CP 100 and CP 200™ 12-Lead
Resting Electrocardiograph
TM
CP 100TM 12- Lead Resting Electrocardiograph
CP 200
12- Lead Resting Electrocardiograph
Service Manual
iiWelch Allyn CP 100 and CP 200 Electrocardiograph
Copyright 2005, Welch Allyn, Inc. All rights are reserved. No one is permitted to reproduce or duplicate, in
any form, this manual or any part thereof without permission from Welch Allyn.
Caution: Federal US law restricts sale of the device identified in this manual to, or on the order of, a
licensed physician.
Welch Allyn assumes no responsibility for any injury, or for any illegal or improper use of the product, that
may result from failure to use this product in accordance with the instructions, cautions, warnings, or
indications for use published in this manual.
Welch Allyn is a registered trademark of Welch Allyn, Inc., and CP 200, CP100, and CardioPerfect are
trademarks of Welch Allyn, Inc.
SD is a trademark of Toshiba.
Software in this product is Copyright 2005, Welch Allyn, Inc., or its vendors. All rights are reserved. The
software is protected by United States of America copyright laws and international treaty provisions
applicable worldwide. Under such laws, the licensee is entitled to use the copy of the software
incorporated within this instrument as intended in the operation of the product in which it is embedded.
The software may not be copied, decompiled, reverse-engineered, disassembled or otherwise reduced to
human-perceivable form. This is not a sale of the software or any copy of the software; all right, title and
ownership of the software remains with Welch Allyn or its vendors.
For information about any Welch Allyn product, please call Welch Allyn Technical Support:
USA 1 800 535 6663
+1 315 685 4560
Canada 1 800 561 8797China +86 216 327 9631
European Call Center +353 46 906 7790France +33 15 569 5849
Germany +49 747 792 7186Japan +81 33 219 0071
Latin America +315 685 2644Netherlands +31 15 750 5000
Singapore +65 6419 8100South Africa +27 11 777 7555
United Kingdom +44 207 365 6780Sweden +46 85 853 6551
Australia +61 29 638 3000
800 074 793
Part Number (manual only): 704340 Rev D
Welch Allyn
PO Box 220
Skaneateles Falls, NY 13153-0220
2Safety SummaryCP 100 and CP 200 Electrocardiograph
Introduction
The safety summary, and all additional specific warnings and cautions located throughout
the documentation, must be read and understood by all users of the CP 100 and CP 200
Electrocardiograph.
Symbols
The symbols illustrated on the following pages may appear on the electrocardiograph, the
packaging, the shipping container, or in this manual.
Documentation Symbols
WARNING Indicates conditions or practices that could lead to illness, injury, or
death.
Caution In the manual, indicates conditions or practices that could damage the
equipment or other property.
Caution On the product, means “Consult accompanying documentation.”
-20°C
EC REP
Shipping, Storing, and Environment Symbols
This end upKeep dry
FragileAltitude limits
+49°C
Temperature limitsRelative humidity limit
Certification Symbols
Meets essential requirements of European
Medical Device Directive 93/42/EEC
European Regulatory Manager
Welch Allyn LTD.
Navan Business Park
Dublin Road
Navan, County Meath, Republic of Ireland
Tel.: 353-46-90-67700
Fax: 353-46-90-67756
N344
Complies with applicable U.S. and Canadian
medical safety standards
Australian registered importer
Directions for UseSafety Summary3
Operation Symbols
For details on the keys, see “Keyboard Functions” on page 14.
On/standby (off)Help
Navigation arrowsBackspace
EnterShift
Auto ECGMenu
Rhythm ECGStop/Cancel
SD memory card slotCom port A
(for patient cable)
Alternating currentBattery charge level
Dangerous voltageDirect current
Defibrulation-proof type CF applied parts
Ground equipotential
(while the electrocardiograph is safety-rated
“CF” for direct cardiac contact it is not
intended to be connected directly to the
patient’s heart; only surface contact with the
patient’s skin is intended).
Do not reuse.Sealed lead-acid battery
Type BF Defibrillator proofRecycle
Do not dispose of this product as unsorted
municipal waste. Prepare this product for
reuse or separate collection as specified by
T2.0A/250V
AC fuse replacement information
Directive 2002/96/EC of the European
Parliament and the Council of the European
Union on Waste Electronic and Electrical
Equipment (WEEE). If this product
is
contaminated, this directive does not apply.
For more specific disposal information, see
www.welchallyn.com/weee,
or contact Welch Allyn Customer Service at
+44 207 365 6780.
Additional Operation Symbols for CP 200
Spirometry KeyOK
Spirometry portCom port B
(for USB cable)
4Safety SummaryCP 100 and CP 200 Electrocardiograph
Servicing the Electrocardiograph Safely
Before using or servicing the electrocardiograph, you must read and understand the
following safety-related information.
General Warnings
The following warning statements apply to electrocardiograph use in general. Warning
statements that apply specifically to particular procedures, such as connecting the patient
cable or performing an ECG test, appear in the corresponding sections of the manual.
Warning statements indicate conditions or practices that could lead to illness, injury, or
death.
Warnings Related to the Environment
WARNING To avoid a possible explosion, do not use the electrocardiograph in
the presence of flammable anesthetics.
WARNING When transporting the electrocardiograph on a cart, tuck the
patient cable away from the wheels so that it does not present a hazard.
WARNING Do not use the CP 100 and CP 200 Electrocardiograph in an MRI
suite or hyperbaric chamber.
WARNING Do not autoclave the CP 100 and CP 200 Electrocardiograph or
patient cable.
Warnings Related to Accessories and Other Equipment
WARNING For operator and patient safety, peripheral equipment and
accessories that can come in direct patient contact must be in compliance with
all appropriate safety, EMC, and regulatory requirements. See “EMC Guidance
and Manufacturer’s Declarations” on page 79
WARNING All signal input and output (I/O) connectors are intended for
connection of only devices complying with IEC 60601-1, or other IEC standards
(for example, IEC 60950), as appropriate to the device. Connecting additional
devices to the electrocardiograph may increase chassis or patient leakage
currents. To maintain operator and patient safety, consider the requirements of
IEC 60601-1-1. Measure the leakage currents to confirm that no electric shock
hazard exists.
WARNING The electrocardiograph has not been designed for use with highfrequency (HF) surgical equipment and does not protect against hazards to the
patient.
Directions for UseSafety Summary5
Warnings Related to Using the Electrocardiograph
WARNING Avoid positioning any leads or cables so that they could easily trip
someone.
WARNING Satisfactory maintenance procedures must be implemented, or
equipment failure and health hazards may result.
Warnings Related to Repairing the Electrocardiograph
WARNING Only qualified service personnel should attempt to repair the
electrocardiograph. In case of a malfunction, call Technical Support and precisely
describe the problem. For phone numbers, see page ii.
WARNING While under warranty, the electrocardiograph must be serviced
only by a Welch Allyn service technician.
WARNING Electrostatic discharge (ESD) can damage or destroy electronic
components. Handle static-sensitive components only at static-safe
workstation.
WARNING Consider all electrical and electronic components of the monitor as
static-sensitive.
Electrostatic Discharge (ESD)
Electrostatic discharge is a sudden current flowing from a charged object to another
object or to ground. Electrostatic charges can accumulate on common items such as
foam drinking cups, cellophane tape, synthetic clothing, untreated foam packaging
material, and untreated plastic bags and work folders, to name only a few.
Electronic components and assemblies, if not properly protected against ESD, can be
permanently damaged or destroyed when near or in contact with electrostatically charged
objects. When you handle components or assemblies that are not in protective bags and
you are not sure whether they are static-sensitive, assume that they are static-sensitive
and handle them accordingly.
•Perform all service procedures in a static-protected environment. Always use
techniques and equipment designed to protect personnel and equipment from
electrostatic discharge.
•Remove static-sensitive components and assemblies from their static-shielding bags
only at static-safe workstations—a properly grounded table and grounded floor mat—
and only when you are wearing a grounded wrist strap (with a resistor of at least 1
megohm in series) or other grounding device.
6Safety SummaryCP 100 and CP 200 Electrocardiograph
•Use only grounded tools when inserting, adjusting, or removing static-sensitive
components and assemblies.
•Remove or insert static-sensitive components and assemblies only with monitor
power turned off.
•Insert and seal static-sensitive components and assemblies into their original staticshielding bags before removing them from static-protected areas.
Always test your ground strap, bench mat, conductive work surface, and ground cord
before removing components and assemblies from their protective bags and before
beginning any disassembly or assembly procedures.
General Cautions
The following caution statements apply to electrocardiograph use in general. Caution
statements that apply specifically to particular procedures, such as connecting the patient
cable or performing an ECG test, appear in the corresponding sections of the manual.
Caution statements indicate conditions or practices that could damage the equipment or
other property.
Caution When removing the electrocardiograph from storage, allow it to
thermally stabilize to surrounding environmental conditions before using it.
Caution To prevent possible damage to the keypad, do not use sharp or hard
objects to press keys. Only use fingertips.
Caution Do not expose the patient cable to strong ultra-violet radiation.
Caution Do not pull or stretch the patient cable. Doing so could result in
mechanical or electrical failures. Form the patient cable into a loose loop before
storing.
Caution Avoid positioning the patient cable where it might get pinched or
stepped on. If the cable’s impedance is altered, measurements might no longer
be accurate, and repair might be necessary.
Caution Using the equipotential terminal for anything but grounding purposes
may contribute to damage of the device.
Caution Use only parts and accessories supplied with the device and available
through Welch Allyn. The use of accessories other than those specified may
result in degraded performance of this device.
Caution Portable and mobile RF communications equipment can affect the
performance of the electrocardiograph.
Caution Other medical equipment—including but not limited to defibrillators,
ultrasound machines, pacemakers, and other stimulators—may be used
simultaneously with the electrocardiograph. However, such devices may disturb
the electrocardiograph signal.
8OverviewWelch Allyn CP 100 and CP 200 Electrocardiograph
Purpose and Scope
The service manual is a reference for periodic preventive maintenance and corrective
service procedures for the Welch Allyn CP 100 and CP 200 Electrocardiograph.
Corrective service is supported to the level of field-replaceable units. This includes some
circuit-board assemblies and some subassemblies, case parts, and other parts. See “Field
Replaceable Parts” on page 69 for complete list of user-replaceable service parts.
