or by any means, electronic or mechanical (including photocopying), nor may its
contents be modified, translated, adapted, sold or disclosed to a third party without prior
written permission of the copyright holder. Translated manuals and translated portions
of multilingual documents are based on the original English versions. In ambiguous
cases, the English versions are applicable, not the translations.
The contents of this manual are subject to change without prior notice.
This manual does not create any legally binding obligations for Vaisala towards
customers or end users. All legally binding obligations and agreements are included
exclusively in the applicable supply contract or the General Conditions of Sale and
General Conditions of Service of Vaisala.
Chapter 1 _________________________________________________________ General Information
CHAPTER 1
GENERAL INFORM ATION
About This Manual
This manual provides information for installing, operating, and
maintaining the Vaisala Combined Pressure, Humidity, and Temperature
Transmitter PTU300.
Contents of This Manual
- Chapter 1, General Information, provides general notes for the manual
and the product.
- Chapter 2, Product Overview, introduces the features, advantages, and
the product nomenclature of PTU300.
- Chapter 3, Installation, provides you with information that is intended
to help you install the product.
- Chapter 4, Operation, contains information that is needed to operate
this product.
- Chapter 5, Modbus, contains information that is needed when
operating the transmitter using the Modbus protocol.
- Chapter 6, Maintenance, contains information that is needed in basic
maintenance of the product.
- Chapter 7, Calibration and adjustment, provides information and
instructions concerning calibration and adjustment of PTU300.
- Chapter 8, Technical data, provides the technical data of the product.
- Appendix A, Probe installation kits and installation examples, presents
the installation kits available for PTU300 and provides some
installation examples.
- Appendix B, Calculation Formulas, presents the formulas used for the
calculated output quantities.
- Appendix C, Modbus Reference, describes the Modbus functions and
data of the transmitter.
Chapter 1 _________________________________________________________ General Information
Ground the product, and verify outdoor installation grounding
periodically to minimize shock hazard.
Do not modify the unit. Improper modification can damage the product,
lead to malfunction, or make the product
legislation.
Safety
The PTU300delivered to you has been tested for safety and approved as
shipped from the factory. Note the following precautions:
WARNING
CAUTION
noncompliant with applicable
ESD Protection
Electrostatic Discharge (ESD) can cause immediate or latent damage to
electronic circuits. Vaisala products are adequately protected against
ESD for their intended use. However, it is possible to damage the product
by delivering electrostatic discharges when touching, removing, or
inserting any objects inside the equipment housing.
To make sure you are not delivering high static voltages yourself:
- Handle ESD sensitive components on a properly grounded and
protected ESD workbench. When this is not possible, ground yourself
to the equipment chassis before touching the boards. Ground yourself
with a wrist strap and a resistive connection cord. When neither of the
above is possible, touch a conductive part of the equipment chassis
with your other hand before touching the boards.
- Always hold the boards by the edges and avoid touching the
component contacts.
Chapter 1 _________________________________________________________ General Information
Type
PTU300
Temperature
B
Humidity
B
Vibration
A
EMC
B
Enclosure
B/IP65
DNV Type Approval
The Vaisala Combined Pressure, Humidity, and Temperature Transmitter
PTU300 is found to comply with Det Norske Veritas' Rules for
Classification of Ships, High Speed & Light Craft and Det Norske
Veritas' Offshore standards.
Applicable tests carried out according to Standard for Certification
No. 2.4, April 2006.
Table 3 Application, Location Classes
TYPE APPROVED PRODUCT
CERTIFICATE NO.: A-13529
Transmitters with LAN or WLAN Interface
This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to Part 15 of the FCC Rules. These limits
are designed to provide reasonable protection against harmful
interference in a residential installation. Operation is subject to the
following two conditions: (1) this device may not cause interference, and
(2) this device must accept any interference, including interference that
may cause undesired operation of the device.
This equipment generates, uses and can radiate radio frequency energy
and, if not installed and used in accordance with the instructions, may
cause harmful interference to radio communications. However, there is
no guarantee that interference will not occur in a particular installation. If
this equipment does cause harmful interference to radio or television
reception, which can be determined by turning the equipment off and on,
the user is encouraged to try to correct the interference by one or more of
the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
Transmitters with WLAN Interface
This device has been designed to operate with a 2 dBi half-wave antenna.
Antennas with a gain greater than 2 dBi are prohibited for use with this
device. The antenna impedance is 50 ohms.
To reduce potential radio interference to other users, the antenna type and
its gain should be so chosen that the equivalent isotropically radiated
power (EIRP) is not more than that permitted for successful
communication.
This Class [B] digital apparatus complies with Canadian ICES-003.
Cet appareil numérique de la classe [B] est conforme à la norme NMB-
003 du Canada.
Patent Notic e
The Vaisala HUMICAP® Humidity, and Temperature Transmitter Series
PTU300 is protected by, for example, the following patents and their
corresponding national rights:
Finnish patents 98861 and 99164, French patents 6650303 and 9504397,
German patents 69418174 and 19513274, Japanese patents 3585973 and
2801156, UK patents 0665303 and 2288465, and U.S. patent 5607564.
Trademarks
HUMICAP® and BAROCAP® are registered trademarks of Vaisala Oyj.
All other trademarks are the property of their respective owners.
Chapter 1 _________________________________________________________ General Information
Software License
This product contains software developed by Vaisala. Use of the software
is governed by license terms and conditions included in the applicable
supply contract or, in the absence of separate license terms and
conditions, by the General License Conditions of Vaisala Group.
Warranty
Visit our Internet pages for standard warranty terms and conditions:
www.vaisala.com/warranty.
Please observe that any such warranty may not be valid in case of
damage due to normal wear and tear, exceptional operating conditions,
negligent handling or installation, or unauthorized modifications. Please
see the applicable supply contract or Conditions of Sale for details of the
warranty for each product.
This chapter introduces the features, advantages, and the product
nomenclature of the Vaisala Combined Pressure, Humidity, and
Temperature Transmitter PTU300.
Introduction to PTU300
The PTU300 transmitter provides reliable pressure measurement in a
wide range of applications. PTU300 series transmitters use a
BAROCAP® silicon capacitive absolute sensor developed by Vaisala for
barometric pressure measurement applications. The measurement
principle of the PTU300 series digital transmitters is based on an
advanced RC oscillator and three reference capacitors against which the
capacitive pressure sensor is continuously measured. The microprocessor
of the transmitter performs compensation for pressure linearity and
temperature dependence.
The applications of the PTU300 range from calibration laboratory
environmental condition monitoring to laser interferometer active
wavelength compensation and GPS meterorological measurements.
The PTU300 transmitter combines three measurement parameters:
pressure, temperature and humidity. The quantities measured and
calculated by PTU300 are presented in Table 4 on page 22. The
quantities available as an option are presented in Table 5 on page 22.
Analog outputs can be chosen between current and voltage signals.
Alternatively, digital outputs RS-232 (standard) or RS-422/485 (optional)
can be selected. A local display is also available.
The PTU300 transmitter is available with one or two pressure
transducers. The PTU301, PTU303 and PTU307 probes are available for
the PTU300 transmitter. A PTU30T model with only the temperature
probe (no humidity measurement) is also available.
In outdoor applications it is recommended to use the HMT330MIK
mounting kit with the PTU300 transmitters.
The PTU301 is intended for wall-mounted applications. The standard
version has a fixed probe.
