TEL: (909) 923-9390 January 2009
FAX: (909) 923-9391 Rev. 3
1520 S. Hellman Ave.
Ontario, California 91761
AUTO-OHM 100/200 Series 2 Operating Procedures
SAFETY SUMMARY
NOTICE
This manual applies to Models Auto-Ohm 100 Series 2, and Auto-Ohm 200 Series 2. The
operating procedures are virtually the same for all models. Any differences are clearly
described in the step-by-step procedures.
FOLLOW EXACT OPERATING PROCEDURES
Any deviation from the procedures described in this operator’s manual may create one or more
safety hazards, damage the Auto-Ohm, or cause errors in the test results; Vanguard Instruments
Co., Inc. assumes no liability for unsafe or improper use of the Auto-Ohm.
The following safety precautions must be observed during all phases of test set-up, test hookups,
testing, and test-lead disconnects.
SAFETY WARNINGS AND CAUTIONS
This device shall be used only by trained operators.
All circuit breakers under test shall be off line and fully isolated.
DO NOT MODIFY TEST EQUIPMENT
Because of the risk of introducing unknown hazards, do not install substitute parts or perform
any unauthorized modification to any Model Auto-Ohm Test unit. To ensure that all designed
safety features are maintained, it is recommended that repairs be performed only by Vanguard
Instruments Co. factory personnel or by an authorized repair service. Unauthorized modifications
can cause serious safety hazards and will nullify the manufacturer's warranty.
Figure 50.0 Test Record Readout............................................................................................. 29
5 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
1.0 INTRODUCTION
1.1 Applicability
This manual applies to the Model Auto-Ohm 100 Series 2 and Model Auto-Ohm 200 Series 2
(hereafter, Auto-Ohm), made by Vanguard Instruments Company.
1.2 General Description
The Auto-Ohm-100/200 Series 2 are the third generation micro-ohmmeters made by Vanguard
Instruments Company. The Auto-Ohm 100/200 features microprocessor-controlled measuring of
very low resistances ranging from 1 micro-ohm to 300 milli-ohms with high accuracy. The AutoOhm is field-portable, rugged, and easily operated by first-time users having a minimum of
training. It features one-knob control and an LCD alpha/numeric display of the resistance
measured. The one-knob control operation is logical and simple: Turning the knob scrolls
through a menu of possible options (which display in sequence) and pressing the knob activates
the selected function. As its name implies, the Auto-Ohm operation is automatic, requiring little
more from the user than connecting it to an unknown resistance and selecting the desired
functions and its options. The Auto-Ohm automatically stores the last 3 resistance measurements,
which can be displayed after testing.
1.3 Functional Description
The Auto-Ohm’s operation is based on the electrical relationships described by Ohm’s law:
R=V/I, where I is a known current and V is the dc voltage measured across the unknown
resistance (typically, a circuit breaker’s contacts). Since the current (user selected) through the
unknown resistance is known and the voltage across the unknown resistance is measured by a
precision voltmeter, the resistance read-out can be calculated using Ohm’s law.
The Auto-Ohm test voltage is supplied by a true 5Vdc power supply. The true DC test current is
selectable in 2-amp steps, from 10 to 100 amperes for the Auto-Ohm-100 and from 10 to 200
amperes for Auto-Ohm-200. The test current is automatically ramped up and down slowly. This
current ramp rate is programmable from 5 seconds to 30 seconds.
The voltmeter test leads run separately from the current-bearing test leads to the resistive load
thus eliminating any I•R voltage drop error in the current cables. These Auto-Ohm features make
very precise micro-ohm measurements possible without having to calculate compensations for
current lead resistance errors.
1.4 Furnished Test Accessories
The Auto-Ohm is supplied with two 30-foot test cables with heavy-duty alligator clamps. Both
the current (#1 AWG) and sense leads are combined into one cable (Figure 1.0). A Ground cable,
and power cord are also included with each Auto-Ohm.
