Welch Allyn Mobile Acuity LT User Manual

Page 1
Mobile Acuity
®
LT
Central Monitoring System
Installation Guide
Software Version 7.10.x
© 2010, Welch Allyn
REF 103830 Ver A DIR 80016134 Ver A 2010-06
Printed in USA
®
Welch Allyn
Welch Allyn
8500 SW Creekside Place Beaverton, Oregon 97008-7101 USA
, Acuity®, Propaq® and Micropaq® are registered trademarks of Welch Allyn. Welch Allyn is protected under various patents and patents pending.
Within USA, toll free: Phone: 1 800 535 6663 Phone Tech. Services: 1 800 289 2501
WorldWide: Phone: +1 315 685 4560 Phone Tech. Services: +1 503 530 7500 Fax Tech. Services: +1 503 526 4970
Page 2
Warnings
Note
Note
Note
Note
WARNING Before setup, inspect all equipment for any signs of damage, cracks, improper mechanical
function, exposed wires or bent connectors. Replace any damaged equipment.
WARNING Do not install the Mobile Acuity LT System within 1000 feet of another Acuity System that has the same wireless network name (ESSID). The network name is on the label of the system transport case. If two systems with the same network name are located in close proximity, wireless monitors could connect to either system, resulting in possible confusion about patient monitoring and patient care.
WARNING Position access points at least two feet away from where people are frequently located. Close installation could result in exposure to levels of radio frequency energy that are higher than recommended by the U.S. Federal Communications Commission (FCC).
WARNING Do not attempt to operate the Mobile Acuity LT Central Monitoring System or Micropaq or Propaq monitors without consulting the Directions For Use for those devices.
WARNING Do not mute or turn off the Mobile Acuity LT System laptop computer speakers. Reducing the volume could result in failure to notice life threatening audible alarms on the system.
Caution
Caution For uninterrupted network communication, connect the wireless controller to an uninterruptible power supply (UPS). A UPS lessens the risk of network exposure to electrical power disturbances. Network power disturbances can cause patient monitors to disconnect from the Mobile Acuity LT System.
Notes
If a wireless patient monitor operates outside of its wireless communication area, the monitor aggressively searches for connection with the central station and depletes its battery life at a faster rate. If it searches for an extended period of time (typically more than four hours), it can display a system fault error that necessitates turning the monitor off and on again.
Use only patient monitors that include on their back labels. Patient monitors without this symbol are incompatible with this Mobile Acuity LT System wireless network, and they will not connect to this system.
Place extra Micropaq monitor batteries in the Micropaq monitor battery charger so that they are ready for use when needed. Full charge occurs within four hours at room temperature. Recharge Micropaq monitor batteries at least annually to ensure that they are at least partially charged.
Use a reliable AC power source to help prevent interruption in patient monitoring.
2
Page 3
1. Locate AC power source, and set up and connect equipment
AC power
source
Wireless controller
Laptop computer
Micropaq monitors
Battery charger for Micropaq monitor
Wireless Propaq CS monitor
Laptop power supply
10-foot data
cable to
controller port 7
Propaq LT monitor
Charging cradle for Propaq LT monitor
Access points (APs)
50-foot data cables to controller ports 0-2
Place 6 to 40 feet apart. Place patient monitors no more than 40 feet from an AP.
Battery charger for Wireless Propaq CS monitor
Uninterruptible power supply (UPS)
3
Page 4
2. Apply power
Note
Note
4. Connect patient monitors
a. Verify that an AC power source is on and that
all power cords are plugged in.
b. Apply power to the laptop computer.
Connect the laptop computer to AC power, to avoid loss of monitoring.
c. Apply power to the wireless controller.
Do not yet install batteries in Micropaq monitors or apply power to patient monitors.
3. Check system operation
Confirm that the system is operating properly.
a. Verify that LEDs are active (power is
supplied) on these devices:
• Laptop computer
• Wireless controller
• Micropaq monitor battery charger
WARNING Before you use a Propaq CS monitor on a new patient, always turn off the monitor for a few seconds, and then turn it on again. This clears the prior patient’s trends and alarm limit settings.
a. Apply power to each patient monitor.
• Micropaq monitor: Insert a charged battery.
