TDR sensor for continuous level and
interface measurement of liquids
VEGAFLEX 81
Probus PA
Document ID: 44217
Quick start
Mounting
Electrical connection
Quick start
The quick start procedure enables a quick setup of the instrument
with many applications. You can nd further information in the respective chapters of the operating instructions manual.
1. Distance from the metallic vessel wall > 300 mm. Distance from
non-metallic vessel wall > 500 mm. The probe must not touch any
installations or the vessel wall.
2. In non-metallic vessels, place a metal sheet beneath the process
tting.
12
Fig. 1: Installation in non-metallic vessel
1 Flange
2 Metal sheet
3. If necessary, fasten probe end.
For further information see chapter "Mounting".
1. Make sure that the power supply corresponds to the specications on the type label.
2. Connect the instrument according to the following illustration
2
3
4
Bus
00
1
8
1
0
0
1
1
9
9
2
2
8
3
3
7
7
4
4
6
6
5
5
Set parameters
2
+
( )
(-)
1
2
5
678
5
1
Fig. 2: Electronics and connection compartment, single chamber housing
1 Voltage supply, signal output
2 For display and adjustment module or interface adapter
3 Selection switch for bus address
4 For external display and adjustment unit
5 Ground terminal for connection of the cable screen
For further information see chapter "Connecting to power supply".
For standard applications we recommend selecting the "Quick setup"
in the display and adjustment module.
VEGAFLEX 81 • Probus PA
44217-EN-130910
Quick start
1. In this menu item you can select the application. You can choose
between level and interface measurement.
2. In the menu item "Medium - Dielectric constant" you can dene
the type of medium (medium).
3. Carry out the adjustment in the menu items "Min. adjustment" and
"Max. adjustment".
4. A "Linearization" is recommended for all vessels in which the
vessel volume does not increase linearly with the level - e.g. in a
horizontal cylindrical or spherical tank. Activate the appropriate
curve.
5. A "False signal suppression" detects, marks and saves the false
signals so that they are no longer taken into account for level
measurement. We generally recommend a false signal suppression.
Parameterization example
44217-EN-130910
VEGAFLEX 81 • Probus PA
The sensor measures the distance from the sensor (reference plane)
to the product surface. See also chapter "Parameter adjustment".
1
4
2
Fig. 3: Measuring ranges - VEGAFLEX 81
1 Reference plane
2 Probe length L
3 Measuring range (default setting refers to the measuring range in water)
4 Upper dead band (in this area no measurement is possible)
5 Lower dead band (in this area no measurement is possible)
2
3
5
1
4
3
5
For this adjustment, the distance is entered when the vessel is full
and nearly empty. If these values are not known, an adjustment with
other distances, for example, 10 % and 90 % is also possible. Starting
3
Quick start
Further steps
point for these distance specications is always the seal surface of
the thread or ange.
1. In the menu "Additional settings", menu item "Damping" you can
adjust the requested damping of the output signal.
2. Select the parameter of the current output and the output characteristics in the menu item "Current output".
44217-EN-130910
4
VEGAFLEX 81 • Probus PA
Contents
1 About this document
1.1 Function ........................................................................................................................... 7
1.2 Target group ..................................................................................................................... 7
1.3 Symbolism used ............................................................................................................... 7
Please note the Ex-specic safety information for installation and operation in Ex areas. These safety instructions are part of the operating
instructions manual and come with the Ex-approved instruments.
Editing status: 2013-07-24
6
VEGAFLEX 81 • Probus PA
1 About this document
1 About this document
1.1 Function
This operating instructions manual provides all the information you
need for mounting, connection and setup as well as important instructions for maintenance and fault rectication. Please read this information before putting the instrument into operation and keep this manual
accessible in the immediate vicinity of the device.
1.2 Target group
This operating instructions manual is directed to trained specialist
personnel. The contents of this manual should be made available to
these personnel and put into practice by them.
1.3 Symbolism used
Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can result.
Warning: If this warning is ignored, injury to persons and/or serious
damage to the instrument can result.
Danger: If this warning is ignored, serious injury to persons and/or
destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
List
•
The dot set in front indicates a list with no implied sequence.
Action
→
This arrow indicates a single action.
1 Sequence of actions
Numbers set in front indicate successive steps in a procedure.
Battery disposal
This symbol indicates special information about the disposal of batteries and accumulators.
44217-EN-130910
VEGAFLEX 81 • Probus PA
7
2 For your safety
2 For your safety
2.1 Authorised personnel
All operations described in this operating instructions manual must
be carried out only by trained specialist personnel authorised by the
plant operator.
During work on and with the device the required personal protective
equipment must always be worn.
2.2 Appropriate use
VEGAFLEX 81 is a sensor for continuous level measurement.
You can nd detailed information on the application range in chapter
"Product description".
Operational reliability is ensured only if the instrument is properly
used according to the specications in the operating instructions
manual as well as possible supplementary instructions.
2.3 Warning about incorrect use
Inappropriate or incorrect use of the instrument can give rise to
application-specic hazards, e.g. vessel overll or damage to system
components through incorrect mounting or adjustment.
2.4 General safety instructions
This is a state-of-the-art instrument complying with all prevailing
regulations and guidelines. The instrument must only be operated in a
technically awless and reliable condition. The operator is responsible
for the trouble-free operation of the instrument.
During the entire duration of use, the user is obliged to determine the
compliance of the necessary occupational safety measures with the
current valid rules and regulations and also take note of new regulations.
The safety instructions in this operating instructions manual, the national installation standards as well as the valid safety regulations and
accident prevention rules must be observed by the user.
For safety and warranty reasons, any invasive work on the device
beyond that described in the operating instructions manual may be
carried out only by personnel authorised by the manufacturer. Arbitrary conversions or modications are explicitly forbidden.
The safety approval markings and safety tips on the device must also
be observed.
2.5 CE conformity
The device fullls the legal requirements of the applicable EC guidelines. By axing the CE marking, we conrm successful testing of the
product.
You can nd the CE Certicate of Conformity in the download section
of our homepage.
8
VEGAFLEX 81 • Probus PA
44217-EN-130910
2 For your safety
Electromagnetic compatibility
Instruments with plastic housing as well as in four-wire or Ex-d-ia
version are designed for use in an industrial environment. Nevertheless, electromagnetic interference from electrical conductors and
radiated emissions must be taken into account, as is usual with a
class A instrument according to EN 61326-1. If the instrument is used
in a dierent environment, the electromagnetic compatibility to other
instruments must be ensured by suitable measures.
2.6 NAMUR recommendations
NAMUR is the automation technology user association in the process
industry in Germany. The published NAMUR recommendations are
accepted as the standard in eld instrumentation.
The device fullls the requirements of the following NAMUR recom-
mendations:
NE 21 – Electromagnetic compatibility of equipment
•
NE 43 – Signal level for malfunction information from measuring
•
transducers
NE 53 – Compatibility of eld devices and display/adjustment
•
components
NE 107 – Self-monitoring and diagnosis of eld devices
•
For further information see www.namur.de.
2.7 Environmental instructions
Protection of the environment is one of our most important duties.
That is why we have introduced an environment management system
with the goal of continuously improving company environmental protection. The environment management system is certied according
to DIN EN ISO 14001.
Please help us fulll this obligation by observing the environmental
instructions in this manual:
Chapter "Packaging, transport and storage"
•
Chapter "Disposal"
•
44217-EN-130910
VEGAFLEX 81 • Probus PA
9
3 Product description
Type plate
3 Product description
3.1 Conguration
The nameplate contains the most important data for identication and
use of the instrument:
1
2
3
4
5
6
7
8
9
10
Fig. 4: Layout of the type label (example)
1 Instrument type
2 Product code
3 Approvals
4 Power supply and signal output, electronics
5 Protection rating
6 Probe length
7 Process and ambient temperature, process pressure
8 Material, wetted parts
9 Hardware and software version
10 Order number
11 Serial number of the instrument
12 Symbol of the device protection class
13 ID numbers, instrument documentation
14 Reminder to observe the instrument documentation
15 NotiedauthorityforCEmarking
16 Approval directives
16
15
14
13
12
11
Serial number
10
The type label contains the serial number of the instruments. Hence
you can nd the following data on our homepage:
Product code of the instrument (HTML)
•
Delivery date (HTML)
•
Order-specic instrument features (HTML)
•
Operating instructions at the time of shipment (PDF)
•
Order-specic sensor data for an electronics exchange (XML)
•
Test certicate pressure transmitters (PDF)
•
Go to www.vega.com, "VEGA Tools" and "Serial number search".
As an alternative, you can nd the data via your Smartphone:
Download the smartphone app "VEGA Tools" from the "Apple App
•
Store" or the "Google Play Store"
Scan the Data Matrix code on the type label of the instrument or
•
Enter the serial number manually in the app
•
VEGAFLEX 81 • Probus PA
44217-EN-130910
3 Product description
Scope of this operating
instructions manual
Versions
Scope of delivery
Application area
This operating instructions manual applies to the following instrument
versions:
Hardware from 1.0.0
•
Software from 1.0.0
•
Only for instrument versions without SIL qualication
•
This electronics version can be determined via the product code on
the type label as well as on the electronics.
Standard electronics: Type FX80PA.-
•
The scope of delivery encompasses:
Sensor
•
Documentation
•
– this operating instructions manual
– Test certicate measuring accuracy (optional)
– Operating instructions manual "Display and adjustment mod-
ule" (optional)
– Supplementary instructions "GSM/GPRS radio module"
(optional)
– Supplementary instructions manual "Heating for display and
adjustment module" (optional)
– Supplementary instructions manual "Plug connector for con-
tinuously measuring sensors" (optional)
– Ex-specic "Safety instructions" (with Ex versions)
– if necessary, further certicates
3.2 Principle of operation
The VEGAFLEX 81 is a level sensor with cable or rod probe for
continuous level or interface measurement, suitable for applications
in liquids.
Functional principle level measurement
44217-EN-130910
VEGAFLEX 81 • Probus PA
High frequency microwave pulses are guided along a steel cable or
a rod. Upon reaching the product surface, the microwave pulses are
reected. The running time is evaluated by the instrument and outputted as level.
To increase sensitivity, the probe is equipped with probe end tracking.
In products with a low dielectric constant, this function is very helpful.
This is the case, for example, in plastic granules, packing chips or in
vessels with uidized products.
Between a dielectric constant of 1.5 and 3, the function switches on, if
required. As soon as the level echo can no longer be detected, probe
end tracking is automatically activated. The measurement is continued with the last calculated dielectric constant.
The accuracy thus depends on the stability of the dielectric constant.
If you measure a medium with a dielectric constant below 1.5, probe
end tracking is always active. In this case, you have to enter the
dielectric constant of the medium. A stable dielectric constant is very
important here.
Functional principle - interface measurement
12
High frequency microwave impulses are guided along a steel cable or
rod. Upon reaching the product surface, a part of the microwave impulses is reected. The other part passes through the upper product
and is reected by the interface. The running times to the two product
layers are processed by the instrument.
d1 Distance to the interface
d2 Distance to the level
TS Thickness of the upper medium (d1 - d2)
h1 Height - Interface
h2 Height - Level
L1 Lower medium
L2 Upper medium
L3 Gas phase
Upper medium (L2)
The upper medium must not be conductive
•
The dielectric constant of the upper medium or the actual distance
•
to the interface must be known (input required). Min. dielectric constant: 1.6. You can nd a list of dielectric constants on our home
page: www.vega.com.
The composition of the upper medium must be stable, no varying
•
products or mixtures
The upper medium must be homogeneous, no stratications
•
within the medium
Min. thickness of the upper medium 50 mm (1.97 in)
•
Clear separation from the lower medium, emulsion phase or detri-
•
tus layer max. 50 mm (1.97 in)
If possible, no foam on the surface
•
L1
44217-EN-130910
VEGAFLEX 81 • Probus PA
Lower medium (L1)
The dielectric constant must be 10 higher than the dielectric
•
constant of the upper medium, preferably electrically conductive.
Example: upper medium dielectric constant 2, lower medium at
least dielectric constant 12.
Gas phase (L3)
Air or gas mixture
•
Gas phase - dependent on the application, gas pahse does not
•
always exist (d2 = 0)
13
3 Product description
Output signal
Packaging
Transport
Transport inspection
Storage
Storage and transport
temperature
The instrument is always preset to the application "Level measurement".
For the interface measurement, you can select the requested output
signal with the setup.
3.3 Packaging, transport and storage
Your instrument was protected by packaging during transport. Its
capacity to handle normal loads during transport is assured by a test
based on ISO 4180.
The packaging of standard instruments consists of environmentfriendly, recyclable cardboard. For special versions, PE foam or PE
foil is also used. Dispose of the packaging material via specialised
recycling companies.
Transport must be carried out in due consideration of the notes on the
transport packaging. Nonobservance of these instructions can cause
damage to the device.
The delivery must be checked for completeness and possible transit
damage immediately at receipt. Ascertained transit damage or concealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and
stored according to the orientation and storage markings on the
outside.
Unless otherwise indicated, the packages must be stored only under
the following conditions:
Not in the open
•
Dry and dust free
•
Not exposed to corrosive media
•
Protected against solar radiation
•
Avoiding mechanical shock and vibration
•
Storage and transport temperature see chapter "Supplement -
The display and adjustment module PLICSCOM is used for measured
value indication, adjustment and diagnosis. It can be inserted into the
sensor or the external display and adjustment unit and removed at
any time.
You can nd further information in the operating instructions "Display and adjustment module PLICSCOM" (Document-ID 27835).
The interface adapter VEGACONNECT enables the connection of
communication-capable instruments to the USB interface of a PC. For
parameter adjustment of these instruments, the adjustment software
PACTware with VEGA-DTM is required.
VEGAFLEX 81 • Probus PA
44217-EN-130910
3 Product description
You can nd further information in the operating instructions "Interface
adapter VEGACONNECT" (Document-ID 32628).
VEGADIS 81
PLICSMOBILE T61
Protective cap
Flanges
Electronics module
The VEGADIS 81 is an external display and adjustment unit for VEGA
plics® sensors.
