No part of this manual may be reproduced, published or publicly displayed in any form
or by any means, electronic or mechanical (including photocopying), nor may its
contents be modified, translated, adapted, sold or disclosed to a third party without prior
written permission of the copyright holder. Translated manuals and translated portions
of multilingual documents are based on the original English versions. In ambiguous
cases, the English versions are applicable, not the translations.
The contents of this manual are subject to change without prior notice.
This manual does not create any legally binding obligations for Vaisala towards
customers or end users. All legally binding obligations and agreements are included
exclusively in the applicable supply contract or the General Conditions of Sale and
General Conditions of Service of Vaisala.
September 2007 - Added HUMICAP® 180R and
HUMICAP® 180RC as sensor options.
Manual Code
Manual Name
M210297EN
Vaisala HUMICAP® Hand-Held Humidity and
Temperature Meter HM70 User’s Guide
M210185EN
Vaisala Humidity calibrator HMK15 User’s Guide
Version Information
Table 1 Manual Revisions
functionality applicable from version 2.0.1 onward.
Voltage output added to analog output options, new
voltage parameters in the AMODE command.
Multipoint calibration instructions and MPC
command added. Added information on filter and
sensor options, rain shield, MI70 hand-held
indicator, and USB connection. Transmitter plastic
housing material changed to PPS. Updated serial
terminal instructions, list of optional accessories
and operating voltage and power consumption
specifications. Calculated variables and fuel cell
heating information added. Document template
HUMICAP® 180C and HUMICAP® 180L2 from
Related Manuals
Table 2 Related Manuals
Documentation Conventions
Throughout the manual, important safety considerations are highlighted
as follows:
WARNING
CAUTION
Warning alerts you to a serious hazard. If you do not read and follow
instructions very carefully at this point, there is a risk of injury or even
death.
Caution warns you of a potential hazard. If you do not read and follow
instructions carefully at this point, the product could be damaged or
important data could be lost.
Note highlights important information on using the product.
Chapter 1 _________________________________________________________ General Information
Ground the product and verify outdoor installation grounding
periodically to minimize shock hazard.
Do not modify the unit. Improper modification can damage the product
or lead
Do not touch the sensor plate.
In pressurized processes it is essential to tighten the supporting nuts and
screws very carefully to prevent loosening of the probe by the action of
pressure.
Take care not to da
probe less tight and may prevent it from going through the clasp nut.
Safety
The Humidity and Temperature Transmitter HMT310 delivered to you
has been tested for safety and approved as shipped from the factory. Note
the following precautions:
WARNING
CAUTION
to malfunction.
CAUTION
CAUTION
CAUTION
mage the probe body. A damaged body makes the
ESD Protection
Electrostatic Discharge (ESD) can cause immediate or latent damage to
electronic circuits. Vaisala products are adequately protected against
ESD for their intended use. It is possible to damage the product,
however, by delivering electrostatic discharges when touching,
removing, or inserting any objects inside the equipment housing.
To make sure you are not delivering high static voltages yourself:
- Handle ESD sensitive components on a properly grounded and
protected ESD workbench.
- When an ESD workbench is not available, ground yourself to the
equipment chassis with a wrist strap and a resistive connection cord.
- If you are unable to take either of the above precautions, touch a
conductive part of the equipment chassis with your other hand before
touching ESD sensitive components.
- Always hold component boards by the edges and avoid touching the
component contacts.
Chapter 1 _________________________________________________________ General Information
Trademarks
HUMICAP® is a registered trademark of Vaisala.
Windows® is a registered trademark of Microsoft Corporation in the
United States and/or other countries.
Software License
This product contains software developed by Vaisala. Use of the software
is governed by license terms and conditions included in the applicable
supply contract or, in the absence of separate license terms and
conditions, by the General License Conditions of Vaisala Group.
Warranty
Visit our Internet pages for standard warranty terms and conditions:
www.vaisala.com/warranty.
Please observe that any such warranty may not be valid in case of
damage due to normal wear and tear, exceptional operating conditions,
negligent handling or installation, or unauthorized modifications. Please
see the applicable supply contract or Conditions of Sale for details of the
warranty for each product.
