Supplementary documents appropriate to the ordered version come
with the delivery. You can nd them listed in chapter "Product descrip-tion".
Instructions manuals for accessories and replacement parts
Tip:
To ensure reliable setup and operation of your VEGACAL 63, we oer
accessories and replacement parts. The corresponding documenta-
tions are:
27720 - VEGADIS 61
•
30531 - Electronics module VEGACAL series 60
•
34296 - Protective cover
•
31088 - Flanges according to DIN-EN-ASME-JIS
•
Editing status: 2013-08-21
3
1 About this document
1 About this document
1.1 Function
This operating instructions manual provides all the information you
need for mounting, connection and setup as well as important instructions for maintenance and fault rectication. Please read this information before putting the instrument into operation and keep this manual
accessible in the immediate vicinity of the device.
1.2 Target group
This operating instructions manual is directed to trained specialist
personnel. The contents of this manual should be made available to
these personnel and put into practice by them.
1.3 Symbolism used
Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can result.
Warning: If this warning is ignored, injury to persons and/or serious
damage to the instrument can result.
Danger: If this warning is ignored, serious injury to persons and/or
destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
List
•
The dot set in front indicates a list with no implied sequence.
Action
→
This arrow indicates a single action.
1 Sequence of actions
Numbers set in front indicate successive steps in a procedure.
Battery disposal
This symbol indicates special information about the disposal of batteries and accumulators.
30028-EN-130917
4
VEGACAL 63 • Probus PA
2 For your safety
2 For your safety
2.1 Authorised personnel
All operations described in this operating instructions manual must
be carried out only by trained specialist personnel authorised by the
plant operator.
During work on and with the device the required personal protective
equipment must always be worn.
2.2 Appropriate use
VEGACAL 63 is a sensor for continuous level measurement.
You can nd detailed information on the application range in chapter
"Product description".
Operational reliability is ensured only if the instrument is properly
used according to the specications in the operating instructions
manual as well as possible supplementary instructions.
For safety and warranty reasons, any invasive work on the device
beyond that described in the operating instructions manual may be
carried out only by personnel authorised by the manufacturer. Arbitrary conversions or modications are explicitly forbidden.
2.3 Warning about incorrect use
Inappropriate or incorrect use of the instrument can give rise to
application-specic hazards, e.g. vessel overll or damage to system
components through incorrect mounting or adjustment.
30028-EN-130917
VEGACAL 63 • Probus PA
2.4 General safety instructions
This is a state-of-the-art instrument complying with all prevailing
regulations and guidelines. The instrument must only be operated in a
technically awless and reliable condition. The operator is responsible
for the trouble-free operation of the instrument.
During the entire duration of use, the user is obliged to determine the
compliance of the necessary occupational safety measures with the
current valid rules and regulations and also take note of new regula-
tions.
The safety instructions in this operating instructions manual, the national installation standards as well as the valid safety regulations and
accident prevention rules must be observed by the user.
For safety and warranty reasons, any invasive work on the device
beyond that described in the operating instructions manual may be
carried out only by personnel authorised by the manufacturer. Arbitrary conversions or modications are explicitly forbidden.
The safety approval markings and safety tips on the device must also
be observed.
5
2 For your safety
2.5 Safety label on the instrument
The safety approval markings and safety tips on the device must be
observed.
2.6 CE conformity
This device fullls the legal requirements of the applicable EC guidelines. By attaching the CE mark, VEGA provides a conrmation of
successful testing. You can nd the CE conformity declaration in the
download area of "www.vega.com".
2.7 FulllmentofNAMURrecommendations
NAMUR is the automation technology user association in the process
industry in Germany. The published NAMUR recommendations are
accepted as the standard in eld instrumentation.
The device fullls the requirements of the following NAMUR recommendations:
NE 21 – Electromagnetic compatibility of equipment
•
NE 43 – Signal level for malfunction information from measuring
•
transducers
NE 53 – Compatibility of eld devices and display/adjustment
•
components
For further information see www.namur.de.
2.8 Safety instructions for Ex areas
Please note the Ex-specic safety information for installation and operation in Ex areas. These safety instructions are part of the operating
instructions manual and come with the Ex-approved instruments.
2.9 Environmental instructions
Protection of the environment is one of our most important duties.
That is why we have introduced an environment management system
with the goal of continuously improving company environmental protection. The environment management system is certied according
to DIN EN ISO 14001.
