YOKOGAWA CL360 User Manual

User’s
r
Manual
CL360 Leakage Clamp-on Tester リーククランプテスタ
IM CL360
Contents
Precautions for Safety Use of the Instrument ---------------------------i
1. Instrument Layout ------------------------------------------------------------1
2. Measurement ------------------------------------------------------------------ 3
2.1 Preparation for Measurement ------------------------------------------------ 3
2.2 AC Current Measurement----------------------------------------------------- 3
2.3 How to Use The Frequency Selector Switch----------------------------- 4
2.4 How to Use Peak Hold Function---------------------------------------------5
2.5 How to Use Peak Hold Frequency Selector Switch -------------------- 7
2.6 How to Use Data Hold Function ------------------------------------------- 10
2.7 How to Use OUTPUT Terminal -------------------------------------------- 10
3. Battery Replacement ------------------------------------------------------ 11
4. Specifications ---------------------------------------------------------------- 12
5. Calibration and After-sales Service ------------------------------------ 15
IM CL360
2003.06 2 版(MC)

Precautions for Safe Use of the Instrument

When handling the instrument, ALWAYS observe all of the cautionary notes on safety given below. Yokogawa M&C Corporation is not at all liable for damage resulting from misuse of this product by the user that is contrary to these cautionary notes. Various symbols are used on the instrument and in this manual to ensure the product is used safety and to protect operators and property from possible hazards or damage. The following safety symbols are used where appropri­ate. Read the explanations carefully and familiarize yourself with the symbols before reading the text.
The instrument and this manual use the following safety symbols:
Danger! Handle with Care.
This symbol indicates that the operator must refer to an explanation in the User’s Manual in order to avoid the risk of personal injury o death and/or damage to the instrument.
Double Insulation
This symbol indicates double insulation.
AC Voltage/Current
This symbol indicates AC voltage or current.
Ground
Indicates that there is a possibility of serious personal injury or loss of life if the operating procedure is not followed correctly and describes the precau­tions for avoiding such injury or loss of life.
Indicates that there is a possibility of serious personal injury of damage to the instrument if the operating procedure is not followed correctly and describes the precautions for avoiding such injury or damage.
NOTE Draws attention to information essential for understanding the op­eration and features.
This symbol indicates ground (earth)
WARNING
CAUTION
WARNING
Never make measurement on a circuit above 600V AC.
Do not use the instrument in an atmosphere where any flammable or ex-
plosive gas is present.
Do not attempt to make measurement in the presence of flammable gas­ses, fumes, vapor or dust. Otherwise, the use of the instrument may cause sparking, which can lead to an explosion.
Avoid using the instrument if it has been exposed to rain or moisture or if your hands are wet.
Do not exceed the maximum allowabIe input of any measurement range.
Never open the battery compartment cover when making measurement.
Do not use the instrument if there is any damage to the casing or when the
casing is removed.
Do not install substitute parts or make any modification to the instrument. Return the instrument to Yokogawa M&C or your distributor for repair or re-calibration.
Always switch off the instrument before opening the battery compartment cover for battery replacement.
WARNING
To avoid damage to the instrument or electric shock! The restrictions on the maximum voltage level for which the CL360 testers can be used, depend on the over-voltage categories specified by the safety standards. These category specifications are formulated to protect operators against transient impulse voltage in power lines.
Function
A
Over-voltage category I (CAT.I):
Signal level, special equipment or parts of equipment, telecommunication, electronic
etc., with smaller transient over-voltages than CAT.II.
Over-voltage category II (CAT.II)
Local level, appliance, portable equipment etc., with smaller transient over-voltages
than CAT.III.
Over-voltage category III (CAT.III):
Distribution level, fixed installation, with smaller transient over-voltages than CAT.IV.
Measuring circuit voltage : AC 600V rms
Maximum Allowable Input
OVERVOLTAGE
CATEGORY II
AC 1000A rms
OVERVOLTAGE
CATEGORY III
AC 1000A rms Measuring circuit voltage : AC 300V rms
IM CL360 i
ii
IM CL360
CAUTION
Be sure to set the Range switch to the "OFF" position after use. When the instrument will not be in use for a long period of time, place it in storage after removing the battery.
Do not expose the instrument to the direct sun, extreme temperatures or dew fall.
Never apply voltage to the OUTPUT terminal. The terminal is not electri­cally isolated from the internal circuits of the instrument.
Use a damp cloth and detergent for cleaning the instrument. Do not use abrasives or solvents.
NOTE
Radiation immunity affects the accuracy of CL360 testers under the condi­tions specified in EN 61000-4-3:1997.
If equipment generating strong electromagnetic interference is located nearly, the testers may malfunction.
IM CL360 iii

