Operating Instructions
Submersible pressure transmitter with metal measuring cell
VEGABAR 87
4 … 20 mA
Document ID: 45507
Contents
Contents
1 Zu diesem Dokument............................................................................................................... |
4 |
|
1.1 |
Funktion............................................................................................................................ |
4 |
1.2 |
Zielgruppe......................................................................................................................... |
4 |
1.3 |
Verwendete Symbolik....................................................................................................... |
4 |
2 For your safety.......................................................................................................................... |
5 |
|
2.1 |
Authorised personnel........................................................................................................ |
5 |
2.2 |
Appropriate use................................................................................................................ |
5 |
2.3 |
Warning about incorrect use............................................................................................. |
5 |
2.4 |
General safety instructions................................................................................................ |
5 |
2.5 |
EU conformity................................................................................................................... |
5 |
2.6 |
NAMUR recommendations............................................................................................... |
6 |
2.7 |
Installation and operation in the USA and Canada............................................................ |
6 |
2.8 |
Environmental instructions................................................................................................ |
6 |
3 Product description.................................................................................................................. |
7 |
|
3.1 |
Configuration.................................................................................................................... |
7 |
3.2 |
Principle of operation........................................................................................................ |
8 |
3.3 |
Packaging, transport and storage................................................................................... |
10 |
3.4 |
Accessories.................................................................................................................... |
11 |
4 Mounting.................................................................................................................................. |
13 |
|
4.1 |
General instructions........................................................................................................ |
13 |
4.2 |
Ventilation and pressure compensation.......................................................................... |
15 |
4.3 |
Level measurement......................................................................................................... |
18 |
4.4 |
External housing............................................................................................................. |
18 |
5 Connecting to power supply................................................................................................. |
19 |
|
5.1 |
Preparing the connection................................................................................................ |
19 |
5.2 |
Connecting..................................................................................................................... |
20 |
5.3 |
Single chamber housing................................................................................................. |
21 |
5.4 |
Ex-d-ia double chamber housing.................................................................................... |
22 |
5.5 |
Housing IP66/IP68 (1 bar)............................................................................................... |
23 |
5.6 |
External housing............................................................................................................. |
24 |
5.7 |
Switch-on phase............................................................................................................. |
25 |
6 Set up with the display and adjustment module................................................................. |
26 |
|
6.1 |
Insert display and adjustment module............................................................................. |
26 |
6.2 |
Adjustment system.......................................................................................................... |
27 |
6.3 |
Measured value indication.............................................................................................. |
28 |
6.4 |
Parameter adjustment - Quick setup............................................................................... |
29 |
6.5 |
Parameter adjustment - Extended adjustment................................................................ |
29 |
6.6 |
Menu overview................................................................................................................ |
41 |
6.7 |
Saving the parameterisation data.................................................................................... |
42 |
7 Setup with PACTware............................................................................................................. |
44 |
|
7.1 |
Connect the PC............................................................................................................... |
44 |
7.2 |
Parameter adjustment with PACTware............................................................................ |
44 |
7.3 |
Saving the parameterisation data.................................................................................... |
45 |
8 Set up with other systems..................................................................................................... |
46 |
|
8.1 DD adjustment programs................................................................................................ |
46 |
|
|
|
|
2 |
VEGABAR 87 • 4 … 20 mA |
<![endif]>210430-EN-45507