Users can select to have the current and sense leads separated as shown in Figure 2.0 and Figure
4.0. Optional hand spike sense leads are also available as shown in Figure 3.0.
6 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
Current Jaw
Voltage Jaw
Figure 1.0 Combined Current and Sense Leads
Figure 2.0 Auto-Ohm 100/200 Sense Leads
7 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
Figure 3.0 Auto-Ohm 100/200 Hand Spikes
Figure 4.0 Auto-Ohm 100/200 Current Cable
8 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
1.5 Optional Accessories
1. Heavy-duty welding-type C-clamps (Figure 5.0) are available as optional accessories.
These C-clamps allow test lead connections to a wide variety of bushing sizes, bus bars,
and conductors that require low-resistance test-lead contacts.
2. Light weight (#4 AWG) cables are also available upon request.
3. Custom cable lengths are available upon request.
4. An optional shipping case (which holds the Auto-Ohm and its cables) is also available.
Figure 5.0 Auto-Ohm 100/200 C-Clamp Cable
9 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
2.0 AUTO-OHM SPECIFICATIONS
2.1 Auto-Ohm 100 Series 2 Specifications
Table 1.0 Auto-Ohm 100 Series 2 Specifications
MODEL...................... Auto-Ohm 100 Series 2
TYPE............................. Special-Purpose Test Equipment, Portable, Low Resistance-Ohmmeter
CONFIGURATION...... Third-generation (improved design, superseding original model)
SIZE (inches).............. 16.8 Wide by 12.6 High by 10.6 Deep (42.7 Cm x 32 Cm x 30.5 Cm)
WEIGHT........................ Less than 15 pounds (6.8 Kg)
OPERATING
VOLTAGE.................... 8 amps, 100-240 Vac, 50/60 Hz, with built in 10A circuit breaker
RESISTANCE
RANGE......................... 1 micro-ohm to 300 milli-ohms
TEST CURRENT
RANGE......................... 10 Amperes to 100 Amperes, Selectable in 2 ampere steps
DISPLAY...................... Backlit LCD, 2-lines high by 16 characters Wide
ACCURACY................. ± 1 % of Reading, ± 1 Count
UNIT PROTECTION... thermal-overload sensor and cutoff
INTERFACE................. RS-232C Connector Port for PC Interface
ENVIRONMENT.......... Operating: 0°C to 55°C; Storage: -40°C to 65°C
FURNISHED ITEMS... One power cord, one ground cable, two 30-ft. test lead cables.
WARRANTY................ One-Year Parts & Labor (Post-Warranty Service Contracts Available)
AUTO-OHM SPECIFICATIONS ARE SUBJECT TO UPGRADES AND MAY BE CHANGED WITHOUT PRIOR NOTICE.
10 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
2.2 Auto-Ohm 200 Series 2 Specifications
Table 2.0 Auto-Ohm 200 Series 2 Specifications
MODEL...................... Auto-Ohm 200 Series 2
TYPE............................. Special-Purpose Test Equipment, Portable, Low Resistance-Ohmmeter
CONFIGURATION...... Third-generation (improved design, superseding original model)
SIZE (inches).............. 16.8 Wide by 12.6 High by 10.6 Deep (42.7 Cm x 32 Cm x 30.5 Cm)
WEIGHT........................ Less than 21 pounds (9.5 Kg)
OPERATING
VOLTAGE.................... 8 amps, 100-240 Vac, 50/60 Hz, with built in 10A circuit breaker
RESISTANCE
RANGE......................... 1 micro-ohm to 300 milli-ohms
TEST CURRENT
RANGE......................... 10 Amperes to 200 Amperes, Selectable in 2 ampere steps
DISPLAY...................... Backlit LCD, 2-lines high by 16 characters wide
ACCURACY................. ± 1 % of Reading, ± 1 Count
UNIT PROTECTION... thermal-overload sensor and cutoff
INTERFACE................. RS-232C Connector Port for PC Interface
ENVIRONMENT.......... Operating: 0°C to 55°C; Storage: -40°C to 65°C
FURNISHED ITEMS... One power cord, one ground cable, two 30-ft. test lead cables
WARRANTY................ One-Year Parts & Labor (Post-Warranty Service Contracts Available)
AUTO-OHM SPECIFICATIONS ARE SUBJECT TO UPGRADES AND MAY BE CHANGED WITHOUT PRIOR NOTICE.
