This instrument has been already adjusted at the
factory before shipment.
To ensure correct use of the instrument, please read
this manual thoroughly and fully understand how to
operate the instrument before operating it.
䊏 Regarding This Manual
•This manual should be passed on to the end user.
• Before use, read this manual thoroughly to comprehend its contents.
• The contents of this manual may be changed
without prior notice.
• All rights reserved. No part of this manual may be
reproduced in any form without Yokogawa’s written
permission.
•Yokogawa makes no warranty of any kind with
regard to this material, including, but not limited to,
implied warranties of merchantability and suitability
for a particular purpose.
• All reasonable effort has been made to ensure the
accuracy of the contents of this manual. However, if
any errors are found, please inform Yokogawa.
• Yokogawa assumes no responsibilities for this
product except as stated in the warranty.
• If the customer or any third party is harmed by the
use of this product, Yokogawa assumes no responsibility for any such harm owing to any defects in the
product which were not predictable, or for any
indirect damages.
䊏 Safety Precautions
• The following general safety precautions must be
observed during all phases of operation, service, and
repair of this instrument. Failure to comply with
these precautions or with specific WARNINGS
given elsewhere in this manual violates safety
standards of design, manufacture, and intended use
of the instrument. YOKOGAWA Electric Corporation assumes no liability for the customer’s failure to
comply with these requirements. If this instrument is
used in a manner not specified in this manual, the
protection provided by this instrument may be
impaired.
The following safety symbol marks are used in
this manual and instrument;
WARNING
A WARNING sign denotes a hazard. It calls
attention to procedure, practice, condition or the
like, which, if not correctly performed or adhered
to, could result in injury or death of personnel.
CAUTION
A CAUTION sign denotes a hazard. It calls
attention to procedure, practice, condition or the
like, which, if not correctly performed or adhered
to, could result in damage to or destruction of
part or all of the product.
IMPORTANT
A IMPORTANT sign denotes an attention to
avoid leading to damage to instrument or system
failure.
NOTE
A NOTE sign denotes a information for essential
understanding of the operation and features.
Protective grounding terminal.
Function grounding terminal. This terminal
should not be used as a “Protective grounding
terminal”.
Alternating current.
Direct current.
1-1
IM 1E6D0-01E
䊏 Warranty
• The guaranteed term of this instrument is described
in the quotation. We repair the damages that
occurred during the guaranteed term for free.
• Please contact with our sales office when this
instrument is damaged.
• If the instrument has trouble, please inform us
model code, serial number, and concrete substances
or situations. It is preferable to be attached a outline
or data.
• We decide after the examination if free repair is
available or not.
• Please consent to the followings for causes of
damages that are not available as free repair, even if
it occured during the guaranteed term.
A:Unsuitable or insufficient maintenance by the
customer.
B: The handling, using, or storage that ignore the
design and specifications of the instrument.
C: Unsuitable location that ignore the description in
this manual.
D:Remaking or repair by a person except whom we
entrust.
E: Unsuitable removing after delivered.
F: A natural disaster (ex. a fire, earthquake, storm and
flood, thunderbolt) and external causes.
1. INTRODUCTION
For the safety using;
WARNING
• The Magnetic Flow Tube is a heavy instrument.
Please give attention to prevent that persons
are injured by carrying or installing. It is preferable for carrying the instrument to use a cart
and be done by two or more persons.
• In wiring, please confirm that the cables are not
powered before connecting.
• When removing the instrument from hazardous
processes, avoid contact with the fluid and the
interior of the meter.
• In case of Hazardous duty type instrument,
further requirements and differences are
described in Chapter 6 “HAZARDOUS DUTY
TYPE INSTRUMENT”. The description in
Chapter 6 is prior to other description in this
instruction manual.
1-2
IM 1E6D0-01E
2.HANDLING PRECAUTIONS
2. HANDLING PRECAUTIONS
This instrument has been already tested thoroughly at
the factory. When the instrument is delivered, please
check externals and make sure that no damage
occurred during transportation.
In this chapter, handling precautions are described.
Please read this chapter thoroughly at first. And please
refer to the relative matter about other ones.
If you have any problems or questions, please make
contact with Yokogawa sales office.
2.1Checking Model and
Specifications
The model and specifications are shown on the Data
Plate. Please confirm the specifications between the
instrument that was delivered and the purchase order
(refer to the chapter 5. Outline).
Please let us know Model and Serial No. when making
contact with Yokogawa sales office.
MAGNETIC
FLOW DETECTOR
AM
L
H
mm
2.3Storage Precautions
In case the instrument is expected to be stored over a
long term, please give attention to the followings;
• The instrument should be stored in its original
packing condition.
• The storage location should be selected according to
the following conditions:
1) The location where it is not exposed to rain or
water.
2) The location where there is few vibration or
shock.
3) Temperature and humidity should be:
Temperature: –30 to 60˚C (–22 to 140˚F)
Humidity:5 to 80% RH (no condensation)
Preferable ambient temperature and humidity
are 25˚C(77˚F) and about 65% RH.
2.4Installation Location
Precautions
Please select the installation location considering the
following items to ensure long term stable operation of
the flow tube.
Made in Japan
F0201.EPS
Figure 2.1Data Plate
2.2Accessories
When the instrument is delivered, please make sure
that the following accessories are in the package.
• Centering device 1-set (for wafer type)
•Hexagonal wrench 1-piece (for hazardous duty type
except JIS flameproof type)
•Hexagonal wrench 2-piece (only for JIS flameproof
type)
• Ambient Temperature:
Please avoid to install the instrument at the location
where temperature changes continuously. If the
location receives radiant heat from the plant, provide
heat insulation or improve ventilation.
•Atmospheric Condition:
Please avoid to install the instrument in an corrosive
atmosphere. In case of installing in the corrosive
atmosphere, please keep ventilating sufficiently and
prevent rain from entering the conduit.
•Vibration or shock:
Please avoid to install the instrument at the location
where there is heavy vibration or shock.
2.5Terminal Box Reori-
entation Precautions
Please do not change the terminal box orientation at
the customer’s site. If the terminal box reorientation is
required, please contact Yokogawa office or service
center.
2-1
IM 1E6D0-01E
3.INSTALLATION
3. INSTALLATION
CAUTION
This instrument must be installed by expert
engineer or skilled personnel. The procedures
described in this chapter are not permitted for
operators.
3.1Piping Design Precautions
IMPORTANT
Please design the correct piping referring to the
followings to prevent damage for flow tube and
to keep correct measuring.
(1) Location
Gate valve
fully open
5 D or more 2 D or more2 D or more
Tee
Figure 3.1.1 Minimum Length of Required Straight Run
Reducer pipeExpander pipe
Do not care to 0 D Do not care to 0 D
90° degrees bent
D: Nominal diameter of flow tube
10 D or more
Various types of valves
10 D or more5 D or more5 D or more Do not care to 0 DDo not care to 0 D
NOTE
1. Nothing must be inserted or installed in the
metering pipe than may interfere with the
magnetic field, induced signal voltages, and
flow velocity distribution.
2. These straight runs may not be required on
the downstream side flowmeter. However, if
the downstream valve or other fittings cause
channeling on the upstream side, provide a
straight run of 2 D to 3 D on the downstream
side.
2 D or more
F030101.EPS
IMPORTANT
Please install the flow tube to the location where
it is not exposed to direct sunlight and ambient
temperature is –10 to + 60°C (14 to 140°F).
(2) Noise Rejection
IMPORTANT
The instrument should be installed away from
large electrical motors, transformers and other
power sources in order to avoid interference with
the measurement.
(3) Length of Straight Run
To keep accurate measuring, JIS B7554 “Electro
Magnetic Flow Tubes” explains about upstream piping
condition of Magnetic Flow Tubes.
We recommend to our customers about the piping
conditions shown in Figure 3.1.1 based on JIS B7554
and our piping condition test data.
(4) Liquid Conductivity
IMPORTANT
Please avoid to install the flow tube at location
where liquid conductivity is likely to be nonuniform. Because it is possible to have bad
influences to the flow indication by non-uniform
conductivity when a chemical liquid is injected
from upstream side close to the flow tube. When
this occurs, it is recommended that chemical
application ports are installed on the downstream
side of the flow tube. In case chemicals must be
added upstream side, please keep the pipe
length enough so that liquid is properly mixed.
(BAD)
Upstream side
Figure 3.1.2 Chemical Insertion
(GOOD)
Downstream side
F030102.EPS
3-1
IM 1E6D0-01E
3. INSTALLATION
(5) Liquid Sealing Compound
IMPORTANT
Please give attention in using Liquid Sealing
Compound to the piping, because it brings bad
influences to measurement by flowing out and
cover the surfaces of electrode and earth-ring.
(6) Service Area
Please select the location where there is enough area to
service installing, wiring, overhaul, etc.
(7) Bypass Line
It is recommended to install the Bypass Line to
facilitate maintenance and zero adjustment.
Bypass valve
Block
valve
Block valve
F030103.EPS
Please design the piping that a fluid is always filled in
the pipes. The Vertical Mounting is effective for fluids
that is easily separate or slurry settles within pipes.
In this case, please flow a fluid from bottom to up.
(GOOD)
h
h>0
Horizontal mountingVertical mounting
(BAD)
(GOOD)
Figure 3.1.4 Filling the Pipe with Liquid
h>0
(BAD)
h
F030104.EPS
(10) No Air Bubbles
IMPORTANT
Please give attention to prevent bad influences
or measuring errors from air bubbles that gathers inside measuring pipes.
In case the fluid includes air bubbles, please design the
piping that prevent to gather air bubbles. In case valves
are installed upstream of the flow tube, it is possible
that a valve causes air bubbles, please install the flow
tube upstream side of a valve.
Figure 3.1.3 Bypass Line
(8) Supporting the Flow Tube
CAUTION
Please avoid to support only the flow tube, but
fix pipes at first and support the flow tube by
pipes to protect the flow tube from forces caused
by vibration, shock, expansion and contraction
through piping.
For small sized flow tubes, please provide a
mounting base so thet the tubes are fixed in the
piping. See the section 3.3 Mounting.
(9) Piping Condition
IMPORTANT
The piping should be designed so that a full pipe
is maintained at all times to prevent loss of
signal and erroneous readings.
(GOOD)
(BAD)
(GOOD)
Figure 3.1.5 Avoiding Air Bubbles
(11) Mounting Direction
IMPORTANT
When the electrodes are vertical to ground, the
electrode is covered with air bubbles at upper
side or slurry at downside, and it may cause the
measuring errors.
Please be sure to mount the terminal box upper
side of piping to prevent water penetration into
terminal box.
(Good)
(No good)
Electrodes
Electrodes
(No good)
Air Bubbles
Slurry
Figure 3.1.6 Mounting Direction
3-2
(BAD)
F030105.EPS
Water slowly
percolates
downward into
terminal box.
F030106.EPS
IM 1E6D0-01E
3. INSTALLATION
(12) Grounding
IMPORTANT
Improper grounding can have an adverse affect
on the flow measurement. Please ensure that
the instrument is properly grounded.
The electromotive force of the magnetic flow tube is
minute and it is easy to be affected by noise. And also
that reference electric potential is the same as the
measuring fluid potential. Therefore, the reference
electric potential (terminal potential) of the Flow Tube
and the Converter/Amplifier also need to be the same
as the measuring fluid. And moreover, that the potential must be the same with ground.
Please be sure to ground according to Figure 3.1.7.
600 V vinyl insulated
electric cable
2
or larger)
(2 mm
3.2Handling Precautions
WARNING
The Magnetic Flow tube is a heavy instrument.
Please be careful to prevent persons from
injuring when it is handled.
3.2.1General Precautions
(1) Precaution for Carrying
The Magnetic Flow Tube is packed tightly. When it is
unpacked, please give attention to prevent damages to
the flow tube. And to prevent the accident during carry
to the installing location, please carry it near the
location keeping packed as it delivered.
CAUTION
Earth ring
Grounding resistance 100 Ω or less
(10 Ω or less for JIS flameproof type)
In case earth rings are used.
Note; See "3.4.1 Protective Grounding" for information on
protective grounding.
Figure 3.1.7 Grounding
In case earth rings are not used.
(Available only for metal piping)
F030107.EPS
In case the Magnetic Flow Tube lifts up, please
refer to Figure 3.2.1. Please never lift up by
using a bar through the flow tube. It damages
liner severely.
Horizontal Lifting
Figure 3.2.1 Lifting Flow Tubes
Vertical Lifting Sling
Rigging Method
F030201.EPS
3-3
IM 1E6D0-01E
3. INSTALLATION
(2) Precaution for Shock
CAUTION
Care should be taken not to drop the flow tube
or subject it to excessive shock. This may lead
to liner damage which will cause inaccurate
readings.
(3) Flange Protection Covers
IMPORTANT
Please keep the protection cover (ex. corrugated
paper or anything possible to protect) attached
with flange except when mounting to the pipe.
(4) Terminal Box Cover
IMPORTANT
3.2.2Flow Tube Piping
CAUTION
Mis-aligned or slanted piping can lead to leakage
and damage to flanges.
•Please correct mis-alignment or slanted piping and
improper distance between mounting flanges before
install the flow tube. (Please refer to Figure 3.2.2)
• Inside a pipeline which is newly installed, some
foreign substances (such as welding scrap or wood
chips) may exist. Please remove them by flushing
piping before mounting the flow tube.