Please help us fulll this obligation by observing the environmental
instructions in this manual:
Chapter "Packaging, transport and storage"
•
Chapter "Disposal"
•
30028-EN-130917
6
VEGACAL 63 • Probus PA
Scope of delivery
3 Product description
3 Product description
3.1 Conguration
The scope of delivery encompasses:
Level sensor VEGACAL 63
•
Documentation
•
– this operating instructions manual
– Operating instructions manual 27835 "Display and adjustment
module PLICSCOM" (optional)
– Supplementary instructions manual 31708 "Heating for display
and adjustment module" (optional)
– Supplementary instructions manual "Plug connector for con-
tinuously measuring sensors" (optional)
– Ex-specic "Safety instructions" (with Ex versions)
– if necessary, further certicates
Constituent parts
Type plate
The VEGACAL 63 consists of the components:
Process tting with probe
•
Housing with electronics
•
Housing cover, optionally available with display and adjustment
•
module
1
2
3
Fig. 1: VEGACAL 63 - rod version with plastic housing
1 Housing cover with integrated display and adjustment module (optional)
2 Housing with electronics
3 Processtting
The nameplate contains the most important data for identication and
use of the instrument:
30028-EN-130917
VEGACAL 63 • Probus PA
7
3 Product description
1
2
3
4
5
6
7
8
Fig. 2: Layout of the type label (example)
1 Instrument type
2 Product code
3 Approvals
4 Process and ambient temperature, process pressure
5 Power supply and signal output, electronics
6 Protection rating
7 Probe length
8 Order number
9 Serial number of the instrument
10 Material, wetted parts
11 Symbol of the device protection class
12 Reminder to observe the instrument documentation
13 ID numbers, instrument documentation
14 NotiedauthorityforCEmarking
15 Approval directives
15
14
13
12
11
10
9
With the serial number, you can access the delivery data of the instrument via www.vega.com, "VEGA Tools" and "serial number search".
In addition to the type label outside, you can also nd the serial number on the inside of the instrument.
Area of application
Functional principle
8
3.2 Principle of operation
VEGACAL 63 is a level sensor with fully insulated probe for continuous level measurement.
The electronics functions according to the admittance principle
(phase-selective admittance processing).
It is designed for industrial use in all areas of process technology and
can be applied in all areas of industrial process measurement.
Fully insulated probes such as VEGACAL 63 are preferably used in
conductive liquids.
Probe, measured product and vessel wall form an electrical capacitor.
The capacitance is inuenced by three main factors.
30028-EN-130917
VEGACAL 63 • Probus PA
3 Product description
1
2
3
Fig. 3: Functional principle - Plate capacitor
1 Distance between the electrode surfaces
2 Size of the electrode surfaces
3 Type of dielectric between the electrodes
The probe and the vessel wall are the capacitor plates. The measured
product and the insulation are the dielectric. Due to the higher dielectric constant of the insulation and the conductive product compared to
air, the capacitance increases as the probe is gradually covered.
The capacitance as well as the resistance change are converted by
the electronics module into a level-proportional signal.
Power supply and bus
communication
GSD/EDD
30028-EN-130917
VEGACAL 63 • Probus PA
Power supply via the Probus DP/PA segment coupler or VEGALOG
571 EP cards. A two-wire cable according to Probus specication
serves as carrier of both power and digital data transmission for
multiple sensors. The instrument prole of VEGACAL 63 corresponds
to prole specication version 3.0.
The backlight of the display and adjustment module is powered by the
sensor. Prerequisite is a certain level of operating voltage.
The data for power supply are specied in chapter "Technical data".
The optional heating requires its own operating voltage. You can nd
details in the supplementary instructions manual "Heating for display and adjustment module".
This function is generally not available for approved instruments.
The GSD (instrument master les) and bitmap les necessary for
planning your Probus-DP-(PA) communication network are available
from the download section on the VEGA homepage www.vega.com
under "Services-Downloads-Software-Probus". There you can
also nd the appropriate certicates. In a PDM environment, an EDD
(Electronic Device Description) is also required to enable the full
range of sensor functions (also available as a download).A CD with
the appropriate les can be ordered via e-mail under info@de.vega.
com or by phone from one of the VEGA agencies under the order
number "DRIVER.S".
3.3 Operation
The instrument can be adjusted with the following adjustment media:
With the display and adjustment module
•
9
3 Product description
Packaging
with the suitable VEGA DTM in conjunction with an adjustment
•
software according to the FDT/DTM standard, e.g. PACTware and
PC
with the adjustment program PDM
•
3.4 Packaging, transport and storage
Your instrument was protected by packaging during transport. Its
capacity to handle normal loads during transport is assured by a test
based on ISO 4180.
The packaging of standard instruments consists of environmentfriendly, recyclable cardboard. For special versions, PE foam or PE
foil is also used. Dispose of the packaging material via specialised
recycling companies.