11 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
3.0 CONTROL AND DISPLAY
3.1 Auto-Ohm 100 Series 2 Front Panel
The Auto-Ohm 100 Series 2 controls and displays are shown in the control-panel illustration
below (Figure 6.0). Pointing leader lines reference each item with an index number. Each index
number is cross-referenced to a functional description in Table 3.0, which describes the function
and purpose of each item on the control panel. Although the purpose of these controls and the
display may seem obvious and intuitive, users should become familiar with them before
attempting to use the Auto-Ohm 100 Series 2. First-time users should also review and become
familiar with the Safety Summary on the front page.
Figure 6.0 Auto-Ohm 100 Series 2 Control-Panel Controls and Display
12 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
Table 3.0 Functional Description of Auto-Ohm 100 Series 2 Controls and Display
Figure 6.0
Index #
1 & 8
2
3
4
5
Adjacent Panel Marking
(Wing Nut)
100-240 Vac, 8A, 50-60 Hz
no marking
GROUND
(Wing Nut)
RS-232C
Functional Description
Current lead connectors
Input power connector with third-wire safety
ground and built-in 10A circuit breaker.
LCD; 2-line by 16-character, back-lighted,
displays menus of selections, operator entries,
and test-measurement results.
Auto-Ohm-100 Series 2 ground stud. Connect
ground stud to substation ground using
provided cable.
RS-232C interface port, 9-pin connector,
female DB type. The data rate is set to 19,200
baud, 1 start bit, 8 data bits, and no parity bit.
PIN ................SIGNAL
2 Rx
3 Tx
5 Signal Gnd
This serial port is dedicated for factory
calibration and software update.
LED indicator, red, lights when high-testcurrent is going through the test leads.
One-knob control (all Auto-Ohm menus and
selections are controlled by this one control
knob). Turning this control knob scrolls through different menu options (shown on
6
7
HIGH CURRENT
PRESENT
CHANGE
“PUSH”
TO SELECT
an LCD), which display. Pressing the knob
selects the displayed function, usually
producing a new menu of selectable options.
See Figure 12.0 for a summary of the step-bystep operating procedures.
9 & 10
(resistor symbol)
Voltage-sensing connector jacks (red).
13 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
3.2 Auto-Ohm 200 Series 2 Front Panel
The Auto-Ohm 200 Series 2 controls and displays are shown in the control-panel illustration,
Figure 7.0. Pointing leader lines reference each item with an index number. Each index number
is cross-referenced to a functional description in Table 4.0, which describes the function and
purpose of each item on the control panel. Although the purpose of these controls and the display
may seem obvious and intuitive, users should become familiar with them before attempting to
use the Auto-Ohm 200 Series 2. First-time users should also review and become familiar with
the Safety Summary on the front page.
Figure 7.0 Auto-Ohm 200 Series 2 Control-Panel Controls and Display
14 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
Table 4.0 Functional Description of Auto-Ohm 200 Series 2 Controls and Display
Figure 7.0
Index #
1 & 8
2
3
4
5
Adjacent Panel Marking
(Wing Nut)
100-240 Vac, 8A, 50-60 Hz
no marking
GROUND
(Wing Nut)
RS-232C
Functional Description
Current lead connectors
Input power connector with third-wire safety
ground and built-in 10A circuit breaker.
LCD; 2-line by 16-character; back-lighted;
displays menus of selections, operator entries,
and test-measurement results.
Auto-Ohm 200 Series 2 ground stud. Connect
ground stud to substation ground using
provided cable.