1104-079
Figure 3 PTU301 Fixed Probe
The PTU301 short cable probe is a special version for use with the
WLAN module, and when LAN module and power supply module are
simultaneously installed. It has the general use PTU303 probe on a short
cable (21 cm), and a mounting plate with a probe holder.
Temperature difference between the probe and external environment can
cause a risk of condensation on the sensor. A wet probe cannot observe
the actual humidity in the ambient air. If the condensed water is
contaminated, the life span of the probe may shorten and calibration may
change.
PTU307 probe shall be used in applications where condensation can
occur due to high humidity and rapid humidity changes. The warmed
probe is heated continuously so that its temperature is always higher than
that of the environment. This prevents condensation on the probe. The
power consumption of the warmed probe is slightly higher than that of
the other probes.
2 = Mount the plate to wall with 4 screws M6 (not provided)
3 = Arched side up
4 = Fasten PTU300 to the mounting plate with 4 fixing screws M3
(provided)
5 = Holes for wall/junction box mounting
183 (7.20)
133 (5.24)
116 (4.57)
104 (4.09)
84 (3.30)
59 (2.32)
59 (2.32)
Ø6.2 (0.24)
4.5 (0.18)
10.5 (0.41)
6.5 (0.26)
8.5 (0.33)
Thickness
9.5 (0.37)
Wall Mounting with Wall Mounting Kit
When mounting with wall mounting kit the mounting plate (Vaisala
order code 214829) can be installed directly on wall or onto a standard
wall box (also US junction box). When wiring through back wall, remove
the plastic plug from the wiring hole in the transmitter before mounting.
0503-004
Figure 7 Mounting with Wall Mounting Kit
Figure 8 Dimensions of the Plastic Mounting Plate (mm/inch)
The PTU301 short cable probe is designed to be wall mounted with the
probe holder plate (Vaisala order code 226252). The probe holder plate is
similar to the standard mounting plate, except for the probe holder at the
bottom.
0911-061
Figure 9 Dimensions of the Probe Holder Plate (mm/inch)
1 = U-bolts (2 pcs) M8 (provided) for 30 ... 102 mm poles
2 = Mounting nuts M8 (4 pcs)
Mounting with DIN Rail Installation Kit
DIN rail installation kit includes a wall mounting kit, 2 clip-fasteners and
2 screws M4 x 10 DIN 7985 (Vaisala order code 215094).
1. Attach two spring holders to the plastic mounting plate by using the
screws provided in the installation kit.
2. Fasten PTU300 to the plastic mounting plate with 4 screws
(provided).
3. Press the transmitter onto the DIN rail so that the clip-fasteners
snap into the rail.
0503-002
Figure 10 Mounting with DIN Rail Installation Kit
Pole Installation with Installation Kit for
Pole or Pipeline
Installation kit for pole or pipeline (Vaisala order code: 215108) includes
the metal mounting plate and 4 mounting nuts for pole mounting. When
mounting, the arrow in the metal mounting plate must point upward;
see Figure 13 on page 33.
To enable a neat and dirt free embedded installation of the transmitter, a
panel mounting frame is available as an option (Vaisala order code:
216038). The frame is a thin, flexible plastic frame for the transmitter,
with adhesive tape on one side.
The frame is used to hide any rough edges of the installation hole, and
provide a more finished look. Note that the panel mounting frame is not
intended to bear the weight of the transmitter, and does not include any
mounting supports.
Use the panel mounting frame as follows:
1. Use the frame as a template to mark the required size for the
installation hole in the panel.
2. Cut the hole in the panel.
3. Mount the transmitter through the panel with suitable supports.
4. Remove the paper protecting the adhesive tape on the frame, and
attach the frame around the transmitter. Refer to Figure 16 on page
3 = Cable for optional power supply module Ø8 ... 11 mm
When there is high electric noise level (for example,
electric motor) in the operating environment it is recommended to use
shielded cable or take care that the signal cables are separated from other
cables.
Wiring
Cable Bushings
A single electrical cable with a screen and three to ten wires is
recommended for power and analog/serial connections. The cable
diameter should be 8 ... 11 mm. The number of cable bushings depends
on the transmitter options. See the following recommendations for the
cable bushings:
In case you need to ground the transmitter housing, the grounding
connector is found inside the housing; see Figure 2 on page 25. Make
sure that the probe is connected to the same potential as the housing, and
that different groundings are made to the same potential. Otherwise
harmful ground currents may be generated.
If it is needed to have galvanic isolation of the power supply line from
the output signals, PTU300 can be ordered with optional galvanic
isolation module. This module prevents harmful grounding loops.
Alternate Wiring Systems
There are three optional ways to connect the transmitter: using basic
wiring, using 8-Pin connector, or using D-9 connector.
The wiring system is selected when ordering the device. If a connector is
needed for wiring, it is set at the factory.
- When using basic wiring, see section Signal and Power Supply
Wiring.
- When using 8-Pin connector, see section 8-Pin Connector on page 41.
- When using D-9 connector, see section D-9 Connector on page 42.
When wiring the power supply module, see section Power Supply
Module on page 48.
0506-028
WARNING
Figure 20 Screw Terminal Block on Motherboard
-energized wires.
1. Open the transmitter cover by taking out the four cover screws.
2. Insert the power supply wires and signal wires through the cable
bushing in the bottom of the transmitter; see the grounding
instructions in the previous sections.
3. Connect the analog output cables to terminals: Ch1 +, Ch1-, Ch2+, Ch2-. Connect the RS-232 user port cables to terminals RxD, GND
and TxD. For more information about the RS-232 connection refer
to section Serial Line Communication on page 81.
4. When wiring the optional modules, see the corresponding section
for instructions:
To prev
connected to a "
must connect the same wire on the "
Connections to a 24 VAC Power Supply
Separate floating supply for each transmitter is recommended (see the
upper part of Figure 23 below). If you have to connect several
transmitters or other instruments to one AC supply, the phase (~) must
always be connected to the (+) connector of each transmitter (see the
lower part of Figure 23).
CAUTION
VAC POWER SUPPLY USE
ent fire and/or damage, if either 24 VAC wire is grounded or
-", "0", or "GND" terminal of any other device, you
In humidity measurement and especially in calibration it is essential that
temperature of the probe and measuring environment is the same. Even a
small difference in temperature between the environment and the probe
causes an error. As the curve below shows, if the temperature is +20 °C
and the relative humidity 100 %RH, a difference of ±1 °C between the
environment and the probe causes an error of ±6 %RH.
The graph below illustrates the measurement error at 100 %RH when the
difference between the ambient and sensor temperature is 1 °C.
0507-023
Figure 24 Measurement Error at 100 %RH
General Instruct ions for Probes with
Cable
Mount the probes with a cable with the probe horizontally; this way, any
water condensing on the tube cannot flow onto the sensor.
When there is no alternative but to install the probe in the process
vertically, the point of entry must be carefully insulated. The cable must
also be allowed to hang loosely as this prevents any condensed water
from running onto the probe along the cable.
Please do not attach a heated probe (PTU307) to metal structures to
avoid condensation problems caused by heat conduction along the metal.
NOTE
If the process temperature is much higher than that of the environment,
the whole probe and preferably plenty of cable must be inside the
process. This prevents measuring inaccuracy caused by heat conduction
along the cable.
When mounted on the side of a duct or channel, the probe must be
inserted from the side of the duct. If this is not possible and the probe
must be inserted from the top, the point of entry must be carefully
insulated.
For Vaisala probe installation kits and some installation examples see
Appendix A on page 195.