This chapter introduces the features, advantages, and the product
nomenclature.
Introduction to HMT310
The Vaisala HUMICAP® Humidity and Temperature Transmitter
HMT310 is a small size humidity and temperature transmitter that
powers up with 10 ... 35 VDC (mode-dependent power supply
requirements). HMT310 uses Vaisala’s HUMICAP® capacitive thin-film
polymer sensor for humidity and temperature measurements: for more
information on HMT310 sensors, see section Sensor Options on page 19.
HMT310 supports the following output alternatives:
- Analog outputs 0/4 ... 20 mA and 0 … 5/10 V (1 … 5 V available as
scaled output: see section Setting, Scaling and Testing Analog Outputs
on page 54).
- RS-232 serial line
- USB cable (optional accessory, order code 238607)
There are several filter types for HMT310. All filters are 12 mm in
diameter with a female thread, and are compatible with all HMT310
probe models. The filters recommended for a probe type can be selected
on the corresponding HMT310 order form.
For a list of the available filters and ordering information, see section
Spare Parts and Accessories on page 93.
Sensor Options
HMT310 uses Vaisala’s HUMICAP® capacitive thin-film polymer sensor
for humidity and temperature measurements. The HUMICAP® sensor
options (selected when ordering according to transmitter type and
application, also available as accessories) include:
- HUMICAP®180R: general purpose humidity and temperature sensor
- HUMICAP® 180RC: composite sensor for chemical purge and probe
heating
- HUMICAP®180VHP: catalytic sensor designed for environments
with vaporized hydrogen peroxide (H2O2)
- HUMICAP®180VHPC: catalytic sensor with chemical purge for
H2O2 environments
HMT310 also uses the Pt 100 (Pt 100 RTD Class F0.1 IEC 60751)
temperature sensor.
The Vaisala HUMICAP® sensor is compatible with direct H2O2
exposure even at high concentrations and hundreds of H2O2 cycles up to
saturation. The long term performance is very good at concentrations
higher than typically used for sterilization. However, for applications
where full saturation is possible, Vaisala recommends using either the
HUMICAP®180VHP/180VHPC catalytic sensor, or the catalytic filter
(order code 231865) for HUMICAP®180R/180RC sensors to protect the
sensor and extend the calibration interval.
In pressurized processes it is essential to tighten the supporting nuts and
screws very carefully to
pressure.
CHAPTER 3
INSTALLATION
This chapter provides you with information that is intended to help you
install the HMT310.
Measuring at Overpressure
HMT314 and HMT318 are designed for humidity measurement at
overpressure. The maximum measurement pressures depend on the probe
as follows:
CAUTION
- HMT314: 0 ... 100 bar (10 MPa), for pressurized rooms and processes,
probe is provided with a nut, fitting screw and sealing washer
- HMT318: 0 ... 40 bar (4 MPa), for pressurized pipelines, ball valve set
available
The actual pressure in the process or sampling cell needs to be set to the
transmitter by using the serial line command PRES (Set Ambient
Pressure for Calculations).
prevent loosening of the probe by the action of
Pressure Regulator Recommended
When sampling pressurized processes exceeding the maximum
measurement pressure of the probe, the pressure in the measurement
chamber must be regulated to acceptable level or below. It is
recommended to use a pressure regulator before the measurement
chamber to prevent significant pressure variations.
Selecting Location
Finding a suitable site for the Humidity and Temperature Transmitter
HMT310 is important for getting representative ambient measurements.
The site should represent the general area of interest.
The following letters and numbers refer to Figure 3 above:
A = Mounting with larger mounting plate (mount from flanges)
B = Mounting with smaller mounting plate (remove the transmitter
and mount using the holes in mounting plate base)
1 = Two Allen screws for fastening or removing the transmitter
module (Allen key provided)
2 = Four screw holes (Ø 4.5 mm) for wall mounting (screws not
provided)
3 = Two screw holes (Ø 6.0 mm) on the base of the plate for wall
mounting (screws not provided)
Mounting the Transmitter/Removing the
Transmitter Unit
Select a place with stable conditions for mounting the transmitter. Do not
expose the transmitter to direct sunlight or rain. Always mount the
transmitter housing with the cable bushings pointing downwards.
