REVISION RECORD ......................................................................................................... 1
iii
IM 01E20C01-01E
1.INTRODUCTION
NOTE
1. INTRODUCTION
This instrument has been adjusted at the factory before
shipment.
To ensure correct use of the instrument, please read
this manual thoroughly and fully understand how to
operate the instrument before operating it.
Regarding This User's Manual
• This manual should be provided to the end user.
• Before use, read this manual thoroughly to comprehend its contents.
• The contents of this manual may be changed
without prior notice.
• All rights are reserved. No part of this manual may
be reproduced in any form without Yokogawa's
written permission.
• Yokogawa makes no warranty of any kind with
regard to this material, including, but not limited to,
implied warranties of merchantability and suitability
for a particular purpose.
• All reasonable effort has been made to ensure the
accuracy of the contents of this manual. However,
if any errors or omissions are found, please inform
Yokogawa.
• Yokogawa assumes no responsibilities for this
product except as stated in the warranty.
• Please note that this user's manual may not be
revised for any specification changes, construction
changes or operating part changes that are not
considered to affect function or performance.
• If the customer or any third party is harmed by the
use of this product, Yokogawa assumes no responsibility for any such harm owing to any defects in the
product which were not predictable, or for any
indirect damages.
ments. If this instrument is used in a manner not
specified in this manual, the protection provided by
this instrument may be impaired.
• Yokogawa will not be liable for malfunctions or
damage resulting from any modification made to this
instrument by the customer.
• The following safety symbol marks are used in this
user's manual and instrument.
WARNING
A WARNING sign denotes a hazard. It calls
attention to procedure, practice, condition or the
like, which, if not correctly performed or adhered
to, could result in injury or death of personnel.
CAUTION
A CAUTION sign denotes a hazard. It calls
attention to procedure, practice, condition or the
like, which, if not correctly performed or adhered
to, could result in damage to or destruction of
part or all of the product.
IMPORTANT
An IMPORTANT sign denotes that attention is
required to avoid damage to the instrument or
system failure.
NOTE
Please refer to manual IM 01E20D01-01E for
information of the AXF Remote Flowtube.
Safety and Modification Precautions
• The following general safety precautions must be
observed during all phases of operation, service, and
repair of this instrument. Failure to comply with
these precautions or with specific WARNINGS
given elsewhere in this manual violates safety
standards of design, manufacture, and intended use
of the instrument. Yokogawa assumes no liability for
the customer's failure to comply with these require-
A NOTE sign denotes information necessary for
essential understanding of operation and features.
Protective grounding terminal
Functional grounding terminal
(This terminal should not be used as a protective
grounding terminal.)
Alternating current
Direct current
1-1
IM 01E20C01-01E
1. INTRODUCTION
1.1Using the Magnetic
Flowmeter Safely
WARNING
(1) Installation
• Installation of the magnetic flowmeter must be
performed by expert engineer or skilled personnel. No operator shall be permitted to perform
procedures relating to installation.
• The magnetic flowmeter is a heavy instrument.
Be careful that no damage is caused to personnel through accidentally dropping it, or by
exerting excessive force on the magnetic
flowmeter. When moving the magnetic flowmeter, always use a trolley and have at least two
people carry it.
• When the magnetic flowmeter is processing hot
fluids, the instrument itself may become extremely hot. Take sufficient care not to get
burnt.
• Where the fluid being processed is a toxic
substance, avoid contact with the fluid and
avoid inhaling any residual gas, even after the
instrument has been taken off the piping line for
maintenance and so forth.
• Do not apply excessive weight, for example, a
person sttepping on the magnetic flowmeter.
• All procedures relating to installation must
comply with the electrical code of the country
where it is used.
(2) Wiring
• The wiring of the magnetic flowmeter must be
performed by expert engineer or skilled personnel. No operator shall be permitted to perform
procedures relating to wiring.
• When connecting the wiring, check that the
supply voltage is within the range of the voltage
specified for this instrument before connecting
the power cable. In addition, check that no
voltage is applied to the power cable before
connecting the wiring.
• The protective grounding must be connected
securely at the terminal with the mark to
avoid danger to personnel.
(3) Operation
• When opening the cover, wait for more than 10
minutes after turning off the power. Only
expert engineer or skilled personnel are permitted to open the cover.
• Be sure to set parameters as "Protect" on the
write protect function after finish of parameter
setting work.
Under extremely rare case, the infra-red
switches may respond unexpectedly in such
conditions as sticking ball of water or
extraneous substances on the surface of
display panel glass according to the principle of
infra-red switch operation.
Its probability rises in such cases as sticking
rain water by storm or other similar situation
and washing up work near flowmeter installation place.
Either to illuminate or stop illuminating the infrared switches by the flashlight may cause the
mis-reaction.
Refer to Chapter 6 "Menu P: Parameter Protection Items" and section "10.2.3" how to use the
write protect function in detail.
(4) Maintenance
• Maintenance of the magnetic flowmeter should
be performed by the trained personnel having
knowledge of safety standard. No operator
shall be permitted to perform any operations
relating to maintenance.
• When opening the cover, wait for more than 10
minutes after turning off the power.
• Always conform to maintenance procedures
outlined in this manual. If necessary, contact
Yokogawa.
• Care should be taken to prevent the build up of
dirt, dust or other substances on the display
panel glass or data plate. If these surfaces do
get dirty, wipe them clean with a soft dry cloth.
1.2Warranty
• The terms of this instrument that are guaranteed are
described in the quotation. We will make any repairs
that may become necessary during the guaranteed
term free of charge.
• Please contact our sales office if this instrument
requires repair.
• If the instrument is faulty, contact us with concrete
details about the problem and the length of time it
has been faulty, and state the model and serial
number. We would appreciate the inclusion of
drawings or additional information.
• The results of our examination will determine
whether the meter will be repaired free of charge or
on an at-cost basis.
The guarantee will not apply in the following
cases:
• Damage due to negligence or insufficient maintenance on the part of the customer.
1-2
IM 01E20C01-01E
1. INTRODUCTION
• Problems or damage resulting from handling,
operation or storage that violates the intended use
and specifications.
• Problems that result from using or performing
maintenance on the instrument in a location that
does not comply with the installation location
specified by Yokogawa.
• Problems or damage resulting from repairs or
modifications not performed by Yokogawa or
someone authorized by Yokogawa.
• Problems or damage resulting from inappropriate
reinstallation after delivery.
• Problems or damage resulting from disasters such as
fires, earthquakes, storms, floods, or lightning strikes
and external causes.
Trademarks:
ADMAG, AXF and ADMAG AXF are registered
trademarks of Yokogawa Electric Corporation.
Company names and product names used in this
material are registered trademarks or trademarks of
their respective owners.
maintenance, and repair are strictly restricted,
and non-observance or negligence of these
restriction would result dangerous condition.
1.3Combination Remote
Flowtubes
IMPORTANT
• The AXFA11 Magnetic Flowmeter Converter
should be used in combination with the following remote flowtubes:
AXFA11G ⇔AXF002-N to AXF26L-N
Contact Yokogawa before using it in combination with flowtubes other than those listed
above.
• The model AXFC remote flowtube with
optional code JF3 (TIIS flame proof type)
cannot be combined with the AXFA11 converter. In this case, use the AXFA14 converter.
• If the converter combined with the AXF magnetic flowmeter remote flowtube is changed
from the AXFA11 to AXFA14 or vice versa, the
meter factor of the remote flowtube must be
readjusted according to its flow calibration.
CAUTION
In case of combination with the explosion proof
type remote flowtube (AXFC-N) for
ATEX, FM, and CSA certification, please see the
manual IM 01E20D01-01E. The construction of
the instrument, installation, external wiring,
1-3
IM 01E20C01-01E
2.HANDLING PRECAUTIONS
2. HANDLING PRECAUTIONS
This instrument has been inspected carefully at the
factory before shipment. When the instrument is
delivered, visually check that no damage has occurred
during transportation.
Read this section carefully as it contains important
information on handling this instrument. Refer to the
relevant sections for information not contained in this
section. If you have any problems or questions, please
contact Yokogawa sales office.
2.1Checking Model and
Specifications
The model code and specifications are found on the
data plate located on the outside of the case. Check
that the model code and specifications match what you
have ordered.
Be sure you have your model number and serial
number available when contacting Yokogawa.
2.3Storage Precautions
If the instrument is to be stored for a long period of
time after delivery, observe the following points.
The instrument should be stored in its original
packing condition in the storage location.
Select a storage location that fulfils the following
conditions:
• A place where it will not be exposed to rain or
water
• A place subject to minimal vibrations or shocks
• Temperature and humidity levels should be as
follows:
Temperature: -30 to 70°C
Humidity: 5 to 80% RH (no condensation)
The preferred ambient temperature and
humidity levels are 25°C and approximately
65% RH.
If the AXFA11 is transferred to the installation site
and stored without being installed, its performance
may be impaired due to the infiltration of rainwater
and so forth. Be sure to install and wire the
AXFA11 as soon as possible after transferring it to
the installation location.
F0201.EPS
2.2Accessories
Check that the parts shown below are included in the
package:
Mounting hardware: 1 set
2.4Installation Location
Precautions
Select the installation location with consideration to the
following items to ensure long-term stable operation of
the instrument.
Ambient Temperature:
Avoid installing the instrument in locations with
constantly fluctuating temperatures. If the location is
subject to radiant heat from the plant, provide heat
insulation or improve ventilation.
Atmospheric Condition:
Avoid installing the instrument in a corrosive
atmosphere. In situations where this is unavoidable,
consider ways to improve ventilation and to prevent
rainwater from entering and being retained in the
conduit pipes.
Vibrations or Shocks:
Avoid installing the instrument in a place subject to
shocks or vibrations.
2-1
IM 01E20C01-01E
3. INSTALLATION
3.INSTALLATION
WARNING
Installation of the magnetic flowmeter must be performed by expert engineer or skilled personnel. No
operator shall be permitted to perform procedures relating to installation.
3.1Installation Location
IMPORTANT
Install the instrument in a location where it is not exposed to direct sunlight. For ambient temperature,
refer to Chapter 11 “OUTLINE”.
The instrument may be used in an ambient humidity where the RH ranges from 0 to 100%. However,
avoid long-term continuous operation at relative humidity above 95%.
3.2Mounting
This instrument can be mounted using surface mounting, 2-inch pipe mounting, or panel
mounting.
Surface Mounting (Wall Mounting)
Unit: mm
(approx. inch)
For surface mounting, use the mounting
fixture provided, using M6 screws.
These M6 screws must be provided
by the user.
194 (7.64)
4-6 Hole or
M6 Screw
65
(2.56)
F0301.EPS
Figure 3.2.1 Surface Mounting
IMPORTANT
Mounting fixture on equipment intended to be mounted on a wall or ceiling shall withstand a force of four
times the weight of the equipment (AXFA11: 3.3kg (7.3 lb)).
3-1
IM 01E20C01-01E
2-inch Pipe Mounting
2-inch pipe
Mounting fixture
Washer
Clamp screw
Pass the four clamp screws through the mounting fixture, position it on the 2-inch pipe,
and then fasten the AXFA11 in place.
3. INSTALLATION
Figure 3.2.2 2-inch Pipe Mounting
Panel Mounting
F0302.EPS
Unit: mm
(approx. inch)
Panel cutout
203 (8.0)
Mounting fixture
172 (6.8)
R3MAX
Washer
Screw
Fit the AXFA11 into the panel. Then attach the mounting fixture to the AXFA11 using the screw and the
washer, and secure the instrument with the two clamp screws.
Figure 3.2.3 Panel Mounting
3-2
Clamp screw
F0303.EPS
IM 01E20C01-01E
4.WIRING
4. WIRING
This section describes the wiring on the converter side
only. For information relating to wiring on the
flowtube side, refer to the user’s manual of the AXF
Remote Flowtube (IM 01E20D01-01E).
WARNING
The wiring of the magnetic flowmeter must be
performed by expert engineer or skilled personnel. No operator shall be permitted to perform
procedures relating to wiring.
CAUTION
Once all wiring is complete, check the connections before applying power to the instrument.
Improper arrangements or wiring may cause a
unit malfunction or damage.
4.1Wiring Precautions
Be sure to observe the following precautions when
wiring:
four-core cables are used for wiring. Keep
conduits or flexible tubes watertight using
sealing tape.
• Ground the remote flowtube and the converter
separately.
• Cover each shield of the signal cable with vinyl
tube or vinyl tape to avoid contact between two
shields or between a shield and a case.
• When waterproof glands or union equipped
waterproof glands are used, avoid tightening
the glands with an excessive torque.
• Be sure to turn power off before opening the
cover.
• Before turning the power on, tighten the cover
securely.
4.2Cables
(1) Dedicated Signal Cable (AXFC)
Conductors (A and B)
Shields (SA and SB)
Tape
Outer jacket
10.5 (0.413")
CAUTION
• In cases where the ambient temperature
exceeds 50°C (122°F), use external heatresistant wiring with a maximum allowable
temperature of 70°C (158°F) or above.
• Do not connect cables outdoors in wet weather
in order to prevent damage from condensation
and to protect the insulation.
• Do not splice the cable between the flowtube
terminal and the converter if it is too short.
Replace the short cable with a cable that is the
appropriate length.
• All the cable ends must be provided with round
crimp-on terminals and be securely wired.
• The signal cables must be routed in separate
steel conduit tubes 16 (JIS C 8305) or flexible
conduit tubes 15 (JIS C 8309).
• Always route the power and output signal
cables in separate steel conduit tubes, except
when the power supply voltage is 24 V and
Shield (C)
Insulation
Insulation
Figure 4.2.1 Dedicated Signal Cable AXFC
F0401.EPS
The flow signal is transmitted via this dedicated cable.
The cable is constructed with double shielding over the
two conductors, and heat-resistant vinyl is used for the
outer jacket material.
Finished diameter:10.5 mm (0.413")
Maximum length:200 m (660 ft)
Maximum temperature:80°C (176°F)
IMPORTANT
If the cable is longer than required, cut off any
extra length rather than coiling it up, and terminate the conductors as shown in Figure 4.2.2.
Avoid using junction terminal boards to extend
the cable length, as this will interrupt the shielding.
4-1
IM 01E20C01-01E
4. WIRING
NOTE
Unit : mm
(approx. inch)
ACB
SA
White Black
On the
converter
side
25 (0.98)
5
150
8(0.3) max.
(5.9)
20 (0.8)
SB
50 (1.97)
Red
60 (2.36)
70 (2.76)
L (Specified Dimensions)
AXFC
ACB
55 (2.17)
90 (3.54)
White Black Red
8 (0.3) max.
5
On the
(5.9)
150
flowtube
side
10.5 (0.4)
F0402.EPS
Figure 4.2.2 Treatment of Dedicated Signal Cables
CAUTION
• As crimp terminals A, B, SA, SB and C have
their own electrical potentials, securely insulate
them so as not to come in contact with one
another.
• To prevent a shield from coming in contact with
another shield or the case, cover each shield
with a vinyl tube or wrap it in vinyl tape.
90 (3.54)
With gland options EG, EU and EW;
• Excitation cable;
10.5 or 11.5 mm (0.41 or 0.45 in.)
• Power and output cable;
7.5 to 12 mm (0.30 to 0.47 in.)
With gland options EP;
6 to 12 mm (0.24 to 0.47 in.)
Nominal Cross Section:
Single wire; 0.5 to 2.5 mm
Stranded wire; 0.5 to 1.5 mm
2
2
In case of power cable, Green/Yellow covered conductor shall be used only for connection to PROTECTIVE
CONDUCTOR TERMINALS. Conform to IEC227,
IEC245 or equivalent national authorization.
Unit : mm
(approx. inch)
Crimp Terminal
EX2
EX1
85 (3.35)
On the converter side
85 (3.35)
On the flowtube side
EX1
EX2
NOTE
Conductors A and B carry the signal from the
electrodes, and C is at the potential of the liquid
(signal common). Shields SA and SB are kept
at the same potentials as the individual electrodes (these are actively driven shields.) This is
done to reduce the effect of the distributed
capacitance of the cable at long cable length.
Note that, since the signals from the individual
electrodes are impedance converted inside the
converter, errors will result if they come in
contact with any other component. Great care
must be taken in the cable end treatment.
(2) Excitation Cable/Power Cable/Output
Cable
JIS C3401 control cable equivalent
JIS C3312 power cable equivalent
14 AWG Belden 8720 equivalent
Outer Diameter:
With no gland option;
6.5 to 12 mm (0.26 to 0.47 in.)
F0403.EPS
Figure 4.2.3 End Treatment of Excitation Cable
• For excitation and power cables, always use a
crimp terminal with an insulation cover.
• Use crimp tools from the manufacturer of the
crimp terminal you want to use to connect the
crimp terminal and cable.
• Use crimp tools that are appropriate for the
diameter of the cable to be connected.
4-2
IM 01E20C01-01E
4. WIRING
4.3Wiring Ports
This instrument is of watertight construction as
stipulated in JIS C0920. It is shipped with a wiring
bracket (waterproof gland or waterproof gland with
union) or a plastic gland attached, only in cases where
an optional specification is selected for the wiring port.
IMPORTANT
The wiring port is sealed with a cap (not waterproof). Do not remove the cap from the unused
wiring port. If waterproof property is necessary,
please use waterproof glands.
(1) When waterproof property is unnecessary
(When there are no particular optional
specifications)
The wiring port is sealed with a cap (not water-proof)
that must be removed before wiring. At this time,
handle the wiring port in accordance with the JIS
C0920 mentioned above. Do not remove the cap from
the unused wiring port.
For working on the electric wire tubes or the flexible
tubes (PF1/2), remove the waterproof gland and attach
them directly to the wiring port.
Washer
PF1/2
Gasket
Tightening gland
Cable
When working on conduit pipes or flexible pipes (PF1/2 only)
F0405.EPS
Figure 4.3.2 Waterproof Gland with Union Joint
(Optional Code EU)
(2) When waterproof property is necessary
(Wiring using waterproof glands)
IMPORTANT
To prevent water or condensation from entering
the converter housing, waterproof glands are
recommended. Do not over-tighten the glands or
damage to the cables may result. Tightness of
the gland can be checked by confirming that the
cable is held firmly in place.
Washer
Gasket
Waterproof gland
Cable
F0404.EPS
Plastic gland
Figure 4.3.3 Plastic Gland (Optional Code EP)
PF1/2
PF3/4
Extension plug (x2)
Conversion plug (x5)
Cable
When working on electric wire tube or flexible tube (PF3/4)
* When connecting PF1/2, remove the conversion plug and
connect directly to wiring port.
When wiring the conduits, pass the conduit through the
wiring connection port, and utilize the waterproof
gland to prevent water from flowing in. Place the
conduit pipe on an angle as shown in Figure 4.3.5.
Install a drain valve at the low end of the vertical pipe,
and open the valve regularly.
Drain valve
F0408.EPS
Figure 4.3.5 Conduit Wiring
4. WIRING
4-4
IM 01E20C01-01E
4.4Wiring Connections
4.4.1Removing Cover
While supporting the front of the cover with your hand, flip the connecting screw protective cover over, and remove the four connecting screws.
Figure 4.4.1 Removing the Front Cover
4.4.2Terminal Configuration
When the cover is removed, the connection terminals will be visible. The terminal configuration labels are attached in the position shown in Figure 4.4.2.
4. WIRING
F0409.EPS
SO2+
N/– L/+EX2EX1P– SI1+ SI2+ COMP+
AL+ AL– CSAAB SB
ALARM OUT
EX2EX1
PULSE OUTSTATUS IN
SIGNAL
P– SI1+ SI2+ COMP+
I+I–
CURRENT OUT
POWER SUPPLY
SO1+COMSO2+
STATUS OUT
N/– L/+
I+ I–AL+ AL– C SA AB SBSO1+COM
EXCIT ATION
Figure 4.4.2 Terminal Layout Labels Position
The description of the terminal symbols is shown in Table 4.4.1.
Table 4.4.1 Terminal Symbols
Terminal Symbols
SIGNAL
ALARM OUT
STATUS OUT
CURRENT OUT
C
SA
A
B
SB
AL+
AL-
SO1+
SO2+
COM
I+
I-
Description
Flow signal input
Alarm output
Status output
(Two output)
Current output
4 to 20mA DC
Terminal Symbols
STATUS IN
PULSE OUT
EXCITATION
POWER SUPPLY
Description
Sl1+
Status input
Sl2+
(Two input)
COM
P+
Pulse output
P-
Excitation current
EX1
EX2
output
L /+
Power supply
N/-
Functional grounding
Protective grounding
(Outside of the terminal)
T0401.EPS
F0410.EPS
4-5
IM 01E20C01-01E
4. WIRING
IMPORTANT
Do not wire the terminal without terminal symbols in terminal layout labels.
4.4.3Precautions for Wiring of Power
Supply Cables
When connecting to the power supply, observe the
points below. Failure to comply with these warnings
may result in an electric shock or damage to the
instrument.
WARNING
• Ensure that the power supply is OFF in order to
prevent electric shocks.
• Ensure the protective grounding terminal is
grounded before turning the power on.
• Use insulating sleeve crimp terminals (for 4-mm
screws) for the power supply wiring and protective grounding wiring.
• To prevent electric shocks, ensure the electrical
wiring cover (transparent) is attached.
• Install an external switch or circuit breaker as a
means to turn the powe off (capacitance; 15A,
conforming to IEC947-1 and IEC947-3). Locate
this switch either near the instrument or in other
places facilitating easy operation. Affix a
“Power Off Equipment” label to this external
switch or circuit breaker.
4.4.4DC Power Connection
When using DC power as the power supply for the
converter, give attention to the following points.
(1) Connecting Power Supply
IMPORTANT
Do not connect power supply with reversed
polarities.
L/+ terminal: connect +
N/– terminal: connect
IMPORTANT
Do not connect power supply with 100 to 240 V
AC or 100 to 120 V DC in the case of a 24 V
power supply version (power supply code 2).
It will give a damage to the converter.
(2) Required Power Supply Voltages
IMPORTANT
When using a 24 V power supply, the specification for the supply voltage is 24 V (–15% to
+20%), but the input voltage of the converter
drops due to cable resistance therefore it must
be used within the following ranges.
Supply Voltage and Cable Length
–
Wiring Procedure
1. Turn the instrument's power off, and remove the
wiring cover (transparent).
2. Wire the power supply cable and the functional
grounding cable to the power supply terminals.
CUR OUT STATUS OUT ALARM OUTSIGNAL
Functional
grounding cable
Figure 4.4.3 Electric Cable Wiring
POWER SUPPLY EXCITER
P–P+
PULSE OUTSTATUS IN
Power supply cable
SBBASACAL–AL+COMSO2+SO1+I+ I–
COMSI2+SI1+EX2EX1L/+N/–
F0411.EPS
3. Reattach the electrical wiring cover (transparent).
420 (1380)
280 ( 920)
200 ( 660)
100 ( 330)
Allowable cable length m(ft)
0
20.422242628.8
Usable range E (V)
Cable cross section area: 1.25 mm
Cable cross section area: 2 mm
(3) Setting Power Supply Frequency
IMPORTANT
Set the local commercial power frequency in
order to eliminate the effect of induction noise
from the commercial power supply.
Refer to “Chapter 6: Parameter Description” in
this manual. Parameter No. J30 and J31.
4-6
2
2
F0411-2.EPS
IM 01E20C01-01E
4. WIRING
4.4.5Grounding
CAUTION
Be sure to connect the protective grounding of
the AXFA11 with a cable of 2mm2 or larger cross
section in oder to avoid electrical shock to the
operators and maintenance engineers and to
prevent the influence of external noise. Connect
the grounding wire to the mark (100 or
less).
IMPORTANT
When optional code A (lighting protector) is
selected, the ground should satisfy Class C
requirements (grounding resistance, 10 or
less).
• The protective grounding terminals are located
on the inside and outside of the terminal area.
Either terminal may be used.
• Use 600 V vinyl insulation wires as the grounding
wires.
4.4.6Wiring the Remote Flowtube
with the AXFA11 Converter
WARNING
Before wiring, be sure that the power supply for
AXFA11 converter has been turned off to
prevent an electrical shock.
(1) Connection with the Remote Flowtube
(General-Purpose Use, Submersible Type,
Sanitary Type, Size 2.5 to 400 mm (0.1 to
16 in.))
Connect wiring as shown in the figure below.
I+ I–
CURRENT OUT
N/– L/+
POWER SUPPLY
A
B
C
Figure 4.4.5 Wiring Diagram
AL+ AL– C SA ABSB
SO1+COMSO2+
ALARM OUT
STATUS OUT
EX2EX1
P– SI1+ SI2+ COMP+
EXCIT ATION
PULSE OUT STATUS IN
Excitation cable
Remote flowtube
EX2
EX1
SIGNAL
AXFA11 Converter
AXFC Dedicated signal
cable
Converter
SA
A
B
SB
C
EX1
EX2
* Individually tape and insulate the
shields corresponding to SA and
SB on the remote flowtube side.
(Explosion proof Type, Size 2.5 to 400 mm
(0.1 to 16 in.))
IMPORTANT
In case of TIIS Flame proof type, a remote
flowtube cannot be combined with AXFA11
converter. In this case, use the AXFA14 converter.
4-7
IM 01E20C01-01E
4. WIRING
In case of explosion proof type for ATEX, FM, and
CSA certification, connect wiring as shown in the
figure below.
In case of the explosion proof type, the protective
grounding
of remote flowtube must be connected to
a suitable IS grounding system. In that case,
(functional grounding terminal) need not be connected.
I+ I–
CURRENT OUT
N/– L/+
POWER SUPPLY
A
B
C
Figure 4.4.6 Wiring Diagram
AL+ AL– C SA ABSB
SO1+COMSO2+
ALARM OUT
STATUS OUT
EX2EX1
P– SI1+ SI2+ COMP+
EXCIT ATION
PULSE OUT STATUS IN
Excitation cable
Remote flowtube
EX2
EX1
SIGNAL
AXFA11 Converter
AXFC Dedicated signal
cable
Converter
SA
A
B
SB
C
EX1
EX2
* Individually tape and insulate the
shields corresponding to SA and
SB on the remote flowtube side.
Remote
flowtube
Taping*
A
B
Taping*
C
EX1
EX2
F0416.EPS
(4) Connection with the Remote Flowtube
(General-Purpose Use, Submersible Type,
Size 1100 to 2600 mm (44 to 104 in.))
Connect wiring as shown in the figure below.
SO1+COMSO2+
I+
Shortbar
CURRENT OUT
4-20mA out
N/– L/+
POWER SUPPLY
I–
AL+ AL– C SA ABSB
STATUS OUT
ALARM OUT
EX2EX1
P– SI1+ SI2+ COMP+
EXCIT ATION
PULSE OUT STATUS IN
SIGNAL
AXFA11 converter
AXFC dedicated
signal cable
Excitation
cable
A
EX2
B
C
EX1
Converter
SA
A
B
SB
C
EX1
EX2
* Individually tape and insulate
the shields corresponding to SA
and SB on the remote flowtube
side.
Remote flowtube
Remote
flowtube
Taping*
A
B
Taping*
C
EX1
EX2
F040218.EPS
Figure 4.4.8 Wiring Diagram
4.4.7Connecting to External Instruments
(3) Connection with the Remote Flowtube
(General-Purpose Use, Submersible Type,
Size 500 to 1000 mm (20 to 40 in.))
Connect wiring as shown in the figure below.