PTU303 for General Use
The PTU303 is a small size (d=12mm) probe for general use, up to
+80 °C (+176 °F). The probe is suitable for weather stations,
environmental compensations, laser interferometers and test benches. It
is suitable for ducts and channels with the installation kit available from
Vaisala.
See Appendix A on page 195 for the following probe installation kits for
PTU303 and installation examples.
- Duct mounting kit
- Cable gland
PTU307 for High Humidi t ies
The PTU307 is for environment where relative humidity is very high,
near saturation. The warmed probe prevents the saturation of the sensor.
An additional temperature probe is also available.
The PTU307 RH+T probe is suitable for temperatures up to +180°C
(+356°F). Note that the operational temperature limit for the PTU307
probe is higher than for the PTU300 transmitter itself. The upper
temperature limit for barometric pressure measurement is +60°C (140°F).
See Appendix A on page 195 for a presentation of the following probe
installation kits for PTU307 with installation examples:
- Duct mounting kit
- Cable gland
- Pressure tight Swagelok connector
- Vaisala's Meteorological Installation kit
The installation kits are available for both humidity and temperature
probe.
Temperature Probe (Optional)
An additional temperature probe is available to measure the ambient
temperature on the PTU307 (with probe warming) and PTU30T
transmitter models. The additional temperature probe allows you to
measure other humidity quantities apart from dewpoint and mixing ratio.
The temperature probe must be connected to the transmitter at the
factory. Do not cut and re-connect the cable yourself.
You must install the additional temperature probe in the same
measurement environment as the PTU307 probe. Make sure that heat
does not transfer from the warmed probe to the temperature probe.
1 = Connect AC (mains) voltage wires to these terminals
2 = Grounding terminal
3 = In case the module is not installed in the factory: Connect wires
from these terminals to the
terminals of
the motherboard.
4 = + 5 =
-
Optional Modules
Power Supply Module
The AC (mains) power connection may be connected to the power supply
module only by an authorized electrician. A readily accessible disconnect
device shall be incorporated in the fixed wiring.
G
supply module.
incompatible to be installed at the same time. If both modules are
installed, there is no room to close the transmitter cover completely.
Number refers to Figure 28 above:
1 = Galvanic isolation module
The following numbers refer to Figure 29 above:
1 = Flat cable pins
2 = Screw terminals for signal line
3 = DIP switches to select the output mode and range
Galvanic Isolation of the Power Supply
If galvanic isolation of the power supply line from the output signals is
needed, PTU300 can be ordered with optional galvanic isolation module.
This module prevents harmful grounding loops.
NOTE
alvanic isolation module is not needed when using the AC power
Only one of the switches 1
Only one of the switches 3
Installation and Wiring
1. Disconnect the power. In case the analog output module is installed
in the factory, continue with the step 4.
2. To attach the module, open the transmitter cover and fasten the
analog output module to the position for MODULE 1 with four
screws. Refer to Figure 2 on page 25.
3. Connect the flat cable between the analog output module and the
motherboard's connector for MODULE 1.
4. Take out the protective plug from the cable gland and thread the
wires.
5. Connect the wires to the screw terminals marked with Ch+ and
Ch- .
6. Select the current/voltage output by setting ON either of the
switches 1 or 2.
7. Select the range by setting ON one of the switches 3...7.
NOTE
... 2 can be ON at a time.
... 7 can be ON at a time.
0508-029
Figure 30 Third Analog Output Selection
8. Connect the power.
9. Select the quantity and scale the channel via the serial line or
display/keypad, see section Analog Output Quantities on page 128.
For testing the analog output, see section Analog Output Tests on
page 130. For fault indication setting, see section Analog Output
Fault Indication Setting on page 131.
PTU300 can be equipped with one configurable relay module. Each
module contains two configurable relays. See the contact ratings in
section Technical Specifications of Optional Modules on page 187.
Installation and Wiring
1. Disconnect the power and open the transmitter cover. In case the
relay-module is installed in the factory, continue with step 5.
2. To attach the module fasten the relay module to the bottom of the
housing with four screws. See the position in Figure 2 on page 25.
3. When the mains power is in use attach the grounding wire to the
grounding terminal.
4. Connect the flat cable between the relay module and the
MODULE 1 pins of the motherboard.
5. Take out the protective plug from the cable gland and thread the
relay wires.
NOTE
6. Connect the wires to the screw terminals: NO, C, and NC. Refer to
section Selecting the Activation State of the Relay.
7. Connect the power and close the cover.
Selecting the Activation State of the Relay
The middlemost C terminal and either one of the terminals NO/NC must
be connected. The polarity can be freely selected.
For instructions on how to operate the relay (for example, selecting
quantity for the relay output and setting the relay setpoints), see section
Operation of Relays on page 132.
The relay module may contain dangerous voltages even if the
transmitter power has been disconnected. Before opening the transmitter
you must switch off
the relay
Do not connect the mains power to relay unit without grounding the
transmitter.
The data lines on earlier versions of the RS
A and B instead of D1+ and D0
voltage compared to D0
When connecting the module, be prepared to swap the D1+ and D0
wires if you have a communication problem.
1
2
3
RS-422/485 Interf ace
The RS-422/485 interface enables communication between an RS-485
network and the PTU300 transmitter. The RS-485 interface is isolated
and offers a maximum communications rate of 115 200 bits/s. (For
maximum bus length of 1 km, use bit rate 19200 b/s or less.)
When selecting an RS-232 to RS-485 converter for the network, avoid
self-powered converters, as they do not necessarily support the needed
power consumption.
NOTE
-232 User Port on PTU300 main board cannot be used and connected
-485 module is connected. Service port operates normally.
If you use the internal termination of the transmitter at the end of the RS
485 bus (instead of using
will block the bus operation.
Installation and Wiring
1. Disconnect the power. In case the RS-422/485 module is installed
in the factory, continue with the item 4.
2. To attach the module, open the transmitter cover and fasten the
RS-422/485 module to the bottom of the housing with four screws.
3. Connect the flat cable between the RS-422/485 module and the
motherboard's pins MODULE1 (Communications).
4. Pull the network wirings through the cable gland.
5. Connect the twisted pair wires (1 or 2 pairs) to the screw terminals
as presented in Table 9 below:
Table 9 Connecting the Twisted Pair Wires to the Screw
Terminals
NOTE
To avoid problems in the RS-422/485 network, the Common
terminal should be wired to the corresponding terminals of the
other devices.
6. If you use RS-422 or RS-485 to connect just one PTU300 to a
master computer, enable the internal termination of PTU300 by
switching switches 1 and 2 ON. Make sure that the master's end of
the line is also terminated (by using master's internal termination or
with a separate terminator).
If you are connecting many transmitters to the same RS-485 bus,
make sure that switches 1 and 2 are OFF and terminate the bus with
separate terminators at both ends. This allows removing any
transmitter without blocking the bus operation.
-
separate terminators) removing that transmitter
7. Use the bus type (4-wire/2-wire) to select the selection switch 3.
In 4-wire mode RS-485 master sends data to the PTU300 through
terminals Rx D1+ and Rx D0- and receives data from PTU300
through terminals Tx D1+ and Tx D0-.
The LAN interface is designed to be used
environments (trusted corporate LAN or VPN
internet). Avoid connecting the transmitter directly to a public network
because the device can be attacked by a malicious user through the
network.
1 =
Flat cable connector
2 =
RJ45 connector with
indicator LEDs for
link and activity
LAN Interface
The optional LAN interface enables an Ethernet connection to the
transmitter. The user can establish a virtual terminal session using a
Telnet client program such as PuTTY or by using Modbus TCP protocol.
