Vaisala HUMICAP, HMP155 User Manual

HMP155 with an additional temperature probe and optional Stevenson screen installation kit.

Features

Vaisala HUMICAPâ180R sensor: superior long-term stability

Optional warmed humidity probe and chemical purge

Plug-and-play

USB connection for service use

Use with DTR13 and DTR503 radiation shields and a Stevenson screen

Weather-proof housing IP66

Optional, fast temperature probe

Different output possibilities: voltage, RS-485, resistive Pt100

Applications: meteorology, aviation and road weather, instrumentation

HUMICAPâ Humidity and

Temperature Probe HMP155

Vaisala HUMICAPâ Humidity and Temperature Probe HMP155 provides reliable humidity and temperature measurement. It is designed especially for demanding outdoor applications.

Long-term stability

HMP155 uses the proven Vaisala HUMICAPâ180R sensor that has excellent stability and withstands well harsh environments. The probe structure is solid and the sensor is protected by default with a sintered teflon filter, which gives maximum protection against water, dust, and dirt.

Warmed probe and highhumidity environment

Measuring humidity reliably is challenging in environments where humidity is near saturation. Measurements may be corrupted by fog, mist, rain, and heavy dew. A wet probe may not give an accurate measurement in the ambient air.

This is an environment to which Vaisala has designed this patented, warmed probe for reliable measurements. As the sensor head is warmed continuously, the humidity level inside it stays below the ambient level. Thus, it also reduces the risk of condensation forming on

the probe.

Fast measurements

With its fast response time,

the additional temperature probe for HMP155 is ideal for measuring in environments with changing temperatures. The membrane filter speeds up the relative humidity measurement.

Long lifetime

Protecting the sensor from precipitation, and scattered and direct solar radiation increases its lifetime. Thus, Vaisala recommends installing HMP155 in one of the following radiation shields: DTR503, DTR13, or Stevenson screen. For

the additional temperature probe,

an installation kit is available for Vaisala DTR502 Radiation Shield.

Calibration

The probe can be calibrated using a computer with a USB cable, with the push buttons, or with the MI70 indicator.

Vaisala HUMICAP, HMP155 User Manual

Technical Data

HMP155 humidity measurement performance

Sensor

HUMICAPâ180R and 180R2 for typical

 

applications

 

HUMICAPâ180C and 180R2C for

 

applications with chemical purge

 

and/or warmed probe

 

 

Observation range

0 … 100 %RH

Response time at +20 °C (+68 °F) in

63 %: 20 s

still air with sintered Teflon filter

90 %: 60 s

 

 

Factory calibration uncertainty at

±0.6 %RH (0 … 40 %RH)

+20 °C (+68 °F)1)

±1.0 %RH (40 … 95 %RH)

Accuracy (including non-linearity, hysteresis, and repeatability)

At +15 … +25 °C (+59 … +77 °F)

±1 %RH (0 … 90 %RH)

 

±1.7 %RH (90 … 100 %RH)

At −20 … +40 °C (−4 … +104 °F)

±(1.0 + 0.008 × reading) %RH

 

 

At −40 … −20 °C (−40 … −4 °F)

±(1.2 + 0.012 × reading) %RH

At +40 … +60 °C (+104 … +140 °F)

±(1.2 + 0.012 × reading) %RH

 

 

At −60 … −40 °C (−76 … −40 °F)

±(1.4 + 0.032 × reading) %RH

 

 

1)Defined as ±2 standard deviation limits. Small variations possible (see also the calibration certificate).

HMP155 temperature measurement performance

Sensor

Pt100 RTD element, Class F 0.1

 

IEC 60751

 

 

Observation range

−80 … +60 °C (−112 … +140 °F)

Response time for additional

63 %: < 20 s

temperature probe in 3 m/s (7 mph) air

90 %: < 35 s

flow

 

 

 

Other measured variables

Dew point / frost point temperature,

 

wet bulb temperature, mixing ratio

Accuracy with voltage output

 

 

 

At −80 … +20 °C (−112 … +68 °F)

±(0.226 - 0.0028 × temperature) °C

At +20 … +60 °C (+68 … +140 °F)

±(0.055 + 0.0057 × temperature) °C

 

 

Accuracy with passive (resistive) output

 

According to Tolerance Class AA

±(0.1 + 0.0017 × |temperature|) °C

IEC 607511)

 

Accuracy with RS 485 output

 

At −80 … +20 °C (−112 … +68 °F)

±(0.176 - 0.0028 × temperature) °C

 

 

At +20 … +60 °C (+68 … +140 °F)

±(0.07 + 0.0025 × temperature) °C

 

 

1)Tolerance Class AA IEC 60751 corresponds to IEC 751 1/3 Class B

HMP155 accuracy over temperature range: voltage and RS 485

HMP155 operating environment

Operating temperature for humidity

−80 … +60 °C (−112 … +140 °F)

measurement

 

 

 

Storage temperature

−80 … +60 °C (−112 … +140 °F)

Operating humidity

0 … 100 %RH

 

 

EMC (industrial environment)

EN/IEC 61326-1, EN 55022

IP rating

IP66

HMP155 inputs and outputs

Operating voltage

7 … 28 VDC

 

 

 

Minimum operating voltage

0

… 1 V output or RS 485: 7 V

 

0

… 5 V output, or warmed probe:

 

12 V

 

0

… 10 V output, chemical purge, or

 

XHEAT: 16 V

Outputs

Voltage output: 0 … 1 V, 0 … 5 V,

 

0

… 10 V

 

Resistive Pt100 4 wire connection

 

RS 485

 

 

 

Average power consumption (+15 VDC,

0

… 1 V output: < 3 mA

load 100 kΩ)

0

… 10 V output: +0.5 mA

 

RS-485: < 4 mA

 

During chemical purge: Maximum

 

110 mA

 

With warmed probe: Maximum

 

150 mA

Settling time at startup

Voltage output: 2 s

 

RS-485: 3 s

HMP155 mechanical specifications

Dimensions (H × W)

 

279 × 40 mm (10.9 × 1.6 in)

Weight

 

86 g (3.0 oz)

Length of additional T probe cable

2 m (6 ft 7 in)

Connection

 

8-pin male M12 connector

Connection cables

 

3.5 m (11 ft 6 in), 10 m (32 ft 10 in),

 

 

30 m (98 ft 5 in)

Maximum wire size

 

0.129 mm2 (26 AWG)

Service cables

 

USB connection cable

 

 

MI70 connection cable

Materials

 

 

Filter

 

Sintered Teflon or membrane

Housing

 

Polycarbonate (PC)

Additional temperature probe

Stainless steel AISI 316L

Cable

 

PUR

 

279 [10.98]

 

 

266 [10.47]

<![if ! IE]>

<![endif]>[1.57]

 

 

<![if ! IE]>

<![endif]>40

<![if ! IE]>

<![endif]>24 [0.94]

86 [3.39]

 

Dimensions in mm (inches)

Published by Vaisala | B210752EN-H © Vaisala Oyj 2020