TEL: (909) 923-9390 May 2008
FAX: (909) 923-9391 Rev 2
1520 South Hellman Avenue
Ontario, California 91761, USA
CT-6500 Series 2 Operating Instructions
SAFETY WARNINGS AND CAUTIONS
Only trained operators shall use the device. All circuit breakers under test shall be off line and
fully isolated.
DO NOT SERVICE OR TEST ALONE
Do not perform test procedures or service unless another person is also present who is
capable of rendering aid and resuscitation.
DO NOT MODIFY TEST EQUIPMENT
Due to the added risk of introducing additional or unknown hazards, do not install substitute
parts or perform any unauthorized modifications to any CT-6500 test unit. To ensure that all
designed safety features are maintained, it is recommended that repairs be performed only by
Vanguard Instruments Company’s factory personnel or by an authorized repair and service
center. Unauthorized modifications can cause serious safety hazards and will nullify the
manufacturer's warranty.
FOLLOW EXACT OPERATING PROCEDURES
Any deviation from the procedures described in the operator’s manual may create one or more
safety hazards, damage the CT-6500, the test circuit breaker or cause errors in the test
results. Vanguard Instruments Company, Incorporated assumes no liability for unsafe or
improper use of the CT-6500.
WARRANTY
The CT-6500’s warranty is only valid to the original purchaser to be free from defects in
material and workmanship for a period of one year. The warranty does not apply to normal
wear or damage from misuse, abuse, improper storage, installation, accident, unauthorized
repair or alterations.
Table 25.0 Print Test Record Data ......................................................................................... 64
Table 26.0 Set Transducer Filter ............................................................................................ 65
Table 27.0 Set Contact Filter .................................................................................................. 66
V
CT-6500 Series 2 Operating Instructions
Appendix
APPENDIX A ITE Circuit Breaker model 14.4K Timing Chart ................................................... 71
APPENDIX B CLOSE-OPEN Timing Chart for ITE Circuit Breaker Model 14.4K ..................... 72
APPENDIX C CLOSE-OPEN Tabulated Report for ITE Circuit Breaker Model 14.4K .............. 73
APPENDIX D OPEN-CLOSE Timing Chart for ITE Circuit Breaker Model 14.4K ..................... 74
APPENDIX E OPEN-CLOSE Tabulated Report for ITE Circuit Breaker Model 14.4K .............. 75
APPENDIX F SIEMENS TCP Breaker Velocity Calculation ...................................................... 76
APPENDIX G SIEMENS SPS2 Breaker Velocity Calculation ................................................... 77
APPENDIX H CT-6500 SIEMENS SPS2 Test Plan .................................................................. 78
APPENDIX I CT-6500 Transducer Illustration ........................................................................... 79
VI
CT-6500 Series 2 Operating Instructions
1.0 Introduction
The CT-6500 is the fifth generation microprocessor-based, digital time travel circuit breaker
analyzer developed by Vanguard Instruments Co. The CT-6500 can fully analyze a utility circuit
breaker’s performance by measuring: the contact time, stroke, velocity, over-travel, bounce-back
and contact wipe.
The circuit breaker contact motion analysis includes: Open, Close, Open-Close, Close-Open
and Open-Close-Open operations. A timing window of one second, ten second or twenty second
can be selected. The ten and twenty second windows allow timing of long timing events such as
circuit switcher contact activities.
The conventional Time-Travel test mode allows the user to fully analyze the circuit breaker
contact time and contact travel information. Additional circuit breaker information such as:
Operating-coil current reading, breaker DC coil power supply voltage reading are also
available.
The CT-6500 is available with the following configurations:
CT-6500-3:
- Three Contact-Input channels
- Three Digital Travel Transducer Input channels
- One Analog Voltage Monitor Input channel
- One Digital Voltage Monitor Input channel
- One Trip/Close Current Monitor channel
CT-6500-6:
- Six Contact-Input channels
- Three Digital Travel Transducer Input channels
- One Analog Voltage Monitor Input channel
- One Digital Voltage Monitor Input channel
- One Trip/Close Current Monitor channel
CT-6500-12:
- Twelve Contact-Input channels
- Three Digital Travel Transducer Input channels
- One Analog Voltage Monitor Input channel
- One Digital Voltage Monitor Input channel
- One Trip/Close Current Monitor channel
There are two different modes of operation for the CT-6500 which are Stand Alone Mode and
Computer Control Mode. The manual will focus on the Stand Alone Mode. The Computer Control
Mode is covered under the Circuit Breaker Analyzer Software Series 2 manual.
7
CT-6500 Series 2 Operating Instructions
2.0 CT-6500 Description
2.1 Contact Timing Channels
A breaker's interruption of contacts are timed with the dry-contact input channels. Each contact
channel can detect the main contacts and the insertion resistor contact times in milliseconds and
cycles.
The CT-6500 applies a test voltage of 35Vdc to each of the contact channels thus allowing an
analog to digital converter (A/D) to determine if a close, an open or an insertion resistor was the
state of the contact. The CT-6500 records into the memory 20,000 readings from the A/Ds and
the transducer position counters. The contact time, circuit breaker stroke and velocity is then
derived from the data that was stored in the memory.
NOTE
The CT-6500 can detect insertion resistors ranging from 10 to 7,000 Ohms.
Resistances over 7,000 Ohms are detected as an open contact.
2.2 Analog Voltage Monitoring Channel
One analog voltage, input channel, “V1” is dedicated to monitoring the breaker's DC power
supply or coil voltages from 0 to 255Vdc or peak Vac.
NOTE
Maximum input voltage is 300Vdc or peak Vac. Recording voltage readings are from 0
to 255Vdc.
2.3 Digital Voltage Monitoring Channel
One digital voltage, input channel, “V2” monitors the status voltage as “ON” or “OFF.” Voltages
from 30 to 255Vdc are considered as “ON” status. Voltages less than 30Vdc are considered as
“OFF” status.
2.4 Trip/Close Current Monitoring
A built-in, hall-effect, current sensor measures the trip and close coil current levels and
duration. The coil current waveform can be graphically plotted onto a printout. The coil current
reading is also printed in the tabulated report. The CT-6500 initiate circuit is capable of carrying
up to 30 amperes. Maximum coil current reading is 20 amperes.
2.5 Travel Transducer Input Channels
Each CT-6500 has three digital transducer input channels. The CT-6500 uses up-down
counters to sense the breaker’s transducer reading. The CT-6500 travel transducer employs
optical encoders to send quadrature signals to the CT-6500’s counters. With the use of digital
transducers and counters, the need to set up or calibrate the transducers is eliminated.
The CT-6500 will interface with a linear, a rotary and other special transducers developed by
Vanguard Instruments Company. Please refer to Appendix I for more details.
2.6 Breaker Initiate Capability
A built-in, solid-state, initiate device allows the user to operate the circuit breaker from the CT-
6500. Operational modes include: Open, Close, Open-Close, Close-Open and Open-CloseOpen. The multiple operations of: Open-Close, Close-Open and Open-Close-Open can be
initiated with a programmable delay time or by sensing the breaker's contact state.
The solid-state switching implemented in the initiate circuitry allows the CT-6500 to switch a
8
CT-6500 Series 2 Operating Instructions
breaker's AC or DC control circuit. . The CT-6500 initiate circuit is capable of carrying up to 30
amperes. Separate 5 Ampere, fast-blow fuses protect each trip and close circuits. An interlock
“ARM” switch action is required by the operator to operate the initiate circuit.
2.7 Built-in Thermal Printer
The breaker's contact analysis results can be printed in both tabular and graphical formats by
a built-in, 4.5 inch wide, thermal printer. Refer to paragraph 7.5 for ordering thermal paper.
2.8 Computer Interface Capabilities
Built-in, RS-232C and USB ports allow the CT-6500 to be interfaced with an IBM-compatible
personal computer. A Windows XP Circuit Breaker Analyzer software is supplied with each CT-
6500. The software allows the user to:
- Control CT-6500 from a personal computer.
- Retrieve timing records stored in the CT-6500.
- Transfer circuit breaker test plans to CT-6500.
- Recall timing records from a hard drive for reanalysis at an office PC.
2.9 Timing Records Storage Capabilities
The CT-6500 uses Electrically Erasable Programmable Read-Only-Memory (EEPROM) to
store timing-shot data. Unlike other media, EEPROMs are immune to temperature, shock and
humidity. Stored shots can be recalled to reanalyze the test data, to reprint test reports or to
transfer data to a personal computer for record keeping. Up to 100 timing records can be stored
in the EEPROMs. The number of test records stored in the EEPROMs may vary depending upon
the size of the records.
2.10 Diagnostic Capabilities
The CT-6500 is designed with self-diagnostic capability to check the integrity of the electronics.
Self testing of the contact cable hook-up and transducers permits the user to examine the CT6500‘s components whenever verification is required.
2.11 CT-6500 Display and Control Switches
An alphanumeric keypad enables the users to operate the CT-6500 and to enter the breaker’s
nameplate data such as: the substation name, breaker model and other descriptions. A 4-line by
20-character LCD readout displays user messages and menus. The LCD is back-lighted to allow
messages to be viewed in low light conditions.
