Welch Allyn 1500 User Manual

Page 1
Welch Allyn
®
1500
Patient Monitor
Directions for use
Software version 1.3.X
Page 2
ii Welch Allyn 1500 Patient Monitor
© 2011 Welch Allyn. All rights are reserved. To support the intended use of the product described in this publication, the purchaser of the product is permitted to copy this publication, for internal distribution only, from the media provided by Welch Allyn. No other use, reproduction, or distribution of this publication, or any part of it, is permitted without written permission from Welch Allyn.
Welch Allyn, Acuity, Smartcuf, FlexNet, and Flexible Monitoring are registered trademarks of Welch Allyn.
Masimo, SET, and Signal Extraction Technology are registered trademarks of Masimo Corporation
NO IMPLIED LICENSE: Possession or purchase of this device does not convey any express or implied license to use the device with unauthorized sensors or cables which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device.
Nellcor and Oximax are trademarks of Nellcor Puritan Bennett.
NOTICE: Purchase of this instrument confers no express or implied license under any Nellcor Puritan Bennett patent to use this instrument with any oximetry sensor that is not manufactured or licensed by Nellcor Puritan Bennett.
Microstream is a registered trademark of Oridion.
NO IMPLIED LICENSE: Possession or purchase of this device does not convey any express or implied license to use the device with unauthorized consumable CO scope of one or more of the patents relating to this device and/or CO2 sampling consumable products
Software in this product is copyrighted by Welch Allyn or its vendors. All rights are reserved. The software is protected by United States of America copyright laws and international treaty provisions applicable worldwide. Under such laws, the licensee is entitled to use the copy of the software incorporated with this instrument as intended in the operation of the product in which it is embedded. The software may not be copied, decompiled, reverse-engineered, disassembled or otherwise reduced to human-perceivable form. This is not a sale of the software or any copy of the software; all right, title and ownership of the software remain with Welch Allyn or its vendors.
For information about any Welch Allyn product, call the nearest Welch Allyn representative.
:
USA + 1 315 685 4560
Canada 800 561 8797 China + 86 216 327 9631
European Call Center
Germany + 49 7477 92 71 86 Japan + 81 3 3219 0071
Latin America + 1 305 669 9003 Netherlands + 31 15 750 5000
Singapore + 65 6419 8100 South Africa + 27 11 777 7555
United Kingdom + 44 20 7365 6780 Sweden + 46 8 58 53 65 51
sampling consumable products which would, alone, or in combination with this device, fall within the
2
800 535 6663
Australia + 61 2 9638 3000
800 074 793
+ 35 3 46 906 7790 France + 33 1 60 09 33 66
REF 104592 (CD)
Manual 2.510865 Ver. c, 2011-09
SCHILLER AG Altgasse 68 CH-6341 Baar Switzerland
www.welchallyn.com
SCHILLER AG Altgasse 68 CH-6341 Baar Switzerland
Page 3
Directions for use Contents iii

Contents

1 - Safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
User responsibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Intended use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Organizational measures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Safety equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Operation with other devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Networks and internet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Additional terms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2 - Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Standard features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
The Welch Allyn
Function buttons. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Setup menu overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Display overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
®
1500 Patient Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3 - Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Startup and initial preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Inserting recorder paper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Switching the monitor on or off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Initial settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Saving the user-defined settings as default . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Standby mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Defining parameter fields . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Defining display waveforms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Trend data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Settings via a parameter field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4 - Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Display of alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Silencing an alarm. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Suspend all alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Turning off an individual parameter alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Alarm Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Alarm limit setting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Physiological alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Page 4
iv Contents Welch Allyn 1500 patient monitor
5 - Monitoring and Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
ECG. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Arrhythmia . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
ST measurement (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Respiration rate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Capnography. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
NIBP monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
SpO
monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
2
IBP monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Temperature monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Cardiac output (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
6 - Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Maintenance interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Button check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Speaker check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
LED check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Alarm check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Battery maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Inspecting and cleaning the monitor and accessories . . . . . . . . . . . . . . . . . . . . 92
Inspection and checklist report. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Replacing the fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
EMC compliance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Mounting on a wall or stand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Recycling monitor components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
7 - Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Setup menu overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Parameter settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
8 - Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Miscellaneous. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
Nellcor SpO Masimo SpO
ECG. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
IBP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
NIBP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Cardiac output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
2
A - Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
System data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Safety standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
Measured values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
Page 5
1
©
1

Safety

User responsibility

The numerical and graphical results and any interpretation given must be examined with respect to the overall clinical condition of the patient and the general recorded data quality.
The indications given by this equipment are not a substitute for regular checking of vital functions.
This monitor is only to be used by those trained in its operation or repair.
Ensure that the personnel have read and understood these operating instructions and in particular this “Safety” section.
Damaged or missing components must be replaced immediately.
It is the owner's responsibility that the valid regulations for safety and prevention of accidents are observed.

Intended use

The Welch Allyn® 1500 Patient Monitor patient monitoring unit is designed for the monitoring of vital parameters such as ECG, SpO pressure (NIBP), invasive blood pressure (IBP), temperature and respiration of a patient. Cardiac output and hemodynamic calculations are also possible.
The device is intended to be used by qualified doctors or trained medical personnel.
The device is not suitable for transport.
There is no danger for patients with pacemaker.
The device is intended for the monitoring of one patient at a time.
The device is not designed for sterile use nor is it designed for outdoor use.
Do not use this monitor in areas where there is any danger of explosion or in the presence of flammable gases.
The device is classified CF. It is defibrillation protected when the original accessories are used. However, as a safety precaution when possible, remove the electrodes before defibrillation.
This product is not designed for direct cardiac application.
The arrhythmia module is not intended for use with neonatal patients.
The ST-analysis module is not intended for use with neonatal patients.
, etCO2, non invasive blood
2
Page 6
2 Safety Welch Allyn 1500 Patient Monitor

Organizational measures

Before using the monitor, ensure that an introduction regarding the monitor functions and the safety precautions have been provided by a medical product representative.
Observe the operating instructions and maintenance instructions.
These operating instructions do not override any statutory or local regulations, or procedures for the prevention of accidents and environmental protection.

Safety

WARNING Mount the monitor securely so that there is no possibility of it falling
on the patient or on the floor.
WARNING If uncertain about the accuracy of any measurement, first check the patient’s vital signs by alternate means, and then make sure the monitor is functioning correctly.
WARNING Do not touch the monitor during defibrillation.
WARNING To ensure patient safety, none of the ECG electrodes including the
neutral electrode, nor the patient or any person with simultaneous patient contact, must come in contact with conductive parts, even when these are grounded.
WARNING If you notice any changes that impair safety (including operating behavior) remove the monitor from service and report it to the person responsible for servicing the monitor.
WARNING Do not place any liquids on the monitor. If liquid is spilled over the monitor, immediately disconnect the monitor from the mains and dry. The monitor must be serviced before reusing.
Caution This manual, and especially these safety notes, must be read and observed.
Caution Electrical installation of the room or the building in which the monitor is to be used must comply with regulations specified by the country in which the equipment is to be used
Caution Ensure the monitor is always mounted on a Welch Allyn approved bracket or stand. The monitor is unstable when the unit is not secured to an approved Welch Allyn mounting system.
Page 7
Directions for use Safety 3

Safety equipment

Operating the monitor without the correctly rated fuse, or with defective cables, constitutes a danger to patient safety. Therefore:
Caution Do not operate the monitor if the ground connection is suspect or if the mains lead is damaged or suspected of being damaged.
Caution Damaged cables and connections must immediately be replaced.
Caution Electrical safety devices, such as fuses, must not be modified.
Caution Fuses must only be replaced with the same type and rating as the
original.

Alarms

WARNING Do not silence the audible alarm if patient safety could be
compromised.
WARNING Always respond immediately to an equipment alert because the patient may not be monitored during certain alarm conditions.
WARNING Before each use, verify that the alarm limits are appropriate for the patient being monitored.
WARNING Check the audible alarm silence duration before temporarily silencing the audible alarms.
Page 8
4 Safety Welch Allyn 1500 Patient Monitor

Operation with other devices

Caution Do not use the monitor in or near an MRI suite.
Only use accessories and other parts recommended or supplied by Welch Allyn. Use of other than recommended or supplied parts may result in injury, inaccurate information or damage to the monitor.
Accessory equipment connected to the analogue and digital interfaces must be certified according to the respective IEC standards (e.g. IEC/EN 60950 for data processing equipment and IEC/EN 60601-1 for medical equipment). Furthermore all configurations shall comply with the current version of the system standard IEC/EN 60601-1-1. Anyone who connects additional equipment to the signal input part or signal output part configures a medical system, and is therefore responsible that the system complies with the requirements of the valid version of the system standard IEC/EN 60601-1-1. If in doubt, consult the technical service department or your local representative.
Any other equipment used with the patient must use the same common ground as the monitor.
Precautions must be observed when using high frequency devices. Operating high frequency electro-surgical equipment in the vicinity of the monitor can produce interference in the monitor and cause incorrect measurements. Only use patient cables recommended by Welch Allyn to avoid possible signal interference during ECG acquisition.
There is no danger when using the ECG monitor simultaneously with electrical stimulation equipment. However, during defibrillation, keep discharge paddles away from the monitor ECG lead wires, electrodes, any other monitor sensors, and other conductive parts in contact with the patient.
If the patient cable should become defective after defibrillation, a lead-off indication is displayed and an audible alarm is issued.
Portable communication equipment, HF two-way radios and devices marked with the
symbol can affect this monitor (see “EMC compliance” on page 98).

Networks and internet

When the monitor is part of a network, (LAN, HIS, etc.), transmitting over a telephone network or any other transmission /reception medium, or if exposed to the Internet or other networks that are not secure, appropriate security measures must be provided to protect the patient information stored.
Patient security and security of the network is the sole responsibility of the user.
Page 9
Directions for use Safety 5

Maintenance

WARNING Danger of electric shock. Do not open the monitor case. There are no
user serviceable parts inside. Servicing may only be performed by a qualified technician authorized by Welch Allyn.
WARNING Before cleaning and to isolate the mains power supply, switch the monitor off and disconnect it from the mains by removing the plug.
Caution Do not use high temperature sterilization processes (such as autoclaving). Do not use E-beam or gamma radiation sterilization.
Caution Do not use solvent or abrasive cleaners on either the monitor or cable assemblies.
Caution Do not immerse the monitor or cable assemblies in liquid.

Symbols

These symbols appear in this user guide.
WARNING Warning statements in this user guide identify conditions or practices that could result in personal injury.
Caution Caution statements in this user guide identify conditions or practices that could result in damage to the equipment or other property.
Page 10
6 Safety Welch Allyn 1500 Patient Monitor
©
The following symbols appear on the monitor, or accessories.
Symbol Definition Symbol Definition
Potential equalization (earth ground) CF symbol. This monitor is classified safe for
The monitor can be recycled. Recycle the monitor and battery
Notified body of the CE certification (TÜV P.S.).
Altitude limits Humidity limits
Keep away from sunlight Stacking limit
Temperature limits Keep away from rain
This way up Fragile
internal and external use. However, it is only defibrillation protected when used with the original Welch Allyn patient cable!
separately from other waste. Refer to
www.welchallyn.com/weee for collection
point and additional information.
Note accompanying documents.
CO
in CO2 out
2
Temperature NIBP
The following symbols appear on the screen.
Symbol Definition Symbol Definition
Alarm off Patient mode symbols; neonate, pediatric,
adult
Page 11
Directions for use Safety 7
Note
Note

Additional terms

Implied authorization
Possession or purchase of this monitor does not convey any express or implied license to use the monitor with replacement parts which would alone, or in combination with this monitor, fall within the scope of one or more patents relating to this monitor.
Terms of warranty
Your monitor is warranted against defects in material and manufacture for the duration of one year (from date of purchase). Excluded from this guarantee is damage caused by an accident or as a result of improper handling. The warranty entitles free replacement including labor, of the defective part. Any liability for subsequent damage is excluded. The warranty is void if unauthorized or unqualified persons attempt to make repairs.
In case of a defect, send the apparatus to your dealer or an authorized Welch Allyn service center. The manufacturer can only be held responsible for the safety, reliability, and performance of the apparatus if:
assembly operations, extensions, readjustments, modifications, or repairs are carried out by persons authorized by the manufacturer.
the monitor and approved attached equipment is used in accordance with the manufacturer's instructions.
There are no express or implied warranties which extend beyond the warranties hereinabove set forth. Welch Allyn makes no warranty of merchantability or fitness for a particular purpose with respect to the product or parts thereof.
This equipment has been tested and found to comply with the limits for a class A digital device, pursuant to both Part 15 of the FCC (Federal Communications Commission) rules and the radio interference regulations of the Canadian Department of Communications. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with this instruction user guide, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
WHEN USED IN CANADA: To prevent radio interference to the licensed service, this device is intended to be operated indoors and away from windows to provide maximum shielding. Equipment (or its transmit antenna) that is installed outdoors is subject to licensing.
Page 12
8 Safety Welch Allyn 1500 Patient Monitor
Page 13
9
2

Introduction

The monitor is designed for adult, pediatric and neonatal use. It has a 15-inch screen for comprehensive vital data monitoring. The monitor can be used with mains power (100 – 240 VAC) or with an internal battery.

