Weidmuller PMX400TMP User Manual [en, fr, es]

en
Operating instructions
fr
Mode d'emploi
Indicateur de température et de fréquence à 4 caractères, avec sortie des valeurs analogiques et surveillance des seuils 25
es
Instrucciones de empleo
Indicador de temperatura y frecuencia de 4 dígitos con salida analógica del valor y monitorizado de valor límite 49
PMX400/TMP 61001001/00/01.08
en
PMX400/TMP
Contents
Operation 16 Connections 18
tion 20
Calibra Setup Sequence 22
15
en
Temperature Sensor Inputs
Analogue Output (Optional)
Alarm 3
Alarm 4
Alarm 1
Alarm 2
(Optional)
(Optional)
Power
Supply
°C, °F or mV
PMX400/TMP
Operation
Warning: Check the power supply against the model number before applying power to the instrument.
Reviewing the setup
For review mode, disconnect security link and press PGM.
10 s timeout applies for review mode.
A
Changing the setup
For set-up mode, connect security link and press version will be displayed. If you wish to continue, press PGM again within 10 s.
Clearing tripped alarms
You can acknowledge a tripped alarm by simply pressing the key for that channel. 16
PGM.
The software
Siren and Group alarms
Channel four has two additional alarms types:
• Siren alarm, which you clear by acknowledging all tripped alarms
• Group alarm, which will only clear when you have acknowledged all tripped alarms outside the deadband)
Sensor break alarms
You can set channel one up as a sensor break alarm, which is triggered if the thermocouple input is open circuit or if any of the RTD sensor legs are disconnected.
Displaying the alarm setpoints
To check a setpoint, simply press the key for that alarm. If the display shows
button, you know it is set up as a sensor break alarm. If the display shows A4=Sr o
channel four button, you know it is set up as a siren or group alarm.
Adjusting the alarm setpoints
To change a setpoint, press the PGM k display. You will then be able to change the value using the arrows and save the change using the ENT key. Note that setpoint security must disabled for this to work.
Viewing the Cold Junction temperature
To view the CJC temperature, hold down the
PGM.
Temperature range limits
Choosing a smaller displayed temperature range improves the resolution for the analogue outputs. However it also limits the range for setpoint and display values. So, unless you are using the analogue outputs, you should choose the maximum range. This will give sufficient resolution for effective alarm and display operation.
and
all the trip conditions have cleared (and are
A1=OC w
hen you press the alarm channel one
r A4=gr when you press the alarm
ey while the value is still on
@
key and then press
17
• Connections
+ve--ve
White
mark
8 9 10 11
Terminal Signal
1 Neutral / –
2 Li
3 Ou
4 Current –
5 Voltage –
6 Securit y Link
7 Securit y Link
8
9
10
11
12
13 Normall y Closed
14 Common
15 Normall y Open
16 Normall y Closed
17 Common
18 Normall y Open
19 Normall y Closed
10 Common
21 Normall y Open
22 Normall y Closed
23 Common
24 Normall y Open
ve / +
tput +
Thermocouple in puts (via CJC Board) RTD inputs
Power suppl y
Anal ogue output (/AO option only)
Link t o allow access to the set-up mode (norma lly not connected)
Millivolt inputs
Not used Not used
B mV inpu t –
B
sense
A Not used
Alar m Channel O ne (/4R O option only)
Alar m Channel Two (/4R O option only)
Alar m Channel T hree (/4R O option only)
Alar m Channel F our (/4R O option only)
mV inpu t +
18
92.0 mm (+0.8/-0.0 mm)
CUTOUT
45.0 mm (+0.6/-0.0 mm)
PANEL THICKNESS Max 18.0 mm
18.0 mm
GASKET
Mounting Dimensions
19
• Calibration
The PMX400/TMP is factory calibrated. Once calibrated the PMX400/ TMP will operate with any sensor type or range, and any output type or range, without recalibration. You should not have to recalibrate the
instrument before installation.
Equipment requirements
• A suitable millivolt source (accurate to 5 μV or better)
• An accurate digital multimeter (accurate to 0.05 mV and ±0.1 μA)
• A Pt100 Rtd simulator (accurate to 0.05 ˚C)
Terminal Connections for calibration
Calibr ation Stage Signal t ype Terminal
milliVolt inputs
CJC Circu it
RTD Input s
Anal ogue Current Ou tput
Anal ogue Voltage Ou tput
Summary
There are four stages to calibrating the PMX400/TMP
• Calibrate the millivolt inputs
• Calibrate the CJC input
• Calibrate the RTD input
• Calibrate the analogue outputs
20
Millivolt s ource (+) 11 Millivolt s ource (–) 10 Millivolt s ource (+) 9 Millivolt s ource (–) 10
Rtd simulator B 10
Rtd S imulator B
sense
Rtd S imulator A 12
mA ou tput + 3 mA ou tput – 4
V output + 3 V output – 5
11
Note: the analogue output calibration sequence is not shown if the analogue output is disabled.
You can calibrate each stage in sequence or skip sections by pressing
When th e
display shows
CALIB ECALn
E=0 E=10 E=20 E=40 E=80
E=200 jCALn
CJ=0
Cj=60 rCALn
-200C 0C
400C 800C
OCALn
I=0
I=20
E=0
E=10
CALIB
SAvE
utton.
