Thank you for purchasing Model CM6G Gas Calorimeter.
In order to have the product deliver its full capabilities, read this instruction manual thoroughly
before you use it.
On how to use the Gas density meter, read the instruction manual which comes with the product.
WARNING
Since the ignition starts until it completes, or right after the ame is extinguished, sample gas
mixed with the air will be released from the top of the equipment into the installation space. Even
though the sample gas will be diluted suciently by the air, great attention should be paid to the
following points if the sample gas contains harmful gases such as CO.
• Conrm that the air is supplied before introducing the sample gas.
• Use the equipment in a well-ventilated environment equipped with a ventilation system.
• Do not expose your face above the top of the calorie detector.
<INTRODUCTION>
i
n About This Manual
• This manual should be passed on to the end user so that the user can refer to the manual
whenever they need.
• If the product is not used in a manner specied in this manual, the safety of this product may
be impaired.
• The contents of this manual are subject to change without prior notice.
• The contents of this manual shall not be reproduced or copied, in part or in whole,without
permission.
• This manual explains the functions contained in this product, but does not warrant that they
are suitable for the particular purpose of the user.
• Every eort has been made to ensure accuracy in the preparation of this manual.
However, when you realize mistaken expressions or omissions, please contact the nearest
Yokogawa Electric representative or sales oce.
n Drawing Conventions
Some drawings may be partially emphasized, simplied, or omitted, for the convenience of
description.
n Inspection on delivery
After the delivery, open the product package carefully and inspect for damages caused during the
transport. Should there be any damage or breakage on the product, please contact YOKOGAWA
immediately. Retain all the packing materials, containers and boxes that came with the product.
After unpacking, make sure the delivered products include all of the component equipment you
ordered. When conrming the specications, refer to the model and sux codes indicated on the
nameplate on the product. The models are indicated on the name plates. Check also those of
accessories. For further information, please read subsection 1.2 of this document.
• In order to protect the system controlled by the product and the product itself and ensure
safe operation, observe the safety precautions described in this user’s manual. We assume
no liability for safety if users fail to observe these instructions when operating the product.
• If this instrument is used in a manner not specied in this user’s manual, the protection
provided by this instrument may be impaired.
• If any protection or safety circuit is required for the system controlled by the product or for
the product itself, prepare it separately.
• Be sure to use the spare parts approved by Yokogawa Electric Corporation (hereafter
simply referred to as YOKOGAWA) when replacing parts or consumables.
• Modication of the product is strictly prohibited.
• The following safety symbols are used on the product as well as in this manual.
WARNING
This symbol indicates that an operator must follow the instructions laid out in this manual in
order to avoid the risks for the human body and health including risk of injury, electric shock, or
fatalities. or the damages to instruments. The manual describes what special care the operator
must take to avoid such risks.
<INTRODUCTION>
ii
CAUTION
This symbol indicates that the operator must refer to the instructions in this manual in order to
prevent the operator or instrument from being injured, or damaged.
The following are signal words to be found only in our instruction manuals.
CAUTION
This symbol gives essential information to avoid damages of soft or hardware, or system failure.
NOTE
This symbol indicates information that complements the present topic.
n Warning and Disclaimer
• The product is provided on an “as is” basis. YOKOGAWA shall have neither liability nor
responsibility to any person or entity with respect to any direct or indirect loss or damage
arising from using the product or any defect of the product that YOKOGAWA can not predict
in advance.
n Trademark policy
All company or product and product names mentioned in this document are trade names,
trademarks or registered trademarks of their representative companies. In this document,
trademarks or trade names are not indicated with TM or ® as trademark symbols.
IM 11R02A01-02E6th Edition :May 10, 2019-00
u After-sales Warranty
n Do not modify the product.
n Yokogawa warrants the product for the period stated in the pre-purchase quotation Yokogawa
shall conduct dened warranty service based on its standard. When the customer site is located
outside of the service area, a fee for dispatching the maintenance engineer will be charged to the
customer.
n During the warranty period, for repair under warranty, carry or send the product to the local
sales representative or service oce. Yokogawa will replace or repair any damaged parts and
return the product to you.
• Before returning a product for repair under warranty, provide us with the model name
and serial number and a description of the problem. Any diagrams or data explaining the
problem would also be appreciated.
