal Section

Measure the output of each components by using digital voltmeter.
1) Set the “METER” switch on the front panel of the calorie detector to “P.MEAS” and measure X5 of the computing station. If preheating is not required, the appropriate output is approximately 0 to 20 mV, which is nearly proportional to 0 to (WI value, namely the highest value in the measurement range). If preheating is required, the appropriate output is approximately 4 to 20 mV, which is nearly proportional to 0 to (WI value, namely the highest
value in the measurement range).
For example:
When the WI value of the highest level in the measuring range is 100 WI, if the sample gas is 80 WI, the voltage output is about 16 mV (when no preheating circuit).
→Although both the gas pressure control section and the di󰀨erential pressure are normal,
the voltage output might be much varied. In this case there is a high chance that some
blockade exist in the air ow distribution line of the burner unit. Check the burner ame as
mentioned in paragraph 7.3.1.
8-3
→When the gas ame is coming out of the burner but the voltage output is very low , there
is a high chance that the wire is disconnected in thermocouple.
2) Measure 3 (+) and 4 (-) on the terminal board of the calorie detector. The normal voltage
output, when preheating circuit does not exist, is 1 to 5 V almost in proportion to 0 to (WI value, namely the highest value in the measurement range). If preheating circuit exist, approx. 1.8 to 5 V is almost in proportion to 0 to (WI value, namely the highest value in the
measurement range).
3) Measure 17 (+) and 18 (-) on the terminal block of the calorie detector. The appropriate value is 23 V ±1%.
→It is due to a defect of the distributor in the calorie detector that the value is not normal.
4) Measure 15 (+) and 16 (-) on the terminal block of the calorie detector. The appropriate value is 23 V ±1%.
→It is due to a defect of the distributor in the calorie detector that the value is not normal.
5) Measure 8 (+) and 7 (-) on the terminal block of the calorie detector. The appropriate value is 3 V ± 0.2 V.
→If the measured value varies much in spite of the air pressure control section being
normal, there is a high chance that the di󰀨erential pressure transmitter has a defect.
6) Measure 6 (+) and 7 (-) on the terminal block of the calorie detector. The appropriate value is 3 V ± 0.2 V.
→ If the measurement value is greater than the appropriate value, even though the gas
pressure controller is normal, the di󰀨erential pressure transmitter is considered to have a
defect.
7) When the density meter is attached, measure 3 (+) and 4 (-) of the density meter terminals. Compare the gas density meter and the voltage output, then obtain an appropriate value , while reviewing the characteristics of gas density meters.
. →If the appropriate value is out of data in spite of no leakage, calibrate as according to the
paragraph 6.4.
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<8. Troubleshooting>
8-4
CALORIE DETECTOR
DIFFERENTIAL PRESSURE TRANSMITTER (GAS)
DC 4~20mA
(+)
S
(–)
DC 4~20mA
(+)
S
(–)
DIFFERENTIAL PRESSURE TRANSMITTER (AIR)
15(+) 16(–)
17(+) 18(–)
DENSITY METER
Figure 8.1 Signal Circuit Diagram
1 2 3 4 5 6 7 8 9
3(+) 4(–)
5(+)
6(+) 7(–)
8(+)
3(+)
4(–)
DC 1~5V
DC 0~20mA
DC 1~5V
DC 1~5V
DC 1~5V
250 Ω
COMPUTING STATION
1(+) 2(–)
20(–) 19(+)
22(+) 23(–)
5(+)
24(+)
6(–)
25(–)
4(–)
26(+)
3(+)
27(–)
7(+)
8(–)
DC 4~20mA
DC 1~5V
DC 1~5V
CALORIFIC OUTPUT
WI OUTPUT
OPTIONAL OUTPUT
10 11121314 15 16 17 18
19 20 212223242526 27
Figure 8.2 Terminal Block of the Calorie Detector
IM 11R02A01-02E 5th Edition :Dec. 28, 2016-00
<8. Troubleshooting>

8.5 Computing Station

If the value is all normal, each input of the computing station shows the following normal values in the table.
Mark of input data Normal indication
NON-CORR. WI % *
A-PRESS Pa 500 ± 50
G-PRESS Pa 500 ± 50
SQT. DENSITY Same as density meter
*: When pre-heating circuit does not exist, 0 to 100% is almost in proportion to 0 to “the highest level WI value in the measurement
range.”
When preheating exists, about 20 to 100% is almost in proportion to 0 to “the highest level WI value of the measurement range.”.
The lamps of both alarm and fail of the computing station are lighted in the following case.
1) Lighting of the alarm lamp. (Yellow color)
It lights when the input or output signal is cut o󰀨. But in this case, the computing inside the station keeps working. Change to Alarm display and examine the cause of the alarm.
2) Lighting of the fail lamp (Red color) The lighting of the fail lamp means an occurrence of an abnormal trouble produced inside the instrument. When a fail occurs, the analog output and status output reserve the value just before the occurrence of the failure. Such hold output has a tendency of decreasing gradually as time proceeds. When the fail lamp is on, the display of fail will be shown. The alarm lamp lights for the abnormal operation of the pressure adjustment for the gas or air line. The lighting when other than under the time of measurement is not abnormal.
8-5