RS-232C interface port; 9-pin connector;
female DB type. The data rate is set to 19,200
baud, 1 start bit, 8 data bits, and no parity bit.
PIN ................SIGNAL
2 Rx
3 Tx
5 Signal Gnd
This serial port is dedicated for factory
calibration and software update.
LED indicator, red, lights when high-testcurrent is going through the test leads.
One-knob control (all Auto-Ohm menus and
selections are controlled by this one control
knob). Turning this control knob scrolls through different menu options (shown on
6
7
HIGH CURRENT
PRESENT
CHANGE
“PUSH”
TO SELECT
an LCD), which display. Pressing the knob
selects the displayed function, usually
producing a new menu of selectable options.
See Figure 12.0 for a summary of the step-bystep operating procedures.
9 & 10
(resistor symbol)
Voltage-sensing connector jacks (red).
15 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
4.0 OPERATING VOLTAGES
The Auto-Ohm operates on voltages between 100-240Vac, 50/60Hz.
5.0 CABLE CONNECTION
The Auto-Ohm is supplied with 30-foot test cables with heavy-duty alligator clamps. Both the
current (#1 AWG) and sense leads are combined into one cable (Figure 1.0). A typical cable
connection for the Auto-Ohm to a device under test (using the combined test leads) is shown in
Figure 9.0 and Figure 11.0. Figure 8.0 and Figure 10.0 which illustrate the connection using
separate current and sense leads.
To protect the Auto-Ohm against static discharge in the substation, always connect the unit’s
ground stud to the substation ground. It is also highly recommended to ground one side of the
circuit breaker bushing during test to eliminate any static discharge to the Auto-Ohm.
NOTE
The sense input is not polarity sensitive. The sense cables may be
connected to either input without affecting circuit operation.
Figure 12.0 Step-by- Step Procedures for Auto-Ohm Operation
18 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
6.0 OPERATING THE AUTO-OHM
The Auto-Ohm is operated with one dual-function control knob. The operator turns the control
knob to scroll through different menu selections on the display. When the desired option appears,
it is selected by pressing the control knob like a pushbutton. Review Figure 12.0 before
proceeding with the step-by-step procedures that follow.
6.1 Step-by-Step Procedures
The following sections present step-by-step procedures for the operation of the Auto-Ohm 100
and Auto-Ohm 200 Series 2.
6.2 Precautions
Do not measure the resistance of inductive devices. This can generate unsafe high-voltage spikes
(created by a collapsing magnetic field) if the test current is interrupted by detaching a test lead
during a test. Do not touch or disconnect any test lead that is connected to a device under test
while current is being conducted. Failure to heed this warning can result injury to the user and/or damage to the Auto-Ohm. The Auto-Ohm measures low, non-inductive resistances
(e.g., breaker contacts and bus-bar junctions). If the resistance of an inductive device needs to be
measured, then the use of an instrument designed for that purpose is recommended (such as the
WRM made by the Vanguard Instruments Company).
6.3 Preparations
a. Ground Auto-Ohm to Substation ground.
b. Plug the Auto-Ohm power cable into a power outlet.
c. Connect current-cable lugs and voltage-sensing cable plugs to control-panel (Figure 6.0 and
Figure 7.0).
d. Attach current test-cable clamps to opposite terminals of the resistive load being tested (Figure
8.0 to Figure 11.0).e. If separate cables are used, attach voltage-sensing clamps to the terminals of the resistive load.
Sensing voltage clamps should be inside the current clamps.
f. To turn on Auto-Ohm power, press the rocker switch to ON.
NOTE
All Auto-Ohm operations begin at the MAIN MENU, which appears after the initial
boot-up (after configuration and software revision data display briefly.) The Main
Menu display is shown below.
MAIN MENU
<RUN TEST>
Figure 13.0 Main Menu
g. The main menu displays a list of four options, which appear in sequence as the control knob is
turned. The four options of the MAIN MENU list are RUN TEST, ADJ CONTRAST, 100A CAL. CHECK, and PREVIOUS RESULTS.