• Propaq LT monitor: Press power
switch, and then press .
• Wireless Propaq CS monitor: Press the right-side power switch.
b. Confirm that within two minutes, the Mobile
Acuity LT System screen displays a yellow Waveform Window for each patient monitor.
c. Click the ID icon in the Enter or Confirm
Patient ID Waveform Window.
• Propaq LT monitor power adapter and cradle
• Propaq CS monitor power adapters
b. Wait five minutes. On the front and left side
of each AP, confirm that the first three LED lights are on continuously.
c. Confirm that the Mobile Acuity LT System
display shows a window similar to the following, and verify that a Check Wireless message is not displayed.
.
d. Verify that this four-digit number matches the
last four digits of the patient monitor serial number.
e. Type the patient ID, patient name and room
number.
f. Click Confirm.
4
Page 5
5. Shut down and remove power
Note
a. Notify staff members, so that procedural
changes in patient care can be put into place.
b. Move the pointer into the gray area in the
upper right corner of the screen, hold down
the key on the keyboard, and press the right mouse key.
The System Menu appears.
c. Click Shutdown Workstation.
d. After the system shuts down, turn off power
to all equipment.
Troubleshooting
If the following suggestions do not solve a problem, please contact the Welch Allyn Technical Support department.
Symptom Possible Cause and Suggested Response
There is no display on the Mobile Acuity LT System screen.
A Check Wireless message appears on the Mobile Acuity LT System screen.
OR
Patient monitors drop out in particular locations in the facility.
Proceed through these steps, as necessary.
1. Check power connections to the laptop computer.
2. Ensure that the laptop computer power switch is in the on position.
3. Shut down and restart the entire system. For instructions, see “5. Shut down and remove power”.
• There are wireless connectivity problems.
Proceed through these steps, as necessary.
1. Ensure that the wireless controller is connected to AC power.
2. Ensure that the AP data cables are firmly connected to access points and the wireless controller.
3. Ensure that the AP antennae are oriented vertically.
4. On the front and left side of each AP, confirm that the first three LED lights are on continuously.
5. Turn off and on again the wireless controller.
• Particular obstructions can block wireless communication, or a patient monitor can be beyond the range of an AP.
Typical obstructions include large metal objects, such as cabinets, vehicles or metal-reinforced walls, or large concentrations of water (such as water storage containers).
1. Mount the APs in a higher location so
that they have an unobstructed “line-of-sight” path to each patient monitor.
2. Move patient monitors closer to access
points.
Symptom Possible Cause and Suggested Response
A patient monitor does not connect to the Mobile Acuity LT System.
Proceed through these steps, as necessary.
1. Check the monitor’s back label for this symbol, which indicates that the patient monitor contains the proper
radio card. If a patient monitor does not include this symbol, it is incompatible with this Mobile Acuity LT System wireless network.
2. If a Propaq LT monitor does not display the wireless connection icon (either or ), the monitor might be configured so that the wireless connection is disabled (the radio card is off). Refer to the Propaq LT Monitor Directions For Use for information about enabling it.
3. Ensure that each patient monitor is within 40 feet of an access point. If the monitor is not communicating with an access point, appears on the monitor.
4. Proceed through the steps in the prior row of this table to ensure that the access points are functional.
5. The patient monitor might not be configured with the wireless network name (ESSID) for the Mobile Acuity LT System.
a. On the laptop computer, find the label
indicating the network name for the Mobile Acuity LT System (similar to “com8.protocol”).
b. Check the patient monitor network
name, and confirm that it matches the laptop computer network name. Change it if necessary. For instructions, see “Identify or change a monitor’s network name (ESSID)”.
c. Retest wireless communication with
the patient monitor.
6. If any nearby Acuity System has the same wireless network name (ESSID), the patient monitor might be connected to the other system.
a. If you suspect that another nearby
Acuity System has the same network name, immediately notify the system administrator about the problem. The Mobile Acuity LT System must be moved to a different location where there is no conflict in monitoring.
b. Notify staff members. Make appropriate
changes in patient care for discontinued monitoring.
c. Follow appropriate procedures to shut
down your Mobile Acuity LT System and remove power. For instructions, see “5. Shut down and remove power”.
