Thank you for your purchase of the YELLOW JACKET® EVac I Digital Vacuum Gauge
Utilizing patented technology, the EVac I is the most accurate vacuum gauge available.
With the EVac I , you can accurately measure vacuum pressure in Microns, Pascals,
Millibar, Millitorr, and mmHg with a resolution of 1 micron. Ideal for the HVAC/R
professional, the EVac I is small, lightweight, rugged and easy to use.
FEATURES
•Large, high-visibility back-lit LCD display
•Measures Vacuum in Microns, Pascals, Millibar, Millitorr and mmHg
•0 to 25,000 Micron Range with 1 Micron Resolution
•“Analog” Vacuum Level Bar Graph
•Automatic Oil Sensor
•Buil
•Rugged, Compact Design – About the Size of a Manifold Gauge
•Long Battery Life
•Calibration Self Test -- Can be field calibrated with no special equipment
To avoid personal injury and to prevent damage to the EVac I Digital
1(
yrettaB enilaklA V9 :rewoP
)gHmm 000.52 ,rroTm 000,52 ,raBm 033.33 ,aP 3333(
)gHmm 100.0 ,rroTm 1 ,raBm 100.0 ,aP
tarepo erutarepmet wol rof dednemmocer muihtiL V9(ion)
H leviwS dna yrettaB gnidulcni )g071( .zo 6 :thgieWook
Page 3
PARTS DIAGRAM
Page 4
LCD DISPLAY
Indicates that Temperature is Displayed on the Alternate NumericDisplay(during
Indicates degrees Celsius on the Alternate Numeric Display (duringcalibration)
Item Function
A Main Numeric Display
B Alternate Numeric Display
C “Analog” Vacuum Level Bar Graph
Backlight Indicator. Flashing: Temporary, Solid: Always On
Sound Indicator
Battery Level Indicator
Oil Sensor Indicator
Calibration Mode Indicator
calibration)
Pascal Units Indicator
Millibar Units Indicator
Millitorr Units Indicator
Millimeters of Mercury Units Indicator
Micron Units Indicator
Page 5
KEYPAD
to indicatepressure
stem to be evacuated with a high quality vacuumhose.Start the
by compressing tab atthebaseof the
Attach battery clip to new battery and insert into battery compartment. Replacebatterycoverby
IMPORTANT: TO PREVENT DAMAGE FROM LEAKING BATTERIES, DO NOTLEAVEADEAD
IS NOT TOBEUSEDFORAN
strength of the battery. Four barsindicatefullpower.
As the battery is depleted, the number of bars displayed decreases, until no barsareleft.At this
Item Function
Press to Turn Power On, Press and Hold to Turn Power Off
Press to Change Display Units
Press and Release to Activate Backlight
Press and Hold to Mute/Un-mute Sound
QUICK START
To operate the EVac I as a basic vacuum gauge:
1.Install the battery as described in the “Battery Installation” section below.
2.Turn the power on by pressing. The display will show
greater than 25,000 Microns.
3.Select the desired units by repeatedly pressing .
4.Attach the EVac I to the sy
vacuum pump. Read the vacuum level from the Main Display.
5.Turn the power off by pressing and holding.
BATTERY INSTALLATION & REPLACEMENT
1.Remove the battery cover from the rear of the EVac I
battery cover.
2.If necessary, remove and detach old battery from battery clip.
3.
by aligning tab and snapping back into place.
BATTERY INSIDE THE EVac I. REMOVE BATTERY IF THE EVac I
AN EXTENDED PERIOD OF TIME.
BATTERY LEVEL INDICATOR
The battery level indicator shows the relative
Page 6
point, it is necessary to replace the batteries (refer to the Battery Installation sectionabove)
If the battery power drops to the point where the EVac I
can no longer function accurately, thealarmwill
the alarm will beep 10 times and the power will turn off automatically. When the battery is
If the sensed vacuum pressure exceeds 25,000 Microns (3333.1 Pa, 33.331 mBar,25,000mTorr, or
helps to conserve battery life throughadvanced
will enter“SleepMode”in
in which the sensor and backlight are turned off, thus reducing the load on the batteryandextending the
will automatically turn itself off to furtherconservebattery
It is necessary to prevent oil from being drawn into the Vacuum Sensor. For HVAC/Rservice, always
-
on the vacuum pump prior to turning the vacuum pump off. Over time, oil vapor and other
has a built in Oil Sensor that will detect thiscondition.If the
key on the keypad. The appropriateunitsindicator
replaced, and the EVac I is turned back on, the EVac I will resume operation with all previous
settings intact.