Note
This guide provides troubleshooting information, assembly procedures, and instructions
for functional testing and performance verification. It is intended for use only by
technically qualified service personnel.
Repair and replacement of the main board is not supported. All service work on
the main board must be performed by certified and qualified service personnel at
an authorized Welch Allyn service center.
Caution No component-level repair of circuit boards and subassemblies is
supported. Use only the repair procedures described in this manual.
WARNING When performing a service procedure, follow the instructions
exactly as presented in this manual. Failure to do so could damage the monitor,
invalidate the product warranty, and lead to serious personal injury.
Service Options
Warranty Service
All repairs on products under warranty must be performed and/or approved by Welch
Allyn. Refer all warranty service to Welch Allyn Factory Service or another authorized
Welch Allyn Service Center. Obtain an RMA number for all returns to Welch Allyn Factory
Service – see Returning Products (page 9).
Caution Unauthorized repairs will void the product warranty.
Non-Warranty Service
Welch Allyn Factory Service and authorized Service Centers support non-warranty repairs.
Contact any Welch Allyn regional service center for pricing and service options.
Welch Allyn offers modular repair parts for sale to support non-warranty service. This
service must be performed only by qualified end-user biomedical/clinical engineers using
this service manual.
The Welch Allyn Electrocardiograph Service Information supports certain service
functions. For information about any Welch Allyn product, please call Welch Allyn
Technical Support: (page ii).
Service ManualCP 100 and CP 200 Electrocardiograph9
Technical Support Services
Welch Allyn offers the following technical support services:
Telephone support
Loaner equipment
Service agreements
Replacement service parts
Factory Service
For information on any of these services, contact Welch Allyn at the numbers listed on
page ii
Returning Products
To return a product for service, contact Welch Allyn Technical Support and request a
Return Material Authorization (RMA) number.
Note
To ship the unit, please observe these packing guidelines:
Welch Allyn does not accept returned products without an RMA.
•When requesting an RMA, please have the following information available:
•Product name, model number, and serial number
•A complete return shipping address, including a contact name and phone
number; include any special shipping instructions
•A purchase-order number or credit-card number if the product is not covered by
warranty
•A full description of the problem or service request
•Remove from the package all hoses, connectors, cables, sensors, power cords,
and other ancillary products and equipment, except those items that might be
associated with the problem.
•Use the original shipping carton and packing materials, or as close an
approximation as possible.
•Include a packing list.
•Write the Welch Allyn RMA number with the Welch Allyn address on the outside
of the shipping carton.
United States federal regulations require that any unit received by Factory Service must
be free from blood-borne pathogens before processing. All incoming products are cleaned
as well as possible, but products that cannot be effectively cleaned cannot be accepted
for repair. Please thoroughly clean all organic residues from the product before shipment.
This will ensure safe receipt, processing and repair, and will help expedite the return of
your device.
10OverviewWelch Allyn CP 100 and CP 200 Electrocardiograph
Product Configurations
Model numbers for the configurations are as follows:
Troubleshooting and repair followed by functional test
Return to authorized service center
Replacing the Battery or Fuse (page 58)
Troubleshooting (page 41)
Disassembly and Repair Procedures (page
47)
Functional Verification (page 29)
Returning Products (page 9)
Service ManualCP 100 and CP 200 Electrocardiograph11
Controls, Indicators, and Connectors
This section describes the controls, indicators, and connectors that are part of the
electrocardiograph.
Figure 1. Top of CP 200 Electrocardiograph
Softkeys and functions keys
See Figure 10 on page 15.
Keyboard
See Figure 8 on page 14.
Figure 2. Top of CP 100 Electrocardiograph
Keyboard
See Figure 9 on page 14.
12OverviewWelch Allyn CP 100 and CP 200 Electrocardiograph
Figure 3. Back of Electrocardiograph
AC power inlet
Equipotential stud
AC fuses
Figure 4. Right Side of Electrocardiograph
Com port B
(for USB cable)
SD memory card
slot
(Not functional on CP 100)
Figure 5. Left Side of Electrocardiograph
Paper tray latch
Spirometry port
(Not functional on CP 100)
Reset
Button
Service ManualCP 100 and CP 200 Electrocardiograph13
Figure 6. Front of CP 200 Electrocardiograph
Com port A
(for patient cable)
Figure 7. Front of CP 100 Electrocardiograph
Com port A
(for patient cable)
14OverviewWelch Allyn CP 100 and CP 200 Electrocardiograph
Keyboard Functions
Figure 8. Keyboard for CP 200 Electrocardiograph
On/Off
See “ Powering the
Electrocardiograph” in the
Directions of Use manual.
Green LED
Lights up when the
electrocardiograph is
connected to AC
power.
Ta b
Moves through the
data-entry fields.
Shift
Capitalizes letters.
Space
Enters a space.
Figure 9. Keyboard for CP 100 Electrocardiograph
Backspace
Deletes the character to the
left of the cursor.
Enter
See “Moving Through the Menus” on
page 18.
Menu
See “About the Main Menu”
on page 17.
Help
See the Directions For
Use manual.
Navigation arrows
See “Moving Through the
Menus” on page 18.
Green LED
Lights up when the
electrocardiograph is
connected to AC
power.
Ta b
Moves through the
data-entry fields.
Help
See the Directions
for Use manual.
Shift
Capitalizes letters.
On/Off
See “ Powering the
Electrocardiograph” in the
Directions of Use manual.
Space
Enters a space.
Stop/Cancel
Stops any current activity.
See “Moving Through the
Menus” on page 18.
Backspace
Deletes the character to the
left of the cursor.
Enter
See “Moving Through the
Menus” on page 18.
Menu
See “About the Main
Menu” on page 17.
Auto ECG
Begins Auto ECGs,
normal and stat.
Rhythm ECG
Begins a Rhythm ECG.
.
Navigation arrows
See “Moving Through the
Menus” on page 18.
Service ManualCP 100 and CP 200 Electrocardiograph15
Figure 10. Softkeys and Function Keys for the CP 200 Electrocardiograph
Softkeys
These softkeys display text or images that
correspond to the unlabeled buttons below
them. The content changes from screen to
screen.
Softkey buttons
These buttons activate the functions displayed
above them. If a softkey is blank, pressing its
Auto ECGSpirometryRhythm ECGStop/CancelOK
button has no effect.
Function Keys
Auto ECG
Begins Auto ECGs, normal and stat.
Spirometry
Begins spirometry tests.
See spirometry manual.
Rhythm ECG
Begins Rhythm ECGs.
Stop/Cancel
Stops any current activity.
See “Moving Through the Menus” on page 18.
OK
Accepts data that you have entered, or chooses a highlighted item.
See “Moving Through the Menus” on page 18.
16OverviewWelch Allyn CP 100 and CP 200 Electrocardiograph
“System Settings” Menu Tree
P 200 Menu Tree
System Settings
Device
Configuration
Set Date/Time
Language
Date Format
Time Format
Weight Unit
Height Unit
Power-Save
Device ID
Audio Beeper
Flow Unit
Pressure Unit
Temperature
Increase Display Contrast
Decrease Display Contrast
Contact Information
Previous Menu
* Applicable for spirometry only.
*
*
*
CP 100 Menu Tree
Device
Info
About
Print Settings
Enable Options
Upgrade Software
Service Info
Previous Menu
System Settings
Device
Administration
See “Managing Data” in the
Directions for Use manual.
Service Info
Print Log
Battery Life Test
Printer Test
Export Files
Import Files
Delete Files
Serial Number
Lead Off Thershold
Service Tests
LCD Test
Previous Menus
Device
Configuration
Set Date/Time
Language
Date Format
Time Format
Weight Unit
Height Unit
Power-Save
Device ID
Audio Beeper
Increase Contrast
Decrease Contrast
Contact Information
Previous Menu
Device
Info
About
Print Settings
Enable Options
Upgrade Software
Service Info
Previous Menu
Service Info
Print Log
Battery Life Test
Printer Test
Export Files
Import Files
Delete Files
Serial Number
Lead Off Thershold
Service Tests
Previous Menus
Service ManualCP 100 and CP 200 Electrocardiograph17
About the Main Menu
The main menu appears when you press the Menu key.
Figure 11. Main Menu CP 200
Main Menu
1 Test Directory
2 Scheduled Patients
3 ECG Settings
4 Spirometry Settings
5 System Settings
6 Edit Medication List
7 Edit History List
0 Exit
Figure 12. Main Menu CP 100
Main Menu
1 ECG Settings
2 System Settings
0 Exit
9:17AM Mar 16 05
9:17AM Mar 16 05
SubmenuPurposeProcedure
Test DirectoryView, change, print, or send saved tests. (CP
200 Only)
Scheduled PatientsView the scheduled patients list, add
patients to the list, or delete patients from
See the Directions for Use manual on your
specific model.
See the Directions for Use manual on your
specific model.
the list. (CP 200 Only
ECG SettingsReview or change ECG settings: Auto
Report format, Rhythm Report format, and
See the Directions for Use manual on your
specific model.
so on.
Spirometry SettingsReview or change spirometry settings:
display settings, print settings, and so on. (CP
See the Directions for Use manual on your
specific model.
200 Only
System SettingsReview or change system settings: device
configuration, device info, user setup, and
See the Directions for Use manual on your
specific model.
so on.
Edit Medication ListEdit the list of medication choices available
to choose during patient data entry. (CP 200
See the Directions for Use manual on your
specific model.
Only
Edit History ListEdit the list of clinical conditions available to
choose during patient data entry. (CP 200
See CP 200 Electrocardiograph Directions for Use manual.
Only
18OverviewWelch Allyn CP 100 and CP 200 Electrocardiograph
Moving Through the Menus
Figure 13. CP 200 Standard MenuFigure 15. CP 200 Parent Menu With Submenu
ECG Setup : Edit Auto Report 1
1 Format
2 Interp Settings
3 Patient Data
0 Previous Menu
9:17AM Mar 16 05
Figure 14. CP 100 Standard Menu
Edit Auto Report
1 Format
2 Interp Settings
3 Patient Data
0 Previous Menu
9:17AM Mar 16 05
Edit Auto Report 1 : Format
1 Lead Arrangement
2 Rhythm Lead 1
3 Rhythm Lead 2
4 Rhythm Lead 3
5 Extended Measurements
6 Average Cycles
0 Previous Menu
9:17AM Mar 16 05
3x4
3x4 +1R
3x4 +3R
6x2
12x1
6x2 50 mm/s
6x2 10 Ext.
