The 2015
workmanship within a period of one (1) year following the date of purchase of the
multimeter by the original purchaser or original user.
with proof of purchase to an authorized Wavetek Meterman Service Center or to the local
Wavetek Meterman dealer or distributor where your multimeter was purchased. See
maintenance section for details.
including but not limited to implied warranties of merchantability and fitness for a
particular purpose, are limited in duration to the above stated one (1) year period.
Wavetek Meterman shall not be liable for loss of use of the multimeter or other
incidental or consequential damages, expenses, or economical loss or for any claim or
claims for such damage, expenses or economical loss.
exclusion or limitation of incidental or consequential damages, so the above limitations
or exclusions may not apply to you.
which vary from state to state.
Digital Multimeter is warranted against any defects of material or
Any multimeter claimed to be defective during the warranty period should be returned
Any implied warranties arising out of the sale of a W
Some states do not allow limitations on how long implied warranties last or the
This warranty gives you specific legal rights, and you may also have other rights
avetek Meterman multimeter,
DGEWÄHRLEISTUNG
WARRANTY
Die Digitale Multimeter Modelle 2015 ist ab Kaufdatum für ein (1) Jahr
gegen Material- und Herstellungsfehler gewährleistet. Siehe Kapitel "Reparatur"
für Einzelheiten.
Implizierte Schadeforderungen sind auch auf ein Jahr beschränkt. W
Meterman ist nicht ansprechbar für Gebrauchsverluß oder Folgeschäden,
Ausgaben, Gewinnverluß, usw.
avetek
EGARANTIA
Este Multímetro Digital Modelo 2015 está garantizado contra cualquier defecto de
material o de mano de obra durante un periodo de un (1) año contado a partir de la
fecha de adquisición. En la sección de "Reparación" se explican los detalles relativos a
reparaciones en garantía.
Cualquier otra garantía implícita está también limitada al periodo citado de un (1)
año. Wavetek Meterman no se hará responsable de pérdidas de uso del multímetro, ni
de ningún otro daño accidental o consecuencial, gastos o pérdidas económicas, en
ninguna reclamación a que pudiera haber lugar por dichos daños, gastos o pérdidas
económicas.
FGARANTIE
Le multimètre digitaux, Modèle 2015 est garanti pour un (1) an à partir de la date
d'achat contre les défauts de matériaux et de fabrication. Voir chapitre "Réparation" pour
plus de détails.
T
oute garantie impliquée est également limitée à un an. Wavetek Meterman ne peut
être tenu responsable pour perte d'utilisation ou autres préjudices indirects, frais, perte
de bénéfice, etc.
CONTENTS
D • Inhalt
E • Contenido
F • Contenu
Safety Information ............................................... 2
Dépannage et Maintenance ................................ 37
EXPLANATION OF SYMBOLS
D • Erklärung der Symbole = E • Significado de los símbolos = F • Explication des Symboles
Direct current •D•
•F•
Courant continu.
Alternating current •D•
•F•
Courant alternatif.
Ground connection •D•
•F•
Connection de terre.
Attention! Refer to Operating Instructions •D•
lesen
Consultez le manuel.
Dangerous voltage may be present at terminals •D•
Spannung kann an den Eingängen anliegen
peligrosa en los terminales
présente aux entrées.
This instrument has double insulation •D•
geisoliert
appareil est prévu d’une double isolation.
Gleichstrom
•E•
Corriente continua
Wechselstrom
•E• ¡
Atención! Consulte las Instrucciones de Uso
•E•
Este instrumento tiene doble aislamiento
•E•
Erdanschluß
•E•
•F•
Une tension dangereuse peut être
Corriente alterna
Conexión a tierra
Achtung! Bitte Anleitung
•E•
Puede haber tensión
Dieses Gerät ist doppelt
– 1 –
•F•
Attention!