<![endif]>45507-EN-210430
|
|
|
|
Contents |
|
|
8.2 |
Field Communicator 375, 475......................................................................................... |
46 |
||
9 |
Diagnostics and servicing..................................................................................................... |
47 |
|||
|
9.1 |
Maintenance................................................................................................................... |
47 |
||
|
9.2 |
Diagnosis function.......................................................................................................... |
47 |
||
|
9.3 |
Rectify faults................................................................................................................... |
49 |
||
|
9.4 |
Exchange process module on version IP68 (25 bar)....................................................... |
50 |
||
|
9.5 |
Exchanging the electronics module................................................................................ |
51 |
||
|
9.6 |
Software update.............................................................................................................. |
51 |
||
|
9.7 |
How to proceed if a repair is necessary.......................................................................... |
52 |
||
10 |
Dismount................................................................................................................................. |
53 |
|||
|
10.1 |
Dismounting steps.......................................................................................................... |
53 |
||
|
10.2 |
Disposal.......................................................................................................................... |
53 |
||
11 |
Supplement............................................................................................................................. |
54 |
|||
|
11.1 |
Technical data................................................................................................................. |
54 |
||
|
11.2 |
Calculation of the total deviation..................................................................................... |
63 |
||
|
11.3 |
Practical example............................................................................................................ |
64 |
||
|
11.4 |
Dimensions..................................................................................................................... |
66 |
||
|
11.5 |
Gewerbliche Schutzrechte.............................................................................................. |
71 |
||
|
11.6 |
Trademark....................................................................................................................... |
71 |
|
Safety instructions for Ex areas |
|
Take note of the Ex specific safety instructions for Ex applications. |
|
These instructions are attached as documents to each instrument |
|
with Ex approval and are part of the operating instructions. |
|
Editing status: 2021-03-31 |
|
|
VEGABAR 87 • 4 … 20 mA |
3 |
1 Zu diesem Dokument
Die vorliegende Anleitung liefert Ihnen die erforderlichen Informationen für Montage, Anschluss und Inbetriebnahme sowie wichtige Hinweise für Wartung, Störungsbeseitigung, den Austausch von Teilen und die Sicherheit des Anwenders. Lesen Sie diese deshalb vor der Inbetriebnahme und bewahren Sie sie als Produktbestandteil in unmittelbarer Nähe des Gerätes jederzeit zugänglich auf.
Diese Betriebsanleitung richtet sich an ausgebildetes Fachpersonal. Der Inhalt dieser Anleitung muss dem Fachpersonal zugänglich gemacht und umgesetzt werden.
Document ID
Dieses Symbol auf der Titelseite dieser Anleitung weist auf die Document ID hin. Durch Eingabe der Document ID auf www.vega.com kommen Sie zum Dokumenten-Download.
Information, Hinweis, Tipp: Dieses Symbol kennzeichnet hilfreiche Zusatzinformationen und Tipps für erfolgreiches Arbeiten.
Hinweis: Dieses Symbol kennzeichnet Hinweise zur Vermeidung von Störungen, Fehlfunktionen, Geräteoder Anlagenschäden.
Vorsicht: Nichtbeachten der mit diesem Symbol gekennzeichneten Informationen kann einen Personenschaden zur Folge haben.
Warnung: Nichtbeachten der mit diesem Symbol gekennzeichneten Informationen kann einen ernsthaften oder tödlichen Personenschaden zur Folge haben.
Gefahr: Nichtbeachten der mit diesem Symbol gekennzeichneten Informationen wird einen ernsthaften oder tödlichen Personenschaden zur Folge haben.
Ex-Anwendungen
Dieses Symbol kennzeichnet besondere Hinweise für Ex-Anwendun- gen.
•Liste
Der vorangestellte Punkt kennzeichnet eine Liste ohne zwingende Reihenfolge.
1Handlungsfolge
Vorangestellte Zahlen kennzeichnen aufeinander folgende Handlungsschritte.
Batterieentsorgung
Dieses Symbol kennzeichnet besondere Hinweise zur Entsorgung von Batterien und Akkus.
4 |
VEGABAR 87 • 4 … 20 mA |
<![endif]>210430-EN-45507
<![endif]>45507-EN-210430
2 For your safety
All operations described in this documentation must be carried out only by trained, qualified personnel authorised by the plant operator.
During work on and with the device, the required personal protective equipment must always be worn.
Model VEGABAR 87 is a pressure transmitter for level and gauge measurement.
You can find detailed information about the area of application in chapter " Product description".
Operational reliability is ensured only if the instrument is properly used according to the specifications in the operating instructions manual as well as possible supplementary instructions.
Inappropriate or incorrect use of this product can give rise to applica- tion-specific hazards, e.g.vessel overfill through incorrect mounting or adjustment. Damage to property and persons or environmental contamination can result. Also, the protective characteristics of the instrument can be impaired.
This is a state-of-the-art instrument complying with all prevailing regulations and directives. The instrument must only be operated in a technically flawless and reliable condition.The operator is responsible for the trouble-free operation of the instrument. When measuring aggressive or corrosive media that can cause a dangerous situation if the instrument malfunctions, the operator has to implement suitable measures to make sure the instrument is functioning properly.
