VEGA FLEX81 User Manual

Operating Instructions
TDR sensor for continuous level and interface measurement of liquids
VEGAFLEX 81
4 … 20 mA/HART two-wire
Coax probe
Document ID: 42279
Mounting
Electrical connection
Quick start
The quick start procedure enables a quick setup with many applica-
tions. You can nd further information in the respective chapters of the
operating instructions manual.
Distance from the metallic vessel wall > 300 mm. Distance from
non-metallic vessel wall > 500 mm. The probe must not touch any installations or the vessel wall.
For further information see chapter "Mounting".
1. Make sure that the power supply corresponds to the specica-
tions on the type label.
2. Connect the instrument according to the following illustration
2
Set parameters
4...20mA
+
( )
(-)
1
2
5
678
1
Fig. 1: Electronics and connection compartment, single chamber housing
1 Voltage supply, signal output 2 For display and adjustment module or interface adapter 3 For external display and adjustment unit 4 Ground terminal for connection of the cable screen
3
4
See also chapter "Connecting to power supply"
For standard applications we recommend selecting the "Quick setup"
in the display and adjustment module.
1. In this menu item you can select the application. You can choose
between level and interface measurement.
2. In the menu item "Medium - Dielectric constant" you can dene
the medium type (medium).
3. Carry out the adjustment in the menu items "Min. adjustment" and "Max. adjustment".
42279-EN-130612
2
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
4. A "Linearization" is recommended for all vessels in which the vessel volume does not increase linearly with the level - e.g. in a horizontal cylindrical or spherical tank. Activate the appropriate curve.
5. A "False signal suppression" detects, marks and saves the false signals so that they are no longer taken into account for level measurement. We generally recommend a false signal suppres-
sion.
Parameterization example
Fur
ther steps
The sensor measures the distance from the sensor (reference plane) to the product surface.
1
4
2
Fig. 2: Measuring ranges - VEGAFLEX 81
1 Reference plane 2 Probe length L 3 Measuring range (default setting refers to the measuring range in water) 4 Upper dead band (in this area no measurement is possible) 5 Lower dead band (in this area no measurement is possible)
2
3
5
1
4
3
5
For this adjustment, the distance is entered when the vessel is full and nearly empty. If these values are not known, an adjustment with other distances, for example, 10 % and 90 % is also possible. Starting point for these distance specications is always the seal surface of the thread or ange.
1. In the menu "Additional settings", menu item "Damping" you can adjust the requested damping of the output signal.
2. Select the parameter of the current output and the output charac­teristics in the menu item "Current output".
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
3
Contents
Contents
1 About this document
1.1 Function ........................................................................................................................... 6
1.2 Target group ..................................................................................................................... 6
1.3 Symbolism used ............................................................................................................... 6
2 For your safety
2.1 Authorised personnel ....................................................................................................... 7
2.2 Appropriate use ................................................................................................................ 7
2.3 Warning about incorrect use ............................................................................................. 7
2.4 General safety instructions ............................................................................................... 7
2.5 CE conformity ................................................................................................................... 7
2.6 NAMUR recommendations .............................................................................................. 8
2.7 Environmental instructions ............................................................................................... 8
3 Product description
3.1 Conguration .................................................................................................................... 9
3.2 Principle of operation...................................................................................................... 10
3.3 Packaging, transport and storage ................................................................................... 13
3.4 Accessories and replacement parts ............................................................................... 13
4 Mounting
4.1 General instructions ....................................................................................................... 15
4.2 Mounting instructions ..................................................................................................... 15
5 Connecting to power supply
5.1 Preparing the connection ............................................................................................... 19
5.2 Connecting ..................................................................................................................... 20
5.3 Wiring plan, single chamber housing.............................................................................. 22
5.4 Wiring plan, double chamber housing ............................................................................ 22
5.5 Wiring plan, Ex-d-ia double chamber housing ................................................................ 24
5.6 Wiring plan - version IP 66/IP 68, 1 bar ........................................................................... 25
5.7 Supplementary electronics ............................................................................................. 26
5.8 Switch-on phase............................................................................................................. 26
6 Set up with the display and adjustment module
6.1 Insert display and adjustment module ............................................................................ 28
6.2 Adjustment system ......................................................................................................... 29
6.3 Parameter adjustment - Quick setup .............................................................................. 30
6.4 Parameter adjustment - Extended adjustment................................................................ 33
6.5 Saving the parameter adjustment data ........................................................................... 52
7 Setup with PACTware
7.1 Connect the PC .............................................................................................................. 53
7.2 Parameter adjustment with PACTware ............................................................................ 54
7.3 Set up with the quick setup ............................................................................................. 55
7.4 Saving the parameter adjustment data ........................................................................... 59
8 Set up with other systems
8.1 DD adjustment programs ............................................................................................... 60
8.2 Field Communicator 375, 475 ........................................................................................ 60
9 Diagnostics and service
9.1 Maintenance .................................................................................................................. 61
42279-EN-130612
4
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
9.2 Diagnosis memory ......................................................................................................... 61
9.3 Status messages ............................................................................................................ 62
9.4 Rectify faults ................................................................................................................... 66
9.5 Exchanging the electronics module ................................................................................ 69
9.6 Software update ............................................................................................................. 70
9.7 How to proceed in case of repair .................................................................................... 70
10 Dismounting
10.1 Dismounting steps.......................................................................................................... 71
10.2 Disposal ......................................................................................................................... 71
11 Supplement
11.1 Technical data ................................................................................................................ 72
11.2 Dimensions .................................................................................................................... 82
Contents
Safety instructions for Ex areas
Please note the Ex-specic safety information for installation and op­eration in Ex areas. These safety instructions are part of the operating instructions manual and come with the Ex-approved instruments.
42279-EN-130612
Editing status: 2013-06-11
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
5

1 About this document

1 About this document

1.1 Function

This operating instructions manual provides all the information you need for mounting, connection and setup as well as important instruc­tions for maintenance and fault rectication. Please read this informa­tion before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device.

1.2 Target group

This operating instructions manual is directed to trained specialist
personnel. The contents of this manual should be made available to
these personnel and put into practice by them.

1.3 Symbolism used

Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can result. Warning: If this warning is ignored, injury to persons and/or serious
damage to the instrument can result.
Danger: If this warning is ignored, serious injury to persons and/or destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
List
The dot set in front indicates a list with no implied sequence.
Action
→
This arrow indicates a single action.
1 Sequence of actions
Numbers set in front indicate successive steps in a procedure.
Battery disposal
This symbol indicates special information about the disposal of bat­teries and accumulators.
42279-EN-130612
6
VEGAFLEX 81 • 4 … 20 mA/HART two-wire

2 For your safety

2 For your safety

2.1 Authorised personnel

All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator.
During work on and with the device the required personal protective
equipment must always be worn.

2.2 Appropriate use

VEGAFLEX 81 is a sensor for continuous level measurement. You can nd detailed information on the application range in chapter
"Product description". Operational reliability is ensured only if the instrument is properly
used according to the specications in the operating instructions
manual as well as possible supplementary instructions.

2.3 Warning about incorrect use

Inappropriate or incorrect use of the instrument can give rise to application-specic hazards, e.g. vessel overll or damage to system
components through incorrect mounting or adjustment.

2.4 General safety instructions

This is a state-of-the-art instrument complying with all prevailing
regulations and guidelines. The instrument must only be operated in a
technically awless and reliable condition. The operator is responsible for the trouble-free operation of the instrument.
During the entire duration of use, the user is obliged to determine the compliance of the necessary occupational safety measures with the current valid rules and regulations and also take note of new regula-
tions.
The safety instructions in this operating instructions manual, the na­tional installation standards as well as the valid safety regulations and accident prevention rules must be observed by the user.
For safety and warranty reasons, any invasive work on the device
beyond that described in the operating instructions manual may be
carried out only by personnel authorised by the manufacturer. Arbi­trary conversions or modications are explicitly forbidden.
The safety approval markings and safety tips on the device must also be observed.

2.5 CE conformity

The device fullls the legal requirements of the applicable EC guide­lines. By axing the CE marking, we conrm successful testing of the
product.
You can nd the CE Certicate of Conformity in the download section of our homepage.
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
7
2 For your safety
Electromagnetic compatibility
Instruments with plastic housing as well as in four-wire or Ex-d-ia version are designed for use in an industrial environment. Neverthe­less, electromagnetic interference from electrical conductors and radiated emissions must be taken into account, as is usual with a class A instrument according to EN 61326-1. If the instrument is used in a dierent environment, the electromagnetic compatibility to other
instruments must be ensured by suitable measures.

2.6 NAMUR recommendations

NAMUR is the automation technology user association in the process industry in Germany. The published NAMUR recommendations are
accepted as the standard in eld instrumentation. The device fullls the requirements of the following NAMUR recom-
mendations:
NE 21 – Electromagnetic compatibility of equipment
NE 43 – Signal level for malfunction information from measuring
transducers
NE 53 – Compatibility of eld devices and display/adjustment
components
NE 107 – Self-monitoring and diagnosis of eld devices
For further information see www.namur.de.

2.7 Environmental instructions

Protection of the environment is one of our most important duties. That is why we have introduced an environment management system with the goal of continuously improving company environmental pro­tection. The environment management system is certied according
to DIN EN ISO 14001.
Please help us fulll this obligation by observing the environmental
instructions in this manual:
Chapter "Packaging, transport and storage"
Chapter "Disposal"
42279-EN-130612
8
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
Type plate

3 Product description

3 Product description
3.1 Conguration
The nameplate contains the most important data for identication and use of the instrument:
1
2
3
4 5
6
7 8 9
10
Fig. 3: Layout of the type label (example)
1 Instrument type 2 Product code 3 Approvals 4 Power supply and signal output, electronics 5 Protection rating 6 Probe length 7 Process and ambient temperature, process pressure 8 Material, wetted parts 9 Hardware and software version 10 Order number 11 Serial number of the instrument 12 Symbol of the device protection class 13 ID numbers, instrument documentation 14 Reminder to observe the instrument documentation
15 NotiedauthorityforCEmarking
16 Approval directives
16
15
14
13
12
11
Serial number
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
With the serial number of the instrument on the type label you can nd the following data on our homepage:
Article number of the instrument (HTML)
Delivery date (HTML)
Order-specic instrument features (HTML)
Operating instructions at the time of shipment (PDF)
Order-specic sensor data for an electronics exchange (XML)
Test certicate pressure transmitters (PDF)
Go to www.vega.com, "VEGA Tools" and "Serial number search".
As an alternative, you can nd the data via your Smartphone:
Download the Smartphone-App "VEGA Tools" from the "Apple
App Store " or the "Google Play Store" Scan the Data-Matrix-Code on the type label of the instrument or
Enter the serial number manually into the App
9
3 Product description
Scope of this operating instructions manual
Versions
Scope of deliv
Application area
ery
This operating instructions manual applies to the following instrument versions:
Hardware from 1.0.0
Software from 1.0.0
Only for instrument versions without SIL qualication
The instrument and the electronics version can be determined via the
product code on the type label as well as on the electronics.
Standard electronics: Type FX80H.-
The scope of delivery encompasses:
Sensor
Documentation
– this operating instructions manual – Test certicate measuring accuracy (optional)Operating instructions manual "Display and adjustment mod-
ule" (optional)
Supplementary instructions "GSM/GPRS radio module"
(optional)
Supplementary instructions manual "Heating for display and
adjustment module" (optional)
Supplementary instructions manual "Plug connector for con-
tinuously measuring sensors" (optional)
Ex-specic "Safety instructions" (with Ex versions)if necessary, further certicates

3.2 Principle of operation

The VEGAFLEX 81 is a level sensor with coax probe for continuous level or interface measurement, suitable for applications in liquids.
Functional principle ­level measurement
10
High frequency microwave pulses are guided along a steel cable or a rod. Upon reaching the product surface, the microwave pulses are reected. The running time is evaluated by the instrument and output­ted as level.
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
3 Product description
1
d
h
Fig. 4: Level measurement
1 Sensorreferenceplane(sealsurfaceoftheprocesstting)
d Distance to the interface (HART value 1) h Height - Level
Probe end tracking
To increase sensitivity, the probe is equipped with probe end tracking. In products with a low dielectric constant, this function is very helpful. This is the case, for example, in plastic granules, packing chips or in vessels with uidized products.
Between a dielectric constant of 1.5 and 3, the function switches on, if required. As soon as the level echo can no longer be detected, probe end tracking is automatically activated. The measurement is contin-
ued with the last calculated dielectric constant.
The accuracy thus depends on the stability of the dielectric constant. If you measure a medium with a dielectric constant below 1.5, probe
end tracking is always active. In this case, you have to enter the dielectric constant of the medium. A stable dielectric constant is very
important here.
Functional principle - in­ter
f
ace measurement
High frequency microwave impulses are guided along a steel cable or rod. Upon reaching the product surface, a part of the microwave im­pulses is reected. The other part passes through the upper product and is reected by the interface. The running times to the two product
layers are processed by the instrument.
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
11
3 Product description
L3
L2
1
d2
d1
TS
Prerequisites for inter­face measurement
h2
h1
Fig. 5: Interface measurement
1 Sensorreferenceplane(sealsurfaceoftheprocesstting)
d1 Distance to the interface (HART value 1) d2 Distance to the level (HART value 3) TS Thickness of the upper medium (d1 - d2) h1 Height - Interface h2 Height - Level L1 Lower medium L2 Upper medium L3 Gas phase
Upper medium (L2)
The upper medium must not be conductive
The dielectric constant of the upper medium or the actual distance
to the interface must be known (input required). Min. dielectric con­stant: 1.6. You can nd a list of dielectric constants on our home page: www.vega.com. The composition of the upper medium must be stable, no varying
products or mixtures The upper medium must be homogeneous, no stratications
within the medium
Min. thickness of the upper medium 50 mm (1.97 in)
Clear separation from the lower medium, emulsion phase or detri-
tus layer max. 50 mm (1.97 in) If possible, no foam on the surface
L1
12
Lower medium (L1)
The dielectric constant must be 10 higher than the dielectric
constant of the upper medium, preferably electrically conductive. Example: upper medium dielectric constant 2, lower medium at
least dielectric constant 12.
Gas phase (L3)
Air or gas mixture
Gas phase - dependent on the application, gas pahse does not
always exist (d2 = 0)
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
42279-EN-130612
3 Product description
Output signal
Pack
aging
Transport
Transport inspection
Storage
Storage and transport temperature
The instrument is always preset to the application "Level measure­ment".
For the interface measurement, you can select the requested output
signal with the setup.