Eine gefährliche
•F•
Cet
WARNINGS AND PRECAUTIONS
n This instrument is EN61010-1 certified for Installation Category III. It
is recommended for use in distribution level and fixed installations, as
well as lesser installations, and not for primary supply lines, overhead
lines and cable systems. n All inputs are protected against continuous
overload conditions up to the limits of each function's stated input
protection (see Specifications). Never exceed these limits
nor the ratings
marked on the instrument itself. n Voltages applied to COM terminal
may be present at all other terminals n Exercise extreme caution when:
measuring voltage >20V, current >10mA, AC power line with inductive
loads, AC power line during electrical storms. High voltages can be lethal
and high voltage transients may occur at any time. n Operator injury or
damage to the multimeter may occur during current measurements if the
fuse blows in a circuit with open circuit voltage >600V (250 V in mA
input). n Always inspect your DMM, test leads and accessories for
sign of damage or abnormality before every use. If abnormal condition
exist (broken or damaged test leads, cracked case, display not reading,
etc.), do not use. n When testing for voltage or current, make sure these
ranges function correctly. Take a reading of a known voltage or current
first. n Never ground yourself when taking measurements. Do not
touch exposed metal pipes, outlets, fixtures, etc., which might be at
ground potential. Keep your body isolated from ground and never touch
exposed wiring, connections, test probe tips, or any live circuit
conductors. n Always measure current in series with the load – NEVER
connect the multimeter ACROSS a voltage source. Check fuse first. n
Never replace a fuse with one of a different rating. n Do not operate
instrument in an explosive atmosphere (flammable gases, fumes, vapor,
dust.) n Do not use this or any piece of test equipment without proper
training n CRT SERVICE SAFETY REMINDER : A potential danger
exists when measuring voltages in the horizontal output and damper
stages of CRT equipment. (High voltage transients greater than 6,000 V).
Refer to your CRT service manual for proper servicing instructions.
D • Warnungen und Vorsichtsmaßnahmen
n Dieses Gerät ist EN61010-1 zertifiziert für Installationsklasse III.
Anwendung ist empfohlen auf Verteilerebene und festen Anlagen sowie
untergeordneten Systemen, jedoch nicht für Starkstromnetze und
Hochspannungsanlagen. n Überschreiten Sie nie die kontinuierlichen
Überlastgrenzen der verschiedenen Meßfunktionen (siehe Spezifikationen)
– 2 –
oder andere Grenzen welche auf dem Gerät markiert sind. n Spannungen
welche am COM Eingang anliegen können sich auch auf andere Eingänge
übertragen nVorsicht beim Messen von Spannungen >20V // Strömen
>10mA // Netzstrom/-spannung bei induktiever Last oder bei Gewittern //
Strom, wenn die Sicherung durchbrennt in einem Schaltkreis mit
Leerlaufspannung >600V (>250V beim mA Eingang) // beim Messen an
Bildröhrgeräten (hohe Spannungsspitzen) n Unsersuchen Sie Gerät,
Meßkabel, Verbinder, usw. vor jeder Messung. Beschädigte Teile nicht
verwenden n Meßspitzen und Stromkreis während der Messung nicht
berühren • Sich selbst isolieren ! n Bei Strommessung, Multimeter
immer in Serie mit Schaltkreis verbinden – Nie in parallel mit
Spannungsquelle. n Sicherung immer mit gleichwertiger ersetzen. n
Gerät nicht in explosiver Umgebung verwenden.
E • Advertencias y Precauciones
n Este instrumento está homologado según EN61010-1 para la
Categoría de Instalación III. Su uso está recomendado en el nivel de
distribución y en instalaciones fijas, así como en instalaciones menores,
pero no en líneas principales de suministro, líneas aéreas ni sistemas de
cable. n No supere nunca los límites de entrada para las diferentes
funciones (vea Especificaciones), ni los límites marcados en el instrumento.
n Tensiones presentes en COM puedan estar presentes en otras entradas.
n Tenga especial cuidado al: medir tensión >20 V // corriente >10 mA //
tensión de red de CA con cargas inductivas // tensión de red de CA durante
tormentas eléctricas // corriente, si salta el fusible en un circuito con
tensión de circuito abierto >600 V (250 V en la entrada de mA) // trabajar
con pantallas TRC n Inspeccione siempre el multímetro, las puntas de
prueba, los conectores y los accesorios antes de cada uso. No utilice
ningún componente que esté dañado. n No se ponga Ud. a tierra cuando
esté tomando medidas, y no toque nunca partes expuestas de los circuitos.
n Al medir corriente, conecte siempre el multímetro EN SERIE con la carga
- NUNCA EN PARALELO con una fuente de tensión. n Nunca sustituya un
fusible con otro que no tenga las mismas especificaciones. n No utilice el
instrumento en ambientes potencialmente explosivos.
