The following safety symbols are used on
the product as well as in this manual.
DANGER
This symbol indicates that an operator
must follow the instructions laid out in
this manual in order to avoid the risks, for
the human body, of injury, electric shock
or fatalities. The manual describes what
special care the operator must take to
avoid such risks.
WARNING
This symbol indicates that an operator
must refer to the instructions in this
manual in order to prevent the instrument
(hardware) or software from being
damaged, or a system failure from
occurring.
This symbol gives information essential
for understanding the operations and
functions.
CAUTION
NOTE
This symbol indicates information that
complements the present topic.
The Advanced Reux Sampler (YARS100) has been developed specically to meet the needs of modern petrochemical
analyzer sampling, especially ethylene cracked gas euents.
1.1 Features
Based on a proven simpler concept, the YARS includes several technologically advanced features outlined below:
• Self-acting temperature controller requiring no electrical power supply – Adjustable across the selected control
range.
• Outlet lter mesh/screen for mist trap to deter liquid carry over
• Sample outlet over temperature shut-o valve – Factory set for a pre-determined shut-o temperature and
typically with a 10oF span between fully open and fully closed.
• Enhanced, high stability temperature control system design
• High sample ow rate 3-5 liters per minute for reduced lag time
• A multitude of options congured specically for your application
SAFETY should be considered rst and foremost importance when working on the
equipment described in this manual. All persons using this manual in conjunction
with the equipment must evaluate all aspects of the task for potential risks, hazards
and dangerous situations that may exist or potentially exist. Please take appropriate
action to prevent ALL POTENTIAL ACCIDENTS.
Product development is a continuous policy of Yokogawa and therefore specications may be subject to change without
notice
The following generic specications apply however; please check exact features and model number with serial numbers
on the units provided.
5
A) Process Connection:
Process ange ratings are ASME B16.6-1996
B) Sample Outlet Connection:
Standard 3/8” o.d. Sawgelock® stainless steel
Flow Rate: Typically 3 Lts./min conditioned at 15°C or
4 Lts./min conditioned at 20°C, pending application
Pressure: ΔP across ARS, < 1psi
NOTE
The sample transport line should be heat traced and
insulated to prevent subsequent sample condensation.
C) Mounting:
Vertical on top of customer supplied process isolation
valve for horizontal process line.
NOTE
If mounting on vertical process line, install appropriate
elbow or Y-piece prior to process isolation valve to
ensure ARS mounted in vertical position – Refer to
Installation details.
D) Instrument Air Connection:
Standard 1/2” o.d. Sawgelock®, stainless steel
Supply Pressure: 60-110 psig (~ 54-7.5 Bar)
Flow Rate: Dual 10 SCFM Vortex Tubes: Standard
Dual 15 SCFM Vortex Tubes: Optional
Grade: Clean, Dry, Oil Free and <5μ particles with
This section of the manual describes the basic theory of operation, designed to give the user a better understanding and
help in operation and maintenance.
3.1 Brief History of Reux Samplers
The concept of the Reux Sampler was introduced in mid 1970’s and until the inception of the Advanced Reux Sampler
(YARS), it had changed very little. Process analysis needs during the mid 1970’s were not as critical as they are now and
consequently, the Reux Samplers designed previously no longer meet today’s demands that are more stringent. The
reux sampler was originally designed to overcome the problem of constantly plugging sample probes installed on the
Transfer Line Exchangers (TLE) of ethylene cracking furnace euents. This high temperature cracked gas euent is high
in particulate and condensables (moisture and heavy hydrocarbons) content and therefore prone to plugging. The idea
that the condensables could be used to backwash a particulate lter was conceived and a Reux Sampler was then
developed. A simple lter mesh was used to trap the particulate matter.
3.2 Filter Section
The lower section of the YARS is for particulate ltration and provides some primary sample cooling. Typically ange
mounted directly to a customer supplied process isolation valve and ange mounted to the heat exchanger section of
the YARS. The lter media is nominally stainless steel wire mesh that traps any particulate matter such as coke from the
ethylene furnaces and/or catalyst nes. During normal operation, the temperature at the lter section outlet should be
in the order of 120oF (50oC), indicating a balanced reuxing condition is established. This lter section is cooled by the
condensate that has dropped out from the heat exchanger mounted directly above the lter. The condensate also acts as
a trap for ner particulate matter. In certain environmental conditions where sub-zero temperatures are anticipated, it may
be desirable to insulate the lter section. This will prevent premature condensing and allow for a balanced reuxing action.
