AC/DC Medical Open Frame
65 Watts
PM65 Series
Specification
Input
Input Voltage • 85 to 264 VAC
Input Frequency • 47 to 440 Hz
Input Current • 1.6 A (rms) for 115 VAC
1.0 A (rms) for 230 VAC
Safety Ground • 50 µA max at 110 VAC 60 Hz
Leakage Current • 100 µA max at 240 VAC 50 Hz
Output
Total output power: • 0 to 65 watts
Ripple and noise: • 1% pk-pk maximum
Overvoltage • Provided on output #1 only;
Protection: set at 112-132% of its nominal
output voltage
Overcurrent • All outputs protected to short
Protection: circuit conditions
Temperature • All outputs ±0.04%/°C
Coefficient: maximum
Transient Response • Maximum excursion of 4% or
better on all models, recovering
to 1% of final value 500 µs after
a 25% step load change
Line Regulation • ±0.5% max at full load
XPiQ inc.
Intelligent Design Quality Product
4000 VAC Isolation
•
100% Burn-in
Low Safety Leakage Current
Wide Input Voltage 85 to 264 VAC
Meets EMI Requirements of
Medical Equipment
Both Line and Neutral Fused for
European Requirements
General
Efficiency • 70% minimum at 65 watt output
Hold-up time • 10 msec minimum at nominal input
Inrush Current • 15 A at 115 VAC or 30 A at
Withstand Voltage • 4000 VAC from input to output
Insulation Resistance • 10 Mohm minimum from
Switching Frequency • 32 kHz (±5 kHz)
Environmental
Operating • 0 °C to +70 °C, derate linearly from
Temperature 100% load at +50 °C to 50% load
Storage • -40 °C to +85 °C
Temperature
Relative Humidity • 5% to 95% non-condensing
MTBF • 100,000 hours minimum at full load
EMC & Safety
EMI Requirement • Meets the conducted limits of
Safety Requirements • Approved to:
•
•
•
•
•
230 VAC at +25 °C cold start
1500 VAC from input to ground
500 VAC from output to ground
output to ground
at +70 °C
at +25 °C ambient
EN 55011 (CISPR11) Level B
(Not for patient contact)
a) UL2601
b) CSA C22.2 No.601.1-M90 per CUL
c) IEC 601-1/EN60601-1 (per TUV)
•
www.XPiQ.com
PH: 508 429.9883
FAX: 800 226.2100
Email: sales@xpiq.com
Holliston, MA 01746 USA
•
OUTPUT VOLTAGE & CURRENT RATINGS PM65
Maximum Output #1 Output #2 Output #3 Output #4 Model
Output Number
Power Vnom Imin Imax Tol. Vnom Imin Imax Tol. Vnom Imin Imax Tol. Vnom Imin Imax Tol.
50 W 5 V 0 A 10.0 A 2% PM65-10A
65 W 12 V 0 A 5.5 A 1% PM65-12A
65 W 15 V 0 A 4.5 A 1% PM65-13A
65 W 24 V 0 A 3.0 A 1% PM65-14A
65 W 30 V 0 A 2.5 A 1% PM65-16A
65 W 36 V 0 A 2.2 A 1% PM65-17A
65 W +5 V 1 A 6.0 A 3% +12 V 0.5 A 3 A 5% PM65-23A
65 W +5 V 1 A 6.0 A 3% +24 V 0.3 A 2 A 5% PM65-25A
65 W +5 V 1 A 6.0 A 3% +12 V 0.5 A 3 A 5% -5 V 0.1 A 0.5 A 10% PM65-30A
65 W +5 V 1 A 6.0 A 3% +12 V 0.5 A 3 A 5% -12 V 0.1 A 0.5 A 10% PM65-31A
65 W +5 V 1 A 6.0 A 3% +15 V 0.4 A 3 A 5% -15 V 0.1 A 0.5 A 10% PM65-32A
65 W +5 V 1 A 6.0 A 3% +12 V 0.5 A 3 A 5% -12 V 0.1 A 0.5 A 10% -5 V 0.1 A 0.5 A 0.1% PM65-40A
Notes
All multi-output units may be operated at no-load without damage. At no-load, output tolerance increases to 10%.
•
Pin Chart
Model 1 2 3 4 5 6 7 8
PM65-10A PM65-12A OUTPUT OUTPUT OUTPUT OUTPUT RETURN RETURN RETURN RETURN
PM65-13A PM65-14A #1 #1 #1 #1
PM65-16A PM65-17A
PM65-23A OUTPUT OUTPUT COMMON COMMON OUTPUT OUTPUT N.C. N.C.
PM65-25A #1 #1 RETURN RETURN #2 #2
PM65-30A PM65 -31A OUTPUT OUTPUT COMMON COMMON OUTPUT OUTPUT OUTPUT N.C.
PM65-32A #1 #1 RETURN RETURN #2 #2 #3
PM65-40A OUTPUT OUTPUT COMMON COMMON OUTPUT OUTPUT OUTPUT OUTPUT
#1 #1 RETURN RETURN #2 #2 #3 #4
Mechanical Details
www.XPiQ.com
Notes:
1. Dimensions shown in inches (mm).
2. Tolerance 0.02 (0.5) maximum.
3. Input connector mates with Molex housing 09-50-3031 and
Molex 2878 series crimp terminal.
4. Output connector mates with Molex housing 09-50-3081
and Molex 2878 series crimp terminal.
5. Weight: 380 grams approx.
PH: 508 429.9883
FAX: 800 226.2100
Email: sales@xpiq.com
Holliston, MA 01746 USA