VEGA FLEX81 User Manual

Operating Instructions
TDR sensor for continuous level and interface measurement of liquids
VEGAFLEX 81
Probus PA
Document ID: 44217

Quick start

Mounting
Electrical connection
Quick start
The quick start procedure enables a quick setup of the instrument with many applications. You can nd further information in the respec­tive chapters of the operating instructions manual.
1. Distance from the metallic vessel wall > 300 mm. Distance from non-metallic vessel wall > 500 mm. The probe must not touch any installations or the vessel wall.
2. In non-metallic vessels, place a metal sheet beneath the process tting.
1 2
Fig. 1: Installation in non-metallic vessel
1 Flange 2 Metal sheet
3. If necessary, fasten probe end.
For further information see chapter "Mounting".
1. Make sure that the power supply corresponds to the specica­tions on the type label.
2. Connect the instrument according to the following illustration
2
3
4
Bus
00
1
8
1
0
0
1
1
9
9
2
2
8
3
3
7
7
4
4
6
6
5
5
Set parameters
2
+
( )
(-)
1
2
5
678
5
1
Fig. 2: Electronics and connection compartment, single chamber housing
1 Voltage supply, signal output 2 For display and adjustment module or interface adapter 3 Selection switch for bus address 4 For external display and adjustment unit 5 Ground terminal for connection of the cable screen
For further information see chapter "Connecting to power supply".
For standard applications we recommend selecting the "Quick setup" in the display and adjustment module.
44217-EN-130910
Quick start
1. In this menu item you can select the application. You can choose between level and interface measurement.
2. In the menu item "Medium - Dielectric constant" you can dene the type of medium (medium).
3. Carry out the adjustment in the menu items "Min. adjustment" and "Max. adjustment".
4. A "Linearization" is recommended for all vessels in which the vessel volume does not increase linearly with the level - e.g. in a horizontal cylindrical or spherical tank. Activate the appropriate curve.
5. A "False signal suppression" detects, marks and saves the false signals so that they are no longer taken into account for level measurement. We generally recommend a false signal suppres­sion.
Parameterization example
44217-EN-130910
The sensor measures the distance from the sensor (reference plane) to the product surface. See also chapter "Parameter adjustment".
1
4
2
Fig. 3: Measuring ranges - VEGAFLEX 81
1 Reference plane 2 Probe length L 3 Measuring range (default setting refers to the measuring range in water) 4 Upper dead band (in this area no measurement is possible) 5 Lower dead band (in this area no measurement is possible)
2
3
5
1
4
3
5
For this adjustment, the distance is entered when the vessel is full and nearly empty. If these values are not known, an adjustment with other distances, for example, 10 % and 90 % is also possible. Starting
3
Quick start
Further steps
point for these distance specications is always the seal surface of the thread or ange.
1. In the menu "Additional settings", menu item "Damping" you can adjust the requested damping of the output signal.
2. Select the parameter of the current output and the output charac­teristics in the menu item "Current output".
44217-EN-130910
4
Contents
1 About this document
1.1 Function ........................................................................................................................... 7
1.2 Target group ..................................................................................................................... 7
1.3 Symbolism used ............................................................................................................... 7
2 For your safety
2.1 Authorised personnel ....................................................................................................... 8
2.2 Appropriate use ................................................................................................................ 8
2.3 Warning about incorrect use ............................................................................................. 8
2.4 General safety instructions ............................................................................................... 8
2.5 CE conformity ................................................................................................................... 8
2.6 NAMUR recommendations .............................................................................................. 9
2.7 Environmental instructions ............................................................................................... 9
3 Product description
3.1 Conguration .................................................................................................................. 10
3.2 Principle of operation...................................................................................................... 11
3.3 Packaging, transport and storage ................................................................................... 14
3.4 Accessories and replacement parts ............................................................................... 14
4 Mounting
4.1 General instructions ....................................................................................................... 17
4.2 Mounting instructions ..................................................................................................... 17
5 Connecting to power supply
5.1 Preparing the connection ............................................................................................... 24
5.2 Connecting ..................................................................................................................... 24
5.3 Wiring plan, single chamber housing.............................................................................. 26
5.4 Wiring plan, double chamber housing ............................................................................ 26
5.5 Wiring plan, Ex-d-ia double chamber housing ................................................................ 28
5.6 Wiring plan - version IP 66/IP 68, 1 bar ........................................................................... 29
5.7 Supplementary electronics ............................................................................................. 30
5.8 Set instrument address .................................................................................................. 30
5.9 Switch-on phase............................................................................................................. 31
6 Set up with the display and adjustment module
6.1 Insert display and adjustment module ............................................................................ 32
6.2 Adjustment system ......................................................................................................... 33
6.3 Parameter adjustment - Quick setup .............................................................................. 34
6.4 Parameter adjustment - Extended adjustment................................................................ 37
6.5 Saving the parameter adjustment data ........................................................................... 55
7 Setup with PACTware
7.1 Connect the PC .............................................................................................................. 57
7.2 Parameter adjustment with PACTware ............................................................................ 57
7.3 Set up with the quick setup ............................................................................................. 58
7.4 Saving the parameter adjustment data ........................................................................... 63
8 Set up with other systems
8.1 DD adjustment programs ............................................................................................... 64
9 Diagnostics and service
9.1 Maintenance .................................................................................................................. 65
44217-EN-130910
Contents
5
Contents
9.2 Diagnosis memory ......................................................................................................... 65
9.3 Status messages ............................................................................................................ 66
9.4 Rectify faults ................................................................................................................... 70
9.5 Exchanging the electronics module ................................................................................ 75
9.6 Exchanging the cable/rod ............................................................................................... 75
9.7 Software update ............................................................................................................. 77
9.8 How to proceed in case of repair .................................................................................... 78
10 Dismounting
10.1 Dismounting steps.......................................................................................................... 79
10.2 Disposal ......................................................................................................................... 79
11 Supplement
11.1 Technical data ................................................................................................................ 80
11.2 Communication Probus PA ........................................................................................... 90
11.3 Dimensions .................................................................................................................... 95
44217-EN-130910
Safety instructions for Ex areas
Please note the Ex-specic safety information for installation and op­eration in Ex areas. These safety instructions are part of the operating instructions manual and come with the Ex-approved instruments.
Editing status: 2013-07-24
6

1 About this document

1 About this document

1.1 Function

This operating instructions manual provides all the information you need for mounting, connection and setup as well as important instruc­tions for maintenance and fault rectication. Please read this informa­tion before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device.

1.2 Target group

This operating instructions manual is directed to trained specialist personnel. The contents of this manual should be made available to these personnel and put into practice by them.

1.3 Symbolism used

Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can result. Warning: If this warning is ignored, injury to persons and/or serious
damage to the instrument can result. Danger: If this warning is ignored, serious injury to persons and/or
destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
List
The dot set in front indicates a list with no implied sequence.
Action
→
This arrow indicates a single action.
1 Sequence of actions
Numbers set in front indicate successive steps in a procedure.
Battery disposal
This symbol indicates special information about the disposal of bat­teries and accumulators.
44217-EN-130910
7

2 For your safety

2 For your safety

2.1 Authorised personnel

All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator.
During work on and with the device the required personal protective equipment must always be worn.

2.2 Appropriate use

VEGAFLEX 81 is a sensor for continuous level measurement. You can nd detailed information on the application range in chapter
"Product description". Operational reliability is ensured only if the instrument is properly
used according to the specications in the operating instructions manual as well as possible supplementary instructions.

2.3 Warning about incorrect use

Inappropriate or incorrect use of the instrument can give rise to application-specic hazards, e.g. vessel overll or damage to system components through incorrect mounting or adjustment.

2.4 General safety instructions

This is a state-of-the-art instrument complying with all prevailing regulations and guidelines. The instrument must only be operated in a technically awless and reliable condition. The operator is responsible for the trouble-free operation of the instrument.
During the entire duration of use, the user is obliged to determine the compliance of the necessary occupational safety measures with the current valid rules and regulations and also take note of new regula­tions.
The safety instructions in this operating instructions manual, the na­tional installation standards as well as the valid safety regulations and accident prevention rules must be observed by the user.
For safety and warranty reasons, any invasive work on the device beyond that described in the operating instructions manual may be carried out only by personnel authorised by the manufacturer. Arbi­trary conversions or modications are explicitly forbidden.
The safety approval markings and safety tips on the device must also be observed.

2.5 CE conformity

The device fullls the legal requirements of the applicable EC guide­lines. By axing the CE marking, we conrm successful testing of the product.
You can nd the CE Certicate of Conformity in the download section of our homepage.
8
44217-EN-130910
2 For your safety
Electromagnetic compatibility
Instruments with plastic housing as well as in four-wire or Ex-d-ia version are designed for use in an industrial environment. Neverthe­less, electromagnetic interference from electrical conductors and radiated emissions must be taken into account, as is usual with a class A instrument according to EN 61326-1. If the instrument is used in a dierent environment, the electromagnetic compatibility to other instruments must be ensured by suitable measures.

2.6 NAMUR recommendations

NAMUR is the automation technology user association in the process industry in Germany. The published NAMUR recommendations are accepted as the standard in eld instrumentation.
The device fullls the requirements of the following NAMUR recom-
mendations:
NE 21 – Electromagnetic compatibility of equipment
NE 43 – Signal level for malfunction information from measuring
transducers
NE 53 – Compatibility of eld devices and display/adjustment
components
NE 107 – Self-monitoring and diagnosis of eld devices
For further information see www.namur.de.

2.7 Environmental instructions

Protection of the environment is one of our most important duties. That is why we have introduced an environment management system with the goal of continuously improving company environmental pro­tection. The environment management system is certied according to DIN EN ISO 14001.
Please help us fulll this obligation by observing the environmental instructions in this manual:
Chapter "Packaging, transport and storage"
Chapter "Disposal"
44217-EN-130910
9

3 Product description

Type plate
3 Product description
3.1 Conguration
The nameplate contains the most important data for identication and use of the instrument:
1
2
3
4 5
6
7 8 9
10
Fig. 4: Layout of the type label (example)
1 Instrument type 2 Product code 3 Approvals 4 Power supply and signal output, electronics 5 Protection rating 6 Probe length 7 Process and ambient temperature, process pressure 8 Material, wetted parts 9 Hardware and software version 10 Order number 11 Serial number of the instrument 12 Symbol of the device protection class 13 ID numbers, instrument documentation 14 Reminder to observe the instrument documentation
15 NotiedauthorityforCEmarking
16 Approval directives
16
15
14
13
12
11
Serial number
10
The type label contains the serial number of the instruments. Hence you can nd the following data on our homepage:
Product code of the instrument (HTML)
Delivery date (HTML)
Order-specic instrument features (HTML)
Operating instructions at the time of shipment (PDF)
Order-specic sensor data for an electronics exchange (XML)
Test certicate pressure transmitters (PDF)
Go to www.vega.com, "VEGA Tools" and "Serial number search".
As an alternative, you can nd the data via your Smartphone:
Download the smartphone app "VEGA Tools" from the "Apple App
Store" or the "Google Play Store" Scan the Data Matrix code on the type label of the instrument or
Enter the serial number manually in the app
44217-EN-130910
3 Product description
Scope of this operating instructions manual
Versions
Scope of delivery
Application area
This operating instructions manual applies to the following instrument
versions:
Hardware from 1.0.0
Software from 1.0.0
Only for instrument versions without SIL qualication
This electronics version can be determined via the product code on the type label as well as on the electronics.
Standard electronics: Type FX80PA.-
The scope of delivery encompasses:
Sensor
Documentation
this operating instructions manualTest certicate measuring accuracy (optional)Operating instructions manual "Display and adjustment mod-
ule" (optional)
– Supplementary instructions "GSM/GPRS radio module"
(optional)
Supplementary instructions manual "Heating for display and
adjustment module" (optional)
Supplementary instructions manual "Plug connector for con-
tinuously measuring sensors" (optional)
Ex-specic "Safety instructions" (with Ex versions)if necessary, further certicates

3.2 Principle of operation

The VEGAFLEX 81 is a level sensor with cable or rod probe for continuous level or interface measurement, suitable for applications in liquids.
Functional principle ­level measurement
44217-EN-130910
High frequency microwave pulses are guided along a steel cable or a rod. Upon reaching the product surface, the microwave pulses are reected. The running time is evaluated by the instrument and output­ted as level.
11
3 Product description
1
d
h
Fig. 5: Level measurement
1 Sensorreferenceplane(sealsurfaceoftheprocesstting)
d Distance to the interface h Height - Level
Probe end tracking
To increase sensitivity, the probe is equipped with probe end tracking. In products with a low dielectric constant, this function is very helpful. This is the case, for example, in plastic granules, packing chips or in vessels with uidized products.
Between a dielectric constant of 1.5 and 3, the function switches on, if required. As soon as the level echo can no longer be detected, probe end tracking is automatically activated. The measurement is contin­ued with the last calculated dielectric constant.
The accuracy thus depends on the stability of the dielectric constant. If you measure a medium with a dielectric constant below 1.5, probe
end tracking is always active. In this case, you have to enter the dielectric constant of the medium. A stable dielectric constant is very important here.
Functional principle - in­terface measurement
12
High frequency microwave impulses are guided along a steel cable or rod. Upon reaching the product surface, a part of the microwave im­pulses is reected. The other part passes through the upper product and is reected by the interface. The running times to the two product layers are processed by the instrument.
44217-EN-130910
L3
L2
3 Product description
1
d2
d1
TS
Prerequisites for inter­face measurement
h2
h1
Fig. 6: Interface measurement
1 Sensorreferenceplane(sealsurfaceoftheprocesstting)
d1 Distance to the interface d2 Distance to the level TS Thickness of the upper medium (d1 - d2) h1 Height - Interface h2 Height - Level L1 Lower medium L2 Upper medium L3 Gas phase
Upper medium (L2)
The upper medium must not be conductive
The dielectric constant of the upper medium or the actual distance
to the interface must be known (input required). Min. dielectric con­stant: 1.6. You can nd a list of dielectric constants on our home page: www.vega.com. The composition of the upper medium must be stable, no varying
products or mixtures The upper medium must be homogeneous, no stratications
within the medium
Min. thickness of the upper medium 50 mm (1.97 in)
Clear separation from the lower medium, emulsion phase or detri-
tus layer max. 50 mm (1.97 in) If possible, no foam on the surface
L1
44217-EN-130910
Lower medium (L1)
The dielectric constant must be 10 higher than the dielectric
constant of the upper medium, preferably electrically conductive. Example: upper medium dielectric constant 2, lower medium at least dielectric constant 12.
Gas phase (L3)
Air or gas mixture
Gas phase - dependent on the application, gas pahse does not
always exist (d2 = 0)
13
3 Product description
Output signal
Packaging
Transport
Transport inspection
Storage
Storage and transport temperature
The instrument is always preset to the application "Level measure­ment".
For the interface measurement, you can select the requested output signal with the setup.