3.0 Test Result Tabulation
3.1 Timing Test Measurement Results
In the Time Travel Mode, each CT-6500 contact channel provides the following test results:
- Contact Close time
- Contact Open time
- Insertion Resistor Contact Open and Close times
- Contact Bounce time
- Resistor On time
- Contact Spread time
9
CT-6500 Series 2 Operating Instructions
NOTE
The CT-6500 can detect insertion resistors ranging from 10 to 7,000 Ohms.
Resistances over 7,000 Ohms are detected as an open contact.
3.2 Mechanical Measurement Results
In the Time Travel Mode, the digital transducer channels provide the following test results:
- CB contact stroke
- CB contact over travel
- CB Contact bounce back
- CB Contact wipe
- CB Contact velocity
Breaker contact speed calculations are based upon a contact's travel distance over a period of
time, which is defined by the manufacturer's specifications.
NOTE
When the resistor type transducer channel is used, the test results are shown as
transducer channel 1.
CB contact mechanical measurement can be shown in “English” or “Metric” units.
3.3 Analog Voltage Input Results
The analog voltage, input channel, “V1” measures the breaker's DC control voltage during an
operation. Printed voltage readings include nominal voltage (when the supply voltage has no
load) and minimum voltage (when the supply voltage is under load or when a coil is energized).
The “V1” connection allows the user to monitor the DC power supply’s voltage fluctuations during
a breaker operation. This feature is effective in detecting breaker power supply problems, which
normally can not be seen by using a voltmeter.
NOTE
Printed voltage readings can range from 0 to 255Vdc.
3.4 Digital Voltage Input Results
The digital voltage, input channel, “V2” monitors the voltage status as “ON” or “OFF”
conditions and graphs the result. A voltage level above 30Vdc is considered as the “ON” state.
The input channel is dedicated to monitor the breaker’s auxiliary switch action during a breaker
operation.
NOTE
The input voltage ranges from 0 to 300Vdc or peak ac.
3.5 Initiate Current Results
The CT-6500 not only plots the trip or close current waveforms, but also prints the steady-state
current readings of the trip and close coils during an operation. A typical circuit breaker timing
report is shown in Figure 11.0.
10
CT-6500 Series 2 Operating Instructions
3.6 Slow-Close Test Results
A Slow-Close test determines the contact touch distance and penetration by using the travel
transducer and dry contact channels. See section 9.13 for Slow-Close test Procedure.
4.0 Test Result Graphics
4.1 Thermal Printer Graphic Printout
A 4.5 inch wide, built-in, thermal printer can output graphic waveforms of the contact channels,
travel curves and actuator-coil currents. Using the Zoom Graphics Mode, the user can enlarge
the graphical outputs by specifying the timing zone. Typical graphical results are shown in
Figures 12.0 and 13.0.
11
CT-6500 Series 2 Operating Instructions
5.0 CT-6500 Controls and Display
Before using the CT-6500, users should become familiar with all of the controls and display
indications. The keypad and display are used to operate the CT-6500. Figure 1.0 represents the
control panel and the numbered lines pointing to each control and indicator refer to Table 1.0,
which describes the function of each control and indicator.
Figure 1.0 CT-6500 Control Panel
12
CT-6500 Series 2 Operating Instructions
Table 1.0 CT-6500 Controls and Displays
Fig. 1.0
Index
no.
1-11 &
31
12
13, 16,
19
14
15
17
PANEL MARKING FUNCTIONAL DESCRIPTION
CONTACT INPUT
Female connectors for the contact channels.
(C1–C12)
0-300V
V1
D1, D2, D3
TRANSDUCER INPUT
0-300V
V2
0-300V
TRIGGER
GROUND
3-pin connector. V1 voltage input channel is dedicated to
monitor circuit breaker DC power supply or coil voltages.
Voltage sensing range is from 0-255 volts, dc or peak
ac.
16-pin connectors. Digital travel transducer input
channels.
3-pin connector. V2 voltage input channel is dedicated to
detect voltage on/off status (present or absent) of an A/B
switch. Voltage input ranges from 0-300 volts, dc or
peak ac.
3-pin connector. Triggers voltage input for external
trigger application. Voltage levels ranging from 30 to 300
V, dc or peak ac.
Safety, ground terminal. 5/16-18 thread stud, with wing
nut.
18
20
21
22
23
24
120/240Vac, 2A, 50-60Hz
Fuse: 250V 3A, Fast Blow
INITIATE
OPEN
250V, 5A FAST-BLOW
CLOSE
250V, 5A FAST-BLOW
“PUSH” TO ARM
(Printer; No panel marking)
3-wire power plug. 3 Ampere, AC fuses. The ON/OFF
switch is a 2-pole rocker.
4-pin connector used for the switching circuit for
operating the circuit breaker under test.
Open circuit fuse: 5 Ampere, 250V, Fast-Blow.
Close circuit fuse: 5 Ampere, 250V, Fast-Blow.
Spring-loaded, pushbutton switch. Press and hold to
complete the Trip or Close circuits for breaker tests.
Thermal printer. Microprocessor controlled printer that
uses specially treated, 4.5 inch wide paper.
13
CT-6500 Series 2 Operating Instructions
Table 1.0 CT-6500 Controls and Displays (Continued)
Fig. 1.0
Index no.
25
26
27
28
PANEL MARKING FUNCTIONAL DESCRIPTION
RS-232C
1-9, 0, ENTER, START,
STOP, CLEAR,
↑ PAPER LCD,
RS-232C serial computer interface port.
16 button keypad. The keys are momentary-contact,
pushbutton switches. Allows users to make menu
selections, enter alphanumeric data, adjust the LCD
contrast and reposition the printer paper.
↓ PAPER LCD
(Display; no panel marking)
4-line by 20-character, backlit LCD. Displays menus,
options, prompts and test result data.
USB
Computer interface, USB Port.
14
CT-6500 Series 2 Operating Instructions
6.0 CT-6500 Analyzer Specifications
PHYSICAL SPECIFICATIONS 16” W by 11”H by 14”D (40.6 cm x 29.9 cm x 35.6 cm)
Weight: 25lbs (11.3Kg).
INPUT POWER 8 Amperes, 90-120Vac or 200-240Vac (selectable), 50/60 Hz.
DRY-CONTACT INPUT 3 or 6 or 12 contact channels. All contact inputs are grounded until testing
is started. Each contact detects main and insertion resistor contacts.
TIMING WINDOW Selectable between: 1, 10, or 20 second.
TIMING RESOLUTION ±50 microseconds at 1 second duration.
±0.5 millisecond at 10 second duration.
±1.0 millisecond at 20 second duration.
TIMING ACCURACY 0.05% of reading, ±50 micro-seconds at 1-second duration.
CONTACT DETECTION RANGE CLOSED: less than 20 Ohms.
RANGE OPEN: greater than 10,000 Ohms.
RESISTOR DETECTION RANGE Ranges from 50 to 5,000 Ohms.
EXTERNAL TRIGGER INPUT Open/Close: 30 to 300Vdc, peak Vac.
VOLTAGE SENSING INPUTS V1: analog input, 0-255V dc or peak ac. Sensitivity: ±1V;
V2: voltage detector (present/absent).Input: 30-300V, dc or peak ac.
BREAKER OPERATION Open, Close, Open-Close, Close-Open, and Open-Close-Open.
BREAKER INITIATE CAPACITY 30A/250V ac/dc max (100A inrush).
INITIATE CURRENT READING One, Non-contact, hall-effect sensor ranging from 0-20 Amperes, DC to 5 kHz.
RANGE
TRAVEL TRANSDUCER 3 travel-transducer channels. Linear motion from 0.0 to 30.0 inches (±0.01) inch.
INPUT Rotary ranges from 0 to 360
BREAKER SLOW-CLOSE Measures “Slow-Close” the contact point distance (hard copy printout).
TEST
DISPLAY Back-lit LCD screen, 4-lines by 20-characters, sunlight viewable.
PRINTOUT Both graphic contact-travel waveforms & tabulated results are printed on 4.5 inch
thermal paper.
STORAGE CAPABILITIES Store up to 100 timing records.
COMPUTER INTERFACE RS-232C Port, USB Port.
PC SOFTWARE (included) Furnished Circuit Breaker Analyzer PC software. Windows XP and Windows Vista
compatible. Th e analysis softw are can be installed on an IBM-compatible computer
for graphical display, numerical reports and database utility for office use.
SAFETY Designed to meet UL 61010A-1 Certification and CAN/CSA C22.2 No 1010.1-92.
o
(± 0.36º).
15
CT-6500 Series 2 Operating Instructions
ENVIROMENT Operating: -10°C to 50°C (+15°F to +122°F).
Storage: -30°C to 70°C (-22°F to +158°F).
OPTIONS Hard shipping case for CT-6500.
Hard shipping case for Travel Transducers.
WARRANTY One year warranty on parts and labor.
