Vaisala HMT360Series User Manual

Page 1
HMT360 Series Transmitters
for Hazardous Areas
USER'S GUIDE
M010056EN-D
May 2004
Page 2
PUBLISHED BY
Vaisala Oyj Phone (int.): (+358 9) 894 91 P.O. Box 26 Fax: (+358 9) 8949 2227 FIN-00421 Helsinki Finland
Visit our Internet pages at http://www.vaisala.com/
© Vaisala 2004
No part of this manual may be reproduced in any form or by any means, electronic or mechanical (including photocopying), nor may its contents be communicated to a third party without prior written permission of the copyright holder.
The contents are subject to change without prior notice.
Page 3
Table of Contents
CHAPTER 1
GENERAL INFORMATION ............................................................................ 6
Safety.........................................................................................6
ESD Protection...................................................................... 6
Version Information ................................................................. 7
Warranty .................................................................................... 7
CHAPTER 2
GENERAL ...................................................................................................... 8
Output quantities...................................................................... 8
Probe options ...........................................................................8
Maintenance.............................................................................. 9
Compliance with the requirements ...................................... 10
European requirements.......................................................10
US requirements (FM).........................................................10
Canadian requirements (CSA) ............................................ 10
Australian requirements (TestSafe) ....................................10
CHAPTER 3
INSTALLATION............................................................................................ 11
General instructions .............................................................. 11
Mounting the probe cable in gas group IIC spaces............. 11
Selecting a place for the transmitter ................................... 11
Temperature differences ..................................................... 12
Checking the temperature reading...................................... 12
Mounting the transmitter ....................................................... 13
Mounting the probes.............................................................. 14
The HMP362 - small pressure-tight probe .......................... 14
The HMP363 probe for confined spaces ............................ 15
The HMP365 probe for high temperatures ......................... 16
HMP364 for pressurized spaces ......................................... 16
The HMP368 probe for measuring humidity in pressurized
pipelines or moisture in liquids ............................................ 19
Installing the probe through the ball valve assembly ..........21
CHAPTER 4
ELECTRICAL CONNECTIONS ................................................................... 23
Installation in hazardous locations ......................................24
US and Canadian requirements.......................................... 24
European requirements....................................................... 24
Maximum load calculation for the barrier............................ 24
HMT360 connected to a galvanic isolator............................ 25
HMT360 connected to a zener barrier .................................. 26
Examples of connections ...................................................... 27
Grounding ............................................................................... 28
Page 4
_________________________________________________________________________________
CHAPTER 5
OPERATION ................................................................................................. 29
HMT360 with display ..............................................................29
HMT360 without display......................................................... 29
Dip switch functions ..............................................................30
CHAPTER 6
OPERATION ERRORS ................................................................................32
Cleaning the sensor from glycol in a natural gas
application...............................................................................32
CHAPTER 7
DISPLAY/KEYPAD COMMANDS................................................................34
Setting the pressure for calculations ...................................34
Selecting the output quantities ............................................. 35
Upper half of the display.................................................35
Lower half of the display.................................................35
Selecting the analog outputs ................................................35
Scaling the analog outputs.................................................... 37
CHAPTER 8
SERIAL INTERFACE ...................................................................................38
Serial communication settings .............................................38
Selecting the analog outputs ................................................39
Scaling the analog outputs.................................................... 40
Adjustment commands..........................................................40
Relative humidity adjustment ..............................................40
Temperature adjustment .....................................................41
Analog output calibration .....................................................41
Output commands ..................................................................42
Testing the analog outputs ..................................................42
Outputting the measurement values ...................................42
Activating a continuous output ............................................42
Stopping the continuous output...........................................43
Setting the output interval.................................................... 43
Setting the ambient pressure for calculations ....................43
CHAPTER 9
MEASURING AT OVERPRESSURE ...........................................................44
CHAPTER 10
CALIBRATION AND ADJUSTMENT...........................................................45
Calibration interval .................................................................45
Factory calibration and adjustment......................................45
Calibration and adjustment by the user ............................... 45
Calculating the correspondence of current values and
output quantities.....................................................................47
Relative humidity calibration................................................. 48
Automatic two-point calibration (only with HMT360 with
display) ................................................................................48
Page 5
_________________________________________________________________________________
Vaisala __________________________________________________________________________ 5
Manual calibration ...............................................................50
Low end adjustment ....................................................... 50
High end adjustment (two-point calibration) .................. 50
Temperature calibration in one point................................... 51
CHAPTER 11
SPECIFICATIONS........................................................................................ 52
Performance ........................................................................... 52
Calculated variables available (typical ranges).............. 53
Outputs.................................................................................... 54
Classification with current outputs ...................................... 54
General .................................................................................... 55
Probes ..................................................................................... 56
Options and accessories....................................................... 56
ACCURACIES OF THE CALCULATED VARIABLES (NOT VALID WITH
HMT360 FOR NATURAL GAS) ................................................................... 57
APPENDIX 1: DIMENSIONS ....................................................................... 59
HMT361............................................................................... 59
HMP362 .............................................................................. 59
HMP363 .............................................................................. 60
HMP364 .............................................................................. 60
HMP365 .............................................................................. 60
HMP368 .............................................................................. 60
APPENDIX 2: WIRING FOR INTRINSICALLY SAFE OPERATION, FM ...63
APPENDIX 3: WIRING FOR INTRINSICALLY SAFE OPERATION, CSA. 64
CERTIFICATES............................................................................................ 65
Page 6
General Information _________________________________________________________________
CHAPTER 1
GENERAL INFORMATION
Safety
Throughout the manual, important safety considerations are highlighted as follows:
WARNING
Warning alerts you to a serious hazard. If you do not read and follow instructions very carefully at this point, there is a risk of injury or even death.
CAUTION
Caution warns you of a potential hazard. If you do not read and follow instructions carefully at this point, the product could be damaged or important data could be lost.
NOTE
Note highlights important information on using the product.
ESD Protection
Electrostatic Discharge (ESD) can cause immediate or latent damage to electronic circuits. Vaisala products are adequately protected against ESD for their intended use. However, it is possible to damage the product by delivering electrostatic discharges when touching, removing, or inserting any objects inside the equipment housing.
To make sure you are not delivering high static voltages yourself:
- Handle ESD sensitive components on a properly grounded and protected ESD workbench. When this is not possible, ground yourself to the equipment chassis before touching the boards. Ground yourself with a wrist strap and a resistive connection cord. When neither of the above is possible, touch a conductive part of
Page 7
General Information _________________________________________________________________
Vaisala __________________________________________________________________________ 7
the equipment chassis with your other hand before touching the
boards.
- Always hold the boards by the edges and avoid touching the component contacts.
Version Information
Table 1 Manual Revisions
Manual Code Description
U336en-1.1 1st version, March 1999 M010056en-A 2nd version, January 2001 M010056en-B 3rd version, March 2002 M010056en-C 4th version, November 2002 M010056EN-D 5th version, May 2004
Warranty
For certain products Vaisala normally gives a limited one-year warranty. Please observe that any such warranty may not be valid in case of damage due to normal wear and tear, exceptional operating conditions, negligent handling or installation, or unauthorized modifications. Please see the applicable supply contract or Conditions of Sale for details of the warranty for each product.
Page 8
General ___________________________________________________________________________
CHAPTER 2
GENERAL
The HMT360 series transmitters are microprocessor based two-wire transmitters for measuring relative humidity and temperature in hazardous areas requiring EX classification. Connect the transmitter always via galvanic isolators or zener barriers in hazardous environments.
Output quantities
The transmitter has a local display and two output current channels. The available quantities are described in Table 2 on page 35.
Probe options
The HMT360 transmitter series has various options for transmitter units, sensor heads and cable lengths (2.5 and 10 m). The available probe types are the following:
HMP361 HMP362 HMP363 HMP364 HMP365 HMP368
Page 9
General___________________________________________________________________________
Vaisala __________________________________________________________________________ 9
HMP361: probe for wall mounting HMP362: probe for pressurized spaces up to 167 bars HMP363: probe head ( 13.5 mm) for tight spaces. HMP364: probe for pressurized spaces up to 100 bars. HMP365 probe for high temperatures up to 180°C HMP368 probe for installations in pressurized pipelines
up to 40 bars; pressure tight sliding clasp nut
Maintenance
The transmitter's electronics unit can be removed and replaced in the field, as well as the probe; see page 44 for detailed information.
All other maintenance is performed at the factory. If a transmitter is damaged, contact Vaisala or Vaisala distributor and send the instrument to the addresses below:
Please, include a short description of the fault and the date and place of purchase.
NORTH AMERICAN SERVICE CENTER Vaisala Inc., 100 Commerce Way, Woburn, MA 01801-1068, USA.