Standard features

Function buttons and trim knob for easy operation
15-inch color screen
Vital parameters:
ECG (3, 5 or 12 lead)
Heart rate
Respiration
Non invasive blood pressure
•SpO
Temperature (x1)
Drug calculations
(Masimo or Nellcor)
2

Options

Printer
•etCO
12-lead resting ECG with measurements
12-lead resting ECG with measurements and interpretation
Arrhythmia analysis
ST analysis
Cardiac output and hemodynamic calculations
Invasive blood pressure (x2 or x4)
Temperature (x2)
and Integrated Pulmonary Index™ (IPI)
2
Page 14
10 Introduction Welch Allyn 1500 Patient Monitor
Visual alarm indicator
Patient connection panel
Function button panel
Trim knob
Ethernet (RJ45) connector
Potential equalization (ground)
Nurse call connector*
Mains connector (100-240 VAC) and fuse panel
Printer
USB connector (lift door)

The Welch Allyn® 1500 Patient Monitor

Front panel
Back panel
*The nurse call can be used to give an external indication of a parameter alarm.
Page 15
Directions for use Introduction 11
Note
Temperature 1
etCO2 sampling input
etCO2 sampling exhaust
ECG
NIBP
IBP 2
SVC
(service only)
SpO
2
IBP 1
IBP 3
IBP 4
Temperature 2
Cardiac Output
Connection panel
The connection panel layout will vary according to the options installed.
Page 16
12 Introduction Welch Allyn 1500 Patient Monitor

Function buttons

Print Printout of three waveforms and all parameters. The waveforms and print settings are defined in the printer can also be obtained when a limit is violated. This is also defined in system setup.
Alarm silence Silence / resume an audible alarm, or confirmation of displayed messages. The silence time is defined in the Setup/Administrator menu (see “Administrator” on page 108).
NIBP measurement interval Interval setup for non-invasive blood pressure measurement or switch-off of the interval measurement (see “Automatic blood pressure measurement” on page 64). Saves patient information.
NIBP measurement Start or stop of the non-invasive blood pressure measurement (see “Taking a single NIBP
measurement” on page 64).
Standby In standby mode patient monitoring is interrupted and the screen is blank. Monitoring is resumed when any button is pressed.
Setup Display of the Setup menu. The required menu item can be selected by turning the trim knob and pressing (see next page).
Trend Displays trend data (see “Trend data” on page 29).
menu (see “Recorder” on page 103). Note that an auto printout
Home Pressing this button closes opened dialogues and returns to the monitoring screen. Any settings that were changed in the opened dialogue screen are saved. Pressing this button is the same as selecting
ON/OFF Press to switch the monitor on. Press and hold for 4 seconds to switch the monitor off. The LEDs below this button indicate:
Left LED - mains power is connected to the monitor.
Right LED - mains connected to the monitor and internal battery being charged.
(see “Switching the monitor on or off” on page 19).
Trim knob The trim knob is used for navigation, value selection and value change. Use as follows:
1. Turn the trim knob to the left or right to select a field or value. A white frame appears around the field.
2. Press the trim knob to open the menu of the selected parameter field or value.
3. Turning the trim knob to the left or right to select the desired value.
4. Press the trim knob to apply the changed value.
OK on the opened dialogue screen.
Page 17
Directions for use Introduction 13
Note

Setup menu overview

Press to enter the setup menu and adjust the following settings and options:
The following is an overview of the setup menu options. Further details are given in section 8 (see “Settings” on page 101).
Parameter Settings/Submenus
Alarm Suspend Silences all alarms for a set period. The silence time is defined in the
Arrhythmia
Alarms Alarm overview. All alarm limits and print on alarm settings.
Speaker Volume Speaker volume.
HR/PR Tone Volume Heart beat volume.
Waveform area Defines the waveforms to be displayed and the size and sweep speed.
Recorder
Parameters
12-lead Resting ECG
Hemodynamic Calculations
Drug Calculations Screen for entry of drug parameters with dose and titration calculations based
Patient Information Enter/edit patient ID and patient information.
Patient Mode Neonatal:
Restore User Defaults Reset all settings to user defaults (see administrator > system, below).
1
2
3
5
Administrator menu (alarms).
Arrhythmia limits and alarm levels. Pacer Display and analysis (on/off).
Defines the data on the printout.
Enable/disable any combination of the following: ST measurements, etCO2, Masimo SpO2 settings, cardiac output, invasive blood pressure display and temperature display options.
4
View electrode status, and take a resting ECG. After the resting ECG has been taken, the option to obtain a printout is given.
Screen for entry of hemodynamic measurement parameters with automatic hemodynamic calculations based on entered parameters.
on entered parameters.
Birth through 28 days.
Pediatric:
Between 29 days and 12 years.
Adult:
13 years and older.
Page 18
14 Introduction Welch Allyn 1500 Patient Monitor
Parameter Settings/Submenus
Administrator Configuration
Display of monitor ID, network settings, options, etc. This is for information only.
Alarms
Alarm settings - silence time, suspend time, etc. This requires a password to enter (see “Settings” on page 101).
System
Time and date settings and unit preference (cm/in, kg/lb). This screen also gives the option to save the current settings as the user default settings, and the option to display the event log screen. This screen requires a password to enter (see “Settings” on page 101).
Communications, Service and Factory
These menu options are for service and factory personnel and can only be accessed by password only. Details are given in the service handbook.
Close Exits the setup menu.
1. The full arrhythmia option is only viewable when the full arrhythmia option is installed.
2. The recorder option is only viewable when the printer is installed.
3. The parameter options will vary according to the monitor configuration and licensed options.
4. The 12-lead resting ECG option is only viewable when the resting ECG option is installed.
5. The hemodynamics calculations option is only viewable when the cardiac output option is installed.
Page 19
Directions for use Introduction 15
Waveform fields Parameter fields
Time of last
blood pressure
measurement
Blood pressure
Alarm and system
messages
Temperature
Information field See“Information
field” following
Note
Note

Display overview

The waveform display is changed in setup menu see “Defining display
waveforms” on page 27).
The parameter fields can be changed according to licensed options, parameter selection (see“Defining parameter fields” on page 25), and patient panel options.
Page 20
16 Introduction Welch Allyn 1500 Patient Monitor
a b
Information field
Top line
The left box on the top line (a) displays the patient’s room number (entered in the patient information screen).
The middle box (b) displays the patient mode (Neonatal, Pediatric, or Adult) indicated by the highlighted icon.
.
Middle line
Displays the patient name.
Bottom line
Displays the current time. When mains is not connected, a battery symbol is also displayed to the left of the time (see “Power supply” on page 22).
Page 21
17
a
cb
3

Operation

Startup and initial preparation

WARNING Danger of electrical shock. Do not operate the monitor if the ground
connection is suspect or if the mains lead is damaged or suspected of being damaged.
WARNING Network the monitor to an Acuity Central Station only. Connecting to
other networks could damage the monitor or injure the patient. If in doubt about the network jacks or devices, consult your facility’s Biomedical Engineering Department.
Connections
1. Connect the mains cable to the rear of the monitor (a).
2. Ethernet connection (b) (not enabled).
3. Connect the potential equalization (ground) cable to the central potential equalization
lug (c).
Caution Ensure that the patient or any person with simultaneous patient
contact does not come in contact with conductive parts of any connectors including the RJ45 connector and the USB connector when the cover is opened.
Page 22
18 Operation Welch Allyn 1500 Patient Monitor
Note
a
b
c

Inserting recorder paper

The monitor is delivered without printing paper installed. Only use original Welch Allyn printing paper. The thermal paper is sensitive to heat, humidity, and chemical vapors. Store the paper in a cool, dry and chemical free area.
1. Pull the locking catch (a) to the front. The paper tray is unlocked.
2. Pull the paper tray out.
3. Insert paper and pull the beginning of the paper out. Make sure that the paper mark
(b) is facing to the top.
4. Reinsert and close the tray. Be sure that the paper lies exactly between the rails (c)
Page 23
Directions for use Operation 19

Switching the monitor on or off

Switching the monitor on
To turn the monitor on, press the On/Off button .
Switching the monitor off
To turn the monitor off, press the On/Off button for approximately 4 seconds. The
following message is displayed when the monitor is shutting down.
Initial Power up
1. Press the On/Off button (confirmed by a beep).
2. Confirm the New Patient dialogue with Yes or No.
Yes: Previous patient information is deleted. The patient information can be
No: Previous patient information, if any, is used.
3. Check the settings.
entered via the setup menu (Setup > Patient information (see “Patient
information” on page 107).
Page 24
20 Operation Welch Allyn 1500 Patient Monitor
Note
Note

Initial settings

Caution Only authorized personnel, trained in the operation of this monitor, are
qualified to do the setups in the following menu.
Alarm and general settings are given in the setup menu. Initial monitor settings may include general alarm settings (alarm silence time, alarm delay time, etc.), and general monitor settings (height and weight units, time and date, etc.).
Access the setup as follows:
1. Pr e s s t h e Setup button .
2. Use the trim knob to select parameters and change values. Press to confirm the selection and settings
The alarm settings are given in the setup menu: Administrator > Alarm
The system settings are given in the setup menu: Administrator > System
The alarm and the system sub-menu are password protected. The password for
both of these is 49, 48, 46.
Details of the setup menu and the passwords are given in the settings section (see “Settings” on page 101).
Page 25
Directions for use Operation 21
Note

Saving the user-defined settings as default

All monitor settings, including alarms settings, are stored until the monitor is switched off. To save the user defined settings as default,
1. Press the Setup button .
2. Select Administrator > System > Save User Defaults.
The system menu is password protected. The password is 49, 48, 46.
3. Confirm with ok:
Restoring the user settings
1. Press the Setup button .
2. Select Restore User Defaults. You are prompted to confirm:
Page 26
22 Operation Welch Allyn 1500 Patient Monitor
a b
Note

Power supply

Mains connected
When the mains supply is connected, the mains LED is illuminated (a). When the mains supply is connected, and the battery is recharging both mains LED (a) and the battery LED (b) are illuminated.
For battery recharging see “Recharging the battery” on page 91.
Mains interrupted
If the mains supply is interrupted, the monitor automatically switches over to battery operation. The user settings are maintained.
Disconnect from the mains
To isolate the monitor from the mains, disconnect the mains cable.
Page 27
Directions for use Operation 23
Full = between 87.5% and 100% capacity.
3/4 full = between 62.5% and 87.5% capacity.
Half full = between 37.5% and 62.5% capacity.
1/4 full = between 12.5% and 37.5% capacity
Empty = between 0% and 12.5% capacity.
Visual Alarm Indicator
Battery operation
Two batteries are available for the monitor:
Lithium-Ion battery: This type of battery will provide power for approximately two
hours when fully charged.
Lead acid battery: This type of battery will provide power for approximately one hour
when fully charged.
When running on battery power the battery symbol is displayed next to the time. The battery indicator gives an approximate guide to the capacity of the battery:
When the battery capacity is close to depletion:
the alarm message Battery low appears
the battery symbol flashes
an audible alarm beep is heard
the visual alarm indicator flashes blue
After a few minutes if the monitor is not connected to the mains supply, the message Battery nearly depleted is displayed and a continuous beep is heard; the monitor
switches off. If mains is connected during this period the monitor remains on.
Connect the monitor to the mains supply. For battery recharging, see “Recharging the
battery” on page 91.
Page 28
24 Operation Welch Allyn 1500 Patient Monitor
Note
Note
Note