• If we replace the product with a new one, we won’t provide you with a repair report.
n In the following cases, customer will be charged repair fee regardless of warranty period.
• Failure of components which are out of scope of warranty stated in instruction manual.
• Failure caused by usage of software, hardware or auxiliary equipment, which Yokogawa
Electric did not supply.
• Failure due to improper or insucient maintenance by user.
• Failure due to modication, misuse or outside-of-specications operation which Yokogawa
does not authorize.
• Failure or damage caused by relocation of the instrument.
• Failure due to power supply (voltage, frequency) being outside specications or abnormal.
• Failure caused by any usage at location or its maintenance that Yokogawa does not
authorize.
• Any damage from re, earthquake, storms and oods, lightning, disturbances, riots, warfare,
radiation and other natural changes.
n Yokogawa does not warrant conformance with the specic application at the user site.
Yokogawa will not bear direct / indirect responsibility for damage due to a specic application.
<INTRODUCTION>
iii
n Yokogawa Electric will not bear direct / indirect responsibility for any damage an end user might
incur, when the user congures the product into systems or resells the product.
n Maintenance service and supplying repair parts will be covered for ve years after the
production ends. For repair for the product, please contact the nearest sales oce described in
this instruction manual.
IM 11R02A01-02E6th Edition :May 10, 2019-00
Blank Page
Model CM6G
Gas Calorimeter
IM 11R02A01-02E 6th Edition
CONTENTS
u Introduction ....................................................................................................i
u Safety Precautions .......................................................................................ii
8.4 Signal Section ................................................................................................... 8-3
8.5 Computing Station ............................................................................................ 8-5
8.6 Other Troubleshooting ..................................................................................... 8-5
Revision Information ...............................................................................................i
IM 11R02A01-02E6th Edition :May 10, 2019-00
<1. Outline>
1. Outline
Model CM6G Gas calorimeter measures and controls a caloric value of gases, Wobbe-Index
(hereinafter referred to as WI), a theoretical air requirement, and a heat input for various kind of
gas burning furnaces.
It detects a temperature rise of a sample gas, whose pressure is normally controlled, by burning it
at a burner through the medium of air.
It picks up the ow rate of the sample gas and the air as the dierential pressure signal and
outputs a signal of WI after compensating calculation of the indication dierence caused by the
ow rate variation.
It also detects density of the sample gas by a density meter, and add to WI signal density
compensation, then, outputs a caloric signal.
1.1StandardSpecications
EMC : EMC Regulatory Arrangement in Australia and New Zealand
Korea Electromagnetic Conformity Standard Class A
A급 기기 (업무용 방송통신기자재)
이 기기는 업무용(A급) 전자파적합기기로서 판매자 또는
사용자는 이 점을 주의하시기 바라며, 가정외의 지역에서
사용하는 것을 목적으로 합니다.
한국 전자파적합성 기준
1-1
n Town Gas Application
Measurement Object : Measurement and control of WI or the caloric value of fuel gas
Measuring Range : 3 to 62 MJ/Nm
Sample Conditions :
Dust ; 5 mg/Nm
Temperature ; 50°C or less
Humidity ; dew point of 0°C or less
Pressure ; (1) 10 to 20 kPa
Range : Select scale range (Span) :
General Gas ; 30 to 50% of maximum value of the span
Butane or Butene + Air ; 20 to 30% of maximum value of the span
Propane or Propylene + Air ;25 to 40% of maximum value of the span
Output : 1 to 5 V DC, 4 to 20 mA DC (simultaneously), non-isolated,
load resistance; 750 Ω or less
Alarm Contact Output :
Flame o alarm; 100 V AC, 5 A, closed when alarm occurs (resistance
Orice Temperature alarm; 100 V AC, 3 A, closed when alarm occurs
Contact Input : Remote ignition (Custom order); 24 V DC, 0.1A or more
for town gas.
3
3
or less
(2) 10 kPa or under: with pump
(3) 100 to 600 kPa: with pressure reducing valve
load) when contact is opened, the leakage current is 2
mA or less (100 V AC)
(resistance load)
IM 11R02A01-02E6th Edition :May 10, 2019-00
<1. Outline>
Repeatability
MeasurementMeasuring range (Note 1)Repeatability
WI
Caloric value
3
MJ/Nm
Note 1: High caloric value means 6.3 MJ/Nm3 or more. Low caloric value means below 6.3 MJ/Nm3.