16
7.0 CT-6500 Special Features
7.1 CT-6500 Supplied Cables
ITEM DESCRIPTION QTY
1 GND Cable 1
2 Power Cord 1
3 Contact Cable 3 or 6 or 12
4 Contact Extension Cable 3 or 6 or 12
5 Transducer cable 1
6 Voltage & Trigger Leads 3
7 Voltage & Trigger Extension cables 3
8 Initiate Lead 1
The CT-6500 operating voltage is selectable between 110-120Vac, 50/60 Hz or 220-240Vac,
50/60 Hz. Operating voltage is set by the voltage selection switch as shown in Figure 2.0.
To change the voltage setting, remove the CT-6500 from the case, locate the voltage setting
switch and set the new operating voltage.
Figure 2.0 CT-6500 Operating Voltage Setting
7.3 CT-6500 Main Power Switch & Circuit Breaker
The CT-6500 uses an AC input module that contains the AC receptacle, power switch and
fuse holder. A 3AG, 250Vac, 3 ampere, fast-blow type fuse is recommend.
7.4 Initiate Circuit Protection Fuses
The OPEN and CLOSE initiate circuits are protected by 5 Ampere fuses (3AG, 250Vac, fastblow types).
18
CT-6500 Series 2 Operating Instructions
7.5 CT-6500 Printer and Printer Paper
The built-in, thermal printer uses 4.5 inch wide, thermal paper for printing test results. To
maintain the highest, quality printing and to avoid paper jamming, it is highly recommended
that the thermal paper be supplied by the factory. Additional paper can be ordered from either
of the two sources listed below:
Vanguard Instruments Co, Inc.
1520 S. Hellman Ave.
Ontario, CA 91761
Tel: 909-923-9390
Fax: 909-923-9391
Part Number: TP-4 Paper
BG Instrument Co.
13607 E. Trent Avenue
Spokane, WA 99216
Tel: 509-893-9881
Fax: 509-893-9803
Part Number: TP-4 paper
7.6 Replacing CT-6500 Thermal Paper
The roll of thermal paper resides inside a pocket underneath the printer cover. To replace
the paper, follow the steps below:
- Remove the printer cover.
- Remove the leftover thermal paper roll from the paper holder.
- Unroll the new thermal paper.
- Feed the thermal paper into the slot between the paper pocket and the rubber
roller. The printer will automatically pull the paper under the thermal head.
7.7 CT-6500 Printer Paper Control
To advance the paper from the printer, press and release the “↑ PAPER LCD” button. To
retract the thermal paper from the printer, press and release the “↓ PAPER LCD” button.
7.8 CT-6500 LCD Contrast Control
To darken the LCD screen, press and hold the “↑ PAPER LCD” button to decrease the
contrast of the LCD screen, press and hold the “↓ PAPER LCD” until the desired contrast is
reached.
- Place the paper roll into the paper holder.
- Lift the thermal head and align the thermal paper if necessary.
- Replace the printer cover back.
NOTE
The paper should feed from the bottom of the roll.
The thermal paper will show a colored stripe in the margin to indicate that the roll
is about to run out of paper.
19
CT-6500 Series 2 Operating Instructions
8.0 Test Hook-up Connections
8.1 Contact Cable Hookup
A typical, contact cable connection to a circuit breaker is shown in Figure 3.0. Red and black
clips are connected across the circuit breaker contact phase A, B and C.
A typical circuit breaker with series contacts is shown in Figure 4.0. In the below example,
the circuit breaker has 2 contacts in series per phase. Contact timing channel 1 and 2 are
used in the illustration.
NOTE
It is advisable to ground one side of the contacts for most testing purposes. If a breaker
is floating or ungrounded, ensure that the contact channel inputs are protected against
static discharge.
Figure 3.0 Contact Cable Connection
20
CT-6500 Series 2 Operating Instructions
Figure 4.0 Series Contact Cable Connection
21
CT-6500 Series 2 Operating Instructions
8.2 Initiate Cable Hookup
The CT-6500 will trip or close breakers through a solid-state device, which will operate on any
AC or DC control voltage ranging from 10 to 300 Volts. Both the trip and close circuits are
protected by 5 Ampere fuses.
A typical DC trip and DC close, control circuit test hookup is shown in Figure 5.0.
A typical DC trip and AC close, control circuit test hookup is shown in Figure 6.0.
Figure 5.0 DC Trip and DC Close, Initiate Circuit Cable Hook-Up
Figure 6.0 DC Trip and AC Close, Initiate Circuit Cable Hook-Up
22
CT-6500 Series 2 Operating Instructions
8.3 Analog Voltage Monitor Hookup
The analog voltage input, “V1” permits the user to monitor a breaker's DC control voltage
during an operation. The analog voltage input will record the nominal DC voltage at no load and
the minimum DC voltage while the Trip or Close coil is energized.
The nominal and minimal voltage readings will be printed on a tabulated report. Analog
waveforms will also be plotted in a graphical format. Thus, the user is able to see the breaker's
DC control voltage "dip" under load conditions. Problems, such as a poor connection or an
excessive voltage drop, during the circuit breaker operation can be easily detected.
A typical voltage monitoring hook up scheme is shown on Figure 7.0.
NOTE
The maximum voltage that can be recorded is 255 Vdc.
Figure 7.0 Voltage Monitoring Cable Hook-Up
8.4 Digital Voltage Monitor Hookup
The digital, voltage input channel, “V2” permits the user to monitor the voltage status as “ON”
or “OFF” states. The voltage “ON” or “OFF” states will be plotted on the graphical report. A typical
hook up is shown in Figure 7.0.
23
CT-6500 Series 2 Operating Instructions
8.5 External Trigger Input
The External Trigger Mode enables the user to start recording data when the CT-6500 senses
a voltage. A typical application for the External Trigger Mode is to time a circuit breaker in a close
operation and to start timing only when the close coil is energized, thus bypassing the 52X relay
delay time.
Since the 52X relay carries the close coil current, the user will need to connect the CT-6500’s
initiate cable to the close terminal shown in Figure 5.0. The CT-6500 will energize the 52X relay
to start the close operation, which will then start the timing when the CT-6500 senses the voltage
across the closing coil. See Figure 8.0 for a typical test hookup. Another application for external
trigger is to start timing the breaker when the user trips or closes the breaker remotely.
NOTE
Minimum trigger voltage is set for 30 Vac/dc. Maximum, continuous voltage is limited to
300 Vac/dc. Different trigger voltages can be set at the factory, based upon specific
requests.
The CT-6500 will start looking for the external trigger voltage when the message,
“AWAITING TRIGGER…” is shown on the LCD screen. The external trigger voltage
needs to be sensed by the CT-6500 within 15 seconds after the initiate sequence has
begun. The CT-6500 will return to the main menu if no voltage is sensed.
Figure 8.0 External Trigger Cable Hook-Up
24
CT-6500 Series 2 Operating Instructions
8.6 Digital Transducer Connection
A typical transducer connection is shown on Figure 9.0. See Appendix I for more transducer
connections.
Figure 9.0 Transducer Connection
25
CT-6500 Series 2 Operating Instructions
9.0 Operating Procedures
The main steps to time a circuit breaker are listed in Figure 10.0.
.
To reduce the possibility of damaging the CT-6500 contact channels due to static
discharge in high voltage substations, it is recommend to ground one side of the
breaker’s bushings.
Always connect the CT-6500 ground stud to the substation ground before
connecting any cables to the unit.
HOOKUP CT-6500
1. GROUND THE CT-6500 (See NOTE)
2. CONNECT CT-6500 CABLES TO BREAKER
3. INSTALL TRANSDUCER WITH CABLE
APPLY POWER
1. TURN THE CT-6500 ON
2. PERFORM SELF TESTS: HOOK-UP TEST,
CABLE TEST, IF NECESSARY
TIME BREAKER
1. SELECT THE CT-6500 TEST
2. OPERATE THE BREAKER
GET RESULTS
1. READ OR PRINT THE TABULATED RESULTS
2. PRINT THE GRAPHICAL RESULTS
3. SAVE THE TIMING SHOT
Figure 10.0 CT-6500 Operating Steps
NOTE
26
CT-6500 Series 2 Operating Instructions
9.1 Timing a Circuit Breaker
The CT-6500 will initiate the breaker operation and do a timing test on the following
operations:
- OPEN
- CLOSE
- OPEN-CLOSE
- CLOSE-OPEN
- OPEN-CLOSE-OPEN
NOTE
1. The CT-6500 can start the OPEN-CLOSE operations without a delay or by using a
programmable delay between the OPEN and CLOSE commands.
2. The CLOSE-OPEN operation can be started by the CT-6500, using several options.
- Contact #1 Close
The CT-6500 can initiate a CLOSE command and then an OPEN command after
detecting the breaker’s contact was closed. The CT-6500 detects that the
breaker’s contact was closed through contact channel #1. The selection is
recommended for the CLOSE-OPEN operation since it truly represents when the
circuit breaker closed and then opened.
- Set DELAY
The CT-6500 can initiate a CLOSE command and then an OPEN command after
a programmable delay that can be set in milliseconds.
- No DELAY
The CT-6500 can initiate a CLOSE command and then an OPEN command
without any delay. Since the circuit breaker is in the open state, the circuit
breaker’s 52B contact allows the close coil to be energized, which can start the
close command. When the circuit breaker is making the transition from an open
state to a closed state, the 52A contact will close allowing the breaker to initiate
the open command.