Standby mode

WARNING In standby mode, vital signs data and alarms are no longer displayed
or collected.
In standby mode, patient monitoring is temporarily interrupted. Confirmed patient information is saved.
Patient information that has not been confirmed is lost when standby mode is entered.
1. Pr e s s t h e Standby button . The following screen is displayed:
The standby message is continuously displayed while the unit is in standby mode.
2. To exit standby mode press any button. You are prompted to confirm the same patient or enter a new patient.
After exiting standby mode, ensure that the NIBP intervals are re-armed by manually starting an NIBP measurement.
Page 29
Directions for use Operation 25

Defining parameter fields

1. Press the Setup button .
2. Select Parameters.
The screen will vary according to the licensed options:
etCO
•The Masimo settings enabled (yes / no), gives extra settings for SpO2
•The Temperature options are only available when two temperature connectors are
•The IBP options are only available when four IBP connectors are installed on the
, ST and CO are only displayed when the options are licensed.
2
measurement (see “SpO2 monitoring” on page 67) and is only displayed when the Masimo module is installed.
installed on the patient panel.
patient panel.
Page 30
26 Operation Welch Allyn 1500 Patient Monitor
RR or when enabled, etCO
2,
measurement, ST measurement, or both ST and etCO
2
measurements.
Two or four IBP measurements (or this area is left blank when no IBP measurement are selected).
Cardiac Output measurement (when enabled).
T1, T1 and T2, or T1 and T
Parameter field display
When etCO2 is enabled, the CO2 parameter replaces the RR parameter below the heart rate.
When the ST parameter is enabled, the ST parameter replaces the RR value and the RR value is moved below the ST parameter.
When both etCO
and ST parameters are enabled, both values are displayed below
2
the heart rate (RR is not displayed).
When IPI is enabled, the value is displayed below the CO
.
2
When four IBP connectors are available the number of IBP measurements displayed can be set, that is, no IBP measurements displayed, two measurements displayed, or four IBP measurements displayed (see previous page).
When two temperature connectors are available, one (T1), two (T1 and T2) or T1 plus the temperature difference can be displayed (T1 plus T).
Page 31
Directions for use Operation 27
Note
Note

Defining display waveforms

1. Press the Setup button .
2. Select Waveform Area.
3. The waveforms are configured through the pull-down menus.
4. Set the amplitude for each waveform according to preference and signal strength. Set the sweep speed (for all waveforms) according to preference and patient.
The RESP and SpO2 sweep speed values are not configurable.
5. Select OK to save.
The number of waveforms that can be displayed will depend on the configuration of the monitors and the options licensed and enabled. Either 4, 5, 7 or 8. waveforms can be displayed.
Waveforms P1, P2, P3, P4, CO the display parameter fields (see “Defining parameter fields” on page 25).
and ST are only displayed when selected in
2
Page 32
28 Operation Welch Allyn 1500 Patient Monitor
Waveform 1 - ECG (I, II, III, V, aVL, aVR, aVF)
Waveform 2 - ECG, RESP, CO2, ST or Trend (“Trend data” on page 29)
Waveform 3 - ECG, RESP, Trend, CO
2
, or SpO2
(Waveform 4- ECG, Trend, or SpO
2
)
Waveform 4 to 7 (5 to 8) - P1 to P4, or P1 and P2, or no IBP waveform displayed. P1, P2, P3, P4 will be displayed as selected for parameter measurement (see
“Defining parameter fields” on
page 25.)
Waveform display
The waveforms that can be configured are as follows:
Waveform position
1
2 •••••
3 ••••
4 ••
5 ••
6 ••
7 ••
8
RESP in waveform 2 can only be displayed when etCO
When ST is enabled SpO SpO
2
ECG RESP CO2 Trend ST SpO2 P1 P2 P3 P4
is not enabled.
2
is available for waveform 4 only. When ST is not enabled
2
is available for waveform 3 only.
When resting ECG is enabled, ECG waveform can also be set in waveform 4.
Waveforms P1, P2, P3, P4 are displayed when selected in the parameter measurements fields and shown in waveform fields 4, 5, 6 and 7 or 5, 6, 7 and 8 only.
The size of the waveforms are automatically adjusted for the number of waveforms displayed.
Page 33
Directions for use Operation 29
Note

Trend data

The monitor records up to 24 hours of trend data at one-minute intervals. The trend values are displayed at the user defined interval (see below) and additionally after every manual NIBP measurement.
Trend data is deleted when a new patient is entered.
When the memory is full, the oldest trend data is overwritten.
The display interval for the trend table can selected for intervals of one minute, five minutes, 15 minutes, one hour, or four hours.
Trend data display and settings
Press the Trend button .
Previous measurements are displayed using the up and down icons.
Use the trim knob to select the trend display interval with the pull down menu in the
Interval setting.
•The Clear option deletes all stored trend data.
•The Print option prints the trend data.
The data recorded will depend on licensed and enabled options and the number of IBP and temperature connectors installed on the monitor (see “Defining
parameter fields” on page 25).
Page 34
30 Operation Welch Allyn 1500 Patient Monitor
Note
Displaying trend data in the measurements screen
The HR trend can also be displayed in the waveforms:
Press the Setup button . and select Waveforms. Scroll down for more options.
The trend data can be displayed in waveform 2 or 3 (or 4 if ST is enabled):
Trend data can be displayed for any of the following:
Heart rate
RR (see example above)
•etCO
•fiCO
IPI when selected in parameter display)
•SpO
P1, P2, P3 P4
•NIBP
•T1
T2 (when selected in parameter display)
T (when selected in parameter display)
CO or CI (when selected in parameter display)
2
2
2
When the trend waveform is displayed and it is highlighted and selected with the trim knob (see next page), options appear at the bottom of the waveform (as shown on the example above) to:
Change the interval (15 mins to 24 hours)
Change the waveform scale
Obtain a printout.
Page 35
Directions for use Operation 31

Settings via a parameter field

1. Select the desired parameter measurement field using the trim knob. A white frame
appears around the selected field.
2. Press the trim knob to display the menu. The following example is displayed when
the heart rate settings screen is selected. Other setting screens are similar:
3. The settings can be saved as default (see “Saving the user-defined settings as
default” on page 21).
Page 36
32 Operation Welch Allyn 1500 Patient Monitor
Page 37
33
Visual alarm indicator
Text Display
Parameter Field
4

Alarms

Display of alarms

During initial powerup
No alarms are displayed if no patient is being monitored.
During monitoring
There are three alarm priorities:
Alarm type Priority LED visual alarm
indicator
Technical Low Blue Single deep tone
Parameter Medium Yellow (flashes with
parameter field)
Parameter High Red (flashes with
parameter field)
Lethal High Red (flashes with
parameter field)
Audible signal Display
every 2 seconds
Two tone high/low every second.
Three high tones every second.
Three high tones every second.
Text display in the alarm status field at the bottom.
Text display in the alarm status field at the bottom.
Yellow flashing parameter field.
Text display in the alarm status field at the bottom
Red flashing parameter field.
Text display in the alarm status field at the bottom
Red flashing parameter field.
Page 38
34 Alarms Welch Allyn 1500 Patient Monitor

Silencing an alarm

Acknowledging an alarm
Alarm Limit
Press the Alarm button to silence the alarm. The audible alarm is silenced for 1, 1.5
or 2 minutes. The visual parameter alarm continues to be displayed.
Press the Alarm button again to resume the alarm.
After the defined silence time, the audible alarm is reactivated. The silence time is
defined in Setup > Setup Administrator> Alarms > Alarm Silence Time (see
“Administrator” on page 108).
Technical Alarm
A technical alarm can be acknowledged by pressing the Alarm Silence button . This
alarm is not reactivated.

Suspend all alarms

The Alarm suspend is used to deactivate all alarms caused by for example, disconnecting patient cables, loose electrodes or relocation of the patient. The alarm is suspended for a
duration of 1, 1.5, or 2 minutes. During this time the message Alarms Suspended is
displayed.
The alarm suspension time is defined in the menu Setup > Administrator> Alarms > Alarm Suspend Time (see “Administrator” on page 108).
1. Pr e s s t h e Setup button .
2. Select Alarm Suspend.
A message is given in the message bar indicating that the alarms have been suspended.
If you wish to reactivate the alarms before the set duration, press the Setup button again
. The menu entry is changed to Alarm Resume. Select this option to reactivate the
alarms.
Page 39
Directions for use Alarms 35
Note

Turning off an individual parameter alarm

WARNING The audible alarm is silenced permanently. The settings are not
reset. Physiological alarms of the patient are silenced. Use this function only if disconnecting a sensor from the patient for a long period of time.
1. Individual alarms can be inhibited via the Alarms menu (see below) and in any parameter measurement field by using the trim knob to select a parameter (a white frame appears around the selected field) and pressing the trim knob to display the
menu for that parameter.
2. Switch off an individual limits by selecting the limit setting and rotating the trim knob to the maximum limit until off is selected.
3. The alarm off symbol is displayed in the respective measurement field.

Alarm Volume

The alarm volume is set in the setup menu.
1. Press the Setup button .
2. With the trim knob, select the menu option.
3. Press the trim knob to display the speaker volume.
The speaker volume is set on a scale of 1 to 10. The volume is heard when scrolling through the values.
A setting is available in the administrator menu, that prevents the HR / PR alarm from being switched off (see “Administrator” on page 108).
Page 40
36 Alarms Welch Allyn 1500 Patient Monitor
Note
Note
Note

Alarm limit setting

All alarm limits are reset to the default settings after confirming a new patient (see “Standby mode” on page 24).
1. Press the Setup button .
2. Select the menu item Alarms.
3. Use the trim knob to scroll through the alarm settings and select the limits.
Individual parameter alarm limits can be set in the parameter menu (see
“Monitoring and Measurements” on page 39).
The Alarm settings for arrhythmia are detailed in the Arrhythmia menu option in the setup menu.
Page 41
Directions for use Alarms 37
Note

Physiological alarms

Alarm abbreviation Description Priority
SpO
low/high Oxygen saturation of the blood Medium
2
PP low/high Peripheral pulse of SpO
2
RRECG low/high Respiration rate impedance Medium
Apnea limit Apnea time limit exceeded Medium
low/high Inspiratory CO
CO
2
low/high Capnographic respiration rate Medium
RRCO
2
etCO
low/high End-tidal expiratory CO
2
2
2
NIBPs low/high Systolic blood pressure Medium
NIBPm low/high Mean average blood pressure Medium
NIBPd low/high Diastolic blood pressure Medium
HR low/high Heart rate Medium
Pxs Art low/high Invasive systolic blood pressure Medium
Medium
Medium
Medium
Pxm Art low/high Invasive mean blood pressure Medium
Pxd low/high Invasive diastolic blood pressure Medium
Temp low/high Temperature in degrees Fahrenheit or degrees Celsius. Medium
All technical alarms are low priority.
Page 42
38 Alarms Welch Allyn 1500 Patient Monitor
Page 43
39
Note
Note
5