1. Instrument Layout

(1)
(2)
(3) Two-way Analogue OUTPUT Terminal : AC current picked up by trans-
former jaws (7) is converted and output as AC and DC voltage output.
(7)
(6)
OFF
(5)
(See Analogue Output in chapter 4. Specifications) Insert output cord Model 91020 into this terminal for monitoring waveform with an oscillo­scope, making RMS measurements or connecting to a recorder.
(4) Safety Hand Strap : Prevents the instrument from slipping off the hand
during use. (5) Range Switch : Selects range. It is also used to turn power on or off. (6) Open/Close Lever : Operates the transformer jaws. Press to open the
transformer Jaws. (7) Transformer Jaws : Pick up current flowing through the conductor. (8) Peak Hold Selector Switch : Selects 10ms or 100ms response time. Set
the switch back to the OFF position to release peak hold or make normal
measurements. (9) Frequency Selector Switch : Makes frequency response selection. Selects
"WIDE" or "50/60Hz".
(9)
(3)
(1) LCD Display : Field effect type of liquid crystal display with maximum
counts of 1999. Range symbol (mA, A) and decimal point automatically appear as the Range switch is turned. "B" is displayed on the lower left corner for low battery warning and "1" is displayed only at the higher digit for overrange indication (except for 1000A range).
(2) Data Hold Button : Allows for easy reading in dimly lit or hard-to-reach
locations. The display can be observed away from the conductor after pushing in the button.
IM CL360 1
(8)
(4)

2. Measurement

2.1 Preparation for Measurement

CAUTION
The jaw section is a delicate, precision sensor. Do not subject the jaw to unreasonably strong shock, vibration, or force when using it.
If dust gets into the tops of the jaws, remove it immediately. Do not close
the jaws when dust is trapped in its joints as the sensor may break.
Please check that the range is set to the desired setting before measure-
ment.

2.2 AC Current Measurement

WARNING
Do not make measurement on a circuit above 600V AC. This may cause shock hazard or damage to the instrument or equipment under test.
(1) Set the range switch to the desired position. Do not exceed the maximum
allowable input current for the selected range.
(2) For normal measurement, press the open/close lever to open the trans-
former jaws and clamp onto one conductor only. Earth leakage current or small current that flow through a grounded wire can also be measured by this method. It is recommended that the conductor is placed at the center of the closed transformer jaws.
Correct Wrong
IM CL360 3
2 IM CL360
(3) To measure out of balance leakage current, clamp onto all conductors
except a grounded wire. The leakage current measured will be indicated on the display.
3-phase system
LoadLoad
In case of 1-phase 3-wire system, clam four wires for measurement.
In case of 3-phase 4-wire system, clamp three wires for measurement.
CAUTION
When measuring large current, observe the time limit specified in chapter 4, Specifications. Otherwise, the transformers jaws may overheat, resulting in damage to the instrument.
NOTE
During current measurement, keep the transformer jaws fully closed. Oth­erwise, accurate measurement cannot be made. The maximum measur­able conductor size is approx. 68mm in diameter.
Frequency selector switch is designed to select the "50/60Hz " and "WIDE" frequency ranges. For further details, refer to section 2.3 for operation of frequency selector switch.
The transformer jaws may buzz when measuring large current. This has no effect on the instrument’s performance or safety.

