Waldmann PULSE PIR User Manual

ENG Operating Manual
Light Control
PULSE HFMD PULSE PIR
Version IV
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Dear customer,
these operating instructions apply in addition to the instructions for use of the light. Observe in particular the safety instructions stated in the instructions for use!
The operating instructions give information on additional setting options on lights equipped with the PULSE HFMD or PULSE PIR control of version IV. The functions described here go beyond the general operating options on the light and are usually rarely required.
Please note that the instructions may also describe functions that are not available on each light.
2
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Operating units
Fig. 1
Fig. 2a
Sensor operating unit "PULSE HFMD"
Fig. 2b
Column operating unit
Sensor operating unit "PULSE PIR"
3
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
General information
Validity of these instructions
These instructions apply to the fol­lowing lights:
- Lights equipped with Pulse HFMD from firmware vers. VFP 04.5 l
- Lights equipped with PULSE PIR from firmware vers. VFS 04.4 E
Safety instructions
Make sure to read the instructions for use of the light, before putting the light into operation!
Make sure to observe the safety instructions given in the instruc­tions for use of the light!
The manufacturer cannot be held liable for damage caused by using the unit for purposes contrary to the designated use or by ignoring safety instructions and warnings.
Abbreviations and symbols:
Savety or warning information!
Important information
HFMD
High-frequency motion sensor
PIR
Passive infrared sensor
LED
Light-emitting diode
Light bar LED constantly lit
Light bar LED flashing
4
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Operation
Fig. 3 (column operating unit)
Key I
Light bar
Key II
Fig. 4 (sensor operating unit)
Key I + LED I
Key II + LED II
Sensor
Operating and display elements
All operating and setting functions be set from the column operating unit [Fig. 3] or from the sensor operating unit [Fig. 4].
In the column operating unit, the fun ctions and settings are dis­played via the light bar.
In the sensor operating unit, all fun ctions and settings are dis­played via the two LEDs.
Note: If a column operating unit is available, the LED disp lay in the sensor is automatically sup­pressed.
In-factory setting
The lights are set in-factory such that initially they can be operated without changing the settings.
If you want to adapt the light to individual requirements, you can change the parameterising func­tions accordingly (see page 10-35).
Important note!
Both LEDs on the sensor are lit in green -> Burn-in function is active.
¾ Ligh t contro l and pre sence
control are disabled.
¾ Dimming is not possible.
¾ Wait for the burn-in time (100
operating hours) to elapse.
Restoring the in-factory setting
The default settings are restored by pressing one of the two keys [I or II] when connecting the light to the mains.
The following functions are re­stored:
- Parameterising functions are set to the default setting.
- All control parameters are set to the default setting.
- All control and follow-up times are set to the default setting.
- The burn-in time is deleted (func-The burn-in time is deleted (func­tion „OFF“)
5
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
General operation
Fig. 3 Keys
Key I
Light bar
Key II
Operating elements
By pressing keys I and II, differ­ent functions can be executed, de pen d ing on whi ch key was pressed:
- Switching on and off
- Changing the light level
- Changing the basic settings
The light is switched on and off by pressing key II.
The light level can be changed by pressing key I.
Display
The light bar in the column oper­ating unit is only activated when basic settings are changed.
In normal operation, the light bar remains switched off.
In the Parameter Setting mode, the loaded function and its setting are displayed.
6
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Menu structure
Functions
ON/OFF
Changing the light level
Changing the parameterising Level 1 Light bar functions External switching Burn-In function Write protection Light control Presence control Demo mode
Level 2 Sensitivity high Automatic switch-on Special function Sensor LED Continuous mode Parameter setting protection
Special function
Level 3 Follow-up time (5, 10, 15, 20, 25, 30 min)
Level 4 Presence range (10, 20, 40, 60, 80, 100%)
Note: The functions written
in blue are not active in all versions.
7
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Switching On/Off
Fig. 3
Actuation Consequence
Switching on
The light is switched on by
Key I
Light bar
Key II
8
pressing key II.
Note: The light can also be switched on by pressing key I.
Switching off
The light is switched off by pressing key II.
1 x
1 x
Press key II
Press key II
The light is switched on to the light level set last.
The light is switched off.
The light level set last is saved automatically.
