yellow jacket EVac I Operation And Maintenance Manual

Page 1
EVac I
Digital Vacuum Gauge
99 Washington Street Melrose, MA 02176
Phone 781-665-1400 Toll Free 1-800-517-8431
Page 2
INTRODUCTION
Thank you for your purchase of the YELLOW JACKET® EVac I Digital Vacuum Gauge Utilizing patented technology, the EVac I is the most accurate vacuum gauge available.
With the EVac I , you can accurately measure vacuum pressure in Microns, Pascals, Millibar, Millitorr, and mmHg with a resolution of 1 micron. Ideal for the HVAC/R professional, the EVac I is small, lightweight, rugged and easy to use.
FEATURES
Large, high-visibility back-lit LCD display
Measures Vacuum in Microns, Pascals, Millibar, Millitorr and mmHg
0 to 25,000 Micron Range with 1 Micron Resolution
“Analog” Vacuum Level Bar Graph
Automatic Oil Sensor
Buil
Rugged, Compact Design – About the Size of a Manifold Gauge
Long Battery Life
Calibration Self Test -- Can be field calibrated with no special equipment
Ideal for HVAC/R Service and Industrial Use
t-in memory retains all previous settings
SPECIFICATIONS
snorciM 000,52 – 0 :egnaR
Vacuum Accuracy: 5% of Reading +/- 5 Microns
Vacuum Resolution: 1 Micron
Warm-up
Response Time: Instant
Battery Life: Up to 300 Hours
Operating Temperature: 10°F – 122°F (-12°C – 50°C)
Dimensions: 3.5” x 3” x 1.25” (9cm x 7.5cm x 3cm)
War Vacuum Gauge, never exceed 500 PSI.
Time: Instant
ning:
To avoid personal injury and to prevent damage to the EVac I Digital
1(
yrettaB enilaklA V9 :rewoP
)gHmm 000.52 ,rroTm 000,52 ,raBm 033.33 ,aP 3333(
)gHmm 100.0 ,rroTm 1 ,raBm 100.0 ,aP
tarepo erutarepmet wol rof dednemmocer muihtiL V9( ion)
H leviwS dna yrettaB gnidulcni )g071( .zo 6 :thgieW ook
Page 3
PARTS DIAGRAM
Page 4
LCD DISPLAY
Indicates that Temperature is Displayed on the Alternate Numeric Display (during
Indicates degrees Celsius on the Alternate Numeric Display (during calibration)
Item Function
A Main Numeric Display
B Alternate Numeric Display
C “Analog” Vacuum Level Bar Graph
Backlight Indicator. Flashing: Temporary, Solid: Always On
Sound Indicator
Battery Level Indicator
Oil Sensor Indicator
Calibration Mode Indicator
calibration)
Pascal Units Indicator
Millibar Units Indicator
Millitorr Units Indicator
Millimeters of Mercury Units Indicator
Micron Units Indicator
Page 5
KEYPAD
to indicate pressure
stem to be evacuated with a high quality vacuum hose. Start the
by compressing tab at the base of the
Attach battery clip to new battery and insert into battery compartment. Replace battery cover by
IMPORTANT: TO PREVENT DAMAGE FROM LEAKING BATTERIES, DO NOT LEAVE A DEAD
IS NOT TO BE USED FOR AN
strength of the battery. Four bars indicate full power.
As the battery is depleted, the number of bars displayed decreases, until no bars are left. At this
Item Function
Press to Turn Power On, Press and Hold to Turn Power Off
Press to Change Display Units
Press and Release to Activate Backlight
Press and Hold to Mute/Un-mute Sound
QUICK START
To operate the EVac I as a basic vacuum gauge:
1. Install the battery as described in the “Battery Installation” section below.
2. Turn the power on by pressing . The display will show greater than 25,000 Microns.
3. Select the desired units by repeatedly pressing .
4. Attach the EVac I to the sy vacuum pump. Read the vacuum level from the Main Display.
5. Turn the power off by pressing and holding .
BATTERY INSTALLATION & REPLACEMENT
1. Remove the battery cover from the rear of the EVac I battery cover.
2. If necessary, remove and detach old battery from battery clip.
3. by aligning tab and snapping back into place.
BATTERY INSIDE THE EVac I. REMOVE BATTERY IF THE EVac I AN EXTENDED PERIOD OF TIME.
BATTERY LEVEL INDICATOR
The battery level indicator shows the relative
Page 6
point, it is necessary to replace the batteries (refer to the Battery Installation section above)
If the battery power drops to the point where the EVac I
can no longer function accurately, the alarm will
the alarm will beep 10 times and the power will turn off automatically. When the battery is
If the sensed vacuum pressure exceeds 25,000 Microns (3333.1 Pa, 33.331 mBar, 25,000 mTorr, or
helps to conserve battery life through advanced
will enter Sleep Mode in
in which the sensor and backlight are turned off, thus reducing the load on the battery and extending the
will automatically turn itself off to further conserve battery
It is necessary to prevent oil from being drawn into the Vacuum Sensor. For HVAC/R service, always
-
on the vacuum pump prior to turning the vacuum pump off. Over time, oil vapor and other
has a built in Oil Sensor that will detect this condition. If the
key on the keypad. The appropriate units indicator
replaced, and the EVac I is turned back on, the EVac I will resume operation with all previous settings intact.