2.3 How to Use The Frequency Selector Switch

Model CL360 has a very good frequency response because of the electro­magnetic property of its transformer jaws. Therefore, it measures AC current not only of fundamental frequency of 50Hz or 60Hz, but of high frequencies and harmonics superimposed on the fundamental frequency. To eliminate these superimposed components and measure only in the fundamental fre­quency, Model CL360 has a high-cut filter circuit, which can be activated by setting the frequency selector switch to the 50/60Hz position.
The high-cut filter has a cut-off frequency of approx. 100Hz and an attenua­tion characteristics of approx. -24dB/octave.
NOTE
-24dB/octave means that the magnitude of a signal declines by a factor of 16
when its initial frequency doubles.
4 IM CL360
The frequency selector switch has the following two positions.
WIDE (40Hz to over 1kHz) : Covers a wide frequency band from mains supply to high frequencies generated by such equipment as inverters.
50/60Hz (40 to Approx. 100Hz) : Filters out high frequency components to restrict measurement in mains frequency band.
NOTE
Selection with the frequency selector switch does not apply to AC output of the two-way analogue output.
2.4 How to Use Peak Hold Function
10ms or 100ms response time can be selected for peak hold measurement. Make selection according to your application needs.
(1) With the transformer jaws clamped onto the conductor under test, slide
the peak hold switch from the OFF position to the desired peak response time position.
2
(2) The peak hold display reads 1/
an RMS reading will be obtained when the current under test has a sinu­soidal waveform.
Input Current
Peak Hold Current Output (display value & DC output)
(3) Slide the peak hold switch back to the OFF position for a reset.
of the peak current value. Therefore,
IP
IP
NOTE
Model CL360 uses an analogue peak hold circuit to ensure a quick re­sponse to input current. Because of the nature of this circuit, the peak hold reading may gradually fall, or in a rare case, rise with time. This is likely to be apparent when the instrument is used in a high temperature and high humidity environment. Therefore, the instrument will not be suitable for making peak measurement over an extended period of time. In case of such a need, connect a recorder to the instrument via the analogue output terminal.
If it is necessary to read the display away from the conductor in a peak hold measurement, press the data hold switch first and then remove the instrument from the conductor. Otherwise, the peak hold reading may be higher than the actual value due to the electrical noise caused by the opening and closing of the transformer jaws. Press the data hold switch again for a reset.
(4) Difference between 10ms and 100ms Peak Response Time. The peak
hold circuit in this instrument charges the peak-hold capacitor after recti­fying the input waveform. The time for the voltage of the capacitor to reach its peak value varies according to its capacitance and the output impedance of the charging circuit. Model CL360 sets the time for the voltage of the capacitor to reach 90% of its peak value to 10ms or 100ms by switching between two output impedances.
10ms Peak Response Time 100ms Peak Response Time
Input
Waveform
100
90
10ms
For instance, select the 10ms response time when measuring a surge current that will occur when a power supply device is switched on. The 100ms re­sponse time is recommended for measuring the starting current of a motor or similar equipment. A stable measurement can be made on the 100ms second response time setting as the peak hold circuit does not readily respond to the surge current.
Hold Capacitor Voltage
t
100
90
100ms
t
IM CL360 5