The safety instructions in this operating instructions manual, the national installation standards as well as the valid safety regulations and accident prevention rules must be observed by the user.
For safety and warranty reasons, any invasive work on the device beyond that described in the operating instructions manual may be carried out only by personnel authorised by the manufacturer. Arbitrary conversions or modifications are explicitly forbidden.For safety reasons, only the accessory specified by the manufacturer must be used.
To avoid any danger, the safety approval markings and safety tips on the device must also be observed.
The device fulfils the legal requirements of the applicable EU directives.By affixing the CE marking, we confirm the conformity of the instrument with these directives.
VEGABAR 87 • 4 … 20 mA |
5 |
2 For your safety
The EU conformity declaration can be found on our homepage.
NAMUR is the automation technology user association in the process industry in Germany. The published NAMUR recommendations are accepted as the standard in field instrumentation.
The device fulfils the requirements of the following NAMUR recommendations:
•
•
•
•
NE 21 – Electromagnetic compatibility of equipment
NE 43 – Signal level for fault information from measuring transducers
NE 53 – Compatibility of field devices and display/adjustment components
NE 107 – Self-monitoring and diagnosis of field devices
For further information see www.namur.de.
This information is only valid for USA and Canada. Hence the following text is only available in the English language.
Installations in the US shall comply with the relevant requirements of the National Electrical Code (ANSI/NFPA 70).
Installations in Canada shall comply with the relevant requirements of the Canadian Electrical Code
A Class 2 power supply unit has to be used for the installation in the USA and Canada.
Protection of the environment is one of our most important duties. That is why we have introduced an environment management system with the goal of continuously improving company environmental protection.The environment management system is certified according to DIN EN ISO 14001.
Please help us fulfil this obligation by observing the environmental instructions in this manual:
•Chapter "Packaging, transport and storage"
•Chapter "Disposal"
<![if ! IE]><![endif]>210430-EN-45507
6 |
VEGABAR 87 • 4 … 20 mA |
|
|
|
3 Product description |
|
|
3 |
Product description |
|
|
3.1 Configuration |
|
Scope of delivery |
The scope of delivery encompasses: |
||
|
|
• |
VEGABAR 87 pressure transmitter |
|
|
• |
Ventilation valves, closing screws – depending on version (see |
|
|
|
chapter "Dimensions") |
The further scope of delivery encompasses:
•Documentation
––Quick setup guide VEGABAR 87
––Test certificate for pressure transmitters
––Instructions for optional instrument features
––Ex-specific "Safety instructions" (with Ex versions)
––If necessary, further certificates
Information:
Optional instrument features are also described in this operating instructions manual. The respective scope of delivery results from the order specification.
Scope of this operating |
|
This operating instructions manual applies to the following instrument |
|||
instructions |
|
versions: |
|||
|
|
|
|
• |
Hardware from 1.0.0 |
|
|
|
|
• |
Software from 1.3.5 |
|
|
|
|
Note: |
|
|
|
|
|
You can find the hardware and software version of the instrument as |
|
|
|
|
|
follows: |
|
|
|
|
|
||
|
|
|
|
• |
On the type plate of the electronics module |
|
|
|
|
• |
In the adjustment menu under "Info" |
Type label |
|
The type label contains the most important data for identification and |
|||
|
|
|
|
use of the instrument: |
<![endif]>45507-EN-210430
VEGABAR 87 • 4 … 20 mA |
7 |
3 Product description
Serial number - Instrument search
Application area
Measured products
Measured variables
8
1 |
2 |
|
7 |
3 |
6 |
|
5 |
4 |
Fig. 1: Layout of the type label (example)
1Product code
2Field for approvals
3Technical data
4Serial number of the instrument
5QR code
6Symbol of the device protection class
7ID numbers, instrument documentation
The type label contains the serial number of the instrument. With it you can find the following instrument data on our homepage:
•Product code (HTML)
•Delivery date (HTML)
•Order-specific instrument features (HTML)
•Operating instructions and quick setup guide at the time of shipment (PDF)
•Order-specific sensor data for an electronics exchange (XML)
•Test certificate (PDF) - optional
Move to "www.vega.com" and enter in the search field the serial number of your instrument.
