Portacal 1000 61001066/00/01.08
en
Portacal 1000
Inhaltsverzeichnis
Description 17
Operation 19
Special functions 21
Calibration 25
16
en
Portacal 1000
1-key recall of all commonly used calibration points
Operating mode change switch
On/Of f Switch
27 S
torage locations for use r defined values (9 for each mode)
+ +Iext
Operating mode indicators
Increment/dec rement mode
indicators
Increment/dec rement mode chan ge
switch
Special mode select
Increment/dec rement by 1 V/mA
Increment/dec rement by 0.1 V/mA
Increment/dec rement by 0.01 V/mA
• Description
The Portacal 1000 is a microprocessor based calibrator for milliamp and voltage signals. It combines signal source and signal measuring features for diagnostic testing of analogue current and voltage loops.
Operation
The Portacal 1000 has three basic operating modes:
voltage source, which simulates auxiliary powered transmitters with
proportional voltage outputs;
current source, which simulates auxiliary powered transmitters with
proportional current outputs;
current sink mode,
outputs.
which simulates 2-wire (loop powered) transmitter
17
In addition, it has four special functions:
1. Two wire test
monitors its output level.
2. Automatic ramp/step,
the values stored for the operating mode at a user defined interval.
3. External Voltage measurement,
signals
4. External Current measurement,
current signals
Stored values
For each output mode there are nine memory locations for your favourite calibration points. In addition, most commonly used values are available by pressing one key.
Continuous Ramp/Step function
You can set the Portacal 1000 to repeatedly ramp or step between up to nine stored values in any output mode with programmable time delay.
Error messages
The display will flash when the battery is low. It also shows the following messages:
SHrt - short circuit output for voltage outputs OPEn - indicates open circuit output for current outputs
-OL- - indicates that the output is overloaded
, which powers a loop powered transmitter and
which ramps or steps through a selection of
for monitoring analogue voltage
for monitoring analogue milliamp
18
• Operation
I+ I­V
INC
DEC
00.01
Device with
Voltage Inputs
+
+Iext
+
Voltage Source Mode
Connections
+
­V
NC
DEC
00.02
Device with
mA Inputs
+Iext
Current Source Mode
Connections
+
+
Current/Voltage/Current Sink Mode selection
Ò to toggle the operating mode. The LED to the right of the
Press display shows the mode selected:
I+ for current source
I- for current sink
V for voltage output When you switch on the calibrator it will be in voltage source mode
with the output set to 0 V.
Output value selection
Press any of the numbered keys to output that value. The value in the top left corner is for current modes. The value in the bottom right is for voltage mode.
To select a specific value
Press the numbered key to give a value close to the value you require. Select increment or decrement mode using the Step to the value you want using the
Õ button.
Ù, â and á buttons.
19
+
­V
NC
DEC
00.02
2-wire Simulation
(Current Sink) Mode
Connections
Instrument
Power Supply
+Iext
Device with
Current Inputs
++
+
I+ I­V
NC
DEC
00.61
Loop Powered
Transmitter
+Iext
2-wire Test mode
Connections
Fn=1
+
+
To store a value for later use
Select the value you require.
Ó and then a numbered key. The value will be associated with
Press that key for the selected mode.
To recall a value
Ô and then the appropriate numbered key.
Press
20
• Special functions
Function Selection
Press Ô and then Õ the display will show Fn=0 Choose a mode using the Ù key.
Press Ó to select. Note: to exit any mode press the Ò key.
1. Loop powered transmitter test mode
In this mode the Portacal 1000 provides power for, and measures the output from, the transmitter. Select Fn=1 for this function.
2. Automatic ramp/step mode
This mode is designed for remote testing, it allows you to set the Portacal 1000 to repeatedly ramp or step between stored values in any output mode with programmable time delay.
First select the mode required (voltage source, current source or 2-wire simulation).
Select Fn=2. Press Ó, the display will show nP=. Press Ó, the display will show the number of points.
Set this to the number you require using the Õ and the Ù key. Select the time delay in seconds (dL=) and step (StEP) or ramp
(rAnP) mode in the same way. When you press Ó to accept the values the INC and DEC LEDs will
toggle to indicate that the unit is in ramp mode.
21
0 mA
20 mA
A1
(0 mA)
A2
(4 V)
B1
(8 V)
A3
(8 V)
B2
(6 V)
Waveform repeats until stopped
Time
Output
Step Mode Operation (Shown for Current outputs)
nP=5
18
9
20
10
10
5
8
16
4
8
0 V
10 V
A1
(0 V)
A2
(4 V)
B1
(8 V)
B3
(8 V)
C2
(10 V)
A1
(0 V)
A3
(8 V)
B2
(6 V)
C1
(10 V)
C3
(10 V)
Delay in Sec
Time
Output
Ramp Mode Operation (Shown for Voltage outputs)
18
9
20
10
10
5
12
6
8
16
4
8 2
1
4
2
0
0
Delay in Sec
Fn=2
22
3. External Voltage Measurement Mode
In this mode the Portacal 1000 measures external voltages. You can either display the measurement as a voltage or percentage. To display percentages you must set up upper and lower values for the measured range.