For more information about troubleshooting a patient monitor, refer to the monitor Directions for Use.
5
Page 6
Identify or change a monitor’s
Note
network name (ESSID)
Identify or change a Micropaq monitor network name
This section describes how to identify or change the extended service set identifier (ESSID) for each type of wireless patient monitor that can connect to an Acuity System wireless local area network (WLAN).
This information is useful if you want to do these things:
Set up a Mobile Acuity LT System network.
Move a patient monitor from one Acuity System network to another.
Add a patient monitor to an Acuity System network.
Especially among networked Acuity Systems, verify a monitor’s ESSID.
An Acuity System’s network name appears on a label in these places:
For a Mobile Acuity LT System: on the system case
For an Acuity System: on the CPU
Changing a monitor’s network name causes the monitor to restart, erase all patient data and restore the monitor settings to default settings.
WARNING Do not attempt to change a monitor’s network name unless you are a qualified biomedical service engineer or technician.
Confirm that the network name that you select matches the network name of the appropriate Acuity System. When you select a new ESSID, the patient monitor seeks a connection with any nearby Acuity System network with that same ESSID.
If you select a wrong network name, patient alarms and data for any patient connected to the monitor would occur at an unexpected Acuity Central Station.
To identify a Micropaq monitor network name
1. Press to highlight
SYSTEM INFORMATION, and then press
.
The System Information menu appears.
2. Press to highlight NETWORK STATUS,
and then press .
The Network Status screen appears.
3. Locate the Network Name in the list.
4. Press to exit.
To change Micropaq monitor network name
1. Press repeatedly until SERVICE MENU
is highlighted, and then press .
The Service menu appears.
2. In this sequence and not too rapidly, press and hold these keys:
, and .
An ARE YOU SURE? prompt appears.
3. Press to highlight YES, and then press
.
A Network Name menu appears.
4. Press or to highlight the desired network name.
5. Carefully confirm that the highlighted name matches the appropriate Acuity System network name.
6. Press .
The Micropaq monitor turns off and on again, and then it attempts to connect to the Acuity System network.
6
Page 7
Identify or change a Propaq LT monitor network name
Identify or change a wireless Propaq CS monitor network name
To identify a Propaq LT monitor network name and prepare to change it (if desired)
1. Highlight SpO2, and then press .
2. Highlight Setup, and then press .
The Setup menu appears.
3. Highlight Service, and then press .
The Service menu appears.
4. Highlight Radio, and then press .
The Radio menu appears, and Show Info is highlighted
5. Take one of these steps:
If you want to change the network name: Continue below to “To proceed to change a Propaq LT monitor network name”.
If you do not want to change the network name:
a. Press .
The current network name is shown.
b. Highlight Exit, and then press .
You have completed the network name identification process.
To proceed to change a Propaq LT monitor network name
1. Highlight NetName and press .
To identify a wireless Propaq CS monitor network name and prepare to change it (if desired)
1. Touch , SETUP, MORE, MORE, SERVICE, YES, MORE, MORE, RADIO.
The Network Status screen opens, and the NET NAME is listed.
2. Take one of these steps:
• If you want to change the network name, continue below to “To proceed to change a wireless Propaq CS monitor network name”.
• If you do not want to change the network
name, touch to return to the main menu.
To proceed to change a wireless Propaq CS monitor network name
1. Touch CHANGE NET NAME.
The Change Net Name screen appears.
2. Touch NEXT repeatedly until the new
network name is highlighted.
3. Carefully confirm that the highlighted name matches the appropriate Acuity System network name.
4. Touch SELECT.
2. Highlight the new network name.
3. Carefully confirm that the highlighted name matches the appropriate Acuity System network name.
4. Simultaneously press and , and hold for one second.
The Propaq LT monitor turns off and on again, and then it attempts to connect to the Acuity System network.
An ARE YOU SURE? confirmation screen appears.
5. Respond to the confirmation prompt.
If you touch YES, the Propaq CS monitor turns off and on again, and then it attempts to connect to the Acuity System network.
Technical support
United States 1 800 289 2501
Worldwide +1 503 530 7500 Ask for Technical Support.
7
Page 8
8
Loading...