Welch Allyn CP-100, CP-200 Service manual

CP 100 and CP 200™ 12-Lead Resting Electrocardiograph
TM
CP 100TM 12- Lead Resting Electrocardiograph
CP 200
Service Manual
ii Welch Allyn CP 100 and CP 200 Electrocardiograph
Copyright 2005, Welch Allyn, Inc. All rights are reserved. No one is permitted to reproduce or duplicate, in any form, this manual or any part thereof without permission from Welch Allyn.
Caution: Federal US law restricts sale of the device identified in this manual to, or on the order of, a licensed physician.
Welch Allyn assumes no responsibility for any injury, or for any illegal or improper use of the product, that may result from failure to use this product in accordance with the instructions, cautions, warnings, or indications for use published in this manual.
Welch Allyn is a registered trademark of Welch Allyn, Inc., and CP 200, CP100, and CardioPerfect are trademarks of Welch Allyn, Inc.
SD is a trademark of Toshiba.
Software in this product is Copyright 2005, Welch Allyn, Inc., or its vendors. All rights are reserved. The software is protected by United States of America copyright laws and international treaty provisions applicable worldwide. Under such laws, the licensee is entitled to use the copy of the software incorporated within this instrument as intended in the operation of the product in which it is embedded. The software may not be copied, decompiled, reverse-engineered, disassembled or otherwise reduced to human-perceivable form. This is not a sale of the software or any copy of the software; all right, title and ownership of the software remains with Welch Allyn or its vendors.
For information about any Welch Allyn product, please call Welch Allyn Technical Support:
USA 1 800 535 6663
+1 315 685 4560
Canada 1 800 561 8797 China +86 216 327 9631
European Call Center +353 46 906 7790 France +33 15 569 5849
Germany +49 747 792 7186 Japan +81 33 219 0071
Latin America +315 685 2644 Netherlands +31 15 750 5000
Singapore +65 6419 8100 South Africa +27 11 777 7555
United Kingdom +44 207 365 6780 Sweden +46 85 853 6551
Australia +61 29 638 3000
800 074 793
Part Number (manual only): 704340 Rev D
Welch Allyn PO Box 220 Skaneateles Falls, NY 13153-0220
www.welchallyn.com
Printed in USA
0297

Contents

1 - Safety Summary. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
iii
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Servicing the Electrocardiograph Safely . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
General Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Electrostatic Discharge (ESD). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
General Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2 - Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Purpose and Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Service Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Warranty Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Non-Warranty Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Technical Support Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Returning Products. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Product Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Recommended Service Intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Controls, Indicators, and Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
“System Settings” Menu Tree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
About the Main Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Moving Through the Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Reviewing the Device Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
CP 100 and CP 200 Electrocardiograph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Service Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Entering the Service Info Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Print Log. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Battery Life Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Printer Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Export Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Import Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Delete Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Serial Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Lead Off Threshold. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Service Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3 - Functional Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Functional Verification Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Full Functional Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
iv Contents Welch Allyn CP 100 and CP 200 Electrocardiograph
Operating/Non-Printing Mode Current Test . . . . . . . . . . . . . . . . . . . . . . . . . 30
Battery Accuracy Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Sleep/Standby Mode Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Charger Detect Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
LED Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
LCD Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Set Unit Initial Configuration/SD Slot Test . . . . . . . . . . . . . . . . . . . . . . . . . . 32
ECG Communication Test (ECG Input test and lead off) . . . . . . . . . . . . . . . 33
Spirometer Communication Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
USB Communications Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Beeper Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Keyboard Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Printer Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Dielectric Strength Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Hi-Pot Test Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Chassis, Earth and Patient Leakage Current Tests. . . . . . . . . . . . . . . . . . . . 39
Checklist and Test Results Report Form . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
4 - Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Problem-Solving Suggestions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Service Policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
5 - Disassembly and Repair Procedures . . . . . . . . . . . . . . . . . . . . . . . 47
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Screws . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Torque Specs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Disassembly Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Opening the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Replacing the Printer Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Replacing the Paper Tray Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Replacing the Platen Roller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Replacing the Printhead . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Replacing the Stepper Motor and Gear Box. . . . . . . . . . . . . . . . . . . . . . . . . 57
Replacing the Battery or Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Replacing the Battery (DC) Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Replacing the Main Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Replacing the Power Supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Replacing the IEC (AC input) Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Replacing the LCD Inverter and LCD Assembly (CP 200) . . . . . . . . . . . . . . 64
Replacing the Function Keypad Assembly (CP 200) . . . . . . . . . . . . . . . . . . . 65
Replacing the Keyboard Assembly (CP 200) . . . . . . . . . . . . . . . . . . . . . . . . 66
Replacing the LCD Assembly (CP 100) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Replacing the Keyboard Assembly (CP 100). . . . . . . . . . . . . . . . . . . . . . . . . 68
6 - Field Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Field Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Service Manual CP 100 and CP 200 Electrocardiograph v
A - Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
B - EMC Guidance and Manufacturer’s Declarations . . . . . . . . . . . . . 79
C - Software Updates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
D - Interconnect Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
CP 200 Interconnect Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
CP 100 Interconnect Diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
vi Contents Welch Allyn CP 100 and CP 200 Electrocardiograph
1
1

