The LPD450F is a field mount, 4+1/2 Digit, LCD Indicator for 4-20 mA
signals.
Loop powered operation
The LPD450F draws its’ power from the 4-20 mA signal current,
resulting in a voltage drop across the unit. The voltage drop
(2.5 V @ 20 mA) is equivalent to an increase in loop load of 125 Ω.
Cleaning
The case can be wiped with a damp cloth. De-energise the unit before
cleaning.
Using the peak and valley hold feature
To see the highest displayed value, hold down the ì key.
To see the lowest displayed value, hold down the Å ke y.
To reset either value, press P while it is on display.
Peak and Valley values are not stored when the instrument is switched
off.
Reviewing the set-up
To review the set-up, press the P key (run select switch must be in the
“normal” position). Press the Å key to scroll through the settings. A ten
second timeout applies.
Symbols
– Documentation must be consulted in all cases where this symbol
Caution: In order to meet product safety requirements, these units must
only be installed, by qualified staff, in accordance with the information
given in this manual, and all relevant national electrical wiring and
safety rules must be followed.
Industrial signals are frequently floating with respect to local earth,
so take care while connecting the signal wires. The LPD450F is not
intended to be connected to signals floating above 300 Vrms with
respect to local earth.
Location and mounting
Locate the instrument in an area that is free from dust, moisture and
corrosive gases.
The standard mounting method uses four 4 mm screws through
he housing. Mounting hole spacings are 120 x 60 mm. Remove the
t
front cover to gain access to the holes. Note: do not drill holes in the
enclosure as this will compromise the IP67 rating.
The optional pipe mount bracket can be attached to suit fitting to a
50 mm pipe. See below for assembly.
Access to instrument terminals
1. Remove the transparent front cover.
2. Release the thumbscrews from the either side of the LCD display.
3. Lift the electronics from the housing using the thumbscrews.
The connections for the instrument are on the underside of the
electronics and are marked for polarity. Use a suitable crimp lug for
connections.
Breakpoint linearisation gives a display of the measurement from nonlinear signals by breaking the cur ve up into a number of segments.
The breakpoints are spread evenly throughout the input range. So, for
example, if you set up a unit for nine breakpoints you will have to set
display values for 4, 6, 8, 10, 12, 14, 16, 18 and 20 mA.
Mann Series LPD450F
P
Switch positions (shown in correct position for installation).
General
The LPD450F is factor y calibrated. You should not have to recalibrate the
unit before installation.
Allow the instrument 15 minutes of powered operation (to reach a
stable temperature) before calibration. During calibration allow a few
seconds for signals to stabilise before accepting the value.
Equipment requirements
A suitable 4-20 mA current source
•
Input calibration procedure
When the display shows Action/Description
Connec t the cur rent source to the in puts an d set to 4.0 0 mA.
Put the r un/cali brate sw itch in ca librate p ositi on, selec t setup m ode.
CALb
CALn
CALy
InH=
UPCn
UPCy
Press P
Press Å
Press P three tim es
Set the cu rrent source to 2 0.00 m A
Press P three tim es
Press Å
Press P
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.