XPiQ inc PM40-12AL, PM40-13AL, PM40-14AL, PM40-23AL, PM40-25AL Datasheet

...
40 Watts
PM40 Series
4000 VAC Isolation
100% Burn-in
Low Safety Leakage Current
Wide Input Voltage 85 to 264 VAC
Meets EMI Requirements of
Both Line and Neutral Fused for
European Requirements
Specification
All specifications typical at nominal line, full load and 25°C
Intelligent Design Quality Product
XPiQ inc.
Medical Switching Power Supply
Input
Input Voltage 85 to 264 VAC Input Frequency 47 to 440 Hz Input Current 1.20A (rms for 115 VAC
0.70A (rms) for 230 VAC Safety Ground 50µA max.at 110 VAC 60 Hz Leakage Current 100µA max. at 240 VAC 50 Hz
Output
Total output power: 0 to 40 watts Ripple and noise: 1% peak to peak maximum Overvoltage Provided on output #1 only;
Protection: set at 112-132% of its nominal
output voltage
Overcurrent All outputs protected to short Protection: circuit conditions
Temperature All outputs ±0.04%/°C coefficient: maximum
Transient response: Maximum excursion of 4% or
better on all models, recovering to 1% of final value 500us after a 25% step load change
Line Regulation ±0.5% max. at full load
General
Efficiency 65% minimum at 40 watt output Hold-up time 10 msec minimum at nominal input Inrush Current 15 amps at 115 VAC or 30 amps at
240 VAC at 25°C cold start
Withstand Voltage 4000 VAC from input to output
1500 VAC from input to ground 500 VAC from output to ground
Insulation Resistance • 10 Mohm minimum from
output to ground
Environmental
Operating 0°C to +70°C Temperature
Storage -40°C to +85°C Temperature
Relative Humidity 5% to 95% non-condensing MTBF 100,000 hours minimum at full load
at 25°C ambient
EMI Requirement Meets the conducted limits of
EN 55011 (CISPR11) Level B
Safety Requirements • Approved to: (Not for patient contact)
a) UL2601 b) CSA C22.2 No.601.1-M90 per CUL c) IEC 601-1/EN60601-1 (per TUV)
* Derate linearly from 100% load at 50°C to 50% load at 70°C
PH: 508 429.9883
FAX: 800 226.2100
Email: sales@xp-iq.com
Holliston, MA 01746 USA
www.XP-iQ.com
151
OUTPUT VOLTAGE & CURRENT RATINGS PM40
All multi-output units may be operated at no-load without damage. At no-load, output tolerance increases to 10%.
Notes
Mechanical Details
Pin Chart
Output #1 Output #2 Output #3 Maximum
Output
Model Vnom Imin Imax Tol. Vnom Imin Imax Tol. Vnom Imin Imax Tol. Power PM40-10AL 5 V 0.0 A 8.0 A 2% 40 W PM40-12AL 12 V 0.0 A 3.5 A 1% 40 W PM40-13AL 15 V 0.0 A 3.0 A 1% 40 W PM40-14AL 24 V 0.0 A 2.0 A 1% 40 W PM40-23AL +5 V 0.5 A 3.0 A 3% +12 V 0.2 A 2 A 5% 40 W PM40-25AL +5 V 0.5 A 3.0 A 3% +24 V 0.1 A 1 A 5% 40 W PM40-30AL +5 V 0.5 A 3.0 A 3% +12 V 0.2 A 2 A 5% -5 V 0.05 A 0.3 A 10% 40 W PM40-31AL +5 V 0.5 A 3.0 A 3% +12 V 0.2 A 2 A 5% -12 V 0.05 A 0.3 A 10% 40 W PM40-32AL +5 V 0.5 A 3.0 A 3% +15 V 0.2 A 2 A 5% -15 V 0.05 A 0.3 A 10% 40 W
Model 1 2 3 4 5 6
PM40-10AL PM40-12AL OUTPUT OUTPUT OUTPUT RETURN RETURN RETURN PM40-13AL PM40-14AL #1 #1 #1 PM40-23AL OUTPUT OUTPUT OUTPUT COMMON COMMON N.C. PM40-25AL #2 #1 #1 RETURN RETURN PM40-30AL PM40-31AL OUTPUT OUTPUT OUTPUT COMMON COMMON OUTPUT PM40-32AL #2 #1 #1 RETURN RETURN #3
Notes:
1. Dimensions shown in inches (mm)
2. Tolerance 0.02 (0.5) maximum
3. Input connector mates with Molex housing 09-50-3031 and Molex 2878 series crimp terminal
4. Output connector mates with Molex housing 09-50-3061 and Molex 2878 series crimp terminal
5. Weight: 300 grams approx.
PH: 508 429.9883
FAX: 800 226.2100
Email: sales@xp-iq.com
Holliston, MA 01746 USA
www.XP-iQ.com
152
Loading...