2.5 How to Use Peak Hold Frequency Selector Switch

Indicated value may vary according on input signal (waveform) when the frequency selector switch is set to 50/60Hz. In this case, make measurement referring to the block diagram below.
Circuit flowchart is shown below.
Frequency
WIDE
selection
I - V
CT
Conversion
(50/60Hz)
SW
High cut
Filter
(1) (2) (3)
Selection switch
Rectifying
Peak hold
circuit
(100ms or 10ms
(OFF)
Peak hold
circuit
divider
Voltage
circuit
1
)
2
Display
The signal (current) picked up by transformer jaws is converted from current signal to voltage signal in I-V conversion circuit, and it enters the frequency selector switch. If the frequency selector switch is set to WIED, it enters the rectifying circuit. If set to 50/60Hz, enters the rectifying circuit after passing through high-cut filter (cut-off frequency of approx. 100Hz and attenuation characteristics of approx. -24dB/octave). If the peak hold selector switch is set to OFF, the rectified signal enters the indication circuit. If set to 100ms or 10ms, passes through the peak hold circuit, picks up the peak value of input
2
signal, and outputs 1/
of the peak value into the indication circuit. Different examples of peak hold statuses when the frequency selector switch is set to WIDE or 50/60Hz are described below.
NOTE
If the frequency selector switch is switched after the peak hold switch is set to 100ms or 10ms, the over indication appears. Make sure that first set the frequency selector switch, then, set the peak hold switch to 100ms or 10ms.
6 IM CL360
Example 1 When measuring the peak of starting current of motor:
(1) (2) (3)
Input waveform
(I-V circuit output)
Frequency
Selector Wave at
WIDE position
Surge pulse at switching on (switching noise)
Frequency
Selector Wave at
50/60Hz posi tion
Waveform after passing through high-cut filter
Example 2 When measuring the peak current of thyristor:
(1) (2) (3)
Input waveform
(I-V circuit output)
Frequency
Selector Wave at
WIDE position
Rectifying
circuit in put
Rectifying
circuit input
Peak hold circuit
(Hold value)
Peak hold circuit
(Hold value)
IM CL360 7
Frequency
Selector Wave at
50/60Hz position
Waveform after passing through high-cut filter
8 IM CL360
Example 3 When measuring the peak current of inverter (where harmonic are
superimposed on the fundamental frequency):
(1) (2) (3)
Frequency
Selector Wave at
WIDE position
Frequency
Selector Wave at
50/60Hz posi tion
Input waveform
(I-V circuit output)
Waveform after passing through high-cut filter
Rectifying
circuit input
Peak hold circuit
(Hold value)
NOTE
The indicated value at 50/60Hz may be higher than that at WIDE because of the phase difference between superimposed harmonics and fundamental frequency.

2.6 How to Use Data Hold Function

Push in the Data Hold button to freeze the reading. This is especially useful for taking a reading a dimly lit or hard-to-reach locations. The display can be observed away from the conductor. Push the button again to release the reading.

2.7 How to Use OUTPUT terminal

AC and DC outputs can be obtained by inserting optional Model 91020 output cord into the two-way analogue output terminal.
AC Output :
Can be monitored by connecting a digital multimeter to analogue output terminal or observed as waveform by connecting an oscilloscope.
DC Output :
Can be monitored by connecting a digital multimeter, or a recorder, which enables many hours of monitoring. With the CL360 in the peak hold mode, DC voltage corresponding to 1/ and outputted.
OFF
Model CL360
(connect to oscilloscope or multimeter)
AC output
2
of a peak current value can be held
Model 91020 (optional)
Note : Put a cap on the plug when not in use.
Connect to multimeter or recorder
DC outputAC output
DC output
Black(COM)
Red(DC)
CAP
Blue(AC)
IM CL360 9

3. Battery Replacement

WARNING
To avoid electric shock hazard, make sure to set the Range switch to "OFF" and remove the test leads from the instrument before trying to replace bat­tery.
CAUTIOIN
Make sure to install battery in correct polarity as indicated in battery com­partment.
Replace the battery when "B" symbol appears on the LCD display.
(1) Set the Range switch to the "OFF" position.
(2) Unscrew and remove the battery compartment on the bottom of the in-
strument.
(3) Replace the battery observing correct polarity. Use a new 6F22 or 6LR61
battery.
(4) Re-place and screw the battery compartment cover.
Multimeter
RecorderOscilloscope
10 IM CL360