Alternatively, you can access the data via your smartphone:
•Download the VEGA Tools app from the "Apple App Store" or the "Google Play Store"
•Scan the DataMatrix code on the type label of the instrument or
•Enter the serial number manually in the app
The VEGABAR 87 is a pressure transmitter for pressure and level measurements of liquids with higher temperatures in the chemical, food processing and pharmaceutical industry.
Measured products are liquids.
Depending on the instrument version and the measurement setup, the measured products can be also viscous.
The VEGABAR 87 is suitable for the measurement of the following process variables:
• Level
VEGABAR 87 • 4 … 20 mA
<![endif]>210430-EN-45507
3 Product description
<![endif]>45507-EN-210430
Fig. 2: Level measurement with VEGABAR 87
Electronic differential Depending on the version, the VEGABAR 87 is also suitable for pressure electronic differential pressure measurement.For this, the instrument
is combined with a Secondary Device.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Fig.3:Electronic differential pressure measurement via a Primary/Secondary |
||||||||
|
combination |
||||||||
|
You can find detailed information in the operating instructions of the |
||||||||
|
respective Secondary Device. |
||||||||
Measuring system |
The process pressure acts on the sensor element via the stainless |
||||||||
|
steel diaphragm and an internal transmission liquid. The process |
||||||||
|
|
|
|
|
|
|
|
|
|
VEGABAR 87 • 4 … 20 mA |
9 |
3 Product description
pressure causes a resistance change which is converted into a corresponding output signal and output as measured value.
The METEC® measuring cell is the measuring unit. It consists of the ceramic-capacitive CERTEC® measuring cell and a special, temperature-compensated isolating system.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
1 |
2 |
3 |
|
|
|
4 |
|
|
|
|
|
|
||||
|
Fig.4:Configuration of the METEC® measuring cell with VEGABAR 87 |
||||||||
|
1 |
Process diaphragm |
|
|
|
|
|
|
|
|
2 |
Isolating liquid |
|
|
|
|
|
|
|
|
3 |
FeNi adapter |
|
|
|
|
|
|
|
|
4 CERTEC® measuring cell |
|
|
|
|
|
|||
Pressure types |
The measuring cell design depends on the selected pressure type. |
||||||||
|
Relative pressure: the measuring cell is open to the atmosphere. |
||||||||
|
The ambient pressure is detected in the measuring cell and compen- |
||||||||
|
sated.It thus has no influence on the measured value. |
||||||||
|
Absolute pressure: the measuring cell contains vacuum and is |
||||||||
|
encapsulated. The ambient pressure is not compensated and does |
||||||||
|
hence influence the measured value. |
|
|||||||
|
Relative pressure, climate-compensated: the measuring cell is |
||||||||
|
evacuated and encapsulated. The ambient pressure is detected |
||||||||
|
through a reference sensor in the electronics and compensated. It |
||||||||
|
thus has no influence on the measured value. |
|
|||||||
Seal concept |
The measuring system is completely welded and hence sealed |
||||||||
|
against the process.The sealing of the process fitting against the |
||||||||
|
process is carried out by a seal provided on site. |
||||||||
|
3.3 Packaging, transport and storage |
||||||||
Packaging |
Your instrument was protected by packaging during transport. Its |
||||||||
|
capacity to handle normal loads during transport is assured by a test |
||||||||
|
based on ISO 4180. |
|
|
|
|
|
|
The packaging consists of environment-friendly, recyclable cardboard. For special versions, PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies.
10 |
VEGABAR 87 • 4 … 20 mA |
<![endif]>210430-EN-45507
<![endif]>45507-EN-210430
|
|
3 Product description |
Transport |
Transport must be carried out in due consideration of the notes on the |
|
|
|
transport packaging. Nonobservance of these instructions can cause |
|
|
damage to the device. |
Transport inspection |
The delivery must be checked for completeness and possible transit |
|
|
|
damage immediately at receipt. Ascertained transit damage or con- |
|
|
cealed defects must be appropriately dealt with. |
Storage
Storage and transport temperature
Up to the time of installation, the packages must be left closed and stored according to the orientation and storage markings on the outside.
