Vaisala PTB330 Technical data

Vaisala Instruments Catalog 2009
Ref. B210768EN rev. A
P.O. Box 26, FI-00421 Helsinki, FINLAND
Email: industrialsales@vaisala.com
Fax: +358 9 8949 2485
PTB330 Digital Barometer for Professional Meteorology, Aviation, and Industrial Users
Features/Benefits
Vaisala BAROCAP® sensor
Accurate measurement
Excellent long-term stability
Added reliability through redundancy
Graphical trend display with 1-year history data
Height and altitude corrected pressure (QFE, QNH)
For professional meteorology and aviation, laboratories, demanding industrial applications
Tel: +358 9 894 91
Vaisala BAROCAP® Digital Barometer PTB330 with a new trend display.
Vaisala BAROCAP® Digital Barometer PTB330 is a new generation barometer, designed for a wide range of high-end atmospheric pressure measurement.
e pressure measurement of the PTB330 is based on the Vaisala in-house, silicon capacitive, absolute pressure sensor - the Vaisala BAROCAP® Sensor. It provides high measurement accuracy and excellent long-term stability.
High accuracy
e PTB330 series features extremely high accuracy. Class A barometers for the most demanding applications are ne-adjusted and calibrated against a High Precision Pressure Calibrator. Class B barometers are adjusted and calibrated using electronic working standard. All the PTB330 barometers come with a NIST traceable, factory calibration certicate.
Reliability through redundancy
According to customers’ choice, the PTB330 can incorporate one, two or
three BAROCAP® sensors. When two or three sensors are used, the barometer continuously compares the readings of the pressure sensors against one another and provides information on whether these are within the set internal dierence criteria. is unique feature provides redundancy in pressure measurement.
us, users also get a stable and reliable pressure reading at all times as well as a pre-indication of when to service or recalibrate the barometer.
In addition to instant pressure, the PTB330 also provides the WMO pressure trend and tendency code.
QNH and QFE
e PTB330 can be set to compensate for QNH and QFE pressure used especially in aviation. e QNH represents the pressure reduced to sea level, based on the altitude and temperature of the observation site. e QFE represents the height corrected pressure of small dierences in altitude, for example, the air pressure
at the aireld elevation.
Data transfer
e data can be transferred to a PC by using MI70 Link Interface Software. You can examine the recorded data easily in a Microsoft® Windows environment and transfer it further to a spreadsheet program, such as Microsoft® Excel, in numeric or graphical format.
Graphical display
e PTB330 features a multi-lingual, graphical display allowing users to monitor measurement trends. e graph is updated automatically while measuring and it provides a one-year measurement history.
Applications
e PTB330 can be used successfully for aviation, professional meteorology, and for demanding industrial pressure measurement applications such as accurate laser interferometric measurement and exhaust gas analysis in engine test benches.
114
Vaisala Instruments Catalog 2009
Ref. B210768EN rev. A
PTB330
Technical Data
BAROMETRIC PRESSURE
Performance
Barometric pressure range 500 ... 1100 hPa
Class A Class B Linearity* ±0.05 hPa ±0.10 hPa Hysteresis* ±0.03 hPa ±0.03 hPa Repeatability* ±0.03 hPa ±0.03 hPa Calibration uncertainty** ±0.07 hPa ±0.15 hPa Accuracy at +20 °C (+68 °F) *** ±0.10 hPa ±0.20 hPa
Barometric pressure range 50 ... 1100 hPa
Class B Linearity* ±0.20 hPa Hysteresis* ±0.08 hPa Repeatability* ±0.08 hPa Calibration uncertainty** ±0.15 hPa Accuracy at +20 °C *** ±0.20 hPa
Temperature dependence****
500 ... 1100 hPa ±0.1 hPa 50 ... 1100 hPa ±0.3 hPa
Total accuracy -40 ... +60 °C (-40 ... +140 °F)
Class A Class B 500 ... 1100 hPa ±0.15 hPa ±0.25 hPa 50 ... 1100 hPa ±0.45 hPa
Long-term stability
500 ... 1100 hPa ±0.1 hPa/year 50 ... 1100 hPa ±0.2 hPa/year
* Dened as ±2 standard deviation limits of endpoint non­linearity, hysteresis or repeatability error. ** Dened as ±2 standard deviation limits of inaccuracy of the working standard including traceability to NIST. *** Dened as the root sum of the squares (RSS) of endpoint non-linearity, hysteresis error, repeatability error and calibration uncertainty at room temperature. **** Dened as ±2 standard deviation limits of temperature dependence over the operating temperature range.
Operating environment
Pressure range 500 ... 1100 hPa, 50 ... 1100 hPa Temperature range
operating -40 ... +60 °C (-40 ... +140 °F) with local display 0 ... +60 °C (+32 ... +140 °F)
Pressure tting barbed tting for 1/8” I.D. tubing
Maximum pressure limit 5000 hPa abs. Compliance EMC standard EN61326-1:1997+
or quick connector with shuto
valve for 1/8” hose
Am1:1998 + Am2:2001: Industrial
Environment
Mechanics
Housing material G AlSi10 Mg (DIN 1725) Housing classication IP65 Weight 1 - 1.5 kg
Analog output (optional)
Current output 0 ... 20 mA, 4 ... 20 mA Voltage output 0 ... 1 V, 0 ... 5 V, 0 ... 10 V
Accuracy at pressure range
at +20 °C ±0.30 hPa ±0.40 hPa at -40 ... +60 °C ±0.60 hPa ±0.75 hPa
500 ... 1100 hPa 50 ... 1100 hPa
Accessories
Serial interface cable 19446ZZ USB-RJ45 serial connection cable 219685 Software interface kit 215005 Wall mounting kit 214829 Outdoor installation kit (weather shield) Installation kit for pole or pipeline 215108 Power supply module POWER-1 Temperature compensated analog output module Isolated RS-485 module RS485-1
215109
AOUT-1T
Dimensions
in mm (inches)
Inputs and outputs
Supply voltage 10 ... 35 VDC Supply voltage sensitivity negligible Typical power consumption at +20 °C
24 VDC, one pressure sensor)
(U
in
RS-232 25 mA RS-485 40 mA U
out
I
out
display and backlight +20 mA
Serial I/O RS232C, RS485/422 Pressure units hPa, mbar, kPa, Pa inHg,
Resolution 0.01 hPa 0.1 hPa Settling time at power-up (one sensor) 4 s 3 s Response time (one sensor) 2 s 1 s Acceleration sensitivity negligible Pressure connector M5 (10-32) internal thread
0, mmHg, torr, psia
mmH
2
Class A Class B
25 mA 40 mA
BAROCAP® is a registered trademark of Vaisala. Specications are subject to change without prior notice. ©Vaisala Oyj
115
Loading...