For sensors with double chamber housing the interface adapter
"DISADAPT" is also required for VEGADIS 81.
You can nd further information in the operating instructions "VE-
GADIS 81" (Document-ID 43814).
The PLICSMOBILE T61 is an external GSM/GPRS radio unit for
transmission of measured values and for remote parameter adjustment of plics® sensors. The adjustment is carried out via PACTware/
DTM by using the integrated USB connection.
You can nd further information in the supplementary instructions
"PLICSMOBILE T61" (Document-ID 37700).
The protective cover protects the sensor housing against soiling and
intense heat from solar radiation.
You will nd additional information in the supplementary instructions
manual "Protective cover" (Document-ID 34296).
Screwed anges are available in dierent versions according to the
following standards: DIN 2501, EN 1092-1, BS 10, ANSI B 16.5,
JIS B 2210-1984, GOST 12821-80.
You can nd additional information in the supplementary instructions
manual "Flanges according to DIN-EN-ASME-JIS" (Document-ID
31088).
The electronics module VEGAFLEX series 80 is a replacement part
for TDR sensors of VEGAFLEX series 80. There is a dierent version
available for each type of signal output.
You can nd further information in the operating instructions manual
"Electronics module VEGAFLEX series 80".
Display and adjustment
module with heating
Rod extension
44217-EN-130910
VEGAFLEX 81 • Probus PA
The display and adjustment module can be optionally replaced by a
display and adjustment module with heating function.
You can use this display and adjustment module in an ambient tem-
perature range of -40 … +70 °C.
You can nd further information in the operating instructions "Display
and adjustment module with heating" (Document-ID 31708).
If you are using an instrument with rod version, you can extend the
rod probe individually with curved segments and rod extensions of
dierent lengths.
All extensions used must not exceed a total length of 6 m (19.7 ft).
The extensions are available in the following lengths:
Rod: ø 12 mm (0.472 in)
Basic segments: 20 … 5900 mm (0.79 … 232 in)
•
15
3 Product description
Rod segments: 20 … 5900 mm (0.79 … 232 in)
•
Curved segments: 100 x 100 mm (3.94 … 3.94 in)
•
You can nd further information in the operating instructions manual
"Rod extension VEGAFLEX series 80".
Bypass tube
Spacer
The combination of a bypass tube and a VEGAFLEX 81 enables con-
tinuous level measurement outside the vessel. The bypass consists
of a standpipe which is mounted as a communicating container on
the side of the vessel via two process ttings. This kind of mounting
ensures that the level in the standpipe and the level in the vessel are
the same.
The length and the process ttings can be congured individually. No
dierent connection versions available.
You can nd further information in the operating instructions manual
"Bypass tube VEGAPASS 81".
If you mount the VEGAFLEX 81 in a bypass tube or standpipe, you
have to avoid contact to the bypass tube by using a spacer at the
probe end.
You can nd additional information in the operating instructions
manual "Centering".
16
44217-EN-130910
VEGAFLEX 81 • Probus PA
Screwing in
4 Mounting
4 Mounting
4.1 General instructions
On instruments with process tting thread, the hexagon must be tightened with a suitable screwdriver. Wrench size see chapter "Dimen-sions".
Warning:
The housing must not be used to screw the instrument in! Applying
tightening force can damage internal parts of the housing.
Protection against moisture
Protective caps
Suitability for the process
conditions
Protect your instrument further through the following measures
against moisture penetration:
Use the recommended cable (see chapter "Connecting to power
•
supply")
Tighten the cable gland
•
Loop the connection cable downward in front of the cable gland
•
This applies particularly to:
Outdoor mounting
•
Installations in areas where high humidity is expected (e.g. through
•
cleaning processes)
Installations on cooled or heated vessels
•
In the case of instrument housings with self-sealing NPT threads, it is
not possible to have the cable entries screwed in at the factory. The
openings for the cable glands are therefore covered with red protective caps as transport protection.
Prior to setup you have to replace these protective caps with approved cable glands or close the openings with suitable blind plugs.
The suitable cable glands and blind plugs come with the instrument.
Make sure that all parts of the instrument exposed to the process are
suitable for the existing process conditions.
These are mainly:
Active measuring component
•
Process tting
•
Process seal
•
Process conditions are particularly:
Process pressure
•
Process temperature
•
Chemical properties of the medium
•
Abrasion and mechanical inuences
•
You can nd the specications of the process conditions in chapter
"Technical data" as well as on the nameplate.
Installation position
44217-EN-130910
VEGAFLEX 81 • Probus PA
4.2 Mounting instructions
Mount VEGAFLEX 81 in such a way that the distance to vessel installations or to the vessel wall is at least 300 mm (12 in). In non-metallic
17
4 Mounting
vessels, the distance to the vessel wall should be at least 500 mm
(19.7 in).
During operation, the probe must not touch any installations or the
vessel wall. If necessary, fasten the probe end.
In vessels with conical bottom it can be advantageous to mount the
sensor in the center of the vessel, as measurement is then possible
nearly down to the lowest point of the bottom. Keep in mind that
measurement all the way down to the tip of the probe may not be possible. The exact value of the min. distance (lower dead band) is stated
in chapter "Technical data".
Fig. 7: Vessel with conical bottom
Type of vessel
18
Plastic vessel/Glass vessel
The guided microwave principle requires a metal surface on the process tting. Therefore use in plastic vessels etc. an instrument version
with ange (from DN 50) or place a metal sheet (ø > 200 mm/8 in)
beneath the process tting when screwing it in.
Make sure that the plate has direct contact with the process tting.
When installing rod or cable probes in vessels without metal walls,
e.g. in plastic vessels, the measured value can be inuenced by
strong electromagnetic elds (emitted interference according to
EN 61326: class A). In this case, use a probe with coaxial version.
44217-EN-130910
VEGAFLEX 81 • Probus PA
12
dh
Fig. 8: Installation in non-metallic vessel
1 Flange
2 Metal sheet
4 Mounting
Socket
If possible, avoid sockets. Mount the sensor ush with the vessel top.
If this is not possible, use short sockets with small diameter.
Higher sockets or sockets with a bigger diameter can generally be
used. They can, however, increase the upper blocking distance (dead
band). Check if this is relevant for your measurement.
In such cases, always carry out a false signal suppression after installation. You can nd further information under "Setup procedure".
_
<
150
_
<
100
Fig. 9: Mounting socket
DN40 ... DN150
> DN150 ... DN200
d
h
When welding the socket, make sure that the socket is ush with the
vessel top.
44217-EN-130910
VEGAFLEX 81 • Probus PA
19
4 Mounting
12
Fig.10:Socketmustbeinstalledush
1 Unfavourable installation
2 Socketush-optimuminstallation
Welding work
Inowingmedium
Measuring range
Before beginning the welding work, remove the electronics module
from the sensor. By doing this, you avoid damage to the electronics
through inductive coupling.
Do not mount the instruments in or above the lling stream. Make sure
that you detect the product surface, not the inowing product.
The reference plane for the measuring range of the sensors is the
sealing surface of the thread or ange.
Keep in mind that a min. distance must be maintained below the refer-
ence plane and possibly also at the end of the probe - measurement
in these areas is not possible (dead band). The length of the cable
can be used all the way to the end only when measuring conductive
44217-EN-130910
20
VEGAFLEX 81 • Probus PA
4 Mounting
products. These blocking distances for dierent mediums are listed
in chapter "Technical data". Keep in mind for the adjustment that the
default setting for the measuring range refers to water.
Pressure
Standpipes or bypass
tubes
The process tting must be sealed if there is gauge or low pressure in
the vessel. Before use, check if the seal material is resistant against
the measured product and the process temperature.
The max. permissible pressure is specied in chapter "Technical data" or on the type label of the sensor.
Standpipes or bypass tubes are normally metal tubes with a diameter
of 30 … 200 mm (1.18 … 7.87 in). In measurement technology, such
a tube corresponds to a coax probe. It does not matter if the standpipe is perforated or slotted for better mixing. Lateral inlets in bypass
tubes also do not inuence the measurement.
For bypass tubes, select the probe length such that the blocking
distance (dead band) of the probe is above or below the lateral lling
openings. You can thus measure the complete range of the medium in
the bypass tube. When designing the bypass tube, keep the blocking
distance of the probe in mind and select the length above the upper
lateral lling opening accordingly.
Microwaves can penetrate many plastics. For process technical reasons, plastic standpipes are problematic. If durability is no problem,
then we recommend the use of metal standpipes.
When the VEGAFLEX 81 is used in standpipes or bypass tubes,
contact with the tube wall must be avoided. We recommend for this
purpose a cable probe with centering weight.
With rod probes, a spacer is generally not required. However, if there
is a risk of the rod probe being pressed against the tube wall by inowing medium, you should mount a spacer at the probe end to avoid
contact with the tube wall. In the case of cable probes, the cable can
be strained.
Keep in mind that buildup can form on the spacers. Strong buildup
can inuence the measurement.
44217-EN-130910
VEGAFLEX 81 • Probus PA
21
4 Mounting
Fasten
1
Fig. 12: Position of the spacer or centering weight
1 Rod probe with spacer (PEEK)
2 Cable probe with centering weight
3 Spacer (PEEK) on the gravity weight of a cable probe
Note:
Measurement in a standpipe is not recommended for extremely
adhesive products.
Instructions for the measurement:
The 100 % point should not be above the upper tube connection
•
to the vessel
The 0 % point should not be below the lower tube connection to
•
the vessel
A false signal suppression with installed sensor is generally rec-
•
ommended to achieve maximum possible accuracy.
If there is a risk of the cable probe touching the vessel wall during
operation due to product movements or agitators, etc., the measuring
probe should be securely xed.
In the gravity weight there is an internal thread (M8), e.g. for an eyebolt (optional) - (article no. 2.1512).
Make sure that the probe cable is not completely taut. Avoid tensile
loads on the cable.
Avoid undened vessel connections, i.e. the connection must be
either grounded reliably or isolated reliably. Any undened change of
this requirement can lead to measurement errors.
23
44217-EN-130910
22
VEGAFLEX 81 • Probus PA
4 Mounting
Lateral installation
1
1
2
2
Fig. 13: Fasten the probe
1 Probe
2 Retaining sleeve
In case of dicult installation conditions, the probe can be also
mounted laterally. For this purpose, adapt the rod with rod extensions
or bow-shaped segments.
Let the probe length determine automatically by the instrument to
compensate the resulting running time changes.
The determine probe length can deviate from the actual probe length
when using bow-shaped segments.
If installations such as struts, ladders, etc. exist on the vessel wall,
then the probe should have a distance to the vessel wall of at least
300 mm (11.81 in).
You can nd further information in the supplementary instructions of
the rod extension.
Rod extension
44217-EN-130910
VEGAFLEX 81 • Probus PA
In case of dicult installation conditions, for example in a socket, the
probe can be adapted respectively with a rod extension.
Let the probe length determine automatically by the instrument to
compensate the resulting running time changes.
You can nd further information in the supplementary instructions of
the rod extension.
23
5 Connecting to power supply
Safety instructions
5 Connecting to power supply
5.1 Preparing the connection
Always keep in mind the following safety instructions:
Connect only in the complete absence of line voltage
•
If overvoltage surges are expected, overvoltage arresters should
•
be installed
Voltage supply
Connection cable
Cable gland ½ NPT
Cable screening and
grounding
The voltage supply is provided by a Probus DP /PA segment coupler.
The voltage supply range can dier depending on the instrument
version. You can nd the data for voltage supply in chapter "Technical data".
Connection is made with screened cable according to the Probus
specication. Power supply and digital bus signal are carried over the
same two-wire connection cable.
Use cable with round cross-section. A cable outer diameter of
5 … 9 mm (0.2 … 0.35 in) ensures the seal eect of the cable gland.
If you are using cable with a dierent diameter or cross-section,
exchange the seal or use a suitable cable gland.
Please make sure that your installation is carried out according to the
Probus specication. In particular, make sure that the termination of
the bus is done with appropriate terminating resistors.
You can nd detailed information of the cable specication, installation and topology in the "ProbusPA-UserandInstallationGuide-
line" on www.probus.com.
With plastic housing, the NPT cable gland or the Conduit steel tube
must be screwed without grease into the threaded insert.
Max. torque for all housings see chapter "Technical data".
Make sure that the cable screen and grounding are carried out according to Fieldbus specication. We recommend to connect the
cable screen to ground potential on both ends.
With systems with potential equalisation, connect the cable screen
directly to ground potential at the power supply unit, in the connection
box and at the sensor. The screen in the sensor must be connected
directly to the internal ground terminal. The ground terminal outside
on the housing must be connected to the potential equalisation (low
impedance).
Connection technology
24
5.2 Connecting
The voltage supply and signal output are connected via the spring-
loaded terminals in the housing.
The connection to the display and adjustment module or to the inter-
face adapter is carried out via contact pins in the housing.
VEGAFLEX 81 • Probus PA
44217-EN-130910
5 Connecting to power supply
Information:
The terminal block is pluggable and can be removed from the
electronics. To do this, lift the terminal block with a small screwdriver
and pull it out. When reinserting the terminal block, you should hear it
snap in.
Connection procedure
Proceed as follows:
1. Unscrew the housing cover
2. If a display and adjustment module is installed, remove it by turning it slightly to the left.
3. Loosen compression nut of the cable entry
4. Remove approx. 10 cm (4 in) of the cable mantle, strip approx.
1 cm (0.4 in) of insulation from the ends of the individual wires
5. Insert the cable into the sensor through the cable entry
Fig. 14: Connection steps 5 and 6 - Single chamber housing
6. Insert the wire ends into the terminals according to the wiring plan
Information:
Solid cores as well as exible cores with wire end sleeves are inserted directly into the terminal openings. In case of exible cores without
end sleeves, press the terminal from above with a small screwdriver;
the terminal opening is freed. When the screwdriver is released, the
terminal closes again.