SO1+COMSO2+
I+
I–
CURRENT OUT
4-20mA out
N/– L/+
POWER SUPPLY
Shortbar
A
B
C
Figure 4.4.7 Wiring Diagram
AL+ AL– C SA ABSB
STATUS OUT
ALARM OUT
EX2EX1
P– SI1+ SI2+ COMP+
EXCIT ATION
PULSE OUT STATUS IN
Excitation
cable
EX2
EX1
Remote flowtube
SIGNAL
AXFA11 converter
AXFC dedicated
signal cable
Converter
SA
A
B
SB
C
EX1
EX2
* Individually tape and insulate
the shields corresponding to SA
and SB on the remote flowtube
side.
Remote
flowtube
Taping*
Taping*
EX1
EX2
F040217.EPS
WARNING
Before wiring with external instrument, be sure to
turn off the power supply for AXFA11 converter
and any external instruments.
Connect the AXFA11 terminal to external instruments,
giving attention to the following points.
4 to 20 mA DC Current Output
A
B
C
AXFA11
l+
Receiver
l-
Instrument
Resistive Load Max. 1 kΩ.
(When using BRAIN/ HART communication,
more than 250 Ω, less than 600 Ω).
Figure 4.4.9 4 to 20 mA DC Output Connection
F0417.EPS
4-8
IM 01E20C01-01E
4. WIRING
Pulse Output
IMPORTANT
• As this is a transistor contact (insulated type),
give attention to proper voltage and polarity
when wiring.
• Do not apply a voltage larger than 30V DC or a
current larger than 0.2A in order to prevent
damage to the instrument.
• When input filter constant of the electronic
counter is large in relation to the pulse width,
the signal will decrease and the count will not
be accurate.
• If the input impedance of the electronic counter
is large, an induction noise from the power
supply may result in inaccurate counts. Use a
shield cable or sufficiently reduce the input
impedance of the electronic counter within the
electromagnetic flowmeter pulse output specification range.
• The active pulse output (Optional Code EM)
cannot be used in conjunction with the standard
pulse output.
• When the active pulse output (Optional Code
EM) is selected, do not be short-circuit between
the P+ and P– terminals to avoid damaging the
instrument.
• When the active pulse output (Optional code
EM) is selected, the range of pulse rate must
be set to 2 pps maximum.
• To avoid communication (BRAIN/ HART)
failure, it is recommended to use the shield
cable.
AXFA11
PULSE OUT
AXFA11
PULSE OUT
Output voltage: 24 V DC 20%
• Current: 150 mA or less
Pulse rate: 0.0001 to 2 pps
Pulse width: 20, 33, 50, 100 ms
Figure 4.4.11 Active Pulse Output Connection
P+
P-
P+
P-
(Optional code EM)
Load
Protective diode
Mechanical Counter
Electronic Counter
F0419.EPS
Status Input
IMPORTANT
Status inputs are designed for use with novoltage (dry) contacts. Be careful not to connect
the status to any signal source carrying voltage.
Applying voltage may damage the input circuit.
Closed: Less than 200 Ω
Open: More than 100 kΩ
+
SI1
+
SI2
AXFA11
AXFA11
PULSE OUT
AXFA11
PULSE OUT
Figure 4.4.10 Pulse Output Connection
P+
P-
P+
P-
Load
30V DC, 0.2A. max
Protective diode
Mechanical Counter
Electronic Counter
F0418.EPS
COM
No-voltage status input
Figure 4.4.12 Status Input Connection
4-9
F0420.EPS
IM 01E20C01-01E
Status Output / Alarm Output
IMPORTANT
Since this is an isolated transistor output, be
careful of voltage and polarity when wiring.
Do not apply a voltage larger than 30V DC or a
current larger than 0.2A in order to prevent
damage to the instrument.
This output cannot switch an AC load. To switch
an AC load, an intermediate relay must be
inserted as shown in Figure 4.4.13 or Figure
4.4.14.
*The alarm output operates from closed (normal)
to open (alarm occurrence) in the default value
(as setup upon plant shipment). Changes can
be made via the parameter settings.
AXFA11
AXFA11
+
AL
Load
-
AL
This connection is not possible.
+
AL
-
AL
External power supply
30V DC, 0.2A. max
Figure 4.4.14 Alarm Output Connection
4. WIRING
Protective diode
Relay
Electromagnetic
AC power supply
valve
F0422.EPS
AXFA11
+
SO1
Load
This connection is not possible.
Load
This connection is not possible.
AXFA11
+
SO2
COM
External power supply
30V DC, 0.2A. max
+
SO1
Protective
diode
+
SO2
COM
External power supply
30V DC, 0.2A. max
Figure 4.4.13 Status Output Connection
Protective diode
Relay
Electromagnetic
Electromagnetic
AC power supply
valve
valve
F0421.EPS
4-10
IM 01E20C01-01E
5. BASIC OPERATING PROCEDURES
5.
BASIC OPERATING PROCEDURES
The modification of data settings from the display unit can be carried out using the three
setting switches (infra-red switches) - namely, the , , and switches.
The infra-red switches enable the user to set parameters without opening the cover.
This chapter will provide a description of basic data configuration and the methods to be
used with the three setting switches. The AXFA11 can also be operated using a handheld
Brain Terminal (BT200) or a HART Communicator. (Please refer to Chapter 7 for operation via Brain Terminal and Chapter 8 for operation via HART Communicator.)
WARNING
Be sure to set parameters as "Protect" on the write protect function after finish of
parameter setting work.
Under extremely rare case, the infra-red switches may respond unexpectedly in
such conditions as sticking ball of water or extraneous substances on the surface
of display panel glass according to the principle of infra-red switch operation.
Its probability rises in such cases as sticking rain water by storm or other similar
situation and washing up work near flowmeter installation place.
Either to illuminate or stop illuminating the infra-red switches by the flashlight may
cause the mis-reaction.
Refer to Chapter 6 "Menu P: Parameter Protection Items" and section "10.2.3"
how to use the write protect function in detail.
(USING THE DISPLAY UNIT)
IMPORTANT
Operate the display unit under the condition where direct sunlight, etc... do not shine
to the setting switches directly when the parameter setting operation is carried out.
NOTE
(1) Always use the setting switches with the cover of the AXFA11 closed.
(2) Use these switches with them covered by the glass window.
(3) If dirt, dust or other substances surfaces on the display panel glass, wipe
them clean with a soft dry cloth.
(4) The operation with dirty gloves may cause a switch response error.
5-1
IM 01E20C01-01E
5. BASIC OPERATING PROCEDURES
5.1Operating Panel Configuration and Functions
(4)
(3) Display items
Reversed character display
(1) Data display area
: : :
0
(2) Setting switches
(Infra-red switches)
3274m
3
3
/h
m
100
FTL
FR
F0501.EPS
(1) Data display area
1st line (Display Select1), 2nd line (Display
Select2), and 3rd line (Display Select3) can be
displayed using parameter settings. The content
corresponding to selected item is shown with the
reversed-character on the right of the line.
(2) Setting switch operations
: Move the layer down, select, and confirm
+ : Move the layer up
(Press the switch while
holding down the switch)
: Move the cursor down (for selection-type
parameters) or increase values (for numeric-type parameters)
: Move the cursor to the right (for numeric-
type parameters)
(3) Display items
Displayed items and
reversed-character indication
Instantaneous flow rate: %
Actual instantaneous flow rate
Instantaneous flow rate: mA
Bar graph indicating
instantaneous flow rate
Totalized forward-direction flow rate
Totalized reverse-direction flow rate
Totalized differential flow rate
Tag number
Diagnosis of electrode adhesion
Communication type
+ : Move the cursor up (for selec-
tion-type parameters)
Content
Displays the instantaneous flow rate for the span as a percentage.
FR
Displays the actual reading for instantaneous flow rate.
FR
Displays the instantaneous flow rate for the span as a current output value.
FR
Displays the instantaneous flow rate for the span as a percentage
None
using bar graph.
Displays the totalized value for flow rate in the forward direction.
FTL
Displays the totalized value for flow rate in the reverse direction.
RTL
Displays the differential totalized value for flow rate between
DTL
forward totalization and reverse totalization.
Display the tag number (using up to 16 characters).
TA G
Displays the adhesion condition in the form of a bar graph.
ADH
(See the description for parameters K10 through K15 from
Chapter 6: Parameter Description for more details.)
Displays the communication type.
COM
(4)
Disp Select1
Disp Select2
Disp Select3
T0501.EPS
5-2
IM 01E20C01-01E
5.2Display Unit Setting Methods
NOTE
Before changing any settings, be sure to check the corresponding setting details
in Chapter 6: Parameter Description.
5.2.1Display Mode → Setting Mode
Display Mode will be adopted when the power is turned on, and the Setting Mode can be
activated using the following procedure.
Parameter Search Mode
(2 seconds or more)
Display Mode
Entry Mode
Select “Yes”
Major Item Parameter
Search Mode
5. BASIC OPERATING PROCEDURES
Sub-item Parameter
Search Mode
Parameter
Replacement Mode
Start:
Display Mode
(*1)
Entry Mode
Select “No”
* The term "mode" is used to refer to a situation
where display and setting are possible.
+
Press the SET switch while holding down the SHIFT switch.
++
Setting modes
F0502.EPS
Sample Display: Procedure for moving from Display Mode to Setting Mode
1st line: Actual instantaneous flow rate [FR]
2nd line: Bar graph indicating instantaneous flow
rate
3rd line: Totalized forward flow rate [FTL]
Setting switches
: : :
0
3274m
m
3
FTL
FR
3
/h
100
Hold the switch for 2 seconds.
Touch and switches simultaneously.
Touch SHIFT + key
NOTE
F0503-1.EPS
When other operations are carried out, the
system will automatically return to the Display
Mode.
When no operations in this display are carried
out for 20 seconds, the system will automatically return to the Display Mode.
(*1)
NOTE
The Major Item Parameter which is set just
before will be shown when entrying the Setting
Mode again within 1 minute after returning
from Setting Mode to Display Mode.
5-3
IM 01E20C01-01E
Setting Mode
No
Yes
Setting Mode
No
Yes
5. BASIC OPERATING PROCEDURES
A screen is displayed to confirm whether or not the
system is to enter Setting Mode.
Press the
The reversed-character (i.e. the cursor position)
indicates the item that is currently selected.
switch and select [Yes].
When [Yes] has been selected, touch the
switch.
NOTE
When the operations except and
switches are carried out, the system will
automatically return to the Display Mode.
When no operations in this display are carried
out for 20 seconds, the system will automatically return to the Display Mode.
Setting Mode
Setting Mode
Yes
Setting Mode
P:Protect
B:Easy Setup
C:Basic Setup
In order to request confirmation, the entire display
flashes on and off.
Touch the switch once again at this time to
fix your selection.
NOTE
When no operations in the flashed display are
carried out for 20 seconds, the system will
automatically return to the Display Mode.
When the operations except are carried
out, the system will automatically return to the
Display Mode.
The system enters Setting Mode.
Parameters to be set can be selected.
To Parameter
Search Mode
F0503-2.EPS
This completes the procedure for changing from the
Display Mode to the Parameter Search Mode.
5-4
IM 01E20C01-01E
5.2.2Setting Mode
When the Setting Mode has been activated using the procedure from Section 5.2.1, parameters can be selected for setting.
Format for Parameter Data
Depending on the type of parameter, data is formatted in one of the following three ways.
5. BASIC OPERATING PROCEDURES
NOTE
If no operations are carried out for a period of 10 minutes in Setting Mode, the
system will automatically return to the Display Mode.
FormatTypical displayContent
(i) Selection-type
(ii) Numeric-type
(iii) Alphanumeric-type
Regarding the alphanumeric-type format (iii), the following alphanumerics are displayed in the following sequence:
The desired data item is selected from a predefined list.
Data is set using the values in each digit and using
the decimal point.
Data is composed using alphanumeric characters
(in the case of tag numbers, special units, and the like).
With this format, setting can be carried out using up to
16 of the characters shown below.
T0502.EPS
5.3Parameter Setting Procedure
Once the system is in Setting Mode, the parameters for setting can be selected. On the
AXFA11, parameters that are frequently used have been grouped together in Easy Setup in
Menu B. This section provides a description of the parameter setting procedure using B:Easy Setup and C: Basic Setup.
For more details regarding parameter content, please refer to Section 6: Parameter Description.
5.3.1Setting Example for Selection-Type Data: Flow rate units
This example describes the setting of the flow rate units for the selection-type parameter
B21: Base Flow Unit from m3 to l (Liter).
Start:
Major Item
Parameter
Search Mode
Setting Mode
Major item
parameter
Setting Mode
P:Protect
B:Easy Setup
C:Basic Setup
F0504-1.EPS
Major Item Parameter Search Mode has been
accessed in this screen.
Touch the switch to access B: Easy Setup.
NOTE
The and symbols to the left of the param-
eters indicate that additional setting items to
those being currently displayed may also be
selected. Use the switch to cycle
through these items.
5-5
IM 01E20C01-01E
5. BASIC OPERATING PROCEDURES
Sub-item
Parameter
Search Mode
Parameter
Replacement Mode
Current unit setting
Available selections
Current unit setting
Available selections
Major item
Sub-item
parameter
Major item
Sub-item
selection (A)
Sub-item
Sub-item
B:Easy Setup
50:Auto Zero Exe
10:Language
20:Flow Damping
B:Easy Setup
20:Flow Damping
21:Base Flow Unit
22:Base Time Unit
B21:Base Flow Unit
3
m
3
m
kl(Kiloliter)
B21:Base Flow Unit
3
m
l(Liter)
3
cm
Sub-item Parameter Search Mode has been accessed
in this screen.
Touch the
X
2
B21: Base Flow Unit.
switch to move the cursor to
The cursor has been moved to B21: Base FlowUnit in this screen. (Sub-item selection screen (A))
In this screen, Parameter Replacement Mode has
been called up using the switch.
Touch the switch to move the cursor to the
unit item for selection. In this example, the
switch is touched twice to select l (Liter)
X
2
B21:Base Flow Unit
l(Liter)
When l (Liter) has been selected, touch the
switch.
In order to request confirmation, the entire display
flashes on and off.
Touch the switch once again at this time to
fix your selection.
NOTE
When no operations are carried out for 20
seconds in the flashing state, the system will
automatically return to the Sub-item Parameter
Search Mode.
When the operations except are carried
out, the parameter cannot be set.
Sub-item
Parameter
Search Mode
Major item
Sub-item
selection (A)
B:Easy Setup
20:Flow Damping
21:Base Flow Unit
22:Base Time Unit
F0504-2.EPS
The system returns automatically to sub-item
selection screen (A).
5-6
IM 01E20C01-01E
5. BASIC OPERATING PROCEDURES
5.3.2Setting Example for Numeric-Type Data: Flow rate span
This example describes the setting of the flow rate span for the numeric-type parameter
B23: Flow Span from 100 l/min to 120 l/min.
Start:
Major Item
Parameter
Search Mode
Sub-item
Parameter
Search Mode
Setting mode
Major item
parameter
Major item
Sub-item
parameter
Major item
Sub-item
selection (B)
Setting Mode
P:Protect
B:Easy Setup
C:Basic Setup
B:Easy Setup
50:Auto Zero Exe
10:Language
20:Flow Damping
X
4
B:Easy Setup
22:Base Time Unit
23:Flow Span
24:Flow Decimal Pnt
Selection of the appropriate parameter
Setting Mode Condition
Touch the switch to access B: Easy Setup.
NOTE
The and symbols to the left of the
parameters indicate that additional setting
items to those being currently displayed may
also be selected. Use the
switch to
cycle through these items.
Sub-item Parameter Search Mode has been accessed
in this screen.
Touch the switch to move the cursor to
B23: Flow Span.
The cursor has been moved to B23: Flow Span in
this screen. (Sub-item selection screen (B))
Touch the switch to access Parameter
Replacement Mode.
Parameter
Replacement Mode
B23:Flow Span
100 l/min
000100. l/min
Rng:0.00001 32000
B23:Flow Span
100 l/min
0000100. l/min
Rng:0.00001
32000
X
X
F0505-1.EPS
4
2
Once Parameter Replacement Mode has been
selected, the digit that can be replaced will be flashed
on and off. When in this condition, confirm the
relevant setting range as displayed at the bottom of
the screen and then set the parameter as required.
In this example, the parameter will be set to “120 l/
min”.
NOTE
When setting a new value, use the switch
to move from digit to digit, and use the
switch to cycle through values for each individual digit. In addition to digit, it is also possible
to select a decimal point, and this allows the
position of the decimal point to be changed.
Modify the value to “120 l/min” as follows:
Touch the switch to move the cursor to the
position for multiples of 10. Then, touch the
switch to change the value at this position from “0” to
“2”. When the value of “120” has been setup, touch
display flashes on and off. Confirm that the setting
has been correctly changed to “120”, and then fix
this value by touching the switch once again.
NOTE
When no operations are carried out for 20
seconds in the flashing state, the system will
automatically return to the Sub-item Parameter
Search Mode.
When the operations except are carried
out, the parameter cannot be set.
The system automatically returns to sub-item
selection screen (B).
NOTE
If the input value is outside the valid selection
range, the message “Out of range. Touch any
key.” will be displayed. In such a case, touch
any switch to return to Parameter Replacement
Mode and redo the setting.
Incorrect setting:
B23:Flow Span
100 l/min
0001.0. l/min
Rng:0.00001
Invalid value.
Touch any key.
32000
NOTE
If more than one decimal point has been input,
the message “Invalid value. Touch any key.”
will be displayed. In such a case, touch any
switch to return to Parameter Replacement
F0505-2.EPS
Mode and redo the setting.
5.3.3Setting Example for Alphanumeric-Type Data: Tag number
This example describes the setting of the tag number for the alphanumeric-type parameter
C10: Tag No. from “FI-1101” to “FI-1201.”
Start:
Major Item
Parameter
Search Mode
Setting mode
Major item
parameter
Setting Mode
P:Protect
B:Easy Setup
C:Basic Setup
F0506-1.EPS
Setting Mode Condition
Touch the switch to access C: Basic Setup.
NOTE
The and symbols to the left of the parameters indicate that additional setting items to
those being currently displayed may also be
selected. Use the switch to cycle through
these items.
5-8
IM 01E20C01-01E
5. BASIC OPERATING PROCEDURES
Sub-item
Parameter
Search Mode
Parameter
Replacement Mode
Current setting value
Modified value
Current setting value
Setting mode
Major item
parameter
Major item
Sub-item
selection (C)
Setting Mode
B:Easy Setup
C:Basic Setup
D:Total Set
C:Basic Setup
49:Flow User Span
10:Tag No
11:Flow Damping
Selection of the appropriate parameter
C10:Tag No
Fl-1101
l-1101
X
4
C10:Tag No
Fl-1101
Fl-1 01
The cursor has been moved to C: Basic Setup in
this screen.
Touch the
switch to enter C: Basic Setup.
Upon selection of C: Basic Setup, the cursor will
be positioned at C10: Tag No. (Sub-item selection
screen (C))
Touch the switch to access Parameter
Replacement Mode.
The cursor will flash on and off on the left of the tag
number. As “FI-1201” is to be setup in this example,
touch the switch to move the cursor to the
position for multiples of 100.
At the position for multiples of 100, touch the
switch to change the “1” to “2”. When the
setting has been changed to “FI-1201”, touch the
switch.
Sub-item
Parameter
Search Mode
Major item
Sub-item
selection (C)
C10:Tag No
Fl-1201
C:Basic Setup
49:Flow User Span
10:Tag No
11:Flow Damping
108.0
: : :
0
Fl-1201 TAG
FR
l/min
100
F0506-2.EPS
When the switch is touched, the entire
display flashes on and off. Confirm that the setting
has been correctly changed to “FI-1201”, and then
fix this setting by touching the switch once
again.
NOTE
When no operations are carried out for 20
seconds in the flashing state, the system will
automatically return to the Sub-item Parameter
Search Mode.
When the operations except are carried
out, the parameter cannot be set.
The system automatically returns to sub-item
selection screen (C).
After returning to Display Mode by touching using
the ( and ) twice as escape switches,
it will be possible to confirm the modified content (if
Display Select has been setup to display the tag
number.)
Indicates selection of Tag No display for B42: Display Select3.
5-9
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
NOTE
6.PARAMETER DESCRIPTION
6.1Parameters
With the exception of parameters that were specified
by the customer upon ordering, all of the AXFA11’s
internal parameters will initially be set to default
values. Actions such as the modification of display
details can then be carried out whenever necessary.
IMPORTANT
Make sure to keep the AXFA11’s power on at
least for 30 seconds after you set the parameters. If you turn the power off immediately after
the parameters are set, the settings will be
canceled.
6.2Parameter Lists
Parameter lists are comprised of the following items.
In order to ensure that correct flow rate data can
be acquired, it is crucial that the nominal size,
flow rate span, and meter factor of the combined
remote flowtube are setup. In cases where a
remote flowtube is ordered at the same time as
the AXFA11, the nominal size and meter factor
will be setup upon shipment from the manufacturing plant, and these will not require additional
setting. If the AXFA11 is ordered individually, the
default value will be setup for the meter factor;
accordingly, it will be necessary to change this
setting to the meter factor indicated on your
remote flowtube data plate.
If a flow rate span is specified upon ordering,
this will be set before shipment. If this is not the
case, however, it will be necessary for the
appropriate value to be set up by the user.
Item Name
R/W
Data
range
Units
Position of
decimal point
Default
value
Description
Describes the parameter content.
The default value (as setup upon plant shipment).
If marked “(*)”, it indicates that the appropriate
setting has been made in accordance with
ordering information and data for a remote
flowtube to be combined.
The position of the decimal point within the data
range and also the allowed relocation range in
terms of places of decimals.
The units corresponding to the data range.
This item defines the range of data to be set for
numeric parameters and also defines selection of
data for selection-type parameters.
R: Read only
W: Writable
The name of the parameter. Entries in parentheses
indicate the parameter name displayed on the
handheld BRAIN Terminal (BT200).
Parameter number.
T0601.EPS
6-1
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
6.3Parameter List Overview
(1) Item A (Menu A): Display items
Menu A contains the instantaneous flow rate, totalization values, and other items relevant to display.
Item
Display unit
(BRAIN)
A00
Display
(DISPLAY)
FR
A10
(FLOW RATE (%))
FR
A20
(FLOW RATE)
FR
A21
(FLW RATE (mA))
FTL
A30
(TOTAL)
RTL
A31
(REV TOTAL)
DTL
A32
(DIF TOTAL)
—
A60
(SELF CHECK)
Name
R/W
R
R
R
R
R
R
R
Data range
Display unit
/BRAIN
-110.0 to 110.0
-999999 to 999999
2.400 to 21.600
0 to 99999999
0 to 99999999
-99999999
to 99999999
Good
Error
Units
%
B21/B22
(C40/C41)
mA
B30
(D10)
B30
(D10)
B30
(D10)
Position
of decimal
point
1
0 to 3
3
0 to 7
0 to 7
0 to 7
Default value
(*): Indicated item
Description
For Display Mode only
For Display Mode only
For Display Mode only
For Display Mode only
For Display Mode only
For Display Mode only
See “6.5 Alarm Functions”.
T0602.EPS
(2) Item B (Menu B): Easy Setup items
Those parameters with a high frequency of use have been grouped together in Menu B. All basic functions can be
controlled using only the parameters from this block.
Name
Item
Display unit
(BRAIN)
B00 Easy Setup
(EASY SETUP)
B10 Language
(LANGUAGE)
B20 Flow Damping
(FLOW DAMPING)
B21 Base Flow Unit
(FLOW UNIT)
R/W
WEnglish
W0.1 to 200.0s13.0 sSets damping time. Linked with
WMl(Megaliter)
Data range
Display unit
/BRAIN
Japanese
French
German
Italian
Spanish
3
m
kl (Kiloliter)
l (Liter)
cm3
m
t
kg
g
kcf
cf
mcf
Mgal (US)
kgal (US)
gal (US)
mgal (US)
kbbl (US Oil)
bbl (US Oil)
mbbl (US Oil)
ubbl (US Oil)
kbbl (US Beer)
bbl (US Beer)
mbbl (US Beer)
ubbl (US Beer)
ft
klb (US)
lb (US)
Units
Position
of decimal
point
Default value
(*): Indicated item
EnglishSelects the language used for the
display unit. Linked with H30.
C11.
m (*)Selects flow units for the flow
rate span. Linked with C40.
Description
T0603-1.EPS
6-2
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
Name
Item
Display unit
(BRAIN)
B22 Base Time Unit
(TIME UNIT)
B23 Flow Span
(FLOW SPAN)
B24 Flow Decimal Pnt
(FLOW DECIMAL)
B30 Total Unit
(TOTAL UNIT)
B31 Total Scale
(TOTAL SCALE)
B32 Pulse Unit
(PULSE UNIT)
B33 Pulse Scale
(PULSE SCALE)
B40 Display Select1
(DISP SELECT1)
B41 Display Select2
(DISP SELECT2)
B42 Display Select3
(DISP SELECT3)
B50 Auto Zero Exe
(AUTOZERO EXE)
B60 —
(SELF CHECK)
Data range
R/W
Display unit
Units
/BRAIN
W/d
/h
/min
/s
W0.0001 to 32000B21
/B22
(C40/C41)
WAuto
0
1
2
3
Wn Unit/P
u Unit/P
m Unit/P
Unit/P
k Unit/P
M Unit/P
Pulse/s
W0 to 32000B30
(D10)
Wn Unit/P
u Unit/P
m Unit/P
Unit/P
k Unit/P
M Unit/P
Pulse/s
W0 to 32000B32
(E10)
WFlow Rate(%)
Flow Rate
Flow Rate(mA)
Forward Total
Reverse Total
Dif Total
WOff
Flow Rate(%)
Flow Rate
Flow Rate(mA)
Flow Rate(Bar)
Forward Total
Reverse Total
Dif Total
Tag No
Adhesion Check
Communication
WSame as B41
(Display Select2)
WNo Execution
Execution
RGood
Error
Position
of decimal
point
Default value
(*): Indicated item
Description
/s (*)Selects time units for the flow
rate span. Linked with C41.
0 to 41 m/s (*)Sets flow rate span (with units
from B21 and B22). Linked with
C42.
Auto (*)Selects decimal point position
for the display unit's
instantaneous flow rate. Linked
with C43.
Pulse/s (*)Selects the flow rate unit per one
pulse as used for totalization
display. Linked with D10.
0 to 40 (*)Sets the flow rate per one pulse
for the totalization display.
Linked with D11.
Pulse/s (*)Selects the flow rate unit per one
pulse as used for pulse output.
Linked with E10.
0 to 40 (*)Sets the flow rate per one pulse
as used for pulse output.
Linked with E11.
Flow RateSelects content of the first line
for Display Mode. Linked with
H10.
OffSelects content of the second line
for Display Mode. Linked with
H11.
OffSelects content of the third line
for Display Mode. Linked with
H12.
No ExecutionSelects whether or not automatic
zero adjustment is carried out.
Linked with M10.
See “6.5 Alarm Functions”.
T0603-2.EPS
6-3
IM 01E20C01-01E
(3) Item C (Menu C): Basic Setting items
Menu C principally contains the basic setting items for the flowtube.