3.3 Packaging, transport and storage

Your instrument was protected by packaging during transport. Its capacity to handle normal loads during transport is assured by a test based on ISO 4180.
The packaging of standard instruments consists of environment­friendly, recyclable cardboard. For special versions, PE foam or PE foil is also used. Dispose of the packaging material via specialised
recycling companies.
Transport must be carried out under consideration of the notes on the transport packaging. Nonobservance of these instructions can cause damage to the device.
The delivery must be checked for completeness and possible transit
damage immediately at receipt. Ascertained transit damage or con-
cealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and
stored according to the orientation and storage markings on the outside.
Unless otherwise indicated, the packages must be stored only under the following conditions:
Not in the open
Dry and dust free
Not exposed to corrosive media
Protected against solar radiation
Avoiding mechanical shock and vibration
Storage and transport temperature see chapter "Supplement -
Technical data - Ambient conditions" Relative humidity 20 … 85 %

3.4 Accessories and replacement parts

PLICSCOM
VEGACONNECT
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
The display and adjustment module PLICSCOM is used for measured value indication, adjustment and diagnosis. It can be inserted into the sensor or the external display and adjustment unit and removed at
any time.
You can nd further information in the operating instructions "Display and adjustment module PLICSCOM" (Document-ID 27835).
The interface adapter VEGACONNECT enables the connection of communication-capable instruments to the USB interface of a PC. For parameter adjustment of these instruments, the adjustment software
PACTware with VEGA-DTM is required.
13
3 Product description
You can nd further information in the operating instructions "Interface adapter VEGACONNECT" (Document-ID 32628).
VEGADIS 81
VEGADIS 62
PLICSMOBILE
Protective cap
Flanges
T61
The VEGADIS 81 is an external display and adjustment unit for VEGA
®
plics
sensors.
For sensors with double chamber housing the interface adapter "DISADAPT" is also required for VEGADIS 81.
You can nd further information in the operating instructions "VE- GADIS 81" (Document-ID 43814).
VEGADIS 62 is suitable for measured value indication and adjustment of sensors with HART protocol. It is looped into the 4 … 20 mA/HART
signal cable.
You can nd further information in the operating instructions "VE­GADIS 62" (Document-ID 36469).
The PLICSMOBILE T61 is an external GSM/GPRS radio unit for transmission of measured values and for remote parameter adjust­ment of plics
®
sensors. The adjustment is carried out via PACTware/
DTM by using the integrated USB connection.
You can nd further information in the supplementary instructions "PLICSMOBILE T61" (Document-ID 36849).
The protective cover protects the sensor housing against soiling and intense heat from solar radiation.
You will nd additional information in the supplementary instructions manual "Protective cover" (Document-ID 34296).
Screwed anges are available in dierent versions according to the following standards: DIN 2501, EN 1092-1, ANSI B 16.5, JIS B 2210­1984, GOST 12821-80.
You can nd additional information in the supplementary instructions manual "Flanges according to DIN-EN-ASME-JIS" (Document-ID
31088).
Electronics module
Display and adjustment module with heating
14
The electronics module VEGAFLEX series 80 is a replacement part
for TDR sensors of VEGAFLEX series 80. There is a dierent version available for each type of signal output.
You can nd further information in the operating instructions manual "Electronics module VEGAFLEX series 80".
The display and adjustment module can be optionally replaced by a
display and adjustment module with heating function.
You can use this display and adjustment module in an ambient tem-
perature range of -40 … +70 °C. You can nd further information in the operating instructions "Display
and adjustment module with heating" (Document-ID 31708).
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
42279-EN-130612
Screwing in

4 Mounting

4 Mounting

4.1 General instructions

On instruments with process tting thread, the hexagon must be tight­ened with a suitable screwdriver. Wrench size see chapter "Dimen- sions".
Warning:
The housing must not be used to screw the instrument in! Applying
tightening force can damage internal parts of the housing.
Protection against mois­ture
Protective caps
Suitability for the process conditions
Protect your instrument further through the following measures
against moisture penetration:
Use the recommended cable (see chapter "Connecting to power
supply")
Tighten the cable gland
Loop the connection cable downward in front of the cable gland
This applies particularly to:
Outdoor mounting
Installations in areas where high humidity is expected (e.g. through
cleaning processes)
Installations on cooled or heated vessels
In the case of instrument housings with self-sealing NPT threads, it is not possible to have the cable entries screwed in at the factory. The openings for the cable glands are therefore covered with red protec­tive caps as transport protection.
Prior to setup you have to replace these protective caps with ap­proved cable glands or close the openings with suitable blind plugs.
The suitable cable glands and blind plugs come with the instrument.
Make sure that all parts of the instrument exposed to the process are suitable for the existing process conditions.
These are mainly:
Active measuring component
Process tting
Process seal
Process conditions are particularly:
Process pressure
Process temperature
Chemical properties of the medium
Abrasion and mechanical inuences
You can nd the specications of the process conditions in chapter "Technical data" as well as on the nameplate.

4.2 Mounting instructions

Installation position
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
Mount VEGAFLEX 81 in such a way that the distance to vessel instal­lations or to the vessel wall is at least 300 mm (12 in). In non-metallic
15
4 Mounting
vessels, the distance to the vessel wall should be at least 500 mm
(19.7 in).
During operation, the probe must not touch any installations or the vessel wall. If necessary, fasten the probe end.
In vessels with conical bottom it can be advantageous to mount the sensor in the center of the vessel, as measurement is then possible nearly down to the lowest point of the bottom. Keep in mind that measurement all the way down to the tip of the probe may not be pos­sible. The exact value of the min. distance (lower dead band) is stated in chapter "Technical data".
Fig. 6: Vessel with conical bottom
Welding work
Inowingmedium
16
Before beginning the welding work, remove the electronics module from the sensor. By doing this, you avoid damage to the electronics through inductive coupling.
Do not mount the instruments in or above the lling stream. Make sure that you detect the product surface, not the inowing product.
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
Fig.7:Mountingofthesensorwithinowingmedium
4 Mounting
Measuring range
Pr
essure
Steam boiler applications
The reference plane for the measuring range of the sensors is the sealing surface of the thread or ange.
Keep in mind that a min. distance must be maintained below the refer­ence plane and possibly also at the end of the probe - measurement
in these areas is not possible (dead band). These dead bands are
listed in chapter "Technical data". Keep in mind for the adjustment that the default setting for the measuring range refers to water.
The process tting must be sealed if there is gauge or low pressure in the vessel. Before use, check if the seal material is resistant against
the measured product and the process temperature.
The max. permissible pressure is specied in chapter "Technical data" or on the type label of the sensor.
Vapours, superimposed gases, high pressures and temperature dif­ferences can change the spreading speed of radar impulses.
There are two possibilities to correct these deviations.
Corrective value in the process control system
In the technical data under "Inuence of superimposed gas and pres­sure on accuracy" you can nd a table with deviation values in some
typical gases or in steam.
In the control system (DCS) you can correct the measurement results
of the VEGAFLEX 81 with these values. The prerequisite is constant temperature and pressure in the vessel.
Automatic correction via the reference distance
The VEGAFLEX 81 can be equipped optionally with a running time correction via reference distance. The probe can carry out an auto­matic running time correction with it.
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
17
4 Mounting
The reference point must hence not be overlled. The upper dead band is hence 450 mm (17.7 in).
1
7
2
Fig. 8: Measuring ranges - VEGAFLEX 81 with steam compensation
1 Reference plane 2 Probe length L 3 Measuring range 4 Upper dead band 5 Lower dead band 6 Additional upper dead band by steam compensation 7 Reference measurement distance to steam compensation
6
4
3
5
18
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
Safety instructions

5 Connecting to power supply

5 Connecting to power supply

5.1 Preparing the connection

Always keep in mind the following safety instructions:
Connect only in the complete absence of line voltage
If overvoltage surges are expected, overvoltage arresters should
be installed
V
oltage supply
Connection to signal con ditioning instruments
Connection cable
Power supply and current signal are carried on the same two-wire
cable. The operating voltage can dier depending on the instrument version.
The data for power supply are specied in chapter "Technical data". Provide a reliable separation between the supply circuit and the
mains circuits according to DIN EN 61140 VDE 0140-1.
Keep in mind the following additional factors that inuence the operat­ing voltage:
Lower output voltage of the power supply unit under nominal load
(e.g. with a sensor current of 20.5 mA or 22 mA in case of fault
message)
Inuence of additional instruments in the circuit (see load values in
chapter "Technical data")
The signal conditioning instruments VEGAMET and VEGASCAN
-
have digital sensor recognition. When connecting VEGAFLEX 81, an up-to-date version of the signal conditioning instrument software is required. For a software update go to "www.vega.com/downloads" and "Software".
The instrument is connected with standard two-wire cable without
screen. If electromagnetic interference is expected which is above the test values of EN 61326-1 for industrial areas, screened cable should
be used.
Use cable with round cross section for instruments with housing and cable gland. To ensure the seal eect of the cable gland (IP protection rating), nd out which cable outer diameter the cable gland is suitable for.
5 … 9 mm (0.20 … 0.35 in)
6 … 12 mm (0.24 … 0.47 in)
10 … 14 mm (0.40 … 0.55 in)
Use a cable gland tting the cable diameter. We generally recommend the use of screened cable for HART multi-
drop mode.
Cable gland ½ NPT
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
With plastic housing, the NPT cable gland or the Conduit steel tube
must be screwed without grease into the threaded insert.
Max. torque for all housings see chapter "Technical data".
19
5 Connecting to power supply
Cable screening and grounding
If screened cable is necessary, connect the cable screen on both ends to ground potential. In the sensor, the screen must be connected
directly to the internal ground terminal. The ground terminal on the outside of the housing must be connected to the potential equalisa­tion (low impedance).
If potential equalisation currents are expected, the connection on the processing side must be made via a ceramic capacitor (e. g. 1 nF, 1500 V). The low-frequency potential equalisation currents are thus suppressed, but the protective eect against high frequency interfer-
ence signals remains.
Warning:
Signicant potential dierences exist inside galvanization plants as well as on vessels with cathodic corrosion protection. Considerable equalisation currents can ow over the cable screen if the screen is
grounded on both ends. To avoid this in such applications, the cable screen must be connect-
ed to ground potential only at one end (in the switching cabinet). The cable screen must not be connected to the internal ground terminal in the sensor and the outer ground terminal on the housing must not be connected to the potential equalisation!
Information:
The metallic parts of the instrument (process tting, transmitter, con­centric tube, etc.) are conductively connected with the inner and outer ground terminal on the housing. This connection exists either directly via connecting metallic parts or, in case of instruments with external electronics, via the screen of the special connection cable.
You can nd specications on the potential connections inside the instrument in chapter "Technical data".
Connection technology
Connection procedure
20

5.2 Connecting

The voltage supply and signal output are connected via the spring-
loaded terminals in the housing.
The connection to the display and adjustment module or to the inter-
face adapter is carried out via contact pins in the housing.
Information:
The terminal block is pluggable and can be removed from the electronics. To do this, lift the terminal block with a small screwdriver and pull it out. When reinserting the terminal block, you should hear it
snap in.
Proceed as follows:
1. Unscrew the housing cover
2. If a display and adjustment module is installed, remove it by turn­ing it slightly to the left.
3. Loosen compression nut of the cable entry
4. Remove approx. 10 cm (4 in) of the cable mantle, strip approx. 1 cm (0.4 in) of insulation from the ends of the individual wires
5. Insert the cable into the sensor through the cable entry
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
42279-EN-130612
5 Connecting to power supply
Fig. 9: Connection steps 5 and 6 - Single chamber housing
Fig. 10: Connection steps 5 and 6 - Double chamber housing
6. Insert the wire ends into the terminals according to the wiring plan
Information:
Solid cores as well as exible cores with wire end sleeves are insert­ed directly into the terminal openings. In case of exible cores without end sleeves, press the terminal from above with a small screwdriver; the terminal opening is freed. When the screwdriver is released, the
terminal closes again.
You can nd further information on the max. wire cross-section under "Technical data/Electromechanical data"
7. Check the hold of the wires in the terminals by lightly pulling on
them
8. Connect the screen to the internal ground terminal, connect the
outer ground terminal to potential equalisation
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
21
5 Connecting to power supply
2
2
Electronics and connec­tion compartment
9. Tighten the compression nut of the cable entry. The seal ring must
completely encircle the cable
10. Reinsert the display and adjustment module, if one was installed
11. Screw the housing cover back on
The electrical connection is hence nished.