HI-PRESSURE INDICATION
mTorr, or25.000 mmHg), the display will show .
SLEEP MODE
Unlike other Digital Vacuum Gauges, the EVac I
power management. After 5 minutes of displaying , the EVac I
extending the battery life up to 300 hours or more. While in this mode, the display will show
and exit Sleep Mode if the pressure is less than 25,000 Microns. The gauge can be manually
brought out of Sleep Mode by pressing any key on the keypad. While in Sleep, the gauge will
beep twice every 5 minutes to remind you that it is on. The EVac I will not sleep in the
Calibration Mode (see the Calibration section below).
. Approximately every 35 seconds, the gauge will automatically check the pressure
AUTO POWER-OFF
After 1 hour in “Sleep Mode” the EVac I
battery power.
OIL SENSOR
always recover refrigerant prior to attaching gauge. If possible, always close the blank
materials may contaminate the sensor. The EVac I
condition. If the OIL indicator in the display illuminates, this is an indication that the vacuum
sensor has been contaminated and is no longer functioning accurately. If the sensor becomes
completely saturated with oil to the extent that it cannot function properly at all, the message
described in the “Cleaning the Vacuum Sensor” section.
will show on the display. Prior to further use of the
EVac I, clean the sensor as
off valve
UNITS
Change the displayed units by pressing the
indicator on the LCD display will show Pa, mBar, mTorr, mmHg, or Microns depending on
the units set.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Page 7
RANGE AND RESOLUTION
Vacuum
Units
Bar
The EVac I has a broad vacuum pressure measurement range and very high resolution
according to the table below:
Range Resolution
Pascals (Pa)
—
3,333.1 0.1 Pa
0
Millibar (mBar)
Millitorr (mTorr)
mmHg (mmHg)
Microns (Microns)
—
33.331 0.001 mBar
0
—
25,000 1 mTorr
0
—
25.000 0.001 mmHg
0
—
25,000 1 Micron
0
“ANALOG” VACUUM LEVEL BAR GRAPH
The EVac I also displays a Vacuum Level Bar Graph allowing for a quick visual determination
of the vacuum level achieved. Each bar corresponds to a vacuum pressure range depending
on units according to the following table:
and the backlight will turn off automatically after 1 minute. To activate the Backlightpermanently,
indicator on the display will show solid. Turn off the backlightbypressing
indicator turns off. During Sleep Mode, the backlightwill turnoff tohelp
conserve battery power, but will turn on again automatically upon resumption of normaloperationifit
nternal speaker that will emit a beep for each valid key press, andalsofunctions
as an alarm in the Calibration mode (see the Calibration section below). It will alsobeepevery5
ressingandholding the
indicatoron theLCD
f the gaugeandwillallowitto
swivel freely in any direction. The gauge may be operated with or without the hookattached.When
fastening the hook to the gauge, turn it finger tight only. Use of a tool to tighten thehookmayresultin
should provide many years of service with no maintenance required.Whennotinuse, the
dust cap should remain in place over the sensor port. Clean the plastic enclosurewithadamp(not
ptable, but use no solvents. Take care not to expose thevacuum
sensor to oil. If the Oil Sensor (described above) indicates a contaminated sensor, follow theSensor
aminated with oil (as indicated by the Oil Sensor),carefully follow
OBJECT INTO THE PORT, AS THIS WILL PERMANENTLY DAMAGE THESENSOR
ir dry over at least an hour, or pull a vacuum on the sensortocleanitmore
Replace the battery and turn on the gauge. The Oil Indicator should be off.If it isstillon, repeat
alibration cycle as detailed in the Calibrationsectionbelow.
: It is important to remove all alcohol vapors from the sensor, either throughair
To activate the Backlight temporarily, press once. The indica
press again. The
repeatedly until the
was previously set in the permanent-on mode.
SOUND
The EVac I has an i
minutes in Sleep Mode. For silent operation, the sound can be muted by p
key. The Calibration mode alarm is not affected by the mute status. The
display indicates that the sound is on (not muted).
SWIVEL HOOK
The EVac I’s removable stainless steel swivel hook enables hanging o
damage to the EVac I case.
MAINTENANCE
The EVac I
wet) rag. Mild detergent is acce
Cleaning Procedure below.