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Changing the light level (dimming)
Fig. 3
Setting a new light level
Actuation Consequence
Key I
Light bar
Key II
Th e l ight le vel ca n b e changed by pressing key I.
Press key I and hold it down.
Release key I and press it again.
After about 1 second, the light will change its intensity.
The dimming direction will change.
9
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Changing the parameterising functions
Fig. 3
Switching to the parameterisation mode
Actuation Consequence
Key I
Light bar
Key II
Fig. 6
LED 1 2 3 4 5 6
10
The parameterisation mode can be loaded by pressing both keys simultaneously for 10 seconds.
Leaving the parameterisation mode
The parameterisation mode can be left by not pressing both keys for 10 seconds.
Press both keys for at least 10 seconds
Do not press any key for at least 10 seconds
The operating unit switches to the parameterisation mode (first to setting level 1).
The LEDs valid for the level will be lit on the light bar:
LED 1 Level 1
LEDs 1+2 Level 2
LEDs 1+2+3 Level 3
LEDs 1+2+3+4 Level 4
The parameterisation mode is left.
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Changing the parameterising functions (Continued)
Fig. 3
Changing the setting level
Actuation Consequence
Key I
Light bar
Key II
Fig. 6
LED 1 2 3 4 5 6
The setti ng l evel can be loa ded by pressing bot h keys sim ul tan eo us ly for 10 seconds.
Press both keys for 5 seconds each as often as necessary for reaching the desired setting level.
n x
Release both keys
Level 2 has been reached when LEDs 1 + 2 are lit on the light bar.
Level 3 has been reached when LEDs 1 to 3 are lit on the light bar.
Level 4 has been reached when LEDs 1 to 2 are lit on the light bar.
Once the desired setting level has been reached, the setting of the first function (x.1) is shown on the light bar as soon as the keys are released (LED 1 is lit or flashes dark green).
11
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Changing the parameterising functions (Continued)
Fig. 3
Changing the function
Actuation Consequence
Key I
Light bar
Key II
Fig. 6
LED 1 2 3 4 5 6
12
The desired function can be loaded by pressing key II.
Press key II as often as necessary for reaching the desired function.
n x
The LED valid for the function will be lit or flash on the light bar:
LED 1 Function x.1
LED 2 Function x.2
LED 3 Function x.3
LED 4 Function x.4
LED 5 Function x.5
LED 6 Function x.6
LEDs 5+6 Function x.7
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Changing the parameterising functions (Continued)
Fig. 3
Changing the setting
Actuation Consequence
Key I
Light bar
Key II
Fig. 6
LED 1 2 3 4 5 6
The function can be toggled between active and inactive by pressing key I.
Press key I as often as
n x
necessary for reaching the desired setting.
The LED on the light bar is con­tinuously lit or flashes to indicate whether a function is active (ON) or inactive (OFF), for example:
- LED constantly lit:
Function active (ON)
- LED flashing:
Function inactive (OFF)
13
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
List of the parameterising functions
Fig. 3
Key I
Light bar
Key II
Fig. 6
LED 1 2 3 4 5 6
The following functions can be selected:
The bold-printed values are the in-factory defaults
Level Function Light bar (Fig. 6) Designation
1 Funct. 1.1: LED 1 (dark green) Light bar ON/OFF
1 Funct. 1.2: LED 2 (light green) External switching ON/OFF
1 Funct. 1.3: LED 3 (yellow) Burn-In function ON/OFF
1 Funct. 1.4: LED 4 (orange) Write protection ON/OFF
1 Funct. 1.5: LED 5 (light red) Light control ON/OFF
1 Funct. 1.6: LED 6 (dark red) Presence control ON/OFF
1 Funct. 1.7: LED 6+5 Demo mode ON/OFF
2 Funct. 2.1: LED 1 (dark green) Sensitivity high ON/OFF
2 Funct. 2.2: LED 2 (light green) Automatic switch-on ON/OFF
2 Funct. 2.3: LED 3 (yellow) Special function ON/OFF
2 Funct. 2.4: LED 4 (orange) Sensor LED ON/OFF
2 Funct. 2.5: LED 5 (light red) Continuous mode ON/OFF
2 Funct. 2.6: LED 6 (dark red) Parameter setting protection ON/OFF
2 Funct. 2.7: LEDs 6+5 Special function ON/OFF
3 Funct. 3.1 - 3.6: LEDs 1-6 Follow-up time 5, 10, 15,20, 25, 30 min
4 Funct. 4.1 - 4.6: LEDs 1-6 Range 10, 20, 40 ,60, 80, 100 %
Note: The functions written in blue are not active in all versions. Please note the explanations of the func-
tions on the following pages.