Slant
Mis-alignment
F030202.EPS
Please never leave the terminal box cover open
until wiring to prevent insulation deterioration.
(5) Long-term Non-use
IMPORTANT
It is not preferable to leave the flow tube for long
term non-use after installation.
In case the flow tube is compelled to do that,
please take care of the flow tube by the followings.
• Confirmation of Sealing Condition for the Flow
Tube
Please confirm the sealing conditions of the terminal
box screw and wiring ports.
In case of the Conduit Piping, please provide the
drain plugs or waterproof glands to it to prevent that
moisture or water penetrates into the flow tube
through the conduit.
• Regular Inspections
Please inspect the sealing condition (as above
mentioned) and inside of the terminal box. And
when it is suspect that water penetration into the
inside flow tube (ex. rain fall), please inspect when
it happened.
Figure 3.2.2 Slant and Mis-alignment of Flow Tube Piping
3-4
IM 1E6D0-01E
3. INSTALLATION
3.3Mounting
3.3.1Nominal Diameter 2.5 mm (0.1˝)
to 10 mm (0.4˝) Union Joint
Type
NOTE
Please ensure to use the attached connecting
fittings.
Ceramic linings with a diameter of 2.5, 5 or 10 mm are
connected using union joints.
Use the connecting fittings according to the table
below. Depending on whether the fitting is to be
welded to or screwed on to the piping.
Table 3.3.1Fitting Dimensions
Screw joint (Code: U2, U3)
D
(1) Mounting Direction
Please mount the Magnetic Flow Tube matching the
flow direction of the fluid to be measured with the
direction of the arrow mark on the flow tube.
IMPORTANT
If it is impossible to match the direction, please
never remodel by changing direction of the
terminal box. In case the measuring fluid flows
against the arrow direction, please refer to the
section “Reversing Flow Direction” in the Instruction Manual of AM11 Magnetic Flow Converter.
(2) Connecting Process Piping
Weld or screw the connecting fittings to the process
piping.
IMPORTANT
ØA
(0.16")
Size
mm(inch)
2.5(0.1)
5(0.2)
10(0.4)
ØAØBØB
(0.16")
Size
mm(inch)
2.5(0.1)
5(0.2)
10(0.4)
0
–0.1
ØC
11.5(0.4")
4
30(1.2")
CodeABCD
ØC
R1/4(PT1/4)
R1/4(PT1/4)
R3/8(PT3/8)
0.3
0
ØD
18.5(0.73")
18.5(0.73")
22.5(0.89")
22(0.87")
U2
22(0.87")
U3
22(0.87")
U2
22(0.87")
U3
25(0.98")
U2
25(0.98")
U3
Weld joint (Code: U1)
4
CodeABCD
22(0.87")
U1
22(0.87")
U1
25(0.98")
U1
8(0.31")
8(0.31")
8(0.31")
8(0.31")
10(0.39")
10(0.39")
35(1.38")
8(0.31")
8(0.31")
10(0.39")
18.5(0.73")
18.5(0.73")
18.5(0.73")
18.5(0.73")
22.5(0.89")
22.5(0.89")
10(0.39")
14.3(0.56")
14.3(0.56")
17.8(0.70")
NPT1/4
NPT1/4
NPT3/8
0
–0.1
T030301.EPS
• Please be sure to pass the connecting fittings
through the union joint nuts in advance. Then
connect the connecting fitting to the piping by
screwing or welding the connecting fitting to the
piping (see Figure 3.3.1).
• In case of weld joint type, please pay attention
the welding condition to avoid deforming piping
or making the stagnant portion of the fluid;
joint preparation, level defference in butt joint,
welding current.
(3) Positioning the Flow Tube
Install the magnetic flow tube on a mounting base and
position it so that the center axis of the flow tube is
aligned with that of the piping. Then mount the flow
tube to union joint nuts by screwing the nuts to the
connecting ports of the flow tube.
CAUTION
The ceramic pipe will be damaged if they are
tightened when they are not properly aligned.
3-5
IM 1E6D0-01E
3. INSTALLATION
(4) Tightening Nuts
CAUTION
Tighten the union joint nuts according to Torque
Values in Table 3.3.1 using a torque wrench. As
the gasket material is Fluorocarbon PTFE, it is
possible that nuts may loose by it’s character as
time passes. Please tighten the nuts regularly.
The table below shows the tightening torque
values. Be sure to use gasket : t=1.5 attached.
3.3.2Nominal Diameter 2.5mm (0.1˝)
to 40mm (1.5˝) Wafer Type
IMPORTANT
Please use appropriate bolts and nuts according
to process connection. In case stud type of
through bolts are used, be sure outside diameter
of a shank is smaller than a thread ridge’s one.
Please use compressed non-asbestos fiber
gasket, PTFE gasket or the gasket which has
equal elasticity. In case of optional code/FRG,
please use rubber gasket or others which has
equal elasticity. Be sure the inner diameter of
the gasket does not protrude to inner piping.
(Refer to Table 3.3.8)
Horizontal Mounting
(1) Mounting Direction
Please mount the Magnetic Flow Tube matching the
flow direction of the fluid to be measured with the
direction of the arrow mark on the flow tube.
IMPORTANT
If it is impossible to match the direction, please
never remodel by changing direction of the
terminal box. In case the measuring fluid flows
against the arrow direction, please refer to the
section “Reversing Flow Direction” in the Instruction Manual of AM11 Magnetic Flow Converter.
(2) Mounting Centering Devices
To keep concentricity of the Flow Tube with pipes,
please mount centering devices on the Mini-Flanges of
the Flow Tube.
Please give attention to the nominal diameter and
flange ratings of the centering devices.
Vertical Munting
Figure 3.3.1 Mounting Procedure (Size: 2.5 mm(0.1") to
10 mm(0.4"))
F030301.EPS
3-6
IM 1E6D0-01E
3. INSTALLATION
(3) Positioning Flow Tube
Please pass two through-bolts to adjacent holes of both
flanges and mount the Flow Tube, and pass other
through-bolts to other holes. (Refer to Figure 3.3.2/
3.3.3) In case stud type of through-bolts are used,
position them coming in contact centering devices with
thread of bolts.
(4) Tightening Nuts
Please tighten the bolts according to Torque Values in
Table 3.3.3. In case of PVC piping, please select
optional code /FRG, use rubber gasket and tighten with
the torque value in Table 3.3.4.
Through-bolts (four)
Nuts
Please use appropriate bolts and
nuts according to process
connection.
CAUTION
In case of PFA lining type, as the lining material
is Fluorocarbon PFA, it is possible that nuts may
loose by its character as time passes. Please
tighten the nuts regularly.
Please be sure to tighten the bolts following
prescribed torque values. Please tighten the
flange bolts diagonally with the same torque
values, step by step up to the prescribed torque
value.
Gasket (two)
Mini-flange
Centering devices (two)
Flange
Nuts (eight)
Horizontal MountingVertical Mounting
Figure 3.3.2 Mounting Procedure (Size: 2.5 mm(0.1") to 15 mm(0.5"))
Through-bolts (four)
Gasket (two)
Mini-flange
Nuts (eight)
Please use appropriate
bolts and nuts according
to process connection.
Table 3.3.3Tightening Torque Values for Metal Piping
Tightening Torque Values for PFA lining / Polyurethane lining Type N-m {kgf-cm} [in-lbf]
Size
mm(inch)
2.5(0.1), 5(0.2)
10(0.4), 15(0.5)
25(1)
40(1.5)
Flange
Rating
6 to 9
{61 to 92}
[53 to 80]
15 to 22
{153 to 224}
[133 to 195]
21 to 32
{214 to 327}
[186 to 283]
JISANSIDIN
10K
20K
6 to 9
{61 to 92}
[53 to 80]
15 to 22
{153 to 224}
[133 to 195]
21 to 32
{214 to 327}
[186 to 283]
150
6 to 9
{61 to 92}
[53 to 80]
12 to 18
{122 to 184}
[106 to 159]
17 to 26
{173 to 265}
[150 to 230]
300
6 to 9
{61 to 92}
[53 to 80]
15 to 22
{153 to 224}
[133 to 195]
25 to 38
{255 to 388}
[221 to 336]
{112 to 173}
{214 to 327}
[186 to 283]
Maximum Tightening Torque Values for Ceramic lining Type N-m {kgf-cm} [in-lbf]
Size
mm(inch)
15(0.5)
25(1)
40(1.5)
Flange
Rating
{143}
[124]
{306}
[265]
{449}
[389]
10K
14
30
44
JIS
20K
14
{143}
[124]
30
{306}
[265]
44
{449}
[389]
150
14
{143}
[124]
22
{224}
[195]
33
{337}
[292]
ANSI
300
14
{143}
[124]
30
{306}
[265]
51
{520}
[451]
3. INSTALLATION
PN10/16
6 to 9
{61 to 92}
[53 to 80]
11 to 17
[97 to 150]
21 to 32
DIN
PN10/16
14
{143}
[124]
25
{255}
[221]
50
{510}
[442]
*Please use compressed non-asbestos fiber gasket, PTFE gasket or the gasket which has equal elasticity.
T030303.EPS
Table 3.3.4Tightening Torque Values for PVC Piping
Tightening Torque Values for PFA lining / Polyurethane lining Type N-m {kgf-cm} [in-lbf]
Size
mm(inch)
2.5(0.1), 5(0.2)
10(0.4), 15(0.5)
25(1)
40(1.5)
Flange
Rating
10K
2.0
{20}
[18]
5.2
{53}
[46]
7.4
{76}
[65]
JIS
20K
—
—
—
150
2.1
{21}
[19]
4.2
{43}
[37]
6.0
{61}
[53]
ANSI
300
—
—
—
DIN
PN10/16
2.0
{20}
[18]
4.0
{41}
[35]
7.4
{76}
[65]
Maximum Tightening Torque Values for Ceramic Type N-m {kgf-cm} [in-lbf]
Size
mm(inch)
15(0.5)
40(1.5)
25(1)
Flange
Rating
10K
1.3
{13}
[12]
3.5
{36}
[31]
5.7
{58}
[50]
JIS
20K
—
—
—
150
1.3
{13}
[12]
2.8
{29}
[25]
4.6
{47}
[41]
ANSI
300
—
—
—
DIN
PN10/16
1.3
{13}
[12]
2.7
{28}
[24]
5.7
{58}
[50]
*Please select optional code/FRG and use rubber gasket or others which has equal elasticity.
T030304.EPS
3-8
IM 1E6D0-01E
3. INSTALLATION
3.3.3Nominal Diameter 50 mm(2˝) to
200 mm(8˝) Wafer Type
IMPORTANT
Please use appropriate bolts and nuts according
to process connection. In case stud type of
through bolts are used, be sure outside diameter
of a shank is smaller than a thread ridge’s one.
Please use compressed non-asbestos fiber
gasket, PTFE gasket or the gasket which has
equal elasticity. In case of optional code/FRG,
please use rubber gasket or others which has
equal elasticity. Be sure the inner diameter of
the gasket does not protrude to inner piping.
(Refer to Table 3.3.8)
(1) Mounting Direction
Please mount the Magnetic Flow Tube matching the
flow direction of the fluid to be measured with the
direction of the arrow mark on the flow tube.
IMPORTANT
NOTE
When installing a size 150mm with JIS F12
(JIS75M) flange, please displace in circumferential direction slightly because the cover of
electrode chamber will interfere against the bolts.
(4) Tightening Nuts
Please tighten the bolts according to Torque Values in
Table 3.3.5. In case of PVC piping, please select
optional code/FRG, use rubber gasket and tighten with
the torque value in Table 3.3.6.
CAUTION
In case of PFA lining type as the lining material
is Fluorocarbon PFA, it is possible that nuts
loose by its character as time passes. Please
tighten the nuts regularly.
Please be sure to tighten the bolts following
prescribed torque values. Please tighten the
flange bolts diagonally with the same torque
values, step by step up to the prescribed torque
value.
If it is impossible to match the direction, please
never remodel to change direction of the terminal box. In case the measuring fluid flows
against the arrow direction, please refer to the
section “Reversing Flow Direction” in the Instruction Manual of AM11 Magnetic Flow Converter.
(2) Mounting Centering Devices
To keep concentricity between the Flow Tube and
pipes, centering devices must be used. Pass two
through-bolts through the four centering devices (two
for each) and lower adjacent holes of both flanges.
(Refer to Figure 3.3.4)
Please give attention to the nominal size and flange
ratings of the centering devices. (Refer to Table 3.3.7)
(3) Positioning Flow Tube
Position the Flow Tube coming in contact four centering devices with Mini-Flanges. At this time, pay
attention to avoid four centering devices come in
contact with Housing. In case stud type of throughbolts are used, position them coming in contact four
centering devices with thread of the bolts. (Refer to
Figure 3.3.4) After positioning the Flow Tube, pass
remaining through-bolts to remaining holes.
Nuts
Please use appropriate
bolts and nuts
according to Process
connection.
Horizontal MountingVertical Mounting
Figure 3.3.4 Mounting Procedure
Through bolts
Gaskets (two)
Flanges
Housing
Mini-Flange
Centering devices (four)
F030304.EPS
3-9
IM 1E6D0-01E
Table 3.3.5Tightening Torque Values for Metal Piping
Tightening Torque Values for PFA lining / Polyurethane lining Type N-m {kgf-cm} [in-lbf]
Size
mm(inch)
Maximum Tightening Torque Values for Ceramic lining Type N-m {kgf-cm} [in-lbf]
Size
mm(inch)
*Please use compressed non-asbestos fiber gasket, PTFE gasket or the gasket which has equal elasticity.
