Welch Allyn assumes no responsibility for any injury to anyone, or for any illegal or improper use of the product, that may result from
failure to use this product in accordance with the instructions, cautions, warnings, or statement of intended use published in this
manual.
Welch Allyn, Propaq, Acuity, and Smartcuf are registered trademarks of Welch Allyn.
SET and Masimo are registered trademarks of Masimo Corporation. Possession or purchase of a Masimo SpO
does not convey any express or implied license to use the device with unauthorized sensors or cables which would, alone or in
combination with this device, fall within the scope of one or more of the patents relating to this device.
Nellcor and Oximax are registered trademarks of Nellcor Puritan Bennett.
Software in this product is copyright by Welch Allyn or its vendors. All rights are reserved. The software is protected by United States
of America copyright laws and international treaty provisions applicable worldwide. Under such laws, the licensee is entitled to use
the copy of the software incorporated with this instrument as intended in the operation of the product in which it is embedded. The
software may not be copied, decompiled, reverse-engineered, disassembled or otherwise reduced to human-perceivable form. This
is not a sale of the software or any copy of the software; all right, title and ownership of the software remain with Welch Allyn or its
vendors.
For information about any Welch Allyn product, call the nearest Welch Allyn representative:
-equipped monitor
2
USA+ 1 315 685 4560
800 535 6663
Australia+ 61 2 9638 3000
800 074 793
Canada 800 561 8797China+ 86 216 327 9631
European Call Center + 35 3 46 906 7790France+ 33 1 60 09 33 66
Germany+ 49 7477 92 71 86Japan+ 81 3 3219 0071
Latin America+ 1 305 669 9003Netherlands+ 31 15 750 5000
This Safety Summary should be read by all Propaq Encore users. Specific warnings and
cautions will be found throughout the Propaq Encore documentation where they apply.
WARNING This monitor is to be operated by qualified personnel only. The
operator of this monitor should read this entire manual, the monitor reference
guide or directions for use, and all accessory directions for use before operating
the monitor.
WARNING Place the Propaq monitor and accessories in locations where they
cannot harm the patient if they fall from their shelf or mount. Lift the monitor only
by its handle; do not lift it by any attached cables.
WARNING Do not connect more than one patient to a monitor. Do not connect
more than one monitor to a patient.
WARNING Do not use the Propaq Encore in an MRI suite or a hyperbaric
chamber.
WARNING Do not autoclave the Propaq. Autoclave accessories only if the
manufacturer's instructions clearly approve it. Many accessories can be severely
damaged by autoclaving.
WARNING Inspect the power adapter cord periodically for fraying or other
damage, and replace the adapter as needed. Do not operate the apparatus from
mains power with a damaged power adapter cord or plug.
WARNING When using a power adapter with this monitor, be sure to connect
the power adapter to a three-wire, grounded, hospital-grade receptacle. Do not
under any circumstances attempt to remove the grounding conductor from the
power plug of the power adapter. Do not plug the power adapter into an
extension cord. If there is any doubt about the integrity of the protective earth
ground of the receptacle for the power adapter, do not plug in the power adapter;
operate the monitor only on battery power. Contact your biomedical engineering
department for assistance in identifying the proper power receptacle and making
appropriate power connections.
WARNING Make frequent electrical and visual checks on cables and electrode
wires.
WARNING Avoid electrosurgery burns at monitoring sites by ensuring proper
connection of the electrosurgery return circuit so that the return paths cannot be
made through monitoring electrodes and probes.
2General informationWelch Allyn Propaq Encore Vital Signs Monitor
WARNING During defibrillation, keep the discharge paddles away from ECG and
other electrodes, as well as other conductive parts in contact with the patient.
Avoid contact with any accessories connected to the Propaq’s left side panel.
WARNING To ensure patient safety, the conductive parts of the ECG electrodes
(including associated connectors) and other patient-applied parts should not
contact other conductive parts, including earth ground, at any time.
WARNING Do not operate this product in the presence of flammable
anesthetics or other flammable substance in combination with air, oxygenenriched environments, or nitrous oxide; explosion can result.
WARNING Within certain governmental jurisdictions, all interconnected
accessory equipment must be labeled by an approved testing laboratory. After
interconnection with accessory equipment, risk (leakage) current and grounding
requirements must be maintained.
WARNING To ensure conformance to risk (leakage) current requirements when
operating from an ac mains power source, use only a Welch Allyn
®
503-0054
series power adapter.
WARNING This monitor should only be repaired by qualified service personnel.
The operator should not attempt to open the monitor case or perform any
maintenance on the monitor except for procedures explicitly described in this
manual that can be performed by operators such as inspection and cleaning.
WARNING To ensure patient safety, use only accessories recommended or
supplied by Welch Allyn. For a list of those accessories, see Products and Accessories (810-0409-XX), which accompanies this manual. Accessories must
be used according to your hospital’s standards and the manufacturer’s
recommendations. Always refer to the manufacturer’s directions for use. Use of
accessories not recommended by Welch Allyn may result in increased
electromagnetic emissions or decreased electromagnetic immunity of the
equipment.
WARNING Safe interconnection between the Propaq monitor and other devices
must comply with applicable medical systems safety standards such as IEC
60601-1-1. Within certain governmental jurisdictions, all interconnected accessory
equipment must be labeled by an approved testing laboratory. After
interconnection with accessory equipment, risk (leakage) current and grounding
requirements must be maintained.