14
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Changing the parameterising functions (Example)
Example
How to load and change the func­tion x.y is shown using the func­tion 2.5 "Continuous mode" as an example.
a) To load the parameterisation
function:
press both keys for at least
10 seconds
b) To load setting level x:
press both keys (x-1) times
for 2 seconds
c) To load function y:
press key II (y-1) times
d) To change the function:
press key I
e) To leave the parameterisation
mode:
do not press any key for at
least 10 seconds.
Loading the parameterisation mode
Loading setting level 2
Function 2.5 Continuous mode
Changing the function from ON to OFF or vice versa
Actuation Display
Press both keys for at least 10 seconds
Press both keys again for 5 seconds
Press key II 4 times
4 x
Press key I as often as
n x
necessary for reaching the desired value
Light bar LED 1 (dark green) Parameterisation mode ON
Light bar LEDs 1+2 Level 2
Light bar LED 5 (light red) Function 2.5
LED changes from constantly lit to flashing or vice versa:
- constant = Function ON
- flashing = Function OFF
15
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Setting function 1.1
Light bar
Function for switching the light bar on and off.
By default, the function is set to “ON”. The light bar is on.
When the function is changed, the light bar remains switched off in normal operation.
The function can be changed on
- Lights equipped with HFMD sensor
- Lights equipped with PIR sensor
- Dimmable lights
Note: In lights without column operating unit, this function is not active.
Actuation Display
Loading the parameterisation mode
Loading setting level 1
Function 1.1 Light bar
Changing the function from ON to OFF or vice versa
Press both keys for at least 10 seconds
Release both keys
Jump to function 1.1 takes place automatically
Press key I as often as
n x
necessary for reaching the desired value
Light bar LED 1 (dark green) Parameterisation mode ON
Light bar LED 1 (dark green) Level 1
Light bar LED 1 (dark green) Function 1.1
LED changes from constantly lit to flashing or vice versa:
- constant = Function ON
- flashing = Function OFF
16
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Setting function 1.2
External switching
Function for operating the light via external switching elements.
By default, the function is set to "OFF". When power is supplied from an external source, the light remains switched off.
When the function is on, the light will switch to the level set last as soon as the power supply is connected (e.g. via a switched socket).
The function can be changed on
- Lights equipped with HFMD sensor
- Lights equipped with PIR sensor
- Dimmable lights
Actuation Display
Loading the parameterisation mode
Loading setting level 1
Function 1.2 External switching
Changing the function from ON to OFF or vice versa
Press both keys for at least 10 seconds
Release both keys
Press key II once
1 x
Press key I as often as
n x
necessary for reaching the desired value
Light bar LED 1 (dark green) Parameterisation mode ON
Light bar LED 1 (dark green) Level 1
Light bar LED 2 (light green) Function 1.2
LED changes from constantly lit to flashing or vice versa:
- constant = Function ON
- flashing = Function OFF
17
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Setting function 1.3
Burn-in function
Function for burning in the lamps
By default, the function is set ex works to "OFF". When the burn-in function is activated, the following functions are disabled:
- Light control
- Presence function
- Dimming function
Afte r 100 operating hours, the function is automatically set to "OFF" again, i.e., the light can now be operated without restrictions.
The function can be changed on
- Lights equipped with HFMD sensor
- Lights equipped with PIR sensor
Actuation Display
Loading the parameterisation mode
Loading setting level 1
Function 1.3 Burn-In function
Changing the function from ON to OFF or vice versa
Press both keys for at least 10 seconds
Release both keys
Press key II twice
2 x
Press key I as often as
n x
necessary for reaching the desired value
Light bar LED 1 (dark green) Parameterisation mode ON
Light bar LED 1 (dark green) Level 1
Light bar LED 3 (yellow) Function 1.3
LED changes from constantly lit to flashing or vice versa:
- constant = Function ON
- flashing = Function OFF
18
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Setting function 1.4
Write protection
Function for allowing or preventing setting changes.