You can nd further information on the max. wire cross-section under
"Technical data/Electromechanical data"
7. Check the hold of the wires in the terminals by lightly pulling on
them
8. Connect the screen to the internal ground terminal, connect the
outer ground terminal to potential equalisation
9. Tighten the compression nut of the cable entry. The seal ring must
completely encircle the cable
10. Reinsert the display and adjustment module, if one was installed
11. Screw the housing cover back on
The electrical connection is hence nished.
5.3 Wiring plan, single chamber housing
The following illustration applies to the non-Ex, Ex-ia and Ex-d ver-
sion.
3
00
0
1
1
9
9
2
2
1
8
8
3
1
0
Bus
3
7
7
4
4
6
6
5
5
4
26
+
( )
(-)
1
2
5
678
5
1
Fig. 16: Electronics and connection compartment, single chamber housing
1 Voltage supply, signal output
2 For display and adjustment module or interface adapter
3 Selection switch for bus address
4 For external display and adjustment unit
5 Ground terminal for connection of the cable screen
5.4 Wiring plan, double chamber housing
The following illustrations apply to the non-Ex as well as to the Ex-ia
1 Internal connection to the connection compartment
2 Contact pins for the display and adjustment module or interface adapter
3 Selection switch for bus address
Information:
The connection of an external display and adjustment unit is not possible with this double chamber housing.
1 Internal connection to the connection compartment
2 Contact pins for the display and adjustment module or interface adapter
3 Selection switch for bus address
4 Internal connection to the plug connector for external display and adjust-
ment unit (optional)
3
7
7
4
4
6
6
5
5
(-)
2
5
678
41
28
Bus
( )
(-)
+
1
2
2
1
Fig. 21: Connection compartment, double chamber housing Ex d
1 Voltage supply, signal output
2 Ground terminal for connection of the cable screen
1 brown (+) and blue (-) to power supply or to the processing system
2 Shielding
29
5 Connecting to power supply
Supplementary electronics - Radio module
PLICSMOBILE
5.7 Supplementary electronics
The radio module PLICSMOBILE is an external GSM/GPRS radio
unit for transmission of measured values and for remote parameter
adjustment.
Instrument address
SIM-Card
( )
+
1
2
Status
Test
USB
(-)
1
Fig. 25: Radio module PLICSMOBILE integrated in the connection compartment
1 Voltage supply
You can nd detailed information on connection in the supplementary
instructions "PLICSMOBILE GSM/GPRS radio module".
5.8 Set instrument address
An address must be assigned to each Probus PA instrument. The
approved addresses are between 0 and 126. Each address must
only be assigned once in the Probus PA network. The sensor is only
recognized by the control system if the address is set correctly.
When the instrument is shipped, address 126 is adjusted. This address can be used for function test of the instrument and for connection to a Probus PA network. Then address must be changed to
integrate additional instruments.
The address setting is carried out either via:
The address selection switch in the electronics compartment of
•
the instrument (address setting via hardware)
The display and adjustment module (address setting via software)
•
PACTware/DTM (address setting via software)
•
Hardware addressing
30
The hardware addressing is eective if an address <126 is adjusted
with the address selection switches on the instrument. Hence the
software addressing is no longer eective, the adjusted hardware
address is valid.
44217-EN-130910
VEGAFLEX 81 • Probus PA
5 Connecting to power supply
1
2
3
00
0
1
1
9
9
2
2
1
8
8
3
1
0
Bus
( )
+
1
2
Fig. 26: Address selection switch
1 Addresses <100 (selection 0), addresses >100 (selection 1)
2 Decade of the address (selection 0 to 9)
3 Unit position of the address (selection 0 to 9)
3
7
7
4
4
6
6
5
5
(-)
5
678
Software addressing
The software addressing is only eective if address 126 or higher is
adjusted on the instrument with the address selection switches.
The addressing procedure is described in the operating instructions
manual "Display and adjustment module.
5.9 Switch-on phase
After VEGAFLEX 81 is connected to the bus system, the instrument
carries out a self-test for approx. 30 seconds. The following steps are
carried out:
Internal check of the electronics
•
Indication of the instrument type, hardware and software version,
•
measurement loop name on the display or PC
Indication of the status message "F 105 Determine measured
•
value" on the display or PC
Status byte goes briey to fault value
•
As soon as a plausible measured value is found, it is outputted to the
signal cable. The value corresponds to the actual level as well as the
settings already carried out, e.g. factory settings.
44217-EN-130910
VEGAFLEX 81 • Probus PA
31
6 Set up with the display and adjustment module
6 Set up with the display and adjustment
module
6.1 Insert display and adjustment module
The display and adjustment module can be inserted into the sensor
and removed any time. Four positions displaced by 90° can be selected. It is not necessary to interrupt the power supply.
Proceed as follows:
1. Unscrew the housing cover
2. Place the display and adjustment module in the requested position onto the electronics and turn to the right until it snaps in
3. Screw housing cover with inspection window tightly back on
Removal is carried out in reverse order.
The display and adjustment module is powered by the sensor, an ad-
ditional connection is not necessary.
32
Fig. 27: Insertion of the display and adjustment module with single chamber
housing into the electronics compartment
44217-EN-130910
VEGAFLEX 81 • Probus PA
6 Set up with the display and adjustment module
12
Fig. 28: Insertion of the display and adjustment module into the double chamber
housing
1 In the electronics compartment
2 In the connection compartment (with Ex-d-ia version not possible)
Note:
If you intend to retrot the instrument with a display and adjustment
module for continuous measured value indication, a higher cover with
an inspection glass is required.
Key functions
44217-EN-130910
VEGAFLEX 81 • Probus PA
6.2 Adjustment system
Fig. 29: Display and adjustment elements
1 LC display
2 Adjustment keys
[OK] key:
•
– Move to the menu overview
1
2
33
6 Set up with the display and adjustment module
– Conrm selected menu
– Edit parameter
– Save value
[->] key:
•
– Presentation, change measured value
– Select list entry
– Select editing position
[+] key:
•
– Change value of the parameter
[ESC] key:
•
– Interrupt input
– Jump to next higher menu
Adjustment system
Switch-on phase
Measured value indication
The sensor is adjusted via the four keys of the display and adjust-
ment module. The LC display indicates the individual menu items. The
functions of the individual keys are shown in the above illustration.
Approx. 10 minutes after the last pressing of a key, an automatic reset
to measured value indication is triggered. Any values not conrmed
with [OK] will not be saved.
After switching on, the VEGAFLEX 81 carries out a short self-test
where the device software is checked.
The output signal transmits a fault signal during the switch-on phase.
The following information is displayed on the display and adjustment
module during the startup procedure:
Instrument type
•
Device name
•
Software version (SW-Ver)
•
Hardware version (HW-Ver)
•
With the [->] key you can move between three dierent indication
modes.
In the rst view, the selected measured value is displayed in large
digits.
In the second view, the selected measured value and a correspond-
ing bar graph presentation are displayed.
In the third view, the selected measured value as well as a second
selectable value, e.g. the temperature, are displayed.
Quick setup
34
44217-EN-130910
6.3 Parameter adjustment - Quick setup
To quickly and easily adapt the sensor to the application, select
the menu item "Quick setup" in the start graphic on the display and
adjustment module.
VEGAFLEX 81 • Probus PA
General information
6 Set up with the display and adjustment module
You can nd "Extended adjustment" in the next sub-chapter.
Sensor address
In the rst menu item you have to enter a sensor address. The selection switches on the electronics module are preset to sensor address
126. This means that the sensor address can be changed via the
display and adjustment module.
If you set a sensor address with the selection switches which is lower
than 126, the set value is applicable. In such case, the address setting
via the display and adjustment module has no eect.
Measurement loop name
In the next menu item you can assign a suitable measurement loop
name. You can enter a name with max. 19 characters.
Application
In this menu item, you can select the application. You can choose
between level measurement and interface measurement. You can
also choose between measurement in a vessel or in a bypass or
standpipe.
Level measurement
44217-EN-130910
VEGAFLEX 81 • Probus PA
Medium - dielectric constant
In this menu item, you can dene the type of medium (product).
Max. adjustment
In this menu item, you can enter the max. adjustment for the level.
Enter the appropriate distance value in m (corresponding to the
percentage value) for the full vessel. The distance refers to the sensor
reference plane (seal surface of the process tting). Keep in mind that
the max. level must lie below the dead band.
Min. adjustment
In this menu item, you can enter the min. adjustment for the level.
Enter the suitable distance value in m for the empty vessel (e.g.
distance from the ange to the probe end) corresponding to the percentage value. The distance refers tot he sensor reference plane (seal
surface of the process tting).
35
6 Set up with the display and adjustment module
Interface measurement
Dielectric constant - upper medium
In this menu item, you can dene the type of medium (product).
Max. adjustment
In this menu item, you can enter the max. adjustment for the level.
Enter the appropriate distance value in m (corresponding to the
percentage value) for the full vessel. The distance refers to the sensor
reference plane (seal surface of the process tting). Keep in mind that
the max. level must lie below the dead band.
Min. adjustment
In this menu item, you can enter the min. adjustment for the level.
Enter the suitable distance value in m for the empty vessel (e.g.
distance from the ange to the probe end) corresponding to the percentage value. The distance refers tot he sensor reference plane (seal
surface of the process tting).
Max. adjustment - Interface
Carry out the max. adjustment for the interface.
To do this, enter the percentage value and the suitable distance value
in m for the full vessel.
Min. adjustment - Interface
Carry out the min. adjustment for the interface.
To do this, enter the percentage value and the suitable distance value
in m for the empty vessel.
Linearization
36
Linearization
A linearization is necessary for all vessels in which the vessel volume
does not increase linearly with the level - e.g. a horizontal cylindrical or spherical tank, when the indication or output of the volume is
required. Corresponding linearization curves are preprogrammed
for these vessels. They represent the correlation between the level
percentage and vessel volume.
The linearization applies for the measured value indication and the
current output. By activating the suitable curve, the percentage vessel
volume is displayed correctly.
VEGAFLEX 81 • Probus PA
44217-EN-130910
6 Set up with the display and adjustment module
False signal suppression
High sockets and internal vessel installations cause interfering reections and can inuence the measurement.
A false signal suppression detects, marks and saves these false
signals so that they are no longer taken into account for the level and
interface measurement. We generally recommend carrying out a false
signal suppression to achieve the best possible accuracy. This should
be done with the lowest possible level so that all potential interfering
reections can be detected.
Enter the actual distance from the sensor to the product surface.
All interfering signals in this section are detected by the sensor and
stored.
The instrument carries out an automatic false signal suppression
as soon as the probe is uncovered. The false signal suppression is
always updated.
AI FB1 Channel
In this menu item you can select the function of the rst Function
Block. AI stands for Analog Input.
With this you can adjust the value for the Primary Value (PV). Further
settings (SV, TV) must be carried out via PACTware.
Main menu
44217-EN-130910
VEGAFLEX 81 • Probus PA
6.4 Parameter adjustment - Extended adjustment
For technically demanding measurement loops you can carry out
extended settings in "Extended adjustment".
The main menu is divided into ve sections with the following func-
tions:
Setup: Settings, e.g. measurement loop name, medium, application,
vessel, adjustment, AI FB 1 Channel - Scaling - Damping, device
units, false signal suppression, linearization
Display: Language setting, settings for the measured value indication
as well as lighting
Diagnosis: Information, for example on the instrument status, pointer,
reliability, AI FB 1 simulation, echo curve
Info: Instrument name, hardware and software version, date of manufacture, instrument features
Note:
For optimum adjustment of the measurement, the individual submenu
items in the main menu item "Setup" should be selected one after
the other and provided with the correct parameters. If possible, go
through the items in the given sequence.
The procedure is described below.
The following submenu points are available:
The submenu points described below.
Setup - Instrument address
Hardware addressing
Software addressing
Setup - Measurement
loop name
An address must be assigned to each Probus PA instrument. Each
address may only be assigned once in the Probus PA network. The
sensor is only recognized by the control system if the address is set
correctly.
When the instrument is shipped, address 126 is adjusted. This address can be used for function test of the instrument and for connection to a Probus PA network. Then address must be changed to
integrate additional instruments.
The address setting is carried out either via:
The address selection switch in the electronics compartment of
•
the instrument (address setting via hardware)
The display and adjustment module (address setting via software)
•
PACTware/DTM (address setting via software)
•
The hardware addressing is eective if an address <126 is set
with the address selection switches on the electronics module of
VEGAFLEX 81. Software addressing thus has no eect - only the set
hardware address applies.
The software addressing is only eective if address 126 or higher is
adjusted on the instrument with the address selection switches.
Here you can assign a suitable measurement loop name. Push the
"OK" key to start the processing. With the "+" key you change the sign
and with the "->" key you jump to the next position.
You can enter names with max. 19 characters. The character set
comprises:
44217-EN-130910
38
VEGAFLEX 81 • Probus PA
6 Set up with the display and adjustment module
Capital letters from A … Z
•
Numbers from 0 … 9
•
Special characters + - / _ blanks
•
Setup - Units
Setup - Units (2)
Setup - Probe length
Setup - Application - Type
of medium
In this menu item you select the distance unit and the temperature
unit.
With the distance units you can choose between m, mm and ft and
with the temperature units betwenn °C, °F and K.
In this menu item, you select the unit of the Secondary Value (SV2).
It can be selected from the distance units such as for example m, mm
and ft.
In this menu item you can enter the probe length or have the length
determined automatically by the sensor system.
When choosing "Ye s ", then the probe length will be determined
automatically. When choosing "No", you can enter the probe length
manually.
In this menu item you can select which type of medium you want to
measure. You can choose between liquid or bulk solid.
Setup - Application - Application
44217-EN-130910
VEGAFLEX 81 • Probus PA
In this menu item, you can select the application. You can choose
between level measurement and interface measurement. You can
also choose between measurement in a vessel or in a bypass or
standpipe.
39
6 Set up with the display and adjustment module
Note:
The selection of the application has a considerable inuence on all
other menu items. Keep in mind that as you continue with the parameter adjustment, individual menu items are only optionally available.
You have the option of choosing the demonstration mode. This mode
is only suitable for test and demonstration purposes. In this mode, the
sensor ignores the parameters of the application and reacts immediately to each change.