Transport
Transport inspection
Storage
Storage and transport
temperature
PLICSCOM
VEGACONNECT
Transport must be carried out in due consideration of the notes on the
transport packaging. Nonobservance of these instructions can cause
damage to the device.
The delivery must be checked for completeness and possible transit
damage immediately at receipt. Ascertained transit damage or concealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and
stored according to the orientation and storage markings on the
outside.
Unless otherwise indicated, the packages must be stored only under
the following conditions:
Not in the open
•
Dry and dust free
•
Not exposed to corrosive media
•
Protected against solar radiation
•
Avoiding mechanical shock and vibration
•
Storage and transport temperature see chapter "Supplement -
The display and adjustment module PLICSCOM is used for measured
value indication, adjustment and diagnosis. It can be inserted into the
sensor or the external display and adjustment unit and removed at
any time.
You can nd further information in the operating instructions "Display
and adjustment module PLICSCOM" (Document-ID 27835).
The interface adapter VEGACONNECT enables the connection of
communication-capable instruments to the USB interface of a PC. For
parameter adjustment of these instruments, the adjustment software
PACTware with VEGA-DTM is required.
30028-EN-130917
10
VEGACAL 63 • Probus PA
3 Product description
You can nd further information in the operating instructions "Interface adapter VEGACONNECT" (Document-ID 32628).
VEGADIS 62
Protective cap
Flanges
Screening tube adapter
VEGADIS 62 is suitable for measured value indication and adjustment
of sensors with HART protocol. It is looped into the 4 … 20 mA/HART
signal cable.
You can nd further information in the operating instructions "VE-GADIS 62" (Document-ID 36469).
The protective cover protects the sensor housing against soiling and
intense heat from solar radiation.
You will nd additional information in the supplementary instructions
manual "Protective cover" (Document-ID 34296).
Screwed anges are available in dierent versions according to the
following standards: DIN 2501, EN 1092-1, BS 10, ANSI B 16.5,
JIS B 2210-1984, GOST 12821-80.
You can nd additional information in the supplementary instructions
manual "Flanges according to DIN-EN-ASME-JIS" (Document-ID
31088).
Condensation
In case of strong condensation, the draining of condensed water
can change the measurement accuracy. The suitable version is the
Screening against condensation. The condensation can drain o
outside on the screening tube adapter.
Typical applications of the screening tube adapters are e.g. for
condensation or sockets. Apart from the standard version, there is
a second version for vacuum with a sepcial seal. When the screening tube adapter is submerged in liquid, we recommend the use of a
vacuum-tight version.
Socket
In case of long sockets, the screening tube can increase the sensitivity of the probe by compensating the inuences of the socket. The
suitable version is Capacitive screening, vacuum-tight.
When the probe is mounted laterally, buildup can accumulate in the
socket. A screening tube makes the covered part of the probe inactive
and hence insensitive to inuence from buildup and socket. Hence,
the screening tube adapter excludes changing inuences caused by
the medium and ensures stable measurement conditions. The suitable version is Capacitive screening, vacuum-tight.
30028-EN-130917
VEGACAL 63 • Probus PA
11
4 Mounting
Suitability for the process
conditions
4 Mounting
4.1 General instructions
Make sure that all parts of the instrument coming in direct contact
with the process, especially the sensor element, process seal and
process tting, are suitable for the existing process conditions, such
as process pressure, process temperature as well as the chemical
properties of the medium.
You can nd the specications in chapter "Technical data" and on the
nameplate.
Installation position
Welding work
Handling
Moisture
Select an installation position you can easily reach for mounting and
connecting as well as later retrotting of a display and adjustment
module. The housing can be rotated by 330° without the use of any
tools. You can also install the display and adjustment module in four
dierent positions (each displaced by 90°).
Before beginning the welding work, remove the electronics module
from the sensor. By doing this, you avoid damage to the electronics
through inductive coupling.
With threaded versions, the housing must not be used to screw in the
instrument. Applying tightening forces on the housing can damage its
internal parts.
Use the hexagon for screwing in.
Use the recommended cables (see chapter "Connecting to power supply") and tighten the cable gland.
You can give your instrument additional protection against moisture
penetration by leading the connection cable downward in front of the
cable entry. Rain and condensation water can thus drain o. This applies mainly to outdoor mounting as well as installation in areas where
high humidity is expected (e.g. through cleaning processes) or on
cooled or heated vessels.
Pressure/Vacuum
12
Fig. 4: Measures against moisture penetration
The process tting must be sealed if there is gauge or low pressure in
the vessel. Before use, check if the seal material is resistant against
the measured product and the process temperature.