5.3 Wiring plan, single chamber housing

The following illustration applies to the non-Ex, Ex-ia and Ex-d ver­sion.
Electronics compartment
+
( )
4...20mA
1
(-)
2
5
678
3
4
1
Fig. 11: Electronics and connection compartment, single chamber housing
1 Voltage supply, signal output 2 For display and adjustment module or interface adapter 3 For external display and adjustment unit 4 Ground terminal for connection of the cable screen

5.4 Wiring plan, double chamber housing

The following illustrations apply to the non-Ex as well as to the Ex-ia version.
4...20mA
( )
+
(-)
1
2
Fig. 12: Electronics compartment, double chamber housing
1 Internal connection to the connection compartment 2 For display and adjustment module or interface adapter
567
11
8
42279-EN-130612
22
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
2
Connection compartment
5 Connecting to power supply
Connection compartment
- Radio module PLICS­MOBILE
+
( )
4...20mA
1
(-)
2
5
678
3
4
1
Fig. 13: Connection compartment, double chamber housing
1 Voltage supply, signal output 2 For display and adjustment module or interface adapter 3 For external display and adjustment unit 4 Ground terminal for connection of the cable screen
Information:
The use of an external display and adjustment unit and a display and
adjustment module in parallel in the connection compartment is not supported.
Status
SIM-Card
Test
Bus
( )
+
(-)
1
2
USB
1
Fig. 14: Connection compartment radio module PLICSMOBILE
1 Voltage supply
You can nd detailed information on connection in the supplementary instructions "PLICSMOBILE GSM/GPRS radio module".
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
23
5 Connecting to power supply
2
Electronics compartment
Connection compartment

5.5 Wiring plan, Ex-d-ia double chamber housing

4...20mA
+
( )
(-)
1
2
Fig. 15: Electronics compartment, double chamber housing
1 Internal connection to the connection compartment 2 For display and adjustment module or interface adapter 3 Internal connection to the plug connector for external display and adjust-
ment unit (optional)
Note:
If an instrument with Ex-d-ia approval is used, HART multidrop op­eration is not possible.
567
31
8
24
4...20mA
+
( )
(-)
1
2
2
1
Fig. 16: Connection compartment, Ex-d double chamber housing
1 Voltage supply, signal output 2 Ground terminal for connection of the cable screen
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
42279-EN-130612
DIS-ADAPT with plug connector M12 x 1 for external display and adjustment unit
5 Connecting to power supply
1
2
3
Fig. 17: View to the electronics compartment
1 DIS-ADAPT 2 Internal plug connection 3 Plug connector M12 x 1
34
Fig. 18: Top view of the plug connector
1 Pin 1 2 Pin 2 3 Pin 3 4 Pin 4
Contact pin Colour connection ca-
Pin 1 Brown 5
Pin 2 White 6
Pin 3 Blue 7
Pin 4 Black 8

5.6 Wiring plan - version IP 66/IP 68, 1 bar

Wire assignment, con­nection cable
Fig.19:Wireassignmentx-connectedconnectioncable
1 brown (+) and blue (-) to power supply or to the processing system 2 Shielding
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
1
ble in the sensor
2
Terminal, electronics module
1
2
25
5 Connecting to power supply
Supplementary electron­ics - Second current output

5.7 Supplementary electronics

You can use the supplementary electronics - second current output to
provided a second measured value. Both current outputs are passive and need a power supply.
Supplementar
y elec­tronics - Radio module PLICSMOBILE
I
4...20mA
( )
+
(-)
1
2
1 2
3
I I
4...20mA
( )
+
(-)
7
8
Fig. 20: Connection compartment double chamber housing, supplementary electronics - second current output
1 First current output (I) - Voltage supply and signal output (HART) 2 Second current output (II) - Voltage supply and signal output (without HART) 3 Ground terminal for connection of the cable screen
The radio module PLICSMOBILE is an external GSM/GPRS radio
unit for transmission of measured values and for remote parameter
adjustment.
Status
SIM-Card
Test
Bus
( )
+
(-)
1
2
USB
1
Fig. 21: Radio module PLICSMOBILE integrated in the connection compart­ment
1 Voltage supply
You can nd detailed information on connection in the supplementary instructions "PLICSMOBILE GSM/GPRS radio module".
26

5.8 Switch-on phase

After connecting the instrument to power supply or after a voltage recurrence, the instrument carries out a self-check for approx. 30 s:
Internal check of the electronics
Indication of the instrument type, hardware and software version,
measurement loop name on the display or PC
Indication of the status message "F 105 Determine measured
value" on the display or PC
The output signal jumps to the set error current
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
42279-EN-130612
5 Connecting to power supply
As soon as a plausible measured value is found, the corresponding current is outputted to the signal cable. The value corresponds to the actual level as well as the settings already carried out, e.g. factory
setting.
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
27

6 Set up with the display and adjustment module

6 Set up with the display and adjustment
module

6.1 Insert display and adjustment module

The display and adjustment module can be inserted into the sensor and removed any time. Four positions displaced by 90° can be se­lected. It is not necessary to interrupt the power supply.
Proceed as follows:
1. Unscrew the housing cover
2. Place the display and adjustment module in the requested posi­tion onto the electronics and turn to the right until it snaps in
3. Screw housing cover with inspection window tightly back on
Removal is carried out in reverse order. The display and adjustment module is powered by the sensor, an ad-
ditional connection is not necessary.
28
Fig. 22: Insertion of the display and adjustment module with single chamber housing into the electronics compartment
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
6 Set up with the display and adjustment module
1 2
Fig. 23: Insertion of the display and adjustment module into the double chamber housing
1 In the electronics compartment 2 In the connection compartment (with Ex-d-ia version not possible)
Note:
If you intend to retrot the instrument with a display and adjustment module for continuous measured value indication, a higher cover with
an inspection glass is required.

6.2 Adjustment system

Fig. 24: Display and adjustment elements
1 LC display 2 Adjustment keys
Key functions
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
[OK] key:
Move to the menu overview
1
2
29
6 Set up with the display and adjustment module
Conrm selected menu – Edit parameter – Save value
[->] key:
Presentation, change measured value – Select list entry – Select editing position
[+] key:
Change value of the parameter
[ESC] key:
– Interrupt input – Jump to next higher menu
Adjustment system
Switch-on phase
ed value indica-
Measur tion
The sensor is adjusted via the four keys of the display and adjust-
ment module. The LC display indicates the individual menu items. The functions of the individual keys are shown in the above illustration. Approx. 10 minutes after the last pressing of a key, an automatic reset to measured value indication is triggered. Any values not conrmed with [OK] will not be saved.
After switching on, the VEGAFLEX 81 carries out a short self-test where the device software is checked.
The output signal transmits a fault signal during the switch-on phase. The following information is displayed on the display and adjustment
module during the startup procedure:
Instrument type
Device name
Software version (SW-Ver)
Hardware version (HW-Ver)
With the [->] key you can move between three dierent indication
modes.
In the rst view, the selected measured value is displayed in large
digits. In the second view, the selected measured value and a correspond-
ing bar graph presentation are displayed.
In the third view, the selected measured value as well as a second selectable value, e.g. the temperature, are displayed.
Quick setup
30
42279-EN-130612

6.3 Parameter adjustment - Quick setup

To quickly and easily adapt the sensor to the application, select
the menu item "Quick setup" in the start graphic on the display and
adjustment module.
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
General information
Le
vel measurement
6 Set up with the display and adjustment module
You can nd "Extended adjustment" in the next sub-chapter.
Measurement loop name
In the rst menu item you can assign a suitable measurement loop name. You can enter a name with max. 19 characters.
Medium type
In the next menu item you can see which product type the instrument is suitable for. If your instrument is only suitable for a certain medium, this menu item is not visible.
Application
In this menu item, you can select the application. You can choose between level measurement and interface measurement. You can also choose between measurement in a vessel or in a bypass or
standpipe.
Medium - dielectric constant
In this menu item, you can dene the medium type (product).
Max. adjustment
In this menu item, you can enter the max. adjustment for the level. Enter the appropriate distance value in m (corresponding to the
percentage value) for the full vessel. The distance refers to the sensor reference plane (seal surface of the process tting). Keep in mind that the max. level must lie below the dead band.
Min. adjustment
In this menu item, you can enter the min. adjustment for the level. Enter the suitable distance value in m for the empty vessel (e.g.
distance from the ange to the probe end) corresponding to the per­centage value. The distance refers tot he sensor reference plane (seal surface of the process tting).
Interface measurement
Dielectric constant - upper medium
In this menu item, you can dene the medium type (product).
Max. adjustment
In this menu item, you can enter the max. adjustment for the level. Enter the appropriate distance value in m (corresponding to the
percentage value) for the full vessel. The distance refers to the sensor
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
31
6 Set up with the display and adjustment module
reference plane (seal surface of the process tting). Keep in mind that the max. level must lie below the dead band.
Min. adjustment
In this menu item, you can enter the min. adjustment for the level. Enter the suitable distance value in m for the empty vessel (e.g.
distance from the ange to the probe end) corresponding to the per­centage value. The distance refers tot he sensor reference plane (seal surface of the process tting).
Max. adjustment - Interface
Carry out the max. adjustment for the interface. To do this, enter the percentage value and the suitable distance value
in m for the full vessel.
Min. adjustment - Interface
Carry out the min. adjustment for the interface. To do this, enter the percentage value and the suitable distance value
in m for the empty vessel.
Linearization
32
Linearization
A linearization is necessary for all vessels in which the vessel volume does not increase linearly with the level - e.g. a horizontal cylindri­cal or spherical tank, when the indication or output of the volume is required. Corresponding linearization curves are preprogrammed for these vessels. They represent the correlation between the level percentage and vessel volume.
The linearization applies for the measured value indication and the current output. By activating the suitable curve, the percentage vessel volume is displayed correctly.
False signal suppression
High sockets and internal vessel installations cause interfering reec­tions and can inuence the measurement.
A false signal suppression detects, marks and saves these false signals so that they are no longer taken into account for the level and interface measurement. We generally recommend carrying out a false signal suppression to achieve the best possible accuracy. This should be done with the lowest possible level so that all potential interfering reections can be detected.
Enter the actual distance from the sensor to the product surface. All interfering signals in this section are detected by the sensor and
stored.
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
42279-EN-130612
6 Set up with the display and adjustment module
The instrument carries out an automatic false signal suppression as soon as the probe is uncovered. The false signal suppression is
always updated.