General

Monitoring and Measurements

Caution The guidelines in this section are given as an overview only. They are
not a substitute for, nor do they overrule manufacturer documentation and instructions or departmental procedures.
Values are only displayed when the ECG cable or at least one sensor is connected. If a sensor is disconnected, a technical alarm is issued. The measured value will no longer be displayed if the sensor is disconnected and the alarm is acknowledged.
Connect the ECG electrodes, the NIBP cuff, the SpO2 sensor, the CO2 sensor, and the temperature sensor to the patient as required.
As soon as the sensors are connected, the corresponding indication appears on the display.
Check or set the alarm limits (see “Alarms” on page 33).
This section gives a general overview of the parameters that can be measured with the monitor. It is aimed at medical professionals and no specific medical direction is given or implied; any instructions given here do not overrule local medical directives.
The individual parameter menu settings are selected with the trim knob, described previously (see “Settings via a parameter field” on page 31).
Page 44
40 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
Note
Note
©
ECG
WARNING In order to minimize interference and the danger of burns to the
patient, only use Welch Allyn ECG cables. Keep the ECG cable as far away as possible from any electrosurgical cables. Make sure that the electrosurgical return conductor (neutral) is properly attached to the patient and that good contact is made.
Patient preparation
The quality of the ECG reading is dependent on the degree of contact resistance between the electrode and the skin. To ensure the lowest resistance consider the following actions:
1. Shave the areas where the electrodes are to be placed.
2. Use alcohol to thoroughly clean the areas where the electrodes are to be placed.
3. When applying the electrodes, make sure that there is a layer of gel between the electrode and the skin.
To maintain the quality of signals during long-term monitoring, the electrodes should be replaced at least every 48 hours. Over longer periods, the electrode gel can dry out and the patient’s skin can be irritated by the gel or adhesive. When replacing electrodes, do not position the new electrodes on exactly the same locations, but a little to the side of the original positions.
Connecting the ECG patient cable
When an electrode falls off or the resistance of an electrode is too high, a lead-off indication is displayed and an audible alarm is issued.
Color code: the colors shown here are according to IEC requirements. The AHA color configuration is shown in “Electrode identification and color code
IEC/AHA” on page 43.
WARNING Patient harm. The monitor is type CF and protected only when
approved Welch Allyn patient cables are used.
Caution Only use approved Welch Allyn patient cables. Use of other cables can
damage the monitor.
Page 45
Directions for use Monitoring and Measurements 41
3-lead
5-lead
Page 46
42 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
Note
12-lead
This graphic shows the IEC color configuration. The AHA (U.S.) color configuration is shown in “Electrode identification and color code IEC/AHA” on page 43.
Page 47
Directions for use Monitoring and Measurements 43
Electrode identification and color code IEC/AHA
The electrode placements shown in this manual are labelled with the colors according to IEC requirements. The equivalent AHA colors are given below.
IEC (Europe) AHA (U.S.)
System Electrode
identifier
Limb R Red RA (right arm) White
L Yellow LA (left arm) Black
F Green LL (left leg) Red
Chest C1 White/Red V1 Brown/Red
C2 White/Yellow V2 Brown/Yellow
C3 White/Green V3 Brown/Green
C4 White/Brown V4 Brown/Blue
C5 White/Black V5 Brown/Orange
C6 White/Violet V6 Brown/Violet
Neutral N Black RL (right leg) Green
Color Electrode
identifier
Color
Page 48
44 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
Red (R)
Green (F)
Yellow (L)
White (C)
Black (N)
Pacemaker monitoring
WARNING Patients with a pacemaker must be observed continuously because
the heart rate from the pacemaker might still be registered in case of a cardiac arrest or some arrhythmias. See specification “Technical data” on page 117 for disclosure of the pacemaker pulse rejection capability of this monitor.
WARNING Pacemaker monitoring is not possible with ECG cables that have un-
shielded lead wires. Ensure that only shielded lead wire ECG cables are used when monitoring patients that have a pacemaker.
WARNING Welch Allyn recommends using an SpO
ECG measurement and to set the alarm range for the peripheral pulse (PP) in the range of the heart rate (HR), or to set the HR source in the SpO (see “SpO
settings” on page 69.)
2
3- and 5-lead cables for pacemaker patients
The following illustration shows the electrode placement with a 5-lead patient cable for optimum results for patients with an implanted pacemaker.
With a 3-lead patient cable, only R, L and F are connected.
sensor in addition to the
2
menu to SpO2
2
IEC AHA
Black (N) Green (RL)
Red (R) White (RA)
Yellow (L) Black (LA)
Green (F) Red (LL)
White (C) Brown (V)
Page 49
Directions for use Monitoring and Measurements 45
Note
aa
Activating the pacer display
1. Select the HR measurement field using the trim knob. A white frame appears around the measurement field.
2. Press the trim knob to display the menu.
3. Scroll down to the pacer display option and select yes.
Pacemaker spikes are presented as vertical lines (a) on the ECG trace. These vertical lines represent neither magnitude nor duration of the pacemaker pulse but are purely time relative.
The Analyze Pacer setting is not applicable.
Page 50
46 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
ECG display
ECG traces can be displayed in waveforms 1, 2 and 3. The waveform lead and size is defined in the waveform display menu (see “Settings via a parameter field” on page 31).
1. P r e s s t h e Setup button .
2. Select Waveforms.
Page 51
Directions for use Monitoring and Measurements 47
Note
12-lead resting ECG (option)
With this option it is possible to record a 12-lead resting ECG. One resting ECG can be stored at a time. The resting ECG cannot be viewed on the monitor but can be printed on the internal printer.
The 12-lead resting ECG is an option and only appears in the setup menu when enabled.
Taking a resting ECG
1. Press the Setup button .
2. Select 12-lead ECG.
3. Check electrode status. Ensure the green On is displayed for all electrodes - this indicates that the electrode resistance is within acceptable range to obtain a valid reading.
4. Select Start 12-lead Resting ECG. The message Rest ECG Analysis - in Progress is displayed while the resting ECG is being taken. This is followed by Rest ECG -
Complete.
5. The resting ECG is saved in memory until a new patient is defined or the ECG is overwritten with a new recording.
Page 52
48 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
Note
Note
ECG menu settings
The parameter settings are selected with the trim knob. (See “Settings via a
parameter field” on page 31).
The default settings are in bold.
Main menu Parameter Description
Setup HR ECG Lead Lead selection I,
Size 0.5, 1, 2, 4 mV/cm
HR/PR source
HR/PR Tone ECG Filter Select Diagnostic or Monitor
1
1
ECG, SpO
2
Off/ on
Two predefined filter settings can be selected. These defined filter options define the cut off frequency for the Myogram, Baseline and Mains filters. The filter definitions and the corresponding filter cut-offs are defined as follows:
II, III, V, AVL, AVR, AVF
, P1
2
Baseline Myogram Mains
Diagnostic 0.05 Hz 150 Hz as set Monitor 0.50 35 Hz as set
Single ECG Yes/No, select lead I, II, III, or V. Select this option to analyze one
ECG lead.
Pacer Display Yes/No displays pacer pulses relative to time but not
representative of either amplitude, duration or polarity.
Analyzer Pacer Yes/No - not applicable.
HR lower / upper
3
limit
Cal Generates a simulated 1 mV calibration impulse on the curve.
1. This can also be set in the setup SpO2 menu.
2. If SpO2 is selected, the pitch of the beep corresponds to the SpO2 saturation. A high pitched beep indicates a high
saturation.
3. The lower limit cannot be set to a higher value than the value set for the upper limit, and vice versa.
Range: 25 - 250 Lower limit default 50 Upper limit default 120
A setting is available in the administrator menu that prevents the HR / PR alarm (upper / lower limit) from being switched off (see “Administrator” on page 108).
Page 53
Directions for use Monitoring and Measurements 49
ECG alarms and messages
Message Possible cause Suggested action
HR – asystole/ASY No QRS detected for 4 seconds. Check the patient.
ECG signal lower than 0.5 mV.
HR – ventricular fibrillation/VF
HR – artifact Patient has moved. Calm the patient.
HR > [upper limit] HR < [lower limit]
HR – lead off Electrode lose/defective. Check and reapply/replace electrodes.
No organized ventricular rhythm detected.
Bad electrode. Checking the electrode pads.
Interferences by other devices. Remove source of the interference.
Heart rate higher/lower than alarm limit.
Patient cable defective. Replace the patient cable.
Check the patient. ECG signal lower than 0.5 mV.
Check the patient.
Page 54
50 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
Note

Arrhythmia

Arrhythmia settings
The full arrhythmia menu is only displayed when the full arrhythmia option is licensed. When the full arrhythmia option has not been enabled, settings are made for VFib, Asystol and Vtach only.
1. Press the Setup button .
2. Select Arrhythmia.
The default settings are in bold.
Main menu Parameter Description
Setup Arrhythmia VFib Alarm Level - Lethal (cannot be changed), Print
Asystole Alarm Level - Lethal (cannot be changed), Print
VTach VTach Limit 100 to 200 (125), Alarm Level - Lethal
PVC run PVC run 3 to 6 (6) Alarm Level (High, Medium,
VRyhthm Alarm Level (High, Medium, Low, Off), Print on
Couplet Alarm Level (High, Medium, Low, Off), Print on
PVC/min PVC/min 1to 30 (30) Alarm Level (High, Medium,
Bigeminy Alarm Level (High, Medium, Low, Off), Print on
on alarm Yes/No.
on alarm Yes/No.
(cannot be changed), Print on alarm Yes/No.
Low, Off), Print on alarm Yes/No.
alarm Yes/No.
alarm Yes/No.
Low, Off), Print on alarm Yes/No.
alarm Yes/No.
Page 55
Directions for use Monitoring and Measurements 51
Main menu Parameter Description
Setup Arrhythmia (continued)
Options Irregular Limit Yes/No, set limit between 45 and 120 (45)
Trigeminy Alarm Level (High, Medium, Low, Off), Print on
alarm Yes/No.
Tachycardia Tachycardia 150 to 250 (180) Alarm Level (High,
Medium, Low, Off), Print on alarm Yes/No.
Bradycardia Bradycardia 20 to 100 (45) Alarm Level (High,
Medium, Low, Off), Print on alarm Yes/No.
Pause Alarm Level (High, Medium, Low, Off), Print on
alarm Yes/No.
Irregular (irregular rhythm) Alarm Level (High, Medium, Low, Off), Print on
alarm Yes/No. This is an irregularity in the R to R interval over a series of at least 16 non­ventricular beats. The number of beats analyzed is given in options.
Non-capture (pacemaker non-capture)
Alarm Level (High, Medium, Low, Off), Print on alarm Yes/No. This is for pacemaker patients with the analyze pacers option enabled (see options below) - a beat does not directly follow a pacer.
Single ECG Yes/No, select lead
option to analyze one ECG lead.
Pacer Display Yes/No displays pacer pulses relative to time but
not representative of either amplitude, duration or polarity.
Analyze Pacers Yes/No enables non-capture (see above).
Arrhythmia OFF Sets all alarms to off except VFib, Asystole and
VTach which remain set at the highest alarm level.
Presets Resets all arrhythmia settings to the default.
ST / ARR Relearn Approximately 15 - 20 complexes are used to set
the parameters (duration, amplitude, etc.) for Arrhythmia analysis. Select this option to redefine the template used.
I, II, III, or V. Select this
Page 56
52 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
Note

ST measurement (option)

The ST segment represents the period from the end of ventricular depolarization to the beginning of ventricular repolarization. The ST segment lies between the end of the QRS complex and the initial deflection of the T-wave. It is normally isoelectric.
ST Analysis is a useful diagnostic tool because it may provide an early indication of myocardial ischemia or infarction.
ST Analysis is an algorithm that analyzes the offset of the ST segment from the ECG signal of normal beats. The offset of a reference beat is measured in millivolts (mV) (or millimeters) with respect to the isoelectric level of the ECG waveform; the offset of subsequent beats is measured relative to the reference beat.
The reference beat is originally obtained by learning the patient’s normal morphology, but it can be modified by the clinician at any time using the re-learn option in the ST Analysis Setup Window (see next page). When re-learn is selected reference beat is updated to the current morphology.
ST re-learn is also initiated by arrhythmia re-learn.
The ST measurement is made at a point 60 or 80 milliseconds (ms) after the J-point (see next page).
The J-point is the point on the ST segment where the slope changes (marking the end of the QRS and the beginning of the ST segment.
Page 57
Directions for use Monitoring and Measurements 53
Note
ST settings
ST measurements can only be displayed when the full arrhythmia and ST option is licensed, and when ST is enabled in setup parameters (See “Defining parameter
fields” on page 25).
The parameter settings are selected with the trim knob, described previously (see
“Settings via a parameter field” on page 31).
The default settings are in
Main menu Parameter Description
Setup ST ST Relearn Reanalyzes the QRS complexes to determine
bold.
base measurement levels.
ST Lead 1 Define the lead for measurement 1 (default lead
II).
ST lead 2 Define the lead for measurement 2 (default lead
V).
ST Measurement point Auto, 60, 80. Define the measurement point (ms
after the j-point). Note: When Auto is defined, the measurement point is set to either 60 or 80 ms. This is dependent on the results of an algorithm that builds a hysteresis counter based on the patient's average heart rate - a faster heart rate is set at 60 ms.
Lower limit / upper limit Range: -10 to +10
Lower limit default -1
Upper limit default +1 Alarm level Off, Low, Medium, High. Print on alarm Yes, No.
Presets Select to return all ST settings to the defaults.
ST alarm messages
Alarm Possible cause Suggested action
ST > [upper limit] ST< [lower limit]
ST is higher or lower than the alarm limit.
Check the patient.
Page 58
54 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
Note
Note

