Velleman DCM266 Operation Manual

DCM266 Digital Clamp meter
OPERATION MANUAL
GEBRUIKERSHANDLEIDING
MANUEL D’UTILISATEUR
DCM266 GB1
Digital Clamp Multimeter
1. Safety Information
The meter is completely portable, LCD, 3 1/2 digit clamp with insulation test function (with option 500V insulation tester unit). It has been designed according to IEC-1010 concerning electronic measuring instruments with an overvoltage category (CAT II) and pollution 2 and safety requirements for handheld current clamps for electrical measurement and test. Follow all safety and operating instructions to ensure that the meter is used safely and is kept in good condition.
1.1 Preliminary
l When using this meter, the user m ust observe all normal safety rules
concerning :
- Protection against the dangers of electronic current.
- Protection of the meter against misuse.
l Full compliance with safety standards can be guaranteed only if used with
test leads supplied. If necessary, they must be replaced with the same model or same electronic ratings. Measuring leads must be in good condition.
1.2 During Use
l Never exceed the protection limit values indicated in specifications for
each range of measurement.
l When the meter is linked to measurement circuit, do not touch unused
terminals.
l When the value scale to be measured is unknown beforehand, set the range
selector at the highest position.
l Before rotating the range selector to change function, disconnect test leads
from the circuit under test.
l When carrying out measurements on TV or switching power circuits,
always remember that there may be high amplitude voltage pulses at test points which can damage the me ter.
l Never perform resistance measurements on live circuits. l Always be careful when working with voltage above 60Vdc or 30V ac rms.
DCM266 GB2
l When the jaw are open and will be connected to the conductor under test,
your fingers must keep behind the barrier indicator.
l Never ground yourself when taking electrical measurements. Do not touch
exposed metal pipes, outlets, fixtures, etc. which might be at ground potential. Keep your body isolated from ground by using dry clothing, rubber shoes, or any approved insulating material.
1.3 Symbols
1.4 Maintenance
l Before opening the meter, always disconnect test leads from all sources of
electric current.
l If any faults or abnormalities are observed, the meter can not be used any
more and it has to be checked out.
l Never use the meter unless the back cover and the battery cover are in
place and fastened fully.
l Do not use abrasives or solvents on the meter, use a damp cloth and mild
detergent only.
Important safety information, refer to the operating manual
Dangerous voltage may be present
Earth ground
Double insulation (Protection class II)
DCM266 GB3
2. Description of the meter
This meter is a handheld 3 1/2 digital clamp meter for measuring DC and AC voltage, AC current, resistance, continuity test and insulation test. Full overload protection, low battery indication and over-range indication are provided.
Transformer jaws ó Trigger ì Data hold switch ö Rotary switch ú LCD display ÷ Input jacks ø Drop-proof wrist strap í Barrier indicator
DCM266 GB4
2.1 Function and range selector
A rotary switch is used to measurement Functions and Ranges. When the switch is set to OFF position, the meter does not operate.
2.2 Transformer jaws
Pick up the AC current flowing through the conductor. Press the TRIGGER to open the transformer jaws. When the finger press on the TRIGGER is released, the jaws will close again.
2.3 Data hold
A push switch (Push ON, Push OFF. A ll ACA, ACV, DCV ranges with this feature)
2.4 Input jacks
This meter has three input jacks that are protected against overload to the limits. During use connect the black test lead to COM jack and connect red test
lead to V jack. The red test lead is depended on function selected.
The EXT jack is used for accept insulation tester unit EXT banana plugs, when measurement insulation resistance.
DCM266 GB5
3. Operating instruction
3.1 Measuring current
1. Set the rotary switch at desired AC range position. The transformer jaws pick up the AC current flowing through the conductor.
2. When only the figure " 1" displayed, it indicates overrange situation and the higher range has to be selected.
3.2 Insulation test
(Option 500V insulation tester unit (not available in France and Benelux market))
1. Connect the insulation test unit V , COM, EXT three bananaplugs to the clamp meter V , COM, EXT.
2. Set the rotary switch of the clamp meter at 2000M position.
3. Set the insulation tester unit range switch to the 2000M position.
4. Use the insulation tester unit of the test leads connect its L, E input connect to being tested installation’s. ( Test installation’s must be power OFF.)
