Caution: Federal (United States) law restricts this
device to sale by or on the order of a physician.
Page 4
CONTACT INFORMATION
To obtain additional information regarding your AortaScan system,
please contact Verathon® Customer Care or visit verathon.com/contact‑us.
Verathon Inc.
20001 North Creek Parkway
Bothell, WA 98011 U.S.A.
800.331.2313 (US and Canada only)
425.867.13 48
Fax: 425.883.2896
verathon.com
Verathon Medical (Europe) B.V.
Willem Fenengastraat 13
1096 BL Amsterdam
The Netherlands
Tel: +31 (0) 20 210 30 91
Fax: +31 (0) 20 210 30 92
verathon.com
0123
Copyright 2009, 2015 Verathon Inc. All rights reserved. No part of this manual may be copied or transmitted by any method without
the express written consent of Verathon Inc.
AortaScan, the AortaScan symbol, Vmode, Verathon, and the Verathon Torch symbol are trademarks of Verathon Inc. Bluetooth® word
mark and logos are owned by the Bluetooth SIG, Inc. and any use of such marks by Verathon is under license. All other brand and
product names are trademarks or registered trademarks of their respective owners.
Information in this manual may change at any time without notice. For the most up‑to‑date information, see the documentation
available at verathon.com.
Page 5
TABLE OF CONTENTS
IMPORTANT INFORMATION ................................................................................................................................. 1
Notice to All Users ...........................................................................................................................................2
Statement of Prescription .................................................................................................................................2
Statement of Intended Use ..............................................................................................................................2
Safety Information ..............................................................................................................................................3
Ultrasound Energy Safety ................................................................................................................................. 3
Components & Features ......................................................................................................................................8
Operations & Maintenance Manual: Table of Contents
i
Page 6
SETTING UP ......................................................................................................................................................... 33
Procedure 1. Perform the Initial Inspection ...............................................................................................33
Procedure 2. Set Up the Battery ...............................................................................................................34
Procedure 3. Attach the Probe to the Console ..........................................................................................36
Procedure 4. Program the Clinic Name .....................................................................................................37
Procedure 5. Set the Date & Time .............................................................................................................39
Procedure 6. Load the Thermal Paper .......................................................................................................40
Procedure 7. Attach the Instrument to a Medical Cart (Optional) ..............................................................41
Procedure 8. Install ScanPoint with QuickPrint (Optional) .........................................................................43
Procedure 9. Watch the Training Video .....................................................................................................43
USING THE DEVICE .............................................................................................................................................44
Procedure 1. Prepare for the Exam ...........................................................................................................46
Help Resources .................................................................................................................................................64
Bluetooth Wireless Technology .......................................................................................................................... 72
Accessory Conformance to Standards.............................................................................................................75
SYMBOL DIRECTORY ..........................................................................................................................................76
Operations & Maintenance Manual: Table of Contents
iii
Page 8
Page 9
IMPORTANT INFORMATION
OVERVIEW
PRODUCT DESCRIPTION
The AortaScan AMI9700 is a portable ultrasound instrument that provides noninvasive measurements of
abdominal aortic diameter. The device consists of an ultrasound probe that scans the patient’s aorta, and a
compact, battery‑operated console that provides measurement‑related information.
The AortaScan AMI9700 can measure diameters ranging between 3 and 12.4 cm with a diameter accuracy
of ±(15%+0.5cm). This error‑range data (Table 1) indicates a range of values obtained by the device
relative to follow up and clinical significance, specifically with respect to risk vs. diameter.
Table 1. Expected Aortic Measurement Ranges
ACTUAL AORTIC DIAMETER
3.0 cm3.5 cm4.0 cm4 .1 cm5.0 cm5.3 cm
Average
estimated risk
of rupture for
actual aortic
diameter
Aortic diameter as reported by the device based on allowable tolerances
± 15%2.553.452.984.033.404.603.494.724.255.754.516.10
With additional
±0.5 cm
Average
estimated risk
of rupture for
reported aortic
diameter
0%0%0%1%11%11%
MinMaxMinMaxMinMaxMinMaxMinMaxMinMax
2.053.952.484.532.905.102.995.223.756.254.016.60
0%0%0%1%0%1%0%11%0%26%
0.5–
5.0%
26%
Note: The AortaScan AMI9700 is not intended to detect, identify, screen for, or diagnose abdominal aortic
aneurysms (AAAs).
The AortaScan AMI9700 is quick and easy to use. When the user releases the scan button, the AortaScan
AMI9700 uses patented Vmode® technology to measure ultrasound reflections on multiple planes inside
the body and produces a three‑dimensional image within seconds. Based on this image, the AortaScan
AMI9700 calculates and displays the approximate abdominal aortic diameter. A sonographer is not required.
If needed, after a scan has been taken, a unique aiming icon guides the operator to optimal probe placement
with a comprehensive, three‑dimensional display showing the aorta in two cross‑sectional images verifying
that a complete scan has been achieved.
Operations & Maintenance Manual: Important Information
1
Page 10
AortaScan AMI9700 measurements can be printed via an onboard thermal printer or transmitted, via
a proprietary wireless connection, to a personal computer running HIPAA‑compliant ScanPoint® with
QuickPrint software. ScanPoint with QuickPrint software enables the user to print detailed reports for medical
records or reimbursement, calibrate the instrument (requires a calibration kit), download and install software
updates, and archive data on Verathon® servers.
The optional calibration kit, consisting of a spiral‑shaped calibration target along with a special calibration
container, allows the user to easily calibrate the device at their facility by scanning a known target.
The AortaScan AMI9700 system also includes a battery charger/wireless hub for the custom,
user‑replaceable lithium‑ion battery that mounts in the console.
The AortaScan AMI9700 may be mounted on a cart which holds the instrument securely in place and
provides a holder for ultrasound gel and other accessories.
NOTICE TO ALL USERS
The AortaScan AMI9700 should be used only by individuals who have been trained and authorized by a
physician or the institution providing patient care. All users must read this entire manual prior to using the
AMI9700. Do not attempt to operate this instrument until you thoroughly understand all instructions and
procedures in this manual. Failure to comply with these instructions may compromise the performance of the
device and the reliability of its measurements.
For the most current version of this manual, please visit verathon.com.
STATEMENT OF PRESCRIPTION
Caution: Federal (United States) law restricts this device to sale by or on the order of a physician.
STATEMENT OF INTENDED USE
The AortaScan AMI9700 is an ultrasound device that projects ultrasound energy into the mid‑abdomen to
obtain an image of the abdominal aorta for aortic diameter measurements.
ESSENTIAL PERFORMANCE
Essential performance is the system performance necessary to achieve freedom from unacceptable risk. The
essential performance of the AortaScan AMI9700 system is to produce ultrasonic output energy, display
ultrasonic images, and display numerical values for aortic diameter. The system has a temperature‑controlled
transducer assembly.
2
Page 11
SAFETY INFORMATION
ULTRASOUND ENERGY SAFETY
To date, exposure to pulsed diagnostic ultrasound has not been shown to produce adverse effects. However,
ultrasound should be used prudently, and total patient exposure should be kept as low as reasonably achievable (ALARA). Following the ALARA principle, ultrasound should only be used by medical professionals
when clinically indicated, using the lowest possible exposure times necessary to obtain clinically useful
information. For more information on ALARA, please refer to the American Institute of Ultrasound in
Medicine publication, Medical Ultrasound Safety.
The ultrasound output power of the system is not user adjustable and is limited to the minimum
level necessary for effective performance. Data on acoustic output levels can be found in the Product
Specifications section of this manual.
CONTRAINDICATIONS
The AortaScan AMI9700 is not intended for fetal use or for use on pregnant patients.
The AortaScan AMI9700 is not intended for acute events such as aortic dissection, ulcer, or rupture.
PRECAUTIONS & WARNINGS
Warnings indicate that injury, death, or other serious adverse reactions may result from use or misuse of
the device. Cautions indicate that use or misuse of the device may cause a problem, such as a malfunction,
failure, or damage to the product. Throughout the manual, pay attention to sections labeled Important, as
these contain reminders or summaries of the following cautions as they apply to a specific component or
usesituation. Please heed the following warnings and cautions.
PRECAUTIONS
CAUTION
The AortaScan AMI9700 and related devices may contain mineral oils, batteries, and other
environmentally hazardous materials. When the instrument and/or accessories have reached the
end of their useful service life, see the section Device Disposal on page63.
CAUTION
When using the AortaScan AMI9700 with optional ScanPoint® software, your computer must
be minimally certified to EN / IEC / CSA / UL60950 or 60101‑1 standards. This configuration
ensures that compliance to the EN/IEC 60601‑1‑1 system standard is maintained. Anyone
connecting additional equipment to the AortaScan AMI9700 signal input port or signal
output port configures a medical system, and is therefore responsible for ensuring that the
system complies with EN/IEC 60601‑1‑1. If you need assistance, contact your biomedical staff,
Verathon
®
representative, or Verathon Customer Care.
Operations & Maintenance Manual: Important Information
3
Page 12
CAUTION
Potential Device Interference. Bluetooth and Wireless LAN devices operate within the same
radio frequency range and may interfere with one another.
If you are using the AortaScan AMI9700 Bluetooth link and wireless LAN devices
simultaneously, you may experience less‑than‑optimal network performance or even lose
your network connection. If this happens, you may need to move the AortaScan AMI9700
and ScanPoint
®
host computer to an area away from the 2.4 GHz wireless LAN devices (40
meters/44 yards, or more).
CAUTION
Medical electrical equipment requires special precautions regarding electromagnetic
compatibility (EMC) and must be installed and operated according to the instructions in this
manual. For more information, see the Electromagnetic Compatibility section on page72.
To maintain electromagnetic interference (EMI) within certified limits, the AortaScan AMI9700
system must be used with the cables, components, and accessories specified or supplied
by Verathon
®
. For additional information, see the System Components & Accessories and
Component Specifications sections. The use of accessories and/or cables other than those
specified or supplied may result in increased emissions and/or decreased immunity of the
system.
The AortaScan AMI9700 system should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the system should be observed to verify
normal operation in the configuration in which it will be used.
This device can radiate radio frequency energy and is very unlikely to cause harmful interference
with other devices in the vicinity. There is no guarantee that interference will not occur in a
particular installation. Evidence of interference may include degradation of performance in
this device or other devices when operated simultaneously. If this occurs, try to correct the
interference by using the following measures:
• Turn devices on and off in the vicinity to determine the source of interference
• Reorient or relocate this device or other devices
• Increase the separation between devices
• Connect the device to an outlet on a circuit different than the other device(s)
• Eliminate or reduce EMI with technical solutions (such as shielding)
• Purchase medical devices that comply with IEC 60601‑1‑2 EMC Standards
Be aware that portable and mobile radio frequency communications equipment (cellular phones,
etc.) may affect medical electrical equipment; take appropriate precautions during operation.