Safety Summary

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Servicing the Electrocardiograph Safely . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2 Safety Summary CP 100 and CP 200 Electrocardiograph

Introduction

The safety summary, and all additional specific warnings and cautions located throughout the documentation, must be read and understood by all users of the CP 100 and CP 200 Electrocardiograph.

Symbols

The symbols illustrated on the following pages may appear on the electrocardiograph, the packaging, the shipping container, or in this manual.
Documentation Symbols
WARNING Indicates conditions or practices that could lead to illness, injury, or death.
Caution In the manual, indicates conditions or practices that could damage the equipment or other property.
Caution On the product, means “Consult accompanying documentation.”
-20°C
EC REP
Shipping, Storing, and Environment Symbols
This end up Keep dry
Fragile Altitude limits
+49°C
Temperature limits Relative humidity limit
Certification Symbols
Meets essential requirements of European Medical Device Directive 93/42/EEC
European Regulatory Manager Welch Allyn LTD. Navan Business Park Dublin Road Navan, County Meath, Republic of Ireland Tel.: 353-46-90-67700 Fax: 353-46-90-67756
N344
Complies with applicable U.S. and Canadian medical safety standards
Australian registered importer
Directions for Use Safety Summary 3
Operation Symbols
For details on the keys, see “Keyboard Functions” on page 14.
On/standby (off) Help
Navigation arrows Backspace
Enter Shift
Auto ECG Menu
Rhythm ECG Stop/Cancel
SD memory card slot Com port A
(for patient cable)
Alternating current Battery charge level
Dangerous voltage Direct current
Defibrulation-proof type CF applied parts
Ground equipotential (while the electrocardiograph is safety-rated “CF” for direct cardiac contact it is not intended to be connected directly to the patient’s heart; only surface contact with the patient’s skin is intended).
Do not reuse. Sealed lead-acid battery
Type BF Defibrillator proof Recycle
Do not dispose of this product as unsorted municipal waste. Prepare this product for reuse or separate collection as specified by
T2.0A/250V
AC fuse replacement information
Directive 2002/96/EC of the European Parliament and the Council of the European Union on Waste Electronic and Electrical Equipment (WEEE). If this product
is
contaminated, this directive does not apply.
For more specific disposal information, see www.welchallyn.com/weee, or contact Welch Allyn Customer Service at +44 207 365 6780.
Additional Operation Symbols for CP 200
Spirometry Key OK
Spirometry port Com port B
(for USB cable)
4 Safety Summary CP 100 and CP 200 Electrocardiograph

Servicing the Electrocardiograph Safely

Before using or servicing the electrocardiograph, you must read and understand the following safety-related information.