High caloric value
Low caloric value
High caloric value
Low caloric value
± 0.5% of measured value
± 1.0% of measured value
± 1.0% of measured value
± 1.5% of measured value
Sample Gas Flow Rate : Approx. 10 l/min
Response Time (Note 2):
Max.WImeasuredDead timeResponse time (63.2%)
50 or more
32 or more, less than 50
13 or more, less than 32
Less than 13
Note 2: Response time varies depending on the WI of a sample gas. This is due to the dierent sample gas ow rate of the
calorimeter. The ow rate is preset depending on the WI of the sample gas to prevent the caloric value at the detector
burner from exceeding the upper limit.
Approx. 30 sec or less
Approx. 27 sec or less
Approx. 23 sec or less
Approx. 21 sec or less
Approx. 60 sec or less
Approx. 53 sec or less
Approx. 47 sec or less
Approx. 41 sec or less
Utility :
Instrument Air ; Approx. 50 Nl/min, pressure 300 to 700 kPa, dew point of 0°C or less
Power Supply ; 100 V AC ± 10%, single phase, 50/60 Hz (Note 3), 860 VA max.
Note 3: When you perform a low caloric value measurement, the frequency variation should be within ±0.4%. If the frequency variation
exceeds ±0.4% (especially for exporting overseas), consult with Yokogawa.
1-2
n Steel Mill Application
Measurement Object : Measurement and control of WI or the caloric value of fuel gas for a
Measuring range : 3 to 62 MJ/Nm
Sample Conditions:
Dust; 100 mg/Nm
Temperature; 50°C or less
Pressure; (1) 8 to 15 kPa
(2) 8 kPa or under: with pump
Range : Select scale range (Span):
General Gas; 30 to 50% of maximum value of the span
Butane or Butene + Air; 20 to 30% of maximum value of the span
Propane or Propylene + Air; 25 to 40% of maximum value of the span
Output : 1 to 5 V DC, 4 to 20 mA DC (simultaneously), non-isolated, load
resistance 750Ω or less
Alarm Contact Output :
Flame o alarm;
Temperature alarm; 100 V AC, 3 A, closed when alarm occurs (resistance load)
Contact Input: Remote ignition (Custom order); 24 V DC, 0.1A or more
Repeatability :
MeasurementMeasuring range (Note 1)Repeatability
WI
Caloric value
3
MJ/Nm
steel mill.
High caloric value
Low caloric value
High caloric value
Low caloric value
3
3
or less
100 V AC, 5 A, closed when alarm occurs (resistance load) when
contact is opened, the leakage current is 2 mA or less (100V AC)
0.5% of measured value
1.0% of measured value
1.0% of measured value
1.5% of measured value
Note 1: High caloric value means 6.3 MJ/Nm3 or more. Low caloric value means below 6.3 MJ/Nm3.
IM 11R02A01-02E6th Edition :May 10, 2019-00
<1. Outline>
Sample Gas Flow Rate :Approx. 10 l/min.
Response Time (Note 2):
Max.WImeasuredDead timeResponse time (63.2%)
50 or more
32 or more, less than 50
13 or more, less than 32
Less than 13
Note 2: Response time varies depending on the WI of a sample gas. This is due to the dierent sample gas ow rate of the
calorimeter. The ow rate is preset depending on the WI of the sample gas to prevent the caloric value at the detector
burner from exceeding the upper unit.
42 sec or less
39 sec or less
36 sec or less
30 sec or less
70 sec or less
60 sec or less
50 sec or less
45 sec or less
Utility :
Water ; Approx. 0.2 l/min, pressure 200 to 600 kPa
Instrument Air ; Approx. 50 Nl/min, pressure 300 to 700 kPa, dew point of 0°C or less
Power Supply ; 100 V AC ± 10%, single phase, 50/60 Hz (Note 3), 1100 VA max.
Note 3: In case of low caloric value measurement, frequency variation should be within ± 0.4%. If frequency variation exceeds ± 0.4%,
consult with Yokogawa.