3.3 Steam Supplement
For the YARS to function reliably there must be sucient quantity of condensable media in the sample. For applications
that have high particle loading and limited condensables present, steam should be added at the inlet. The YARS lter
section has a steam inlet facility located at the inlet. A suitably rated needle valve is provided for regulating a nominal ow
of steam in to the lter section via an inverted tube such that the steam is directed up in to the lters. Note that the steam
is added directly to the sample and therefore the composition will be aected the H2O addition. As each application has
many dierent operating parameters, there are no specic settings for the quantity of steam addition and each application
is set-up individually.
3.4 Inlet Temperature Section
Mounted between the lter and heat exchanger section is the inlet temperature thermowell. As standard, a suitable
temperature gauge is installed in the thermowell. The purpose of temperature measurement at this point in the YARS is
to help establish the correct temperature gradient required for reuxing. The two undesirable conditions within the ler
section are as below:
•Temperature reading too high – will cause the heat exchanger to work excessively and indicates a problem
elsewhere in the system
•Temperature reading too low – will cause the condensable media to dropout early in the lter section and not
provide proper lter cleansing by reux.
When troubleshooting the YARS, consider the correct temperature at this point.
3.5 Heat Exchanger Section
This is truly unique to the YARS in many respects. This section provides the cooling of the sample and is most critical to
the overall operation. As the sample gas rises from the lter section, through the inlet temperature section, it is channeled
in to the heat exchanger chamber. The sample gas rises through a chamber in which the cooling coils are concentrically
mounted. The design of the heat exchanger is such that maximum surface area contact between the cooling coils and
the sample gas is achieved. There are two cooling coils located in the same heat exchanger chamber, which helps
provide more accurate and ecient cooling. In the head-space of the heat exchanger, a stainless steel ne lter mesh is
installed to act as a mist trap and nal liquid sample droplet trap prior to existing the sampler. Flow of the cooling media
through each coil is automatically controlled to ensure a constant outlet temperature is maintained. The sample is normally
saturated at the outlet temperature and pressure. The cooling media is supplied by vortex air cooler(s) via the self-acting
controller.
3.6 Outlet Temperature Section
Located on the outlet of the heat exchanger is the outlet temperature indicator. For self-acting controller units, a
temperature sensor is installed. The self-acting temperature controller is located (screwed) directly in to the top of the
heat exchanger in to the headspace area. This provides the most accurate temperature reading and control of the actual
sample outlet temperature. The purpose of temperature measurement at this point in the YARS is to control the nal outlet
temperature (or maximum dewpoint). Depending upon the application, this outlet temperature may be typically between
41-80oF (5-27oC).
3.7 Self Acting Temperature Controller
The operation and maintenance of the adjustable self-acting temperature controller is described herein.
3.8 Instrument Air for the Vortex Tube
The standard media used by the Vortex air tube(s) is ISA grade Instrument air. It is possible to change the cool air
outlet temperature by adjusting the recessed screw (needle valve) located in the hot air exhaust. A counter-clockwise
adjustment of the valve will produce colder temperatures while a clockwise adjustment will produce warmer temperatures.
This ability to adjust the temperature may be considered when ne-tuning the temperature control system to ensure
optimum temperature stability. The performance of the Vortex tubes is also based on the incoming instrument air
temperature. For more ecient operation, avoid direct sun light (and other heat emitting sources) when running the air line.
7
3.9 Vortex Theory of Operation:
Vortex tubes are simple devices that contain no moving parts and
therefore may be used as a reliable device for generating a cold air supply.
Compressed air (clean, dry, ISA Grade) will enter the cylindrical chamber
(to the left of the diagram) and is caused to rotate internally. This rotating
air moves down the inner walls of the hot air exhaust chamber at extremely
high velocity. When exiting the hot exhaust (to the top of the diagram), a
portion of the air is directed back through the center at lower speed than
the exiting air. Heat transfer from the inner, slow moving air to the exiting,
fast moving air causes cold air to be generated. This cold air exits via the
cold air exhaust (shown at the bottom of the diagram).