19 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
h. When the option of choice appears in the MAIN MENU, press (or “Push”) the control knob
to enter the selection and start that sequence. The step-by-step operating procedures to follow
describe each of the selected options in the order listed above.
6.4 Running a Normal Test Procedure
To run a test, turn the control knob until RUN TEST appears on the display, then press the
control knob to begin the procedures for running a test.
The TEST TYPE menu allows the user to select NORMAL Test or AUTOMATIC Test mode.
NORMAL test mode allows the user to select test current, ramp time, and burn-in time. Press the control knob when the Auto-Ohm displays <NORMAL> under the TEST TYPE menu.
Figure 14.0 Test Type Menu <NORMAL>
The SELECT TEST CUR (current) menu will now appear on the Auto-Ohm’s display.
Figure 15.0 Select 10A Menu
a. This screen prompts the user to select the desired test current. Test current options are 10, 25,
50, and 100 Amp (for Auto-Ohm 100) or 10, 25, 50, 100, and 200 Amp (for Auto-Ohm 200),
CUSTOM, and ABORT TEST. Select the desired test current option by turning the control knob
until the proper current level displays, then press the knob to enter that choice (in which case go
to step d). If a smaller increment of current is desired, select CUSTOM (Figure 16.0) and press
the control knob and go to the next step.
Figure 16.0 Select Custom Current Menu
b. This menu allows the user to select any test current from 10 to 100 amperes (200 amperes for
Auto-Ohm 200) in 2 amperes steps.
SELECT TEST CUR
Figure 17.0 Custom Current Menu
TEST TYPE
<NORMAL>
SELECT TEST CUR
<10 AMPS>
<CUSTOM>
SET CURRENT:
50 AMPS
20 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
c. Custom current is defaulted to 50 amperes for the Auto-Ohm 100 (100A for Auto-Ohm 200)
in this menu. Turning the control knob clockwise or counter-clockwise will increment or
decrement the current reading on the display. Turn the control knob until the desired current is
displayed, then press the control knob to enter the desired current. The display will now show the
BURN IN TIME menu (Figure 18.0).
NOTE
Selecting ABORT TEST causes the display to return to
the Main Menu without making any selections.
BURN IN TIME
<5 SEC>
Figure 18.0 The BURN-IN TIME Menu
d. This menu selects the amount of time the burn-in current will be run through the test load
before its resistance is measured. Burn-in time range is from 5 to 60 seconds (in 5-second steps).
The default burn-in time is set for 5 seconds. Turn the control knob until the desired burn-in time
appears in the display, then press the control knob to enter that time. Go to the next step.
RAMP TIME:
< 5 SEC>
Figure 19.0 The RAMP TIME Menu
e. The display above shows the ramp time. The ramp time range is 5 to 30 seconds. Turn the
control knob until the desired ramp time appears in the display, then press the control knob to
select that time. Go to the next step.
50 AMPS 5 SEC
“PRESS” TO RUN
Figure 20.0 Test Current & Burn-In Time
f. The display above shows the current and time that have been selected. If these are the intended
current and time, press the control knob to accept the selections and begin the burn-in and
measurement sequence. Go to the next step. Turning the knob in either direction will return the
display to the MAIN MENU.
CUR RAMP: 20%
Figure 21.0 Current Ramp Message
21 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
g. The display above shows the current ramp percentage of the test-level current. The Auto-Ohm
will ramp the test current from 0 (zero) amperes to the selected test current level in the selected
ramp time. The test current is automatically ramped up and down. This current ramp rate is
programmable from 5 seconds to 30 seconds. When the current reaches 100%, the next display
(shown in Figure 22.0) appears automatically. This display shows the resistance reading and the
remaining burn-time of the test.