CLEANING THE VACUUM SENSOR
If the vacuum sensor becomes cont
this procedure:
1. Turn the EVac I power off.
2. Shake the gauge to remove any large quantities of oil from the sensor.
3. Apply a few drops of rubbing alcohol inside the sensor vacuum port. (DO
4. Place your finger over the port and shake for a few moments.
5. Remove your finger and shake out the alcohol.
6. Repeat steps (3) – (5) at least three times.
7. Allow the sensor to a
quickly.
8.
the cleaning procedure.
9. If full accuracy is desired, perform a c
Note
vacuum. Any remaining vapors will cause an incorrect vacuum reading.
NOT INSERT ANY
-drying or via
).
Page 9
CALIBRATION TEST
should rarely require recalibration, though it may be necessary to know that yourgaugeis
calibrated properly for full accuracy. The Calibration Test mode assures you thatthe
if the instrument is calibrated properly.
if the instrument requires calibration.Pleaseseethe
If the Calibration Test indicates recalibration is required, the gauge may be recalibrated.Unlikeany
ed to factory specificationswithout any
For best results, clean the sensor with alcohol prior to calibration. Ensure thesensoris
y, install a fully charged battery intothegauge.
indicator should illuminate, an
should show on the Main Numeric Display. If not, turn the power off again andrepeat.
in a clear Ziploc (resealable zipper storage) bag, press outanyextraair, and
°
F). At this point, the alarm willsoundand the
mperature of at least23
F). At this point, the alarm will sound, and
is now calibrated to factory
to warm slowly. Attemptingto
accelerate the warming by using a heat source will not provide satisfactory results.Duringthe
ill be indicated on the Alternate NumericDisplayin
degrees Celsius. The calibration process may be canceled at any time by turningoff the
°
The EVac I
calibrated to factory specifications. Test the calibration as follows:
1.Turn the EVac I power off.
2.Important: Expose the EVac I to atmospheric pressure.
3.Press and hold (do not release) the key for about 5 seconds.
4.The display will show
5.The display will show
Calibration section below.
CALIBRATION
other vacuum gauge, the EVac I can be easily recalibrat
special equipment, with the following procedure:
1.
completely dry before proceeding.
2.Turn the EVac I power off. If necessar
3.Place the dust cap over the vacuum fitting.
4.Hold while pressing . As soon as the power turns on, release
press quickly at least three times. The
5.Place the EVac I
seal.
6.Place the bagged EVac I into a freezer with a temperature of less than –5
7.Allow the EVac I to cool to below -2°C (28.4°
display will change to .
8.Remove from freezer and press any key to silence the alarm.
9.Place the EVac I undisturbed in an area with a room te
no greater than 30°C (86° F).
10. Allow the EVac I to warm to 20° C (68°
show on the display.
11. Press any key to silence the alarm. The EVac I
EVac I is
and and then
d
C (23°F).
°C (73.4° F) but
will
specifications.
Note: For accurate calibration, it is necessary to allow the EVac I
cooling/warming process, the temperature w
previous calibration will be unchanged.
EVac I. The
Page 10
RESTORING FACTORY CALIBRATION
calibration can be restored at any time by the following procedure:
indicator should illuminate, a
should show on the Main Numeric Display. If not, turn the power off again andrepeat.
isnowreset to the
. Pleaseusethe table
Solution
Clean theVacuumSensor
Vac
vacuum portwithyourhand,
and turn the
Clean theVacuumSensor
and RestartCalibration
Restart Calibration.Allow the
Gauge to
Restart Calibration.Leave
the GaugeUndisturbed
During WarmingPhase.
C). Whileoperatingbelow
C), the display update rate will slow from 3.5 readings every second toonereading
A standard alkaline battey may not provide acceptable battery life at temperaturesbelow0
The original factory
1.Turn the EVac I power off.
2.Hold while pressing . As soon as the power turns on, release
press quickly at least three times. The
3.Press five times followed by .
4.The gauge will beep 5 times and return to normal operation. The EVac I
original factory calibration.
TROUBLESHOOTING
Under certain conditions, the display may read or
below to determine and fix the problem:
Display Mode Possible Problem
Normal
Operation
Calibration Sensor Contaminated
Calibration
Sensor Contaminated
Ambient Temperature too
Low
Gauged Warmed too
Quickly
Gauge Disturbed During
Calibration
Turn the e
and and then
nd
I off, warm the
eVac I back on.
Warm Slowly
LOW TEMPERATURE OPERATION
The EVac I can operate accurately at temperatures as low as 10°F (-12°
freezing (32°F/0°
every two seconds.
For low temperature operation, a 9V Lithium battery is recommended.
°C (32° F).
Page 11
NOTES
Page 12
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
4
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.