Flange
Rating
Flange
Rating
{286 to 429}
[248 to 372]
{214 to 316}
[186 to 274]
{357 to 520}
[310 to 451]
{663 to 959}
[575 to 832]
{582 to 857}
[504 to 743]
50(2)
80(3)
100(4)
150(6)
200(8)
50(2)
80(3)
100(4)
150(6)—
200(8)
JISANSIDIN
10K20K150300PN10
28 to 42
21 to 31
35 to 51
65 to 94
57 to 84
10K20K
50
{510}
[442]
36
{367}
[319]
48
{490}
[425]
79
{806}
[699]
70
{714}
[619]
{143 to 214}
[124 to 186]
{265 to 398}
[230 to 345]
{439 to 653}
[381 to 566]
{439 to 694}
[381 to 602]
{622 to 939}
[540 to 814]
JIS
14 to 21
26 to 39
43 to 64
43 to 68
61 to 92
27
{276}
[239]
44
{449}
[389]
58
{592}
[513]
55
{561}
[487]
76
{776}
[673]
28 to 42
{286 to 429}
[248 to 372]
41 to 61
{418 to 622}
[363 to 540]
35 to 51
{357 to 520}
[310 to 451]
63 to 89
{643 to 908}
[558 to 788]
93 to 120
{949 to 1224}
[823 to 1062]
ANSI
150300
50
{510}
[442]
75
{765}
[664]
49
{500}
[434]
66
{673}
[584]
102
{1041}
[903]
14 to 21
{143 to 214}
[124 to 186]
25 to 37
{255 to 378}
[221 to 327]
41 to 61
{418 to 622}
[363 to 540]
41 to 60
{418 to 612}
[363 to 531]
65 to 93
{663 to 949}
[575 to 823]
27
{276}
[239]
44
{449}
[389]
56
{571}
[496]
43
{439}
[381]
76
{776}
[673]
{663 to 959}
[575 to 832]
94 to 125
{959 to 1276}
[832 to 1106]
65 to 94
DIN PN
PN10PN16
—
—
—
103
{1051}
[911]
Table 3.3.6 Tightening Torque Values for PVC Piping
Tightening Torque Values for PFA lining / Polyurethane lining Type N-m {kgf-cm} [in-lbf]
Size
mm(inch)
Maximum Tightening Torque Values for Ceramic lining Type N-m {kgf-cm} [in-lbf]
Size
mm(inch)
*Please select optional code/FRG and use rubber or gasket or others which has equal elasticity.
Flange
Rating
50(2)———
80(3)
100(4)
150(6)
200(8)
Flange
Rating
50(2)
80(3)
100(4)
150(6)
200(8)
JISANSIDIN
10K20K150300PN10
9.8
{100}
[87]
7.2
{73}
[64]
12.2
{124}
[108]
21.4
{218}
[189]
19.3
{197}
[171]
JIS
10K20K
8.2
{84}
[73]
6.2
{63}
[55]
8.0
{82}
[71]
19.8
{202}
[175]
17.5
{179}
[155]
———
——
——
——
————
———
———
——
——
9.8
{100}
[87]
14.4
{147}
[127]
12.2
{124}
[108]
20.5
{209}
[181]
27.6
{282}
[244]
ANSI
150300
8.2
{84}
[73]
12.4
{127}
[110]
8.1
{83}
[72]
18.9
{193}
[167]
25.1
{256}
[222]
—
21.4
{218}
[189]
28.9
{295}
[256]
PN10PN16
19.8
{202}
[175]
26.2
{267}
[232]
DIN
PN16
28 to 42
{286 to 429}
[248 to 372]
21 to 31
{214 to 316}
[186 to 274]
35 to 52
{357 to 531}
[310 to 460]
65 to 94
{663 to 959}
[575 to 832]
58 to 84
{592 to 857}
[513 to 743]
63
{643}
[558]
36
{367}
[319]
48
{490}
[425]
76
{776}
[673]
67
{684}
[593]
PN16
9.8
{100}
[87]
7.2
{73}
[64]
12.2
{124}
[108]
21.4
{218}
[189]
19.3
{197}
[171]
8.2
{84}
[73]
6.2
{63}
[55]
8.0
{82}
[71]
19.8
{202}
[175]
17.5
{179}
[155]
JIS G3451
F12
41 to 62
{418 to 633}
[363 to 549]
68 to 102
{694 to 1041}
[602 to 903]
68 to 100
{694 to 1020}
[602 to 885]
69 to 101
{704 to 1031}
[611 to 894]
JIS G3451
F12
80
{816}
[708]
105
{1071}
[929]
84
{857}
[743]
102
{1041}
[903]
T030305.EPS
JIS G3451
F12(75M)
—
14.4
{147}
[127]
24.4
{249}
[216]
23.4
{239}
[207]
31.6
{322}
[280]
JIS G3451
F12(75M)
12.3
{126}
[109]
16.1
{164}
[142]
21.6
{220}
[191]
28.7
{293}
[254]
T030306.EPS
3. INSTALLATION
3-10
IM 1E6D0-01E
Table 3.3.7Centering Device Identification
Flange Rating
Liningmm(inch)
50(2)
PFA,
polyurethane
80(3)
100(4)
150(6)
200(8)
50(2)
80(3)
Ceramics
100(4)
150(6)
200(8)
JIS 10K
JIS 20K
B
B
B
C
C
B
B
B
B
B
*Each Centering Devices is engraved a character as identifidation.
*Please ensure that the I.D. of the gasket does not protrude into the I.D. of the Earth Ring.
(This dimension is also applied when no earth ring is used)
*[ ] ; for Process connection code G1
3. INSTALLATION
T030308.EPS
3-11
IM 1E6D0-01E
3. INSTALLATION
3.3.4Nominal Diameter 15 mm (0.5˝)
to 400 mm (16˝) Flange Type
IMPORTANT
Please use appropriate bolts and nuts according
to process connection. Please use compressed
non-asbestos fiber gasket, PTFE gasket or the
gasket which has equal elasticity. In case of
optional code/FRG, please use rubber gasket or
others which has equal elasticity. Be sure the
inner diameter of the gasket does not protrude to
inner piping.(Refer to Table 3.3.8)
(1) Mounting Direction
Please mount the Magnetic Flow Tube matching the
flow direction of the fluid to be measured with the
direction of the arrow mark on the flow tube.
IMPORTANT
(2) Tightening Nuts
Please tighten the bolts according to Torque Values in
Table 3.3.9. In case of PVC piping, please select
optional code/FRG, use rubber gasket and tighten with
the torque value in Table 3.3.10.
CAUTION
In case of PFA lining type as the lining material
is Fluorocarbon PFA, it is possible that bolts
loose by its character as time passes. Please
tighten the nuts regularly.
Please be sure to tighten the bolts following
prescribed torque values. Please tighten the
flange bolts diagonally with the same torque
values, step by step up to the prescribed torque
value.
If it is impossible to match the direction, please
never remodel to change direction of the terminal box. In case the measuring fluid flows
against the arrow direction, please refer to the
section “Reversing Flow Direction” in the Instruction Manual of AM11 Magnetic Flow Converter.
Gasket (two)
Flange (Flow tube)
Bolts
Please use appropriate bolts and
nuts according to process
conneciton.
Figure 3.3.5 Mounting Procedure (Size: 15 mm (0.5") to 400 mm (16"))
Nuts
Flange
(Piping side)
3-12
F030305.EPS
IM 1E6D0-01E
Table 3.3.9Tightening Torque Values for Metal Piping in N-m{kgf-cm} [in-lbf]
15(0.5mm)
25(1mm)
40(1.5mm)
50(2mm)
80(3mm)
100(4mm)
150(6mm)
200(8mm)
250(10mm)
300(12mm)
350(14mm)
400(16mm)
Flange
Rating
{31 to 51}
[27 to 44]
{82 to 133}
[71 to 115]
{184 to 265}
[159 to 230]
{224 to 316}
[195 to 274]
{153 to 224}
[133 to 195]
{214 to 357}
[186 to 310]
{663 to 959}
[575 to 832]
{582 to 857}
[504 to 743]
142 to 174
{1449 to 1776}
[1257 to 1540]
114 to 138
{1163 to 1408}
[1009 to 1221]
158 to 183
{1612 to 1867}
[1398 to 1619]
243 to 261
{2480 to 2663}
[2150 to 2310]
Size
mm(inch)
*Please use compressed non-asbestos fiber gasket, PTFE gasket or the gasket which has equal elasticity.
JISANSIDIN
10K20K150300PN10
3 to 5
8 to 13
18 to 26
22 to 31
15 to 22
21 to 35
65 to 94
57 to 84
3 to 5
{31 to 51}
[27 to 44]
8 to 13
{82 to 133}
[71 to 115]
18 to 26
{184 to 265}
[159 to 230]
11 to16
{112 to 163}
[97 to 142]
18 to 27
{184 to 276}
[159 to 239]
25 to 44
{255 to 449}
[221 to 389]
43 to 68
{439 to 694}
[381 to 602]
61 to 92
{622 to 939}
[540 to 814]
154 to 182
{1571 to 1857}
[1363 to 1611]
125 to 145
{1276 to 1480}
[1106 to 1283]
—
—
3 to 5
{31 to 51}
[27 to 44]
6 to 10
{61 to 102}
[53 to 88]
15 to 21
{153 to 214}
[133 to 186]
23 to 31
{235 to 316}
[204 to 274]
35 to 43
{357 to 439}
[310 to 381]
22 to 35
{224 to 357}
[195 to 310]
63 to 89
{643 to 908}
[558 to 788]
93 to 120
{949 to 1224}
[823 to 1062]
145 to 177
{1480 to 1806}
[1283 to 1566]
164 to 187
{1673 to 1908}
[1451 to 1655]
245 to 284
{2500 to 2898}
[2168 to 2513]
253 to 275
{2582 to 2806}
[2239 to 2434]
3 to 5
{31 to 51}
[27 to 44]
8 to 13
{82 to 133}
[71 to 115]
21 to 31
{214 to 316}
[186 to 274]
11 to 16
{112 to 163}
[97 to 142]
18 to 26
{184 to 265}
[159 to 230]
25 to 42
{255 to 429}
[221 to 372]
41 to 59
{418 to 602}
[363 to 522]
65 to 93
{663 to 949}
[575 to 823]
126 to 151
{1286 to 1541}
[1115 to 1336]
154 to 180
{1571 to 1837}
[1363 to 1593]
—
—
65 to 94
{663 to 959}
[575 to 832]
94 to 125
{959 to 1276}
[832 to 1106]
136 to 164
{1388 to 1673}
[1204 to 1451]
155 to 199
{1582 to 2031}
[1372 to 1761]
152 to 192
{1551 to 1959}
[1345 to 1699]
248 to 331
{2531 to 3378}
[2195 to 2929]
—
—
—
—
—
—
PN16
3 to 5
{31 to 51}
[27 to 44]
6 to 10
{61 to 102}
[53 to 88]
18 to 26
{184 to 265}
[159 to 230]
23 to 31
{235 to 316}
[204 to 274]
16 to 22
{163 to 224}
[142 to 195]
21 to 35
{214 to 357}
[186 to 310]
65 to 94
{663 to 959}
[575 to 832]
58 to 84
{592 to 857}
[513 to 743]
154 to 175
{1571 to 1786}
[1363 to 1549]
175 to 213
{1786 to 2173}
[1549 to 1885]
—
—
JIS G3451
F12(75M)
—
—
—
—
33 to 47
{337 to 480}
[292 to 416]
47 to 75
{480 to 765}
[416 to 664]
65 to 100
{663 to 1020}
[575 to 885]
69 to 101
{704 to 1031}
[611 to 894]
215 to 270
{2194 to 2755}
[1903 to 2389]
190 to 249
{1939 to 2541}
[1681 to 2204]
274 to 325
{2796 to 3316}
[2425 to 2876]
313 to 395
{3194 to 4031}
[2770 to 3496]
T030309.EPS
3. INSTALLATION
Table 3.3.10 Tightening Torque Values for PVC Piping in N-m{kgf-cm} [in-lbf]
Flange
mm(inch)
*Please select optional code/FRG and use rubber or gasket or others which has equal elasticity.
RatingSize
15(0.5)
25(1)
40(1.5)
50(2)
80(3)
100(4)
150(6)
200(8)
JIS
10K20K
1.1
{11}
[10]
2.9
{30}
[26]
5.9
{60}
[52]
7.2
{73}
[64]
4.9
{50}
[43]
8.1
{83}
[72]
29.8
{304}
[264]
26.3
{268}
[233]
ANSI
150300
—
—
—
—
—
—
—
—
1.1
{11}
[10]
2.3
{23}
[20]
4.7
{48}
[42]
7.2
{73}
[64]
9.9
{101}
[88]
8.2
{84}
[73]
28.5
{291}
[252]
37.7
{385}
[334]
—
—
—
—
—
—
—
—
DIN
PN10PN16
—
—
—
—
—
—
29.8
{304}
[264]
39.4
{4,2}
[349]
1.1
{11}
[10]
2.2
{22}
[19]
5.9
{60}
[52]
7.2
{73}
[64]
4.9
{50}
[43]
8.1
{83}
[72]
29.8
{304}
[264]
26.3
{268}
[233]
JIS G3451
F12(75M)
—
—
—
—
9.9
{101}
[88]
16.3
{166}
[144]
32.6
{33}
[288]
43.1
{440}
[381]
T030310.EPS
3-13
IM 1E6D0-01E
3. INSTALLATION
3.3.5Mominal Diameter 500 mm (20")
to 2600 (104") Flange Type
(1) Mounting Direction
Please mount the Magnetic Flow Tube matching the
flow direction of the fluid to be measured with the
direction of the arrow mark on the flow tube.