3.3 Packaging, transport and storage

Your instrument was protected by packaging during transport. Its
capacity to handle normal loads during transport is assured by a test based on ISO 4180.
The packaging of standard instruments consists of environment­friendly, recyclable cardboard. For special versions, PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies.
Transport must be carried out in due consideration of the notes on the
transport packaging. Nonobservance of these instructions can cause damage to the device.
The delivery must be checked for completeness and possible transit
damage immediately at receipt. Ascertained transit damage or con­cealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and stored according to the orientation and storage markings on the outside.
Unless otherwise indicated, the packages must be stored only under the following conditions:
Not in the open
Dry and dust free
Not exposed to corrosive media
Protected against solar radiation
Avoiding mechanical shock and vibration
Storage and transport temperature see chapter "Supplement -
Technical data - Ambient conditions" Relative humidity 20 … 85 %
PLICSCOM
VEGACONNECT
14

3.4 Accessories and replacement parts

The display and adjustment module PLICSCOM is used for measured
value indication, adjustment and diagnosis. It can be inserted into the sensor or the external display and adjustment unit and removed at any time.
You can nd further information in the operating instructions "Display and adjustment module PLICSCOM" (Document-ID 27835).
The interface adapter VEGACONNECT enables the connection of
communication-capable instruments to the USB interface of a PC. For parameter adjustment of these instruments, the adjustment software PACTware with VEGA-DTM is required.
44217-EN-130910
3 Product description
You can nd further information in the operating instructions "Interface adapter VEGACONNECT" (Document-ID 32628).
VEGADIS 81
PLICSMOBILE T61
Protective cap
Flanges
Electronics module
The VEGADIS 81 is an external display and adjustment unit for VEGA
plics® sensors. For sensors with double chamber housing the interface adapter
"DISADAPT" is also required for VEGADIS 81. You can nd further information in the operating instructions "VE-
GADIS 81" (Document-ID 43814).
The PLICSMOBILE T61 is an external GSM/GPRS radio unit for
transmission of measured values and for remote parameter adjust­ment of plics® sensors. The adjustment is carried out via PACTware/ DTM by using the integrated USB connection.
You can nd further information in the supplementary instructions "PLICSMOBILE T61" (Document-ID 37700).
The protective cover protects the sensor housing against soiling and
intense heat from solar radiation. You will nd additional information in the supplementary instructions
manual "Protective cover" (Document-ID 34296).
Screwed anges are available in dierent versions according to the following standards: DIN 2501, EN 1092-1, BS 10, ANSI B 16.5, JIS B 2210-1984, GOST 12821-80.
You can nd additional information in the supplementary instructions manual "Flanges according to DIN-EN-ASME-JIS" (Document-ID
31088).
The electronics module VEGAFLEX series 80 is a replacement part
for TDR sensors of VEGAFLEX series 80. There is a dierent version available for each type of signal output.
You can nd further information in the operating instructions manual "Electronics module VEGAFLEX series 80".
Display and adjustment module with heating
Rod extension
44217-EN-130910
The display and adjustment module can be optionally replaced by a
display and adjustment module with heating function. You can use this display and adjustment module in an ambient tem-
perature range of -40 … +70 °C. You can nd further information in the operating instructions "Display
and adjustment module with heating" (Document-ID 31708).
If you are using an instrument with rod version, you can extend the rod probe individually with curved segments and rod extensions of dierent lengths.
All extensions used must not exceed a total length of 6 m (19.7 ft). The extensions are available in the following lengths:
Rod: ø 12 mm (0.472 in)
Basic segments: 20 … 5900 mm (0.79 … 232 in)
15
3 Product description
Rod segments: 20 … 5900 mm (0.79 … 232 in)
Curved segments: 100 x 100 mm (3.94 … 3.94 in)
You can nd further information in the operating instructions manual "Rod extension VEGAFLEX series 80".
Bypass tube
Spacer
The combination of a bypass tube and a VEGAFLEX 81 enables con-
tinuous level measurement outside the vessel. The bypass consists of a standpipe which is mounted as a communicating container on the side of the vessel via two process ttings. This kind of mounting ensures that the level in the standpipe and the level in the vessel are the same.
The length and the process ttings can be congured individually. No dierent connection versions available.
You can nd further information in the operating instructions manual "Bypass tube VEGAPASS 81".
If you mount the VEGAFLEX 81 in a bypass tube or standpipe, you have to avoid contact to the bypass tube by using a spacer at the probe end.
You can nd additional information in the operating instructions manual "Centering".
16
44217-EN-130910
Screwing in

4 Mounting

4 Mounting

4.1 General instructions

On instruments with process tting thread, the hexagon must be tight­ened with a suitable screwdriver. Wrench size see chapter "Dimen- sions".
Warning:
The housing must not be used to screw the instrument in! Applying tightening force can damage internal parts of the housing.
Protection against mois­ture
Protective caps
Suitability for the process conditions
Protect your instrument further through the following measures against moisture penetration:
Use the recommended cable (see chapter "Connecting to power
supply") Tighten the cable gland
Loop the connection cable downward in front of the cable gland
This applies particularly to:
Outdoor mounting
Installations in areas where high humidity is expected (e.g. through
cleaning processes) Installations on cooled or heated vessels
In the case of instrument housings with self-sealing NPT threads, it is not possible to have the cable entries screwed in at the factory. The openings for the cable glands are therefore covered with red protec­tive caps as transport protection.
Prior to setup you have to replace these protective caps with ap­proved cable glands or close the openings with suitable blind plugs.
The suitable cable glands and blind plugs come with the instrument.
Make sure that all parts of the instrument exposed to the process are suitable for the existing process conditions.
These are mainly:
Active measuring component
Process tting
Process seal
Process conditions are particularly:
Process pressure
Process temperature
Chemical properties of the medium
Abrasion and mechanical inuences
You can nd the specications of the process conditions in chapter "Technical data" as well as on the nameplate.
Installation position
44217-EN-130910

4.2 Mounting instructions

Mount VEGAFLEX 81 in such a way that the distance to vessel instal­lations or to the vessel wall is at least 300 mm (12 in). In non-metallic
17
4 Mounting
vessels, the distance to the vessel wall should be at least 500 mm (19.7 in).
During operation, the probe must not touch any installations or the vessel wall. If necessary, fasten the probe end.
In vessels with conical bottom it can be advantageous to mount the sensor in the center of the vessel, as measurement is then possible nearly down to the lowest point of the bottom. Keep in mind that measurement all the way down to the tip of the probe may not be pos­sible. The exact value of the min. distance (lower dead band) is stated in chapter "Technical data".
Fig. 7: Vessel with conical bottom
Type of vessel
18
Plastic vessel/Glass vessel
The guided microwave principle requires a metal surface on the pro­cess tting. Therefore use in plastic vessels etc. an instrument version with ange (from DN 50) or place a metal sheet (ø > 200 mm/8 in) beneath the process tting when screwing it in.
Make sure that the plate has direct contact with the process tting. When installing rod or cable probes in vessels without metal walls,
e.g. in plastic vessels, the measured value can be inuenced by strong electromagnetic elds (emitted interference according to EN 61326: class A). In this case, use a probe with coaxial version.
44217-EN-130910
1 2
dh
Fig. 8: Installation in non-metallic vessel
1 Flange 2 Metal sheet
4 Mounting
Socket
If possible, avoid sockets. Mount the sensor ush with the vessel top. If this is not possible, use short sockets with small diameter.
Higher sockets or sockets with a bigger diameter can generally be used. They can, however, increase the upper blocking distance (dead band). Check if this is relevant for your measurement.
In such cases, always carry out a false signal suppression after instal­lation. You can nd further information under "Setup procedure".
_
<
150
_
<
100
Fig. 9: Mounting socket
DN40 ... DN150
> DN150 ... DN200
d
h
When welding the socket, make sure that the socket is ush with the vessel top.
44217-EN-130910
19
4 Mounting
1 2
Fig.10:Socketmustbeinstalledush
1 Unfavourable installation
2 Socketush-optimuminstallation
Welding work
Inowingmedium
Measuring range
Before beginning the welding work, remove the electronics module from the sensor. By doing this, you avoid damage to the electronics through inductive coupling.
Do not mount the instruments in or above the lling stream. Make sure that you detect the product surface, not the inowing product.
Fig.11:Mountingofthesensorwithinowingmedium
The reference plane for the measuring range of the sensors is the
sealing surface of the thread or ange. Keep in mind that a min. distance must be maintained below the refer-
ence plane and possibly also at the end of the probe - measurement in these areas is not possible (dead band). The length of the cable can be used all the way to the end only when measuring conductive
44217-EN-130910
20
4 Mounting
products. These blocking distances for dierent mediums are listed in chapter "Technical data". Keep in mind for the adjustment that the default setting for the measuring range refers to water.
Pressure
Standpipes or bypass tubes
The process tting must be sealed if there is gauge or low pressure in
the vessel. Before use, check if the seal material is resistant against the measured product and the process temperature.
The max. permissible pressure is specied in chapter "Technical data" or on the type label of the sensor.
Standpipes or bypass tubes are normally metal tubes with a diameter of 30 … 200 mm (1.18 … 7.87 in). In measurement technology, such a tube corresponds to a coax probe. It does not matter if the stand­pipe is perforated or slotted for better mixing. Lateral inlets in bypass tubes also do not inuence the measurement.
For bypass tubes, select the probe length such that the blocking distance (dead band) of the probe is above or below the lateral lling openings. You can thus measure the complete range of the medium in the bypass tube. When designing the bypass tube, keep the blocking distance of the probe in mind and select the length above the upper lateral lling opening accordingly.
Microwaves can penetrate many plastics. For process technical rea­sons, plastic standpipes are problematic. If durability is no problem, then we recommend the use of metal standpipes.
When the VEGAFLEX 81 is used in standpipes or bypass tubes, contact with the tube wall must be avoided. We recommend for this purpose a cable probe with centering weight.
With rod probes, a spacer is generally not required. However, if there is a risk of the rod probe being pressed against the tube wall by in­owing medium, you should mount a spacer at the probe end to avoid contact with the tube wall. In the case of cable probes, the cable can be strained.
Keep in mind that buildup can form on the spacers. Strong buildup can inuence the measurement.
44217-EN-130910
21
4 Mounting
Fasten
1
Fig. 12: Position of the spacer or centering weight
1 Rod probe with spacer (PEEK) 2 Cable probe with centering weight 3 Spacer (PEEK) on the gravity weight of a cable probe
Note:
Measurement in a standpipe is not recommended for extremely adhesive products.
Instructions for the measurement:
The 100 % point should not be above the upper tube connection
to the vessel The 0 % point should not be below the lower tube connection to
the vessel A false signal suppression with installed sensor is generally rec-
ommended to achieve maximum possible accuracy.
If there is a risk of the cable probe touching the vessel wall during operation due to product movements or agitators, etc., the measuring probe should be securely xed.
In the gravity weight there is an internal thread (M8), e.g. for an eye­bolt (optional) - (article no. 2.1512).
Make sure that the probe cable is not completely taut. Avoid tensile loads on the cable.
Avoid undened vessel connections, i.e. the connection must be either grounded reliably or isolated reliably. Any undened change of this requirement can lead to measurement errors.
2 3
44217-EN-130910
22
4 Mounting
Lateral installation
1
1
2
2
Fig. 13: Fasten the probe
1 Probe 2 Retaining sleeve
In case of dicult installation conditions, the probe can be also mounted laterally. For this purpose, adapt the rod with rod extensions or bow-shaped segments.
Let the probe length determine automatically by the instrument to compensate the resulting running time changes.
The determine probe length can deviate from the actual probe length when using bow-shaped segments.
If installations such as struts, ladders, etc. exist on the vessel wall, then the probe should have a distance to the vessel wall of at least 300 mm (11.81 in).
You can nd further information in the supplementary instructions of the rod extension.
Rod extension
44217-EN-130910
In case of dicult installation conditions, for example in a socket, the probe can be adapted respectively with a rod extension.
Let the probe length determine automatically by the instrument to compensate the resulting running time changes.
You can nd further information in the supplementary instructions of the rod extension.
23

5 Connecting to power supply

Safety instructions
5 Connecting to power supply

5.1 Preparing the connection

Always keep in mind the following safety instructions:
Connect only in the complete absence of line voltage
If overvoltage surges are expected, overvoltage arresters should
be installed
Voltage supply
Connection cable
Cable gland ½ NPT
Cable screening and grounding
The voltage supply is provided by a Probus DP /PA segment coupler.
The voltage supply range can dier depending on the instrument version. You can nd the data for voltage supply in chapter "Technical data".
Connection is made with screened cable according to the Probus specication. Power supply and digital bus signal are carried over the same two-wire connection cable.
Use cable with round cross-section. A cable outer diameter of 5 … 9 mm (0.2 … 0.35 in) ensures the seal eect of the cable gland. If you are using cable with a dierent diameter or cross-section, exchange the seal or use a suitable cable gland.
Please make sure that your installation is carried out according to the Probus specication. In particular, make sure that the termination of the bus is done with appropriate terminating resistors.
You can nd detailed information of the cable specication, installa­tion and topology in the "ProbusPA-UserandInstallationGuide- line" on www.probus.com.
With plastic housing, the NPT cable gland or the Conduit steel tube
must be screwed without grease into the threaded insert. Max. torque for all housings see chapter "Technical data".
Make sure that the cable screen and grounding are carried out ac­cording to Fieldbus specication. We recommend to connect the cable screen to ground potential on both ends.
With systems with potential equalisation, connect the cable screen directly to ground potential at the power supply unit, in the connection box and at the sensor. The screen in the sensor must be connected directly to the internal ground terminal. The ground terminal outside on the housing must be connected to the potential equalisation (low impedance).
Connection technology
24

5.2 Connecting

The voltage supply and signal output are connected via the spring-
loaded terminals in the housing. The connection to the display and adjustment module or to the inter-
face adapter is carried out via contact pins in the housing.
44217-EN-130910
5 Connecting to power supply
Information:
The terminal block is pluggable and can be removed from the electronics. To do this, lift the terminal block with a small screwdriver and pull it out. When reinserting the terminal block, you should hear it snap in.
Connection procedure
Proceed as follows:
1. Unscrew the housing cover
2. If a display and adjustment module is installed, remove it by turn­ing it slightly to the left.
3. Loosen compression nut of the cable entry
4. Remove approx. 10 cm (4 in) of the cable mantle, strip approx. 1 cm (0.4 in) of insulation from the ends of the individual wires
5. Insert the cable into the sensor through the cable entry
Fig. 14: Connection steps 5 and 6 - Single chamber housing
44217-EN-130910
Fig. 15: Connection steps 5 and 6 - Double chamber housing
25
5 Connecting to power supply
2
Electronics and connec­tion compartment
6. Insert the wire ends into the terminals according to the wiring plan
Information:
Solid cores as well as exible cores with wire end sleeves are insert­ed directly into the terminal openings. In case of exible cores without end sleeves, press the terminal from above with a small screwdriver; the terminal opening is freed. When the screwdriver is released, the terminal closes again.
You can nd further information on the max. wire cross-section under "Technical data/Electromechanical data"
7. Check the hold of the wires in the terminals by lightly pulling on
them
8. Connect the screen to the internal ground terminal, connect the outer ground terminal to potential equalisation
9. Tighten the compression nut of the cable entry. The seal ring must completely encircle the cable
10. Reinsert the display and adjustment module, if one was installed
11. Screw the housing cover back on
The electrical connection is hence nished.

5.3 Wiring plan, single chamber housing

The following illustration applies to the non-Ex, Ex-ia and Ex-d ver-
sion.
3
00
0
1
1
9
9
2
2
1
8
8
3
1
0
Bus
3
7
7
4
4
6
6
5
5
4
26
+
( )
(-)
1
2
5
678
5
1
Fig. 16: Electronics and connection compartment, single chamber housing
1 Voltage supply, signal output 2 For display and adjustment module or interface adapter 3 Selection switch for bus address 4 For external display and adjustment unit 5 Ground terminal for connection of the cable screen

5.4 Wiring plan, double chamber housing

The following illustrations apply to the non-Ex as well as to the Ex-ia
version.
44217-EN-130910
2
2
Electronics compartment
Connection compartment
5 Connecting to power supply
3
00
0
1
1
9
9
2
2
1
8
8
3
1
0
Bus
( )
+
1
3
7
7
4
4
6
6
5
5
(-)
2
5
678
11
Fig. 17: Electronics compartment, double chamber housing
1 Internal connection to the connection compartment 2 Contact pins for the display and adjustment module or interface adapter 3 Selection switch for bus address
Information:
The connection of an external display and adjustment unit is not pos­sible with this double chamber housing.
Bus
Connection compartment
- Radio module PLICS­MOBILE
44217-EN-130910
(-)
( )
+
1
2
1
Fig. 18: Connection compartment, double chamber housing
1 Voltage supply, signal output 2 For display and adjustment module or interface adapter 3 Ground terminal for connection of the cable screen
SIM-Card
( )
+
1
2
Status
Test
USB
(-)
1
Fig. 19: Connection compartment radio module PLICSMOBILE
1 Voltage supply
3
27
5 Connecting to power supply
2
Electronics compartment
Connection compartment
You can nd detailed information on connection in the supplementary instructions "PLICSMOBILE GSM/GPRS radio module".