3. The OPEN-CLOSE-OPEN operation can be initiated by the CT-6500, by using a
programmable delay between each of the operations, where the delays can be set
between the OPEN to CLOSE and between the CLOSE to OPEN commands.
27
CT-6500 Series 2 Operating Instructions
9.1.1 Timing an OPEN Operation (With No Insertion Resistors)
The following example shows the steps for performing an Open Test on a circuit breaker.
Table 3.0 Open Operation With No Insertion Resistors
5 SELECT OPEN OPERATION. PRESS #1 KEY. OPEN TIMING
Hold "ARM" Switch,
Then "START".
"STOP" to ABORT
6 INITIATE OPEN COMMAND. HOLD THE "ARM"
SWITCH AND
PRESS THE "START"
KEY.
7 OBSERVE BREAKER OPERATED.
ANALYZER IS READY TO OUTPUT RESULTS.
NONE. 1.RUN TEST 06/12/07
TEST IN PROGRESS
Hold "ARM" Until
Test complete.
(Up to 25 seconds)
2.GET RSLT 19:20:01
3.SET-UP
4.DIAGNOSTIC
Go to section 9.2 to print a test report.
NOTE
One-second timing windows are used for breaker timing. The 10 and 20 second
timing windows are used for long timing events (i.e. timing circuit-switchers).
The user can set the CT-6500 to print tabulated and graphical reports
automatically after it completes a test. See Paragraph 9.17 for the automatic print
setting.
28
CT-6500 Series 2 Operating Instructions
9.1.2 Timing an OPEN Operation with Insertion Resistors
The CT-6500 can time breakers with insertion resistors. The insertion resistor can
range from 10 to 7,000 Ohms. Any insertion resistor more than 7,000 Ohms is detected as
an open circuit. The timing results will show the main contact time and the insertion
resistor contact time. Graphic reports will show the main contact and the resistor contact
activities on each of the channels.
6 SELECT OPEN OPERATION. PRESS #1 KEY. OPEN TIMING
7 INITIATE OPEN COMMAND. HOLD THE "ARM"
8 ANALYZER IS READY TO OUTPUT RESULTS.
PRESS #1 KEY. TIMING WINDOWS:
1.WINDOW = 1 SEC
2.WINDOW = 10 SEC
3.WINDOW = 20 SEC
1. Internal Trigger
2. External Trigger
1. OPEN 2. CLOSE
3. O-C 4. C-O
5. O-C-O
Hold "ARM" Switch,
Then "START".
"STOP" to ABORT
TEST IN PROGRESS
SWITCH AND
PRESS THE "START"
KEY.
NONE. 1.RUN TEST 06/12/07
Hold "ARM" Until
Test complete.
(Up to 25 seconds)
2.GET RSLT 19:20:01
3.SET-UP
4.DIAGNOSTIC
Go to section 9.2 to print a test report.
29
CT-6500 Series 2 Operating Instructions
NOTE
One-second timing windows are used for breaker timing. The 10 and 20 second
timing windows are used for slower timing events (i.e. timing circuit-switchers).
The user can set the CT-6500 to print tabulated and graphical reports
automatically after it completes a test. See Paragraph 9.17 for the automatic print
setting.
9.1.3 Timing CLOSE-OPEN Operation Using Contact Channel #1
The CLOSE-OPEN operation of the breaker simulates a condition where a breaker is
closed on a fault. There are 3 choices for the user to setup the CLOSE-OPEN operation:
1. Contact #1 CLOSE
The CT-6500 will initiate a CLOSE operation. The Open operation is then
initiated after contact channel # 1 is closed.
2. Set DELAY
The user can set a delay from 10 to 500 ms between a CLOSE command
to the OPEN command. The CT-6500 will initiate the CLOSE operation and
start the delay counter. The OPEN operation will be initiated when the delay
time completes.
3. No DELAY
Both the CLOSE and OPEN commands are initiated by the CT-6500
simultaneously. The OPEN coil is energized when the Open auxiliary switch
makes.
NOTE
The CLOSE-OPEN command option “Contact #1 CLOSE” closely simulates the
breaker closing on a fault condition in the field.
30
CT-6500 Series 2 Operating Instructions
Table 5.0 Close-Open Operation Using Contact Channel #1
8 ANALYZER IS READY TO OUTPUT RESULTS. NONE. 1.RUN TEST 06/12/07
TEST IN PROGRESS
Hold "ARM" Until
Test complete.
(Up to 25 seconds)
2.GET RSLT 19:20:01
3.SET-UP
4.DIAGNOSTIC
NOTE
1. Users can program a delay from 10 to 500 ms between the CLOSE and
OPEN commands. When a delay is used in the operation, it may not
represent a true breaker operating condition.
2. Users can also program in no delay for the operation in which case, the
CLOSE and OPEN commands are initiated at the same time by the CT-
6500.
31
CT-6500 Series 2 Operating Instructions
9.1.4 Timing OPEN-CLOSE-OPEN Operation
The OPEN-CLOSE-OPEN operation requires the user to enter two time-delays
between the circuit breaker operations. The time delays are in millisecond intervals.
The first delay is from the first OPEN command to the CLOSE command. The second
delay is from the CLOSE command to the second OPEN command. Refer to Table 6.0
for setting up the OPEN-CLOSE-OPEN operation.
9 ANALYZER IS READY TO OUTPUT RESULTS. NONE. 1.RUN TEST 06/12/07
ENTER DELAY USING
0-9 KEYS, PRESS THE
“ENTER” KEY TO
CONFIRM.
ENTER DELAY USING
0-9 KEYS, PRESS THE
“ENTER” KEY TO
CONFIRM.
SWITCH AND
PRESS THE "START"
KEY.
C-O Delay in mS:
(10 – 350)
mSec
ENTER when done
OPEN-CLOSE-OPEN TIMING
Hold "ARM" Switch,
Then "START".
"STOP" to ABORT
TEST IN PROGRESS
Hold "ARM" Until
Test complete.
(Up to 25 seconds)
2.GET RSLT 19:20:01
3.SET-UP
4.DIAGNOSTIC
32
CT-6500 Series 2 Operating Instructions
9.2 Printing Tabulated Timing Results Using the Thermal Printer
The circuit breaker timing results can be printed on the thermal printer by using the steps listed
in Table 7.0.
A typical circuit breaker timing result printed on thermal printer is shown in Figure 11.0.
Table 7.0 Printing Tabulated Results
STEP OPERATION ACTION DISPLAY
1 GET TIMING RESULTS. PRESS #2 KEY. 1.PRINT TEST RESULTS
TEST: CLOSE
DATE: 05/23/07 12:36
"ENTER" TO PROCEED
4 CONFIRM SAVING SHOT. PRESS THE
"ENTER" KEY.
5 OBSERVE SHOT BEING SAVED. NONE. SHOT SAVE
6 RETURN TO MAIN MENU. PRESS ANY KEY.
SAVE IN PROCESS
PLEASE WAIT
1.RUN TEST 06/10/07
2.GET RESULTS 20:10:01
3.SETUP
4.DIAGNOSTIC
NOTE
Timing test record storage capability varies, depending upon the size of the records.
The CT-6500 will assign an identification number for each of the records stored in
the EEPROM.
Shot ID #2 was assigned in the example above.
41
CT-6500 Series 2 Operating Instructions
9.5 Printing Timing Test Record Directory Stored in the EEPROM
The user can obtain a timing-record directory that was stored in the CT-6500’s FLASH
EEPROM by using the steps in Table 10.0.
Table 10.0 Print Timing Test Record Directory
STEP OPERATION ACTION DISPLAY
1 SELECT SETUP MENU. PRESS #3 KEY. 1.ANALYSIS POINT
2.MEASUREMENT UNITS
3.SAVE / RESTORE
4.NEXT PAGE
2 SELECT SAVE/RESTORE MENU.
3 SELECT SHORT DIRECTORY.
4 SELECT SHORT DIRECTORY. PRESS #1 KEY. PRINTING ABBREVIATED
5 RETURN TO MAIN MENU. NO ACTION.
PRESS #3 KEY. 1.SAVE SHOT
2.RESTORE SHOT
3.SHOT DIRECTORY
4.ERASE SHOT
PRESS # 3 KEY. 1.SHORT DIRECTORY
2.FULL DIRECTORY
"STOP" TO QUIT
DIRECTORY
“STOP” TO QUIT
1.RUN TEST 06/10/07
2.GET RESULTS 20:10:01
3.SETUP
4.DIAGNOSTIC
NOTE
1. If the "Short Directory" was selected, the CT-6500 will print out the shot
identifications of the last ten timing shots stored in the EEPROM.
2. If the "Full Directory" was selected, the CT-6500 will print out the shot
identifications of all the timing shots stored in the EEPROM.
3. A typical shot directory printout is shown on in Figure 17.0.
42
CT-6500 Series 2 Operating Instructions
Figure 17.0 Printout of the Timing Shot Directory From the Thermal Printer
43
CT-6500 Series 2 Operating Instructions
9.6 Recalling a Timing Test Record
The following steps in Table 11.0 are used to recall a timing test record stored in the EEPROM
to the working memory. Once a test record is recalled, the user can get a printout of the test
results from the thermal printer.