This menu item is only available, if you have chosen interface meas-
urement under the menu item "Application". In this menu item you can
enter if there is a superimposed gas phase in your application.
Only set the function to "Ye s ", if the gas phase is permanently present.
This menu item is only available if you have selected interface meas-
urement under the menu item "Application". In this menu item you can
choose the type of medium of the upper medium.
44217-EN-130910
40
VEGAFLEX 81 • Probus PA
6 Set up with the display and adjustment module
You can enter the dielectric constant of the upper medium directly or
have the value determined by the instrument. To do this you have to
enter the measured or known distance to the interface.
Setup - Max. adjustment
Level
Setup - Min. adjustment
Level
In this menu item you can enter the max. adjustment for the level. With
interface measurement this is the maximum total level.
Adjust the requested percentage value with [+] and store with [OK].
Enter the appropriate distance value in m (corresponding to the
percentage value) for the full vessel. The distance refers to the sensor
reference plane (seal surface of the process tting). Keep in mind that
the max. level must lie below the dead band.
In this menu item you can enter the min. adjustment for the level. With
interface measurement this is the minimum total level.
Adjust the requested percentage value with [+] and store with [OK].
44217-EN-130910
VEGAFLEX 81 • Probus PA
Enter the suitable distance value in m for the empty vessel (e.g.
distance from the ange to the probe end) corresponding to the percentage value. The distance refers tot he sensor reference plane (seal
surface of the process tting).
41
6 Set up with the display and adjustment module
Setup - Max. adjustment Interface
Setup - Min. adjustment Interface
This menu item is only available if you have selected interface meas-
urement under the menu item "Application".
You can accept the adjustment of the level measurement also for the
interface measurement. If you select "Yes", the current setting will be
displayed.
If you have selected "No", you can enter the adjustment for the interface separately. Enter the requested percentage value.
For the full vessel, enter the distance value in m matching the percentage value.
This menu item is only available if you have selected interface meas-
urement under the menu item "Application". If you have selected "Yes"
in the previous menu item (accept adjustment of the level measurement), the current setting will be displayed.
Setup - False signal suppression
42
If you have selected "No", you can enter the adjustment for the interface measurement separately.
Enter the respective distance value in m for the empty vessel corresponding to the percentage value.
The following circumstances cause interfering reections and can
inuence the measurement:
High sockets
•
Vessel installations such as struts
•
Note:
A false signal suppression detects, marks and saves these false
signals so that they are no longer taken into account for the level and
interface measurement. We generally recommend carrying out a false
signal suppression to achieve the best possible accuracy. This should
VEGAFLEX 81 • Probus PA
44217-EN-130910
6 Set up with the display and adjustment module
be done with the lowest possible level so that all potential interfering
reections can be detected.
Proceed as follows:
Enter the actual distance from the sensor to the product surface.
All interfering signals in this section are detected by the sensor and
stored.
Note:
Check the distance to the product surface, because if an incorrect
(too large) value is entered, the existing level will be saved as a false
echo. The lling level would then no longer be detectable in this area.
If a false signal suppression has already been created in the sensor,
the following menu window appears when selecting "False signal suppression":
Setup - Linearization
44217-EN-130910
VEGAFLEX 81 • Probus PA
The instrument carries out an automatic false signal suppression
as soon as the probe is uncovered. The false signal suppression is
always updated.
The menu item "Delete" is used to completely delete an already created false signal suppression. This is useful if the saved false signal
suppression no longer matches the metrological conditions in the
vessel.
A linearization is necessary for all vessels in which the vessel volume
does not increase linearly with the level - e.g. a horizontal cylindrical or spherical tank, when the indication or output of the volume is
required. Corresponding linearization curves are preprogrammed
for these vessels. They represent the correlation between the level
percentage and vessel volume.
The linearization applies to the measured value indication and the
current output. By activating the appropriate curve, the volume percentage of the vessel is displayed correctly. If the volume should not
be displayed in percent but e.g. in l or kg, a scaling can be also set in
the menu item "Display".
43
6 Set up with the display and adjustment module
Warning:
If a linearization curve is selected, the measuring signal is no longer
necessarily linear to the lling height. This must be considered by the
user especially when adjusting the switching point on the limit signal
transmitter.
In the following, you have to enter the values for your vessel, for
example the vessel height and the socket correction.
For non-linear vessel forms, enter the vessel height und the socket
correction.
For the vessel height, you have to enter the total height of the vessel.
For the socket correction you have to enter the height of the socket
above the upper edge of the vessel. If the socket is lower than the upper edge of the vessel, this value can also be negative.
+ h
- h
Setup - AI FB1
44
D
Fig. 30: Vessel height und socket correction value
D Vessel height
+h Positive socket correction value
-h Negative socket correction value
44217-EN-130910
Since the adjustment is very comprehensive, the menu points of
Function Blocks 1 (FB1) were put together in a submenu.
VEGAFLEX 81 • Probus PA
6 Set up with the display and adjustment module
Setup - AI FB1 - Channel
Setup - AI FB1 - Scaling
unit
Setup - AI FB1 - Scaling
In menu item"Channel" you determine which measured value the
output refers to.
In menu item "Scaling unit" you dene the scaling variable and the
scaling unit for the level value on the display, e.g. volume in l.
In menu item "Scaling" you dene the scaling format on the display
and the scaling of the measured level values for 0 % and 100 %.
Level measured value min.
Measured level value max.
Setup - AI FB1 - Damping
Lock/release setup - Adjustment
44217-EN-130910
VEGAFLEX 81 • Probus PA
To damp process-dependent measured value uctuations, you can
set a time of 0 … 999 s in this menu item.
The damping applies to the level and interface measurement.
The default setting is a damping of 0 s.
In the menu item "Lock/unlock adjustment", you can protect the
sensor parameters against unauthorized modication. The PIN is
activated/deactivated permanently.
45
6 Set up with the display and adjustment module
The following adjustment functions are possible without entering the
PIN:
Select menu items and show data
•
Read data from the sensor into the display and adjustment mod-
•
ule.
Caution:
With active PIN, adjustment via PACTware/DTM as well as other
systems is also blocked.
You can change the PIN number under "Additional adjustments - PIN".
Display
Display - Menu language
Display - Displayed value
1
In the main menu point "Display", the individual submenu points
should be selected subsequently and provided with the correct
parameters to ensure the optimum adjustment of the display options.
The procedure is described in the following.
The following submenu points are available:
The submenu points described below.
This menu item enables the setting of the requested national lan-
guage.
In the delivery status, the sensor is set to the ordered national language.
In this menu item, you dene the indication of the measured value
on the display. You can display two dierent measured values. In this
menu item, you dene measured value 1.
44217-EN-130910
The default setting for the displayed value 1 is "Filling height Level".
Display - Displayed value
2
46
In this menu item, you dene the indication of the measured value
on the display. You can display two dierent measured values. In this
menu item, you dene measured value 2.
VEGAFLEX 81 • Probus PA
6 Set up with the display and adjustment module
The default setting for the displayed value 2 is the electronics temperature.
Display - Backlight
Diagnostics - Device
status
Diagnostics - Peak values
Distance
The optionally integrated background lighting can be adjusted via the
adjustment menu. The function depends on the height of the supply
voltage, see "Technical data".
The lighting is switched o in the delivery status.
In this menu item, the device status is displayed.
The respective min. and max. measured value is saved in the sen-
sor. The two values are displayed in the menu item "Peak values, distance".
If you have selected interface measurement under the menu item
"Setup - Application", the peak values of the interface measurement
are displayed in addition to the peak values of the level measurement.
In another window you can carry out a reset of the two peak values
separately.
Diagnostics - Peak values
Measurement certainty
44217-EN-130910
VEGAFLEX 81 • Probus PA
The respective min. and max. measured values are saved in the
sensor. The two values are displayed in the menu item "Peak values, measurement certainty".
The measurement can be inuenced by the process conditions. In
this menu item, the measurement certainty of the level measurement
is displayed as percentage value. The higher the value, the more reliable the measurement. Values > 90 % indicate reliable measurement.
If you have selected interface measurement under the menu item
"Setup - Application", the peak values of the interface measurement
are displayed in addition to the peak values of the level measurement.
47
6 Set up with the display and adjustment module
In another window you can carry out a reset of the two peak values
separately.
Diagnostics - Peak values
Additional
Diagnostics - Echo curve
The respective min. and max. measured values are saved in the
sensor. The values are displayed in the menu item "Peak values Ad-ditional".
This menu item displays the peak values of the electronics temperature as well as the dielectric constant.
In another window you can carry out a reset of the two peak values
separately.
The menu item "Echo curve" shows the signal strength of the echoes
over the measuring range in V. The signal strength enables an evaluation of the quality of the measurement.
With the following functions you can zoom part sections of the echo
curve.
"X-Zoom": Zoom function for the meas. distance
•
"Y-Zoom": 1, 2, 5 and 10x signal magnication in "V"
•
"Unzoom": Reset the presentation to the nominal measuring range
•
with single magnication
Diagnosis - Simulation
48
44217-EN-130910
In this menu item you can simulate measured values via the current
output. This allows the signal path to be tested, e.g. via downstream
indicating instruments or the input card of the control system.
VEGAFLEX 81 • Probus PA
6 Set up with the display and adjustment module
Select the requested simulation size and adjust the requested value.
Caution:
During simulation, the simulated value is outputted as 4 … 20 mA current value and digital HART signal.
Push the [ESC] key to deactivate the simulation.
Information:
The simulation is terminated automatically 60 minutes after the last
key has been pushed.
Diagnostics - Echo curve
memory
Additional settings - PIN
With the menu item "Setup" the echo curve it is possible to save at
the time of setup. This is generally recommended; for using the Asset
Management functions it is necessary. If possible, the curve should
be saved with a low level in the vessel.
With this, you can detect signal changes over the operating time. With
the adjustment software PACTware and the PC, the high-resolution
echo curve can be displayed and used to compare the echo curve of
the setup with the actual echo curve.
The function "Echo curve memory" enables storing echo curves of
the measurement.
Under the sub-menu item "Echo curve memory" you can store the
current echo curve.
Parameter settings for recording the echo curve and the settings of
the echo curve itself can be carried out in the adjustment software
PACTwa r e.
With the adjustment software PACTware and the PC the high-resolution echo curve can be displayed and used later on to assess the
quality of the measurement.
Entering a 4-digit PIN protects the sensor data against unauthorized
access and unintentional modication. In this menu item, the PIN is
displayed or edited and changed. However, this menu item is only
44217-EN-130910
VEGAFLEX 81 • Probus PA
49
6 Set up with the display and adjustment module
available if adjustment is enabled in the menu "Lock/Release setup/adjustment".
In delivery status, the PIN is "0000".
Additional adjustments Date Time
Additional adjustments
- Reset
In this menu item, the internal clock of the sensor is adjusted.
With a reset, certain parameter adjustments carried out by the user
are reset.
The following reset functions are available:
Delivery status: Restoring the parameter settings at the time of ship-
ment from the factory incl. the order-specic settings. A created false
signal suppression, user-programmable linearization curve as well as
the measured value memory will be deleted.
Basic settings: Resetting of the parameter settings incl. special
parameters to the default values (presettings) of the respective instrument. Any created false signal suppression or user-programmable linearization curve as well as the measured value memory are deleted.
The following table shows the default values of the instrument. Depending on the instrument version or application, all menu items may
not be available or some may be dierently assigned:
Min. adjustment - InterfaceDistance: Probe length - take dead
Integration time - Level0.0 s
Integration time - Interface0.0 s
Linearization typeLinear
Linearization - Socket correction0 mm
Linearization - Vessel heightProbe length
ing distances
band into account
ing distances
band into account
AI FB1 Tag Descriptor
AI FB1 ChannelPrimary Value (lin. percent level)
AI FB1 scaling PV Scale (min.)0 %
AI FB1 scaling PV Scale (max.)100 %
AI FB1 Lin. TypeLinear
AI FB1 Out Scale Unit%
AI FB1 Out Scale Decimal Point#.##
AI FB1 Out Scale (min.)0 %
AI FB1 Out Scale (max.)100 %
AI FB1 PV FTime0 s
AI FB1 Hi Hi Limit3.402823E+38 %
AI FB1 Hi Limit3.402823E+38 %
AI FB1 Lo Lo Limit-3.402823E+38 %
AI FB1 Lo Limit-3.402823E+38 %
AI FB1 Hysteresis0.50 %
44217-EN-130910
VEGAFLEX 81 • Probus PA
51
6 Set up with the display and adjustment module
Menu itemDefault valueModiedvalue
AI FB1 Fail Safe Mode (behaviour in case of malfunction)
AI FB1 Fail Safe Value0.00 %
AI FB1 Target ModeAuto
Last Valid Out Value (last valid
measured value)
Display
Menu itemDefault valueModiedvalue
LanguageOrder-specic
Displayed value 1Filling height Level
Displayed value 2Electronics temperature
BacklightSwitched o
Diagnostics
Menu itemDefault valueModiedvalue
Status signals - Function controlSwitched on
Status signals - Out of specicationSwitched o
Status signals - MaintenanceSwitched o
Device memory - Echo curve memoryStopped
Device memory - Measured value memoryStarted
Device memory - Measured value memory Measured values
Device memory - Measured value memory - Recording in time interval
Device memory - Measured value memory - Recording with measured value dierence
Device memory - Measured value memory - Start
with measured value
Device memory - Measured value memory - Stop
with measured value
Device memory - Measured value memory - Stop
recording when memory is full
Distance level, percentage value
level, reliability level, electronics
temperature
3 min.