WARNING As with all medical equipment, carefully route the patient cabling to
reduce the possibility of patient entanglement or strangulation.
WARNING A product that has been dropped or severely abused should be
checked by qualified service personnel to verify proper operation and acceptable
risk (leakage) current values.
WARNING The pulse oximetry channel should NOT be used as an apnea
monitor.
WARNING Do not use the pulse oximeter as a replacement or substitute for
ECG-based arrhythmia analysis.
Reference GuideGeneral information3
WARNING If the monitor detects an unrecoverable problem, an error message
window appears containing an error number and a short message. Report such
errors to Welch Allyn.
WARNING When taking NIBP measurements, periodically observe the patient’s
limb to make sure that the circulation is not impaired for a prolonged period of
time. Also make sure the blood pressure cuff is properly placed according to
Propaq Encore Directions for Use or Propaq Encore Reference Guide. Be
especially careful when using the short-term automatic mode (TURBOCUF).
Prolonged impairment of circulation or improper cuff placement can cause
contusions.
WARNING The range of values measured by the monitoring parameters is
provided in the Specifications section of Propaq Encore Directions for Use or Propaq Encore Reference Guide. Operation of the monitor outside the range of
specified values is not recommended and may cause inaccurate results.
WARNING Electronic equipment that emits very strong electromagnetic or radio
frequency signals can cause electrical interference with monitor operation,
including causing the monitor to turn off power. Avoid operating this monitor near
such equipment. For guidance about electromagnetic emissions and the
recommended separation distance between the monitor and such equipment,
refer to the specifications section of this manual.
Symbols
Caution Changes or modifications not expressly approved by Welch Allyn could
void the purchaser’s authority to operate the equipment.
Caution Federal (U.S.A.) law restricts this device to sale, distribution, or use by
or on the order of a licensed medical practitioner.
The Propaq Encore should be serviced only by a Welch Allyn service technician while
under warranty. Propaq Encore Service Manual (810-0696-XX) is available from Welch
Allyn to assist the biomedical engineer during post-warranty period service.
The following symbols may appear on the Propaq Encore monitor or accessories or
documentation. These internationally recognized symbols are defined by the International
Electrotechnical Commission, IEC 878 and IEC 417A.
WARNING Indicates conditions that could lead to illness, injury, or death.
Caution In this manual, indicates conditions that could damage equipment or other property.
Caution On the product, means “Consult the accompanying documentation.”
Off (Standby)Patient connections are
Type CF, isolated for direct
cardiac application, and
protected against
defibrillation.
OnAlternating current
Transformer meets
requirements of a
short-circuit-proof
safety-isolating power
transformer.
4General informationWelch Allyn Propaq Encore Vital Signs Monitor
For continued fire
protection, use only
the specified fuse.
Patient connections are
Type BF, and protected
against defibrillation.
Direct currentPatient connections are
Typ e B.
Battery charging when
green indicator
illuminated
The CE Mark and
Notified Body
Registration Number
signify the device has
met all essential
requirements of
European Medical
Device Directive
93/42/EEC.
IPX1
NRTL/C
Evaluated to CSA 601-1
and UL2601-1
Enclosure Protection Drip
proof: Classification IPX1
per IEC Publication 529
The Canadian
StandardsAssociation has
evaluated this device
according to CSA 601-1
and Underwriters
Laboratory Standard UL
2601-1. (This symbol is on
the Universal Power
Adapter.)
Input portTemperature sensor
Two way
Output portSingle-use only (not
communication port
NIBP cuff sizes:
Thigh
Large adult
Adult
Small adult
Child
Infant
For indoor use only (on
power adapter only)
Caution: On the
product, means
“Consult
accompanying
documentation.”
The CE Mark signifies
the device has met all
essential
requirements of
European Medical
Device Directive
93/42/EEC for a Class
1 product. (This
symbol is on the
Universal Power
Adapter.)
input
reusable).
Apply the NIBP cuff as
shown.
Non-ionizing
Fuse
electromagnetic
radiation
Pb
n%
Separate lead acid
battery from other
disposables for
recycling
Temperature limitsStacking limit (by number)Altitude limit
n
Humidity limitKeep away from rainFragile
Recycle the monitor and battery separately from other
disposables.
www.welchallyn.com/weee
n
This device has been tested and certified by the Canadian Standards Association International to comply
with applicable U.S. and Canadian medical safety standards.
US
C
Reference GuideGeneral information5
Propaq Encore documentation
The documentation set
The Propaq Encore documentation set consists of documents for the clinician, the
biomedical technician, and the department head or purchaser of accessories for the
Propaq Encore monitors.
This Propaq Encore Reference Guide contains important safety and operating information
for the clinician.
Propaq Encore Service Manual (810-0696-XX) contains information on how to properly
maintain the Propaq Encore through routine calibration, inspection, and maintenance.
Welch Allyn Products & Accessories (810-0409-XX) provides a comprehensive list of
accessories recommended for Propaq Encore monitors and options.
About this reference guide
This Reference Guide provides descriptions and operating information for the Propaq
Encore models 202EL, 204EL, and 206EL, including all available options at the time of this
manual's printing.
Statement of expectations of the reader
This Reference Guide was written for the clinician. Although this guide may describe
some monitoring techniques, Welch Allyn expects that you are a trained clinician who
knows how to take and interpret a patient’s vital signs. The Propaq Encore has been
designed as a quality monitor; however, inherent limitations require that good clinical
judgment always prevails.