No Print
Figure 16. CP 100 Parent Menu with Submenu
Format
1 Lead Arrangement
2 Rhythm Lead 1
3 Rhythm Lead 2
4 Rhythm Lead 3
5 Extended Measurements
6 Average Cycles
0 Previous Menu
To move up or down a list or (keyboard or softkey arrows)
To open a standard menu (Figure 13)
or or
or item’s number or letter
To move from parent menu to submenu on same
screen (Figure 15)
To perform an action
or
To accept data
To check or uncheck a checkbox
To return to parent menu from submenu on same
screen (Figure 15) or
(To select the highlighted submenu
item.)
or
(To make no change.)
To move back through the menus or zero key
To move through data-entry fields
To return to the “Main Screen” from a standard menu
(Figure 13)
Note
Keyboard and softkey arrows work the same way.
Service ManualCP 100 and CP 200 Electrocardiograph19
Reviewing the Device Information
1.Press the Menu key .
2. Choose System Settings > Device Info.
The following screen appears.
Figure 17. “Device Info” CP 200 Screen
System Setup : Device Info
1 About
2 Print Settings
3 Enable Options
4 Upgrade Software
5 Service Info
0 Previous Menu
9:17AM Mar 16 05
Figure 18. “Device Info” CP 100 Screen
Device Info
1 About
2 Print Settings
3 Enable Options
4 Upgrade Software
5 Service Info
0 Previous Menu
9:17AM Mar 16 05
3. Select the desired item:
ItemDescription
AboutDisplays the following information about the electrocardiograph:
• serial number
• modules configured
• version numbers
Print SettingsCP 200 = Prints your ECG, spirometry, and system settings as well as medication &
history lists.
CP 100 = Prints your ECG and system settings.
Enable OptionsContact Technical Support. For phone numbers, see page ii.
Upgrade SoftwareContact Technical Support. For phone numbers, see page ii.
Service InfoAccessible to service support only. See Service Menus (page 20)
20OverviewWelch Allyn CP 100 and CP 200 Electrocardiograph
CP 100 and CP 200 Electrocardiograph
Service Menus
Entering the Service Info Screen
This menu is password protected and only Welch Allyn Authorized Technician can access
the Service Menus,)
CP 200 Service Info Screen
ProcedureFigure 19. CP 200 “User Login”
1. Power ON unit.
2. Press Menu key.
3. Choose System Setting>Device Info> Service Info
The “UserLogin” screen appears. See Figure 19.
UserLogin
9:17AM Mar 16 05
Use the “Tab” or “Arrow” keys to move the cursor from field to field.
Enter the following:
User ID: Service (case sensitive)
Password:normser (case sensitive)
4. Press “Done” using the softkey. Do not use the“Enter” key.
The “Service Info” screen appears see Figure 20
User ID
Password
Enter valid user ID and password
Change
Password
Cancel
Figure 20. CP 200 “Service Info” Screen
9:17AM Mar 16 05
Service Info
1.Print Log
2.Battery Life Test
3.Printer Test
4.Export Files
5.Import Files
6. Delete Files
7.Serial Number
8.Lead Off Thershold
9.Service Tests
A LCD Test
0.Previous Menus
Done
Service ManualCP 100 and CP 200 Electrocardiograph21
CP 100 Service Info Screen
ProcedureFigure 21. CP 100 “UserLogin” Screen
1. Power ON unit.
2. Press Menu key.
3. Choose System Setting > Device Info > Service Info
User Login
9:17AM Mar 16 05
The “UserLogin” screen appears.See Figure 21.
User ID
Use the “Arrow” keys to move the cursor from field to field.
Password
Enter the following:
User ID: Service (case sensitive)
Done
Password:normser (case sensitive)
4. Highlight “Done” using the down arrow key.
5. Press “Enter” key.
The “Service Info” screen appears. See Figure 22
Figure 22. CP 100 “Service Info” Screen
9:17AM Mar 16 05
Service Info
1. Print Log
2. Battery Life Test
3. Printer Test
4. Export Files
5. Import Files
6. Delete Files
7. Serial Number
7. Lead Off Thershold
8. Service Tests
0. Previous Menus
Print Log
Caution These printouts can be very long. Only use the printouts if necessary.
To stop the printouts press Stop/Cancel.
FunctionDescription
Print Events:This prints all events stored in the unit in descending
order by date and time. Used for development of the
software. Has no Service value.
Print Errors:This prints any errors that occurred in the unit, in
descending order by date and time, during normal
operation. This can be valuable to use when the enduser says he/she has been having problems.
Print All:This prints the Print Events and Print Errors as
described above.
Figure 23. “Print Log” Screen
Print Log
1.Print Events
2.Print Errors
3.Print All
0.Previous Menus
9:17AM Mar 16 05
22OverviewWelch Allyn CP 100 and CP 200 Electrocardiograph
Battery Life Test
This displays the percentage of battery life remaining on the battery, based on the number
of charges it has incurred. This percentage approximates the life left in the battery based
on this information.
Caution Only reset when the battery has been replaced and charged for at
least 12hrs continuously.
FunctionDescription
ResetResets the battery life to 100%. This is used only
when the battery has been replaced
Reset Pages Resets page count for the number of prints made on
the printer. Should only be reset when print head has
been replaced.
Printer Test
FunctionDescription
Figure 24. “Battery Life Test” Screen
9:17AM Mar 16 05
Battery Life Test
Battery Life Remaining: 99%
Reset
Reset
Pages
Figure 25. Printer Test” Screen
Exit
Test PagePrints a test page which repeats until the “Cancel”
button is pressed. Refer to Functional Verification test
for explanation. See Printer Test (page 35).
Ramp Test Prints a one page ramp test. Refer to Functional
Verification test for explanation. See Printer Test
(page 35).
Printer Test
Test Page
9:17AM Mar 16 05
Select a print test
Ramp
Page
Exit
Service ManualCP 100 and CP 200 Electrocardiograph23
Export Files
Use the “Export Files” function when the Main PCB is replaced, to reinstall the
customer’s saved ECG information and spirometry (CP 200 Only) or the unit’s original
configuration after repair.
Caution The tests exported to the SD memory card contain patients’ medical
information and should be treated the same as any confidential medical
information.
Note
If you are performing all functions you only need one SD memory card. Use a
formatted blank SD memory card installed in the unit for each function.
FunctionDescription
Export Log FilesExports the Event log to a SD memory card in a folder
called Log. Once complete, a message appears at
bottom of the screen stating if the export was
successful or not.
Export Audit Trail
(CP 200 ONLY)
Export Tests (CP 200
ONLY)
Export
Configuration Files:
Exports the audit information of the unit to a SD
memory card in a folder called Audit. Once complete
a message appears at bottom of screen stating if
export was successful or not.
Note: Audit information is available if Audit trail
is turned on in Device Administration. See the CP 200 Directions for Use manual
Exports all tests in the memory to a SD memory card
in a folder called Tests for importing after repair.
Exports the configuration of the unit (i.e. print
formats, patient data settings, logins, etc.) to a SD
memory card in a folder called Config for importing
after repair.
Figure 26. “Export Files” Screen
Export Files
1.Export Log Files
2.Export Audit Files
3.Export Tests
4.Export Configuration Files
0.Previous Menus
9:17AM Mar 16 05
24OverviewWelch Allyn CP 100 and CP 200 Electrocardiograph
Import Files
FunctionDescription
Import Tests (CP 200
ONLY)*
Imports previously exported tests from the SD
memory card.
*Before importing tests, you must delete the previous
copies from the unit using the “Delete Files” menu.
Import
Configuration Files
Imports previously exported configuration files from
SD memory card.
After the import configurations files process is
complete, the screen displays a reset prompt. Press
any key to reset, and then follow the prompts to set
the date and time.
Delete Files
Figure 27. “Import Files” Screen
Import Files
1.Import Tests
2.Import Configuration Files
0.Previous Menus
Caution No warning is displayed when you delete files/tests.
9:17AM Mar 16 05
FunctionDescription
Delete Log FilesDeletes log files (event and error) from unit.
Delete Audit Files:
Deletes the Audit Log file from the unit.
(CP 200 ONLY)
Delete Tests: (CP
200 ONLY)*
Delete
Deletes ALL saved ECG and Spirometry tests in
memory.
Deletes the Configuration files from the unit.
Configuration Files:
Note
*Before importing the tests file, you must delete any previous tests from the unit using
the “Delete Files” menu.
You do not have to delete the Configuration files on the unit before importing a
Configuration files from SD memory card. All previous Configurations Files on the unit will
be overwritten during the importing process.
Figure 28. “Delete Files” Screen
Delete Files
1.Delete Log Files
2.Delete Audit Files
3.Delete Tests
4.Delete Configuration Files
0.Previous Menus
9:17AM Mar 16 05
Service ManualCP 100 and CP 200 Electrocardiograph25
Serial Number
Used to enter the S/N of the unit after a Main Board has been replaced.
Note
For use by Welch Allyn Authorized Repair facilities.
Lead Off Threshold
The Lead Off Threshold setting provides a mechanism to increase/decrease the sensitivity
of the device to lead impedance. Lead impedance is the electrical resistance of a lead.
Lead impedance can rise if there is poor electrode/skin contact. As lead impedance rises,
the input signals become imbalanced, resulting in a poor quality signal.
If the lead impedance for any of the leads is above the threshold, the device replaces the
poor quality signal received from the patient cable with a square wave to indicate to the
user that the affected signals are not usable. The lead impedance of each of the leads is
independent, but there is a single threshold value. The threshold can be any of 15 values,
from 1 (lowest threshold) to 15 (highest threshold). The default threshold is 7. If the
threshold is changed to 1, it causes Lead Off all of the time, and setting it to 15 causes the
leads to never appear “Off”
FunctionDescription
Lead Off ThresholdThe Lead Off Threshold setting is provided to diagnose problems with the front end. By increasing the Lead
Off Threshold setting, the device accepts higher impedance signals, instead of replacing them with square
waves. Although the device displays poorer quality signals, they may be unsuitable for diagnostic purposes,
because they could be significantly distorted.