6. PARAMETER DESCRIPTION
Name
Item
Display unit
(BRAIN)
C00
Basic Setup
(BASIC SETUP)
C10
Tag No
(TAG NO)
C11 Flow Damping
(FLOW DAMPING)
C20 Measure Mode
(MEASURE MODE)
C21 Low MF
(LOW MF)
(HIGH MF)
(LOW MF(EDF))
(HIGH MF(EDF))
C30 Select Flow Tube
(FLOW TUBE)
C31 Nominal Size Unit
(SIZE UNIT)
(NOMINAL SIZE)
C40 Base Flow Unit
(FLOW UNIT)
C41 Base Time Unit
(TIME UNIT)
(FLOW SPAN)
C43 Flow Decimal Pnt
(FLOW DECIMAL)
R/W
Data range
Display unit
/BRAIN
ASCII 16 charactersW
Units
Position
of decimal
point
Default value
(*): Indicated item
Description
Sets Tag number up to 16
characters.
0.1 to 200.0Ws
143.0 s
Sets damping time. Linked with
B20.
Standard DF
WStandard DF
Enhanced DF
0.0100 to 3.0000W1.0000 (*)
Selects measurement mode for
dual frequency excitation
Sets low-frequency meter factor for
standard dual frequency excitation
0.0100 to 3.0000W1.0000 (*)
4C22 High MF
Sets high-frequency meter factor for
standard dual frequency excitation
0.0000 to 3.0000W1.0000 (*)
4C23 Low MF(EDF)
Sets low-frequency meter factor for
enhanced dual frequency excitation
0.0000 to 3.0000W1.0000 (*)
4C24 High MF(EDF)
Sets high-frequency meter factor for
enhanced dual frequency excitation
W
ADMAG AXF
ADMAG
ADMAG AE
ADMAG SE
ADMAG AXF
Selects the flowtube's model
name.
Note: Select “ADMAG AXF” for
ADMAG AXW flowtube.
YEWMAG
Calibrator
Other
mm
Wmm
inch
0.99 to 3000.1
W100 (*)
0.01 to 120.1
Wm (*)
Ml(Megaliter)
3
m
mm
inch
inch
0 to 2C32 Nominal Size
Selects the nominal size units for
the flowtube.
Sets flowtube nominal size in
selected unit at C31.
Selects flow units for the flow
rate span. Linked with B21.
kl(Kiloliter)
l(Liter)
3
cm
m
t
kg
g
kcf
cf
mcf
Mgal (US)
kgal (US)
gal (US)
mgal (US)
kbbl (US Oil)
bbl (US Oil)
mbbl (US Oil)
ubbl (US Oil)
kbbl (US Beer)
bbl (US Beer)
mbbl (US Beer)
ubbl (US Beer)
ft
klb (US)
lb (US)
/d
W/s (*)
/h
Selects time units for the flow
rate span. Linked with B22.
/min
/s
0.0001 to 32000W1 m/s (*)
C40/C41
(B21
/B22)
Auto
WAuto (*)
0
1
0 to 4C42 Flow Span
Sets flow rate span (with units
from C40 and C41). Linked with
B23.
Selects decimal point position for
the display unit's instantaneous
flow rate. Linked with B24.
2
3
T0604-1.EPS
6-4
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
Name
Item
Display unit
(BRAIN)
(VELOCITY CHK)
C45 Density Unit
(DENSITY UNIT)
R/W
Wkg/m
Data range
Display unit
/BRAIN
0.000 to 99.999R
3
kg/m
lb/gal
Units
m/s3C44 Velocity Check
Position
of decimal
point
Default value
(*): Indicated item
3
Description
Display of the span setting using
flow velocity (m/s).
Sets units for density when mass
flow rate is selected.
lb/cf
(MASS DENSITY)
C47 User Span Select
(USER SPN SEL)
C48 Flow User Unit
(FL USER UNIT)
(FL USER SPAN)
—
C60
(SELF CHECK)
0 to 32000W0
No
WNo
C450 to 4C46 Mass Flow Density
Yes
W
characters
0.0001 to 32000W100
C480 to 4C49 Flow User Span
R
Error
Sets density when mass flow rate
is selected (with units from C45).
Selects whether or not special
units are used for flow rate units.
Sets the special flow rate units.8 alphanumeric
Sets span when using special flow
rate units.
See “6.5 Alarm Functions”.Good
T0604-2.EPS
(4) Item D (Menu D): Total Setting items
Menu D contains setting items such as the totalization scale and the forward/reverse totalized values.
Name
Item
Display unit
(BRAIN)
D00 Total Set
(TOTAL SET)
D10 Total Unit
(TOTAL UNIT)
D11 Total Scale
(TOTAL SCALE)
D12 Total Decimal Pnt
(TL DECIMAL)
D13 Total Low Cut
(TOTAL LOWCUT)
D20 Total Execution
(TOTAL EXEC)
D21 Ttl Set Val Lower
(TL SET VAL L)
D22 Ttl Set Val Upper
(TL SET VAL U)
D23 Ttl Switch Lower
(TL SWITCH LO)
D24 Ttl Switch Upper
(TL SWITCH UP)
D30 Ttl User Select
(TL USER SEL)
D31 Ttl User Unit
(TL USER UNIT)
D60 —
(SELF CHECK)
R/W
Wn Unit/P
W0 to 32000D10
W0
W0 to 100%03 %
WStart
W0 to 99999900
W0 to 9900
W0 to 99999900
W0 to 9900
WNo
W8 alphanumeric
RGood
Data range
Display unit
/BRAIN
u Unit/P
m Unit/P
Unit/P
k Unit/P
M Unit/P
Pulse/s
1
2
3
4
5
6
7
Stop
Preset Total
Preset Rev Total
Yes
characters
Error
Units
(B30)
Position
of decimal
point
Default value
(*): Indicated item
Description
Pulse/s (*)Selects the flow rate unit per one
pulse as used for totalization
display. Linked with B30.
0 to 40 (*)Sets the flow rate per one pulse
for the totalization display.
Linked with B31.
0Selects position of decimal point
for totalization display
Sets the range in vicinity of 0%
within which the totalization
display will be halted.
Start
Executes “Start” or “Stop” of the
totalization function, or executes
“Preset Total” or “Preset Rev
Total”.
Sets the totalization preset value
in the lower 6 digits of the 8-digit
totalized value.
Sets the totalization preset value
in the upper 2 digits of the 8-digit
totalized value.
Sets the totalization switch value
in the lower 6 digits of the 8-digit
totalized value.
Sets the totalization switch value
in the upper 2 digits of the 8digits totalized value.
No
Selects whether or not special
units are used as totalized units.
Sets the special totalized units.
See “6.5 Alarm Functions”.
6-5
T0605.EPS
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
(5) Item E (Menu E): Pulse Setting items
Menu E contains items relevant to pulse output. This is used to set parameters such as the pulse scale and width.
Name
Item
Display unit
(BRAIN)
E00 Pulse Set
(PULSE SET)
E10 Pulse Unit
(PULSE UNIT)
E11 Pulse Scale
(PULSE SCALE)
E12 Pulse Width
(PULSE WIDTH)
E13 Pulse Low Cut
(PULSE LOWCUT)
E20 Pulse Active Mode
(PLS ACT MODE)
E60 —
(SELF CHECK)
R/W
Wn Unit/P
W0 to 32000E10
W50% Duty
W0 to 100%03 %Sets the range in vicinity of 0%
WClosed(On)
RGood
Data range
Display unit
/BRAIN
u Unit/P
m Unit/P
Unit/P
k Unit/P
M Unit/P
Pulse/s
0.05 ms
0.1 ms
0.5 ms
1 ms
20 ms
33 ms
50 ms
100 ms
Open(Off) Act
Error
Act
Units
(B32)
Position
of decimal
point
0 to 40 (*)Sets the flow rate per one pulse
Default value
(*): Indicated item
Pulse/s (*)Selects the flow rate unit per one
50% DutySelects the pulse width for pulse
Closed(On)
pulse as used for pulse output.
Linked with B32.
as used for pulse output.
Linked with B33.
output.
within which pulse output will be
halted.
Act
Selects whether pulse output will
be set to “On Active” or “Off
Active.”
See “6.5 Alarm Function”.
(6) Item F (Menu F): Status Functions Setting items
Menu F contains items relevant to multiplex range output and other status Input/Output.
Description
T0606.EPS
Name
Item
Display unit
(BRAIN)
F00 Status Function
(STATUS FUNC)
F10 SO1 Function
(SO1 FUNCTION)
F10 SO2 Function
(SO2 FUNCTION)
Data range
R/W
WNo Function
WNo Function
Display unit
/BRAIN
Warning Output
Total Switch
H/L Alarm
HH/LL Alarm
Fwd/Rev Ranges
Auto 2 Ranges
Auto 3 Ranges
Auto 4 Ranges
Ext 2 Answer
Ext 3 Answer
Ext 4 Answer
Warning Output
Total Switch
H/L Alarm
HH/LL Alarm
Fwd/Rev Ranges
Auto 2 Ranges
Auto 3 Ranges
Auto 4 Ranges
Ext 2 Answer
Ext 3 Answer
Ext 4 Answer
Units
Position
of decimal
point
Default value
(*): Indicated item
No FunctionSelects function for the SO1
terminal
No FunctionSelects function for the SO2
terminal
Description
T0607-1.EPS
6-6
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
Name
Item
Display unit
(BRAIN)
F12 SI1 Function
(SI1 FUNCTION)
R/W
WNo Function
Data range
Display unit
/BRAIN
0% Singal Lock
Units
Position
of decimal
point
Default value
(*): Indicated item
No FunctionSelects function for the SI1
terminal
Ext Auto Zero
Ext Ttl Preset
Ext Rev Ttl Set
Ext 2 Ranges
Ext 3 Ranges
Ext 4 Ranges
F13 SI2 Function
(SI2 FUNCTION)
WNo Function
0% Singal Lock
No FunctionSelects function for the SI2
terminal
Ext Auto Zero
Ext Ttl Preset
Ext Rev Ttl Set
Ext 2 Ranges
Ext 3 Ranges
Ext 4 Ranges
F14
SO1/2 Active Mode
(SO ACT MODE)
WClosed(On)
Open(Off) Act
Act
Closed(On)
Act
Selects whether SO1/SO2 output
will be set to “On Active” or
“Off Active.”
F15 SI1/2 Active Mode
(SI ACT MODE)
WShort Active
Open Active
Short ActiveSelects whether SI1/SI2 input
will be set to “Short Active” or
“Open Active.”
F30 Forward Span2
(FWD SPAN2)
F31 Forward Span3
(FWD SPAN3)
F32 Forward Span4
(FWD SPAN4)
F33 Reverse Span1
(REV SPAN1)
F34 Reverse Span2
(REV SPAN2)
F35 Reverse Span3
(REV SPAN3)
F36 Reverse Span4
(REV SPAN4)
F40 Auto Range Hys
(AUTO RNG HYS)
F41 Bi Direction Hys
(BI DIREC HYS)
F60 —
(SELF CHECK)
WC40
0 to 40.0001 to 320001
/C41
WC40
0 to 40.0001 to 320001
/C41
WC40
0 to 40.0001 to 320001Sets flow rate span for forward
/C41
WC40
0 to 40.0001 to 320001Sets flow rate span for reverse
/C41
WC40
0 to 40.0001 to 320001
/C41
WC40
0 to 40.0001 to 320001
/C41
WC40
0 to 40.0001 to 320001Sets flow rate span for reverse
/C41
W%00 to 1510 %Sets hysteresis width for
W%00 to 82 %
RGood
Error
Sets flow rate span for forward
No. 2 range
Sets flow rate span for forward
No. 3 range
No. 4 range
No. 1 range
Sets flow rate span for reverse
No. 2 range
Sets flow rate span for reverse
No. 3 range
No. 4 range
automatic range switching
Sets hysteresis width for
forward/reverse switching
See “6.5 Alarm Function”.
(7) Item G (Menu G): Alarm Setting items
Menu G contains setting items relevant to alarm output, burnout, alarm record, etc.
Description
T0607-2.EPS
Name
Item
Display unit
(BRAIN)
G00 Alarm
(ALARM)
G10 Low Alarm
(LOW ALARM)
G11 High Alarm
(HIGH ALARM)
G12 Low Low Alarm
(LO LO ALARM)
G13 High High Alarm
(HI HI ALARM)
R/W
Data range
Display unit
/BRAIN
Units
Position
of decimal
point
Default value
(*): Indicated item
Description
W%0-110 to 110-110Sets level setting value for low
flow rate limit (L)
W%0-110 to 110110Sets level setting value for high
flow rate limit (H)
W%0-110 to 110-110Sets level setting value for low-
low flow rate limit (LL)
W%0-110 to 110110Sets level setting value for high-
high flow rate limit (HH)
T0608-1.EPS
6-7
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
Name
Item
Display unit
(BRAIN)
G14 H/L Alarm Hys
(H/L ALM HYS)
G20 Alm Out Act Mode
(
ALM OUT ACT)
G21 4-20mA Alarm Out
(4-20 ALM OUT)
G22 4-20mA Burn Out
(4-20 BURNOUT)
G30 Alm-Setting
(ALM-SETTING)
G31 Alm-Sig Over
(ALM-SIG OVER)
G32 Alm-Emp Pipe
(ALM-EMP PIPE)
G33 Alm-HH/LL
(ALM-HH/LL)
G34 Alm-Adhesion
(ALM-ADHESION)
G40 Operation Time
(OPERATE TIME)
G41 Alm Record1
(ALM RECORD1)
G42 Alm Record Time1
(ALM TIME 1)
G43 Alm Record2
(ALM RECORD2)
G44 Alm Record Time2
(ALM TIME 2)
G45 Alm Record3
(ALM RECORD3)
G46 Alm Record Time3
(ALM TIME 3)
G47 Alm Record4
(ALM RECORD4)
G48 Alm Record Time4
(ALM TIME 4)
G60 —
(SELF CHECK)
R/W
Data range
Display unit
/BRAIN
Units
Position
of decimal
point
Default value
(*): Indicated item
Description
W%00 to 105 %Sets hysteresis width for high-low
flow rate limit alarm
WClosed(On) Act
Open(Off) Act
W2.4mA or Less
4.0mA
Open(Off)
Act
21.6mA or
More
Selects whether alarm output will
be set to “On Active” or “Off Active.”
Selects the current output during
alarm occurrence.
Hold
21.6mA or More
RHigh
Low
WNo
Yes
WNo
Yes
WNo
Yes
WNo
Yes
—
Yes
Yes
Yes
No
Displays the current output
during a CPU failure.
Selects whether a setting alarm is
to be specified as an alarm.
Selects whether a signal overflow
alarm is to be specified as an alarm.
Selects whether an empty pipe alarm
is to be specified as an alarm.
Selects whether a flow rate
high-high or low-low alarm is
to be specified as an alarm.
WNo
Yes
R0D 00:00 to
No
Selects whether an electrode adhesion
alarm is to be specified as an alarm.
Operation time
99999D 23:59
R
10:uP Fault
Displays the content of the most
recent alarm.
11:EEPROM Fault
12:A/D(H) Fault
13:A/D(L) Fault
14:A/D(Z) Fault
15:Coil Open
16:
Displays the operation time at the
occurrence of the most recent alarm.
Displays the content of the second
most recent alarm.
Displays the operation time at occurrence
of the second most recent alarm.
Displays the content of the third most
recent alarm.
Displays the operation time at the
occurrence of the third most recent alarm.
Displays the content of the fourth
most recent alarm.
Displays the operation time at the
occurrence of the fourth most recent alarm.
See “6.5 Alarm Functions”.
Error
T0608-2.EPS
6-8
IM 01E20C01-01E
(8) Item H (Menu H): Display Setting items
Menu H contains setting items that are relevant to display on the display unit.
6. PARAMETER DESCRIPTION
Name
Item
Display unit
(BRAIN)
H00 Display Set
(DISP SET)
H10 Display Select1
(DISP SELECT1)
H11 Display Select2
(DISP SELECT2)
H12 Display Select3
(DISP SELECT3)
H20 Display Cycle
(DISP CYCLE)
H30 Language
(LANGUAGE)
H60 —
(SELF CHECK)
R/W
WFlow Rate(%)
WOff
WSame as H11
W200ms
WEnglish
RGood
Data range
Display unit
/BRAIN
Flow Rate
Flow Rate(mA)
Forward Total
Reverse Total
Dif Total
Flow Rate(%)
Flow Rate
Flow Rate(mA)
Flow Rate(Bar)
Forward Total
Reverse Total
Dif Total
Tag No
Adhesion Check
Communication
(Display Select2)
400ms
1s
2s
4s
8s
Japanese
French
German
Italian
Spanish
Error
Units
Position
of decimal
point
Default value
(*): Indicated item
Flow RateSelects content of the first line
for Display Mode. Linked with
B40.
OffSelects content of the second line
for Display Mode. Linked with
B41.
OffSelects content of the third line
for Display Mode. Linked with
B42.
400msSelects the display cycle.
EnglishSelects the language used by the
display unit. Linked with B10.
See “6.5 Alarm Functions”.
(9) Item J (Menu J): Auxiliary Function Setting items
Menu J contains setting items such as the flow direction, rate limits, and low cut.
Name
Item
Display unit
(BRAIN)
J00 Aux
(AUX)
J10 4-20mA Low Cut
(4-20 LOW CUT)
J11 4-20mA Low Lmt
(4-20 LOW LMT)
J12 4-20mA High Lmt
(4-20 HI LMT)
J20 Flow Direction
(FLOW DIRECT)
J21 Rate Limit
(RATE LIMIT)
J22 Dead Time
(DEAD TIME)
J23 Pulsing Flow
(PULSING FLOW)
J24 T/P Damp Select
(T/P DAMP SEL)
R/W
W%0 to 1000%
W-20.0 to 100.0%1-20.0%Sets the low limit for current
W0.0 to 120.0%1120.0%Sets the high limit for current
WForward
W0 to10%05%Sets the level to reduce output
Ws0 to 1500s
WNo
WDamping
Data range
Display unit
/BRAIN
Reverse
Yes
No Damping
Units
Position
of decimal
point
Default value
(*): Indicated item
Sets the range in vicinity of 0%
within which the current output
will be 4 mA
output
output
ForwardSelects the flow direction.
fluctuation.
Sets the dead time to reduce output
fluctuation. When “0” is set, rate
limit function is not available.
NoSelects whether pulsing flow is to
Damping
be supported.
Selects whether the flow rate value
obtained through damping calculation
for total/pulse or the instantaneous
flow rate value (no damping) for
total/pulse is to be used.
Description
T0609.EPS
Description
T0610-1.EPS
6-9
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
Name
Item
Display unit
(BRAIN)
J30 Power Synch
(POWER SYNCH)
J31 Power Frequency
(POWER FREQ)
J40 Memo 1
(MEMO 1)
J41 Memo 2
(MEMO 2)
J42 Memo 3
(MEMO 3)
J50 Software Rev No
(SOFTWARE REV)
J60 —
(SELF CHECK)
R/W
WNo
R/WHz47.00 to 63.00250.00
W
W
W
R—Software revision number
RGood
Data range
Display unit
/BRAIN
Yes
ASCII 16 characters
ASCII 16 characters
ASCII 16 characters
Error
Units
Position
of decimal
point
Default value
(*): Indicated item
YesSelects whether or not the internal
frequency is to be synchronized
with the power supply frequency.
Displays the power-supply
frequency (for Power Synch =
“Yes”), or sets the power-supply
frequency (for Power Synch=“No”).
Memo field
Memo field
Memo field
See “6.5 Alarm Functions”.
Description
(10) Item K (Menu K): Diagnostic Function Setting items
Menu K contains items that are relevant to the diagnosis of insulation adhesion to the electrode.
Name
Item
Display unit
(BRAIN)
K00 Diagnosis
(DIAGNOSIS)
K10 Adhesion Check
(ADHESION CHK)
K11 Adhesion Level1
(ADH LEVEL1)
K12 Adhesion Level2
(ADH LEVEL2)
K13 Adhesion Level3
(ADH LEVEL3)
K14 Adhesion Level4
(ADH LEVEL4)
K15
Adh Measure Value
(ADH MEAS VAL)
K60 —
(SELF CHECK)
R/W
WNo
WM ohm0.00 to 100.0020.10
WM ohm0.00 to 100.0020.50
WM ohm0.00 to 100.0021.00
WM ohm0.00 to 100.0023.00
RM ohm—2
RGood
Data range
Display unit
/BRAIN
Yes
Error
Units
Position
of decimal
point
Default value
(*): Indicated item
No
Description
Selects whether or not to perform diagnosis
of adhesion to the electrode.
Sets the resistance value for adhesion
Level 1 to the electorode.
Sets the resistance value for adhesion
Level 2 to the electorode.
Sets the resistance value for adhesion
Level 3 to the electorode.
Sets the resistance value for adhesion
Level 4 to the electorode.
Displays the resistance value for
adhesion to the electrode.
See “6.5 Alarm Functions”.
(11) Item M (Menu M): Automatic Zero-Adjustment Function Setting items
Menu M contains items that are relevant to automatic zero adjustment.
T0610-2.EPS
T0611.EPS
Name
Item
Display unit
(BRAIN)
M00 Adjustment
(ADJUSTMENT)
M10 Auto Zero Exe
(AUTOZERO EXE)
M11 Magflow Zero
(MAGFLOW ZERO)
—
M60
(SELF CHECK)
Data range
R/W
Display unit
/BRAIN
WNo Execution
Units
Position
of decimal
point
Default value
(*): Indicated item
No ExecutionSelects whether or not automatic
Execution
R/W-99.999 to 99.99930.000
RGood
Error
6-10
Description
zero adjustment is carried out.
Linked with B50.
Displays the result of the automatic
zero adjustment, or sets the zero point.
See “6.5 Alarm Functions”.
T0612.EPS
IM 01E20C01-01E
(12) Item N (Menu N): Loop Test Setting items
Menu N contains items that are relevant to the execution of loop testing.
6. PARAMETER DESCRIPTION
Name
Item
Display unit
(BRAIN)
N00 Test
(TEST)
N10 Test Mode
(TEST MODE)
N11 Test Output Value
(TEST OUT VAL)
N20 Test SO1
(TEST SO1)
N21 Test SO2
(TEST SO2)
N22 Test Alarm Out
(TEST ALM OUT)
N23 Test SI1
(TEST SI1)
N24 Test SI2
(TEST SI2)
N60 —
(SELF CHECK)
R/W
WNormal
W%-10 to 11000%Sets the test output value.
WOpen(Off)
WOpen(Off)
WOpen(Off)
ROpen
ROpen
RGood
Data range
Display unit
/BRAIN
Test
Closed(On)
Closed(On)
Closed(On)
Short
Short
Error
Units
Position
of decimal
point
Default value
(*): Indicated item
NormalSelects whether mode will be set
Open(Off)Selects the test condition for
Open(Off)Selects the test condition for
Closed(On)Selects the test condition for
(13) Item P (Menu P): Parameter Protection items
Menu P contains items that are relevant to write protection and passwords.
Name
Item
Display unit
(BRAIN)
P00 Protect
(PROTECT)
P10 Key Code
(KEY CODE)
P20 Write Protect
(W PROTECT)
P21 Enable Wrt Passwd
(ENABLE WRITE)
P22 New Password
(NEW PASSWORD)
P23 Software Seal
(SOFT SEAL)
P60 —
(SELF CHECK)
R/W
W0 to 99990Parameter of the display restriction
RNo
WASCII 8 charactersSets the correct password so that
WASCII 8 characters
RBreak
RGood
Data range
Display unit
/BRAIN
Yes
Keep
Error
Units
Position
of decimal
point
Default value
(*): Indicated item
No
Keep
Description
to “Normal” or “Test”.
SO1 terminal.
SO2 terminal.
alarm output terminal.
Displays the test condition for
SI1 terminal.
Displays the test condition for
SI2 terminal.
See “6.5 Alarm Functions”.
T0613.EPS
Description
Displays whether or not overwriting
of parameter data is prohibited.
write protection function will be
released.
Sets the password for write protection
funcion
Displays whether or not a Joker
password was used (Break).
See “6.5 Alarm Functions”.
T0613-1.EPS
6-11
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
NOTE
NOTE
6.4Parameter Description
(1) Menu B: Easy Setup items
Those parameters with a high frequency of use have
been grouped together in Easy Setup. All basic
functions can be controlled using only the parameters
from this block. Parameters from Menu B share
identical names with those from other menus; however,
modification of one such parameter will result in the
other being automatically modified.
[B10: Language] Selection of language used for the
display unit
→ This setting is linked with that of parameter H30.
One of the following languages can be selected for the
display unit.
Data Range
Setting itemDescription
English
Japanese
French
German
Italian
Spanish
[B20: Flow Damping] Setting of the damping time
constant
→ This setting is linked with that of parameter C11.
The damping time constant should be modified to
suppress an output fluctuation or to change the response time. This time constant has an effect on analog
output and on the flow rate display (i.e., actual instantaneous flow rate, %, current value), and in addition, it
also affects pulse output and totalization. However,
when “No Damping” has been set for J24: T/P DampSelect, there will be no effect on pulse output or
totalization.
* Time constant: The time required for the output to
reach 63.2% from 0%.
[B21: Base Flow Unit] Selection of flow units for the
flow rate span
→ This setting is linked with that of parameter C40.
This parameter selects the flow units for the flow rate
span. (In case of mass flow, the setting of density is
also required. Refer to C46: Mass Flow Density
for more details.)
All parameters, alarm messages, etc. displayed in
English.
All parameters, alarm messages, etc. displayed in
Japanese katakana.
All parameters, alarm messages, etc. displayed in
French.
All parameters, alarm messages, etc. displayed in
German.
All parameters, alarm messages, etc. displayed in
Italian.
All parameters, alarm messages, etc. displayed in
Spanish.
T0614.EPS
[B22: Base Time Unit] Selection of time units for
the flow rate span
→ This setting is linked with that of parameter C41.
This parameter selects the time units for the flow rate
span; however, if “m” or “ft” has been selected for the
flow rate units, “/s” is automatically set for this
parameter.
[B23: Flow Span] Setting of the flow rate span
→ This setting is linked with that of parameter C42.
The span can be set for the forward flow rate in the
range 0 to 32,000 (although this does not include 0). The
units set using B21/C40: Base Flow Unit and B22/C41: Base Time Unit will be displayed at this time.
If the flow rate units, time units, and flow rate
span are specified upon ordering, these parameters will be setup before shipment; however, if
this is not the case, it will be necessary for the
appropriate values to be set up by the user.
Flow rate span is the value for instantaneous
flow rate that corresponds to a current output of
20 mA. The following factors should be taken
into consideration when deciding on the flow rate
span.
• In the case of applications with large variations
in flow rate, the maximum flow rate should be
set. If a flow rate in excess of the flow rate
span was to occur, output would be possible
up to an upper limit of 108%, and beyond this,
error would occur. Note that the same applies
to pulse output and totalization.
• In the case of applications that have a relatively stable flow rate, a flow rate span of 1.5 to
2.0 times larger than the normal flow rate may
be considered suitable.
• The flow rate to be adopted should - upon
conversion to flow velocity - correspond to a
value within the range of 0.1 to 10 m/s. (0.3 to
10 m/s for size 1100 mm or larger) The flow
velocity can be confirmed using sizing data or
with parameter C44: Velocity Check, and in
the latter case, the value obtained when span is
converted to flow velocity will be displayed.
• Regardless of the position of the decimal point,
the largest value that can be set on the display
unit is 32,000. Furthermore, it is not possible to
set a number of 4 or greater for the highestorder digit. Similarly, if 3 is set for this highestorder digit, it will not be possible to set a
6-12
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
NOTE
NOTE
number of 2 or greater for the next digit to the
right, regardless of the position of the decimal
point.