IMPORTANT
If it is impossible to macth the direction, please
never remodel to change direction of the terminal box. In case the measuring fluid flows
against the arrow direction, please refer to the
section “Reversing Flow Direction” in the Instruction Manual of AM11 Magnetic Converter.
CAUTION
When lifting the flow tube, please use the lifting
rings (eye bolts or shackle).
To assure safety, please keep lifting angle less
than 90 degrees as shown in Figure 3.3.6.
Please do not shock the installation foundation
bracket when installing the flowmeter as the
bracket may be damaged.
Note 1: See the external dimensions for the support dimensions.
Note 2: The value marked as * is reference dimensions.
T
3-14
IM 1E6D0-01E
3. INSTALLATION
(4) Connecting to Pipe
Please pass bolts from pipe line side, not flow tube
side, and tighten the bolts according to Torque Values
in Table 3.3.9.
CAUTION
• Please be sure to tighten the bolts following
prescribed torque values. Please tighten the
flange bolts diagonally with the same torque
values, step by step up to the prescribed torque
value.
• Please completely cover the brackets and liners
with mortar.
Figure 3.3.7 Mounting Flow Tube
Table 3.3.13 Tightening Torgue Values for Metal Piping
* Please use rubber gasket or the gasket which has equal elasticity.
N-mN-mN-mkgf-cmkgf-cmkgf-cmin-Ibfin-Ibfin-Ibf
121
1234
to 1530
1183
to 1479
1765
to 2193
1836
to 2285
1040
to 1295
2356
to 2938
3040
to 3794
3070
to 3835
4406
to 5498
4723
to 5896
4774
to 5967
4947
to 6171
6273
to 7834
6997
to 8752
7742
to 9670
10261
to 12821
1071
to 1327
1027
to 1283
1531
to 1903
1593
to 1982
903
to 1124
2044
to 2549
2637
to 3292
2664
to 3327
3823
to 4770
4097
to 5115
4142
to 5177
4292
to 5354
5442
to 6796
6071
to 7593
6717
to 8389
8903
to 11124
to 150
116
to 145
173
to 215
180
to 224
102
to 127
231
to 288
298
to 372
301
to 376
432
to 539
463
to 578
468
to 585
485
to 605
615
to 768
686
to 858
759
to 948
1006
to 1257
63
78
104
115
175
204
to 79
to 97
to 130
to 144
to 182
to 254
—
—
—
—
—
—
—
—
—
—
JIS10K
643
to 806
796
to 989
1061
to 1326
1173
to 1469
1785
to 1856
2081
to 2591
—
—
—
—
—
—
—
—
—
—
558
to 699
690
to 858
920
to 1150
1018
to 1274
1549
to 1611
1805
to 2248
—
—
—
—
—
—
—
—
—
—
59 to 73
67 to 84
—
—
—
—
—
—
—
—
—
—
—
—
—
—
ANSI150
602 to 745
683 to 857
—
—
—
—
—
—
—
—
—
—
—
—
—
—
522 to 646
593 to 743
—
—
—
—
—
—
—
—
—
—
—
—
—
—
T030313.EPS
IMPORTANT
If it is impossible to macth the direction, please
never remodel to change direction of the terminal box. In case the measuring fluid flows
against the arrow direction, please refer to the
section “Reversing Flow Direction” in the Instruction Manual of AM11 Magnetic Converter.
(2) Welding of Mating Ferrule to Piping
IMPORTANT
Weld the attached ferrules to the piping. Please
pay attention to the welding condition to avoid
deforming piping or making the stagnant portion
of the fluid; joint preparation, level difference in
butt joint, welding current.
(3) Mounting Gasket
Mount the attached gasket so it fits into the groove on
the ferrule.
(4) Positioning Flow Tube
Position the flow tube between the mating ferrule.
(5) Tightening Clamp
Mount the clamp so they cover the flow tube and
mating ferrule tapered part.
Clamp
Ferrule
Pipe
3.3.6Mounting Procedure for Sanitary Type
The sanitary type is mounted using IDF clamps.
(1) Mounting Direction
Please mount the Magnetic Flow Tube matching the
flow direction of the fluid to be measured with the
direction of the arrow mark on the flow tube.
Figure 3.3.8 Mounting Procedure
3-15
Welding
F030308.EPS
IM 1E6D0-01E
3. INSTALLATION
3.4Wiring Precautions
This section is described wiring only for flow tube
side. Please see "Wiring" in AM11 Magnetic Flow
Converter Instruction Manual for converter side.
3.4.1Protective Grounding
CAUTION
Please be sure to connect protective grounding
of ADMAG with cable of 2mm2 or larger cross
section in order to avoid the electrical shock to
the operators and maintenance engineers and
prevent the influence of external noise. And
further connect the grounding wire to the
mark (100Ω or less).
3.4.2General Precautions
Please give attention to the followings in wiring.
• Please be sure to fully tighten the terminal box
cover before the power is turned on.
• Please be sure to turn off the power before
opening the terminal box cover.
3.4.3Cable Types
(1) Dedicated Signal Cable(AM011)
The flow signal is transmitted via this dedicated cable.
The cable is constructed with double shielding over the
two conductors, and used heat-resistant vinyl as the
outer jacket material.
Finished diameter:10.5 mm (0.413”)
Maximum length:200 m (660 ft)
Maximum temperature: 80°C (176°F)
CAUTION
• In case the ambient temperature exceeds 50°C
(122°F), please use heat-resistant cable with
maximum allowable temperature of 70°C
(158°F) or above for JIS flameproof type.
• Please pay attention to avoid the cable is
bended excessively.
• Please do not connect cables outdoors in case
of rain to prevent damages from dew formation
and to keep insulation inside the terminal box
of the flow tube.
• Please do not splice the cable between a flow
tube and a converter if it is too short. Please
replace the short cable with the cable which is
appropriate length wholly.
• The all cable ends are to be provided with
round crimp-on terminal.
• The signal cables must be routed in separate
steel conduit tubes or flexible tubes.
•Please keep conduit or flexible tube water-tight
using sealing tape.
•Please ground each of a flow tube and a
converter separately.
• Please cover each shield of the signal cable
with PVC tube or PVC tape to avoid contacting
between two shields; shield and case.
•When waterproof glands or union equipped
waterproof glands are used, the glands must be
properly tightened to keep the box watertight.
Figure 3.4.1 Dedicated Signal Cable AM011
IMPORTANT
If the cable is longer than required, cut off any
extra length, rather coiling it up, and terminate
the conductors as shown in Figure 3.4.2. Avoid
using intermediate terminal boards to extend the
cable length, or this will interrupt the shielding.
ACB
SA
On the converter
Figure 3.4.2 Treatment of Dedicated Signal Cable
30 (1.2)
(WHITE)(BLACK)(RED)
side
150
(5.9)
SB
70 (2.8)
8 (0.3) Max.
20 (0.8)
120 (4.7)
130 (5.1)
L (SPECIFIED LENGTH)
AM011*A
ACB
90 (3.5)
(WHITE)(BLACK)(RED)
8 (0.3) Max.
150
(5.9)
On the flow tube
ø10.5 (0.4)
30 (1.2)
90 (3.5)
side
F030402.EPS
3-16
IM 1E6D0-01E
3. INSTALLATION
CAUTION
Since A, B, SA, SB, and C all operate at different electrical potentials, securely insulate them
from each other so they do not touch.
The shields must not be allowed to touch each
other or to touch the case.
Cover each shield with vinyl tube or wrap in vinyl
tape.
NOTE
Conductors A and B carry the signal from the
electrodes, and C is at the potentials of the liquid
it self (signal common) . Shields SA and SB are
kept at the same potentials as the individual
electrodes (these are actively driven shields).
This is done to reduce the effect of the distributed capacitance of the cable at long cable
length. Note that, since the signals from the
individual electrodes are impedance converted
inside the converter, errors will result if they
come in contact with any other component.
Great care must be taken in the cable end
treatment.
3.4.4Connnection to AM11Magnetic
Flow Converter
Connect the flow tube and converter in the following
method.
EX1
EX2
Dedicated signal cable
AM011-2
A
B
C
Flow tube
Figure 3.4.4 Connection
CSAAB
EX2
EX1
Note: Terminate those shielding wire terminal
SB
Converter
Terminal
SA
A
B
SB
C
EX1
EX2
using insulating tape.
AM11
converter
Flow tube
(See Note below)
A
B
(See Note below)
C
EX1
EX2
F030404.EPS
(2) Excitation Cable
Please use Polyvinyl chloride insulated and sheathed
control cables (JIS C3401) or Polyvinyl chloride
insulated and sheathed portable power cables (JIS
C3312) or equivalents.
Outer Diameter
• 6.5 to 12mm in diameter (7.5 to 12 mm for waterproof gland / ECG, /ECU, /ECW)
Nominal Cross Section
•Single wire; 0.5 to 2.5mm2 , Stranded wire; 0.5 to
2
2.5mm
EX2
EX1
85
(3.3)
On the converter sideOn the flow tube side
(3.3)
85
EX1
Unit: mm (inch)
EX2
Figure 3.4.3 End Treatment of Excitation Cable
F030403.EPS
3-17
IM 1E6D0-01E
3. INSTALLATION
3.4.5Wiring Ports
Please select the most suitable standard of wiring
procedure for the wiring ports by customer’s own.
A : Using the Waterproof Gland
IMPORTANT
To prevent water or condensate from entering
the converter housing, waterproof glands are
recommended. Do not over-tighten the glands or
damage to the cables may result. Tightness of
the gland can be checked by confirming that the
cable is held firmly in place.
Waterproof gland
Optional code : /ECG
Gasket
Tightening
gland
Waterproof gland with union joint
Optional code : /ECU
G1/2 female
Gasket
Tightening gland
G1/2 female
B : Insulation Check
After wiring is completed, check insulation of the
following terminals together with the wiring under the
condition of the converter side wiring terminals
disconnected.
Between terminal EX1 and terminal A, B, C
Between terminal C and terminal A, B
Between terminal A and terminal B
Between terminal EX1 and ground
Between terminal EX2 and ground
All insulation measurements must be performed with a
500V megger. Insulation resistances must be 100MΩ
or more each.
IMPORTANT
Be sure to disconnect the cables at the terminal
of the converter when checking insulation.
C : Conduit Wiring
In case of conduit wiring, please use the waterproof
gland to prevent water flowing through the conduit
pipe into the wiring connection.
Please slope the conduit pipe down, and install a drain
valve at the low end of the vertical pipe.
Waterproof gland
Optional code : /ECW
Gasket
Tightening
gland G3/4
Figure 3.4.6 Waterproof Gland
Please open the drain valve regularly.
G1/2 female
F030406.EPS
Drain valve
F030407.EPS
Figure 3.4.7 Conduit Wiring
3-18
IM 1E6D0-01E
4.MAINTENANCE
4. MAINTENANCE
WARNING
This instrument must be repaired or maintenance-serviced by expert engineer or skilled
personnel. The procedures described in this
chapter are not permitted for operators.
Regular maintenance and inspection should be carried
out to fully utilize all functions and to obtain maximum
performance from the magnetic flowmeter.
4.1Regular Inspection
Items
(1)Inspection of moisture-proofing inside the
terminal box: Once/year
(2)Refastening of piping joint screws: About
twice/year
(3)Inspection of electrodes and lining (in case
of adhesive and/or abrasive fluid, etc.)
Determine the period of regular inspection as
necessary.
(2) Insulation Resistance Check
Check the insulation resistances in accordance with the
tables below. If one of them falls below the value in
the tables, replacement or repair of the flow tube is
needed. In case of submersible type flowmeter, release
wiring connection of converter side and measure
resistance at cable terminations.
Coil Circuit
Checking is possible even if the pipe is filled with
fluid.
Test T erminals
Between Terminals EX1 and C
Test V oltageSpecification
500 V DC1MΩ or more
T0401.EPS
Signal Circuit
Be sure to empty and dry the pipe inside, and release
wiring connection of converter side before checking.
Test T erminals
Between Terminals A and C
Between Terminals B and C
Test V oltage
500 V DC100MΩ or more for each
Specification
T0402.EPS
4.2Trouble Shooting
Since the ADMAG magnetic flowmeter has “self
diagnostic functions”, if a failure occurs, it is displayed
on the AM11 magnetic flow converter. Please refer to
Instruction Manual of AM11 magnetic flow converter.
IMPORTANT
Be sure to disconnect the cables at the terminals
of the flow tube when checking.
(1) Excitation Coil Check
Check that the resistance between terminals EX1 and
EX2 in the terminal box is 150 Ω or less with a
multimeter. If it is not, coils may be broken down, and
replacement or repair of the flow tube is needed.