The Default setting for the Lead Off Threshold to 7
To access and change the Lead Off Threshold
Select Main Menu > Service Info > Device Info > Lead Off Threshold
Service Tests
Only authorized Welch Allyn Authorized Technicians are to perform the service tests.
Note
Function Description
CPWSTEST.EXEUsed to help troubleshoot USB issues.
MEANSMDWTESTER.EXE:Strictly used for development purposes.
To exit any of the applications press the “Enter” or”OK” key.
Enable USB: Turns USB port on to be recognized by computer.
Disable USB: Turns USB port off so it is not recognized by computer.
26OverviewWelch Allyn CP 100 and CP 200 Electrocardiograph
Function Description
MMCSDTEST.EXE:Used to test the SD memory card write/read function of the unit. Can also be used to test an SD memory card
for problems.
GO: Starts the SD memory card test by pressing “G”
STOP: Stops the SD memory card test by pressing “S”
ERASE ALL: Erases all files written by this tester by pressing “E”
Card Present: If checked the unit detects the SD memory card inserted.
Write Protect: If checked the unit detects the lock is enabled on the SD memory card.
Pulse: Flashes when test is active
Note: Once the test is started it runs continuously. Press the “S” key to stop the test.
NORMANDY.EXE:Used to return to the ECG program.
NORMTESTER.EXEStrictly used for development purposes.
PPPTEST.EXEStrictly used for development purposes.
PPPTEST2.EXE Used to see battery and charger status/state. There are a number of selections on this screen. Listed below
are the most commonly used functions. To activate a selection press the letter that coincides with the
selection.
Real-time: Activates a real-time monitoring of the readings from the battery circuit.
Battery Voltage: Unplugged from AC this reflects the Voltage of the battery in millivolts; plugged into
AC this reflects the charge voltage in millivolts.
Battery Current: Indicates charge current in mA when charger is on.
Charger State: Indicates if charge circuit is on or off
Charge State: Indicates if the Fast Charge is on or off
Fast Charge On: Turns fast charge on. If battery is fully charged the sensing circuit can work against
you. It only changes the reading for a brief second before switching back.
Fast Charge Off: Turns fast charge off.
Charger On: Turns charger on when plugged into AC.
Charger Off: Turns charger off.
Service ManualCP 100 and CP 200 Electrocardiograph27
Function Description
PPPTEST3.EXEUsed to see printer assembly status/state. There are a number of selections on this screen. Listed below are
the most commonly used functions. To activate a selection press the letter that coincides with the selection.
Real-time: Activates a real-time monitoring of the readings from the printer assembly.
+24 V State: Status of voltage applied to printhead for printing. Only turns on when motor is on.
Step Motor State: Status of paper drive motor.
Paper Exhaust State: Status of paper in unit.
OK indicates paper is loaded.
Exhausted indicates out of paper.
Paper Status: 'JAM' indicates printer was unable to feed paper and detect a cue mark in the required
time. 'OK' indicates no jam.
Door State: This is equivalent to the Paper Exhaust State. There is no separate door sensor/switch.
Printhead Temperature State: State of printhead temperature. If temperature is too high, the printer
does not print.
Printhead Temperature (C): Indicates temperature of the printhead in Celsius.
Turn ON Motor: Turns stepper motor on to drive paper.
Turn OFF Motor: Turns stepper motor off.
PPPTEST4.EXEUsed to see printer Assembly status/state. There are a number of selections on this screen. Listed below are
the most commonly used functions. To activate a selection press the letter that coincides with the selection.
Real-time: Activates a real-time monitoring of the readings from the printer assembly.
Feed to Next Page: Advances paper to next page.
Set Speed to 10 mm/s: Sets paper speed to 10mm/s.
Set Speed to 25 mm/s: Sets paper speed to 25mm/s.
Set Speed to 50 mm/s: Sets paper speed to 50mm/s.
Turn On Motor: Turns stepper motor on to drive paper.
Turn Off Motor: Turns stepper motor off.
Get Paper Registration State: Indicates if the sensor has identified the installed paper. Note: If you
open the door and close it, you must feed the paper to next page so unit can detect a cue mark. When the
unit detects a cue mark, the state changes to YES.
Get Cue Levels: Indicates sensing level for the cue sensor for different paper states.
< 10: Indicates paper door open or out of paper.
10 - 90: Indicates cue sensor over a paper cue mark.
> 190: Indicates cue sensor over white paper.
28OverviewWelch Allyn CP 100 and CP 200 Electrocardiograph
30Functional VerificationWelch Allyn CP 100 and CP 200 Electrocardiograph
Functional Verification Overview
This section describes the procedure for a complete functional test to support
recommended preventive-maintenance schedules.
A checklist of the functional tests is provided on page 40. It is recommended that you
print a copy of the checklist each time you perform the functional verification procedure,
so you can record and save the test results. If the electrocardiograph ever requires
service, the records of test results often facilitate troubleshooting.
Functional verification does not require opening the device case.
Full Functional Check
WARNING Only qualified service personnel should perform a full functional
checkout procedure.
Whenever the electrocardiograph is serviced or problems are suspected, Welch Allyn
recommends a Functional Check.
Equipment Required
Commercially Available General-purpose/Medical Test Equipment
ItemManufacturer
Power SupplyVariable, 0 – 8 VDC, 5.0A
AC withstand voltage (hi-pot) tester Associated Research 3605 or equivalent
Electrical Safety Analyzer w/ECGFluke DNI 232D or equivalent
ComputerPIII or better, CardioPerfect installed, available USB port
SD Memory Card64Mb or Greater
USB CableA – A type (WA P/N 401435)
Spirometry DongleSerial Version (W/A P/N 703552)
Operating/Non-Printing Mode Current Test
When the unit is powered on, the AC is disconnected and the LCD on, the unit’s current
is:
CP 100 CurrentCP 200 Current
125 to 250 mA @ 6.4V DC ± .05675 to 825 mA @ 6.4V DC ± .05
(minimum), with voltage and current indicators
1.Unplug unit from line power
2. Connect a DC power supply, adjusted for 6.4V DC ± .05, in place of the battery.
3. Verify the current draw is within limits as indicated for unit after powering the unit up.
Service ManualFunctional Verification31
Battery Accuracy Test
When a voltage of 6.2 ± 0.05 VDC is supplied to the unit battery terminals, and the AC is
disconnected from the unit, the unit reports a “Normal to High” voltage.
When a voltage of 5.45 ± 0.05VDC is supplied to the unit battery terminals, and the AC is
disconnected from the unit, the unit reports a “critically low” voltage.
Note
1.Unplug unit from line power.
2. Connect a DC power supply in place of the battery.
3. Adjust power supply to 6.2VDC.
4. Turn the unit on, and then verify the unit does not indicate a low or critically low
5. Adjust power supply to 5.45 VDC.
6. Verify the unit indicates a critically low power and will shut down.
The unit may have a 5 to 10 second delayed response.
battery.
Sleep/Standby Mode Current Test
When the unit is powered on and the AC is disconnected, and the unit is in low power
standby mode, the unit’s current is:
CP 100 CurrentCP 200 Current
< 5 mA< 5 mA
1.Unplug the unit from line power and adjust the power supply to 6.4V DC ± .05.
2. Connect the DVM in series with the positive battery connection from the power
supply.
3. Press the on/standby button.
4. Verify the LCD turns off and that the current draw is less than indicated for the unit.
5. Disconnect the power supply.
Note
There may be up to a 10 sec delay in the drop of current due to the progressive
hibernation mode.
Charger Detect Test
1.Connect a DVM to the battery harness while observing correct polarity.
2. Plug the unit into the AC power, and power ON.
3. Choose Main Menu > Systems Setting > Device info > Service Info > Service
Tests>PPPTEST2.EXE.See “Service Tests” on page 25.
32Functional VerificationWelch Allyn CP 100 and CP 200 Electrocardiograph
4. Turn fast charge ON. Check DVM for 6.8 to 7.3 VDC.
5. Turn fast charge off. Check DVM for 6.3 to 6.8 VDC.
6. Unplug the unit from the AC power.
7.Disconnect the DVM and install the battery.
LED Test
Verify the Power LED light is on when unit is plugged in, and verify the light is off when
unit is unplugged.
1.Plug the unit into AC power.
2. Verify the AC indicator LED under the On/Standby button is lit.
3. Unplug the AC power cord.
4. Verify the LED is off.
5. Plug the unit back into AC power.
LCD Test
CP 200
CP 100
Note
1.Choose Main Menu > Systems Setting > Device info > Service Info > LCD Test.
2. Verify the test pattern is visible without breaks and is consistent over the entire
3. Verify Contrast can be adjusted to adequately see the display.
1.Disconnect the patient cable.
2. Turn ON unit. The patient hook up screen appears.
3. Verify the display is visible without breaks.
Note
There may be a few seconds delay in the LED response to the line being removed
and added.
screen.
If contrast adjustment is needed refer to the Directions for Use manual for
CP 100.
Set Unit Initial Configuration/SD Slot Test
Initial Configuration of Unit
1.Verify the Serial Number on the unit is the same as displayed on the unit’s screen. If
incorrect, enter correct S/N. See ““System Settings” Menu Tree” on page 16, Service
Info.
Service ManualFunctional Verification33
Select Main Menu > System Settings > Device Info > Service Info > Serial
Number.
2. If the main board was replaced and a configuration files were exported, import the
configuration files, see “Import Files” on page 24.
3. Print the Configuration of the unit.
Select Main Menu > System Settings > Device Info > Print Settings
4. Verify the settings from the printout.
a. Check for software version installed.
b. S/N of the unit.
c. Interpretation (if applicable).
If imported, check any customized customer settings.
SD Slot Test
1.Insert a formatted SD memory card into the unit.
2. Choose Main Menu > System Settings > Device info > Service Info > Service Tests>MMCSDTEST.EXE file.
3. Press G to start the test. After 30 seconds the MyFile.TXT increases one increment.
4. Press S to stop the test and confirm that no errors are reported.
ECG Communication Test (ECG Input test and lead off)
Test the patient cable and lead off indications connection.
1.Connect a known good Patient Cable w/leads
2. Connect to ECG simulator with V1/C1 lead disconnected.
3. Power ON unit.
4. Verify the lead placement screen shows V1/C1 lead flashing and all other leads
should appear not flashing.