6.4 Parameter adjustment - Extended adjustment

For technically demanding measurement loops you can carry out
extended settings in "Extended adjustment".
Main menu
The main menu is divided into ve sections with the following func-
tions:
Setup: Settings, e.g. measurement loop name, medium, vessel, adjustment, signal output, device unit, false signal suppression, linearization curve
Display: Settings, e.g., for language, measured value display, lighting Diagnosis: Information, e.g. on instrument status, pointer, measure-
ment reliability, simulation, echo curve
Additional adjustments: Reset, date/time, reset, copy function Info: Instrument name, hardware and software version, date of manu-
facture, instrument features
Note:
For optimum adjustment of the measurement, the individual submenu items in the main menu item "Setup" should be selected one after the other and provided with the correct parameters. If possible, go through the items in the given sequence.
The procedure is described below.
The following submenu points are available:
The submenu points described below.
Setup - Measurement loop name
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
Here you can assign a suitable measurement loop name. Push the
"OK" key to start the processing. With the "+" key you change the sign and with the "->" key you jump to the next position.
You can enter names with max. 19 characters. The character set
comprises:
Capital letters from A … Z
Numbers from 0 … 9
33
6 Set up with the display and adjustment module
Special characters + - / _ blanks
Setup - Units
Setup - Probe length
Setup - Application ­Product type
Setup - Application - Ap­plication
In this menu item you select the distance unit and the temperature unit.
With the distance units you can choose between m, mm and ft and with the temperature units betwenn °C, °F and K.
In this menu item you can enter the probe length or have the length
determined automatically by the sensor system.
When choosing "Ye s ", then the probe length will be determined automatically. When choosing "No", you can enter the probe length
manually.
In this menu item you can select which type of medium you want to
measure. You can choose between liquid or bulk solid.
In this menu item, you can select the application. You can choose between level measurement and interface measurement. You can also choose between measurement in a vessel or in a bypass or
standpipe.
Note:
The selection of the application has a considerable inuence on all other menu items. Keep in mind that as you continue with the param­eter adjustment, individual menu items are only optionally available.
You have the option of choosing the demonstration mode. This mode is only suitable for test and demonstration purposes. In this mode, the sensor ignores the parameters of the application and reacts immedi-
ately to each change.
42279-EN-130612
34
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
6 Set up with the display and adjustment module
Setup - Application - Me­dium, dielectric constant
Setup - Application - Gas phase
Setup - Application - Di­electric constant
In this menu item, you can dene the medium type (product). This menu item is only available if you have selected level measure-
ment under the menu item "Application".
You can choose between the following medium types:
Dielectricgure Medium type Examples
> 10 Water-based liq-
uids
3 … 10 Chemical mix-
tures
< 3 Hydrocarbons Solvents, oils, liquid gas
Acids, alcalis, water
Chlorobenzene, nitro lacquer, aniline, isocyanate, chloroform
This menu item is only available, if you have chosen interface meas-
urement under the menu item "Application". In this menu item you can enter if there is a superimposed gas phase in your application.
Only set the function to "Ye s ", if the gas phase is permanently pre-
sent.
This menu item is only available if you have selected interface meas-
urement under the menu item "Application". In this menu item you can choose the medium type of the upper medium.
You can enter the dielectric constant of the upper medium directly or have the value determined by the instrument. To do this you have to enter the measured or known distance to the interface.
Setup - Max. adjustment
evel
L
In this menu item you can enter the max. adjustment for the level. With interface measurement this is the maximum total level.
Adjust the requested percentage value with [+] and store with [OK].
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
35
6 Set up with the display and adjustment module
Enter the appropriate distance value in m (corresponding to the percentage value) for the full vessel. The distance refers to the sensor reference plane (seal surface of the process tting). Keep in mind that the max. level must lie below the dead band.
Setup - Min. adjustment
el
Lev
Setup - Max. adjustment ­Interface
In this menu item you can enter the min. adjustment for the level. With interface measurement this is the minimum total level.
Adjust the requested percentage value with [+] and store with [OK].
Enter the suitable distance value in m for the empty vessel (e.g. distance from the ange to the probe end) corresponding to the per­centage value. The distance refers tot he sensor reference plane (seal surface of the process tting).
This menu item is only available if you have selected interface meas-
urement under the menu item "Application".
You can accept the adjustment of the level measurement also for the interface measurement. If you select "Yes", the current setting will be
displayed.
42279-EN-130612
36
If you have selected "No", you can enter the adjustment for the inter­face separately. Enter the requested percentage value.
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
6 Set up with the display and adjustment module
For the full vessel, enter the distance value in m matching the per­centage value.
Setup - Min. adjustment ­Interf
ace
Setup - Damping
This menu item is only available if you have selected interface meas-
urement under the menu item "Application". If you have selected "Ye s " in the previous menu item (accept adjustment of the level measure­ment), the current setting will be displayed.
If you have selected "No", you can enter the adjustment for the inter­face measurement separately.
Enter the respective distance value in m for the empty vessel cor­responding to the percentage value.
To damp process-dependent measured value uctuations, set an
integration time of 0 … 999 s in this menu item. If you have selected interface measurement under the menu item
"Application", you can adjust the damping for the level and the inter­face separately.
The default setting is a damping of 0 s.
Setup - Linearization
A linearization is necessary for all vessels in which the vessel volume does not increase linearly with the level - e.g. a horizontal cylindri­cal or spherical tank, when the indication or output of the volume is required. Corresponding linearization curves are preprogrammed for these vessels. They represent the correlation between the level percentage and vessel volume.
The linearization applies to the measured value indication and the current output. By activating the appropriate curve, the volume per­centage of the vessel is displayed correctly. If the volume should not be displayed in percent but e.g. in l or kg, a scaling can be also set in the menu item "Display".
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
37
6 Set up with the display and adjustment module
Warning:
If a linearization curve is selected, the measuring signal is no longer necessarily linear to the lling height. This must be considered by the
user especially when adjusting the switching point on the limit signal transmitter.
In the following, you have to enter the values for your vessel, for example the vessel height and the socket correction.
For non-linear vessel forms, enter the vessel height und the socket
correction.
For the vessel height, you have to enter the total height of the vessel. For the socket correction you have to enter the height of the socket
above the upper edge of the vessel. If the socket is lower than the up­per edge of the vessel, this value can also be negative.
+ h
- h
38
D
Fig. 25: Vessel height und socket correction value
D Vessel height +h Positive socket correction value
-h Negative socket correction value
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
42279-EN-130612
6 Set up with the display and adjustment module
Setup - Scaling Level
Setup - Scaling Le
vel 1
Setup - Scaling Level 2
Setup/Scaling Interface
Since scaling is very extensive, scaling of the level value was divided
into two menu items.
In menu item "Level 1" you dene the scaling variable and the scaling unit for the level value on the display, e.g. volume in l.
Since scaling is very extensive, scaling of the level value was divided
into two menu items.
In menu item "Level 2" you dene the scaling format on the display and the scaling of the measured level value for 0 % and 100 %.
Since scaling is very extensive, scaling of the interface value was divided into two menu items.
Setup - Scaling Interface (1)
In menu item "Interface 1" you dene the scaling size and the scaling unit for the interface value on the display, e.g. volume in l.
You can accept the scaling of the level measurement also for the interface measurement. If you select "Ye s", the current setting is
displayed.
If you have selected "No", you can enter the scaling for the interface
separately.
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
39
6 Set up with the display and adjustment module
Setup - Scaling Interface (2)
Setup - Current output Size
Setup - Current output, adjustment
In menu item "Interface (2)" you dene the scaling format on the display and the scaling of the interface measured value for 0 % and 100 %.
In menu item"Current output, size" you determine which measured value the current output refers to.
In menu item "Current output, adjustment" you can assign a respec­tive measured value to the current output.
Setup/Current output mode
40
In the menu item "Current output mode" you determine the output characteristics and reaction of the current output in case of failure.
42279-EN-130612
The default setting is output characteristics 4 … 20 mA, failure mode
< 3.6 mA.
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
6 Set up with the display and adjustment module
Setup - Current output Min./Max.
Setup - Current output 2
Setup - HART variables (1)
In the menu item "Current output Min./Max.", you determine the reac­tion of the current output during operation.
The default setting is min. current 3.8 mA and max. current 20.5 mA.
If a supplementary electronics with a second current output is im­plemented in the instrument, then you can adjust the second current
output separately.
In menu item"Current output 2" you specify which measured value the second current output refers to.
The procedure corresponds to the previous settings of the standard current output. See "Setup - Current output".
Since the HART variables are very extensive, the indication was divided into two menu items.
In the menu "HART variables 1" you can have the rst HART value PV (primary value) and the second HART value SV (secondary value) of
the sensor displayed.
The values cannot be changed in the display and adjustment module. To change these values, you have to use the adjustment software
PAC Tw a re .
Setup - HART variables (2)
In the menu "HART variables 2" you can have the third HART value TV (third value) and the fourth HART value QV (quarter value) of the
sensor displayed.
The values cannot be changed in the display and adjustment module. To change these values, you have to use the adjustment software
PAC Tw a re .
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
41
6 Set up with the display and adjustment module
Setup - False signal sup­pression
The following circumstances cause interfering reections and can
inuence the measurement:
High sockets
Vessel installations such as struts
Note:
A false signal suppression detects, marks and saves these false signals so that they are no longer taken into account for the level and interface measurement. We generally recommend carrying out a false signal suppression to achieve the best possible accuracy. This should be done with the lowest possible level so that all potential interfering reections can be detected.
Proceed as follows:
Enter the actual distance from the sensor to the product surface.
All interfering signals in this section are detected by the sensor and
stored.
Note:
Check the distance to the product surface, because if an incorrect (too large) value is entered, the existing level will be saved as a false echo. The lling level would then no longer be detectable in this area.
If a false signal suppression has already been created in the sensor, the following menu window appears when selecting "False signal suppression":
Lock/release setup - Ad­justment
42
The instrument carries out an automatic false signal suppression as soon as the probe is uncovered. The false signal suppression is
always updated.
The menu item "Delete" is used to completely delete an already cre­ated false signal suppression. This is useful if the saved false signal
suppression no longer matches the metrological conditions in the
vessel.
In the menu item "Lock/unlock adjustment", you can protect the sensor parameters against unauthorized modication. The PIN is activated/deactivated permanently.
The following adjustment functions are possible without entering the
PIN:
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
42279-EN-130612
6 Set up with the display and adjustment module
Select menu items and show data
Read data from the sensor into the display and adjustment mod-
ule.
Caution:
With active PIN, adjustment via PACTware/DTM as well as other
systems is also blocked.
You can change the PIN number under "Additional adjustments ­PIN".
Display
Display - Menu language
Display - Displayed value 1
In the main menu point "Display", the individual submenu points should be selected subsequently and provided with the correct parameters to ensure the optimum adjustment of the display options. The procedure is described in the following.
The following submenu points are available:
The submenu points described below.
This menu item enables the setting of the requested national lan-
guage.
In the delivery status, the sensor is set to the ordered national lan­guage.
In this menu item, you dene the indication of the measured value on the display. You can display two dierent measured values. In this menu item, you dene measured value 1.
The default setting for the displayed value 1 is "Filling height Level".
Display - Displayed value 2
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
In this menu item, you dene the indication of the measured value on the display. You can display two dierent measured values. In this menu item, you dene measured value 2.
43
6 Set up with the display and adjustment module
The default setting for the displayed value 2 is the electronics tem­perature.
Display - Backlight
Diagnostics - Device status
Diagnostics - P
eak values
Distance
The optionally integrated background lighting can be adjusted via the
adjustment menu. The function depends on the height of the supply voltage, see "Technical data".
The lighting is switched o in the delivery status.
In this menu item, the device status is displayed.
The respective min. and max. measured value is saved in the sen-
sor. The two values are displayed in the menu item "Peak values, distance".
If you have selected interface measurement under the menu item "Setup - Application", the peak values of the interface measurement are displayed in addition to the peak values of the level measurement.
In another window you can carry out a reset of the two peak values
separately.
Diagnostics - Peak values Measurement certainty
44
The respective min. and max. measured values are saved in the
sensor. The two values are displayed in the menu item "Peak values, measurement certainty".
The measurement can be inuenced by the process conditions. In this menu item, the measurement certainty of the level measurement is displayed as percentage value. The higher the value, the more reli­able the measurement. Values > 90 % indicate reliable measurement.
If you have selected interface measurement under the menu item "Setup - Application", the peak values of the interface measurement are displayed in addition to the peak values of the level measurement.
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
42279-EN-130612
6 Set up with the display and adjustment module
In another window you can carry out a reset of the two peak values
separately.
Diagnostics - Peak values Additional
Diagnostics - Echo cur
ve
The respective min. and max. measured values are saved in the
sensor. The values are displayed in the menu item "Peak values Ad- ditional".
This menu item displays the peak values of the electronics tempera-
ture as well as the dielectric constant.
In another window you can carry out a reset of the two peak values
separately.
The menu item "Echo curve" shows the signal strength of the echoes
over the measuring range in V. The signal strength enables an evalua­tion of the quality of the measurement.
With the following functions you can zoom part sections of the echo curve.
"X-Zoom": Zoom function for the meas. distance
"Y-Zoom": 1, 2, 5 and 10x signal magnication in "V"
"Unzoom": Reset the presentation to the nominal measuring range
with single magnication
Diagnosis - Simulation
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
In this menu item you can simulate measured values via the current output. This allows the signal path to be tested, e.g. via downstream indicating instruments or the input card of the control system.
45
6 Set up with the display and adjustment module
Select the requested simulation size and adjust the requested value.
Caution:
During simulation, the simulated value is outputted as 4 … 20 mA cur­rent value and digital HART signal.
Push the [ESC] key to deactivate the simulation.
Information:
The simulation is terminated automatically 60 minutes after the last
key has been pushed.
Diagnostics - Echo curve memory
Additional settings - PIN
With the menu item "Setup" the echo curve it is possible to save at
the time of setup. This is generally recommended; for using the Asset Management functions it is necessary. If possible, the curve should be saved with a low level in the vessel.
With this, you can detect signal changes over the operating time. With the adjustment software PACTware and the PC, the high-resolution echo curve can be displayed and used to compare the echo curve of the setup with the actual echo curve.
The function "Echo curve memory" enables storing echo curves of
the measurement.
Under the sub-menu item "Echo curve memory" you can store the current echo curve.
Parameter settings for recording the echo curve and the settings of the echo curve itself can be carried out in the adjustment software
PAC Tw a re .
With the adjustment software PACTware and the PC the high-reso­lution echo curve can be displayed and used later on to assess the quality of the measurement.
42279-EN-130612
Entering a 4-digit PIN protects the sensor data against unauthorized
access and unintentional modication. In this menu item, the PIN is displayed or edited and changed. However, this menu item is only
46
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
6 Set up with the display and adjustment module
available if adjustment is enabled in the menu "Lock/Release setup/ adjustment".
In delivery status, the PIN is "0000".
Additional adjustments ­Date Time
Additional adjustments
- Reset
In this menu item, the internal clock of the sensor is adjusted.
With a reset, certain parameter adjustments carried out by the user
are reset.
The following reset functions are available: Delivery status: Restoring the parameter settings at the time of ship-
ment from the factory incl. the order-specic settings. A created false signal suppression, user-programmable linearization curve as well as the measured value memory will be deleted.
Basic settings: Resetting the parameter settings incl. special pa-
rameters to the default values of the respective instrument. A created false signal suppression, user programmable linearization curve as well as the measured value memory will be deleted.
The following table shows the default values of the instrument. De­pending on the instrument version or application, all menu items may not be available or some may be dierently assigned:
Setup
Menu item Default value Modiedvalue
Block adjustment Released
Measurement loop name Sensor
Units Distance unit: mm
Temperature unit: °C
Probe length Length of the probe Ex factory
Medium type Liquid
Application Level, vessel
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
47
6 Set up with the display and adjustment module
Menu item Default value Modiedvalue
Medium, dielectric constant Water-based, > 10
Superimposed gas phase Ye s
Dielectric gure, upper medium (TS) 1.5
Tube inner diameter
Max. adjustment - Level 100 %
Max. adjustment - Level Distance: 0.000 m(d) - note block-
Min. adjustment - Level
Min. adjustment - Level Distance: Probe length - take dead
Accept adjustment of the level measurement? Yes
Max. adjustment - Interface
Max. adjustment - Interface Distance: 0.000 m(d) - note block-
Min. adjustment - Interface
Min. adjustment - Interface Distance: Probe length - take dead
Integration time - Level
Integration time - Interface 0.0 s
Linearization type Linear
Linearization - Socket correction
Linearization - Vessel height Probe length
Scaling size - Level Volume in l
Scaling unit - Level Litres
Scaling format - Level Without decimal positions
Scaling level - 100 % correspond to 100
Scaling level - 0 % correspond to 0
Accept scaling of the level measurement Ye s
Scaling variable - Interface Volume
Scaling unit - Interface Litres
Scaling format - Interface Without decimal positions
Scaling interface - 100 % correspond to 100
Scaling interface - 0 % correspond to 0
Current output, output variable First HART variable (PV)
Current output - Output characteristics
Current output - Reaction in case of failure ≤ 3.6 mA
Current output - Min. 3.8 mA
200 mm
ing distances
0 %
band into account
100 %
ing distances
0 %
band into account
0.0 s
0 mm
Lin. percent - Level
0 … 100 % correspond to
4 … 20 mA
42279-EN-130612
48
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
6 Set up with the display and adjustment module
Menu item Default value Modiedvalue
Current output - Max. 20.5 mA
Current output 2 - Output variable Second HART variable (SV)
Current output 2 - Output characteristics
Current output 2 - Reaction in case of failure ≤ 3.6 mA
Current output - Min. 3.8 mA
Current output - Max. 20.5 mA
Third HART variable (TV) Measurement certainty, level
Fourth HART variable (QV) Electronics temperature
Distance - Level
0 … 100 % correspond to
4 … 20 mA
Display
Menu item Default value Modiedvalue
Language Order-specic
Displayed value 1 Filling height Level
Displayed value 2 Electronics temperature
Backlight Switched o
Diagnostics
Menu item Default value Modiedvalue
Status signals - Function control Switched on
Status signals - Out of specication Switched o
Status signals - Maintenance Switched o
Device memory - Echo curve memory Stopped
Device memory - Measured value memory Started
Device memory - Measured value memory ­Measured values
Device memory - Measured value memory - Re­cording in time interval
Device memory - Measured value memory - Re­cording with measured value dierence
Device memory - Measured value memory - Start with measured value
Device memory - Measured value memory - Stop with measured value
Device memory - Measured value memory - Stop recording when memory is full
Distance level, percentage value level, reliability level, electronics
temperature
3 min.
15 %
Not active
Not active
Not active
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
49
6 Set up with the display and adjustment module
Additional adjustments
Menu item Default value Modiedvalue
PIN 0000
Date Actual date
Time Actual time
Time - Format
Probe type Device-specic
24 hours
Additional adjustments ­Copy instrument settings
Additional adjustments ­Probe type
The instrument settings are copied with this function. The following
functions are available:
Read from sensor: Read data from sensor and store into the
display and adjustment module
Write into sensor: Store data from the display and adjustment
module back to the sensor
The following data or settings for adjustment of the display and ad­justment module are saved:
All data of the menu "Setup" and "Display"
In the menu "Additional adjustments" the items "Reset, Date/Time"
Special parameters
The copied data are permanently saved in an EEPROM memory in the display and adjustment module and remain there even in case of power failure. From there, they can be written into one or more sen­sors or kept as backup for a possible electronics exchange.
Note:
Before the data are stored in the sensor, a check is carried out to determine if the data t the sensor. If the data do not t, a fault signal is triggered or the function is blocked. When data are being written into the sensor, the display shows which instrument type the data originate from and which TAG-no. this sensor had.
In this menu item you can select the type and size of your probe from a list of all possible probes. This is necessary to adapt the electronics
optimally to the probe.
Additional adjustments ­Special parameters
50
42279-EN-130612
In this menu item you gain access to the protected area where
you can enter special parameters. In exceptional cases, individual parameters can be modied in order to adapt the sensor to special
requirements.
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
6 Set up with the display and adjustment module
Change the settings of the special parameters only after having con­tacted our service sta.
Additional adjustments ­HART mode
o - Instrument name
Inf
The sensor oers the HART modes "Analogue current output" and
"Fix current (4 mA)". In this menu item you determine the HART mode
and enter the address with Multidrop mode.
In the mode "Fixed current output" up to 63 sensors can be operated
on one two-wire cable (Multidrop operation). An address between 0 and 63 must be assigned to each sensor.
If you select the function "Analogue current output" and also enter an address number, you can output a 4 … 20 mA signal in Multidrop
mode.
With the mode "
Fixed current (4 mA)" a xed 4 mA signal is outputted
independently of the actual level.
The default setting is "Analogue current output" and the address 00.
In this menu, you read out the instrument name and the instrument
serial number:
Info - Instrument version
In this menu item, the hardware and software version of the sensor is
displayed.
Info - Factory calibration date
In this menu item, the date of factory calibration of the sensor as well as the date of the last change of sensor parameters are displayed via the display and adjustment module or via the PC.
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
51
6 Set up with the display and adjustment module
Info - Sensor character­istics
In this menu item, the features of the sensor such as approval, pro­cess tting, seal, measuring range, electronics, housing and others
are displayed.