Phone: +1 781 933 4500, Fax +1 781 933 8029 Email: us-customersupport@vaisala.com
EUROPEAN SERVICE CENTER Vaisala Instruments Service, Vanha Nurmijärventie 21 FIN-01670 Vantaa,
FINLAND. Phone: +358 9 8949 2758, Fax +358 9 8949 2295 E-mail: instruments.service@vaisala.com
ASIAN SERVICE CENTER Vaisala KK, 42 Kagurazaka 6-Chome, Shinjuku-Ku, Tokyo 162-0825, JAPAN.
Phone: +81 3 3266 9611, Fax +81 3 3266 9610 E-mail: aftersales.asia@vaisala.com
www.vaisala.com
Page 10
General ___________________________________________________________________________
Compliance with the requirements
European requirements
EU directive 94/9/EC (ATEX 100a) Standards EN 50014 and EN 50020 Standard EN 50284 for apparatus of equipment group II, category 1 G (intended for hazardous locations) Marking EEx ia IIC T4: at temperature range -20...+60 °C and pressure range 0.8...1.1 bar.
US requirements (FM)
FM approved for Classes I, II & III, Division 1, Groups A, B, C, D, E, F & G and Division 2, Groups A, B, C, D, F and G. Intrinsically safe when connected as shown in Appendix 2.
Canadian requirements (CSA)
CSA approved for Class I, Division 1 and 2, Groups A, B, C, D; Class II Division 1 and 2, Groups G and Coal Dust; Class III. Intrinsically safe when connected as shown in Appendix 3.
Australian requirements (TestSafe)
TestSafe approved for Ex ia IIC T5 IP65 (Zone 0).
Page 11
Installation ________________________________________________________________________
Vaisala _________________________________________________________________________ 11
CHAPTER 3
INSTALLATION
General instructions
NOTE
Mounting the probe cable in gas group IIC spaces
The following instructions shall be followed to fulfil the specifications of the EN 50284 for non conductive layer of the probe cable:
Never mount or handle the probe cable when hazardous gases are
present !
For achieving a conductive shield, cover the probe cable with
conductive material like metal or conductive tape or mount the probe cable in a metal conduit !
Assure that the conductive shield fulfils requirements of the
standard EN50284 (resistance less than 1 G) and make sure that it can not come loose in any operation situation !
NOTE
During the installation work of the sensor heads in gas group IIC areas (requiring category I devices), it has to be guaranteed that even in fault cases sparks generated by impacts or friction on the housing's surface can never occur.
Selecting a place for the transmitter
Select a place with stable conditions for mounting the transmitter. Do not expose the transmitter to direct sunlight or rain. A rain shield for HMT360 transmitter is available. When mounting the sensor head, select a place representing the process conditions.
Page 12
Installation_________________________________________________________________________
Temperature differences
If there is a difference between the process and ambient temperatures, it is recommended to mount the sensor head and at least one meter of cable inside the process. This is to avoid incorrect relative humidity readings due to temperature difference. If this installation is not possible and the sensor head must be installed through the wall (e.g. with a flange), the sensor head and the cable must be insulated carefully.
Checking the temperature reading
The actual temperature of the process can be measured with a reference instrument to be compared with the transmitter reading. The heat transfer is less evident if you remove the protective filter of the sensor for a short-term test. However, never use the sensor long periods without the filter as this may cause a faster contamination of the sensor. The transmitter fulfils the specified EMC regulations with the protective filter on the sensor head.
Page 13
Installation ________________________________________________________________________
Vaisala _________________________________________________________________________ 13
Mounting the transmitter
FIGURE 3-1 Mounting the transmitter and detaching the probe
FIGURE 3-2 Different parts of the transmitter
1. Attach the mounting plate to the wall with 4 screws.
2. Press down the transmitter so that it slides along the rails of the mounting plate.
3. Fasten the transmitter to the mounting plate with the Allen screw.*
4. The probe can be detached and replaced when needed by simply unfastening the two Allen screws.
* 3 mm Allen key provided
1. electronics unit
2. transmitter base
3. probe; including a part o
f
measurement electronics (e.g. calibration memory and the sensor head)
4. protective covers
5. flat cable
6. sensor head
7. cable glands
8. grounding terminals
9. RS232 connector
8
7
9
Page 14
Installation_________________________________________________________________________
Mounting the probes
NOTE
- Do not unsolder and then again resolder the sensor head cable
from and to the printed board during installation.
- Do not shorten or lengthen the probe cable
These procedures may alter the humidity calibration of the transmitter.
The HMP362 - small pressure-tight probe
The HMP362 probe is a small pressure-tight probe equipped with installation flange. It is suitable for dewpoint measurements in natural gas. When sampling in pressurized processes, the sampling cell HMP302SC is available as an optional accessory.
HMP362 installation (without the sampling cell)
or chamber
3.2
m
i
n
.
ø
6
0
.
3
3
(
2
"
)
m
i
n
.
2
-
3
m
m
0
.
1
"
)
Screws, hexagon socked head (provided)
HMP362 Probe
Process pipe
Threaded sleeve
(not provided)
Threaded sleeve
O-Ring (provided)
ø
2
4
.
5
M6
9.5 (3/8")
(
3
1
/
3
2
"
)
Page 15
Installation ________________________________________________________________________
Vaisala _________________________________________________________________________ 15
CAUTION
In pressurized processed it is essential to tighten the supporting nuts and screws very carefully to prevent loosening of the probe by the action of pressure.
NOTE
If the HMP362 is installed in a process with a pressure differing from normal atmospheric pressure, please enter the pressure value of the process (in hPa or mbar) into the transmitter memory via the serial line (see page 42) or via the display/keypad (see page
33) .
The HMP363 probe for confined spaces
The HMP363 is a small size general-purpose probe suitable for installation in ducts and channels with the installation kit available. The installation kit includes a flange, a supporting bar for the sensor head cable and screws for attaching the flange to the duct wall. The distance between the sensor and the channel wall can be adjusted. The adjustment range is 100 - 320 mm, measured from the tip of the sensor head to the flange.
HMP302SC (optional sampling cell)
Probe
Gas in
Gas out
Probe
Sampling cell
Clamp ( not needed if sampling cell is suspended on the piping)
Page 16
Installation_________________________________________________________________________
FIGURE 3-3 Installation of the sensor head HMP363 in a channel
with a flange and a supporting bar
The HMP365 probe for high temperatures
When installing the HMP365 sensor head in a duct or channel, there should not be remarkable difference between temperatures inside and outside of the duct. This can cause incorrect humidity readings.
FIGURE 3-4 Mounting the HMP365 sensor head in a duct or
channel.
HMP364 for pressurized spaces
The HMP364 pressure tight probe is designed for the humidity measurements in pressurized rooms and industrial processes. This probe is also available for dewpoint measurements in natural gas. The probe is provided with a nut, a fitting screw and a sealing washer to make the installation easier.
1. duct wall
2. plugged hole for reference measurements is recommended
3. flange
4. seal
5. supporting bar
6. sensor head
1. a plugged hole for reference measurements
2. sensor head
3. mounting flange
Page 17
Installation ________________________________________________________________________
Vaisala _________________________________________________________________________ 17
CAUTION
In pressurized processed it is essential to tighten the supporting nuts and screws very carefully to prevent loosening of the probe by the action of pressure.
1. Unscrew the fitting screw from the nut and the sensor head.
2. Fasten the fitting screw to the chamber wall with a sealing washer.
Tighten the fitting screw into the threaded sleeve with a torque spanner. The tightening torque is 150±10 Nm (110±7 ft-lbs)
3. Insert the body of the sensor head into the fitting screw and screw
the nut manually to the fitting screw until the connection feels tight.
4. Mark both the fitting screw and the nut hex.
5. Tighten the nut a further 30° (1/12 turn) or if you have a torque
spanner tighten it with a torque of 80 ±10 Nm (60±7 ft-lbs).
NOTE
After detachment the nut must be tightened without increased effort.
6. Clean and grease the tightening cone of the fitting screw after every
tenth detachment. Change the sealing washer every time the fitting screw is detached. Use high-vacuum grease (e.g. Down Corning, Europe) or a similar grease.
1. tightening cone
2. nut
3. fitting screw
4. sealing washer
5. sensor head
;
13.5 mm
1. fitting screw
2. sealing washer
3. tightening cone
4. clean cotton stick
Page 18
Installation_________________________________________________________________________
NOTE
If the HMP364 is installed in a process with a pressure differing from normal atmospheric pressure, please enter the pressure value of the process (in hPa or mbar) into the transmitter memory via the serial line (see page 42) or via the display/keypad (see page 33) .
Page 19
Installation ________________________________________________________________________
Vaisala _________________________________________________________________________ 19
The HMP368 probe for measuring humidity in pressurized pipelines or moisture in liquids
The HMP368 probe is available with two lengths, standard 178 mm and optional 400 mm. The probe is provided with a sliding fit which enables different installation depths and makes the installation and removal easier in pressurized processes. The probe is specially designed for the measurements in pipelines. This probe is also available for dewpoint measurements in natural gas.