5. Remove LL from the Simulator and verify LL flashes with all V/C leads grayed, RA,LA
& RL should not be flashing.
6. Connect all leads to the simulator.
7.Verify all leads stop flashing and the next screen appears in approximately three
seconds.
Spirometer Communication Test
CP 200 (Only)
Verify and test the connection of a Spirometer Dongle to the serial port.
1.Attach a known good spirometer dongle to the spirometer port on the side of the unit.
2. Power ON unit.
34Functional VerificationWelch Allyn CP 100 and CP 200 Electrocardiograph
3. Press the Spirometry button
4. Verify that the unit opens the Spirometry module without errors.
USB Communications Test
Verify the test connection between the unit and the CardioPerfect Workstation computer.
1.Connect the unit to the computer using a USB A-A Cable.
2. The computer should recognize the CP 200 connected and load the driver for
communication.
3. Transfer an ECG record to the CardioPerfect Workstation.
a. Press the Menu button.
b. Choose Test Directory.
c. Press the Search softkey.
d. Press the WorkStation softkey.
Beeper Test
4. Verify the record was transferred into the workstation.
Turn on and test the Audible beeper by using the keyboard keys.
1.P o w e r On unit.
2. Choose Main Menu > System Settings > Device Configuration >Audio Beeper > set to ON.
3. Exit the Main Menu
4. Verify that an Audible beep is heard when invalid keys are pressed on the keyboard.
5. If the Audio Beeper was originally set to off, choose Main Menu >
System Settings > Device Configuration > Audio Beeper > set to OFF to return
the Audio Beeper to its original setting.
Service ManualFunctional Verification35
Keyboard Test
The unit detects all the keys pressed listed in Table 1, “Keyboard Keys”.
Table 1. Keyboard Keys
KeyCP 100CP 200
’5’ 55
‘R’RR
‘D’DD
‘X’XX
BackspaceBack a spaceBack a space
‘P’PP
DownDown a fieldDown a field
UpUp a fieldUp a field
1.P o w e r On unit.
2. Press Auto ECG. See Figure 9 on page 14 for button location.
3. The patient information screen appears.
4. Press the Keys indicated in Table 1, “Keyboard Keys” and verify the corresponding
action.
Printer Test
5. Use the softkeys to verify the correct response for the keys (Only for the CP 200 unit).
See Figure 10 on page 15.
Note
1.Plug unit into AC power to perform the printer test.
2. Choose Main Menu > System Settings > Device Info > Service Info > Printer
3. Press the “Test Page” softkey. The test print starts after a few seconds.
4. Press the “Cancel” softkey when the second page has completed printing.
5. Use the second test print page to verify correct operation. See Table 2, “Print
Please note these measurements are dependent upon the tracking of the paper
being in the center. Please make sure there is no curling/crinkling of the paper at
the top or bottom when printing.
Test.
Verification Check List” and Figure 29 on page 36.
36Functional VerificationWelch Allyn CP 100 and CP 200 Electrocardiograph
Table 2. Print Verification Check List
PointDescription
Point AVerify the distance of the arc on the circle from the top edge of the paper is 37mm +/-
Point BVerify the distance of the arc on the circle from the left edge of the paper is 45mm +/-
Point CVerify line is straight and without breaks from top to bottom
Point DVerify thick black line is Uniform in Blackness without breaks from top to bottom
Point EVerify line is straight and measures 200mm +/- 4mm
Point FVerify the print is readable
Point GVerify the lead in and output trail of the cue sensor are between 19 and 82mm down
Point HVerify the cue sense pattern is between 97 and 135mm down from the 200mm line
3mm
5mm
from the 200mm line measured in point E
measured in point E
Figure 29. Test Printout Example
Service ManualFunctional Verification37
Ramp Page Test Printout
1.Press the Ramp Page key.
2. Verify the top of the ramp cuts off between 250 and 550 on the gauge printed.
Figure 30. Ramp Printout Example
38Functional VerificationWelch Allyn CP 100 and CP 200 Electrocardiograph
Dielectric Strength Test
It is important to complete the dielectric strength test after any procedure in which the
monitor is opened to verify proper patient isolation.
WARNING Failure to run the dielectric test when indicated could cause serious
injury to patients, and could lead to damage to the monitor.
WARNING Always follow the tester manufacturer’s operation instructions
exactly. Failure to perform this test properly can result in serious injury or death.
WARNING This test must be performed by qualified service personnel only.
WARNING Run this test, only, on an insulated table top, and away from other
people and equipment.
WARNING Must avoid connector shell and pe ground because pe is only
resistant to 500 Vac.
WARNING The patient isolation test involves exposure to extremely high
voltages and is extremely hazardous.
The patient isolation test requires an AC Withstand Voltage (hi-pot) Tester, such as the AR
3605 or equivalent. If this equipment is not available, Welch Allyn can perform the patient
isolation test for you quickly, for a nominal fee
.
To create test cables for the patient isolation test. See “Hi-Pot Test Connections” on
page 39.
Set up the test parameters on the AC Withstand Voltage Tester as follows:
Note
Refer to the tester manufacturer’s operation manual for complete details on set
up and use.
ParameterSpecification
Voltage1500Vac
Maximum Limit5.00mA
Minimum Limit0.15 mA
Ramp Up10 s
Dwell1 min.
Delay0.5 s
Ramp Down0.0 s
Arc Sense0
FrequencyLocal line (50 Hz or 60 Hz)
ContinuityOff
ConnectOff
To create test cables for this procedure. See “Hi-Pot Test Connections” on page 39.
1.Connect the modified power cord between the return jack on the tester and the AC
input of the unit.
2. Connect the high voltage cable between the equipotential stud and the high voltage
on the tester.
3. Press the test button to run the AC withstand voltage test.
Service ManualFunctional Verification39
4. Verify the electrocardiograph passed the test. (If the unit current never exceeds
Hot and neutral (black and white) wires connected together and
terminated appropriately for your hi-pot tester. Cut and terminate
the ground (green) wire to prevent possible injury.
Return
Chassis, Earth and Patient Leakage Current Tests
Verify continued electrical safety of the device, using IEC 60601-1 or ANSI/AAMI ES1
methods and limits. Test for the following:
•Earth leakage
•Chassis leakage
•Patient leakage
Note
Refer to the Electrical Safety Analyzer manual for complete details on set-up and
use for testing.
40Functional VerificationWelch Allyn CP 100 and CP 200 Electrocardiograph
Checklist and Test Results Report Form
Use a copy of the form to track your progress though the validation tests.
Electrocardiograph Serial #:
Tested by:
Test date:
42TroubleshootingWelch Allyn CP 100 and CP 200 Electrocardiograph
Problem-Solving Suggestions
Caution Replace parts, components, or accessories only with parts supplied or
approved by Welch Allyn. The use of any other parts can lead to inferior device
performance and will void the product warranty.
This section includes several tables:
•Lead quality problems (Ta ble 3)
•System failure problems (Tab le 4, on page 43)
•System messages (Tab le 5, on page 44)
If you try these suggestions and still have problems, contact Technical Support. For phone
numbers, see “For information about any Welch Allyn product, please call Welch Allyn
Technical Support:” on page ii
Table 3. Lead Quality Problems
ConditionCausesActions
A red dot is flashing on the Lead Off
screen. (CP 200)
A dot is flashing on the Lead Off screen.
(CP 100)
OR
• Electrode contact may be poor.
• A lead may be loose.
• Reattach the lead.
• Replace the electrode.
• Verify that the electrode area has been properly
prepared: shaved, cleaned with alcohol or
acetone, allowed to dry.
• Verify that electrodes have been properly stored
and handled.
Lead-off information is displayed on the
screen.
OR
One or more leads prints as a square
wave:
Wandering baseline (an upward and
downward fluctuation of the waveforms):
Muscle tremor interference (random,
irregular voltage superimposed on the
waveforms). May resemble or coincide
with AC interference:
• Electrodes that are dirty, corroded,
loose, or positioned on a bony area.
• Insufficient or dried electrode gel.
• Oily skin or body lotions.
• Rising and falling of chest during rapid
or apprehensive breathing.
• Patient is uncomfortable, tense,
nervous.
• Patient is cold and shivering.
• Exam bed is too narrow or short to
comfortably support arms and legs.
• Arm or leg electrode straps are too
tight.
• Clean skin with alcohol or acetone.
• Reposition or replace electrodes.
• Help patient relax.
• If wandering baseline persists, turn the baseline
filter on. See “Base Line Filter” in the CP 200
Directions for Use Manual.
See “Adjust Baseline Filter” in the CP 100
Directions for Use Manual.
• Help patient get comfortable.
• Check all electrode contacts.
• If interference persists, turn the muscle-tremor
filter on. See “Muscle Filter” in the CP 200
Directions for Use Manual.
See “Adjust Muscle Filter” in the CP 100
Directions for Use Manual.
• If interference still persists, the problem is
probably electrical in nature. See the following
suggestions for reducing AC interference.
Service ManualTroubleshooting43
Table 3. Lead Quality Problems (continued)
ConditionCausesActions
AC interference (even-peaked, regular
voltage superimposed on the waveforms).
May resemble or coincide with muscletremor interference.
• Electrodes that are dirty, corroded,
loose, or positioned on a bony area.
• Insufficient or dried electrode gel.
• Patient or technician touching an
electrode during recording.
• Patient touching any metal parts of an
exam table or bed.
• Broken lead wire, patient cable, or
power cord.
• Electrical devices in the immediate
area, lighting, concealed wiring in
walls or floors.
• Improperly grounded electrical outlet.
• Incorrect AC filter frequency setting or
AC filter is turned off.
• Check all electrode contacts and lead wires.
• Verify that the patient is not touching any metal.
• Verify that the AC power cable is not touching
the patient lead cable.
• Verify that the proper AC filter is selected. See
“Mains Filter” in the Directions For Use manual.
• If interference persists, unplug the
electrocardiograph from AC power and run it on
the battery. If this solves the problem, you’ll
know that the noise was introduced through the
power line.
• If interference still persists, the noise may be
caused by other equipment in the room or by
poorly grounded power lines. Try moving to
another room.
Table 4. System Failure Problems
ConditionCausesActions
Won't turn on when plugged into AC
power.
Won't turn on when unplugged from AC
power.
Shuts down during printingBattery low or bad.• Recharge the battery. See “Recharging a Fully
Prints fewer than 10 reports on a full
battery charge.