By default, the function is set to "OFF", i.e., set levels can always be overwritten.
The levels cannot be changed if the write protection is active.
The function can be changed on
- Lights equipped with HFMD sensor
- Lights equipped with PIR sensor
Actuation Display
Loading the parameterisation mode
Loading setting level 1
Function 1.4 Write protection
Changing the function from ON to OFF or vice versa
Press both keys for at least 10 seconds
Release both keys
Press key II 3 times
3 x
Press key I as often as
n x
necessary for reaching the desired value
Light bar LED 1 (dark green) Parameterisation mode ON
Light bar LED 1 (dark green) Level 1
Light bar LED 4 (orange) Function 1.4
LED changes from constantly lit to flashing or vice versa:
- constant = Function ON
- flashing = Function OFF
19
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Setting function 1.5
Light control
Function for the automatic readjust­ment of the set brightness in case the ambient brightness changes.
By d efault, the function is set to "ON", i.e., the light control is functioning.
If the light control is off, the light functions as a dimmable light.
The function can be changed on
- Lights equipped with HFMD sensor
- Lights equipped with PIR sensor
Actuation Display
Loading the parameterisation mode
Loading setting level 1
Function 1.5 Light control
Changing the function from ON to OFF or vice versa
Press both keys for at least 10 seconds
Release both keys
Press key II 4 times
4 x
Press key I as often as
n x
necessary for reaching the desired value
Light bar LED 1 (dark green) Parameterisation mode ON
Light bar LED 1 (dark green) Level 1
Light bar LED 5 (light red) Function 1.5
LED changes from constantly lit to flashing or vice versa:
- constant = Function ON
- flashing = Function OFF
20
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Setting function 1.6
Presence control
Function for detecting the pres­ence of persons.
By default, the function is set to "ON", i.e., the presence control is functioning.
When the function is on, the light will be automatically switched off when no presence is detected over a certain period of time.
If the presence control is switched off, the light functions as a light with light control or as a dimmable light.
The function can be changed on
- Lights equipped with HFMD sensor
- Lights equipped with PIR sensor
Actuation Display
Loading the parameterisation mode
Loading setting level 1
Function 1.6 Presence control
Changing the function from ON to OFF or vice versa
Press both keys for at least 10 seconds
Release both keys
Press key II 5 times
5 x
Press key I as often as
n x
necessary for reaching the desired value
Light bar LED 1 (dark green) Parameterisation mode ON
Light bar LED 1 (dark green) Level 1
Light bar LED 6 (dark red) Function 1.6
LED changes from constantly lit to flashing or vice versa:
- constant = Function ON
- flashing = Function OFF
21
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Setting function 1.7
Demo mode
Function for demonstration and for testing the light functions.
By default, the function is set to "OFF", i.e., the demo mode is not functioning.
When the demo mode is on, all control times are effective within seconds.
Note: This mode is left automati­cally 10 minutes after the last key press.
The function can be changed on
- Lights equipped with HFMD sensor
- Lights equipped with PIR sensor
Actuation Display
Loading the parameterisation mode
Loading setting level 1
Function 1.7 Demo mode
Changing the function from ON to OFF or vice versa
Press both keys for at least 10 seconds
Release both keys
Press key II 6 times
6 x
Press key I as often as
n x
necessary for reaching the desired value
Light bar LED 1 (dark green) Parameterisation mode ON
Light bar LED 1 (dark green) Level 1
Light bar LEDs 5+6 Function 1.7
LEDs change from constantly lit to flashing or vice versa:
- constant = Function ON
- flashing = Function OFF
22
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
23
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Setting function 2.1
Sensitivity high
Function for changing the sensitiv­ity of the presence control.
By default, the function is set to "ON", i.e., the presence has its normal sensitivity (100%). Small movem ents are als o detect ed within the detection range of the sensor.
When the function is set to "OFF", the presence sensor works with reduced sensitivity (approx. 50%). Small movements are no longer de­tected within the detection range of the sensor.