4-1
IM 1E6D0-01E
5.OUTLINE
■STANDARD SPECIFICATIONS
Protection:
General Purpose /Explosion proof /Sanitary Type:
IP67, NEMA4X, JIS C0920 water tight
type.
Submersible Type: JIS C0920 submersible type, IP68.
(Test Condition: 50m below the surface of
the water, equivalent to 0.5MPa
pressure, for one month. If always under
water or corrosion fluid, contact
Yokogawa office.)
(equivalent to Hastelloy C-276), Titanium,
Tantalum, Platinum-Iridium, Platinumalumina cermet only for ceramic lining.
*For size over 500mm and sanitary type,
SUS316L only
276), Titanium,SUS304,(Size 500 to
1000mm (20 to 40in.), SS400 lined with
SUS316 (Size 1100 to 2600mm (40 to
104in.) ), PFA lining+Earth electrodes
(Tantalum, Platinum-Iridium).
Note: Hastelloy is a registered trademark of
Ha ynes International Inc.
Gasket:• VALQUA#7020;Fluoro resin with filler
(between flow tube body and earth ring;
for ceramic lining)
• VALQUA#4010;Fluoro rubber, viton
(between flow tube body and earth ring;
for optional code /FRG)
• Non-asbestos joint sheet sheathed with
fluoro resin (between earth ring and
process flange; for optional code /BCF or
/BSC)
• Chloroprene rubber (between earth ring
and process flange; for optional code
/BCC or /BSC)
* Other gaskets between flow tube body and earth ring;
• VALQUA#7026;corrosion resistance gasket
(Fluoro resin with carbon)
• VALQUA#4010(Mixing#RCD970);Alkali
resistance gasket for PVC piping (Fluoro
rubber)
• VALQUA#4010(Mixing#RCD470);Acid
resistance gasket for PVC piping (Fluoro
rubber)
Contact Yokogawa office. (Refer to TI
1E6A0-06E)
Union Joint:Stainless steel (SUS316L), for ceramic
lining size 10mm (0.4in.) or less*
*Contact Yokogwa office if PVC union joints are required.
Ferrule(for sanitary type):SCS13
5-1
IM 1E6D0-01E
5. OUTLINE
Electrode Construction:
• General Purpose/Ex-proof/Submersible type
PFA, Polyurethane Linings: External Insertion type.
Ceramic Lining: Integral type.
• Sanitary type
Internal Insertion type.
* Only PFA lining is available for sanitary type
Electrical Connection:
JIS G1/2(PF1/2)female, ANSI 1/2NPT female,
DIN Pg13.5 female, ISO M20X1.5 female
Terminal Connection:
M4 size screw terminal.
Grounding:
100Ω or less
* In case of JIS Flameproof type, JIS Class C(10Ω or
less) or JIS Class A(10Ω or less)
■ HAZARDOUS AREA CLASSIFICATION
AM100 / 200DN
FM:
• Explosion proof for Class I, Division 1, Groups A, B,
C & D. Dust-ignitionproof for Class II/III, Division 1,
Groups E, F &G. Intrinsically safe (electrode) for
Class I, Division 1, Groups A, B, C & D.
Electrode Circuit Vmax: 250V ac/dc
Temp. Class. T6
Enclosure: NEMA 4X
• Nonincendive for Class I, Division 2, Groups A, B, C
& D; Suitable for Class II, Division 2, Groups F &G;
Class III, Division 1 and 2. Intrinsically safe (electrode) for Class I, Division 1, Groups A, B, C & D.
• Nonincendive for Class I, Division 2, Groups A, B, C
& D; Suitable for Class II, Division 2, Groups F &G;
Class III, Division 2. Intrinsically safe (electrode) for
Class I, Division 1, Groups A, B, C & D.
Electrode Circuit Vmax: 250V ac/dc
Excitation Circuit: 130Vp 72/75 Hz, 11 VAmax
Temp. ClassT5T4T3
Process Temp. 85120150˚C
Enclosure: Type 4
JIS:
• Exde[ia]II C T4X
Converter, Terminal box ; Explosion proof
Flow Tube ; Increased Safety
Signal Circuit ; Intrinsically Safety(ia)
Ambient Temperature : -10˚C to 60 ˚C*
Fluid Temperature : 120˚Cmax.
Electrode Circuit : 250V AC/DC
Grounding : JIS Class C(10Ω or less) or JIS Class
A(10Ω or less)
* In case the ambient temperature exceeds 50˚C,
use heat-resistant cables with maximum allowable
temperature of 70˚C or above.
■STANDARD PERFORMANCE
Accuracy:
PFA and Ceramic Lining
Vs: Span setting value (m/s)
Size in
mm (inch)
2.5 to 15
(0.1 to 0.5)
25 to 400
(1 to 16)
Polyurethane Lining
Size in
mm (inch)
25 to 400
(1 to 16)
500 to 1000
(20 to 40)
1100 to 2000
(44 to 80)
2200 to 2600
(88 to 104)
Repeatability:
0.1% of flowrate (1mm/s Minimum )
Maximum Power Consumption:
28W (for combination with AM11 converter).
Insulation Resistance:
• 100MΩ between excitation terminal and each signal
terminal at 500V DC.
• 100MΩ among signal terminals at 500V DC.
Withstand Voltage
(Except for ex-proof type):
• 1000V AC or 1400 V DC between excitation
terminal and ground terminal for 1minute.
(Ex-proof type):
• 1500V AC between excitation terminal and ground
terminal for 1 minute.
• 500V AC between each signal terminal and ground
terminal for 1 minute.
• 1760V AC between each signal terminal and each
excitation terminal for 1 minute.
Span in m/s
(ft/s)
0.1 to 0.3(0.3 to 1)
0.3 to 1(1 to 3)
1 to 3 (3 to 33)
(both included)
0.1 to 0.3(0.3 to 1)
0.3 to 1
(1 to 3)
1 to 10(3 to 33)
(both included)
Span in m/s
(ft/s)
0.1 to 0.3(0.3 to 1)
0.3 to 1(1 to 3)
1 to 10 (3 to 33)
0.1 to 0.3(0.3 to 1)
0.3 to 1(1 to 3)
1 to 10 (3 to 33)
0.3 to 1(1 to 3)
1 to 10(3 to 33)
0.3 to 1(1 to 3)
1 to 10(3 to 33)
Accuracy
0.15/Vs % of span
0.5% of span
0.25% of span (at indications below
50%)
0.5% of flowrate indication
(at indications 50% of span or more)
0.075/Vs % of span
0.25% of span
(at indications below 50% of span)
0.5% of flowrate indication
(at indications 50% of span or more)
0.1% of span
(at indications below 20% of span)
0.5% of flowrate indication
(at indications 20% of span or more)
Vs: Span setting value (m/s)
Accuracy
0.15/Vs % of span
0.5% of span
0.25% of span (at indications
below 50% of span)
0.5% of flowrate indication
(at indications 50% of span or
more)
0.225/Vs % of span
0.75 % of span
0.5% of span
0.75 % of span
0.5% of span
1/Vs % of span
1% of span
T02.EPS
T03.EPS
5-2
IM 1E6D0-01E
5. OUTLINE
CAUTION
When performing the Voltage Breakdown Test,
Insulation Resistance Test or any unpowered
electrical test, wait 10 seconds after the power
supply is tumed off before removing the housing
cover. Be sure to remove the Short Bar at terminal “G”. After testing, return the Short Bar to its
correct position. Screw tightening torque should
be 12kg-cm(0.88ft-lb) or more, becase the Gterminal is thought as a protective grounding and
should conform to the Safety Requirements.
Protective grounding
(outside of the terminal box)
T06.EPS
■GASKET
Please use compressed non-asbestos fiber gasket,
PTFE gasket or the gasket which has equal elasticity.
In case of optional code/FRG, please use rubber
gasket or others which has equal elasticity.
■SIZING DATA
10.0
5.0
Flow velocity
(m/s)
1.0
0.5
0.3
0.1
0.001
33
Flow velocity
(ft/s)
10
5.0
2.0
1.0
0.01
SI Units
Size: mm
2.5
0.1
English Units
0.1
0.2
5
1
Flowrate (m /h)
Size: inch
0.4
1,350
2,000
1,600
25
15
10
10
3
1
0.5
1.5
65
40
2
80
50
100
3
4
2.5
100
5
200
250
150
125
1,000
6
8
12
10
500
700
400
10,000
20
800
900
600
1,000
1,200
1,100
1,500
100,000
60
40
24
28
54
32
48
44
36
64
350
300
16
14
2,400
2,200
2,600
1,800
72
04
88
80
96
1
0.33
0.01
0.1
1.0
10
Flowrate (GPM)
100
1,000
10,000
100,000
1,000,000
F04.EPS
* Measurable flow velocity is from 0 m/s.
5-4
IM 1E6D0-01E
■MODEL AND SUFFIX CODE
•General purpose/Explosion proof/Submersible Type AM100D, 200D
Note 1: Process connection for size 2.5 to 10mm(0.1 to 0.4in.), wafer type is equal to one for 15mm(0.5in.).
Note 2: Select PN16 when requirement of PN10 for size 100mm(4in.)or less because of same mating dimensions.
Note 3: Flange type is available for size 15mm or larger, PFA or Polyurethane lining.
Note 4: Select ANSI 1/2NPT female electrical connection(code A) in case of requirement of FM/CSA explosion proof type.
Note 5: Select JIS G1/2 female electrical connection (code J) and optional code /JF3 with /G12 in case of requirement
Note 6: Platinum-iridium and Tantalum earth electrodes are not available with 65, 125mm and ceramics 150, 200mm.
. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . .
W
. . . . . . . . . . . . . . . . . . .
N
. . . . . . . . . . . . . . . . . . .
C
. . . . . . . . . . . . . . . . .
-A
. . . . . . . . . . . . . . . . .
-U
. . . . . . . . . . . . . . . . .
-C
. . . . . . . . . . . . . .
K1
. . . . . . . . . . . . . .
K2
. . . . . . . . . . . . . .
B1
. . . . . . . . . . . . . .
B2
. . . . . . . . . . . . . .
E2
. . . . . . . . . . . . . .
H1
. . . . . . . . . . . . . .
J1
. . . . . . . . . . . . . .
J2
. . . . . . . . . . . . . .
A1
. . . . . . . . . . . . . .
A2
. . . . . . . . . . . . . .
D1
. . . . . . . . . . . . . .
D2
. . . . . . . . . . . . . .
G1
. . . . . . . . . . . . . .
U1
. . . . . . . . . . . . . .
U2
. . . . . . . . . . . . . .
U3
. . . . . . . . . . .
-L
. . . . . . . . . . .
-P
. . . . . . . . . . .
-E
. . . . . . . . . . .
-H
. . . . . . . . . . .
-T
. . . . . . . . . . .
-V
. . . . . . . . .
S
. . . . . . . . .
P
. . . . . . . . .
N
. . . . . . . . .
H
. . . . . . . . .
T
. . . . . . . . .
V
Provide 15mm process flanges for these flow tubes.
For CENELEC explosion proof type, select ANSI 1/2NPT (code A), DIN Pg13.5(code D) or ISO M20⫻1.5(code M)
electrical connection.
of JIS Flameproof type.
Size 2.5 mm (0.1in.)
Size 5 mm (0.2in.)
Size 10 mm (0.4in.)
Size 15 mm (0.5in.)
Size 25 mm (1in.)
Size 40 mm (1.5in.)
Size 50 mm (2in.)
Size 65 mm (2.5in.)
Size 80 mm (3in.)
Size 100 mm (4in.)
Size 125 mm (5in.)
Size 150 mm (6in.)
Size 200 mm (8in.)
General purpose
Submersible
CENELEC, FM, CSA Ex.proof
JIS Flameproof
Fluorocarbon PFA
Polyurethane
Ceramic
JIS 10K wafer
JIS 20K wafer
ANSI 150 wafer
ANSI 300 wafer
DIN PN10/16 wafer
JIS F12 (JIS75M) wafer
JIS 10K flange
JIS 20K flange
ANSI 150 flange
ANSI 300 flange
DIN PN10 flange
DIN PN16 flange
JIS F12 (JIS75M) flange
Union joint (weld joint)
Union joint (R1/4 or R3/8)
Union joint (1/4NPT or 3/8NPT)
SUS316L
Platinum-iridium
Platinum-alumina cermet
Hastelloy C 276 equivalent
Tantalum
Titanium
Size 250 mm (10in.)
Size 300 mm (12in.)
Size 350 mm (14in.)
Size 400 mm (16in.)
General purpose
Submersible
Fluorocarbon PFA
Polyurethane
JIS 10K flange
JIS 20K flange
ANSI 150 flange
ANSI 300 flange
DIN PN10 flange
DIN PN16 flange
JIS G3451 F12 flange
SUS316L
Platinum-iridium
Hastelloy C 276 equivalent
Tantalum
Titanium
SUS316
Hastelloy C 276 equivalent
Titanium
JIS G1/2(PF1/2)female
ANSI 1/2NPT female
DIN Pg13.5 female
ISO M20X1.5 female
*1 Size 500 to 1000mm(20 to 40in.)
*2 Size 500/600mm(20/24in.)
*3 Size 500 to 1500mm(20 to 60in.)