General Warnings

The following warning statements apply to electrocardiograph use in general. Warning statements that apply specifically to particular procedures, such as connecting the patient cable or performing an ECG test, appear in the corresponding sections of the manual.
Warning statements indicate conditions or practices that could lead to illness, injury, or death.
Warnings Related to the Environment
WARNING To avoid a possible explosion, do not use the electrocardiograph in
the presence of flammable anesthetics.
WARNING When transporting the electrocardiograph on a cart, tuck the patient cable away from the wheels so that it does not present a hazard.
WARNING Do not use the CP 100 and CP 200 Electrocardiograph in an MRI suite or hyperbaric chamber.
WARNING Do not autoclave the CP 100 and CP 200 Electrocardiograph or patient cable.
Warnings Related to Accessories and Other Equipment
WARNING For operator and patient safety, peripheral equipment and
accessories that can come in direct patient contact must be in compliance with all appropriate safety, EMC, and regulatory requirements. See “EMC Guidance
and Manufacturer’s Declarations” on page 79
WARNING All signal input and output (I/O) connectors are intended for connection of only devices complying with IEC 60601-1, or other IEC standards (for example, IEC 60950), as appropriate to the device. Connecting additional devices to the electrocardiograph may increase chassis or patient leakage currents. To maintain operator and patient safety, consider the requirements of IEC 60601-1-1. Measure the leakage currents to confirm that no electric shock hazard exists.
WARNING The electrocardiograph has not been designed for use with high­frequency (HF) surgical equipment and does not protect against hazards to the patient.
Directions for Use Safety Summary 5
Warnings Related to Using the Electrocardiograph
WARNING Avoid positioning any leads or cables so that they could easily trip
someone.
WARNING Satisfactory maintenance procedures must be implemented, or equipment failure and health hazards may result.
Warnings Related to Repairing the Electrocardiograph
WARNING Only qualified service personnel should attempt to repair the
electrocardiograph. In case of a malfunction, call Technical Support and precisely describe the problem. For phone numbers, see page ii.
WARNING While under warranty, the electrocardiograph must be serviced only by a Welch Allyn service technician.
WARNING Electrostatic discharge (ESD) can damage or destroy electronic components. Handle static-sensitive components only at static-safe workstation.
WARNING Consider all electrical and electronic components of the monitor as static-sensitive.

Electrostatic Discharge (ESD)

Electrostatic discharge is a sudden current flowing from a charged object to another object or to ground. Electrostatic charges can accumulate on common items such as foam drinking cups, cellophane tape, synthetic clothing, untreated foam packaging material, and untreated plastic bags and work folders, to name only a few.
Electronic components and assemblies, if not properly protected against ESD, can be permanently damaged or destroyed when near or in contact with electrostatically charged objects. When you handle components or assemblies that are not in protective bags and you are not sure whether they are static-sensitive, assume that they are static-sensitive and handle them accordingly.
Perform all service procedures in a static-protected environment. Always use techniques and equipment designed to protect personnel and equipment from electrostatic discharge.
Remove static-sensitive components and assemblies from their static-shielding bags only at static-safe workstations—a properly grounded table and grounded floor mat— and only when you are wearing a grounded wrist strap (with a resistor of at least 1 megohm in series) or other grounding device.
6 Safety Summary CP 100 and CP 200 Electrocardiograph
Use only grounded tools when inserting, adjusting, or removing static-sensitive components and assemblies.
Remove or insert static-sensitive components and assemblies only with monitor power turned off.
Insert and seal static-sensitive components and assemblies into their original static­shielding bags before removing them from static-protected areas.
Always test your ground strap, bench mat, conductive work surface, and ground cord before removing components and assemblies from their protective bags and before beginning any disassembly or assembly procedures.

