Welch Allyn Fiber optic laryngoscope handles User Manual

©2014 Welch Allyn, Inc. MM 721462 Ver.
1
C
Intended
About this
use
The laryngoscope blades which are used to examine placement of a tracheal
document
This directions for use apply to Welch Allyn reusable fiber optic laryngoscope handles: 60813, 60814, 60815. Welch Allyn with MacIntosh
Warnings and
WARNING: R use.
WARNING: The reprocessing procedure and the equipment and materials described must be followed and conducted by persons trained and familiar reprocessing.
WARNING: Consult cleaning and disinfecting agent manufacturer proper preparation and use.
WARNING: Repeated inspection procedures to assure damage has not occurred to the handle.
WARNING: High level disinfection and/or sterilization are not achieved by these methods.
WARNING: WARNING: Discard any component that shows evidence of damage or
deterioration. WARNING: Do not modify this equipment. Any modification of this
lead to patient injury. Any modification of this equipment voids the WARNING: Personnel shall
appropriate personal protective equipment when handling equipment.
WARNING:
fiber optic laryngoscope handle is an accessory used w i t h co m p a t i b l e ri g i d f i b e r op t i c
and visualize
tube.
reusable
Welch Allyn fiber optic laryngoscope blades MacIntosh
REF:
6921X, and Miller REF: 6806X.
fiber optic laryngoscope handles may be used
a patient’s airway
REF:
6906X, English
and
aid
cautions
eusable
fiber optic
laryngoscope handles
must be reprocessed
after
with
medical device
instructions
reprocessing may
Lamps, if left illuminated, could generate sufficient heat to cause burns
follow their facility policies
Laryngoscope equipment is not suitable for use in intense
degrade
elements of the handle.
product warranty.
and
procedures
potentially
and
contaminated
magnetic
for their
Follow
equipment may
wear
fields
each
CAUTION: damage o r poor performance of the handle.
CAUTION: If the device will be unused for several months or longer, remove batteries prior to storing the device.
Only use lamp specified. Failure
to follow these instructions may cause
the
s
2
Directions for use
Welch Allyn fiber optic laryngoscope handle
Reprocessing
instructions
These reprocessing instructions refer to procedures for cleaning and intermediate level disinfection. Fiber optic laryngoscope handles must be reprocessed prior to first use and between each use using
• Cleaning
the
following method as outlined in this
and intermediate level
disinfection
Welch Allyn has validated the above instruction as being capable of preparin g these laryngoscope handles for re-use. The user must ensure that the reprocessing as actually performed by the user's personnel, with the user's equipment and materials, achieves the desired result. This may require validation and routine monitoring of the user's actual proce ss
NOTE: The main handle and bottom cap components are compatible with the disinfection solution and autoclave methods identified which are provided for facilities who wish to perform either method after cleaning and intermediate level disinfection.
Cleaning and intermediate level disinfection instructions
Point of use: 1.
Preparation
for
decontamination:
Separate
blade assembly from handle and place handle into suitable containment for subsequent reprocessing (per f igure 1). Do not place handle with sharp devices.
2. Prevent the handle from drying (i.e. wrap/cover in moist germicidal
wipe).
1. Select an appropriate quaternary ammonium isopropanol based germicidal cleaner labeled suitable for use on healthcare equipment and capable intermediate level disinfection. Reference EPA­registered disinfectants: http://www.epa.gov/oppad001/chemregindex.htm Outside of the U.S., please consult applicable regulatory body for equivalent quaternary ammonium isopropanol germicidal cleaner.
2. Remove batteries and lamp cartridge (per figure 2).
document:
Directions for use
Cleaning intermediate level disinfection:
and
Welch Allyn fiber optic laryngoscope handles
3
1. Follow the ge r m i c i dal w i p e manufacturer’s instructions to clean all exposed surfaces of the main handle, end cap, and lamp cartridge.
2. If necessary, brush with a dry, soft-bristled brush and re-wipe to loosen/remove excessive visible soil.
3. After all visible soil is removed, re-wipe to wet all surfaces and allow adequate contact time for disinfection as directed by the germicidal wipe manufacturer.
CAUTION: Only use quaternary ammonium
isopropanol based germicidal wipes.
Drying:
Maintenance, Inspection and Testing
Storage
End of
reprocessing
Allow components to air dry.
1. Inspect
each
component
area (per f
igure 3) for
damage
or
deterioration.
WARNING: Discard any component that
shows
evidence of damage or deterioration.
Contact Welch Allyn for component replacement.
2. Reassemble handle (per figure 4) with new or batteries in known good condition.
3 . Attach handle to a clean and disinfected test blade
assembly in known working condition. Verify
Blade assembly engages and locks onto handle.
Blade assembly deploys into its locked position on
handle
Light output is satisfactory.
If the la mp fails to replace the lamp and/or
AND lamp illuminates.
light or output is low
batteries.
that:
, check or
Store handle per facility practice to allow device to remain clean, dry, and ready for service.
instructions for intermediate level disinfection.
Loading...
+ 4 hidden pages