Table 11.0 Recalling a Timing Test Record
STEP OPERATION ACTION DISPLAY
1 SELECT SETUP MENU. PRESS #3 KEY. 1.ANALYSIS POINT
4 SELECT ERASE SHOT. PRESS # 1 KEY. ENTER SHOT NUMBER
TO BE ERASED
XXXX
"ENTER" TO CONFIRM
5 ENTER SHOT NUMBER “1” TO BE
ERASED.
6 CONFIRM TO ERASE SHOT. PRESS THE
7 WAIT FOR SHOT BEING ERASED. NO ACTION. ERASE COMPLETE
8 RETURN TO MAIN MENU. PRESS ANY KEY.
PRESS KEYS 0-9 TO
SELECT SHOT
NUMBER, PRESS
THE "ENTER" KEY
TO CONFIRM.
"ENTER" KEY.
ERASE SHOT # 0001
TEST: OPEN
DATE: 03\12\07 09:12
"ENTER" TO CONFIRM
ERASE IN PROCESS
PLEASE WAIT
1.RUN TEST 06/10/07
2.GET RESULTS 20:10:01
3.SETUP
4.DIAGNOSTIC
45
CT-6500 Series 2 Operating Instructions
9.8 Test Record Description
The CT-6500 has the capability for users to enter the breaker and other identification
information. The data usually consists of the name of the company, station, circuit and
manufacturer. Other information could be the breaker's model, serial number and the operator's
name. The identification data will print out in the header section of the tabulated report.
If the user saves the timing shot, the identification data will also be saved with the timing shot,
along with the time and date.
The user can input letters of the alphabet by utilizing the alphanumeric keys. To select the
letter "A," the user will need to press key numbered “2,” twice. To select the letter "C," the user
will need to press the key numbered “2,” four times and so on.
To move the cursor forward, press the “UP ARROW” key. To move the cursor backwards, use
the “DOWN ARROW” key. Use the “CLEAR” key to enter a space.
Table 13.0 Entering Test Record Identification Header
STEP OPERATION ACTION DISPLAY
1 SELECT SET UP MENU.
2 SELECT NEXT PAGE.
3 SELECT SHOT DESCRIPTION.
4 ENTER COMPANY NAME.
5 ENTER STATION NAME. PRESS KEYS 0-9 AND
6 ENTER CIRCUIT NAME. PRESS KEYS 0-9 AND
PRESS #3 KEY. 1.ANALYSIS POINTS
PRESS # 4 KEY. 1.SHOT DESCRIPTION
PRESS # 1 KEY. COMPANY:
PRESS KEYS 0-9 AND
PRESS "ENTER" KEY.
PRESS "ENTER" KEY.
PRESS "ENTER" KEY.
2.MEASUREMENT UNIT
3.SAVE / RESTORE
4.NEXT PAGE
2.NUMBER OF CHANNELS
3.SET DATE & TIME
4.SET PRINT MODE
XXXXXX
↑/↓ TO POSITION
"ENTER" TO ACCEPT
STATION
XXXXXX
↑/↓ TO POSITION
"ENTER" TO ACCEPT
CIRCUIT
XXXXXX
↑/↓ TO POSITION
"ENTER" TO ACCEPT
MANUFACTURER
XXXXXX
↑/↓ TO POSITION
"ENTER" TO ACCEPT
7 ENTER MANUFACTURER NAME. PRESS KEYS 0-9 AND
PRESS "ENTER" KEY.
46
BREAKER
XXXXXX
↑/↓ TO POSITION
"ENTER" TO ACCEPT
CT-6500 Series 2 Operating Instructions
Table 13.0 Entering Test Record Identification Header (Continued)
STEP OPERATION ACTION DISPLAY
8 ENTER BREAKER MODEL. PRESS KEYS 0-9 AND
PRESS "ENTER" KEY.
9 ENTER SERIAL NUMBER. PRESS KEYS 0-9 AND
PRESS "ENTER" KEY.
10 RETURN TO MAIN MENU. NO ACTION.
SERIAL
XXXX
↑/↓ TO POSITION
"ENTER" TO ACCEPT
OPERATOR
XXXX
↑/↓ TO POSITION
"ENTER" TO ACCEPT
1.RUN TEST 06/10/07
2.GET RESULTS 20:10:01
3.SETUP
4.DIAGNOSTIC
47
CT-6500 Series 2 Operating Instructions
9.9 Computer Interface
Each CT-6500 has one RS-232C and one USB port, where the Circuit Breaker Analyzer
Software Series 2 package provided with each CT-6500, enables an IBM-compatible personal
computer to communicate with the unit. The software operating under Windows XP environments
allows the user to transfer timing records stored in the EEPROM into the computer. From the
user's computer, the user can now reanalyze the breaker’s timing shots and then print the timing
reports through an office printer. Since timing shots are stored in the office computer, users can
create a database of a breaker's timing history.
This PC program allows the user to generate and transfer circuit breaker test plans to the CT6500 EEPROM. Up to 99 circuit breaker test plans can be stored in the CT-6500 EEPROM.
Using the PC program, the user can control the CT-6500 to time circuit breakers.
Note
1. There is no requirement to place the CT-6500 under computer control to access the RS232C or USB port.
2. An interconnect cable must be run between the CT-6500’s RS-232C port or USB port
to the user’s PC.
3. The CT-6500's RS-232 connector pin layout is shown in the table below.
Table 14.0 RS-232C Connector Pin Definition
Pin No. Signal Name
2 Tx
3 Rx
5 Gnd
48
CT-6500 Series 2 Operating Instructions
9.10 Breaker Travel Analysis
The CT-6500’s travel analysis includes the breaker’s stroke, over-travel, bounce-back, contactwipe distances and contact velocity calculations.
9.10.1 Breaker Stroke
The CT-6500 uses a digital transducer to measure a breaker's contact stroke, over-travel and
bounce-back. The digital transducer output is 200 counts per linear inch of travel, therefore the
resolution is accurate to about 1/200 inch. The output resolution is ±0.01 inch on the test result
report.
Unlike slide-wire transducers, the CT-6500’s digital transducer needs no calibration or setup.
A user verifies the transducer’s functionality by selecting a diagnostic test for the transducer,
which is further explained in Paragraph 9.14.
9.10.2 Breaker Over-Travel Distance
Over-travel is the distance the contact moves beyond the resting position. Over-travel is
typically found in the close operation. (See Appendix A for more details).
9.10.3 Breaker Bounce-Back Distance
Bounce-back is the distance the breaker contact moves before the resting position after the
over- travel. Again, bounce-back is typically found in the close operation. (See appendix A for
more details).
9.10.4 Contact Wipe
Contact wipe is the distance measured from the close position to the contacts touching or
parting positions. In the close operation, contact wipe is measured from the contacts touching
position to the final close position.
In the open operation, the contact wipe is measured from the close position to the contact
break or parting position.
NOTE
Contact wipe is measured during an operation. The measurement may not be as
accurate as the measurement done using the Slow-Close Mode. It is recommended
that the user use the Slow-Close Test to verify the contact wipe measurement, if
required. Please refer to Paragraph 9.13 for further information on the Slow-Close
Mode.
9.10.5 Breaker Velocity
When used with a travel transducer, the CT-6500 can calculate the breaker contact velocity
through the arc zone. Users will need to program the calculation points or analysis points on the
travel curve for the unit to calculate the contact velocity. Analysis point selections and setup are
discussed in the following paragraphs.
9.10.6 Analysis Point Selections
Analysis points are used to calculate the average contact speed through the contact's arc
zone. The analysis points are usually specified by the breaker’s manufacturers. The analysis
points can be programmed into the CT-6500 by the user. The CT-6500 will store the two setup
points for calculating velocity one set each for the open and another set for the close operations.
These setups remain in the CT-6500’s memory until it is changed by the user.
49
CT-6500 Series 2 Operating Instructions
9.10.7 Analysis Point No.1 (AP1)
Three selections are available to the user for setting analysis point No. 1:
1. PERCENTAGE OF STRK
2. DISTANCE FROM CLOSE
3. CONTACT #1
a. Percentage of Stroke
Percentage of stroke is the distance based upon the percentage of the total breaker’s stroke
distance. The distance is always measured from the starting point at the fully closed position
of the breaker contacts.
b. Distance From Close
Distance from Close range is selectable from 00.0 inch to 99.99 inches or from 0.0
centimeter to 99.90 centimeters. Again, the distance is referenced from the contact's closed position. See Appendix A for more detail.
c. Contact Point #1
Contact point is the distance from the contact's closed position to the point where it is in
transition from the close-to-open or the open-to-close position. Contact channel #1 will be
used for the selection.
9.10.8 Analysis Point No. 2 (AP2)
There are three selections for analysis point 2:
1. PERCENTAGE OF STRK
2. DISTANCE FROM CLOSE
3. CONTACT +/- TIME
a. See Paragraph 9.10.7.a for “PERCENTAGE OF STROKE” description.
b. See Paragraph 9.10.7.b for “DISTANCE FROM CLOSE” description.
c. “CONTACT +/- TIME”, is available for both OPEN and CLOSE operation under analysis
point #2 only. For “CONTACT + TIME”, the user enters the millisecond time after the
contact channel #1 made the transition from OPEN to CLOSE or CLOSE to OPEN to define
the Analysis Point No. 2.