• Faulty AC power connection.
• Blown AC fuses.
• No AC power.
• Battery disconnected or incorrectly
connected.
• Battery low, not charging, depleted, or
bad.
• Blown battery fuse.
Degraded battery.Replace battery. See“Replacing the Battery or Fuse”
• Check the AC power source.
• Check the AC fuses. See “Replacing the AC
Fuses” in the Directions for Use manual.
• Check battery connections. See“Replacing the
Battery or Fuse” on page 58.
• Recharge the battery. See “Recharging a Fully
Discharged Battery” in the Directions for Use
manual.
• Replace battery. See“Replacing the Battery or
Fuse” on page 58.
• Replace battery fuse. See “Replacing the Battery
(DC) Fuse” on page 60.
Discharged Battery” in the Directions for Use
manual.
• Replace battery. See“Replacing the Battery or
Fuse” on page 58.
on page 58.
44TroubleshootingWelch Allyn CP 100 and CP 200 Electrocardiograph
Table 5. System Messages (alphabetical order)
System MessageProblemsActions
“A test is being printed that has not been
saved and will be lost. Continue
shutdown?” (CP 200)
“Audit Trail is too large. Please print and
purge.” (CP 200)
“Battery extremely low. Please connect
AC power.”
“Insufficient space available”Not enough space on memory card.• Delete some tests from the card at a PC.
“Memory card error”Problem writing to memory card.• Verify that the write-protect tab is in the
“Out Of Paper”• Printer is out of paper.
was pressed while printing an
Auto ECG.
Audit information is approaching its
maximum allotted storage.
Battery low or bad.• Recharge the battery. See “Recharging a Fully
• Printer door is open.
To shut down without saving the test, press .
To cancel the shutdown, press .
Print and purge the audit trail. See “Working With the
Audit Trail”.
Discharged Battery” in the Directions for Use
manual.
• Replace battery. See“Replacing the Battery or
Fuse” on page 58.
• Use a different card.
unprotected position.
• Reseat the card in its slot.
• Use a different card.
• Load paper. See “Loading the Thermal Chart
Paper” in the Directions for Use.
• Close the printer door. See “Loading the Thermal
Chart Paper” in the Directions for Use.
“Paper Error” (CP 200)
“Paper Jam” ( CP 100)
“Powering down”Low battery.Recharge the battery. See “Recharging a Fully
“Problem loading the following settings:
<System>
<ECG>
*
*
*
Using default settings.”
“Shutdown?” was pressed while printing a
“Temperature Error”Printer head temp is too high.Allow to cool, then try again.
“Test has not been saved. Continue
shutdown?” (CP 200)
“Unable to communicate with
workstation“ (CP 200)
Paper was loaded incorrectly.• Reload the paper. See “Loading the Thermal
Problem loading your settings at startup.
May indicate memory problems.
Rhythm ECG.
was pressed after printing an
Auto ECG.
• Configuration not correct.
• Cable not properly connected.
Chart Paper” in the Directions for Use.
Discharged Battery”.
Contact Technical Support. For phone numbers, see
page ii.
To shut down, press .
To cancel the shutdown, press .
To shut down without saving the test, press .
To cancel the shutdown, press .
• Verify communication settings on the PC.
• Check cable connections.
Service ManualTroubleshooting45
Limited Warranty
Welch Allyn, Inc., warrants that the Cardiopulmonary line of electrocardiographs, including
the CP 100 and CP 200 models (the Products) meet the labeled specifications of the
Products and will be free from defects in materials and workmanship that occur within 3
years after the date of purchase, except that accessories used with the Products are
warranted for 90 days after the date of purchase. Such accessories include: lead wires,
cabling, electrodes, and battery.
The date of purchase is: 1) the date specified in our records, if you purchased the Product
directly from us, 2) the date specified in the warranty registration card that we ask you to
send to us, or 3) if you don’t return the warranty registration card, 30 days after the date
on which the Product was sold to the dealer from whom you bought the Product, as
documented in our records.
This warranty does not cover damage caused by: 1) handling during shipping, 2) use or
maintenance contrary to labeled instructions, 3) alteration or repair by anyone not
authorized by Welch Allyn, and 4) accidents.
If a Product or accessory covered by this warranty is determined to be defective because
of defective materials, components, or workmanship, and the warranty claim is made
within the warranty period described above, Welch Allyn will, at its discretion, repair or
replace the defective Product or accessory free of charge. If your Product requires repairs
covered by this warranty, upon your request Welch Allyn will loan to you, at no cost, a
substitute Product for use until your repaired Product is returned.
You must obtain a return authorization from Welch Allyn to return your Product before you
send it to Welch Allyn’s designated service center for repair. Contact Welch Allyn
Technical Support. For phone numbers, see page ii.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED,
INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILTY
AND FITNESS FOR A PARTICULAR PURPOSE. WELCH ALLYN’S OBLIGATION UNDER
THIS WARRANTY IS LIMITED TO REPAIR OR REPLACEMENT OF PRODUCTS
CONTAINING A DEFECT. WELCH ALLYN IS NOT RESPONSIBLE FOR ANY INDIRECT OR
CONSEQUENTIAL DAMAGES RESULTING FROM A PRODUCT DEFECT COVERED BY
THE WARRANTY.
46TroubleshootingWelch Allyn CP 100 and CP 200 Electrocardiograph
Service Policy
All repairs on products under warranty must be performed or approved by Welch Allyn.
Unauthorized repairs will void the warranty. In addition, whether or not covered under
warranty, any product repair shall exclusively be performed by Welch Allyn certified
service personnel.
If the product fails to function properly—or if you need assistance, service, or spare
parts—contact the nearest Welch Allyn Technical Support Center. For phone numbers,
see page ii.
Before contacting Welch Allyn, try to duplicate the problem, and check all accessories to
ensure that they are not causing the problem. When calling, please be prepared to
provide:
•Product name and model number and complete description of the problem.
•Serial number of your product (if applicable).
•Complete name, address and phone number of your facility.
•For out-of-warranty repairs or spare parts orders, a purchase order (or credit card)
number.
•For parts orders, the required spare or replacement part numbers.
If your product requires warranty, extended warranty, or non-warranty repair service,
please call first the nearest Welch Allyn Technical Support Center. A representative will
assist you troubleshooting the problem and will make every effort to solve it over the
phone, avoiding potential unnecessary returns.
In case a return cannot be avoided, the representative will record all necessary
information and will provide a Return Material Authorization (RMA) number, as well as the
appropriate return address. An RMA number must be obtained prior to any return.
If you have to return goods for service, follow these recommended packing instructions:
•Remove all hoses, cables, sensors, power cords, and ancillary products (as
appropriate) before packing, unless you suspect they are associated with the
problem.
•Wherever possible use the original shipping carton and packing materials.
•Include a packing list and the Welch Allyn Return Material Authorization (RMA)
number.
It is recommended that all returned goods be insured. Claims for loss or damage to the
product must be initiated by the sender.
48Disassembly and Repair ProceduresWelch Allyn CP 100 and CP 200 Electrocardiograph
Introduction
These procedures cover CP 200 and CP 100 electrocardiograph disassembly for subassemblies and board removal. Unless otherwise noted:
•The procedure is the same for both units.
•The assembly procedure is the reverse of the disassembly procedure.
Caution Perform all repair procedures at a static-protected station. Observe
recommended screw torque specifications, especially with screws that secure
directly into plastic standoffs.
WARNING The battery can deliver currents sufficient to cause serious personal
injury and to damage the electrocardiograph. When opening the
electrocardiograph for any reason, always remove and disconnect the battery
immediately, before proceeding with disassembly.
Screws
Caution To avoid mismatching screws and holes, keep the screws for each
piece with the piece as you remove modules, circuit assemblies, and other
components.
Torque Specs
The Following are recommended torque specs for all screws and hex nuts..
Service ManualCP 100 and CP 200 Electrocardiograph49
Disassembly Procedures
Opening the Unit
WARNING Disconnect ALL power from the unit before repairing or
disassembling unit. Unplug the power cord and disconnect the battery at the
fuse in the battery compartment. See Replacing the Battery or Fuse (page 58)
ProcedureFigure 31. Removing the Screws
1.Place unit on a clean, static mat.
2. Turn the unit up-side-down
3. Remove the 8 screws around the outside
rim of the unit. SeeFigure 31.
Screws
4. Using both hands, hold the upper case and
lower case of the unit together, turn the
unit upright.
5. Slowly lift the upper case, and disconnect
the ribbon cable from the upper case.
6. Lift off the top cover. See Figure 32.
Figure 32. Disconnecting the Ribbon Cable
Upper Case
Ribbon Cable
Lower Case
50Disassembly and Repair ProceduresWelch Allyn CP 100 and CP 200 Electrocardiograph
Replacing the Printer Assembly
Procedure
1.Open unit using the Disassembly
Procedures (page 49) directions.
2. Open the paper door and remove all paper
in the paper tray.
3. Insert a flat blade screwdriver into the slit
opening located on the side of the auxiliary
paper cover. Figure 33.
4. Apply outward pressure on the screwdriver
while pressing upward on the side of the
cover. Repeat process for the other side of
the cover. Lift the cover out of unit.
5. Remove the 4 screws in the bottom of the
printer tray. See Figure 34.
Figure 33. Inserting a Flat blade Screwdriver
Auxiliary Paper Cover
Slit Openings
Figure 34. Removing the Screws in the Paper Tray
4 Screws
AAMIV1V2V3V4V5V6RALARLLL
IEC C1C2C3C4C5C6R L N F
Selectto end
Change settings as desired
LL
LA
V6
V5
V4
Select
RL
Rhythm ECG
R4
ECG
Select
Print
VerifyECG quality
V3
Select
DONE
V2
VI
Enter Patient Data
Select
Automatic ECG
Hold downfor 1 second
Stat ECGStat ECG
Quick Step Operating Instructions
Service ManualCP 100 and CP 200 Electrocardiograph51
Procedure
6. Slide the paper tray cover until it locks into
place before lifting the paper tray assembly.
7.Slowly lift the paper tray assembly and
remove the connections. See Figure 35
Reassembly Note
1.Keep the black printhead harness wires
behind the tabs on the auxiliary tray cover as
you place the cover into the assembly.