Respiration rate

The RR measurement field is not displayed if the etCO2 field is enabled.
If the RR should be measured via the ECG instead of etCO2, the etCO2 measurement field must be deactivated as follows:
Press the Setup button , select Parameters and deactivate etCO
The RR signal is measured via the R (RA) and F (LL) electrodes of the ECG cable (impedance measurement). After the patient is connected, about 30 seconds can elapse before a reliable value is displayed.
Respiration rate settings
The parameter settings are selected with the trim knob, described previously (see
“Settings via a parameter field” on page 31).
The default settings are in bold.
Main menu Parameter Description RESP Enabled Display RR measurement Yes, No. When the respiration is not
Apnea Apnea time 6, 10, 15, 20, 25, 30 seconds.
Setup RR Lower limit / upper limit Range: 2- 150
(No).
2
enabled, (Disabled) is displayed in the measurement screen.
Lower limit default 5 Upper limit default 30
Page 59
Directions for use Monitoring and Measurements 55
Respiration rate alarms and messages
Message Possible cause Suggested action
RR out of range (too high)
RR lead off Electrode loose/defective. Check and reapply/replace electrodes.
RR artifact Patient has moved. Calm the patient.
RR > [upper limit] RR < [lower limit]
The patient's RR is too high for accurate measurement.
Electrical interferences from other devices.
Signal disturbed due to frequent artifacts caused by bad electrode contact.
Interferences by other devices. Remove source of the interference.
Bad electrode. Check/replace electrodes.
RR is higher or lower than alarm limit.
Check the patient.
Remove source of the interference.
Check and reapply/replace electrodes if required.
Check the patient.
Page 60
56 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
Note

Capnography

The capnography module is intended to provide professionally trained health care providers with continuous, non-invasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and respiration rate. It is intended for use with neonatal, pediatric, and adult patients.
The etCO2 menu is only displayed when the etCO2 option is enabled. If the etCO2 measurement field is not displayed ensure it is enabled in the parameter settings.
Safety
Press the Setup button , enter the menu Setup/Parameters and activate etCO
Only use Welch Allyn approved accessories for etCO2 monitoring.
WARNING Carefully route the sampling line to reduce the possibility of patient
entanglement or strangulation.
WARNING The sampling line may ignite in the presence of oxygen when
directly exposed to laser, electro-surgical devices, or high heat. When performing head and neck procedures involving laser, electrosurgical devices or high heat, use with caution to prevent flammability of the sampling line or surrounding surgical drapes
.
2
WARNING When using a sampling line for intubated patients with a closed
suction system, do not place the airway adapter between the suction catheter and endotracheal tube. This is to ensure that the airway adapter does not interfere with the functioning of the suction catheter.
WARNING Loose or damaged connections may compromise ventilation or
cause an inaccurate measurement of respiratory gases. Securely connect all components and check connections for leaks according to standard clinical procedures.
WARNING Do not cut, remove any part, bend or crush the sampling line. This
could lead to erroneous readings.
Page 61
Directions for use Monitoring and Measurements 57
WARNING If too much moisture enters the sampling line (i.e., from ambient
humidity or breathing of unusually humid air), and the sampling line cannot be cleared, the message Blockage appears in the message area. Replace the sampling line once the sampling line blockage message appears.
Caution In high-altitude environments, etCO
values may be lower than values
2
observed at sea level. When using the monitor in high altitude environments, it is advisable to consider adjusting EtCO
Caution Microstream
®
etCO2 sampling lines are designed for single patient
alarm settings accordingly.
2
use, and are not to be reprocessed. Do not attempt to clean, disinfect, sterilize or flush any part of the sampling line as this can cause damage to the monitor.
Caution Dispose of sampling lines according to standard operating procedures
or local regulations for the disposal of contaminated medical waste.
Caution Before use, carefully read the Microstream etCO
sampling lines
2
Directions for Use.
Caution Only use Microstream etCO
sampling lines to ensure the monitor
2
functions properly.
Caution Dispose of Microstream etCO
sampling lines according to standard
2
operating procedures or local regulations for the disposal of contaminated medical waste.
Caution During nebulization or suction for Intubated patients, in order to avoid
moisture buildup and sampling line occlusion, remove the sampling line luer connector from the monitor.
Caution Replace the sampling line according to hospital protocol or when a
blockage is indicated by the monitor. Excessive patient secretions or a build-up of liquids in the airway tubing may occlude the sampling line, requiring more frequent replacement.
Page 62
58 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
Preparing the Oridion sensor
Basic principles for choosing microstream CO
When choosing Microstream CO2 sampling lines, the following should be considered:
The condition of the patient (ventilated or not ventilated)
If the patient is ventilated, whether ventilation is humidified or non-humidified
Patient’s size and weight
The probability that the patient will switch between oral and nasal breathing
Duration of use
For best results, for short term monitoring, use Microstream CO orange connectors. For long term monitoring, use Microstream CO with yellow connectors.” Products that include an “H” in the name are intended for long term use.
Select the appropriate sampling line and connect it to the monitor before putting on the patient. Be sure to follow the directions for provided with the sampling line.
For further information, please contact your local representative.
Connecting a sampling line
The appropriate sampling line must be connected to the monitor and to the patient. Connect as follows:
1. Slide open the sampling line input connector shutter and connect the appropriate sampling line.
sampling lines
2
sampling lines with
2
sampling lines
2
2. Screw the sampling line connector into the monitor clockwise until it can no longer be turned.
3. Connect the sampling line to the patient as described in the Directions for Use supplied with the sampling line.
When the sampling line is connected, the monitor will immediately begin to search for breaths, but it will not indicate a No Breath condition before any valid breaths have occurred.
CO2 data displayed by the monitor
The monitor Home screen displays real time CO
•Real time etCO2 values along with selected units and alarm settings
Respiration rate (RR) in breaths per minute and alarm settings
Real-time FiCO
•CO
Waveform if enabled (see “Waveform display” on page 28)
2
IPI value if enabled (see following).
values along with selected units and alarm settings
2
data. The displayed data includes:
2
Page 63
Directions for use Monitoring and Measurements 59
etCO2 settings
The parameter settings are selected with the trim knob, described previously (see
“Settings via a parameter field” on page 31).
The default settings are in bold.
Parameter Description Size 0 to 40 mmHg, 0 to 60mmHg, 0 to 80mmHg. Units mmHg/kPa. Apnea time 6, 10, 15, 20, 25, 30 seconds.
EtCO2 lower /upper limit Range: 0- 99 mmHg
RR lower limit / upper limit Range: 2- 150
FiCO
upper limit 2 - 25 mmHg (5 mmHg / 0.7 kPa).
2
Lower limit default 25 mmHg / 3 kPa) Upper limit default 60 mmHg / 8 kPa.
Lower limit default 5 Upper limit default 30.
Page 64
60 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
Integrated Pulmonary Index (IPI) settings and measurement
The capnography module provides the clinician with an integrated pulmonary index (IPI). The IPI is based on end tidal carbon dioxide, respiration rate, oxygen saturation and heart rate. The IPI is a single index of an adult or pediatric patient's ventilatory status displayed on a scale of 1 - 10, where 10 indicates optimal pulmonary status. IPI monitoring displays a single value that represents the patient's pulmonary parameters and alerts clinicians to changes in the patient's pulmonary status.
IPI provides an uncomplicated, inclusive assessment of a patient’s ventilatory and oxygenation status. By following the trend of the IPI, a clinician can quickly assess the inter-relations of a patient’s respiratory parameters.
IPI also provides an early indication of changes in a patient’s respiratory status that may not be indicated by the values of the individual parameters.
IPI is only supported in adult and pediatric mode. When patient mode is set for neonatal, IPI cannot be enabled. Patient mode is defined in the setup menu (mode see “Patient mode” on page 107). When pediatric is set, there are three pediatric sub­modes based on age. It is important to choose the patient age group to which the current patient belongs (see below).
IPI settings
Since the IPI uses data from the monitoring of both CO
and SpO2,
2
it will only be available when both parameters are available and RR and PR are being calculated from these parameters.
The parameter settings are selected with the trim knob, described previously (see
“Settings via a parameter field” on page 31).
The default settings are in bold.
Parameter Description
IPI Enabled
1
No/yes. When enabled the measurement appears below the etCO measurement. The lower limit is displayed after the measurement.
2
Age range
Lower limit 1 to 9 (default
1. IPI can only be enabled when patient mode is set to either adult or pediatric
2. Age range is only applicable when the patient mode is set to pediatric. When patient mode is set to adult this option
2
is greyed and cannot be entered.
1 to 3, 3 to 6, 6 to 12.
4)
Page 65
Directions for use Monitoring and Measurements 61
Note
Assessing the IPI value
IPI can provide an early indication of a change in ventilatory status which may not be shown by the current value of any of the four parameters individually. The IPI is designed to summarize information regarding patient status, possibly before etCO PR values reach levels of clinical concern. While the IPI highlights a change in ventilatory status, the root parameters are still required to understand patient status. It is important to note that different conditions can cause similar IPI values.
The IPI trend graph displays the patient ventilatory status trend in one graph, and thus can alert you to changes in patient status. The importance of the IPI therefore, lies not only in its absolute numeric number, but also in its relationship to previous values, so that a graph can display an upward or downward trend in patient status and indicates that attention or intervention may be required.
The IPI is modeled on normal healthy patients (ASA Physical Status value of 1 according to the ASA Physical Status Classification System, as published in Relative Value Guide® ­2008 of the American Society of Anesthesiologists). Patients with ASA Physical Status values of 3 or higher are expected to have low IPI values by definition. Therefore, for patients with an ASA Physical Status value of 3 or higher, the IPI may have no added value.
, RR, SpO2, or
2
The range of the IPI is 1-10; values should be understood as seen in the table below.
Index range Patient Status
10 Normal
8 - 9 Within normal range
7 Close to normal range: requires attention
5 - 6 Requires attention and may require intervention
3 - 4 Requires intervention
1 - 2 Requires immediate intervention
A graphical trend view of the IPI can be displayed. This is selected in waveforms (see “Displaying trend data in the measurements screen” on page 30).
If patient mode is changed, all IPI trend data is purged.
Page 66
62 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
etCO2 alarms and messages
l
Message Possible cause Suggested action
etCO
> [upper limit]
2
< [lower limit]
etCO
2
RR > [upper limit] RR < [lower limit]
FiCO
> [upper limit] FiCO2 has exceeded alarm limit. Check the patient and ventilator.
2
needs calibration CO2 measurement reading too
CO
2
etCO
is higher / lower than
2
alarm limit.
RR is higher / lower than alarm limit.
low due to:
Check the patient.
Check the patient.
Carry out calibration.
Incorrect calibration
Module requires calibration
IPI cannot measure CO2 measurement reading too
low. SpO
measurement too low or
2
absent.
Check the patient.
Check SpO
measurement and sensor.
2
Page 67
Directions for use Monitoring and Measurements 63