The function can only be changed on
- Lights equipped with HFMD sensor
Actuation Display
Loading the parameterisation mode
Loading setting level 2
Function 2.1 Sensitivity high
Changing the function from ON to OFF or vice versa
Press both keys for at least 10 seconds
Press both keys again for 5 seconds
Release both keys
Press key I as often as
n x
necessary for reaching the desired value
Light bar LED 1 (dark green) Parameterisation mode ON
Light bar LEDs 1+2 Level 2
Light bar LED 1 (dark green) Function 2.1
LED changes from constantly lit to flashing or vice versa:
- constant = Function ON
- flashing = Function OFF
24
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Setting function 2.2
Automatic switch-on
Function for allowing or suppress­ing automatic switch-on when light is insufficient or in case of presence (semi-automatic).
By default, the function is set to "ON", i.e., the light is automatically switched on in case of presence or when light is insufficient.
When automatic switch-on is off, the light has to be switched on manually even in case of pres­ence.
The function can be changed on
- Lights equipped with HFMD sensor
- Lights equipped with PIR sensor
Actuation Display
Loading the parameterisation mode
Loading setting level 2
Function 2.2 Automatic switch-on
Changing the function from ON to OFF or vice versa
Press both keys for at least 10 seconds
Press both keys again for 5 seconds
Press key II once
1 x
Press key I as often as
n x
necessary for reaching the desired value
Light bar LED 1 (dark green) Parameterisation mode ON
Light bar LEDs 1+2 Level 2
Light bar LED 2 (light green) Function 2.2
LED changes from constantly lit to flashing or vice versa:
- constant = Function ON
- flashing = Function OFF
25
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Setting function 2.3
Special function
The special function cannot be used.
The function cannot be changed.
Actuation Display
Loading the parameterisation mode
Loading setting level 2
Function 2.3 Special function
Changing the function from ON to OFF or vice versa
Press both keys for at least 10 seconds
Press both keys again for 5 seconds
Press key II twice
2 x
Press key I as often as
n x
necessary for reaching the desired value
Light bar LED 1 (dark green) Parameterisation mode ON
Light bar LEDs 1+2 Level 2
Light bar LED 3 (yellow) Function 2.3
LED changes from constantly lit to flashing or vice versa:
- constant = Function ON
- flashing = Function OFF
26
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Setting function 2.4
Sensor LED
Function for switching both LEDs in the operating unit sensor on and off (see Fig. 4).
By default, the function is set to "OFF", i.e., the LEDs in the light head are off.
For lights without column operating unit, the function is set to "ON".
If the function is on, the LEDs in the light head are lit in addition to the light bar display.
The function can be changed on
- Lights equipped with HFMD sensor
- Lights equipped with PIR sensor
Actuation Display
Loading the parameterisation mode
Loading setting level 2
Function 2.4 Sensor LED
Changing the function from ON to OFF or vice versa
Press both keys for at least 10 seconds
Press both keys again for 5 seconds
Press key II 3 times
3 x
Press key I as often as
n x
necessary for reaching the desired value
Light bar LED 1 (dark green) Parameterisation mode ON
Light bar LEDs 1+2 Level 2
Light bar LED 4 (light green) Function 2.4
LED changes from constantly lit to flashing or vice versa:
- constant = Function ON
- flashing = Function OFF
27
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Setting function 2.5
Continuous operation
Function for setting the light to continuous mode.
By default, the function is set to "OFF".
When the function is switched on, the light will work in continu­ous mode. Even if no presence is detected, the light will be lit at the lowest level.
The function can be changed on
- Lights equipped with HFMD sensor
- Lights equipped with PIR sensor
Actuation Display
Loading the parameterisation mode
Loading setting level 2
Function 2.5 Continuous mode
Changing the function from ON to OFF or vice versa
Press both keys for at least 10 seconds
Press both keys again for 5 seconds
Press key II 4 times
4 x
Press key I as often as
n x
necessary for reaching the desired value
Light bar LED 1 (dark green) Parameterisation mode ON
Light bar LEDs 1+2 Level 2
Light bar LED 5 (light red) Function 2.5
LED changes from constantly lit to flashing or vice versa:
- constant = Function ON
- flashing = Function OFF
28
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Setting function 2.6
Parameter setting protection
Function for activating the param­eter setting protection (required for final inspection)
By default, the function is set to "OFF".
When the function is switched on, a factory reset will not overwrite the set parameters. The parameter setting of the electronic ballast units will be carried out.