When the probe is used in a pressurized pipeline or in liquids (e.g. feeder routes) the sensor head should preferably be installed directly in the process through a ball valve assembly, see page 20.
CAUTION
In pressurized processed it is essential to tighten the supporting nuts and screws very carefully to prevent loosening of the probe by the action of pressure.
NOTE
When the HMP368 is installed in a process with a pressure
differing from normal atmospheric pressure, please enter the pressure value of the process (in hPa or mbar) into the transmitter memory via the serial line (see page 42) or via the display/keypad (see page 33).
Make sure that the temperature at the measurement point is equal
to that of the process, otherwise the humidity reading may be incorrect.
Page 20
Installation_________________________________________________________________________
FIGURE 3-5 Probe dimensions: standard 178 mm and optional
400 mm
CAUTION
Take care not to damage the probe body. A damaged body makes the probe head less tight and may prevent it from going through the clasp nut.
Tightening the clasp nut
1. Adjust the probe to a suitable depth according to the type of
installation.
2. Tighten the clasp nut first manually.
3. Mark the fitting screw and the clasp nut.
4. Tighten the nut a further 50 - 60° (ca.1/6" turn) with a fork spanner.
If you have a suitable torque spanner, tighten the nut to 45±5 Nm (33±4 ft-lbs).
1. clasp nut, 24 mm hex nut
2. fitting screw, 27 mm hex head
A: probe 178 mm: adjustment range 120 mm; A: probe 400 mm: adjustment range 340 mm
Page 21
Installation ________________________________________________________________________
Vaisala _________________________________________________________________________ 21
FIGURE 3-6 Tightening the clasp nut.
NOTE
Take care not to tighten the clasp nut more than 60° to avoid difficulties when opening it.
Installing the probe through the ball valve assembly
The best way for installing the sensor head to pressurized process or pipeline is to install it through the ball valve assembly. This installation makes it unnecessary to empty or shut down the process for installing or removing the sensor head. Use the Vaisala DMP248BVS ball valve set or a 1/2" ball valve assembly with a ball hole of 14 mm or more. If you install the sensor head ( 13.5 mm) in a process pipe, please note that the nominal size of the pipe must be at least 1 inch (2.54 cm). The manual press tool can be used to press the sensor head into the pressurized (< 10 bar) process or pipeline.
1. probe
2. clasp nut
3. pen
4. fitting screw
Page 22
Installation_________________________________________________________________________
FIGURE 3-7 Installing the sensor head through the
DMP248BVS ball valve assembly.
Install the sensor head transversely against the direction of the process flow.
NOTE
Installation or removing the probe through a ball valve assembly is possible without shutting the process when the pressure is less than 10 bar. With 10...40 bar pressures you must shut down the process before installing or removing the probe. The maximum operation pressure for the probe HMP368 is 40 bar.
1. sensor head
2. handle of the ball valve
3. ball of the ball valve
4. process chamber / pipeline
5. manual press tool
6. the groove on the probe indicates the upper adjustment limit
7. filter
1
2
3
4
5
6
7
Page 23
Electrical connections________________________________________________________________
Vaisala _________________________________________________________________________ 23
CHAPTER 4
ELECTRICAL CONNECTIONS
Refer to local requirements regarding cabling, grounding and galvanic isolator or barrier connections.
WARNING
Be sure that the main power switch of the transmitter is set off before making any electrical installations in hazardous areas.
1. Open the transmitter cover and remove the protective cover of
the transmitter base.
2. Thread the power supply wires through the cable gland (Figure
3-2, item 7)
3. Connect the unpowered power supply wires to the connectors:
CH1 (humidity) and CH2 (temperature). Both channels require an own power supply. NOTE! As CH1 is a main output, the transmitter does not operate if only CH2 is connected (CH2 is optoisolated from transmitter electronics).
4. Replace the protective cover. Turn the transmitter on ON with
the ON/OFF switch (Figure 9-1, item 2).
5. Close the cover. The transmitter is ready for use.
When using the transmitter in hazardous places, the use of galvanic isolators or barriers is essential. The following barrier & isolator are available in Vaisala: barrier No. 210664 (STAHL 9001/51-280-091-141) for USA&Canada and galvanic isolator No. 212483 (STAHL 9160/13-11-11) or barrier No. 210664 for EU. Examples of connections are presented on page 25.
Page 24
Electrical connections ________________________________________________________________
Vaisala _________________________________________________________________________ 24
Installation in hazardous locations
US and Canadian requirements
USA (FM): Wiring for intrinsically safe operation is shown in APPENDIX 2. Canada (CSA): Wiring for intrinsically safe operation is shown in APPENDIX 3.
European requirements
CATEGORY 1 (zone 0)
The HMT360 has to be connected to Exia-certified associated apparatus with galvanic isolation, gas group IIB or IIC. NOTE! If both analog outputs are in use, the channel 1 (-) and channel 2 (-) must be short circuited (see figure 4-3).
CATEGORY 2 or 3 (zone 1 or 2)
The HMT360 has to be connected either to a zener barrier or galvanic isolator NOTE! If both analog outputs are in use with a galvanic isolator, the channel 1 (-) and channel 2 (-) must be short circuited (see figure 4-3).
Figures 4-1 and 4-2 present examples of galvanic isolators and zener barrier connections (only CH1 connected).
Maximum load calculation for the barrier
Always check the technical data of galvanic isolator or zener barrier to define maximum load. If the maximum load is not included in specifications it can be determined as follows.
Example: Stahl 9002/13-280-093-00 (zener barrier)
General specifications of the HMT360 Supply voltage Uin = 24 V (12...35 V) Maximum current I
max
= 20 mA (100%RH)
Minimum operating voltage for HMT360 U
min
= 12 V (15 V with serial port )
Cable resistance (as an example) R
cable
= 0.085 ohm/m/core (2*0.085 ohm/m/pair)
Cable lenght =250 m
R
Ctot
= 2*0.085 ohm/m*250m= 45 ohm
Stahl 9002/13-280-093-00 (values taken from the specifications): Typical end to end resistance R
barrier
= 340 ohm
Max. additional voltage drop (20mA) U
add
= 1 VDC
Voltage drop U
drop
= I
max
*(R
Ctot
+ R
barrier
) + U
add
= 20 mA*(45+340)ohm + 1 V = 8.7 V
Voltage for R
l
U
Rl
= U
in
- U
min
- U
drop
= 24 V - 12 V - 8.7 V = 3.3 V
Maximum load R
lmax
= URl/I
max
= 3.3 V / 20 mA = 165ohm , R
lmax
< 165ohm
Page 25
Electrical connections________________________________________________________________
Vaisala _________________________________________________________________________ 25
HMT360 connected to a galvanic isolator
a) current signal controller
b) voltage signal controller
FIGURE 4-1. HMT360 connected to a galvanic isolator.
C H 1
+ +-
-
C H 2
Rl
U
Power Supply
Galvanic Separator
Controller
U = I*Rl
L+ L-
C H 1
+
+
--
C H 2
I
Power Supply
Galvanic Separator
Controller
L+ L-
I = 4...20mA
HAZARDOUS AREA SAFE AREA
HAZARDOUS AREA SAFE AREA
Page 26
Electrical connections ________________________________________________________________
Vaisala _________________________________________________________________________ 26
HMT360 connected to a zener barrier
a) current signal controller
b) voltage signal controller
FIGURE 4-2 HMT360 connected to a zener barrier.
C H 1
+
+-
-
C H 2
Power Su pply
Zener Barrier
I = 4...20mA
I
Controller
C H 1
+
+
-
Rl
U
Power Supply
Zener Barrier
Controller
U = R x I
+
-
-
C H 2
HAZARDOUS AREA SAFE AREA
HAZARDOUS AREA
SAFE AREA
Page 27
Electrical connections________________________________________________________________
Vaisala _________________________________________________________________________ 27
Examples of connections
FIGURE 4-3 STAHL 9160/13-11-11 (galvanic isolator).
FIGURE 4-4 STAHL 9001/51-280-091-141 (zener barrier).
Make the connection between the
- terminals always when using a galvanic isolator !
STAHL 9160/13-11-11
-Ch2+
HMT360 Connection
HAZARDOUS AREA
-Ch1+
board
SAFE AREA
Load Ch1
Load Ch2
V+
V-
18...31,2VDC
V-
V+
18...31,2VDC
12
10
10
12
7 9 1 2
9
2
1
7
(+)
(-)
(-)
(+)
HAZARDOUS AREA SAFE AREA
HMT360 Connection board
-Ch1+ -Ch2+
Load Ch2
20...35 VDC
20...35 VDC
Load Ch1
STAHL 9001/51-280-091-141
1
2
2
1
4
3
3
4
Page 28
Electrical connections ________________________________________________________________
Vaisala _________________________________________________________________________ 28
FIGURE 4-5 MTL5044 (galvanic isolator).