Panel:
Construction : For indoor installation, rack panel
Paint Color: Munsell 3.2PB7.4/1.2 (inside and outside)
Ambient Temperature:
0 to 40°C (little temperature variation,particularly no
rapid change )
1-3
1.2 ModelandSuxCodes
n CM6G Gas Calorimeter
ModelSuxCodeOption CodeDescription
CM6G
Gas
Pressure
Measurement00
Power supply-5
RangeR- - - - - - - - - -Measuring range
Style*C - - - - - - - - - -Style C
Note: Measuring range and unit must be specied.
- - - - - - - - - - - - - - - - -
-S6- - - - - - - - - -Always - S6
1
2
3
4
5
6
7
8
10
-6
- - - - - - - - - -Gas calorimeter
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
- - - - - - - - - -
Gas pressure 10 to 20 kPa for town gas, quake-proof
Gas pressure 10 to 20 kPa for town gas
Gas pressure 10 kPa or under for town gas
Gas pressure 100 ro 600 kPa for town gas
Gas pressure 8 to 15 kPa for steel mill, without preheating
Gas pressure 8 to 15 kPa for steel mill, with preheating
Gas pressure 8 kPa or under for steel mill, without preheating
Gas pressure 8 kPa or under for steel mill, with preheating
WI measurement
Caloric value measurement (GD400G should be purchased
separately)
100 V AC 50 Hz
100 V AC 60 Hz
l Standard Accessories
Followings are the standard accessories supplied.
• Calorie Detector
NameQ’tyRemarks
Mirror
Fuse
For burner ame inspection
1
2
3.15 A (A1113EF)
IM 11R02A01-02E6th Edition :May 10, 2019-00
<1. Outline>
• Orice Assembly
NameQ’tyRemarks
O-Ring
O-Ring
Hexagon Wrench
Hexagon Wrench
1
P16 (Viton) (Y9114XB)
3
P20 (Silicon) (L9817MT)
1
Nominal size 1.5 mm
1
Nominal size 2.5 mm
n Gas Density Meter
Gas Density Meter compensates a density value in caloric value measurement.
It is not required for WI measurement.
Converter: GD400G-N-10-N-/PA
Detector: GD300S-J-/KU
Measuring range and unit (specic gravity or density) should be specied. See GS 11T3E1-01E
* The code is basic for gas pressure and measurement.
Note: A wet sample gas in the town gas application is outside the scope of the standard specications. Consult
with Yokogawa.
Without
density
meter
With
density
meter
Without
density
meter
With
density
meter
Gas pressure 10 to 20 kPa: Standard
Gas pressure 10 kPa or under: With pump
Gas pressure 100 to 600 kPa: With pressure
reducing value
Gas pressure 10 to 20 kPa: Quake-proof
Gas pressure 10 to 20 kPa: Standard
Gas pressure 10 kPa or under: With pump
Gas pressure 100 to 600 kPa: With pressure
reducing value
Gas pressure 8 to 15 kPa: Without preheating
Gas pressure 8 to 15 kPa: With preheating
Gas pressure 8 kPa or under: Without preheating
Gas pressure 8 kPa or under: With preheating
Gas pressure 8 to 15 kPa: Without preheating
Gas pressure 8 to 15 kPa: With preheating
Gas pressure 8 kPa or under: Without preheating
Gas pressure 8 kPa or under: With preheating
-S6200
-S6300
-S6400
-S6110
-S6210
-S6310
-S6410
-S6500
-S6600
-S6700
-S6800
-S6510
-S6610
-S6710
-S6810
IM 11R02A01-02E6th Edition :May 10, 2019-00
<1. Outline>
n Instructions for System Selection
● The quake-proof type gas calorimeter is always equipped with the density meter.
● The CM6G Gas Calorimeter controls the ow rate under a constant dierential pressure. In
the caloric value measurement, if the density of a sample gas changes, a ow rate error
proportional to the reciprocal of the square root of the density of the sample gas,
will be generated, which directly aects the caloric value. Therefore, density compensation
is required using a density meter.