Unless otherwise indicated, the packages must be stored only under the following conditions:
•Not in the open
•Dry and dust free
•Not exposed to corrosive media
•Protected against solar radiation
•Avoiding mechanical shock and vibration
•Storage and transport temperature see chapter "Supplement - Technical data - Ambient conditions"
•Relative humidity 20 … 85 %
Lifting and carrying |
With instrument weights of more than 18 kg (39.68 lbs) suitable and |
|
approved equipment must be used for lifting and carrying. |
|
3.4 Accessories |
|
The instructions for the listed accessories can be found in the down- |
|
load area on our homepage. |
PLICSCOM |
The display and adjustment module is used for measured value indi- |
|
cation, adjustment and diagnosis. |
|
The integrated Bluetooth module (optional) enables wireless adjust- |
|
ment via standard adjustment devices. |
VEGACONNECT |
The interface adapter VEGACONNECT enables the connection of |
|
communication-capable instruments to the USB interface of a PC. |
VEGADIS 82 |
The VEGADIS 82 is suitable for measured value indication of |
|
4 … 20 mA and 4 … 20 mA/HART sensors. It is looped into the signal |
|
cable. |
Overvoltage protection |
The overvoltage arrester B81-35 is used instead of the terminals in |
|
the single or double chamber housing. |
Protective cover |
The protective cover protects the sensor housing against soiling and |
|
intense heat from solar radiation. |
Flanges |
Screwed flanges are available in different versions according to the |
|
following standards: DIN 2501, EN 1092-1, BS 10, ASME B 16.5, |
|
JIS B 2210-1984, GOST 12821-80. |
|
|
VEGABAR 87 • 4 … 20 mA |
11 |
3 Product description
Welded socket, threaded Welded sockets are used to connect the devices to the process, and hygienic adapter threaded and hygienic adapters for easy adaptation of devices with
standard threaded fittings, e.g.to hygienic connections on the process side.
<![if ! IE]><![endif]>210430-EN-45507
12 |
VEGABAR 87 • 4 … 20 mA |
<![endif]>45507-EN-210430
4 Mounting
Process conditions
Protection against moisture
Screwing in
Vibrations
VEGABAR 87 • 4 … 20 mA
Note:
For safety reasons, the instrument must only be operated within the permissible process conditions.You can find detailed information on the process conditions in chapter "Technical data" of the operating instructions or on the type label.
Hence make sure before mounting that all parts of the instrument exposed to the process are suitable for the existing process conditions.
These are mainly:
•Active measuring component
•Process fitting
•Process seal
Process conditions in particular are:
•Process pressure
•Process temperature
•Chemical properties of the medium
•Abrasion and mechanical influences
Protect your instrument against moisture ingress through the following measures:
•Use a suitable connection cable (see chapter "Connecting to power supply")
•Tighten the cable gland or plug connector
•Lead the connection cable downward in front of the cable entry or plug connector
This applies mainly to outdoor installations, in areas where high humidity is expected (e.g. through cleaning processes) and on cooled or heated vessels.
Note:
Make sure that during installation or maintenance no moisture or dirt can get inside the instrument.
To maintain the housing protection, make sure that the housing lid is closed during operation and locked, if necessary.
Devices with threaded fitting are screwed into the process fitting with a suitable wrench via the hexagon.
See chapter "Dimensions" for wrench size.
Warning:
The housing or the electrical connection may not be used for screwing in! Depending on the device version, tightening can cause damage, e. g. to the rotation mechanism of the housing.
Avoid damages on the device by lateral forces, for example by vibrations.It is thus recommended to fix the devices with process fitting
13
4 Mounting
Permissible process pressure (MWP) - Device
Permissible process pressure (MWP) - Mounting accessory
Temperature limits
thread G½ of plastic at the installation site via a suitable measuring instrument holder.
If there is strong vibration at the mounting location, the instrument version with external housing should be used. See chapter "External housing".
The permissible process pressure range is specified on the type label with "MWP" (Maximum Working Pressure), see chapter "Configuration". This applies even if a measuring cell with a measuring range (order-related) higher than the permissible pressure range of the process fitting is installed.