15 %
Not active
Not active
Not active
Additional adjustments
Menu itemDefault valueModiedvalue
PIN0000
DateActual date
TimeActual time
Time - Format24 hours
Probe typeDevice-specic
52
VEGAFLEX 81 • Probus PA
44217-EN-130910
6 Set up with the display and adjustment module
Menu itemDefault valueModiedvalue
AI FB2 Tag Descriptor
AI FB 2 ChannelPrimary Value (lin. percent level)
AI FB2 scaling PV Scale (min.)0 %
AI FB2 scaling PV Scale (max.)100 %
AI FB2 Lin. TypeLinear
AI FB2 Out Scale Unit%
AI FB2 Out Scale Decimal Point#.##
AI FB2 Out Scale (min.)0 %
AI FB2 Out Scale (max.)100 %
AI FB2 PV FTime0 s
AI FB2 Hi Hi Limit3.402823E+38 %
AI FB2 Hi Limit3.402823E+38 %
AI FB2 Lo Lo Limit-3.402823E+38 %
AI FB2 Lo Limit-3.402823E+38 %
AI FB2 Hysteresis0.50 %
AI FB2 Fail Safe Mode (behaviour in case of malfunction)
AI FB2 Fail Safe Value0.00 %
AI FB2 Target ModeAuto
Last Valid Out Value (last valid
measured value)
AI FB3 Tag Descriptor
AI FB3 ChannelPrimary Value (lin. percent level)
AI FB1 scaling PV Scale (min.)0 %
AI FB3 scaling PV Scale (max.)100 %
AI FB3 Lin. TypeLinear
AI FB3 Out Scale Unit%
AI FB3 Out Scale Decimal Point#.##
AI FB3 Out Scale (min.)0 %
AI FB3 Out Scale (max.)100 %
AI FB3 PV FTime0 s
AI FB3 Hi Hi Limit3.402823E+38 %
AI FB3 Hi Limit3.402823E+38 %
AI FB3 Lo Lo Limit-3.402823E+38 %
AI FB3 Lo Limit-3.402823E+38 %
AI FB3 Hysteresis0.50 %
AI FB3 Fail Safe Mode (behaviour in case of malfunction)
AI FB3 Fail Safe Value0.00 %
Last Valid Out Value (last valid
measured value)
44217-EN-130910
VEGAFLEX 81 • Probus PA
53
6 Set up with the display and adjustment module
Menu itemDefault valueModiedvalue
AI FB3 Target ModeAuto
Additional adjustments Copy instrument settings
Additional adjustments Probe type
The instrument settings are copied with this function. The following
functions are available:
Read from sensor: Read data from sensor and store into the
•
display and adjustment module
Write into sensor: Store data from the display and adjustment
•
module back to the sensor
The following data or settings for adjustment of the display and adjustment module are saved:
All data of the menu "Setup" and "Display"
•
In the menu "Additional adjustments" the items "Reset, Date/Time"
•
Special parameters
•
The copied data are permanently saved in an EEPROM memory in
the display and adjustment module and remain there even in case of
power failure. From there, they can be written into one or more sensors or kept as backup for a possible electronics exchange.
Note:
Before the data are stored in the sensor, a check is carried out to
determine if the data t the sensor. If the data do not t, a fault signal
is triggered or the function is blocked. When data are being written
into the sensor, the display shows which instrument type the data
originate from and which TAG-no. this sensor had.
In this menu item you can select the type and size of your probe from
a list of all possible probes. This is necessary to adapt the electronics
optimally to the probe.
Additional adjustments Special parameters
54
In this menu item you gain access to the protected area where
you can enter special parameters. In exceptional cases, individual
parameters can be modied in order to adapt the sensor to special
requirements.
Change the settings of the special parameters only after having contacted our service sta.
VEGAFLEX 81 • Probus PA
44217-EN-130910
6 Set up with the display and adjustment module
Info - Instrument name
Info - Instrument version
Info - Factory calibration
date
Info - Sensor characteristics
In this menu, you read out the instrument name and the instrument
serial number:
In this menu item, the hardware and software version of the sensor is
displayed.
In this menu item, the date of factory calibration of the sensor as well
as the date of the last change of sensor parameters are displayed via
the display and adjustment module or via the PC.
In this menu item, the features of the sensor such as approval, process tting, seal, measuring range, electronics, housing and others
are displayed.
44217-EN-130910
VEGAFLEX 81 • Probus PA
6.5 Saving the parameter adjustment data
We recommended noting the adjusted data, e.g. in this operating
instructions manual, and archiving them afterwards. They are thus
available for multiple use or service purposes.
If the instrument is equipped with a display and adjustment module,
the data in the sensor can be saved in the display and adjustment
module. The procedure is described in the operating instructions
manual "Display and adjustment module" in the menu item "Copy sensor data". The data remain there permanently even if the sensor
power supply fails.
The following data or settings for adjustment of the display and adjustment module are saved:
All data of the menu "Setup" and "Display"
•
In the menu "Additional adjustments" the items "Sensor-specic
•
units, temperature unit and linearization"
The values of the user programmable linearization curve
•
The function can also be used to transfer settings from one instrument to another instrument of the same type. If it is necessary to
exchange a sensor, the display and adjustment module is inserted
55
6 Set up with the display and adjustment module
into the replacement instrument and the data are likewise written into
the sensor via the menu item "Copy sensor data".
56
44217-EN-130910
VEGAFLEX 81 • Probus PA
2
Via the interface adapter
directly on the sensor
Prerequisites
7 Setup with PACTware
7 Setup with PACTware
7.1 Connect the PC
1
3
Fig. 31: Connection of the PC directly to the sensor via the interface adapter
1 USB cable to the PC
2 Interface adapter VEGACONNECT
3 Sensor
7.2 Parameter adjustment with PACTware
For parameter adjustment of the sensor via a Windows PC, the conguration software PACTware and a suitable instrument driver (DTM)
according to FDT standard are required. The up-to-date PACTware
version as well as all available DTMs are compiled in a DTM Collection. The DTMs can also be integrated into other frame applications
according to FDT standard.
Note:
To ensure that all instrument functions are supported, you should
always use the latest DTM Collection. Furthermore, not all described
functions are included in older rmware versions. You can download
the latest instrument software from our homepage. A description of
the update procedure is also available in the Internet.
Further setup steps are described in the operating instructions manual "DTM Collection/PACTware" attached to each DTM Collection and
which can also be downloaded from the Internet. Detailed descriptions are available in the online help of PACTware and the DTMs.
44217-EN-130910
VEGAFLEX 81 • Probus PA
57
7 Setup with PACTware
Fig. 32: Example of a DTM view
Standard/Full version
General information
All device DTMs are available as a free-of-charge standard version
and as a full version that must be purchased. In the standard version,
all functions for complete setup are already included. An assistant for
simple project conguration simplies the adjustment considerably.
Saving/printing the project as well as import/export functions are also
part of the standard version.
In the full version there is also an extended print function for complete
project documentation as well as a save function for measured value
and echo curves. In addition, there is a tank calculation program as
well as a multiviewer for display and analysis of the saved measured
value and echo curves.
The standard version is available as a download under
www.vega.com/downloads and "Software". The full version is available on CD from the agency serving you.
7.3 Set up with the quick setup
The quick setup is another option for parameter adjustment of the
sensor. It allows fast, convenient adjustment of the most important
parameters to adapt the sensor quickly to standard applications. To
use it, select the function "Quick setup" in the start screen.
44217-EN-130910
58
VEGAFLEX 81 • Probus PA
7 Setup with PACTware
Fig. 33: Select quick setup
1 Quick setup
2 Extended adjustment
3 Maintenance
Quick setup
With quick setup you can carry out the parameter adjustment of
VEGAFLEX 81 for your application in just a few simple steps. The
assistant-driven adjustment includes the basic settings for simple,
reliable setup and commissioning.
Information:
If the function is inactive, then possibly no instrument is connected.
Check the connection to the instrument.
Extended adjustment
With the extended adjustment, you carry out the parameter adjustment for the instrument via the clear menu structure in the DTM
(Device Type Manager). This enables additional and special settings
over and above those oered by quick setup.
Maintenance
Under the menu item "Maintenance" you get comprehensive and
important support for servicing and maintenance. You can call up
diagnostic functions and carry out an electronics exchange or a
software update.
Start quick setup
Step 1
Instrumentconguration
44217-EN-130910
VEGAFLEX 81 • Probus PA
Click to the button "Quick setup", to start the assistant-driven adjustment for a simplied and reliable setup.
Device name
Here, you can nd the instrument name. You cannot change this line
because the instrument name is unmodiably saved in the instrument.
59
7 Setup with PACTware
Serial number
Here, you can nd the serial number of your instrument. You cannot
change this line because the serial number is unmodiably saved in
the instrument.
Measurement loop name
Here you can enter a suitable measurement loop name for your VEGAFLEX 81. You can enter a name with max. 19 characters. You can
use capital and small letters as well as numbers. The following special
characters are also possible: + - . : , ( ) / < >
Probelengthmodied?
If you have modied the probe length, this must be entered in the
selection eld.
If you select "No", then the instrument uses automatically the
•
preset length of the default setting.
If you select "Yes ", then you can enter in another eld the modied
•
length of the instrument.
Instrument address
An address must be assigned to each Probus PA instrument. Each
address may only be assigned once in the Probus PA network. The
sensor is only recognized by the control system if the address is set
correctly.
When the instrument is shipped, address 126 is adjusted. This address can be used for function test of the instrument and for connection to a Probus PA network. Then address must be changed to
integrate additional instruments.
The address setting is carried out either via:
The address selection switch in the electronics compartment of
•
the instrument (address setting via hardware)
The display and adjustment module (address setting via software)
•
PACTware/DTM (address setting via software)
•
Hardware addressing - The hardware addressing is eective if an
•
address <126 is set with the address selection switches on the
electronics module of VEGAFLEX 81. The software addressing is
then no longer eective, only the set hardware address is valid.
Software addressing - The software addressing is only eective
•
if address 126 or higher is set on the instrument with the address
selection switches.
60
Probe length L from seal surface
If you have modied the length of the probe, you can enter in this eld
the modied probe length. Keep the selected unit in mind.
Determine probe length automatically
If you do not know the probe length, you can have the length of the
probe determined automatically. The requirement for this is a probe
unrestricted and not covered by the medium.
Click to "Carry out now", to start the automatic length determination.
VEGAFLEX 81 • Probus PA
44217-EN-130910
7 Setup with PACTware
Step 2
Application
Type of medium
Here you can see which type of medium your instrument is suitable
for. If this function is inactive, the medium type your instrument is suitable for was already preset.
Information:
In special cases you can change the type of medium. This setting can
be changed under "Extended adjustment".
Application
In this eld you can select the application you want to use your instrument for. You have the following selection options:
Level in the vessel
•
Level in the bypass/standpipe
•
Interface in the vessel
•
Interface in the bypass/standpipe
•
Demonstration mode
•
Level measurement: If you select "Level", you can select the properties of the medium in another eld.
Interface measurement: If you select "Interface", the instrument needs
more information, such as the distance to the interface, the dielectric
constant of the upper medium or whether or not there is a superimposed gas phase.
Demonstration mode: This mode is only suitable for test and demonstration purposes. In this mode, the sensor ignores all parameters
and reacts immediately to all measured value changes within the
measuring range.
Application - Level measurement
The level measurement refers to the product surface which is the limit
to the gas phase.
The interface measurement refers to the phase limit between two
liquids. The total level is also available as a measured value.
Superimposed gas phase present
•
– Check if there is a superimposed gas phase in the vessel. This
is always the case if the total level never touches the process
tting.
Properties
•
– Here you can enter the dielectric constant of the upper medium
– As an alternative you can enter the distance to the interface
61
7 Setup with PACTware
Step 3
Adjustment
Step 4
Linearization
Step 5
Setting - AI FB1
Adjustment for the level measurement
If you have selected level measurement in the previous menu, then
you can enter the values for the min. and max. adjustment. The value
to be entered refers to the distance from the sealing surface of the
process tting (sensor reference plane) to the surface of the product.
Adjustment for the level and interface measurement
If you have selected interface measurement in the previous menu,
then you can enter the values for the min. and max. adjustment of
level and interface or accept the values of the level measurement. The
entered value refers to the distance from the sealing surface of the
process tting (sensor reference plane) to the total level or interface.
Linearization is required if the measured value should be outputted
in proportion to the volume and not the level. The linearization acts
identically on the level and the interface measurement. You can nd
further linearization types in the extended adjustment.
If you have a non-linear vessel, you can select here the respective
linearization curve.
Linear
•
Spherical tank
•
Horizontal cylindrical tank
•
You must enter the following vessel dimensions with non-linear conditions:
Height of the socket h
•
Vessel height D
•
In this window you can adjust the output signal. When the function is
inactive, you can change the settings via the "Extended adjustment".
Scaling unit
In the menu item "Out Scale Unit (scaling unit)" you dene the scaling
unit, for example volume in l for the specied channel (measured
value).
Scaling
In the menu items "Out Scale (min.)" and "Out Scale (max.)" you
dene the scaling of the measured values for 0 % and 100 %.
Step 6
Sensor optimization
62
These settings allow you to optimize the sensor. With them you can
compare a sounded distance with the indicated value and correct it, if
necessary.
Probe immersed in the liquid (covered)
Select whether or not the probe is immersed in the medium.
Measured distance to the medium
If the probe is immersed in the medium, you can enter here the measured distance to the medium.
VEGAFLEX 81 • Probus PA
44217-EN-130910
7 Setup with PACTware
Displayeddistancecorrect?
Is the displayed distance value correct? If you have the possibility, you
can enter here the sounded distance to the medium.
False signal suppression
With this function you can carry out the automatic false signal suppression. We recommend carrying out false signal suppression in any
case.
Step 7
Additional settings
When the setup of the instrument is nished, additional settings can
be made. These are various backups and the locking of the instrument against unauthorised or inadvertent adjustment.
For backup purposes, the current parameter adjustment of the instrument is stored in a le. You can use this le later on to restore the
instrument parameter adjustment. The complete data set is downloaded from the device. This procedure can last several minutes.
Createinstrumentdocumentation?
This function is used to print or create a PDF le of the current parameter adjustment. To read the PDF le, you need a suitable program
(for example Acrobat Reader). To print or create the PDF le, all data
are downloaded from the device. For this function, the full version
of the DTM Collection is required. This procedure can last several
minutes.
Storeechocurveofthesetupinthesensor?
Have you completed the initial setup of the instrument? In such case,
we recommend storing the current signal conditions in the device for
later instrument tests and diagnostics.
LockadjustmentwithPINaftersetting?