6General informationWelch Allyn Propaq Encore Vital Signs Monitor
7
2
Getting started
Introducingthe Propaq Encore
Intended use
Before using thePropaq Encore on a patient, be sure you understand the Safety
Summary at the front of this book. It provides important information about safely using
the Propaq Encore. The Propaq Encore monitor is intended to be used by skilled clinicians
for multiparameter vital signs monitoring of neonatal, pediatric, and adult patients in
health care facility bedside applications; as well as for intra- and interfacility transport.
Propaq Encore monitors that do not include CO
light rain exposure over short periods of time (uniform distribution of approximately 1 mm
of water/ minute for 10 minutes or less).
Propaq Encore models and options
Three models of Propaq Encore monitors are available.
Features common to all modelsECG, 3-lead or 5-lead configurations, 0.05-40/0.5-40 Hz
NIBP, with neonatal, pediatric and adult modes
Temperature, 2 channels: YSI 400 and 700 series-compatible connectors
Defibrillator Synchronization
Real-time Analog output of ECG
Electrocautery noise suppression on all channels except Impedance
Pneumography
or printer options are able to withstand
2
206ELTwo Invasive Pressure Channels
204ELOne Invasive Pressure Channel
202ELNo Invasive Pressure
Options available for each modelPulse Oximetry (SpO
Capnography (CO
Mainstream Capnography (MCO
Sidestream Capnography (SCO
Dualstream Capnography (Both MCO
Impedance Pneumography (RESP) (available only with SpO
Printer
HP-compatible side panel
)
2
) (available only with SpO2):
2
)
2
)
2
and SCO2)
2
)
2
8Getting startedWelch Allyn Propaq Encore Vital Signs Monitor
Expansion module
The Propaq Encore Expansion Module attaches to the monitor and houses additional
capabilities. The Expansion Module can be fitted with the SpO
ALARM(S)
OFF Light
ALARM Light
Option
SpO
2
, CO2, and Printer options.
2
Mainstream
CO
Option
2
Printer Option
Sidestream
CO
Option
2
Propaq Encore pulse oximetry option (SpO2)
The Propaq Encore Pulse Oximetryoption (SpO2) is installed in the Expansion Module or
in a smaller unit that attaches to the rear of the monitor:
Masimo
Nellcor
®
Pulse Oximetry option
®
Pulse Oximetry option
Capnography (CO2) options
The Propaq Encore CO
option and sidestream CO
a ventilator. The sidestream CO
patients through a cannula. The CO
Dualstream CO
(SpO
) option.
2
in the Expansion Module. These options require the Pulse Oximetry
2
options allow carbon-dioxide monitoring. The mainstream CO
2
option allow CO2 monitoring directly in the breathing circuit of
2
option also allows CO2 monitoring of non-intubated
2
options can be installed separately, or together as
2
Expansion
Module
2
Impedance pneumography (RESP) option
The RESP option detects the rate or absence of respiratory effort, and is configured with
the Pulse Oximetry option.
Printer option
The Expansion Module withPrinter (EMP) provides a lightweight 3-channel recorder.
Reference GuideGetting started9
Propaq-to-Acuity® option
This option allows communication between the Propaq Encore and the Acuity Central
Monitoring System by means of an ethernet network system installed in your facility. The
Acuity System operator can view the patient data and control most of the bedside Propaq
functions. The Propaq Encore connects to the Acuity System through an Acuity network
cable that plugs into the Propaq right side panel.
Modem-Propaq option
This option allows telecommunication between a Propaq Encore and the Acuity System
by means of external modems. This option is configured with the Propaq-to-Acuity option.
For more information refer to Modem-Propaq Reference Guide.
HP-compatible side panel option
The HP connector-compatible option makes the Propaq Encore compatible with many
Hewlett-Packard sensors and accessories used with the Hewlett Packard Component
Monitoring System. This option replaces the standard Propaq Encore left side panel.
Using the Propaq Encore
System controls (right side panel)
On/Off Switch
Input Fuse
3A
2AG
Power Input
Connector
WARNING Safe interconnection between the Propaq Encore and other devices
must comply with applicable medical systems safety standards such as
IEC 60101 -1.
12-28V, 3A
DEFIB SYNCHROMONITOR
!
!
EKG x 1000
!
Defib Sync Connector
Contact Welch Allyn technical
support for information (for
specifications, see page 101)
Real-time ECG Output
Connector
Speaker
Battery Charging Light
!
Connector for Acuity or
Modem-Propaq
On/Off switch
This switch turns the monitor on and off. The switch is recessed to prevent accidentally
turning off the monitor, which would result in losing patient data.
10Getting startedWelch Allyn Propaq Encore Vital Signs Monitor
Input fuse
The input fuse, which protects the Propaq Encore against power surges, is a 3-Ampere
fuse, externally replaceable by qualified service personnel. See “Replacing the fuse” on
page 92 for fuse replacement instructions.
Power input connector
This receptacle accepts the Welch Allyn ac power adapter, which must be used for ac
mains operation and battery charging. The Propaq Encore is also designed to operate with
other 12-28 volt, dc-only power sources, such as a vehicle battery system.
Defib sync connector
This connector allows connection with a LIFEPAK 5 or LIFEPAK 6s defibrillator for
synchronized cardioversion. See “Real-time ECG analog/defib sync” on page 101.
Real-time ECG output connector
This connector provides a real-time analog ECG signal output.