*4 Size 1600mm(64in.) or larger
*5 Size 1100mm(44in.) or larger
U
J1
A1
G1
W1
*1
*2
*3
*4
L
S
J
A
Size: 500 mm (20in.)
Size: 600 mm (24in.)
Size: 700 mm (28in.)
Size: 800 mm (32in.)
Size: 900 mm (36in.)
Size: 1000 mm (40in.)
Size: 1100 mm (44in.)
Size: 1200 mm (48in.)
Size: 1350 mm (54in.)
Size: 1500 mm (60in.)
Size: 1600 mm (64in.)
Size: 1800 mm (72in.)
Size: 2000 mm (80in.)
Size: 2200 mm (88in.)
Size: 2400 mm (96in.)
Size: 2600 mm (104in.)
General purpose
Submersible
Polyurethane
JIS 10K Flange
ANSI Class 150 Flange
JIS G3451 F12 Flange
JIS 7.5K Flange
SUS316L
SUS304(for AM400)
SS400 lined with
SUS316
JIS G1/2(PF1/2) female
*1
ANSI 1/2NPT female
-000
Always -000
Style A
*A
*5
Style C
*C
Refer to optional
specification table
A: Available
N: Not Available
UUA
Only for size 250 or 300 mm(10 or 12in.)
Only for size 250 or 300 mm(10 or 12in.)
Only for size 250 or 300 mm(10 or 12in.)
(for AM500)
T09.EPS
5. OUTLINE
T08.EPS
5-6
IM 1E6D0-01E
•Sanitary Type
5. OUTLINE
Model
AM202DH
AM204DH
AM205DH
AM206DH
AM208DH
AM210DH
Lining
Process connection
Electrode material
Earthring
Wiring port
Style code
Optional specification
Suffix code
. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . .
-A
. . . . . . . . . . . . . .
C1
. . . . . . . . . . .
-L
. . . . . . . . .
N
. . . . . .
J
. . . . . .
A
. . . . . .
D
. . . . . .
M
*A
/
Description
Size 25 mm (1in.)
Size 40 mm (1.5in.)
Size 50 mm (2in.)
Size 65 mm (2.5in.)
Size 80 mm (3in.)
Size 100 mm (4in.)
Fluorocarbon PFA
IDF clamp
SUS316L
None
JIS G1/2(PF1/2)female
ANSI 1/2NPT female
DIN Pg13.5 female
ISO M20X1.5 female
Style A
Refer to optional specification code
T10.EPS
Optional Specifications
Item
Stainless steel tag plate
Waterproof Gland
Waterproof Gland with
Union Joint
Gasket for PVC Piping
DHC use (for size 100 to
400mm)
Epoxy Coating
High Anti-corrosion
Coating
Material Certificate
(Note 2)
Bolt & Nut Assembly
(Note 3)
FM Approval
CSA Certification
(Note 4)
CENELEC Certification
(KEMA Approval)(Note 4)
JIS Flameproof
(Note 4)(Note 5)
Flameproof Packing
Adapter for JIS
Flameproof(Note 5)
Silicon Rubber Gasket for
sanitary type
GOST Certification
C-Tick mark
Mirror Finished Ceramics
Oil-prohibited Use
Oil-prohibited Use with
Dehydrating T reatment
Hydrostatics Test
Certificate
Calibration Cretificate
Note1No need to select because waterproof glands(union joint)are attached for submersible type as standard.
Note2In case of requirement of EN10204 3.1B for Material Certificate, contact Yokogawa office.
Note3It is available only for JIS 10K wafer, JIS 20K wafer or ANSI 150 wafer type, not available with 65, 125mm.
Note4It is available only for size 2.5 to 200mm(0.1 to 8in.).
Note5Select optional code /JF3 with /G12 in case of requirement of JIS Flameproof type.
Stainless steel (SUS304) tag plate is fixed with
screws.
Waterproof glands are attached to all wiring ports. For
JIS G1/2 only.
Waterproof glands (union joint) are attached to all
wiring ports. For JIS G1/2 only.
Gasket are attached between earth ring and flow
tube. For 200mm or smaller size. (except 65,125mm)
and Cooling use. Terminal box; urethane resin potting.
Wired 30m signal cable at factory.
Coating is changed to epoxy coating.
Coating is changed to three-layer coating (Urethane
coating on two-layer Epoxy coating)
Reproduced material certificate for pipe, electrode,
earth ring, mini-flange and flange (depends on spec.).
Carbon steel bolts/nuts and chloroprene gaskets
assembly for wafer type
Carbon steel bolts/nuts and non-asbestos PTFE-
wrapped gaskets assembly for wafer type
Stainless steel bolts/nuts and chloroprene gaskets
assembly for wafer type
Stainless steel bolts/nuts and non-asbestos PTFE-
wrapped gaskets assembly for wafer type
FM Explosion proof/Nonincendive,
CSA Nonincendive
For 200mm or smaller size.
CENELEC Explosion proof
For 200mm or smaller size.
JIS Flameproof type
For 200mm or smaller size.
Two Flameproof Packing Adapters are attached.
In case of fluid temp. over 120 degC(248 degF);
160 degC (320 degF) max. (Not available with 65mm)
Calibration Certificate for GOST (only for products
produced at YMF)
AS/NZS2064 compliance
Only for Australia or New Zealand
Mirror finishing on the inside of ceramic tube(Rmax.<=
1micro-meter)
Degreased cleansing treatment
Degreased cleansing treatment; Packaging with
desiccant
Test pressure depends on process connection (Test
duration 10minutes) Test result is full in NOTE of QIC.
Level2:Declaration and Calibration Equipment List
Level3:Declaration and Primary Standard List
Level4:Declaration and YOKOGAWA Measuring
* The face-to-face length is longer by approx.22mm (0.89")
** The height is shorter by 3mm (0.11") for ex-proof type.
øD
W
H
H1
H2
for earth ring P,T. The face-to-face length is longer by
approx.2mm (0.08")for PFA lining with optional code/FRG.
G
K
115
AMD
C-C
B
N
E
General (G), Ex-proof(C,N)
102105110115
2.5(0.1)
5(0.2)10(0.4)
A
L
85(3.3)*
44(1.7)
72(2.8)
214(8.4)**
66(2.6)
128(5.0)**
2.4(5.3)
15(0.5)
A.C
SD1.eps
Ground Terminal
(M4)
H*
102
G
105
AMD
C-C
U
110
N
Process Connection
Size code
U1
U2
U3
Type
Size code
Size
Lining code
Process connection
Face-to-face
Width
Height
Height
Height
Weight kg(lb)
Lining code C : Ceramics pipe
*The height is shorter by 3mm(0.12") for ex-proof type.
102
L
W
H
H1
H2
105110
Weld joint
R1/4
1/4NPT
General (G), Ex-proof(C,N)
102
2.5(0.1)
U1
140(5.5)
105
5(0.2)
C
U2, U3
130(5.1)U1140(5.5)
72(2.8)
202(8.0)*
60(2.4)
116(4.6)*
2.4(5.3)
R3/8
3/8NPT
U2, U3
130(5.1)
110
10(0.4)
U1
140(5.5)
SD2c.eps
ø86(3.4)
48
(1.9)
L
U2, U3
130(5.1)
øA
4
(0.16)
Size
2.5
(0.1)
5
(0.2)
10
(0.4)
36 (1.4)
H2*
H1
11.5(0.45)
A
22(0.87)
22(0.87)
22(0.87)
22(0.87)
25(0.98)
25(0.98)
Shroud
(for ex-proof)
30(1.2)
B
8(0.31)
8(0.31)
8(0.31)
8(0.31)
10(0.39)
10(0.39)
C
18.5(0.73)
18.5(0.73)
18.5(0.73)
18.5(0.73)
22.5(0.89)
22.5(0.89)
58
(2.3)
NPT1/4
NPT1/4
NPT3/8
øB
D
R1/4
R1/4
R3/8
109
(4.3)
W
4-ø5.8(0.2)
D
øD
10
(0.39)
C
øC
D
18.5(0.73)
18.5(0.73)
22.5(0.89)
SD2.eps
øC
øB
øA
(0.16)
2.5(0.1)
4
Size
5(0.2)
10(0.4)
A
22(0.87)
22(0.87)
25(0.98)
35
(1.38)
B
8(0.31)
8(0.31)
10(0.39)
14.3(0.56)
14.3(0.56)
17.8(0.70)
5-8
IM 1E6D0-01E
5. OUTLINE
Ground Terminal
(M4)
H
202
204
205
AMD
206
208
210
Type
Size code
Size
Lining code
Face-to-face
Outer dia
Height
Height
Weight kg(lb)
Lining code A : PFA lining
*The face-to-face length is longer by approx.22mm(0.89")for earth ring P,T.
The face-to-face length is longer by approx.2mm(0.08")for PFA or Polyuretane lining with optional code/FRG.
**Submersible flowtubes are provided with 30m(98ft) of signal and excitation cables, potted terminal box, and waterproof gland fittings.
Please add 9.5kg(20.9lb) for submersible type.
o86(3.4)
48
(1.9)
L*
K
G
AB
W
-
UE
C
CH
N
204
202
25(1)
40(1.5)
A.U.C
93(3.7)*
68(2.7)
197(7.8)
111(4.4)
2.3(5.5)
106(4.2)*
86(3.4)
217(8.5)
131(5.2)
3.2(7.1)
L*
D
H
H1
U : Polyurethane rubber lining
C : Ceramics pipe
Water-proof gland
(for submersible type**)
36 (1.4)
H1
General(G), Submersible(W)
205
50(2)
120(4.7)*
120(4.7)
99(3.9)
117(4.6)
230(9.1)
248(9.8)
144(5.7)
162(6.4)
4.2(9.3)
5.2(11.5)
206
65(2.5)
Shroud
(for ex-proof)
208
210
80(3)
100(4)
A.UCA.CAAC
160(6.3)*
180(7.1)*
129(5.1)
155(6.1)
270(10.6)
184(7.2)
7.3(16.1)
296(11.7)
210(8.3)
10.2(22.5)
260(10.2)
208
80(3)
160(6.3)*
129(5.1)
174(6.9)
6.9(15.2)
109(4.3)
210
100(4)
180(7.1)*
155(6.1)
296(11.7)
210(8.3)
10(22.1)
oD
202
25(1)
93(3.7)*
68(2.7)
195(7.7)
109(4.3)
2.3(5.5)
EX-proof(C,N)
204
40(1.5)
106(4.2)*
86(3.4)
214(8.4)
128(5.0)
3.2(7.1)
205
50(2)
120(4.7)*
99(3.9)
228(9.0)
142(5.6)
4.2(9.3)
208
80(3)
160(6.3)*
129(5.1)
268(10.6)
182(7.2)
7.3(16.1)
210
100(4)
180(7.1)*
155(6.1)
296(11.7)
208(8.2)
10.2(22.5)
208
80(3)
160(6.3)*
129(5.1)
258(10.2)
172(6.8)
6.9(15.2)
210
100(4)
180(7.1)*
155(6.1)
294(11.6)
208(8.2)
10(22.1)
Ground Terminal
(M4)
36 (1.4)
H1
212
G
215
AM
D
W
220
N
C
Type
Size code
Size
Lining code
Face-to-face
Outer dia.
Height
Height
Width
Weight kg(lb)**
Lining code A : PFA lining
U : Polyurethane rubber lining
36 (1.4)
H1
-
AB
UE
K
H
L
D
H
H1
W
ø86(3.4)
15.7(34.6)
48
(1.9)
o86(3.4)
48
(1.9)
L*
General(G), Submersible(W)
215
212
150(6)
125(5)
A
230(9.1)*
230(9.1)
200(7.9)
175(6.9)
339(13.3)
313(12.3)
253(9.9)
227(8.9)
256(10.1)
230(9.1)
17.9(39.4)
Water-proof gland
(for submersible type**)
H
220
200(8)
150(6)
A, U
300(11.8)*
230(9.1)*
250(9.8)
200(7.9)
389(15.3)
336(13.2)
303(11.9)
250(10.1)
307(12.1)
256(10.1)
28.4(62.6)
20.9(46.1)
Shroud(for ex-proof)
Ground Terminal
(M4)
H
Shroud(for ex-proof)
oD
Ex-proof(C, N)
215
220
200(8)
300(11.8)*
250(9.8)
386(15.2)
300(11.8)
307(12.1)
33.4(73.6)
109(4.3)
109(4.3)
W
* The face-to-face length is longer by approx.30mm(1.2")for earth ring P,T.
The face-to-face length is longer by approx.2mm(0.08")for optional code/FRG.
**Submersible flowtubes are provided with 30m (98ft) of signal and excitation calbles,
potted terminal box, and waterproof gland fittings.
Please add 9.5kg(20.9lb) for submersible type.
SD4.eps
K
215
G
B
-
DCAM
220
N
E
C
Type
Size code
Size
øD
Lining code
Face-to-face
Outer dia
Height
Height
Weight kg(lb)
Type
Size code
Size
Lining code
Process Connection
Outer dia.
h
-ø
4
D
Thickness
ø
Pitch circle dia.
øC
Hole dia.
Weight kg(lb)
Process connection : J1 ; JIS10K J2 ; JIS20K A1 ; ANSI 150
A2 ; ANSI 300 D2 ; DIN PN16
Lining code A : PFA lining
)
.23
.8(0
5
* The face-to-face length is longer by approx.22mm (0.89") for earth ring P,T.