Example: A value of 333.33 is represented by
the character string 33333, and since
this exceeds 32000, it cannot be set.
In such a case, the value 333.3
should be set instead.
[B24: Flow Decimal Pnt] Setting of the decimal
point position for the instantaneous flow rate
→ This setting is linked with that of parameter C43.
This parameter sets the position of the decimal point
for instantaneous flow rate values in terms of the
number of digits. When set using “Auto”, the decimal
point position will be automatically determined in
accordance with the setting value for B23/C42: FlowSpan as shown below.
Flow Span 9Decimal point position: 3 digits
9 < Flow Span 90Decimal point position: 2 digits
90 < Flow Span 900Decimal point position: 1 digit
900 < Flow SpanDecimal point position: no
digits (i.e., no decimal point)
ItemDescription
-
n Unit/P
u Unit/P
m Unit/P
Unit/P
k Unit/P
M Unit/P
Pulse/s
FU: Flow rate unit selected in B21/C40: Base Flow Unit.
9
10
FU
-
6
10
FU
-
3
10
FU
FU
3
10
FU
6
10
FU
Number of pulses to be counted for one second at 100% output.
T0616.EPS
[B31: Total Scale] Setting of the totalization scale
→ This setting is linked with that of parameter D11.
The flow rate is totalized in individual counts in
accordance with this parameter’s setting. If 0 is
selected, it indicates that the totalization function is not
to be used.
If a totalization scale is specified upon ordering,
this parameter is set up before shipment; however, if this is not the case, it will be necessary
for the appropriate value to be set up by the
user.
When an item other than “Auto” is set, the selected
number of digits for the decimal point position is used.
With the decimal point removed, 6 digits are available
for the instantaneous flow rate value, and display is
possible up to 999,999. If an overflow occurs as a
result of the setting adopted for decimal point position,
the warning 84: Disp Over Wng will be displayed to
provide notification of this condition.
Example: When 1000 m3/h is set for B23/C42:
Flow Span
Item
Auto
0
1
2
3
Display content for instantaneous flow rate value
1000 m3/h
3
1000 m
/h
3
1000.0 m
1000.00 m3/h
With the decimal point removed, 7 digits are not
available for the instantaneous flow rate value;
therefore, a warning is displayed.
/h
T0615.EPS
[B30: Total Unit] Setting of units for totalization
scale
→ This setting is linked with that of parameter D10.
This parameter selects the flow rate units for use in
totalization.
• By setting the totalization scale, the totalized
value is displayed on the display unit. The
totalization scale is determined in accordance
with the settings of B30/D10: Total Unit and
B31/D11: Total Scale.
• The maximum value that can be displayed is
99999999, and if this is exceeded, the value 0
is counted once again. However, counting
stops at 99999999 when the totalization switch
function is used.
• If multiple ranges are being used, the flow rate
span for the smallest range becomes the
standard for the D13: Total Low Cut setting
value.
• Totalization for the reverse flow rate and for the
differential flow rate is carried out only when
“Fwd/Rev Ranges” is selected for F10: SO1Function or F11: SO2 Function.
• The totalized units are indicated on the display
unit when B31/D11 is 0.001, 0.01, 0.1, 1, 10,
100, or 1000. In the case of other setting
values, the totalized units are not indicated.
6-13
IM 01E20C01-01E
Example 1: To count in 1 Ml (mega-liter) steps
NOTE
NOTE
with flow rate span = 100 m3/h
Since 1 Ml = 103 x m3, k Unit/P is
set for B30/D10, and 1 is set for
B31/D11. “x103 m3” is indicated for
the totalized units in the Display
Mode.
Example 2: To count in 10 l (liter) steps with
flow rate span = 100 m3/h
Since 1 l = 10-3 x m3, m Unit/P is
set for B30/D10, and 10 is set for
B31/D11. “x10-2 m3” is indicated for
the totalized units in the Display
Mode.
Example 3: To count in 5 l (liter) steps with
flow rate span = 100 m3/h
Since 1 l = 10-3 x m3, m Unit/P is
set for B30/D10, and 5 is set for
B31/D11. Since B31/D11 is not
0.001, 0.01, 0.1, 1, 10, 100, or
1000, there is no indication of
totalized units in the Display Mode.
• Setting of totalization scale is not possible
when specific selections have been made for
B30/D10: Total Unit, B31/D11: Total Scale,
and B23/C42: Flow Span. In such a case, a
setting alarm will be displayed, and parameters
should be changed in accordance with the
instructions given.
[B32: Pulse Unit] Setting of the pulse units
→ This setting is linked with that of parameter E10.
This parameter selects the flow rate units to be used for
pulse output.
ItemDescription
-
n Unit/P
u Unit/P
m Unit/P
Unit/P
k Unit/P
M Unit/P
Pulse/s
FU: Flow rate unit selected in B21/C40: Base Flow Unit.
9
10
FU
-
6
10
FU
-
3
10
FU
FU
3
10
FU
6
10
FU
Number of pulses to be output for one second at 100% output.
T0617.EPS
6. PARAMETER DESCRIPTION
If a pulse scale is specified upon ordering, this
parameter is setup before shipment; however, if
this is not the case, it will be necessary for the
appropriate value to be setup by the user.
• By setting the pulse scale, pulse output performs. The pulse scale is determined in accordance with the settings of B32/E10: Pulse Unit
and B33/E11: Pulse Scale.
• If multiple ranges are being used, the flow rate
span for the smallest range becomes the
standard for the E13: Pulse Low Cut setting
value.
• Pulse output for the reverse flow rate is carried
out only when “Fwd/Rev Ranges” is selected
for F10: SO1 Function or F11: SO2 Function.
• Setting of pulse scale is not possible when
specific selections have been made for B32/
E10: Pulse Unit, B33/E11: Pulse Scale, E12:
Pulse Width and B23/C42: Flow Span. In
such a case, a setting alarm will be displayed,
and parameters should be changed in accordance with the instructions given.
Example 1: To perform pulse output in 1 Ml
(mega-liter) steps with flow rate
span = 100 m3/h
Since 1 Ml = 103 x m3, k Unit/P is
set for B32/E10, and 1 is set for
B33/E11.
Example 2: To perform pulse output in 10 l
(liter) steps with flow rate span =
100 m3/h
Since 1 l = 10-3 x m3, m Unit/P is
set for B32/E10, and 10 is set for
B33/E11.
Example 3: To perform pulse output in 5 l
(liter) steps with flow rate span =
100 m3/h
Since 1 l = 10-3 x m3, m Unit/P is
set for B32/E10, and 5 is set for
B33/E11.
[B33: Pulse Scale] Setting of pulse scale
→ This setting is linked with that of parameter E11.
Pulse output is performed in individual counts in
accordance with this parameter’s setting. If 0 is
selected, it indicates that the pulse output function is
not to be used.
[B40: Display Select1] Setting of the first line for
display unit
→ This setting is linked with that of parameter H10.
This parameter selects the display content of the first
line for display unit. The size of the characters which
are displayed will depend on the selections made for
6-14
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
NOTE
NOTE
B41/H11: Display Select2 and B42/H12: Display
Select3 as described below. (For more details, refer to
Chapter 5: Basic Operating Procedures.)
CAUTION
It is not possible to set Display Select1 to “Off”.
[B41: Display Select2] Setting of the second line
for display unit
→ This setting is linked with that of parameter H11.
This parameter selects the display content of the
second line for display unit. When “Off” is selected,
one-line display will be adopted regardless of the
setting made for B42/H12: Display Select 3.
[B42: Display Select3] Setting of the third line for
display unit.
→ This setting is linked with that of parameter H12.
This parameter selects the display content of the third
line for display unit. When “Off” is selected for this
parameter, two-line display is adopted.
[B50: Auto Zero Exe] Execution of the automatic
zero adjustment function
→ This setting is linked with that of parameter M10.
This parameter executes the automatic zero adjustment
function: If “Execution” is selected, this function will
be started. “Now Auto Zero Executing...” is indicated
while the Auto Zero function is being carried out. The
result of the automatic zero adjustment is confirmed
using M11: Magflow Zero, and if the result exceeds
the rated value, the warning 82: Auto Zero Wng will
be displayed. (For more details, refer to Chapter 9:
Operation.)
SettingFunction
No Execution
Execution
No execution
Automatic zero adjustment is started.
T0643.EPS
(2) Menu C: Basic Setting items
Menu C principally contains the basic setting items for
the flowtube.
NOTE
will be setup upon shipment from the manufacturing plant, and these will not require additional
setting. If the AXFA11 is ordered individually, the
default value will be setup for the meter factor;
accordingly, it will be necessary to set the meter
factor indicated on your remote flowtube data
plate.
If a flow rate span is specified upon ordering,
this will be set before shipment. If this is not the
case, however, it will be necessary for the
appropriate value to be set up by the user.
[C10: Tag No] Setting of the tag number
→ The setting for this parameter corresponds to one of
the ordered items.
Up to a maximum of 16 characters can be entered for
the display unit. For more details regarding the actual
characters that can be used, refer to Section 5.2.2:
Setting Mode.
If a tag number is specified upon ordering, this
parameter is set up before shipment; however, if
this is not the case, it will be necessary for the
appropriate value to be set up by the user.
[C11: Flow Damping] Setting of the damping time
→ Refer to the description of parameter B20.
[C20: Measure Mode] Selection of dual frequency
excitation mode
ItemDescription
Standard DF Standard dual frequency excitation
Enhanced DF Enhanced dual frequency excitation
T0618.EPS
Enhanced DF (i.e., enhanced dual frequency excitation)
is set to stabilize measurement for difficult applications, such as for high concentration slurries or low
conductivity fluid. Note that this parameter is only
valid when this product is used in combination with an
AXF flowtube, as only supports enhanced dual
frequency excitation. (Optional code HF1 or HF2)
In order to ensure that correct flow rate data can
be acquired, it is crucial that the nominal size,
flow rate span, and meter factor of the combined
remote flowtube are setup. In cases where a
remote flowtube is ordered at the same time as
the AXFA11, the nominal size and meter factor
•When this product is used in combination with
any flowtube that does not support enhanced
dual frequency excitation and “Enhanced DF” is
6-15
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
selected, a setting alarm will be displayed. The
setting should be returned to “Standard DF” in
such a case.
•When dual frequency excitation mode is
changed, perform zero adjustment. For details
on zero adjustment, refer to chapter 9.
[C21: Low MF] Setting of the low-frequency meter
factor
This parameter sets the low-frequency meter factor for
standard dual frequency excitation.
This parameter is also used for pulsed DC excitation in
which AXFA11 is combined with a remote flowtube
the size of 500 mm (20 in.) or larger.
[C22: High MF] Setting of the high-frequency meter
factor
This parameter sets the high-frequency meter factor for
standard dual frequency excitation.
If AXFA11 combined with a remote flowtube the size
of 500 mm or larger, high-frequency meter factor need
not be set.
NOTE
Meter Factor Settings
[C23: Low MF (EDF)] Setting of the low-frequency
meter factor for EDF
This parameter sets the low-frequency meter factor as
required when Enhanced DF (i.e., enhanced dual
frequency excitation) is selected. If “Standard DF” has
been selected for C20: Measure Mode, neither C23:Low MF (EDF) nor C24: High MF (EDF) is
displayed, and if “Enhanced DF” is selected, the four
parameters from C21 to C24 will be displayed.
[C24: High MF (EDF)] Setting of the high-frequency
meter factor for EDF
This parameter sets the high-frequency meter factor as
required when Enhanced DF (i.e., enhanced dual
frequency excitation) is selected.
[C30: Select Flow Tube] Setting of the flowtube’s
model
This parameter sets the models of flowtube. When
combining this product with an AXF Remote
Flowtube, “ADMAG AXF” should be selected.
[C31: Nominal Size Unit] Setting of the nominal
size units
This parameter selects the units used for setting of the
nominal size.
Meter factors of flowtube
Flowtube data plate
Converter data plate
F0601.EPS
(1) Confirm that the serial number indicated by
COMB.NO. on the AXFA11 converter’s data
plate corresponds with the AXF Remote
Flowtube’s serial number.
(2) Set the values that are marked in the
METER FACTOR fields on the data plate for
the Remote Flowtube.
(3) The meter factors are crucial in ensuring that
the electromotive force is correctly in proportion to the flow velocity and are determined
at the manufacturing plant by actual-flow
calibration.
[C32: Nominal Size] Setting of the nominal size
This parameter sets the nominal size of flowtube.
[C40: Base Flow Unit] Selection of flow units for
the flow rate span
→ Refer to the description of parameter B21.
[C41: Base Time Unit] Selection of time units for
the flow rate span
→ Refer to the description of parameter B22.
[C42: Flow Span] Setting of the flow rate span→ Refer to the description of parameter B23.
[C43: Flow Decimal Pnt] Setting of the decimal
point position for the instantaneous flow rate
→ Refer to the description of parameter B24.
[C44: Velocity Check] Display of the flow rate span
velocity
This parameter displays the flow rate span for the
maximum range in m/s units.
6-16
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
[C45: Density Unit] Setting of the density units for
mass flow rate
This parameter selects the units for density as required
when making settings using C46: Mass Density.
[C46: Mass Density] Setting of the density for mass
flow rate
This parameter is necessary in situations where t, kg, g,
klb or lb has been selected as the mass unit in B21/
C40: Base Flow Unit. If a mass unit is selected in
B21/C40: Base Flow Unit and a value of 0 is set for
this parameter, the setting alarm “57: Dens Set Err”
will be displayed. In such a case, ensure that the
density is set correctly.
[C47: User Span Select] Selection of the use of
special flow rate units.
This parameter selects whether or not special units are
used for instantaneous flow rate. Actual setting of these
units is carried out using C48: Flow User Unit and
C49: Flow User Span.
[C48: Flow User Unit] Setting of the special flow
rate units
This parameter is used to select the special units (up to
maximum of 8 characters in length). These units are
displayed when instantaneous flow rate is selected in
the Display Mode, and they are displayed for A20:FLOW RATE when BRAIN communication is being
carried out.
[C49: Flow User Span] Setting of the special flow
rate span
This parameter sets the special flow rate span to be
displayed for 100% output in the maximum range.
Span.
“100 dl/s” is indicated for 100%
output in the Display Mode.
(3) Menu D: Total Setting items
Menu D contains parameters that are relevant to
totalization function settings.
[D10: Total Unit] Setting of units for totalization
scale
→ Refer to the description of parameter B30.
[D11: Total Scale] Setting of the totalization scale→ Refer to the description of parameter B31.
[D12: Total Decimal Pnt] Setting of the decimal
point position for the totalization display
This parameter sets the position of the decimal point
for totalization display in terms of the number of digits.
Except in cases where 0 is selected, the totalized units
are not displayed.
Example: When totalized value is 12345678 m
Item
0
1
2
3
4
5
6
7
Totalization display
12345678 m
1234567.8
123456.78
12345.678
1234.5678
123.45678
12.345678
1.2345678
3
3
T0619.EPS
NOTE
Example :To set the special flow rate span to
100 dl/s
Since 100 dl (deci-liter)=10 l (liter),
“l (Liter)” is set for B21/C40: BaseFlow Unit,
“/s” is set for B22/C41: Base TimeUnit,
“10” is set for B23/C42: FlowSpan,
“Yes” is set for C47: User SpanSelect,
“dl/s” is set for C48: Flow UserUnit,
“100” is set for C49: Flow User
[D13: Total Low Cut] Setting of the totalization stop
range
This parameter allows the settings to be made that
prevent totalization when the flow rate is at or below
the low-cut setting value. In cases where there are
multiple ranges or forward/reverse ranges, low cut is
carried out at the setting value for the smallest span
(i.e., an integer between 0 and 10%).
Example: When the first range = 10 m3/h, the second
range = 100 m3/h, and the Total Low Cut =
3%, no totalization is carried out at flow
rates of 0.3 m3/h or lower.
6-17
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
NOTE
[D20: Total Execution] Operation setting for the
totalization function
This parameter sets “Start” and “Stop” of the totalization function, in addition to performing the preset
function for the forward totalized value and the reverse
totalized value.
*: The preset function starts the count for totalization
from the set value.
ItemDescription
Start (initial value)
Stop
Preset Total
Preset Rev Total
NOTE
Starts totalization
Stop totalization
Sets the preset value for totalization display that has
been specified as the forward totalized value. Preset
value are determined using D21: Ttl Set Val Lower
and D22: Ttl Set Val Upper.
In case that “Start” is sellected, the count for
totalization starts from the preset value. Setting of zero
as the preset value allows the zero-reset function to be
implemented.
Sets the preset value for totalization display that has
been specified as the reverse totalized value. Preset
value are determined using D21: Ttl Set Val Lower
and D22: Ttl Set Val Upper.
In case that “Start” is sellected, the count for
totalization starts from the preset value. Setting of zero
as the preset value allows the zero-reset function to be
implemented.
T0620.EPS
Totalization presets can also be set up by using
status input. For details regarding the setting
method, refer to the descriptions of F12 and
F13.
[D23: Ttl Switch Lower] Setting of the totalization
switch value (lower 6 digits)
The totalization switch function operates to set the
status output terminal (i.e., SO1 or SO2) to “Closed
(On)” when the forward internal totalized value reaches
or exceeds the totalization switch value. (For details
regarding the setting method for the status output, refer
to the descriptions of parameters F10 and F11.)
If this function is set up, the totalization count will stop
at 99999999.
D23 sets the lower 6 digits of the 8-digit totalization
switch value.
[D24: Ttl Switch Upper] Setting of the totalization
switch value (upper 2 digits)
This parameter sets the upper 2 digits of the 8-digit
totalization switch value.
[D30: Ttl User Select] Selection of the use of
special totalization unit
This parameter specifies whether or not special units
are used for totalized units. Actual setting of these
units is carried out using D31: Ttl User Unit.
[D31: Ttl User Unit] Setting of special totalization
units
Units of up to maximum 8 characters in length can be
specified using this parameter. The units set with this
parameter are displayed whenever totalization (i.e.,
FTL, RTL, DTL) is selected in the Display Mode, and
they are displayed for A30: TOTAL, A31: REVTOTAL, and A32: DIF TOTAL when BRAIN
communication is being carried out.
[D21: Ttl Set Val Lower] Setting of the totalization
preset value (lower 6 digits)
This parameter sets a totalization preset value in the
lower 6 digits of the 8-digit totalized value. If zero is
to be set as the preset value, “000000” should be set
here.
[D22: Ttl Set Val Upper] Setting of the totalization
preset value (upper 2 digits)
This parameter sets a totalization preset value in the
upper 2 digits of the 8-digit totalized value. If zero is
to be set as the preset value, “00” should be set here.
Upper 2 digits
Lower 6 digits
F0602.EPS
Example:To count in 1 dl (deci-liter) steps
with flow rate span=10 l/s.
Since 1 dl (deci-liter) = 0.1 l (liter),
“l (Liter)” is set for B21/C40: BaseFlow Unit,
“/s” is set for B22/C41: Base TimeUnit,
“10” is set for B23/C42: FlowSpan,
“Unit/P” is set for B30/D10: TotalUnit,
“0.1” is set for B31/D11: TotalScale,
“Yes” is set for D30: Ttl UserSelect,
6-18
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
and “dl” is set for D31: Ttl User
Unit.
“dl” is indicated for the totalized
units in the Display Mode and is
counted in 1 dl steps.
(4) Menu E: Pulse Setting items
Menu E contains items relevant to pulse output.
[E10: Pulse Unit] Setting of the pulse units
→ Refer to the description of parameter B32: Pulse
Unit
[E11: Pulse Scale] Setting of the pulse scale→ Refer to the description of parameter B33: Pulse
Scale
[E12: Pulse Width] Setting of the pulse width
This parameter selects the pulse width (i.e., ms :
millisecond) that is output.
[E13: Pulse Low Cut] Setting of the pulse output
stop range
This parameter allows the settings to be made which
prevent pulse output when the flow rate is at or below
the low-cut setting value. In cases where there are
multiple ranges or forward/reverse ranges, low cut is
carried out at the setting value for the smallest span
(i.e., an integer between 0 and 10%).
Example: When the first range = 10 m3/h, the second
range = 100 m3/h, and the Pulse Low Cut =
3%, no pulse output is carried out at flow
rates of 0.3 m3/h or lower.
[E20: Pulse Active Mode] Setting of active mode
for pulse output
This parameter sets whether pulse-output active mode
is to be “On Active” or “Off Active”.
ItemDescription
Closed (On) Act
Open (Off) Act
When pulse output is carried out, contacts
are Closed (On).
When pulse output is carried out, contacts
are Open (Off).
T0622.EPS
Data Range
Maximum Value Minimum Value
(0) 50% Duty
(1) 0.05ms
(2) 0.1ms
(3) 0.5ms
(4) 1ms
(5) 20ms
(6) 33ms
(7) 50ms
(8) 100ms
Pulse Rate (pps)
Setting
110000.0001
10000
5000
1000
500
25
15
10
5
NOTE
(pps: pulses per second)
*:The pulse width with the exception of “50%
Duty” is the “Closed (On)” time for each pulse
in case that “Closed (On) Act” is sellected for
E20: Pulse Active Mode.
Closed(On)
Example: When E12: Pulse Width is 1 ms, output
takes place as follows in accordance with the
setting for E20: Pulse Active Mode.
Closed(On)
Open(Off)
1 ms1 ms
Closed (On) ActiveOpen (Off) Active
F0604.EPS
T0621.EPS
Open(Off)
Pulse width
F0603.EPS
A limit applies to the maximum pulse scale that
can be set with respect to the pulse width. If a
value in excess of this limit is set, a setting
alarm will be displayed.
6-19
IM 01E20C01-01E
(5) Menu F: Status Functions Setting items
Menu F contains setting items relevant to status Input/Output functions.
[F10: SO1 Function] Setting of the function for the SO1 status output terminal
This parameter sets the function for the SO1 (status output 1) terminal.
Setting
No FunctionStop output (i.e., inactive condition) As no function is set, there is no output.
Warning Output Output upon warning
Total SwitchTotalization switch output
H/L AlarmH/L alarm output
HH/LL AlarmHH/LL alarm output
Fwd/Rev Ranges
Auto 2 RangesAutomatic 2 ranges switching
Auto 3 Ranges
Auto 4 Ranges
Ext 2 Answer
Ext 3 Answer
Ext 4 Answer
Note 1: When these functions are selected, two terminals become necessary for status output.
Accordingly, the setting for F10: SO1 Function is automatically adopted as the setting for F11: SO2 Function.
(Setting of these functions is not possible using F11: SO2 Function.)
Note 2: Function-specific SO1 and SO2 operations
Operations are performed in accordance with the content of the following table when the active mode has been
set to "Closed (On) Act" using F14: SO1/2 Active Mode.
Operating patterns are reversed when the active mode has been set to "Open (Off) Act" using this parameter.
Forward/reverse flow rate measurement
Automatic 3 ranges switching (Note 1)
Automatic 4 ranges switching (Note 1)
Answerback: 2 ranges switching via
external status input
Answerback: 3 ranges switching via
external status input (Note 1)
Answerback: 4 ranges switching via
external status input (Note 1)
Refer to Alarms (Section 6.5)
Status output is carried out when the forward internal
totalized value reaches or exceeds the totalization switch value.
The totalization switch value is determined using D23: Ttl
Switch Lower and D24: Ttl Switch Upper.
Status output is carried out when the instantaneous flow rate
equals or falls below the low flow rate limit (L), or when it equals
or exceeds the high flow rate limit (H). These limit values are
determined using G10: Low Alarm and G11: High Alarm.
Status output is carried out when the instantaneous flow rate
equals or falls below the low-low flow rate limit (LL), or when it
equals or exceeds the high-high flow rate limit (HH). These limit
values are determined using G12: Low Low Alarm and G13:
High-High Alarm. To output “HH/LL Alarm” as an alarm, set
G33: Alm-HH/LL to “Yes”.
When flow is in the reverse direction, switching to the reverse
range is carried out automatically, measurement is performed,
and status output is carried out.
This function ensures that when the
exceeds 100% of the range, transition to the next range is
carried out automatically. Status output is carried out
upon range switching.
When range switching is carried out in response to external
status input, status output is performed in the form of an
answerback to indicate the range currently being used.
3 Answer,” and “Ext 4 Answer,” see the Multiple ranges setting section.
Condition of SO1 or SO2 terminal
Good (normal)
Below setting value
Normal
Normal
Forward direction
Warning status
Equal or above setting value
H/L alarm status
HH/LL alarm status
Reverse direction
Closed (On)Open (Off)
T0624.EPS
[F11: SO2 Function] Setting of function for the SO2 status output terminal
This parameter sets the function for the SO2 (status output 2) terminal. Selectable functions are the same as for F10:SO1 Function; however, the selection of “Auto 3 Ranges”, “Auto 4 Ranges”, “Ext 3 Answer”, or “Ext 4 Answer”
is not possible for F11.
6-20
IM 01E20C01-01E
[F12: SI1 Function] Setting of the funtion for the SI1 status input terminal
This parameter sets the function for the SI1 (status input 1) terminal.
Setting
No FunctionNo input function
Based on the external status input, the instantaneous flow rate
indication is forcibly set to 0% (i.e., 4 mA), and both totalization
0% Signal Lock
Ext Auto Zero
Ext Ttl Preset
Ext Rev Ttl Set
Ext 2 Ranges
Ext 3 Ranges
Ext 4 Ranges
Note 1: When these functions are selected, two terminals become necessary for status input.
Accordingly, the setting for F12: SI1 Function is automatically adopted as the setting for F13: SI2 Function.
(Setting of these functions is not possible using F13: SI2 Function.)
Note 2: Function-specific SI1 and SI2 operations
Operations are performed in accordance with the content of the following table when the active mode has
been set to "Short Active" using F15: SI1/2 Active Mode.
Operating patterns are reversed when the active mode has been set to "Open Active" using this parameter.
Selected function
0% Signal Lock
Ext Auto Zero
Ext Ttl Preset
Ext Rev Ttl Set
Note: For “Ext 2 Ranges,” “Ext 3 Ranges,” and “Ext 4 Ranges,” see the Multiple
ranges setting section.
0% signal lock via external status
input.
Automatic zero adjustment via
external status input
Forward totalization preset via
external status input
Reverse totalization preset via
external status input
2 ranges switching via external
status input
3 ranges switching via external
status input (Note 1)
4 ranges switching via external
status input (Note 1)
Condition of SI1 or SI2
Normal
Normal
Normal
Normal
and pulse outputs are halted. This setting has precedence over
the output signal whenever an alarm occurs. When 0% signal lock
is canceled, the instantaneous flow rate is restored in accordance
with the time constant originally set using B20/C11:Flow Damping.
Automatic zero adjustment is carried out in response to
external status input. For more details regarding automatic
zero adjustment, refer to Chapter 9: Operation.
The totalization display value is preset or reset to zero in
accordance with the external status input. The preset value is
determined using D21: Ttl Set Val Lower and D22: Ttl Set Val Upper.
In case that “Start” is selected for D20: Total Execution , the
count for totalization starts from the preset value.
This function allows switching of up to 4 ranges in
response to status input for a single direction only.
ShortOpen
Signal lock status
Start of automatic zero adjustment
Forward totalization preset
Reverse totalization preset
6. PARAMETER DESCRIPTION
DescriptionFunction
T0625.EPS
T0626.EPS
[F13: SI2 Function] Setting of the function for the
SI2 status input terminal
This parameter sets the function for the SI2 (status
input 2) terminal. Selectable functions are the same as
for F12: SI1 Function; however, the selection of Ext
3 Ranges and Ext 4 Ranges is not possible for F13.
[F14: SO1/2 Active Mode] Setting of the active
mode for status output
This parameter sets the active mode for the terminals
SO1 and SO2. Active modes cannot be set individually
for these two terminals.