The instrument is locked with the current PIN. A parameter adjustment is possible only after the PIN is entered again.
7.4 Saving the parameter adjustment data
We recommend documenting or saving the parameter adjustment
data via PACTware. That way the data are available for multiple use or
service purposes.
44217-EN-130910
VEGAFLEX 81 • Probus PA
63
8 Set up with other systems
8 Set up with other systems
8.1 DD adjustment programs
Device descriptions as Enhanced Device Description (EDD) are
available for DD adjustment programs such as, for example, AMS™
and PDM.
The les can be downloaded at www.vega.com/downloads under
"Software".
64
44217-EN-130910
VEGAFLEX 81 • Probus PA
9 Diagnostics and service
9 Diagnostics and service
9.1 Maintenance
If the device is used correctly, no maintenance is required in normal
operation.
9.2 Diagnosis memory
The instrument has several memories which are available for diagnosis purposes. The data remain even with voltage interruption.
Measured value memory
Event memory
Up to 100,000 measured values can be stored in the sensor in a ring
memory. Each entry contains date/time as well as the respective
measured value. Storable values are for example:
Distance
•
Filling height
•
Percentage value
•
Lin. percent
•
Scaled
•
Current value
•
Meas. reliability
•
Electronics temperature
•
When the instrument is shipped, the measured value memory is
active and stores distance, measurement certainty and electronics
temperature every 3 minutes.
In "Extended adjustment" you can select the respective measured
values.
The requested values and recording conditions are set via a PC with
PACTware/DTM or the control system with EDD. Data are thus read
out and also reset.
Up to 500 events are automatically stored with a time stamp in the
sensor (non-deletable). Each entry contains date/time, event type,
event description and value. Event types are for example:
Modication of a parameter
•
Switch-on and switch-o times
•
Status messages (according to NE 107)
•
Error messages (according to NE 107)
•
The data are read out via a PC with PACTware/DTM or the control
system with EDD.
Echo curve memory
44217-EN-130910
VEGAFLEX 81 • Probus PA
The echo curves are stored with date and time and the corresponding
echo data. The memory is divided into two sections:
Echo curve of the setup: This is used as reference echo curve for
the measurement conditions during setup. Changes in the measurement conditions during operation or buildup on the sensor can thus
be recognized. The echo curve of the setup is stored via:
PC with PACTware/DTM
•
Control system with EDD
•
65
9 Diagnostics and service
Display and adjustment module
•
Further echo curves: Up to 10 echo curves can be stored in a ring
buer in this memory section. Further echo curves are stored via:
PC with PACTware/DTM
•
Control system with EDD
•
Display and adjustment module
•
9.3 Status messages
The instrument features self-monitoring and diagnostics according
to NE 107 and VDI/VDE 2650. In addition to the status messages in
the following tables there are more detailed error messages available
under the menu item "Diagnostics" via the display and adjustment
module, PACTware/DTM and EDD.
Status messages
The status messages are divided into the following categories:
Failure
•
Function check
•
Out of specication
•
Maintenance requirement
•
and explained by pictographs:
4123
Fig. 34: Pictographs of the status messages
1 Failure - red
2 Outofspecication-yellow
3 Function check - orange
4 Maintenance - blue
Failure: Due to a malfunction in the instrument, a failure message is
outputted.
This status message is always active. It cannot be deactivated by the
user.
Function check: The instrument is in operation, the measured value
is temporarily invalid (for example during simulation).
This status message is inactive by default. It can be activated by the
user via PACTware/DTM or EDD.
Outofspecication: The measured value is unstable because the
instrument specication is exceeded (e.g. electronics temperature).
This status message is inactive by default. It can be activated by the
user via PACTware/DTM or EDD.
Maintenance: Due to external inuences, the instrument function
is limited. The measurement is aected, but the measured value is
44217-EN-130910
66
VEGAFLEX 81 • Probus PA
9 Diagnostics and service
still valid. Plan in maintenance for the instrument because a failure is
expected in the near future (e.g. due to buildup).
This status message is inactive by default. It can be activated by the
user via PACTware/DTM or EDD.
Failure (failure)
The following table shows the codes and text messages of the status
message "Failure" and provides information on causes as well as
corrective measures.
Code
Text message
F013
no meas-
ured value
available
F017
Adjustment
span too
small
F025
Error in the
linearization table
F036
No op-
erable
software
F040
Error in the
electronics
F080 – General software error – Separate operating
F105
Determine
measured
value
F113
Communi-
cation error
CauseRecticationPA DevS-
– Sensor does not
detect an echo during
operation
– Antenna system con-
taminated or defective
– Adjustment not within
specication
– Index markers are not
continuously rising, for
examle unlogical value
pairs
– Failed or interrupted
software update
– Hardware defect – Exchanging the elec-
– The instrument is still
in the start phase, the
measured value could
not yet be determined
– Error in the internal
instrument communication
– Check or correct
installation and/or
parameter adjustment
– Clean or exchange
process component or
antenna
– Change adjustment
according to the limit
values (dierence
between min. and
max. ≥ 10 mm)
– Check linearization
table
– Delete table/Create
new
– Repeat software
update
– Check electronics
version
– Exchanging the elec-
tronics
– Send instrument for
repair
tronics
– Send instrument for
repair
voltage briey
– Wait for the end of the
switch-on phase
– Duration depending
on the version and
parameter adjustment
up to approximately
3 min.
– Separate operating
voltage briey
– Send instrument for
repair
pec
Diagnosis
Bit 0
Bit 1
Bit 2
Bit 3
Bit 4
Bit 5
Bit 6
Bit 7
44217-EN-130910
VEGAFLEX 81 • Probus PA
67
9 Diagnostics and service
Code
Text message
F125
Unper-
missible
electronics
temperature
F260
Error in the
calibration
F261
Error in the
conguration
F264
Installa-
tion/Setup
error
F265
Meas-
urement
function
disturbed
F267
No execut-
able sensor
software
CauseRecticationPA DevS-
– Temperature of the
electronics in the nonspecied section
– Error in the calibra-
tion carried out in the
factory
– Error in the EEPROM
– Error during setup
– False signal suppres-
sion faulty
– Error when carrying
out a reset
– Adjustment not within
the vessel height/
measuring range
– Max. measuring range
of the instrument not
sucient
– Sensor no longer
carries out a measurement
– Operating voltage
too low
– Sensor cannot start – Exchanging the elec-
– Check ambient tem-
perature
– Isolate electronics
– Use instrument with
higher temperature
range
– Exchanging the elec-
tronics
– Send instrument for
repair
– Repeat setup
– Repeat reset
– Check or correct
installation and/or
parameter adjustment
– Use an instrument
with bigger measuring
range
– Check operating
voltage
– Carry out a reset
– Separate operating
voltage briey
tronics
– Send instrument for
repair
pec
Diagnosis
Bit 8
Bit 9
Bit 10
Bit 11
Bit 12
-
Function check
Outofspecication
68
The following table shows the error codes and text messages in the
status message "Function check" and provides information on causes
as well as corrective measures.
Code
Text message
C700
Simulation ac-
tive
CauseRectication
– A simulation is active – Finish simulation
– Wait for the automatic end
after 60 mins.
The following table shows the error codes and text messages in the
status message "Outofspecication" and provides information on
causes as well as corrective measures.
VEGAFLEX 81 • Probus PA
44217-EN-130910
9 Diagnostics and service
Maintenance
Code
Text message
S600
Unpermissi-
ble electronics
temperature
S601
Overlling
S602
Level with-
in the search
range, compensation
echo
S603
Impermissi-
ble operating
voltage
CauseRectication
– Temperature of the pro-
cessing electronics in the
non-specied section
– Level echo in the close
range not available
– Compensation echo super-
imposed by medium
– Operating voltage below
specied range
– Check ambient temperature
– Isolate electronics
– Use instrument with higher
temperature range
– Reduce level
– 100 % adjustment: Increase
value
– Check mounting socket
– Remove possible interfering
signals in the close range
– Use coaxial probe
– 100 % adjustment: Increase
value
– Check electrical connection
– if necessary, increase
operating voltage
The following table shows the error codes and text messages in the
status message "Maintenance" and provides information on causes
as well as corrective measures.
Code
Text message
M500
Error in the
delivery status
M501
Error in the
non-active
linearization
table
M502
Error in the
event memory
M503
Reliability too
low
CauseRectication
– With the reset to delivery
status, the data could not
be restored
– Hardware error EEPROM – Exchanging the electronics
– Hardware error EEPROM – Exchanging the electronics
– Measurement certainty is
too low for a reliable measurement
– Process component or
probe contaminated or
defective
– Repeat reset
– Load XML le with sensor
data into the sensor
– Send instrument for repair
– Send instrument for repair
– Check installation and
process conditions
– Clean or exchange process
component or probe
44217-EN-130910
VEGAFLEX 81 • Probus PA
69
9 Diagnostics and service
Reaction when malfunctions occur
Procedure for fault recti-
cation
Code
Text message
M504
Error on an
device inter-
face
M505
no measured
value available
M506
Installation/
Setup error
M507
Error in the
instrument
settings
CauseRectication
– Hardware defect – Exchanging the electronics
– Sensor does not detect an
echo during operation
– Process component or
probe contaminated or
defective
– Error during setup – Check and correct instal-
– Error during setup
– Error when carrying out a
reset
– False signal suppression
faulty
– Send instrument for repair
– Check and correct instal-
lation and/or parameter
adjustment
– Clean or exchange process
component or probe
lation and/or parameter
adjustment
– Check probe length
– Carry out reset and repeat
setup
9.4 Rectify faults
The operator of the system is responsible for taking suitable meas-
ures to rectify faults.
The rst measures are:
Evaluation of fault messages, for example via the display and
•
adjustment module
Checking the output signal
•
Treatment of measurement errors
•
Further comprehensive diagnostics options oer a PC with the software PACTware and the suitable DTM. In many cases, the reasons
can be determined in this way and faults can be rectied.
Treatment of measurement errors
70
The below tables show typical examples for application-relevant
measurement errors. There are two measurement errors:
Constant level
•
Filling
•
Emptying
•
The images in column "Error pattern" show the real level with a broken line and the level displayed by the sensor as a continuous line.
VEGAFLEX 81 • Probus PA
44217-EN-130910
Level
time
time
0
time
0
9 Diagnostics and service
1
2
0
Fig. 35: The broken line 1 shows the real level, the continuous line 2 shows the
level displayed by the sensor
Note:
Wherever the sensor displays a constant value, the reason could
•
also be the fault setting of the current output to "Hold value"
In case of a too low level indication, the reason could be a line
•
resistance that is too high
Measurement error with constant level
Fault descriptionError patternCauseRectication
1. Measured value
shows a too low or too
high level
false echo has changed (e.g.
condensation, buildup); false
signal suppression no longer
matches
– Adapt min./max. adjustment
– Carry out a false signal sup-
pression
– Determine the reason for the
changed false echoes, carry
out false signal suppression,
e.g. with condensation
time
44217-EN-130910
VEGAFLEX 81 • Probus PA
73
9 Diagnostics and service
time
0
time
0
time
0
time
Measurementerrorduringlling
Fault descriptionError patternCauseRectication
3. Measured value
jumps towards 0 %
during lling
4. Measured value uctuates around
10 … 20 %
5. Measured value
jumps sporadically to
100 % during lling
Level
Level
Level
– Amplitude of a multiple echo
(vessel top - product surface) is
larger than the level echo
– The level echo cannot be distin-
guished from the false echo at
a false echo position (jumps to
multiple echo)
– Transverse reection from an
extraction funnel, amplitude of
the transverse reection larger
than the level echo
– Various echoes from an uneven
product surface, e.g. a material
cone
– Reections from the product
surface via the vessel wall
(deection)
– Varying condensation or con-
tamination on the antenna
– Check parameter "Application",
especially vessel top, type
of medium, dished end, high
dielectric constant, adapt if
necessary
– In case of interferences due to
installations in the close range:
Change polarisation direction
– Chose a more suitable installa-
tion position
– Direct sensor to the opposite
funnel wall, avoid crossing with
the lling stream
– Check parameter "Type of
medium" and adapt, if neces-
sary
– Optimize installation position
and sensor orientation
– Select a more suitable installa-
tion position, optimize sensor
orientation, e.g. with a swivel-
ling holder
– Carry out a false signal sup-
pression or increase false
signal suppression with con-
densation/contamination in the
close range by editing
– With bulk solids use radar sen-
sor with purging air connection
or exible antenna cover
Measurement error during emptying
Fault descriptionError patternCauseRectication
6. Measured value remains unchanged in
Level
the close range during
emptying
0
74
– False echo larger than the level
echo
– Level echo too small
– Remove false echoes in the
close range. Check: Antenna
must protrude out of the socket
– Remove contamination on the
antenna
– In case of interferences due to
installations in the close range:
Change polarisation direction
– After removing the false ech-
oes, the false signal suppres-
sion must be deleted. Carry out
a new false signal suppression
VEGAFLEX 81 • Probus PA
44217-EN-130910
9 Diagnostics and service
time
0
time
Fault descriptionError patternCauseRectication
7. Measured value
jumps sporadically towards 100 % during
emptying
8. Measured value uctuates around
10 … 20 %
Level
Level
0
– Varying condensation or con-
tamination on the antenna
– Various echoes from an uneven
product surface, e.g. an extraction funnel
– Reections from the product
surface via the vessel wall
(deection)
– Carry out false signal suppres-
sion or increase false signal
suppression in the close range
by editing
– With bulk solids use radar sen-
sor with purging air connection
or exible antenna cover
– Check parameter "Type of
medium" and adapt, if neces-
sary
– Optimize installation position
and sensor orientation
Reaction after fault recti-
cation
24 hour service hotline
Depending on the reason for the fault and the measures taken, the
steps described in chapter "Setup" must be carried out again or must
be checked for plausibility and completeness.
Should these measures not be successful, please call in urgent cases
the VEGA service hotline under the phone no. +49 1805 858550.
The hotline is also available outside normal working hours, seven
days a week around the clock.