-ø
4
The thickness(t) is longer by approx.11mm (0.43")for earth ring P,T.
The face-to-face length is longer by approx.2mm (0.08") for optional code/FRG.
The thickness(t) is longer by approx.1mm (0.04") for optional code/FRG.
** The height is shorter by 3mm (0.12") for ex-proof type.
109(4.3)
h
-ø
N
øD
ø
C
⬚
* The face-to-face length is longer
by approx.22mm (0.89") for earth ring P,T.
The thickness(t) is longer
by approx.11mm (0.43") for earth ring P,T.
The face-to-face length is longer
D2
by approx.2mm (0.08") for optional code/FRG.
The thickness(t) is longer by approx.1mm (0.04")
263(10.4)**
for optional code/FRG.
177(7.0)**
** The height is shorter by 2.5mm (0.1") for ex-proof type.
*** Submersible meters are provided with 30m (98ft) of
165(6.5)
signal and excitation cable, potted terminal box, and
24(0.9)
waterproof fitlings. Please add 9.5kg (20.9lbs) in
~30(1.2)*
weight for submersible type.
125(4.92)
45
18(0.71)
4
10.2(22.5)
SD12.eps
J1
258(10.2)**
172(6.8)**
155(6.1)
23(0.9)
~29(1.1)*
120(4.72)
45
19(0.75)
4
9.0(19.8)
J2
258(10.2)**
172(6.8)**
94.5(3.7)**
155(6.1)
25(1)
~31(1.2)*
120(4.72)
22.5
19(0.75)
8
9.1(20)
205
50(2)
A, U
A1
257(10.1)**
171(6.7)**
152(6.0)
26(1)
~33(1.3)*
121(4.76)
45
19(0.75)
4
9.4(20.7)
A2
263(10.4)**
177(7.0)**
165(6.5)
30(1.2)
~37(1.5)*
127(5.00)
22.5
19(0.75)
8
10.8(23.8)
109(4.3)
D2
A2
95(3.7)
20(0.8)~
26(1.0)*
65(2.6)
14(0.6)
4.2(9.3)
SD11.eps
(H2)**
36 (1.4)
H**
t*
H1**
G
J
W
A
A
208
-
AM D
C
U
D
210
N
G
Type
Size code
Size
Lining code
Process Connection
Height
Height
Height
Outer dia.
Thickness
Pitch circle dia.
Bolt hole pitch
Hole dia.
Number of holes
Face-to-face
Weight kg(lb)***
Process connection : J1 ; JIS10K J2 ; JIS20K A1 ; ANSI 150 A2 ; ANSI 300 D2 ; DIN PN16 G1 ; JIS F12
Lining code A : PFA Lining
J1
H
288(11.3)**
H1
202(8.0)**
H2
øD
185(7.2)
25(1)
t
~31(1.2)*
øC
150(5.1)
⬚
22.5
øh
19(0.7)
8
N
L
13(28.6)
U : Polyurethane lining
J2
296(11.7)**
210(8.3)**
200(7.9)
29(1.1)
~35(1.4)*
160(6.3)
22.5
23(0.8)
8
15(33.0)
*L
General(G), Submersible(W), Ex-proof(C,N)
208
80(3)
A, U
A1
A2
D2
8
296(11.7)**
210(8.3)**
200(7.9)
26(1)
~32(1.3)*
160(6.3)
22.5
18(0.7)
8
15(33.1)
G1
301(11.9)**
215(8.5)**
211(8.3)
25(1)
~32(1.3)*
168(6.6)
45
19(0.7)
4
15.5(34.2)
291(11.4)*
205(8.1)*
109.5(4.3)**
191(7.5)
31(1.2)
~38(1.5)*
152(6.0)
45
19(0.7)
4
200(7.9)*
15.3(33.7)
300(11.8)**
214(8.4)*
210(8.3)
36(1.4)
~43(1.7)*
168(6.6)
22.5
22(0.8)
18.3(40.3)
J1
324(12.8)**
238(9.4)**
210(8.3)
25(1)
~31(1.2)*
175(6.9)
22.5
19(0.7)
8
21.0(46.3)
J2
331(13.0)**
245(9.6)**
225(8.9)
31(1.2)
~37(1.5)*
185(7.3)
22.5
23(0.9)
8
24.2(53.4)
210
100(4)
A, U
A1
333(13.1)*
247(9.7)*
132.5(5.2)**
229(9.1)
31(1.2)
~38(1.5)*
191(7.5)
22.5
19(0.7)
8
250(9.8)*
25.0(55.1)
346(13.6)**
260(10.2)*
254(10)
~46(1.8)*
200(7.9)
32.4(71.4)
A2
39(1.5)
22.5
22(0.9)
8
D2
329(13.0)**
243(9.6)**
220(8.7)
26(1)
~32(1.3)*
180(7.1)
22.5
18(0.7)
8
22.7(50.1)
-øh
N
D
ø
ø
C
⬚
* The face-to-face length is longer
by approx.22mm (0.89") for earth ring P,T.
The thickness(t) is longer by approx.11mm (0.43")
for earth ring P,T.
The face-to-face length is longer
G1
by approx.2mm (0.08") for optional code/FRG.
The thickness(t) is longer by approx.1mm (0.04")
338(13.3)**
for optional code/FRG.
252(9.9)**
** The height is shorter by 2.5mm (0.1") for ex-proof type.
*** Submersible meters are provided with 30m (98ft) of
238(9.4)
signal and excitation cable, potted terminal box, and
25(1)
waterproof fitlings. Please add 9.5kg (20.9lbs) in weight
* The face-to-face length is longer by approx.28mm (1.1") for earth ring P,T.
The thickness(t) is longer by approx.14mm (0.55") for earth ring P,T.
The face-to-face length is longer by approx.2mm (0.08") for optional code/FRG.
The thickness(t) is longer by approx.1mm (0.04") for optional code/FRG.
** The height is shorter by 2.5mm (0.1") for ex-proof type.
*** Submersible meters are provided with 30m (98ft) of signal and excitation cable,
potted terminal box, and waterproof gland fittings.
*Packing
(Choose from the table below
depend on cable outside diameter)
Dimension
T1T2CD
G 1/2G 1/2
(1.38)
35
L
39
(1.54)
T2
T1
16.5(0.65)
Cable outer diameter
L
8.0 to 10.0 (0.31 to 0.39)
94.5
(3.72)
10.0 to 12.0 (0.39 to 0.47)
Adapter body(M. Screw)
O-Ring
Packing case
Hexagon socket set screw
Packing *
Hexagon socket set screw
O-Ring
O-Ring
Washer
Union nut
Packing gland
Clamp ring
Clamp nut
O-Ring
B.coupling
Cable(user's scope)
Packing diameter
F
10.0(0.39)
12.0(0.47)
G
20.0
(0.79)
Unit : mm
(Approx. inch)
C
Identification
mark
16 8-10
16 10-12
D
Weight
kg
(lb)
0.26
(0.57)
5-11
IM 1E6D0-01E
5. OUTLINE
ø86
Eye bolt
(3.4)
48
(1.9)
50 (2.0)
H
2
L
325
-
G
J
A
330
AMD
W
G
U
335
340
Type
Size code
Size
Lining code
Face-to-face
Eye bolt hole dia.
Height
Outer dia.
JIS 10K
JIS 20K
JIS F12
Lining code A : PFA lining
U : Polyuretane lining
** Submersible meters are provided with 30m (98ft) of signal and excitation cable,
potted terminal box, and waterproof fitlings.
Please add 9.5kg (20.9lbs) in weight for submersible type.
Lining code A : PFA lining
U : Polyuretane lining
** Submersible meters are provided with 30m (98ft) of signal and excitation cable,
potted terminal box, and waterproof fitlings.
Please add 9.5kg (20.9lbs) in weight for submersible type.
Hole dia.
Outer dia.
Pitch circle dia.
Bolt hole pitch
Number of holes
JIS F12
Hole dia.
Thickness
Weight*kg(lb)
*Submersible flowtubes are provided with 30m (98ft) of signal and excitation
cables, potted terminal box, and waterproof gland fittings.
Please add 9.5kg(20.9lb) for submersible type.
*Submersible flowtubes are provided with 30m (98ft)
signal and excitation cables, potted termianl box, and
980(38.6)
1500(59.1)
1700(66.9)
5-13
waterproof gland fittings. Please add 9.5kg(20.9lb)
for submersible type.
IM 1E6D0-01E
5. OUTLINE
A1
16-R
T
S
L1
L
T
S
N- h
H
H1
A2
C
P
Q
Y
AM516/518/520/522/524/526
Model
Size
Face-to-face
Height
Height
Interval1
Interval2
Eye nut
Size
Width
Support
Length
Interval
Interval
Mounting
bolt
Hole dia.
Outer dia.
Pitch circle dia.
Bolt hole pitch
Number of holes
Hole dia.
JIS G5527
7.5K Flange
Thickness
Weigjt* kg(lb)
*Submersible flowtubes are provided with 30m (98ft) of signal and excitation cables, potted
terminal box, and waterproof gland fittings. Please add 9.5kg(20.9lb) for submersible type.
In this section, further requirements and differences for
explosion proof type instrument are described. For
explosion proof type instrument, the description in this
chapter is prior to other description in this Instruction
Manual.
NOTE
The terminal box cover is locked by special
screw. In case of opening the terminal box
cover, please use the Hexagonal Wrench
attached.
CAUTION
Be sure to lock the cover with the special screw
using the Hexagonal Wrench attached after
tightening the cover.
6.1FM
(2) Installation
WARNING
• All wiring shall comply with national electrical
code ANSI/NFPA 70 and local electrical code.
• In hazardous locations, wiring to be in conduit
as shown in the figure.
(3) Operation
WARNING
• Open circuit before opening the covers and
seal all conduits with in 18 inches in hazardous
locations.
• Take care not to generate mechanical spark
when access to the instrument and peripheral
devices in hazardous locations.
(4) Maintenance and Repair
(1) Technical Data
• Explosionproof for Class I Division 1 Groups A, B,
C & D. Dust-ignitionproof for Class II/III Division 1
Groups E, F & G. Intrinsically safe (electrode) for
Class I Division 1 Groups A, B, C & D
The instrument modification or parts replacement
by other than authorized representative of
Yokogawa Electric Corporation is prohibited and
will void the approval of Factory Mutual Research Corporation.
HAZARDOUS LOCATIONS
Magnetic Flow Converter
Conduit
18" (457 mm) Max.
Sealing Fitting
F0601.EPS
6-1
IM 1E6D0-01E
6. HAZARDOUS DUTY TYPE INSTRUMENT
6.2CSA
(1) Technical Data
• Nonincendive for Class I, Division 2, Groups A, B,
C & D ; Suitable for Class II, Division 2, Groups E,
F and G ; Class III, Division 2
Intrinsically safe (electrode) for Class I Division 1
Groups A, B, C & D
Electrode Circuit Vmax: 250V ac/dc
Excitation Circuit: 130Vp 72/75Hz, 11VAmax
Temp. CodeT5T4T3
Process Temp85120150°C
Enclosure: Type 4
(2) Installation
WARNING
All wiring shall comply with Canadian Electrical
Code Part I and Local Electrical Codes.
(3) Operation
WARNING
6.3CENELEC (KEMA)
WARNING
Only trained persons use this instrument in
industrial locations.
•All wiring shall comply with local installation
requirements and local electrical code.
• In hazadous locations, the cable entry devices
shall be of a certified flameproof type, suitable
for the conditions of use and correctly installed.
CAUTION : OPEN CIRCUIT BEFORE REMOV-
ING COVER.
OUVRIR LE CIRCUIT AVANT
D’ENLEVER LE COUVERCLE.
Take care not to generate mechanical spark
when access to the instrument and peripheral
devices in hazardous location.
(4) Maintenance and Repair
WARNING
The instrument modification or parts replacement
by other than authorized representative of
YOKOGAWA Electric Corporation or
YOKOGAWA Corporation of AMERICA is
prohibited and will void Canadian Standards
Explosionproof Certification.
(3) Operation
WARNING
• Open circuit before opening the covers.
• Take care not to generate mechanical spark
when access to the instrument and peripheral
devices in hazardous locations.
(4) Maintenance and Repair
WARNING
The instrument modification or parts replacement
by other than authorized representative of
Yokogawa Electric Corporation is prohibited and
will void the certification.
6-2
IM 1E6D0-01E
6.4JIS
WARNING
Care should be taken to install, wiring, piping to
keep safety. This instrument is restricted to
maintenance and repair. Please read "INSTALLATION AND OPERATING PRECAUTIONS
FOR JIS FLAMEPROOF EQUIPMENT" in the
end of this manual.
(1) Technical Data
• Construction : Exde [ia] II CT4X
: Terminal box ; Explosion proof
Flow Tube ; Increased Safety
Signal Circuit ; Intrinsically Safety (ia)
Um=250 V AC 50/60 Hz, 250 V DC,
Uo=250 V*, Io=3.37 mA*, Po=0.211 W
* Uo and Io are rms value.
6. HAZARDOUS DUTY TYPE INSTRUMENT
:Ignition and Explosion Class of gas or
vapour ; II CT4X
• Ambient Temperature : –10 to 60°C
• Fluid Temperature : 120°C or less
• Current Signal Circuit Allowable Voltage :
250V AC/DC
• Grounding: 10 Ω or less
WARNING
In case the ambient temperature exceeeds 50°C,
use heat-resistant cables with maximum allowable temperature of 70°C or above.
