The HMT330 transmitter family has the solution for demanding industrial humidity measurements.
Features/Benefits
Six models for demanding
industrial applications
Full 0 ... 100 % RH
measurement, temperature range up to +180 °C (+356 °F) (depending on model)
Pressure tolerance up to
100 bar (depending on model)
Next generation Vaisala
HUMICAP® Sensor for excellent accuracy and stability
Graphical display of
measurement trends and over four-year history
Multilingual user interface
Excellent performance in harsh
chemical concentrations
Corrosion resistant IP65
housing
NIST traceable (certificate
included)
Analog outputs, WLAN/LAN
The Vaisala HUMICAP® Humidity and Temperature Transmitter Series HMT330 is designed for demanding industrial applications where stable measurement and wide customization is important.
Vaisala HUMICAP® performance
The HMT330 series incorporates Vaisala’s 30 years of experience in industrial humidity measurement. An updated, even more stable and chemical resistant HUMICAP is available, next to the proven reliable and accurate sensor.
®
sensor
Chemical purge minimizes eects of contaminants
In environments with high concentrations of chemicals and cleaning agents, chemical purge helps to maintain measurement accuracy between calibration intervals.
Chemical purge involves heating the sensor to remove harmful chemicals. The function can be initiated manually or programmed to occur at set intervals.
Graphical display of history and measurement trends
The HMT330 can be ordered with a large numerical and graphical display, which allows the user to clearly monitor operational data, measurement trends and 1-year measurement history. The optional data logger with real-time clock makes it possible to generate over four years of measured history, and zoom in on any desired time or time frame. The battery backup of the clock guarantees a reliable logging of measured data.
The display alarm allows tracking of any measured parameter, with a freely congurable low and high limit.
Data collection and (wireless) transfer to PC
The recorded measurement data can be viewed on the display or transferred to a PC with Microsoft Windows can also be connected to a network with an optional (W)LAN interface, which enables a (wireless) Ethernet connection. A USB-RJ45 cable makes it easy to connect the service port of the HMT330 to a PC.
®
software. The transmitter
Many ways to install and versatile outputs
Mains and DC power options, and several mounting accessories make the instrument easy to install.
The HMT330 can have up to three
analog outputs. Isolated galvanic power supply and analog outputs are also available. For serial interface the USB connection, RS232 and RS485 can be used.
Flexible calibration
The HMT330 instruments are calibrated at six humidity points at the factory. A quick, one-point eld calibration can be performed with the hand-held HM70 meter. A two­point calibration can be performed with the HMK15 salt bath calibrator in a controlled environment. The transmitter can be sent to Vaisala for recalibration. Accredited calibrations and maintenance contracts are also available.
°C
°C
0
0-80 -60 -40 -20 20 40 60 80 100 120 140 160
180
-0.7
-0.6
-0.5
-0.4
-0.3
-0.2
-0.1
0.1
0.2
0.3
0.4
0.5
0.6
0.7
Technical data
Performance
RELATIVE HUMIDITY
Measurement range 0 ... 100 % RH Accuracy (including non-linearity, hysteresis, and repeatability) with Vaisala HUMICAP with Vaisala HUMICAP
at +15 ... +25 °C (59 ... +77 °F) ±1 %RH (0 ... 90 %)
at -20 ... +40 °C (-4 ... +104 °F) ±(1.0 + 0.008 x reading) %RH at -40 ... +180 °C (-40 ... +356 °F) ±(1.5 + 0.015 x reading) %RH
Factory calibration uncertainty (+20 °C) ± 0.6 % RH (0 ... 40 %RH)
Response time (90 %) at +20 °C (+68 °F) 8 s / 17 s* with grid filter in still air 20 s / 50 s* with grid + steel netting filter
* with HUMICAP
TEMPERATURE
Accuracy at +20 °C (+68 °F) ± 0.2 °C (± 0.36 °F) Accuracy over temperature range (measurement range depends on model)
®
180 or 180R for typical applications
®
180C or 180RC for applications with
chemical purge/warmed probe
±1.7 %RH (90 ... 100 %RH)
± 1.0 % RH (40 ... 97 %RH)
(Defined as ±2 standard deviation limits. Small
variations possible, see also calibration certificate.)
40 s / 60 s* with sintered filter
®
180R or 180RC sensor
Inputs and outputs
Operating voltage 10 ... 35 VDC, 24 VAC
with optional power supply module 100 ... 240 VAC, 50/60 HZ
Power consumption at +20 °C (U
RS-232 max. 25 mA
2 x 0 ... 1 V/0 ... 5 V/0 ... 10 V max. 25 mA
U
out
2 x 0 ... 20 mA max. 60 mA
I
out
display and backlight + 20 mA during chemical purge max. 110 mA during probe heating (HMT337) + 120 mA
Analog outputs (2 standard, 3rd optional)
current output 0 ... 20 mA, 4 ... 20 mA
voltage output 0... 1 V, 0... 5 V, 0... 10 V Accuracy of analog outputs at +20 °C ±0.05% full scale Temperature dependence of the analog outputs ±0.005%/°C full scale External loads
current outputs R
0 ... 1 V output R
0 ... 5 V and 0 ... 10 V outputs R Max. wire size 0.5 mm
Digital outputs RS-232, RS-485 (optional)
24 VDC)
in
L
L
stranded wires recommended
< 500 ohm
> 2 kohm
L
> 10 kohm
2
(AWG 20)
Service connection RS-232, USB Relay outputs (optional) 0.5 A, 250 VAC Ethernet interface (optional)
Supported standards 10/100Base-T
Connector RJ45
Protocols Telnet
Software support Vaisala MI70 link WLAN interface (optional)
Supported standards 802.11b
Antenna connector type RP-SMA
Protocols Telnet
Security WEP 64/128,WPA
Software support Vaisala MI70 link Authentication / Encryption (WLAN)
Temperature sensor PT 100 RTD 1/3 Class B IEC 751 Other variables available (depends on model)
dewpoint temperature, mixing ratio, absolute humidity, wet bulb temperature, enthalpy, water vapor pressure
Open / no encryption
Open / WEP
WPA Pre shared key / TKIP
WPA Pre shared key / CCMP (a.k.a. WPA2) Optional data logger with real-time clock
Logged parameters max. three with trend/min/max values
Logging interval 10 sec (fixed)
Max. logging period 4 years 5 months
Logged points 13.7 million points per parameter
Battery lifetime min. 5 years Display LCD with backlight, graphic trend display
of any parameter
Menu languages English, Chinese, Japanese, Spanish, German,
French, Swedish, Russian, Finnish