4
Page 13
WARNINGS
CAUTION
Use of the following cleaning methods or solutions may cause device damage not covered by
the AortaScan AMI9700 warranty.
• Do not immerse the instrument in disinfectant solution.
®
• Do not use Cidex Plus
to disinfect the instrument. Cidex Plus will damage the plastic enclosure.
• Do not subject any part of the instrument to steam sterilization or ethylene oxide sterilization.
WARNING
The aortic diameter measurement function provides images that may be used for diagnosis and
screening. If clinically indicated, appropriate patients should be referred for additional diagnostic
testing.
WARNING
The AortaScan system is designed to detect the fluid (blood) filled region of the abdominal
aorta only. The system cannot detect the presence of a blood clot (thrombus) and therefore may
provide a false negative result.
WARNING
The AortaScan system is an ultrasound‑based device and is subject to all limitations of this
method. If clinically indicated, appropriate patients should be referred for a diagnostic standard
(confirmatory) test, regardless of test results.
WARNING
Risk of explosion. If you use the AortaScan AMI9700 in the presence of flammable
anesthetics, the hazard of potential explosion exists.
WARNING
Risk of electric shock or burns. Do not use the AortaScan instrument in conjunction with HF
surgical equipment.
WARNING
Ensure proper distance from patient. The AMI9700 battery charger/wireless hub and the
computer used to access online ScanPoint
®
image archives (if used) must be placed outside the
patient vicinity (more than 2m [6ft] from the patient).
Operations & Maintenance Manual: Important Information
5
Page 14
WARNING
Risk of explosion, fire, or serious injury. The AortaScan AMI9700 is powered by a lithium‑ion
battery. Failure to note the following when handling the battery may result in seriousinjury:
• Never short‑circuit the battery by either accidentally or intentionally bringing the battery
terminals into contact with any other conductive object. This could cause serious injury or fire
and could also damage the battery and/or the instrument.
• Never expose the battery to abnormal shock, vibration, or pressure. The battery’s internal
protective covering could fail, causing it to overheat or ignite, resulting in caustic liquid
leakage, explosion, or fire.
• Do not disassemble, heat above 60°C (140°F), or incinerate the battery. Keep battery out
of reach of children and in original package until ready to use. Dispose of used batteries
promptly according to local recycling or waste regulations.
• If the battery is leaking or its case is cracked, put on protective gloves to handle it, and
discard it immediately. Always dispose of used batteries in compliance with all applicable
laws and regulations. Put insulating tape, such as cellophane tape, on the electrodes during
transportation in order to avoid a possible short circuit, fire, or electrical shock.
WARNING
Do not use the AortaScan AMI9700 on:
• A patient who has open skin or wounds in the mid‑abdominal area.
• A patient with ascites.
• A pregnant patient.
WARNING
Potential patient hazard. To date, exposure to low‑power, pulsed diagnostic ultrasound has not
been shown to produce adverse effects. However, medical professionals should use ultrasound
only when clinically indicated, using the lowest exposure times possible to obtain proper
measurements. The ultrasonic output of the AortaScan AMI9700 is not user adjustable and is
limited to the minimum level necessary for effective performance. For more information about the
acoustic output levels of this device, see the section Product Specifications on page68.
WARNING
Ensure that you follow the manufacturer’s instructions for handling and disposing of the
cleaning and disinfection solutions provided in this manual.
6
Page 15
WARNING
Risk of inaccurate measurements/results. When using the instrument, be aware of the
following conditions that can decrease the accuracy of exam results:
• In some cases, the normal operating tolerances of the instrument can cause it to report a
falsely normal or abnormal measurement. For more information, see Interpret the Aortic
Measurement Results on page51.
• Visual verification that the aorta position is fully within the scan cone on the displayed
images is important.
• A thrombus (blood clot) can complicate aortic measurements. A soft, blood‑like thrombus
may appear as part of the lumen. However, a calcified thrombus may appear as part of the
aorta’s wall, resulting in a measurement of lumen diameter that is smaller than the aorta
diameter. Accordingly, in patients where thrombus is known or suspected, other imaging
methods should be used prior to ruling out an aneurysm.
• Use care when scanning patients who have had abdominal surgery. Scar tissue, surgical
incisions, sutures, and staples can affect ultrasound transmission and accuracy.
• Ensure that the patient fasts for 12 hours prior to undergoing an aortic diameter
measurement in order to minimize the presence of bowel gas, which may obstruct proper
measurement.
• Obesity may affect ultrasound aortic diameter measurements. For more information, see
Obesity on page54.
Accuracy is compromised if the user does not obtain an optimal, repeatable image.
WARNING
Cleaning is critical to ensuring the component is ready for disinfection. Failure to
properly clean the device could result in a contaminated instrument after completing the
disinfectionprocedure.
WARNING
This product may only be cleaned and disinfected by using the approved processes provided in
this manual. Cleaning and disinfection methods listed are recommended by Verathon
®
based on
compatibility with component materials.
WARNING
Availability of cleaning and disinfection products varies by country, and Verathon is unable
to test products in every market. For more information, please contact Verathon Customer
Care at 1.800.331.2313 or your local representative. For additional contact information, visit
verathon.com/contact‑us.
Operations & Maintenance Manual: Important Information
7
Page 16
INTRODUCTION
COMPONENTS & FEATURES
The AortaScan AMI9700 is designed for simple, intuitive operation. However, to ensure safe and effective
operation, before using the device:
• Familiarize yourself with the contents of this manual.
• Watch the training video provided on the instrument.
The AortaScan AMI9700 has two main components: the console and the probe. The console and probe are
linked by a detachable cable.
Figure 1. AortaScan AMI9700 System Components
Console
Battery packs
Battery charger/
wireless hub
Probe
8
Page 17
PROBE COMPONENTS
The probe transmits and receives ultrasound waves, automatically moving its internal probe 360º to scan
twelve different planes to produce a three‑dimensional image of the aorta. The probe is attached to the
console by a cable.
Figure 2. Probe Components
Microphone (above aiming display)
Aiming display
Scan button
Table 2. Probe Components
PART NAMEPURPOSE
Scan buttonWhen pressed, a scan is performed.
Aiming displayDisplays directional arrows to ensure the aorta is centered within the scanning cone.
MicrophoneRecords voice annotations.
Operations & Maintenance Manual: Introduction
9
Page 18
CONSOLE COMPONENTS
The console provides all operating controls for the scanning process by means of five variable‑function
buttons. The measured aortic diameter and target‑shaped aiming icons are clearly displayed on the LCD
screen. The console also provides controls for adjusting brightness and volume, turning the power on/off,
interfacing with a ScanPoint® host computer (optional), and adjusting user settings and preferences. The
console also houses the battery and the printer.
Figure 3. Console Components
Main display
Battery pack
Five variable
function buttons
Brightness button
Power button
Battery status indicator
Printer (behind display)
Volume but ton
Table 3. Console Components
PART NAMEPURPOSE
Main display
Displays the aortic volume measurement, patient type, settings, and instrument
status.
Power on/offToggles main power on/off.
Volume
Adjusts volume up/down on voice annotation playback, start up sound, and “scan
complete” tone.
Provides access to all instrument functions for scanning, recording annotations,
printing, connecting to ScanPoint® (optional), accessing the training video, and
setting user preferences.
Printer/printer doorPrints the scan results.
10
Page 19
BATTERY CHARGER/WIRELESS HUB
The AortaScan AMI9700 is powered by a lithium‑ion battery. The battery charger provided with the
AMI9700 can charge two lithium‑ion batteries while simultaneously functioning as the wireless hub linking
the AMI9700 to the ScanPoint® host computer. A battery icon on the instrument display is always present
indicating how much power remains and when the battery needs to be changed. The user can change the
battery whenever necessary. Removing a discharged battery and replacing it with a fully charged battery will
not erase any saved exams or user settings.
To provide power to the batteries, the battery charger/wireless hub must be plugged into a wall outlet by
using the power cord provided. Use only the battery charger provided with the AMI9700. Any other battery
charger may damage the battery. The battery charger automatically detects whether a lithium‑ion battery is
being charged.
To provide wireless communication between the AMI9700 and the ScanPoint host computer, plug the
battery charger/wireless hub USB connector into a USB port on the ScanPoint host computer. The battery
charger/wireless hub maintains an operating distance of up to 120 feet (36 meters) between the ScanPoint
computer and the AMI9700, regardless of barriers such as walls, ceilings, or windows.
Note: Use of ScanPoint with QuickPrint software is optional.
Figure 4. Battery Charger/Wireless Hub
Table 4. Battery Charger/Wireless Hub Components
PART NAMEDESCRIPTION
Battery charger/
wireless hub
Lithium‑ion batteriesWhen charged, provides power to the AMI9700 device.
Power cordConnects the battery charger/wireless hub to the wall outlet.
Wireless hub USB
cable
Operations & Maintenance Manual: Introduction
Charges the lithium‑ion batteries and receives and sends information to/from the
AMI9700 instrument within communication range.
Connects the battery charger/wireless hub to the ScanPoint host computer.
11
Page 20
SYSTEM COMPONENTS & ACCESSORIES
Table 5. Components and Accessories
COMPONENTS
AMI9700 console
AMI9700 probe
Battery charger/wireless hub with AC power cord
ACCESSORIES
Lithium‑ion battery (2 provided)
AortaScan AMI9700 In‑Service CD or USB, containing the Operations & Maintenance Manual
Thermal paper roll for the printer
Ultrasound gel
Mobile cart (Optional)
Universal accessory basket (Optional)
ScanPoint® with QuickPrint software install CD (Optional)
ScanPoint with QuickPrint user’s manual (Optional)
Calibration kit (Includes calibration container, calibration target, etc.) (Optional)
To order any of the above parts or accessories, contact your authorized Verathon® sales representative or
contact Verathon Customer Care.
12
Page 21
ICONS & BUTTONS
The console LCD presents user information and prompts that vary depending on the current device function.
The five buttons below the display have variable functions according to device mode. Button functions are
indicated by icons in the display footer, immediately above each button.
CONSOLE DISPLAY ICONS
The following icons may appear on the console main display.
ICONPURPOSE
A fully charged battery.
A battery 50% to 75% charged.
