Welch Allyn Cardio Perfect Workstation Spirometer Installation Guide

Page 1
CARDIOPERFECT
WORKSTATION
INSTALLATION GUIDE
Windows® XP
INSTALLING THE SPIROMETER ON A CARDIOPERFECT WORKSTATION
1 Assemble the spirometer, and then plug it into
any USB port.
Caution: With the spirometer, always use the USB extension cable, as shown. This cable helps to prevent damage to the spirometer.
2 Set up the USB ports.
This screen might not appear.
(next page)
Page 2
CARDIOPERFECT
WORKSTATION
INSTALLATION GUIDE
Caution: When the Found New Hardware wizard appears this second time, do not cancel. You must complete the wizard twice per USB port. Otherwise, the software will not be installed properly.
This screen might not appear.
Set up other USB ports for possible future use with this device:
• Remove the cable, and insert it into the next available USB port.
• Repeat Step 2 for that port.
• Continue repeating Step 2 for all available USB ports.
2
Page 3
3 Verify the installation of the spirometer.
If your spirometer is not already plugged into the desired USB port, plug it in now. Attach calibration syringe as illustrated.
Welch Allyn
CardioPerfect
c
a
b
(a) Click the Go button. (b) Select your test patient. (c) Click the Spirometry button.
The lot code and calibration code appear on the flow transducer box:
(next page)
3
Page 4
CARDIOPERFECT
WORKSTATION
INSTALLATION GUIDE
b
a
(a) Read the instructions on this screen. (b) Click the Calibrate button.
Follow the instructions on this screen.
..3.. ..2.. ..1..
Don’t worry if your screen says ATS standard has NOT been met. You don’t need to meet the standard for this verification.
b
a
(a) Verify that you see a flow loop. Each flow loop
looks slightly different. (b) Click Done. You have verified that your spirometer is communicating with the Workstation. Before performing spirometry tests, be sure to perform a successful calibration.
Push and then pull the plunger in sync with the side-to-side motion of the blue bar. Move steadily, all the way in and then all the way out, without hesitating between push and pull.
4341 State Street Road, PO Box 220, Skaneateles Falls, NY 13153-0220 USA (p) 800.535.6663 (f) 315.685.2174 www.welchallyn.com