Figure 35. Lifting the Paper Tray Assembly
Printhead
Harness
Connections
Figure 36.
Tab s
Paper Tray
Assembly
Ground
2. When pressing the auxiliary paper cover into
place, listen for two distinct clicks when the
tabs lock into place.
Printhead Harness Wires
52Disassembly and Repair ProceduresWelch Allyn CP 100 and CP 200 Electrocardiograph
Replacing the Paper Tray Cover
Procedure
1.Open the unit using the Disassembly
Procedures (page 49) directions.
2. Next, remove the paper tray cover. See
Replacing the Printer Assembly (page 50).
3. Slide the paper tray cover out until it stops.
See Figure 37.
4. Turn paper tray assembly up-side-down.
5. Remove the 3 screws holding the paper
guide. See Figure 38.
Figure 37. Sliding the Paper Tray Cover
Paper Tray Cover
LL
RL
LA
R4
ECG
Print
V3
DONE
V2
V6
VI
V5
V4
Figure 38. Bottom of the Paper Tray Assembly
Bottom of paper tray cover
AAMI V1 V2 V3 V4 V5 V6 RA LA RL LL
IEC C1C2C3C4C5C6RL N F
Selectto end
Change settings as desired
Select
Rhythm ECG
Select
VerifyECG quality
Select
Enter Patient Data
Select
Automatic ECG
Hold downfor 1 second
Stat ECGStatECG
Quick Step Operating Instructions
Screws holding paper guide
Service ManualCP 100 and CP 200 Electrocardiograph53
Procedure
6. Turn the paper tray assembly upright.
7.Slide the paper tray cover in half way until
the paper guide drops into the paper tray.
8. Slide the paper tray cover out to remove it.
Figure 39.
Figure 39. Removing the Paper Tray Cover
Direction to Slide Paper Tray Cover to
V
3
V
4
V
5
V
6
R
A
L
A
R
L
L
Printhead
Paper Guide
L
LL
L
A
V
6
V
5
V
4
Direction to Slide Tray Cover to
Drop Paper Guide into the Tray.
C
3
C
4
C
5
C
6
R
L
N
F
to end
settings as desired
R
L
R
4
E
C
G
Print
G
quality
V
3
D
O
N
E
V
2
V
I
for 1
second
Operating
Instructions
down
A
A
M
I
V
1
V
2
IE
C
C
1
C
2
Select
Change
S
elect
Rhythm
E
CG
S
elect
Verify EC
Select
Enter Patient Data
Select
Automatic ECG
Hold
Stat ECGStat ECG
Q
uick
Step
Reassembly Note
1.Slide the paper tray cover half way into the
paper tray assembly.
2. Place the paper guide under the roller on the
paper tray cover.
3. While holding the paper guide in place, slide
the paper tray cover out until it hits the stop.
4. Turn the paper assembly up-side-down.
5. Replace the 3 screws in the holes located in
the bottom of the paper tray assembly. See
Figure 38.
54Disassembly and Repair ProceduresWelch Allyn CP 100 and CP 200 Electrocardiograph
Replacing the Platen Roller
Procedure
Figure 40. Platen Roller
1.Open unit using the Disassembly
Procedures (page 49) directions.
2. Remove printer assembly using the
Replacing the Printer Assembly (page 50)
directions.
3. Remove the paper tray cover using the
Replacing the Paper Tray Cover (page 52)
directions.
To remove the platen roller
4. Loosen the set screw using a T5 Torx on the
drive gear and slide the drive gear off the
platen roller shaft.
5. Remove the retaining ring from both ends of
the platen roller.
6. Lift the platen roller shaft out of the paper
tray cover on the end opposite the drive
gear.
7.Slide the roller out.
Bearings
Spring
Retaining Ring
Drive Gear
Set Screw
Spacers
Paper Tray Cover
IEC
A
AM
I
V1
V2
V3
V4
V5
V6
R
A
LA RL L
L
Rhythm ECG
Select
Change settings as desired
Select
Select
Print
E
CG
R
4
R
L
LL
LA
C
1
C2
C
3
C
4
C
5
C6
to end
R
L
N
F
Verify ECG quality
Automatic ECG
Stat ECGStatECG
Select
Hold down
for 1 second
Quick Step Operating Instructions
Select
Enter Patient Data
DO
N
E
V3
V2
V
I
V
V
V4
6
5
Bearings
Platen Roller
Spacers
Retaining Ring
Place the bearings in this side first for ease of
assembly.
Reassemble Note
First, place the hardware into the side opposite
the drive gear for ease of assembly.
Service ManualCP 100 and CP 200 Electrocardiograph55
Replacing the Printhead
Procedure
1.Open unit using the Disassembly
Procedures (page 49) directions.
2. Remove printer assembly using the
Replacing the Printer Assembly (page 50)
directions.
3. Lift the printhead spring (metal rod behind
printhead assembly) out of unit. See
Figure 41.
4. Slide and lift out the printhead assembly.
See Figure 42.
Figure 41. Lifting the Printhead Spring
Printhead Spring
Figure 42. Sliding and Lifting Out Printhead
Assembly
Rotate Direction
Slide Directions
Connections
Printhead Assembly
Connections
56Disassembly and Repair ProceduresWelch Allyn CP 100 and CP 200 Electrocardiograph
Procedure
5. Disconnect the paper detector harness
from the back of the printhead assembly.
See Figure 43
6. Disconnect the ground connection using a
¼” nut driver. See Figure 43
7.Remove the 2 shoulder screws from the
opposite end of the printhead using a 3/16”
nut driver.
8. Remove the 2 screws holding the Que
sensor Assembly. See Figure 43.
9. Carefully remove the printhead connectors
from both ends of the printhead. Figure 42
on page 55.
Figure 43. Disconnecting Paper Detector Harness
and Ground Connections
Ground Connection
Screws holding the Que
Sensor Assembly
Paper Detector Harness
Note
The printhead does not come with the Antistatic brushes. Order the Antistatic
brushes separately. See “Field Replaceable Parts” on page 69
Service ManualCP 100 and CP 200 Electrocardiograph57
Replacing the Stepper Motor and Gear Box
Procedure
1.Open unit using the Disassembly
Procedures (page 49) directions.
2. Remove printer assembly using the
Replacing the Printer Assembly (page 50)
directions.
3. Open the paper tray cover about half way.
4. Turn paper tray assembly up-side-down.
5. Remove the 3 screws to remove the gear
box and stepper motor. See Figure 44
6. Lift out the gear box and stepper motor.
Replacing the Stepper Motor
Procedure
Figure 44. Bottom of the Printer Assembly
Screws
1.Remove the 2 screw holding the motor.
See Figure 45.
2. Remove the motor.
Figure 45. Removing the Stepper Motor
Stepper Motor
Screws
58Disassembly and Repair ProceduresWelch Allyn CP 100 and CP 200 Electrocardiograph
Replacing the Battery or Fuse
Procedure
1.Turn the electrocardiograph up-side-down.
2. Unscrew and remove the battery door. See
Figure 46.
3. Lift out the battery. See Figure 47.
Figure 46. Removing the Battery Door
Battery door
Figure 47. Lifting Out the Battery
Battery
4. Inspect the fuse. See Figure 48.
•If the “Z” wire is intact, go to Step 5.
•If the “Z” wire is broken or dark,
replace the fuse. See “Replacing the
Battery (DC) Fuse” on page 60.
Figure 48. Battery Fuse Plugged In
Fuse
Service ManualCP 100 and CP 200 Electrocardiograph59
Procedure
5. Unplug the two battery connectors. See
Figure 49.
•You may need to pull hard. Use pliers
gently if necessary.
6. Connect a new battery by matching
connector sizes.
Do not force connectors to mismatch.
7.Place the battery into the battery
compartment. Replace the battery door
and screws. Turn the electrocardiograph
right-side-up.
It turns on automatically and displays a
prompt to enter date and time.
8. Reset the unit after you replace the battery.
See Battery Accuracy Test (page 31) for
directions.
9. Enter the date and time.
The electrocardiograph is ready to use.
Figure 49. Battery Connectors
Battery
connectors
10. Discard the old battery appropriately.
The battery is a sealed, lead-acid battery and must be recycled or
disposed of properly according to national or local regulations.
60Disassembly and Repair ProceduresWelch Allyn CP 100 and CP 200 Electrocardiograph
Replacing the Battery (DC) Fuse
If the battery (DC) fuse requires replacing, as described in “Replacing the Battery or Fuse”
on page 58. For fuse value, see “Fuses” on page 77.
Procedure
1.Remove and discard the fuse. See
Figure 50.
You may need to pull hard. Use pliers
gently if necessary.
2. Connect a new fuse. It goes in either way.
3. Go to Step 7 on page 59.
Figure 50. Battery Fuse Remove
Service ManualCP 100 and CP 200 Electrocardiograph61
Replacing the Main Board
Procedure
1.Open unit using the Disassembly
Procedures (page 49) directions.
2. Remove printer assembly using the
Replacing the Printer Assembly (page 50)
directions.
3. Disconnect the locking connection for the
power supply and the battery.
4. Remove the ribbon cable that attaches the
top case.
5. Remove the 5 self taping screws holding
the board into place. See Figure 51.
6. Carefully lift and slide the board toward the
power supply.
Figure 51. Removing the Main Board
Connection
Ribbon Cable
Locking Connection
Screws
Power
Supply
62Disassembly and Repair ProceduresWelch Allyn CP 100 and CP 200 Electrocardiograph
Replacing the Power Supply
Procedure
1.Open unit using the Disassembly
Procedures (page 49) directions.
2. Remove printer assembly using the
Replacing the Printer Assembly (page 50)
directions.
3. Disconnect connections from the power
supply to main board. See Figure 52.
4. Disconnect Ground wire and AC input
connections. See Figure 52.
5. Remove the 4 screws holding the power
supply in place.
6. Lift out the power supply.
Figure 52. Replacing the Power Supply
Connection
Power Supply
Ground and AC Connections
Service ManualCP 100 and CP 200 Electrocardiograph63
Replacing the IEC (AC input) Connector
Procedure
1.Open unit using the Disassembly
Procedures (page 49) directions.
2. Remove printer assembly using the
Replacing the Printer Assembly (page 50)
directions.
3. Disconnect connections from the back of
the IEC connector.
4. Press in locking devices located on each
side of the IEC connector.