6.5 Saving the parameter adjustment data

We recommended noting the adjusted data, e.g. in this operating instructions manual, and archiving them afterwards. They are thus available for multiple use or service purposes.
If the instrument is equipped with a display and adjustment module, the data in the sensor can be saved in the display and adjustment
module. The procedure is described in the operating instructions
manual "Display and adjustment module" in the menu item "Copy sensor data". The data remain there permanently even if the sensor
power supply fails. The following data or settings for adjustment of the display and ad-
justment module are saved:
All data of the menu "Setup" and "Display"
In the menu "Additional adjustments" the items "Sensor-specic
units, temperature unit and linearization" The values of the user programmable linearization curve
The function can also be used to transfer settings from one instru­ment to another instrument of the same type. If it is necessary to exchange a sensor, the display and adjustment module is inserted
into the replacement instrument and the data are likewise written into
the sensor via the menu item "Copy sensor data".
52
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
2
Via the interface adapter directly on the sensor
Via the inter
face adapter
and HART

7 Setup with PACTware

7 Setup with PACTware

7.1 Connect the PC

1
3
Fig. 26: Connection of the PC directly to the sensor via the interface adapter
1 USB cable to the PC 2 Interface adapter VEGACONNECT 3 Sensor
2 4
Fig. 27: Connecting the PC via HART to the signal cable
1 Sensor
2 HARTresistance250Ω(optionaldependingonprocessing)
3 Connection cable with 2 mm pins and terminals 4 Processing system/PLC/Voltage supply 5 Interface adapter, for example VEGACONNECT 4
Note:
With power supply units with integrated HART resistance (internal
resistance approx. 250 Ω), an additional external resistance is not necessary. This applies, e.g. to the VEGA instruments VEGATRENN 149A, VEGAMET 381, VEGAMET 391. Common Ex separators are also usually equipped with a sucient current limitation resistance. In
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
1
LOCK
TWIST
3
USB
OPEN
5
53
7 Setup with PACTware
Prerequisites
such cases, the interface converter can be connected parallel to the 4 … 20 mA cable (dashed line in the previous illustration).

7.2 Parameter adjustment with PACTware

For parameter adjustment of the sensor via a Windows PC, the con­guration software PACTware and a suitable instrument driver (DTM)
according to FDT standard are required. The up-to-date PACTware
version as well as all available DTMs are compiled in a DTM Collec­tion. The DTMs can also be integrated into other frame applications
according to FDT standard.
Note:
To ensure that all instrument functions are supported, you should always use the latest DTM Collection. Furthermore, not all described functions are included in older rmware versions. You can download the latest instrument software from our homepage. A description of the update procedure is also available in the Internet.
Further setup steps are described in the operating instructions manu-
al "DTM Collection/PACTware" attached to each DTM Collection and which can also be downloaded from the Internet. Detailed descrip­tions are available in the online help of PACTware and the DTMs.
Standard/Full version
54
Fig. 28: Example of a DTM view
All device DTMs are available as a free-of-charge standard version and as a full version that must be purchased. In the standard version, all functions for complete setup are already included. An assistant for simple project conguration simplies the adjustment considerably. Saving/printing the project as well as import/export functions are also part of the standard version.
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
42279-EN-130612
General information
7 Setup with PACTware
In the full version there is also an extended print function for complete project documentation as well as a save function for measured value and echo curves. In addition, there is a tank calculation program as well as a multiviewer for display and analysis of the saved measured value and echo curves.
The standard version is available as a download under www.vega. com/downloads and "Software". The full version is available on CD from the agency serving you.

7.3 Set up with the quick setup

The quick setup is another option for parameter adjustment of the
sensor. It allows fast, convenient adjustment of the most important
parameters to adapt the sensor quickly to standard applications. To
use it, select the function "Quick setup" in the start screen.
Fig. 29: Select quick setup
1 Quick setup 2 Extended adjustment 3 Maintenance
Quick setup
With quick setup you can carry out the parameter adjustment of VEGAFLEX 81 for your application in just a few simple steps. The assistant-driven adjustment includes the basic settings for simple,
reliable setup and commissioning.
Information:
If the function is inactive, then possibly no instrument is connected.
Check the connection to the instrument.
Extended adjustment
With the extended adjustment, you carry out the parameter adjust­ment for the instrument via the clear menu structure in the DTM
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
55
7 Setup with PACTware
(Device Type Manager). This enables additional and special settings over and above those oered by quick setup.
Maintenance
Under the menu item "Maintenance" you get comprehensive and important support for servicing and maintenance. You can call up diagnostic functions and carry out an electronics exchange or a software update.
Start quick setup
Step 1
Instrumentconguration
Step 2 Application
Click to the button "Quick setup", to start the assistant-driven adjust­ment for a simplied and reliable setup.
Device name
Here, you can nd the instrument name. You cannot change this line because the instrument name is unmodiably saved in the instrument.
Serial number
Here, you can nd the serial number of your instrument. You cannot change this line because the serial number is unmodiably saved in
the instrument.
Measurement loop name
Here you can enter a suitable measurement loop name for your VE­GAFLEX 81. You can enter a name with max. 19 characters. You can use capital and small letters as well as numbers. The following special characters are also possible: + - . : , ( ) / < >
Probelengthmodied?
If you have modied the probe length, this must be entered in the selection eld.
If you select "No", then the instrument uses automatically the
preset length of the default setting. If you select "Yes ", then you can enter in another eld the modied
length of the instrument.
Probe length L from seal surface
If you have modied the length of the probe, you can enter in this eld the modied probe length. Keep the selected unit in mind.
Determine probe length automatically
If you do not know the probe length, you can have the length of the probe determined automatically. The requirement for this is a probe unrestricted and not covered by the medium.
Click to "Carry out now", to start the automatic length determination.
Medium type
Here you can see which medium type your instrument is suitable for. If this function is inactive, the medium type your instrument is suitable for was already preset.
Information:
In special cases you can change the medium type. This setting can be
changed under "Extended adjustment".
42279-EN-130612
56
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
7 Setup with PACTware
Application
In this eld you can select the application you want to use your instru­ment for. You have the following selection options:
Level in the vessel
Level in the bypass/standpipe
Interface in the vessel
Interface in the bypass/standpipe
Demonstration mode
Level measurement: If you select "Level", you can select the proper­ties of the medium in another eld.
Interface measurement: If you select "Interface", the instrument needs more information, such as the distance to the interface, the dielectric constant of the upper medium or whether or not there is a superim-
posed gas phase.
Demonstration mode: This mode is only suitable for test and demon­stration purposes. In this mode, the sensor ignores all parameters and reacts immediately to all measured value changes within the
measuring range.
Application - Level measurement
The level measurement refers to the product surface which is the limit
to the gas phase.
Liquids
Solvents, oils, LPG - dielectric constant < 3Chemical mixtures - dielectric constant 3 … 10Water-based - dielectric constant > 10
Bulk solids
Dusts, wood chips - dielectric constant < 1.5Granules, dusts, powders - dielectric constant 1.5 … 3Cereals, our - dielectric constant > 3
Application - Interface measurement
The interface measurement refers to the phase limit between two liquids. The total level is also available as a measured value.
Superimposed gas phase present
Check if there is a superimposed gas phase in the vessel. This
is always the case if the total level never touches the process tting.
Properties
Here you can enter the dielectric constant of the upper mediumAs an alternative you can enter the distance to the interface
Step 3 Adjustment
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
Adjustment for the level measurement
If you have selected level measurement in the previous menu, then you can enter the values for the min. and max. adjustment. The value to be entered refers to the distance from the sealing surface of the process tting (sensor reference plane) to the surface of the product.
Adjustment for the level and interface measurement
If you have selected interface measurement in the previous menu, then you can enter the values for the min. and max. adjustment of
57
7 Setup with PACTware
level and interface or accept the values of the level measurement. The entered value refers to the distance from the sealing surface of the process tting (sensor reference plane) to the total level or interface.
Step 4 Linearization
Step 5 Setting - Output signal
Step 6 Sensor optimiz
ation
Linearization is required if the measured value should be outputted in proportion to the volume and not the level. The linearization acts identically on the level and the interface measurement. You can nd further linearization types in the extended adjustment.
If you have a non-linear vessel, you can select here the respective linearization curve.
Linear
Spherical tank
Horizontal cylindrical tank
You must enter the following vessel dimensions with non-linear condi­tions:
Height of the socket h
Vessel height D
In this window you can adjust the output signal. When the function is inactive, you can change the settings via the "Extended adjustment".
These settings allow you to optimize the sensor. With them you can
compare a sounded distance with the indicated value and correct it, if
necessary.
Probe immersed in the liquid (covered)
Select whether or not the probe is immersed in the medium.
Measured distance to the medium
If the probe is immersed in the medium, you can enter here the meas­ured distance to the medium.
Displayeddistancecorrect?
Is the displayed distance value correct? If you have the possibility, you
can enter here the sounded distance to the medium.
False signal suppression
With this function you can carry out the automatic false signal sup­pression. We recommend carrying out false signal suppression in any
case.
Step 7 Additional settings
58
When the setup of the instrument is nished, additional settings can
be made. These are various backups and the locking of the instru­ment against unauthorised or inadvertent adjustment.
Prepareabackupleoftheinstrumentparameteradjustment?
For backup purposes, the current parameter adjustment of the instru­ment is stored in a le. You can use this le later on to restore the
instrument parameter adjustment. The complete data set is down-
loaded from the device. This procedure can last several minutes.
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
42279-EN-130612
7 Setup with PACTware
Createinstrumentdocumentation?
This function is used to print or create a PDF le of the current param­eter adjustment. To read the PDF le, you need a suitable program (for example Acrobat Reader). To print or create the PDF le, all data are downloaded from the device. For this function, the full version of the DTM Collection is required. This procedure can last several
minutes.
Storeechocurveofthesetupinthesensor?
Have you completed the initial setup of the instrument? In such case, we recommend storing the current signal conditions in the device for
later instrument tests and diagnostics.
LockadjustmentwithPINaftersetting?
The instrument is locked with the current PIN. A parameter adjust-
ment is possible only after the PIN is entered again.

7.4 Saving the parameter adjustment data

We recommend documenting or saving the parameter adjustment data via PACTware. That way the data are available for multiple use or service purposes.
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
59

8 Set up with other systems

8 Set up with other systems

8.1 DD adjustment programs

Device descriptions as Enhanced Device Description (EDD) are available for DD adjustment programs such as, for example, AMS™
and PDM.
The les can be downloaded at www.vega.com/downloads under "Software".