For the WI measurement, a density meter is not required since the WI is a value
proportional to
ρ
.
g
1/
1-5
ρ
g
1/
,
IM 11R02A01-02E6th Edition :May 10, 2019-00
<1. Outline>
1.4ExternalDimensions
n For Town Gas Applications
CM6G-S6200, S6210, S6300, S6310, S6400, S6410
1-6
Unit: mm
Mark Name Connection
A SAMPLE GAS IN Rc1/4
SAMPLE GAS OUT
B
C INST.AIR IN Rc1/4
D STD.GAS IN Rc1/4
E STD.GAS IN Rc1/4
B
D
E
A
C
250
View X
130
900
700
550
400
300
0
OR VENT Rc1/2
30
740
800
Density meter
converter
Computing
station
Approx.70
30
*
1800100
Approx.1970
100
4-Ø14 holes
X
* CM6G-S610 (with density meter) only
Density meter
detector
700
900
*
Calorie detector
Flowmeter for
density meter
100
*
Maintenance Space
100
800800500
900
Back
Front
500
Weight: Approx. 300kg
Dim-6210.ai
IM 11R02A01-02E6th Edition :May 10, 2019-00
<1. Outline>
n For Town Gas Application (Quake-proof Type)
CM6G-S6110
1-7
3064030
700
Ø120 hole
Density meter
converter
1500100
Approx. 1670
X
D
400
Computing station
FAIL
---CM6G---
ALM
TREND 3
90.00
18.00
Y2Y1X3X1500.0
35.00
C
1000.0
37.67
100.00
23.02
A
M
▼▼▼
△
▽
PF
0.00
14.65
0.0
23.02
▼
C
O
SHIFT
<
<<<
YOKOGAWA◇
1000
4-Ø15 holes
100600
Calorie detector
Flowmeter for
density meter
100800100
300700
90
A
B
C2
View X
C1
90
Mark Name Connection
A SAMPLE GAS IN Rc1/4
B INST.AIR IN Rc1/4
C1 STD.GAS IN (ZERO) Rc1/4
C2 STD.GAS IN (SPAN) Rc1/4
D SAMPLE GAS OUT OR VENT Rc1/2
Wiring to switch box should be made
through the bottom.
Mark Name Connection
A SAMPLE GAS IN Rc1/2
B SAMPLE GAS OUT Rc1/2
C SAMPLE GAS OUT Rc1/2
D STD.GAS IN Rc1/4
E WATER IN Rc1/2
F DRAIN OUT Rc1/2
G INST.AIR IN Rc1/4
H SAMPLE GAS VENT Rc1/4
500
600
700
800
80
1900
1650
1500
150
750
View X
60
100
300
200
100
0
G
D
Fulflo filter
Dehumidifier
E
F
74030(30)
800
0
350
Washing bubbler
Drain pot
Pressure regulating
pot
Density meter
converter
Approx. 70
H
B
C
A
**
Pump
*
Computing
station
4-Φ14 holes
80
60
40
Approx.1160
(100)
* CM6G-S610 (with density meter and flowmeter)
** CM6G-S670, CM6G-S680 (with pump)
Density meter
detector
700
900
100
500
800
800
Maintenance Space
540
900260
Back
Front
500
Weight: Approx. 350kg
Dim-6810.ai
IM 11R02A01-02E6th Edition :May 10, 2019-00
<1. Outline>
n Dimensions of Options
l Probe
Unit: mm
Part number L
H7800HA Approx. 650
H7800HB Approx. 1150
H7800HC Approx. 1650
)( 350Insertion length L
18
Φ34
1-9
JIS 10K-80-FF Equiv. SUS304
l FuloFilter(Partno.:G7043XJ)
108
Rc1/2Rc1/2
Approx. 350
Unit: mm
l PressureReducingValve(Partno.:G7008XF)
Secondary pressure
gauge
0.3MPaG
Rc1
Primary pressure
gauge
25MPaG
H7800HA.ai
Unit: mm
Approx. 119
Approx. 38
Handle for
pressure control
Approx. 62
Approx. 120
Outlet
Handle for
stop valve
Stop valve
Rc1/4
Outlet
H
C
I
A
H
K
O
A
T
開
閉
Safety valve
Approx. 166
Daiflon packing
W22 Left-hand thread
Inlet
Nut
Approx. 81
F2-8.ai
IM 11R02A01-02E6th Edition :May 10, 2019-00
<1. Outline>
1.5 Principle of Measurements
The instrument detects the temperature dierence, using a thermocouple, between the exhaust
combustion gas made after sample gas is burnt in the burner, and the feed air at the inlet of
the burner. Then the instrument amplies the value of temperature dierence. It adds the
compensating calculation to the output signal and measures the caloric value of WI.