5.5 Wiring plan, Ex-d-ia double chamber housing

3
00
0
1
1
9
9
2
2
1
8
8
3
1
0
Bus
+
( )
1
Fig. 20: Electronics compartment, double chamber housing
1 Internal connection to the connection compartment 2 Contact pins for the display and adjustment module or interface adapter 3 Selection switch for bus address 4 Internal connection to the plug connector for external display and adjust-
ment unit (optional)
3
7
7
4
4
6
6
5
5
(-)
2
5
678
41
28
Bus
( )
(-)
+
1
2
2
1
Fig. 21: Connection compartment, double chamber housing Ex d
1 Voltage supply, signal output 2 Ground terminal for connection of the cable screen
44217-EN-130910
Electronics compartment
Assignment of the plug connector
5 Connecting to power supply
1
2
3
Fig. 22: View to the electronics compartment
1 DIS-ADAPT 2 Internal plug connection 3 Plug connector M12 x 1
34
Wire assignment, con­nection cable
44217-EN-130910
1
Fig. 23: Top view of the plug connector
1 Pin 1 2 Pin 2 3 Pin 3 4 Pin 4
Contact pin Colour connection ca-
ble in the sensor
Pin 1 Brown 5
Pin 2 White 6
Pin 3 Blue 7
Pin 4 Black 8
2
Terminal, electronics module

5.6 Wiring plan - version IP 66/IP 68, 1 bar

1
2
Fig.24:Wireassignmentx-connectedconnectioncable
1 brown (+) and blue (-) to power supply or to the processing system 2 Shielding
29
5 Connecting to power supply
Supplementary elec­tronics - Radio module PLICSMOBILE

5.7 Supplementary electronics

The radio module PLICSMOBILE is an external GSM/GPRS radio
unit for transmission of measured values and for remote parameter adjustment.
Instrument address
SIM-Card
( )
+
1
2
Status
Test
USB
(-)
1
Fig. 25: Radio module PLICSMOBILE integrated in the connection compart­ment
1 Voltage supply
You can nd detailed information on connection in the supplementary instructions "PLICSMOBILE GSM/GPRS radio module".

5.8 Set instrument address

An address must be assigned to each Probus PA instrument. The approved addresses are between 0 and 126. Each address must only be assigned once in the Probus PA network. The sensor is only recognized by the control system if the address is set correctly.
When the instrument is shipped, address 126 is adjusted. This ad­dress can be used for function test of the instrument and for con­nection to a Probus PA network. Then address must be changed to integrate additional instruments.
The address setting is carried out either via:
The address selection switch in the electronics compartment of
the instrument (address setting via hardware) The display and adjustment module (address setting via software)
PACTware/DTM (address setting via software)
Hardware addressing
30
The hardware addressing is eective if an address <126 is adjusted
with the address selection switches on the instrument. Hence the software addressing is no longer eective, the adjusted hardware address is valid.
44217-EN-130910
5 Connecting to power supply
1
2
3
00
0
1
1
9
9
2
2
1
8
8
3
1
0
Bus
( )
+
1
2
Fig. 26: Address selection switch
1 Addresses <100 (selection 0), addresses >100 (selection 1) 2 Decade of the address (selection 0 to 9) 3 Unit position of the address (selection 0 to 9)
3
7
7
4
4
6
6
5
5
(-)
5
678
Software addressing
The software addressing is only eective if address 126 or higher is
adjusted on the instrument with the address selection switches. The addressing procedure is described in the operating instructions
manual "Display and adjustment module.

5.9 Switch-on phase

After VEGAFLEX 81 is connected to the bus system, the instrument carries out a self-test for approx. 30 seconds. The following steps are
carried out:
Internal check of the electronics
Indication of the instrument type, hardware and software version,
measurement loop name on the display or PC Indication of the status message "F 105 Determine measured
value" on the display or PC Status byte goes briey to fault value
As soon as a plausible measured value is found, it is outputted to the signal cable. The value corresponds to the actual level as well as the settings already carried out, e.g. factory settings.
44217-EN-130910
31

6 Set up with the display and adjustment module

6 Set up with the display and adjustment
module

6.1 Insert display and adjustment module

The display and adjustment module can be inserted into the sensor and removed any time. Four positions displaced by 90° can be se­lected. It is not necessary to interrupt the power supply.
Proceed as follows:
1. Unscrew the housing cover
2. Place the display and adjustment module in the requested posi­tion onto the electronics and turn to the right until it snaps in
3. Screw housing cover with inspection window tightly back on
Removal is carried out in reverse order. The display and adjustment module is powered by the sensor, an ad-
ditional connection is not necessary.
32
Fig. 27: Insertion of the display and adjustment module with single chamber housing into the electronics compartment
44217-EN-130910
6 Set up with the display and adjustment module
1 2
Fig. 28: Insertion of the display and adjustment module into the double chamber housing
1 In the electronics compartment 2 In the connection compartment (with Ex-d-ia version not possible)
Note:
If you intend to retrot the instrument with a display and adjustment module for continuous measured value indication, a higher cover with an inspection glass is required.
Key functions
44217-EN-130910

6.2 Adjustment system

Fig. 29: Display and adjustment elements
1 LC display 2 Adjustment keys
[OK] key:
Move to the menu overview
1
2
33
6 Set up with the display and adjustment module
Conrm selected menu – Edit parameter – Save value
[->] key:
Presentation, change measured valueSelect list entrySelect editing position
[+] key:
Change value of the parameter
[ESC] key:
– Interrupt input – Jump to next higher menu
Adjustment system
Switch-on phase
Measured value indica­tion
The sensor is adjusted via the four keys of the display and adjust-
ment module. The LC display indicates the individual menu items. The functions of the individual keys are shown in the above illustration. Approx. 10 minutes after the last pressing of a key, an automatic reset to measured value indication is triggered. Any values not conrmed with [OK] will not be saved.
After switching on, the VEGAFLEX 81 carries out a short self-test where the device software is checked.
The output signal transmits a fault signal during the switch-on phase. The following information is displayed on the display and adjustment
module during the startup procedure:
Instrument type
Device name
Software version (SW-Ver)
Hardware version (HW-Ver)
With the [->] key you can move between three dierent indication
modes. In the rst view, the selected measured value is displayed in large
digits. In the second view, the selected measured value and a correspond-
ing bar graph presentation are displayed. In the third view, the selected measured value as well as a second
selectable value, e.g. the temperature, are displayed.
Quick setup
34
44217-EN-130910

6.3 Parameter adjustment - Quick setup

To quickly and easily adapt the sensor to the application, select
the menu item "Quick setup" in the start graphic on the display and adjustment module.
General information
6 Set up with the display and adjustment module
You can nd "Extended adjustment" in the next sub-chapter.
Sensor address
In the rst menu item you have to enter a sensor address. The selec­tion switches on the electronics module are preset to sensor address
126. This means that the sensor address can be changed via the
display and adjustment module.
If you set a sensor address with the selection switches which is lower than 126, the set value is applicable. In such case, the address setting via the display and adjustment module has no eect.
Measurement loop name
In the next menu item you can assign a suitable measurement loop name. You can enter a name with max. 19 characters.
Application
In this menu item, you can select the application. You can choose between level measurement and interface measurement. You can also choose between measurement in a vessel or in a bypass or standpipe.
Level measurement
44217-EN-130910
Medium - dielectric constant
In this menu item, you can dene the type of medium (product).
Max. adjustment
In this menu item, you can enter the max. adjustment for the level. Enter the appropriate distance value in m (corresponding to the
percentage value) for the full vessel. The distance refers to the sensor reference plane (seal surface of the process tting). Keep in mind that the max. level must lie below the dead band.
Min. adjustment
In this menu item, you can enter the min. adjustment for the level. Enter the suitable distance value in m for the empty vessel (e.g.
distance from the ange to the probe end) corresponding to the per­centage value. The distance refers tot he sensor reference plane (seal surface of the process tting).
35
6 Set up with the display and adjustment module
Interface measurement
Dielectric constant - upper medium
In this menu item, you can dene the type of medium (product).
Max. adjustment
In this menu item, you can enter the max. adjustment for the level. Enter the appropriate distance value in m (corresponding to the
percentage value) for the full vessel. The distance refers to the sensor reference plane (seal surface of the process tting). Keep in mind that the max. level must lie below the dead band.
Min. adjustment
In this menu item, you can enter the min. adjustment for the level. Enter the suitable distance value in m for the empty vessel (e.g.
distance from the ange to the probe end) corresponding to the per­centage value. The distance refers tot he sensor reference plane (seal surface of the process tting).
Max. adjustment - Interface
Carry out the max. adjustment for the interface. To do this, enter the percentage value and the suitable distance value
in m for the full vessel.
Min. adjustment - Interface
Carry out the min. adjustment for the interface. To do this, enter the percentage value and the suitable distance value
in m for the empty vessel.
Linearization
36
Linearization
A linearization is necessary for all vessels in which the vessel volume does not increase linearly with the level - e.g. a horizontal cylindri­cal or spherical tank, when the indication or output of the volume is required. Corresponding linearization curves are preprogrammed for these vessels. They represent the correlation between the level percentage and vessel volume.
The linearization applies for the measured value indication and the current output. By activating the suitable curve, the percentage vessel volume is displayed correctly.
44217-EN-130910
6 Set up with the display and adjustment module
False signal suppression
High sockets and internal vessel installations cause interfering reec­tions and can inuence the measurement.
A false signal suppression detects, marks and saves these false signals so that they are no longer taken into account for the level and interface measurement. We generally recommend carrying out a false signal suppression to achieve the best possible accuracy. This should be done with the lowest possible level so that all potential interfering reections can be detected.
Enter the actual distance from the sensor to the product surface. All interfering signals in this section are detected by the sensor and
stored. The instrument carries out an automatic false signal suppression
as soon as the probe is uncovered. The false signal suppression is always updated.
AI FB1 Channel
In this menu item you can select the function of the rst Function Block. AI stands for Analog Input.
With this you can adjust the value for the Primary Value (PV). Further settings (SV, TV) must be carried out via PACTware.
Main menu
44217-EN-130910

