Vaisala HUMICAP HMP5, HUMICAP HMP4, HUMICAP HMP8, HUMICAP TMP1 Quick Manual

Page 1
M211982EN-C
EN
Quick Guide
Vaisala HUMICAPâ Humidity and
Temperature Probes
JA
Page 2
PUBLISHED BY
Vaisala Oyj Street address: Vanha Nurmijärventie 21, FI-01670 Vantaa, Finland Mailing address: P.O. Box 26, FI-00421 Helsinki, Finland Phone: +358 9 8949 1
Visit our Internet pages at www.vaisala.com.
No part of this manual may be reproduced, published or publicly displayed in any form or by any means, electronic or mechanical (including photocopying), nor may its contents be modified, translated, adapted, sold or disclosed to a third party without prior written permission of the copyright holder. Translated manuals and translated portions of multilingual documents are based on the original English versions. In ambiguous cases, the English versions are applicable, not the translations.
The contents of this manual are subject to change without prior notice.
Local rules and regulations may vary and they shall take precedence over the information contained in this manual. Vaisala makes no representations on this manual’s compliance with the
local rules and regulations applicable at any given time, and hereby disclaims any and all responsibilities related thereto.
This manual does not create any legally binding obligations for Vaisala towards customers or end users. All legally binding obligations and agreements are included exclusively in the applicable supply contract or the General Conditions of Sale and General Conditions of Service of Vaisala.
This product contains software developed by Vaisala or third parties. Use of the software is governed by license terms and conditions included in the applicable supply contract or, in the absence of separate license terms and conditions, by the General License Conditions of Vaisala Group.
Page 3

Table of Contents

English............................................................................................................................ 5
日本語........................................................................................................................... 17
3
Page 4
4 M211982EN-C
Page 5

Product Overview

2
1
3
4
5
6
7
8
HMP4, HMP5, HMP7, and HMP8 are humidity and temperature measurement probes with a digital (Modbus) output. TMP1 is a temperature-only version. The probes are designed for demanding humidity and temperature measurement applications. Each probe model has a dierent probe head, but the same probe body and a 2 m (6.5 ft) cable between the probe body and the probe head.
The probes are compatible with Vaisala Indigo™ transmitters. They can also be connected to Vaisala Insight software for
configuration, diagnostics, and temporary online monitoring.

Probe Structure

Figure 1 Probe Parts
1 Protection cap (remove before use) 2 5-pin M12 connector 3 Probe body with type label 4 Status indicator:
Green Power on and probe online,
Red Error
O Power o or indicator
5 Probe cable. Do not cut. 6 Probe head (HMP7 model shown) 7 Sensor(s) under protective filter 8 Protection cap (remove before use)
flashes when communicating
disabled

ENGLISH

More Information

For detailed instructions for installing, configuring, and maintaining the probes, see HMP4, 5, 7, 8, and TMP1 User Guide in English M212022EN available at www.vaisala.com .
5
Page 6

HMP4 Probe

183 [7.20]
120 [4.72]
Fitting body M22x1.5 or NPT1/2”
41 [1.61]
Ø 12
[0.47]
32 mm 27 mm
mm [in]
41 [1.61]
Ø 12
[0.47]
Ø 13.5
[0.53]
253 [9.96]
243 [9.57]
mm [in]
Figure 2 HMP4 Probe Head Dimensions
Vaisala HUMICAPâ Humidity and Temperature Probe HMP4 is designed for high-pressure applications such as compressed air systems in maritime, breathing air, and industrial applications, where measurement performance and chemical tolerance are essential.
• Operating pressure 0 ... 10 MPa (0 ... 100 bar)
• M22x1.5 and NPT1/2” probe fittings included
• Sintered AISI316L filter standard in delivery

HMP5 Probe

Figure 3 HMP5 Probe Head Dimensions
Vaisala HUMICAPâ Humidity and Temperature Probe HMP5 is designed for high-temperature applications such as baking ovens, pasta dryers and industrial drying kilns where measurement performance and chemical tolerance are essential.
• Temperature measurement range -70 ... +180 °C (-94 ... +356 °F)
• 250 mm (9.84 in) probe allows easy process installation through insulation
• Sintered AISI316L
6 M211982EN-C
filter standard in delivery
Page 7