NOTE
If the transmitter is mounted outdoors cover it with a shelter (purchased
by customer). A rain shield designed for HMT310 is available as an
optional accessory. For information on ordering accessories, see section
Spare Parts and Accessories on page 93.
1. Mount the plate onto the wall with four/two screws
(Ø 4.5 mm/6.0 mm).
2. Place the transmitter onto the mounting plate and fasten it with two
Allen screws.
The transmitter module can be unfastened for calibration by releasing the
two Allen screws on the left side.
1 = Transmitter’s larger mounting plate: attaches to rain shield from
flanges with four Ø 4.5 mm screws (not provided).
2 = Rain shield: Four Ø 8.5 mm holes for (U-bolt) mounting. The
maximum diameter of the pole in U-bolt mounting is 60 mm.
3 = Rain shield: Four Ø 4.5 mm screw holes for mounting.
Mounting the Transmitter with Optional
Rain Shield
If you mount HMT310 outdoors, use a shelter to protect the transmitter.
An HMT310 rain shield is available from Vaisala as an optional
accessory (order code ASM211103).
NOTE
Vaisala does not recommend installing the transmitter outdoors with the
smaller mounting plate (without flanges). Use the larger mounting plate
with flanges for outdoor installations.
Mounting the Rain Shield with Larger
Mounting Plate
1311-251
Figure 5 Rain Shield with Larger Mounting Plate
Figure 5 above:
The transmitter’s larger mounting plate (see Figure 3 on page 22) is
attached to the rain shield from the mounting plate’s flanges using four
Ø 4.5 mm screws. The rain shield has Ø 4.5 mm and Ø 8.5 mm holes for
mounting with screws or U-bolts.
For rain shield measurements, see section Rain Shield Dimensions on
page 95.
Do not unsolder and then again resolder the probe cable from the mother
board during installation; this may alter the humidity calibration of the
transmitter.
Beware of Temperature Differences
In humidity measurement and especially in calibration it is essential that
the temperature of the probe and the measuring environment is the same.
Even a small difference in temperature between the environment and the
probe causes an error. As the curve in Figure 6 below shows, if the
temperature is +20 °C and the relative humidity 100 %RH, a difference
of ±1 °C between the environment and the probe causes an error of
±6 %RH.
Figure 6 Measurement Error at 100 %RH when Difference
The following numbers refer to Figure 7 above:
1 = To be sealed
2 = To be insulated
3 = Insulate
4 = Let the cable hang loosely. This prevents condensed water
running to the sensor along the cable.
General Instructions for Probes with Cable
It is recommended that the probes with a cable are mounted with the
probe horizontal; this way, any water condensing on the tube cannot
flow onto the sensor.
0507-024
Figure 7 Horizontal Mounting of Probe
When there is no alternative but to install the probe in the process
vertically, the point of entry must be carefully insulated. The cable must
also be allowed to hang loosely as this prevents any condensed water
from running onto the probe along the cable.
The following numbers refer to Figure 8 above:
1 = To be sealed
2 = Insulate the cable
3 = To be insulated
4 = Let the cable hang loosely. This prevents condensed water
running to the sensor along the cable.
D
condensation problems caused by heat conduction along the metal.
0507-022
Figure 8 Vertical Mounting of Probe
CAUTION
o not attach a heated probe (HMT317) to metal structures to avoid
If the process temperature is much higher than that of the environment,
the whole probe and preferably a piece of the cable must be inside the
process. This prevents measuring inaccuracies caused by heat conduction
along the cable.
When mounted on the side of a duct or channel, the probe must be
inserted from the side of the duct. If this is not possible and the probe
must be inserted from the top, the point of entry must be carefully
insulated.
HMT313 for General Use
The HMT313 is a small size (d = 12mm) general-purpose probe suitable
for ducts and channels with the installation kit available from Vaisala.