6.4 Parameter adjustment - Extended adjustment

For technically demanding measurement loops you can carry out extended settings in "Extended adjustment".
The main menu is divided into ve sections with the following func-
tions:
Setup: Settings, e.g. measurement loop name, medium, application, vessel, adjustment, AI FB 1 Channel - Scaling - Damping, device units, false signal suppression, linearization
Display: Language setting, settings for the measured value indication as well as lighting
Diagnosis: Information, for example on the instrument status, pointer, reliability, AI FB 1 simulation, echo curve
37
6 Set up with the display and adjustment module
Additional adjustments: Sensor address, PIN, date/time, reset,
copy sensor data
Info: Instrument name, hardware and software version, date of manu­facture, instrument features
Note:
For optimum adjustment of the measurement, the individual submenu items in the main menu item "Setup" should be selected one after the other and provided with the correct parameters. If possible, go through the items in the given sequence.
The procedure is described below. The following submenu points are available:
The submenu points described below.
Setup - Instrument ad­dress
Hardware addressing
Software addressing
Setup - Measurement loop name
An address must be assigned to each Probus PA instrument. Each address may only be assigned once in the Probus PA network. The sensor is only recognized by the control system if the address is set correctly.
When the instrument is shipped, address 126 is adjusted. This ad­dress can be used for function test of the instrument and for con­nection to a Probus PA network. Then address must be changed to integrate additional instruments.
The address setting is carried out either via:
The address selection switch in the electronics compartment of
the instrument (address setting via hardware) The display and adjustment module (address setting via software)
PACTware/DTM (address setting via software)
The hardware addressing is eective if an address <126 is set
with the address selection switches on the electronics module of VEGAFLEX 81. Software addressing thus has no eect - only the set hardware address applies.
The software addressing is only eective if address 126 or higher is
adjusted on the instrument with the address selection switches.
Here you can assign a suitable measurement loop name. Push the "OK" key to start the processing. With the "+" key you change the sign and with the "->" key you jump to the next position.
You can enter names with max. 19 characters. The character set
comprises:
44217-EN-130910
38
6 Set up with the display and adjustment module
Capital letters from A … Z
Numbers from 0 … 9
Special characters + - / _ blanks
Setup - Units
Setup - Units (2)
Setup - Probe length
Setup - Application - Type of medium
In this menu item you select the distance unit and the temperature unit.
With the distance units you can choose between m, mm and ft and with the temperature units betwenn °C, °F and K.
In this menu item, you select the unit of the Secondary Value (SV2).
It can be selected from the distance units such as for example m, mm and ft.
In this menu item you can enter the probe length or have the length determined automatically by the sensor system.
When choosing "Ye s ", then the probe length will be determined automatically. When choosing "No", you can enter the probe length manually.
In this menu item you can select which type of medium you want to measure. You can choose between liquid or bulk solid.
Setup - Application - Ap­plication
44217-EN-130910
In this menu item, you can select the application. You can choose between level measurement and interface measurement. You can also choose between measurement in a vessel or in a bypass or standpipe.
39
6 Set up with the display and adjustment module
Note:
The selection of the application has a considerable inuence on all other menu items. Keep in mind that as you continue with the param­eter adjustment, individual menu items are only optionally available.
You have the option of choosing the demonstration mode. This mode is only suitable for test and demonstration purposes. In this mode, the sensor ignores the parameters of the application and reacts immedi­ately to each change.
Setup - Application - Me­dium, dielectric constant
Setup - Application - Gas phase
Setup - Application - Di­electric constant
In this menu item, you can dene the type of medium (product). This menu item is only available if you have selected level measure-
ment under the menu item "Application".
You can choose between the following medium types:
Dielectric con­stant
> 10 Water-based liq-
3 … 10 Chemical mix-
< 3 Hydrocarbons Solvents, oils, liquid gas
Type of medium Examples
uids
tures
Acids, alcalis, water
Chlorobenzene, nitro lacquer, aniline, isocyanate, chloroform
This menu item is only available, if you have chosen interface meas-
urement under the menu item "Application". In this menu item you can enter if there is a superimposed gas phase in your application.
Only set the function to "Ye s ", if the gas phase is permanently pre­sent.
This menu item is only available if you have selected interface meas-
urement under the menu item "Application". In this menu item you can choose the type of medium of the upper medium.
44217-EN-130910
40
6 Set up with the display and adjustment module
You can enter the dielectric constant of the upper medium directly or have the value determined by the instrument. To do this you have to enter the measured or known distance to the interface.
Setup - Max. adjustment Level
Setup - Min. adjustment Level
In this menu item you can enter the max. adjustment for the level. With interface measurement this is the maximum total level.
Adjust the requested percentage value with [+] and store with [OK].
Enter the appropriate distance value in m (corresponding to the percentage value) for the full vessel. The distance refers to the sensor reference plane (seal surface of the process tting). Keep in mind that the max. level must lie below the dead band.
In this menu item you can enter the min. adjustment for the level. With interface measurement this is the minimum total level.
Adjust the requested percentage value with [+] and store with [OK].
44217-EN-130910
Enter the suitable distance value in m for the empty vessel (e.g. distance from the ange to the probe end) corresponding to the per­centage value. The distance refers tot he sensor reference plane (seal surface of the process tting).
41
6 Set up with the display and adjustment module
Setup - Max. adjustment ­Interface
Setup - Min. adjustment ­Interface
This menu item is only available if you have selected interface meas-
urement under the menu item "Application".
You can accept the adjustment of the level measurement also for the interface measurement. If you select "Yes", the current setting will be displayed.
If you have selected "No", you can enter the adjustment for the inter­face separately. Enter the requested percentage value.
For the full vessel, enter the distance value in m matching the per­centage value.
This menu item is only available if you have selected interface meas-
urement under the menu item "Application". If you have selected "Yes" in the previous menu item (accept adjustment of the level measure­ment), the current setting will be displayed.
Setup - False signal sup­pression
42
If you have selected "No", you can enter the adjustment for the inter­face measurement separately.
Enter the respective distance value in m for the empty vessel cor­responding to the percentage value.
The following circumstances cause interfering reections and can
inuence the measurement:
High sockets
Vessel installations such as struts
Note:
A false signal suppression detects, marks and saves these false signals so that they are no longer taken into account for the level and interface measurement. We generally recommend carrying out a false signal suppression to achieve the best possible accuracy. This should
44217-EN-130910
6 Set up with the display and adjustment module
be done with the lowest possible level so that all potential interfering reections can be detected.
Proceed as follows:
Enter the actual distance from the sensor to the product surface.
All interfering signals in this section are detected by the sensor and stored.
Note:
Check the distance to the product surface, because if an incorrect (too large) value is entered, the existing level will be saved as a false echo. The lling level would then no longer be detectable in this area.
If a false signal suppression has already been created in the sensor, the following menu window appears when selecting "False signal suppression":
Setup - Linearization
44217-EN-130910
The instrument carries out an automatic false signal suppression as soon as the probe is uncovered. The false signal suppression is always updated.
The menu item "Delete" is used to completely delete an already cre­ated false signal suppression. This is useful if the saved false signal suppression no longer matches the metrological conditions in the vessel.
A linearization is necessary for all vessels in which the vessel volume does not increase linearly with the level - e.g. a horizontal cylindri­cal or spherical tank, when the indication or output of the volume is required. Corresponding linearization curves are preprogrammed for these vessels. They represent the correlation between the level percentage and vessel volume.
The linearization applies to the measured value indication and the current output. By activating the appropriate curve, the volume per­centage of the vessel is displayed correctly. If the volume should not be displayed in percent but e.g. in l or kg, a scaling can be also set in the menu item "Display".
43
6 Set up with the display and adjustment module
Warning:
If a linearization curve is selected, the measuring signal is no longer necessarily linear to the lling height. This must be considered by the user especially when adjusting the switching point on the limit signal transmitter.
In the following, you have to enter the values for your vessel, for example the vessel height and the socket correction.
For non-linear vessel forms, enter the vessel height und the socket correction.
For the vessel height, you have to enter the total height of the vessel. For the socket correction you have to enter the height of the socket
above the upper edge of the vessel. If the socket is lower than the up­per edge of the vessel, this value can also be negative.
+ h
- h
Setup - AI FB1
44
D
Fig. 30: Vessel height und socket correction value
D Vessel height +h Positive socket correction value
-h Negative socket correction value
44217-EN-130910
Since the adjustment is very comprehensive, the menu points of Function Blocks 1 (FB1) were put together in a submenu.
6 Set up with the display and adjustment module
Setup - AI FB1 - Channel
Setup - AI FB1 - Scaling unit
Setup - AI FB1 - Scaling
In menu item"Channel" you determine which measured value the output refers to.
In menu item "Scaling unit" you dene the scaling variable and the scaling unit for the level value on the display, e.g. volume in l.
In menu item "Scaling" you dene the scaling format on the display and the scaling of the measured level values for 0 % and 100 %.
Level measured value min.
Measured level value max.
Setup - AI FB1 - Damping
Lock/release setup - Ad­justment
44217-EN-130910
To damp process-dependent measured value uctuations, you can
set a time of 0 … 999 s in this menu item. The damping applies to the level and interface measurement.
The default setting is a damping of 0 s.
In the menu item "Lock/unlock adjustment", you can protect the sensor parameters against unauthorized modication. The PIN is activated/deactivated permanently.
45
6 Set up with the display and adjustment module
The following adjustment functions are possible without entering the
PIN:
Select menu items and show data
Read data from the sensor into the display and adjustment mod-
ule.
Caution:
With active PIN, adjustment via PACTware/DTM as well as other systems is also blocked.
You can change the PIN number under "Additional adjustments - PIN".
Display
Display - Menu language
Display - Displayed value 1
In the main menu point "Display", the individual submenu points should be selected subsequently and provided with the correct parameters to ensure the optimum adjustment of the display options. The procedure is described in the following.
The following submenu points are available:
The submenu points described below.
This menu item enables the setting of the requested national lan-
guage.
In the delivery status, the sensor is set to the ordered national lan­guage.
In this menu item, you dene the indication of the measured value on the display. You can display two dierent measured values. In this menu item, you dene measured value 1.
44217-EN-130910
The default setting for the displayed value 1 is "Filling height Level".
Display - Displayed value 2
46
In this menu item, you dene the indication of the measured value on the display. You can display two dierent measured values. In this menu item, you dene measured value 2.
6 Set up with the display and adjustment module
The default setting for the displayed value 2 is the electronics tem­perature.
Display - Backlight
Diagnostics - Device status
Diagnostics - Peak values Distance
The optionally integrated background lighting can be adjusted via the
adjustment menu. The function depends on the height of the supply voltage, see "Technical data".
The lighting is switched o in the delivery status.
In this menu item, the device status is displayed.
The respective min. and max. measured value is saved in the sen-
sor. The two values are displayed in the menu item "Peak values, distance".
If you have selected interface measurement under the menu item "Setup - Application", the peak values of the interface measurement are displayed in addition to the peak values of the level measurement.
In another window you can carry out a reset of the two peak values separately.
Diagnostics - Peak values Measurement certainty
44217-EN-130910
The respective min. and max. measured values are saved in the
sensor. The two values are displayed in the menu item "Peak values, measurement certainty".
The measurement can be inuenced by the process conditions. In this menu item, the measurement certainty of the level measurement is displayed as percentage value. The higher the value, the more reli­able the measurement. Values > 90 % indicate reliable measurement.
If you have selected interface measurement under the menu item "Setup - Application", the peak values of the interface measurement are displayed in addition to the peak values of the level measurement.
47
6 Set up with the display and adjustment module
In another window you can carry out a reset of the two peak values separately.
Diagnostics - Peak values Additional
Diagnostics - Echo curve
The respective min. and max. measured values are saved in the
sensor. The values are displayed in the menu item "Peak values Ad- ditional".
This menu item displays the peak values of the electronics tempera­ture as well as the dielectric constant.
In another window you can carry out a reset of the two peak values separately.
The menu item "Echo curve" shows the signal strength of the echoes
over the measuring range in V. The signal strength enables an evalua­tion of the quality of the measurement.
With the following functions you can zoom part sections of the echo curve.
"X-Zoom": Zoom function for the meas. distance
"Y-Zoom": 1, 2, 5 and 10x signal magnication in "V"
"Unzoom": Reset the presentation to the nominal measuring range
with single magnication
Diagnosis - Simulation
48
44217-EN-130910
In this menu item you can simulate measured values via the current output. This allows the signal path to be tested, e.g. via downstream indicating instruments or the input card of the control system.
6 Set up with the display and adjustment module
Select the requested simulation size and adjust the requested value.
Caution:
During simulation, the simulated value is outputted as 4 … 20 mA cur­rent value and digital HART signal.
Push the [ESC] key to deactivate the simulation.
Information:
The simulation is terminated automatically 60 minutes after the last key has been pushed.
Diagnostics - Echo curve memory
Additional settings - PIN
With the menu item "Setup" the echo curve it is possible to save at
the time of setup. This is generally recommended; for using the Asset Management functions it is necessary. If possible, the curve should be saved with a low level in the vessel.
With this, you can detect signal changes over the operating time. With the adjustment software PACTware and the PC, the high-resolution echo curve can be displayed and used to compare the echo curve of the setup with the actual echo curve.
The function "Echo curve memory" enables storing echo curves of the measurement.
Under the sub-menu item "Echo curve memory" you can store the current echo curve.
Parameter settings for recording the echo curve and the settings of the echo curve itself can be carried out in the adjustment software PACTwa r e.
With the adjustment software PACTware and the PC the high-reso­lution echo curve can be displayed and used later on to assess the quality of the measurement.
Entering a 4-digit PIN protects the sensor data against unauthorized access and unintentional modication. In this menu item, the PIN is displayed or edited and changed. However, this menu item is only
44217-EN-130910
49
6 Set up with the display and adjustment module
available if adjustment is enabled in the menu "Lock/Release setup/ adjustment".
In delivery status, the PIN is "0000".
Additional adjustments ­Date Time
Additional adjustments
- Reset
In this menu item, the internal clock of the sensor is adjusted.
With a reset, certain parameter adjustments carried out by the user
are reset.
The following reset functions are available: Delivery status: Restoring the parameter settings at the time of ship-
ment from the factory incl. the order-specic settings. A created false signal suppression, user-programmable linearization curve as well as the measured value memory will be deleted.
Basic settings: Resetting of the parameter settings incl. special parameters to the default values (presettings) of the respective instru­ment. Any created false signal suppression or user-programmable lin­earization curve as well as the measured value memory are deleted.
The following table shows the default values of the instrument. De­pending on the instrument version or application, all menu items may not be available or some may be dierently assigned:
Setup
Menu item Default value Modiedvalue
Block adjustment Released
Measurement loop name Sensor
Units Distance unit: mm
Temperature unit: °C
Probe length Length of the probe Ex factory
Type of medium Liquid
Application Level, vessel
50
44217-EN-130910
6 Set up with the display and adjustment module
Menu item Default value Modiedvalue
Medium, dielectric constant Water-based, > 10
Superimposed gas phase Yes
Dielectric constant, upper medium (TS) 1.5
Tube inner diameter 200 mm
Max. adjustment - Level 100 %
Max. adjustment - Level Distance: 0.000 m(d) - note block-
Min. adjustment - Level 0 %
Min. adjustment - Level Distance: Probe length - take dead
Accept adjustment of the level measurement? Yes
Max. adjustment - Interface 100 %
Max. adjustment - Interface Distance: 0.000 m(d) - note block-
Min. adjustment - Interface 0 %
Min. adjustment - Interface Distance: Probe length - take dead
Integration time - Level 0.0 s
Integration time - Interface 0.0 s
Linearization type Linear
Linearization - Socket correction 0 mm
Linearization - Vessel height Probe length
ing distances
band into account
ing distances
band into account
AI FB1 Tag Descriptor
AI FB1 Channel Primary Value (lin. percent level)
AI FB1 scaling PV Scale (min.) 0 %
AI FB1 scaling PV Scale (max.) 100 %
AI FB1 Lin. Type Linear
AI FB1 Out Scale Unit %
AI FB1 Out Scale Decimal Point #.##
AI FB1 Out Scale (min.) 0 %
AI FB1 Out Scale (max.) 100 %
AI FB1 PV FTime 0 s
AI FB1 Hi Hi Limit 3.402823E+38 %
AI FB1 Hi Limit 3.402823E+38 %
AI FB1 Lo Lo Limit -3.402823E+38 %
AI FB1 Lo Limit -3.402823E+38 %
AI FB1 Hysteresis 0.50 %
44217-EN-130910
51
6 Set up with the display and adjustment module
Menu item Default value Modiedvalue
AI FB1 Fail Safe Mode (behaviour in case of mal­function)
AI FB1 Fail Safe Value 0.00 %
AI FB1 Target Mode Auto
Last Valid Out Value (last valid measured value)
Display
Menu item Default value Modiedvalue
Language Order-specic
Displayed value 1 Filling height Level
Displayed value 2 Electronics temperature
Backlight Switched o
Diagnostics
Menu item Default value Modiedvalue
Status signals - Function control Switched on
Status signals - Out of specication Switched o
Status signals - Maintenance Switched o
Device memory - Echo curve memory Stopped
Device memory - Measured value memory Started
Device memory - Measured value memory ­Measured values
Device memory - Measured value memory - Re­cording in time interval
Device memory - Measured value memory - Re­cording with measured value dierence
Device memory - Measured value memory - Start with measured value
Device memory - Measured value memory - Stop with measured value
Device memory - Measured value memory - Stop recording when memory is full
Distance level, percentage value level, reliability level, electronics
temperature
3 min.
15 %
Not active
Not active
Not active
Additional adjustments
Menu item Default value Modiedvalue
PIN 0000
Date Actual date
Time Actual time
Time - Format 24 hours
Probe type Device-specic
52
44217-EN-130910
6 Set up with the display and adjustment module
Menu item Default value Modiedvalue
AI FB2 Tag Descriptor
AI FB 2 Channel Primary Value (lin. percent level)
AI FB2 scaling PV Scale (min.) 0 %
AI FB2 scaling PV Scale (max.) 100 %
AI FB2 Lin. Type Linear
AI FB2 Out Scale Unit %
AI FB2 Out Scale Decimal Point #.##
AI FB2 Out Scale (min.) 0 %
AI FB2 Out Scale (max.) 100 %
AI FB2 PV FTime 0 s
AI FB2 Hi Hi Limit 3.402823E+38 %
AI FB2 Hi Limit 3.402823E+38 %
AI FB2 Lo Lo Limit -3.402823E+38 %
AI FB2 Lo Limit -3.402823E+38 %
AI FB2 Hysteresis 0.50 %
AI FB2 Fail Safe Mode (behaviour in case of mal­function)
AI FB2 Fail Safe Value 0.00 %
AI FB2 Target Mode Auto
Last Valid Out Value (last valid measured value)
AI FB3 Tag Descriptor
AI FB3 Channel Primary Value (lin. percent level)
AI FB1 scaling PV Scale (min.) 0 %
AI FB3 scaling PV Scale (max.) 100 %
AI FB3 Lin. Type Linear
AI FB3 Out Scale Unit %
AI FB3 Out Scale Decimal Point #.##
AI FB3 Out Scale (min.) 0 %
AI FB3 Out Scale (max.) 100 %
AI FB3 PV FTime 0 s
AI FB3 Hi Hi Limit 3.402823E+38 %
AI FB3 Hi Limit 3.402823E+38 %
AI FB3 Lo Lo Limit -3.402823E+38 %
AI FB3 Lo Limit -3.402823E+38 %
AI FB3 Hysteresis 0.50 %
AI FB3 Fail Safe Mode (behaviour in case of mal­function)
AI FB3 Fail Safe Value 0.00 %
Last Valid Out Value (last valid measured value)
44217-EN-130910
53
6 Set up with the display and adjustment module
Menu item Default value Modiedvalue
AI FB3 Target Mode Auto
Additional adjustments ­Copy instrument settings
Additional adjustments ­Probe type
The instrument settings are copied with this function. The following
functions are available:
Read from sensor: Read data from sensor and store into the
display and adjustment module Write into sensor: Store data from the display and adjustment
module back to the sensor
The following data or settings for adjustment of the display and ad­justment module are saved:
All data of the menu "Setup" and "Display"
In the menu "Additional adjustments" the items "Reset, Date/Time"
Special parameters
The copied data are permanently saved in an EEPROM memory in the display and adjustment module and remain there even in case of power failure. From there, they can be written into one or more sen­sors or kept as backup for a possible electronics exchange.
Note:
Before the data are stored in the sensor, a check is carried out to determine if the data t the sensor. If the data do not t, a fault signal is triggered or the function is blocked. When data are being written into the sensor, the display shows which instrument type the data originate from and which TAG-no. this sensor had.
In this menu item you can select the type and size of your probe from a list of all possible probes. This is necessary to adapt the electronics optimally to the probe.
Additional adjustments ­Special parameters
54
In this menu item you gain access to the protected area where you can enter special parameters. In exceptional cases, individual parameters can be modied in order to adapt the sensor to special requirements.
Change the settings of the special parameters only after having con­tacted our service sta.
44217-EN-130910
6 Set up with the display and adjustment module
Info - Instrument name
Info - Instrument version
Info - Factory calibration date
Info - Sensor character­istics
In this menu, you read out the instrument name and the instrument serial number:
In this menu item, the hardware and software version of the sensor is displayed.
In this menu item, the date of factory calibration of the sensor as well as the date of the last change of sensor parameters are displayed via the display and adjustment module or via the PC.
In this menu item, the features of the sensor such as approval, pro­cess tting, seal, measuring range, electronics, housing and others are displayed.
44217-EN-130910