Action/Description
Pres s ENT to enter ca librate mo de Pres s ì or Å and then EN T to select millivolt input c alibration (or P GM to skip)
Appl y a 0.00 mV signal and press ENT Appl y a 10.00 mV signal and press ENT Appl y a 20.00 mV signal and p ress ENT Appl y a 40.00 mV signal and press ENT Appl y a 80.00 mV signal and press ENT Appl y a 200.0 0 mV signal and pre ss ENT Pres s ì or Å and then EN T to select CJC input calib ration (or PGM t o skip) Appl y a 0.00 mV signal to the CJC input and press ENT Appl y a 600.0 mV signal and press ENT Pres s ì or Å and then EN T to select Rtd i nput calib ration (or PGM to skip) Set the RT D simulato r to -200 ˚C and press E NT Set the RT D simulato r to 0 ˚C and press ENT Set the RT D simulato r to 400 ˚C and press E NT Set the RT D simulato r to 800 ˚C and press E NT Pres s ì or Å and then EN T to select analogue output calibration (or PGM t o skip)
Meas ure the current (mA) o utput. Adjust using the ì and Å keys un til the out put is at 0.00 m A. Pres s ENT.
Adjust using the ì and Å keys un til the out put is at 20.00 mA. Pres s ENT. Meas ure the out put volta ge. Adjust using the ì and Å keys un til the out put is at 0.00 V. Pres s ENT.
Adjust using the ì and Å keys un til the out put is at 10. 00 V. Pres s ENT.
You are now back in the main menu, press PGM Inst rument will save the new value s and return to r un mode
the PGM b
Put the ins trument in setup mode and scroll t hrough the main menu
21
• Setup Sequence
1(.
%.4
1(.
1(.
%.4
1(.
1(.
%.4
1(.
1(.
%.4
1(.
1(.
Menu Sett ing Options Description
v1.11
INP
dISP
S/ W Version, e.g., 1.11 No te: this guide is f or versions 1.10 to 1.19
Input menu
J typ e thermocouple
IP=_1
K typ e thermocouple
IP=_2
N typ e thermocouple
IP=_3
T typ e thermocouple
IP=_4
Input
type
Dampi ng
fact or
Display
units
Display
range
E typ e thermocouple
IP=_5
B typ e thermocouple
IP=_6
S typ e thermocouple
IP=_7
R typ e thermocouple
IP=_8 IP=_9
milliVolt s
IP=10
Pt10 0 RTD
Intr oduces the damping factor - - Next
dF=
Value e.g., 2
2
(minimum = 1 / maximu m = 99)
Display menu
Degrees Ce lsius
degC
Degrees Fa hrenheit
degF
(not shown f or milliVolt inputs) Display range upper limit
dHI=
Select the highe st value available
870.0
Display range lower limit - - Next
dLO=
-220.0
For informatio n only - Accept
ì Å
Inc Dec Accept
ENT
Accept
Next
Previous
Toggle A ccept
- - Next Toggle A ccept
22
%.4
1(.
1(.
%.4
%.4
1(.
1(.
%.4
1(.
1(.
1(.
%.4
1(.
%.4
1(.
1(.
Setup Sequence (continued)
1 & 2
AL1-4
AL1
Select
3 & 4
Select
Pres s PGM to skip to analog ue output menu.
Alarm channel setup menus (repeated for each enabled alarm channel)
Coil stat e
Alar m
type
Note: the following se ttings ar e not shown for group siren or sens or break al arms
Setp oint
Deadband
Timer
delay
Alarms 1- 4 enable/disable menu
Alar ms one and t wo enabled
A12y
Alar ms one and t wo disabled
A12n
Alar ms three an d four enabled
A34y
Alar ms three an d four disa bled
A34n
Normally energised
A1nE
Normally de-energis ed
A1nd
Low alarm
A1=lO a1=HI a1=OC A4=Sr A4=gr
High a larm Open c ircuit ala rm (ch 1 only ) Siren alarm (ch 4 only) Group alar m (ch 4 only)
Intr oduces setpoint value
SP1=
Any value witin the display limits-Inc-Dec
50.0
Intr oduces deadband value
db1=
Any value within the display span-Inc-Dec
10.0
Intr oduces tim er delay - - Next
dl1=
(0s to 420 0 s) Inc Dec Accept
0
Toggle A ccept
Toggle A ccept
Toggle A ccept
Toggle A ccept
Next
Accept
Next
Accept
AL2 , A L3 , AL4
A-ALL
Alarm Channel t wo, three and fo ur setup menus
All Alarms Subm enu (reset mode and setpoint se curity settings)
Rese t
seque nce
Setp oint
secur ity
Auto matic reset
nor
Manual res et
reS
Change set up mode only
SECy
Change in run or setup mode
SECn
Toggle A ccept
Toggle A ccept
23
%.4
1(.
1(.
%.4
1(.
1(.
Setup Sequence (continued)
Outp ut
selec t
Note: the following se ttings ar e not shown if the analogue outputs are disabled
Type
AOP
Input
range
Outp ut
range
Outp ut
Action
Analogue Outputs setup menu
Enabled
AOPy
Disabled
AOPn
Current (m A) output
OP=I
Voltage ou tput
OP=E
Input that gives minimum output
Inl=
e.g., 0 .0 °C
0.0
Input that gives maximum output
InH=
e.g.,100 .0 °C
100.0
Outp ut signal zero level
OPL=
e.g., 4 .00 mA
4.00
Outp ut signal f ull scale
OPH=
e.g., 2 0.00 mA
20.00
Direct act ing outpu t
OP=d
Rever se acting output
OP=r
Toggle A ccept
Toggle A ccept
-
Inc-Dec
Accept
-
Inc-Dec
Accept
-
Inc-Dec
Accept
-
Inc-Dec
Accept
Toggle A ccept
Next
Next
Next
Next
24
CalIB
SAVE
Calibr ation Menu (pr ess PGM to skip)
Inst rument is s aving the change s to the set up
fr
PMX400/TMP
Table des matières
Fonctionnement 26 Connexions 29
ge 31
Etalonna Séquence de configuration 34
25