User’s
Manual
CL220
Clamp-on Tester
IM CL220
Contents
Precautions for Safety Use of the Instrument
1. Instrument Layout
2. Measurement
2.1 Preparation for Measurement
2.2 DC Current Measurement
2.3 AC Current Measurement
3. Other Functions
3.1 Sleep Function
3.2 Data Hold Function
3.3 Optional Accessories
4. Battery Replacement
5. Specifications
6. Calibration and After-sales Service
2nd Edition: June 2003 (MC)
IM CL220
Precautions for Safe Use of the Instrument
When handling the instrument, ALWAYS observe all of the cautionary
notes on safety given below. Yokogawa M&C Corporation is not at all
liable for damage resulting from misuse of this product by the user
that is contrary to these cautionary notes.
Various symbols are used on the instrument and in this manual to
ensure the product is used safety and to protect operators and property from possible hazards or damage. The following safety symbols
are used where appropriate. Read the explanations carefully and
familiarize yourself with the symbols before reading the text.
The instrument and this manual use the following safety symbols:
Danger! Handle with Care.
This symbol indicates that the operator must refer to an explanation in the User’s Manual in order to avoid the risk o
personal injury or death and/or damage to the instrument.
Double Insulation
This symbol indicates double insulation.
AC Voltage/Current
This symbol indicates AC voltage or current.
DC Voltage/Current
This symbol indicates DC voltage or current.
AC/DC Voltage/Current
This symbol indicates AC/DC voltage or current.
Ground
Indicates that there is a possibility of serious personal injury or loss of
life if the operating procedure is not followed correctly and describes
the precautions for avoiding such injury or loss of life.
Indicates that there is a possibility of serious personal injury of damage to the instrument if the operating procedure is not followed correctly and describes the precautions for avoiding such injury or damage.
NOTE
Draws attention to information essential for understanding the operation and features.
This symbol indicates ground (earth)
WARNING
CAUTION
WARNING
● Never make measurement on a circuit above 300V AC/DC.
● Do not use the instrument in an atmosphere where any flammable
or explosive gas is present.
● Do not attempt to make measurement in the presence of flammable gasses, fumes, vapor or dust. Otherwise, the use of the instrument may cause sparking, which can lead to an explosion.
● Avoid using the instrument if it has been exposed to rain or moisture or if your hands are wet.
● Do not exceed the maximum allowabIe input of any measurement
range.
● Never open the battery compartment cover when making measurement.
● Do not use the instrument if there is any damage to the casing or
when the casing is removed.
● Do not install substitute parts or make any modification to the instrument. Return the instrument to Yokogawa M&C or your distributor for repair or re-calibration.
● Always switch off the instrument before opening the battery compartment cover for battery replacement.
WARNING
To avoid damage to the instrument or electric shock!
The restrictions on the maximum voltage level for which the CL220
testers can be used, depend on the over-voltage categories specified
by the safety standards. These category specifications are formulated
to protect operators against transient impulse voltage in power lines.
Function
A, A
Over-voltage category I (CAT.I):
Signal level, special equipment or parts of equipment, telecommu-
nication, electronic etc., with smaller transient over-voltages than
CAT.II.
Over-voltage category II (CAT.II)
Local level, appliance, portable equipment etc., with smaller tran-
sient over-voltages than CAT.III.
Over-voltage category III (CAT.III):
Distribution level, fixed installation, with smaller transient
over-voltages than CAT.IV.
Maximum Allowable Input
OVERVOLTAGE CATEGORY III
Measuring circuit voltage : 300Vrms AC/300V DC
300Arms AC/300A DC
CAUTION
● Be sure to set the Function Selector switch to the "OFF" position
after use. When the instrument will not be in use for a long period of
time, Place it in storage after removing the battery.
● Use a damp cloth and detergent for cleaning the instrument. Do not
use abrasives or solvents.
NOTE
● Radiation immunity affects the accuracy of CL220 testers under the
conditions specified in EN 61000-4-3: 1997.
● If equipment generating strong electromagnetic interference is
located nearly, the testers may malfunction.
1. Instrument Layout
Open/Close Lever
Transformer Jaws
LCD Display
DC Current
AC Current
Data Hold Button
Low Battery Warning
Negative Sign
Function Selector Switch
Zero ADJ. Button
Safety Hand Strap
Display
Data Hold Mode
Current
2. Measurement
2.1 Preparation for Measurement
CAUTION
● The jaw section is a delicate, precision sensor. Do not subject the
jaw to unreasonably strong shock, vibration, or force when using it.