Battery charginglight
This green light turns on when a power source (ac power adapter or external dc source) is
connected and the battery is charging. Although the monitor may be turned off, battery
charging continues when an external power source is connected.
Connector for Acuity or Modem-Propaq
This connector allows either direct connection to an Acuity System, or connection to an
external modem for telecommunication to an Acuity System. For more information about
the Acuity System, see “Acuity Central Monitoring system” on page 85. For more
information about the Modem-Propaq, refer to Modem-Propaq Reference Guide.
Alarm lights
Alarms and limits are described in detail beginning on “Alarms and limits” on page 69.
ALARM light
When an alarm limit is violated, the red ALARM light turns on.
ALARM(S) OFF light
When any alarm limit is turned off, the yellow ALARM(S) OFF light turns on.
Reference GuideGetting started11
Power-up screen
WELCH ALLYN
MODEL PROPAQ 204
DIAGNOSTICS IN PROGRESS
BATTERY: 8.3 VOLTS
(c) WELCH ALLYN PROTOCOL INC. 1988-2007
A few seconds later, the top two lines of the screen are
replaced with text indicating the current patient mode
(adult, pediatric, or neonatal).
SOFTWARE VERSION 2.50.00
When you first turn on the monitor, the power-up
screen displays information about the Propaq Encore
and the monitor runs diagnostic tests to ensure proper
functioning.
PEDIATRIC MODE
DIAGNOSTICS IN PROGRESS
BATTERY: 8.3 VOLTS
(c) WELCH ALLYN PROTOCOL INC. 1988-2007
SOFTWARE VERSION 2.50.00
WARNING Before you use a Propaq on a new patient, always turn off the Propaq
for a few seconds, then turn it on again. This clears the prior patient’s trend
values, alarm limit settings, and NIBP cuff inflation target.
1.If the Propaq Encore has been used for a previous patient, switch the monitor off,
then on again. The monitor will turn on in the powerup patient mode with the
associated settings.
Note
Verify that the powerup tone is produced. If the monitor has SpO
tones are produced to make sure that both speakers are working.
2. Verify the monitor is in the correct patient mode according to the patient’s age. If the
patient mode is not correct, change it. (See “Monitor setup” on page 20 to change
the patient mode.)
WARNING Always check the patient mode when monitoring a new patient. The
patient mode determines default alarm limits, maximum cuff inflation pressure,
and internal algorithm settings.
3. Verify the battery voltage is sufficient for monitoring. If it is less than 7.4 V, connect to
a power adapter (see “Power adapter intended use” on page 89 for information about
the power adapter).
Power-up equipment alert: program fault, settings lost
If a PROGRAM FAULT: SETTINGS LOST, TIME/DAY RESET equipment alert appears when
you turn on the monitor, the monitor cannot recall the programmed custom settings and
current time and date. This can occur if the battery is drained or after new software has
been installed.
If this occurs, the monitor provides a special sequence of display windows to help you
regain use of your monitor as quickly as possible. Do the following:
, verify two
2
1.Connect an ac power adapter to recharge the battery (if the battery is drained).
12Getting startedWelch Allyn Propaq Encore Vital Signs Monitor
2. Press any button below the equipment alert screen to acknowledge the alert. The
monitor will display the Mode Setup window (shown on page 23).
3. Press these buttons to select one of the Factory patient modes for use:
Factory Adult modePOWERUP*, YES.
Factory Pediatric modeNEXT, POWERUP*, YES.
Factory Neonatal modeNEXT, NEXT, POWERUP*, YES.
After you press YES, the monitor will display the Time/Day window.
4. Press NEXT, UP, and DOWN as needed to set the time and date. Then press ENTER
to store the new time and date.
Note
These display screens are only displayed in this order if the PROGRAM FAULT
equipment alert occurs.
5. Turn off the monitor, then turn it on again so the settings will take effect.
The monitor is ready for use. If you want to store some customized patient mode program
settings, refer to page 23.
If you follow these steps and the equipment alert reappears at powerup, the monitor may
need to be serviced and the battery replaced. Contact a qualified service person.
Reference GuideGetting started13
Patient connections
The left side panels differ depending on the Propaq Encore model. All models have ECG,
NIBP, and two temperature connectors. The Propaq Encore 204 left side panel includes
one invasive-pressure connector, and the Propaq Encore 206 includes two invasivepressure connectors.
ECG / EKG RESPINV. BP
NIBP
PSNI
ECG / EKG RESP
!
P1
NIBP
T1
PSNI
!
NIBP
PSNI
T1
ECG / EKG RESPINV. BP
P1
T1
!
!
COMPATIBLE
HP
202 HP
ECG / EKG RESP
P
S
N
I
NIBP
ECG / EKG RESP
P
S
N
I
NIBP
T2
INV. BP
P2
Propaq Encore 206 EL
INV. BP
COMPATIBLE
INV. BP
P1
!
HP
P2
206 HP
T2
Propaq Encore 204 ELPropaq Encore 202 EL
INV. BP
!
COMPATIBLE
HP
P1
204 HP
T2
ECG / EKG RESP
P
S
N
I
NIBP
On Propaqs with the Hewlett-Packard connector option, all models have only one
temperature connector, the YSI 400 connector.
14Getting startedWelch Allyn Propaq Encore Vital Signs Monitor
Option connectors
CO
2
CO
2
!
!