8.2 Field Communicator 375, 475

Device descriptions for the instrument are available as EDD for pa­rameter adjustment with the Field Communicator 375 or 475.
60
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire

9 Diagnostics and service

9 Diagnostics and service

9.1 Maintenance

If the device is used correctly, no maintenance is required in normal
operation.

9.2 Diagnosis memory

The instrument has several memories which are available for diagno­sis purposes. The data remain even with voltage interruption.
Measured value memory
Ev
ent memory
Up to 100,000 measured values can be stored in the sensor in a ring memory. Each entry contains date/time as well as the respective measured value. Storable values are for example:
Distance
Filling height
Percentage value
Lin. percent
Scaled
Current value
Meas. reliability
Electronics temperature
When the instrument is shipped, the measured value memory is active and stores distance, measurement certainty and electronics temperature every 3 minutes.
In "Extended adjustment" you can select the respective measured values.
The requested values and recording conditions are set via a PC with
PACTware/DTM or the control system with EDD. Data are thus read out and also reset.
Up to 500 events are automatically stored with a time stamp in the sensor (non-deletable). Each entry contains date/time, event type, event description and value. Event types are for example:
Modication of a parameter
Switch-on and switch-o times
Status messages (according to NE 107)
Error messages (according to NE 107)
The data are read out via a PC with PACTware/DTM or the control
system with EDD.
Echo curve memory
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
The echo curves are stored with date and time and the corresponding
echo data. The memory is divided into two sections: Echo curve of the setup: This is used as reference echo curve for
the measurement conditions during setup. Changes in the measure­ment conditions during operation or buildup on the sensor can thus
be recognized. The echo curve of the setup is stored via:
PC with PACTware/DTM
Control system with EDD
61
9 Diagnostics and service
Display and adjustment module
Further echo curves: Up to 10 echo curves can be stored in a ring buer in this memory section. Further echo curves are stored via:
PC with PACTware/DTM
Control system with EDD
Display and adjustment module

9.3 Status messages

The instrument features self-monitoring and diagnostics according to NE 107 and VDI/VDE 2650. In addition to the status messages in the following tables there are more detailed error messages available under the menu item "Diagnostics" via the display and adjustment module, PACTware/DTM and EDD.
Status messages
The status messages are divided into the following categories:
Failure
Function check
Out of specication
Maintenance requirement
and explained by pictographs:
41 2 3
Fig. 30: Pictographs of the status messages
1 Failure - red
2 Outofspecication-yellow
3 Function check - orange 4 Maintenance - blue
Failure: Due to a malfunction in the instrument, a failure message is
outputted.
This status message is always active. It cannot be deactivated by the
user.
Function check: The instrument is in operation, the measured value is temporarily invalid (for example during simulation).
This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD.
Outofspecication: The measured value is unstable because the instrument specication is exceeded (e.g. electronics temperature).
This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD.
Maintenance: Due to external inuences, the instrument function is limited. The measurement is aected, but the measured value is
42279-EN-130612
62
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
9 Diagnostics and service
still valid. Plan in maintenance for the instrument because a failure is expected in the near future (e.g. due to buildup).
This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD.
Failure
The following table shows the error codes in the status message
"Failure" and gives information on the reason and rectication. Keep in mind that some information is only valid with four-wire instruments.
Code
Text mes­sage
F013
no measured value avail­able
F017
Adjustment span too small
F025
Error in the linearization table
F036
No operable
software
F040
Error in the electronics
F041
Probe loss
F080 General soft-
ware error
F105
Measured value is deter­mined
F113
Communica­tion error
F125
Unpermissi­ble electronics temperature
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
Cause Rectication
– Sensor does not detect an
echo during operation
– Process component or
probe contaminated or
defective
– Adjustment not within
specication
– Index markers are not con-
tinuously rising, for examle unlogical value pairs
Failed or interrupted soft-
ware update
Hardware defect Exchanging the electronics
– Cable probe broken or rod
probe defective
General software error Separate operating voltage
– The instrument is still in the
start phase, the measured value could not yet be
determined
EMC interferences – Transmission error with the
external communication
with 4-wire power supply unit
Temperature of the elec-
tronics in the non-specied
section
– Check or correct installation
and/or parameter adjust­ment
Clean or exchange process
component or probe
– Change adjustment accord-
ing to the limit values (dif­ference between min. and max. ≥ 10 mm)
Check values of the lineari-
zation table
– Delete/create a new lineari-
zation table
Repeat software updateCheck electronics versionExchanging the electronicsSend instrument for repair
Send instrument for repair
Check probe and exchange,
if necessary
briey
Wait for the end of the
switch-on phase
– Duration depending on
the version and parameter adjustment max. 5 min.
Remove EMC inuencesExchange 4-wire power
supply unit or electronics
– Check ambient temperature – Isolate electronics – Use instrument with higher
temperature range
63
9 Diagnostics and service
Function check
Code
Text mes­sage
F260
Error in the calibration
F261
Error in the instrument settings
F264
Installation/ Setup error
F265
Measurement function dis­turbed
F266
Impermissi­ble operating
voltage
F267
No executable sensor soft-
ware
Cause Rectication
– Error in the calibration car-
ried out in the factory
– Error in the EEPROM
– Error during setup – Error when carrying out a
reset
– False signal suppression
faulty
– Error during setup – Check or correct installation
– Sensor no longer carries
out a measurement
Operating voltage below
specied range
– Sensor cannot start Exchanging the electronics
Exchanging the electronicsSend instrument for repair
– Carry out a reset – Repeat setup
and/or parameter adjust­ment
– Check probe length
– Carry out a reset – Separate operating voltage
briey
– Check electrical connection – if necessary, increase
operating voltage
– Send instrument for repair
The following table shows the error codes and text messages in the
status message "Function check" and provides information on causes as well as corrective measures.
Code
Text mes­sage
C700
Simulation ac-
tive
Cause Rectication
A simulation is active – Finish simulation
Wait for the automatic end
after 60 mins.
Outofspecication
64
The following table shows the error codes and text messages in the
status message "Outofspecication" and provides information on causes as well as corrective measures.
Code
Text mes­sage
S600
Unpermissi­ble electronics temperature
Cause Rectication
Temperature of the pro-
cessing electronics in the
non-specied section
– Check ambient temperature – Isolate electronics – Use instrument with higher
temperature range
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
42279-EN-130612
9 Diagnostics and service
Maintenance
Code
Text mes­sage
S601
Overlling
S602 Level with-
in the search range, com­pensation echo
S603
Impermissi­ble operating
voltage
Cause Rectication
– Level echo in the close
range not available
– Compensation echo super-
imposed by medium
Operating voltage below
specied range
Reduce level100 % adjustment: Increase
value
– Check mounting socket – Remove possible interfering
signals in the close range
Use coaxial probe
100 % adjustment: Increase
value
– Check electrical connection – if necessary, increase
operating voltage
The following table shows the error codes and text messages in the
status message "Maintenance" and provides information on causes as well as corrective measures.
Code
Text mes­sage
M500
Error in the delivery sta­tus
M501
Error in the
non-active
linearization table
M502
Error in the event mem­ory
M503
Reliability too low
M504
Error on an
device inter­face
Cause Rectication
– With the reset to delivery
status, the data could not
be restored
– Hardware error EEPROM Exchanging the electronics
– Hardware error EEPROM Exchanging the electronics
– Measurement certainty is
too low for a reliable meas­urement
– Process component or
probe contaminated or
defective
Hardware defect Exchanging the electronics
– Repeat reset – Load XML le with sensor
data into the sensor
Send instrument for repair
Send instrument for repair
– Check installation and
process conditions
Clean or exchange process
component or probe
Send instrument for repair
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
65
9 Diagnostics and service
Reaction when malfunc­tions occur
Procedure for fault recti-
cation
Code
Text mes­sage
M505
no measured value avail­able
M506
Installation/ Setup error
M507
Error in the instrument settings
Cause Rectication
– Sensor does not detect an
echo during operation
– Process component or
probe contaminated or
defective
– Error during setup – Check and correct instal-
– Error during setup – Error when carrying out a
reset
– False signal suppression
faulty
– Check and correct instal-
lation and/or parameter adjustment
Clean or exchange process
component or probe
lation and/or parameter adjustment
– Check probe length
– Carry out reset and repeat
setup

9.4 Rectify faults

The operator of the system is responsible for taking suitable meas-
ures to rectify faults.
The rst measures are:
Evaluation of fault messages, for example via the display and
adjustment module Checking the output signal
Treatment of measurement errors
Further comprehensive diagnostics options oer a PC with the soft­ware PACTware and the suitable DTM. In many cases, the reasons can be determined in this way and faults can be rectied.
Check the 4 … 20 mA signal
66
Connect a multimeter in the suitable measuring range according to
the wiring plan. The following table describes possible errors in the current signal and helps to remove them:
Error Cause Rectication
4 … 20 mA signal not stable
Level uctua-
tions
– Set damping according to the
instrument via the display and
adjustment module or PACTware/ DTM
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
42279-EN-130612
9 Diagnostics and service
Error Cause Rectication
4 … 20 mA signal missing
Current sig­nal greater than 22 mA or less than 3.6 mA
– Electrical con-
nection faulty
– Voltage supply
missing
Operating volt-
age too low or load resistance too high
– Electronics
module in the
sensor defec­tive
– Check connection according to
chapter "Connection steps" and if necessary, correct according to chapter "Wiring plan"
Check cables for breaks; repair if
necessary
Check, adapt if necessary
Exchange the instrument or send it
in for repair
Treatment of measure­ment errors
The below tables show typical examples for application-relevant
measurement errors. There are two measurement errors:
Constant level
Filling
Emptying
The images in column "Error pattern" show the real level with a bro­ken line and the level displayed by the sensor as a continuous line.
Level
1
2
0
Fig. 31: The broken line 1 shows the real level, the continuous line 2 shows the level displayed by the sensor
Note:
Wherever the sensor displays a constant value, the reason could
also be the fault setting of the current output to "Hold value" In case of a too low level indication, the reason could be a line
resistance that is too high
time
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
67
9 Diagnostics and service
time
0
time
time
time
time
0
Measurement error with constant level
Fault description Error pattern Cause Rectication
1. Measured value
shows a too low or too
high level
Level
Min./max. adjustment not
correct
Wrong linearization curve Adapt linearization curve
– Running time error (small
Adapt min./max. adjustment
– Repeat setup
measurement error close to
100 %/serious error close to 0 %)
2. Measured value jumps towards 100 %
Level
– Due to the process, the ampli-
tude of the product echo sinks
Carry out a false signal sup-
pression
A false signal suppression was
0
not carried out
Amplitude or position of a
false signal has changed (e.g. buildup); false signal suppres-
sion no longer matches
Determine the reason for the
changed false signals, carry out false signal suppression, e.g.
with buildup
Measurementerrorduringlling
Fault description Error pattern Cause Rectication
3. Measured value re­mains in the area of the bottom during lling
4. Measured value re-
mains momentarily unchanged during ll­ing and then jumps to
the correct level
5. Measured value
jumps sporadically to
100 % during lling
Level
– Echo from the probe end larger
than the product echo, for example, with products with
0
Level
0
Level
εr < 2.5 oil-based, solvents, etc.
– Turbulence on the product
surface, quick lling
– Changing condensation or
contamination on the probe
Check parameter "Medium"
and "Vessel height", adapt if
necessary
Check parameters, change if
necessary, e.g. in dosing ves­sel, reactor
Carry out a false signal sup-
pression
6. Measured value jumps to ≥ 100 % or
0 m distance
Level
Level echo is no longer
detected in the close range
due to false signals in the close
range. The sensor goes into
overll protection mode. The
Eliminate false signals in the
close range
– Check installation conditions – If possible, switch o the func-
tion "Overll protection"
max. level (0 m distance) as
well as the status message "Overll protection" are output­ted.
42279-EN-130612
68
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
9 Diagnostics and service
time
0
time
0
Measurement error during emptying
Fault description Error pattern Cause Rectication
7. Measured value re­mains unchanged in the close range during emptying
8. Measured value re­mains reproducible in one position during emptying
Level
Level
False echo larger than the level
echo
Level echo too small
Stored false signals in this
position are larger than the
level echo
Eliminate false signals in the
close range
Remove contamination on the
probe. After having removed the source of the false signals, the false signal suppression
must be deleted.
Carry out a new false signal
suppression
Delete false signal memoryCarry out a new false signal
suppression
Reaction after fault recti-
cation
24 hour service hotline
Depending on the reason for the fault and the measures taken, the steps described in chapter "Setup" must be carried out again or must be checked for plausibility and completeness.
Should these measures not be successful, please call in urgent cases the VEGA service hotline under the phone no.
+49 1805 858550.
The hotline is also available outside normal working hours, seven
days a week around the clock.
Since we oer this service worldwide, the support is provided in English. The service itself is free of charge, the only costs involved are
the normal call charges.

9.5 Exchanging the electronics module

If the electronics module is defective, it can be replaced by the user. In Ex applications only one instrument and one electronics module
with respective Ex approval may be used. If there is no electronics module available on site, the electronics
module can be ordered through the agency serving you. The electron­ics modules are adapted to the respective sensor and dier in signal output or voltage supply.
The new electronics module must be loaded with the default settings of the sensor. These are the options:
In the factory
Or on site by the user
In both cases, the serial number of the sensor is needed. The serial numbers are stated on the type label of the instrument, on the inside of the housing as well as on the delivery note.
When loading on site, rst of all the order data must be downloaded from the Internet (see operating instructions manual "Electronics module").
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
69
9 Diagnostics and service

9.6 Software update

The following components are required to update the sensor soft­ware:
Sensor
Voltage supply
Interface adapter VEGACONNECT
PC with PACTware
Current sensor software as le
You can nd the actual sensor software as well as detailed infor­mation of the procedure under "www.vega.com/downloads" and "Software".
You can nd information about the installation in the download le.
Caution:
Instruments with approvals can be bound to certain software ver­sions. Therefore make sure that the approval remains eective with a software update.
You can nd detailed information on www.vega.com/downloads and "Approvals".