HMP7 Probe

99.5 [3.92]
79.5 [3.13]
37.5 [1.48]
Ø 12
[0.47]
Ø 12
[0.47]
Groove for lock ring
mm
[in]
Ø 12
[0.47]
232 [9.13]
194 [7.64]
41 ... 149 [1.16 ... 5.87]
41 [1.61]
Ø 13.5
[0.53]
Fitting body ISO1/2” or NPT1/2”
mm
[in]
Figure 4 HMP7 Probe Head Dimensions
Vaisala HUMICAPâ Humidity and Temperature Probe HMP7 is designed for applications which involve constant high humidity or rapid changes in humidity such as drying and test chambers, combustion air and other humidifiers and meteorological measurements where measurement performance and chemical tolerance are essential.
• Probe and sensor warming functions minimize condensation on probe
• Vapor and pressure proof construction
• Stainless steel mesh
Probe Heating
HMP7 supports probe heating. Probe heating heats up not only the sensor, but the entire probe head. When probe temperature is heated above dew point temperature, condensation on the probe can be avoided while measuring the dew point temperature of the process.
Probe heating is disabled by default. If you enable probe heating, output parameters that are dependent on temperature measurement (such as relative humidity) are disabled unless the true temperature of the measured environment is updated to the temperature compensation register of the probe from another measurement instrument (for example, the TMP1 probe). Output parameters such as dew point temperature that can be measured or calculated without this external temperature information are available even without the temperature input.
filter standard in delivery
ENGLISH

HMP8 Probe

Figure 5 HMP8 Probe Head Dimensions
7
Page 8
Vaisala HUMICAPâ Humidity and Temperature Probe HMP8 is designed for pressurized
130 [5.12]
Ø 3.2 [0.13]
Ø 6 [0.24]
mm [in]
applications in compressed air systems, refrigerant dryers, environmental chambers and other pressurized industrial applications where easy insertion and removal of the probe and adjustable installation depth into the pipeline are needed.
• Probe installation depth can be freely adjusted and probe can be hot-swapped from pressurized pipelines with an installation valve
• Operating pressure 0 ... 4 MPa (0 ... 40 bar)
• ISO1/2" and NPT1/2" probe
• Sintered AISI316L filter standard in delivery
fittings and press handle included

TMP1 Probe

Figure 6 TMP1 Probe Head Dimensions
Vaisala Temperature Probe TMP1 is designed for demanding temperature measurements in industrial applications such as pharmaceutical industry and calibration laboratories, where accuracy and robustness are essential.
• Temperature measurement range -70 ... +180 °C (-94 ... +356 °F)
• Temperature accuracy up to 0.1 °C (0.18 °F)
8 M211982EN-C
Page 9

Probe Body Dimensions and Pinout

136 [5.35]
Probe cable 2000 [78.74]
Ø 25
[0.98]
M12/5
1
5
34
2
M12/5
Male Connector
mm
[in]
Figure 7 Probe Body
ENGLISH
Pin # Function Notes Wire Colors in Vaisala
1 Power supply Supply voltage:
• HMP7: 18 ... 30 VDC
• Other models: 15 ... 30 VDC
Typical current 10 mA, max. 500 mA.
2 RS-485- White
3 Power GND Blue
4 RS-485+ Black
5 Not connected Gray
Cables
Brown
Recommended maximum length of the RS-485 line is 30 m (98 ft).
9
Page 10

Attaching Probes to Indigo Transmitters

Figure 8 Attaching the Probe to Indigo 200 Series
1. Insert the probe or the connection cable into the transmitter's connector. Use of connection cable is recommended for strain relief.
2. Turn the locking wheel of the transmitter to lock the probe or cable in place. Do not turn the probe or the cable itself, as that will damage the connectors.
3. If you are using a connection cable, connect the probe to the cable.
4. When the transmitter recognizes the connected probe, it shows a the display.
10 M211982EN-C
notification message on
Page 11

Installation

-40 ... +80 °C (-40 ... +176 °F)
-70 ... +180 °C (-94 ... +356 °F)
1
2
43
Figure 9 Example Installation of HMP8 Model
1 Mount the probe head horizontally to prevent any water condensing on the probe head
from running to the sensors.
2 Let the cable hang loosely to prevent condensed water from running along the cable to
the probe head.
3 Attach the probe body to a wall or other surface using the two mounting clips (Vaisala
item 243257SP) supplied with the probe. Each clip attaches to the installation surface with one screw (screw hole Ø 4.2 mm).
4
Cable to Modbus master or Indigo™ transmitter.
ENGLISH
If the temperature of the measured environment diers greatly from ambient temperature, the whole probe and preferably plenty of cable must be inside the process. This prevents measurement inaccuracy caused by heat conduction along the cable.
11
Page 12
Configuration