The HMT313 has two probe versions for different measurement
environments:
- Probe with a flexible cable, suitable for environments up to 80 ºC
The following numbers refer to Figure 9 above:
1 = Tightening cone
2 = Nut
3 = Fitting screw
4 = Sealing washer
5 = Probe; Ø 12 mm
6 = M22×1.5 or NPT 1/2"
See Appendix A on page 99 for the following probe installation kits for
HMT313 and installation examples:
- Duct mounting kit
- Cable gland
HMT314 for Pressurized Spaces
HMT314 probe is for humidity measurements in pressurized rooms and
industrial processes. The probe is provided with a nut, a fitting screw and
a sealing washer. Keep the fitting screw and nut in place on the body of
the probe during handling to prevent damage to the highly polished
surface of the probe. Follow the instructions below to achieve a leak-tight
assembly:
1. Remove the fitting screw from the nut and the probe.
2. Attach the fitting screw to the chamber wall with a sealing washer.
Tighten the fitting screw into the threaded sleeve with a torque
wrench. The tightening torque is 150 ± 10 Nm (110 ± 7 ft-lbs).
3. Insert the body of the probe into the fitting screw and attach the nut
manually to the fitting screw so that the connection feels tight.
In pressurized processes it is essential to tighten the supporting nuts and
screws very carefully to prevent loosening of the probe by the action of
pressure.
When HMT314 is installed in a process with pressure differing from
normal atmospheric pressure, enter the pressure value of the process (in
hPa or mbar) into the transmitter memory via the serial line (see
Ambient Pressure for Calculations
4. Mark both the fitting screw and the nut hex.
5. Tighten the nut a further 30° (1/12 turn) or if you have a torque
wrench tighten it with a torque of 80 ± 10 Nm (60 ± 7 ft-lbs).
0505-273
Figure 10 Marking the Nut
NOTE
CAUTION
NOTE
-tightening the nut after detachment the nut must be tightened
6. Clean and grease the tightening cone of the fitting screw after every
tenth detachment. Change the sealing washer every time the fitting
screw is detached. Use high-vacuum grease (for example; Down
Corning, Europe) or a similar grease.
See also section Measuring at Overpressure, on page 21.
Set
on page 51).
HMT315 for High Temperatures
HMT315 is installed similarly than the HMT313 probe but without the
supporting bar. Refer to Appendix A on page 99 for more information on
the duct installation kit for HMT315.
To avoid incorrect humidity readings, the temperature differences
between inside and outside of the duct must not be remarkable.
The following numbers refer to Figure 11 above:
1 = Clasp nut; 24 mm hex nut
2 = Fitting body; 27 mm hex head
HMT317 for Demanding Processes
The HMT317 is for environments where relative humidity is very high,
near saturation. The warmed probe prevents the saturation of the sensor.
See Appendix A on page 99 for a presentation of the following probe
installation kits for HMT317 with installation examples:
- Duct mounting kit
- Cable gland
- Pressure tight Swagelok connector
HMT318 for Pressurized Pipelines
Due to the sliding fit the HMT318 is easy to install into and remove from
the pressurized process. The probe is especially suitable for the
measurements in pipelines. See section Ball Valve Installation kit for
HMT318 on page 102.
The following numbers refer to Figure 13 above:
1 = Probe
2 = Clasp nut
3 = Fitting screw
4 = Pen (for marking)
0505-276
Figure 13 Tightening the Clasp Nut
NOTE
CAUTION
CAUTION
NOTE
Take care not to overtighten the clasp nut to avoid difficulties when
opening it.
Take care not to damage the probe body. A damaged body makes the
probe less tight and may prevent it from going through the clasp nut.
In pressurized processes it is essential to tighten the supporting nuts and
screws very carefully to prevent loosening of the probe by the action of
pressure.
When HMT318 is installed in a process with pressure differing from
normal atmospheric pressure, enter the pressure value of the process (in
hPa or mbar) into the transmitter memory via the serial line (see section
Set Ambient Pressure for Calculations on page 51).
When the HMT310 leaves the factory, the measurement ranges, output
scaling and quantities have already been set according to the customer
order. The unit is calibrated at the factory and ready for use.
The transmitter is delivered with either a screw terminal connector or
with a detachable 5m cable with eight wires for serial port, analog
outputs and 10 … 35 VDC power supply (requirements vary
dependending on the operating environment, functions and output type).