NIBP monitoring

Safety
WARNING To prevent extensive pressure on the extremity, it is very important
to:
Choose the correct cuff size.
Check the initial pressure in the NIBP menu. The correct initial pressure for
adults is 160 mmHg, for pediatric patients 120 mmHg, and for neonates 90 mmHg.
WARNING In case of long-term monitoring or automatic operation, the
connected body areas of the patient and the extremity to which the cuff is attached must be checked regularly for signs of ischaemia, purpuras or neuropathy.
WARNING The cuff must not be attached to a limb that is already used for
interventions such as infusions or SpO
measurement.
2
WARNING To prevent incorrect measurement results, make sure that the NIBP
tube is not compressed.
WARNING To achieve correct arterial pressure measurement, the cuff must
always be installed on the level of the right atrium.
Page 68
64 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
Note
Note
Note
Taking a single NIBP measurement
1. P r e s s t h e NIBP start/stop button .
2. The measurement can be stopped at any time by pressing the button again.
Automatic blood pressure measurement
1. P r e s s t h e NIBP measurement interval button .
2. Select the interval between 3 minutes and 60 minutes, and confirm your selection
with OK.
3. The message NIBP interval – xx minutes is displayed.
4. The first measurement is started by pressing the NIBP start/stop button .
After exiting the standby mode, ensure that the NIBP intervals are re-armed by manually starting an NIBP measurement.
These settings are reset when the monitor is switched off and automatic measurement must again be defined when the monitor is switched on.
The monitor sets the maximum pressure as follows:
Adults: 270 mmHg.
Pediatric: 180 mmHg.
Neonate: 150 mmHg.
The patient mode is defined in the setup menu (“Patient mode” on page 107).
Page 69
Directions for use Monitoring and Measurements 65
NIBP settings
The parameter settings are selected with the trim knob, described previously (see
“Settings via a parameter field” on page 31).
The default settings are in bold.
Parameter Description NIBP Interval Off, 3, 5, 10, 15, 30, 60 Format Sys/Dia or Mean. This defines the main measurement to be
displayed, that is the larger measurement displayed in the NIBP box. The secondary measurement is displayed smaller by the side of the defined measurement. Note that the mean measurement is shown in brackets.
Initial Inflation Pressure Adult: 160
Pediatric: 120 Neonate: 90
NIBPs (systolic blood pressure) upper/lower limit
NIBPd (diastolic blood pressure) upper/lower limit
Range: 30 - 300 Adult 220/75 Pediatric: 145/75 Neonate: 100/50
Range: 20 - 235 Adult 110/35 Pediatric: 100/35 Neonate: 70/30
NIBPm (mean pressure) upper/ lower limit
Range: 20 - 255 Adult 120/50 Pediatric: 110/50 Neonate: 80/35
Page 70
66 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
NIBP alarms and messages
Message Possible cause Suggested action
NIBP needs service No NIBP module detected. Switch off and restart.
Replace monitor.
NIBP artifact cannot measure
Cannot measure NIBP
NIBP cuff leak No cuff connected, or cuff or
NIBP signal low Blocked tube; kink in the tube. Check and replace the tube if required.
Patient has moved. Calm the patient.
Max. required pressure is higher than the initial pressure of 160 mmHg.
Patient has moved. Check and calm patient.
Very unsteady pulse. Apply cuff to another extremity with less
Air tube plugged or leaking. Check tube and cuff.
insufficiently fitted or defective.
Cuff not applied correctly. Reposition/check the cuff.
1
Repeat measurement. The monitor will automatically increase the initial pressure.
movement and steady pulse.
Check cuff position.
Check cuff for tightness.
Check if the cuff is connected to the monitor.
Pulse too low for good measurement.
NIBP time too long Inflation time exceeded (max.
135 sec.) due to interferences because the patient has moved
NIBPs < [lower limit] Systolic pressure too low. Check the patient and alarm limits.
NIBPs > [upper limit] Systolic pressure too high. Check the patient and alarm limits.
NIBPd < [lower limit] Diastolic pressure too low. Check the patient and alarm limits.
NIBPd > [upper limit] Diastolic pressure too high. Check the patient and alarm limits.
NIBPm < [lower limit] Mean pressure too low. Check the patient and alarm limits.
NIBPm > [upper limit] Mean pressure too high. Check the patient and alarm limits.
1. If the initial pressure is too low, the measurement is immediately restarted and the pressure is increased by
60 mmHg.
Apply the cuff to another extremity where the pulse measurement is easier.
Check the patient (see also message "cannot measure").
Repeat the measurement.
Page 71
Directions for use Monitoring and Measurements 67
Masimo module connector
Nellcor module connector

SpO2 monitoring

Two SpO2 modules are available with the monitor - Masimo or Nellcor. The two modules are distinguished by the connectors. The display data and waveform are similar for both modules. The Masimo module has extra clinical settings for signal processing (see
“Masimo settings” on page 70).
Pulse oximetry enables the continuous non-invasive monitoring of the functional oxygen saturation of the arterial hemoglobin and the peripheral pulse rate.
The display shows the continuous progress of the numeric SpO waveform and signal quality values.
The displayed plethysmographic waveform is not proportional to the pulse volume.
The update period of the measurement readings on the display is approximately 2 seconds.
In accordance with the relevant standards, the temporary alarm silence period can be set to a maximum of 2 minutes.
The peak wavelength and maximum optical power of the light emitted by the pulse oximeter probes can be especially useful to clinicians e.g. performing photodynamic therapy. They are as follows:
, plethysmographic
2
Range of peak wavelengths: 600 nm to 900 nm.
Maximum light power output: <15 mW.
Page 72
68 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
Safety
WARNING Only use sensors recommended from Welch Allyn for SpO
measurement with the monitor. Only use Masimo sensors when the monitor has a Masimo module. Only use Nellcor sensors when the monitor has a Nellcor module. Other sensors can impact the performance and give incorrect measurement readings.
WARNING The information in this manual does not overrule any instructions
given in the SpO read the sensor directions for use.
WARNING Do not use the pulse oximeters or sensors during magnetic
resonance image scanning. Induced current could potentially cause burns, and the pulse oximetry may affect the image and the accuracy of the measurements.
WARNING Do not use the pulse oximeter or sensors in or near the presence of
MRI equipment or in an MRI suite.
WARNING Tissue damage can be caused by incorrect application or use of a
sensor. Inspect the sensor site as described in the sensor directions for use to ensure skin integrity and correct positioning of the sensor.
sensor directions for use. Before using the sensor, carefully
2
2
WARNING Do not use damaged patient cables, damaged sensors or a sensor
with exposed optical components.
WARNING Substances causing disturbances: Carboxyhemoglobin can lead to
falsely high measurement readings. Colors or substances containing colors that influence the natural blood pigments can also lead to incorrect measurement readings.
WARNING Exposure to excessive illumination, such as surgical lamps
(especially those with xenon light source), bilirubin lamps, fluorescent lights, infrared heating lamps or direct sunlight, can affect the performance of an SpO sensor. To prevent exposure to excessive illumination, ensure that the sensor is correctly applied and that it is covered with an opaque material, if required. If these measures are neglected, excessive illumination can lead to incorrect measurements.
WARNING Change the sensor's position at least every 4 hours.
2
Page 73
Directions for use Monitoring and Measurements 69
SpO2 settings
The parameter settings are selected with the trim knob, described previously (see
“Settings via a parameter field” on page 31).
The default settings are in bold.
Parameter Description
alarm pause When the alarm is paused the message SpO2 Alarms Paused
SpO
2
HR/PR source
HR/PR tone
1
1
SpO2 lower/upper limit Range: 50- 100
1. This can also be set in the Setup HR menu.
2. If SpO2 is the HR/PR source, the pitch of the beep corresponds to the SpO2 saturation. A high-pitched beep
indicates a high saturation.
appears in the message line. The alarm is paused until the sensor is reinstalled.
ECG, SpO
displayed in brackets next to the HR parameter. Off, on
, P1 (selection of the heart rate source). The source is
2
2
Lower limit default 50 Upper limit default 100
Page 74
70 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
Note
SIQ values
PI value
Masimo settings
When the monitor has a Masimo module installed, the following settings are additionally available:
The extra settings for masimo modules are only available when Masimo settings enabled in the parameter settings (see “Defining parameter fields” on page 25).
Parameter Description
Display Signal IQ (SIQ) Waveform
1
On, Off. This value shows the acquired signal quality and timing of the pulse relative to the plethysmograph. A vertical line is displayed corresponding with the pulse beep: a tall vertical line indicates a high quality signal, a small line indicates a low quality signal.
Display Perfusion Index (PI)
Averaging Time 2, 4, 8, 10, 12, 14, 16. Averaging time in seconds. It is recommended
Sensitivity Normal, Max, APOD. The normal mode provides the best
Fast SAT Off, on. Fast SAT mode enables rapid tracking of arterial oxygen
2
On, Off. The perfusion index indicates the percentage of pulsatile to non-pulsatile signal. The range of signal strength is 0.02 to 20%. A value of 0 indicates that no measurement is available. This setting allows clinicians to place sensors on optimal sites and can also be used as a troubleshooting tool.
that the default of 8 seconds is suitable for most applications.
combination of sensitivity and probe off detection performance. This mode is recommended for the majority of patients. The max mode can be used where obtaining a reading is most difficult and the signal may be very weak. The adaptive probe off detection (APOD) mode is the least sensitive in picking up a reading on patients with low perfusion but has the best detection for off-probe conditions. This mode can be used for patients that are at particular risk of the sensor becoming detached (pediatric, combative, etc).
saturation changes by minimizing the averaging. This mode is clinically applicable during procedures when detecting rapid changes in oxygen saturation is paramount such as induction, intubation and sleep studies. Fast SAT is always on when averaging time (above) is set to 2 or 4.
1. It is necessary to reboot the device after changing this setting.
2. It is necessary to reboot the device after changing this setting.
Page 75
Directions for use Monitoring and Measurements 71
SpO2 messages
Message Possible cause Suggested action
SpO
– check sensor Defective SpO2 sensor Replace the sensor.
2
Incorrect settings in the monitor Check the monitor settings.
SpO
– check sensor
2
placement
Poor sensor contact or the sensor has fallen off
Check the contact between the sensor and the patient.
Sensor is disturbed by ambient
Cover the sensor.
light
Sensor defective (red light on
Replace the sensor.
the sensor is not lit)
Low Perfusion
SpO
2
1
Sensor not properly applied Check the sensor and reapply.
Fingernail varnish on the finger Remove fingernail varnish.
Thick skin Change finger.
Sensor failed Change sensor.
Weak Signal see SpO
unplugged SpO2 not connected to the
SpO
2
monitor
Low Perfusion above see SpO2 Low Perfusion above.
2
Connect sensor.
SpO2 artifact Patient has moved Calm the patient.
Hemodynamic interference Apply sensor to another extremity.
Too thin skin Apply sensor to a larger finger.
< [lower limit] SpO2 too low Check the patient and alarm limits.
SpO
2
SpO
> [upper limit] SpO2 too high Check the patient and alarm limits.
2
2
SpO2 low signal IQ
Poor quality signal Check the sensor and reapply. Move sensor
to obtain a better quality signal.
PR < [lower limit] Pulse rate too low Check the patient and alarm limits.
PR > [upper limit] Pulse rate too high Check the patient and alarm limits.
1. When the monitor has a Masimo module, the message SpO2 low perfusion only occurs when the perfusion index
feature is enabled.
2. The SpO2 low signal IQ message is only given when the monitor has a Masimo module and only occurs if Signal IQ
waveform feature is enabled.
Page 76
72 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
Note
Note
Note