SettingFunction
Closed (On) Act
Open (Off) Act
Status output becomes "Closed (On)" when an
event occurs.
Status output becomes "Open (Off)" when an
event occurs.
Example: When the “Total Switch” function is selected
for SO1 or SO2
Closed(On)
Open(Off)
Equal or above setting value
Closed (On) Active
6-21
Equal or above setting value
Open (Off) Active
IM 01E20C01-01E
T0627.EPS
F0605.EPS
6. PARAMETER DESCRIPTION
[F15: SI1/2 Active Mode] Setting of the active mode
for status input
This parameter sets the active mode for the terminals
SI1 and SI2. Active modes cannot be set individually
for these two terminals.
SettingFunction
Short Active
Open Active
When the status input is set to "Short", occurrence
of the selected event will be recognized.
When the status input is set to "Open", occurrence
of the selected event will be recognized.
T0628.EPS
Example: When the “0% signal lock” function is
selected for SI1 or SI2
Short
Open
Signal lock status
Short Active
Signal lock status
Open Active
F0606.EPS
NOTE
Multiple ranges setting
Parameters from F30 to F41 are used with the
automatic multiple ranges and the multiple
ranges switching via external status input. The
followings will describe the setting method for
each range.
The multiple ranges use the following parameters:
[B23: Flow Span ] Setting of the flow rate span
(Setting of the forward No.1 range)
[F30: Forward Span 2] Setting of the forward No. 2
range
[F31: Forward Span 3] Setting of the forward No. 3
range
[F32: Forward Span 4] Setting of the forward No. 4
range
[F33: Reverse Span 1] Setting of the reverse No. 1
range
[F34: Reverse Span 2] Setting of the reverse No. 2
range
[F35: Reverse Span 3] Setting of the reverse No. 3
range
[F36: Reverse Span 4] Setting of the reverse No. 4
range
[F40: Auto Range Hys] Setting of the automatic
multiple range hysteresis width
[F41: Bi Direction Hys] Setting of the forward/
reverse flow measurement hysteresis width
Multiple Ranges Setting 1:
Automatic multiple ranges switching
When the instantaneous flow rate exceeds 100% of
the range, transition to the next range (up to four
ranges) is carried out automatically. Furthermore,
when the flow is in reverse, the reverse range can
also be automatically selected.
Range switching can be confirmed according to the
status of the SO1 and SO2 status output terminals.
Refer to Table 6.4.1: Status Output for Automatic
Multiple Ranges Switching for details of status
output conditions for each range.
Status Output for Automatic Multiple Ranges Switching
Operations are performed in accordance with the
following table when the active mode has been set to
“Closed (On) Act” using F14: SO1/2 Active Mode.
Operating patterns are reversed when the active mode
has been set to “Open (Off) Act”.
Table 6.4.1 Status Output for Automatic Multiple Ranges
Switching
Parameter setting
F10
No Function
Fwd/Rev
Ranges
No Function
Auto
2 Ranges
No Function
Auto
3 Ranges
Auto
4 Ranges
Fwd/Rev
Ranges
Auto
2 Ranges
*: "No Function" is the default value. Only SO1 or SO2 terminals are
used for single or dual ranges; accordingly, the unused terminal
remains at the default value (i.e., No Function) and can therefore
be used for other functions.
No Function
No Function
Fwd/Rev
Ranges
No Function
Auto
2 Ranges
Auto
3 Ranges
Auto
4 Ranges
Auto
2 Ranges
Fwd/Rev
Ranges
Function
Forward single range
Auto forward/reverse
1 range (SO1)
Auto forward/reverse
1 range (SO2)
Auto forward
2 ranges (SO1)
Auto forward
2 ranges (SO2)
Auto forward
3 ranges
Auto forward
4 ranges
Auto forward/reverse
2 ranges
Auto forward/reverse
2 ranges
Forward
Reverse
Forward
Reverse
Forward 1 range
Forward 2 range
Forward 1 range
Forward 2 range
Forward 1 range
Forward 2 range
Forward 3 range
Forward 1 range
Forward 2 range
Forward 3 range
Forward 4 range
Forward 1 range
Forward 2 range
Reverse 1 range
Reverse 2 range
Forward 1 range
Forward 2 range
Reverse 1 range
Reverse 2 range
6-22
Status output
SO2SO1F11
–
–
Open
–
Closed
Closed
Closed
Closed
Closed
Closed
Closed
Closed
Closed
IM 01E20C01-01E
–
–
Open
–
–
Open
Open
Open
Open
Open
Open
Open
Open
–
Open
Closed
–
–
Open
Closed
Open
Open
Closed
Open
Open
Closed
Closed
Open
Closed
Open
Closed
Open
Open
Closed
Closed
T0629.EPS
6. PARAMETER DESCRIPTION
NOTE
Parameter setting sequence
(for automatic multiple ranges switching)
F10: SO1 Function
Select a function
F14: SO1/2 Active Mode
Select whether SO1 and SO2 output is to
be “Closed (On) Act” or “Open (Off) Act”
Function Selection
No Function: Output is stopped (*: other functions may be selected)
Fwd/Rev Ranges: Forward and reverse flow rate measurement
Auto 2 Ranges: Automatic 2-ranges switching
Auto 3 Ranges: Automatic 3-ranges switching
Auto 4 Ranges: Automatic 4-ranges switching
B23: Flow Span
Forward No. 1 range
F32: Forward
Span 4
Forward No. 4 range
If “instantaneous flow rate % display (FR)” is selected for
Display Mode and A10: FLOW RATE(%) is selected for
BRAIN communication, the instantaneous flow rate %
and following symbols will be displayed only for multiple
ranges and forward/reverse flow measurement.
Forward No. 1 range : [F1] Reverse No. 1 range : [R1]
Forward No. 2 range : [F2] Reverse No. 2 range : [R2]
Forward No. 3 range : [F3]
Forward No. 4 range : [F4]
F30: Forward
Forward No. 2 range
F33: Reverse
Reverse No. 1 range
Set the spans for the ranges to be used.
No. 1 range No. 2 range
No. 3 range No. 4 range
F11: SO2 Function
Select a function
Refer to Table 6.4.1
and set F10 or F11.
Span 2
Span 1
F31: Forward
Span 3
Forward No. 3 range
F34: Reverse
Span 2
Reverse No. 2 range
F0607.EPS
Output
Reverse
Figure 6.4.1 Multiple Ranges and Hysteresis Widths
No. 1 range
Minimum span % within multiple and
forward/reverse (set using F41)
No. 2 range
10% of No. 1 range
(set using F40)
No. 3 range
10% of No. 3 range
10% of No. 2 range
(set using F40)
(set using F40)
Forward
No. 4 range
100%
F0608.EPS
For more details regarding the setting of hysteresis width, refer to the description of setting
parameter for F40: Auto Range Hys and F41:Bi Direction Hys.
Input
Multiple Ranges Setting 2:
Multiple ranges switching via external
status input
For both the forward and reverse directions, switch-
ing of up to four ranges can be carried out based on
status input; however, switching between directions
is not possible. Switching between forward and
reverse ranges is carried out automatically only
when the flow direction reverses.
SI1 and SI2 status input terminals are used for
multiple range switching. For more details, refer to
Table 6.4.2: Multiple Ranges Switching via External
Status Input.
6-23
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
T0630.EPS
Parameter setting
Function
Status inputStatus output
SI2(F13)
–
–
Open
Short
Open
Open
Short
Open
Open
Short
Short
Open
Short
Open
Short
Open
Short
Open
Short
Open
Open
Short
Open
Open
Short
Open
Open
Short
Open
Open
Short
Open
Open
Short
Short
Open
Open
Short
Short
Open
Open
Short
Short
Open
Open
Short
Short
Open
Short
Open
Short
Open
Open
Short
Open
Open
Short
Short
Open
Short
Open
Short
Open
Short
Open
Short
SO1(F10)
–
–
–
–
–
–
–
–
–
–
–
Open
Open
Closed
Closed
Open
Open
Closed
Closed
–
–
–
–
–
–
–
–
Open
Open
Open
Closed
Closed
Closed
–
–
–
–
–
–
Open
Open
Open
Open
Closed
Closed
Closed
Closed
–
–
–
–
–
–
–
–
Open
Closed
Open
Closed
–
–
–
–
Open
Closed
Open
Open
Closed
Open
Closed
Open
Open
Closed
Closed
Open
Open
Closed
Closed
Open
Closed
Open
Closed
Open
Closed
Open
Closed
SO2(F11)
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
Open
Open
Closed
Closed
Open
Open
Closed
Closed
–
–
–
–
–
–
Open
Open
Open
Closed
Closed
Closed
–
–
–
–
–
–
–
–
Open
Open
Open
Open
Closed
Closed
Closed
Closed
–
–
–
–
Open
Closed
Open
Closed
Open
Open
Closed
Open
Open
Closed
Closed
Open
Closed
Open
Closed
Open
Closed
Open
Closed
Open
Open
Closed
Closed
Open
Open
Closed
Closed
SI1(F12)
Open
Short
–
–
Open
Short
Open
Open
Short
Open
Short
Open
Short
Open
Short
Open
Short
Open
Short
Open
Short
Open
Open
Short
Open
Open
Short
Open
Open
Short
Open
Open
Short
Open
Short
Open
Short
Open
Short
Open
Short
Open
Short
Open
Short
Open
Short
Open
Short
Open
Short
Open
Short
Open
Open
Short
Open
Short
Open
Short
Open
Short
Open
Short
Open
Short
F13
No Function
F12
Ext 2 Ranges
F11
No Function
F10
No Function
External forward
2 ranges
External
forward/reverse
4 ranges
External
forward/reverse
2 ranges, with
answerback
External forward
4 ranges, with
answerback
External forward
3 ranges, with
answerback
External forward
2 ranges, with
answerback
External
forward/reverse
3 ranges
External
forward/reverse
2 ranges
External forward
4 ranges
External forward
3 ranges
Ext 2 RangesNo FunctionNo FunctionNo Function
Ext 3 RangesExt 3 RangesNo FunctionNo Function
Ext 4 RangesExt 4 RangesNo FunctionNo Function
No FunctionExt 2 RangesNo Function
Fwd/Rev
Ranges
Ext 2 RangesNo FunctionNo Function
Fwd/Rev
Ranges
No FunctionExt 2 Ranges
Fwd/Rev
Ranges
No Function
Ext 2 RangesNo Function
Fwd/Rev
Ranges
No Function
Ext 3 RangesExt 3 RangesNo Function
Fwd/Rev
Ranges
Ext 3 RangesExt 3 Ranges
Fwd/Rev
Ranges
No Function
Ext 4 RangesExt 4 RangesNo Function
Fwd/Rev
Ranges
Ext 4 RangesExt 4 Ranges
Fwd/Rev
Ranges
No Function
No FunctionExt 2 RangesNo FunctionExt 2 Answer
Ext 2 RangesNo FunctionNo FunctionExt 2 Answer
No FunctionExt 2 RangesExt 2 AnswerNo Function
Ext 2 RangesNo FunctionExt 2 AnswerNo Function
Ext 3 RangesExt 3 RangesExt 3 AnswerExt 3 Answer
Ext 4 RangesExt 4 RangesExt 4 AnswerExt 4 Answer
No FunctionExt 2 RangesExt 2 Answer
Fwd/Rev
Ranges
Ext 2 RangesNo FunctionExt 2 Answer
Fwd/Rev
Ranges
No FunctionExt 2 Ranges
Fwd/Rev
Ranges
Ext 2 Answer
Ext 2 RangesNo Function
Fwd/Rev
Ranges
Ext 2 Answer
Forward 1 range
Forward 2 range
Forward 1 range
Forward 2 range
Forward 1 range
Forward 2 range
Forward 3 range
Forward 1 range
Forward 2 range
Forward 3 range
Forward 4 range
Forward 1 range
Forward 2 range
Reverse 1 range
Reverse 2 range
Forward 1 range
Forward 2 range
Reverse 1 range
Reverse 2 range
Forward 1 range
Forward 2 range
Reverse 1 range
Reverse 2 range
Forward 1 range
Forward 2 range
Reverse 1 range
Reverse 2 range
Forward 1 range
Forward 2 range
Forward 3 range
Reverse 1 range
Reverse 2 range
Reverse 3 range
Forward 1 range
Forward 2 range
Forward 3 range
Reverse 1 range
Reverse 2 range
Reverse 3 range
Forward 1 range
Forward 2 range
Forward 3 range
Forward 4 range
Reverse 1 range
Reverse 2 range
Reverse 3 range
Reverse 4 range
Forward 1 range
Forward 2 range
Forward 3 range
Forward 4 range
Reverse 1 range
Reverse 2 range
Reverse 3 range
Reverse 4 range
Forward 1 range
Forward 2 range
Forward 1 range
Forward 2 range
Forward 1 range
Forward 2 range
Forward 1 range
Forward 2 range
Forward 1 range
Forward 2 range
Forward 3 range
Forward 1 range
Forward 2 range
Forward 3 range
Forward 4 range
Forward 1 range
Forward 2 range
Reverse 1 range
Reverse 2 range
Forward 1 range
Forward 2 range
Reverse 1 range
Reverse 2 range
Forward 1 range
Forward 2 range
Reverse 1 range
Reverse 2 range
Forward 1 range
Forward 2 range
Reverse 1 range
Reverse 2 range
*: "No Function" is the default value. Only SI1 or SI2 terminals are used for single or dual ranges; accordingly, the
unused terminal remains at the default value (i.e., No Function) and can therefore be used for other functions.
Multiple Ranges Switching via External Status Input
Operations are performed in accordance with the following table when the active mode has been set to “Closed (On) Act” using
F14: SO1/2 Active Mode and when the active mode has been set to “Short Active” using F15: SI1/2 Active Mode.
Operating patterns are reversed when the active mode has been set to “Open (Off) Act” using F14; and “Open Active”, using
F15.
Table 6.4.2 Multiple Ranges Switching via External Status Input
6-24
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
Parameter setting sequence (for multiple ranges switching via external status input)
F10: SO1 Function
Select a function
F14: SO1/2 Active Mode
Select whether SO1 and SO2 output is to be “Closed (on) Act” or
“Open (off) Act.”
Function Selection
No Function: Output is stopped (*: other functions may be selected)
Fwd/Rev Ranges: Forward and reverse flow rate measurement
Ext 2 Answer: Answerback; 2-ranges switching via external status input
Ext 3 Answer: Answerback; 3-ranges switching via external status input
Ext 4 Answer: Answerback; 4-ranges switching via external status input
B23:Flow Span
Forward No. 1 range
F33:Reverse
Span1
Reverse No. 1 range
If “instantaneous flow rate % display (FR)” is selected for the Display
Mode and A10: FLOW RATE(%) is selected for BRAIN communication,
the instantaneous flow rate % and following symbols will be displayed
only for multiple ranges and forward/reverse flow measurement.
Forward No. 1 range : [F1]
Forward No. 2 range : [F2]
Forward No. 3 range : [F3]
Forward No. 4 range : [F4]
F11: SO2 Function
Select a function
Refer to Table 6.4.2 and set F10
and F11 or F12 and F13.
F30:Forward
Span2
Forward No. 2 range
F34:Reverse
Span2
Reverse No. 2 range
Reverse No. 1 range : [R1]
Reverse No. 2 range : [R2]
Reverse No. 3 range : [R3]
Reverse No. 4 range : [R4]
F12: SI1 Function
Select a function
F15: SI1/2 Active Mode
Select whether SI1 and SI2 input is to be “Short Active”
or “Open Active.”
Function Selection
No Function: No input function (*: other functions may be
selected)
Ext 2 Ranges: 2-ranges switching via external status input
Ext 3 Ranges: 3-ranges switching via external status input
Ext 4 Ranges: 4-ranges switching via external status input
F31:Forward
Span3
Forward No. 3 range
F35:Reverse
Span3
Reverse No. 3 range
Set the spans for the ranges to be used.
No. 1 range No. 2 range No. 3 range No. 4 range
F32:Forward
Span4
Forward No. 4 range
F36:Reverse
Span4
Reverse No. 4 range
F13: SI2 Function
Select a function
F0609.EPS
[F40: Auto Range Hys] Setting of automatic rangeswitching hysteresis width
Automatic switching takes place for multiple ranges
switching when 100% of the range is exceeded, and
this parameter allows a hysteresis width to be set for
this switching.
Refer to Figure 6.4.1: Multiple Ranges and Hysteresis
Widths.
[F41: Bi Direction Hys] Setting of forward/reverse
flow measurement hysteresis width
This parameter sets the hysteresis for forward/reverse
flow rate measurement as a % value of the minimum
flow span.
Refer to Figure 6.4.1: Multiple Ranges and Hysteresis
Widths.
(6) Menu G: Alarm Setting items
(Refer to Section 6.5: Alarm Functions for more
details.)
Menu G principally contains setting items relevant to
alarms.
[G10: Low Alarm] Low alarm setting
This parameter sets the low limit (L) alarm value, and
this is done using a % value of the maximum span.
A setting value of -110% indicates that the alarm is
disabled.
[G11: High Alarm] High alarm setting
This parameter sets the high limit (H) alarm value, and
this is done using a % value of the maximum span.
A setting value of 110% indicates that the alarm is
disabled.
[G12: Low Low Alarm] Low-low alarm setting
This parameter sets the low-low limit (LL) alarm
value, and this is done using a % value of the maximum span.
A setting value of -110% indicates that the alarm is
disabled.
6-25
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
[G13: High High Alarm] High-high alarm setting
This parameter sets the high-high limit (HH) alarm
value, and this is done using a % value of the maximum span.
A setting value of 110% indicates that the alarm is
disabled.
NOTE
Setting of -110% or 110% results in the corresponding function being disabled; accordingly,
settings can be combined to implement only high
alarms or low alarms, etc.
Output Example 1
The high-high alarm (HH) is set to 90% or more of the
flow rate span; the low-low alarm (LL), to 20% or less;
the high alarm (H), to 80% or more; and the low alarm
(L), to 30% or less.
Settings are:
G10: Low Alarm = 30%
G11: High Alarm = 80%
G12: Low Low Alarm = 20%
G13: High High Alarm = 90%
[%]
100
90
80
50
30
20
0Time (t)
SO1
HL
High High Alarm
High Alarm
Instantaneous flow rate
Low Alarm
Low Low Alarm
Output Example 2
The high alarm (H) is set to 80% or more of the flow
rate span; the low-low alarm (LL), to 20% or less.
Settings are:
G10: Low Alarm = -110%
G11: High Alarm = 80%
G12: Low Low Alarm = 20%
G13: High High Alarm = 110%
[%]
100
80
High Alarm
Instantaneous flow rate
50
20
Low Low Alarm
0Time (t)
SO1
SO2
AL
Select “H/L Alarm” for F10: SO1 Function
Select “HH/LL Alarm” for F11: SO2 Function
Select “Closed (On) Act” for F14: SO1/2 Active Mode
Select “Open (Off) Act” for G20: Alm Out Act Mode
Select “Yes” for G33: Alm-HH/LL
H
LL
HH Alarm set to 110% indicates
that the alarm is disabled.
LL
L Alarm set to
-110% indicates
that the alarm is
disabled.
F0611.EPS
SO2
AL
Select “H/L Alarm” for F10: SO1 Function
Select “HH/LL Alarm” for F11: SO2 Function
Select “Closed (On) Act” for F14: SO1/2 Active Mode
Select “Open (Off) Act” for G20: Alm Out Act Mode
Select “Yes” for G33: Alm-HH/LL
HH
HH
LL
LL
F0610.EPS
6-26
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
Output Example 3
The high alarm (H) is set to 80% or more of the flow
rate span; the high-high alarm (HH), to 90% or more.
Settings are:
G10: Low Alarm = -110%
G11: High Alarm = 80%
G12: Low Low Alarm = -110%
G13: High High Alarm = 90%
[%]
100
90
80
50
0Time (t)
SO1
SO2
AL
Select “H/L Alarm” for F10: SO1 Function
Select “HH/LL Alarm” for F11: SO2 Function
Select “Closed (On) Act” for F14: SO1/2 Active Mode
Select “Open (Off) Act” for G20: Alm Out Act Mode
Select “Yes” for G33: Alm-HH/LL
H
HH
LL Alarm set to -110% indicates
that the alarm is disabled.
HH
High High Alarm
High Alarm
Instantaneous flow rate
L Alarm set to
-110% indicates
that the alarm is
disabled.
F0612.EPS
[G14: H/L Alarm Hys] Setting of upper/lower alarm
value hysteresis width
This parameter sets the hysteresis width for upper and
lower alarm value, using a % value of the maximum
span.
Output Example
The hysteresis width is set to 5%.
Settings are:
G10: Low Alarm = 30%
G11: High Alarm = 80%
G12: Low Low Alarm = 20%
G13: High High Alarm = 90%
G14: H/L Alarm Hys = 5%
[%]
100
90
80
50
30
20
0Time (t)
SO1
SO2
AL
85%
75%
75%
HL
85%
HH
HH
85%25%
High High Alarm
High Alarm
Instantaneous flow rate
Low Alarm
35%
25%
Low Low Alarm
35%
25%
LL
LL
NOTE
• Although the same items can be selected using
the SO1 output terminal (selected for F10) and
the SO2 output terminal (selected for F11),
output is identical for both.
• Setting values of -110% and 110% are used to
disable corresponding functions; and accordingly, status output can be customized for
specific purposes.
Select “H/L Alarm” for F10: SO1 Function
Select “HH/LL Alarm” for F11: SO2 Function
Select “Closed (On) Act” for F14: SO1/2 Active Mode
Select “Open (Off) Act” for G20: Alm Out Act Mode
Select “Yes” for G33: Alm-HH/LL
[G20: Alm Out Act Mode] Setting of the active
mode for the alarm output.
This parameter selects the active mode for the alarm
output terminal.
SettingFunction
Closed (On) Act
Open (Off) Act
When an alarm occurs, alarm output
becomes "Closed (On)."
When an alarm occurs, alarm output
becomes "Open (Off)."
6-27
F0613.EPS
T0631.EPS
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
NOTE
NOTE
Example
Closed(On)
Open(Off)
Alarm occurs
Closed (On) Active
Alarm occurs
Open (Off) Active
F0614.EPS
[G21: 4-20mA Alarm Out] Setting of the current
output during an alarm occurrence
This parameter can be used to set up the current output
during alarm occurrence in advance.
SettingFunction
2.4mA or Less
4.0mA
Hold
21.6mA or More
Fixed at 2.4 mA or less
Fixed at 4 mA
Fixed current value when an alarm occured.
Fixed at 21.6 mA or more
T0632.EPS
[G22: 4-20mA Burn Out] Display of the current
output during a CPU failure
This parameter displays the current output direction for
a CPU failure (i.e., burnout). Note that communication
will not be possible if such a failure occurs.
With the standard specification, this is set to High and
25mA is output when a failure occurs. Low is setup for
optional code C1, and in such a case, 0mA is output
when a failure occurs.
NOTE
SettingFunction
No
Ye s
Not recognized as an alarm
Recognized as an alarm
T0634.EPS
[G32: Alm-Emp Pipe] Alarm recognition of “Empty
Pipe Alarm”
This parameter specifies whether the empty pipe
(flowtube is not filled with fluid) in process alarms will
be recognized as an alarm.
SettingFunction
No
Ye s
Not recognized as an alarm
Recognized as an alarm
T0635.EPS
[G33: Alm-HH/LL] Alarm recognition of “HH/LL
Alarm”
(Refer to the descriptions of G12 and G13 for more
details regarding HH and LL alarms.)
This parameter specifies whether HH/LL alarm in
process alarms will be recognized as an alarm.
SettingFunction
No
Ye s
Not recognized as an alarm
Recognized as an alarm
T0636.EPS
To set “HH/LL Alarm” as an alarm, it is necessary to
set “HH/LL Alarm” according to F10: SO1 Function
or F11: SO2 Function, and set G12: Low LowAlarm or G13: High High Alarm as well.
The current output direction for a CPU failure
(i.e., burnout) can be changed. Refer to selection 10.2.1: Setting of Burnout Switch.
[G30: Alm-Setting] Alarm recognition of “Setting
Alarm”
This parameter specifies whether the setting alarm will
be recognized as an alarm.
SettingFunction
No
Ye s
Not recognized as an alarm
Recognized as an alarm
T0633.EPS
[G31: Alm-Sig Over] Alarm recognition of “Signal
Overflow Alarm”
This parameter specifies whether the signal overflow in
process alarms will be recognized as an alarm. A signal
overflow occur when there is an error in the input
signal.
[G34: Alm-Adhesion] Alarm recognition of “Adhesion Alarm”
This parameter specifies whether the electrode adhesion alarm in process alarms will be recognized as an
alarm.
SettingFunction
No
Ye s
Not recognized as an alarm
Recognized as an alarm
T0636-1.EPS
The AXFA11 has three different types of alarm
(i.e., system alarms, process alarms and setting
alarms). For setting alarms and process alarms,
settings are made with G30, G31, G32, G33 andG34 to specify whether these will be recognized
as an alarm.
Refer to Section 6.5: Alarm Functions for more
details regarding the content of each alarm and
the effect of alarm recognition on output.
6-28
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
[G40: Operation Time] Display of operation time
This parameter is used to display the operation time.
The operation time is the total time that is counted
while the device works actually.
When the power supply is off, the operation time is not
counted.
For example, “1D23:45” indicates an operation time of
1 day, 23 hours, and 45 minutes.
IMPORTANT
Use this value as a rough guideline because the
operation time has an error.
[G41: Alm Record1] Alarm record1
This parameter is used to display the most-recent
alarm, and the alarms that can be displayed are as
follows.
Alarm Items
Item
: 16 space characters
(i.e., no display)
10 : uP Fault
11 : EEPROM Fault
12 : A/D(H) Fault
13 : A/D(L) Fault
14 : A/D(Z) Fault
15 : Coil Open
16 : EEPROM Dflt
18 : Power off
19 : Inst Pwr Fail
28 : WDT
30 : Sig Overflow
31 : Empty Pipe
33 : Adhesion Alm
NOTE
No issuing of alarms
Microprocessor (CPU) failure
EEPROM failure
A/D converter
A/D converter
A/D converter
Flowtube coil is open-circuit
EEPROM returns to default values
Power supply is off.
Instantaneous power fail for tens of
milliseconds.
After this fail is released, outputs reach
the previous value immediately.
The return from excessive instantaneous
noise.
After the noise is released, output return
the normal condition.
Input signal error
Flowtube is not filled with fluid
Insulation adhered to electrode
Description
failure
(at high frequency side)
failure
(at low frequency side)
failure
(detection of adhesion)
T0637.EPS
Records for “30: Sig Overflow” are kept only
when G31 specifies that this condition is to be
recognized as an alarm (i.e., “Yes” is selected).
Records for “31: Empty Pipe” are kept only when
G32 specifies that this condition is to be recognized as an alarm (i.e., “Yes” is selected).
Records for “33: Adhesion Alm” are kept only
when G34 specifies that this condition is to be
recognized as an alarm (i.e., “Yes” is selected).