9.7 How to proceed in case of repair

You can nd a repair form as well as detailed information on how to
proceed under www.vega.com/downloads and "Formsandcerti- cates".
By doing this you help us carry out the repair quickly and without hav­ing to call back for needed information.
If a repair is necessary, please proceed as follows:
Print and ll out one form per instrument
Clean the instrument and pack it damage-proof
Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
Please contact for the return shipment the agency serving you. You
can nd the agency on our home page www.vega.com.
70
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire

10 Dismounting

10 Dismounting

10.1 Dismounting steps

Warning:
Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel, high temperatures, corrosive or toxic
products etc.
Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order.

10.2 Disposal

The instrument consists of materials which can be recycled by spe­cialised recycling companies. We use recyclable materials and have
designed the parts to be easily separable.
Correct disposal avoids negative eects on humans and the environ­ment and ensures recycling of useful raw materials.
Materials: see chapter "Technical data" If you have no way to dispose of the old instrument properly, please
contact us concerning return and disposal.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/EG and the respective national laws. Pass the instrument directly on to a spe-
cialised recycling company and do not use the municipal collecting
points. These may be used only for privately used products according to the WEEE directive.
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
71

11 Supplement

11 Supplement

11.1 Technical data

General data
316L corresponds to 1.4404 or 1.4435 Materials, wetted parts
Ʋ Process tting 316L and Aluminium oxide-ceramic 99.7 % (Al2O3),
Ʋ Process seal on the instrument side
(rod leadthrough)
Ʋ Process tting - for ammonia applica-
tions
Ʋ Process seal on the instrument side -
for ammonia applications
Ʋ Process seal On site (instruments with thread: Klingersil C-4400 is
Ʋ Inner conductor (up to the separation
rod)
Ʋ Tube: ø 21.3 mm (0.839 in) 316L or Hastelloy C22 (2.4602) Ʋ Tube: ø 42.2 mm (1.661 in) 316L or Hastelloy C22 (2.4602)
Materials, non-wetted parts
Ʋ Plastic housing plastic PBT (Polyester) Ʋ Aluminium die-casting housing Aluminium die-casting AlSi10Mg, powder-coated - basis:
Ʋ Stainless steel housing - precision
casting
Ʋ Stainless steel housing, electropol-
ished
Ʋ Second line of defence (optional) Borosilicate glass GPC 540 Ʋ Seal between housing and housing
cover
Ʋ Inspection window in housing cover
(optional)
Ʋ Ground terminal 316L
Ohmic contact Between ground terminal, process tting and probe
Process ttings - tube: ø 21.3 mm (0.839 in)
Ʋ Pipe thread, cylindrical (ISO 228 T1) G¾, G1, G1½ according to DIN 3852-A Ʋ American pipe thread, conical
(ASME B1.20.1)
Ʋ Flanges e.g. DIN from DN 25, ANSI from 1"
Process ttings - tube: ø 42.2 mm (1.661 in)
Hastelloy C22 (2.4602) and Aluminium oxide-ceramic
99.7 % (Al2O3) FKM (SHS FPM 70C3 GLT), FFKM (Kalrez 6375),
EPDM (A+P 75.5/KW75F)
316L
Borosilicate glass GPC 540
1)
attached)
316L
Polyester
316L
316L
NBR (stainless steel housing, precision casting), silicone (aluminium/plastic housing; stainless steel housng,
electropolished)
Polycarbonate (with Ex d version: glass)
¾ NPT, 1 NPT, 1½ NPT
42279-EN-130612
1)
Not suitable for steam applications
72
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
Ʋ Pipe thread, cylindrical (ISO 228 T1) G1½ according to DIN 3852-A Ʋ American pipe thread, conical
1½ NPT
(ASME B1.20.1)
Ʋ Flanges DIN from DN 50, ANSI from 2"
Weight
Ʋ Instrument weight (depending on
approx. 0.8 … 8 kg (0.176 … 17.64 lbs)
process tting)
Ʋ Tube: ø 21.3 mm (0.839 in) approx. 1110 g/m (11.9 oz/ft) Ʋ Tube: ø 42.2 mm (1.661 in) approx. 3100 g/m (33.3 oz/ft)
Probe length L (from seal surface)
Ʋ Tube: ø 21.3 mm (0.839 in) up to 6 m (19.69 ft) Ʋ Tube: ø 42.2 mm (1.661 in) up to 6 m (19.69 ft) Ʋ Trimming accuracy - tube ±1 mm
Lateral load
Ʋ Tube: ø 21.3 mm (0.839 in) 60 Nm (44 lbf ft) Ʋ Tube: ø 42.2 mm (1.661 in) 300 Nm (221 lbf ft)
Torque for NPT cable glands and Conduit tubes
Ʋ Plastic housing max. 10 Nm (7.376 lbf ft) Ʋ Aluminium/Stainless steel housing max. 50 Nm (36.88 lbf ft)
Input variable
Measured variable Level of liquids Min. dielectric constant of the medium ε
≥ 1.4
r
11 Supplement
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
73
11 Supplement
1
4
2
2
3
5
Fig. 40: Measuring ranges - VEGAFLEX 81
1 Reference plane 2 Probe length L 3 Measuring range (default setting refers to the measuring range in water) 4 Upper dead band (see diagrams under Accuracy - grey section) 5 Lower dead band (see diagrams under Accuracy - grey section)
Output variable
Output signal
Range of the output signal 3.8 … 20.5 mA/HART (default setting) Fullled HART specication 7
Signal resolution Failure signal current output (adjustable) Last valid measured value, ≥ 21 mA, ≤ 3.6 mA
Max. output current 21.5 mA
Starting current ≤ 10 mA for 5 ms after switching on, ≤ 3.6 mA
4 … 20 mA/HART
0.3 µA
1
4
3
5
42279-EN-130612
74
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
11 Supplement
Load see load under Power supply
Damping (63 % of the input variable),
0 … 999 s
adjustable
HART output values according to HART 7 (default setting)
2)
Ʋ First HART value (PV) Linearised percentage value, level (current output 1) Ʋ Second HART value (SV) Distance to level (current output 2 - if available) Ʋ Third HART value (TV) Measurement certainty, level Ʋ Fourth HART value (QV) Electronics temperature
Ex-d-ia versions are not suitable for HART multidrop operation Indication value - DIsplay and adjustment module
3)
Ʋ Displayed value 1 Filling height Level Ʋ Displayed value 2 Electronics temperature
Resolution, digital
< 1 mm (0.039 in)
Output variable - Second current output
For details on the operating voltage see chapter "Voltage supply"
Output signal
4 … 20 mA (passive)
Range of the output signal 3.8 … 20.5 mA (default setting)
Signal resolution 0.3 µA Failure signal current output (adjustable) Last valid measured value, ≥ 21 mA, ≤ 3.6 mA
Max. output current 21.5 mA
Starting current ≤ 10 mA for 5 ms after switching on, ≤ 3.6 mA Load Load resistor, see chapter "Voltage supply"
Damping (63 % of the input variable),
0 … 999 s
adjustable
Indication value - DIsplay and adjustment module
4)
Ʋ Displayed value 1 Filling height Level Ʋ Displayed value 2 Electronics temperature
Resolution, digital
< 1 mm (0.039 in)
Accuracy (according to DIN EN 60770-1)
Process reference conditions according to DIN EN 61298-1
Ʋ Temperature +18 … +30 °C (+64 … +86 °F) Ʋ Relative humidity 45 … 75 % Ʋ Air pressure +860 … +1060 mbar/+86 … +106 kPa
(+12.5 … +15.4 psig)
Installation reference conditions
Ʋ Min. distance to installations > 500 mm (19.69 in)
2)
The output values can be assigned individually
3)
The indication values can be assigned individually
4)
The indication values can be assigned individually
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
75
11 Supplement
Ʋ Vessel metallic, ø 1 m (3.281 ft), centric installation, process
tting ush with the vessel ceiling
Ʋ Medium Water/Oil (dielectric constant ~2.0)
5)
Ʋ Installation Probe end does not touch the vessel bottom
Sensor parameter adjustment No gating out of false signals carried out
Typical deviation - Interface measure-
± 5 mm (0.197 in)
ment
Typical deviation - Total level interface
± 5 mm (0.197 in)
measurement
Typical deviation - Level measurement Max. deviation - Version with reference
6)7)
See following diagrams
± 10 mm (0.39 in)
distance
Deviating upper dead band - Version with
450 mm (17.7 in)
reference distance
L
5 mm (0.197")
2 mm (0.079")
-2 mm (-0.079")
0
1
-10 mm (0.394")
0,03 m (1.181
")
0,15 m (5.906")
Fig. 41: Deviation VEGAFLEX 81 in coax version in water
1 Dead band - no measurement possible in this area L Probe length
0,05 m
(1.969
")
5)
With interface measurement = 2.0
6)
Depending on the installation conditions, there can be deviations which can be rectied with an adaptation of
the adjustment or a change of the measured value oset in the DTM service mode
7)
The dead bands can be optimizes by a false signal suppression.
76
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
42279-EN-130612
10 mm (0.394")
5 mm (0.197")
2 mm (0.079")
-2 mm (-0.079
-5 mm (-0.197")
L
0
")
1 1
11 Supplement
0,1 m (3.937
Fig. 42: Deviation VEGAFLEX 81 in coax version in oil
1 Dead band - no measurement possible in this area L Probe length
")
0,15 m (5.906")
0,1 m
(3.937
")
0,05 m (1.97
")
Repeatability ≤ ±1 mm
Variablesinuencingmeasurementaccuracy
Specicationsforthedigitalmeasuredvalue
Temperature drift - Digital output ±3 mm/10 K relating to the max. measuring range or
max. 10 mm (0.394 in)
Additional deviation through electromag-
< ±10 mm (< ±0.394 in)
netic interference acc. to EN 61326
Specicationsapplyalsotothecurrentoutput
8)
Temperature drift - Current output ±0.03 %/10 K relating to the 16 mA span max. ±0.3 % Deviation on the current output through
< ±15 µA
analogue/digital conversion Additional deviation through electromag-
< ±150 µA
netic interference acc. to EN 61326
Inuenceofthesuperimposedgasandpressuretotheaccuracy
The propagation speed of the radar impulses in gas or vapour above the medium is reduced by high pressure. This eect depends on the superimposed gas or vapour and is especially large at
low temperatures.
The following table shows the resulting deviation for some typical gases and vapours. The specied values refer to the distance. Positive values mean that the measured distance is too large, negative values that the measured distance is too small.
Gas phase Temperature Pressure
1 bar (14.5 psig) 10 bar (145 psig) 50 bar (725 psig)
Air 20 °C/68 °F 0.00 % 0.22 % 1.2 %
200 °C/392 °F -0.01 % 0.13 % 0.74 %
400 °C/752 °F -0.02 % 0.08 % 0.52 %
8)
Also for the second current output (optional)
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
77
11 Supplement
Gas phase Temperature Pressure
1 bar (14.5 psig) 10 bar (145 psig) 50 bar (725 psig)
Hydrogen 20 °C/68 °F -0.01 % 0.10 % 0.61 %
200 °C/392 °F -0.02 % 0.05 % 0.37 %
400 °C/752 °F -0.02 % 0.03 % 0.25 %
Steam (saturated steam)
100 °C/212 °F 0.26 % - -
180 °C/356 °F 0.17 % 2.1 % -
264 °C/507 °F 0.12 % 1.44 % 9.2 %
366 °C/691 °F 0.07 % 1.01 % 5.7 %
Characteristics and performance data
Measuring cycle time
Step response time
9)
< 500 ms ≤ 3 s
Max. lling/emptying speed 1 m/min
Ambient conditions
Ambient, storage and transport tempera-
-40 … +80 °C (-40 … +176 °F)
ture
Process conditions
For the process conditions, please also note the specications on the type label. The lowest value
always applies.
The measurement error from the process conditions is in the specied pressure and temperature range of below 1 %.
Process pressure
Ʋ Standard version -1 … +40 bar/-100 … +4000 kPa (-14.5 … +580 psig),
depending on the process tting
Ʋ with borosilicate glass leadthrough -1 … +100 bar/-100 … +10000 kPa
(-14.5 … +1450 psig), depending on the process tting
Vessel pressure relating to the ange
nominal stage
see supplementary instructions manual "Flanges ac­cording to DIN-EN-ASME-JIS"
Process temperature (thread or ange temperature)
Ʋ FKM (SHS FPM 70C3 GLT) -40 … +150 °C (-40 … +302 °F) Ʋ EPDM (A+P 75.5/KW75F) -40 … +150 °C (-40 … +302 °F) Ʋ FFKM (Kalrez 6375) -20 … +150 °C (-4 … +302 °F) Ʋ FFKM (Kalrez 6375) -20 … +200 °C (-4 … +392 °F) Ʋ with borosilicate glass leadthrough -60 … +150 °C (-76 … +302 °F)
9)
Time span after a sudden measuring distance change by max. 0.5 m in liquid applications, max 2 m with bulk
solids applications, until the output signal has taken for the rst time 90 % of the nal value (IEC 61298-2).
78
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
42279-EN-130612
1
80°C
(176°F)
65°C
(149°F)
11 Supplement
3 4
-40°C
(-104°F)
0°C
(32°F)
-40°C
(-104°F)
80°C
(176°F)
130°C
(266°F)
150°C
(302°F)
2
Fig. 43: Ambient temperature - process temperature, standard version
1 Ambient temperature 2 Process temperature (depending on the seal material) 3 Maximum permissible temperature - standard 4 Limited temperature range - plastic housing and stainless steel housing, electropolished
1
80°C
(176°F)
55°C
(131°F)
0°C
-20°C
(-68°F)
(32°F)
-40°C
(-104°F)
150°C
(302°F)
Fig. 44: Ambient temperature - process temperature, version with temperature adapter
1 Ambient temperature 2 Process temperature (depending on the seal material) 3 Maximum permissible temperature - standard 4 Limited temperature range - plastic housing and stainless steel housing, electropolished
200°C
(392°F)
2
3 4
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
79
11 Supplement
1
80°C
(176°F)
65°C
(149°F)
3 4
0°C
-60°C
(-140°F)
Fig. 45: Ambient temperature - process temperature, elastomer-free version for ammonia applications
1 Ambient temperature 2 Process temperature (depending on the seal material) 3 Maximum permissible temperature - standard 4 Limited temperature range - plastic housing and stainless steel housing, electropolished
(32°F)
-40°C
(-104°F)
130°C
(266°F)
150°C
(302°F)
2
Viscosity - dynamic 0.1 … 500 mPa s (requirement: with density 1)
Vibration resistance
Ʋ Instrument housing 4 g at 5 … 200 Hz according to EN 60068-2-6 (vibration
with resonance)
Ʋ Coax probe 1 g with 5 … 200 Hz according EN 60068-2-6 (vibration
at resonance) with tube length 50 cm (19.69 in)
Shock resistance
Ʋ Instrument housing 100 g, 6 ms according to EN 60068-2-27 (mechanical
shock)
Ʋ Coax probe 25 g, 6 ms according to EN 60068-2-27 (mechanical
shock) with tube length 50 cm (19.69 in)
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Options of the cable entry
Ʋ Cable gland M20 x 1.5 Ʋ Cable diameter (options) 5 … 9 mm (0.20 … 0.35 in)
6 … 12 mm (0.24 … 0.47 in)
10 … 14 mm (0.39 … 0.55 in)
Ʋ Cable entry ½ NPT Ʋ Blind plug M20 x 1.5; ½ NPT Ʋ Closing cap M20 x 1.5; ½ NPT
Plug options
Ʋ Signal circuit Plug M12 x 1, according to ISO 4400, Harting HAN,
7/8" FF
Ʋ Indication circuit M12 x 1 plug
Wire cross-section (spring-loaded terminals)
Ʋ Massive wire, cord 0.2 … 2.5 mm² (AWG 24 … 14)
42279-EN-130612
80
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
Ʋ Stranded wire with end sleeve 0.2 … 1.5 mm² (AWG 24 … 16)
Electromechanical data - version IP 66/IP 68 (1 bar)
Options of the cable entry
Ʋ Cable gland M20 x 1.5 Ʋ Cable diameter (options) 5 … 9 mm (0.20 … 0.35 in)
6 … 12 mm (0.24 … 0.47 in)
10 … 14 mm (0.39 … 0.55 in)
Ʋ Cable entry ½ NPT Ʋ Blind plug M20 x 1.5; ½ NPT
Connection cable
Ʋ Wire cross-section 0.5 mm² (AWG 20) Ʋ Wire resistance < 0.036 Ω/m Ʋ Tensile strength < 1200 N (270 lbf) Ʋ Standard length 5 m (16.4 ft) Ʋ Max. length 180 m (590.6 ft) Ʋ Min. bending radius 25 mm (0.984 in) with 25 °C (77 °F) Ʋ Diameter approx. 8 mm (0.315 in) Ʋ Colour - Non-Ex version Black Ʋ Colour - Ex-version Blue
Display and adjustment module
Display element Display with backlight
Measured value indication
Ʋ Number of digits 5 Ʋ Size of digits W x H = 7 x 13 mm
Adjustment elements 4 keys
Protection rating
Ʋ unassembled IP 20 Ʋ mounted into the housing without
IP 40
cover
Materials
Ʋ Housing ABS Ʋ Inspection window Polyester foil
11 Supplement
Integrated clock
Date format Day.Month.Year Time format
12 h/24 h
Time zone Ex factory CET
Measurement electronics temerature
Resolution
1 °C (1.8 °F)
Accuracy ±1 °C (1.8 °F)
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
81
11 Supplement
Voltage supply
Operating voltage
Ʋ Non-Ex instrument, Ex-d instrument 9.6 … 35 V DC Ʋ Ex-ia instrument 9.6 … 30 V DC Ʋ Ex-d-ia instrument 15 … 35 V DC
Operating voltage with illuminated display and adjustment module
Ʋ Non-Ex instrument, Ex-d instrument 16 … 35 V DC Ʋ Ex-ia instrument 16 … 30 V DC Ʋ Ex-d-ia instrument 20 … 35 V DC
Interpolation protection Integrated
Permissible residual ripple - Non-Ex, Ex-ia instrument
Ʋ for 9.6 V< UN < 14 V ≤ 0.7 Ve (16 … 400 Hz) Ʋ for 18 V< UN < 36 V ≤ 1 Ve (16 … 400 Hz)
Permissible residual ripple - Ex-d-ia instrument
Ʋ for 18 V< UN < 35 V ≤ 1 Ve (16 … 400 Hz)
Load resistor
Ʋ Calculation (UB - U Ʋ Example - Non-Ex instrument with
(24 V - 9.6 V)/0.0215 A = 670 Ω
)/0.0215 A
min
UB= 24 V DC
Electrical protective measures
Protection, depending on housing version
Ʋ Plastic housing IP 66/IP 67 Ʋ Aluminium housing, stainless steel
IP 66/IP 68 (0.2 bar)
10)
housing - investment casting, stain­less steel housing - electro-polished
Ʋ Aluminium and stainless housing, in-
IP 66/IP 68 (1 bar)
vestment casting (optionally available)
Overvoltage category III
Protection class III
Approvals
Instruments with approvals can have dierent technical data depending on the version. For that reason the associated approval documents of these instruments must be carefully noted.
They are part of the delivery or can be downloaded under www.vega.com and "VEGA Tools" as well as under "Downloads" and "Approvals".