Vaisala Insight Software

Vaisala Insight software is a configuration software for Indigo-compatible devices. The supported operating systems are Windows 7 (64-bit), Windows 8.1 (64-bit), and Windows 10 (64-bit).
With the Insight software, you can:
• See probe information and status
• See real-time measurement
• Calibrate and adjust the probe Configure probe features such as measurement filtering, chemical purge, heating, and
• serial communication
Download Vaisala Insight software at www.vaisala.com/insight.
The probe can be connected to Vaisala Insight software using a Vaisala USB cable (no.
242659).

Connecting to Insight Software

• Computer with Vaisala Insight software installed
• USB connection cable (no. 242659)
Figure 10 Connecting Probe to Insight
1. Open the Insight software.
2. Connect the USB cable to a free USB port on the PC.
3. Connect the probe to the USB cable.
4. Wait for Insight software to detect the probe.
12 M211982EN-C
Page 13

Modbus

Serial Interface Settings

Table 1 Default Serial Interface Settings
Property Description/Value
Bit rate 19200
Parity None
Data bits 8
Stop bit 2
Flow control None
ENGLISH
You can use up to ten probes on the same RS-485 line. You must configure each probe on the line to have a
dierent Modbus address.

Measurement Data Registers

Table 2 Modbus Measurement Data Registers (Read-Only)
Register Number (Decimal)
Floating Point Values
1 0000
3 0002
7 0006
9 0008
11 000A
13 000C
14 000E
16 0010
19 0012
21 0014
23 0016
Address (Hexadecimal)
hex
hex
hex
hex
hex
hex
hex
hex
hex
hex
hex
Register Description Data Format Metric Unit
Relative humidity 32-bit float %RH
Temperature 32-bit float °C
Dew point temperature 32-bit float °C
Dew/frost point
32-bit float °C
temperature
Dew/frost point
32-bit float °C
temperature at 1 atm
Dew point temperature
32-bit float °C
at 1 atm
Absolute humidity 32-bit float
g/m
3
Mixing ratio 32-bit float g/kg
Wet-bulb temperature 32-bit float °C
Water vapor concentration
32-bit float ppm
v
Water vapor pressure 32-bit float hPa
13
Page 14
Register Number (Decimal)
Floating Point Values
Address (He
25 0018
27 001A
31 001E
33 0020
65 0040
xadecimal)
he
x
hex
he
x
hex
he
x
Register Description Data Format Metric Unit
Water vapor saturation
32-bit float hPa
pressure
Enthalpy 32-bit float kJ/kg
Dew point temperature
32-bit float °C
dierence
Absolute humidity at
32-bit float
g/m
3
NTP
Water vapor mass fraction
32-bit float ppm
w
Configuration
Table 3 Modbus
Register Number (Decimal)
General
1289 0508
Compensation Setpoints
769 0300
Configuration Data Registers (Writable)
Register Address (Hexadecim al)
hex
he
x
Registers
Register Description Data
Probe heating Available on HMP7 model only.
When on, relative humidity measurement is disabled unless temperature is written to register 0334 from an external source.
Pressure compensation setpoint
on/o.
hex
Format
16-bit boolean
Unit / Valid Range
0 = o (default)
1 = on
32-bit float Unit: hPA
Default: 1013.25 hPa
14 M211982EN-C
Page 15
Compensation Setpoints
821 0334
Communication
1537 0600
hex
hex
Temperature
32-bit float Unit: °C compensation setpoint. If a value is written to this register, probe uses it when calculating relative humidity instead of its own measurement.
When probe heating is in use, temperature must be written to this register to enable relative humidity measurement.
Modbus address 16-bit
integer
ENGLISH
1 ... 247
Default: 240
15
Page 16
16 M211982EN-C
Page 17