When the LAN interface is in use, serial communication using the RS232 User Port is disabled.
The LAN interface module must be installed at the factory (when
ordering the transmitter), or by a Vaisala Service Center. Once installed,
the module is automatically used by the transmitter. The physical
connection to the network is made to the RJ45 connector on the LAN
interface module, using a standard twisted pair Ethernet cable
(10/100Base-T). Transmitters with the optional LAN interface are
delivered pre-installed with a suitable cable and cable gland.
The LAN interface can use both static and dynamic network settings. If
the interface is configured to use dynamic settings, the network where the
LAN interface is connected must have a DHCP server that provides the
settings.
CAUTION
The network configuration can be done using the optional display and
keypad, or by using the service port. For instructions, see section LAN
Communication on page 84. The LAN interface also provides a web
configuration interf ace , whi ch you can acc ess b y entering the IP address
of the LAN interface in the address field of a web browser. See section
Web Configuration for LAN and WLAN on page 90.
Connector for antenna
cable (connected to
transmitter cover)
WLAN Interface
The optional WLAN interface enables a wireless Ethernet connection
(IEEE 802.11b) to the transmitter. The user can establish a virtual
terminal session using a Telnet client program such as PuTTY or by
using Modbus TCP protocol. When the WLAN interface is in use, serial
communication using the RS-232 User Port is disabled.
The interface supports Wired Equivalent Privacy (WEP) and Wi-Fi
Protected Access (WPA, WPA2). For WEP, 64 and 128 bit encryption is
supported, with open system or shared key authentication. WPA is used
in the Pre-Shared Key (PSK) mode, with either TKIP or CCMP (also
known as AES) encryption.
Similarly to the LAN interface, the WLAN interface can use both static
and dynamic network settings. If the interface is configured to use
dynamic settings, the network where the WLAN interface is connected
must have a DHCP server that provides the settings.
The WLAN interface also provides a web configuration interface, which
you can access by entering the IP address of the WLAN interface in the
address field of a web browser. See section Web Configuration for LAN
and WLAN on page 90.
The WLAN interface is designed to be used in trusted network
environments (trusted corporate LAN or VPN-based connection over
internet). Avoid connecting the transmitter directly to a public network
because the device can be attacked by a malicious user through the
network.
The default country setting for the WLAN interface is United States,
which limits the operation to WLAN channels 1 ... 11. To enable
additional channels (12 ... 14), you can change the country setting using
the web configuration interface.
The WLAN interface is not recommended for use with the
probe model
probe is designed to be used with the WLAN interface.
Observation Period
Period for Trend/Max/Min
Calculations (Resolution)
20 minutes
10 seconds
3 hours
90 seconds
1 day
12 minutes
10 days
2 hours
2 months
12 hours
1 year
3 days
4 years
12 days
NOTE
PTU301 fixed
(for wall-mounted applications). The PTU301 short cable
Attaching the WLAN Antenna
The WLAN interface module must be installed at the factory (when
ordering the transmitter), or by a Vaisala Service Center. Before taking
the transmitter into use, you must attach the antenna of the WLAN
interface into the RP-SMA connector on the transmitter cover. The
location of the antenna is shown in Figure 80 on page 192.
Data Logger Module
The optional data logger module extends the data storage for the
measurement data. When the data logger is present, this storage is
automatically used by the transmitter. The stored data can be browsed
using the optional display module, and accessed through the serial
connections. See sections Graphic History on page 70 and Data
Recording on page 123.
The data logger module contains non-volatile flash memory providing 4
years 5 months of storage for 4 parameters at a 10 second sampling
interval. When the memory is full, the data recording will not stop.
Instead, the oldest data is overwritten. For each parameter and
observation period, the module stores the minimum and maximum values
during the interval, as well a data trend value that is averaged from
samples taken during the interval (see Table 12 below).
Table 12 Observartion Periods and Resolution
The quantities that are logged are the same that have been selected for
measurement using the display/keypad or the serial line. When taking the
transmitter into use, verify that the desired quantities are selected. If you
change the quantities later, the transmitter will start logging the new
quantities, and stop logging the quantities that are no longer selected.
Changing the quantities does not delete any measurement data that is
already in memory. However, the data of any removed quantity cannot be
accessed unless the quantity is re-activated first.
The data logger module has a real time clock with a battery back-up. The
clock has been set to the Coordinated Universal Time (UTC) at the
factory, and its time cannot be set by the user. The data that is stored in
the logger's memory is timestamped using the logger's clock.
When date and time are set on the transmitter, they are stored to the
transmitter's memory as an offset from the time on the logger's clock.
When browsing the stored data, the time offset is applied to the
timestamps shown in the graphical history, and data outputted from the
serial port. The timestamps in the data logger's memory remain as they
were originally stored.
You can compensate for the clock drift (less than ±2 min/year) by setting
the time on the transmitter. This updates the time offset used on the
display and the serial port. You can set the time by using the
keypad/display or the serial commands.
0706-068
Figure 37 Data Logger Module
After a reset or a power up, it will usually take at least 10 seconds before
the data logger module is initialized. The instrument will not start up
before the data logger module is ready.
The indicator LED on the module will blink green during normal
operation. If the LED is lit in red color, there is a problem with the
module. The transmitter will also indicate the problem by activating the
"Internal error on add-on module" error. If the module is not operating
correctly, the transmitter must be sent to Vaisala for maintenance.
The data logger module must be installed at the factory (when ordering
the transmitter), or by a Vaisala Service Center. Once installed, the
module is automatically used by the transmitter. When the module
requires a new battery, the transmitter must be sent to Vaisala for
maintenance.
This chapter contains information that is needed to operate this product.
Getting Started
Within a few seconds after power-up the led on the cover of the
transmitter is lit continuously indicating normal operation. If the data
logger module is installed, the startup may take up to 18 seconds. When
using the optional display and turning the transmitter on the first time, the
language selection menu window opens. Select the language with the
▼▲ arrow buttons and press the SELECT button (the left-hand
button).
The pressure has an effect on humidity calculations and accuracy.
Therefore, accurate calculations can be achieved only when the ambient
pressure is taken into consideration. PTU300 uses measured pressure for
compensation by default.
See section Pressure Compensation Settings on page 109 for instructions
on how to set the pressure.
1 = INFO shortcut key; see section Information Display on
page 72.
2 = GRAPH shortcut key; see section Graphic History on page 70.
3 = Quantities selected for display
From any view, even in the absence of an
press on the right
display.
Display/Keypad (Optional)
The optional display and keypad combination enables shortcuts for
viewing current settings and status of the device, current measurement
values, and graph of the recent measurement history. Additionally the
device has user friendly, visible menu system for adjusting settings and
turning functions on or off.
Basic Display
Display shows you the measurement values of the selected quantities in
the selected units. You can select 1 ... 4 quantities for the basic display
(see section Changing Quantities and Units on page 101).
NOTE
0601-014
Figure 38 Basic Display
-hand function button takes you directly to the basic
1 = Tendency: Increasing/decreasing graph symbol with the code
number (for more information, see Figure 40 on page 68)
2 = P3h symbol
3 = Trend (middlemost numeric value)
Pressure 3h Trend and Tendency
Reading
Using Basic Display
0604-056
Figure 39 Basic Display with P
Trend and Tendency
3H
Pressure tendency graphics and codes
The characteristic symbols of pressure tendency during the three hours
preceding the time of observation are described in Figure 40 on page 68.