General Cautions

The following caution statements apply to electrocardiograph use in general. Caution statements that apply specifically to particular procedures, such as connecting the patient cable or performing an ECG test, appear in the corresponding sections of the manual.
Caution statements indicate conditions or practices that could damage the equipment or other property.
Caution When removing the electrocardiograph from storage, allow it to thermally stabilize to surrounding environmental conditions before using it.
Caution To prevent possible damage to the keypad, do not use sharp or hard objects to press keys. Only use fingertips.
Caution Do not expose the patient cable to strong ultra-violet radiation.
Caution Do not pull or stretch the patient cable. Doing so could result in
mechanical or electrical failures. Form the patient cable into a loose loop before storing.
Caution Avoid positioning the patient cable where it might get pinched or stepped on. If the cable’s impedance is altered, measurements might no longer be accurate, and repair might be necessary.
Caution Using the equipotential terminal for anything but grounding purposes may contribute to damage of the device.
Caution Use only parts and accessories supplied with the device and available through Welch Allyn. The use of accessories other than those specified may result in degraded performance of this device.
Caution Portable and mobile RF communications equipment can affect the performance of the electrocardiograph.
Caution Other medical equipment—including but not limited to defibrillators, ultrasound machines, pacemakers, and other stimulators—may be used simultaneously with the electrocardiograph. However, such devices may disturb the electrocardiograph signal.
7
2

Overview

Purpose and Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Service Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Product Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Recommended Service Intervals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Controls, Indicators, and Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
“System Settings” Menu Tree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
About the Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Moving Through the Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Reviewing the Device Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
CP 100 and CP 200 Electrocardiograph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
8 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph

Purpose and Scope

The service manual is a reference for periodic preventive maintenance and corrective service procedures for the Welch Allyn CP 100 and CP 200 Electrocardiograph.
Corrective service is supported to the level of field-replaceable units. This includes some circuit-board assemblies and some subassemblies, case parts, and other parts. See “Field
Replaceable Parts” on page 69 for complete list of user-replaceable service parts.
Note
This guide provides troubleshooting information, assembly procedures, and instructions for functional testing and performance verification. It is intended for use only by technically qualified service personnel.
Repair and replacement of the main board is not supported. All service work on the main board must be performed by certified and qualified service personnel at an authorized Welch Allyn service center.
Caution No component-level repair of circuit boards and subassemblies is supported. Use only the repair procedures described in this manual.
WARNING When performing a service procedure, follow the instructions exactly as presented in this manual. Failure to do so could damage the monitor, invalidate the product warranty, and lead to serious personal injury.

Service Options

Warranty Service

All repairs on products under warranty must be performed and/or approved by Welch Allyn. Refer all warranty service to Welch Allyn Factory Service or another authorized Welch Allyn Service Center. Obtain an RMA number for all returns to Welch Allyn Factory Service – see Returning Products (page 9).
Caution Unauthorized repairs will void the product warranty.

Non-Warranty Service

Welch Allyn Factory Service and authorized Service Centers support non-warranty repairs. Contact any Welch Allyn regional service center for pricing and service options.
Welch Allyn offers modular repair parts for sale to support non-warranty service. This service must be performed only by qualified end-user biomedical/clinical engineers using this service manual.
The Welch Allyn Electrocardiograph Service Information supports certain service functions. For information about any Welch Allyn product, please call Welch Allyn
Technical Support: (page ii).
Service Manual CP 100 and CP 200 Electrocardiograph 9

Technical Support Services

Welch Allyn offers the following technical support services:
Telephone support Loaner equipment Service agreements Replacement service parts Factory Service
For information on any of these services, contact Welch Allyn at the numbers listed on
page ii

Returning Products

To return a product for service, contact Welch Allyn Technical Support and request a Return Material Authorization (RMA) number.
Note
To ship the unit, please observe these packing guidelines:
Welch Allyn does not accept returned products without an RMA.
When requesting an RMA, please have the following information available:
Product name, model number, and serial number
A complete return shipping address, including a contact name and phone
number; include any special shipping instructions
A purchase-order number or credit-card number if the product is not covered by
warranty
A full description of the problem or service request
Remove from the package all hoses, connectors, cables, sensors, power cords,
and other ancillary products and equipment, except those items that might be associated with the problem.
Use the original shipping carton and packing materials, or as close an
approximation as possible.
Include a packing list.
Write the Welch Allyn RMA number with the Welch Allyn address on the outside
of the shipping carton.
United States federal regulations require that any unit received by Factory Service must be free from blood-borne pathogens before processing. All incoming products are cleaned as well as possible, but products that cannot be effectively cleaned cannot be accepted for repair. Please thoroughly clean all organic residues from the product before shipment. This will ensure safe receipt, processing and repair, and will help expedite the return of your device.
10 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph

Product Configurations

Model numbers for the configurations are as follows:
Model Number Description
CP 1 CP 100 Electrocardiograph
CP 1A CP 100 Electrocardiograph w/Interpretation
CP 2 CP 200 Electrocardiograph
CP 2A CP 200 Electrocardiograph w/Interpretation
CP 2S CP 200 Electrocardiograph w/Spirometry
CP 2AS CP 200 Electrocardiograph w/Interpretation & Spirometry

Recommended Service Intervals

Interval or Condition Action Recommended Procedure
Every 6 – 24 months (per hospital protocols)
Battery does not hold a charge Check battery capacity
Unit has been dropped or otherwise damaged
Unit malfunctioning Complete Functional test Functional Verification (page 29)
Unit does not pass Functional Verification
Complete Functional test Functional Verification (page 29)
Battery Accuracy Test (page 31)
Replace battery
Complete Functional test Functional Verification (page 29)
Troubleshooting and repair followed by functional test
Return to authorized service center
Replacing the Battery or Fuse (page 58)
Troubleshooting (page 41) Disassembly and Repair Procedures (page
47) Functional Verification (page 29)
Returning Products (page 9)
Service Manual CP 100 and CP 200 Electrocardiograph 11

Controls, Indicators, and Connectors

This section describes the controls, indicators, and connectors that are part of the electrocardiograph.
Figure 1. Top of CP 200 Electrocardiograph
Softkeys and functions keys See Figure 10 on page 15.
Keyboard See Figure 8 on page 14.
Figure 2. Top of CP 100 Electrocardiograph
Keyboard See Figure 9 on page 14.
12 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph
Figure 3. Back of Electrocardiograph
AC power inlet
Equipotential stud
AC fuses
Figure 4. Right Side of Electrocardiograph
Com port B
(for USB cable)
SD memory card
slot
(Not functional on CP 100)
Figure 5. Left Side of Electrocardiograph
Paper tray latch
Spirometry port
(Not functional on CP 100)
Reset
Button
Service Manual CP 100 and CP 200 Electrocardiograph 13
Figure 6. Front of CP 200 Electrocardiograph
Com port A
(for patient cable)
Figure 7. Front of CP 100 Electrocardiograph
Com port A
(for patient cable)
14 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph

Keyboard Functions

Figure 8. Keyboard for CP 200 Electrocardiograph
On/Off
See “ Powering the Electrocardiograph” in the Directions of Use manual.
Green LED
Lights up when the electrocardiograph is connected to AC power.
Ta b
Moves through the data-entry fields.
Shift
Capitalizes letters.
Space
Enters a space.
Figure 9. Keyboard for CP 100 Electrocardiograph
Backspace
Deletes the character to the left of the cursor.
Enter
See “Moving Through the Menus” on page 18.
Menu
See “About the Main Menu” on page 17.
Help
See the Directions For Use manual.
Navigation arrows
See “Moving Through the
Menus” on page 18.
Green LED
Lights up when the electrocardiograph is connected to AC power.
Ta b
Moves through the data-entry fields.
Help
See the Directions for Use manual.
Shift
Capitalizes letters.
On/Off
See “ Powering the Electrocardiograph” in the Directions of Use manual.
Space
Enters a space.
Stop/Cancel
Stops any current activity. See “Moving Through the
Menus” on page 18.
Backspace
Deletes the character to the left of the cursor.
Enter
See “Moving Through the
Menus” on page 18.
Menu
See “About the Main
Menu” on page 17.
Auto ECG
Begins Auto ECGs, normal and stat.
Rhythm ECG
Begins a Rhythm ECG. .
Navigation arrows
See “Moving Through the
Menus” on page 18.
Service Manual CP 100 and CP 200 Electrocardiograph 15
Figure 10. Softkeys and Function Keys for the CP 200 Electrocardiograph
Softkeys
These softkeys display text or images that correspond to the unlabeled buttons below them. The content changes from screen to screen.
Softkey buttons
These buttons activate the functions displayed above them. If a softkey is blank, pressing its
Auto ECG Spirometry Rhythm ECG Stop/Cancel OK
button has no effect.
Function Keys
Auto ECG
Begins Auto ECGs, normal and stat.
Spirometry
Begins spirometry tests. See spirometry manual.
Rhythm ECG
Begins Rhythm ECGs.
Stop/Cancel
Stops any current activity. See “Moving Through the Menus” on page 18.
OK
Accepts data that you have entered, or chooses a highlighted item. See “Moving Through the Menus” on page 18.
16 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph

“System Settings” Menu Tree

P 200 Menu Tree
System Settings
Device
Configuration
Set Date/Time Language Date Format Time Format Weight Unit Height Unit Power-Save Device ID Audio Beeper Flow Unit Pressure Unit Temperature Increase Display Contrast Decrease Display Contrast Contact Information Previous Menu * Applicable for spirometry only.
*
*
*
CP 100 Menu Tree
Device
Info
About Print Settings Enable Options Upgrade Software Service Info Previous Menu
System Settings
Device
Administration
See “Managing Data” in the Directions for Use manual.
Service Info
Print Log Battery Life Test Printer Test Export Files Import Files Delete Files Serial Number Lead Off Thershold Service Tests LCD Test Previous Menus
Device
Configuration
Set Date/Time Language Date Format Time Format Weight Unit Height Unit Power-Save Device ID Audio Beeper Increase Contrast Decrease Contrast Contact Information Previous Menu
Device
Info
About Print Settings Enable Options Upgrade Software Service Info Previous Menu
Service Info
Print Log Battery Life Test Printer Test Export Files Import Files Delete Files Serial Number Lead Off Thershold Service Tests Previous Menus
Service Manual CP 100 and CP 200 Electrocardiograph 17

About the Main Menu

The main menu appears when you press the Menu key .
Figure 11. Main Menu CP 200
Main Menu
1 Test Directory 2 Scheduled Patients 3 ECG Settings 4 Spirometry Settings 5 System Settings 6 Edit Medication List 7 Edit History List 0 Exit
Figure 12. Main Menu CP 100
Main Menu
1 ECG Settings 2 System Settings 0 Exit
9:17AM Mar 16 05
9:17AM Mar 16 05
Submenu Purpose Procedure
Test Directory View, change, print, or send saved tests. (CP
200 Only)
Scheduled Patients View the scheduled patients list, add
patients to the list, or delete patients from
See the Directions for Use manual on your specific model.
See the Directions for Use manual on your specific model.
the list. (CP 200 Only
ECG Settings Review or change ECG settings: Auto
Report format, Rhythm Report format, and
See the Directions for Use manual on your specific model.
so on.
Spirometry Settings Review or change spirometry settings:
display settings, print settings, and so on. (CP
See the Directions for Use manual on your specific model.
200 Only
System Settings Review or change system settings: device
configuration, device info, user setup, and
See the Directions for Use manual on your specific model.
so on.
Edit Medication List Edit the list of medication choices available
to choose during patient data entry. (CP 200
See the Directions for Use manual on your specific model.
Only
Edit History List Edit the list of clinical conditions available to
choose during patient data entry. (CP 200
See CP 200 Electrocardiograph Directions for Use manual.
Only
18 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph

Moving Through the Menus

Figure 13. CP 200 Standard Menu Figure 15. CP 200 Parent Menu With Submenu
ECG Setup : Edit Auto Report 1
1 Format
2 Interp Settings 3 Patient Data 0 Previous Menu
9:17AM Mar 16 05
Figure 14. CP 100 Standard Menu
Edit Auto Report
1 Format
2 Interp Settings 3 Patient Data 0 Previous Menu
9:17AM Mar 16 05
Edit Auto Report 1 : Format
1 Lead Arrangement 2 Rhythm Lead 1 3 Rhythm Lead 2 4 Rhythm Lead 3 5 Extended Measurements 6 Average Cycles 0 Previous Menu
9:17AM Mar 16 05
3x4 3x4 +1R 3x4 +3R 6x2
12x1
6x2 50 mm/s 6x2 10 Ext. No Print
Figure 16. CP 100 Parent Menu with Submenu
Format
1 Lead Arrangement
2 Rhythm Lead 1 3 Rhythm Lead 2 4 Rhythm Lead 3 5 Extended Measurements 6 Average Cycles 0 Previous Menu
9:17AM Mar 16 05
3x4 +1R 3x4 +3R 6x2 12x1 6x2 50 mm/s 6x2 10 Ext. No Print
Desired Actions Keys to Press
To move up or down a list or (keyboard or softkey arrows)
To open a standard menu (Figure 13)
or or
or item’s number or letter
To move from parent menu to submenu on same screen (Figure 15)
To perform an action
or
To accept data To check or uncheck a checkbox
To return to parent menu from submenu on same screen (Figure 15) or
(To select the highlighted submenu item.)
or
(To make no change.)
To move back through the menus or zero key
To move through data-entry fields
To return to the “Main Screen” from a standard menu (Figure 13)
Note
Keyboard and softkey arrows work the same way.
Service Manual CP 100 and CP 200 Electrocardiograph 19

Reviewing the Device Information

1. Press the Menu key .
2. Choose System Settings > Device Info.
The following screen appears.
Figure 17. “Device Info” CP 200 Screen
System Setup : Device Info
1 About 2 Print Settings 3 Enable Options 4 Upgrade Software 5 Service Info 0 Previous Menu
9:17AM Mar 16 05
Figure 18. “Device Info” CP 100 Screen
Device Info
1 About 2 Print Settings 3 Enable Options 4 Upgrade Software 5 Service Info 0 Previous Menu
9:17AM Mar 16 05
3. Select the desired item:
Item Description
About Displays the following information about the electrocardiograph:
• serial number
• modules configured
• version numbers
Print Settings CP 200 = Prints your ECG, spirometry, and system settings as well as medication &
history lists. CP 100 = Prints your ECG and system settings.
Enable Options Contact Technical Support. For phone numbers, see page ii.
Upgrade Software Contact Technical Support. For phone numbers, see page ii.
Service Info Accessible to service support only. See Service Menus (page 20)
20 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph

CP 100 and CP 200 Electrocardiograph

Service Menus

Entering the Service Info Screen

This menu is password protected and only Welch Allyn Authorized Technician can access the Service Menus,)
CP 200 Service Info Screen Procedure Figure 19. CP 200 “User Login”
1. Power ON unit.
2. Press Menu key.
3. Choose System Setting> Device Info> Service Info
The “UserLogin” screen appears. See Figure 19.
UserLogin
9:17AM Mar 16 05
Use the “Tab” or “Arrow” keys to move the cursor from field to field.
Enter the following: User ID: Service (case sensitive) Password:normser (case sensitive)
4. Press “Done” using the softkey. Do not use the“Enter” key.
The “Service Info” screen appears see Figure 20
User ID
Password
Enter valid user ID and password
Change
Password
Cancel
Figure 20. CP 200 “Service Info” Screen
9:17AM Mar 16 05
Service Info
1.Print Log
2.Battery Life Test
3.Printer Test
4.Export Files
5.Import Files
6. Delete Files
7.Serial Number
8.Lead Off Thershold
9.Service Tests A LCD Test
0.Previous Menus
Done
Service Manual CP 100 and CP 200 Electrocardiograph 21
CP 100 Service Info Screen Procedure Figure 21. CP 100 “UserLogin” Screen
1. Power ON unit.
2. Press Menu key.
3. Choose System Setting > Device Info > Service Info
User Login
9:17AM Mar 16 05
The “UserLogin” screen appears. See Figure 21.
User ID
Use the “Arrow” keys to move the cursor from field to field.
Password
Enter the following: User ID: Service (case sensitive)
Done
Password:normser (case sensitive)
4. Highlight “Done” using the down arrow key.
5. Press “Enter” key.
The “Service Info” screen appears. See Figure 22
Figure 22. CP 100 “Service Info” Screen
9:17AM Mar 16 05
Service Info
1. Print Log
2. Battery Life Test
3. Printer Test
4. Export Files
5. Import Files
6. Delete Files
7. Serial Number
7. Lead Off Thershold
8. Service Tests
0. Previous Menus

