www.vaisala.com
HMT330 Series Humidity and Temperature
Transmitters for demanding humidity measurement
The HMT330 transmitter family has the solution for demanding industrial humidity
measurements.
Features/Benefits
Six models for demanding
▪
industrial applications
Full 0 ... 100 % RH
▪
measurement, temperature
range up to +180 °C (+356 °F)
(depending on model)
Pressure tolerance up to
▪
100 bar (depending on model)
Next generation Vaisala
▪
HUMICAP® Sensor for excellent
accuracy and stability
Graphical display of
▪
measurement trends and over
four-year history
Multilingual user interface
▪
Excellent performance in harsh
▪
chemical concentrations
Corrosion resistant IP65
▪
housing
NIST traceable (certificate
▪
included)
Analog outputs, WLAN/LAN
▪
The Vaisala HUMICAP® Humidity
and Temperature Transmitter
Series HMT330 is designed for
demanding industrial applications
where stable measurement and wide
customization is important.
Vaisala HUMICAP®
performance
The HMT330 series incorporates
Vaisala’s 30 years of experience in
industrial humidity measurement.
An updated, even more stable and
chemical resistant HUMICAP
is available, next to the proven
reliable and accurate sensor.
®
sensor
Chemical purge minimizes
eects of contaminants
In environments with high
concentrations of chemicals and
cleaning agents, chemical purge
helps to maintain measurement
accuracy between calibration
intervals.
Chemical purge involves heating the
sensor to remove harmful chemicals.
The function can be initiated
manually or programmed to occur at
set intervals.
Graphical display of
history and measurement
trends
The HMT330 can be ordered with
a large numerical and graphical
display, which allows the user to
clearly monitor operational data,
measurement trends and 1-year
measurement history. The optional
data logger with real-time clock
makes it possible to generate over
four years of measured history, and
zoom in on any desired time or time
frame. The battery backup of the
clock guarantees a reliable logging of
measured data.
The display alarm allows tracking
of any measured parameter, with a
freely congurable low and high limit.
Data collection and
(wireless) transfer to PC
The recorded measurement data
can be viewed on the display or
transferred to a PC with Microsoft
Windows
can also be connected to a network
with an optional (W)LAN interface,
which enables a (wireless) Ethernet
connection. A USB-RJ45 cable makes
it easy to connect the service port of
the HMT330 to a PC.
®
software. The transmitter
Many ways to install and
versatile outputs
Mains and DC power options, and
several mounting accessories make
the instrument easy to install.
The HMT330 can have up to three
analog outputs. Isolated galvanic
power supply and analog outputs are
also available. For serial interface the
USB connection, RS232 and RS485
can be used.
Flexible calibration
The HMT330 instruments are
calibrated at six humidity points at
the factory. A quick, one-point eld
calibration can be performed with
the hand-held HM70 meter. A twopoint calibration can be performed
with the HMK15 salt bath calibrator
in a controlled environment. The
transmitter can be sent to Vaisala for
recalibration. Accredited calibrations
and maintenance contracts are also
available.
°C
°C
0
0-80 -60 -40 -20 20 40 60 80 100 120 140 160
180
-0.7
-0.6
-0.5
-0.4
-0.3
-0.2
-0.1
0.1
0.2
0.3
0.4
0.5
0.6
0.7
Technical data
Performance
RELATIVE HUMIDITY
Measurement range 0 ... 100 % RH
Accuracy (including non-linearity, hysteresis, and repeatability)
with Vaisala HUMICAP
with Vaisala HUMICAP
at +15 ... +25 °C (59 ... +77 °F) ±1 %RH (0 ... 90 %)
at -20 ... +40 °C (-4 ... +104 °F) ±(1.0 + 0.008 x reading) %RH
at -40 ... +180 °C (-40 ... +356 °F) ±(1.5 + 0.015 x reading) %RH
Factory calibration uncertainty (+20 °C) ± 0.6 % RH (0 ... 40 %RH)
Response time (90 %) at +20 °C (+68 °F) 8 s / 17 s* with grid filter
in still air 20 s / 50 s* with grid + steel netting filter
* with HUMICAP
TEMPERATURE
Accuracy at +20 °C (+68 °F) ± 0.2 °C (± 0.36 °F)
Accuracy over temperature range (measurement range
depends on model)
®
180 or 180R for typical applications
®
180C or 180RC for applications with
chemical purge/warmed probe
±1.7 %RH (90 ... 100 %RH)
± 1.0 % RH (40 ... 97 %RH)
(Defined as ±2 standard deviation limits. Small
variations possible, see also calibration certificate.)
40 s / 60 s* with sintered filter
®
180R or 180RC sensor
Inputs and outputs
Operating voltage 10 ... 35 VDC, 24 VAC
with optional power supply module 100 ... 240 VAC, 50/60 HZ
Power consumption at +20 °C (U
RS-232 max. 25 mA
2 x 0 ... 1 V/0 ... 5 V/0 ... 10 V max. 25 mA
U
out
2 x 0 ... 20 mA max. 60 mA
I
out
display and backlight + 20 mA
during chemical purge max. 110 mA
during probe heating (HMT337) + 120 mA
Analog outputs (2 standard, 3rd optional)
current output 0 ... 20 mA, 4 ... 20 mA
voltage output 0... 1 V, 0... 5 V, 0... 10 V
Accuracy of analog outputs at +20 °C ±0.05% full scale
Temperature dependence of the
analog outputs ±0.005%/°C full scale
External loads
current outputs R
0 ... 1 V output R
0 ... 5 V and 0 ... 10 V outputs R
Max. wire size 0.5 mm
Digital outputs RS-232, RS-485 (optional)
24 VDC)
in
L
L
stranded wires recommended
< 500 ohm
> 2 kohm
L
> 10 kohm
2
(AWG 20)
Service connection RS-232, USB
Relay outputs (optional) 0.5 A, 250 VAC
Ethernet interface (optional)
Supported standards 10/100Base-T
Connector RJ45
Protocols Telnet
Software support Vaisala MI70 link
WLAN interface (optional)
Supported standards 802.11b
Antenna connector type RP-SMA
Protocols Telnet
Security WEP 64/128,WPA
Software support Vaisala MI70 link
Authentication / Encryption (WLAN)
Temperature sensor PT 100 RTD 1/3 Class B IEC 751
Other variables available (depends on model)
dewpoint temperature, mixing ratio, absolute humidity, wet bulb
temperature, enthalpy, water vapor pressure
Open / no encryption
Open / WEP
WPA Pre shared key / TKIP
WPA Pre shared key / CCMP (a.k.a. WPA2)
Optional data logger with real-time clock
Logged parameters max. three with trend/min/max values
Logging interval 10 sec (fixed)
Max. logging period 4 years 5 months
Logged points 13.7 million points per parameter
Battery lifetime min. 5 years
Display LCD with backlight, graphic trend display
of any parameter
Menu languages English, Chinese, Japanese, Spanish, German,
French, Swedish, Russian, Finnish