Grounding
When grounding the transmitter, follow the local requirements. Use at least 4 mm
2
grounding cable when grounding the transmitter or barrier. Note that the allowed resistance between barrier and system ground must be less than 1 ohm. See the figure below.
FIGURE 4-6 Grounding.
HAZARDOUS AREA
-Ch2+-Ch1+
HMT360 Connection board
5
4
1
2
9
12
11
SAFE AREA
Load Ch1
Load Ch2
20...35VDC
Vs+
Vs-
-
+
-
+
+
-
+
-
14
13
MTL5044
Zener Barrier
Instrument system
Barrier ground
System groun
d
-Ch1+
-Ch2+
HAZARDOUS AREA SAFE AREA
Page 29
Operation _________________________________________________________________________
Vaisala _________________________________________________________________________ 29
CHAPTER 5
OPERATION
Turn the transmitter on ON with the ON/OFF switch (Figure 9-1, item
2).
HMT360 with display
Measurement readings appear on the display after switching power ON. The analog output signals can be read from the system or load resistor.
HMT360 for Natural Gas
When switching on the HMT360 for natural gas, the existing pressure setting appears on the display; the default setting is 1000 hPa (=1 bar or 14.5 psi).
To modify the pressure setting, turn the transmitter's internal
CALIBRATION ENABLED/DISABLED dip switch to position ENABLED
(up). Adjust the pressure reading with buttons and on the display cover; the adjustment step is 100 hPa). Acknowledge the value with button E. To complete the pressure setting, turn the dip switch back to position DISABLED (down). If the pressure setting is not modified, the measurement readings appear automatically on the display after 60 seconds.
NOTE
Avoid static discharge! Always use a wet cloth for wiping the display.
HMT360 without display
If the transmitter does not have a display, the red LED indicator on the cover indicates various phases of calibration and possible errors. In normal operation it is not lit. If the LED is lit and all calibration or test dip switches are disabled, it is an indication of an internal error.
Page 30
Operation _________________________________________________________________________
Vaisala _________________________________________________________________________ 30
Otherwise, the LED indicates the following:
LED blinking = calibration of the dry end (offset)
LED lit = calibration of the wet end (gain)
NOTE
If calibration is concluded but dip switch settings are not restored, the LED indicator keeps blinking.
Dip switch functions
The table below is also printed on the protection board:
n o p q r s t u
n: ANALOG OUTPUT TEST ON/OFF:
if you turn the switch to ON position (up), you can force the outputs to states 4 mA, 12 mA and 20 mA by pressing the buttons and ▼on the cover.
o: CALIBRATION DISABLED/ENABLED:
the EEPROMs are write protected. If this switch is in the DISABLED position (down), it does not allow any calibrations or scalings.
NOTE
Keep this switch always in the disabled position during normal use of the transmitter!
Page 31
Operation _________________________________________________________________________
Vaisala _________________________________________________________________________ 31
p and q CALIBRATION RH, T, ANALOG:
with these combinations you can perform relative humidity, temperature or analog output calibrations with a multimeter or with the transmitter display unit. Turn the dip switches to the desired position according to the table printed on the protective cover.
r OUTPUT QUANTITIES:
determines whether the output units are metric (down) on non-metric.
s, t and u SELECT OUTPUT QUANTITIES:
with the three dip switches on the right, you can select the output quantities according to the table printed on the right side of the protective cover. Note that only the ordered quantities can be selected. The SPECIAL option gives you the choice of setting any ordered quantity to each channel.
NOTE
Always restore the dip switch settings after having tested the analog outputs or performed the calibration.
Page 32
Operation errors ____________________________________________________________________
Vaisala _________________________________________________________________________ 32
CHAPTER 6
OPERATION ERRORS
The following indicates an operation error of the HMT360:
Analog output(s) current drops below 4 mA
Display shows ERROR
LED on the cover is lit and no calibration is going on (transmitters
without a display)
In the case of error:
Check that the sensor is connected properly
Check if there is condensed water in the probe, if yes, let the probe dry
Cleaning the sensor from glycol in a natural gas application
If a glycol layer on the VAISALA HUMICAP® sensor is disturbing the moisture measurement, the sensor can be cleaned on site soaking the sensor to de-ionized water or to IPA (isopropanol, propan-2-ol).
Page 33
Operation errors ____________________________________________________________________
Vaisala _________________________________________________________________________ 33
1. Remove the sintered filter carefully.
2. Dip the sensor in de-ionized water or IPA (max 1 minute).
3. Gently blow the sensor dry with dry gas (Nitrogen) if available.
4. When the sensor and probe are dry, set the filter back on the
probe.
NOTE
Do not touch the sensor surface, especially if removing the HUMICAP® sensor from the probe.
Only dipping to de-ionized water or IPA is allowed, other agents like ethanol cannot be used.
Do not keep the sensor for more than one minute in the water or IPA.
The water or IPA temperature must be below 30 °C (86 °F).
Do not use a cotton stick or any other mechanical means for cleaning and/or drying the sensor.
In case of constant error, please contact Vaisala Service (see Chapter 2: Maintenance).
Page 34
Display/keypad commands____________________________________________________________
Vaisala _________________________________________________________________________ 34
CHAPTER 7
DISPLAY/KEYPAD COMMANDS
With these commands you can scale the outputs and select special output quantities if the transmitter configuration allows that.
NOTE
Chapter 9 describes separately display/keypad commands for the calibration.
Setting the pressure for calculations
In humidity transmitters, ambient pressure is used for the mixing ratio and wet bulb calculations. In dewpoint transmitters (natural gas applications), the process pressure in the measurement point is required to achieve the specified accuracy.
To modify the pressure settings, turn the transmitter's internal
CALIBRATION ENABLED/DISABLED dip switch to position ENABLED
(up). Press button C on the display cover: text "SCAL" appears on the display. Then press button E until the following display appears (the numeric value is always the existing setting, in this example 1013.25):
Adjust the pressure reading with buttons and . Acknowledge the value with button E. To complete the pressure setting, turn the dip switch back to position DISABLED (down).
See the pressure conversion table on page 42.
Page 35
Display/keypad commands ___________________________________________________________
Vaisala _________________________________________________________________________ 35
Selecting the output quantities
HMT360 for Natural Gas
Two quantities are shown on the LCD. The upper half of the display shows the moisture, whereas the lower half is reserved for the temperature reading.
Upper half of the display
To modify the displayed quantity, turn the transmitter's CALIBRATION
ENABLED
/DISABLED dip switch to position ENABLED (up). Select the required quantity with button on the display cover, and acknowledge the value with button E. To complete the selections, turn the dip switch back to position DISABLED (down).
When the OUTPUT QUANTITIES METRIC/NON METRIC dip switch is in the
METRIC (down) position, the available quantities are dew/frost point
temperature (Td °C), parts per million (ppm), and absolute humidity in mg/m3 (W). When the dip switch is in the NON METRIC (up) position, the available quantities are dew/frost point temperature (Td °F), parts per million (ppm), and absolute humidity in lb/mmscf (W).
Lower half of the display
The OUTPUT QUANTITIES METRIC/NON METRIC dip switch can be used to select between °C and °F.
It is possible to check the pressure setting of the transmitter by pressing button . To return to the temperature reading, press button again.
Selecting the analog outputs
You can select the output quantities for channels 1 and 2 by turning the transmitter's CALIBRATION ENABLED/DISABLED dip switch to position ENABLED (up) and the three OUTPUT SELECTION dip switches to position
SPECIAL (all up).
Page 36
Display/keypad commands____________________________________________________________
Vaisala _________________________________________________________________________ 36
Press button C on the display cover: text "SCAL" appears on the display. Then press button E until the following display appears:
The characters on the second line of the display in this menu correspond to the quantities according to the following table:
Table 2 Available Quantities
character quantity abbreviation availability
0 relative humidity RH A, D 1 temperature T A, D, F, H, N 2 dewpoint temperature Td D, N 3 absolute humidity a D 4 mixing ratio x D 5 wet bulb temperature Tw D 6
volume concentration
ppm H, N 7 water activity aw H, F 8 relative humidity of saturation RS H 9 saturation temperatur Ts H A water content w N
A: available for HMT360A D: available for HMT360D F: available for HMT360F H: available for HMT360H N: available for HMT360N
Select the quantity for channel 1 with buttons and and acknowledge the selection with button E.
If the transmitter is equpped with two analog channels, select the quantity for channel 2 in the same way, for example:
Press button C to exit the display command mode or continue by setting the pressure.
NOTE
Remember to restore the dip switch settings.