Mainstream CO2
Connector
Sidestream CO2
Connector
Cable: see Products and
Accessories (810-0409-XX)
Masimo SpO2 Connector
(motion tolerant)
Nellcor SpO2 Connector
Reference GuideGetting started15
Propaq Encore display
The display shows waveforms, vital sign numeric values, Propaq Encore status, and alarm
information in different windows. Different vital sign numeric values (such as heart rate
and blood pressures) have upper and lower range limits. If the Propaq Encore detects a
vital sign value outside of the Propaq's measurable range, the monitor displays – – –
(below the range) or + + + (above the range) instead of the vital sign value.
WARNING The Propaq Encore will show + + + for HR numerics between 301350 beats per minute. Above 350 beats per minute, it may display incorrectly low
heart rates, due to intermittent picking of R-waves.
Note
Due to differences in software versions and standards required by different
countries, the displays shown in this reference guide may be slightly different
than the display on your Propaq Encore.
Time of day, caution and status
messages.
Waveform window
Status window
ETCO2 concentration
Breath Rate from CO
Oxygen saturation is a
percentage value.
2
Temperature values are displayed
here.
Noninvasive blood
pressure values can
appear here.
18:45:28
III
mV
1
cm
MON
2
CO
GAS COMPENSATION: OFF
RESPONSE : NORMAL
2
SOURCE : MAINSTREAM
CO
SWEEP SPEED : 6.25 mm/s
BAROMETER : 762.2 mmHg
GAS COMPPREV MENUSOURCERESPONSE
NIBP
122
S
Systolic, diastolic, and (if space permits)
Heart Rate Source:
HR indicates ECG;
PR indicates blood pressure
102.3
1
mean pressures are displayed.
or SpO
F
Bells indicate alarm limit status
Heart Rate in beats per
minute.
.
2
HRT
60
ART
122
58
( 85)
PA
35/ 18
35
mmH
SpO
BR
12
g
2
92
MCO2
The screenspace is reallocated when vital signs are added or removed. By changing the
size of the numeric windows below the heart rate, the Propaq Encore provides the best
possible view of all numerics for vitals signs being monitored.
16Getting startedWelch Allyn Propaq Encore Vital Signs Monitor
You can select up to three waveforms to be shown on the Propaq. When only one
waveform is selected, a trend window automatically appears below the waveform. While
changing Propaq Encore settings, a status window may appear below the waveform.
12:45:28
III
mV
1
cm
MON
TIMEINCO2ETCO2HR/PR
10:00
9:58
9:56
9:54
PRINTOXYCRGNXT TRND
BPMHH:MM
101
88
93
91
NIBP
RESP TREND
SpO2
%
98
SRCH
OFF
100
Propaq Encore buttons
The four buttons at the sides of the screen are reserved for the most commonly used
functions.
ALARMSSilences or resumes alarm tone.
START/STOPStarts and stops NIBP measurements. The STOP function will
NET OFFDisconnects the monitor from the Acuity network (if
FREEZE/UNFREEZE Freezes or “unfreezes” the waveforms. If only one or two
102.3
122
S
1
F
HRT
80
ART
122
58
( 85)
PA
BR
Br/M
11
11
5
10
mmHgmmH
30
31
30
30
3
3
10
2
g
MCO
35
mmH
SpO
35/ 18
2
g
2
BR
12
92
automatically vent the cuff.
connected).
waveforms are displayed and you press FREEZE, the frozen
waveform(s) are shown along with an active waveform so you
can continue to monitor the patient’s condition.
Patient waveform and trend information
can be simultaneously displayed, while
numeric values are continuously updated.
MAIN MENUPressing MAIN MENU always returns the monitor to the top
level menu.
NET OFF
ALARMS
START/STOP
(NIBP)
FREEZE/
UNFREEZE
MAIN MENU
The five labels
above the
associated
buttons
The five buttons below the screen, and their associated labels located on the screen,
provide access to the menus.
Later in this manual, the notation A , B , C is used as a shorter way to say “Press Button
A, then B, then C.”
Reference GuideGetting started17
Propaq Encore menus
Menus for some patient vital signs are displayed only if that option is included in your
Propaq.
MAIN MENU
SETUPNIBPECG/RESPINV PRSSpO2/CO2
SpO2 MENU 1
SpO2 MENU 2
INV PRS MENU 1
RANGE
INV PRS MENU 2
LABEL P1FORMAT 1LABEL P2FORMAT 2
ECG/RESP MENU 1
ECG SIZE
ECG/RESP MENU 2
NEXTCHANGE
NIBP MENU
STARTAUTO/MA NTURBOCUF
STOP
INTERVAL
SpO2CO2
CO2 MENU 1
CO2 MENU 2A (NO CO2 SENSORS)
GAS COMPRESPONSE
CO2 MENU 2B (MAINSTREAM CO2 ACTIVE)
GAS COMPRESPONSESOURCE
CO2 MENU 2C (SIDESTREAM CO2 ACTIVE)
GAS COMPRESPONSESOURCE
SIZEMORE
RESPONSEC-LOCK
RESCALE
ECG LEADRESP SZE
ZERO P1
CANCEL
PREV MENU
ZERO P2
CANCEL
MORE
PREV MENU
MORE
PREV MENU
PREV MENU
FLOWRATE
PREV MENU
PREV MENU
PREV MENURANGEmm/sMORE
PREV MENU
PREV MENU
PREV MENU
18Getting startedWelch Allyn Propaq Encore Vital Signs Monitor
Key-press route to setup menu 1
SETUPNIBPECG/RESPINV PRSSpO2/CO2
SETUP MENU 1
STATSCALE
ALARMS
ALARMS MENU
STAT SET
TRENDSWAVE SEL
4 SUSPND
RESUME
MORE
(MORE button takes you to
next Setup Menu)
TRENDS MENU
PRINTNXT TRND
LIMITS
ALARM LIMITS MENU*
NEXT
PREV MENU
UPDOWNON/OFF
(*ON/OFF button is not displayed for HR/PR alarm limits
if the HR/PR ALARM LIMITS setting is set to CANNOT
OXYCRG
NEXT PAGE
TURN OFF.)