IBP monitoring

Safety
WARNING Carefully read the manufacturer's instructions before using the
invasive blood pressure kit.
WARNING When applying the kit to the patient, make sure that absolutely no air
penetrates the system.
WARNING To achieve correct arterial pressure measurement, the pressure
sensor must be installed on the level of the right atrium.
WARNING If the pressure sensor's position is moved after calibration, this may
give inaccurate values.
WARNING If an invasive catheter for blood pressure measurement is introduced
into an arterial vessel, the circulation in the terminal vessels must be checked at regular intervals.
WARNING Single-use accessories must not be reused.
WARNING For patient safety, ensure that neither the sensors nor the patient or
persons touching the patient, come into contact with conducting objects, even if these are grounded.
WARNING Precautions must be observed when using high frequency devices.
To prevent the incorrect IBP measurements, only use sensors that are protected against high-frequency radiation.
The kit and operating procedure vary according to manufacturer. Please consult the manufacturer's documentation for connection.
For warm-up time/ready for measurement and displacement for invasive transducers, refer to the documentation of the transducer manufacturer.
P1 is the only connection that can determine the HR/PR source.
Page 77
Directions for use Monitoring and Measurements 73
Note
Preparing IBP measurement
Refer to the manufacturer’s directions for use for operating information for the IBP sensor.
IBP settings
The parameter settings are selected with the trim knob, described previously (see
“Settings via a parameter field” on page 31).
When changing the IBP label in the setup window, the upper and lower alarm limits will change to match the label (see“IBP alarm defaults” on page 74). The IBP default size (scale) will also change according to the label set.
Parameter Description
Zero Set Zero set the selected IBP1, IBP2, IBP3, IBP4
Label P1, P2, P3, P4
ART
PA
RA
LA
CVP
ICP
UA
UV
Size -10 to 20 mmHg
-10 to 60 mmHg
0 to 150 mmHg
0 to 200 mmHg
0 to 250 mmHg
0 to 300 mmHg
Format Sys/Dia or Mean. This defines the main measurement to be
displayed, that is the larger measurement displayed in the IBP box. The secondary measurement is displayed smaller below the main measurement. Note that the mean measurement is shown in brackets.
Page 78
74 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
IBP alarm defaults
Label / Parameter Adult
Lower Limit
P1, P2, P3, P4 P1, P2, P3, P4 P1, P2, P3, P4
ART SYS
DIA
ART ART Mean
CVP SYS CVP DIA CVP Mean
PA SYS PA
DIA Mean
PA
SYS
ICP ICP
DIA Mean
ICP
SYS
RA RA
DIA Mean
RA
SYS
LA LA
DIA Mean
LA
SYS
UA UA
DIA Mean
UA
SYS
UV UV
DIA Mean
UV
SYS DIA Mean
6 14 2 10 2 10
-4 6 -4 2 -4 2 0 10 0 4 0 4 75 220 75 145 50 100 35 110 35 100 30 70 50 120 50 110 35 80 6 14 2 10 2 10
-4 6 -4 2 -4 2 0 10 0 4 0 4 10 34 10 34 10 34 016016016 020020020 614210210
-4 6 -4 2 -4 2 0100404 6 14 2 10 2 10
-4 6 -4 2 -4 2 0 10 0 4 0 4 6 14 2 10 2 10
-4 6 -4 2 -4 2 0 10 0 4 0 4
Adult Upper Limit
Pediatric Lower Limit
Pediatric Upper Limit
Neonate Lower Limit
50 100 30 70 35 80 2 10
-4 2 0 4
Neonate Upper Limit
Page 79
Directions for use Monitoring and Measurements 75
Note
IBP zero set
Zero Set must be carried out before every application.
To prevent incorrect measurement readings due to the sensor's physical null drift, calibrate the sensor every 24 hours.
Ensure the sensor is kept still during zero set. If the pressure sensor's position is moved during zero set, this can lead to incorrect values.
1. Move to the desired IBP measurement field (P1, P2) using the trim knob.
2. Press the trim knob to display the IBP menu.
3. Select Zero Set with the trim knob and press to carry out the zeroing.
4. A straight line is displayed in the IPB waveform.
IBP alarms and messages
Message Possible cause Suggested action
IBP needs service No IBP module detected Switch off/on.
Replace monitor.
IBP needs calibration Zero-point sensor too high/low
by more than ± 30 mmHg or unsteady pressure
IBP artifact Loose sensor contact Inspect the sensor and cable connection.
A manipulation at the sensor, such as rinsing, has caused variation peaks of ± 150 mmHg
Incorrect IBP value displayed
IBPs < [lower limit] Systolic pressure too low Check the patient and alarm limits.
IBPs > [upper limit] Systolic pressure too high Check the patient and alarm limits.
IBPm < [lower limit] Mean pressure too low Check the patient and alarm limits.
IBPm > [upper limit] Mean pressure too high Check the patient and alarm limits.
IBPd < [lower limit] Diastolic pressure too low Check the patient and alarm limits.
IBPd > [upper limit] Diastolic pressure too high Check the patient and alarm limits.
Constant pressure (± 30 mmHg) during the calibration in the system
Check tube system, sensor and valves.
Re-calibrate the sensor.
After rinsing, calibrate the sensor.
Check tube system, sensor and valves. Set three-way valve to ambient pressure.
Re-calibrate the sensor.
Page 80
76 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
Note

Temperature monitoring

Depending on the sensor type, the sensor can be applied to the ear, the skin or the rectum.
To achieve a reliable measured value, independent of the measuring site, the measurement duration must be at least 2 minutes.
Temperature settings
The parameter settings are selected with the trim knob, described previously (see
“Settings via a parameter field” on page 31).
The default settings are in bold.
Parameter Description TEMP Display Mode T1 and T2, T1 only, or T1 and Temp Units °C or °F
T1 and T2 lower / upper limit Range: 15°C to 45°C (59°F and 113°F)
T lower / upper limit Range: 15°C to 45°C (59°F and 113°F)
Alarms can only be set for the temperature parameters that are currently on display. So for example, if T1 and delta T were selected for display, there would be no T2 alarm settings.
Temperature alarm and messages
Message Possible cause Suggested action
TEMP unplugged TEMP not connected to the
monitor
TEMP needs service The monitor has detected an
error
TEMP out of range The temperature is outside the
measuring range of the monitor.
T
Lower limit default: 35°C (95°F) Upper limit default: 37.8°C (100°F)
Lower limit default: 0.0°C (0.0°F) Upper limit default: 0.1°C (0.2°F)
Connect sensor.
Switch monitor Off/On or replace monitor.
Check the patient or alarm limits.
Sensor or monitor problem Check the sensor and monitor. Switch the
monitor off/on.
TEMP < [lower limit] Temperature too low Check the patient and alarm limits.
TEMP > [upper limit] Temperature too high Check the patient and alarm limits.
Page 81
Directions for use Monitoring and Measurements 77

Cardiac output (option)

Introduction
The WA 1500 patient monitor uses the thermodilution method of measuring cardiac output. A pulmonary artery catheter (PAC) is inserted into the right side of the heart. The PAC is balloon tipped and is inflated to help it through the right ventricle to occlude a smaller branch of the pulmonary artery system. The balloon is deflated and the PAC injects a small amount (10ml) of cold injectate at a known temperature into the pulmonary artery. The temperature is then measured at a known distance using the same catheter.
The cardiac output can be calculated from the measured temperature curve. Low cardiac output will change the temperature slowly, and high cardiac output will change the temperature rapidly. The degree of change in temperature is directly proportional to the cardiac output.
Usually three to five repeated measures are averaged to improve accuracy.
CO Units
Cardiac output is expressed as liters / minute:
Cardiac Output (CO) = Stroke volume × Heart rate
where: Stroke volume = End Diastolic Volume (EDV) - End Systolic Volume (ESV)
Heart rate = beats per minute
The cardiac index is expressed as liters / minute / m
Cardiac Index (CI) = Cardiac output/ BSA
where: BSA is Body Surface Area in square meters.
2
:
Page 82
78 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
Note
Note
Safety
WARNING Carefully read the manufacturer's catheter instructions before
carrying out measurements.
WARNING When an invasive catheter is introduced into an arterial vessel, the
circulation in the terminal vessels must be checked at regular intervals.
WARNING When applying the kit to the patient, make sure that no air
penetrates the system.
WARNING PAC use is complicated by arrhythmias, infection, pulmonary artery
rupture, and right heart valve damage.
WARNING Single-use accessories must not be reused.
WARNING It is important that the CO Computation Constant (see “Cardiac
output settings” on page 80) of the catheter is entered correctly to ensure
accurate measurements for the specific Edwards-compatible catheter used.
Caution Only use Edwards-compatible catheters
Caution Only use Baxter/Edwards/Abbott "in-line" injectate sensors
Caution It is important that the injectate temperature is in the range specified
for the computation constant.
Caution It is important that the injectate volume is the same as that specified in
the computation constant.
The kit and operating procedure vary according to manufacturer. Please consult the manufacturer's documentation for connection.
For warm-up time/ready for measurement and displacement for invasive transducers, refer to the documentation of the transducer manufacturer.
Page 83
Directions for use Monitoring and Measurements 79
Note
Note
Approved catheters and injectate sensor type
Catheters
The monitor will work with any blood temperature sensor that is Edwards-compatible.
Injectate Sensor
The injectate sensor must be a Baxter/Edwards/Abbott "in-line" thermistor.
The monitor does not support bath probe or arrow flow-through sensor.
Preparing for cardiac output measurement
The CO is only displayed when the CO option is enabled. If the CO measurement field is not displayed, ensure that it is enabled in the parameter settings.
1. Press the Setup button , enter the menu Setup/Parameters and activate CO
enabled (yes).
2. Connect the catheter to the CO connector on the side panel.
3. Setup the catheter according to the instructions given by the manufacturer.
Page 84
80 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
Note
Taking a CO measurement
Cardiac output settings
1. Use the trim knob to select and highlight the CO measurement field. Press the trim knob to display the settings.
Parameter selection with the trim knob is described previously (see “Settings via
a parameter field” on page 31).
2. The settings screen appears:
3. Make the settings as shown on the following page.
Page 85
Directions for use Monitoring and Measurements 81
CC =
1. 0 8 * C
t
* 60 * V
I
1000
The default settings are in bold.
Parameter Description Mode Auto, manual. In auto mode the unit monitors the blood temperature and
indicates that a measurement can be commenced only when the temperature is stable. In manual mode, the user monitors starts the measurement at any time.
Computation constant Between 000 and 999 (starting figure 400). This is a constant based on the
temperature coefficient of the catheter, temperature and volume of the injectate, and catheter dimensions.
The formula used is as follows:
Where:
1.08 = density times specific heat (5% dextrose / blood) C
= catheter thermal loss coefficient (specified by the catheter
t
manufacturer) 60 = seconds V
= volume of the injectate
I
1000 = conversion factor (cubic centimeters to liter)
Note: The C temperature of the injectate solution and separate C different temperature ranges, for example 0
value specified by the manufacturer, will differ according the
t
o
values will be stated for
t
to 5o, 19o to 22o , and 23o to 25o.
Since all factors are constants except the injection volume and the injection temperature, many manufactures will specify the computation constant CC for specific volumes and temperature range instead of the thermal loss coefficient C
in their instruction leaflet. This can be entered directly.
t
Display parameter CO, CI. Measurement shown as cardiac output (liters per minute) or cardiac
index (cardiac output / body surface area (m2)). For CI the patient data must have been entered.
Caution It is essential that the computation constant is entered correctly so
that the monitor knows the volume and temperature range of the injectate, and the thermal loss of the catheter. Inaccurate results will be obtained if the computation constant is not correct.
Page 86
82 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
Note
Real-time blood temperature
Injectate temperature
See Note below
Taking a CO measurement
1. Insert the catheter and prepare injectate solution in accordance with the instructions provided by the manufacturer.
2. On the monitor make the settings and enter the computation constant (see previous page).
3. The blood temperature is monitored and when the temperature is stable, Ready is
indicated at the top of the screen.
In manual mode, (see the mode parameter in“Cardiac output settings” on
page 80), a start button is positioned in the lower right of the screen. The Ready
indication is given when this is selected.
Page 87
Directions for use Monitoring and Measurements 83
Note
Temperature waveform
CO Measurements 1 to 5
Averaged value of all measurements
View all temperature curves
4. Inject the solution. The temperature waveform is displayed and the CO value given in measurement 1.
Follow you facility’s guidelines for injecting the solution.
5. Repeat to obtain 3- 5 measurements.
Editing the measurements
The average value is displayed in the bottom middle of the screen and is updated after every measurement or if a measurement is deleted.
After five measurements, when another measurement is taken, the latest measurement takes the place of measurement 5 and all measurements are shifted left (and measurement 1 is deleted).
Any measurement can be deleted by selecting and pressing the trim knob. If another measurement is taken, the later measurements are shifted left as above and the measurement deleted is replaced by the one higher.
•Select Review curves to view the temperature waveforms of all measurements.
Page 88
84 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
Note
Saving a measurement
When three or more consistent measurements have been obtained, Save Average can
be selected to save the measurement. You are prompted to confirm.
When confirmed the averaged measurement is entered in the CO field at the bottom of the screen.
The time when the measurement was made appears in the top right corner and the measurement parameter (CO or CI) appears in the left corner.
The measurement remains until overwritten or a new patient is confirmed.
A graphical trend view of the CO can be displayed. This is selected in waveforms (see “Displaying trend data in the measurements screen” on page 30).
Page 89
Directions for use Monitoring and Measurements 85
CO messages
The following self-clearing alert messages may be displayed during the CO measurement process:
Message Message trigger and when displayed Suggested action
BT out of range (too high or BT out of range (too low)
IT out of range (too high) or IT out of range (too low)
CO measurement Time-out
Blood temperature out of range (High or Low), displayed when:
detected during a CO measurement session.
CO start has been selected for manual mode.
Injectate temperature out of range (high or low), displayed when:
detected during a CO measurement session.
detected after CO start has been selected for manual mode.
No blood temperature change, displayed when:
CO start has been selected for manual mode and no reading is detected in the time-out period.
Verify correct catheter position.
Verify that injectate volume and temperature corresponds for the catheter used and that the computation constant has been entered correctly.
Check injectate temperature.
Check injectate probe and catheter probe connection.
Ensure the injectate is injected soon after CO start has been selected. Reselect Start.
Verify correct catheter position.
BT unplugged Blood temperature transducer unplugged,
IT unplugged Injectate temperature transducer unplugged,
BT check sensor Blood temperature transducer fault, displayed
IT check sensor Injectate temperature transducer fault,
CO computation constant not yet entered
displayed when:
detected during a CO measurement session.
detected after CO start has been selected for manual mode.
displayed when:
detected during a CO measurement session.
detected after CO start has been selected for manual mode.
when:
detected during a CO measurement session.
detected after CO start has been selected for manual mode.
displayed when:
detected during a CO measurement session.
detected after CO start has been selected for manual mode.
Computation constant not yet entered, displayed when:
detected during a CO measurement session.
Check connections and catheter.
Check connections.
Check connections.
Check catheter.
Check connections.
Check injectate sensor.
Enter the computation constant.
detected after CO start has been selected for manual mode.
Page 90
86 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
Entered Val ues
Calculated Values
Hemodynamic calculations
1. Press the Setup button .
2. Select Hemodynamic Calculations.
Enter the following parameters to calculate the hemodynamic calculations (see next page). As soon as the calculated values have sufficient data to for the entered parameters, the calculated values are shown.
Parameter Description
CO Cardiac Output
HR Heart rate (bpm)
ARTs Arterial pressure systolic
ARTd Arterial pressure Diastolic
ARTm Mean arterial pressure
PAs Pulmonary artery pressure systolic
PAd Pulmonary artery pressure diastolic
PAm Mean pulmonary artery pressure
CVP Central venous pressure
PAWP Pulmonary artery wedge pressure - the pressure measured by
Height Height in cm or inches
Weight Weight in Kg or lbs
wedging the pulmonary catheter with the inflated balloon in the small pulmonary arterial branch.
Page 91
Directions for use Monitoring and Measurements 87
Calculations
Parameter Units Description
Time Time when the calculations were made
BSA m
2
SV ml Stroke volume (milliliters / stroke):
SVR dyne-sec/cm
PVR dyne-sec/cm
LCW kg-m Left cardiac work:
Body surface area in m2 calculated as follows:
BSA = ([Height(cm) x Weight(kg)] / 3600 )
½
SV = (C0*1000) / HR
5
Systemic Vascular Resistance (SVR). Represents the load applied to the left ventricular muscle during ejection.
SRV = (ARTm - CVPm) * 80 / CO
5
Pulmonary vascular resistance (PVR):
PRV = (PAm - PAWP) * 80 / CO
LCW = CO * (ARTm - PAWP) * 0.0136
LVSW gm-m Left ventricular stroke work:
LVSW = SV * (ARTm - PAWP) * 0.0136
RCW kg-m Right Cardiac Work:
LVSW = SV * (ARTm - PAWP) * 0.0136
RVSW gm-m Right Ventricular Stroke Work:
RVSW = SV * (PAm - CVPm) * 0.0136
CI l/min/m
2
Cardiac Index:
CI = CO / BSA
SVI ml/m
2
Stroke Volume Index:
SVI = SV / BSA
Page 92
88 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
Parameter Units Description
5/m2
SVRI dyne-sec/cm
PVRI dyne-sec/cm
LCWI kg-m/m
LVSWI gm-m/m
RCWI kg-m/m
RVSWI gm-m/m
2
2
2
2
Systemic Vascular Resistance Index:
SVRI = SRV * BSA
5/m2
Pulmonary Vascular Resistance Index:
PVRI = PVR * BSA
Left Cardiac Work Index:
LCWI = LCW / BSA
Left Ventricular Stroke Work Index:
LVSWI = LVSW / BSA
Right Cardiac Work Index:
RCWI = RCW / BSA
Right Ventricular Stroke Work Index:
RCSWI = RVSW / BSA
Page 93
89
6