6.5 Saving the parameter adjustment data

We recommended noting the adjusted data, e.g. in this operating instructions manual, and archiving them afterwards. They are thus available for multiple use or service purposes.
If the instrument is equipped with a display and adjustment module, the data in the sensor can be saved in the display and adjustment module. The procedure is described in the operating instructions manual "Display and adjustment module" in the menu item "Copy sensor data". The data remain there permanently even if the sensor power supply fails.
The following data or settings for adjustment of the display and ad­justment module are saved:
All data of the menu "Setup" and "Display"
In the menu "Additional adjustments" the items "Sensor-specic
units, temperature unit and linearization" The values of the user programmable linearization curve
The function can also be used to transfer settings from one instru­ment to another instrument of the same type. If it is necessary to exchange a sensor, the display and adjustment module is inserted
55
6 Set up with the display and adjustment module
into the replacement instrument and the data are likewise written into the sensor via the menu item "Copy sensor data".
56
44217-EN-130910
2
Via the interface adapter directly on the sensor
Prerequisites

7 Setup with PACTware

7 Setup with PACTware

7.1 Connect the PC

1
3
Fig. 31: Connection of the PC directly to the sensor via the interface adapter
1 USB cable to the PC 2 Interface adapter VEGACONNECT 3 Sensor

7.2 Parameter adjustment with PACTware

For parameter adjustment of the sensor via a Windows PC, the con­guration software PACTware and a suitable instrument driver (DTM) according to FDT standard are required. The up-to-date PACTware version as well as all available DTMs are compiled in a DTM Collec­tion. The DTMs can also be integrated into other frame applications according to FDT standard.
Note:
To ensure that all instrument functions are supported, you should always use the latest DTM Collection. Furthermore, not all described functions are included in older rmware versions. You can download the latest instrument software from our homepage. A description of the update procedure is also available in the Internet.
Further setup steps are described in the operating instructions manu­al "DTM Collection/PACTware" attached to each DTM Collection and which can also be downloaded from the Internet. Detailed descrip­tions are available in the online help of PACTware and the DTMs.
44217-EN-130910
57
7 Setup with PACTware
Fig. 32: Example of a DTM view
Standard/Full version
General information
All device DTMs are available as a free-of-charge standard version and as a full version that must be purchased. In the standard version, all functions for complete setup are already included. An assistant for simple project conguration simplies the adjustment considerably. Saving/printing the project as well as import/export functions are also part of the standard version.
In the full version there is also an extended print function for complete project documentation as well as a save function for measured value and echo curves. In addition, there is a tank calculation program as well as a multiviewer for display and analysis of the saved measured value and echo curves.
The standard version is available as a download under www.vega.com/downloads and "Software". The full version is avail­able on CD from the agency serving you.

7.3 Set up with the quick setup

The quick setup is another option for parameter adjustment of the
sensor. It allows fast, convenient adjustment of the most important parameters to adapt the sensor quickly to standard applications. To use it, select the function "Quick setup" in the start screen.
44217-EN-130910
58
7 Setup with PACTware
Fig. 33: Select quick setup
1 Quick setup 2 Extended adjustment 3 Maintenance
Quick setup
With quick setup you can carry out the parameter adjustment of VEGAFLEX 81 for your application in just a few simple steps. The assistant-driven adjustment includes the basic settings for simple, reliable setup and commissioning.
Information:
If the function is inactive, then possibly no instrument is connected. Check the connection to the instrument.
Extended adjustment
With the extended adjustment, you carry out the parameter adjust­ment for the instrument via the clear menu structure in the DTM (Device Type Manager). This enables additional and special settings over and above those oered by quick setup.
Maintenance
Under the menu item "Maintenance" you get comprehensive and important support for servicing and maintenance. You can call up diagnostic functions and carry out an electronics exchange or a software update.
Start quick setup
Step 1
Instrumentconguration
44217-EN-130910
Click to the button "Quick setup", to start the assistant-driven adjust­ment for a simplied and reliable setup.
Device name
Here, you can nd the instrument name. You cannot change this line because the instrument name is unmodiably saved in the instrument.
59
7 Setup with PACTware
Serial number
Here, you can nd the serial number of your instrument. You cannot change this line because the serial number is unmodiably saved in the instrument.
Measurement loop name
Here you can enter a suitable measurement loop name for your VE­GAFLEX 81. You can enter a name with max. 19 characters. You can use capital and small letters as well as numbers. The following special characters are also possible: + - . : , ( ) / < >
Probelengthmodied?
If you have modied the probe length, this must be entered in the selection eld.
If you select "No", then the instrument uses automatically the
preset length of the default setting. If you select "Yes ", then you can enter in another eld the modied
length of the instrument.
Instrument address
An address must be assigned to each Probus PA instrument. Each address may only be assigned once in the Probus PA network. The sensor is only recognized by the control system if the address is set correctly.
When the instrument is shipped, address 126 is adjusted. This ad­dress can be used for function test of the instrument and for con­nection to a Probus PA network. Then address must be changed to integrate additional instruments.
The address setting is carried out either via:
The address selection switch in the electronics compartment of
the instrument (address setting via hardware) The display and adjustment module (address setting via software)
PACTware/DTM (address setting via software)
Hardware addressing - The hardware addressing is eective if an
address <126 is set with the address selection switches on the electronics module of VEGAFLEX 81. The software addressing is then no longer eective, only the set hardware address is valid. Software addressing - The software addressing is only eective
if address 126 or higher is set on the instrument with the address selection switches.
60
Probe length L from seal surface
If you have modied the length of the probe, you can enter in this eld the modied probe length. Keep the selected unit in mind.
Determine probe length automatically
If you do not know the probe length, you can have the length of the probe determined automatically. The requirement for this is a probe unrestricted and not covered by the medium.
Click to "Carry out now", to start the automatic length determination.
44217-EN-130910
7 Setup with PACTware
Step 2 Application
Type of medium
Here you can see which type of medium your instrument is suitable for. If this function is inactive, the medium type your instrument is suit­able for was already preset.
Information:
In special cases you can change the type of medium. This setting can be changed under "Extended adjustment".
Application
In this eld you can select the application you want to use your instru­ment for. You have the following selection options:
Level in the vessel
Level in the bypass/standpipe
Interface in the vessel
Interface in the bypass/standpipe
Demonstration mode
Level measurement: If you select "Level", you can select the proper­ties of the medium in another eld.
Interface measurement: If you select "Interface", the instrument needs more information, such as the distance to the interface, the dielectric constant of the upper medium or whether or not there is a superim­posed gas phase.
Demonstration mode: This mode is only suitable for test and demon­stration purposes. In this mode, the sensor ignores all parameters and reacts immediately to all measured value changes within the measuring range.
Application - Level measurement
The level measurement refers to the product surface which is the limit to the gas phase.
Liquids
Solvents, oils, LPG - dielectric constant < 3Chemical mixtures - dielectric constant 3 … 10Water-based - dielectric constant > 10
Bulk solids
Dusts, wood chips - dielectric constant < 1.5Granules, dusts, powders - dielectric constant 1.5 … 3Cereals, our - dielectric constant > 3
44217-EN-130910
Application - Interface measurement
The interface measurement refers to the phase limit between two liquids. The total level is also available as a measured value.
Superimposed gas phase present
Check if there is a superimposed gas phase in the vessel. This
is always the case if the total level never touches the process tting.
Properties
Here you can enter the dielectric constant of the upper mediumAs an alternative you can enter the distance to the interface
61
7 Setup with PACTware
Step 3 Adjustment
Step 4 Linearization
Step 5 Setting - AI FB1
Adjustment for the level measurement
If you have selected level measurement in the previous menu, then you can enter the values for the min. and max. adjustment. The value to be entered refers to the distance from the sealing surface of the process tting (sensor reference plane) to the surface of the product.
Adjustment for the level and interface measurement
If you have selected interface measurement in the previous menu, then you can enter the values for the min. and max. adjustment of level and interface or accept the values of the level measurement. The entered value refers to the distance from the sealing surface of the process tting (sensor reference plane) to the total level or interface.
Linearization is required if the measured value should be outputted in proportion to the volume and not the level. The linearization acts identically on the level and the interface measurement. You can nd further linearization types in the extended adjustment.
If you have a non-linear vessel, you can select here the respective linearization curve.
Linear
Spherical tank
Horizontal cylindrical tank
You must enter the following vessel dimensions with non-linear condi­tions:
Height of the socket h
Vessel height D
In this window you can adjust the output signal. When the function is inactive, you can change the settings via the "Extended adjustment".
Scaling unit
In the menu item "Out Scale Unit (scaling unit)" you dene the scaling unit, for example volume in l for the specied channel (measured value).
Scaling
In the menu items "Out Scale (min.)" and "Out Scale (max.)" you dene the scaling of the measured values for 0 % and 100 %.
Step 6 Sensor optimization
62
These settings allow you to optimize the sensor. With them you can
compare a sounded distance with the indicated value and correct it, if necessary.
Probe immersed in the liquid (covered)
Select whether or not the probe is immersed in the medium.
Measured distance to the medium
If the probe is immersed in the medium, you can enter here the meas­ured distance to the medium.
44217-EN-130910
7 Setup with PACTware
Displayeddistancecorrect?
Is the displayed distance value correct? If you have the possibility, you can enter here the sounded distance to the medium.
False signal suppression
With this function you can carry out the automatic false signal sup­pression. We recommend carrying out false signal suppression in any case.
Step 7 Additional settings
When the setup of the instrument is nished, additional settings can
be made. These are various backups and the locking of the instru­ment against unauthorised or inadvertent adjustment.
Prepareabackupleoftheinstrumentparameteradjustment?
For backup purposes, the current parameter adjustment of the instru­ment is stored in a le. You can use this le later on to restore the instrument parameter adjustment. The complete data set is down­loaded from the device. This procedure can last several minutes.
Createinstrumentdocumentation?
This function is used to print or create a PDF le of the current param­eter adjustment. To read the PDF le, you need a suitable program (for example Acrobat Reader). To print or create the PDF le, all data are downloaded from the device. For this function, the full version of the DTM Collection is required. This procedure can last several minutes.
Storeechocurveofthesetupinthesensor?
Have you completed the initial setup of the instrument? In such case, we recommend storing the current signal conditions in the device for later instrument tests and diagnostics.
LockadjustmentwithPINaftersetting?
The instrument is locked with the current PIN. A parameter adjust­ment is possible only after the PIN is entered again.

7.4 Saving the parameter adjustment data

We recommend documenting or saving the parameter adjustment data via PACTware. That way the data are available for multiple use or service purposes.
44217-EN-130910
63

8 Set up with other systems

8 Set up with other systems

8.1 DD adjustment programs

Device descriptions as Enhanced Device Description (EDD) are available for DD adjustment programs such as, for example, AMS™ and PDM.
The les can be downloaded at www.vega.com/downloads under "Software".
64
44217-EN-130910

9 Diagnostics and service

9 Diagnostics and service

9.1 Maintenance

If the device is used correctly, no maintenance is required in normal operation.

9.2 Diagnosis memory

The instrument has several memories which are available for diagno­sis purposes. The data remain even with voltage interruption.
Measured value memory
Event memory
Up to 100,000 measured values can be stored in the sensor in a ring memory. Each entry contains date/time as well as the respective measured value. Storable values are for example:
Distance
Filling height
Percentage value
Lin. percent
Scaled
Current value
Meas. reliability
Electronics temperature
When the instrument is shipped, the measured value memory is active and stores distance, measurement certainty and electronics temperature every 3 minutes.
In "Extended adjustment" you can select the respective measured values.
The requested values and recording conditions are set via a PC with PACTware/DTM or the control system with EDD. Data are thus read out and also reset.
Up to 500 events are automatically stored with a time stamp in the sensor (non-deletable). Each entry contains date/time, event type, event description and value. Event types are for example:
Modication of a parameter
Switch-on and switch-o times
Status messages (according to NE 107)
Error messages (according to NE 107)
The data are read out via a PC with PACTware/DTM or the control system with EDD.
Echo curve memory
44217-EN-130910
The echo curves are stored with date and time and the corresponding
echo data. The memory is divided into two sections: Echo curve of the setup: This is used as reference echo curve for
the measurement conditions during setup. Changes in the measure­ment conditions during operation or buildup on the sensor can thus be recognized. The echo curve of the setup is stored via:
PC with PACTware/DTM
Control system with EDD
65
9 Diagnostics and service
Display and adjustment module
Further echo curves: Up to 10 echo curves can be stored in a ring buer in this memory section. Further echo curves are stored via:
PC with PACTware/DTM
Control system with EDD
Display and adjustment module