Page 37
Display/keypad commands ___________________________________________________________
Vaisala _________________________________________________________________________ 37
Scaling the analog outputs
Turn the transmitter's internal dip switch CALIBRATION ENABLED/
DISABLED to upward position (ON). Press button C on the display
cover and a text similar to the following appears:
Numbers on the second line indicate the low end scaling of channel 1 currently stored in the transmitter memory. The text SET LO on the lower left corner indicates that you can now change the low end scaling with buttons and . Acknowledge the value with button E and a text similar to the following appears:
Numbers on the second line indicate the high end scaling of channel 1. The text SET HI on the lower left-hand corner indicates that you can now change the high end scaling with buttons and ▼. Acknowledge with button E.
If there is another channel available, the display shifts to the scaling menu of channel 2. You can now scale the analog outputs for the channel 2 in the same way as described above.
Press button C to exit the display command mode or continue by selecting the output quantities. This menu starts automatically after the scaling menu only if the
OUTPUT SELECTION dip switches are on the
position SPECIAL (all up) from the beginning.
NOTE
Remember to restore the dip switch settings.
LO
Page 38
Serial interface _____________________________________________________________________
Vaisala _________________________________________________________________________ 38
CHAPTER 8
SERIAL INTERFACE
WARNING
The serial interface MUST NOT be used in hazardous areas.
Use the serial interface for calibration and testing purposes in safe areas only. Always use the serial interface cable (optional accessory, order code: 25905ZZ). Connect one end of the cable to the serial port of your computer and the other to the connector marked 'RS 232C' on the transmitter's electronics unit (see Figure 3-2, item 9).
NOTE
With serial communication the current consumption increases approximately to 7 mA and the transmitter is not able to operate with 4 mA. Therefore, it is recommended to use serial communication only temporarily for changing settings or for calibrating the transmitter in a safe area. With serial port, the minimum supply voltage is 15 VDC.
NOTE
The transmitter incorporates a serial interface detector. However, not all terminals or PC serial ports (e.g. opto-isolated or ports not meeting RS232C standards) recognize this standard. If the communication is not possible via a serial interface, use a forced activation by pressing simultaneously buttons and E on the transmitter cover. To deactivate the forced activation, press these buttons again or reset the transmitter. If the transmitter does not receive any commands for half an hour, it automatically closes the serial communication.
Serial communication settings
TABLE 8-1 Serial communication settings
parameter value
bauds 2400 parity none data bits 8 stop bits 1
Page 39
Serial interface _____________________________________________________________________
Vaisala _________________________________________________________________________ 39
To start giving commands, make sure that the HMT360 is connected to a serial port of your computer and that the terminal session is open. Give commands by typing them on your computer according to the following instructions. In these commands, stands for pressing
ENTER.
NOTE
Each time the HMT360 transmitter sends something to the terminal, it also sends line feed and carriage return.
Selecting the analog outputs
Turn the internal dip switch CALIBRATION ENABLED/DISABLED to ON-position before the selecting the analog outputs and return it to OFF- position after selecting .
ASEL xxx yyy
where: xxx = quantity of channel 1 and yyy = quantity of channel 2
output quantity abbreviation
relative humidity RH temperature T dewpoint temperature Td absolute humidity a mixing ratio x wet bulb temperature
1
Tw
mass proportion
1
ppm
water activity
1
aw relative humidity of saturation1RS saturation temperature
1
Ts water content
2
w
1
Available only for HMT368H for moisture measurements in liquid.
2
Available only for HMT360N for natural gas
Example: >asel rh t
Ch1 RH lo : 0.00 %RH Ch1 RH hi : 100.00 %RH Ch2 T lo : -40.00 'C Ch2 T hi : 60.00 'C >
Page 40
Serial interface _____________________________________________________________________
Vaisala _________________________________________________________________________ 40
Scaling the analog outputs
Turn the internal dip switch CALIBRATION ENABLED/DISABLED to ON-position before the scaling the analog outputs and return it to OFF- position after scaling.
Szz aa.a bb.b
where: zz = quantity (RH, T, Td, x, a, Tw, ppm, aw, RS, Ts)
aa.a = lower limit of the quantity bb.b = upper limit of the quantity
Example: Scaling the relative humidity:
>SRH 0 100 RH lo 0.00 %RH RH hi 100.00 %RH >
Exception: When you have natural gas transmitter and you scale the quantity of water content w (mg/kg or lb/mscf) give a command
SNG1.
Adjustment commands
Turn the internal dip switch CALIBRATION ENABLED/DISABLED to ON-position before the calibration and return it to OFF- position after calibration.
Note that HMT360 for natural gas can be checked but not adjusted by a user.
Relative humidity adjustment
NOTE
Read also the calibrator e.g. HMK15 manual and refer to page 47 of this manual for more detailed instructions on salt bath calibration.
CRH
Page 41
Serial interface _____________________________________________________________________
Vaisala _________________________________________________________________________ 41
The transmitter asks and measures relative humidity and calculates the calibration coefficients, for example:
>CRH RH : 1.82 1. ref ? 0 Press any key when ready... RH : 74.22 2. ref ? 75
OK >
The OK indicates that the calibration has succeeded.
Temperature adjustment
CT
The transmitter asks and measures temperature readings and calculates the calibration coefficients, for example:
>CT T : 0.81 1. ref ? 0.5 Press any key when ready... T : 99.12 2. ref ? 99.5
OK >
The OK indicates that the calibration has succeeded. When performing one-point calibration press only for 2. reference.
Analog output calibration
Connect the HMT360 to a multimeter. Give the following command.
ACAL
1. Disconnect the serial cable from the transmitter while reading the
multimeter value for the CH1 (I1).
2. Reconnect the serial cable. Type the multimeter reading and press
↵.
3. Type the higher current multimeter reading and press ↵.
>ACAL Ch1 I1 (mA) ? 4.846 Ch1 I2 (mA) ? 19.987
>
Page 42
Serial interface _____________________________________________________________________
Vaisala _________________________________________________________________________ 42
Output commands
Testing the analog outputs
This command outputs the current value of each channel and the corresponding control signal of the digital-to-analog converter.
ITEST aa.aaa bb.bbb
where: aa.aaa = current value to be set for channel 1 (mA)
bb.bbb = current value to be set for channel 2 (mA)
For example:
>ITEST 8 12
8.000 3F8 12.000 70O >
>ITEST
6.349 25A 19.001 E93 >
The set current values remain valid until you give the command ITEST without readings. With this command the desired outputs of the transmitter are shown.
Note! When outputting low currents from channel 1, remember to remove RS232C-cable while reading the current output, because of the increased current consumption for using RS port.
Outputting the measurement values
SEND
This command outputs the measured values in one point.
Activating a continuous output
R
With the command R the transmitter outputs measured values continuously.
Page 43
Serial interface _____________________________________________________________________
Vaisala _________________________________________________________________________ 43
Stopping the continuous output
S
The continuous outputting is stopped with the command S. If outputting is active, this command is not echoed.
Setting the output interval
INTV n xxx
where: n = 1 - 255 xxx = S, MIN or H
Sets the output interval when the transmitter outputs measurement readings. The time interval is used when the continuous output is active. For example, the output interval is set to 10 minutes:
>INTV 10 min Output intrv. : 10 min >
Setting the ambient pressure for calculations
Turn the internal dip switch CALIBRATION ENABLED/DISABLED to ON-position before setting the pressure and return it to OFF- position after setting.
PRES aaaa.a
where: aaaa.a = pressure (hPa)
>PRES 2400
Pressure : 2400 >
Pressure conversion chart
mmHg/Torr inHg atm bar psi
hPa/mbar
1 1.333224 33.86388 1013.25 1000 68.94757
mmHg/Torr
0.7500617 1 25.40000 760 750.0617 51.71493
inHg
0.02952999 0.03937008 1 29.921 29.52999 2.036021
atm
0.00098692 0.00131597 0.033422 1 0.98692 0.068046
bar
0.001 0.001333224 0.03386388 1.01325 1 0.06894757
psi
0.01450377 0.01933678 0.4911541 14.6962 14.50377 1
Example: 29.9213 inHg = 29.9213 x 33.86388 = 1013.25 hPa / mbar
NOTE
Conversions from mmHg and inHg are defined at 0 °C and for mmH2O and inH2O at 4 °C.
FROM
TO
Page 44
Measuring at overpressure ____________________________________________________________
Vaisala _________________________________________________________________________ 44
CHAPTER 9
MEASURING AT OVERPRESSURE
The probes HMP362, HMP364 and HMP368 are designed for humidity measurement at overpressure. The maximum measurement pressures depend on the probe as follows:
HMP362: 0...167 bar (16,7 MPa), for natural gas, sample cell
HMP302SC available
HMP364: 0...100 bar (10 MPa), for pressurized rooms and processes,
probe is provided with a nut, fitting screw and sealing washer
HMP368: 0...40 bar (4 MPa), for pressurized pipelines, ball valve set
available
The accuracy of the dewpoint measurement is affected by pressure in the measurement chamber. The actual pressure in the sampling cell shall be set to the transmitter by using the serial line command PRES (Setting the ambient pressure for calculations), page 42 or by using the keypad commands, page 33.
