User’s
Manual
CL130/CL135
Clamp-on Tester
IM CL130
Contents
Precautions for Safety Use of the Instrument
1. Instrument Layout
2. Measurement
2.1 Preparation for Measurement
2.2 AC Current Measurement
2.3 AC Voltage Measurements
2.4 Resistance Measurement and Continuity Check
3. Other Functions
3.1 Data Hold Function
3.2 Optional Accessories
4. Battery Replacement
5. Specifications
6. Calibration and After-sales Service
2nd Edition: June 2003 (MC)
IM CL130
Precautions for Safe Use of the Instrument
When handling the instrument, ALWAYS observe all of the cautionary notes on
safety given below. Yokogawa M&C Corporation is not at all liable for damage
resulting from misuse of this p roduct by the user that is contra ry to these
cautionary notes.
Various symbols are used on the instrument and in this manual to ensure the
product is used safety and to protect operators and property from possible
hazards or damage. The following safety symbols are used where appropriate.
Read the explanations carefully and familiarize yourself with the symbols before
reading the text.
The instrument and this manual use the following safety symbols:
Danger! Handle with Care.
This symbol indicates that the operator must refer to an explanation in
the User’s Manual in order to avoid the risk of personal injury or death
and/or damage to the instrument.
Double Insulation
This symbol indicates double insulation.
AC Voltage/Current
This symbol indicates AC voltage or current.
Ground
Indicates that there is a possibility of serious personal injury or loss of life if the
operating procedure is not followed correctly and describes the precautions for
avoiding such injury or loss of life.
Indicates that there is a possibility of serious personal injury of damage to the
instrument if the operating procedure is not followed correctly and describes the
precautions for avoiding such injury or damage.
NOTE
Draws attention to information essential for understanding the operation and
features.
This symbol indicates ground (earth)
WARNING
CAUTION
WARNING
● Never make measurement on a circuit above 600V AC.
● Do not use the instrument in an atmosphere where any flammable or explosive
gas is present.
● The transformer jaws are made of metal and their tips are not insulated. Be
especially careful about the hazard of possible shorting where the equipment
under test has exposed metal parts.
● Avoid using the instrument if it has been exposed to rain or moisture or if your
hands are wet.
● Do not exceed the maximum allowabIe input of any measurement range.
● Never open the battery compartment cover when making measurement.
● Do not use the instrument if there is any damage to the casing or when the
casing is removed.
● Do not turn the Function Selector switch with plugged in test leads connected
to the circuit under test.
● Do not install substitute parts or make any modification to the instrument.
Return the instrument to Yokogawa M&C or your distributor for repair or
re-calibration.
● Always switch off the instrument before opening the battery compartment cover
for battery replacement.
WARNING
To avoid damage to the instrument or electric shock!
The restrictions on the maximum voltage level for which the CL130/CL135 testers
can be used, depend on the over-voltage categories specified by the safety
standards. These category specifications are formulated to protect operators
against transient impulse voltage in power lines.
Model Function
CL130 A, V
CL135 A, V
Over-voltage category I (CAT.I):
Signal level, special equipment or parts of equipment, telecommunication,
electronic etc., with smaller transient over-voltages than CAT.II.
Over-voltage category II (CAT.II)
Local level, appliance, portable equipment etc., with smaller transient
over-voltages than CAT.III.
Over-voltage category III (CAT.III):
Distribution level, fixed installation, with smaller transient over-voltages than
CAT.IV.
Maximum Allowable Input
OVERVOLTAGE CATEGORY III
Measuring circuit voltage : 600Vrms AC
Measuring circuit voltage : 600Vrms AC
600A AC
600A AC
CAUTION
● Always make sure to insert each plug of the test leads fully into the appropriate
terminal on the instrument.
● Make sure to remove the test leads from the instrument before making current
measurement.
● Be sure to set the Function Selector switch to the "OFF" position after use.
When the instrument will not be in use for a long period of time, Place it in
storage after removing the battery.
● Use a damp cloth and detergent for cleaning the instrument. Do not use
abrasives or solvents.
NOTE
● Radiation immunity affects the accuracy of CL130/CL135 testers under the
conditions specified in IEC 801-3: 1984 (CL130)/IEC 801-3: 1993 (CL135).
● If equipment generating strong electromagnetic interference is located nearly,
the testers may malfunction.
1. Instrument Layout
Transformer Jaws
Open/Close Lever
Data Hold Switch
Lo Terminal
LCD Display
Data Hold Mode
Low Battery Warning
Function Selector
Switch
Display
Hi Terminal
Safety Hand Strap
2. Measurement
2.1 Preparation for Measurement
CAUTION
● The jaw section is a delicate, precision sensor. Do not subject the jaw to
unreasonably strong shock, vibration, or force when using it.