9.3 Status messages

The instrument features self-monitoring and diagnostics according to NE 107 and VDI/VDE 2650. In addition to the status messages in the following tables there are more detailed error messages available under the menu item "Diagnostics" via the display and adjustment module, PACTware/DTM and EDD.
Status messages
The status messages are divided into the following categories:
Failure
Function check
Out of specication
Maintenance requirement
and explained by pictographs:
41 2 3
Fig. 34: Pictographs of the status messages
1 Failure - red
2 Outofspecication-yellow
3 Function check - orange 4 Maintenance - blue
Failure: Due to a malfunction in the instrument, a failure message is outputted.
This status message is always active. It cannot be deactivated by the user.
Function check: The instrument is in operation, the measured value is temporarily invalid (for example during simulation).
This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD.
Outofspecication: The measured value is unstable because the instrument specication is exceeded (e.g. electronics temperature).
This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD.
Maintenance: Due to external inuences, the instrument function is limited. The measurement is aected, but the measured value is
44217-EN-130910
66
9 Diagnostics and service
still valid. Plan in maintenance for the instrument because a failure is expected in the near future (e.g. due to buildup).
This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD.
Failure (failure)
The following table shows the codes and text messages of the status
message "Failure" and provides information on causes as well as corrective measures.
Code
Text mes­sage
F013
no meas-
ured value available
F017
Adjustment
span too
small
F025
Error in the
lineariza­tion table
F036
No op-
erable software
F040
Error in the
electronics
F080 – General software error – Separate operating
F105
Determine measured
value
F113 Communi-
cation error
Cause Rectication PA DevS-
– Sensor does not
detect an echo during
operation
– Antenna system con-
taminated or defective
– Adjustment not within
specication
– Index markers are not
continuously rising, for examle unlogical value
pairs
Failed or interrupted
software update
Hardware defect Exchanging the elec-
The instrument is still
in the start phase, the measured value could not yet be determined
Error in the internal
instrument communi­cation
Check or correct
installation and/or parameter adjustment
Clean or exchange
process component or antenna
Change adjustment
according to the limit values (dierence between min. and max. ≥ 10 mm)
– Check linearization
table
Delete table/Create
new
Repeat software
update
Check electronics
version
Exchanging the elec-
tronics
Send instrument for
repair
tronics
Send instrument for
repair
voltage briey
Wait for the end of the
switch-on phase
Duration depending
on the version and
parameter adjustment up to approximately 3 min.
– Separate operating
voltage briey
Send instrument for
repair
pec
Diagnosis
Bit 0
Bit 1
Bit 2
Bit 3
Bit 4
Bit 5
Bit 6
Bit 7
44217-EN-130910
67
9 Diagnostics and service
Code
Text mes­sage
F125 Unper-
missible electronics
tempera­ture
F260
Error in the
calibration
F261
Error in the congura­tion
F264 Installa-
tion/Setup
error
F265 Meas-
urement
function disturbed
F267 No execut-
able sensor software
Cause Rectication PA DevS-
– Temperature of the
electronics in the non­specied section
Error in the calibra-
tion carried out in the
factory
Error in the EEPROM
Error during setupFalse signal suppres-
sion faulty
Error when carrying
out a reset
Adjustment not within
the vessel height/ measuring range
– Max. measuring range
of the instrument not sucient
Sensor no longer
carries out a measure­ment
Operating voltage
too low
Sensor cannot start Exchanging the elec-
Check ambient tem-
perature
Isolate electronicsUse instrument with
higher temperature range
Exchanging the elec-
tronics
Send instrument for
repair
Repeat setupRepeat reset
Check or correct
installation and/or parameter adjustment
– Use an instrument
with bigger measuring range
– Check operating
voltage
Carry out a resetSeparate operating
voltage briey
tronics
Send instrument for
repair
pec
Diagnosis
Bit 8
Bit 9
Bit 10
Bit 11
Bit 12
-
Function check
Outofspecication
68
The following table shows the error codes and text messages in the
status message "Function check" and provides information on causes as well as corrective measures.
Code
Text mes­sage
C700 Simulation ac-
tive
Cause Rectication
A simulation is active Finish simulation
Wait for the automatic end
after 60 mins.
The following table shows the error codes and text messages in the
status message "Outofspecication" and provides information on causes as well as corrective measures.
44217-EN-130910
9 Diagnostics and service
Maintenance
Code
Text mes­sage
S600 Unpermissi-
ble electronics
temperature
S601 Overlling
S602 Level with-
in the search range, com­pensation echo
S603
Impermissi-
ble operating voltage
Cause Rectication
– Temperature of the pro-
cessing electronics in the non-specied section
– Level echo in the close
range not available
– Compensation echo super-
imposed by medium
– Operating voltage below
specied range
Check ambient temperatureIsolate electronicsUse instrument with higher
temperature range
Reduce level100 % adjustment: Increase
value
Check mounting socketRemove possible interfering
signals in the close range
Use coaxial probe
100 % adjustment: Increase
value
Check electrical connectionif necessary, increase
operating voltage
The following table shows the error codes and text messages in the
status message "Maintenance" and provides information on causes as well as corrective measures.
Code
Text mes­sage
M500
Error in the delivery sta­tus
M501
Error in the non-active
linearization table
M502
Error in the event memory
M503 Reliability too
low
Cause Rectication
– With the reset to delivery
status, the data could not be restored
Hardware error EEPROM Exchanging the electronics
Hardware error EEPROM Exchanging the electronics
Measurement certainty is
too low for a reliable meas­urement
– Process component or
probe contaminated or defective
Repeat resetLoad XML le with sensor
data into the sensor
Send instrument for repair
Send instrument for repair
Check installation and
process conditions
Clean or exchange process
component or probe
44217-EN-130910
69
9 Diagnostics and service
Reaction when malfunc­tions occur
Procedure for fault recti-
cation
Code
Text mes­sage
M504
Error on an device inter-
face
M505
no measured
value avail­able
M506 Installation/
Setup error
M507
Error in the instrument
settings
Cause Rectication
Hardware defect Exchanging the electronics
Sensor does not detect an
echo during operation
– Process component or
probe contaminated or defective
Error during setup Check and correct instal-
Error during setupError when carrying out a
reset
False signal suppression
faulty
Send instrument for repair
Check and correct instal-
lation and/or parameter adjustment
– Clean or exchange process
component or probe
lation and/or parameter adjustment
Check probe length
Carry out reset and repeat
setup

9.4 Rectify faults

The operator of the system is responsible for taking suitable meas-
ures to rectify faults.
The rst measures are:
Evaluation of fault messages, for example via the display and
adjustment module Checking the output signal
Treatment of measurement errors
Further comprehensive diagnostics options oer a PC with the soft­ware PACTware and the suitable DTM. In many cases, the reasons can be determined in this way and faults can be rectied.
Treatment of measure­ment errors
70
The below tables show typical examples for application-relevant
measurement errors. There are two measurement errors:
Constant level
Filling
Emptying
The images in column "Error pattern" show the real level with a bro­ken line and the level displayed by the sensor as a continuous line.
44217-EN-130910
Level
time
time
0
time
0
9 Diagnostics and service
1
2
0
Fig. 35: The broken line 1 shows the real level, the continuous line 2 shows the level displayed by the sensor
Note:
Wherever the sensor displays a constant value, the reason could
also be the fault setting of the current output to "Hold value" In case of a too low level indication, the reason could be a line
resistance that is too high
Measurement error with constant level
Fault description Error pattern Cause Rectication
1. Measured value shows a too low or too high level
2. Measured value jumps towards 100 %
Level
0
Level
Min./max. adjustment not
correct
Wrong linearization curve Adapt linearization curve
Running time error (small
measurement error close to 100 %/serious error close to 0 %)
– Due to the process, the ampli-
tude of the product echo sinks
A false signal suppression was
not carried out
Amplitude or position of a
false signal has changed (e.g. buildup); false signal suppres­sion no longer matches
Adapt min./max. adjustment
Repeat setup
Carry out a false signal sup-
pression
Determine the reason for the
changed false signals, carry out false signal suppression, e.g. with buildup
time
44217-EN-130910
Measurementerrorduringlling
Fault description Error pattern Cause Rectication
3. Measured value re­mains in the area of the bottom during lling
Level
– Echo from the probe end larger
than the product echo, for example, with products with εr < 2.5 oil-based, solvents, etc.
– Check parameter "Medium"
and "Vessel height", adapt if
necessary
71
9 Diagnostics and service
time
0
time
0
time
0
time
0
time
0
Fault description Error pattern Cause Rectication
4. Measured value re­mains momentarily unchanged during ll-
Level
– Turbulence on the product
surface, quick lling
– Check parameters, change if
necessary, e.g. in dosing ves-
sel, reactor ing and then jumps to the correct level
5. Measured value jumps sporadically to
Level
– Changing condensation or
contamination on the probe
Carry out a false signal sup-
pression
100 % during lling
6. Measured value jumps to ≥ 100 % or
0 m distance
Level
– Level echo is no longer
detected in the close range due to false signals in the close range. The sensor goes into overll protection mode. The
– Eliminate false signals in the
close range
Check installation conditionsIf possible, switch o the func-
tion "Overll protection"
max. level (0 m distance) as well as the status message "Overll protection" are output­ted.
Measurement error during emptying
Fault description Error pattern Cause Rectication
7. Measured value re­mains unchanged in the close range during emptying
Level
False echo larger than the level
echo
Level echo too small
Eliminate false signals in the
close range
– Remove contamination on the
probe. After having removed
the source of the false signals,
the false signal suppression
must be deleted.
Carry out a new false signal
suppression
8. Measured value re­mains reproducible in one position during
Level
– Stored false signals in this
position are larger than the level echo
Delete false signal memoryCarry out a new false signal
suppression
emptying
Treatment of measure­ment errors with bulk solids
The below tables show typical examples for application-relevant
measurement errors with bulk solids. There are two measurement
errors:
Constant level
Filling
Emptying
The images in column "Error pattern" show the real level with a bro­ken line and the level displayed by the sensor as a continuous line.
44217-EN-130910
72
Level
time
time
9 Diagnostics and service
1
2
0
1 Real level 2 Level displayed by the sensor
Notes:
Wherever the sensor displays a constant value, the reason could
also be the fault setting of the current output to "Hold value" In case of a too low level indication, the reason could be a line
resistance that is too high
Measurement error with constant level
Fault description Error pattern Cause Rectication
1. Measured value shows a too low or too high level
2. Measured value jumps towards 100 %
Level
0
Level
0
Min./max. adjustment not
correct
Wrong linearization curve Adapt linearization curve
Due to the process, the ampli-
tude of the product echo sinks
A false signal suppression was
not carried out
Amplitude or position of a
false echo has changed (e.g. condensation, buildup); false signal suppression no longer
matches
Adapt min./max. adjustment
Carry out a false signal sup-
pression
Determine the reason for the
changed false echoes, carry
out false signal suppression,
e.g. with condensation
time
44217-EN-130910
73
9 Diagnostics and service
time
0
time
0
time
0
time
Measurementerrorduringlling
Fault description Error pattern Cause Rectication
3. Measured value jumps towards 0 % during lling
4. Measured val­ue uctuates around 10 … 20 %
5. Measured value jumps sporadically to 100 % during lling
Level
Level
Level
– Amplitude of a multiple echo
(vessel top - product surface) is larger than the level echo
– The level echo cannot be distin-
guished from the false echo at a false echo position (jumps to multiple echo)
– Transverse reection from an
extraction funnel, amplitude of the transverse reection larger than the level echo
– Various echoes from an uneven
product surface, e.g. a material
cone
Reections from the product
surface via the vessel wall (deection)
Varying condensation or con-
tamination on the antenna
Check parameter "Application",
especially vessel top, type
of medium, dished end, high
dielectric constant, adapt if
necessary
In case of interferences due to
installations in the close range:
Change polarisation direction
Chose a more suitable installa-
tion position
– Direct sensor to the opposite
funnel wall, avoid crossing with
the lling stream
– Check parameter "Type of
medium" and adapt, if neces-
sary
Optimize installation position
and sensor orientation
Select a more suitable installa-
tion position, optimize sensor
orientation, e.g. with a swivel-
ling holder
– Carry out a false signal sup-
pression or increase false
signal suppression with con-
densation/contamination in the
close range by editing
– With bulk solids use radar sen-
sor with purging air connection
or exible antenna cover
Measurement error during emptying
Fault description Error pattern Cause Rectication
6. Measured value re­mains unchanged in
Level
the close range during emptying
0
74
False echo larger than the level
echo
Level echo too small
Remove false echoes in the
close range. Check: Antenna
must protrude out of the socket
Remove contamination on the
antenna
In case of interferences due to
installations in the close range:
Change polarisation direction
– After removing the false ech-
oes, the false signal suppres-
sion must be deleted. Carry out
a new false signal suppression
44217-EN-130910
9 Diagnostics and service
time
0
time
Fault description Error pattern Cause Rectication
7. Measured value jumps sporadically to­wards 100 % during emptying
8. Measured val­ue uctuates around 10 … 20 %
Level
Level
0
Varying condensation or con-
tamination on the antenna
Various echoes from an uneven
product surface, e.g. an extrac­tion funnel
– Reections from the product
surface via the vessel wall (deection)
– Carry out false signal suppres-
sion or increase false signal
suppression in the close range
by editing
– With bulk solids use radar sen-
sor with purging air connection
or exible antenna cover
– Check parameter "Type of
medium" and adapt, if neces-
sary
Optimize installation position
and sensor orientation
Reaction after fault recti-
cation
24 hour service hotline
Depending on the reason for the fault and the measures taken, the steps described in chapter "Setup" must be carried out again or must be checked for plausibility and completeness.
Should these measures not be successful, please call in urgent cases the VEGA service hotline under the phone no. +49 1805 858550.
The hotline is also available outside normal working hours, seven days a week around the clock.
Since we oer this service worldwide, the support is provided in English. The service itself is free of charge, the only costs involved are the normal call charges.

9.5 Exchanging the electronics module

If the electronics module is defective, it can be replaced by the user. In Ex applications only one instrument and one electronics module
with respective Ex approval may be used. If there is no electronics module available on site, the electronics
module can be ordered through the agency serving you. The electron­ics modules are adapted to the respective sensor and dier in signal output or voltage supply.
The new electronics module must be loaded with the default settings of the sensor. These are the options:
In the factory
Or on site by the user
In both cases, the serial number of the sensor is needed. The serial numbers are stated on the type label of the instrument, on the inside of the housing as well as on the delivery note.
When loading on site, rst of all the order data must be downloaded from the Internet (see operating instructions manual "Electronics module").
44217-EN-130910
Exchanging the cable/rod

9.6 Exchanging the cable/rod

If necessary, the cable or rod (measuring part) of the probe can be exchanged.
75
9 Diagnostics and service
Loosen the rod or cable with a fork wrench, wrench size 7 (rod ø 8, cable ø 2 and 4) or wrench size 10 (rod ø 12).
1. Loosen the rod or cable with a fork wrench applied to the at surface, provide counterforce with another fork wrench.
2. Unscrew the loosened rod or cable manually.
3. Insert the new measuring rod carefully by hand with a twisting motion into the opening of the process tting.
4. Continue screwing in the rod manually into the opening of the pro­cess tting.
5. Exert counterforce with the second fork spanner and tighten the rod or cable on the at surfaces with the following torque.
Rod ø 8, cable ø 2 and 4: 6 Nm (4.43 lbf ft) Rod ø 12: 10 Nm (7.37 lbf ft)
Shorten cable/rod
76
Fig. 53: Exchange cable or rod
Information:
Please keep the stated torque so that the max. tensile strength remains.
6. Enter new probe length and if necessary the new probe type and then carry out a fresh adjustment (see "Setup procedure, Carry- ing out min. adjustment - Carrying out max. adjustment").
The rod or cable of the probe can be shortened individually.
1. Mark the requested length with mounted measuring rod.
44217-EN-130910
9 Diagnostics and service
2. Cable: Loosen the pins on the gravity weight (hexagon 3)
3. Cable: remove the pins
4. Cable: Pull the cable out of the gravity weight
5. Shorten the cable/rod with a cut-o wheel or metal saw at the marking. Take note of the specications in the following illustration when shortening the cable.
6. Cable with gravity weight: Shift the cable according to the drawing into the gravity weight
7. Cable with gravity weight: Fasten cable with the pins, torque 7 Nm (5.16 lbf ft)
Cable with centering weight: Fasten cable with the pins, torque 7 Nm (5.16 lbf ft) and x the clamping part on the centering weight.
8. Enter new probe length and then carry out a fresh adjustment (see "Setup procedure, Carrying out min. adjustment - Carrying out max. adjustment").
1
(1.77")
45 mm
(3.94")
100 mm
(0.98")
25 mm
(1.97")
50 mm
A
B
1
4
C
2 2 3
44217-EN-130910
Fig. 54: Shortening the cable probe
A Gravity weight - cable ø 4 mm B Gravity weight - cable ø 2 mm C Centering weight - cable ø 2 mm 1 Pins 2 Thread M8 for eye-bolt 3 Fixing screw - centering weight

9.7 Software update

The following components are required to update the sensor soft­ware:
Sensor
Voltage supply
Interface adapter VEGACONNECT
PC with PACTware
Current sensor software as le
You can nd the actual sensor software as well as detailed infor­mation of the procedure under "www.vega.com/downloads" and "Software".
77
9 Diagnostics and service
You can nd information about the installation in the download le.
Caution:
Instruments with approvals can be bound to certain software ver­sions. Therefore make sure that the approval remains eective with a software update.
You can nd detailed information on www.vega.com/downloads and "Approvals".

9.8 How to proceed in case of repair

You can nd a repair form as well as detailed information on how to
proceed under www.vega.com/downloads and "Formsandcerti- cates".
By doing this you help us carry out the repair quickly and without hav­ing to call back for needed information.
If a repair is necessary, please proceed as follows:
Print and ll out one form per instrument
Clean the instrument and pack it damage-proof
Attach the completed form and, if need be, also a safety data
sheet outside on the packaging Please contact for the return shipment the agency serving you. You
can nd the agency on our home page www.vega.com.
78
44217-EN-130910

10 Dismounting

10 Dismounting

10.1 Dismounting steps

Warning:
Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel or pipeline, high temperatures, cor­rosive or toxic products etc.
Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order.