A battery 25% to 50% charged.
An almost depleted battery. Can power a few more scans.
A fully discharged battery. Replace immediately.
Empty exam folder
Current exam folder
Saved exam folder
(Solid arrow) The aorta is contained within the image cone (cone‑shaped area in which
the probe transmits ultrasound waves), but the presence of bowel gas prevents a
proper measurement. You may be able to obtain a more appropriate measurement by
moving the probe 1/2 to 1 inch (1 to 2 cm) in the direction indicated by the arrow.
(Flashing arrow) The aorta is obstructed by bowel gas and the probe needs to be
repositioned. The arrow shows the direction to move or tilt the probe to improve the
measurement. Try moving the probe 1/2 to 1 inch (1 to 2 cm) in the direction of the
arrow.
Operations & Maintenance Manual: Introduction
13
Page 22
VARIABLE BUTTON FUNCTIONS
ICONPURPOSE
Go to the Home screen.
View the training video.
Go to the Settings screen.
Go to the Review screen. If there are no saved exams, this button is disabled.
Initiate communication with the ScanPoint® host computer. Saved and annotated exams
will be automatically uploaded to the host computer.
Note: ScanPoint software must be previously installed and the computer connected to
the wireless hub. Use of the ScanPoint software is optional.
Record a voice annotation.
Play a previously recorded voice annotation. If no voice annotations are recorded, this
button is disabled.
Print exam results from the onboard printer. While printing is in progress, an hourglass
icon appears on the display, and most buttons are disabled.
Move down an item.
Move up an item.
Move right an item.
Delete an exam or cancel the current action.
Select the highlighted item.
Stop recording a voice annotation.
14
Play video playback.
Pause video playback.
Page 23
ICONPURPOSE
Add and/or toggle characters, as appropriate.
Remove and/or toggle characters, as appropriate.
Select the Axis on which to manually measure the aorta.
Go to the manual measurement tool and select appropriate cursor while manually
measuring the aorta.
No function.
Operations & Maintenance Manual: Introduction
15
Page 24
BUTTON FUNCTIONS FOR EACH DISPLAY SCREEN
The Power, Brightness, and Volume buttons are constant buttons on the body of the console and can be
pressed at any time. The five buttons below the LCD have variable functions according to device mode. The
Scan button is located on the underside of the probe.
SCREEN/MODEACTIVE BUTTONS
(1) No function.
Home screen
Appears when the instrument is turned on and set
to the AortaScan mode.
Scan screen
Appears when the operator presses and releases
the Scan button.
As the aortic diameter is calculated, the display
refreshes and updates until the scan is complete.
Results screen
Appears when a scan is complete. Prominently
displays calculated aortic diameter in centimeters,
the ultrasound image of the scan, and available
memory. An hourglass icon appears when the
device is printing.
(2) Tutorial: opens Tutorial screen.
(3) Settings: opens Settings screen.
(4) Review: opens the Review screen.
(5) ScanPoint®: transmits saved exams to ScanPoint.
Scan button: Press and release to take a scan.
(1) ‑ (4): No function.
(5): Home: Return to Home screen.
(1) Record: press to record, changes to a stop button
during recording.
(2) Print: print to onboard printer.
(3) Listen: press to listen to the voice annotations. If
no scans are saved, the button is disabled.
(4) If in B‑mode: Manual Measure: allows for a
manual measurement of the aorta.
(4) If in C‑mode: Review: opens the Review screen.
Review screen
Allows the user to review saved exams. Saved
exam folders are on the left side of the screen with
the currently selected saved exam being an open
folder icon. The ultrasound images associated with
selected exam are on the main display.
Tutorial screen
View the training modules menu.
16
(5) Home: return to Home screen.
(1) Down Arrow: select the next saved exam.
(2) Print: print to onboard printer.
(3) Listen: replay voice annotation for selected exam.
(4) Delete: delete selected exam.
(5) Home: return to Home screen.
(1) Down Arrow: skip to next video.
(2) Up Arrow: select previous video.
(3) Select: play selected video.
(4) No function.
(5) Home: return to Home screen.
Page 25
SCREEN/MODEACTIVE BUTTONS
Video Viewing screen
Plays the selected tutorial video.
Settings screen
Start screen for editing clinic name, date and time,
general preferences, savings preferences, and self
test options.
(1) No function.
(2) Play: plays selected video, changes to a pause
button when video is playing.
(3) Up Arrow: return to Tutorial screen.
(4) No function.
(5) Home: return to Home screen.
(1) Down Arrow: select next setting in list.
(2) Up Arrow: select previous setting in list.
(3) Select: proceed to the selected screen.
(4) No function.
(5) Home: return to Home screen.
(1) Down Arrow: move to the character below.
Name screen
Displays alpha numeric characters for entering
information.
Date and Time screen
Allows the user to set the date and time.
General Preferences screen
List of available settings and their current values.
Self Test screen
Displays test progress and results. It begins testing
as soon as Self Test option is selected.
ScanPoint® screen
Displays status information about the ScanPoint
communication.
(2) Right Arrow: move to the character to the right.
(3) Plus Sign: add currently selected character.
(4) Minus Sign: delete currently selected character.
(5) Settings: return to main settings screen.
(1) Down Arrow: move forward to next editable
unit.
(2) Up Arrow: move back to previous editable unit.
(3) Plus Sign: add/toggle units.
(4) Minus Sign: decrease/toggle units.
(5) Settings: save current date/time entries and
return to main settings screen.
(1) Down Arrow: select next setting in list.
(2) Up Arrow: select previous setting in list.
(3) Plus Sign: select next option.
(4) Minus Sign: select previous option.
(5) Settings: return to main settings screen.
(1) ‑ (4) No function.
(5) Settings: go back to main settings screen.
(1) ‑ (3) No function.
(4) Cancel: cancels connection to ScanPoint.
Note: Available only when ScanPoint is installed on
instrument.
Operations & Maintenance Manual: Introduction
(5) No function.
17
Page 26
DISPLAY SCREENS
HOME SCREEN
The Home screen appears when the AortaScan is first powered on. It serves as a starting point for all of the
main functions of the device.
The Home screen displays:
• In the header: Your clinic’s name, the battery status indicator and the current date and time.
• On the left side: A list of saved exam results (10 maximum) saved in chronological order. Yellow
folders hold saved exams. Grey folders represent empty spaces still available for saving exam results.
Table 6. Battery Power Level
BAT TERY ICONPOWER LEVEL
Indicates a fully charged battery.
Indicates a battery 50% to 75% charged.
Indicates a battery 25% to 50% charged.
Indicates a battery is almost depleted. Can power a few more scans.
Indicates that a battery should be replaced immediately.
18
Page 27
Table 7. Home Screen Button Functions
BUTTONFUNCTION
SCAN SCREEN
The Scan screen appears after you press the Scan button on the probe and displays a progressively updating
image of the aorta outline. When the ultrasound measurement is complete, the Results screen opens
automatically. Four buttons below the display do not function during the scan.
No function.
View the training video.
Go to the Settings screen (set the time, date, institution name, and user preferences).
Review a previously saved exam.
Initiate communication with the ScanPoint® host computer. Saved and annotated exams will
be automatically uploaded to the host computer.
Note: ScanPoint software must be previously installed, and the computer must be connected
to the wireless hub. Use of the ScanPoint software is optional.
Operations & Maintenance Manual: Introduction
19
Page 28
RESULTS SCREEN
The Results screen appears automatically when an ultrasound scan is complete. The display presents the
result of the exam: crosshairs, aorta outline, and the calculated aorta diameter. You may choose to print this
result to the onboard printer and/or to record a voice annotation to save the exam. After the annotation is
recorded, the Play and Review buttons become active, and the newly recorded exam appears on the main
display as an yellow folder icon.
Depending on the Print Report Type selected (see General Preferences Screen), the Results screen will show
one of two displays.
Figure 5. B‑Mode Report Type
Table 8. Results Screen Button Functions
BUTTONFUNCTION
Record a voice annotation (up to 10 seconds long).
Figure 6. C‑Mode Report Type
20
Print exam results to the onboard printer.
Play a previously recorded voice annotation. If no voice annotations are recorded, this
button is disabled.
B‑Mode report type: Manually measure the aortic diameter.
C‑Mode report type: Review a previously saved exam.
Return to the Home screen.
Page 29
TUTORIAL SCREEN
To open the Tutorial screen, press the Tutorial button from the Home screen. The Tutorial screen
presents a menu of training modules.
Note: When this screen is open, the Scan button on the probe is disabled.
Table 9. Tutorial Screen Button Functions
BUTTONFUNCTION
Move down one title or skip back one chapter in the training module.
Move up one title or skip forward one module.
Begin module playback. While the module is playing, press to pause. Press again to
resume play.
No function.
Return to the Home screen.
Operations & Maintenance Manual: Introduction
21
Page 30
VIDEO VIEWING SCREEN
The Video Viewing screen is activated by pushing the Enter button on the Tutorial screen.
Press the Play button to begin the desired tutorial.
Note: When this screen is open, the Scan button on the probe is disabled.
Table 10. Video Viewing Screen Button Functions
BUTTONFUNCTION
No function.
Play or pause video playback.
Return to the screen showing the list of titles..
No function.
Return to the Home screen.
22
Page 31
SETTINGS SCREEN
To open the Settings screen, push the Settings button on the Home screen. The display presents a list of
user‑configurable settings: Name, Date & Time, General Preferences, and Self Test.
Note: When this screen is open, the Scan button on the probe is disabled.
Table 11. Settings Screen Button Functions
BUTTONFUNCTION
Move down one setting in the list.
Move up one setting in the list.
Select the highlighted setting.
No function.
Return to the Home screen.
Operations & Maintenance Manual: Introduction
23
Page 32
NAME SCREEN
This screen allows you to select the appropriate alpha numeric characters for entering your health care
institution’s name.
For more information, see Program the Clinic Name on page37.
Table 12. Name Screen Button Functions
BUTTONFUNCTION
Move down in the grid.
Move right in the grid.
Add the highlighted character to the name.
Delete one character from the name.
Save the current name setting and return to the Settings screen.
24
Page 33
DATE & TIME SCREEN
This screen allows you to adjust the date and time.
For more information see Set the Date & Time on page39.
Note: If the time display is set to show a 24‑hour clock, the hour units are 0–23. If the clock is set to show a
12‑hour clock, the hour units are 01‑12.
Table 13. Date & Time Screen Button Functions
BUTTONFUNCTION
Move back one changeable unit.