F • Avertissements et Précautions
n Cet instrument est certifié EN61010-1 catégorie d’installation III. Son
utilisation est recommandée pour le niveau de distribution de réseau, les
installations fixes et systèmes subordonnés, et non pour les installations de
puissance et lignes de transmission et câblages à haute tension. n
N’excédez jamais les limites de surcharge continues (voir spécifications) ou
d’autres limites marquées sur l’appareil. n Des tensions appliquées à
l’entrée COM peuvent être également présentes aux autres bornes d’entrée.
n Soyez très prudent quand vous mesurez des tensions >20V ou des
– 3 –
courants >10mA // tension ou courant de secteur avec charge inductive ou
par temps de tempête // du courant quand le fusible saute dans un circuit
avec tension en circuit ouvert de >600 volts (>250V pour l’entrée mA) //
dans des appareils à tube cathodique (transitoires à haute tension) n
Inspectez appareil, câbles, connecteurs avant chaque mesure. N’utilisez pas
des pièces endommagées n Ne touchez pas les pointes de touche ou le
3 3/4 digit LCD with Unit, Status and Function Indicators
3 3/4Digit LCD ; Dezimalpunkt, Funktions- und Einheitsanzeigen
LCD de
3
3/4
dígitos con indicadores de unidades, estado y función
LCD 3 3/4digits ; point décimal, annonciateurs de fonctions et d’unités
Low Battery
Batterie entladen
Pila baja
Pile déchargée
40-Segment Bargraph
40-Segment Bargraf
Barra analógica
Bargraphe 40-segments
Function/Range/Off
Selector
Funktion-/Bereich-
/Aus Schalter
Selector de
Función/Escala/Off
Sélecteur fonctions/
calibres/marche-arrêt
20 A Input
20 A Eingang
Entrada 20A
Entrée 20 A
400mA Input
400mA Eingang
Entrada 400mA
Entrée 400mA
COM Input – common or low input for all
measurements
COM Eingang – Referenzpunkt für alle Messungen
COM- entrada común o “baja” para todas las
medidas
Entrée COM – commun ou bas pour toutes mesures
Special Function
Buttons
Spezielle Funktionstasten
Teclas de función
especial
Boutons de
fonctions spéciales
High input for
voltage, resistance,
frequency,
capacitance
V HzC Eingang. Hoch
für Spannung
Widerstand,
Frequenz, Kapazität
Entrada “alta” para
tensión, resistencia y
capacidad
Entrée V HzC. Haut
pour tension,
résistance,
fréquence, capacité
– 4 –
circuit pendant les mesures. Isolez-vous ! n Pour la mesure de courant,
connectez l’appareil en série avec le circuit – JAMAIS en parallèle avec une
source de tension. n Remplacez les fusibles toujours par des fusibles
équivalents n N’utilisez pas cet appareil dans des atmosphères
explosives.
PREPARATION FOR USE – UNPACKING
Your shipping carton should include the multimeter, a holster, one test
lead set (one black, one red), one 9V battery (installed), a warranty card
and this manual. If any of the items are damaged or missing, immediately
return the complete package to the place of purchase for an exchange.
D • Gebrauchsvorbereitung - Auspacken
Die Verpackung sollte enthalten: ein Multimeter, ein Meßkabelsatz (ein
schwarz, ein rot), eine 9V Batterie (im Gerät), eine Garantiekarte und diese
Anleitung. Wenn ein Teil fehlt oder beschädigt ist, bitte bei der
Verkaufsstelle umtauschen.
E • Preparación del multímetro para su uso - Desembalaje
El embalaje debe contener: el multímetro, una funda protectora, un juego de
puntas (una negra y otra roja), una pila de 9 V (instalada), una tarjeta de
garantía y este manual. Si falta algún componente u observa daños,
devuelva el conjunto al lugar donde lo adquirió para que se lo cambien.