For “CONTACT – TIME”, the user enters the millisecond time before the contact channel #1
makes the transition from OPEN to CLOSE or CLOSE to OPEN to define the Analysis Point
No. 2.
See appendix A for more detail.
50
CT-6500 Series 2 Operating Instructions
NOTES
1. Average velocity through the arc zone is calculated by using the following
formula:
V
= Distance÷ Time
avg
2. The breaker contact velocity can be recalculated based upon the data stored
inside the memory after any new analysis points were selected. The user does not
need to operate the breaker again to acquire new contact velocity data after
changing the analysis points because the new velocity is then calculated from the
travel data stored in memory from the last operation.
3. For complex velocity calculations, the user can create a test plan using the CT6500 Breaker Analysis Software from a PC. The test plan then can be downloaded
to the CT-6500. The complex breaker velocity calculation can now be executed
easily by recalling the test plan before running the timing test.
51
CT-6500 Series 2 Operating Instructions
9.10.9 Set-Up Open Analysis Points
The following procedure allows the user to set the analysis points to calculate velocity of the circuit
breaker in the open operation. The open analysis point #1 is 1.00 inch from the fully closed
position of the circuit breaker. The open analysis point #2 is 2.00 inches from the fully closed
position.
Table 15.0 Setting Up "OPEN" Analysis Points Using Distance
STEP OPERATION ACTION DISPLAY
1 SELECT SET UP MENU.
2 SELECT ANALYSIS POINT.
3 SELECT OPEN TIMING.
4 SELECT DISTANCE FROM CLOSE. PRESS # 2 KEY. OPEN ANALYSIS PT 1
5 ENTER 1 INCH. PRESS KEYS 0-9 TO
6 CONFIRM ENTRY. PRESS THE "ENTER"
PRESS # 3 KEY. 1.A NALYSIS POINT
2.MEASUREMENT UNITS
3.SAVE / RESTORE
4.NEXT PAGE
PRESS # 1 KEY. 1.OPEN TIMING
2.CLOSE TIMING
3.PRINT SETTING
4.TEST PLAN
PRESS # 1 KEY. OPEN ANALYSIS PT 1
1.PERCENT OF STROKE
2.DISTANCE FROM CLOSE
3.CONTACT #1
DISTANCE FROM CLOSE
VALUE: XX.XX in
"ENTER" TO CONFIRM
OPEN ANALYSIS PT 1
SET 1.00. USE THE
“ENTER” KEY TO
CONFIRM.
KEY.
DISTANCE FROM CLOSE
VALUE: 01.00 in
"ENTER" TO CONFIRM
OPEN ANALYSIS PT 2
1.PERCENTAGE OF STROKE
2.DISTANCE FROM CLOSE
3.CONTACT +/- TIME
7 SELECT DISTANCE FROM CLOSE. PRESS # 2 KEY. OPEN ANALYSIS PT 2
DISTANCE FROM CLOSE
VALUE: XX.XX in
"ENTER" TO CONFIRM
8 ENTER 2 INCHES. PRESS KEYS 0-9 TO
SET 2.00 INCHES.
9 RETURN TO MAIN MENU. PRESS THE "ENTER"
KEY.
OPEN ANALYSIS PT 2
DISTANCE FROM CLOSE
VALUE: 02.00 in
"ENTER" TO CONFIRM
1.RUN TEST 06/10/07
2.GET RESULTS 20:10:01
3.SETUP
4.DIAGNOSTIC
52
CT-6500 Series 2 Operating Instructions
NOTES
1. If an entry is not correct, re-enter the digits until the screen shows the desired
numbers, then press the “ENTER” key.
2. Analysis Point #1 is 1.00 inch from the fully closed position.
3. Analysis Point #2 is 2.00 inches from the fully closed position.
53
CT-6500 Series 2 Operating Instructions
9.11 Select English or Metric Measurement Unit.
The CT-6500 supports and displays both English and Metric calculations from the breaker's
travel parameters such as the stroke, velocity and other parameters. The user can switch from
one system to the other by following the steps in Table 16.0.
Table 16.0 Select English or Metric Procedure
STEP OPERATION ACTION DISPLAY
1 SELECT SET UP MENU.
2 MEASUREMENT UNITS.
3 SELECT METRIC UNIT
PRESS # 3 KEY. 1.ANALYSIS POINT
2.MEASUREMENT UNITS
3.SAVE / RESTORE
4.NEXT PAGE
PRESS # 2 KEY. SELECT UNITS
1.ENGLISH
2.METRIC
3.ROT ENC 4.RES ENC
PRESS # 2 KEY.
1.RUN TEST 06/10/07
2.GET RESULTS 20:10:01
3.SETUP
4.DIAGNOSTIC
9.12 Digital Rotary Transducer Selection
A rotary transducer requires the user to enter the defined linear distance in millimeters or
inches, per one degree of rotary motion as seen in Table 17.0.
Table 17.0 Select Rotary Transducer
STEP OPERATION ACTION DISPLAY
1 SELECT SET UP MENU. PRESS # 3 KEY. 1. ANALYSIS POINT
2. MEASUREMENT UNITS
3. SAVE / RESTORE
4. NEXT PAGE
2 SELECT MEASUREMENT UNITS. PRESS # 2 KEY. SELECT UNITS
4 SELECT INCH PER DEGREE. PRESS #1 KEY. INCHES/DEGREE
0.014 in/deg
5 ENTER 0.50 INCH PER DEGREE. ENTER 0.50 USING
0-9 KEYS.
6 RETURN TO MAIN MENU. PRESS THE
“ENTER” KEY.
INCHES/DEGREE
0.500 in/deg
1.RUN TEST 06/10/07
2.GET RESULTS 20:10:01
3.SETUP
4.DIAGNOSTIC
54
CT-6500 Series 2 Operating Instructions
9.13 Slow-Close Test
The CT-6500 offers a unique feature called a "Slow-Close" test. The test permits users to
measure the distance a breaker's contact travels from the fully opened position to the point of
contact or "touch" position, and the contact wipe or penetration distance. The test requires the
operator to manually jack the breaker's contact from the fully open position to the fully closed
position.
Contact channel 1, 2, 3 and transducer channel 1 will be used for “Slow-Close” testing. The
CT-6500 will measure the contact touch and contact wipe for each contact channel.
Contact touch is the distance measured from the contact fully open to the first position where
the contact channel is electrically closed.
Contact wipe, or penetration, is the distance measured from the contact’s first, electrically
closed position to the final closed, latching position.
Table 18.0 Performing a Slow-Close Test
STEP OPERATION ACTION DISPLAY
1 SELECT DIAGNOSTIC MENU. PRESS # 4 KEY. 1. SLOW CLOSE TEST
2. CHECK HOOKUP
3. TEST TRANSDUCER
4. NEXT PAGE
2 SELECT SLOW CLOSE TEST. PRESS # 1 KEY. BREAKER SLOW CLOSE TEST
RMV CB GND (1 SIDE)
"ENTER" TO BEGIN
3 START SLOW CLOSE TEST.
4 CONTINUE TO CLOSE BREAKER. NO ACTION. LIFT BRKR TO CLS POS
5 RETURN TO MAIN MENU. NO ACTION.
PRESS THE "ENTER"
KEY.
CONTACT 1:OPEN
CONTACT 2:OPEN
CONTACT 3:OPEN
LIFT TO CLS BREAKER NOW
THEN PRESS "ENTER"
TO FIND CONTACT WIPE
1.RUN TEST 06/10/07
2.GET RESULTS 20:10:01
3.SETUP
4.DIAGNOSTIC
NOTE
The user can also save a Slow-Close test report in the EEPROM. A typical report of
a Slow-Close test is shown on in Figure 18.0.
55
CT-6500 Series 2 Operating Instructions
Figure 18.0 Typical Slow-Close Test Report
NOTE
The breaker’s stroke is the sum of the contact touch and contact wipe distances.
56
CT-6500 Series 2 Operating Instructions
9.14 Transducer Self-Test
The operator can check the transducer electronics by connecting the transducer to the CT6500 (D1 or D2 or D3) and running a “Test Transducer” diagnostics.
Table 19.0 Performing a Transducer Self Test
STEP OPERATION ACTION DISPLAY
1 SELECT DIAGNOSTIC.
2 SELECT TRANSDUCER. PRESS # 3 KEY. TRANSDUCER TEST
3 MOVE TRANSDUCER SLIDER TO A
KNOWN DISTANCE & OBSERVE
DISPLAY
2 RETURN O MAIN MENU PRESS # STOP
PRESS # 4 KEY. 1. SLOW-CLOSE TEST
2. CHECK HOOKUP
3. TEST TRANSDUCER
4. NEXT PAGE
1= 0.00 IN / 000.0°
2= 0.00 IN / 000.0°
3= 0.00 IN / 000.0°
MOVE
TRANSDUCER
SLIDER
KEY.