1-10
DIFFERENTIAL
PRESSURE
DETECTION
GAS
AIR
Figure 1.1 Measurement principle diagram
BURNING
PRIMARY
MIXING
SECONDARY
TERTIARY
AIR
DIVIDING
BLOW
COMPENSATING
CALCULATION
OUTPUT
SIGNAL
When the sample gas is completely burnt by the air, the formula of the increased temperature is
as follows:
∆θ =(1.1)
K • Fg
Cps • Fs
where K: Caloric value of the sample gas
Fg: Flow rate of the sample gas
Fs: Air-diluted combustion exhaust gas ow rate
Cps: Constant pressure heat ratio of air-diluted combustion exhaust gas
Air ow rate Fa is big enough compared with the sample gas ow rate Fg (Fa:Fg=50-200:1)
and Cps Cpa, Fs Fa+Fg=Fa ( 1 + g ), therefore, the formula (1.1) is as follows:
∆θ =(1.2)
K • Fg
Cpa • Fa (1 + g)
where Fa: Air ow rate
Cpa: Air constant pressure heat ratio
g: Fg/Fa
When using orice and take out Fa, Fg as a dierential pressure of before and after orice, Fa
and Fg are represented by the following formula:
Fa = Ka(1.3)
∆ Pa
ρ
a
Fg = Kg
∆ Pg
ρ
g
where: ∆Pa, ∆Pg: Air, gas dierential pressure between before and after orice
ρ
a, ρg: Density of the air and the gas
ka, kg: Orice constant gure of the air and the gas
(Orice coecient orice sectional area)
IM 11R02A01-02E6th Edition :May 10, 2019-00
<1. Outline>
If insert formula (1.3) into formula (1.2), ∆θ is represented by the following formula:
1-11
∆θ = C.K.(1.4)
1
ρ
•
g
(C =
∆ Pg
∆ Pa
1
Cpa (1+g)
Kg
••)
Ka
ρ
a
According to formula (1.4), if ∆Pa, ∆Pg are constant, temperature dierence ∆θ is in proportion
ρ
g
to WI (
K/
) or caloric value (K). Thus we can measure continuously WI after ∆θ
measurement, and the caloric value after the measurement and calculation of the density.
IM 11R02A01-02E6th Edition :May 10, 2019-00
Blank Page
<2. Installation, Piping, Wiring>
2. Installation, Piping, Wiring
2.1 Installations
Observe the following conditions when you install instruments.
(1) Adequate space for maintenance should be provided around the gas calorimeter.
* See “External Dimention” on 1.4.
(2) The base should be horizontal..
(3) Ambient temperature is 0 to 40°C and no rapid change in ambient temperature is allowed.
*Rapid change here means a change of approximately 10°C within 30 minutes.
(4) Install the instrument in the place where it is not directly exposed to the current of a
conditioned air.
(5) Minimal vibration( If much vibration is unavoidable, take an appropriate measure to absorb
shock, e.g, use of vibration-proof robber).
(6) A ventilation system should be provided.
(7) Corrosive gases and dust are present in small quantities and humidity is low.
(8) The water of the sampling system and the drain line do not freeze up.
2-1
WARNING
From the moment the ignition starts until it completes, or right after the ame extinction, sample
gas mixed with the air will be released from the top of the equipment into the installation space of
the equipment. Use the equipment in a well ventilated environment equipped with a ventilation
system.
2.2 Externalpiping
Refer to the drawing Figure 1.4. on the connections on the panel
The principal points to be taken care are as follows:
• Bent the outlet of the bent piping as U shape, and prevent it from the rain penetration. Set
the outlet at higher position as much as possible where there is little fear of physical danger.
• Drain piping shall be conducted below the level of drain outlet on the panel and conrm no
drain is accumulated on the panel bed.
• It is necessary for the sample line of the instrument for steel mill use to provide a slope of
more than 1/3, so that there is no blockade to the gas line by the drain at the bending part of
the piping. Make piping as short as possible. Equip the sample line with thermal insulation
so as to prevent the drain in the pipe line from freezing.
• Locate the standard gas cylinders at the place where they are comparatively cool and not
exposed to direct sunshine.
IM 11R02A01-02E 5th Edition : Dec. 28, 2016-00
Loading...
+ 45 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.