E • Préparation pour l’Emploi - Déballage
Votre emballage doit contenir: un multimètre, un jeu de câbles de mesure
(un rouge, un noir), une pile 9V (installée), une carte de garantie et ce
manuel. Si une pièce manque ou est endommagée, retournez à votre point
de vente pur un échange.
PROTECTIVE HOLSTER
The holster/tilt stand protects the meter from accidental falls and provides
greater ease of use. The tilt stand has two positions, and a velcro-strap
allows to hang the meter from a fixture. Both test lead probes can be
attached to the holster for storage. One probe can be attached for
measurement, holding the meter with probe in one hand and the second
probe in the other hand.
– 5 –
D • Schutz-Holster
Das Schutzholster
beschirmt das Gerät
vor Stürzen und
Stößen. Der
Kippständer hat zwei
Positionen, und mit
der Schlaufe können
Sie das Gerät
aufhängen. Die Meßsonden können am
Holster befestigt
werden.
E • Funda Protectora
La funda con pie integrado protege el medidor en caso de caídas
accidentales. El pie inclinado tiene dos posiciones, y una banda con Velcro
permite colgar el multímetro. Las puntas de prueba pueden fijarse a la
funda.
F • Holster de Protection
Le holster protège l’appareil contre des chutes accidentelles et des chocs. Il
est muni d’une béquille à deux positions et d’une boucle de suspension.
Les sondes des cordons peuvent être attachées au holster.
OVERLOAD INDICATION
Range Overload is indicated by a flashing “4”
(MSD) and three non-flashing zeros (plus a
beep tone for voltage, capacitance and
current). Take immediate steps to remove the
cause of the overload. Select the next higher
range until a value is displayed. If overload still exists in the highest
range, remove test leads from the measurement setup as it is beyond the
range of the meter. Overload indication is normal in the OHMS and
continuity ranges when the leads are not connected to anything or when
the measured value is higher than the selected resistance range.
– 6 –
D • Überlastanzeige
Bei Bereichsüberschreitung blinkt die höchste Ziffer “4”; andere Digits
werden als “0” angezeigt. In Spannungs-, Kapazitäts- und Strombereichen
ertönt gleichzeitig ein Summton. Ursache der Bereichsüberschreitung sofort
entfernen (Höheren Bereich wählen oder Messung unterbrechen).
Überlastanzeige ist normal in Widerstands- und Durchgangsbereichen bei
offener Schaltung oder zu hohem Widerstand.
E • Indicación de Sobrecarga
La sobrecarga de escala se indica mediante el parpadeo del dígito más
significativo (“4”) y tres ceros fijos (más un bip o tono audible en tensión,
capacidad y corriente). Para eliminar la causa de la sobrecarga, seleccione
una escala más ata o interrumpa la medida. La indicación de sobrecarga es
normal en medida de OHMS y prueba de continuidad, cuando el circuito
está abierto o la resistencia es demasiado alta.
F • Indication de Surcharge
Un dépasssement de calibre est indiqué par le clignottement du chiffre le
plus significatif “4”; les autres chiffres étant “0”. Cet affichage est
accompagné d’un signal sonore dans les gammes de tension, de capacité et
de courant. Ecartez immédiatement la cause du dépassement – choisissez
une gamme plus élevée ou interrompez la mesure. L’indication de
dépassement est normale dans les gammes de résistance ou de continuité
avec circuit ouvert ou résistance trop élevée.
– 7 –
MEASURING PROCEDURES
GENERAL: Turn instrument on by turning function/range switch away
from OFF and selecting the parameter you want to measure. Six switch
positions have dual functions (resistance/continuity, diode/capacitance,
four AC/DC current ranges). How to select either is explained under the
respective function.
measurement unit displayed. In the example shown you would be in the
40V AC range. Hold a range by pressing the RANGE button. Cycle
through ranges by continuing to depress RANGE. Press and hold RANGE
button for two seconds to return to Autoranging.
Bargraph: The 40-segment bargraph indicates at which percentage of
the range you presently measure. It updates 20 times per second, and,
like an analog needle display, it is ideal for circuit adjustments.
