Vaisala Ozone Interface Kit RSA921
www.vaisala.com
Features / Benefits
Built around Vaisala Digital
▪
Interface OIF92
Four-channel digital interface
▪
for connecting digital Vaisala
Radiosonde RS92 models to
ECC-type ozone sensors for
ozone sounding
Enables accurate
▪
measurements, sharp vertical
resolution and a fast sampling
rate
Can be used for 1-2 other
▪
sensor measurements within
specified ranges
The Vaisala Ozone Interface Kit
RSA921, featuring the Vaisala Digital
Interface OIF92, is used for ozone
sounding with the digital versions
of the Vaisala Radiosonde RS92 and
an ECC-type ozone sensor (Science
Pump Corporation model ECC-6A or
EN-SCI Corporation model Z). The
OIF92 is powered by the radiosonde
battery and has four channels. Two
channels are dedicated to the ozone
sensor current and temperature. Two
channels are dedicated to voltage
measurement for other purposes.
With the OIF92, an ECC-type
ozone sensor and a digital RS92
radiosonde it is possible to measure
humidity, pressure, temperature
and geopotential height while
measuring the vertical distribution
of atmospheric ozone. Winds in the
upper atmosphere are detected using
GPS navigation signals.
Ground equipment
The OIF92 and digital RS92
radiosondes are used with the
Vaisala DigiCORA® Sounding System
MW31. The ozone data is stored to
le for further processing by the
user. It is easy to perform simulations
and post-ascent processing with the
METGRAPH module of the DigiCOR A®
Sounding Software. System
maintenance can be performed by
Vaisala under the terms of a Vaisala
Service Contract.
Technical data
Digital Interface OIF92
For use with Science Pump Corporation ECC-6A
ozone sensor and
EN-SCI Corporation Model Z ozone sensor
Dimensions 40 (l) x 35 (w) mm
Cable length 380 mm
Weight with cable Max. 20 g
Number of analog input channels 4
Analog-to-digital converter resolution 16-bit
Measurement temperature range for all
channels -5 ... +60 °C
Operating temperature range for all
channels - 40 ... +85 °C
Sampling rate for all channels Dependent upon the
radiosonde:
e.g. once per second with the RS92-SGP
Synchronization All channels are measured in synchronization
with the meteorological measurements
(pressure, temperature, humidity, wind)
Ozone current measurement range 0 ... 20 µA
Ozone current measurement uncertainty *)
0.05% of the reading,
minimum 1 nA
Accuracy examples:
– for 5 µA measurement the accuracy is 2.5 nA
– for 10 µA measurement the accuracy is 5 nA
– for 20 µA measurement the accuracy is 10 nA
Resolution 1 nA (LSB = 0.35 nA)
Typical noise 1 nA (standard deviation)
Ozone temperature measurement range -5 ... +60 °C
Temperature measurement uncertainty*) ±0.2 °C
Temperature resolution 0.01 °C
Typical noise 0.015 °C (standard deviation)
Voltage measurement range
(both voltage channels) 0 ... 8 V
Voltage measurement uncertainty*) 0.1 % of the reading,
minimum 1 mV
Resolution 1 mV (LSB = 0.13 mV)
Typical noise 0.5 mV (standard deviation)
Input resistance 20 kΩ
Power consumption < 3 mA (from radiosonde battery)
* 2-sigma (k=2) confidence level (95.5 %)
For more information, visit
www.vaisala.com or contact
us at sales@vaisala.com
Ref. B210447EN-B ©Vaisala 2010
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.