5. While pressing in locking devices, apply
pressure out on the connector.
Note Do one side at a time.
6. After the locking devices are released, pull
the connector out of the housing. See
Figure 53.
Figure 53. IEC (AC input) Connector
Housing
IEC
Connector
64Disassembly and Repair ProceduresWelch Allyn CP 100 and CP 200 Electrocardiograph
Replacing the LCD Inverter and LCD Assembly (CP 200)
Procedure
1.Open unit using the Disassembly
Procedures (page 49) directions.
2. Disconnect cable between the Inverter
board and keyboard.
3. Remove the 2 T4 torx screws holding the
LCD inverter in place. See Figure 54.
4. Disconnect the cable from the inverter to
LCD backlight. See Figure 54.
5. Disconnect the flat cable from the LCD at
the keyboard. See Figure 54.
6. Remove the 4 phillips screws in each
corner of the LCD. See Figure 54
7.Remove the LCD.
Figure 54. CP 200 Top Cover
Inverter
Board
T4 Torx
Screws
Phillips Screws
LCD
Assembly
Flat Cable from
the LCD to
Keyboard
Service ManualCP 100 and CP 200 Electrocardiograph65
Replacing the Function Keypad Assembly (CP 200)
Procedure
1.Open unit using the Disassembly
Procedures (page 49) directions.
2. Disconnect the cables to the LCD and the
Function Keyboard. See Figure 56.
3. Remove the 5 screws holding the Function
Keypad Assembly. See Figure 58.
4. Gently lift out the Function Keypad
Assembly.
Figure 55. Replacing the Keypad Assembly
LCD Assembly
Cable connection LCD
the and Function
Keypad
Phillips Screws
66Disassembly and Repair ProceduresWelch Allyn CP 100 and CP 200 Electrocardiograph
Replacing the Keyboard Assembly (CP 200)
Procedure
1.Open unit using the Disassembly
Procedures (page 49) directions.
2. Disconnect the ground and the flat data
cable to the LCD and the flat data cable,
the cables to the Inverter, to the Keypad
Assembly.
3. Remove the 13 screws holding in the
Keyboard Assembly.
4. Gently lift out the Keyboard Assembly.
Reassembly Note
1.Install and tighten the 3 screws first
(shown in Figure 56) to align the keyboard.
DO NOT tighten the screws past the
specified torque.
2. Install remaining 10 screws using the
specified torque.
Figure 56. Replacing the Keyboard Assembly (CP 200)
LCD Assembly
Cable between Inverter
board and Keyboard
Flat Cable from the
LCD to Keyboard
Keypad
Assembly
Screws
Service ManualCP 100 and CP 200 Electrocardiograph67
Replacing the LCD Assembly (CP 100)
Procedure
1.Open unit using the Disassembly
Procedures (page 49) directions.
2. Disconnect the flat cable from the
keyboard. See Figure 57.
3. Remove the 4 screws securing the LCD
bracket. See Figure 57.
4. Gently lift out the LCD bracket.
5. Gently lift out the LCD assemby.
Figure 57. CP 100 Top Cover
LCD Assembly
LCD Bracket
Screws
Flat Cable
68Disassembly and Repair ProceduresWelch Allyn CP 100 and CP 200 Electrocardiograph
Replacing the Keyboard Assembly (CP 100)
Procedure
1.Open unit using the Disassembly
Procedures (page 49) directions.
2. Disconnect the flat cable from the
keyboard.
3. Remove the 13 screws holding in the
keyboard assembly.
4. Gently lift out the keyboard assembly.
Reassembly Note
1.Install and tighten the 3 screws first
(shown in Figure 58) to align the keyboard.
DO NOT tighten the screws past the
specified torque.
2. Install remaining 10 screws using the
specified torque.
Figure 58. Replacing the Keyboard Assembly (CP 100)
Flat Cable
Screws
69
6
Field Replaceable Parts
Introduction
The field replaceable parts list is intended as a guide for ordering parts for both the CP 100
and CP 200 Electrocardiographs along with the optional Cart. The list is organized, by
assembly, with all parts to that assembly listed including quantity per assembly.
Note
“For information about any Welch Allyn product, please call Welch Allyn Technical
Support:” on page ii. or www.welchAllyn.com/medical/support.
Hardware pieces such as screws, washers, bearings, nuts, setscrews, and some
labels may only be available in kits. However, all such pieces are listed with sizes,
in the description, to allow for possible purchase from your local hardware store.
70Field Replaceable PartsWelch Allyn CP 100 and CP 200 Electrocardiograph
Field Replacement Parts
Table 7. MISCELLANEOUS PARTS LISTING
Part NumberDescriptionQuantity
703247SCREW, 6-32 X 375LG PHPS, V (Housing)8
76400POWER CORD, DETACH.(DOM/JAPAN)1
76402POWER CORD, DETACH.(EUROPEAN)1
76404POWER CORD, DETACH.(UK)1
76406POWER CORD, DETACH.(AUSTRALIA)1
761076-8POWER CORD, DETACH.(SWISS)1
761076-9POWER CORD, DETACH.(SOUTH AMERICA)1
761076-20POWER CORD, DETACH.(SOUTH AFRICA)1
701586DUST COVER, CP 100/CP 2001
701557DIRECTIONS FOR USE MANUAL, CP 200, English (All languages available on CD)1
401151CP 200 PRODUCT INFORMATION CD1
703228DIRECTIONS FOR USE, CP 100, ENGLISH (Other languages available on CD)1
401150CP 100 PRODUCT INFORMATION CD1
401126ECG PATIENT LEAD CABLE, AHA, CP 100/2001
401127ECG PATIENT LEAD CABLE, IEC, CP 100/2001
401128ECG PATIENT LEAD CABLE, IEC, Vac, CP 100/2001
Supported through the Welch Allyn CardioPerfect workstation software
exceed all AAMI standards
+10° C to +40° C (+50° F to +104° F)
15 – 95% noncondensing
700 – 1060 hPa
-20° C to +49° C (-4° F to +120° F)
15 – 95% noncondensing (30 – 70% for printing)
500 – 1060 hPa
Protection against electric shockClass I, internally powered
Typ e CF
Protection against ingress of water,
per IEC 60529:
Electrocardiograph
Spirometry Component
Mode of operationContinuous
*Per AAMI EC11:1991/(R)2001 Diagnostic Electrocardiographic Devices, Section 3.1.2.1 Disclosure of cautionary
information/ performance characteristics paragraph c) Accuracy of input signal reproduction, the manufacturer shall
disclose the methods used to establish overall system error and frequency response. Welch Allyn has used methods A & D,
as prescribed in section 3.2.7.2 and 4.2.7.2 of this same standard, to verify overall system error and frequency response.
Because of the sampling characteristics and the asynchronism between sample rate and signal rate, digital ECG systems
such as the CP 100 and CP 200 may produce a noticeable modulating effect from one cycle to the next, particularly in
pediatric recordings. This phenomenon is not physiologic.
Specifications are subject to change without notice.
IPXØ
IPXØ
EMC Guidance and Manufacturer’s
79
B
Table 15. Electromagnetic Emissions
The electrocardiograph is intended for use in the electromagnetic environment specified below. The customer or user of the electrocardiograph
should assure that it is used in such an environment.
Group 1The electrocardiograph uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby electronic
equipment.
Class AThe electrocardiograph is suitable for use in all establishments other than domestic and
those directly connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Class A
Complies
80Appendix B EMC Guidance and Manufacturer’s DeclarationsWelch Allyn CP 100 and CP 200 Electrocardiograph
Table 16. Electromagnetic Immunity
The electrocardiograph is intended for use in the electromagnetic environment specified below. The customer or user of the electrocardiograph
should assure that it is used in such an environment.
Immunity TestIEC 60601
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions, and
voltage variations on
power supply input
lines.
IEC 61000-4-11
Power frequency
(50/60Hz)
magnetic field
Test Level
± 6 kV contact
± 8 kV air
±2 kV for power supply
lines
±1 kV for input/output
lines
±1 kV differential mode
±2 kV common mode
>95% dip in 0.5 cycle
60% dip in 5 cycles
30% dip for 25 cycles
>95% dip in 5 seconds
3 A/m3 A/mPower frequency magnetic fields should be at levels
Floors should be wood, concrete, or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30%.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment. If the user of the electrocardiograph
requires continued operation during power mains
interruptions, it is recommended that the electrocardiograph
be powered from an uninterruptible power supply or battery.
characteristic of a typical location in a typical commercial or
hospital environment.
IEC 61000-4-8
Service ManualAppendix B EMC Guidance and Manufacturer’s Declarations81
Table 17. Electromagnetic Immunity
The electrocardiograph is intended for use in the electromagnetic environment specified below. The customer or user of the electrocardiograph
should assure that it is used in such an environment.
Immunity TestIEC 60601
Test Level
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Compliance
Electromagnetic Environment - Guidance
Level
Portable and mobile RF communications equipment should be used no closer
to any part of the electrocardiograph, including cables, than the
recommended separation distance calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
3 Vrmsd = (1.17)
3 V/md = (1.17) 80 MHz to 800 MHz
d = (2.33) 800 MHz to 2.5 GHz
P
P
P
where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended
separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,
each frequency range.
a
should be less than the compliance level in
b
Interference may occur in the vicinity of equipment marked with the
following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the
electrocardiograph is used exceeds the applicable RF compliance level above, the electrocardiograph should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the
electrocardiograph.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
82Appendix B EMC Guidance and Manufacturer’s DeclarationsWelch Allyn CP 100 and CP 200 Electrocardiograph
Table 18. Recommended Separation Distances Between Portable and Mobile RF Communications Equipment
and the Electrocardiograph
The electrocardiograph is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer
or user of the electrocardiograph can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the electrocardiograph as recommended below, according to the maximum output
power of the communications equipment.
Separation Distance According to Frequency of Transmitter (m)
Rated Max. Output
Power of Transmitter
(W)
0.010.1170.1170.233
0.10.370.370.74
11.171.172.33
103.703.707.37
10011.7011.7023.30
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects, and people.
150 kHz to 80 MHz
d = (1.17)
PPP
80 MHz to 800 MHz
d = (1.17)
800 MHz to 2.5 GHz
d = (2.33)
83
C
Software Updates
Caution Do not unplug or turn unit off during this process. It will cause the
system to be inoperable.