[G42: Alm Record Time1] Display the operation
time of alarm record1
This parameter is used to display the operation time at
which the alarm indicated by G41: Alm Record1
was occurred. For example, “1D23:45” indicates that
an alarm was occurred at the operation time of 1 day,
23 hours, and 45 minutes.
IMPORTANT
Use these values as a reference because the
time of alarm records (1 to 4) have an error.
[G43: Alm Record2] Alarm record2
This parameter is used to display the second mostrecent alarm, and the alarms that can be displayed are
the same as those for G41: Alm Record1.
[G44: Alm Record Time2] Display the operation
time of alarm record2
This parameter is used to display the operation time at
which the alarm indicated by G43: Alm Record2
was occurred. For example, “1D23:45” indicates that
an alarm was occurred at the operation time of 1 day,
23 hours, and 45 minutes.
[G45: Alm Record3] Alarm record3
This parameter is used to display the third most-recent
alarm, and the alarms that can be displayed are the
same as those for G41: Alm Record1.
[G46: Alm Record Time3] Display the operation
time of alarm record3
This parameter is used to display the operation time at
which the alarm indicated by G45: Alm Record3
was occurred. For example, “1D23:45” indicates that
an alarm was occurred at the operation time of 1 day,
23 hours, and 45 minutes.
[G47: Alm Record4] Alarm record4
This parameter is used to display the fourth mostrecent alarm, and the alarms that can be displayed are
the same as those for G41: Alm Record1.
[G48: Alm Record Time4] Display the operation
time of alarm record4
This parameter is used to display the operation time at
which the alarm indicated by G47: Alm Record4
was occurred. For example, “1D23:45” indicates that
an alarm was occurred at the operation time of 1 day,
23 hours, and 45 minutes.
6-29
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
NOTE
NOTE
(7) Menu H: Display Setting items
Menu H contains setting items relevant to the display
unit.
[H10: Display Select1] Setting of the first line for
display unit
→ Refer to the description for parameter B40
This parameter selects the display content of the first
line for display unit.
[H11: Display Select2] Setting of the second line
for display unit
→ Refer to the description for parameter B41
This parameter selects the display content of the
second line for display unit.
[H12: Display Select3] Setting of the third line for
display unit
→ Refer to the description for parameter B42
This parameter selects the display content of the third
line for display unit.
[H20: Display Cycle] Setting of the display cycle
This parameter sets the cycle for the display-response
speed of display unit. Settings should be made in
accordance with the measurement environment by, for
example, setting a longer display cycle when using the
equipment in low temperatures.
Example: Situation where low cut is set to 10%
OutputSetting value
5.6mA(10%)
4mA
Setting range: 0 to 10%
Input
0.5% 0.5%
Hysteresis
fixed at 1%
F0615.EPS
[J11: 4-20mA Low Lmt] Setting of the low limit for
current output
This parameter is used to restrict low current portions
of current (4 to 20mA) output, and it is initially set to
-20%. Setting should be performed when a higher
value is required for the lower limit.
The indications of the instantaneous flow rates (%,
Actual instantaneous flow rate, mA, Bar graph) on the
display unit are the same action.
Example: Situation where low limit is set to 10%
Output
5.6mA(10%)
4mA
Input
F0655.EPS
[H30: Language] Selection of language used for the
display unit.
→ Refer to the description for parameter B10
This parameter can be used to select the language for
display unit.
(8) Menu J: Auxiliary Function Setting items
Menu J contains setting items such as the flow direction, rate limits, and current output limits.
[J10: 4-20mA Low Cut] Setting of the low-cut range
for current output.
This parameter is used to force current output to
0%(i.e., 4mA) in the vicinity of 0% and setting for the
current (4 to 20mA) output low cut is made using a
percentage of the smallest flow rate span. However, the
low cut function will be terminated if this parameter is
set to 0%.
The indications of the instantaneous flow rates (%,
Actual instantaneous flow rate, mA, Bar graph) on the
display unit are the same action.
If “2.4mA or less” has been set for G21:4-20mA
Alarm Out, 2.4mA or less will be output upon an
alarm occurrence, regardless of the low limit
setting.
• If the setting value for the low limit is not less
than the high limit value (as set using J12: 4-20mA High Lmt), the setting alarm “54: 4-20
Lmt Err” will be displayed.
• This parameter has no effect on pulse output
or the totalization function.
[J12: 4-20mA High Lmt] Setting of the high limit
for current output
This parameter is used to restrict high current portions
of current (4 to 20mA) output, and it is initially set to
120%. Setting should be performed when a lower value
is required for the higher limit.
The indications of the instantaneous flow rates (%,
Actual instantaneous flow rate, mA, Bar graph) on the
display unit are the same action.
6-30
IM 01E20C01-01E
Example: Situation where high limit is set to 90%
NOTE
Output
18.4mA(90%)
4mA
Input
F0656.EPS
NOTE
If “21.6mA or more” has been set for G21:4-20
mA Alarm Out, 21.6mA or more will be output
upon an alarm occurrence, regardless of the
high limit setting.
[J20: Flow Direction] Setting of the flow direction
Upon shipment from the manufacturing plant, the
system is setup such that flow in the same direction, as
shown by the direction of the arrow mark on the
flowtube, will be measured as forward flow; however,
this parameter can be used to set “Reverse” so that
flow in the opposite direction to the arrow mark will be
treated as forward.
Note: This function does not apply to measurement in
both the forward and reverse directions, although this can be setup using by selecting
“Fwd/Rev Ranges” from either F10: SO1Function or .
SettingFunction
Forward
Reverse
Forward direction corresponds with arrow mark
Forward direction is opposite to arrow mark
T0638.EPS
6. PARAMETER DESCRIPTION
Determining rate limit value and dead time
Rate limit value:
Determines the level for output
fluctuation cutoff. For example,
if this is set to 2%, noise above
2% will be eliminated as shown
in the diagram.
Dead time (T
This is to be determined using the
output fluctuation width. If noise
exceeds the dead time as shown in
the diagram below, the dead time
should be made longer.
2%
2%
0
):
T
0
T
0
F0616.EPS
Signal processing method:
A fixed upper and lower limit value is setup with
respect to the primary delay response value for the
flow rate value obtained during the previous sampling, and if the currently sampled flow rate is
outside these limits, then the corresponding limit is
adopted as the current flow rate value. In addition, if
signals which breach the limits in the same direction
occur over multiple samples (i.e., within the dead
time), it is concluded that the corresponding signal is
a flow rate signal.
Example 1: Step input
Input: 0 to 10%
Damping time constant: 3 s
Dead time: 3 s
10
Step signal
%
Rate limit value: 1%
[J21: Rate Limit] Setting of the rate limit value
This parameter is used in situations where sudden
noise cannot be eliminated by increasing the
damping time constant.
In situations where step signals or sudden noise
signals caused by slurries or the like are entered, this
parameter is used to set the standard for determining
whether an input corresponds to a flow measurement
or noise. Specifically, this determination is made
using upper and lower rate limits and using the dead
time.
Rate limit values are set using a percentage of the
smallest range. The range of deviation per one
calculation cycle should be input.
[J22: Dead Time] Setting of dead time
This parameter sets the time for application of the rate
limit, and if a value of 0 is set, the rate limit function
will be terminated.
Flow rate value
after rate limit
processing
(a)
1
%
(b)
100225
Dead time: 3 s
(c)
Flow rate value after
(d)
damping
Number of signal samples
(1) In comparison with the previous value at (a), it is
determined that the signal is in excess of the rate
limit value and the response becomes 1%. However, the actual output applies damping, and
therefore the output turns out to be as indicated by
the solid line.
(2) Subsequent flow values within the dead time zone
correspond to signals of post-damping flow value
+ rate limit value (1%).
(3) Since input signals do not return to within the rate
limit value during the dead time, it is determined at
(c) that this signal is a flow rate signal.
6-31
IM 01E20C01-01E
63.2
F0617.EPS
%
6. PARAMETER DESCRIPTION
(4) The output signal becomes a damped curve and
compliance with the step signal begins.
Three seconds after determination of a flow rate
signal in the above figure, a level of 63.2% is
reached.
Example 2: Slurry noise
Input: 0 to 10%
Slurry noise
Flow rate value
after rate limit
+1
%
-1
%
Flow rate value after damping
Dead time: 1 s
processing
Damping time constant: 1 s
Dead time: 1 s
Rate limit value: 1%
In the figure on the left, it is
determined that the slurries
noise signal is not a flow rate
signal.
Time
F0618.EPS
[J23: Pulsing Flow] Selection of pulsing flow support
In a situation where pulsating flow causes error in the
average flow value, due to the application of a plunger
pump, this parameter provides functionality whereby
calculation is controlled and variations in flow rate are
followed.
SettingFunction
No
Ye s
Normal
Support for pulsing flow
T0639.EPS
[J24: T/P Damp Select] Setting of damping operation
This parameter is used to select that the flow rate value
obtained through damping calculation for totalization
and pulse output or the instantaneous flow rate value
(no damping) for totalization and pulse output.
SettingFunction
Damp
No Damp
Damping
No damping
T0640.EPS
[J30: Power Synch] Setting of power synchroniza-
tion
This parameter sellects whether or not the internal
frequency is to be synchronized with that of the power
supply.
SettingFunction
No
Ye s
Not synchronized
Synchronized
T0641.EPS
IMPORTANT
In situations where a DC power supply is used
for converters, set the local commercial power
frequency in area where the converter is installed (size 2.5mm to 1000mm). Set “No” for
J30: Power Synch and the local commercial
power frequency for J31: Power Frequency.
IMPORTANT
In situations where AXFA11 is combined with a
remote flowtube the size of 1100 mm (44 in.) to
2600 mm (104 in.), set the fixed frequency
(49.00 Hz) in case that either AC or DC power
supplies is used for converters.
Set “No” for J30: Power Synch and “49.00” for
J31: Power Frequency.
Following settings are necessary by power supply and
by flow tube size.
Power Supply Code 1
(100 to 240 V AC or 100 to 120 V DC)
Size 2.5 mm (0.1 in.)
to 1000 mm (40 in.)
AC power
supply
DC power
supply
Power Supply Code 2 (24 V AC/DC)
AC power
supply
DC power
supply
Power synchronous (Yes)
J30
J31
No setting
Power asynchronous (No)Power asynchronous (No)
J30
J31
Local commercial power
frequency
Size 2.5 mm (0.1 in.)
to 1000 mm (40 in.)
Power asynchronous (No)
J30
J31
Local commercial power
frequency
Power asynchronous (No)Power asynchronous (No)
J30
J31
Local commercial power
frequency
Size 1100 mm (44 in.)
to 2600 mm (104 in.)
Power asynchronous (No)
49.00 Hz
49.00 Hz
T0642-1.EPS
Size 1100 mm (44 in.)
to 2600 mm (104 in.)
Power asynchronous (No)
49.00 Hz
49.00 Hz
T0642-2.EPS
[J40: Memo 1] Setting of memo 1
[J41: Memo 2] Setting of memo 2
[J42: Memo 3] Setting of memo 3
These parameters are used with the memo function,
and up to 16 characters can be set for each.
[J31: Power Frequency] Setting of power frequency
When “Yes” (i.e., in synchrony) has been selected for
J30: Power Synch, this parameter is used to display
the power supply frequency. If “No” (i.e., not synchronized) has been selected, the power supply frequency is
to be specified.
[J50: Software Rev No] Display of software
revision
This parameter is used to display the software’s
revision number.
6-32
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
(9) Menu K: Diagnostic Function Setting items
Menu K contains items that are relevant to the diagnosis of insulation adhesion to the electrode.
[K10: Adhesion Check] Setting of adhesion
diagnostic function
This parameter selects whether or not the adhesion
diagnostic function will be carried out.
SettingFunction
Halt the adhesion diagnostic functionNo
Ye s
NOTE
Carry out the adhesion diagnostic function
T0642.EPS
Adhesion Diagnostic Function
This function diagnose adhesion using elec-
trode resistance values.
When “Adhesion check” has been set for B41/
H11: Display Select 2 or B42/H12: Display
Select 3, the diagnose adhesion is indicated
on the display unit using four different levels.
If the judgment value for Level 3 is exceeded,
a warning is displayed; and if the value for
Level 4 is exceeded, an alarm is displayed.
Available conductivity for this function is
limited to:
Nominal size 10 mm or smaller: 30 S/cm
Nominal size 15 mm or larger: 10 S/cm
Make sure to use the adhesion diagnostic
function with the greater conductivity than the
above mentioned value.
[K13: Adhesion Level3] Setting the resistance value
for adhesion diagnostic level 3.
This parameter sets the resistance value (in M ohm) for
judgment of Level 3.
*: The warning 80: Adhesion Wng is displayed
when the adhension level reaches Level 3.
*: If “Warning Output” has been selected for F10:
SO1 Function or F11: SO2 Function, then
status output will be performed when the adhesion
level reaches Level 3.
[K14: Adhesion Level4] Setting the resistance value
for adhesion diagnostic level 4.
This parameter sets the resistance value (in M ohm) for
judgment of Level 4.
*: The process alarm 33: Adhesion Alm is displayed
when the adhension level reaches Level 4.
*: Alarm output will be performed if “Yes” has been
selected for G34: Alm-Adhesion.
[K15: Adh Measure Value] Displays the resistance
value for adhesion diagnose
This parameter displays the value measured using the
adhesion diagnostic function (in M ohm).
When “No” is selected for K10: Adhesion Check,
this parameter displays the indetermination value.
(10) Menu M: Automatic Zero Adjustment
Function Setting items
Menu M contains items that are relevant to automatic
adjustment of the zero point.
Warning
Level: 2Level: 1
Level: 3
Alarm
Level: 4
F0619.EPS
[K11: Adhesion Level1] Setting the resistance value
for adhesion level 1.
This parameter sets the resistance value (in M ohm) for
judgment of Level 1.
[K12: Adhesion Level2] Setting the resistnce value
for adhesion diagnostic level 2.
This parameter sets the resistance value (in M ohm) for
judgment of Level 2.
[M10: Auto Zero Exe] Execution of automatic zero
adjustment function
→ Refer to the description of parameter B50.
[M11: Magflow Zero] Results of automatic zero
adjustment
This parameter is used to display the results obtained
from B50/M10: Auto Zero Exe. Specifically, the
correction value displayed, and it is also possible to
directly enter correction value.
(11) Menu N: Loop Test Setting items
Menu N contains items that are relevant to loop testing.
[N10: Test Mode] Setting for loop test execution
SettingFunction
Normal
Test
6-33
No execution of loop testing
Loop testing is started
IM 01E20C01-01E
T0644.EPS
6. PARAMETER DESCRIPTION
NOTE
IMPORTANT
(1) Test output has priority over flow rate mea-
surement signals. When carrying out flow
rate measurements, be sure to always return
to “Normal”.
(2) Upon entry to the Test Mode, all output
terminals will simultaneously adopt test
condition.
(3) “Normal” will be restored when the power is
turned off or when 30 minutes have elapsed
since entry to Test Mode.
(4) In Test Mode, the warning 83: Fix Cur Wng
will be displayed as a warning message. (For
more details, refer to Section 6.5 Alarm
Functions.)
[N11: Test Output Value] Setting for test output
values
During loop testing, current output (4 to 20mA),
totalization, and pulse will be output in accordance
with this parameter’s setting, and values can be set
when “Test” has been selected for N10: Test Mode.
With multiple ranges or when performing forward/
reverse flow measurements, setting should be done
using a percentage of the maximum range.
[N20: Test SO1] Setting for SO1 output terminal
condition during testing
This parameter sets the condition of the SO1 status
output terminal during loop testing. Setting is possible
when “Test” has been selected for N10: Test Mode.
SettingFunction
Open (Off)
Closed (On)
SO1 output terminal in Open (Off) condition
SO1 output terminal in Closed (On) condition
T0645.EPS
[N21: Test SO2] Setting for SO2 output terminal
condition during testing
This parameter sets the condition of the SO2 status
output terminal during loop testing. Setting is possible
when “Test” has been selected for N10: Test Mode.
SettingFunction
Open (Off)
Closed (On)
SO2 output terminal in Open (Off) condition
SO2 output terminal in Closed (On) condition
T0645-2.EPS
[N22: Test Alarm Out] Setting for alarm terminal
condition during testing
This parameter sets the condition of the alarm output
terminal (AL) during loop testing.
SettingFunction
Open (Off)
Closed (On)
AL output terminal in Open (Off) condition
AL output terminal in Closed (On) condition
T0645-3.EPS
[N23: Test SI1] Display for SI1 status input terminal
condition during testing
This parameter is used to display the condition of the
SI1 status input terminal during loop testing.
SettingFunction
Open
Short
SI1 input terminal in Open condition
SI1 input terminal in Short condition
T0646.EPS
[N24: Test SI2] Display for SI2 status input terminal
condition during testing
This parameter is used to display the condition of the
SI2 status input terminal during testing.
SettingFunction
Open
Short
SI2 input terminal in Open condition
SI2 input terminal in Short condition
T0646-2.EPS
(12) Menu P: Parameter Protection items
Menu P contains items that are relevant to write
protection and passwords.
[P10: Key Code] Parameter of the display restriction
This parameter restricts access to the Service Mode.
Write Protect function
The parameters P20 through P23 are set
when using the write protect function. Specifically, this function responds to a hardware
switch or the setting of a software password,
and it protects parameters from being overwritten.
If the hardware switch is set to “Protect”, it will
not be possible to overwrite parameters;
furthermore, this condition will be maintained
until the switch is set to “Enable”.
For more details regarding hardware switch
settings, refer to Section 10.2.3: Setting of
Write Protect Switch.
6-34
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
NOTE
[P20: Write Protect] Status indication for write
protection
This parameter is used to indicate whether or not write
protection is currently on.
Default setting (Enable)
P20:Write Protect
No
F0620.EPS
Write protection (Protect)
P20:Write Protect
Yes
F0621.EPS
[P21: Enable Wrt Passwd] Setting of password to
release the write protection function
When the correct password is input, write protection
will be released for a period of 10 minutes; furthermore, this period will be extended by a further 10
minutes each time a parameter is overwritten.
P21:Enable Wrt Passwd
********
To change a password, first of all use the password
originally set with P21: Enable Wrt Passwd to
release the write protect function, and then set the new
password. Alternatively, if it is desired to return to the
condition where no password is set, enter a string of 8
spaces.
[P23: Software Seal] Display the software seal
When the joker password has been used to release
write protection, this parameter displays “Break”, and
when protection is cancelled using the password set
using P22: New Password, it returns to “Keep”.
If you should forget your password, the joker
password can be used to temporarily release
write protection function. To obtain the joker
password, please contact your nearest
YOKOGAWA sales office.
F0622.EPS
The cursor will flash when entering Parameter Replacement Mode, and the password set with P22: New
Password should be input at this time.
[P22: New Password] Setting of a new password
This parameter sets the password required for the
release of write protection. When set, it will be
possible to make write protect settings on the software
side.
Default setting
P22:New Password
F0623.EPS
The default setting for this parameter is a string of 8
spaces (i.e., Enable), and thus, the password field will
be empty. When the cursor is flashing, the password
should be input. Press the SET key twice to confirm
the password. The display will then change to
“********”.
After password setting
P22:New Password
********
F0624.EPS
6-35
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
6.5Alarm Functions
6.5.1Alarm Levels
Alarms are classified into the following four different
types based on level.
AlarmDescriptionLevel
Major
System
breakdown
alarm
Intermediate
Process
level break-
alarm
down
Minor
Setting
breakdown
alarm
Warning
Warning
When an alarm has been occurred, the corresponding
alarm name, description, and suitable countermeasure
will be displayed on the display unit. The normal
Display Mode and Alarm Mode may be displayed
alternatively. When a warning has been issued, the
corresponding content will be shown in the third line in
the Display Mode.
Device breakdown or inability to obtain correct
measurements. Replacement will be required.
Device is normal but process-related errors
make correct measurement impossible.
Maintenance or the like will be required.
Device is normal but errors have been made
in the setting of parameters. Functions not
related to the incorrect settings are operating
normally.
The incorrect settings must be corrected.
Device and measurements are normal but a
warning is occurred.
T0647.EPS
6.5.2Alarm Selection
The display and output differs depending on the alarm
levels. Certain types of alarm may or may not be
recognized as alarms, according to the settings of
certain parameters. The parameters that are relevant to
this function as follows.
[G20: Alm Out Act Mode] Setting of the active
mode for the alarm output
[G21: 4-20mA Alarm Out] Setting of the current
output during an alarm occurring.
[G30: Alm-Setting] Alarm recognition of “Setting
Alarm”
[G31: Alm-Sig Over] Alarm recognition of “Signal
Overflow Alarm”
[G32: Alm-Emp Pipe] Alarm recognition of “Empty
Pipe alarm”
[G33: Alm-HH/LL] Alarm recognition of “HH/LL
Alarm” (Refer to the descriptions of G12 and G13 for
more details regarding HH and LL alarms.)
[G34: Alm-Adhesion] Alarm recognition of “Adhesion Alarm”
[G41: Alm Record1] Alarm record1
Upon occurrence of an alarm
(example: a System alarm)
Upon issuance of a warning
(a description appears in the 3rd line only
when a warning is occurred)
Alarm name
Description
Countermeasure
message
Description of
warning
F0625.EPS
[G43: Alm Record2] Alarm record2
[G45: Alm Record3] Alarm record3
[G47: Alm Record4] Alarm record4
6-36
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
(1) Display and output condition for system alarms
Alarm description
NormalNormal
P Fault
10
EEPROM Fault
11
A/D(H) Fault
12
A/D(L) Fault
13
A/D(Z) Fault
14
Coil Open
15
EEPROM Dflt
16
Microprocessor(CPU)failure
EEPROM failure
A/D converter failure
Flowtube coil is open-circuit
EEPROM default values
Alarm
output
Closed
(On)
Open
(Off)
Open
(Off)
Note: • Operations are performed in accordance with above table, when “Open (Off) Act” is set for G20: Alm
Out Act Mode.
• 4-20 mA output upon the occurrence of an alarm will be fixed at the value selected with G21: 4-20mA Alarm Out.
*The output value is performed in accordance with the setting of the burnout switch. For information about
this switch, see Section 10.2.1.
4-20 mA
output
0mA or
(*)
25mA
Fixed
(G21 selection)
Totali-
Pulse Display unit
zation
Normal Normal
Indetermination
Stopped StoppedRecorded
Display ModeNo
Stopped Indetermination
Alarm Mode
(display of system
Alarm
record
Indetermination
alarm message)
T0648.EPS
(2) Display and output condition for process alarms
Selection
(parameter
number)
YES
(G31)
NO
(G31)
YES
(G32)
NO
(G32)
YES
(G33)
NO
(G33)
YES
(G34)
NO
(G34)
Alarm
output
Open
(Off)
Closed
(On)
Open
(Off)
Closed
(On)
Open
(Off)
Closed
(On)
Open
(Off)
Closed
(On)
4-20 mA
output
Fixed
Continu-
ous (*)
Fixed
Continu-
ous (*)
Normal
operation
Fixed Stopped Stopped
Continu-
ous (*)
30
Sig Overflow
31
Empty Pipe
32
HH/LL Alm
33
Adhesion Alm
Alarm
description
Input signal error
Flowtube is not
filled with fluid
HH/LL Alarm
Electrode
adhesion alarm
Note: • Operations are performed in accordance with above table, when “Open (Off) Act” is set for G20: Alm
Out Act Mode.
• 4-20 mA output upon the occurrence of an alarm will be fixed at the value selected with G21: 4-20mA Alarm Out.
*: Although outputs are continuous, output values are not guaranteed.
Totalization
Stopped
Continu-
ous (*)
Stopped
Continu-
ous (*)
Normal
operation
Continu-
ous (*)
Pulse
output
Stopped
Continu-
ous (*)
Stopped
Continu-
ous (*)
Normal
operation
Continu-
ous (*)
Display
unit
Alarm Mode
(Message)
Display Mode
Alarm Mode
(Message)
Display Mode
Alarm Mode
(Message)
Display Mode
Alarm Mode
(Message)
Display Mode
Alarm
record
Recorded
No
Recorded
No
No
Recorded
No
T0649.EPS
6-37
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
(3) Display and output condition for setting alarm occurrences
50
Span > 10m/s
51
Span < 0.1m/s
52
TTL>10000p/s
53
TTL<0.0001p/s
54
4-20 Lmt Err
55
Multi Rng Err
56
H/L HH/LL Set
57
Dens Set Err
60
PLS >10000p/s
61
PLS > 5000p/s
62
PLS > 1000p/s
63
PLS > 500p/s
64
PLS > 25p/s
65
PLS > 15p/s
66
PLS > 10p/s
67
PLS > 5p/s
70
PLS<0.0001p/s
71
Meas Mod Set
72
Size Set Err
73
Adh Set Err
Occurring of
any alarm from
50 through 73
Alarm description
Span flow velocity setting
is 11 m/s or more
Span flow velocity setting
is 0.05 m/s or less
Totalization rate is 11000
pps or more
Totalization rate is
0.00005 pps or less
The condition [4-20 low
limit (J11) < 4-20 high
limit (J12)] is not satisfied.
The condition [No. 1 range
< No. 2 range < No. 3 range
< No. 4 range] is not
satisfied for multiple ranges.
The condition [High Alarm
(G11) – Low Alarm (G10) >
H/L Alarm Hys (G14)] or the
condition [High High Alarm
(G13) – Low Low Alarm
(G12) > H/L Alarm Hys
(G14)] is not satisfied.
Mass units have been
selected for Base Flow Unit
(C40) but density is set to 0.
Pulse rate is 11000 pps
or more with 50% duty
selection.
Pulse rate is 10000 pps
or more with 0.05 ms
selection.
Pulse rate is 5000 pps or
more with 0.1 ms selection.
Pulse rate is 1000 pps or
more with 0.5 ms selection.
Pulse rate is 500 pps or
more with 1 ms selection.
Pulse rate is 25 pps or
more with 20 ms selection.
Pulse rate is 15 pps or
more with 33 ms selection.
Pulse rate is 10 pps or
more with 50 ms selection.
Pulse rate is 5 pps or
more with 100 ms selection.
Pulse rate is 0.00005 pps
or less.
Measure Mode (C20) is
set to Enhanced DF
without selecting an
optional code HF1 or
HF2.
A value of 3000.1 mm or
more is set for Nominal
Size (C32).
The condition [Level:1<
Level:2<Level:3<Level:4]
is not satisfied for
adhesion diagnostic level.
–
Selection
(parameter
number)
(G30)
(G30)
NO
YES
Alarm
4-20 mA
output
output
Closed
Fixed Stopped Stopped
(On)
Normal
Closed
operation
(On)
Closed
Fixed
(On)
Closed
Fixed Stopped Stopped
(On)
Closed
Normal
(On)
operation
Closed
Fixed Stopped Stopped
(On)
Closed
Normal
(On)
operation
Closed
Fixed Stopped Stopped
(On)
Closed
Fixed Stopped Stopped
(On)
Normal
Closed
operation
(On)
Open
Fixed Stopped Stopped
(Off)
Totalization
Stopped
Normal
operation
Normal
operation
Normal
operation
Normal
operation
Pulse
output
Normal
operation
Normal
operation
Normal
operation
Stopped
Normal
operation
Display
Alarm Mode
(message)
Alarm Mode
(message)
Alarm Mode
(message)
Alarm Mode
(message)
Alarm Mode
(message)
Alarm Mode
(message)
Alarm Mode
(message)
Alarm Mode
(message)
Alarm Mode
(message)
Alarm Mode
(message)
Alarm Mode
(message)
Note: • Operations are performed in accordance with above table, when “Open (Off) Act” is set for G20: Alm
Out Act Mode.