8.6 Other Troubleshooting

Check and deal with measures mentioned in paragraph 8.1 to 8.5. As to the trouble not mentioned in this manual, contact our service personnel and describe how the inspection goes, and if you have any record..
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
Blank Page
<Revision>

Revision Information

l Title : Model CM6G Gas Calorimeter
l Manual No. : IM 11R02A01-02E
May 2019/6th Edition
Corrections (p. 1-2. 1-3)
Dec. 2016/5th Edition
Full revision. Changed chapter order. Deleted contents described in the former Chapter 12 of and
CMPL.
Oct. 2011/4th Edition
All over revised Chapter 2 Subsection 2.1.1 :”Town Gas Application,” Alarm Contact Output: Changed of
description. Added the item of “Contact Input”. Panel: Changed of panel color.
Subsection 2.1.2, “Town Gas Application,” Range: Changed of description. Alarm
Contact Output: Changed of description. Added the item of “Contact Input”.
Panel: Chaned of panel color.
Subsection 2.2.1, “Gas Calorimeter”: Changed of style code.
Nov. 2008/3rd Edition
Revised with the change of style code *B Chapter 2 Subsection 2.1.1 :Rated current of alarm contact output is corrected to 3A. Subsection 2.1.2 :Description about Range is revised. Subsection 2.2.2 :Change to style *B in MS code table
Subsection 2.3.1 :Su󰀩x code “-S3610” is corrected to “-S6310”. Section 2.4 :The item “Hexagon Wrentch” is deleted from table list of calorie detector
standard accessory.
Section 2.5 (page2-6 to 2-8): Drawings of Computing Station change to Model
YS1700 Subsection 2.5.2 :In item 1, length of H7800HC is corrected. Chapter 3 Section 3.1 :Description about “conditioned air” is added Section 3.2 :Explanation about “piping” is added to body. Thermal insulation line and
steam line is added to Figure 3.1 . Section 3.3: In Figure 3.4, drawings of Computing Station change to Model YS1700 Chapter 5 In Figure 5.1 and 5.5, drawings of Computing Station change to Model YS1700 Subsection 5.5.1: Complete revision (Application of Model YS1700 operation)
Subsection 5.5.2: In table 5.3, items X6 to 8, Y5 to 6 and P08 to 30 are added Subsection 5.5.3: Equation (5.2) and (5.3) are revised. Figure numbers of 5.17 and
5.18 are changed.
Chapter 6 Explanation about computing station operation is revised in; subsection 6.1.7 a)&b),
6.1.8 town gas g)&h), steel mill f)&g) and 6.2.3 both of c)
Range of ow rate is expanded (0.5-1 l/min) in;
subsection 6.1.8 town gas f), steel mill d), and 6.2.3 town gas b)
Chapter 7 Explanation about computing station operation is revised in; subsection 7.1.1 f), 7.1.2
f)&g), 7.2.1 a)&b), 7.2.2 h), 7.3.1 c), 7.3.2 b)&d)
Description about “indication check” is added to; subsection 7.1.1 d) and 7.1.2 d) Title is changed and description with “See also” is inserted to subsection 7.1.1 d) and
7.1.2 d).
Subsection 7.3.3: P05 is corrected to P02. Chapter 9 Explanation about computing station operation is revised in subsection 9.1.1 (7) and
9.1.2 (10).
Range of ow rate is expanded (0.5-1 l/min) in subsection 9.1.1 (6) and 9.1.2 (9).
Description about “indication check” is added to 9.1.1 (4).
Subsection 9.1.2: Unit “Pa” is corrected to “kPa”.
Section 9.4: In item 2), The value “200” is corrected to “20” Section 9.4: In Figure 9.1, alphabet marks of the computing station terminal are
corrected to number marks. No.19(+) and No.20(-) are written in the station
terminal, and connected to No.1 and No.2 of the calorie detector,
respectively.
Terminal Numbers of distoributor are revised in the body and Figure 9.1.
i
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
<Revision>
Section 9.5: Item 1) is revised. 2) is deleted. 3) is partially changed and printed as
item 2).
Chapter 10 Subsection 10.1.1 : In item 8, part No. is corrected to “Y9107XB.”
Aug 2007/2nd Edition
All over revised
Jun.2009/1stEdition
Newly published
ii
IM 11R02A01-02E 1st Edition :Dec. 28, 2016-00
Loading...