4. Specifications

Instrument Specifications
Measuring Ranges and Accuracy (at 23 ±10°C, 85% relative humidity or less)
AC Current
Ranges
200mA 0 to 199.9mA
2A 0 to 1.999A
20A 0 to 19.99A
200A 0 to 199.9A
1000A
A
Measuring
range
0 to 500A
501 to 1000A
Accuracy
Frequency
WIDE 50/60Hz
±1.0% rdg ±2dgt (50/60Hz) ±3.0% rdg ±2dgt (40 to 1kHz)
±1.5% rdg ±2dgt (50/60Hz) ±3.5% rdg± 2dgt (40 to 1kHz) ±1.5% rdg ±2dgt (50/60Hz) ±3.5% rdg ±2dgt (40 to 1kHz) ±5% rdg (50/60Hz) ±10% rdg (40 to 1kHz)
±1.5% rdg ±2dgt
±2.0% rdg ±2dgt
±2.0% rdg ±2dgt
±5.5% rdg 10min.
Time limit for Meas­urements
Continuous
Remove
this screw
6F22 Battery
IM CL360 11
1000A range
Attenuation (dB)
Typical frequency characteristics for 2A and higher ranges with the frequency selector switch at WIDE position.
Typical characteristics for 200mA range with the frequency sel ector switch at WIDE position.
All ranges
With the frequency selector switch set at 50/60Hz position to cut off higher frequencies -24dB/oc t
Frequency (Hz)
Errors as aginst reading at 50Hz (%)
12 IM CL360
Analogue Output (Output impedance: Approx. 1kΩ) AC Output
Range
200mA 0 to 200mA
2A 0 to 2A
20A 0 to 20A
200A 0 to 200A
1000A
Voltage proportional to the current under test is output with “WIDE” frequency characteristics regardless of the setting of the frequency selector or peak hold switch.
Measuring range AC Output Voltage
0 to 200mV
0 to 500A 0 to 50mV ±3.0% rdg
501 to 1000A 50 to 100mV ±5.0% rdg
±2.0% rdg
±2.5% rdg
Accuracy
DC Output
Range
Measuring range
200mA 0 to 200mA
2A 0 to 2A
20A 0 to 20A
200A 0 to 200A
0 to 500A 0 to 50mV ±5.0% rdg ±5.5% rdg
1000A
501 to 1000A 50 to 100mV ±7.0% rdg ±7.5% rdg
DC Output
Voltage
0 to 200mV
Accuracy (Frequency Range)
Freq. Selector
Switch at WIDE
position
±3.0% rdg ±3.5% rdg
±3.5% rdg ±4.0% rdg
Freq. Selector
Switch at 50/60Hz
position
General Specifications
Operating System : Dual integration
Measurement Function : AC current
Display : Field effect 3 - 1/2 digit liquid crystal display with maximum count
of 1999
Range selection : Manual
Overrange Indication : "1" is displayed on the highest digit except for
1000A AC range.
Response Time : Approx. 1 seconds.
Sample Rate : Approx. 3 times per second.
Temperature and Humidity for Guaranteed Accuracy : 23°C ±10°C, relative
humidity up to 85% without condensation
Operating Temperature and Humidity : 0 to 40°C, relative humidity up to 85% without condensation
Storage Temperature and Humidity : -10 to 50°C, relative humidity up to 80% without condensation
Effect of conductor position: Within 2% difference between maximum and minimum values to a 10 mm-dia conductor, at every part inside the jaws
Effect of External Magnetic Field : 15mA AC max. in proximity to a 10mm-dia conductor carrying 100A AC.
Effect of Residual Current : 10mA AC max. when clamping on two 10mm-dia conductors, each carrying supply or return 100A AC current.
Power Source : 6LR61 or 6F22 DC9V battery
Battery Life : Approx. 60 hours (continuity)
Current Consumption : Approx. 5mA max.
Withstanding Voltage : 3700V AC, 50/60Hz for 1 minute between electrical
circuit and housing case or metal part of the jaws
Insulation Resistance : 10MΩ or greater at 1000V between electrical circuit and housing case or metal part of the jaws
Conductor Size : Approx. 68mm diameter max.
Dimensions : Approx.130(W) x 250(H) x 50(D) mm
Weight : Approx. 570g
Safety Standard: EN 61010-1
EN 61010-2-032
AC 300V CAT III, AC 600V CAT II, Pollution degree 2, indoor use
EMC Standard: EN 61326 EN 55022
Accessories : 6F22 battery······························ 1
Carrying case Model 93031 ······1
User’s Manual ··························· 1
Optional Accessories : Output cable with banana plug Model 91020
Peak Hold : Response time selectable from approx. 10ms and 100ms.
2
of the peak current is displayed, which means an RMS reading will
1/ be obtained when current having sinusoidal waveform in measured. Peak hold reading varies by approximately ±1% max of full scale in the first minute at 23°C ±10°C and relative humidity of 75% without condensa­tion.
IM CL360 13