In addition, a temperature derating of the process fitting, e.g.with flanges, can limit the permissible process pressure range according to the respective standard.
The permissible process pressure range is stated on the type label. The instrument should only be operated with these pressures if the mounting accessory used also fulfils these values.This should be ensured by suitable flanges, welded sockets, tension rings with Clamp connections, sealings, etc.
Higher process temperatures often mean also higher ambient temperatures. Make sure that the upper temperature limits stated in chapter "Technical data" for the environment of the electronics housing and connection cable are not exceeded.
2
1
Transport and mounting protection
Fig. 5: Temperature ranges
1Process temperature
2Ambient temperature
Depending on the transmitter, the VEGABAR 87 is supplied with a protective cap or a transport and mounting protection.
14 |
VEGABAR 87 • 4 … 20 mA |
<![endif]>210430-EN-45507
4 Mounting
1
2
Fig. 6: VEGABAR 87, transport and mounting protection
1Transmitter
2Transport and mounting protection
Remove this protection after mounting and before setting up the instrument.
In case of slightly contaminated measured media, the transport and mounting protection can remain on the instrument as an impact protection during operation.
Filter element - Function The filter element in the electronics housing has the following functions:
•Ventilation of the electronics housing
•Atmospheric pressure compensation (with relative pressure measuring ranges)
Caution:
The filter element causes a time-delayed pressure compensation.
When quickly opening/closing the housing cover, the measured value can change for approx. 5 s by up to 15 mbar.
For an effective ventilation, the filter element must be always free from buildup. In case of horizontal mounting, turn the housing so that the filter element points downward after the instrument is installed.This provides better protection against buildup.
Caution:
Do not use a high-pressure cleaner.The filter element could be damaged, which would allow moisture into the housing.
The following paragraphs describe how the filter element is arranged in the different instrument versions.
<![if ! IE]><![endif]>45507-EN-210430
VEGABAR 87 • 4 … 20 mA |
15 |
4 Mounting
Filter element - Position
Filter element - Position Ex-d version
16
6 |
6 |
6 |
1 |
2 |
3 |
6 |
|
6 |
4 |
|
5 |
Fig.7:Position of the filter element
1Plastic, stainless steel single chamber (precision casting)
2Aluminium - single chamber
3Stainless steel single chamber (electropolished)
4Plastic double chamber
5Aluminium, stainless steel double chamber housing (precision casting)
6Filter element
With the following instruments a blind plug is installed instead of the filter element:
•Instruments in protection IP66/IP68 (1 bar) - ventilation via capillaries in non-detachable cable
•Instruments with absolute pressure
→Turn the metal ring in such a way that the filter element points downward after installation of the instrument. This provides better protection against buildup.
<![if ! IE]><![endif]>210430-EN-45507
VEGABAR 87 • 4 … 20 mA
4 Mounting
1
2
Filter element - Position Second Line of Defense
Fig.8:Position of the filter element - Ex-d version
1Rotatable metal ring
2Filter element
With absolute pressure measuring ranges, a blind plug is used instead of the filter element.
The Second Line of Defense (SLOD) is a second level of the process separation in form of a gas-tight leadthrough in the housing neck, preventing products from penetrating into the housing.
With these instruments, the process assembly is completely encapsulated. An absolute pressure measuring cell is used so that no ventilation is required.
With relative pressure measuring ranges, the ambient pressure is detected and compensated by a reference sensor in the electronics.
1
1
2
2
<![endif]>45507-EN-210430
Fig.9:Position of the filter element - gastight leadthrough
1Filter element
2Gas-tight leadthrough
VEGABAR 87 • 4 … 20 mA |
17 |
4 Mounting
Filter element - Position IP69K version
1
Fig.10:Position of the filter element - IP69K version
1 Filter element
Instruments with absolute pressure have a blind plug mounted instead of the filter element.