BURNING IN: 02
120.1 MICRO-OHM
Remaining burnin time
Figure 22.0 The BURNING IN Menu
I=100 AMPS
120 MICRO-OHM
Test Current
Final Resistance
Reading
Figure 23.0 Test Current & Resistance Readings
h. The Auto-Ohm then ramps the test current back to zero. Figure 23.0 shows the final resistance
measurement and the test current at which the resistance was measured. After the result is
examined, press or turn the control knob for the REPEAT TEST menu.
REPEAT TEST?
<YES>
Figure 24.0 The REPEAT TEST Menu
i. To repeat the test, press then release the control knob. The Auto-Ohm repeats the test using the
same test current, ramp time, and burn-in time settings.
To exit this test mode, turn the control knob to select <NO>. Press the control knob to return to
the MAIN MENU.
From the Main Menu another test can be run. If all tests have been completed, turn off power to
the Auto-Ohm, disconnect test leads and power cable, and stow them. This completes the
procedure for performing the NORMAL test procedure.
NOTE
A CABLE ERROR message will be displayed on the Auto-Ohm LCD if the current
cables or sense cables are not connected as shown in Figure 8.0 to Figure 11.0.
CABLE ERROR!
Figure 25.0 Cable Error Message
22 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
6.5 Running an Automatic Test Procedure
The TEST TYPE menu allows the user to select the Normal or Automatic test mode. In the
Automatic test mode the user selects the test current and ramp time. The measurement cycle is
initiated when the sense leads are connected to the device under test. This feature is handy when
the user wants to take resistance readings of multiple points.
To use the Automatic Test Mode the user must connect the current cables outside the two points
where the measurement is to be made. The Auto-Ohm is then placed in the Automatic Test
mode. The test will start as soon as the sense cables are connected to the two points of interest.
The Auto-Ohm test current will ramp from zero to the selected test current. After a five second
burn-in time, the Auto-Ohm current is ramped back to zero. The resistance reading is then
displayed on the LCD. To start a new test, disconnect, then reconnect the sense leads between
the same or two new points of interest.
NOTE
•Current cables should be connected across the resistive load to establish the
current path. Removing and reconnecting one or both sense cables starts a new test.
• Hand spikes (see Figure 3.0) are designed specifically for this application.
To select the Automatic mode, turn the control knob in the TEST TYPE menu until
<AUTOMATIC> menu is displayed (Figure 26.0). Press the control knob to select the
Automatic test mode.
Figure 26.0 Test Type Menu <AUTOMATIC>
The SELECT TEST CUR (current) menu will now appear on the Auto-Ohm’s display.
Figure 27.0 Select 10A Menu
a. This screen prompts the user to select the desired test current. Test current options are: 10, 25,
50, 100 Amp (for Auto-Ohm 100) or 10, 25, 50, 100, 200 Amp (for Auto-Ohm 200), CUSTOM,
and ABORT TEST. Select the desired test current option by turning the control knob until the
proper current level displays, then press the knob to enter that choice (in which case, go to step
d). If a smaller increment of current is desired, select <CUSTOM> (Figure 28.0), press the
control knob and go to the next step.
Figure 28.0 Select Custom Current Menu
SELECT TEST CUR
TEST TYPE
<AUTOMATIC>
SELECT TEST CUR
<10 AMPS>
<CUSTOM>
23 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
b. This menu allows the user to select any test current from 10 to 100 amperes (or 200 amperes
for Auto-Ohm 200) in 2 amperes steps.
SET CURRENT:
50 AMPS
Figure 29.0 Custom Current Menu
c. Custom current is defaulted to 50AMPS for the Auto-Ohm 100 (100A for Auto-Ohm 200) in
this menu. Turning the control knob clockwise or counter-clockwise will increment or decrement
the current reading on the display. Turn the control knob until the desired current is displayed,
then press the control knob to enter the desired current. The RAMP TIME menu (Figure 30.0)
will appear.
RAMP TIME:
<5 SEC>
Figure 30.0 The Ramp Time Menu
d. The display above shows the Ramp Time. The Ramp Time range is 5 to 30 seconds. Turn the
control knob until the desired ramp time appears in the display, then press the control knob to
select that time. Go to the next step.