Move to the next changeable unit
Add and/or toggle digits as appropriate. Press and hold the button to move through
options more quickly.
Subtract or toggle digits as appropriate. Press and hold the button to move through
options more quickly.
Save the current date and time settings and return to the Settings screen.
Operations & Maintenance Manual: Introduction
25
Page 34
GENERAL PREFERENCES SCREEN
This screen displays a list of available settings and their current values.
Available settings:
• Language: Multiple languages are available. English is the default setting.
• Date Format: mm/dd/yyyy; dd.mm.yyyy; yyyy‑mm‑dd.
• Time Format: 12 hour or 24 hour.
• Calibration Warning: On (default), Off. When On is selected, a calibration warning will appear in the
display header when the device requires calibration.
• Print Report Type: Toggle between C‑mode images (aorta in crosshairs) and B‑mode images (image
of aorta and abdominal space below probe).
• Enable ScanPoint®: On (default), Off. Select “Off” to disable ScanPoint.
Verathon Medical
AMI 9700
Aortic Measurement Instrument
Language
Date Format
Time Format
Calibration Warning
Print Report Warning
Enable ScanPoint
AMI 9700
English
mm/dd/yyyy
24H
On
B-mode
On
Figure 7. B‑Mode and C‑Mode Print Reports
06/10/2013
12:29
B‑Mode print report
26
C‑Mode print report
Page 35
Table 14. General Preferences Screen Button Functions
BUTTONFUNCTION
Move down a setting in the list.
Move up a setting in the list.
Select the next option. Press and hold to move through options more quickly.
Select the previous option. Press and hold to move through options more quickly.
Save the current settings and return to the Setup screen.
Operations & Maintenance Manual: Introduction
27
Page 36
SELF TEST SCREEN
When you open the Self Test screen, testing begins automatically. Once testing is complete, data on the
screen is printed automatically to the instrument’s onboard printer.
Table 15. Self Test Screen Button Functions
BUTTONFUNCTION
No function.
No function.
No function.
No function.
Return to the Settings screen.
28
Page 37
SCAN COUNTER FEATURE ON THE SELF TEST SCREEN
The AortaScan AMI9700 is equipped with a scan counter feature. It counts all scan button pushes captured
by the console. It is designed to enable clinical users or service technicians to determine the number of scans
the device has performed over its lifetime. It counts all scans taken with the instrument, including air scans
and practice scans. The counter advances automatically after each scan.
Please note that the scan counter feature is available only with software version 3.1.0.0 or higher. Some
AortaScan consoles cannot be upgraded to run software version 3.0 or higher. Software updates may be
performed by either logging on to ScanPoint®, or by contacting Verathon® Customer Care.
The scan counter may be monitored as a part of a regular device maintenance program. The number of scans
appears as a value on the self test screen and the self test printout.
To ensure reliability, a backup copy of the scan count is stored in device memory. If both the scan counter
and its backup copy are corrupted, the scan counter will automatically reset to a zero value.
The scan counter feature is designed so that the value cannot be manually reset or modified by the clinical
user or service partner.
PRINTING THE SCAN COUNT FROM THE SELF TEST SCREEN
Once the self‑test screen is accessed, data on the screen are printed automatically using the instrument’s
onboard printer.
Figure 8. Printout of Self Test Screen
Operations & Maintenance Manual: Introduction
29
Page 38
TROUBLESHOOTING
The scan counter feature is designed for redundancy, so the scan value is stored in multiple locations in the
instrument’s internal memory. If one of the storage locations fails, the text “EEPROM Failed” will be added to
the DSP status line. In the event of an EEPROM failure, the counter will continue to work but will not have a
backup copy stored in the instrument.
Figure 9. Self Test Screen and Printout when EEPROM Has Failed
30
Page 39
SCANPOINT SCREEN
Note: This screen is only available if the optional ScanPoint® software is installed on a PC.
Press the ScanPoint button on the Home screen. The ScanPoint screen displays information about the
status of the link between the AortaScan instrument and the ScanPoint host computer.
Figure 10. ScanPoint Screen (Searching)
Table 16. ScanPoint Screen Button Functions
BUTTONFUNCTION
No function.
Figure 11. ScanPoint Screen (Connected)
No function.
No function.
Cancels the current action and ends communication with ScanPoint®.
No function.
Operations & Maintenance Manual: Introduction
31
Page 40
SLEEP MODE
To conserve battery power, the AortaScan AMI9700 goes into sleep mode by shutting itself down
automatically when not in use.
After four minutes of idle time, a sleep mode alert message displays for 15 seconds. While the message
is displayed, press any button to keep the console awake and dismiss the message. After 15 seconds, the
console goes to sleep. To wake the instrument from sleep, simply press the Power button.
32
Page 41
SETTING UP
To help you get up and running as quickly as possible, the next few pages explain how to:
1. Perform the Initial Inspection
2. Set Up the Battery
3. Attach the Probe to the Console
4. Program the Clinic Name
5. Set the Date & Time
6. Load the Thermal Paper
7. Attach the Instrument to a Medical Cart (Optional)
8. Install ScanPoint with QuickPrint (Optional)
9. Watch the Training Video
PROCEdURE 1. PERFORM THE INITIAL INSPECTION
When you receive the AortaScan AMI9700 system, Verathon® recommends that an operator familiar with
the instrument perform a full visual inspection of the system for any obvious physical damage that may have
occurred during shipment.
1. Carefully open the top flaps of the shipping box. Do not insert anything sharp through the top of
thebox.
2. Remove the contents and verify that you have received the appropriate components for your system.
3. Inspect the components for damage.
4. If any of the components are missing or damaged, notify the carrier and Verathon Customer Care or your
local representative.
Operations & Maintenance Manual: Setting Up
33
Page 42
PROCEdURE 2. SET UP THE BATTERY
WARNING
Risk of explosion, fire, or serious injury. The AortaScan AMI9700 is powered by a
lithium‑ion battery. Failure to note the following when handling the battery may result in serious
injury:
• Never short‑circuit the battery by either accidentally or intentionally bringing the battery
terminals into contact with any other conductive object. This could cause serious injury or fire
and could also damage the battery and/or the instrument.
• Never expose the battery to abnormal shock, vibration, or pressure. The battery’s internal
protective covering could fail, causing it to overheat or ignite, resulting in caustic liquid
leakage, explosion, or fire.
• Do not disassemble, heat above 60°C (140°F), or incinerate the battery. Keep battery out
of reach of children and in original package until ready to use. Dispose of used batteries
promptly according to local recycling or waste regulations.
• If the battery is leaking or its case is cracked, put on protective gloves to handle it, and
discard it immediately. Always dispose of used batteries in compliance with all applicable
laws and regulations. Put insulating tape, such as cellophane tape, on the electrodes during
transportation in order to avoid a possible short circuit, fire, or electrical shock.
WARNING
Ensure proper distance from patient. The AMI9700 battery charger/wireless hub and the
computer used to access online ScanPoint
®
image archives (if used) must be placed outside the
patient vicinity (more than 2m [6ft] from the patient).
Two lithium‑ion batteries are included with the AortaScan AMI9700. One battery can be recharged in the
battery charger/wireless hub while the other is installed in the AortaScan instrument. This ensures there is no
instrument downtime. The charger will bring the batteries to a full charge within 6 hours or less. Before using
the AortaScan AMI9700 for the first time, you need to charge both batteries.
The console draws very little power when it is turned off. However, if you do not plan to use the instrument
for several weeks, you should remove the battery to prevent it from discharging. When batteries are not in
use, they should be stored in the battery charger so they remain fully charged.
CHARGE THE BATTERIES
1. Plug the battery charger/wireless hub unit into a standard wall outlet.
2. Insert the battery into the recess in the battery charger.
Note: Fully charging the battery may take up to 6 hours. Batteries may be stored in the charger. There is
no danger of overcharging the batteries.
34
Page 43
3. Check the colored indicator lights on the battery charger to determine battery status:
Solid green: Battery fully charged.
Amber: Battery charging.
The battery status indicator remains in the top right corner of the screen and indicates the charge level of
the battery.
Table 17. Battery Power Level
BAT TERY ICONPOWER LEVEL
Indicates a fully charged battery.
Indicates a battery 50% to 75% charged.
Indicates a battery 25% to 50% charged.
Indicates a battery is almost depleted. Can power a few more scans.
Indicates that a battery should be replaced immediately.
INSERT A BATTERY INTO THE INSTRUMENT
4. To insert the charged battery into the battery well in the console, slide it under the ledge and push down
gently until the battery clicks into place.
Note: The battery is designed to prevent incorrect installation. If the battery does not slide into the
battery well easily, remove the battery and try again. Do not force the battery into position.
Operations & Maintenance Manual: Setting Up
35
Page 44
PROCEdURE 3. ATTACH THE PROBE TO THE CONSOLE
1. Locate the cable port on the back of the console.
2. Align the silver arrow on the probe cable connector to the top of the cable port.
Arrow on cable connector
3. Gently push the connector ring into the port, until the cable clicks into place and is secure.
The cable can remain attached to the console in between uses.
Note: To remove the cable, pull the connector ring back until the cable disconnects. Do not pull on the cable.
36
Page 45
PROCEdURE 4. PROGRAM THE CLINIC NAME
You can customize your AortaScan AMI9700 by entering your facility’s name. This information will
subsequently be included on the console display and all printouts of exam results.
1. Turn the AortaScan AMI9700 on by pressing the Power button on the front of the console.
2. When the Home screen appears, press the Settings button to open the Settings screen.
3. On the Settings screen, push either the Up Arrow button or Down Arrow button until “Name”
is highlighted. Press the Enter button to open the Name screen.
Operations & Maintenance Manual: Setting Up
37
Page 46
4. On the Name screen, use the Right Arrow button and Down Arrow button to move to the
desired character. When the desired character is highlighted, press the Plus button to add it to your
text. Use the Minus button to delete characters.
To add a space between words, select the blank space below the letter x.
To add a second line of text use the ¶ character
5. When finished, press the Settings button to return to the Settings screen. From the Settings screen,
press the Home button to return to the Home screen. The facility name will now appear in the
display header.
Note for extended‑Latin and/or non‑Latin characters: Extended Latin characters (tilde, umlaut, accents,
circumflex, etc.) and/or non‑Latin characters can be entered only by using ScanPoint® with QuickPrint
software. To enter a name that uses extended or non‑Latin characters, please refer to the instructions in
the ScanPoint with QuickPrint User’s Manual.