CAUTION
In pressurized processed it is essential to tighten the supporting nuts and screws very carefully to prevent loosening of the probe by the action of pressure.
Pressure regulator recommended
When sampling pressurized processes exceeding the maximum measurement pressure of the probe, the pressure in the measurement chamber must be regulated to acceptable level or below. It is recommended to use pressure regulator before the measurement chamber to prevent remarkable pressure variations.
Page 45
Calibration and adjustment____________________________________________________________
Vaisala _________________________________________________________________________ 45
CHAPTER 10
CALIBRATION AND ADJUSTMENT
Calibration interval
HMT360 is calibrated as shipped from the factory. Typical humidity calibration interval is one year. Depending on the application it may be good to make the first calibration check earlier.
Factory calibration and adjustment
The device (or only the electronics unit) can be sent to Vaisala Service Centers for calibration and adjustment, see contact information on page 9. HMT360 for
natural gas shall be adjusted always in Vaisala Service Center.
Calibration and adjustment by the user
Note that HMT360 for natural gas can be checked but not adjusted by a user.
Calibration is carried out by using the keypad or by using serial commands. The following additional equipment is needed in calibration of the HMT360:
power supply for the channel(s) to be calibrated (12-24 VDC)
a multimeter for the HMT360 without a display
the calibrated references
Firstly, the electronics unit is removed from the transmitter base to perform a calibration in safe area. Secondly, the power supply and a multimeter is connected to the electronics unit according to the following instructions.
Page 46
Calibration and adjustment ____________________________________________________________
Vaisala _________________________________________________________________________ 46
Removing the electronics unit for the calibration
Connections
Connections
FIGURE 9-2 Connecting the power supply and multimeter for calibration.
1. Unfasten the screws and open the cover.
2. Switch the transmitter off with the ON/OFF switch.
3. Disconnect the flat cable by lifting it carefully e.g. with a screw-driver.
4. Turn the electronics unit slightly upwards to release it from the hinges. Leave the transmitter base with the cable connections on place. When putting the electronics unit back to place, attach the upper hinge first.
FIGURE 9-1 Detaching the electronics
unit with the probe for calibration.
Page 47
Calibration and adjustment____________________________________________________________
Vaisala _________________________________________________________________________ 47
1. Connect with banana plugs a power supply (12...24 VDC, with a serial port the
minimum supply voltage is 15 VDC ) to the terminals B1(-) and C1(+) (channel 1), see Figure 9-2.
2. Connect the multimeter in series with the supply, B1(-) and A1(+).
3. Follow the same procedure with channel 2 using the terminals B2 (-) and C2 (+). When
calibrating both channels at the same time, use two galvanically separated power supplies.
You can now calibrate or check the humidity and temperature or the analog outputs according to the instructions given in calibration chapters.
If the current measurement is needed in a hazardous area, the multimeter is connected to the terminals A1/A2 (+) and B1/B2 (-). Use only an approved multimeter.
WARNING
The power supply connectors (C1 and C2) MUST NOT be used in hazardous areas. For calibration in a hazardous area use only an approved multimeter, which fulfils the safety factors printed on the protective cover.
Calculating the correspondence of current values and output quantities
When using HMT360 without a display, calibration is carried out by using a multimeter. Use the following equations to calculate the current values corresponding the reference output quantities.
min)max(
min)(
164
QQ
QQref
mAmAI
+= (9-1)
where: Qref = reference value of the calibrated quantity
Qmin = value corresponding to 4 mA Qmax = value corresponding to 20 mA
Example 1:
relative humidity scaling 0 - 100 %RH, reference 11.3 %RH:
mA
RHRH
RHRH
mAmAI 808.5
)%0%100(
)%0%3.11(
164 =
+=
Example 2:
temperature scaling -40 - +120 °C, reference 22.3 °C
mA
CC
CC
mAmAI 230.10
)40120(
)403.22(
164 =
°°
°°
+=
Page 48
Calibration and adjustment ____________________________________________________________
Vaisala _________________________________________________________________________ 48
Relative humidity calibration
Automatic two-point calibration (only with HMT360 with display)
The automatic calibration procedure is a user friendly way to calibrate with salt solutions; user do not have to feed the reference values to the HMT360. The transmitter displays the accurate value based on the measured temperature and the Greenspan table stored into the transmitter memory. The display chart of the calibration procedure is presented in FIGURE 9-3 Display chart of the automatic calibration procedure.
Set the calibration dip switch to ENABLED position and select RH
calibration by using switches 3 and 4 (see Chapter 5: Dip switch functions).
Remove the filter and insert the probe head into a measurement
hole of the LiCl salt chamber in the humidity calibrator. Acknowledge by pressing the button E. Alternatively select the 0.1 (humidity in Nitrogen) with buttons and . Acknowledge with the button E.
The transmitter remains to wait for the process stabilization (10-15
min) and then stores the correction.
Insert the probe head into a measurement hole of the NaCl salt
chamber in the humidity calibrator. Acknowledge by pressing the button E. Alternatively you can select K2SO4 value with buttons and .
The transmitter remains to wait for the stabilization and then stores
the correction. The text CAL PASS is displayed after the calibration is performed.
NOTE
Remember to restore the dip switch settings.
Page 49
Calibration and adjustment____________________________________________________________
Vaisala _________________________________________________________________________ 49
FIGURE 9-3 Display chart of the automatic calibration procedure.
Calibration enabled RH calibration on
Restore switches
Page 50
Calibration and adjustment ____________________________________________________________
Vaisala _________________________________________________________________________ 50
Manual calibration
Low end adjustment
- Remove the filter and insert the probe head into a measurement hole of the dry end reference chamber (e.g. LiCl: 11 %RH) in the humidity calibrator.
NOTE
If you use serial commands, please refer to page 39.
Set the calibration dip switch 2 to ENABLED (up) position and
select RH calibration by using switches 3 and 4 (see Chapter 5: Dip switch functions).
- Press button C to ignore the automatic calibration procedure. The humidity reading measured by the transmitter appears on the display and on the lower left corner you can see the text SET LO. If the transmitter has no display, the LED indicator is blinking.
- According to the calibrator used, wait at least 10 - 15 minutes for the sensor to stabilize; use a multimeter or the transmitter display to monitor the stabilization.
- Press the buttons and to adjust the display reading or multimeter reading to correspond the reference value (when using a multimeter calculate the current value corresponding to the reference humidity by using the equations presented in page 46). Press the button E to conclude the low end adjustment.
- If the low end adjustment is sufficient, press the button E again to conclude the calibration and restore the dip switch settings; set the calibration dip switch to position DISABLED (down). To continue with the high end adjustment follow the instructions in next chapter.
High end adjustment (two-point calibration)
- After having made the low end adjustment, insert the probe head into a measurement hole of the NaCl (75 %RH) salt chamber in the humidity calibrator. Please, note that the difference between the two humidity references must be at least 30%RH.
NOTE
If you use serial commands, please refer to page 39.
- The readings measured by the transmitter appear on the display, as well as the text, SET HI on the lower left hand corner. If the transmitter has no display, the LED indicator lights up.
Page 51
Calibration and adjustment____________________________________________________________
Vaisala _________________________________________________________________________ 51
- Wait at least 10 - 15 minutes for the sensor to stabilize; use a multimeter or the transmitter display to monitor the stabilization.
- Press the buttons and to adjust the display or multimeter reading (when using a multimeter calculate the current value corresponding to the reference humidity by using the equations presented in page 47) and conclude the calibration with the button E.
- Restore the dip switch settings. Set the calibration dip switch to position DISABLED (down).
Temperature calibration in one point
Always use a high quality standard for calibrating the temperature.
NOTE
If you use serial commands, please refer to page 40.
- Set the calibration dip switch to position ENABLED (up).
- Set the calibration dip switches to position 'TEMPERATURE
CALIBRATION
' according to the table printed on the protective cover. The temperature value measured by the transmitter appears on the display as well as the text SET LO on the lower left corner. If the transmitter has no display, the LED indicator starts to blink.
- Let the sensor stabilize; use a multimeter or the transmitter display to monitor the stabilization.
- Press the buttons and to adjust the display or multimeter reading to correspond the reference value (when using a multimeter calculate the current value corresponding to the reference humidity by using the equations presented in page 46). Conclude the one point calibration by pressing the button E twice.