● If dust gets into the tops of the jaws, remove it immediately. Do not close the
jaws when dust is trapped in its joints as the sensor may break.
● Please check that the Function Selector switch is set to the desired position
2.2 AC Current Measurement
WARNING
Never use the instrument on a circuit above 600V AC. This may cause electrical
shock hazard and damage to the instrument or the circuit under test.
(1) Set the Function Selector switch to the " 200A " or " 600A " position.
(2) Press the open/close lever to open the transformer jaws and clamp them onto
a single conductor and take the reading on the display. The most accurate
reading will be obtained by keeping the conductor at the center of the
transformer jaws.
NOTE
● During current measurement, keep the transformer jaws fully closed.
Otherwise, accurate measurements cannot be taken. Maximum conductor size
is 30mm in diameter.
● When measuring larger current, the transformer jaws may buzz. This in not a
fault and does not affect the accuracy either.
WrongCorrect
2.3 AC Voltage Measurements
WARNING
Never use the instrument on a circuit above 600V AC. This may cause electrical
shock hazard and damage to the instrument or the circuit under test.
Red lead
Black lead
(1) Set the Function Selector switch to the " 200V " or " 600V " position.
(2) Plug the red test lead into Hi terminal and the black test lead into the Lo
terminal.
(3) Connect the tip of the red and black test leads to the circuit under test and
take the reading on the display.
2.4 Resistance Measurement and Continuity Check
WARNING
Never use the instrument on an energized circuit.
(1) Set the Function Selector switch to the " " position.
(2) Plug the red test lead into Hi terminal and the black test lead into the Lo
terminal. Check that "1" (over indication) is indicated on the LCD display.
(3) Short the tip of the test leads and check whether the display reads "0".
(4) Connect the tip of the test leads to the circuit under test. If the resistance is
30.0
or less, the buzzer beeps.
Ω
NOTE
When shorting the tip of the
test leads, the display may
read a very small resistance
instead of "0". This is the
resistance of the test leads on
the display.
Black test lead Red test lead
3. Other Functions
3.1 Data Hold Function
This is a function used to freeze the measured value on the display. Press the
Data Hold button to freeze the reading. The reading will be held regardless of
subsequent variation in input. "
display while the instrument is in the Data Hold mode.
To exit the Data Hold mode, press the Data Hold button again.
" is shown on the upper right corner of the
3.2 Optional Accessories
Clamp Adapter Model 99025 (For AC current measurement only)
Clamp Adapter Model 99025 is designed to
increase the measuring capability of a clamp
meter. With the use of the Clamp Adapter, you
can not only extend current range over 3000A,
but also clamp on a large bus-bar or conductor.
(1) Set the Function Selector switch to the
200A " or " 600A " position.
"
(2) As shown in the figure below, clamp Model
CL130/CL135 onto the pickup coil of Model
99025.
(3) Clamp Model 99025 onto the bus-bar or
conductor under test.
(4) Take the reading on Model CL130/CL135
and multiply it by 10.
NOTE
For detailed specification, refer to the Clamp
Adapter User’s Manual.
4. Battery Replacement
WARNING
To avoid electric shock hazard, make sure to set the Function Selector switch to
"OFF" and remove the test leads from the instrument before trying to replace
battery.
CAUTIOIN
Make sure to install battery in correct polarity as indicated in battery
compartment.
If the battery voltage becomes too low for the instrument to operate normally,
" is shown on the display. Then, replace the battery. Note that when the
"
battery is completely exhausted, the display blanks without "
(1) Set the Function Selector switch to the "OFF" position.
(2) Unscrew and remove the battery compartment on the bottom of the
instrument.
(3) Replace the battery observing correct polarity. Use a new 6LR61 (Alkaline
battery) or 6F22 (Manganese battery).
(4) Re-place and screw the battery compartment cover.
Battery
compartment cover
Battery
Screw
" shown.