Display Hold: When you need to concentrate on your circuit, you can
push the HOLD button to keep the reading on the display. “HLD” appears
in the display. Push the HOLD button again to resume normal display of
current value. (Automatic shut-down is disabled while “HLD” is
displayed).
General Measurement Procedures: When connecting or
disconnecting test leads to a circuit, always first turn off power to device
or circuit being tested and discharge all capacitors. Strictly observe
the max input limits. Do not change functions while test leads are
connected to circuit.
Ranging: This instrument is autoranging on
all ranges except
current and 200mV AC.
It automatically selects
the range that gives the
best resolution for the
measured value.
“AUTO” in the display
confirms that you are in
Autoranging mode. You
can tell which range
(and function) you are
in by the position of the
decimal point and the
– 8 –
D • Meßprozeduren
Allgemein: Gerät einschalten durch Funktionsschalter von OFF weg, nach
einer gewünschten Meßfunktion zu drehen.
Bereichswahl: Automatisch für alle Funktionen/Bereiche, ausgenommen
Strom und 400mV AC. “AUTO” wird angezeigt. Aktuelle Funktion und
Bereich werden durch Dezimalpunkt und Meßeinheit angezeigt. Sie können
einen Bereich festhalten durch die RANGE Taste zu drücken. Sie können die
verschiedenen Bereiche manuell durchlaufen durch wiederholt RANGE zu
drücken. RANGE zwei Sekunden lang drücken um erneut automatische
Bereichswahl zu wählen.
Bargraf: Der 40-Segment Bargraf zeigt an an welcher Stelle des Bereiches
(%) Sie messen. Erneuerung: 20 pro Sekunde.
Anzeigenstop: HOLD Taste drücken um einen Meßwert auf der Anzeige
festzuhalten. “HLD” wird angezeigt. HOLD erneut drücken um aktuelle
Meßwerte anzuzeigen. Bei aktiver HOLD Funktion schaltet das Gerät nicht
automatisch ab.
Allgemeine Meßprozeduren: Beim Verbinden/Trennen der Meßkabel
mit dem/vom Schaltkreis, zuerst Schaltkreis abschalten und Kondensatoren
entladen. Eingangsgrenzen beachten. Wahlschalter nicht drehen
wenn Meßkabel mit dem Meßkreis verbunden sind.
E • Procedimientos de medida
EN GENERAL: Encienda el instrumento girando el selector de
función/escala en otra posición que no sea OFF, seleccionando el
parámetro que desee medir.
Selección de escala: Este instrumento selecciona automáticamente la
escala (auto-rango) en todas las medidas excepto corriente y 400mV AC.
Aparece la indicación “AUTO”. La escala y la función seleccionadas en cada
momento se indican mediante el punto decimal y la indicación de unidades.
Para mantener fija una escala, pulse la tecla RANGE. Para recorrer todas las
posibles escalas, pulse repetidamente RANGE. Para volver a AUTO, pulse y
mantenga pulsada RANGE durante dos segundos.
Barra analógica: La barra de 40 segmentos indica el porcentaje del
fondo de escala en el que está midiendo en cada momento. La información
se actualiza 20 veces por segundo.
Congelación de la lectura: Pulse la tecla HOLD para mantener una
lectura congelada en el visualizador. Aparece la indicación “HLD”. Pulse de
nuevo HOLD para volver a la indicación de la lectura real.
– 9 –
Procedimientos generales de medida: Antes de conectar o des-
conectar las puntas de prueba a/de un circuito , apague siempre el dispositivo o circuito sometido a prueba y descargue todos los condensadores.
Observe estrictamente los límites máximos de entrada. No cambie de
función mientras tenga las puntas de prueba conectadas a un circuito.
F • Procédures de Mesure
Général: Mettez l’appareil sous tension en tournant le sélecteur de la
position OFF vers la fonction souhaitée.
Sélection de gammes: Automatique pour toutes les fonctions et
gammes, sauf le courant et 400mV ca. “AUTO” est affiché. La fonction et la
gamme en cours sont affichées par le point décimal et l’unité. Vous pouvez
bloquer une gamme en presssant le bouton RANGE. Continuez à presser
RANGE pour parcourir les diverses gammes. Pressez RANGE pendant deux
secondes pour retourner à la sélectiion automatique.