Since we oer this service worldwide, the support is provided in
English. The service itself is free of charge, the only costs involved are
the normal call charges.
9.5 Exchanging the electronics module
If the electronics module is defective, it can be replaced by the user.
In Ex applications only one instrument and one electronics module
with respective Ex approval may be used.
If there is no electronics module available on site, the electronics
module can be ordered through the agency serving you. The electronics modules are adapted to the respective sensor and dier in signal
output or voltage supply.
The new electronics module must be loaded with the default settings
of the sensor. These are the options:
In the factory
•
Or on site by the user
•
In both cases, the serial number of the sensor is needed. The serial
numbers are stated on the type label of the instrument, on the inside
of the housing as well as on the delivery note.
When loading on site, rst of all the order data must be downloaded
from the Internet (see operating instructions manual "Electronics module").
44217-EN-130910
Exchanging the cable/rod
VEGAFLEX 81 • Probus PA
9.6 Exchanging the cable/rod
If necessary, the cable or rod (measuring part) of the probe can be
exchanged.
75
9 Diagnostics and service
Loosen the rod or cable with a fork wrench, wrench size 7 (rod ø 8,
cable ø 2 and 4) or wrench size 10 (rod ø 12).
1. Loosen the rod or cable with a fork wrench applied to the at
surface, provide counterforce with another fork wrench.
2. Unscrew the loosened rod or cable manually.
3. Insert the new measuring rod carefully by hand with a twisting
motion into the opening of the process tting.
4. Continue screwing in the rod manually into the opening of the process tting.
5. Exert counterforce with the second fork spanner and tighten the
rod or cable on the at surfaces with the following torque.
Rod ø 8, cable ø 2 and 4: 6 Nm (4.43 lbf ft)
Rod ø 12: 10 Nm (7.37 lbf ft)
Shorten cable/rod
76
Fig. 53: Exchange cable or rod
Information:
Please keep the stated torque so that the max. tensile strength
remains.
6. Enter new probe length and if necessary the new probe type and
then carry out a fresh adjustment (see "Setup procedure, Carry-ing out min. adjustment - Carrying out max. adjustment").
The rod or cable of the probe can be shortened individually.
1. Mark the requested length with mounted measuring rod.
VEGAFLEX 81 • Probus PA
44217-EN-130910
9 Diagnostics and service
2. Cable: Loosen the pins on the gravity weight (hexagon 3)
3. Cable: remove the pins
4. Cable: Pull the cable out of the gravity weight
5. Shorten the cable/rod with a cut-o wheel or metal saw at the
marking. Take note of the specications in the following illustration
when shortening the cable.
6. Cable with gravity weight: Shift the cable according to the drawing
into the gravity weight
7. Cable with gravity weight: Fasten cable with the pins, torque 7 Nm
(5.16 lbf ft)
Cable with centering weight: Fasten cable with the pins, torque
7 Nm (5.16 lbf ft) and x the clamping part on the centering
weight.
8. Enter new probe length and then carry out a fresh adjustment
(see "Setup procedure, Carrying out min. adjustment - Carrying out max. adjustment").
1
(1.77")
45 mm
(3.94")
100 mm
(0.98")
25 mm
(1.97")
50 mm
A
B
1
4
C
223
44217-EN-130910
VEGAFLEX 81 • Probus PA
Fig. 54: Shortening the cable probe
A Gravity weight - cable ø 4 mm
B Gravity weight - cable ø 2 mm
C Centering weight - cable ø 2 mm
1 Pins
2 Thread M8 for eye-bolt
3 Fixing screw - centering weight
9.7 Software update
The following components are required to update the sensor software:
Sensor
•
Voltage supply
•
Interface adapter VEGACONNECT
•
PC with PACTware
•
Current sensor software as le
•
You can nd the actual sensor software as well as detailed information of the procedure under "www.vega.com/downloads" and
"Software".
77
9 Diagnostics and service
You can nd information about the installation in the download le.
Caution:
Instruments with approvals can be bound to certain software versions. Therefore make sure that the approval remains eective with a
software update.
You can nd detailed information on www.vega.com/downloads and
"Approvals".
9.8 How to proceed in case of repair
You can nd a repair form as well as detailed information on how to
proceed under www.vega.com/downloads and "Formsandcerti-
cates".
By doing this you help us carry out the repair quickly and without having to call back for needed information.
If a repair is necessary, please proceed as follows:
Print and ll out one form per instrument
•
Clean the instrument and pack it damage-proof
•
Attach the completed form and, if need be, also a safety data
•
sheet outside on the packaging
Please contact for the return shipment the agency serving you. You
•
can nd the agency on our home page www.vega.com.
78
44217-EN-130910
VEGAFLEX 81 • Probus PA
10 Dismounting
10 Dismounting
10.1 Dismounting steps
Warning:
Before dismounting, be aware of dangerous process conditions such
as e.g. pressure in the vessel or pipeline, high temperatures, corrosive or toxic products etc.
Take note of chapters "Mounting" and "Connecting to power supply"
and carry out the listed steps in reverse order.
10.2 Disposal
The instrument consists of materials which can be recycled by specialised recycling companies. We use recyclable materials and have
designed the parts to be easily separable.
Correct disposal avoids negative eects on humans and the environment and ensures recycling of useful raw materials.
Materials: see chapter "Technical data"
If you have no way to dispose of the old instrument properly, please
contact us concerning return and disposal.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/EG and
the respective national laws. Pass the instrument directly on to a specialised recycling company and do not use the municipal collecting
points. These may be used only for privately used products according
to the WEEE directive.
44217-EN-130910
VEGAFLEX 81 • Probus PA
79
11 Supplement
11 Supplement
11.1 Technical data
General data
316L corresponds to 1.4404 or 1.4435
Materials, wetted parts
Ʋ Process tting (version up to 6 bar)316L and PPS GF 40Ʋ Process tting (version up to 40 bar)316L and Aluminium oxide-ceramic 99.7 % (Al2O3),
Ʋ Process seal on the instrument side
(cable/rod leadthrough)
Ʋ Process sealOn site (instruments with thread: Klingersil C-4400 is
Ʋ Rod: ø 8 mm (0.315 in)316L or Hastelloy C22 (2.4602)Ʋ Rod: ø 12 mm (0.472 in)316L or Hastelloy C22 (2.4602)Ʋ Cable: ø 2 mm (0.079 in)316 (1.4401)Ʋ Cable: ø 4 mm (0.157 in)316 (1.4401)Ʋ Inner conductor (up to the cable)316LƲ Gravity weight (optionally available)316LƲ Centering weight (optionally available) 316L
Ʋ Rod: ø 8 mm (0.315 in)approx. 400 g/m (4.3 oz/ft)Ʋ Rod: ø 12 mm (0.472 in)approx. 900 g/m (9.68 oz/ft)Ʋ Cable: ø 2 mm (0.079 in)approx. 16 g/m (0.17 oz/ft)Ʋ Cable: ø 4 mm (0.157 in)approx. 60 g/m (0.65 oz/ft)Ʋ Gravity weight for cable ø 2 mm
100 g (3.22 oz)
(0.079 in)
Ʋ Gravity weight for cable ø 4 mm
200 g (6.43 oz)
(0.157 in)
Ʋ Centering weight ø 40 mm (1.575 in) 180 g (5.79 oz)Ʋ Centering weight ø 45 mm (1.772 in) 250 g (8.04 oz)Ʋ Centering weight ø 75 mm (2.953 in) 825 g (26.52 oz)Ʋ Centering weight (ø 95 mm (3.74 in)1050 g (33.76 oz)
Probe length L (from seal surface)
Ʋ Rod: ø 8 mm (0.315 in)up to 6 m (19.69 ft)Ʋ Rod: ø 12 mm (0.472 in)up to 6 m (19.69 ft)Ʋ Trimming accuracy - rod±1 mmƲ Cable: ø 2 mm (0.079 in)up to 75 m (246.1 ft)Ʋ Cable: ø 4 mm (0.157 in)up to 75 m (246 ft)Ʋ Trimming accuracy - cable±0.05 %
Measured variableLevel of liquids
Min. dielectric constant of the medium
44217-EN-130910
VEGAFLEX 81 • Probus PA
81
11 Supplement
Ʋ Cable probesεr ≥ 1.6
Ʋ Rod probesεr ≥ 1.6
11
4
2
2
4
3
3
5
5
Fig. 55: Measuring ranges - VEGAFLEX 81
1 Reference plane
2 Probe length L
3 Measuring range (default setting refers to the measuring range in water)
4 Upper dead band (see diagrams under Accuracy - grey section)
5 Lower dead band (see diagrams under Accuracy - grey section)
Output variable
Output signaldigital output signal, format according to IEEE-754
Sensor address126 (default setting)
Damping (63 % of the input variable)0 … 999 s, adjustable
Probus PA prole3.02
44217-EN-130910
82
VEGAFLEX 81 • Probus PA
11 Supplement
Number of FBs with AI (function blocks
3
with analogue input)
Default values
Ʋ 1. FBPVƲ 2. FBSV 1Ʋ 3. FBSV 2
Current value
Ʋ Non-Ex and Ex ia instrument10 mA, ±0.5 mAƲ Ex-d instruments16 mA, ±0.5 mA
Resolution, digital< 1 mm (0.039 in)
Accuracy (according to DIN EN 60770-1)
Process reference conditions according to DIN EN 61298-1
Ʋ Min. distance to installations> 500 mm (19.69 in)Ʋ Vesselmetallic, ø 1 m (3.281 ft), centric installation, process
tting ush with the vessel ceiling
Ʋ MediumWater/Oil (dielectric constant ~2.0)
1)
Ʋ InstallationProbe end does not touch the vessel bottom
Sensor parameter adjustmentNo gating out of false signals carried out
Typical deviation - Interface measure-
± 5 mm (0.197 in)
ment
Typical deviation - Total level interface
See following diagrams
measurement
Typical deviation - Level measurement
2)3)
See following diagrams
1)
With interface measurement = 2.0
2)
Depending on the installation conditions, there can be deviations which can be rectied with an adaptation of
the adjustment or a change of the measured value oset in the DTM service mode
3)
The dead bands can be optimizes by a false signal suppression.
44217-EN-130910
VEGAFLEX 81 • Probus PA
83
11 Supplement
15mm
")
(0.591
2mm
")
(0.079
0
-2mm
(-0.079
")
1
-10mm
(-0.394
")
0,08 m
(3.15")
0,3 m
(11.811
")
Fig. 56: Deviation VEGAFLEX 81 in rod version in water
1 Dead band - no measurement possible in this area
L Probe length
15mm
")
(0.591
2mm
(0.079
")
0
-2mm
(-0.079
")
-10mm
(-0.591
1
")
0,15 m
(5.906
0,3 m
(11.811
")
")
L
0,02 m
(0.787
")
L
1
0,07 m
(2.756
")
0,05 m
(1.97
")
Fig. 57: Deviation VEGAFLEX 81 in rod version in oil
1 Dead band - no measurement possible in this area
L Probe length
84
44217-EN-130910
VEGAFLEX 81 • Probus PA
15mm
(0.591
")
2mm
")
(0.079
0
-2mm
")
(-0.079
10mm
(0.394
1
")
0,08 m
(3.15
0,3 m
")
(11.811
")
Fig. 58: Deviation VEGAFLEX 81 in cable version in water
1 Dead band - no measurement possible in this area
L Probe length
15mm
")
(0.591
2mm
")
(0.079
0
-2mm
")
(-0.079
11 Supplement
L
0,1 m
(3.94
")
L
-15mm
")
(-0.591
0,15 m
0,3 m
")
")
(5.906
(11.811
Fig. 59: Deviation VEGAFLEX 81 in cable version ø 2 mm (0.079 in), in oil
1 Dead band - no measurement possible in this area
L Probe length
44217-EN-130910
VEGAFLEX 81 • Probus PA
11
0,05 m
(1.969
")
0,08 m
(3.15
")
85
11 Supplement
20mm
")
(0.787
15mm
(0.591
")
2mm
(0.079
")
0
-2mm
(-0.079
")
L
-15mm
)
(-0.591
"
0,3 m
0,15 m
(5.906
")
(11.811
")
0,05 m
(1.969
")
0,15 m
(5.906
")
Fig. 60: Deviation VEGAFLEX 81 in cable version ø 4 mm (0.157 in), in oil
1 Dead band - no measurement possible in this area
L Probe length
Repeatability≤ ±1 mm
Variablesinuencingmeasurementaccuracy
Temperature drift - Digital output±3 mm/10 K relating to the max. measuring range or
The propagation speed of the radar impulses in gas or vapour above the medium is reduced by
high pressure. This eect depends on the superimposed gas or vapour and is especially large at
low temperatures.
The following table shows the resulting deviation for some typical gases and vapours. The specied
values refer to the distance. Positive values mean that the measured distance is too large, negative
values that the measured distance is too small.
Gas phaseTemperaturePressure
1 bar (14.5 psig)10 bar (145 psig)50 bar (725 psig)
Air20 °C/68 °F0.00 %0.22 %1.2 %
200 °C/392 °F-0.01 %0.13 %0.74 %
400 °C/752 °F-0.02 %0.08 %0.52 %
Hydrogen20 °C/68 °F-0.01 %0.10 %0.61 %
200 °C/392 °F-0.02 %0.05 %0.37 %
400 °C/752 °F-0.02 %0.03 %0.25 %
44217-EN-130910
86
VEGAFLEX 81 • Probus PA
11 Supplement
Gas phaseTemperaturePressure
1 bar (14.5 psig)10 bar (145 psig)50 bar (725 psig)
Steam (saturated
steam)
100 °C/212 °F0.26 %--
180 °C/356 °F0.17 %2.1 %-
264 °C/507 °F0.12 %1.44 %9.2 %
366 °C/691 °F0.07 %1.01 %5.7 %
Characteristics and performance data
Measuring cycle time< 500 ms
Step response time
4)
≤ 3 s
Max. lling/emptying speed1 m/min
Ambient conditions
Ambient, storage and transport tempera-
-40 … +80 °C (-40 … +176 °F)
ture
Process conditions
For the process conditions, please also note the specications on the type label. The lowest value
always applies.