5. Specifications
Instrument Specifications
● Measuring Ranges and Accuracy (at 23±5°C, relative humidity up to 75%)
CL130
Range Frequency Accuracy
200A AC
600A AC
200V AC
600V AC
200Ω
CL135
200A AC
600A AC
200V AC
600V AC
200Ω
Conversion method : AC coupled, true rms responding, calibrated to the rms
50/60Hz 1.5% rdg +6dgt sin
40 to 1kHz 2% rdg +5dgt
50/60Hz 1% rdg +3dgt
40 to 1kHz 2% rdg +5dgt
50/60Hz 1% rdg +2dgt
40 to 1kHz 1.5% rdg +4dgt
50/60Hz 1% rdg +2dgt
40 to 1kHz 1.5% rdg +4dgt
Range Frequency Accuracy
50/60Hz 1.5% rdg +4dgt
40 to 1kHz 2% rdg +5dgt sin
50/60Hz 1.5% rdg +4dgt
40 to 1kHz 2% rdg +5dgt sin
50/60Hz 1% rdg +2dgt
40 to 1kHz 1.5% rdg +4dgt sin
50/60Hz 1% rdg +2dgt
40 to 1kHz 1.5% rdg +4dgt sin
1.2% rdg +4dgt
Beeps below about 30
1.2% rdg +4dgt
Beeps below about 30
Ω
Ω
Crest Factor
↑
↑
↑
↑
↑
↑
↑
Crest Factor
3
≤
3
≤
3
≤
3
≤
General Specifications
● Operating System : Dual integration
● Measurement Function : AC current, DC current, Resistance, Continuity check
● Display : Liquid crystal display with maximum counts of 1999
● Overrange Indication : "1" is displayed on the highest digit.
● Response Time : Approx. 1 second (approx.2 seconds on resistance range)
● Temperature and Humidity for Guaranteed Accuracy : 23°C ±5°C, relative
humidity up to 75% without condensation
● Operating Temperature and Humidity : -10 to 50°C (without condensation)
up to 30°C, 90% relative humidity
up to 40°C, 75% relative humidity
up to 50°C, 45% relative humidity
● Storage Temperature and Humidity : -20 to 60°C, relative humidity up to 75%
without condensation
● Effect of conductor position :
CL130 Within 2% of indicated value at the center to a 10 mm-dia conductor
carrying 100A, at every part inside the jaws
CL135 Within 3% of indicated value at the center to a 10 mm-dia conductor
carrying 100A, at every part inside the jaws
● Effect of external magnetic field : 2 A or less in AC magnetic field of 400 A/m
● Power Source : 6F22(9V DC) or equivalent battery
● Battery Life : Approx. 200 hours (continuous)
● Current Consumption : Approx. 2mA
● Withstanding Voltage : 5550V AC for 1 minute between housing case and
metal part of jaws
● Insulation Resistance : 10M
metal part of jaws
● Conductor Size : Approx. 30mm diameter max.
● Dimensions : Approx. 91(W) x 208(H) x 40(D) mm
● Weight : Approx. 400g (batteries included)
● Safety Standard: EN 61010-1
EN 61010-2-31
EN 61010-2-32
(600V AC CAT.III, Pollution degree2, indoor use)
● EMC Standard: EN 55022
● Radiation immunity : IEC 801-3
● Accessories : Test leads Model 98010····················· 1set
6F22 battery······································· 1
Carrying case Model 93032··············· 1
User’s Manual···································· 1
● Optional Accessories : Clamp adapter Model 99025
or greater at 1000V between housing case and
Ω
6. Calibration and After-sales Service
Should any failure occur while you are using the tester, follow the instructions
given below. If the tester still fails to operate correctly and needs repair, contact
the vendor from whom you purchased the instrument or the nearest Yokogawa
M&C sales office.
● Turn off the POWER switch once, then turn it back on again.
● If the tester does not turn on, replace the battery with a new one.
Calibration
It is recommended that the instrument be calibrated once every year.
YOKOGAWA M&C CORPORATION
International Sales Dept.
2-9-32 Nakacho, Musashino-shi, Tokyo, 180-8750 Japan
Phone: 81-422-52-5716 Facsimile: 81-422-55-8654
YOKOGAWA CORPORATION OF AMERICA (U.S.A.)
Phone: 1-770-253-7000 Facsimile: 1-770-251-2088
YOKOGAWA EUROPE B. V. (THE NETHERLANDS)
Phone: 31-334-64-1611 Facsimile: 31-334-64-16 10
YOKOGAWA AMERICA DO SUL S. A. (BRAZIL)
Phone: 55-11-5681-2400 Facsimile: 55-11-5681-1274
YOKOGAWA ENGINEERING INSTRUMENTS KOREA CORPORATION (KOREA)
Phone: 82-2-551-0660 to -0664 Facsimile: 82-2-551-0665
YOKOGAWA AUSTRALIA PTY. LTD. (AUSTRALIA)
Phone: 61-2-9805-0699 Facsimile: 61-2-9888-1844
YOKOGAWA BLUE STAR LTD. (INDIA)
Phone: 91-80-227-1513 Facsimile: 91-80-227-4270
LTD. YOKOGAWA ELECTRIC (RUSSIAN FEDERATION)
Phone: 7-095-737-7868 Facsimile: 7-095-737-7869
KIM3E-2003.2