● If dust gets into the tops of the jaws, remove it immediately. Do not
close the jaws when dust is trapped in its joints as the sensor may
break.
● Please check that the Function Selector switch is set to the desired
position before measurement.
2.2 DC Current Measurement
WARNING
Do not make measurement on a circuit above 300V DC. This may
cause shock hazard or damage to the instrument or equipment under
test.
(1) Set the Function Selector switch to the “
should be shown on the upper left corner of the display.
(2) With the transformer jaws closed and without clamping them onto
the conductor, press the
zero adjust the display.
(3) Press the open/close lever to open the transformer jaws and
clamp them onto the conductor under test, then take the reading
on the display. The most accurate reading will be obtained by
keeping the conductor at the center of the transformer jaws.
NOTE
● During current measurement, keep the transformer jaws fully
closed. Otherwise, accurate measurement cannot be made. The
maximum measurable conductor size is approx. 24mm in diameter.
● When the current flows from the upside (the display side) to the
underside of the instrument, the polarity of the reading is positive
and vice versa. (See the figure at the below light)
Correct
Wrong
button for about one second to
A” position. "DC"
"+" Current
"-" Current
2.3 AC Current Measurement
WARNING
Never use the instrument on a circuit above 300V AC. This may
cause electrical shock hazard and damage to the instrument or the
circuit under test.
(1) Set the Function Selector switch to the "
should be shown on the upper left corner of the display.
(2) Press the open/close lever to open the transformer jaws and
clamp them onto a single conductor and take the reading on the
display. The most accurate reading will be obtained by keeping
the conductor at the center of the transformer jaws.
NOTE
● During current measurement, keep the transformer jaws fully
closed. Otherwise, accurate measurements cannot be taken.
Maximum conductor size is 24mm in diameter.
● Zero adjustment is not necessary in AC current measurement.
There is no polarity in the reading either.
Correct Wrong
A" position. "AC"
3. Other Functions
3.1 Sleep Function
This is a function to prevent the instrument from being left powered on
in order to conserve battery life. This function causes the instrument
to enter the Sleep (powered-down) mode about 30 minutes after the
last switch or button operation.
To exit the Sleep mode, turn the Function Selector switch back to
"OFF", then to any other position, or press any button.
The current is consumed a little in the Sleep mode.
NOTE
If the instrument in the Data Hold mode goes into “sleep”, the Data
Hold function will remain effective when the instrument is powered on
again.
3.2 Data Hold Function
This is a function used to freeze the measured value on the display.
Press the
regardless of subsequent variation in input. "
upper right corner of the display while the instrument is in the Data
Hold mode.
To exit the Data Hold mode, press the
button to freeze the reading. The reading will be held
" is shown on the
button again.
NOTE
If the instrument in the Data Hold mode goes into “sleep”, the Data
Hold function will remain effective when the instrument is powered on
again.
3.3 Optional Accessories
Clamp Adapter Model 99025 (For AC current measurement only)
Clamp Adapter Model 99025 is designed
to increase the measuring capability of a
clamp meter. With the use of the Clamp
Adapter, you can not only extend current
range over 3000A, but also clamp on a
large bus-bar or conductor.
(1) Set the Function Selector switch to
(2) As shown in the figure below, clamp
(3) Clamp Model 99025 onto the bus-bar
(4) Take the reading on Model CL220
NOTE
For the detailed specification, refer to the
Clamp Adapter User’s Manual.
A " position.
the "
Model CL220 onto the pickup coil of
Model 99025.
or conductor under test.
and multiply it by 10.
4. Battery Replacement
WARNING
To avoid electric shock hazard, never try to replace batteries during
measurement.
CAUTION
● Do not mix new and old batteries.
● Make sure to install battery in correct polarity as indicated in
battery compartment.
If the battery voltage becomes too low for the instrument to operate
normally, "
Note that when the battery is completely exhausted, the display
blanks without "
(1) Set the Function Selector switch to the "OFF" position.
(2) Press in the hole on the battery compartment cover with the tip of
a pointed object, then slide open the cover.
(3) Replace the battery observing correct polarity. Use two new LR44
or SR44 batteries.
(4) Slide the battery compartment cover back in place.
Battery compartment cover
" is shown on the display. Then, replace the battery.
" shown.