38
Page 47
PROCEdURE 5. SET THE dATE & TIME
1. Turn on the device by pressing the Power button .
2. From the Home screen, press the Settings button to open the Settings screen.
3. On the Settings screen, push either the Up Arrow button or Down Arrow button buttons until
“Date & Time” is highlighted. Press the Enter button to open the Date and Time screen.
4. On the Date and Time screen, use the Up Arrow button and Down Arrow button to move to
the desired unit (hours, minutes, month, day, year). When the desired unit is highlighted, press the Plus
button to increase values and the Minus button to decrease values.
Note: If the time display is set to show a 24‑hour clock, the hour units are 0–23. If the clock is set to
show a 12‑hour clock, the hour units are 1‑12.
5. When the time and date are set correctly, press the Settings button to return to the Settings screen.
From the Settings screen, push the Home button to return to the Home screen.
Operations & Maintenance Manual: Setting Up
39
Page 48
PROCEdURE 6. LOAd THE THERMAL PAPER
If paper appears to be stuck in the printer, see the procedure Clear a Paper Jam on page66.
1. Locate the paper compartment door on the base of the console, behind the display.
2. Slide the door to the right, then lift up.
3. If there is an empty paper roll, remove it.
4. In the paper well, insert the end of a new paper roll with the thermal side down.
5. Extend the end of the paper past the side of the unit.
6. Snap the door completely closed, then slide the door back into the console.
7. Tear off any paper extending from the back of the console.
40
Page 49
PROCEdURE 7. ATTACH THE INSTRUMENT TO A MEdICAL CART (OPTIONAL)
The AortaScan AMI9700 is completely portable and can be easily moved and positioned for convenient use.
Installing the AMI9700 on the optional mobile cart will allow you to move the AortaScan along with related
accessories to the patient examining area or bedside, as necessary.
Figure 12. Assembled Medical CartFigure 13. Medical Cart Assembly
Table 18 . Medical Cart Parts List
ITEMQT YPART
11Medical tray
21Post
31Medical cart base
44Screw PH W Lock 25‑20 x 1/2
51Fluted knob 3/8‑16 x 1.00
63Caster, 3 inch
72Caster, 3 inch with brake
—2Loctite® 680 retaining compound (not pictured)
Operations & Maintenance Manual: Setting Up
41
Page 50
ASSEMBLE THE MEdICAL CART
1. Insert the five casters into the medical cart base, positioning the brake casters on opposite ends of the
base.
2. Insert the post into the square relief on the underside of the medical tray.
3. Insert the four screws through the bracket on the top of the post into the molded inserts in the medical
tray and tighten securely.
4. If you want to permanently attach the post to the wheeled base, refer to Step 6 through Step 11.
If you want the ability to disassemble the medical cart at a later date, place the tray assembly with the
post into the wheeled medical cart base.
5. Place the AortaScan into the foot prints on the medical tray.
If you want to secure the instrument to the medical cart, refer to Step 13 through Step 15.
PERMANENTLY ATTACH THE POST TO THE WHEELEd BASE (OPTIONAL)
6. Place medical cart base on level ground.
7. Open the two tubes of Loctite 680 by snapping off the tips of the tubes.
8. Apply the Loctite 680 all around the tapered portion of the post. Use all of the contents of both tubes.
Complete coverage of the tapered portion is not necessary as the Loctite will spread upon insertion into
the base.
9. Slide the post into the hole in the base with a twisting motion and press down firmly.
10. Wipe off excess Loctite with paper towel and discard towel as waste.
11. Allow post and base to sit undisturbed for 3 hours.
ATTACH THE ACCESSORY BASKET (OPTIONAL)
A universal accessory basket is available for the medical cart to provide additional storage capacity.
12. Follow the manufacturer’s instructions for attaching the accessory basket to the pole.
Figure 14. Universal Accessory Basket
42
Page 51
ATTACH THE INSTRUMENT TO THE MEdICAL CART (OPTIONAL)
13. Place the AMI9700 atop the cart, aligning the rubber pads on the bottom of the device to the
corresponding indentations on the tray.
14. On the bottom of the tray, insert the fluted knob into the mounting hole in the center.
15. Screw the fluted knob into the mounting hole until the device is secure on the tray.
Figure 15. Attach the AMI9700 to the Medical Cart
PROCEdURE 8. INSTALL SCANPOINT WITH QUICKPRINT (OPTIONAL)
The optional ScanPoint® with QuickPrint software is designed to work seamlessly with your AortaScan
devices. The AMI9700 automatically downloads exam data to the ScanPoint host computer via a wireless
connection enabled by the battery charger/wireless hub, allowing further viewing, analysis, archiving, and
report generation.
To install ScanPoint with QuickPrint software, insert the ScanPoint with QuickPrint install CD into your
computer’s CD drive and follow the onscreen prompts. Please refer to the separate manual provided with
ScanPoint with QuickPrint software for complete installation and operating instructions.
PROCEdURE 9. WATCH THE TRAINING VIdEO
The training video provides an overview of how to perform an ultrasound scan of the aorta using the
AortaScan AMI9700. The video:
• Is approximately 5 minutes long.
• Is available at the Verathon® Web site: http://verathon.com.
• Is available for review anytime on the instrument by pushing the Tutorial button from the Home
screen.
Operations & Maintenance Manual: Setting Up
43
Page 52
USING THE DEVICE
WARNING
The aortic diameter measurement function provides images that may be used for diagnosis and
screening. If clinically indicated, appropriate patients should be referred for additional diagnostic
testing.
WARNING
The AortaScan system is designed to detect the fluid (blood) filled region of the abdominal
aorta only. The system cannot detect the presence of a blood clot (thrombus) and therefore may
provide a false negative result.
WARNING
The AortaScan system is an ultrasound‑based device and is subject to all limitations of this
method. If clinically indicated, appropriate patients should be referred for a diagnostic standard
(confirmatory) test, regardless of test results.
WARNING
Risk of inaccurate measurements/results. When using the instrument, be aware of the
following conditions that can decrease the accuracy of exam results:
• In some cases, the normal operating tolerances of the instrument can cause it to report a
falsely normal or abnormal measurement. For more information, see Interpret the Aortic
Measurement Results on page51.
• Visual verification that the aorta position is fully within the scan cone on the displayed
images is important.
• A thrombus (blood clot) can complicate aortic measurements. A soft, blood‑like thrombus
may appear as part of the lumen. However, a calcified thrombus may appear as part of the
aorta’s wall, resulting in a measurement of lumen diameter that is smaller than the aorta
diameter. Accordingly, in patients where thrombus is known or suspected, other imaging
methods should be used prior to ruling out an aneurysm.
• Use care when scanning patients who have had abdominal surgery. Scar tissue, surgical
incisions, sutures, and staples can affect ultrasound transmission and accuracy.
• Ensure that the patient fasts for 12 hours prior to undergoing an aortic diameter
measurement in order to minimize the presence of bowel gas, which may obstruct proper
measurement.
• Obesity may affect ultrasound aortic diameter measurements. For more information, see
Obesity on page54.
44
Accuracy is compromised if the user does not obtain an optimal, repeatable image.
Page 53
WARNING
Risk of explosion. If you use the AortaScan AMI9700 in the presence of flammable
anesthetics, the hazard of potential explosion exists.
WARNING
Do not use the AortaScan AMI9700 on:
• A patient who has open skin or wounds in the mid‑abdominal area.
• A patient with ascites.
• A pregnant patient.
The AortaScan AMI9700 provides the capability to measure abdominal aortic diameter noninvasively
using 3D ultrasound. AortaScan ultrasound may be preferred as the initial imaging modality for measuring
abdominal aortic diameter due to its portability, availability, lack of ionizing radiation, and cost when
compared to other alternatives like CT, CTA, MRI, MRA, or standard ultrasound performed by trained
sonographers.
The AortaScan AMI9700 can measure diameters ranging between 3.0 and 12.4 cm with a diameter
accuracy of ± (15% + 0.5 cm). For details about the normal variability of scan results, and the effect of that
variability on potential rupture risk, refer to Table 19 on page 51.
Note: The AortaScan AMI9700 is not intended for screening or diagnosis of abdominal aortic aneurysms
(AAAs), or for use on acute events such as aortic dissection, ulcer, or rupture.
In B‑mode report type, the scan output provides two images: an aiming display on the left and a results
display on the right. The AMI9700 also displays the calculated diameter of the aorta in centimeters (cm)
below the aiming display.
Figure 16. B‑Mode Report Type
Figure 17. C‑Mode Report Type
Operations & Maintenance Manual: Using the Device
45
Page 54
AIMING dISPLAY
The aiming display illustrates the location of the aorta relative to the ultrasound probe. The center of the two
axes represents the center of the probe. The aorta is shown in red. The white lines represent areas of high
reflection, most likely caused by bowel gas.
When aiming, the goal is to angle the probe so the aorta in red is not intersecting any white lines (bowel gas)
on the aiming display. Arrows will appear, which indicate the direction the probe needs to move in order to
produce a better scan.
RESULTS dISPLAY
The Results display provides an image of the cross section of the abdominal space below the probe. Both
axes of the image are in centimeters. The abdominal aorta is seen as a dark circular shadow with a red
outline amidst the black and white “speckling” of the ultrasound image.
PROCEdURE 1. PREPARE FOR THE EXAM
1. Ensure you are familiar with the parts and functions of the instrument. For more information, see the
Introduction chapter on page8.
2. Check the instrument battery icon to ensure the battery has sufficient power.
If the battery icon is ¼ full or less, replace the battery with a freshly charged battery before proceeding.
Ensure the instrument is off before you replace the battery. Place the discharged battery in the battery
charger to recharge.
3. Ensure that the instrument has been properly cleaned according to the instructions in the chapter
Cleaning & Maintenance on page55.
4. Be aware of the following conditions that may affect ultrasound transmission and the accuracy of
theexam:
• Bowel gas is a common problem for abdominal ultrasound measurement and results in unreadable
exams. To avoid bowel gas obstruction of the ultrasound, have patients fast for 12 hours prior to the
exam. When aiming, position the probe so that the image of the aorta on the aiming display does not
intersect with bowel gas.
• Previous abdominal surgery. Scar tissue, surgical incisions, sutures, and staples can affect ultrasound
transmission and reflection.
Do not use the AMI9700 on:
• Patients with ascites.
• Patients with open skin or wounds in the mid‑abdominal region.
• Pregnant patients.
46
Page 55
PROCEdURE 2. MEASURE ABdOMINAL AORTIC dIAMETER
1. Turn on the instrument by pressing the Power button .
2. With the patient in the supine position, identify four equally spaced scan locations between the xiphoid
process and the umbilicus.