50 AMPS AUTO
“PRESS” TO RUN
Figure 31.0 Select 50 Amperes Menu
e. The display above (Figure 31.0) shows the selected test current. Press the control knob to start
the AUTO TEST mode (Figure 32.0).
AUTO TEST MODE
Figure 32.0 Auto Test Mode Menu
f. The Auto-ohm is ready to start testing when the Auto Test Mode menu shown above is
displayed. To start testing, connect the sense leads across the device under test. The test current
is gradually ramped up and down with a 5 second burn-in test time. Test results will be displayed
as shown in Figure 33.0 below.
24 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
I=10 AMPS
101.0 MICRO-OHM
Figure 33.0 Automatic Mode Test results
g. To initiate another test, disconnect and reconnect the sense leads to the device under test. To
terminate the Automatic Test Mode, press the Auto-Ohm control knob.
NOTE
The user should allow two seconds between disconnecting and reconnecting to the test
leads to initiate a new test.
6.6 Auto-Ohm Quick Test Procedure
The Auto-Ohm Quick Test mode is a preset test that runs a test at 100 Amps for 5 seconds. To
start a quick test, go to the RUN TEST option in the MAIN MENU. Press and hold the control
knob for 2 seconds. A short beep will be heard at which point the knob must be quickly released.
The Auto-Ohm will beep twice to acknowledge the quick test command has been accepted. Press
then release the control knob again to initiate the test. The Auto-Ohm will run a 100A test with a
5-second burn in time (Figure 34.0). The displayed test result will resemble Figure 33.0.
100 AMPS 5 SEC
“PRESS” TO RUN
Figure 34.0 Quick Test Menu
6.7 Contrast Adjustment
The purpose of this procedure is to adjust the darkness level of the alpha-numeric characters
shown in the LCD display, in order to produce the best readability for the ambient light in the
testing area. To adjust the contrast, turn the control knob to select the ADJ. CONTRAST option
from the MAIN MENU as shown in Figure 35.0.
MAIN MENU
<ADJ. CONTRAST>
Figure 35.0 Select Adjust Contrast Menu
a. Press the control knob to select ADJ. CONTRAST mode. The following menu (Figure 36.0)
will be shown.
ADJUST CONTRAST
“PRESS”= DONE
Figure 36.0 Contrast Menu
25 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
b. Turn the control knob until the contrast is suitable. Press the control knob to set the contrast.
The display will return to the MAIN MENU. The contrast setting will be recalled each time the
Auto-Ohm is turned on and remain until it is changed again using this procedure.
6.8 Auto-Ohm Calibration Check
The purpose of the Calibration Check is to verify that the Auto-Ohm is operating within
acceptable specifications by running a functional check on the Auto-Ohm electronics.
To run a Calibration Check, turn the control knob to select the 100A CAL CHECK option from
the MAIN MENU (see Figure 37.0).
MAIN MENU
<100A CAL CHECK>
Figure 37.0 100A Cal Check
a. Press the control knob to begin the calibration check. The following display will appear.
ATTACH SHORT BAR
<START CAL CHK>
Figure 38.0 Attach Short Bar Prompt
b. Attach the test leads to a piece of aluminum or copper bar (several inches apart, the spacing is
not critical, since this is a functional check). See Figure 39.0 and Figure 40.0 below for a
connection illustration. Press the control knob to start the test.
c. During the calibration check the display below (Figure 41.0) shows the current ramping status.
When the test current ramps to 100 %, this display is automatically replaced with the test result
display of Figure 42.0 or Figure 43.0.
Figure 41.0 Current Ramp Message
Figure 42.0 Current-Ramp Error Message
d. If the current did not ramp properly (test failed) the CUR RAMP ERROR display above
(Figure 42.0) will appear. Press the control knob to abort the test (restart the CAL CHECK
when the problem is corrected).
