Vaisala RSA921 Technical data

Vaisala Ozone Interface Kit RSA921
www.vaisala.com
Features / Benefits
Built around Vaisala Digital
Interface OIF92
Four-channel digital interface
for connecting digital Vaisala Radiosonde RS92 models to ECC-type ozone sensors for ozone sounding
Enables accurate
measurements, sharp vertical resolution and a fast sampling rate
Can be used for 1-2 other
sensor measurements within specified ranges
The Vaisala Ozone Interface Kit RSA921, featuring the Vaisala Digital Interface OIF92, is used for ozone sounding with the digital versions of the Vaisala Radiosonde RS92 and an ECC-type ozone sensor (Science Pump Corporation model ECC-6A or EN-SCI Corporation model Z). The OIF92 is powered by the radiosonde battery and has four channels. Two channels are dedicated to the ozone sensor current and temperature. Two channels are dedicated to voltage measurement for other purposes.
With the OIF92, an ECC-type ozone sensor and a digital RS92 radiosonde it is possible to measure humidity, pressure, temperature and geopotential height while
measuring the vertical distribution of atmospheric ozone. Winds in the upper atmosphere are detected using GPS navigation signals.
Ground equipment
The OIF92 and digital RS92 radiosondes are used with the Vaisala DigiCORA® Sounding System MW31. The ozone data is stored to le for further processing by the user. It is easy to perform simulations and post-ascent processing with the METGRAPH module of the DigiCOR A® Sounding Software. System maintenance can be performed by Vaisala under the terms of a Vaisala Service Contract.
Technical data
Digital Interface OIF92
For use with Science Pump Corporation ECC-6A
ozone sensor and
EN-SCI Corporation Model Z ozone sensor Dimensions 40 (l) x 35 (w) mm Cable length 380 mm Weight with cable Max. 20 g Number of analog input channels 4 Analog-to-digital converter resolution 16-bit Measurement temperature range for all channels -5 ... +60 °C Operating temperature range for all channels - 40 ... +85 °C Sampling rate for all channels Dependent upon the
radiosonde:
e.g. once per second with the RS92-SGP
Synchronization All channels are measured in synchronization
with the meteorological measurements
(pressure, temperature, humidity, wind)
Ozone current measurement range 0 ... 20 µA
Ozone current measurement uncertainty *)
0.05% of the reading, minimum 1 nA
Accuracy examples: – for 5 µA measurement the accuracy is 2.5 nA – for 10 µA measurement the accuracy is 5 nA – for 20 µA measurement the accuracy is 10 nA Resolution 1 nA (LSB = 0.35 nA) Typical noise 1 nA (standard deviation) Ozone temperature measurement range -5 ... +60 °C Temperature measurement uncertainty*) ±0.2 °C Temperature resolution 0.01 °C Typical noise 0.015 °C (standard deviation) Voltage measurement range (both voltage channels) 0 ... 8 V Voltage measurement uncertainty*) 0.1 % of the reading,
minimum 1 mV
Resolution 1 mV (LSB = 0.13 mV) Typical noise 0.5 mV (standard deviation) Input resistance 20 kΩ Power consumption < 3 mA (from radiosonde battery)
* 2-sigma (k=2) confidence level (95.5 %)
For more information, visit www.vaisala.com or contact us at sales@vaisala.com
Ref. B210447EN-B ©Vaisala 2010
This material is subject to copyright protection, with all copyrights retained by Vaisala and its individual partners. All rights reserved. Any logos and/or product names are trademarks of Vaisala or its individual partners. The reproduction, transfer, distribution or storage of information contained in this brochure in any form without the prior written consent of Vaisala is strictly prohibited. All specifications — technical included — are subject to change without notice.