• 4-20 mA output upon the occurrence of an alarm will be fixed at the value selected with G21: 4-20mA Alarm Out.
unit
Alarm
record
T0650.EPS
No
No
No
No
No
No
No
No
No
No
No
6-38
IM 01E20C01-01E
6.5.3Alarms & Warning Messages
System Alarms (Device breakdown or inability to obtain correct measurements.)
Display unit/BRAIN
(60) content
10:uP Fault
11:EEPROM Fault
12:A/D(H) Fault
13:A/D(L) Fault
14:A/D(Z) Fault
15:Coil Open
16:EEPROM Dflt
Process Alarms (Device is normal but process-related errors make correct measurement impossible.)
Display unit/BRAIN
(60) content
30:Sig Overflow
31:Empty Pipe
32:HH/LL Alm
33:Adhesion Alm
Alarm countermeasure
message on display unit
Contact nearest office or
service center
Contact nearest office or
service center
Contact nearest office or
service center
Contact nearest office or
service center
Contact nearest office or
service center
Cut the power and check coil
& EX cable
Contact nearest office or
service center
Alarm countermeasure
message on display unit
Check signal cable
and grounding
Fill flow tube with
fluid
Check the flow rate and
setting value
Clean electrodes
6. PARAMETER DESCRIPTION
Alarm descriptionCountermeasure
Microprocessor (CPU) failure
EEPROM failure
Contact your nearest Yokogawa office or
service center.
A/D converter failure
Turn off the AXFA11 power supply and
Flowtube coil is open-circuit
EEPROM default values
Alarm descriptionCountermeasure
Input signal error
Flowtube is not filled with
fluid
Flow rate alarm for greater
than High-High limit value
or less than Low-Low limit
value.
Electrode adhesion alarmClean the electrodes.
examine the excitation cable for breakage.
If there is no breakage, contact your nearest
Yo kogawa office or service center.
Contact your nearest Yokogawa office or
service center.
Carry out an investigation as follows:
• Check the signal cable for breakage.
• Check for contact between signal cable,
power cable, and excitation cable.
• Check for stray currents in the fluid.
• Check the grounding.
Fill the flowtube with fluid.
Check the flow rate and setting value of
High High limit and Low Low limit.
T0651.EPS
T0652.EPS
6-39
IM 01E20C01-01E
6. PARAMETER DESCRIPTION
Setting Alarms (Device is normal but errors have been made in the setting of parameters.)
Display unit/BRAIN
(60) content
50:Span > 10m/s
51:Span < 0.1m/s
52:TTL>10000p/s
53:TTL<0.0001p/s
54:4-20 Lmt Err
55:Multi Rng Err
56:H/L HH/LL Set
57:Dens Set Err
60:PLS > 10000p/s
61:PLS > 5000p/s
62:PLS > 1000p/s
63:PLS > 500p/s
64:PLS > 25p/s
65:PLS > 15p/s
66:PLS > 10p/s
67:PLS > 5p/s
70:PLS<0.0001p/s
71:Meas Mod SetCheck parameter C20
72: Size Set ErrCheck parameter C32
73: Adh Set Err
Alarm countermeasure
message on display unit
Check parameter C40, C41,
and C42
Check parameter C40, C41,
and C42
Check parameter D10 and
D11
Check parameter D10 and
D11
Check parameter J11and J12
Check parameter F30 to F36
Check parameter G10 to G14
Check parameter C40, C45,
and C46
Check parameter E10, E11,
and E12
Check parameter E10, E11,
and E12
Check parameter E10, E11,
and E12
Check parameter E10, E11,
and E12
Check parameter E10, E11,
and E12
Check parameter E10, E11,
and E12
Check parameter E10, E11,
and E12
Check parameter E10, E11,
and E12
Check parameter E10, E11,
and E12
Check parameter K11 to K14
Alarm descriptionCountermeasure
Span flow velocity setting is 11 m/s or
more
Span flow velocity setting is 0.05 m/s or
less
Totalization rate is 11000 pps or more
Totalization rate is 0.00005 pps or less
The condition [4-20 low limit (J11) <
4-20 high limit (J12)] is not satisfied.
The condition [No. 1 range < No. 2
range < No. 3 range < No. 4 range] is
not satisfied for multiple ranges.
The condition [High Alarm (G11) –
Low Alarm (G10) > H/L Alarm Hys
(G14)] or the condition [High High
Alarm (G13) – Low Low Alarm (G12)
> H/L Alarm Hys (G14)] is not satisfied.
Mass units have been selected for Base
Flow Unit (C40) but density is set to 0.
Pulse rate is 11000 pps or more with
50% duty selection.
Pulse rate is 10000 pps or more with
0.05 ms selection.
Pulse rate is 5000 pps or more with 0.1
ms selection.
Pulse rate is 1000 pps or more with 0.5
ms selection.
Pulse rate is 500 pps or more with 1 ms
selection.
Pulse rate is 25 pps or more with 20 ms
selection.
Pulse rate is 15 pps or more with 33 ms
selection.
Pulse rate is 10 pps or more with 50 ms
selection.
Pulse rate is 5 pps or more with 100 ms
selection.
Pulse rate is 0.00005 pps or less.
Measure Mode (C20) is set to Enhanced
DF without selecting an optional
code HF1 or HF2.
A value of 3000.1 mm or more is set for
Nominal Size (C32).
The condition in Adhesion detection
level, Level:1<Level:2<Level:3<Level:4
is not satisfied.
Check whether parameters
C40, C41, and C42 are correct.
In case that multiple range or
forward and reverse flow
measurement functions is used,
check whether parameters F30
through F36 are correct.
Check whether parameters
D10 and D11 are correct.
Check whether parameters J11
and J12 are correct.
Check whether parameters
F30 through F36 are correct.
Check whether parameters
G10 through G14 are correct.
Check whether parameters
C40, C45, and C46 are correct.
Check whether parameters
E10, E11, and E12 are correct.
Check whether parameter C20
is correct.
Check whether parameter C32
is correct.
Check whether parameters K11,
K12, K13 and K14 are correct.
T0653.EPS
Setting Alarms (Device and measurements are normal but a warning is issued.)
Display unit/BRAIN
(60) content
80:Adhesion Wng
82:Auto Zero Wng
83:Fix Cur Wng–The current value is fixed.
84:Disp Over Wng
(only for display unit)
90:Disp SW Wng
(only for display unit)
Alarm countermeasure
message on display unit
–
–
–
–
Alarm descriptionCountermeasure
Slight adhesion to
electrodes.
Results of automatic zero
adjustment are higher than
the rated values.
Overflow in the display digits during
instantaneous flow rate display.
Display unit switches are not
operating.
Clean and check the electrodes.
Refer to parameter K13.
Carry out adjustment as follows:
• Check if the flowtube is filled with
fluid.
• Check if the flow velocity is
completely zero.
• Check the condition of grounding.
Confirm whether the flow rate is in
excess of the upper limit (108%) or
below the lower limit (-8%), or
whether upon entry to the Test Mode
or not.
Check whether parameter C43 is
correct.
Investigate whether the display unit
cover is fitted or whether the cover’s
glass surface is dirty.
6-40
T0654.EPS
IM 01E20C01-01E
7. OPERATION VIA BRAIN TERMINAL (BT200)
7.
OPERATION VIA BRAIN TERMINAL (BT200)
7.1BT200 Basic Operations
7.1.1Key Layout and Display
LCD
(21 characters ⴛ 8 lines)
Movement keys
• Select items
• Move the cursor
• Change pages
Power
ON/OFF key
Alphanumeric keys
• Enter digits
• Enter letters in
combination with
the shift keys.
Figure 7.1 Key Layout
Menu Screen
MENU
A:DISPLAY
B:EASY SETUP
C:BASIC SETUP
D:TOTAL SET
Screen titles
E:PULSE SET
F:STATUS FUNC
HOME SETADJ
Parameter Screen
PARAMETER
A10:FLOW RATE(%)
80.0
A20:FLOW RATE
100m3/h
A21:FLW RATE(mA)
1417
DATA DIAG PRNT ESC
If BATTERY is displayed on
the screen, it indicates that the
battery voltage has dropped.
BATTERY
%
Function keys
Used to execute
commands
displayed at the
bottom of the
screen.
(See Table 7.1.)
ENT key
• Enters selected
items
• Sets up input data
on the connected
instrument.
• Runs BT200
functions.
Shift keys
Messages
Menu choices
(up to six are
displayed)
ESC
Function commands
Parameters
(up to three are
displayed)
Function commands
F0701.EPS
F0702.EPS
7.1.2Key Descriptions
(1) Alphanumeric keys and shift keys
You can use the alphanumeric keys in conjunction with
the shift keys to enter letters, digits, and symbols.
Alphanumeric keys
Shift keys
a) Entering digits, symbols, and spaces [i.e., 0-
9, period (.), hyphen (-), underscore (_)]
Simply press the required alphanumeric key.
Entry
–4
0.3
1 –9
Key-in sequence
b) Entering letters (i.e., A through Z)
Press alphanumeric key following a shift key to
enter the letter shown on the same side as the shift
key. The shift key must be pressed for each letter
being input.
Left-side letter on an
alphanumeric key
EntryKey-in sequence
W
IC
J. B
Right-side letter on
an alphanumeric key
F0703.EPS
T0701.EPS
F0704.EPS
T0702.EPS
7-1
IM 01E20C01-01E
7. OPERATION VIA BRAIN TERMINAL (BT200)
Use the function key [F2]
CAPS
to select between
uppercase and lowercase (for letters only). The case
toggles between uppercase and lowercase each time
you press [F2] CAPS.
Entering uppercase
CODE CAPSCLRESC
Entry
Boy
( B )( y )( o )
Use the function key [F1]
Entering lowercase
CODE capsCLRESC
Key-in Sequence
to lower case
CODE
to enter symbols. The
F0807.EPS
following symbols will appear in sequence, one at a
time, at the cursor each time you press [F1] CODE:
/ . – , + * ) ( ’ & % $ # ” !
To enter characters next to these symbols, press [ > ] to
move the cursor.
Entry
symbol command
l/m
Key-in Sequence
( / )
( m )( I )
T0803.EPS
(2) Function Keys
The functions of the function keys depend on the
function commands on display.
Function Command List
CommandFunction
ADJDisplays the ADJ menu
CAPS/caps
CODE
CLR
DATA
DEL
DIAG
ESC
HOME
NO
OK
PARM
SET
SLOT
UTIL
*COPY
*FEED
*LIST
*PON/POFF
*PRNT
*GO
*STOP
* Available on BT200-P00 (with printer).
Selects uppercase or lowercase
Selects symbols
Erases input data or deletes all data
Updates parameter data
Deletes one character
Calls the self-check panel
Returns to the most recent display
Displays the menu panel
Quits setup and returns to the previous display
Proceeds to the next panel
Enters the parameter number setup mode
Displays the SET menu
Returns to the slot selection panel
Calls the utility panel
Prints out parameters on display
Paper feed
Lists all parameters in the menu
Automatic printout mode on or off
Changes to the print mode
Starts printing
Cancels printing
T0804.EPS
MENU
A:DISPLAY
B:EASY SETUP
C:BASIC SETUP
D:TOTAL SET
E:PULSE SET
F:STATUS FUNC
HOME SETADJESC
Function commands
Function keys
F0808.EPS
7-2
IM 01E20C01-01E
7. OPERATION VIA BRAIN TERMINAL (BT200)
NOTE
7.2AXFA11 Operation
Using a BT200
This section describes procedures for setting parameters
using a BRAIN Terminal (BT200). For more details
regarding AXFA11 functions, refer to Chapter 6:
Parameter Description; and for more details regarding
BT200 operation methods, refer to the BT200 User’s
Manual (IM 01C00A11-01E).
7.2.1BT200 Connection
Connection to a 4 to 20 mA DC signal line
The communication signal of the AXFA11 is superimposed onto the 4 to 20mA DC analog signals to be
transmitted.
AXFA11 terminals
IMPORTANT
Restrictions exist with regard to the distance
over which communication is possible. (See
Chapter 11: Outline.)
IMPORTANT
If the power of flowmeter is turned off within 30
seconds after parameters have been set, these
settings will be canceled. Accordingly, please
keep the power on for at least 30 seconds after
setting parameters.
IMPORTANT
After approximately 5 minutes of inactivity, the
Auto Power-Off function will operate to turn your
BT200 off.
4 to 20mA
DC signal
Relaying
terminals
I
I
Figure 7.2Connecting the BT200
transmission
line
BT200BT200BT200
Control room
Te r minal
board
Receiving
instrument
Receiving
resistance:
250 to 450
F0705.EPS
WARNING
Be sure to set parameters as "Protect" on the
write protect function after finish of parameter
setting work.
Under extremely rare case, the infra-red switches
may respond unexpectedly in such conditions as
sticking ball of water or extraneous substances on
the surface of display panel glass according to
the principle of infra-red switch operation.
Its probability rises in such cases as sticking rain
water by storm or other similar situation and
washing up work near flowmeter installation place.
Either to illuminate or stop illuminating the infrared switches by the flashlight may cause the
mis-reaction.
Refer to Chapter 6 "Menu P: Parameter Protection Items" and section "10.2.3" how to use the
write protect function in detail.
In case of BT200, the parameters are displayed
in English only.
Even if the language with the exception of English
is selected at B10/H30: Language, the parameters are displayed in English upon BT200.
7.2.2The data update and upload/
download function of BT200
(1) The data update of BT200
When the following parameters are displayed, the measured data is updated automatically every seven seconds.
Item
A10
A20
A21
A30
A31
A32
C44
G22
G40
G41
G42
Name (BRAIN)
FLOW RATE (%)
FLOW RATE
FLW RATE (mA)
TOTAL
REV TOTAL
DIF TOTAL
VELOCITY CHK
4-20 BURNOUT
OPERATE TIME
ALM RECORD1
ALM TIME 1
Item
G43
G44
G45
G46
G47
G48
N23
N24
P20
P23
Name (BRAIN)
ALM RECORD2
ALM TIME 2
ALM RECORD3
ALM TIME 3
ALM RECORD4
ALM TIME 4
TEST SI1
TEST SI2
W PROTECT
SOFT SEAL
T0703.EPS
7-3
IM 01E20C01-01E
(2) Upload/download function of BT200
NOTE
Upload function is used when the parameters of one
AXFA11 are copied to the BT200. And download
function is used when the parameters copied to the
BT200 are set to another AXFA11.
For details, refer to BT200 User’s Manual (IM
01C00A11-01E). The targeted parameters for upload
and download are following.
Item
C11/B20
C31
C32
C40/B21
C41/B22
C42/B23
C43/B24
Name (BRAIN)
FLOW DAMPING
SIZE UNIT
NOMINAL SIZE
FLOW UNIT
TIME UNIT
FLOW SPAN
FLOW DECIMAL
Item
D10/B30
D11/B31
E10/B32
E11/B33
H10/B40
H11/B41
H12/B42
Name (BRAIN)
TOTAL UNIT
TOTAL SCALE
PULSE UNIT
PULSE SCALE
DISP SELECT1
DISP SELECT2
DISP SELECT3
T0704.EPS
7.2.3BT200 Screens & Flow Rate
Data Display
Use the following procedure to display flow rate data
on the BT200.
• The display of flow rate data is updated every 5 seconds.
–––WELCOME–––
BRAIN Terminal
ID:BT200
Check cable connection
and press the Ent key.
UNTIL FEED
When the BT200 is turned
on, the message “Please
wait ...” is displayed for
several seconds, and then
the screen on the left is
shown.
Press the key.
7. OPERATION VIA BRAIN TERMINAL (BT200)
MENU
A:DISPLAY
B:EASY SETUP
C:BASIC SETUP
D:TOTAL SET
E:PULSE SET
F:STATUS FUNC
HOME SET ADJ ESC
PARAMETER
A10:FLOW RATE(
80.0
A20:FLOW RATE
100m3/h
A21:FLW RATE(mA)
16.800mA
DATA DIAG PRNT ESC
%
)
%
The Menu Screen (Major
Item Parameter Search
Mode) will be displayed.
If the key is
pressed while A: DISPLAY
is selected by the highlight
bar, the Parameter Screen
(Sub-Item Parameter
Search Mode) will be
displayed.
( )
ESC
The Flow Rate Screen
will be displayed.
A maximum of three items
of data can be displayed.
The or
key can be used to move
the cursor up and down.
(Note that the key
can be used to move the
cursor to the top of the
next screen.
Press the
to return to Menu Screen
(Major Item Parameter
Search Mode).
( )
ESC
F0706.EPS
7.3Parameter Setting
Using a BT200
This section describes the procedure for setting of
parameters using a BT200.
key
PARAMETER
01:MODEL
AXFA11-A00
02:TAG NO
FI-1101
03:SELF CHECK
GOOD
OK
Press the key
to display initial data as
shown on the left.
The tag number
specified upon
ordering is found
here.
Press the ( )or
key.
or
OK
IMPORTANT
If the power of flowmeter is turned off within 30
seconds after parameters have been set, these
settings will be canceled. Accordingly, please
keep the power on for at least 30 seconds after
setting parameters.
Before updating any setting, remember to always
check the data content you want to change as
described in Chapter 6: Parameter Description.
7-4
IM 01E20C01-01E
7. OPERATION VIA BRAIN TERMINAL (BT200)
7.3.1BT200 Setting of Selection-Type Data: Flow rate units
In this example, the flow rate units specified by the selection-type parameter B21: Flow Unit are changed from m
to l (Liter).
3
MENU
A:DISPLAY
B:EASY SETUP
C:BASIC SETUP
D:TOTAL SET
E:PULSE SET
F:STATUS FUNC
HOME SET ADJ ESC
MENU
A:DISPLAY
B:EASY SETUP
C:BASIC SETUP
D:TOTAL SET
E:PULSE SET
F:STATUS FUNC
HOME SET ADJ ESC
PARAMETER
B10:LANGUAGE
English
B20:FLOW DAMPING
3.0s
B21:FLOW UNIT
m3
DATA DIAG PRNT ESC
PARAMETER
B10:LANGUAGE
English
B20:FLOW DAMPING
3.0s
B21:FLOW UNIT
m3
DATA DIAG PRNT ESC
When Menu Screen (Major
Item Parameter Search
Mode) has been entered as
shown here, press the
With the cursor aligned with
“l (Liter)”, press the
key.
When selected, “l (Liter)”
will begin to flash on and
off: Press the
key once again to confirm
your selection.
In case of cancel, press the
NO
( ) key.
The units for flow rate have
been changed to l (Liter).
Press ( ) key to
OK
exit Setting Screen
(Parameter Replacement
Mode), and then the display
returns to Parameter
Screen (Sub-Item
Parameter Search Mode) of
Menu B items.
In case of redo of setting,
press the ( ) key
NO
to return the Setting Screen
(Parameter Replacement
Mode).
In the parameter list, the
parameter for setting of flow
rate span is B23: FLOW
SPAN. Press the
key four times to move the
cousor to B23.
(Note that the key
can be used to move the
cursor to the top of the next
screen.
or
After moving the cursor to
B23: FLOW SPAN, press
the key to access
Setting Screen (Parameter
Replacement Mode).
In this mode, new values can
be entered into the
highlighted field using the
alphanumeric keys.
SETTING
B23:FLOW SPAN
100.000l/min
+ 120
DEL CLR ESC
SETTING
B23:FLOW SPAN
100.000l/min
+ 120
PRINTER OFF
F2:PRINTER ON
FEED POFF NO
SETTING
B23:FLOW SPAN
120.000l/min
FEED NO OK
PARAMETER
B22:TIME UNIT
/min
B23:FLOW SPAN
120.000l/min
B24:FLOW DECIMAL
l
DATA DIAG PRNT ESC
In this example, the value
“120” is entered, press the
key.
The data will begin to flash
on and off: Press the
key once again
to confirm your entry.
In case of cancel, press the
( ) key.
NO
The value for flow span has
been changed to
“120.000 l/min.”
Press ( ) key to
OK
exit Setting Screen
(Parameter Replacement
Mode), and then the display
returns to Parameter
Screen (Sub-Item
Parameter Search Mode) of
Menu B items.
In case of redo of setting,
press the ( ) key
to return the Setting Screen
(Parameter Replacement
Mode).
OK
( )
The changed setting for
flow rate span can now be
confirmed.
Press the ( ) key
ESC
to return to Menu Screen
(Major Item Parameter
Search Mode).
NO
F0708.EPS
DEL CLR ESC
7-6
IM 01E20C01-01E
7. OPERATION VIA BRAIN TERMINAL (BT200)
7.3.3BT200 Setting of Alphanumeric-Type Data: Tag number
In this example, the tag number specified by the alphanumeric-type parameter C10: TAG NO is changed from
“FI-1101” to “FI-1201”.
MENU
A:DISPLAY
B:EASY SETUP
C:BASIC SETUP
D:TOTAL SET
E:PULSE SET
F:STATUS FUNC
HOME SET ADJ ESC
MENU
A:DISPLAY
B:EASY SETUP
C:BASIC SETUP
D:TOTAL SET
E:PULSE SET
F:STATUS FUNC
HOME SET ADJ ESC
PARAMETER
C10:TAG NO
FI-1101
C11:FLOW DAMPING
3.0s
C20:MEASURE MODE
Standard DF
DATA DIAG PRNT ESC
When Menu Screen (Major
Item Parameter Search
Mode) has been entered as
shown here, press the
The highlighted data will
begin to flash on and off:
Press the key
once again to confirm your
entry.
In case of cancel, press the
( ) key.
NO
The tag number has been
changed to “FI-1201”.
Press ( ) key to
OK
exit Setting Screen
(Parameter Replacement
Mode), and then the display
returns to Parameter
Screen (Sub-Item
Parameter Search Mode) of
Menu C items.
In case of redo of setting,
press the ( ) key
NO
to return the Setting Screen
(Parameter Replacement
Mode).
( )
OK
SETTING
C10:TAG NO
FI-1101
FI-1101
CODE CAPS CLR ESC
SETTING
C10:TAG NO
FI-1101
FI-1201
CODE CAPS CLR ESC
An underscore (small
cursor) will start to flash on
and off at the furthest left
character in the tag number
as displayed in the dataoverwrite section.
Press the key four
times to move the flashing
underscore (small cursor) to
the position for multiples of
100.
X 4
When the flashing underscore (small cursor) has
been moved to the position
for multiples of 100, use the
alphanumeric keys to enter
the value “2”. (When this
value has been entered, the
underscore will move to the
position for multiples of 10.)
When the tag number has
been corrected to “FI-1201”
in this way, press the
key.
PARAMETER
C10:TAG NO
FI-1201
C11:FLOW DAMPING
3.0s
C20:MEASURE MODE
Standard DF
DATA DIAG PRNT ESC
The changed setting for tag
number can now be
confirmed.
Press the ( ) key
ESC
to return to Menu Screen
(Major Item Rarameter
Search Mode).
F0709.EPS
7-7
IM 01E20C01-01E
8. OPERATION VIA HART COMMUNICATOR TOOL (HART 5)
8.
This chapter describes the AXFA11 as an
example.
Note: HART is a registered trademark of the HART
OPERATION VIA HART COMMUNICATOR TOOL (HART 5)
NOTE
Communication Foundation.
8.1Matching of instrument (AXFA11) DD
and HART Configuration Tool’s DD
Before using the HART Configuration Tool (such as
FieldMate), confirm that the DD (Device Description)
of the AXFA11 is installed in the Configuration Tool
before using.
About the DD, use the device type, device revision and
DD Revision shown in follows.
2. Confirmation of the device revision for the HART
Configuration Tool
(1) Turn on the power of the Configuration Tool
under the standalone condition.
(2) Confirm the installed DD revision in accor-
dance with the procedure of the Configuration
Tool. Refer to its manual how to confirm it in
detail.
The first 2 digits of the DD file are expressed
the device revision, and its last 2 digits are
expressed the DD revision.
01 03.XXX
DD revision
Device revision
F0801.EPS
Model Name DTM Name Device TypeDevice
Revision
AXFA11AXFA11
V1.1
The DD revisions for AXFA11 and Configuration
Tool’s can confirm in accordance with the following
steps.
If the correct DD is not installed in the HART Configuration Tool, download them from the official
HART programming sites, otherwise, contact the
respective vendors of the Configuration Tool for its
upgrade information.
1. Confirmation of the device revision for AXFA11
(1) Connect the Configuration Tool to the
AXFA11. The device revision can be checked
as follows.
DDDevice Setup → Review → Review4
→ Fld dev rev
DTMConfi guration → HART → Fld dev rev
AXFA11
(0x0050)
1
T0801.EPS
T0802.EPS
8-1
IM 01E20C01-01E
8. OPERATION VIA HART COMMUNICATOR TOOL (HART 5)
NOTE
NOTE
8.2Interconnection between AXFA11 and HART
Configuration Tool
The HART Configuration Tool can interface with the AXFA11 from the control room, the
AXFA11 site, or any other wiring termination point in the loop, provided there is a
minimum load resistance of 250 between the connection and the receiving instrument.
To communicate, it must be connected in parallel with the AXFA11, and the connections
must be non-polarized. Figure 8.2.1 illustrates the wiring connections for a direct interface
at the AXFA11 site. The HART Configuration Tool can be used for remote access from
any terminal strip as well.
Control room
I +I -
AXFA11
4 to 20 mA DC signal line
Relaying
terminals
USB
FieldMate
modem
PC/FieldMate
USB
Te r minal board
Receiving instrument
load resistance:
250 Ω to 600 Ω
(including cable
resistance)
F0802.EPS
Figure 8.2.1 Interconnection Diagram
WARNING
Be sure to set parameters as "Protect" on the
write protect function after finish of parameter
setting work.
Under extremely rare case, the infra-red
switches may respond unexpectedly in such
conditions as sticking ball of water or extraneous
substances on the surface of display panel glass
according to the principle of infra-red switch
operation.
Its probability rises in such cases as sticking rain
water by storm or other similar situation and
washing up work near flowmeter installation
place.
Either to illuminate or stop illuminating the infrared switches by the flashlight may cause the
mis-reaction.
Refer to Chapter 6 "Menu P: Parameter Protection Items" and section "10.2.3" how to use the
write protect function in detail. This menu in the
275 is located at [Hot key] → [2 Wrt Protect Menu].
IMPORTANT
If the power of flowmeter is turned off within 30
seconds after parameters have been set, these
settings will be canceled. Accordingly, please
keep the power on for at least 30 seconds after
setting parameters.
Before updating any setting, remember to always
check the data content you want to change as
described in Chapter 6: Parameter Description.
In case of HART Configuration, the parameters
are displayed in English only.
Even if the language with the exception of
English is selected at “Language” setting display,
the parameters are displayed in English upon
HART Configuration Tool.
8-2
IM 01E20C01-01E
8. OPERATION VIA HART COMMUNICATOR TOOL (HART 5)
NOTE
8.3Basic Setup
Tag and Device Information
The tag number and device information can be checked
as follows:
• The location for the tag number and device informa-
tion
(DD)
TagDevice Setup → Detailed Setup
→ Basic setup → Ta g
Descriptor Device Setup → Detailed Setup
→ Device info → Field device info
→ Descriptor
MessageDevice Setup → Detailed Setup
→ Device info → Field device info
→ Message
DateDevice Setup → Detailed Setup
→ Device info → Field device info
→ Date
T0803.EPS
(DTM)
TagEasy Setup → Ta g
or Confi guration → HART → Tag
Descriptor Confi guration → Device infomation
→ Descriptor
MessageConfi guration → Device infomation
→ Message
DateConfi guration → Device infomation
→ Date
T0804.EPS
When changing the tag number or device information,
enter the information directly within the following
limitations.