Bargraphe: Le bargraphe à 40-segments indique à quelle position de la
gamme (%) vous mesurez. Il est raffraichi 20 fois par seconde.
Maintien de lecture: Pressez la touche HOLD pour maintenir un
affichage. “HLD” est affiché. Pressez HOLD à nouveau pour afficher les
valeurs actuelles. La coupure automatique est désactivée pendant que
HOLD est actif.
Procédures Générales: Avant de connecter/déconnecter les câbles
de mesure à/d’un ciruit, coupez l’alimentation du circuit et déchargez les
condensateurs. Ne dépassez pas les limites d’entrée. Ne tournez pas
le sélecteur tandis que les cordons sont raccordés à un circuit.
DC AND AC VOLTAGE MEASUREMENT (See Fig. 1)
Connect test leads as shown in figure 1. Turn function switch to
Vac () or Vdc (). Touch Probe tips across voltage source (in
parallel with circuit). Voltage value will appear on Digital Display
along with the voltage polarity (DC – positive implied).
Note: The 400mV AC range needs to be selected manually. See page 8
“RANGING” for procedure.
Note for AC Measurements: Model 2015 is a True-RMS meter.
It measures the true RMS value of distorted AC voltage or current signals.
The Crest Factor handling capability is shown in table 1. The Crest Factor
is the Peak Voltage divided by the RMS voltage.
Input RMS, Percentage of Full-Scale
Effektiv-Wert Eingang, % vom Endbereich
EntrØe effective, % pleine Øchelle
Additional correction from 1.5 to 5.0
Zus tzliche Korrektur von 1.5 bis 5.0
Correction supplØm. de 1.5 5.0
– 11 –
D • Gleich- und Wechselspannungsmessung (siehe Fig.1)
Meßkabel wie in Figur 1 gezeigt verbinden. Funktionsschalter auf
VDC () oder VAC () stellen. Meßspitzen mit Meßkreis
verbinden – parallel zur Spannungsquelle. Meßwert ablesen
(automatische Polaritätsanzeige bei DC Messungen).
Anmerkung: Der 400mV AC Bereich muß manuell gewählt werden. Siehe
Seite 9 “Bereichswahl” für Prozedur.
Anmerkung für AC Messungen: Modell 2015 ist ein EchtEffektivwertmeßgerät: Es mißt den Effektivwert von verzerrten
Spannungs- und Stromsignalen. Crest Faktor Grenzen sind in Tabelle 1
angezeigt. Der Crest Factor ist die Spitzenspannung dividiert durch die
Effektivwertspannung.
E • Medidas de tensión CC y CA (vea Fig. 1)
Conecte las puntas de prueba como se muestra en la Figura 1. Ponga
el selector de función en ACV o DCV. Toque con las puntas de prueba
los puntos de tensión (en paralelo con el circuito). Aparece el valor de la
tensión en el visualizador digital, junto con la polaridad en l caso de DCV.
Nota: La escala de 400mV AC debe seleccionarse manualmente. Consulte
el procedimiento en página 9, “Selección de escalas”.
Nota para medidas de CA: El Modelo 2015 es un medidor de
verdadero valor eficaz (TRMS). Mide el verdadero valor eficaz de
señales distorsionadas de tensión o corriente CA. En la tabla 1 se muestra
su capacidad de proceso del factor de cresta. El factor de cresta es la
tensión máxima (pico) dividida por la tensión eficaz (RMS).
F • Mesure de Tensions CC et CA (voir fig. 1)
Connectez les câbles comme indiqué en figure 1. Placez le sélecteur
sur VDC () ou VAC (). Connectez les pointes de touche au
circuit – en parallèle avec la source de tension. Lisez la valeur.
Note: Le calibre 400mV ca doit être sélectionné manuellement. Voir page
10 “Changement de calibres” pour procédure.
Notes pour mesure en courant alternatif: Le modèle 2015 est un
appareil à mesure efficace vraie: Il mesure la valeur efficace de toutes
les formes d’ondes, avec ou sans distortion. Les limites de facteur de crête
sont indiquées dans le tableau 1. Le facteur de crête est la valeur crête
divisée par la valeur efficace.
– 12 –