5. Set the insulation tester power switch to the ON position.
6. Depress the PUSH 500V push-push switch, the 500V on red LED lamp will light. Clamp meter display reading is the insulation resistance value. If the
reading is below 19M, change clamp meter and insulation tester unit to 20M range, can be increase the accuracy.
7. If the insulation tester unit is not used, the power switch must shift to power OFF position, and the test leads must leave the E, L input connect. That can be increase battery life and prevent electrical shock hazard.
Wrong Correct
DCM266 GB6
3.3 Measuring voltage
1. Connect the black test lead to the COM jack and the red test lead to the V jack.
2. Set the rotary switch at the desired DCV or AC V range position and connect test leads across the source or load under measurement. The polarity of the red lead connection will be indicated along with the voltage value when making DC voltage measurement.
3. When only the figure " 1" is displayed, it indicates overrange situation and the higher range has to be selected.
3.4 Measuring resistance
1. Connect the black test lead to the COM jack and the red test lead to the V jack.
2. Set the rotary switch at desired position and connect test leads across the resistor under measurement.
NOTE :
1. If the resistance being measured exceeds the maximum value of the range selected or the input is not connected, an overrange indication " 1" will be displayed.
2. When checking in-circuit resistance, be sure the circuit under test has all power removed and that all capacitors have been discharged fully.
3.5 Continuity test
1. Connect the black test lead to the COM jack and the red test lead to the V jack. ( The polarity of the red lead is positive " +")
2. Set the rotary switch at
position and connect test leads across two points
of the circuit under testing. If continuitiy exists ( i.e., resistance less than about 50 ), built-in buzzer will sound.
DCM266 GB7
4. Specifications
Accuracy is specified for a period of one year after calibration and at 18°C to 28°C (64°F to 82°F) with relative humidity to 80%.
4.1 General
DISPLAY 3 1/2 digit LCD, with automatic polarity
indication
MEASURING dual-slope integration A-D converter
system OVERRANGE INDICATION "1" figure only in the display MAXIMUM VOLTAGE BETWEEN TERMINALS AND EARTH GROUND
CAT II 600V
OPERATING TEMPERATURE 0°C to 40°C (32°F to 104°F) STORAGE ENVIRONMENT -10°C to 50°C (14°F to 122°F) POWER SUPPLY 9V battery ACCESSORIES operating manual, set of test leads LOW BATTERY INDICATION " BAT " appears on the display SIZE (H x W x L) 37 x 90 x 230 mm (1.46" x 3.54" x
9.06")
WEIGHT approx. 320g (including battery)
4.2 AC current
Range Resolution Accuracy Note
200A 100mA ±2.0% of rdg ± 5 digits 600A 1A ±2.0% of rdg ± 5 digits
Frequency Range : 50Hz to 60Hz
Response : Average, Calibrated in rms of sine wave
Overload protection : 1200A within 60 seconds. Jaw opening : 2" (5cm )
DCM266 GB8
4.3 Insulation test
With option 500V insulation tester unit.
Range Resolution Accuracy Note 20M 10K
±2.0% of (rdg-50) ± 2 digits
2000M 1M
±4.0% of (rdg-100) ± 2 digits
500M
±5.0% of (rdg-100) ± 2 digits
> 500M
4.4 AC Voltage
Range Resolution Accuracy
600V 1V ±1.0% of rdg ± 4 digits
Input Impedance : 9M on all ranges Overload protection : 600V peak or 600Vrms AC on all ranges Frequency Range : 50Hz to 500Hz Response : Average, calibrated in rms of sine wave
4.5 DC voltage
Range Resolution Accuracy
600V 1V ±0.8% of rdg ± 2 digits
Input impedance : >9M Over protection : 250Vrms AC for 200mV range
600V peak or 600V rms AC for other ranges
DCM266 GB9
4.6 Resistance
Range Resolution Accurcacy
200 0.1
±1.0% of rdg ± 3 digits
20K 10
±1.0% of rdg ± 1 digit
Overload protection : 250V dc or 250Vrms AC on all ranges Open circuit voltage : 700mV
5. Accessories
Test leads Battery Operating manual
6. Battery replacement
If the sign "BAT" appears on the LCD display, it indicates that battery should be replaced. Remove the battery cover of case. Replace the exhausted battery with a new one.
WARNING Before attempting to open the battery cover, be sure that test leads have been disconnected from measurement circuit to avoid electric shock hazard.