HI-PRESSURE INDICATION
mTorr, or25.000 mmHg), the display will show .
SLEEP MODE
Unlike other Digital Vacuum Gauges, the EVac I
power management. After 5 minutes of displaying , the EVac I
extending the battery life up to 300 hours or more. While in this mode, the display will show
and exit Sleep Mode if the pressure is less than 25,000 Microns. The gauge can be manually brought out of Sleep Mode by pressing any key on the keypad. While in Sleep, the gauge will beep twice every 5 minutes to remind you that it is on. The EVac I will not sleep in the Calibration Mode (see the Calibration section below).
. Approximately every 35 seconds, the gauge will automatically check the pressure
AUTO POWER-OFF
After 1 hour in “Sleep Mode” the EVac I battery power.
OIL SENSOR
always recover refrigerant prior to attaching gauge. If possible, always close the blank
materials may contaminate the sensor. The EVac I condition. If the OIL indicator in the display illuminates, this is an indication that the vacuum sensor has been contaminated and is no longer functioning accurately. If the sensor becomes completely saturated with oil to the extent that it cannot function properly at all, the message
described in the “Cleaning the Vacuum Sensor” section.
will show on the display. Prior to further use of the
EVac I, clean the sensor as
off valve
UNITS
Change the displayed units by pressing the indicator on the LCD display will show Pa, mBar, mTorr, mmHg, or Microns depending on the units set.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Page 7
RANGE AND RESOLUTION
Vacuum
Units
Bar
The EVac I has a broad vacuum pressure measurement range and very high resolution according to the table below:
Range Resolution
Pascals (Pa)
3,333.1 0.1 Pa
0
Millibar (mBar)
Millitorr (mTorr)
mmHg (mmHg)
Microns (Microns)
33.331 0.001 mBar
0
25,000 1 mTorr
0
25.000 0.001 mmHg
0
25,000 1 Micron
0
“ANALOG” VACUUM LEVEL BAR GRAPH
The EVac I also displays a Vacuum Level Bar Graph allowing for a quick visual determination of the vacuum level achieved. Each bar corresponds to a vacuum pressure range depending on units according to the following table:
Value Microns Pa Millibar Millitorr mmHg
10K 10,000 — ATM 1,000 — ATM 10 — ATM 10,000 — ATM 10 — ATM
5K 5,000 — 10,000 500 — 1,000 5 — 10 5,000 — 10,000 5 — 10 2K 2,000 — 5,000 200 — 500 2 — 5 2,000 — 5,000 2 — 5
1K 1,000 — 2,000 100 — 200 1 — 2 1,000 — 2,000 1 — 2 500 500 — 1,000 50 — 100 0.5 — 1 500 — 1,000 0.5 — 1 200 200 — 500 20 — 50 0.2 — 0.5 200 — 500 0.2 — 0.5 100 100 — 200 10 — 20 0.1 — 0.2 100 — 200 0.1 — 0.2
50 50 — 100 5 — 10 0.05 — 0.1 50 — 100 0.05 — 0.1
20 20 — 50 2 — 5 0.02 — 0.05 20 — 50 0.02 — 0.05
10 10 — 20 1 — 2 0.01 — 0.02 10 — 20 0.01 — 0.02
Page 8
BACKLIGHT
tor on the display will flash,
and the backlight will turn off automatically after 1 minute. To activate the Backlight permanently,
indicator on the display will show solid. Turn off the backlight by pressing
indicator turns off. During Sleep Mode, the backlight will turn off to help
conserve battery power, but will turn on again automatically upon resumption of normal operation if it
nternal speaker that will emit a beep for each valid key press, and also functions
as an alarm in the Calibration mode (see the Calibration section below). It will also beep every 5
ressing and holding the
indicator on the LCD
f the gauge and will allow it to
swivel freely in any direction. The gauge may be operated with or without the hook attached. When
fastening the hook to the gauge, turn it finger tight only. Use of a tool to tighten the hook may result in
should provide many years of service with no maintenance required. When not in use, the
dust cap should remain in place over the sensor port. Clean the plastic enclosure with a damp (not
ptable, but use no solvents. Take care not to expose the vacuum
sensor to oil. If the Oil Sensor (described above) indicates a contaminated sensor, follow the Sensor
aminated with oil (as indicated by the Oil Sensor), carefully follow
OBJECT INTO THE PORT, AS THIS WILL PERMANENTLY DAMAGE THE SENSOR
ir dry over at least an hour, or pull a vacuum on the sensor to clean it more
Replace the battery and turn on the gauge. The Oil Indicator should be off. If it is still on, repeat
alibration cycle as detailed in the Calibration section below.
: It is important to remove all alcohol vapors from the sensor, either through air
To activate the Backlight temporarily, press once. The indica
press again. The
repeatedly until the
was previously set in the permanent-on mode.