Select Fn=3. Press Ó, the display will show AbS or PEr for absolute value (volts) or
percentage. Select using Ù and accept using Ó.
If you selected absolute value mode the calibrator will now be displaying the measured voltage.
For percentage display the display will show InL=. Press Ó, the display will show the input span lower voltage.
Set this to the voltage you require using the Õ and the Ù key. Press Ó and the display will show I NH=. Press Ó, the display will show the input span upper voltage.
Set this to the voltage you require using the Õ and the Ù key. When you press Ó to accept the values the V, INC and DEC LEDs will
light to indicate that the unit is in external voltage measurement mode.
4. External Current measurement Mode
In this mode the Portacal 1000 measures external currents. It is identical to voltage measurement except you must select Fn=4.
In Current measurement mode the I, INC and DEC LEDs are all lit.
23
A
B
2 3
C
1
5 64 8 9
Stored values are used in this sequence for ramp/step mode
7
2 31
+
-
V
NC
DEC
03.01
Device with
Voltage Outputs
+Iext
Voltage Measurement
Connections
Fn=3
+
+
+ I­V
NC
DEC
00.02
Device with
Current Outputs
+Iext
Current (mA)
Measurement
Connections
Fn=4
+
+
24
CAL
Run
Calibration Jumper
• Calibration
Equipment Requirements
You will need:
• An accurate, 5 + 1/2 digit multimeter
• A suitable 24 V DC power supply
• An accurate current/voltage reference
• Suitable Phillips head screwdriver (for opening the case)
Keyboard operation
During calibration the keys function as follows:
Ó accepts that the output is at the required level. Note: always
allow about 10 s for the output to stabilise before accepting a value.
25
Ò switches operating modes. If you are adjusting a value you will
lose those changes.
Õ toggles between increment and decrement mode.
Ù, â and á adjust the output, á provides the finest adjustment.
Calibrate/Run mode selection jumper
Place the calibration jumper in the CAL position to start calibrating. You will have to open the case to select the jumper and normal precautions against ESD are essential. Take careful note of the position of the wires connecting the front and back of the case together.
Voltage Source Mode
Display I+ I- V Keys Action
Swit ch on Power, C alibrate link I n CA L Posi tion, Mea sure outp ut vol tage
u=0
u=0 Ó
u=10 Õ Ù â á
u=10 Ó
Sur e Ó
Off Off On
Õ Ù â á
Adjust Ou tput V olta ge to 0 V
Acce pt output
Adjust Ou tput V olta ge to 10 V
Acce pt output
To Save V alues (or any othe r key to rest art).
Current Source Mode
Display I+ I- V Keys Action
Swit ch on Power, C alibrate link I n CA L Posi tion
u=0
Conn ect mu ltimeter t o measure o utput cur rent (mA) .
Off Off On
Ò
Sele cts C urrent Sou rce Mode
26
Current Source Mode (continued)
i=2
i=2 Ó
i=20 Õ Ù â á
i=20 Ó
Sur e Ó
On Of f Off
Õ Ù â á
Adjust Ou tput C urrent to 2 mA
Acce pt output
Adjust Ou tput C urrent to 20 mA
Acce pt output
To Save V alues (or any othe r key to rest art).
Current Sink Mode
Display I+ I- V Keys Action
Swit ch on Power, C alibrate link I n CA L Posi tion
u=0
Conn ect P ower S upply and Mu ltimeter t o measure o utput cur rent (mA) .
i=2
i=2 Ó
i=20 Õ Ù â á
i=20 Ó
Sur e Ó
Off Off On
Off On Of f
Ò Ò
Õ Ù â á
Pres s twice to selec t Current Sink m ode
Adjust Ou tput C urrent to 2 mA
Acce pt output
Adjust Ou tput C urrent to 20 mA
Acce pt output
To Save V alues (or any othe r key to rest art).
External Voltage Inputs
Display I+ I- V Keys Action
Note : you mu st calibr ate th e Voltage S ource Mode bef ore co mpleting t his procedure.
Swit ch on Power, C alibrate link I n CA L Posi tion
u=0
Conn ect vo ltage sou rce and Se t to 0 V.
U=0
Off Off On
Off Off On
Ô Õ Ù Ù Ó
Ó
Pres s sequence to sel ect Ext ernal Volt age mode. INC a nd DEC Led s will both light .
Calibrate 0 V poi nt
27