製品概要

2
1
3
4
5
6
7
8
HMP4、HMP5、HMP7、HMP8 は、デジタル(Modbus)出力の湿度・温度計測プローブで す。TMP1 は、温度のみを計測できるバージョンです。これらのプローブは厳しい環境下に おける湿度と温度の計測用に設計されています。プローブモデルごとにプローブヘッドは 異なりますが、プローブ本体や、プローブ本体とプローブヘッド間の 2 m(6.5 ft)ケーブ ルは同じです。
これらのプローブは、ヴァイサラ Indigo™ 変換器と互換性があります。また、Vaisala Insight ソフトウェアに接続して、設定、診断、および一時オンラインモニタリングを行え ます。

プローブ構造

図 11 プローブの部品
1 保護キャップ(使用前に取り外す) 2 5 ピン M12 コネクター 3 プローブ本体(タイプラベル付き) 4 状態インジケーター:
電源がオンでプローブがオン
ライン、通信時は点滅
エラー オフ 電源がオフ、またはインジ
ケーターが無効
5 プローブケーブル。切断しないでく
ださい。
6 プローブヘッド(図は HMP7 モデル) 7 保護フィルター下のセンサ 8 保護キャップ(使用前に取り外す)
日本語
詳細
プローブの取り付け、設定、およびメンテナンスの詳細については、www.vaisala.com 入手可能な『HMP4、5、7、8、および TMP1 取扱説明書(英語版 M212022EN)』を参照し てください。
17
Page 18

HMP4 プローブ

183 [7.20]
120 [4.72]
ࣇ࢕ࢵࢸ࢕ࣥࢢᮏయ
M22x1.
5 ࡲࡓࡣ NPT1/2”
41 [1.61]
Ø 12
[0.47]
32 mm 27 mm
mm [in]
41 [1.61]
Ø 12
[0.47]
Ø 13.5
253 [9.96]
243 [9.57]
mm
[in]
図 12 HMP4 プローブヘッドの寸法
ヴァイサラ HUMICAPâ 湿度 要となる、海洋、呼吸用空気、工業利用における圧縮空気システムなどの高圧用途向けに 設計されています。
• 動作圧力:0 ~ 10 MPa(0 ~ 100 bar)
• M22x1.5 および NPT1/2 プローブフィッティング付属
焼結 AISI316L フィルタは標準付属品
・温度プローブ HMP4 は、計測性能と耐化学薬品性が非常に重

HMP5 プローブ

図 13 HMP5 プローブヘッド寸法
ヴァイサラ HUMICAPâ 湿度 要となる、ベーキングオーブン、パスタ乾燥機、工業用乾燥炉などの高温用途向けに設計 されています。
• 温度計測範囲:-70 ~ +180 °C(-94 ~ +356 °F)
• 250 mm(9.84 in)プローブにより、絶縁体を通したプロセス環境への取り付けが容易
• 焼結 AISI316L フィルタは標準付属品
・温度プローブ HMP5 は、計測性能と耐化学薬品性が非常に重
18 M211982EN-C
Page 19

HMP7 プローブ

99.5 [3.92]
79.5 [3.13]
37.5 [1.48]
Ø 12
[0.47]
Ø 12
[0.47]
ࣟࢵࢡࣜࣥࢢ
ࡢ⁁
mm
[in]
Ø 12
[0.47]
232 [9.13]
194 [7.64]
41 149 [1.16 5.87]
41 [1.61]
Ø 13.5
[0.53]
ࣇ࢕ࢵࢸ࢕ࣥࢢᮏయ
ISO1/2” ࡲࡓࡣ NPT1/2”
mm
[in]
図 14 HMP7 プローブヘッド寸法
ヴァイサラ HUMICAPâ 湿度・温度プローブ HMP7 は、計測性能と耐化学薬品性が非常に重 要となる、乾燥チャンバーやテストチャンバー、燃焼用空気、その他の加湿器や気象計測 器などの常時高湿度または急速な湿度変化が伴う用途向けに設計されています。
• プローブでの結露を最小限に抑えるプローブおよびセンサ加温機能
• 耐蒸気および耐圧力構造
• ステンレス製メッシュフィルタは標準付属品
プローブ加温
HMP7 はプローブ加温をサポートしています。プローブ加温では、センサだけでなく、プ ローブヘッド全体が加温されます。加温されたプローブの温度が露点温度を上回ると、プ ロセスの露点温度の計測時にプローブでの結露を回避できます。
初期設定では、プローブ加温はオフになっています。プローブ加温がオンであれば、計測 環境の実際の温度が別の計測機器(TMP1 プローブなど)からプローブの温度補正レジス ターに更新される場合を除き、温度計測に依存する出力パラメーター(相対湿度など)は 無効になります。この外部温度情報なしで計測または計算できる、露点温度などの出力パ ラメーターは、温度入力がなくても利用できます。
日本語