The function can be changed on
- Lights equipped with HFMD sensor
- Lights equipped with PIR sensor
Actuation Display
Loading the parameterisation mode
Loading setting level 2
Function 2.6 Parameter setting protection
Changing the function from ON to OFF or vice versa
Press both keys for at least 10 seconds
Press both keys again for 5 seconds
Press key II 5 times
5 x
Press key I as often as
n x
necessary for reaching the desired value
Light bar LED 1 (dark green) Parameterisation mode ON
Light bar LEDs 1+2 Level 2
Light bar LED 6 (dark red) Function 2.6
LED changes from constantly lit to flashing or vice versa:
- constant = Function ON
- flashing = Function OFF
29
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Setting function 2.7
Special function
The special function cannot be used.
The function cannot be changed.
Actuation Display
Loading the parameterisation mode
Loading setting level 2
Function 2.7 Special function
Changing the function from ON to OFF or vice versa
Press both keys for at least 10 seconds
Press both keys again for 5 seconds
Press key II 6 times
6 x
Press key I as often as
n x
necessary for reaching the desired value
Light bar LED 1 (dark green) Parameterisation mode ON
Light bar LEDs 1+2 Level 2
Light bar LEDs 5+6 Function 2.7
LED changes from constantly lit to flashing or vice versa:
- constant = Function ON
- flashing = Function OFF
30
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
31
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Setting functions 3.1 to 3.6
Follow-up time
Function for setting the time after which the light is to be switched off if no presence is detected.
By default, the follow-up time is set to 10 min.
6 different follow-up times from 5 to 30 min can be set.
The function can be changed on
- Lights equipped with HFMD sensor
- Lights equipped with PIR sensor
Actuation Display
Loading the parameterisation mode
Loading setting level 3
Show follow-up time
Press both keys for at least 10 seconds
Press both keys again twice for 5 seconds each
2 x
Release both keys
Light bar LED 1 (dark green) Parameterisation mode ON
Light bar LEDs 1+2+3 Level 3
5 min
10 min
15 min
20 min
25 min
30 min
Light bar LEDs 1-6 some are flashing
32
Continued >
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Setting functions 3.1 to 3.6 (Continued)
Follow-up time
(Continued)
Actuation Display
Changing the function - Increasing the follow-up time step-by-step
Press key I as often as
n x
necessary for reaching the desired value
Changing the function - Decreasing the follow-up time step-by-step
Light bar LEDs 1-6 some are flashing Follow-up time level 3 (15 min)
Each LED that is not flashing stands for a follow-up time of 5 min.
5 min
10 min
15 min
20 min
25 min
30 min
Press key II as often as necessary for reaching the desired value
n x
Light bar LEDs 1-6 some are flashing Follow-up time level 1 (5 min)
33
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Setting functions 4.1 to 4.6
Presence range
Function for setting the detection range of the presence sensor.
By default, the range for the pres­ence sensor is set to 80%.
6 different levels from 10% to 100% can be set.
The function can only be changed on
- Lights equipped with HFMD sensor
Note: For lights equipped with PIR
sensor, the range can be changed by turning the sensor ball (see in­structions for use of the light.
Actuation Display
Loading the parameterisation mode
Loading setting level 4
Displaying the presence range
Press both keys for at least 10 seconds
Press both keys again three times for 5 seconds each
3 x
Release both keys
Light bar LED 1 (dark green) Parameterisation mode ON
Light bar LEDs 1+2+3+4 Level 4
10 %
20 %
40 %
60 %
80 %
100 %
Light bar LEDs 1-6 some are flashing
34
Continued >
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
Setting functions 4.1 to 4.6 (Continued)
Presence range
(Continued)
Actuation Display
Changing the function - Increasing the presence range step-by-step
Press key I as often
n x
as necessary for reaching the desired value
Changing function - Decreasing presence range step-by-step
Press key II as often as necessary for reaching the
n x
desired value
Light bar LEDs 1-6 some are flashing Presence range level 6 (100 %)
Light bar LEDs 1-6 some are flashing Presence range level 4 (60 %)
10 %
20 %
40 %
60 %
80 %
100 %
35
ENG
Operating Manual Light Control „PULSE HFMD / PULSE PIR - Vers. IV“
36
Order No. 405 262 110 - © copyright H.Waldmann GmbH & Co. KG - nB/04/2011 - TB - Subject to change - Printed in Germany
Edition: 2011/04/18
Loading...