10.2 Disposal

The instrument consists of materials which can be recycled by spe­cialised recycling companies. We use recyclable materials and have designed the parts to be easily separable.
Correct disposal avoids negative eects on humans and the environ­ment and ensures recycling of useful raw materials.
Materials: see chapter "Technical data" If you have no way to dispose of the old instrument properly, please
contact us concerning return and disposal.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/EG and the respective national laws. Pass the instrument directly on to a spe­cialised recycling company and do not use the municipal collecting points. These may be used only for privately used products according to the WEEE directive.
44217-EN-130910
79

11 Supplement

11 Supplement

11.1 Technical data

General data
316L corresponds to 1.4404 or 1.4435 Materials, wetted parts
Ʋ Process tting (version up to 6 bar) 316L and PPS GF 40 Ʋ Process tting (version up to 40 bar) 316L and Aluminium oxide-ceramic 99.7 % (Al2O3),
Ʋ Process seal on the instrument side
(cable/rod leadthrough)
Ʋ Process seal On site (instruments with thread: Klingersil C-4400 is
Ʋ Rod: ø 8 mm (0.315 in) 316L or Hastelloy C22 (2.4602) Ʋ Rod: ø 12 mm (0.472 in) 316L or Hastelloy C22 (2.4602) Ʋ Cable: ø 2 mm (0.079 in) 316 (1.4401) Ʋ Cable: ø 4 mm (0.157 in) 316 (1.4401) Ʋ Inner conductor (up to the cable) 316L Ʋ Gravity weight (optionally available) 316L Ʋ Centering weight (optionally available) 316L
Materials, non-wetted parts
Ʋ Plastic housing plastic PBT (Polyester) Ʋ Aluminium die-casting housing Aluminium die-casting AlSi10Mg, powder-coated - basis:
Ʋ Stainless steel housing - precision
casting
Ʋ Stainless steel housing, electropol-
ished
Ʋ Second line of defence (optional) Borosilicate glass GPC 540 Ʋ Seal between housing and housing
cover
Ʋ Inspection window in housing cover
(optional)
Ʋ Ground terminal 316L
Ohmic contact Between ground terminal, process tting and probe
Process ttings
Ʋ Pipe thread, cylindrical (ISO 228 T1) G¾, G1, G1½ according to DIN 3852-A Ʋ American pipe thread, conical
(ASME B1.20.1)
Ʋ Flanges e.g. DIN from DN 25, ANSI from 1"
Hastelloy C22 (2.4602) and Aluminium oxide-ceramic
99.7 % (Al2O3) FKM (SHS FPM 70C3 GLT), FFKM (Kalrez 6375),
EPDM (A+P 75.5/KW75F), silicone FEP coated (A+P FEP-O-SEAL)
attached)
Polyester 316L
316L
NBR (stainless steel housing, precision casting), silicone (aluminium/plastic housing; stainless steel housing, electropolished)
Polycarbonate (with Ex d version: glass)
¾ NPT, 1 NPT, 1½ NPT
44217-EN-130910
80
Weight
Ʋ Instrument weight (depending on
approx. 0.8 … 8 kg (0.176 … 17.64 lbs)
process tting)
Ʋ Rod: ø 8 mm (0.315 in) approx. 400 g/m (4.3 oz/ft) Ʋ Rod: ø 12 mm (0.472 in) approx. 900 g/m (9.68 oz/ft) Ʋ Cable: ø 2 mm (0.079 in) approx. 16 g/m (0.17 oz/ft) Ʋ Cable: ø 4 mm (0.157 in) approx. 60 g/m (0.65 oz/ft) Ʋ Gravity weight for cable ø 2 mm
100 g (3.22 oz)
(0.079 in)
Ʋ Gravity weight for cable ø 4 mm
200 g (6.43 oz)
(0.157 in)
Ʋ Centering weight ø 40 mm (1.575 in) 180 g (5.79 oz) Ʋ Centering weight ø 45 mm (1.772 in) 250 g (8.04 oz) Ʋ Centering weight ø 75 mm (2.953 in) 825 g (26.52 oz) Ʋ Centering weight (ø 95 mm (3.74 in) 1050 g (33.76 oz)
Probe length L (from seal surface)
Ʋ Rod: ø 8 mm (0.315 in) up to 6 m (19.69 ft) Ʋ Rod: ø 12 mm (0.472 in) up to 6 m (19.69 ft) Ʋ Trimming accuracy - rod ±1 mm Ʋ Cable: ø 2 mm (0.079 in) up to 75 m (246.1 ft) Ʋ Cable: ø 4 mm (0.157 in) up to 75 m (246 ft) Ʋ Trimming accuracy - cable ±0.05 %
Lateral load
Ʋ Rod: ø 8 mm (0.315 in) 10 Nm (7.38 lbf ft) Ʋ Rod: ø 12 mm (0.472 in) 30 Nm (22.13 lbf ft)
Max. tensile load
Ʋ Cable: ø 2 mm (0.079 in) 1.5 KN (337 lbf) Ʋ Cable: ø 4 mm (0.157 in) 2.5 KN (562 lbf)
Thread in gravity weight, e.g. for eye-bolt
M 8
(cable version) Torque for exchangeable cable or rod probe (in the process tting)
Ʋ Cable: ø 2 mm (0.079 in) 6 Nm (4.43 lbf ft) Ʋ Cable: ø 4 mm (0.157 in) 6 Nm (4.43 lbf ft) Ʋ Rod: ø 8 mm (0.315 in) 6 Nm (4.43 lbf ft) Ʋ Rod: ø 12 mm (0.472 in) 10 Nm (7.38 lbf ft)
Torque for NPT cable glands and Conduit tubes
Ʋ Plastic housing max. 10 Nm (7.376 lbf ft) Ʋ Aluminium/Stainless steel housing max. 50 Nm (36.88 lbf ft)
11 Supplement
Input variable
Measured variable Level of liquids Min. dielectric constant of the medium
44217-EN-130910
81
11 Supplement
Ʋ Cable probes εr ≥ 1.6 Ʋ Rod probes εr ≥ 1.6
1 1
4
2
2
4
3
3
5
5
Fig. 55: Measuring ranges - VEGAFLEX 81
1 Reference plane 2 Probe length L 3 Measuring range (default setting refers to the measuring range in water) 4 Upper dead band (see diagrams under Accuracy - grey section) 5 Lower dead band (see diagrams under Accuracy - grey section)
Output variable
Output signal digital output signal, format according to IEEE-754 Sensor address 126 (default setting) Damping (63 % of the input variable) 0 … 999 s, adjustable Probus PA prole 3.02
44217-EN-130910
82
11 Supplement
Number of FBs with AI (function blocks
3
with analogue input) Default values
Ʋ 1. FB PV Ʋ 2. FB SV 1 Ʋ 3. FB SV 2
Current value
Ʋ Non-Ex and Ex ia instrument 10 mA, ±0.5 mA Ʋ Ex-d instruments 16 mA, ±0.5 mA
Resolution, digital < 1 mm (0.039 in)
Accuracy (according to DIN EN 60770-1)
Process reference conditions according to DIN EN 61298-1
Ʋ Temperature +18 … +30 °C (+64 … +86 °F) Ʋ Relative humidity 45 … 75 % Ʋ Air pressure +860 … +1060 mbar/+86 … +106 kPa
(+12.5 … +15.4 psig)
Installation reference conditions
Ʋ Min. distance to installations > 500 mm (19.69 in) Ʋ Vessel metallic, ø 1 m (3.281 ft), centric installation, process
tting ush with the vessel ceiling
Ʋ Medium Water/Oil (dielectric constant ~2.0)
1)
Ʋ Installation Probe end does not touch the vessel bottom Sensor parameter adjustment No gating out of false signals carried out Typical deviation - Interface measure-
± 5 mm (0.197 in)
ment
Typical deviation - Total level interface
See following diagrams
measurement
Typical deviation - Level measurement
2)3)
See following diagrams
1)
With interface measurement = 2.0
2)
Depending on the installation conditions, there can be deviations which can be rectied with an adaptation of
the adjustment or a change of the measured value oset in the DTM service mode
3)
The dead bands can be optimizes by a false signal suppression.
44217-EN-130910
83
11 Supplement
15mm
")
(0.591
2mm
")
(0.079
0
-2mm
(-0.079
")
1
-10mm
(-0.394
")
0,08 m (3.15")
0,3 m
(11.811
")
Fig. 56: Deviation VEGAFLEX 81 in rod version in water
1 Dead band - no measurement possible in this area L Probe length
15mm
")
(0.591
2mm
(0.079
")
0
-2mm
(-0.079
")
-10mm
(-0.591
1
")
0,15 m
(5.906
0,3 m
(11.811
")
")
L
0,02 m
(0.787
")
L
1
0,07 m
(2.756
")
0,05 m (1.97
")
Fig. 57: Deviation VEGAFLEX 81 in rod version in oil
1 Dead band - no measurement possible in this area L Probe length
84
44217-EN-130910
15mm
(0.591
")
2mm
")
(0.079
0
-2mm
")
(-0.079
10mm
(0.394
1
")
0,08 m (3.15
0,3 m
")
(11.811
")
Fig. 58: Deviation VEGAFLEX 81 in cable version in water
1 Dead band - no measurement possible in this area L Probe length
15mm
")
(0.591
2mm
")
(0.079
0
-2mm
")
(-0.079
11 Supplement
L
0,1 m
(3.94
")
L
-15mm
")
(-0.591
0,15 m
0,3 m
")
")
(5.906
(11.811
Fig. 59: Deviation VEGAFLEX 81 in cable version ø 2 mm (0.079 in), in oil
1 Dead band - no measurement possible in this area L Probe length
44217-EN-130910
1 1
0,05 m
(1.969
")
0,08 m (3.15
")
85
11 Supplement
20mm
")
(0.787
15mm
(0.591
")
2mm
(0.079
")
0
-2mm
(-0.079
")
L
-15mm )
(-0.591
"
0,3 m
0,15 m
(5.906
")
(11.811
")
0,05 m
(1.969
")
0,15 m
(5.906
")
Fig. 60: Deviation VEGAFLEX 81 in cable version ø 4 mm (0.157 in), in oil
1 Dead band - no measurement possible in this area L Probe length
Repeatability ≤ ±1 mm
Variablesinuencingmeasurementaccuracy
Temperature drift - Digital output ±3 mm/10 K relating to the max. measuring range or
max. 10 mm (0.394 in)
1 1
Additional deviation through electromag-
< ±10 mm (< ±0.394 in)
netic interference acc. to EN 61326
Inuenceofthesuperimposedgasandpressuretotheaccuracy
The propagation speed of the radar impulses in gas or vapour above the medium is reduced by high pressure. This eect depends on the superimposed gas or vapour and is especially large at low temperatures.
The following table shows the resulting deviation for some typical gases and vapours. The specied values refer to the distance. Positive values mean that the measured distance is too large, negative values that the measured distance is too small.
Gas phase Temperature Pressure
1 bar (14.5 psig) 10 bar (145 psig) 50 bar (725 psig)
Air 20 °C/68 °F 0.00 % 0.22 % 1.2 %
200 °C/392 °F -0.01 % 0.13 % 0.74 %
400 °C/752 °F -0.02 % 0.08 % 0.52 %
Hydrogen 20 °C/68 °F -0.01 % 0.10 % 0.61 %
200 °C/392 °F -0.02 % 0.05 % 0.37 %
400 °C/752 °F -0.02 % 0.03 % 0.25 %
44217-EN-130910
86
11 Supplement
Gas phase Temperature Pressure
1 bar (14.5 psig) 10 bar (145 psig) 50 bar (725 psig)
Steam (saturated steam)
100 °C/212 °F 0.26 % - -
180 °C/356 °F 0.17 % 2.1 % -
264 °C/507 °F 0.12 % 1.44 % 9.2 %
366 °C/691 °F 0.07 % 1.01 % 5.7 %
Characteristics and performance data
Measuring cycle time < 500 ms Step response time
4)
≤ 3 s
Max. lling/emptying speed 1 m/min
Ambient conditions
Ambient, storage and transport tempera-
-40 … +80 °C (-40 … +176 °F)
ture
Process conditions
For the process conditions, please also note the specications on the type label. The lowest value always applies.
The measurement error from the process conditions is in the specied pressure and temperature range of below 1 %.
Process pressure
Ʋ Process tting with PPS GF 40 -1 … 6 bar/-100 … 600 kPa (-14.5 … 87 psi), depending
on the process tting
Ʋ Process tting with Al2O
3
Vessel pressure relating to the ange nominal stage
-1 … +40 bar/-100 … +4000 kPa (-14.5 … +580 psig), depending on the process tting
see supplementary instructions manual "Flanges ac- cording to DIN-EN-ASME-JIS"
Process temperature (thread or ange temperature)
Ʋ PPS GF 40 -40 … +80 °C (-40 … +176 °F) Ʋ FKM (SHS FPM 70C3 GLT) -40 … +150 °C (-40 … +302 °F) Ʋ EPDM (A+P 75.5/KW75F) -40 … +150 °C (-40 … +302 °F) Ʋ Silicone FEP coated (A+P FEP-O-
-40 … +150 °C (-40 … +302 °F)
SEAL)
Ʋ FFKM (Kalrez 6375) -20 … +150 °C (-4 … +302 °F) Ʋ FFKM (Kalrez 6375) - with tempera-
-20 … +200 °C (-4 … +392 °F)
ture adapter
4)
Time span after a sudden measuring distance change by max. 0.5 m in liquid applications, max 2 m with bulk
solids applications, until the output signal has taken for the rst time 90 % of the nal value (IEC 61298-2).
44217-EN-130910
87
11 Supplement
1
80°C
(176°F)
65°C
(149°F)
3 4
-40°C
(-104°F)
0°C
(32°F)
-40°C
(-104°F)
80°C
(176°F)
130°C
(266°F)
150°C
(302°F)
2
Fig. 61: Ambient temperature - process temperature, standard version
1 Ambient temperature 2 Process temperature (depending on the seal material) 3 Maximum permissible temperature - standard 4 Limited temperature range - plastic housing and stainless steel housing, electropolished
1
80°C
(176°F)
3 4
55°C
(131°F)
-20°C
(-68°F)
0°C
(32°F)
-40°C
(-104°F)
150°C
(302°F)
200°C
(392°F)
2
Fig. 62: Ambient temperature - process temperature, version with temperature adapter
1 Ambient temperature 2 Process temperature (depending on the seal material) 3 Maximum permissible temperature - standard 4 Limited temperature range - plastic housing and stainless steel housing, electropolished
Vibration resistance
Ʋ Instrument housing 4 g at 5 … 200 Hz according to EN 60068-2-6 (vibration
with resonance)
Ʋ Rod probe 1 g with 5 … 200 Hz according EN 60068-2-6 (vibration
at resonance) with rod length 50 cm (19.69 in)
Shock resistance
Ʋ Instrument housing 100 g, 6 ms according to EN 60068-2-27 (mechanical
shock)
Ʋ Rod probe 25 g, 6 ms according to EN 60068-2-27 (mechanical
shock) with rod length 50 cm (19.69 in)
44217-EN-130910
88
11 Supplement
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Options of the cable entry
Ʋ Cable gland M20 x 1.5 (cable: ø 5 … 9 mm) Ʋ Cable entry ½ NPT Ʋ Blind plug M20 x 1.5; ½ NPT Ʋ Closing cap M20 x 1.5; ½ NPT
Plug options
Ʋ Signal circuit Plug M12 x 1, according to ISO 4400, Harting HAN,
7/8" FF
Ʋ Indication circuit M12 x 1 plug
Wire cross-section (spring-loaded terminals)
Ʋ Massive wire, cord 0.2 … 2.5 mm² (AWG 24 … 14) Ʋ Stranded wire with end sleeve 0.2 … 1.5 mm² (AWG 24 … 16)
Electromechanical data - version IP 66/IP 68 (1 bar)
Options of the cable entry
Ʋ Cable gland with integrated connec-
M20 x 1.5 (cable: ø 5 … 9 mm)
tion cable
Ʋ Cable entry ½ NPT Ʋ Blind plug M20 x 1.5; ½ NPT
Connection cable
Ʋ Wire cross-section 0.5 mm² (AWG 20) Ʋ Wire resistance < 0.036 Ω/m Ʋ Tensile strength < 1200 N (270 lbf) Ʋ Standard length 5 m (16.4 ft) Ʋ Max. length 180 m (590.6 ft) Ʋ Min. bending radius 25 mm (0.984 in) with 25 °C (77 °F) Ʋ Diameter approx. 8 mm (0.315 in) Ʋ Colour - Non-Ex version Black Ʋ Colour - Ex-version Blue
Display and adjustment module
Display element Display with backlight Measured value indication
Ʋ Number of digits 5 Ʋ Size of digits W x H = 7 x 13 mm
Adjustment elements 4 keys Protection rating
Ʋ unassembled IP 20 Ʋ mounted into the housing without
IP 40
cover
Materials
Ʋ Housing ABS
44217-EN-130910
89
11 Supplement
Ʋ Inspection window Polyester foil
Integrated clock
Date format Day.Month.Year Time format 12 h/24 h Time zone Ex factory CET Rate deviation max. 10.5 min/year
Measurement electronics temerature
Resolution 1 °C (1.8 °F) Accuracy ±1 °C (1.8 °F)
Voltage supply
Operating voltage
Ʋ Non-Ex instrument 9 … 32 V DC Ʋ EEx-ia instrument - Power supply
9 … 17.5 V DC
FISCO model
Ʋ EEx-ia instrument - Power supply
9 … 24 V DC
ENTITY model
Ʋ EEx-d instrument 14 … 32 V DC
Number of sensors per DP/PA segment coupler, max.
Ʋ Non-Ex 32 Ʋ Ex 10
Electrical protective measures
Protection, depending on housing version
Ʋ Plastic housing IP 66/IP 67 Ʋ Aluminium housing, stainless steel
IP 66/IP 68 (0.2 bar) housing - investment casting, stain­less steel housing - electro-polished
Ʋ Aluminium and stainless housing, in-
IP 66/IP 68 (1 bar) vestment casting (optionally available)
Overvoltage category III Protection class III
5)
Approvals
Instruments with approvals can have dierent technical data depending on the version. For that reason the associated approval documents of these instruments must be carefully noted.
They are part of the delivery or can be downloaded under www.vega.com and "VEGA Tools" as well as under "Downloads" and "Approvals".
11.2 CommunicationProbusPA
Instrumentmasterle
The instrument master le (GSD) contains the characteristic data of the Probus PA instrument.
5)
The prerequisites for maintaining the protection rating are a suitable cable as well as correct mounting.
90
44217-EN-130910
11 Supplement
These data are, e.g. the permissible transmission rates as well as information on diagnostics values and the format of the measured value outputted by the PA instrument.
A bitmap le is also provided for the Probus network planning tool. This le is installed automati­cally when the GSD le is integrated. The bitmap le is used for symbolic indication of the PA instru­ment in the conguration tool.
ID number
Each Probus instrument gets an unambiguous ident number (ID number) from the Probus user organisation (PNO). This ID number is also included in the name of the GSD le. Optionally to this manufacturer-specic GSD le, PNO provides also a general so-called prole-specic GSD le. If the general GSD le is used, the sensor must be set to the prole-specic ident number via the DTM software. By default, the sensor operates with the manufacturer-specic ID number. When using the instruments on a segment coupler SK-2 or SK-3, no special GSD les are required.
VE010BF5 The following table shows the instrument ID and the GSD names for the VEGAFLEX sensor series.
Device name Instrument ID GSDlename
VEGA Instrument class in
prole3.02
VEGAFLEX Serie 80 0xBF5 0x9702 VE010BF5.GSD PA139702.GSD
VEGA Prole-specic
Cyclicaldatatrac
The master class 1 (e.g. PLC) cyclically reads out measured values from the sensor during opera­tion. The below block diagram below shows which data can be accessed by the PLC.
44217-EN-130910
91
11 Supplement
Fig. 63: VEGAFLEX 81: Block diagram with AI FB 1 … AI FB 3 OUT values
TB Transducer Block FB 1 … FB 3 Function Block
Module of the PA sensors
For the cyclic data trac, VEGAFLEX 81 provides the following modules:
AI FB1 (OUT)
Out value of the AI FB1 after scaling
AI FB2 (OUT)
Out value of the AI FB2 after scaling
AI FB3 (OUT)
Out value of the AI FB3 after scaling
Free Place
This module must be used if a value in the data telegram of the cyclical data trac should not
be used (e.g. replacement of temperature and Additional Cyclic Value)
A maximum of three modules can be active. By means of the conguration software of the Probus master you can determine the conguration of the cyclical data telegram with these modules. The procedure depends on the respective conguration software.
Note:
The modules are available in two versions:
Short for Probus master supporting only one "Identier Format" byte, e.g. Allen
Bradley
92
44217-EN-130910
11 Supplement
Sign
-23
Long for Probus master only supporting the "Identier Format" byte, e.g. Siemens
S7-300/400
Examplesoftelegramconguration
In the following you will see how the modules can be combined and how the appendant data tel­egram is structured.
Example 1
AI FB1 (OUT)
AI FB2 (OUT)
AI FB3 (OUT)
Byte­No.
Format IEEE-754-Floating point
Value AI FB1 (OUT) AI FB1 AI FB2 (OUT) AI FB2 AI FB3 (OUT) AI FB3
Example 2
Byte-No. 1 2 3 4 5
Format IEEE-754-Floating point value Status
Value AI FB1 (OUT) AI FB1
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
value
Status IEEE-754-Floating point
value
Status IEEE-754-Floating point
value
AI FB1 (OUT) Free Place Free Place
Note:
Bytes 6-15 are not used in this example.
Status
44217-EN-130910
Data format of the output signal
Byte0
Byte4
Byte3
Status Value (IEEE-754)
Fig. 64: Data format of the output signal
The status byte corresponds to prole 3.02 "Probus PA Prole for Process Control Devices" coded. The status "Measured value OK" is coded as 80 (hex) (Bit7 = 1, Bit6 … 0 = 0).
The measured value is transferred as a 32 bit oating point number in the IEEE-754 format.
Byte n
Bit
Bit
6
7
2
Bit
Bit
4
3
5
6543
2
2
2
Exponent
VZ
2
Bit
7
VZ
Bit
Value = (-1)
Fig. 65: Data format of the measured value
Byte1
Byte2
Bit
Bit
Bit
0
2
1
210-1
2
2
2
(Exponent - 127)
Byte n+1
Bit
7 2
Bit
Bit
Bit
4
6
5
-2-3-4-5-6 -7
2
2
2
Significant
(1 + Significant)
Bit
Byte n+2
Bit
Bit
Bit
Bit
7
3 2
0
2
1
2
2
2
2
Bit
-9
Bit
5
-10 -11
2
Bit
4
3
2
2
Bit
6
-8
2
Bit
Bit
Bit
Bit
0
2
1
-12
2
7
-13
-14 -15
2
2
2
Significant
Byte n+3 Bit
Bit
6
-16
2
Bit
Bit
-17
Bit
Bit
4
5
-19
-18
2
2
Bit
3
2
1
0
-20
-22
-21
2
2
2
2
Significant
93
11 Supplement
Coding of the status byte associated with the PA output value
You can nd further information for the coding of the status byte in the Device Description 3.02 on www.probus.com.
Status code Description according to
Probusstandard
0 x 00 bad - non-specic Flash-Update active
0 x 04 bad - conguration error – Adjustment error
0 x 0C bad - sensor failure – Hardware error
0 x 10 bad - sensor failure – Measured value generation error
0 x 1f bad - out of service con-
stant
0 x 44 uncertain - last unstable
value
0 x 48 uncertain substitute set – Switch on simulation
0 x 4c uncertain - initial value Failsafe replacement value (Failsafe-Mode = "Last valid
0 x 51 uncertain - sensor; con-
version not accurate - low limited
0 x 52 uncertain - sensor; con-
version not accurate - high limited
0 x 80 good (non-cascade) - OK OK
0 x 84 good (non-cascade) - ac-
tive block alarm
0 x 89 good (non-cascade) - ac-
tive advisory alarm - low limited
0 x 8a good (non-cascade) - ac-
tive advisory alarm - high limited
0 x 8d good (non-cascade) - ac-
tive critical alarm - low limited
0 x 8e good (non-cascade) - ac-
tive critical alarm - high limited
Possible cause
– Conguration error with PV-Scale (PV-Span too
small)
Unit irregularityError in the linearization table
Converter errorLeakage pulse errorTrigger error
– Temperature measurement error
"Out of Service" mode switched on
Failsafe replacement value (Failsafe-Mode = "Last val­ue" and already valid measured value since switching on)
– Failsafe replacement value (Failsafe-Mode = "Fsafe
value")
value" and no valid measured value since switching on)
Sensor value < lower limit
Sensor value > upper limit
Static revision (FB, TB) changed (10 sec. active, after the parameter of the static category has been written)
Lo-Alarm
Hi-Alarm
Lo-Lo-Alarm
Hi-Hi-Alarm
44217-EN-130910
94
11 Supplement