Page 52
Specifications ______________________________________________________________________
Vaisala _________________________________________________________________________ 52
CHAPTER 11
SPECIFICATIONS
Performance
Relative humidity
Measurement range 0 - 100 %RH Accuracy (including non-linearity and repeatability); when calibrated against: high quality, certified humidity standards ±1 %RH (0 - 90 %RH)
±2 %RH (90 - 100 %RH) salt solutions ±2 %RH (0 - 90 %RH) (ASTM E104-85) ±3 %RH (90 - 100 %RH)
Response time (90%) at 20 °C in still air (with sintered filter) 15 seconds Sensors: HUMICAP®180 for typical applications HUMICAP®L for applications with a demanding
chemical environment (max. +40 °C
in high humidities)
Moisture in natural gas (HMT362N/364N/368N)
Dewpoint measurement
Measurement range -50...+50 °C (-58...+122 °F) T
d
Accuracy ±2 °C (±3.6 °F) (see the graph on
the next page) Response time 63 % [90 %]
-20 -> +10 °C (-4 -> +50 °F) 6 s [32 s] +10 -> -20 °C (+50 -> -4 °F) 120 s [370 s] Sensor Vaisala HUMICAP® 180M
Page 53
Specifications ______________________________________________________________________
Vaisala _________________________________________________________________________ 53
Calculated variables available (typical ranges)
Water content 0...20 lb/mmscf, 0...326 mg/m
3
Parts per million by volume 0...500 ppm
v
Water activity in jet fuel applications(HMT368H/J)
Measurement range 0...1 aw (@-40...+180 °C (-
40...+356 °F)
Accuracy when calibrated against high-quality, certified humidity standards:
±0.01 (0...0.9) ±0.02 (0.9...1.0)
when calibrated against salt solutions (ASTM E104-85):
±0.02 (0...0.9)
±0.03 (0.9...1.0) Response time (90 %) at +20 °C in still oil (stainless steel filter) 10 min
Humidity sensor HUMICAP
®
Calculated variables available water content
relative humidity for saturation
saturation temperature Accuracy of the water content better than ±15 % of the reading
Temperature
Measurement range -40...+180 °C
(depends on selected probe) Typical accuracy of electronics at +20 °C ±0.1 °C Typical temperature dependence of electronics 0.005 °C/°C Sensor Pt 1000 RTD 1/3 Class B IEC 751
Page 54
Specifications ______________________________________________________________________
Vaisala _________________________________________________________________________ 54
Calculated variables (typical ranges)
With HMP361 probe
dewpoint temperature -40...+60 °C mixing ratio 0...160 g/kg d.a. absolute humidity 0...160 g/m
3
wet bulb temperature 0...+60 °C
With HMP363, HMP364, HMP365 & HMP368
dewpoint temperature -40...+100 °C mixing ratio 0...500 g/kg d.a absolute humidity 0...600 g/m
3
wet bulb temperature 0...+100 °C
Outputs
Two analog outputs Two wire 4...20 mA (one standard, one optional) Typical accuracy of analog Outputs at +20 °C ±0.05% full scale Typical temperature dependence of analog outputs 0.005% / °C full scale RS232C serial output for service use connector type RJ45 only in safe area Output connections via safety barriers
Classification with current outputs
EU (94/9/EC, ATEX100a) II 1 G EEx ia IIC T4
PTB 00 ATEX 2112 X
Safety factors: Ui=28 V, Ii=100 mA, Pi =0.7 W
Ci=1 nF, Li=0 H Environmental specifications: Tamb = -20...+60 °C
Pamb = 0.8...1.1 bar
Dust classification: II 1 D (IP65 T= 70 °C) (with protection cover) VTT 04 ATEX 023X
U.S.A. (FM) Classes I, II, III
Division 1: Groups A-G
Division 2: Groups A-D, F and G
FM Project ID:3010615 Safety factors: Vmax=28VDC,Imax=100mA,Ci=1nF, Li=0,
Pi=0.7 W, T
amb
= 60 ºC, T5
Page 55
Specifications ______________________________________________________________________
Vaisala _________________________________________________________________________ 55
Canada (CSA) Class 1, Divisions 1 and 2, Groups A,B,C,D
Class II, Divisions 1 and 2, Groups G and Coal Dust Class III CSA File No: 213862 0 000; CSA report 1300863 T
amb
= 60 ºC, T4
Japan (TIIS) Ex ia IIC T4
Code number: C15354
Safety factors Ui= 28 VDC, Ii=100 mA, Ci=1nF, Li=0,
Pi=0.7 W T
amb
= 60 ºC
Australia (TestSafe) Certificate No: Ex AUS Ex 3738X
Marking code: Ex ia IIC T5 IP65 AUS Ex 3738X
Safety factors V
max
= 28 VDC, Ii=100 mA, Pi= 0.7 W,
Ci=1nF, Li=0 mH
General
Operating voltage 12...28 V
with serial port (service mode) 15...28 V
Connections screw terminals, 0.33...2.0 mm2 wires
(AWG 14-22) Cable bushing Pg11 (5...12mm) Conduit fitting Pg11/NPT 1/2"-14 Operating temperature range for electronics -40...+60 °C
with display and/or with
Storage temperature range -40...+70 °C Housing material G-A1Si 10 Mg (DIN 1725) Housing classification IP 65 (NEMA 4) Housing dimensions 164 x 115 x 62 mm Housing weight 950g Fully electromagnetically compatible according to standards EN 61326-1:1997 +Am :1998,
Electrical equipment for measurement, control and laboratory use - EMC requirements; Industrial environment Note ! IEC 1000-4-5 complies only when using external EXi approved surge arrester on safe area.
Page 56
Specifications ______________________________________________________________________
Vaisala _________________________________________________________________________ 56
Probes
Probe cable length 2, 5, or 10 m Probe cable diameter 5.5 mm Sensor protection options stainless steel sintered filter
stainless steel filter PPS grid with steel netting filter PPS grid
HMP361 Probe - wall mount
Temperature range -40...+60 °C
HMP362 Probe - small pressure tight
Temperature range -40...+110 °C Pressure range 0...16.7 MPa (0...167 bar)
HMP363 Probe - confined spaces
Temperature range -40...+120 °C
HMP364 Probe - high pressures
Temperature range -40...+180 °C Pressure range 0...10 MPa (0...100 bar)
HMP365 Probe - high temperatures
Temperature range -40...+180 °C
HMP368 Probe - pressurized pipelines
Temperature range -40...+180 °C Pressure range 0...4 MPa (0...40 bar)
Options and accessories
Display two-line LCD
character size (1st line/2nd line) 12 mm/10 mm
Calculated output variables dewpoint temperature
mixing ratio absolute humidity
wet bulb temperature Additional analog output 4...20 mA Duct mounting installation kit (for HMP363 sensor head) Order code: HMP233FAH Installation flange (for HMP365 sensor head)
aluminum HMP235FA stainless steel HMP235 FS
Page 57
Accuracies of the calculated variables (not valid with HMT360 for Natural Gas) __________________
Vaisala _________________________________________________________________________ 57
Ball valve set (for HMP368 sensor head) DMP248BVS
pressure range at +20 °C 0...70 bar Sampling cell HMP302SC Serial interface cable for PC connectors RJ45 - D9 female 25905ZZ Rain shield HMT360SAR Protection cover (for use in the presence of combustible dust) 214101 HMK15 adapter fitting 211011 for 12 mm probes Galvanic isolator (EU) 212483 Barrier (USA&Canada) 210664
ACCURACIES OF THE CALCULATED VARIABLES (NOT VALID WITH HMT360 FOR
NATURAL GAS)
Accuracy of the calculated variables depend on the calibration accuracy of the humidity and temperature sensors; here the accuracy are given for ±2 %RH and ±0.2 °C.