Print Log

Caution These printouts can be very long. Only use the printouts if necessary. To stop the printouts press Stop/Cancel.
Function Description
Print Events: This prints all events stored in the unit in descending
order by date and time. Used for development of the software. Has no Service value.
Print Errors: This prints any errors that occurred in the unit, in
descending order by date and time, during normal operation. This can be valuable to use when the end­user says he/she has been having problems.
Print All: This prints the Print Events and Print Errors as
described above.
Figure 23. “Print Log” Screen
Print Log
1.Print Events
2.Print Errors
3.Print All
0.Previous Menus
9:17AM Mar 16 05
22 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph

Battery Life Test

This displays the percentage of battery life remaining on the battery, based on the number of charges it has incurred. This percentage approximates the life left in the battery based on this information.
Caution Only reset when the battery has been replaced and charged for at least 12hrs continuously.
Function Description
Reset Resets the battery life to 100%. This is used only
when the battery has been replaced
Reset Pages Resets page count for the number of prints made on
the printer. Should only be reset when print head has been replaced.

Printer Test

Function Description
Figure 24. “Battery Life Test” Screen
9:17AM Mar 16 05
Battery Life Test
Battery Life Remaining: 99%
Reset
Reset Pages
Figure 25. Printer Test” Screen
Exit
Test Page Prints a test page which repeats until the “Cancel”
button is pressed. Refer to Functional Verification test for explanation. See Printer Test (page 35).
Ramp Test Prints a one page ramp test. Refer to Functional
Verification test for explanation. See Printer Test
(page 35).
Printer Test
Test Page
9:17AM Mar 16 05
Select a print test
Ramp
Page
Exit
Service Manual CP 100 and CP 200 Electrocardiograph 23

Export Files

Use the “Export Files” function when the Main PCB is replaced, to reinstall the customer’s saved ECG information and spirometry (CP 200 Only) or the unit’s original configuration after repair.
Caution The tests exported to the SD memory card contain patients’ medical information and should be treated the same as any confidential medical information.
Note
If you are performing all functions you only need one SD memory card. Use a formatted blank SD memory card installed in the unit for each function.
Function Description
Export Log Files Exports the Event log to a SD memory card in a folder
called Log. Once complete, a message appears at bottom of the screen stating if the export was successful or not.
Export Audit Trail (CP 200 ONLY)
Export Tests (CP 200 ONLY)
Export Configuration Files:
Exports the audit information of the unit to a SD memory card in a folder called Audit. Once complete a message appears at bottom of screen stating if export was successful or not.
Note: Audit information is available if Audit trail is turned on in Device Administration. See the CP 200 Directions for Use manual
Exports all tests in the memory to a SD memory card in a folder called Tests for importing after repair.
Exports the configuration of the unit (i.e. print formats, patient data settings, logins, etc.) to a SD memory card in a folder called Config for importing after repair.
Figure 26. “Export Files” Screen
Export Files
1.Export Log Files
2.Export Audit Files
3.Export Tests
4.Export Configuration Files
0.Previous Menus
9:17AM Mar 16 05
Loading...
+ 67 hidden pages