VEGACAL 63 • Probus PA
30028-EN-130917
4 Mounting
The max. permissible pressure is specied in chapter "Technical data" or on the type label of the sensor.
Insulating measures, such as e.g. covering the thread with teon tape,
can interrupt the necessary electrical connection with metal vessels.
For this reason, ground the probe on the vessel or use a conductive
seal material.
Vessel material
Vessel forms
Installation position
Metal vessel
Make sure that the mechanical connection of the probe to the vessel
is electrically conductive to ensure sucient grounding.
Use conductive seals, such as those made of copper or lead, etc.
Insulating measures, such as covering the thread with Teon tape,
can interrupt the necessary electrical connection with metal vessels.
For this reason, ground the probe on the vessel or use a conductive
seal material.
Non-conductivevessels
In non-conductive vessels, e.g. plastic tanks, the second pole of the
capacitor must be provided separately, e.g. in the form of a concentric
tube.
If possible, the capacitive probe should be mounted vertically or parallel to the counter electrode. This applies particularly to applications
in non-conductive products.
In cylindrical tanks, spherical tanks or other asymmetrical tank forms,
nonlinear level values are generated due to the varying distance to
the vessel wall.
Use a concentric tube in non-conductive products or linearize the
meas. signal.
4.2 Mounting instructions
During operation, the probe must not touch any installations or the
vessel wall. The measured value can also change if the distance to
the vessel wall changes considerably. If necessary, secure the end of
the probe (insulated).
30028-EN-130917
VEGACAL 63 • Probus PA
13
4 Mounting
Inowingmedium
1
2
Fig. 5: Fasten the probe
1 Probe
2 Plastic socket
1
2
In vessels with conical bottom it can be advantageous to mount the
sensor in the center of the vessel, as measurement is then possible
down to the lowest point of the vessel bottom.
If the instrument is mounted in the lling stream, unwanted false
measurement signals can be generated. For this reason, mount the
instrument at a position in the vessel where no disturbances, e.g. from
lling openings, agitators, etc., can occur.
This applies particularly to instrument versions with a longer probe.
Torque with PTFE plated
anges
14
30028-EN-130917
Fig.6:Inowingmedium
To compensate the normal voltage loss due to sealing materials, you
have to additionally use disc springs for fastening ange screws on
VEGACAL 63 • Probus PA
4 Mounting
PTFE coated anges. Tighten the screws moderately with the torque
stated in the technical data.
30028-EN-130917
VEGACAL 63 • Probus PA
15
5 Connecting to power supply
Safety instructions
5 Connecting to power supply
5.1 Preparing the connection
Always keep in mind the following safety instructions:
Connect only in the complete absence of line voltage
•
If voltage surges are expected, overvoltage arresters should be
•
installed according to Probus specications
Tip:
We recommend VEGA overvoltage arrester B63-32.
In hazardous areas you must take note of the respective regulations,
conformity and type approval certicates of the sensors and power
supply units.
Voltage supply
Connection cable
Cablegland½NPT
Cable screening and
grounding
Power is supplied via a Probus DP/PA segment coupler or a VEGALOG 571 EP input card. The power supply range can dier depending
on the instrument version.
The data for power supply are specied in chapter "Technical data".
Connection is made with screened cable according to the Probus
specication. Power supply and digital bus signal are carried over the
same two-wire connection cable.
Use cable with round cross-section. A cable outer diameter of
5 … 9 mm (0.2 … 0.35 in) ensures the seal eect of the cable gland.
If you are using cable with a dierent diameter or cross-section,
exchange the seal or use a suitable cable gland.
Please make sure that your installation is carried out according to the
Probus specication. In particular, make sure that the termination of
the bus is done with appropriate terminating resistors.
On the instrument with cable entry ½ NPT and plastic housing there is
a metallic ½" threaded insert moulded into the plastic housing.
Caution:
No grease should be used when screwing the NPT cable gland or
steel tube into the threaded insert. Standard grease can contain
additives that corrode the connection between threaded insert and
housing. This would inuence the stability of the connection and the
tightness of the housing.
With systems with potential equalisation, connect the cable screen
directly to ground potential at the power supply unit, in the connection
box and at the sensor. The screen in the sensor must be connected
directly to the internal ground terminal. The ground terminal outside
on the housing must be connected to the potential equalisation (low
impedance).
In systems without potential equalisation, connect the cable screen
directly to ground potential at the power supply unit and at the sensor.
In the connection box or T-distributor, the screen of the short stub to
the sensor must not be connected to ground potential or to another
cable screen. The cable screens to the power supply unit and to the
30028-EN-130917
16
VEGACAL 63 • Probus PA
Loading...
+ 36 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.