Vaisala HMW82 User manual

INSTALLATION

76.9
123.7
9 5
5.5
24
81
127.3
77.9
60.3
59
Ø4.4 Ø4.4
32.4
20
1. Find a suitable mounting location for the transmitter. The conditions at the location
should represent well the area of interest.
NOTE
2. Use the mounting holes to attach the mounting base securely. Use at least two
screws (not included). The arrow on the mounting base must point straight up after installation. Proper orientation is important, so that air will flow through the vents on the bottom and top.
3. Connect the wiring to the screw terminals on the mounting base. The supply voltage
and terminal assignments are model-specific; see Wiring on page 2.
4. After completing the wiring, connect the cover over the mounting base.
Vaisala INTERCAP® Humidity and
Temperature Transmitter Series HMW80
and direct sunlight.

Dimensions (in mm)

Find Quality Products Online at: sales@GlobalTestSupply.com
M211328EN-C _______________________________________________________________________ 1
www.GlobalTestSupply.com
Quick Guide ____________________________________________________________________ HMW80
-T
4...20 mA
(-5 ... +55 ºC)
-RH
4...20 mA
(0 ... 100% RH)
+T +RH
Power supply 10 ... 28 VDC R
L
= 0 ... 500 Ω
Power supply 10 ... 28 VDC R
L
= 0 ... 500 Ω
mA
mA
-T
4...20 mA
(-5 ... +55ºC)
-RH
4...20 mA
(0 ... 100% RH)
+T +RH
Power supply 10 ... 28 VDC R
L
= 0 ... 500 Ω
Power supply 10 ... 28 VDC RL = 0 ... 500 Ω
mA
mA
2 11
Pt100
t+
-T
4...20 mA
(-5 ... +55 ºC)
+T
Power supply 10 ... 28 VDC R
L
= 0 ... 500 Ω
mA

WIRING

HMW82

You must connect the RH channel of the HMW82, even if you only
want to measure temperature.
Connecting the T channel is optional.

HMW82P100

TMW82

HMW82P100
Includes a passive Pt100 temperature sensor (Class F 0.1 IEC 60751) with a 3-wire connection.
Leave the two terminals on the left unconnected.
Find Quality Products Online at: sales@GlobalTestSupply.com
© Vaisala 2016. All rights reserved. _______________________________________________________ 2
www.GlobalTestSupply.com
Quick Guide ____________________________________________________________________ HMW80
-T
0...10 V
(0...100 %RH)
-VS+T +VS
Power supply 15 ... 35 VDC or 24 VAC ±20% R
L
= 10 kΩ min.
-RH +RH
0...10 V (-5...+55 ºC)
V
V
-T -VS+T +VS
Power supply 15 ... 35 VDC or 24 VAC ±20% R
L
= 10 kΩ min.
-RH +RH
V
V
0...10 V
0...10 V
Recommended wiring for long cables.
3-wire connection with -VS as common
more than 2.5 Ω.
-T
0...10 V
(-5 ... +55 ºC)
-VS+T +VS
V
Power supply 15 ... 35 VDC or 24 VAC ±20% R
L
= 10 kΩ min.
-T -VS+T +VS
V
Power supply 15 ... 35 VDC or 24 VAC ±20% R
L
= 10 kΩ min.
0...10 V
Recommended wiring for long cables.
3-wire connection with -VS as common
more than 2.5 Ω.

HMW83

TMW83

ground. Not recommended if cable resistance is
ground. Not recommended if cable resistance is
Find Quality Products Online at: sales@GlobalTestSupply.com
© Vaisala 2016. All rights reserved. _______________________________________________________ 3
www.GlobalTestSupply.com