5. Specifications
Instrument Specifications
Measuring Ranges and Accuracy (at 23±5°C, relative humidity up to 85%)
●
DC Current
Range Measuring Range Accuracy
40A 0 to ±40.00A ±1.0% rdg ±4dgt
300A
AC Current
Range
40A 0 to 40.00A
300A
(Auto-ranging)
±20.0 to ±200.0A ±1.5% rdg ±4dgt
±200.0 to ±300.0A ±3.0% rdg
(Auto-ranging)
Measuring Range
20.0 to 200.0A
200.0 to 300.0A
±1.0% rdg ±4dgt (50/60Hz)
±2.5% rdg ±4dgt (20Hz to 1kHz)
±1.5% rdg ±4dgt (50/60Hz)
±2.5% rdg ±4dgt (20Hz to 1kHz)
±3.5% rdg (50/60Hz)
±4.0% rdg (20Hz to 1kHz)
Accuracy
General Specifications
● Operating System : Dual integration
● Measurement Function : AC current, DC current
● Display : Liquid crystal display with maximum counts of 4000
● Overrange Indication : "OL" is shown on the display
● Response Time : Approx. 2 seconds.
● Sample Rate : Approx. 2.5 times per second.
● Temperature and Humidity for Guaranteed Accuracy : 23°C ±5°C,
relative humidity up to 85% without condensation
● Operating Temperature and Humidity : 0 to 40°C, relative humidity
up to 85% without condensation
● Storage Temperature and Humidity : -20 to 60°C, relative humidity
up to 85% without condensation
Effect of conductor position : Within ±2.0%rdg ±5dgt of indicated value
●
at the center to a 10 mm-dia conductor, at every part inside the jaws
● Effect of external magnetic field : 1A or less in AC or DC magnetic
field of 400 A/m
● Power Source : Tow LR-44 or SR-44 (3V DC) batteries
● Battery Life : Approx. 100 hours (continuous)
● Current Consumption : Approx. 9mA
● Sleep function : Automatically powered down in approx. 5 minutes
after the last switch operation (power consumption : approx. 20µA)
● Withstanding Voltage : 3700V AC for 1 minute between housing
case and metal part of jaws
● Insulation Resistance : 10MΩ or greater at 1000V between hous-
ing case and metal part of jaws
● Conductor Size : Approx. 24mm diameter max.
● Dimensions : Approx. 59(W) x 147(H) x 25(D) mm
● Weight : Approx. 100g (batteries included)
● Safety Standard: EN 61010-1
EN 61010-2-032
● EMC Standard: EN 61326
EN 55022
● Accessories : LR-44 battery ····························2
● Optional Accessories : Clamp adapter Model 99025
(300V AC/DC CAT. III, Pollution degree2, indoor use)
Carrying case Model 93033·······1
User’s Manual ···························1
6. Calibration and After-sales Service
Should any failure occur while you are using the tester, follow the
instructions given below. If the tester still fails to operate correctly and
needs repair, contact the vendor from whom you purchased the instrument or the nearest Yokogawa M&C sales office.
● Turn off the POWER switch once, then turn it back on again.
● If the tester does not turn on, replace the battery with a new one.
Calibration
It is recommended that the instrument be calibrated once every year.
YOKOGAWA M&C CORPORATION
International Sales Dept.
2-9-32 Nakacho, Musashino-shi, Tokyo, 180-8750 Japan
Phone: 81-422-52-5716 F acsimile: 81-422-55-8654
YOKOGAWA CORPORATION OF AMERICA (U.S.A.)
Phone: 1-770-253-7000 F acsimile: 1-770-251-2088
YOKOGAWA EUROPE B. V. (THE NETHERLANDS)
Phone: 31-334-64-1611 Facsimile: 31-334-64-1610
YOKOGAWA AMERICA DO SUL S. A. (BRAZIL)
Phone: 55-11-5681-2400 Facsimile: 55-11-5681-1274
YOKOGAWA ENGINEERING INSTRUMENTS KOREA CORPORATION (KOREA
Phone: 82-2-551-0660 to -0664 Facsimile: 82-2-551-0665
YOKOGAWA AUSTRALIA PTY. LTD. (AUSTRALIA)
Phone: 61-2-9805-0699 F acsimile: 61-2-9888-1844
YOKOGAWA BLUE STAR LTD. (INDIA)
Phone: 91-80-227-1513 F acsimile: 91-80-227-4270
LTD. YOKOGAWA ELECTRIC (RUSSIAN FEDERATION)
Phone: 7-095-737-7868 F acsimile: 7-095-737-7869
)
KIM3E-2003.2