|
4.3 |
Level measurement |
|
Measurement setup |
Keep the following in mind when setting up the measuring system: |
||
|
• |
Do not mount the instrument close to the filling stream or emptying |
|
|
area |
||
|
• |
Mount the instrument so that it is protected against pressure |
|
|
|
shocks from the stirrer |
|
Configuration |
4.4 |
External housing |
|
|
|
|
|
|
|
|
1 |
|
|
|
2 |
3
4
Fig. 11: Arrangement measurement loop, external housing
1Sensor
2Connection cable sensor, external housing
3External housing
4Signal cable
18 |
VEGABAR 87 • 4 … 20 mA |
<![endif]>210430-EN-45507
<![endif]>45507-EN-210430
|
|
|
5 Connecting to power supply |
|
|
|
5 |
Connecting to power supply |
|
|
|
5.1 Preparing the connection |
||
Safety instructions |
Always keep in mind the following safety instructions: |
|||
|
|
• |
Carry out electrical connection by trained, qualified personnel |
|
|
|
authorised by the plant operator |
||
|
|
• |
If overvoltage surges are expected, overvoltage arresters should |
|
|
|
|
be installed |
|
|
|
Warning: |
||
|
|
Only connect or disconnect in de-energized state. |
||
Voltage supply |
Power supply and current signal are carried on the same two-wire |
|||
|
|
cable.The operating voltage can differ depending on the instrument |
||
|
|
version. |
||
|
|
The data for power supply are specified in chapter "Technical data". |
||
|
|
Provide a reliable separation between the supply circuit and the |
||
|
|
mains circuits according to DIN EN 61140 VDE 0140-1. |
||
|
|
Power the instrument via an energy-limited circuit acc. to IEC 61010- |
||
|
|
1, e.g. via Class 2 power supply unit. |
||
|
|
Keep in mind the following additional factors that influence the operat- |
||
|
|
ing voltage: |
||
|
|
• |
Lower output voltage of the power supply unit under nominal load |
|
|
|
(e.g. with a sensor current of 20.5 mA or 22 mA in case of fault) |
||
|
|
• |
Influence of additional instruments in the circuit (see load values in |
|
|
|
|
chapter "Technical data") |
|
Connection cable |
The instrument is connected with standard two-wire cable without |
|||
|
|
shielding. If electromagnetic interference is expected which is above |
||
|
|
the test values of EN 61326-1 for industrial areas, shielded cable |
||
|
|
should be used. |
||
|
|
Use cable with round cross section for instruments with housing and |
||
|
|
cable gland. Use a cable gland suitable for the cable diameter to |
||
|
|
ensure the seal effect of the cable gland (IP protection rating). |
||
Cable screening and |
If shielded cable is required, we recommend connecting the cable |
|||
grounding |
screening on both ends to ground potential. In the sensor, the cable |
|||
|
|
screening must be connected directly to the internal ground terminal. |
||
|
|
The ground terminal on the outside of the housing must be connected |
||
|
|
to the ground potential (low impedance). |
||
|
|
In Ex systems, the grounding is carried out according to the installa- |
||
|
|
tion regulations. |
||
|
|
In electroplating plants as well as plants for cathodic corrosion protec- |
||
|
|
tion it must be taken into account that significant potential differences |
||
|
|
exist. This can lead to unacceptably high currents in the cable screen |
||
|
|
if it is grounded at both ends. |
||
|
|
|
||
VEGABAR 87 • 4 … 20 mA |
|
19 |
5 Connecting to power supply
Note:
The metallic parts of the instrument (process fitting, sensor, concentric tube, etc.) are connected with the internal and external ground terminal on the housing. This connection exists either directly via the conductive metallic parts or, in case of instruments with external electronics, via the screen of the special connection cable.
You can find specifications on the potential connections inside the instrument in chapter "Technical data".