HMP8 プローブ

図 15 HMP8 プローブヘッドの寸法
19
Page 20
ヴァイサラ HUMICAPâ 湿度
130 [5.12]
Ø 3.2 [0.13]
Ø 6 [0.24]
mm [in]
あり、パイプラインへの取り付けの深さを調整できる必要がある、圧縮空気システム、冷 却式ドライヤ、環境チャンバー、その他の加圧工業利用における加圧用途向けに設計され ています。
• プローブは、取り付けの深さを自由に調整でき、取り付けバルブを使用して圧力下パ イプラインからホットスワップ可能
• 動作圧力:0 ~ 4 MPa(0 ~40 bar)
ISO1/2"
• 焼結 AISI316L フィルタは標準付属品
および NPT1/2" プローブフィッティングとプレスハンドル付属
・温度プローブ HMP8 は、プローブの挿入と取り出しが容易で

TMP1 プローブ

図 16 TMP1 プローブヘッドの寸法
ヴァイサラ温度プロー 校正試験所などの工業用途向けに設計されています。
• 温度計測範囲:-70 ~ +180 °C(-94 ~ +356 °F)
• 温度精度:最大 0.1 °C(0.18 °F)
20 M211982EN-C
ブ TMP1 は、幅広い範囲での精度が非常に重要となる、製薬工業や
Page 21

プローブ本体の寸法およびピン配列

136 [5.35]
ࣉ࣮ࣟࣈࢣ࣮ࣈࣝ
2000 [78.74]
Ø 25
[0.98]
M12/5
1
5
34
2
M12/5
࢜ࢫࢥࢿࢡࢱ
mm
[in]
図 17 プローブ本体
日本語
ピン番号機能 備考 ヴァイサラケーブルの線の
1 電源 電源電圧:
• HMP7:18 ~ 30 VDC
• その他のモデル:15 ~ 30 VDC
電流は標準 10 mA、最大 500 mA。
2 RS-485-
3 電源 GND
4 RS-485+
5 接続なし グレー
RS-485 ラインの推奨最大長は 30 m(98 ft)です。
21
Page 22

プローブの Indigo 変換器への取り付け

図 18 プローブの Indigo 200 シリーズへの取り付け
1. プローブまたは接続ケーブルを変換器のコネクタに挿入します。負荷を緩和するため に、接続ケーブルの使用をお勧めします。
2. 変換器の固定用ホイールを回して、プローブまたはケーブルを所定の位置に固定しま す。コネクタが損傷するため、プローブまたはケーブル自体をつかんで回さないでく ださい
3. 接続ケーブルを使用する場合、プローブをケーブルに接続します。
4. 変換器で接続したプローブが認識されると、ディスプレイに通知メッセージが表示さ れます。
22 M211982EN-C
Page 23

取り付け

-40 +80 °C
(-40 +176 °F)
-70 +180 °C
(-94 +356 °F)
1
2
43
図 19 HMP8 モデルの取り付け例
1 プローブヘッドに結露した水がセンサに流れるのを防止するため、プローブヘッドを水
平に取り付けます。
2 結露した水がケーブルを伝わってプローブヘッドに流れるのを防止するため、ケーブル
をゆるく吊します。
3 プローブに付属の 2 つの取り付けクリップ(ヴァイサラ部品:243257SP)を使用して、
プローブ本体を壁や他の表面に取り付けます。クリップはそれぞれ、ネジ 1 本(ネジ穴 Ø 4.2 mm)を使用して取り付け面に取り付けます。
4
Modbus マスターまたは Indigo™ 変換器にケーブルを接続します。
日本語
もし計測環境の温度が周囲温度と大きく異なる場合は、プローブ全体とケーブ ルをなるべく長く、プロセス内に入れてください。これにより、ケーブルの熱 伝導による計測誤差を防ぐことができます。
23
Page 24
設定