Source: The World Meteorological Organization (WMO) publication
Manual on Codes Vol. I.1, International Codes, Part A - Alphanumerical
Codes, 1995 Edition, WMO - No. 306. Section C, Code Table 0200: a.
tendency is chosen as a display quantity, the PTU300 logs
actual measured pressure, instead of trend or
Using Serial Line
Pressure 3h trend and tendency reading is also available through serial
line. Key in the lines below:
>form "P=" p "trend=" p3h "tend=" a3h #r#n
OK
>send
P= 1024.7trend= 0.8tend=1
The last line shows the values.
For more information on the Form command, see the section Changing
Quantities and Units, starting on page 101.
Missing Trend
The PTU300 series barometers output a code "*" when the pressure
tendency has not yet been calculated - that is, less than three hours have
elapsed since the power-up of the barometer. The absence of the pressure
trend is indicated in a similar manner, too.
Period for Trend/Max/Min
Calculations (Resolution)
20 minutes
10 seconds
3 hours
90 seconds
1 day
12 minutes
10 days
2 hours
2 months
12 hours
1 year
3 days
4 year*
12 days
Graphic History
The graphical display shows the data trend or min/max graph of the
selected quantities, one at a time. The graph is updated automatically
while measuring.
0706-052
Figure 41 Graphical Display
Trend graph: Shows you a curve of average values. Each value is a
calculated average over a period. See Table 13 below.
Max/min graph: Shows you the minimum and maximum values in a
form of curve. Each point shows the absolute minimum and maximum
observed during the time period represented by the data point. See Table
13 below.
Table 13 Periods for Trend and Max/Min Calculations
* Shows the maximum logging period of the data logger module (available when data
logger module is installed)
Use the following functions in the graphical display:
- Press the NEXT button to change between the trend graph and
max/min graph for the quantities selected for display.
- Press the EXIT button to return to the basic display.
- Press the ▼▲ arrow buttons to zoom in and out in the graph window.
- Press the ◄► arrow buttons move the cursor (vertical bar) along the
time axis. The cursor mode allows you to observe individual
measuring points. The numerical value at the cursor position is shown
at the left upper corner. The right upper corner shows the time from
the present to the chosen moment (without the logger module), or the
date and time at the cursor position (when the logger module is
installed).
Power failure (marked also with dashed
vertical line)
No data
Quantity has not been selected for the display
System error
General device or power supply problem
T meas. error
Temperature measurement/sensor failure
RH meas. error
Humidity measurement/sensor failure
P meas. error
Pressure measurement/sensor failure
Adj. mode active
Adjustment mode active (data recorded in the
adjustment mode is not displayed)
- If the optional data logger module is installed, you can scroll the
cursor off the screen to move to a new point on the time axis. The new
date will be displayed, and the cursor will be centered at the date
where the cursor scrolled off the screen.
0706-030
Figure 42 Graphical Display with Data Logger
If the data logger module is installed, you can move quickly on the time
axis: zoom out, move the cursor, and zoom back in.
The time that is shown below the graph is adjusted with the current time
offset of the transmitter. If you change the transmitter's date and time
setting, the displayed timestamps in the history graph change
accordingly. For an explanation of the effect of changing the date and
time manually, see section Data Logger Module on page 63.
Table 14 Graph Information Messages in Cursor Mode
A question mark after time indicates that at least one power failure
(dashed vertical line) has occurred after the chosen moment. In this case,
the exact time difference between the present and the cursor position is
not exactly known. If the data logger module is installed, the time is
known and the question marks are not present.
The information display contains current settings and status of the device.
You can open the display by pressing the left function button INFO in
the basic display. The following information will be shown:
- Current sensor operation (for example, chemical purge), if any, in
progress
- Present or past unacknowledged errors, if any
- Device identification; product name, version and serial number
- Adjustment information (the latest date, user made adjustments)
- Measuring settings
- Information on chemical purge settings (when applicable)
- Display Alarm settings
- Serial interface information
- Network settings and status of the LAN or WLAN interface
(when applicable)
- Analog output information
- Relay output information (when applicable)
0706-002
Figure 43 Device Information on Display
Proceed in the information views by pressing the MORE button
repeatedly until you get the desired information. You can also press the
left/right arrow buttons to browse through the information.
To exit the display, press the right function button OK.
5. Select the language with the ▼▲ buttons, and confirm the
selection by pressing the left-hand button.
6. Press the right-hand button to exit to the basic display.
Rounding Setting
Round off one decimal by using the Rounding function . The default
setting is rounding on. Rounding has no effect on quantities without
decimals.
1. Open the Main Menu by pressin g an y of the arrow buttons.
2. Select Display and press the right arrow button.
3. Select Rounding and press the ON/OFF key.
4. Press the EXIT key to return to the basic display.
Display Backlight Setting
As a default the display backlight is always on. In the automatic mode
the backlight stays on for 30 seconds from the last press of any key.
When pressing any key, the light turns on again.
1. Open the Main Menu by pressing any of the arrow buttons.
2. Select Display, press the right arrow button.
3. Select Backlight, press the CHANGE key.
4. Select On/Off/Automatic, press the SELECT key.
5. Press the EXIT key to return to the basic display.
Display Contrast Sett ing
Display contrast is automatically adjusted based on the ambient
temperature. However, depending on the installation location and
viewing direction, it may be necessary to fine-tune the contrast manually.
Using Display/Keypad
1. Open the Main Menu by pressing any of the arrow buttons.
2. Select Display, press the right arrow button.
3. Select Contrast, press the ADJUST key.
4. Adjust the contrast by pressing the left/right arrow buttons.
5. Press the OK key and then EXIT to return to the basic display
Contrast value for display. Recommended range -9 ... 9,
default 0. Negative values make screen lighter, positive
values make it darker.
Using Serial Line
Use the serial line command CON to view or set the display contrast.
CON [xxx]<cr>
Example (view current contrast setting):
>con
Contrast : 0
>
Example (set contrast to 5):
>con 5
Contrast : 5
>
Keypad Lock (Keyguard)
This function locks the keypad and prevents unintentional key presses.
1. Press and hold down the left function button for 4 seconds to lock
the keypad (at any display).
2. To unlock the keypad, press and hold down the same key for 4
seconds.
Menu PIN Lock
You can prevent unauthorized changes of the device settings by
activating the menu PIN lock. When this function is activated, the basic
display, graphical view, and device information displays are available,
but access to the menus is locked. The key symbolindicates the
activation of this feature.
1. Open the Main Menu by pressing any of the arrow buttons.
2. Select System, press the right arrow button.
3. Select Menu PIN, press the ON key.
4. Enter a PIN code by using the up/down arrow buttons. Move to the
next digit by pressing the left/right arrow buttons. Press the OK
key to confirm the setting. Now the PIN lock is on and a key
symbol is shown in a display.
5. Press the EXIT key to return to the basic display. Returning to the
menu is possible only by entering the correct PIN code.
You can also disable the keypad completely with serial command
LOC
118
When you want to turn off the PIN lock, open the menu by entering the
PIN code and select System, Menu PIN, press the OFF key.
In case you have forgotten the PIN code, open the transmitter cover and
press the ADJ button once. Wait for a few seconds and the Adjustmentmenu opens. Select Clear menu PIN, press the CLEAR key.
NOTE
K. See section Locking Menu/Keypad using Serial Line on page
.