Cable glands |
Metric threads |
|
In the case of instrument housings with metric thread, the cable |
|
glands are screwed in at the factory. They are sealed with plastic |
|
plugs as transport protection. |
|
|
|
|
Note: |
|
|
|
|
|
You have to remove these plugs before electrical connection. |
|
|
|
|
|
NPT thread |
|
|
|
|
|
In the case of instrument housings with self-sealing NPT threads, it is |
|
|
|
|
|
not possible to have the cable entries screwed in at the factory. The |
|
|
|
|
|
free openings for the cable glands are therefore covered with red dust |
|
|
|
|
|
protection caps as transport protection. |
|
|
|
|
|
Note: |
|
|
|
|
|
Prior to setup you have to replace these protective caps with ap- |
|
|
|
|
|
proved cable glands or close the openings with suitable blind plugs. |
|
|
|
|
|
On plastic housings, the NPT cable gland or the Conduit steel tube |
|
|
|
|
|
must be screwed into the threaded insert without grease. |
|
|
|
|
|
Max. torque for all housings, see chapter "Technical data". |
|
|
|
|
|
5.2 Connecting |
|
Connection technology |
|
The voltage supply and signal output are connected via the spring- |
|||
|
|
|
|
loaded terminals in the housing. |
|
|
|
|
|
Connection to the display and adjustment module or to the interface |
|
|
|
|
|
adapter is carried out via contact pins in the housing. |
|
|
|
|
|
Information: |
|
|
|
|
|
The terminal block is pluggable and can be removed from the |
|
|
|
|
|
electronics. To do this, lift the terminal block with a small screwdriver |
|
|
|
|
|
||
|
|
|
|
and pull it out. When reinserting the terminal block, you should hear it |
|
|
|
|
|
snap in. |
|
Connection procedure |
|
Proceed as follows: |
|||
|
|
|
|
1. Unscrew the housing lid |
|
|
|
|
|
2. If a display and adjustment module is installed, remove it by turn- |
|
|
|
|
|
ing it slightly to the left |
|
|
|
|
|
3. Loosen compression nut of the cable gland and remove blind |
|
|
|
|
|
plug |
|
|
|
|
|
4. Remove approx. 10 cm (4 in) of the cable mantle, strip approx. |
|
|
|
|
|
1 cm (0.4 in) of insulation from the ends of the individual wires |
|
|
|
|
|
|
|
20 |
|
|
|
VEGABAR 87 • 4 … 20 mA |
<![endif]>210430-EN-45507
5 Connecting to power supply
5. Insert the cable into the sensor through the cable entry
Fig. 12: Connection steps 5 and 6 - Single chamber housing
6. Insert the wire ends into the terminals according to the wiring plan
Information:
Solid cores as well as flexible cores with wire end sleeves are inserted directly into the terminal openings.In case of flexible cores without end sleeves, press the terminal from above with a small screwdriver, the terminal opening is then free. When the screwdriver is released, the terminal closes again.
7.Check the hold of the wires in the terminals by lightly pulling on them
8.Connect the shielding to the internal ground terminal, connect the external ground terminal to potential equalisation
9.Tighten the compression nut of the cable entry gland. The seal ring must completely encircle the cable
10.Reinsert the display and adjustment module, if one was installed
11.Screw the housing lid back on
The electrical connection is finished.
The following illustration applies to the non-Ex, Ex-ia and Ex-d version.
<![if ! IE]><![endif]>45507-EN-210430
VEGABAR 87 • 4 … 20 mA |
21 |
5 Connecting to power supply
Electronics and connection compartment
4...20mA |
|
(+)1 |
2(-) |
1 |
|
2
3
Fig. 13: Electronics and connection compartment - single chamber housing
1Voltage supply, signal output
2For display and adjustment module or interface adapter
3Ground terminal for connection of the cable screening
5.4Ex-d-ia double chamber housing
Electronics compartment
2
4...20mA
(+) |
2 |
(-) |
5 |
6 |
7 |
8 |
1 |
|
1 3
Fig. 14: Electronics compartment - Ex-d-ia double chamber housing
1Internal connection to the connection compartment
2For display and adjustment module or interface adapter
3Internal connection to the plug connector for external display and adjustment unit (optional)
Note:
HART multidrop mode is not possible when using an Ex-d-ia instrument.
<![if ! IE]><![endif]>210430-EN-45507
22 |
VEGABAR 87 • 4 … 20 mA |
5 Connecting to power supply
Connection compartment
Wire assignment, connection cable
4...20mA |
|
|
(+)1 |
2(-) |
2 |
1 |
|
|
Fig. 15: Connection compartment - Ex-d-ia double chamber housing
1Voltage supply, signal output
2Ground terminal for connection of the cable screening
1
2
Fig. 16: Wire assignment in permanently connected connection cable
1Brown (+) and blue (-) to power supply or to the processing system
2Shielding
<![endif]>45507-EN-210430
VEGABAR 87 • 4 … 20 mA |
23 |