TRANSDUCER TEST
1= 5.00 IN / 360.0°
2= 0.00 IN / 000.0°
3= 0.00 IN / 000.0°
1.RUN TEST 06/10/07
2.GET RESULTS 20:10:01
3.SETUP
4.DIAGNOSTIC
NOTES
1. After selecting the transducer test mode, move the transducer slider and observe the
display on the CT-6500 LCD.
2. If the slider is moved 1.00 inch, the CT-6500 LCD should display 1.00 inch.
3. Press the “STOP” key to return back to the main menu.
57
CT-6500 Series 2 Operating Instructions
9.15 Check Cable Hookup
The following steps allow the user to check the CT-6500 contact cable connections to the
circuit breaker.
Table 20.0 Checking the Cable Hookup
STEP OPERATION ACTION DISPLAY
1 SELECT DIAGNOSTIC. PRESS # 4 KEY. 1. SLOW-CLOSE TEST
1. The CT-6500 will also print a test report with the thermal printer.
2. A contact closure is shown as "C" on the LCD and printout.
3. An open contact is shown as "-" on the LCD and printout.
1 2 3
C - - -
“START” to RETEST
“STOP” to EXIT
1.RUN TEST 06/10/07
2.GET RESULTS 20:10:01
3.SETUP
4.DIAGNOSTIC
58
CT-6500 Series 2 Operating Instructions
9.16 Set CT-6500 Time and Date
The CT-6500 has an internal clock with a lithium battery. The clock's time and date is
displayed on the LCD screen. When a timing record is stored in the EEPROM, the time and date
are also stored along with the record. To set the clock, follow the steps in Table 21.0.
Table 21.0 Set CT-6500 Time and Date
STEP OPERATION ACTION DISPLAY
1 SELECT SET UP MENU.
2 SELECT NEXT PAGE.
3 SELECT SET DATE & TIME.
4 ENTER MONTH DATE & YEAR. PRESS KEYS 0-9. ENTER
5 ENTER HOUR MINUTES & SECONDS. PRESS KEYS 0-9. ENTER
6 RETURN TO MAIN MENU. NO ACTION.
PRESS # 3 KEY. 1. ANALYSIS POINT
2. MEASUREMENT UNITS
3. SAVE / RESTORE
4. NEXT PAGE
PRESS # 4 KEY. 1. SHOT DESCRIPTION
2. NUMBER OF CHANNELS
3. SET DATE & TIME
4.SET PRINT MODE
PRESS # 3 KEY. ENTER
MM-DD-YY HH:MM:SS
MM-DD-YY HH:MM:SS
06/10/07
MM-DD-YY HH:MM:SS
06/10/07 20:10:00
1.RUN TEST 06/10/07
2.GET RESULTS 20:10:01
3.SETUP
4.DIAGNOSTIC
NOTE
1. The clock battery is good for three years.
2. Loosing the clock battery will not affect the CT-6500’s performance.
59
CT-6500 Series 2 Operating Instructions
9.17 Automatic Print Setting
The CT-6500 can be set to print graphs and tabulated results automatically after each timing
shot. To turn on the “Auto Print Results” mode, follow the steps in Table 22.0.
Table 22.0 Automatic Print Setting
STEP OPERATION ACTION DISPLAY
1 SELECT SET UP MENU.
2 SELECT NEXT PAGE.
3 SELECT NEXT PAGE. PRESS #4 KEY. 1. SET PRINT MODE
3 SELECT SET PRINT MODE.
4 SELECT “AUTOMATIC PRINT”. PRESS # 1 KEY.
PRESS # 3 KEY. 1. ANALYSIS POINT
2. MEASUREMENT UNITS
3. SAVE / RESTORE
4. NEXT PAGE
PRESS # 4 KEY. 1. SHOT DESCRIPTION
2. NUMBER OF CHANNELS
3. SET DATE & TIME
4. NEXT PAGE
2. SET 50/60 Hz
PRESS # 1 KEY. SET PRINT MODE
1. AUTOMATIC PRINT
2. BY REQUEST ONLY
1.RUN TEST 06/10/07
2.GET RESULTS 20:10:01
3.SETUP
4.DIAGNOSTIC
NOTE
Select print “BY REQUEST ONLY” to disable the “Automatic Print” mode.
60
CT-6500 Series 2 Operating Instructions
9.18 Selecting 50 Hz or 60 Hz for Contact Time Reading
Tabulated contact time is printed in both milli-seconds and cycles. The cycle readings can be
in 50 Hz or 60 Hz. The example below shows the steps to set for 60 Hz readings.
Table 23.0 Selecting 50Hz or 60 Hz for Contact Time Reading
STEP OPERATION ACTION DISPLAY
1 SELECT SET UP MENU.
2 SELECT NEXT PAGE.
3 SELECT NEXT PAGE. PRESS #4 KEY. 1. SET PRINT MODE
4 SELECT50 / 60 Hz MODE.
5 SELECT “60 Hz”. PRESS # 2 KEY.
PRESS # 3 KEY. 1. ANALYSIS POINT
2. MEASUREMENT UNITS
3. SAVE / RESTORE
4. NEXT PAGE
PRESS # 4 KEY. 1. SHOT DESCRIPTION
2. NUMBER OF CHANNELS
3. SET DATE & TIME
4. NEXT PAGE
2. SET 50/60 Hz
PRESS # 2 KEY. 1. SET 50 Hz
2. SET 60 Hz
1.RUN TEST 06/10/07
2.GET RESULTS 20:10:01
3.SETUP
4.DIAGNOSTIC
61
CT-6500 Series 2 Operating Instructions
Cycle Reading at 50
Figure 19.0 Typical 50 Hz Tabulated Report
Cycle Time at 60 Hz
Figure 20.0 Typical 60 Hz Tabulated Report
62
CT-6500 Series 2 Operating Instructions
9.19 CT-6500 Channel Configuration
The CT-6500 is available with 3 or 6 or 12 contact inputs. Most common timing conditions
require the user to use only 3 contact timing channels and one travel transducer channel. There is
no need to print more than 3 timing channels and one transducer channel on the graphic and
tabulated report. The CT-6500 will always default to 3 contact channels (channel 1, 2 and 3) and
one transducer channel (transducer channel #1) when the user first turns on the unit. The user
can always turn on the rest of the timing channels and transducer channels by following the steps
in the table below:
Table 24.0 Channel Configuration
STEP OPERATION ACTION DISPLAY
1 SELECT SET UP MENU.
2 SELECT NEXT PAGE.
3 SELECT NUMBER OF CHANNELS.
4 SELECT 3 CHANNELS OR 6
CHANNELS OR 12 CHANNELS.
5 SELECT 1 TRANSDUCER, OR 2, OR 3. PRESS #1, OR #2,
PRESS # 3 KEY. 1. ANALYSIS POINT
PRESS # 4 KEY. 1. SHOT DESCRIPTION
PRESS # 2 KEY. NUMBER OF CHANNELS
PRESS # 1, OR 2 OR
3 KEY.
OR #3.
2. MEASUREMENT UNITS
3. SAVE / RESTORE
4. NEXT PAGE
2. NUMBER OF CHANNELS
3. SET DATE & TIME
4. NEXT PAGE
1. 3 CONTACT
2. 6 CONTACT
3. 12 CONTACT
1. TRANSDUCER 1
2. TRANSDUCER 1 , 2
3. TRANSDUCER 1 , 2 , 3
1.RUN TEST 06/10/07
2.GET RESULTS 20:10:01
3.SETUP
4.DIAGNOSTIC
63
CT-6500 Series 2 Operating Instructions
9.20 Print Test Record Data
Print data command allows the user to print the hexadecimal timing data collected by the CT-
6500. This feature is mainly for factory testing. The CT-6500 collects 20,000 data points in the
one-second timing window. As a results, each data sample is 0.05 ms apart. By looking at the
time marker and the data point, the user can actually find the channel activities vs time in the
timing record.
The following steps allow the user to print CT-6500 timing data from memory.
Table 25.0 Print Test Record Data
STEP OPERATION ACTION DISPLAY
1 SELECT DIAGNOSTIC. PRESS # 4 KEY. 1. SLOW-CLOSE TEST
As sample of the CT-6500 timing data print out is shown in Figure 21.0.
Note
Pressing key #0 will start printing the data at address location 0 (or the first data point)
Pressing key #1 will start printing the data at address location .10000 second.
Timer marker 0.100000 ms
Time marker 0.10010 ms
Figure 21.0 CT-6500 Timing Data
64
CT-6500 Series 2 Operating Instructions
9.21 Transducer Encoder Filter Setting
In a typical 1 second timing record, the CT-6500 records 20,000 data points for each of the
Contact channels (C1-C6), Digital transducer channels (D1-D3), Voltage input channels (V1 &
V2), CT channel, DCR channel, Resistor transducer channel and Initiate current channel.
In most circuit breaker timing application, the breaker activities end after 200 ms. A
common problem found in the field was the CT-6500 recorded erroneous data on the
transducer channels (after breaker activities ended) due to vibration. The erroneous data is
probably due to the transducer not being secured to the mounting plate properly or a poor
linkage problem between the transducer and the circuit breaker mechanism.