Figure 43.0 Current Ramp Circuit Pass Message
e. The above status display indicates that the current ramped properly. The status displays of
Figure 44.0 through Figure 46.0 show for about 3 seconds each as each test is performed and
passed. If any test fails, the status display will stop at the failed function and indicate “FAIL” on
CUR RAMP: 20%
CUR RAMP ERROR!
CHECK CABLES
CURRENT RAMP CKT
“PASS”
27 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
the display. The Auto-Ohm will continue to beep until the control knob is pressed. Once the
problem has been fixed, the CAL CHECK can be rerun.
Figure 44.0 Zero Circuit Test Message
f. The Auto-Ohm checks the ZERO circuit. The “PASS” message is displayed as shown above
in Figure 44.0.
Figure 45.0 Full-Scale Circuit Test Message
g. The full scale range check (FSCL CKT CHECK) is performed next. The “PASS” message is
displayed as shown above in Figure 45.0.
h. The measure circuit test (MEASURE CKT CHECK) is performed next. The “PASS”
message displays as shown below in Figure 46.0.
Figure 46.0 Measure Circuit Test Message
i. If all of the calibration functions pass, the Auto-Ohm calibration complete m essage appears as
shown in Figure 47.0.
Figure 47.0 Cal Check Complete Message
This completes the Calibration Check procedure. Press the control knob to return to the MAIN
MENU.
ZERO CKT CHECK
“PASS”
FSCL CKT CHECK
“PASS”
MEASURE CKT CHECK
“PASS”
CAL CHECK DONE!
PRESS KEY……....
28 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
6.9 Display Previous Results
The purpose of this procedure is to let an operator view the last 3 readings stored in the AutoOhm. To view previous results, turn the control knob to select the PREV RESULTS option
from the MAIN MENU, as shown in Figure 48.0.
Figure 48.0 Previous Results Menu
a. Press the control knob to select this option. The user now can select any of the last three
readings to be displayed. To select the reading, turn the control knob to one of the menus.
Figure 49.0 Select Reading Menus
b. When one of the prompts above displays, press the control knob to display its reading. The
test results will be displayed as shown in Figure 50.0 below.
Figure 50.0 Test Record Readout
c. The above display shows the recorded test resistance (120.2 Micro-ohms) and the test current
level (100 amps) at which it was measured. When the displayed record of resistance is reviewed
and noted, press the control knob to return to the MAIN MENU. This ends the PREVIOUS RESULTS procedure.
MAIN MENU
<PREV RESULTS>
PREVIOUS RESULTS
<LAST TEST>
PREVIOUS RESULTS
<SECOND TO LAST>
PREVIOUS RESULTS
<THIRD TO LAST>
I= 100AMPS
120.2 MICRO-OHM
NOTE
To display previous test results, the Auto-Ohm must contain tests performed since it was last
(most recently) powered on. When the Auto-Ohm is powered off, it loses all test information.
This concludes the operating procedures for all of the Auto-Ohm functions.
29 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
APPENDIX A
Auto-Ohm Troubleshooting Guide
Item Symptom Possible Problem Solution
1 Reading is
incorrect.
1. Poor connection at
the test clamps.
1. Check connections to
ensure teeth of voltagesensing and current clamps
are firmly in contact with
the device under test.
2 “Cable Error”
Message.
1. No test current
going through the
device under test.
2. Sensing cables
problem.
1. Check current cable
Connection to the device
under test.
2. Check sensing cable
connection.
3. Run Calibration Test.
30 Rev 3, 2009
AUTO-OHM 100/200 Series 2 Operating Procedures
1520 S. Hellman Ave., Ontario, CA 91761, USA
Phone 909-923-9390 Fax 909-923-9391
Web site: http//www.vanguard-instruments.com
Copyright 2009 by Vanguard Instruments Company, Inc.
Auto-Ohm 100/200 Series 2 HPN Jan 2009
31 Rev 3, 2009
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.