11.2 Dimensions

The following dimensional drawings represent only an extract of all possible versions. Detailed
dimensional drawings can be downloaded at www.vega.com/downloads under "Drawings".
10)
The prerequisites for maintaining the protection rating are a suitable cable as well as correct mounting.
82
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
42279-EN-130612
Plastic housing
~ 69 mm
(2.72")
ø 79 mm
(3.11")
~ 84 mm
(3.31")
M16x1,5
11 Supplement
ø 79 mm
(3.11")
(4.41")
112 mm
(4.41")
112 mm
M20x1,5/ ½ NPT
M20x1,5/
1
Fig. 46: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated display and adjustment module the housing is 9 mm/0.35 in higher
1 Single chamber version 2 Double chamber version
½ NPT
2
Aluminium housing
M20x1,5/ ½ NPT
~ 116 mm
(4.57")
ø 86 mm
(3.39")
M20x1,5
(4.57")
116 mm
~ 87 mm
M16x1,5
M20x1,5/ ½ NPT
(3.43")
ø 86 mm
(3.39")
21
(4.72")
120 mm
Fig. 47: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated display and adjustment module the housing is 9 mm/0.35 in higher
1 Single chamber version 2 Double chamber version
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
83
11 Supplement
Aluminium housing in protection rating IP 66/IP 68 (1 bar)
~ 150 mm
(5.91")
ø 84 mm
(3.31")
(4.57")
116 mm
M20x1,5
M20x1,5
1
Fig.48:HousingversionsinprotectionIP␣66/IP␣68(1␣bar)-withintegrateddisplayandadjustmentmodulethe housingis9␣mm/0.35␣inhigher
1 Single chamber version 2 Double chamber version
~ 105 mm
(4.13")
M20x1,5/ ½ NPT
ø 84 mm
(3.31")
2
M16x1,5
(4.72")
120 mm
Stainless steel housing
~ 59 mm
(2.32")
ø 80 mm
(3.15")
~ 69 mm
(2.72")
ø 79 mm
(3.11")
~ 87 mm
(3.43")
M16x1,5
ø 86 mm
(3.39")
(4.41")
112 mm
(4.61")
117 mm
120 mm
M20x1,5/
½ NPT M20x1,5/ ½ NPT
Fig. 49: Housing versions in protection IP 66/IP 68 (0.2 bar) - with integrated display and adjustment module the housing is 9 mm/0.35 in higher
1 Single chamber version, electropolished 2 Single chamber version, precision casting 3 Double chamber version, precision casting
84
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
M20x1,5/ ½ NPT
321
(4.72")
42279-EN-130612
Stainless steel housing in protection rating IP 66/IP 68, 1 bar
~ 93 mm
(3.66")
ø 80 mm
(3.15")
~ 105 mm (4.13")
ø 86 mm
(3.39")
M16x1,5
~ 103 mm
(4.06")
11 Supplement
ø 79 mm
(3.11")
M20x1,5/ ½ NPT
(4.41")
112 mm
1
M20x1,5/ ½ NPT
2
(4.72")
120 mm
117 mm
M20x1,5
3
Fig.50:HousingversionsinprotectionIP␣66/IP␣68(1␣bar)-withintegrateddisplayandadjustmentmodulethe housingis9␣mm/0.35␣inhigher
1 Single chamber version, electropolished 2 Single chamber version, precision casting 3 Double chamber version, precision casting
(4.61")
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
85
11 Supplement
VEGAFLEX 81, coax version
SW 36 (1.
42")
(G¾, ¾ NPT) (G1, 1 NPT)
61")
SW 41 (1.
81")
SW 46 (1.
(G1½, 1½ NPT)
G¾, ¾ NPT, G1, 1 NPT, G1½, 1½ NPT
ø 21,3 mm
1 2
Fig. 51: VEGAFLEX 81, threaded version
L Sensor length, see chapter "Technical data" 1 Coaxial version ø 21.3 mm (0.839 in) 2 Coaxial version ø 42.2 mm (1.661 in)
(0.84")
46 mm
(1.81")
L
SW 55
(2.17")
G1½, 1½ NPT
22mm
ø 42,2 mm
(1.66")
(0.86")
51 mm
L
(2.01")
86
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
11 Supplement

11.3 Industrial property rights

VEGA product lines are global protected by industrial property rights. Further information see www.vega.com.
Only in U.S.A.: Further information see patent label at the sensor housing.
VEGA Produktfamilien sind weltweit geschützt durch gewerbliche Schutzrechte.
Nähere Informationen unter www.vega.com.
Les lignes de produits VEGA sont globalement protégées par des droits de propriété intellec­tuelle. Pour plus d'informations, on pourra se référer au site www.vega.com.
VEGA lineas de productos están protegidas por los derechos en el campo de la propiedad indus­trial. Para mayor información revise la pagina web www.vega.com.
Линии продукции фирмы ВЕГА защищаются по всему миру правами на интеллектуальную собственность. Дальнейшую информацию смотрите на сайте www.vega.com.
VEGA系列产品在全球享有知识产权保护。 进一步信息请参见网站<www.vega.com>。

11.4 Trademark

All the brands as well as trade and company names used are property of their lawful proprietor/
originator.
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
87
INDEX
INDEX
A
Accessories
– Display and adjustment module 13 – External display and adjustment unit 14 – External radio unit 14 – Flanges 14 – Interface adapter 13 – Protective cap 14
Adjustment
Max. adjustment 35, 36 – Min. adjustment 36, 37
Adjustment system 30 Application 34, 35 Application area 10
B
Backlight 44 Block adjustment 42
C
Cable entry 19 Check signal 66 Connection cable 19 Connection procedure 20 Connection technology 20 Copy sensor settings 50 Current output 2 41
Current output, adjustment 40 Current output Min./Max. 41
Current output mode 40
Current output, size 40 Curve indication
– Echo curve 45
D
Damping 37
Date of manufacture 51
Date/Time 47
Default values 47 Deviation 67
E
Echo curve memory 61 Echo curve of the setup 46 EDD (Enhanced Device Description) 60 Electronics and connection compartment,
single chamber housing 22
Electronics compartment, double chamber
housing 22
Electronics compartment with Ex-d double
chamber housing 24 Error messages 62 Event memory 61
F
Factory calibration date 51 False signal suppression 42 Fault rectication 66 Functional principle 10
G
Gas phase 35 Grounding 20
H
HART address 51 HART resistor 53 HART variables 41
I
Inowing medium 16 Installation position 15
K
Key function 29
L
Language 43 Linearization 37
M
Main menu 33 Meas. reliability 44
Measured value indication 43 Measured value memory 61
Measurement loop name 33 Medium type 34 Moisture 15
N
NAMUR NE 107
– Failure 63 – Function check 64 – Maintenance 65Out of specication 64
P
Packaging 13 Peak value 44, 45 PIN 46 Potential equalisation 20
42279-EN-130612
88
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
Probe length 34 Probe type 50
Q
Quick setup 31
R
Read out info 51
Repair 70 Replacement parts
– Display and adjustment module with
heating 14
– Electronics module 14
Reset 47
S
Scaling measured value 39, 40 Sensor characteristics 52 Sensor status 44 Service hotline 69 Shielding 20 Simulation 45 Special parameters 50 Status messages 62 Storage 13
T
Type plate 9
INDEX
U
Units 34
V
Voltage supply 19
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
89
Notes
90
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
Notes
42279-EN-130612
VEGAFLEX 81 • 4 … 20 mA/HART two-wire
91
Printing date:
All statements concerning scope of delivery, application, practical use and operat­ing conditions of the sensors and processing systems correspond to the information available at the time of printing. Subject to change without prior notice
© VEGA Grieshaber KG, Schiltach/Germany 2013
VEGA Grieshaber KG Am Hohenstein 113 77761 Schiltach Germany
Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail: info.de@vega.com www.vega.com
42279-EN-130612
Loading...