Figure 18. Four Scan Locations for Measuring
Scan 1
Scan 2
Scan 3
Scan 4
3. Place an ample amount of ultrasound gel on the patient’s abdomen in the selected scan locations.
4. Standing at the patient’s right side, place the probe on the gel at the first position.
5. Hold the probe with the long axis aligned with the midline of the abdomen. Do not hold the probe with
the handle pointing to either side of the patient.
Figure 19. Correct and Incorrect Orientation of the Probe
6. Press and release the Scan button on the underside of the probe. When you hear the tone, the scan is
complete.
Note: Do not move the probe while the scan is in progress as that will decrease the accuracy of the
measurement.
7. Save the exam results by creating a voice annotation. See Save, Review, & Print Exam Results.
Operations & Maintenance Manual: Using the Device
47
Page 56
8. Perform three more measurements and save and annotate the result of each exam. For more
information, see Save, Review, & Print Exam Results on page50.
Figure 20. Results Screen
9. If you would like to perform another exam, press the Home button , and then repeat this procedure.
If you would like to measure abdominal aortic diameter manually, complete the procedure Measure
Aortic Diameter Manually (Optional) on page49.
10. Once you have completed the exam, wipe the gel off the patient and probe.
For ScanPoint® subscribers, logging onto ScanPoint automatically transfers and saves your annotated
exams to your ScanPoint host computer.
To measure abdominal aortic diameter manually, you must perform a scan in B‑mode. For information about
setting the instrument in B‑mode, see General Preferences Screen on page26.
1. After completing the scan, press the Manual Mode button . The Manual Measurement Mode
screenopens.
Axis Select Manual Mode
Return
2. By using the following button controls, move one cursor to the right edge of the aorta and move the
other cursor to the opposite edge of the aorta:
• Press the Axis Select button to swap between the Up and Down arrows or the Left and Right
arrows.
• Use the Manual Mode button in order to swap between cursors on the results display.
• When you are finished moving the cursors, press the Return button .
This records the measurement and exits Manual Measurement Mode. The manual measurement is
displayed on the Review screen. Record a voice annotation in order to save the manual measurement
result.
Operations & Maintenance Manual: Using the Device
49
Page 58
PROCEdURE 4. SAVE, REVIEW, & PRINT EXAM RESULTS
IMPORTANT
In order to save the scan, you must record an annotation. If you do not record an annotation, the scan
result will be lost, and the next scan you perform will overwrite the non‑annotated scan.
The AMI9700 instrument does not automatically save each scan. It is recommended that you add a voice
annotation or write down the diameter calculated for each location.
SAVE/ANNOTATE AN EXAM
1. On the console, press and release the Record button .
2. Hold the probe approximately six inches (15 cm) from your mouth, and then record the patient
information by speaking clearly into the probe microphone located just above the aiming display on the
probe.
3. When you are finished recording, press the Stop button . An hourglass icon appears to indicate that
the scan is being saved.
4. Press the Listen button . The voice annotation plays.
If you are not satisfied with the recording, press the Record button again to record a new
annotation.
Note: You can make a new recording only if the instrument still displays the aortic diameter for that
particular scan.
If desired, the instrument is ready to perform another scan.
REVIEW AN EXAM
5. On the console, press the Review button .
On the Review Screen, two types of diameter measurements may be displayed:
• Diameter
• Diameter
– Diameter measured automatically by AMI9700
V‑mod e
– Diameter measured manually in Manual Measurement Mode. See Measure Aortic
Manual
Diameter Manually (Optional) for further details.
Note: You must record a voice annotation in order to review the results.
PRINT AN EXAM
6. If you are printing exam results immediately after the measurement is taken, on the Results screen, press
the Print button .
7. If you are printing saved exam results, press the Review button , select the saved exam that you want
to print, then press the Print button .
50
Page 59
PROCEdURE 5. INTERPRET THE AORTIC MEASUREMENT RESULTS
WARNING
The aortic diameter measurement function provides images that may be used for diagnosis and
screening. If clinically indicated, appropriate patients should be referred for additional diagnostic
testing.
WARNING
The AortaScan system is designed to detect the fluid (blood) filled region of the abdominal
aorta only. The system cannot detect the presence of a blood clot (thrombus) and therefore may
provide a false negative result.
WARNING
The AortaScan system is an ultrasound‑based device and is subject to all limitations of this
method. If clinically indicated, appropriate patients should be referred for a diagnostic standard
(confirmatory) test, regardless of test results.
The AortaScan AMI9700 can measure diameters ranging between 3 and 12.4 cm with a diameter accuracy
of ±(15%+0.5cm). This error‑range data (Table 19) indicates a range of values obtained by the device
relative to follow up and clinical significance, specifically with respect to risk vs. diameter.
Table 19. Expected Aortic Measurement Ranges
ACTUAL AORTIC DIAMETER
3.0 cm3.5 cm4.0 cm4 .1 cm5.0 cm5.3 cm
Average
estimated risk
of rupture for
actual aortic
diameter
Aortic diameter as reported by the device based on allowable tolerances
± 15%2.553.452.984.033.404.603.494.724.255.754.516.10
With additional
±0.5 cm
Average
estimated risk
of rupture for
reported aortic
diameter
0%0%0%1%11%11%
MinMaxMinMaxMinMaxMinMaxMinMaxMinMax
2.053.952.484.532.905.102.995.223.756.254.016.60
0%0%0%1%0%1%0%11%0%26%
0.5–
5.0%
26%
Operations & Maintenance Manual: Using the Device
51
Page 60
An unobstructed scan has been achieved when the probe displays eight solid green arrows.
When the scan is complete, the AortaScan AMI9700 shows the aortic diameter and two displays on the
console screen.
Aorta
Bowel gas
Results display
Aiming display
The aiming display on the left side of the screen shows the location of the aorta relative to the probe, as
viewed looking from the probe into the body. The aorta is shown in red and bowel gas is shown as white
lines. The green dot on the left side is the reference mark correlating the aiming display with the results
display on the right side of the screen.
The results display is a cross‑section of the abdomen below the probe. The abdominal aorta is shown as
a dark circular shadow with a red outline. The white dot on the image is a reference mark correlating the
results display with the aiming display.
52
Page 61
MEASURING AORTIC dIAMETERS <3CM
The AortaScan AMI9700 can detect aortas with diameters between 3cm and 12.4cm. Diameters less than
3cm occur in patients who have normal‑sized aortas.
The round shadow at 6cm depth in the results display is the patient’s abdominal aorta. Patients with
aortas less than 3cm in diameter will show no red outline around the aorta in the results display, as the
diameter cannot be measured automatically. However, because of the potential variability between actual
and measured diameters, the absence of a red outline around the aorta cannot be relied upon to identify an
abdominal aorta less than 3 cm in diameter. The diameter can also be estimated using Manual Measurement
Mode.
The “speckled” look to the Results display indicates the ultrasound signal was not blocked by bowel gas. The
clearly visible aorta and the lack of any arrows telling the user to re‑aim the probe mean the user can feel
confident that the lack of diameter information is due to a small aorta and not due to the presence of bowel
gas. In this case, the aortic diameter measurement is valid.
PARTIAL GAS OBSTRUCTION
In some cases, gas or air bubbles may be present but do not block the aorta entirely. In this case, diameter
measurements are still calculated, but they are not typical.
A green arrow on the console and a solid green arrow on the probe indicate the abdominal aorta can be
detected, but the presence of bowel gas prevents a proper measurement.
Figure 21. Partial Gas Obstruction
Moving the probe 1/2to1inch (1to3cm) in the direction of the arrow has a high probability of obtaining a
successful scan.
In this case, the probe should be repositioned and the patient rescanned. Gently but firmly work the probe
into the tissues of the abdomen with a side‑to‑side rocking motion to try and displace any bowel gas
obscuring the aorta. Do not move the probe while the scan is in progress as that will decrease the accuracy
of the measurement.
Operations & Maintenance Manual: Using the Device
53
Page 62
SUBSTANTIAL GAS OBSTRUCTION
A substantial amount of gas in the abdomen can block ultrasound from reaching the aorta and results in an
unreadable or inappropriate scan.
A red arrow on the console and a flashing green arrow on the probe indicate bowel gas has substantially
obscured the aorta. No diameter measurement can be calculated and the results display shows a diameter of
<3 cm, meaning the aorta was not detected.
Figure 22. Substantial Gas Obstruction
Although moving the probe 1/2to1inch (1to3cm) in the direction of the arrow has a low probability
of providing a successful scan, an additional scan should be attempted. In this case, the probe should be
repositioned and the patient rescanned. Gently but firmly work the probe into the tissues of the abdomen
with a side‑to‑side rocking motion to try and displace any bowel gas obscuring the aorta. Do not move the
probe while the scan is in progress as that will decrease the accuracy of the measurement.
If rescanning is not successful, the exam should be postponed and rescheduled. Have the patient fast for
12hours prior to the exam.
OBESITY
Attenuation of the ultrasound signal by excess abdominal fat can result in a poor ultrasound image, which
affects the quality of the diameter measurement.
With obese patients, try pressing the probe firmly into the abdomen to reduce the distance to the aorta as
much as possible, while attempting to minimize patient discomfort.
In rare cases, it is possible for a patient’s abdomen to be too thick for the ultrasound to reach the aorta. If
a patient has an extra‑thick abdomen where the distance from the probe face to the aorta is greater than
18cm (7in), the AortaScan AMI9700 will not detect the aorta appropriately. In these cases, alternative
imaging methods should be used.
54
Page 63
CLEANING & MAINTENANCE
Routine cleaning and maintenance will help ensure safe and effective operation of the AortaScan AMI9700.
For more information, please contact your authorized AortaScan Service Center, your local AortaScan
distributor, or Verathon® Customer Care.
CLEANING & DISINFECTING
Clean and disinfect the instrument before use and between patient exams.
WARNING
This product may only be cleaned and disinfected by using the approved processes provided in
this manual. Cleaning and disinfection methods listed are recommended by Verathon based on
compatibility with component materials.
WARNING
Availability of cleaning and disinfection products varies by country, and Verathon is unable
to test products in every market. For more information, please contact Verathon Customer
Care at 1.800.331.2313 or your local representative. For additional contact information, visit
verathon.com/contact‑us.
WARNING
Ensure that you follow the manufacturer’s instructions for handling and disposing of the
cleaning and disinfection solutions provided in this manual.
PROCEdURE 1. CLEAN THE INSTRUMENT
WARNING
Cleaning is critical to ensuring the component is ready for disinfection. Failure to properly clean the
device could result in a contaminated instrument after completing the disinfectionprocedure.