Factory Settings
Use the display/keypad to restore the factory settings. This operation
does not affect the adjustments. Only settings available in the menus are
restored.
1. Press any of the arrow buttons to open the Main Menu.
2. Select System, press the right arrow button.
3. Select Factory settings and press the REVERT key to confirm
your selection. Press the YES key to reset all settings to the factory
defaults.
In case you change your mind and want to exit the menu without making
any changes, press the NO key.
See section General Settings on page 101 for a description of the other
menu options.
Configuring Display Ala rm s
The Display Alarm feature provides two independently configurable
alarms for transmitters with the display/keypad option. Each alarm tracks
a selected quantity, with freely configurable limit values. Each alarm also
has a configurable hysteresis value to prevent unnecessary triggering
when the measurement fluctuates around an alarm limit. The alarms can
be configured for any quantity supported by the transmitter. The
configuration of the Display Alarms can only be done using the
display/keypad option.
You can have the alarm activate inside or outside the specified range by
setting the limits correctly:
- To have the alarm activate when the measured quantity is outside the
specified range, set the Act. above limit higher than the Act. below
limit.
- To have the alarm activate when the measured quantity is inside the
range, set the Act. above limit lower than the Act. below limit.
The alarm limits are shown on the graph display as thicker dotted lines.
When a display alarm is activated, the automatic scaling of the graph
display always keeps the limits in view.
1102-012
Figure 45 Alarm Limits Shown on Graph Screen
When an alarm is activated, an alarm note is displayed on the display,
and the lights of the display will blink. If the data logger module is
installed, the alarm note includes the time and date of the alarm.
0802-041
Figure 46 Display Alarm Active
Multiple alarms can be active at the same time; the alarm that was
triggered first will be shown on the display. The next active alarm is
revealed when the currently shown alarm is acknowledged by pressing
the OK button.
Activated alarms are only shown on the screen; there are no alarm
messages output to the serial line. After an alarm has been
acknowledged, you must refer to the data graphs to see when the
measured quantities have exceeded the limits. You may use the optional
relay outputs as alarm outputs by configuring them separately; see
section Setting Relay Outputs on page 137.
Using Display/Keypad
1. Enter the Main Menu by pressing an arrow button on the keypad.
2. Use the arrow keys to select Display, followed by Alarms, to open
the Display Alarms menu. The Display Alarms menu shows the
currently enabled and disabled alarms.
Changes you do on the alarm editing page will take effect immediately,
and may cause an alarm to appear on the screen.
0802-069
Figure 47 Display Alarms
3. Use the arrow keys to select an alarm to configure. The alarm
editing page opens.
NOTE
4. To select a quantity for the alarm, press the CHANGE button and
select the quantity from the list.
5. To modify or remove the alarm limit values, move the selection
over the Act. above or Act. below field and press the SET button.
You will be prompted to Modify or Remove the value.
0802-070
Figure 48 Modifying an Alarm Limit
When modifying the value, use the arrow up and down buttons to
change the value under the cursor. Left and right arrow buttons
move the cursor. Select the OK button to accept the modified
value, or Cancel to undo the modification.
6. Set a suitable Hysteresis value to prevent the alarm from being
triggered unnecessarily by small measurement changes that pass
the alarm limit repeatedly.
7. Set or clear the Alarm enable checkbox to enable or disable the
alarm.
8. Press the EXIT button to leave the alarm configuration screen and
return to the basic view.
Quantity for display alarm 1. You can set any quantity
available on your transmitter; see
addition to quantities, you can als
FAULT
ONLINE
measurement output (for example, chemical purge).
quantity2
=
Quantity for display alarm 2. Options are the same as for
quantity1.
Name
Description
above
If specified, the alarm goes off when the quantity value rises above
Cannot be set for fault and online alarms.
below
If specified, the alarm goes off when the quantity value decreases
Cannot be set for fault and online alarms.
hyst
Distance from setpoint amount of extra retractment needed in order
Cannot be set for fault and online alarms.
enabl
Enable (ON) or disable (OFF) the display alarm.
Using Serial Line
Use the serial line command ALSEL to view or set the display alarms.
ALSEL [quantity1 quantity2]<cr>
tables on page 22. In
o select the following:
– Alarm for device errors
– Alarm for sensor operations that freeze the
Example (view current display alarm settings):
>alsel ?
Al1 P above: - ?
Al1 P below: 1013.00 hPa ?
Al1 P hyst : 10.00 hPa ?
Al1 P enabl: ON ?
Al2 T above: 40.00 'C ?
Al2 T below: - ?
Al2 T hyst : 1.00 'C ?
Al2 T enabl: OFF ?
Warning! One or more alarms disabled.
>
Table 15 ALSEL Parameters
the exact value of this setpoint. However, if above < below, the
alarm goes off at (above + hyst) and (below – hyst) instead.
below the exact value of this setpoint. However, if above < below,
the alarm goes off at (above + hyst) and (below – hyst) instead.
to regenerate alarm on the next crossing of the setpoint.
Example (set Tdf and T as display alarm quantities and configure the
alarms as prompted):
>alsel tdf t
Al1 Tdf above: 80.00 'C ? 10
Al1 Tdf below: - ?
Al1 Tdf hyst : 1.00 'C ?
Al1 Tdf enabl: ON ?
Al2 T above: 40.00 'C ? Al2 T below: - ? 0
Al2 T hyst : 1.00 'C ?
Al2 T enabl: OFF ? on
>
MI70 Link Program for Data Handling
MI70 Link is a support program that enables you to handle the
measurement data of your Vaisala transmitter using a Windows PC.
Using the MI70 Link you can, for example:
- Monitor the transmitter’s measurement readings directly using the
real-time window function.
- Transfer recorded data from the main transmitter memory in numeric
or graphical format, for further use in a spreadsheet program (such as
Microsoft Excel) or virtually any other application. Note that the
maximum amount of data points that can be downloaded in one
transfer is 65000.
To use the MI70 Link program, you must connect your Windows PC to
the service port of the transmitter with a serial or USB cable. The MI70
Link program, and the optional connection cables, are available from
Vaisala. See section Spare Parts and Accessories on page 189.
Follow the steps below to connect your transmitter to the MI70 Link
program using the serial interface:
1. Connect your PC to the service port of the transmitter. Refer to
section Service Port Connection on page 82.
2. Check that the PTU300 is powered.
3. Start the MI70 Link program.
4. Start using the program. There is usually no need to select a COM
port manually, the MI70 Link software can detect it automatically.
NOTE
For instructions on using the MI70 Link, see the program’s online help.
Connect the serial interface by using either the user port or the service
port.
For permanent interfacing to host system, use the user port. You can
change the serial settings and operate in RUN, STOP, SEND, POLL, and
MODBUS modes.
For temporary connections use the service port. Service port is always
available with fixed serial settings.
0605-039
Figure 49 Service Port Connector and User Port Terminal on
Motherboard
User Port Connection
Use suitable serial cable between the user port RxD, GND and TxD
screw terminals and the PC serial port, see Figure 50 on page 82.
Table 16 Default Serial Communication Settings for the User
To connect to the service port, you need a suitable cable with an RJ45
connector. Depending on the connections of your PC, you can either use
the Serial Connection Cable (optional accessory 19446ZZ) or the
USB-RJ45 Serial Connection Cable (optional accessory 219685). The
USB cable enables you to connect the transmitter to a PC via a standard
type A USB port. Note that the USB cable does not enable high speed
data transfer, since the bit rate is limited by the serial interface of the
service port.