The measurement error from the process conditions is in the specied pressure and temperature
range of below 1 %.
Process pressure
Ʋ Process tting with PPS GF 40-1 … 6 bar/-100 … 600 kPa (-14.5 … 87 psi), depending
on the process tting
Ʋ Process tting with Al2O
3
Vessel pressure relating to the ange
nominal stage
-1 … +40 bar/-100 … +4000 kPa (-14.5 … +580 psig),
depending on the process tting
see supplementary instructions manual "Flanges ac-cording to DIN-EN-ASME-JIS"
Ʋ FFKM (Kalrez 6375)-20 … +150 °C (-4 … +302 °F)Ʋ FFKM (Kalrez 6375) - with tempera-
-20 … +200 °C (-4 … +392 °F)
ture adapter
4)
Time span after a sudden measuring distance change by max. 0.5 m in liquid applications, max 2 m with bulk
solids applications, until the output signal has taken for the rst time 90 % of the nal value (IEC 61298-2).
44217-EN-130910
VEGAFLEX 81 • Probus PA
87
11 Supplement
1
80°C
(176°F)
65°C
(149°F)
3
4
-40°C
(-104°F)
0°C
(32°F)
-40°C
(-104°F)
80°C
(176°F)
130°C
(266°F)
150°C
(302°F)
2
Fig. 61: Ambient temperature - process temperature, standard version
1 Ambient temperature
2 Process temperature (depending on the seal material)
3 Maximum permissible temperature - standard
4 Limited temperature range - plastic housing and stainless steel housing, electropolished
1
80°C
(176°F)
3
4
55°C
(131°F)
-20°C
(-68°F)
0°C
(32°F)
-40°C
(-104°F)
150°C
(302°F)
200°C
(392°F)
2
Fig. 62: Ambient temperature - process temperature, version with temperature adapter
1 Ambient temperature
2 Process temperature (depending on the seal material)
3 Maximum permissible temperature - standard
4 Limited temperature range - plastic housing and stainless steel housing, electropolished
Vibration resistance
Ʋ Instrument housing4 g at 5 … 200 Hz according to EN 60068-2-6 (vibration
with resonance)
Ʋ Rod probe1 g with 5 … 200 Hz according EN 60068-2-6 (vibration
at resonance) with rod length 50 cm (19.69 in)
Shock resistance
Ʋ Instrument housing100 g, 6 ms according to EN 60068-2-27 (mechanical
shock)
Ʋ Rod probe25 g, 6 ms according to EN 60068-2-27 (mechanical
shock) with rod length 50 cm (19.69 in)
44217-EN-130910
88
VEGAFLEX 81 • Probus PA
11 Supplement
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Options of the cable entry
Ʋ Cable glandM20 x 1.5 (cable: ø 5 … 9 mm)Ʋ Cable entry½ NPTƲ Blind plugM20 x 1.5; ½ NPTƲ Closing capM20 x 1.5; ½ NPT
Plug options
Ʋ Signal circuitPlug M12 x 1, according to ISO 4400, Harting HAN,
IP 66/IP 68 (1 bar)
vestment casting (optionally available)
Overvoltage categoryIII
Protection classIII
5)
Approvals
Instruments with approvals can have dierent technical data depending on the version.
For that reason the associated approval documents of these instruments must be carefully noted.
They are part of the delivery or can be downloaded under www.vega.com and "VEGA Tools" as
well as under "Downloads" and "Approvals".
11.2 CommunicationProbusPA
Instrumentmasterle
The instrument master le (GSD) contains the characteristic data of the Probus PA instrument.
5)
The prerequisites for maintaining the protection rating are a suitable cable as well as correct mounting.
90
VEGAFLEX 81 • Probus PA
44217-EN-130910
11 Supplement
These data are, e.g. the permissible transmission rates as well as information on diagnostics values
and the format of the measured value outputted by the PA instrument.
A bitmap le is also provided for the Probus network planning tool. This le is installed automatically when the GSD le is integrated. The bitmap le is used for symbolic indication of the PA instrument in the conguration tool.
ID number
Each Probus instrument gets an unambiguous ident number (ID number) from the Probus user
organisation (PNO). This ID number is also included in the name of the GSD le. Optionally to this
manufacturer-specic GSD le, PNO provides also a general so-called prole-specic GSD le. If
the general GSD le is used, the sensor must be set to the prole-specic ident number via the DTM
software. By default, the sensor operates with the manufacturer-specic ID number. When using the
instruments on a segment coupler SK-2 or SK-3, no special GSD les are required.
VE010BF5
The following table shows the instrument ID and the GSD names for the VEGAFLEX sensor series.
Device nameInstrument IDGSDlename
VEGAInstrument class in
prole3.02
VEGAFLEX Serie 80 0xBF50x9702VE010BF5.GSDPA139702.GSD
VEGAProle-specic
Cyclicaldatatrac
The master class 1 (e.g. PLC) cyclically reads out measured values from the sensor during operation. The below block diagram below shows which data can be accessed by the PLC.
44217-EN-130910
VEGAFLEX 81 • Probus PA
91
11 Supplement
Fig. 63: VEGAFLEX 81: Block diagram with AI FB 1 … AI FB 3 OUT values
TB Transducer Block
FB 1 … FB 3 Function Block
Module of the PA sensors
For the cyclic data trac, VEGAFLEX 81 provides the following modules:
AI FB1 (OUT)
•
– Out value of the AI FB1 after scaling
AI FB2 (OUT)
•
– Out value of the AI FB2 after scaling
AI FB3 (OUT)
•
– Out value of the AI FB3 after scaling
Free Place
•
– This module must be used if a value in the data telegram of the cyclical data trac should not
be used (e.g. replacement of temperature and Additional Cyclic Value)
A maximum of three modules can be active. By means of the conguration software of the Probus
master you can determine the conguration of the cyclical data telegram with these modules. The
procedure depends on the respective conguration software.
Note:
The modules are available in two versions:
Short for Probus master supporting only one "Identier Format" byte, e.g. Allen
•
Bradley
92
VEGAFLEX 81 • Probus PA
44217-EN-130910
11 Supplement
Sign
-23
Long for Probus master only supporting the "Identier Format" byte, e.g. Siemens
•
S7-300/400
Examplesoftelegramconguration
In the following you will see how the modules can be combined and how the appendant data telegram is structured.
The status byte corresponds to prole 3.02 "Probus PA Prole for Process Control Devices" coded.
The status "Measured value OK" is coded as 80 (hex) (Bit7 = 1, Bit6 … 0 = 0).
The measured value is transferred as a 32 bit oating point number in the IEEE-754 format.
Byte n
Bit
Bit
6
7
2
Bit
Bit
4
3
5
6543
2
2
2
Exponent
VZ
2
Bit
7
VZ
Bit
Value = (-1)
Fig. 65: Data format of the measured value
VEGAFLEX 81 • Probus PA
Byte1
Byte2
Bit
Bit
Bit
0
2
1
210-1
2
2
2
(Exponent - 127)
Byte n+1
Bit
7
2
Bit
Bit
Bit
4
6
5
-2-3-4-5-6 -7
2
2
2
Significant
(1 + Significant)
Bit
Byte n+2
Bit
Bit
Bit
Bit
7
3
2
0
2
1
2
2
2
2
Bit
-9
Bit
5
-10 -11
2
Bit
4
3
2
2
Bit
6
-8
2
Bit
Bit
Bit
Bit
0
2
1
-12
2
7
-13
-14 -15
2
2
2
Significant
Byte n+3
Bit
Bit
6
-16
2
Bit
Bit
-17
Bit
Bit
4
5
-19
-18
2
2
Bit
3
2
1
0
-20
-22
-21
2
2
2
2
Significant
93
11 Supplement
Coding of the status byte associated with the PA output value
You can nd further information for the coding of the status byte in the Device Description 3.02 on
www.probus.com.
Status codeDescription according to
Probusstandard
0 x 00bad - non-specicFlash-Update active
0 x 04bad - conguration error – Adjustment error
0 x 0Cbad - sensor failure – Hardware error
0 x 10bad - sensor failure – Measured value generation error
0 x 1fbad - out of service con-
stant
0 x 44uncertain - last unstable
value
0 x 48uncertain substitute set – Switch on simulation
0 x 4cuncertain - initial valueFailsafe replacement value (Failsafe-Mode = "Last valid
0 x 51uncertain - sensor; con-
version not accurate - low
limited
0 x 52uncertain - sensor; con-
version not accurate - high
limited
0 x 80good (non-cascade) - OK OK
0 x 84good (non-cascade) - ac-
tive block alarm
0 x 89good (non-cascade) - ac-
tive advisory alarm - low
limited
0 x 8agood (non-cascade) - ac-
tive advisory alarm - high
limited
0 x 8dgood (non-cascade) - ac-
tive critical alarm - low
limited
0 x 8egood (non-cascade) - ac-
tive critical alarm - high
limited
Possible cause
– Conguration error with PV-Scale (PV-Span too
small)
– Unit irregularity
– Error in the linearization table
Failsafe replacement value (Failsafe-Mode = "Last value" and already valid measured value since switching
on)
– Failsafe replacement value (Failsafe-Mode = "Fsafe
value")
value" and no valid measured value since switching on)
Sensor value < lower limit
Sensor value > upper limit
Static revision (FB, TB) changed (10 sec. active, after
the parameter of the static category has been written)
Lo-Alarm
Hi-Alarm
Lo-Lo-Alarm
Hi-Hi-Alarm
44217-EN-130910
94
VEGAFLEX 81 • Probus PA
11 Supplement
11.3 Dimensions
The following dimensional drawings represent only an extract of all possible versions. Detailed
dimensional drawings can be downloaded at www.vega.com/downloads under "Drawings".
Plastic housing
~ 69 mm
(2.72")
ø 79 mm
(3.11")
~ 84 mm
(3.31")
M16x1,5
ø 79 mm
(3.11")
M20x1,5/
½ NPT
(4.41")
112 mm
1
M20x1,5/
½ NPT
2
(4.41")
112 mm
Fig. 66: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated display and adjustment module the
housing is 9 mm/0.35 in higher
1 Single chamber version
2 Double chamber version
Aluminium housing
~ 116 mm
(4.57")
ø 86 mm
(3.39")
(4.57")
116 mm
M20x1,5/
½ NPT
M20x1,5
Fig. 67: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated display and adjustment module the
housing is 9 mm/0.35 in higher
1 Single chamber version
2 Double chamber version
~ 87 mm
M16x1,5
M20x1,5/
½ NPT
(3.43")
ø 86 mm
(3.39")
21
(4.72")
120 mm
44217-EN-130910
VEGAFLEX 81 • Probus PA
95
11 Supplement
Aluminium housing in protection rating IP 66/IP 68 (1 bar)
1 Single chamber version, electropolished
2 Single chamber version, precision casting
3 Double chamber version, precision casting
(4.72")
44217-EN-130910
VEGAFLEX 81 • Probus PA
97
11 Supplement
VEGAFLEX 81, cable version with gravity weight
SW 36 (1.
SW 41 (1.
SW 46 (1.
ø 2 mm
(0.08")
42")
(G¾, ¾ NPT)
61")
(G1, 1 NPT)
81")
(G1½, 1½ NPT)
G¾, ¾ NPT,
G1, 1 NPT,
G1½, 1½ NPT
ø 16 mm
(0.63")
ø 20 mm
(0.79")
(1.97")
50 mm
ø 4 mm
(0.16")
ø 16 mm
(0.63")
ø 20 mm
(0.79")
46 mm
(1.81")
(0.87")
22 mm
100 mm (3.94")
L
ø 4 mm
(0.16")
123
ø 16 mm
(0.63")
ø 20 mm
(0.79")
(3.78")
96 mm
(0.87")
22 mm
100 mm (3.94")
L
44217-EN-130910
Fig. 71: VEGAFLEX 81, threaded version with gravity weight (all gravity weights with thread M8 for eye-bolt)
L Sensor length, see chapter "Technical data"
98
VEGAFLEX 81 • Probus PA
1 Cable version ø 2 mm (0.079 in) with gravity weight
2 Cable version ø 4 mm (0.157 in) with gravity weight
3 Cable version with temperature adapter
VEGAFLEX 81, cable version with centering weight
SW 36 (1.
42")
(G¾, ¾ NPT)
61")
SW 41 (1.
SW 46 (1.
(G1, 1 NPT)
(G1½, 1½ NPT)
81")
G¾, ¾ NPT,
G1, 1 NPT,
G1½, 1½ NPT
11 Supplement
(1.81")
22 mm (0.87")46 mm
(1.18")
30 mm
L
ø 2 mm
(0.08")
ø 4 mm
(0.16")
(1.18")
30 mm
X
X
12
Fig. 72: VEGAFLEX 81, threaded version
L Sensor length, see chapter "Technical data"
x ø 40 mm (1.57 in)
ø 45 mm (1.77 in)
ø 75 mm (2.95 in)
ø 95 mm (3.74 in)
1 Cable version ø 2 mm (0.079 in) with centering weight (see supplementary instructions "Centering")
2 Cable version ø 4 mm (0.157 in) with centering weight (see supplementary instructions "Centering")
44217-EN-130910
VEGAFLEX 81 • Probus PA
99
11 Supplement
VEGAFLEX 81, rod version
SW 36 (1.
42")
(G¾, ¾ NPT)
61")
SW 41 (1.
SW 46 (1.
(G1, 1 NPT)
(G1½, 1½ NPT)
81")
G¾, ¾ NPT,
G1, 1 NPT,
G1½, 1½ NPT
ø 8 mm
(0.32")
12
Fig. 73: VEGAFLEX 81, threaded version
L Sensor length, see chapter "Technical data"
1 Rod version ø 8 mm (0.315 in)
2 Rod version ø 12 mm (0.472 in)
46 mm
(1.81")
22 mm (0.87")
SW 55
(2.17")
G1½, 1½ NPT
L
ø 12¨mm
(0.47")
(1.18")
30mm
51 mm
L
(2.01")
100
44217-EN-130910
VEGAFLEX 81 • Probus PA
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.