Reference GuideGetting started19
Key-press route to setup menu 2
SETUPNIBPECG/RESPINV PRSSpO2/CO2
SCATSCALEALARMSTRENDSWAVE SEL
SETUP MENU 2
PATIENT MODE MENU
(Patient Mode menu is
accessed when CHANGE
MODE SETUP MENU
is pressed for PATIENT
MODE.)
NEXTUPDOWNENTER
SERVICE MENU 1*
SERVICE MENU 2*
TEMP TEST
SERVICE MENU 3*
PRINTER SETUP MENUS
(Printer Trend Select status window)
NEXTCHANGEPR TRENDMORE
WAVE SELECT MENU
MORE
MORENEXTCHANGEWAVE SELPRINTER
ON/OFFNEXT
*(Service menu tests are for use by authorized
service personnel only, and are available only
when in the Adult patient mode.)
TIME/DAY MENU
PREV MENU
PIXL TST
NET TEST
SETTINGS MENU
(Printer Setup status
PREV MENUNEXTCHANGEPR TRE NDMORE
PREV MENU
INSERV
NO INSRV
PREV MENUNEONATALPEDADULTSETUP
PREV MENUNEXTPOWERUP*USE NOWSAVE
PREV MENU
SERVICE
MORENIBP TESTIBP TESTSpO2 TSTCO2 TEST
MORE
MORESETTINGSKEY TESTSYSTEM
PREV MENUNEXTCHANGEALL ALRM
20Getting startedWelch Allyn Propaq Encore Vital Signs Monitor
Monitor setup
Setup Menu 1 is accessed by pressing the SETUP button on the Main Menu.
STATSCALEMOREWAVE SEL TRENDSALARMS
STATSCALEAutomatically readjusts all waveform scales.
ALARMSAllows access to the Alarms menu.
WAVE SEL Allows you to turn on and off desired waveforms or NIBP numerics
for display.
TRENDSAllows access to the Trend settings and display.
MOREDisplays the next setup menu and the following status window:
SETUPHR/PRHR/PR RR/BR
CURRENT SOURCE : ECG MCO2
SELECTED SOURCE : ART --SWEEP (mm/s) : 12.5 3.13
ALARM TONE : HIGH
HR/PR TONE : LOW
PATIENT MODE : ADULT
NEXTMOREWAVE SEL PRINTERCHANGE
NEXTSelects the next setting in the status window.
CHANGEChanges the currently selected display setting. (Pressing CHANGE
at PATIENT MODE allows you to choose between Adult, Pediatric,
and Neonatal in a Patient Mode window.)
PRINTERAllows access to the Printer Menu.
MOREAllows access to the Time/Day window.
CURRENT
SOURCE
SELECTED
SOURCE
SWEEP (mm/s) The selectable sweep speeds for HR/PR are 12.5, 25, and 50 mm/
ALARM TONESets the Alarm Tone volume to HIGH, MEDIUM, or LOW.
When the selected HR/PR source is no longer available, the current
source is the active source with highest priority. The RR/BR source
cannot be manually selected. It will always be CO
Otherwise, it will be ECG/RESP.
The user-selected HR/PR source is displayed along with the HR/PR
source currently being used by the monitor.
sec. The sweep speeds for RR/BR are 3.13, 6.25, and 12.5 mm/sec.
( 85)
PA
MCO
2BR
35
mmH
SpO
35/ 18
12
g
2
92
if CO2 is active.
2
Reference GuideGetting started21
HR/PR TONESets the Heart Tone volume to HIGH, MEDIUM, LOW, or OFF.
PATIENT MODE Pressing CHANGE in this selection displays the following Patient
Mode window:
Note
PATIENT MODE
SELECT PATIENT MODE BASED ON AGE:
NEO : < 44 WEEKS GEST. AGE
PED : > 44 WEEKS GEST. AGE,
< 9 YEARS
ADULT: > 9 YEARS
NEONATALPREV MENUADULTPED
BATTERY: 9.2 VOLTS
SETUP
( 85)
PA
2BR
MCO
35
mmH
SpO
35/ 18
g
2
12
92
If you press NEONATAL, PED, or ADULT, a confirmation window
appears, requiring you to confirm your selection:
PATIENT MODE
CHANGING SETTINGS AND MODE
TO:
CUSTOM: ADULT
ADULT ALARM LIMITS WILL BE SET
ARE YOU SURE?
YESNO
( 85)
PA
MCO
2BR
35
mmH
SpO
35/ 18
g
2
12
92
Whenever you change the patient mode, the alarm limit settings are
automatically changed to the defaults for that mode. If Custom
settings have been set for that mode, the defaults are the Custom
mode settings. If no Custom settings have been set, the defaults are
the Factory Mode settings. See page 23 for more information about
patient modes.