Installing the Driver f or t he USB Cable
Before taking the USB cable into use, you must install the provided USB
driver on your PC. When installing the driver, you must acknowledge
any security prompts that may appear.
1. Check that the USB cable is not connected. Disconnect the cable if
you have already connected it.
2. Insert the media that came with the cable, or download the latest
driver from www.vaisala.com.
3. Execute the USB driver installation program (setup.exe), and
accept the installation defaults.
4. After the driver has been installed, connect the USB service cable
to a USB port on your PC. Windows will detect the new device,
and use the driver automatically.
5. The installation has reserved a COM port for the cable. Verify the
port number, and the status of the cable, using the Vaisala USB Instrument Finder program that has been installed in the
Windows Start menu.
Windows will recognize each individual cable as a different device, and
reserve a new COM port. Remember to use the correct port in the
settings of your terminal program.
Using the Service Port
1. Open the screws on the transmitter cover, and open the cover.
2. Connect the desired cable (serial interface cable or USB cable) to
your PC and the service port connector on the transmitter. For the
location of the service port, refer to Figure 49 on page 81.
3. Open a terminal program and set the communication settings as
follows:
Table 17 Communication Settings for the Service Port
For a detailed explanation of using a terminal program, see section
If enabled, the transmitter will retrieve its network
are used instead.
Web configuration
If enabled, the settings of the interface can be
address of the transmitter.
IP Address
The four part network ID of the transmitter. Must
Example value: 192.168.0.222
Netmask
Used together with the IP address to determine
A common netmask is 255.255.255.0
Gateway
IP address of the server that enables the
Example value: 192.168.0.1
MAC
The MAC address is the unique hardware address
changed.
LAN Communication
To enable LAN communication, a LAN or WLAN interface must be
physically connected to the network, and the networking settings must be
suitable for your network. For a description of interfaces, see sections
LAN Interface on page 61 and WLAN Interface on page 62.
The LAN and WLAN interfaces both operate by accessing the serial
interface (User Port) of the transmitter. All commands and protocols that
are available using the serial interface are available through the LAN and
WLAN interfaces; refer to section List of Serial Commands on page 94.
For instructions on how to connect using a terminal program, see section
Terminal Program Settings on page 91.
IP Configuration
The IP settings of the LAN and WLAN interfaces are described in Table
18. The current settings can be viewed on the serial line or using the
device information display; see section Information Display on page 72.
Table 18 IP Settings for the LAN and WLAN Interfaces
(DHCP)
settings (including the IP Address) from a server
in the network. If disabled, static network settings
changed using a web browser. The configuration
page can be accessed by browsing to the IP
be set manually if automatic configuration is not
used.
which network the transmitter is a part of. Must be
set manually if automatic configuration is not
used.
transmitter to access other networks. Must be set
manually if automatic configuration is not used.
You can configure the IP settings of the LAN and WLAN interfaces
using the display/keypad as follows:
1. Press any of the arrow buttons to open the Main Menu.
2. Press the ► arrow button to select Interfaces.
3. Press the ► arrow button to select Network settings. There will be
a delay as the transmitter refreshes the network information.
4. You are now in the Network Interface menu. Selecting the IP configuration option opens the IP configuration menu.
1102-015
Figure 51 Network Interface Menu
The Network Interface menu also allows you to configure the
Communication Protocol, enable or disable the Web
configuration option, or Disconnect all users that are currently
accessing the LAN or WLAN interface.
5. In the IP configuration menu, select Automatic configuration (DHCP), or enter the IP address, Netmask and Gateway
manually. If automatic configuration is enabled, manual
configuration cannot be done.
0709-004
Figure 52 IP Configuration Menu
To enter a value manually, use the ▲▼ arrow buttons to select the
parameter to change, and press CHANGE. A cursor will appear in
the first digit. Move the cursor using the ◄► arrow buttons, and
change the value under the cursor using the ▲▼ arrow buttons.
Confirm the selection by pressing OK.
6. After configuring the desired parameters, press EXIT to apply the
changes and return to the basic display.
ON or OFF. Enables or disables the automatic IP
configuration.
WEB
=
ON or OFF. Enables or disables the Web
Configuration page.
IP = The four part network ID of the transmitter. Must
be set manually if automatic configuration is not
used.
SUBNET
=
Used together with the IP address to determine
which network the transmit
set manually if automatic configuration is not used.
GATEWAY
=
IP address of the server that enables the transmitter
to access other networks. Must be set manually if
automatic configuration is not used.
Using Serial Line
Use the serial line command NET to view or set the network settings for
the LAN and WLAN interfaces. You can also refresh the network
information or disconnect all active connections.
NET [REFRESH][DISCONNECT][DHCP WEB] [DHCP IP SUBNET
GATEWAY WEB]<cr>
Examples:
>net refresh
OK
DHCP : OFF
IP address : 192.168.0.101
Subnet mask : 255.255.255.0
Default gateway: 192.168.0.1
Web config. : OFF
MAC address : 00:40:9d:2c:d2:05
Status : Not connected
>
>net on off
DHCP : ON
IP address : 192.168.0.104
Subnet mask : 255.255.255.0
Default gateway: 192.168.0.1
Web config. : OFF
MAC address : 00:40:9d:2c:d2:05
Status : Connected
OK
>
The service set identifier (that is, network name) of the
wireless network to connect to. 1 … 32 characters.
Security type
The security type of the wireless network. The options
see below.
Security key
The encryption key or passphrase that is used with an
encrypted network.
>net off 192.168.0.101 255.255.255.0 192.168.0.1 off
DHCP : OFF
IP address : 192.168.0.101
Subnet mask : 255.255.255.0
Default gateway: 192.168.0.1
Web config. : OFF
MAC address : 00:40:9d:2c:d2:05
Status : Not connected
OK
>
Wireless LAN Co nfiguration
The settings of the WLAN interface are described in Table 19. The
current settings can be viewed on the serial line or using the device
information display; see section Information Display on page 72.
- CCMP encryption is called AES in some networking products.
- The WLAN interface supports both WPA and WPA2 (also known as
802.11i) security when either WPA-PSK mode is selected.
- If your WLAN network is using some other security method than
listed below, you must set up a temporary WLAN network with any of
these settings and then use the web configuration to set WLAN
security settings for the actual network. See section Web
Configuration for LAN and WLAN on page 90.
Table 19 Wireless LAN Settings
are:
OPEN
OPEN/WEP
WPA-PSK/TKIP
WPA-PSK/CCMP
All other choices except OPEN require a security key;
You can configure the Wireless LAN settings using the display/keypad
as follows:
1. Press any of the arrow buttons to open the Main Menu.
2. Press the ► arrow button to select Interfaces.
3. Press the ► arrow button to select Network settings. There will be
a delay as the transmitter refreshes the network information.
4. Press the ► arrow button to select Wireless LAN settings.
0802-111
Figure 53 Wireless LAN Settings
5. The Name entry on the page shows the SSID of the currently
selected wireless network. To change the SSID, press the SET
button. Use the ▲▼ arrow buttons to change the character under
the cursor, and ◄► arrow buttons to more the cursor. Press the
OK button when done.
0802-110
Figure 54 Entering Network SSID
6. To change the currently selected Network type, select the Type
entry and press the CHANGE button. Select the new type from the
list and press the SELECT button.