Vortex tubes produce less air ow at colder temperatures and have less
BTUH capacity. Tubes are available in a variety of sizes depending upon
the application. Maximum eciency will be achieved when operating
at 100 psig instrument air pressure and while they can operate at lower
pressure, the eciency will be less (typically 60% ecient at 55 psig)
Section 2 – General Specication listed above denes the required utilities needed for YARS operation. Also, refer to the
General Installation drawing provided specically for the YARS by serial number for the hook-up details.
4.1 Instrument Air
The media used by the Vortex air tube(s) is ISA grade Instrument air. Plant Nitrogen may be used in place of the air,
provided it meets the required pressure and ow demands of the YARS. The air shall be clean, dry (-40oC dew point),
oil free and contain particles <5µ. The instrument air pressure at the YARS shall be 60-100 psig (~ 4-7 Barg). Failure to
use a lter will cause potential freezing and clogging of the compressed air path inside the Vortex Tube. Filter elements
must be changed on a regular basis. Frequency of change is determined by the conditions of the compressed air
supply.
At pressure less than this, the Vortex tube performance declines signicantly and may be unable to provide the required
quantity of cooling air. Any instrument air-line run to the YARS in the eld should be 1/2” o.d. to ensure adequate ow at
the desired pressure.
Electronically controlled samplers use two vortex tubes, each typically with a 4 SCFM ow capacity (Check Data Sheets
for your installed model) however, the actual air demand during normal operation will vary depending on the changing
process and climatic conditions at each site.
CAUTION
• COMPRESSED AIR COULD CAUSE DEATH, BLINDNESS OR INJURY
• DO NOT OPERATE A VORTEX TUBE AT AIR PRESSURES ABOVE150PSIG (10.3 BARG)
• DO NOT OPERATE A VORTEX TUBE AT LINE TEMPERATURES ABOVE 110OF (43OC)
• AVOID DIRECT CONTACT WITH COMPRESSED AIR
• DO NOT DIRECT COMPRESSED AIR AT ANY PERSON
• WHEN USING COMPRESSED AIR, WEAR SAFETY GLASSES AND SIDE SHIELDS
The performance of the Vortex tubes is also based on the incoming instrument air temperature and therefore, for more
ecient operation, avoid direct sun light (and other heat emitting sources) when running the air-line.
4.2 Low Pressure Steam
The sampler is equipped with low-pressure steam injection valve located in the lower lter section. Steam may be used
either as an addition at the inlet (for assisted reux action), or for cleansing through either the inlet or outlet.
NOTE
Do not use steam temperatures in excess of 232oC (450oF).
Ensure the needle valves are closed prior to connecting the steam supply and ensure the steam supply line is adequately
insulated. Connect the steam supply directly to the needle valve supplied with the sampler.
NOTE
WHEN OPERATING WITH STEAM, NOTE THAT SURFACES MAY BE EXTREMELY HOT AND CAN CAUSE SEVERE BURNS
AND OTHER PERSONAL INJURIES – ADHERE ALL SAFETY PROCEDURES APPLICABLE AT SITE FOR STEAM SERVICES
- FAILURE TO DO SO COULD RESULT IN INJURY OR DEATH.
Typically, the YARS will be subject to large vibration conditions and therefore a sound mechanical installation is required.
While ange mounting to the customer supplied process isolation valve provides an installation, it is also recommended
that the upper section of the YARS be secured mechanically. The heat exchanger and controller give the YARS ‘top-heavy’
properties that should be considered during installation.
It is also desirable to have ready access to the unit such that routine maintenance checks can be made on the operation
of the unit in a safe manner.
Ensure that an appropriate gasket is used between the ange mounting faces and when tightening the ange mounting
bolts to the process isolation valve, please ensure that the correct torque is applied (typically 100 foot pounds for a 2”
300# ange). As with any anged bolt tightening, ensure that the correct sequence of tightening is followed (typically as
shown below):
1
8
6
9
3
5
2
4
7
5.2 Fast Loop Line Size and Response Times
The following table may be used when estimating sample line sizes, response times and resultant pressure drop through
the line. These calculations are based on a typical ethylene cracked gas euent sample and should be used for estimating