(2) Wiring
Flameproof packing adapter is attached to electrical
connection.
WARNING
Please use G12 (option) flameproof packing
adapter attached.
6-3
IM 1E6D0-01E
6. HAZARDOUS DUTY TYPE INSTRUMENT
18(0.71)
F
G
*Packing
(Choose from the table below
depend on cable outside diameter)
L
T2
Dimension
T1T2CD
G 1/2G 1/2
35
(1.38)
39
(1.54)
Figure 6.4.1 Flame Proof Packing Adapter
T1
16.5(0.65)
L
94.5
(3.72)
O-Ring
Packing case
Hexagon socket set screw
Packing *
Hexagon socket set screw
O-Ring
O-Ring
Washer
Cable outer diameter
8.0 to 10.0 (0.31 to 0.39)
10.0 to 12.0 (0.39 to 0.47)
Adapter body(M. Screw)
Union nut
Packing gland
Clamp ring
Clamp nut
O-Ring
B.coupling
Cable(user's scope)
Packing diameter
F
10.0(0.39)
12.0(0.47)
20.0
(0.79)
Unit : mm
(Approx. inch)
C
D
F0603.EPS
Identification
G
mark
16 8-10
16 10-12
Weight
kg
(lb)
0.26
(0.57)
T0601.EPS
Follow the procedure for flame proof packing adapter setting.
CAUTION
Before tightening, confirm cable length from terminals to flame proof packing adapter when setting. Once
it is tightened, loosening and re-tightening may damage its sealing performance.
(a)Measure the bi-direction of the cable outside diameter to one decimal place in mm.
(b)Calculate the average of these value, and choose the suitable packing from attached two. [Refer to the figure.
6.4.1]
(c)Screw the M.screw into the electrical connection. (Ensure O-ring)
(d)Set union nut, B.coupling, clamp nut, clamp ring, packing gland, packing case with inserted packing and washer
in that order around the cable, and tighten packing gland. And then tighten clamp nut. Confirm packing is
compressed and cable is fixed.
(e)Insert packing gland into M.screw.
(f)Connect each wire of cable to terminal.
(g)Tighten and fix union nut.
(h)Tighten and fix hexagon socket set screw of M.screw and union nut.
6-4
IM 1E6D0-01E
INSTALLATION AND OPERATING PRECAUTIONS FOR JIS FLAMEPROOF EQUIPMENT
INSTALLATION AND OPERATING PRECAUTIONS FOR
JIS FLAMEPROOF EQUIPMENT
Apparatus Certified Under Technical Criteria
(IEC-compatible Standards)
1.General
The following describes precautions on electrical apparatus
of flameproof construction (hereinafter referred to as
flameproof apparatus) in explosion-protected apparatus.
Following the Labour Safety and Health Laws of Japan,
flameproof apparatus is subjected to type tests to meet either
the technical criteria for explosionproof electrical machinery
and equipment (standards notification no. 556 from the
Japanese Ministry of Labour) (hereinafter referred to as
technical criteria), in conformity with the IEC Standards, or
the “Recommended Practice for Explosion-Protected
Electrical Installations in General Industries,” published in
1979. These certified apparatus can be used in hazardous
locations where explosive or inflammable gases or vapours
may be present.
Certified apparatus includes a certification label and an
equipment nameplate with the specifications necessary for
explosion requirements as well as precautions on explosion
protection. Please confirm these precautionary items and use
them to meet specification requirements.
For electrical wiring and maintenance servicing, please refer
to “Internal Wiring Rules” in the Electrical Installation
Technical Standards as well as “USER’S GUIDELINES for
Electrical Installations for Explosive Gas Atmospheres in
General Industry,” published in 1994.
construction is of completely enclosed type and its enclosure
shall endure explosive pressures in cases where explosive
gases or vapours entering the enclosure cause explosion. In
addition, the enclosure construction shall be such that flame
caused by explosion does not ignite gases or vapours outside
the enclosure.
In this manual, the word "flameproof" is applied to the
flameproof equipment combined with the types of protection
"e", "o", "i", and "d" as well as flameproof equipment.
3.Terminology
(1) Enclosure
An outer shell of an electrical apparatus, which encloses live
parts and thus is needed to configure explosion-protected
construction.
(2) Shroud
A component part which is so designed that the fastening of
joint surfaces cannot be loosened unless a special tool is
used.
(3) Enclosure internal volume
This is indicated by:— the total internal volume of the
flameproof enclosure minus the volume of the internal
components essential to equipment functions.
To meet flameproof requirements, equipment that can be
termed “flameproof” must:
(1) Be certified by a Japanese public authority in accordance
with the Labour Safety and Health Laws of Japan and
have a certification label in an appropriate location on its
case, and
(2) Be used in compliance with the specifications marked on
its certification label, equipment nameplate and
precautionary information furnished.
2.Electrical Apparatus of Flame-
proof Type of Explosion-Protected Construction
Electrical apparatus which is of flameproof construction is
subjected to a type test and certified by the Japanese Ministry
of Labour aiming at preventing explosion caused by electrical
apparatus in a factory or any location where inflammable
gases or vapours may be present. The flameproof
(4) Path length of joint surface
On a joint surface, the length of the shortest path through
which flame flows from the inside to outside of the
flameproof enclosure. This definition cannot be applied to
threaded joints.
(5) Gaps between joint surfaces
The physical distance between two mating surfaces, or
differences in diameters if the mating surfaces are cylindrical.
Note: The permissible sizes of gaps between joint surfaces, the path
length of a joint surface and the number of joint threads are
determined by such factors as the enclosure’s internal
volume, joint and mating surface construction, and the
explosion classification of the specified gases and vapours.
1
EX-B03E
INSTALLATION AND OPERATING PRECAUTIONS FOR JIS FLAMEPROOF EQUIPMENT
4.Installation of Flameproof Ap-
paratus
(1) Installation Area
Flameproof apparatus may be installed, in accordance with
applicable gases, in a hazardous area in Zone 1 or 2, where
the specified gases are present. Those apparatus shall not be
installed in a hazardous area in Zone 0.
Note: Hazardous areas are classified in zones based upon the
frequency of the appearance and the duration of an explosive
gas atmosphere as follows:
Zone 0: An area in which an explosive gas atmosphere is
present continuously or is present for long periods.
Zone 1: An area in which an explosive gas atmosphere is
likely to occur in normal operation.
Zone 2: An area in which an explosive gas atmosphere is not
likely to occur in normal operation and if it does
occur it will exist for a short period only.
(2) Environmental Conditions
The standard environmental condition for the installation of
flameproof apparatus is limited to an ambient temperature
range from –20°C to +40°C (for products certified under
Technical Criteria). However, some field-mounted
instruments may be certified at an ambient temperature up to
+60°C as indicated on the instrument nameplates. If the
flameproof apparatus are exposed to direct sunshine or
radiant heat from plant facilities, appropriate thermal
protection measures shall be taken.
5.External Wiring for Flameproof
Apparatus
Flameproof apparatus require cable wiring or flameproof
metal conduits for their electrical connections. For cable
wiring, cable glands (cable entry devices for flameproof type)
to wiring connections shall be attached. For metal conduits,
attach sealing fittings as close to wiring connections as
possible and completely seal the apparatus. All non-live
metal parts such as the enclosure shall be securely grounded.
For details, see the “USER’S GUIDELINES for Electrical
Installations for Explosive Gas Atmospheres in General
Industry,” published in 1994.
(1) Cable Wiring
• For cable wiring, cable glands (cable entry devices for
flameproof type) specified or supplied with the apparatus
shall be directly attached to the wiring connections to
complete sealing of the apparatus.
• Screws that connect cable glands to the apparatus are
those for G-type parallel pipe threads (JIS B 0202) with
no sealing property. To protect the apparatus from
corrosive gases or moisture, apply nonhardening sealant
such as liquid gaskets to those threads for waterproofing.
•Specific cables shall be used as recommended by the
“USER’S GUIDELINES for Electrical Installations for
Explosive Gas Atmospheres in General Industry,”
published in 1994.
•In necessary, appropriate protective pipes (conduit or
flexible pipes), ducts or trays shall be used for
preventing the cable run (outside the cable glands) from
damage.
• To prevent explosive atmosphere from being propagated
form Zone 1 or 2 hazardous location to any different
location or non-hazardous location through the protective
pipe or duct, apply sealing of the protective pipes in the
vicinity of individual boundaries, or fill the ducts with
sand appropriately.
•When branch connections of cables, or cable connections
with insulated cables inside the conduit pipes are made,
a flameproof or increased-safety connection box shall be
used. In this case, flameproof or increased-safety cable
glands meeting the type of connection box must be used
for cable connections to the box.
(2) Flameproof Metal Conduit Wiring
• For the flameproof metal conduit wiring or insulated
wires shall be used as recommended by the USER’S
GUIDELINES for Electrical Installations for Explosive
Gas Atmospheres in General Industry, published in
1994.
• For conduit pipes, heavy-gauge steel conduits
conforming to JIS C 8305 Standard shall be used.
•Flameproof sealing fittings shall be used in the vicinity
of the wiring connections, and those fittings shall be
filled with sealing compounds to complete sealing of the
apparatus. In addition, to prevent explosive gases,
moisture, or flame caused by explosion form being
propagated through the conduit, always provide sealing
fittings to complete sealing of the conduit in the
following locations:
(a) In the boundaries between the hazardous and non-
hazardous locations.
(b) In the boundaries where there is a different
classification of hazardous location.
• For the connections of the apparatus with a conduit pipe
or its associated accessories, G-type parallel pipe threads
(JIS B 0202) shall be used to provide a minimum of
five-thread engagement to complete tightness. In
addition, since these parallel threads do not have sealing
property, nonhardening sealant such as liquid gaskets
shall thus be applied to those threads for ensuring
waterproofness.
• If metal conduits need flexibility, use flameproof flexible
fittings.
2
EX-B03E
INSTALLATION AND OPERATING PRECAUTIONS FOR JIS FLAMEPROOF EQUIPMENT
6.Maintenance of Flameproof
Apparatus
To maintain the flameproof apparatus, do the following. (For
details, see Chapter 10 “MAINTENANCE OF EXPLOSIONPROTECTED ELECTRICAL INSTALLATION” in the
USER’S GUIDELINES for Electrical Installations for
Explosive Gas Atmospheres in General Industry.)
(1) Maintenance servicing with the power on.
Flameproof apparatus shall not be maintenance-serviced with
its power turned on. However, in cases where maintenance
servicing is to be conducted with the power turned on, with
the equipment cover removed, always use a gas detector to
check that there is no explosive gas in that location. If it
cannot be checked whether an explosive gas is present or not,
maintenance servicing shall be limited to the following two
items:
(a) Visual inspection
Visually inspect the flameproof apparatus, metal
conduits, and cables for damage or corrosion, and other
mechanical and structural defects.
(b) Zero and span adjustments
These adjustments should be made only to the extent
that they can be conducted from the outside without
opening the equipment cover. In doing this, great care
must be taken not to cause mechanical sparks with tools.
(2) Repair
If the flameproof apparatus requires repair, turn off the power
and transport it to a safety (non-hazardous) location. Observe
the following points before attempting to repair the
apparatus.
(a) Make only such electrical and mechanical repairs as will
restore the apparatus to its original condition. For the
flameproof apparatus, the gaps and path lengths of joints
and mating surfaces, and mechanical strength of
enclosures are critical factors in explosion protection.
Exercise great care not to damage the joints or shock the
enclosure.
(b) If any damage occurs in threads, joints or mating
surfaces, inspection windows, connections between the
transmitter and terminal box, shrouds or clamps, or
external wiring connections which are essential in
flameproofness, contact Yokogawa Electric Corporation.
requirements for flameproof apparatus (however, bear in
mind that the apparatus must always be restored to its
original condition). If you attempt to repair the
flameproof apparatus, company-specified components
shall be used.
(d) Before starting to service the apparatus, be sure to check
all parts necessary for retaining the requirements for
flameproof apparatus. For this, check that all screws,
bolts, nuts, and threaded connections have properly been
tightened.
(3) Prohibition of specification changes and
modifications
Do not attempt to change specifications or make
modifications involving addition of or changes in external
wiring connections.
7.Selection of Cable Entry Devices for Flameproof Type
IMPORTANT
The cable glands (cable entry devices for flameproof
type) conforming to IEC Standards are certified in
combination with the flameproof apparatus. So,
Yokogawa-specified cable entry devices for flameproof
type shall be used to meet this demand.
References:
(1) Type Certificate Guide for Explosion-Protected Con-
struction Electrical Machinery and Equipment (relating
to Technical Standards Conforming to International
Standards), issued by the Technical Institution of
Industrial Safety, Japan
(2) USER’S GUIDELINES for Electrical Installations for
Explosive Gas Atmospheres in General Industry (1994),
issued by the Japanese Ministry of Labour, the Research
Institute of Industrial Safet
CAUTION
Do not attempt to re-process threaded connections or
refinish joints or mating surfaces.
(c) Unless otherwise specified, the electrical circuitry and
internal mechanisms may be repaired by component
replacement, as this will not directly affect the
3
EX-B03E
Loading...