Security type of the wireless network. The options
are:
OPEN
OPEN/WEP
WPA
WPA-PSK/CCMP
7. If you have selected an encrypted network type (WEP or WPA),
you must enter the security key to be used. Select the
Key/passphrase entry and press the SET button. Enter the key in
the same way as the SSID, and press the OK button. With the WEP
encryption you must enter the encryption key in hexadecimal
(10 hexdecimals for 64-bit encryption or 26 hexadecimals for
128-bit encryption). A WPA key must be 8 … 63 ASCII
characters.
8. After setting the wireless network parameters, press the EXIT
button in the Wireless LAN Settings menu. You will be asked to
confirm the new settings. Note that when new settings are saved,
all currently active WLAN connections are disconnected.
Using Serial Line
Use the serial line command WLAN to view or set the wireless network
settings. If you set an encrypted network type, you will be asked to enter
the security key. With the WEP encryption you must enter the encryption
key in hexadecimal (10 hexdecimals for 64-bit encryption or 26
hexadecimals for 128-bit encryption) or with plain ASCII characters (5
characters for 64-bit encryption or 13 characters for 128-bit encryption).
A WPA key must be 8 … 63 ASCII characters.
WLAN [SSID TYPE]<cr>
-PSK/TKIP
Examples:
>wlan ?
Network SSID : WLAN-AP
Type : OPEN
>
>wlan accesspoint wpa-psk/tkip
Network SSID : accesspoint
Type : WPA-PSK/TKIP
WPA-PSK phrase ? thequickbrownfox
Save changes (Y/N) ? y
OK
>
When a Telnet connection is established via the LAN or WLAN
interface, the session has the same communication mode, run interval,
poll address and echo settings as the serial port (user port) session would
have.
These settings can be changed using the display/keypad, using the serial
line (user port or service port), or on-the-fly during the Telnet session.
The display menu path to the Communication Protocol settings is:
Main Menu ► Interfaces ► Network Interface ► Communication
Protocol.
The commands for changing the settings are SMODE, INTV, ADDR,
and ECHO.
Web Configuration for LAN and WLAN
The LAN and WLAN interfaces both have a web configuration page that
is accessible using a browser. If you have not disabled the page from the
network settings, you can access it with a web browser at the IP address
of the interface. You can check the currently assigned IP address of the
interface from the device information screens (see Device Information on
page 115), or from the serial line by issuing the net ? command.
When accessing the web configuration page, you must first log in.
Username: user
Password: vaisala
The web configuration page provides similar network configuration
options as the serial line and the display/keypad. It also has additional
options for advanced users. For example, there are more options for
securing the wireless network.
If these additional options are used, they will appear as custom
configurations when viewed from the serial line or the display/keypad.
P
the transmitter is configured to use the
can always use PuTTY to access the transmitter through the Service
1102-017
Figure 56 Web Configuration Interface for WLAN
Terminal Program Settings
NOTE
The instructions below describe how to connect to the PTU300 using the
PuTTY terminal application for Windows. Perform the necessary cabling
and configuration of the transmitter before following the instructions.
PuTTY is available for download at www.vaisala.com.
uTTY cannot be used to access the transmitter through the User Port if
Modbus protocol. However, you
Port.
Opening a Serial/USB connection
1. Power up the PTU300 and start the PuTTY application.
2. Select the Serial & USB settings category, and check that the
correct COM port is selected in the Serial or USB line to connect to field. Change the port if necessary.
If you are using a Vaisala USB cable, you can check the port that it
uses by clicking the USB Finder... button. This opens the Vaisala USB Instrument Finder program that has been installed along with
the USB drivers.
3. Check that the other serial/USB line settings are correct for your
connection, and change if necessary. The default settings (shown in
Figure 57 on page 92) are used by the Service Port of the PTU300.
The instructions below assume that the LAN/WLAN interface on the
transmitter is correctly configured and the network connection is already
done.
NOTE
0810-070
Figure 57 Opening a Serial Connection
4. Click the Open button to open the connection window and start
using the serial line.
If PuTTY is unable to open the serial port you selected, it will show
you an error message instead. If this happens, restart PuTTY and
check the settings.
Opening a Telnet Session (LAN/WLAN)
1. Power up the instrument and start the PuTTY application. If your
transmitter acquires a network address using DHCP, wait for a
minute for this to complete, then verify the address from the
transmitter.
2. In the Session window, select the Telnet connection type.
3. Enter the IP address of your transmitter in the Host Name (or IP address) field. Use the default Telnet port 23.
4. Click the Open button to open the connection window and start
using the Telnet session.
If PuTTY is unable to connect the IP address you entered, it will
show you an error message instead. If this happens, check the
IP address and the connections, restart PuTTY, and try again.
Output the reading once from all connected
transmitters (in STOP or POLL mode)
SCOM
Define a user-specific SEND command for one
message output
SMODE [STOP/SEND/RUN/
POLL/MODBUS]
Set the serial interface mode
SDELAY
View or set user port (RS-232 or RS-485)
answer minimum delay
SERI [baud p d s]
User Port settings (Default: 4800 E 7 1) baud:
300 ... 115200
ADDR [0 ... 255]
Set the transmitter address (for POLL and
MODBUS modes)
NET
View or set networking parameters for LAN and
WLAN interfaces
WLAN
View or set wireless network parameters for
WLAN interface
OPEN [0 ... 255]
Open a temporary connection to a POLL mode
device
CLOSE
Close the temporary connection (back to POLL
mode)
Command
Description
FORM
Set the output format of SEND and R
commands
TIME
Set the time
DATE
Set the date
FTIME [ON/OFF]
Add time to SEND and R outputs
FDATE [ON/OFF]
Add date to SEND and R outputs
FST [ON/OFF]
Add the state of probe heating and chemical
commands
SCOM
Assign a new command name that works like
the SEND command
UNIT
Select the metric or non-metric output units
List of Serial Commands
All commands can be issued either in uppercase or lowercase. In the
command examples, the keyboard input by the user is in bold type.
The notation <cr> refers to pressing the carriage return (Enter) key on
your computer keyboard. Enter a <cr> to clear the command buffer
before starting to enter commands.
In the tables below, the bold text in the brackets is the default setting.
Enter the R command to start the continuous output of measurements.
The default output format depends on the quantities your transmitter
supports. You can change the output message format and interval with
the FORM and INTV commands. See sections FORM on page 101 and
INTV on page 122.
Example:
>r
P= 1021.6 hPa T= 23.2 'C RH= 5.8 %RH
>
If a value is too long to fit to the allocated space in the output, or if there
is an error in outputting the quantity, the value is displayed with stars '*'.
Example:
RH=***.* %RH T= 31.0 'C
Stopping Continuous Outputt ing
S
Use the S command to end the RUN mode. After this command all other
commands can be used. You can also press the Esc button or reset the
transmitter to stop the outputting.
See command SMODE to change the default (power-up) operation
mode.
Outputting Reading Once
SEND
Use the SEND command to output the reading once in STOP mode.
Example:
You can remove the SCOM definition by pressing Esc when entering the
command name:
>scom
Send command : p ? <esc>
>
NOTE
Outputting Reading Once Fr om All Trans m it t e rs
Use the DSEND command to output the reading once from all connected
transmitters that are in STOP or POLL modes. In order to avoid data
collisions, the transmitters will send their response after a suitable delay.
The response includes the transmitters’s address, followed by their
measurement message.
The total execution time of the DSEND command depends on the bit rate
of the serial port connection. At slow speeds, it will take a considerable
amount of time for devices with large address number to respond. The
delay will be the same even if there are no other devices at all.
DSEND<cr>
Example (transmitters with addresses 3 and 25 responding):