5. Calibration and After-sales Service

Should any failure occur while you are using the tester, follow the instructions given below. If the tester still fails to operate correctly and needs repair, con­tact the vendor from whom you purchased the instrument or the nearest Yokogawa M&C sales office.
Condition Possible Cause Remedy Display blanks after power-on Display reading remains frozen.
Transformer jaws buzz when measuring large current. The lowest digit of reading is unstable
Output cable (Model 91020) does not output voltage.
Calibration It is recommended that the instrument be calibrated once every year.
Battery is improperly installed. Battery is exhausted. Data hold button is pressed in. Peak hold switch is at ON position.
-----------------
-----------------
The cable is open circuit. Check the cable for an open
Install the battery correctly. Replace the buttery. Release data hold button. Set peak hold switch to OFF position. This is not a failure
This is not a failure. The instrument is highly accu­rate, so it senses slight variations in the current under test.
circuit.
14 IM CL360
IM CL360 15
YOKOGAWA M&C CORPORATION International Sales Dept.
2-9-32 Nakacho, Musashino-shi, Tokyo, 180-8750 Japan Phone: 81-422-52-5716 Facsimile: 81-422-55-8654
YOKOGAWA CORPORATION OF AMERICA (U.S.A.)
Phone: 1-770-253-7000 Facsimile: 1-770-251-2088
YOKOGAWA EUROPE B. V. (THE NETHERLANDS)
Phone: 31-334-64-1611 Facsimile: 31-334-64-1610
YOKOGAWA AMERICA DO SUL S. A. (BRAZIL)
Phone: 55-11-5681-2400 Facsimile: 55-11-5681-1274
YOKOGAWA ENGINEERING INSTRUMENTS KOREA CORPORATION (KOREA)
Phone: 82-2-551-0660 to -0664 Facsimile: 82-2-551-0665
YOKOGAWA AUSTRALIA PTY. LTD. (AUSTRALIA)
Phone: 61-2-9805-0699 Facsimile: 61-2-9888-1844
YOKOGAWA BLUE STAR LTD. (INDIA)
Phone: 91-80-227-1513 Facsimile: 91-80-227-4270
LTD. YOKOGAWA ELECTRIC (RUSSIAN FEDERATION)
Phone: 7-095-737-7868 Facsimile: 7-095-737-7869
計測機器営業部
180-8750 東京都武蔵野市中町 2-9-32 電話:0422-52-5984 ファクシミリ:0422-55-8953
関西支店
564-0063 大阪府吹田市江坂町 1-23-101 大同生命江坂ビル 10 階 電話:06-6368-7041 ファクシミリ:06-6368-7045
中部支店
450-0003 名古屋市中村区名駅南 1-27-2 日本生命笹島ビル 12 階 電話:052-581-7490 ファクシミリ:052-581-7664
広島営業所
730-0037 広島市中区中町 8-12 広島グリーンビル8階 電話:082-240-7676 ファクシミリ:082-541-4567
九州営業所
812-0037 福岡市博多区御供所町 3-21 大博通りビジネスセンター7 階 電話:092-262-5740 ファクシミリ:092-262-5741
本社
180-5679 東京都武蔵野市中町 2-9-32 電話:0422-52-5679 ファクシミリ:0422-51-8455
KIM1JE-2003.2
Loading...