11.3 Dimensions

The following dimensional drawings represent only an extract of all possible versions. Detailed dimensional drawings can be downloaded at www.vega.com/downloads under "Drawings".
Plastic housing
~ 69 mm
(2.72")
ø 79 mm
(3.11")
~ 84 mm
(3.31")
M16x1,5
ø 79 mm
(3.11")
M20x1,5/ ½ NPT
(4.41")
112 mm
1
M20x1,5/ ½ NPT
2
(4.41")
112 mm
Fig. 66: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated display and adjustment module the housing is 9 mm/0.35 in higher
1 Single chamber version 2 Double chamber version
Aluminium housing
~ 116 mm
(4.57")
ø 86 mm
(3.39")
(4.57")
116 mm
M20x1,5/ ½ NPT
M20x1,5
Fig. 67: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated display and adjustment module the housing is 9 mm/0.35 in higher
1 Single chamber version 2 Double chamber version
~ 87 mm
M16x1,5
M20x1,5/ ½ NPT
(3.43")
ø 86 mm
(3.39")
21
(4.72")
120 mm
44217-EN-130910
95
11 Supplement
Aluminium housing in protection rating IP 66/IP 68 (1 bar)
~ 150 mm
(5.91")
ø 86 mm
(3.39")
(4.57")
116 mm
M20x1,5
M20x1,5
1
Fig.68:HousingversionsinprotectionIP␣66/IP␣68(1␣bar)-withintegrateddisplayandadjustmentmodulethe housingis9␣mm/0.35␣inhigher
1 Single chamber version 2 Double chamber version
~ 105 mm
(4.13")
M20x1,5/ ½ NPT
ø 86 mm
(3.39")
2
M16x1,5
(4.72")
120 mm
Stainless steel housing
~ 59 mm
(2.32")
ø 80 mm
(3.15")
~ 69 mm
(2.72")
ø 79 mm
(3.11")
~ 87 mm
M16x1,5
(3.43")
ø 86 mm
(3.39")
M20x1,5/ ½ NPT
(4.41")
112 mm
M20x1,5/ ½ NPT
(4.61")
117 mm
M20x1,5/ ½ NPT
321
(4.72")
120 mm
Fig. 69: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated display and adjustment module the housing is 9 mm/0.35 in higher
1 Single chamber version, electropolished 2 Single chamber version, precision casting 3 Double chamber version, precision casting
96
44217-EN-130910
Stainless steel housing in protection rating IP 66/IP 68, 1 bar
~ 93 mm
(3.66")
ø 80 mm
(3.15")
~ 103 mm
(4.06")
ø 79 mm
(3.11")
~ 105 mm
(4.13")
11 Supplement
ø 86 mm
(3.39")
M16x1,5
(4.61")
117 mm
M20x1,5/ ½ NPT
120 mm
3
M20x1,5/ ½ NPT
(4.41")
112 mm
M20x1,5
21
Fig.70:HousingversionsinprotectionIP␣66/IP␣68(1␣bar)-withintegrateddisplayandadjustmentmodulethe housingis9␣mm/0.35␣inhigher
1 Single chamber version, electropolished 2 Single chamber version, precision casting 3 Double chamber version, precision casting
(4.72")
44217-EN-130910
97
11 Supplement
VEGAFLEX 81, cable version with gravity weight
SW 36 (1. SW 41 (1. SW 46 (1.
ø 2 mm (0.08")
42")
(G¾, ¾ NPT)
61")
(G1, 1 NPT)
81")
(G1½, 1½ NPT)
G¾, ¾ NPT, G1, 1 NPT, G1½, 1½ NPT
ø 16 mm
(0.63")
ø 20 mm
(0.79")
(1.97")
50 mm
ø 4 mm
(0.16")
ø 16 mm
(0.63")
ø 20 mm
(0.79")
46 mm
(1.81")
(0.87")
22 mm
100 mm (3.94")
L
ø 4 mm
(0.16")
1 2 3
ø 16 mm
(0.63")
ø 20 mm
(0.79")
(3.78")
96 mm
(0.87")
22 mm
100 mm (3.94")
L
44217-EN-130910
Fig. 71: VEGAFLEX 81, threaded version with gravity weight (all gravity weights with thread M8 for eye-bolt)
L Sensor length, see chapter "Technical data"
98
1 Cable version ø 2 mm (0.079 in) with gravity weight 2 Cable version ø 4 mm (0.157 in) with gravity weight 3 Cable version with temperature adapter
VEGAFLEX 81, cable version with centering weight
SW 36 (1.
42")
(G¾, ¾ NPT)
61")
SW 41 (1. SW 46 (1.
(G1, 1 NPT)
(G1½, 1½ NPT)
81")
G¾, ¾ NPT, G1, 1 NPT, G1½, 1½ NPT
11 Supplement
(1.81")
22 mm (0.87") 46 mm
(1.18")
30 mm
L
ø 2 mm
(0.08")
ø 4 mm (0.16")
(1.18")
30 mm
X
X
1 2
Fig. 72: VEGAFLEX 81, threaded version
L Sensor length, see chapter "Technical data" x ø 40 mm (1.57 in)
ø 45 mm (1.77 in) ø 75 mm (2.95 in)
ø 95 mm (3.74 in) 1 Cable version ø 2 mm (0.079 in) with centering weight (see supplementary instructions "Centering") 2 Cable version ø 4 mm (0.157 in) with centering weight (see supplementary instructions "Centering")
44217-EN-130910
99
11 Supplement
VEGAFLEX 81, rod version
SW 36 (1.
42")
(G¾, ¾ NPT)
61")
SW 41 (1. SW 46 (1.
(G1, 1 NPT)
(G1½, 1½ NPT)
81")
G¾, ¾ NPT, G1, 1 NPT, G1½, 1½ NPT
ø 8 mm
(0.32")
1 2
Fig. 73: VEGAFLEX 81, threaded version
L Sensor length, see chapter "Technical data" 1 Rod version ø 8 mm (0.315 in) 2 Rod version ø 12 mm (0.472 in)
46 mm
(1.81")
22 mm (0.87")
SW 55
(2.17")
G1½, 1½ NPT
L
ø 12¨mm
(0.47")
(1.18")
30mm
51 mm
L
(2.01")
100
44217-EN-130910
Loading...