8.4Parameters
8.4.1Parameter configuration
The parameter structure of the HART configuration
tool is hierarchical.
Refer to 8.4.6, Menu Tree Example for the corresponding parameters. The menu tree shows a cross-reference
of the parameters for HART and BRAIN.
See “Chapter 6 Parameter Description” for the functions of parameters.
Note that some display parameters of AXFA11 are
different from those of HART configuration tools.
8.4.2Data Renewing
There are two methods to load the AXFA11 data
from/to HART Configuration Tool --- periodic data
renewing and discretionary data renewing. Refer to the
8.4.6 menu tree for each corresponding parameter.
(1) Periodic Data Renewing
The data is updated 0.5 to 2 second cycles. The
parameter of this type is marked as “P” in the 8.4.6
menu tree.
(2) Discretionary Data Renewing
The data can be loaded from/to the AXFA11 when
the data save is finished to the Configuration Tool.
The parameter of this type is marked as “u/d” in
the 8.4.6 menu tree.
ItemNumber and characters
Tag8
Descriptor16
Message32
Date2/2/2 (mm/dd/yy)
• mm : month
• dd : day
• yy : year
T0805.EPS
The AXFA11 is not able to set parameters via its
display unit during a communication of the HART
Configuration Tool.
8-3
IM 01E20C01-01E
8. OPERATION VIA HART COMMUNICATOR TOOL (HART 5)
8.4.3Self-diagnostic
The self-diagnostic function of the AXFA11 is explained in Section 6.5 “Alarm Functions.”
The HART configuration tool can be used to run
self-diagnostics on a device and check for incorrect
data settings.
(1) Using DD
The Self test and Status commands are available for
self-diagnostics. When Self test is run, the display unit
shows an error code and alarm message if the device
detects any illegal parameter settings or functional
faults.
• Procedure to call up the Self test display
Device Setup → Diag/Service → Test → Self test
If no error is detected, “Self test OK” is displayed on
the configuration tool.
If the specific diagnostic item is known for the check,
you can directly call up the item by using the Status
command.
The status is categorized from 1 to 7.
Show an example below to confirm the status of Status
group 1.
• Procedure to call up the Status display
Device Setup → Diag/Service → Status → Status group 1
If no error is detected, “Off” is displayed on the
configuration tool.
If there is an error, “On” is displayed on the configuration tool, and a countermeasure for that error is
necessary.
Example of display: Span > 10 m/s On
Span < 0.3 m/s Off
TTL > 10000 p/s Off
TTL < 0.0001 p/s Off
4-20 Lmt Err Off
The HART configuration tool diagnoses at each
communication.
When an improper operation is performed, the error
message is displayed.
(2) Using DTM
The Device Status commands are used for
self-diagnostics. When Device Status is run, the
display unit shows an error code and alarm message if
the AXFA11 detects any illegal parameter settings or
functional faults.
If no error is detected, “Status: Normal” is displayed
on the configuration tool.
If the specific diagnostic item is known for the check,
you can directly call up the item by using the Diagnostic List in the Device Status display.
The Diagnostic List is categorized to Device Status,
Hardware Failure, Transducer Status, Diag Status, and
Configuration.
If no error is detected, color symbol which shows
Normal State is displayed on top of the error message.
If color symbol which shows Error State is displayed,
there is an error and a countermeasure for that error is
necessary.
The HART configuration tool diagnoses at each
communication.
When an improper operation is performed, the error
message is displayed.
8.4.4HART Specific Functions
Trim Analog Output
This function is used to adjust the analog output at
4 mA and 20 mA with the D/A trim or Scaled D/A trim.
(1) D/A trim
Connect a calibration digital ammeter, and then
enter the read value of the ammeter for each output
of AXFA11.
• Procedure to call up the D/A trim display
DDDevice Setup → Diagnosis/Service →
Adjustment → D/A trim
DTMCalibration → D/A trim
T0808.EPS
(2) Scaled D/A trim
The output is adjusted by using a voltmeter or
other types of meters with 0 to 100% scale.
Example using a voltmeter:
4 mA DC → 1 V
20 mA DC → 5 V
Connect this meter, and enter the read value of the
meter for each output of AXFA11.
• Procedure to call up the Scaled D/A trim display
DDDevice Setup → Diagnosis/Service
→ Calibration
→ Analog output trim →Scaled D/A trim
DTMCalibration → Scaled D/A trim
T0809.EPS
• Procedure to call up the Device Status display
Device Status
IMPORTANT
The D/A trim should be executed only at single
range mode. If the D/A trim is executed at Bi
direction mode, it is possible that the current
output becomes 108%.
8-4
IM 01E20C01-01E
8. OPERATION VIA HART COMMUNICATOR TOOL (HART 5)
IMPORTANT
When “D/A trim” or “Scaled D/A trim” is carried
out, the warning message “83: Fix Cur Wng” is
displayed on the display unit.
CAUTION
The output adjustment function can match the
4mA and 20mA output to the reference meter
such as a voltmeter. In the output adjustment, it
is necessary to use the calibrated voltmeter and
resistance.
Fixed Current Output
This feature can be used to output a fixed current for
loop checks.
In the case of using DD
Call up the test output parameter (Loop test) and select
the output signal.
• Procedure to call up the display
DDDevice Setup → Diagnosis/Service →
Test → Loop test →
→ 4mAOutput a 4mA DC signal
→ 20mAOutput a 20mA DC signal
→ OtherSet a desired output signal value
→ EndExit
T0806.EPS
In the case of using DTM
Call up the test output parameter (Loop test) and select
either manual test or auto test, and set the current
value.
• Procedure to call up the display
DTMDiag and Service → Output Test →
Loop test →
→ Manual TestSet the current value or % value at
Test output value, then click the Start
button.
→ Auto TestSet the interval and rate of change of
current output at Auto Test Setting,
then click the Start button.
T0807.EPS
Burst Mode
AXFA11 continuously sends the data via HART
communication when the burst mode is set on (any one
of PV, % range/current, or process vars/crnt). The data
is sent intermittently as a digital signal at 3 times a
second.
The burst mode is set as follows.
(1) Setting the data to be sent
• Procedure to call up the display
DDDevice Setup → Detailed setup
→ HART output → Burst option
DTMConfi guration → HART → Burst option
T0810.EPS
Select the type of data to be sent from the following
options:
- Instantaneous flow rate (PV)
- Output in % and current output (% range/current)
- Instantaneous flow rate, totalization value* and
current output (Process vars/crnt)
* “Totl,” “Reverse Totl” or “Dif Totl”
(2) Setting the burst mode
• Procedure to call up the display
DDDevice Setup → Detailed setup →
HART output → Burst mode
DTMConfi guration → HART → Burst mode
T0811.EPS
Then, select “On” at the menu to start the burst
mode.
To release from the burst mode, call up the burst
mode display, and set to “Off.”
The default setting is “Off.”
Multidrop Mode
When set in the multidrop mode, up to 15 field devices
in a single communications line can be connected. To
activate multidrop communication, the address of the
field devices must be set to a number from 1 to 15.
This deactivates the 4 to 20 mA analog output, fixing it
to 4 mA. Burn out is also disabled.
Note that the accuracy for multidrop mode is different
from that for other modes. Refer to Chapter 11:
Outline.
The multidrop mode is set as follows.
(1) Setting the polling address
• Procedure to call up the display
DD (HART 5)Device Setup → Detailed setup →
Output condition → HART output →
DTM (HART 5)Confi guration → HART →
→ Poll addrEnter the number from 1 to 15
T0812.EPS
8-5
IM 01E20C01-01E
(2) Activating the multidrop mode
About the procedure to call up the Polling display,
please refer to the User’s Manual of each configuration tool.
NOTE
When the same polling address is assigned for
two or more field devices in multidrop mode,
communication with these field devices is
disabled.
(3) Communication when set in the multidrop mode
1. The HART configuration tool searches a field
device that is set in the multidrop mode when
the HART configuration tool is turned on.
When the HART configuration tool is connected
to the field device, the polling address and the
tag will be displayed.
2. Select the field device, and then communication
with it is possible. The communication speed is
slow in this case.
3. To communicate with another field device, turn
off the power once and then turn it back on, or
call up “Online” menu, and select “Online.”
The polling address and the tag will appear.
Select the field device.
(4) Release of the multidrop mode
Call up the “Poll addr” display, and set the
address to “0.”
8. OPERATION VIA HART COMMUNICATOR TOOL (HART 5)
8.4.5Other operations for the HART
configuration tool
Regarding other operations for the HART configuration tool, refer to the HART configuration tool operations manual.
8-6
IM 01E20C01-01E
8.4.6Menu Tree for DD (HART 5)
8. OPERATION VIA HART COMMUNICATOR TOOL (HART 5)
Offline
Online
Frequency
Utility
1 Device setup
New Configuration
Saved Configuration
1. Device setup
2. PV
3. PV AO
4. PV Span
1 Process Variables PV % rnge
2 Diag/Service
PV
PV AO
To tl
Reverse Totl
Dif Totl
1 Test/Status
2 Adjustment
1 Status
2 Self test
1 Auto Zero Exe
2 Magflow Zero
3 D/A trim
4 Scaled D/A trim
Status group 1(System alarms)
Status group 3(Process alarms)
Status group 5(Setting alarms)
Status group 6(Setting alarms)
Status group 7(Setting alarms)
Status group 8(Warnings)
Read/Write
R
R
R
R
R
R
R
R
R
R
R
R
W
W
R/W
W
W
Parameter
of BRAIN
protocol
See
Section 6.5
“Alarm
Functions”
M10/B50
Data(*1)
A10
A20
A21
A30
A31
A32
–
–
–
–
–
–
––
–
M11
–
–
–
–
–
P
P
P
P
P
P
3 Easy Setup
3 Output Test
4 Diagnosis
1 Language
2 PV Damping
3 Base Flow Unit
4 Base Time Unit
5 PV Span
6 Flow Decimal Pnt
7 Total Unit
8 Total Scale
9 Pulse Unit
Pulse Scale
Display Select1
Display Select2
Display Select3
Auto Zero Exe
4 Select Flow Tube
5 Nominal Size unit
6 Nominal Size
7 Base Flow Unit
8 Base Time Unit
9 PV Span
Flow Decimal Pnt
Velocity Check
Density Unit
Mass Flow Density
User Span Select
Flow User Unit
Flow User Span
1 Total Unit
2 Total Scale
3 Total Decimal Pnt
4 Total Low Cut
5 Total Execution
6 Ttl Set Val Lower
7 Ttl Set Val Upper
8 Ttl Switch Lower
9 Ttl Switch Upper
Ttl User Select
Ttl User Unit
1 SO1 Function
2 SO2 Function
3 SI1 Function
4 SI2 Function
5 SO1/2 Active Mode
6 SI1/2 Active Mode
7 Forward Span2
8 Forward Span3
9 Forward Span4
Reverse Span1
Reverse Span2
Reverse Span3
Reverse Span4
Auto Range Hys
Bi Direction Hys
Ta g
Language
PV Damping
Base Flow Unit
Base Time Unit
PV Span
Flow Decimal Pnt
Total Unit
Total Scale
Pulse Unit
Pulse Scale
Display Select1
Display Select2
Display Select3
Auto Zero Exe
Select Flow Tube
Nominal Size Unit
Nominal Size
Measure Mode
Low MF
High MF
Low MF (EDF)
High MF (EDF)
MF Set
PV Damping
Base Flow Unit
Base Time Unit
PV Span
Flow Decimal Pnt
User Span Select
Flow User Unit
Flow User Span
Velocity Check
Density Unit
Mass Flow Density
Total Unit
Total Scale
Total Decimal Pnt
Total Low Cut
Total Execution
Ttl Set Val Lower
Ttl Set Val Upper
Ttl Switch Lower
Ttl Switch Upper
Ttl User Select
Ttl User Unit
Read/Write
W
W
W
W
W
W
W
W
–
W
W
W
W
W
W
W
W
R
W
W
W
W
W
W
W
W
W
W
W
W
W
Parameter
of BRAIN
protocol
C30
C31
C32
C20
C21
C22
C23
C24
–
C11/B20
C40/B21
C41/B22
C42/B23
C43/B24
C47
C48
C49
C44
C45
C46
D10/B30
D11/B31
D12
D13
D20
D21
D22
D23
D24
D30
D31
Data
Renewing
(*1)
–
u/d
u/d
–
–
–
–
–
–
u/d
u/d
u/d
u/d
u/d
u/d
u/d
u/d
P
u/d
u/d
u/d
u/d
u/d
u/d
u/d
u/d
u/d
u/d
u/d
u/d
u/d
Pulse Set
Status FunctionSO1 Function
(continued on next page)
Pulse Unit
Pulse Scale
Pulse Low Cut
Pulse Width
Pulse Active Mode
SO2 Function
SI1 Function
SI2 Function
SO1/2 Active Mode
SI1/2 Active Mode
Auto Range Hys
Bi Direction Hys
Forward Span2
Forward Span3
Forward Span4
Reverse Span1
Reverse Span2
Reverse Span3
Reverse Span4
Display Select2
Display Select3
Display Cycle
Language
Aux
4-20mA Low Cut
4-20mA Low Lmt
4-20mA High Lmt
Flow Direction
Rate Limit
Dead Time
Pulsing Flow
T/P Damping Select
Power Synch
Power Frequency
Memo1
Memo2
Memo3
Software Rev No
Model Specific
Use
Lining
Electrode Material
Electrode Struct
Grounding Ring
Process Connect
Lay Length
Electrical Conn
Sensor Serial No
Model
Manufacturer
Hardware rev
Software rev
Desciptor
Message
Date
Final asmbly num
Ta g
Poll addr
Dev id
Universal rev
Fld dev rev
Num req preams
Physical signl code
Burst mode
Burst option
R
R
R
R
W
W
W
W
W
W
R
R
R
R
R
W
W
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
u/d
–
–
–
–
–
–
–
–
F0811.EPS
8-15
IM 01E20C01-01E
8. OPERATION VIA HART COMMUNICATOR TOOL (HART 5)
F
Calibration
G
Write Protect
Auto Zero Exe
Magflow Zero
D/A trim
Scaled D/A trim
Write protect
Enable Wrt Password
New Password
Software Seal
*1: Data Renewing
P : Periodic Data Renewing
u/d
: Discretionary Data Renewing
– : Others (Method, etc...)
Read/Write
–
R/W
–
–
R
W
W
R
Parameter
of BRAIN
protocol
M10/B50
M11
–
–
P20
P21
P22
P23
Data
Renewing
(*1)
–
–
–
–
P
–
–
–
F0812.EPS
8-16
IM 01E20C01-01E
9.ACTUAL OPERATION
After you have installed the flowtube into the process
piping, wired the input/output terminals, set up the
required parameters, and performed a pre-operation
zero adjustment, the magnetic flowmeter should output
an accurate flow signal from its terminals as soon as
flow of the fluid to be measured begins. This section
describes zero adjustment and the corresponding
procedures.
9.1Pre-operation Zero
Adjustment
Zero adjustment is carried out to ensure that the output
for zero flow is 0% (i.e., 4 mA). Although adjustment
to zero is performed at the manufacturing plant prior to
shipment, this procedure must be carried out once
again following the installation of piping in order to
match the magnetic flowmeter to its operating conditions.
9. ACTUAL OPERATION
This section describes the zero adjustment procedure
using display unit switches from the converter and
using the external status input; accordingly, one of
these methods should be selected and implemented.
IMPORTANT
• Zero adjustment should be carried out before
actual operation. Note that setting and update
functions cannot be carried out during this
procedure (i.e., for approximately 30 seconds).
• Zero adjustment should only be carried out
when the flowtube has been filled with
mesurement fluid and the fluid velocity is
completely zero by closing the valve.
• Each time that the fluid being measured is
changed, it will be necessary for zero adjustment to be carried out for the new fluid.
9-1
IM 01E20C01-01E
9.1.1Zero Adjustment Using Display Unit Switches
This section describes the procedure for zero adjustment using the display unit switches.
(For more details regarding setting methods using these switches, refer to Chapter 5: Basic
Operating Procedures.)
The parameters for zero adjustment are B50/M10: Auto Zero Exe (and either of these
can be used to carry out this procedure). For more details regarding these parameters, refer
to Chapter 6: Parameter Description.
The parameter M10: Auto Zero Exe will be used in the following description.
Start:
Major Item
Parameter
Search Mode
Once in Setting Mode, use the switch to
move the cursor to M: Ajustment.
9. ACTUAL OPERATION
Sub-item
Parameter
Search Mode
Parameter
Replacement
(execute) Mode
Setting Mode
Major item
parameter
Major item
Sub-item
selection (D)
Setting Mode
K:Diagnosis
M:Adjustment
N:Test
M:Adjustment
10:Auto Zero Exe
11:Magflow Zero
M10:Auto Zero Exe
No Execution
No Execution
Execution
M10:Auto Zero Exe
No Execution
Execution
No Execution
M10:Auto Zero Exe
Execution
Touch the switch to access Sub-item
Parameter Search Mode.
Upon selection of M: Ajustment, the cursor will be
positioned at M10: Auto Zero Exe. (Sub-item
sellection (D))
Touch the switch to access Parameter
Replecement Mode.
Touch the switch to move the cursor to
“Execution”.
Touch the switch to select “Execution”.
In order to request confirmation, the entire display
flashes on and off. Touch the switch once
again at this time to fix selection of the automatic
zero adjustment function.
Now Auto Zero
Executing...
F0901-1.EPS
9-2
NOTE
When no operations are carried out for 20
seconds in the flashing state, the system will
automatically return to the Sub-item Parameter
Search Mode.
When the operations except are carried
out, the parameter cannot be set.
Auto zero adjustment function is being executed
(about 30 seconds).
IM 01E20C01-01E
9. ACTUAL OPERATION
Sub-item
Parameter
Search Mode
Major item
Sub-item
selection (D)
M:Adjustment
10:Auto Zero Exe
11:Magflow Zero
F0901-2.EPS
When zero adjustment function has been completed,
the system automatically returns to the sub-item
selection screen (D).
The results of M10: Auto Zero Exe can be
displayed using M11: Magflow Zero. Alternatively, if the results of the automatic zero
adjustment exceed the rated value, the warning 82: Auto Zero Wng will be displayed.
9.1.2Zero Adjustment via External Status Input
This section describes the procedure for zero adjustment via external status input. (For
more details regarding external status input, refer to Chapter 6: Parameter Description.)
CAUTION
In certain cases where the multiple range function is being used with other status
inputs, it may not be possible to perform settings for automatic zero adjustment.
For more details, refer to the description of multiple ranges from Chapter 6:
Parameter Description.
NOTE
Start:
Major Item
Parameter
Search Mode
Sub-item
Parameter
Search Mode
In order to carry out zero adjustment via external status input, it will be necessary to set
“Ext Auto Zero” using either F12: SI1 Function or F13: SI2 Function. The parameter
F12: SI1 Function will be used in the following description.
Once in Setting Mode, use the switch to
move the cursor to item F: Status Function.
Setting Mode
Major item
parameter
Major item
Sub-item
parameter
Major item
Sub-item
selection (E)
Setting Mode
E:Pulse Set
F:Status Function
G:Alarm
F:Status Function
41:Bi Direction Hys
10:SO1 Function
11:SO2 Function
X
F:Status Function
11:SO2 Function
12:SI1 Function
13:SI2 Function
2
Touch the switch to access Sub-item
Parameter Search Mode.
Touch the switch twice to move the cursor
to F12: SI1 Function.
The cursor has been moved to F12: SI1 Function
in this screen. (Sub-item selection screen (E))
Touch the switch to access Parameter
Replacement Mode.
F0902-1.EPS
9-3
IM 01E20C01-01E
9. ACTUAL OPERATION
Parameter
Replacement
(execute) Mode
F12:SI1 Function
No Function
No Function
0% Signal Lock
F12:SI1 Function
No Function
Ext Auto Zero
Ext Ttl Preset
F12:SI1 Function
Ext Auto Zero
Touch the switch twice to move the cursor
to “Ext Auto Zero”.
X
2
Touch the switch to select “Ext Auto Zero
(Zero adjustment via external status input)”.
In order to request confirmation, the entire display
flashes on and off. Touch the switch once
again at this time to fix selection of the automatic
zero adjustment function.
NOTE
When no operations are carried out for 20
seconds in the flashing state, the system will
automatically return to the Sub-item Parameter
Search Mode.
When the operations except are carried
out, the parameter cannot be set.
Sub-item
Parameter
Search Mode
Major item
Sub-item
selection (E)
F:Status Function
11:SO2 Function
12:SI1 Function
13:SI2 Function
The system automatically returns to sub-item
selection screen (E).
F0902-2.EPS
Zero adjustment will be started if the SI1 terminals are shorted (When the active mode has
been set to “Short Active” using F15: SI1/2 Active Mode).
This process will end after approximately 30 seconds.
Now Auto Zero
Executing...
F0902-3.EPS
Auto zero adjustment function is being executed.
(about 30 seconds.)
NOTE
When the SI1 terminals continue to be shorted, this zero adjustment is automatically repeated.
9-4
IM 01E20C01-01E
10.MAINTENANCE
S
10. MAINTENANCE
10.1 Maintenance
WARNING
• Maintenance work must be carried out by the
trained personnel having knowledge of safety
standard and not by operators.
• When opening the cover, wait for more than 10
minutes after turning off the power. Furthermore, opening of the cover must also be
carried out by the trained personnel having
knowledge of safety standard.
IMPORTANT
• As a rule, maintenance of this flowmeter should
be implemented in a maintenance service shop
where the necessary tools are provided.
• The amplifier assembly contains sensitive parts
that may be damaged by static electricity.
Excercise care so as not to directly touch the
electronic parts or circuit patterns on the board,
for example, by preventing static electrification
by using grounded wrist straps when handing
the assembly.
10.2 Setting of Switches
IMPORTANT
• Lifting of display unit is necessary for setting
switches. Perform display unit lifting as described in section 10.2.1.
• To preserve the safety, do not touch the
electrical circuit and the cables except the
setting switches.
10.2.1 Lifting of display unit
(1) Turn off the power.
(2) Remove the cover as described in section 4.4.1.
(3) Loosen the amplifier assembly’s two screws while
supporting it with your hand (See Figure 10.2.1).
(4) Lift the display unit (See Figure 10.2.2).
At this time, do not remove the connector.
(5) Set the switches as described in Section 10.2.2 and
10.2.3.
(6) Taking care not to entangle the cables, install the
display unit with two mounting screws.
(7) Install the cover.
10.1.1 Fuse Replacement
CAUTION
Please contact Yokogawa's service office for
fuse replacement. Also be sure to use the fuse
that was supplied by Yokogawa’s sales or
service offices.
Display unit mounting
screws (x2)
F1008.EP
Figure 10.2.1 Removing the Front Cover and Mounting
Screws of Display Unit
10-1
IM 01E20C01-01E
10. MAINTENANCE
NOTE
10.2.3 Setting of Write Protect Switch
By setting the write protect function to “Protect” it is
possible to prevent the overwriting of parameters.
Write protection can be carried out using either the
hardware switch on the CPU board (i.e., Switch 2) or
software parameter settings. If either of these items is
set to “Protect”, the overwriting of parameters will be
prohibited.
The burnout function sets the direction of current
output in situations where the CPU has become
damaged. Upon shipment from the manufacturing
plant, the burnout direction is set to High (i.e., 25 mA);
however, in cases where the optional code C1 has been
specified, the output direction will be set to Low (i.e.,
0 mA).
Modification of the burnout direction must be carried
out using the setting switch from the amplifier’s CPU
board (i.e. Switch 1) (See Figure 10.2.2).
Table 10.1Output Setting Pins for Burnout
Position
of Pin
LowHigh
Low
High
Burnout
Direction
High
Low
Output
25 mA
0 mA
Remarks
Set to High before
shipment
Set to Low for
optional code C1
T1001.EPS
If the hardware switch is set to “Protect”, it will
not be possible to overwrite parameters; furthermore, this condition will be maintained until the
switch is set to “Enable”.
For more details regarding usage of the write protect
function and the software’s parameter switches, refer to
Chapter 6: Parameter Description.
NOTE
On the amplifier’s CPU board, the burnout
setting switch (i.e., Switch 1) and the write
protect switch (i.e., Switch 2) are located adjacent to each other. Accordingly, special care
should be taken when making switch settings.
10-2
IM 01E20C01-01E
10.3 Troubleshooting
Although magnetic flowmeters rarely require maintenance, failures may occur when the
instrument is not operated correctly. This section describes troubleshooting procedures
where the cause of the breakdown is identified through receiver indication.
10.3.1 No Indication
START
Is an error being displayed? Check the converter display.
10. MAINTENANCE
Is an error being
displayed?
NO
Is anything else
being displayed?
YES
Does the
arrow on the tube match the
flow direction?
YES
Check whether the flowtube
is filled with fluid.
Is the flowtube
full of fluid?
YES
Check the resistance between
the flowtube’s A, B, and C
terminals using a multimeter.
YES
NO
NO
NO
Examine in accordance with
Section 6.5: Alarms.
Perform either of the following:
a. Reinstall the flowtube in
order to make the arrow and
flow direction match.
b. Reverse the setting for the
flow direction.
Change the piping or the
mounting condition so that
the tube fills with the fluid.
resistance 100 M or more
Remove the tube and
inspect the inside of the pipe.
Remove the foreign matter.
Is the
in each case?
YES
Is foreign
matter adhering to
the walls?
YES
NO
NO
10-3
Contact a service center.
F1005.EPS
IM 01E20C01-01E
10.3.2 Unstable Zero
START
Investigate whether or not the flowtube
is filled with fluid and that it is free of
bubbles.
*1
Is the flowtube
completely full of fluid?
YES
NO
10. MAINTENANCE
*1: When checking for bubbles, it is
convenient if there is a gas vent hole
on the flowtube’s downstream side.
Particular care must be taken in the case
of horizontal mounting. In order to ensure
complete filling of the tube, either adjust
the mounting position or switch to vertical
mounting.
Are there bubbles
in the fluid?
NO
Check for the presence of valve
leakage.
Are valves leaking?
NO
• Confirm the flowmeter grounding.
• Confirm whether or not the
grounding wire is disconnected
from the earth flange.
Is the flowmeter properly
grounded?
YES
Recheck the fluid to be measured.
• Does the fluid’s
conductivity vary?
• Is the fluid conductivity
within the limits?
NO
Is the fluid likely to
contain deposits?
NO
YES
YES
NO
YES
YES
Install a defoaming device on the
flowtube’s upstream side.
Replace the valve(s).
Completely ground the converter and
flowtube (Refer to the section on
grounding in the relevant user’s manual).
• Ensure that the fluid’s conductivity
changes as little as possible.
• Do not use the meter when fluid
conductivity is out of limits.
• Whenever possible, reduce the flow
velocity to being in the vicinity of the
lower conductivity limit.
• Remove dirt and scale of and near
electrodes.
• Use a soft brush or cloth to perform
scale removal, and make sure that
the lining is not damaged.
Is the flowmeter
located near motors, transformers,
or electric equipment?
NO
Contact a service center.
YES
10-4
Provide the meter with sufficient
shielding or move it away from the
electrical equipment
F1006.EPS
IM 01E20C01-01E
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.