Cleaning is the removal of all visible soil or contaminants from the exterior surfaces of the device. The device
must be cleaned after every use, and cleaning is an essential step before disinfection.
1. Wipe the ultrasound gel completely off the probe.
2. Use a moistened, soft cloth to remove particulate matter or body fluids that remain on the instrument.
3. Do not re‑use cloths or wipes.
4. Allow the device to air dry, or towel dry with a clean dry cloth. Next, you must disinfect the instrument.
Failure to heed the following may cause device damage not covered by the warranty:
• Do not immerse the instrument in the disinfectant solution.
• Do not subject any part of the instrument to steam, ethylene oxide, radiation, or similar methods of
sterilization or autoclaving.
• Do not use CidexPlus
Disinfectants and cleaning methods listed are based on compatibility with product materials, not biological
effectiveness. Refer to the instructions from the manufacturer of the disinfectant for guidance on biological
effectiveness of the disinfectant.
The following liquid disinfectants and wipes are compatible with the materials used in the instrument:
• A‑456® II Disinfectant
• Accel
• Cavicide
• CaviWipes
• Chloro‑Sol Spray
• Clorox® Germicidal Wipes
®
to disinfect the instrument. CidexPlus will damage the plastic enclosure.
®
TB Wipes
®
®
®
• Sani‑Cloth
• Sani‑Cloth
• Sani‑Cloth
• Sporicidin
• T‑Spray II
®
Bleach Wipes
®
Germicidal Wipes
®
Plus Germicidal Wipes
®
Disinfecting Towelettes
®
The level of disinfection required for a device is based on the type of tissue it contacts during use. Based on
the intended use of the AortaScan AMI9700, low‑level disinfection is the minimum level required.
1. Ensure the instrument has been properly cleaned according to the procedure Clean the Instrument on
page55.
2. Ensure the disinfectant has not expired.
3. If using a liquid disinfectant, prepare the disinfection solution according to the manufacturer’s label
instructions, ensuring that you are using the proper concentration.
4. Apply the solution to a soft cloth or wipe.
Note: Do not spray or apply liquid disinfectants directly to the surface of the device, and do not soak the
device in liquids.
5. Wipe the surfaces of the device and allow the surface to remain wet for the required contact duration.
Follow the manufacturer’s instructions for the appropriate disinfection level contact duration.
6. Do not re‑use cloths or wipes.
7. If rinsing or removal of the disinfectant solution from the device is required by the disinfectant
manufacturer’s instructions, wipe with a clean soft cloth dampened in sterile water. Verathon®
recommends wiping the device three separate times to remove all residual disinfectant.
8. Allow the device to air dry, or towel dry the device with a clean, dry cloth.
56
Page 65
REGULAR INSPECTIONS
Verathon® recommends that the AortaScan AMI9700 be certified by an authorized Service Center once a
year. Certification service includes comprehensive inspection and testing of the instrument to ensure proper
performance in clinical use. For more information, please contact your authorized Verathon Service Center,
your local Verathon distributor, or Verathon Customer Care.
Note: ScanPoint® Online customers can maintain device certification via the Internet by accessing their
ScanPoint account. For more information about using ScanPoint Online, please refer to the ScanPoint with
QuickPrint user’s manual.
WEEKLY INSPECTIONS
Once a week, you should inspect the probe and cable for physical faults or cracks. Cracks that allow the ingress
of fluid may affect the performance of the instrument. Any apparent cracks or faults in the console, probe, or
the cable that links the console and the probe, must be referred to your authorized AortaScan Service Center,
your local Verathon distributor, your local Verathon representative, or Verathon CustomerCare.
MAINTENANCE
PROCEdURE 1. RUN A SELF TEST
The AMI9700 can perform a number of self‑diagnostic tests. To access the Self Test utility:
1. From the Home screen, press the Settings button .
2. When the Settings screen opens, press the Up Arrow button or Down Arrow button buttons
until Self Test is highlighted in red, then press the Enter button . The Self Test screen opens and
testing begins automatically. The display provides status and results, and when the test in complete, the
printer prints the results.
Note: Make sure the printer is loaded with paper. See Load the Thermal Paper.
3. If the screen indicates any failed tests or abnormal results, contact your authorized AortaScan
representative, or contact the Verathon Customer Care Department.
4. When the test is complete, press the Settings button to return to the Settings screen, then press the
Home button to return to the Home screen.
1. On the instrument, on the Home screen, press the ScanPoint button .
2. On the computer, double‑click the ScanPoint with QuickPrint icon. ScanPoint® opens.
3. On the computer, in the ScanPoint QuickPrint window, click Find New. QuickPrint establishes a
connection with the instrument, and an icon for the device appears in the left pane.
58
Page 67
4. Select the 9000 Series device, verify that the serial number matches the device you are updating, and
then click the Update Instrument button.
5. If any updates are available, the device downloads and installs them. The console displays a progress bar
and automatically restarts when the installation is complete.
If no updates are available, nothing happens.
6. If you would like to view the current software version and verify that the newest software is installed,
complete the procedure Run a Self Test. The results screen displays the software version.
PROCEdURE 3. CALIBRATE THE PROBE USING THE SCANPOINT SYSTEM
If you do not have ScanPoint® with QuickPrint software, you must send your instrument to an authorized
Verathon® service center for calibration. Contact Verathon Customer Care for more information.
At minimum, the AMI9700 must be calibrated every 12 months in order to ensure appropriate results.
Calibrating ensures proper alignment of the instrument’s internal coordinate system. If calibration is not
performed by the prescribed date, the instrument can still be used to take scans but measurements may be
compromised. When calibration is required, a warning appears in the display header.
Figure 23. Calibration Warning
1. Within 10feet of the Battery Charger/Wireless Hub, place the calibration tank on a flat, nonreflective
surface, and then remove the lid.
2. Pour clean, room‑temperature water into the container, filling to the indicator mark. Ensure that there is a
minimal amount of bubbles in the water.
3. Using the notches to position the spiral‑shaped target correctly, place the target in the container.
4. Replace the lid onto the calibration container.
60
Page 69
5. Place the probe into the cutout in the lid. Ensure that the tip of the probe is submerged in the water.
6. On the computer, double‑click the ScanPoint with QuickPrint icon. ScanPoint® opens.
7. On the console, on the Home screen, press the ScanPoint button .
8. On the computer, in the ScanPoint QuickPrint window, click Find New. QuickPrint establishes a
connection with the instrument, and an icon for the device appears in the left pane. On the console, two
arrows appear, confirming that the console is connected to ScanPoint.
9. Select the 9000 Series device, verify that the serial number matches the device you are calibrating, and
then click the calibration tank icon.
10. Click the Calibrate button. ScanPoint® begins to scan and analyze the data in order to ensure that it
meets the calibration parameters. If necessary, the instrument automatically rescans the phantom.
62
Page 71
11. If calibration is successful, a “Calibration Successful” message is displayed on the computer.
If calibration fails, a Calibration Failure message appears. Ensure that the calibration chamber has
sufficient water and that the probe is seated properly in the calibration lid, and then on the Calibration
Failure message, click Yes. ScanPoint® restarts the calibration.
12. On the console, click the Exit button . This terminates the calibration procedure and ends
communication with ScanPoint QuickPrint.
13. Remove the probe from the calibration lid, and then dry it with a clean, soft cloth.
DEVICE DISPOSAL
The AortaScan AMI9700 and related devices may contain mineral oils, batteries, and other environmentally
hazardous materials. When the AortaScan AMI9700 has reached the end of its useful service life, return
the device, battery charger/wireless hub, and related accessories to a Verathon® Service Center for proper
disposal. Alternatively, follow your local protocols for hazardous waste disposal.
Verathon® provides several customer service resources, described in the table below.
RESOURCEDESCRIPTION
In‑serviceCD or USB
Video tutorial
The CD or USB flash drive included with your system that provides instructions
for using the instrument.
The video tutorial explains how to take aortic diameter measurements. The
tutorial is installed on the instrument.
Phone support
DEVICE REPAIR
The AortaScan AMI9700 console, probe, and battery charger/wireless hub are completely sealed. There are
no user‑serviceable components. Verathon does not make available any type of circuit diagrams, component
parts lists, descriptions, or other information that would be required for repairing the device and related
accessories.
Premium Warranty customers have access to a loaner unit and free shipping options that vary according to
the service plan.
If you have any questions, contact your local Verathon representative or Verathon Customer Care.
Please refer to the list of Customer Care resources available at
verathon.com/contact‑us
64
Page 73
TROUBLESHOOTING PROCEDURES
PROCEdURE 1. TROUBLESHOOT SCANPOINT CONNECTION
Complete this procedure if the console cannot connect to ScanPoint®.
1. In ScanPoint, retry the connection by clicking the Find New button. Repeat this step up to 3times.
If the console does not connect, continue to the next step.
2. Turn the console off, turn the console on, and then press the ScanPoint button . On the PC, in
ScanPoint, click Find New.
If the console does not connect, continue to the next step.
3. On the PC, click Find New. While the device is attempting to connect to ScanPoint, remove the battery.
4. Reinsert the battery, allow the device to power on, and then press the ScanPoint button .
5. On the PC, click Find New.
If the console does not connect, contact Verathon® Customer Care.
Operations & Maintenance Manual: Troubleshooting
65
Page 74
PROCEdURE 2. TROUBLESHOOT POWER ISSUES
If the instrument does not turn on, this is usually due to a dead or discharged battery and can be remedied
by replacing the dead battery with a charged battery.
When the battery charge is too low to allow normal operation (but not too low to permit operation of the
internal circuitry) the device displays the following message:
Battery charge level is too low for instrument operation. Recharge before next use.
In this case, the battery must be recharged or replaced with a charged one.
You may find that you need to exchange the battery more often over time due to normal loss of battery
capacity. For this reason, Verathon® recommends replacing the existing batteries with new batteries every
two years.
If the instrument has stopped responding even with a new battery, perform a full reset by removing and
reinserting the battery. If the instrument still does not respond, contact Verathon Customer Care.
PROCEdURE 3. INSTRUMENT TOO HOT
The AMI9700 displays the message “Too hot” if the print head overheats. In this case, turn off the
AMI9700 immediately. This condition may be the result of a paper jam.
PROCEdURE 4. CLEAR A PAPER JAM
Complete this procedure if the paper will not advance through the printer.
1. Open the printer door on the back of the console and clear the paper jam.
2. Ensure that the thermal paper is loaded correctly according to the instructions in the procedure Load the
Thermal Paper on page40.
