Vaisala HUMICAP, HMP155 User Manual

HUMICAPâ Humidity and Temperature Probe HMP155
â
Vaisala HUMICAP HMP155 provides reliable humidity and temperature measurement. It is designed especially for demanding outdoor applications.
Humidity and Temperature Probe
HMP155 with an additional temperature probe and optional Stevenson screen installation kit.
Features
• Vaisala HUMICAPâ180R sensor: superior long-term stability
• Optional warmed humidity probe and chemical purge
• Plug-and-play
• USB connection for service use
• Use with DTR13 and DTR503 radiation shields and a Stevenson screen
• Weather-proof housing IP66
• Optional, fast temperature probe
Dierent output possibilities: voltage, RS-485, resistive Pt100
• Applications: meteorology, aviation and road weather, instrumentation
Long-term stability
HMP155 uses the proven Vaisala HUMICAPâ180R sensor that has excellent stability and withstands well harsh environments. The probe structure is solid and the sensor is protected by default with a sintered teflon filter, which gives maximum protection against water, dust, and dirt.
Warmed probe and high­humidity environment
Measuring humidity reliably is challenging in environments where humidity is near saturation. Measurements may be corrupted by fog, mist, rain, and heavy dew. A wet probe may not give an accurate measurement in the ambient air.
This is an environment to which Vaisala has designed this patented, warmed probe for reliable measurements. As the sensor head is warmed continuously, the humidity level inside it stays below the ambient level. Thus, it also reduces the risk of condensation forming on the probe.
Fast measurements
With its fast response time, the additional temperature probe for HMP155 is ideal for measuring in environments with changing temperatures. The membrane filter speeds up the relative humidity measurement.
Long lifetime
Protecting the sensor from precipitation, and scattered and direct solar radiation increases its lifetime. Thus, Vaisala recommends installing HMP155 in one of the following radiation shields: DTR503, DTR13, or Stevenson screen. For the additional temperature probe, an installation kit is available for Vaisala DTR502 Radiation Shield.
Calibration
The probe can be calibrated using a computer with a USB cable, with the push buttons, or with the MI70 indicator.
Technical Data
24
40 [1.57]
86 [3.39]
279 [10.98]
266 [10.47]
[0.94]
HMP155 humidity measurement performance
Sensor HUMICAPâ180R and 180R2 f
Observation range 0 … 100 %RH
Response time at +20 °C (+68 °F) in still air with sintered Teflon filter
Factory calibration uncertainty at +20 °C (+68 °F)
Accuracy (including non-linearity, hysteresis, and repeatability)
At +15 … +25 °C (+59 … +77 °F) ±1 %RH (0 … 90 %RH)
At −20 … +40 °C (−4 … +104 °F) ±(1.0 + 0.008 × reading) %RH
At −40 … −20 °C (−40 … −4 °F) ±(1.2 + 0.012 × reading) %RH
At +40 … +60 °C (+104 … +140 °F) ±(1.2 + 0.012 × reading) %RH
At −60 … −40 °C (−76 … −40 °F) ±(1.4 + 0.032 × reading) %RH
1) Defined as ±2 s
1)
tandard deviation limits. Small variations possible (see also the calibration certificate).
applications HUMICAPâ180C and 180R2C for applications with chemical purge and/or warmed probe
63 %: 20 s 90 %: 60 s
±0.6 %RH (0 … 40 %RH) ±1.0 %RH (40 … 95 %RH)
±1.
7 %RH (90 … 100 %RH)
or typical
HMP155 temperature measurement performance
Sensor Pt100 RTD element, Class F 0.1
Observation range −80 … +60 °C (−112 … +140 °F)
Response time for additional temperature probe in 3 m/s (7 mph) air
flow
Other measured variables Dew point / frost point temperature,
Accuracy with voltage output
At −80 … +20 °C (−112 … +68 °F) ±(0.226 - 0.0028 × temperature) °C
At +20 … +60 °C (+68 … +140 °F) ±(0.055 + 0.0057 × temperature) °C
Accuracy with passive (resistive) output
According to Tolerance Class AA
1)
IEC 60751
Accuracy with RS485 output
At −80 … +20 °C (−112 … +68 °F) ±(0.176 - 0.0028 × temperature) °C
At +20 … +60 °C (+68 … +140 °F) ±(0.07 + 0.0025 × temperature) °C
1) Tolerance Class AA IEC 60751 corresponds to IEC 751 1/3 Class B
IE
C 60751
63 %: < 20 s 90 %: < 35 s
wet bulb temperature, mixing ratio
±(0.1 + 0.0017 × |temperature|) °C
HMP155 operating environment
Operating temperature for humidity measur
ement
Storage temperature −80 … +60 °C (−112 … +140 °F)
Operating humidity 0 … 100 %RH
EMC (industrial environment) EN/IEC 61326-1, EN 55022
IP rating IP66
−80 … +60 °C (−112 … +140 °F)
HMP155 inputs and outputs
Operating voltage 7 … 28 VDC
Minimum operating voltage 0 … 1 V output or RS485: 7 V
Outputs Voltage output: 0 … 1 V, 0 … 5 V,
Average power consumption (+15 VDC, load 100 kΩ)
Settling time at startup Voltage output: 2 s
HMP155 mechanical specifica
Dimensions (H × W) 279 × 40 mm (10.9 × 1.6 in)
Weight 86 g (3.0 oz)
Length of additional Tpr
Connection 8-pin male M12 connector
Connection cables 3.5 m (11 ft 6 in), 10 m (32 ft 10 in),
Maximum wire size
Service cables USB connection cable
Materials
Filter Sintered Teflon or membrane
Housing Polycarbonate (PC)
Additional temperature probe Stainless steel AISI 316L
Cable PUR
obe cable 2 m (6 ft 7 in)
0 … 5 V output, or w 12 V 0 … 10 V output, chemical purge, or XHEAT: 16 V
0 … 10 V Resistive Pt100 4wire connection RS485
0 … 1 V output: < 3 mA 0 … 10 V output: +0.5 mA RS-485: < 4 mA During chemical purge: Maximum 110 mA With warmed probe: Maximum 150 mA
RS-485: 3 s
armed probe:
tions
30 m (98 ft 5 in)
0.129 mm2 (2
MI70 connection cable
6 AWG)
HMP155 accuracy over temperature range: voltage and RS485
Published by Vaisala | B210752EN-H © Vaisala Oyj 2020
www.vaisala.com
All rights reserved. Any logos and/or product names are trademarks of Vaisala or its individual partners. Any reproduction, transfer, distribution or storage of information contained in this document is strictly prohibited. All
specifications — technical included — are subject to change without notice.
Dimensions in mm (inches)