If you change the patient mode, the CO
mode might vary slightly from the originally-programmed CO
for the new mode. Check the CO
2
alarm limits in the new
2
alarm limits.
limits
2
If you press SETUP in the previous Patient Mode window, the Mode
Setup window appears. This allows you to set custom patient
modes and powerup defaults as described on page 23.
22Getting startedWelch Allyn Propaq Encore Vital Signs Monitor
Selecting waveforms for display
To select waveforms for display, press SETUP, WAVE SEL . Use the NEXT and ON/OFF
buttons to turn on the desired waveforms in the wave select window:
SETUP
ECG : ON RESP : OFF
ART : ON SpO2 : ON
PA : OFF NIBP : ON
CO
2: OFF
NEXTPREV MENUINSERVON/OFF
Display priorities
You can turn on more than three waveforms, but only the first three waveforms listed in
the wave select window that are monitored are displayed. The patient parameters being
monitored are listed in the order they will be displayed if all are turned on.
Because of the critical nature of the ECG waveform, you cannot turn off ECG. However, if
ECG is not monitored, another waveform will occupy its place.
The displayed waveforms are also the ones printed if a printer is attached.
Setting the time and date
To set the time and date, from the Main Menu press SETUP, MORE, MORE. The monitor
displays the Time/Day window:
TIME/DAY
TIME DAY
H:MIN:S MO/DA/YR
07:45:32 06/12/97
NEXTPREV MENUDOWNENTERUP
WAVE SELECT
BATTERY: 9.2 VOLTS
( 85)
PA
2BR
MCO
35
mmH
SpO
( 85)
PA
MCO
35
mmH
SpO
35/ 18
g
2
35/ 18
2
g
2
12
92
BR
12
92
Press NEXT, UP, and DOWN as needed to set the time and date. Then press ENTER to
store the new time and date.
Time/day settings and trends
WARNING Changing the hour/minute/second setting for the monitor in the
Time/Day window can cause the monitor to erase previously stored patient trend
data.
When you change the hour/minute/second setting for the monitor in the Time/Day
window, the monitor deletes any patient trend data that is older than five hours for nonNIBP trends or older than eight hours for NIBP trends according to the new clock setting.
However, if the monitor has not yet stored the full capacity of trends and you change the
hour/minute/second setting to a time that is within the stored trend period, previously
stored trends are not erased.
Changing the day, month, or year setting does not affect the stored patient trends.
Reference GuideGetting started23
Changing the date format, filter, and units
To change the date format, ECG filter, or some measurement units, first make sure you
are in the Adult patient mode. Then press SETUP, MORE, MORE, SERVICE, YES (to
access the Service Menu), MORE, MORE, SETTINGS. The monitor displays the Settings
window:
SETTINGS
DATE : MO/DA/YR
FILTER : 60 Hz
TEMP F/C :
DECIMAL : .
HR/PR ALARM LIMITS: CAN TURN OFF
2UNITS: mmH
CO
NEXTPREV MENUCHANGE
BATTERY: 9.2 VOLTS
C
g
( 85)
PA
2BR
MCO
35
mmH
SpO
35/ 18
12
g
2
92
NEXTSelects the next setting in the status window.
CHANGEChanges the currently selected display setting.
DATESets the date format: Month/Day/Year, Day.Month.Year, or Year/Month/
Day.
FILTERSets the ECG filter frequency. Make sure it is set to your ac mains
frequency.
TEMP F/CSets the temperature display units: either degrees Fahrenheit or
Celsius. If you change the units, the TEMP trends will not be cleared.
DECIMALSets the decimal character as either a period (.) or a comma (,).
HR/PR
ALARM
LIMITS
CO
UNITS Sets the CO2 display units as mmHg, kPa, or percent (%). If you change
2
Allows or prohibits turning off the HR/PR alarm limits. If CANNOT TURN
OFF is selected, the ON/OFF button is not displayed on the HR/PR
Alarm Limits Menu.
the units, the CO2 trends will be cleared and CO2 alarm limit settings
change to the factory default settings for the currently-used patient
mode.
Note
Any time you change the Date, Filter, Temp F/C, Decimal, HR/PR Alarm Limits
(CAN or CANNOT TURN OFF) or CO
Units setting, the new setting also
2
becomes the powerup default setting.
Setting the current, custom, and power-up modes
The Propaq Encore has two sets of patient mode settings:
•Factory patient modes. The powerup settings and alarm limits for these patient
modes are preset and cannot be changed. They are listed in “Factory default
settings” on page 118.
•Custom patient modes. You can customize the power-up settings and alarm limits
for these patient modes. (See SAVE on page 24.)
WARNING If any alarms are set to OFF and you select SAVE to store the
settings for that CUSTOM patient mode, those alarms will be OFF whenever the
Propaq powers up in that CUSTOM patient mode or when that CUSTOM patient
mode is selected. Consider carefully before setting CUSTOM patient mode
powerup alarms to OFF.
24Getting startedWelch Allyn Propaq Encore Vital Signs Monitor
Note
The alarm for apnea cannot be turned off at any time.
You can program the Propaq Encore to power up in any of the Factory patient modes or
the Custom patient modes. You can also change the current patient mode during
operation.
Note
Whenever you change the patient mode, the alarm limit settings automatically
change to the settings for that mode.