66
Page 75
WARRANTY
Verathon® warrants the AortaScan AMI9700 against defects in material and workmanship as long as it is
covered by the Premium Warranty.
Verathon’s policy is to honor product warranties and to perform services only on products purchased from
an authorized Verathon entity. If you purchase a Verathon product or system components from unauthorized
entities or if the original factory serial number has been removed, defaced, altered, or if the product is
past its expiration date, your Verathon warranty will be invalidated. Purchasing Verathon products from
unauthorized entities could result in receipt of products or system components that are counterfeit, used,
expired, defective, or not intended for use in your region.
Pursuant to this warranty, a service center authorized by Verathon will repair or replace units that prove to be
defective during the warranty period.
This warranty does not apply if the unit was misused or modified by anyone other than an authorized
servicecenter.
The unit must be used in accordance with the instructions contained in this manual. Consumable items are
not covered in this warranty and should be used in conformance with Verathon product specifications, as
provided in the Product Specifications chapter.
For further details, consult your Premium Warranty. Warranty conditions may differ in some countries outside
the United States. Contact your local distributor for warranty terms.
DISCLAIMER OF ADDITIONAL WARRANTIES
There are no understandings, agreements, representations of warranties expressed or implied (including
warranties of merchantability or fitness for a particular purpose) other than those set forth in the preceding
Warranty section. The contents of this manual do not constitute a warranty.
Some states disallow certain limitations on applied warranties. The purchaser, user, and patient
should consult state law if there is a question regarding this disclaimer. This information, descriptions,
recommendations, and safety notations in this manual are based upon Verathon experience and judgment
with the AortaScan AMI9700 as of December 2013. The contents of this manual should not be considered
to be all‑inclusive, or to cover all contingencies.
The physician who directs the use of the AortaScan AMI9700 at the institution where it is in use is
responsible for keeping current with clinical research in aortic diameter measurements.
Please direct any questions or problems concerning aortic diameter measurement, using the instrument, or
the interpretation of data to the responsible physician.
Operations & Maintenance Manual: Warranty
67
Page 76
PRODUCT SPECIFICATIONS
COMPONENT SPECIFICATIONS
CONSOLE & PROBE SPECIFICATIONS
Table 20. General Specifications
ITEMSPECIFICATION
InputLithium‑ion battery.
OutputNo load to full load at rated voltage. Refer to unit label.
InsulationThe power supply is Class I with basic insulation to each terminal.
Transient overvoltageCategory II
Weight5.2 lb (2.36 kg) (with battery)
Display13.36 cm W x 10.13 cm H (5.26 in W x 3.99 in H) (640 x 480 pixels, 120 dpi)
Integrated printerThermal printer
Table 21. Ultrasound Acoustic Output Parameters
Values in this table are the maximum readings obtained from three test results.
Each scan point along the scan line consists of two transmit pulses. The first pulse is 1 cycle at 2.95 MHz and the second pulse is 5
cycles at 1.74 MHz. Data for each pulse is provided and separated by a comma.
68
Page 77
Table 22. Accuracy Specifications
SPECIFICATIONDESCRIPTION
Aorta diameter range3–12.4 cm
Diameter accuracy±(15%+0.5cm) on a Verathon® tissue‑equivalent phantom
The accuracy specifications assume the instrument is being used according to the instructions provided by
Verathon while scanning a tissue‑equivalent phantom.
The AortaScan AMI9700 is provided with two lithium‑ion batteries. A battery icon on the instrument display
is always present, indicating how much power remains and when the battery needs to be changed. You can
change the battery whenever necessary.
Removing a discharged battery and replacing it with a fresh battery should not erase any saved exams or user
settings. In the event any user settings change, reset them using the instructions in the Setting Up chapter.
Use only the battery charger provided with the AMI9700. Any other battery charger may damage the
battery.
Table 24. Battery Specifications
CONDITIONDESCRIPTION
Battery typeLithium‑ion
Battery life
Charging time
Rated capacity4800–5200 mAh
Normal voltage11.1 V
Max charging voltage12.6 V
Max weight350 g
Width3.11 in (79 mm)
Length4.65 in (118 mm)
Thickness0.91 in (23 mm)
A fully charged battery can provide approximately 30 exams within a
24‑hour period.
Charging time offline will take no more than six hours from an empty
battery to a full charge.
70
Page 79
BATTERY CHARGER/WIRELESS HUB SPECIFICATIONS
The battery charger/wireless hub is powered from a standard wall outlet (adaptable to international power
standards). The battery charger/wireless hub can charge two batteries simultaneously.
Computer connectionUSB 2.0
ChargerPowered by a desktop DC power supply.
Input voltage100–240 V AC RMS
Input frequency50–60 Hz
Input current1 A max
Input connection2 wire IEC 60320 C7
Output9V at 1A
InsulationClass II with double insulation
Fuses250 VAC, 2A, quick acting
Tes ting CSA 60950‑1‑03/UL 60950‑1
The Bluetooth® technology used in the AortaScan AMI9700 is compliant with:
• Bluetooth Specification as defined and approved by The Bluetooth Special Interests Group.
• Logo certification with Bluetooth wireless technology as defined by The Bluetooth Special Interest
Group.
ELECTROMAGNETIC COMPATIBILITY
The AortaScan AMI9700 system is designed to be in compliance with IEC 60601‑1‑2:2007, which contains
electromagnetic compatibility (EMC) requirements for medical electrical equipment. The limits for emissions
and immunity specified in this standard are designed to provide reasonable protection against harmful
interference in a typical medical installation.
The AortaScan AMI9700 system complies with the applicable essential performance requirements specified
in IEC 60601‑1 and IEC 60601‑2‑37. Results of immunity testing show that the essential performance of
the system is not affected under the test conditions described in the following tables. For more information
about the essential performance of the AortaScan AMI9700 system, see Essential Performance on
page2.
ELECTROMAGNETIC EMISSIONS
Table 26. Guidance and Manufacturer’s Declaration—Electromagnetic Emissions
The AortaScan AMI9700 system is intended for use in the electromagnetic environment specified below.
The customer or the user of the AortaScan AMI9700 system should assure that it is used in such an
environment.
Voltage fluctuations/
flicker emissions
IEC 61000 ‑3‑3
Group 1
Class A
Class A
Complies
The AortaScan AMI9700 system uses RF energy only for its
internal function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic equipment.
The AortaScan AMI9700 system is suitable for use in all
establishments other than domestic and those directly connected
to the public low‑voltage power supply network that supplies
buildings used for domestic purposes.
72
Page 81
ELECTROMAGNETIC IMMUNITY
Table 27. Guidance and Manufacturer’s Declaration —Electromagnetic Immunity
The AortaScan AMI9700 system is intended for use in the electromagnetic environment specified below.
The customer or the user of the AortaScan AMI9700 system should assure that it is used in such an
environment.
IMMUNITY TESTSIEC 60601 TEST LEVEL
Electrostatic discharge
(ESD)
IEC 61000‑4‑2
Electrical fast
transient/burst
IEC 61000‑4‑4
Surge
IEC 61000‑4‑5
± 6kV contact
± 8kV air
± 2kV for power supply
lines
± 1kV for input/output
lines
± 1kV line(s) to line(s)
± 2kV line(s) to earth
<5% Ut (>95% dip in Ut)
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000‑4‑11
for 0.5cycle
40% Ut (60% dip in Ut)
for 5cycles
70% Ut (30% dip in Ut)
for 25cycles
<5% Ut (>95% dip in Ut)
for 5s
Power frequency
(50/60Hz) magnetic
field
3A/mIn compliance
IEC 61000‑4‑8
Conducted RF
IEC61000‑4‑6
3Vrms
150kHz to 80MHz
COMPLIANCE
LEVEL
In compliance
In compliance
In compliance
In compliance
3V
ELECTROMAGNETIC ENVIRONMENT –
GUIDANCE
Floors should be wood, concrete,
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user of
the AortaScan AMI9700 system
requires continued operation during
power mains interruptions, it is
recommended that the AortaScan
AMI9700 system be powered from
an uninterruptible power supply or a
battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
Portable and mobile RF
communications equipment
should be used no closer to any
part of the AortaScan AMI9700
system, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Table 27. Guidance and Manufacturer’s Declaration —Electromagnetic Immunity
The AortaScan AMI9700 system is intended for use in the electromagnetic environment specified below.
The customer or the user of the AortaScan AMI9700 system should assure that it is used in such an
environment.
IMMUNITY TESTSIEC 60601 TEST LEVEL
Radiated RF
IEC61000‑4‑3
3V/m
80MHz to 2.5GHz
COMPLIANCE
LEVEL
3V/m
ELECTROMAGNETIC ENVIRONMENT –
GUIDANCE
d=1.2 √P 80MHz to 800MHz
d=2.3 √P 800MHz to 2.5GHz
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,a should
be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity
of equipment marked with the
following symbol:
Note: Ut is the AC mains voltage prior to application of the test level.
At 80MHz and 800MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which
the AortaScan AMI9700 system is used exceeds the applicable RF compliance level above, the AortaScan AMI9700 system should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re‑orienting or
relocating the AortaScan AMI9700 system.
b. Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.
74
Page 83
RECOMMENDED SEPARATION DISTANCES
Table 28. Recommended Separation Distances between Portable and Mobile RF Communications
Equipment and the AortaScan AMI9700 System
The AortaScan AMI9700 system is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the AortaScan AMI9700 system can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the AortaScan AMI9700 system as recommended below,
according to the maximum output power of the communications equipment.
RATED MAXIMUM
OUTPUT POWER OF
TRANSMITTER (W)
0.010.120 .120.23
0.10.380.380.73
11.21.22.3
103.83.87.3
100121223
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
SEPARATION DISTANCE ACCORDING TO FREQUENCY OF TRANSMITTER (m)
150 kHz to 80 MHz
d=1.2 √P
ACCESSORY CONFORMANCE TO STANDARDS
To maintain electromagnetic interference (EMI) within certified limits, the system must be used with the
cables, components, and accessories specified or supplied by Verathon®. For additional information, see
the System Components & Accessories and Component Specifications sections. The use of accessories or
cables other than those specified or supplied may result in increased emissions or decreased immunity of
thesystem.
80 MHz to 800 MHz
d=1.2 √P
800 MHz to 2.5 GHz
d=2.3 √P
Table 29. EMC Standards for Accessories
ACCESSORYMAX LENGTH
AC power cord2 m (6.6ft)
USB cable1.9m (6.2ft)
Desktop power supply—