SOUND
The EVac I has an i
minutes in Sleep Mode. For silent operation, the sound can be muted by p key. The Calibration mode alarm is not affected by the mute status. The display indicates that the sound is on (not muted).
SWIVEL HOOK
The EVac I’s removable stainless steel swivel hook enables hanging o
damage to the EVac I case.
MAINTENANCE
The EVac I
wet) rag. Mild detergent is acce
Cleaning Procedure below.
CLEANING THE VACUUM SENSOR
If the vacuum sensor becomes cont this procedure:
1. Turn the EVac I power off.
2. Shake the gauge to remove any large quantities of oil from the sensor.
3. Apply a few drops of rubbing alcohol inside the sensor vacuum port. (DO
4. Place your finger over the port and shake for a few moments.
5. Remove your finger and shake out the alcohol.
6. Repeat steps (3) – (5) at least three times.
7. Allow the sensor to a quickly.
8. the cleaning procedure.
9. If full accuracy is desired, perform a c
Note
vacuum. Any remaining vapors will cause an incorrect vacuum reading.
NOT INSERT ANY
-drying or via
).
Page 9
CALIBRATION TEST
should rarely require recalibration, though it may be necessary to know that your gauge is
calibrated properly for full accuracy. The Calibration Test mode assures you that the
if the instrument is calibrated properly.
if the instrument requires calibration. Please see the
If the Calibration Test indicates recalibration is required, the gauge may be recalibrated. Unlike any
ed to factory specifications without any
For best results, clean the sensor with alcohol prior to calibration. Ensure the sensor is
y, install a fully charged battery into the gauge.
indicator should illuminate, an
should show on the Main Numeric Display. If not, turn the power off again and repeat.
in a clear Ziploc (resealable zipper storage) bag, press out any extra air, and
°
F). At this point, the alarm will sound and the
mperature of at least 23
F). At this point, the alarm will sound, and
is now calibrated to factory
to warm slowly. Attempting to
accelerate the warming by using a heat source will not provide satisfactory results. During the
ill be indicated on the Alternate Numeric Display in
degrees Celsius. The calibration process may be canceled at any time by turning off the
°
The EVac I
calibrated to factory specifications. Test the calibration as follows:
1. Turn the EVac I power off.
2. Important: Expose the EVac I to atmospheric pressure.
3. Press and hold (do not release) the key for about 5 seconds.
4. The display will show
5. The display will show Calibration section below.
CALIBRATION
other vacuum gauge, the EVac I can be easily recalibrat special equipment, with the following procedure:
1. completely dry before proceeding.
2. Turn the EVac I power off. If necessar
3. Place the dust cap over the vacuum fitting.
4. Hold while pressing . As soon as the power turns on, release
press quickly at least three times. The 
5. Place the EVac I seal.
6. Place the bagged EVac I into a freezer with a temperature of less than –5
7. Allow the EVac I to cool to below -2°C (28.4°
display will change to .
8. Remove from freezer and press any key to silence the alarm.
9. Place the EVac I undisturbed in an area with a room te no greater than 30°C (86° F).
10. Allow the EVac I to warm to 20° C (68° show on the display.
11. Press any key to silence the alarm. The EVac I
EVac I is
and and then
d
C (23°F).
°C (73.4° F) but
will
specifications.
Note: For accurate calibration, it is necessary to allow the EVac I
cooling/warming process, the temperature w
previous calibration will be unchanged.
EVac I. The
Page 10
RESTORING FACTORY CALIBRATION
calibration can be restored at any time by the following procedure:
indicator should illuminate, a
should show on the Main Numeric Display. If not, turn the power off again and repeat.
is now reset to the
. Please use the table
Solution
Clean the Vacuum Sensor
Vac
vacuum port with your hand,
and turn the
Clean the Vacuum Sensor
and Restart Calibration
Restart Calibration. Allow the
Gauge to
Restart Calibration. Leave
the Gauge Undisturbed
During Warming Phase.
C). While operating below
C), the display update rate will slow from 3.5 readings every second to one reading
A standard alkaline battey may not provide acceptable battery life at temperatures below 0
The original factory
1. Turn the EVac I power off.
2. Hold while pressing . As soon as the power turns on, release
press quickly at least three times. The 
3. Press five times followed by .
4. The gauge will beep 5 times and return to normal operation. The EVac I original factory calibration.
TROUBLESHOOTING
Under certain conditions, the display may read or below to determine and fix the problem:
Display Mode Possible Problem
Normal
Operation
Calibration Sensor Contaminated
Calibration
Sensor Contaminated
Ambient Temperature too
Low
Gauged Warmed too
Quickly
Gauge Disturbed During
Calibration
Turn the e
and and then
nd
I off, warm the
eVac I back on.
Warm Slowly
LOW TEMPERATURE OPERATION
The EVac I can operate accurately at temperatures as low as 10°F (-12° freezing (32°F/0° every two seconds.
For low temperature operation, a 9V Lithium battery is recommended.
°C (32° F).
Page 11
NOTES
Page 12
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
4
Loading...