The Transducer Encoder Filter setting allows the user to stop recording the transducer
channel data after a specific window. This feature allows the user to filter out the unwanted
motion picked up by the transducer after the breaker activities have ended.
The following procedure allows the user to stop recording transducer data after 200ms.
Table 26.0 Set Transducer Filter
STEP OPERATION ACTION DISPLAY
1 SELECT DIAGNOSTIC. PRESS # 4 KEY. 1. SLOW-CLOSE TEST
2. CHECK HOOKUP
3. TEST TRANSDUCER
4. NEXT PAGE
2 SELECT NEXT PAGE. PRESS # 4 KEY. 1. PRINT DATA
2. ENCODER FILTER
3. CONTACT FILTER
3 SELCT ENCODER FILTER. PRESS # 2 KEY. ENTER FILTER TIME
(010 - 999)
END RECORDING DATA A FTER
4
200MS.
5 SELECT FILTER SETTING. PRESS “ENTER” KEY.
ENTER “2, 0, 0”. ENTER FILTER TIME
(010 - 999)
200
1.RUN TEST 06/10/07
2.GET RESULTS 20:10:01
3.SETUP
4.DIAGNOSTIC
Note
The Transducer Encoder Filter setting is default to “No Filter” when the CT-6500 is
powered on.
The filter setting is available from 10 ms to 999 ms.
A setting of 200 ms requires the user to enter 200 in the “FILTER TIME”. This
setting makes the CT-6500 not to record any Transducer Encoder data after 200 ms.
65
CT-6500 Series 2 Operating Instructions
9.22 Contact Filter Setting
The contact filter setting is applicable for timing shots with insertion resistors only. The
CT-6500 automatically detects the contact time using its own algorithm. However, manual
setting for contact time calculation is made available to the user.
The filter setting of 1 allows the CT-6500 to pick up the first contact transition time after the
resistor contact activity is detected. The filter setting of 300 allows the CT-6500 to pick up the
last contact transition time after the resistor contact activity is detected. Use the steps below to
set contact filtering.
Table 27.0 Set Contact Filter
STEP OPERATION ACTION DISPLAY
1 SELECT DIAGNOSTIC. PRESS # 4 KEY. 1. SLOW-CLOSE TEST
The CT-6500 will return to automatic filter setting after power is cycled.
66
CT-6500 Series 2 Operating Instructions
Filter Setting @ 1 will select
this contact time
Filter setting @ 300 will
select this contact time.
Filter Setting @ 1 will select
this contact time
Filter setting @ 300 will
select this contact time.
Figure 22.0 Contact Filter Setting Illustration
67
CT-6500 Series 2 Operating Instructions
9.23 Running CT-6500s in Tandem
For timing application requiring more than 6 timing contact channels, the user can use two or
more CT-6500s to time a circuit breaker. A typical application is to use one CT-6500-6 and one
CT-6500-3 to create a 9 contact timing application. To time this circuit breaker, the user can use
the following steps:
- Connect the CT-6500-6 to phase A and B of the circuit breaker.
- Connect the CT-6500-3 to phase C of the circuit breaker.
- Connect the CT-6500-6 initiate cable to the breaker control circuit. This CT-6500-6 will
operate the circuit breaker.
- Connect the CT-6500-3 external trigger cable to the OPEN coil for the OPEN test and
CLOSE coil for the CLOSE test. This CT-6500-3 will operate as a slave device.
- Set the CT-6500-3 to start a test using “EXTERNAL TRIGGER”.
- Set the CT-6500-6 to start a test using “INTERNAL TRIGGER”.
- Hold the “ARM” switch on the first CT-6500-6 to operate the circuit breaker.
The CT-6500-6 will operate the circuit breaker and start the timing sequence. The CT-6500-3
will start its timing sequence as soon as it senses the voltage applied to the breaker coil.
The timing report of the CT-6500-6 will show the contact time of phase A and B. The timing
report of the CT-6500-3 shows contact time for phase C.
If the user downloads the test results of both the CT-6500s to the PC, the user then can
combine the 2 test reports to generate a 9-channel test report. This report now can be printed on
the PC printer.
68
CT-6500 Series 2 Operating Instructions
10.0 CT-6500 Trouble Shooting Guide
Problems Suggestion
1. CT-6500 will not trip nor close the breaker. - Make sure you hold the “ARM” switch down during
each test.
- Check the OPEN or CLOSE fuses.
- Check the initiate leads. For a “Positive Trip” circuit,
one Close lead and one Open lead should be connected
to the Positive side of the power supply.
- Check the initiate leads. For a “Negative Trip” circuit,
one Close lead and one Open lead should be connected
to the Negative side of the power supply.
2. Getting message “TRIGGER FAULT”
“No Trigger”
When using external trigger.
3. Missing timing or transducer channels on report and
graphic output.
4. No contact time (0 ms) in Trip or Close test. - Make sure one side of the breaker in not grounded.
5. No contact time (0 ms) in Close-Open. - Make sure the breaker contacts complete the transition
6. Select Open-Close test. The breaker opens and will
not close.
7. No data or erratic reading on travel transducer. - Bad transducer. Try “TEST TRANSDUCER”
8. No activity on” V2” voltage input channel. - The “V2” voltage input channel requires a voltage
- Trigger voltage was not detected by the CT-6500.
- Check the external trigger input to the CT-6500.
- Make sure the external trigger signal is presented
when the CT-6500 displays “AWAITING TRIGGER.”
- Timing or transducer channels were not turned on.
Turn on channels by going to “SETUP”, “NEXT PAGE”,
“NUMBER OF CHANNELS”.
- Make sure the breaker operates on the Open or Close
test only. A dual operation OPEN-CLOSE or CLOSEOPEN will result in 0 ms reading.
- Damaged CT-6500 contact channels or bad cables.
Try the “CHECK HOOK-UP” test.
from the Open (initial state) to the Close, then to the
Open (final state). You can see The transition by looking
at the timing chart.
- Make sure the contact close state is longer than
10 ms. The CT-6500 has a 10ms de-bounce time, any
contact state less than or equal to 10ms is invalid, thus
the delay time is probably too slow.
- Try the Close-Open using contact channel #1 option.
- Delay between Open to Close is probably too small.
Increase delay time.
diagnostics.
- Transducer slider is not secured to timing rod.
- Transducer is not secured properly.
change from 0 V to above 30 V, to see activities on the
timing chart.
69
CT-6500 Series 2 Operating Instructions
Problems Suggestion
9. How does one review the current analysis setting? - Go to SETUP (Press key #3).
- Select Analysis (Press key #1).
- Select PRINT SETTING (Press key #3). The CT-6500
will print out the current AP1 & AP2 settings.
10. The CT-6500 will not communicate with the PC. - Make sure your PC communication port is selected
correctly.
- Make sure your serial or USB cable is connected
properly.
11. Turn on Automatic Print Results mode. - See section 9.17.
12. Where can one get some thermal paper? - Call the factory at 909-923-9390 or BG Instruments at
509-893-9881. The CT-6500 uses TP-4 thermal paper.
- In case of an emergency, thermal paper can be
purchased at Radio Shack. Radio Shack carries 4.25
inch wide thermal paper. It is not the right size but it will
work.
13. How does one darken or lighten the LCD?
14. How does one move thermal paper forward or
backward?
- Press and hold the “↑ PAPER LCD” key will darken the
LCD.
- Press and hold the “↓ PAPER LCD” key will lighten the
LCD.
- Press and release the “↑ PAPER LCD” key, the
thermal paper will advance.
- Press and release the “↓ PAPER LCD” key, the
thermal paper will retract.
70
AP2= 3.5”
CT-6500 Series 2 Operating Instructions
APPENDIX A ITE Circuit Breaker model 14.4K Timing Chart
AP1= 1.5”
Bounce Back
71
CT-6500 Series 2 Operating Instructions
APPENDIX B CLOSE-OPEN Timing Chart for ITE Circuit Breaker Model 14.4K
Contact Touch Distance
Peak To Peak Travel
72
CT-6500 Series 2 Operating Instructions
APPENDIX C CLOSE-OPEN Tabulated Report for ITE Circuit Breaker Model 14.4K
73
CT-6500 Series 2 Operating Instructions
APPENDIX D OPEN-CLOSE Timing Chart for ITE Circuit Breaker Model 14.4K
74
CT-6500 Series 2 Operating Instructions
APPENDIX E OPEN-CLOSE Tabulated Report for ITE Circuit Breaker Model 14.4K
75
CT-6500 Series 2 Operating Instructions
APPENDIX F SIEMENS TCP Breaker Velocity Calculation
76
CT-6500 Series 2 Operating Instructions
APPENDIX G SIEMENS SPS2 Breaker Velocity Calculation
77
CT-6500 Series 2 Operating Instructions
APPENDIX H CT-6500 SIEMENS SPS2 Test Plan
78
CT-6500 Series 2 Operating Instructions
APPENDIX I CT-6500 Transducer Illustration
Linear Transducer on an OCB
Rotary Transducer on Alstom HE 1014 Circuit Breaker
79
CT-6500 Series 2 Operating Instructions
Rotary Transducer on ABB HMB Mechanism
Rotary transducer on Federal Pacific Circuit Breaker
80
www.vanguard-instruments.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.