Accuracy of dewpoint temperature °C
Relative humidity
Temp.102030405060708090100
-40 1.86 1.03 0.76 0.63 0.55 0.50 0.46 0.43
-20 2.18 1.19 0.88 0.72 0.62 0.56 0.51 0.48 — 0 2.51 1.37 1.00 0.81 0.70 0.63 0.57 0.53 0.50 0.48 20 2.87 1.56 1.13 0.92 0.79 0.70 0.64 0.59 0.55 0.53 40 3.24 1.76 1.27 1.03 0.88 0.78 0.71 0.65 0.61 0.58 60 3.60 1.96 1.42 1.14 0.97 0.86 0.78 0.72 0.67 0.64 80 4.01 2.18 1.58 1.27 1.08 0.95 0.86 0.79 0.74 0.70 100 4.42 2.41 1.74 1.40 1.19 1.05 0.95 0.87 0.81 0.76 120 4.86 2.66 1.92 1.54 1.31 1.16 1.04 0.96 0.89 0.84 140 5.31 2.91 2.10 1.69 1.44 1.26 1.14 1.05 0.97 0.91 160 5.80 3.18 2.30 1.85 1.57 1.38 1.24 1.14 1.06 0.99
Page 58
Accuracies of the calculated variables (not valid with HMT360 for Natural Gas)___________________
Vaisala _________________________________________________________________________ 58
Accuracy of mixing ratio g/kg
(ambient pressure 1013 mbar)
Relative humidity
Temp.102030405060708090100
-40 0.003 0.003 0.003 0.003 0.003 0.004 0.004 0.004
-20 0.017 0.018 0.019 0.021 0.022 0.023 0.025 0.026 — 0 0.08 0.09 0.09 0.10 0.10 0.11 0.11 0.12 0.13 0.13 20 0.31 0.33 0.35 0.37 0.39 0.41 0.43 0.45 0.47 0.49 40 0.97 1.03 1.10 1.17 1.24 1.31 1.38 1.46 1.54 1.62 60 2.68 2.91 3.16 3.43 3.72 4.04 4.38 4.75 5.15 5.58 80 6.73 7.73 8.92 10.34 12.05 14.14 16.71 19.92 24.01 29.29 100 16.26 21.34 28.89 40.75 60.86 98.85 183.66 438.56 — — 120 40.83 74.66 172.36 —
Accuracy of wet bulb temperature °C
Relative humidity
Temp.102030405060708090100
-40 0.20 0.20 0.20 0.20 0.20 0.20 0.20 0.20
-20 0.21 0.21 0.22 0.22 0.22 0.22 0.23 0.23 — 0 0.27 0.28 0.28 0.29 0.29 0.29 0.30 0.30 0.31 0.31 20 0.45 0.45 0.45 0.44 0.44 0.44 0.43 0.43 0.42 0.42 40 0.84 0.77 0.72 0.67 0.64 0.61 0.58 0.56 0.54 0.52 60 1.45 1.20 1.03 0.91 0.83 0.76 0.71 0.67 0.63 0.60 80 2.23 1.64 1.32 1.13 0.99 0.89 0.82 0.76 0.72 0.68 100 3.06 2.04 1.58 1.31 1.14 1.01 0.92 0.85 0.80 0.75 120 3.85 2.40 1.81 1.48 1.28 1.13 1.03 0.95 0.88 0.83 140 4.57 2.73 2.03 1.65 1.41 1.25 1.13 1.04 0.97 0.91 160 5.25 3.06 2.25 1.82 1.55 1.37 1.24 1.13 1.05 0.99
Accuracy of absolute humidity g/m³
Relative humidity
Temp.102030405060708090100
-40 0.004 0.004 0.005 0.005 0.005 0.006 0.006 0.006
-20 0.023 0.025 0.027 0.029 0.031 0.032 0.034 0.036 — 0 0.10 0.11 0.12 0.13 0.13 0.14 0.15 0.15 0.16 0.17 20 0.37 0.39 0.41 0.43 0.45 0.47 0.49 0.51 0.53 0.55 40 1.08 1.13 1.18 1.24 1.29 1.34 1.39 1.44 1.49 1.54 60 2.73 2.84 2.95 3.07 3.18 3.29 3.40 3.52 3.63 3.74 80 6.08 6.30 6.51 6.73 6.95 7.17 7.39 7.61 7.83 8.05 100 12.2 12.6 13.0 13.4 13.8 14.2 14.6 15.0 15.3 15.7 120 22.6 23.3 23.9 24.6 25.2 25.8 26.5 27.1 27.8 28.4 140 39.1 40.0 41.0 42.0 43.0 44.0 45.0 45.9 46.9 47.9 160 63.5 64.9 66.4 67.8 69.2 70.7 72.1 73.5 74.9 76.4
Page 59
_________________________________________________________________________________
Vaisala _________________________________________________________________________ 59
APPENDIX 1: DIMENSIONS
HMT361
HMP362
Flange intersection
Sample cell HMP302SC
ø36 (1.42)
ø32 (1.26)
41(1.61)
M6
ø12 (0.47)
1/8"-27 NPT
9(0.35)
7 (0.28)
71.5 (2.8 1)
92.5 (3.64)
81 (3.19)
71 (2.80)
58.7 (2.31)
1
6
(0
.
6
3
)
ø
2
4
.
5
4
5
°
4
4
5
°
M
6
ø12.5 (0.49)
1/8"-27 NPT
Page 60
_________________________________________________________________________________
Vaisala _________________________________________________________________________ 60
HMP363
HMP364
HMP365
HMP368
Page 61
_________________________________________________________________________________
Vaisala _________________________________________________________________________ 61
Mounting plate
Rain shield
180
80
15
Page 62
_________________________________________________________________________________
Vaisala _________________________________________________________________________ 62
Protection cover
208
224
160
184
ø
8
.
2
189
96
82
2
Page 63
_________________________________________________________________________________
Vaisala _________________________________________________________________________ 63
APPENDIX 2: WIRING FOR INTRINSICALLY SAFE OPERATION, FM
Use FM approved associated apparatus; zener barriers or galvanic isolators with entity concept parameters: V
oc
< 28 V
I
sc
<100 mA
C
a
> Ci + Ccable
L
a
> Li + Lcable
HMT360-series transmitters are approved for use in Classes I, II ans III, Division 1, groups A-G and Division 2, Groups A-D, F and G.
Safety factors for HMT360-series transmitters are:V
max
= 28 V, I
max
=100 mA,
C
i
= 1 nF, Li=0, Pi=0.7 W.
NOTE:
1. Barrier installation must be completed in accordance with ANSI/ISA RP 12.6 and the National Electrical Code.
2. Intrinsically safe barrier ground must be less than 1 ohm.
3. Maximum safe area voltage is 250 V.
Material Weight Finish
Part list
General tolerance
laces
laced
e
pted
dle
r
ator
cked
Scale
Title
Status
Code
Rev
01-09-04 ARH
1:1
CONTROL DRAWNING
DRW211603 A
Load Ch1
Vmax=28VDC
Vmax=28VDC
Load Ch2
Vaisala Oyj Vanha Nurmijärventie 21 Vantaa Finland
HAZARDOUS AREA SAFE AREA
ch1 ch2
Page 64
_________________________________________________________________________________
Vaisala _________________________________________________________________________ 64
APPENDIX 3: WIRING FOR INTRINSICALLY SAFE OPERATION
, CSA
R
eplaced
by
D
rawn
eview
ppr
esign
eplaces
Scale
id
Title
Arch
no
Serial
Sheet
doc no
Cooper ator`s
Dwg no
R
e
NOTE:
1. Each channel must be supplied through separate shielded cables.
2. When using galvanic separators CH1- and CH2­must be short circuited with an external wire.
3. When using transmitter in Class I, Division 2 the main switch shall not be operated or the unit shall not be disconnected unless power has been switched off, or area is known to be non hazardous.
4. Use only conduit connection in Division 2.
5. Substitution on components may impair intrinsic safety or suitability for Division 2.
6. Only intrinsically safe installation is allowed in Class II and Class III environments.
7. Intrinsically safe barrier ground must be less than 1 ohm.
8. Maximum safe area voltage is 250V.
p
HMT360-series transmitters are approved for use in Division 1 and 2, Class I, Groups A, B, C, and D. Division 1, and 2, Class II, Group G and coal dust. Division 1, and 2, Class III.
Vaisala Oyj Vanhanurmijärventie 21 Vantaa Finland
Installati on Drawing
DRW213478
HAZARDOUS AREA SAFE AREA
HMT360 connec tion board
-Ch1+ -Ch2+
Load Ch2
Vmax=28VD C
Vmax=28VD C
Load Ch1
KKe 2002-08-21
HMT360-series transmitters shall be used with following probes: Probe HMP361 with 116mm lenght pipe. Probe HMP363 with 2, 5 or 10 m lenght cable. Probe HMP364 with 2, 5 or 10 m lenght cable. Probe HMP365 with 2, 5 or 10 m lenght cable. Probe HMP368 with 2, 5 or 10 m lenght cable.
The material for associated cable is FEP (Tetrafluoropropylene).
To avoid static discharge shall the cable cover wi
t
conductive material.
HMT360-series transmitters shall be used with following zener barrier or galvanic separator types:
Manufacturer: Type: R. Stahl 9001/51-280-091-14 R. Stahl 9002/13-280-093-00 R. Stahl 9303/11-22-11 MTL MTL7206 MTL MTL5044
Cable parameters shall be less than 60pF/ft (197pF/m) and 0.2µH/ft (0.67µH/m). The maximum cable lenght is 2000ft (600m).
Max cable lenght 2000ft (600m)
A
Page 65
_________________________________________________________________________________
Vaisala _________________________________________________________________________ 65
CERTIFICATES
Page 66
_________________________________________________________________________________
Vaisala _________________________________________________________________________ 66
Page 67
_________________________________________________________________________________
Vaisala _________________________________________________________________________ 67
Loading...