From the Main Menu, press SETUP, MORE, CHANGE, SETUP. The Mode Setup window
appears:
MODE SETUP
FACTORY:
CUSTOM
:
* = PATIENT MODE ON POWERUP
NEXTPREV MENU
POWERUP*
BATTERY: 9.3 VOLTS
*ADULTPED
ADULTPED
USE NOW
NEO
NEO
SAVE
2
P
MCO
38
mmH
Sp02
ZEROED
NO ZERO
2BR
12
g
97
The asterisk (*) indicates which patient mode is currently selected for powerup.
NEXTSelects the next setting in the status window.
POWERUP* Selects the highlighted patient mode (and its associated settings) as
the powerup mode. The selected powerup mode is marked by an
asterisk (*). (This does not change the current patient mode.)
USE NOWSelects the highlighted patient mode (and its associated settings) as
the current patient mode. (This does not affect the powerup mode.)
SAVEUse this button to reprogram the settings of Custom patient mode:
1.Make sure the patient mode you want to reprogram (ADULT, PED
or NEO) is currently used. (To change patient modes, highlight the
desired mode and press USE NOW, YES.)
2. Exit the Mode Setup window, then use other menus and buttons to
set the monitor settings and alarm limits as desired.
3. Re-enter the Mode Setup window, highlight the desired Custom
mode, and press SAVE, YES.
Reference GuideGetting started25
Printer functions
Press SETUP, MORE, PRINTER to display the printer menu and setup window.
18:45:28
III
mV
1
cm
MON
PRINTERSETUP PAGE
CONTINUOUS : 25.0 mm/s
AUTO PRINT : 15 minutes
ALARM PRINT : OFF
NIBP TICKET
APNEA TICKET : OFF
OXYCRG ON ALARM: OFF
NEXTMOREPREV MENUPR TRENDCHANGE
NIBP
S
: OFF
122
1
102.3
F
HRT
80
ART
122
58
( 85)
PA
35/ 18
2
35
mmH
SpO
BR
12
g
2
92
MCO
NEXTSelects the next setting in the status window.
CHANGEChanges the currently selected display setting.
PR TRENDPrints all trends turned on in the Printer Trend Select Window.
MOREPressing the MORE button displays another menu and status
window.
PREV MENUReturns you to the previous menu.
CONTINUOUSSets the print speed for real time (continuous) measurements to
6.25, 12.5, or 25 mm/sec. This sets the print speed for a printout
obtained by pressing the START/STOP button on the printer.
AUTO PRINTAutomatically prints 8 seconds of patient information every 15
minutes, 30 minutes, 1 hour, 2 hours, or 4 hours. This is the
latest patient information (real time). The print speed is
automatically set to 25 mm/sec.
ALARM PRINTAutomatically prints upon an alarm. The Propaq Encore prints 20
seconds of patient information. The first 12 seconds contain
information prior to the alarm. The print speed is automatically
set to 25 mm/sec.
NIBP TICKETAutomatically prints an NIBP Ticket when the measurement is
taken.
APNEA TICKETWhen turned on, an Apnea Ticket is printed at the conclusion of
an apnea alarm and at the one-minute clock interval if the apnea
alarm does not cease.
OXYCRG ON
ALARM
When turned on, an oxycardiorespirogram will print if an HR/PR,
SpO
, or RR/BR alarm occurs. For more information on OxyCRG,
2
see “OxyCRG” on page 82.
26Getting startedWelch Allyn Propaq Encore Vital Signs Monitor
PRINTER FAULT messages
These PRINTER FAULT messages can appear in an equipment alert window.
LOW BATTERY,
PRINTER DISABLED
This message appears when the Propaq’s battery voltage is
less than 7.6 volts. To continue operation, plug the ac power
adapter into the Propaq.
CHECK DOORThe door on the bottom of the printer is open. Close door to
remove this message.
PAPER OUTTo add printer paper, see “Printer maintenance” on page 96.
OVERHEATINGThe printer is overheating. Service may be required.
The front panel of the printer lets you control the basic printer functions.
Manually starts and stops a printout of patient
information as it is monitored (continuous or real time).
Hold down top and bottom
keys simultaneously to
generate a test strip.
START
STOP
SNAP
SHOT
PRINT
TRENDS
Hold down top two keys simultaneously to generate a
paper feed.
Prints the last 8 seconds of data for nonrespiration
waveforms and 32 seconds of compressed waveform
history for respiration waveforms.
Prints all trends that are enabled in the Printer Trend
Select Window.
If you press FREEZE prior to pressing SNAPSHOT, the printer prints the 8 seconds of
patient information obtained prior to when you pressed FREEZE.
Learning the Propaq Encore
Using in-service mode
You can practice using the Propaq Encore without a patient simulator by using the
Propaq's in-service mode of operation. The in-service mode cannot be activated while you
are monitoring a patient. The message “SIMULATING” alternates with the time of day
and patient mode on the display.
To begin practicing with your Propaq, disconnect all patient cables connected to the
monitor. Leave the cuff connected so you can take NIBP measurements. If you have been
monitoring a patient, turn off the Propaq Encore and turn it back on. From the Main Menu,
press SETUP, WAVE S E L , INSERV.
The Propaq Encore has two sets of simulated patient information—an initial set and an
alternate set. To change between them, press the INSERV button again.
If you connect a patient cable or set the NIBP channel to automatically take pressure
measurements, the Propaq Encore stops simulating, goes through its powerup tests, and
erases any simulated trend data it might have stored.
Loading...
+ 112 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.