Maintenance

Maintenance interval

This software controlled monitor has undergone a software risk analysis to minimize any hazards associated with software defects.
The regular system maintenance must include a functional test according to the manufacturer's instructions. The test results should be recorded (see “Inspection and
checklist report” on page 94).
Maintenance work not described in this section, e.g. battery replacement, may only be accomplished by a qualified technician.
The following table indicates the intervals and responsibilities of the maintenance work required. Local regulations in your country may stipulate additional or different inspection intervals and tests.
Interval Maintenance Responsible
Before use Visual inspection of the monitor
Every 6 months Visual inspection of the monitor
User
and cables
User
and cables
Button check
Speaker check
LED check
Alarm check
Every 12 months Yearly test and test after repair
1. The need for calibration is based upon physical component changes that occur during use. The module requires its first calibration after 1200 operating hours or one calendar year, whichever comes sooner, and then after each 4000 operating hours or once a year, whichever comes sooner. The message Calibration Due appears when the hourly limit is reached. It is advisable to calibrate in the one-year maintenance program especially if the monitor is used for intermittent, short term use typical of patient monitors.
according to IEC/EN 62353.
Calibration
CO
2
1
Qualified technician
Page 94
90 Maintenance Welch Allyn 1500 Patient Monitor

Visual inspection

Defective monitors or damaged cables must be removed from service until repaired or replaced.
Visually inspect the monitor and cables for the following:
Monitor casing damaged or cracked, excessively scratched, etc.
Damage to the LCD screen.
Damage to sensor sheathing, mains, or potential equalization cables.
Damage to connection panels or connectors.
Legibility of the labels on the rear of the monitor.
Legibility of the annotation on the function button panel.

Button check

Press all buttons and trim knob and check that they work properly.

Speaker check

On switch-on, beeps must be audible.

LED check

Connect mains to the monitor and ensure the Mains LED is illuminated.
Disconnect the mains supply and leave the monitor on for 10 minutes. Reconnect the mains supply and ensure both the mains LED and the battery charge LED are illuminated.

Alarm check

The alarm check is performed with the SpO2 sensor. Proceed as follows:
1. Connect the SpO normal range.
sensor to a volunteer and check that the measurement is within
2
2. Set the SpO on page 69), so that the alarm activates.
3. Check that the visual and audible alarms are activated.
alarm to on and set the limit to the lowest setting (see “SpO2 settings”
2
Page 95
Directions for use Maintenance 91

Battery maintenance

The battery is maintenance free during its normal life.
No maintenance is necessary during normal operation.
If the monitor is not used, check and recharge the batteries every three months. The battery should not be allowed to fully discharge during storage.
Replace the battery every 2 to 5 years (depending upon application). When the running time falls substantially under two hours (lithium ion battery), or one hour (lead acid battery), replace the battery.
Recharging the battery
Totally discharged batteries require the following times to charge:
Lead acid battery: 80% capacity - 2.8 hours, 100% capacity 3.5 hours
Li-Ion battery: 80% capacity - 2.5 hours, 100% capacity - 6.5 hours
It is possible to use the monitor when the battery is being charged; however, the charging time of the battery will be extended.
1. Connect the monitor to the mains but do not switch it on.
2. The LEDs for both mains and battery are illuminated.
3. The battery LED is extinguished when the battery is fully charged.
Battery disposal
Batteries must be disposed of in municipally approved areas or sent back to Welch Allyn. See “Recycling monitor components” on page 99.
WARNING Explosion warning. The battery must not be burned or disposed of in
domestic trash.
WARNING Flammability and chemical danger. Do not open the battery.
WARNING Protect the contacts from shorting when disposing of the battery.
Apply non-conducting tape to the contacts.
Page 96
92 Maintenance Welch Allyn 1500 Patient Monitor

Inspecting and cleaning the monitor and accessories

WARNING Do not autoclave the monitor or any accessories.
WARNING Do not immerse the monitor in liquid when cleaning. Do not
immerse accessories in liquid when cleaning unless the accessory manufacturer’s cleaning instructions explicitly instruct you to do so.
WARNING Fire and electrical shock hazard. Always unplug the monitor from the
electrical power outlet before inspecting or cleaning the monitor and accessories. Exposing any of these to liquids, such as cleaning solutions, while they are connected to electrical power could result in electrical shock or fire.
Before cleaning the monitor or any accessories, thoroughly inspect them.
Look for any signs of damage and any improper mechanical function of buttons or
connectors.
Gently bend and flex cables, inspecting them for damage or extreme wear,
exposed wires, or bent connectors.
Confirm that all connectors engage securely.
Ensure that all transducers and accessories are within their expiration date.
Immediately report any sign of damage or malfunction to your service
department, and remove the monitor from service.
To clean the monitor or any accessories, follow these steps:
1. Wipe the equipment with a cloth slightly moistened (not wet) with one of the approved cleaning solutions listed in “Cleaning instructions and cleaning solutions” on page 93.
2. Clean cable assemblies by gently wiping from the center of the cable. Do not allow the sheathing to be displaced.
3. Thoroughly wipe off any excess cleaning solution. Do not let the cleaning solution run into or accumulate in connector openings, latches, or crevices. If liquid gets into connectors, dry the area with warm air, and then check the equipment to confirm that it operates properly.
Caution Use only a cleaning solution recommended by Welch Allyn for this
equipment. Use of any other cleaning solutions can cause damage to the equipment, including cracking and deterioration of the plastic case.
Caution Always follow the mixing/diluting instructions provided by the
manufacturer of the cleaning solution.
Caution Never use any of the following solutions or similar products to clean
the equipment: ethyl alcohol, ethanol, acetone, hexane, abrasive or scouring powder or material.
Page 97
Directions for use Maintenance 93
Cleaning instructions and cleaning solutions
Equipment Cleaning instructions Approved cleaning solutions
1
Monitor
ECG cable, extension cable
SpO
cable, extension
2
cable
Other accessories Consult manufacturer’s instructions. Consult manufacturer’s
1. The equipment can be disinfected to comply with OSHA requirements for cleaning and decontaminating spills of blood
and other body fluids. (Federal OSHA blood borne pathogens standard: 29 CFR 1910.1030, 12/6/91.)
2. If liquid gets into the connectors, dry the area with warm air and then verify all monitoring functions.
Wipe with a nearly dry cloth moistened with cleaning solution. Thoroughly wipe off any excess cleaning solution. Do not let cleaning solution run into connector openings or crevices.
2
70 % solution isopropyl alcohol; neutral mild detergent solution; all products designed for cleaning plastic.
Consult manufacturer’s instructions. Mild detergent solution; also
consult manufacturer’s instructions.
Consult manufacturer’s instructions. Consult manufacturer’s
instructions.
instructions.
Page 98
94 Maintenance Welch Allyn 1500 Patient Monitor

Inspection and checklist report

In accordance with the maintenance interval detailed previously, the following check list should be copied and followed.
Monitor Serial Number: _________________________________
Every six months
Inspection Result Checked
General examination
Visual inspection of the monitor.
Visual inspection of the LCD.
Visual inspection of all cable assemblies and sensors and accessories.
Monitor casing not broken or cracked.
LCD screen undamaged.
Electrode cable sheathing and connectors undamaged. 


No kinks, abrasion or wear in any cable assembly.
All transducers and accessories are within their expiratory date.
Plug and socket connectors.
Button check Buttons work. 
Speaker check. Switch the monitor on by pressing the On button.
LED check Mains LED on when mains
Alarm Check
Connect SpO volunteer.
Set SpO2 limit to lowest setting.
Recurrent test
Confirm the date of last factory inspections and test.
sensor to
2
Input/output connectors undamaged.
Switch-on beeps sounded. The standard screen is displayed. 
connected. Battery LED on when battery charging.
Measurement within range.
Measurement out of range and visual and audible alarm activated.
If the monitor is due for a yearly test, have a qualified technician perform the test.







Date of Inspection:
Inspector:
Page 99
Directions for use Maintenance 95
Every 2 to 5 years
Inspection Result Checked
Internal battery
Replace battery if operation falls substantially under two hours (Lithium ion battery), or one hour (lead acid battery).
Date of Inspection:
Inspector:
Replace battery 
Page 100
96 Maintenance Welch Allyn 1500 Patient Monitor

Replacing the fuses

WARNING Disconnect the monitor from the mains before changing the fuses.
WARNING Blown fuses must only be replaced with the fuse types indicated in
the below table.
Fuse types
Voltage range Number Fuse type WA Part No. Manufacturer Part No.
100-240 VAC 2 M 1.6A E 250V 4.210004 Schurter Inc, FSM 0034.2518
M= Medium time lag E= Enhanced breaking capacity
Changing the fuse
1. Disconnect the monitor from the mains.
2. Release the fuse holder by gently squeezing the side retaining clips and remove the fuse holder.
3. Replace both fuses. Re-insert the fuse holder until the two side clips snap in place.
Loading...