Vaisala Insight ソフトウェア

Vaisala Insight ソフトウェアは、Indigo 互換デバイスの設定ソフトウェアです。サポート されているオペレーティングシステムは、Windows 7(64 ビット)、Windows 8.1(64 ビッ ト)、Windows 10(64 ビット)です。
Insight ソフトウェアを使用すると、以下を行うことができます。
• プローブの情報と状態の表示
• リアルタイム計測の表示
• プローブの校正と調整
• 計測フィルタリング、ケミカルパージ、加温、シリアル通信などのプローブ機能の設 定
Vaisala Insight ソフトウェアは、www.vaisala.com/insight からダウンロードしてくださ い。
プローブは、ヴァイサラ USB ケーブル(注文コード 242659)を使用して、Vaisala Insight ソフトウェアに接続できます。

Insight ソフトウェアへの接続

• Vaisala Insight ソフトウェアをインストールしたコンピューター
• USB 接続ケーブル(注文コード 242659)
図 20 プローブの Insight への接続
1. Insight ソフトウェアを開きます。
2. USB ケーブルを PC の空いている USB ポートに接続します。
3. プローブを USB ケーブルに接続します。
4. Insight ソフトウェアでプローブが検出されるのを待ちます。
24 M211982EN-C
Page 25

Modbus

シリアルインターフェースの設定

表 4 シリアルインターフェースの初期設定
特性 説明/値
ビットレート 19200
パリティ なし
データビット 8
ストップビット 2
フロー制御 なし
同じ RS-485 ラインで最大 10 台のプローブを使用できます。ライン上のプローブごとに異 なる Modbus アドレスを設定する必要があります。

計測データレジスター

表 5 Modbus 計測データレジスター(読み取り専用)
登録番号 (10 進数)アドレス(16 進数)レジスターの説明データ形式 メートル単位
浮動小数点値
1 0000
hex
相対湿度 32 ビット浮動小数点%RH
日本語
3 0002
7 0006
9 0008
11 000A
13 000C
14 000E
16 0010
19 0012
hex
hex
hex
hex
hex
hex
hex
hex
温度 32 ビット浮動小数点°C
露点温度 32 ビット浮動小数点°C
露点/霜点温度 32 ビット浮動小数点°C
1 気圧における露
32 ビット浮動小数点°C
点/霜点温度
1 気圧における露
32 ビット浮動小数点°C
点温度
絶対湿度 32 ビット浮動小
g/m
数点
混合比 32 ビット浮動小数点g/kg
湿球温度 32 ビット浮動小数点°C
3
25
Page 26
登録番号 (10 進数)アドレス(16 進数)レジスターの説明データ形式 メートル単位
浮動小数点値
21 0014
hex
水蒸気濃度 32 ビット浮動小数点ppm
v
23 0016
25 0018
27 001A
31 001E
33 0020
65 0040
hex
hex
hex
hex
hex
hex
水蒸気圧 32 ビット浮動小数点hPa
飽和水蒸気圧 32 ビット浮動小数点hPa
エンタルピー 32 ビット浮動小数点kJ/kg
露点温度差 32 ビット浮動小数点°C
NTP における絶 対湿度
32 ビット浮動小 数点
g/m
水蒸気質量分率 32 ビット浮動小数点ppm
3
w

設定レジスター

表 6 Modbus 設定データレジスター(書き込み可能)
登録番号(10 進数)アドレス(16 進数)レジスターの説明データ形式 単位/有効範囲
設定パラメーター
769 0300
hex
圧力補正設定値 32 ビット浮動小数点単位 hPa。
初期設定
1013.25 hPa。
26 M211982EN-C
Page 27
登録番号(10 進数)アドレス(16 進数)レジスターの説明データ形式 単位/有効範囲
設定パラメーター
821 0334
hex
温度補正設定値。 値がこのレジス
32 ビット浮動小数点単位 °C。
ターに書き込ま れている
場合、プ ローブは相対湿 度を計算する際 に自身の計測値 の代わりにレジ スターの値を使 用します。
プローブ加温を 使用している場 合、温度をこのレ ジスターに書き 込んで、相対湿度 計測を有効にす る必要がありま す。
機能制御
1289 0508
hex
プローブ加温の オン/オフ。 HMP7 モデルの みで使用できま
16 ビットブール値0 = オフ(初期設
定) 1 = オン
す。 オンの場合、温度
が外部ソースか らレジスター
0334
に書き込
hex
まれていなけれ ば、相対湿度計 測が無効になり ます。
通信設定
1537 0600
hex
Modbus アドレス16 ビット整数
範囲 1 ~ 247. 初期設定 240。
日本語
27
Page 28
28 M211982EN-C
Page 29
Page 30
www.vaisala.com
Loading...