The wiring is described in this section. For power supply requirements,
see section Power Supply Requirements on page 34.
Cable Wiring
Figure 14 8-Pin Connector (Left) and Screw Terminal
- Pressures above 10 bara (145 psia): minimum 24 VDC
Table 5 below lists the HMT310 current consumption for different output
types and functions.
Table 5 Current Consumption by Output Type/Function
Channel1 & Channel 2
Channel 1 & Channel 2
NOTE
HMT310 can also be powered by an MI70 indicator or a USB cable
connection to a computer. For information on ordering the optional MI70
and USB connection cables, see section Spare Parts and Accessories on
page 93.
NOTE
When the USB connection is used to power the transmitter, the heating
and purge functions may not work as intended due to the limited USB
powering capability. This depends on the computer used and
environmental conditions. Using the USB connection as a permanent
power supply is not recommended.
This chapter contains information that is needed to operate the HMT310.
Power Supply
Use a 10 … 35 VDC power supply (required minimum voltage varies
depending on the operating environment, functions and output type). For
instructions on selecting the correct operating voltage, see section Power
Supply Requirements on page 34.
Transmitter Start-Up
When the transmitter is connected to a power supply, there is a delay of
three seconds as the transmitter starts up. After the start-up is complete,
the measurement data is available from the analog output and serial line.
If serial line output is used, note that the transmitter will behave
according to the configured serial mode:
- In STOP mode, the transmitter outputs the transmitter model and
software version. This is the default mode.
- In RUN mode a measurement output starts immediately.
- In POLL mode the transmitter does not output anything after
powerup.
For instructions on configuring the serial mode, see section Set Serial
Interface Mode on page 43.
HMT310 measurement data can be accessed with the following serial or
analog options:
- RS-232 serial port connection to a computer
- Analog output (current or voltage)
For temporary access, you can also connect to the transmitter with the
following:
- USB connection to a computer using an optional USB cable (order
code 238607).
- Connection with a handheld MI70 indicator using an optional MI70
calibration cable (DRW216050SP)
When using HMT310 with a serial connection (RS-232 serial cable or
USB), you need a terminal program to operate the HMT310 command
line interface. For instructions on installing and using the PuTTY
terminal application for Windows (available from www.vaisala.com), see
section Terminal Program Settings on page 39.
For information on using the analog outputs, see section Setting, Scaling
and Testing Analog Outputs on page 54.
RS-232 Serial Port Connection
When connecting the HMT310 to a computer with an RS-232 cable, use
the following data format settings (factory settings) in the terminal
program:
- 1 start bit
- 7 data bits
- 1 stop bit
- Even parity
- 4800 bits per second, programmable to 19200
- Full duplex
- Serial asynchronous
- Configured as Data Terminal Equipment (DTE)
Enter the communication parameters when using this terminal session for
the first time; save them for future use.
Before starting to use the USB cable, you must
driver
When the
purge functions may not work as intended due to the limited USB
powering capability. This depends on the computer used and
environmental
power supply is not recommended.
USB Connection
When performing temporary checks or adjustments, HMT310 can be
connected to a computer by using a USB cable (238607). The USB cable
is an optional accessory: for more information on ordering optional
items, see section Spare Parts and Accessories on page 93.
NOTE
NOTE
install the USB cable
(requires Windows).
transmitter is powered by the USB connection, the heating and
conditions. Using the USB connection as a permanent
Installing the Driver for the USB Cable
1. Check that the USB cable is not connected. Disconnect the cable if
you have already connected it.
2. Insert the media that came with the cable, or download the driver
from www.vaisala.com.
3. Execute the USB driver installation program (setup.exe), and
accept the installation defaults. The installation of the driver may
take several minutes.
4. After the driver has been installed, connect the USB cable to a USB
port on your computer. Windows will detect the new device, and
use the driver automatically.
5. The installation has reserved a COM port for the cable. Verify the
port number, and the status of the cable, using the Vaisala USB
Instrument Finder program that has been installed in the Windows
Start menu.
Windows will recognize each individual cable as a different device, and
reserve a new COM port. Remember to use the correct port in the
settings of your terminal program.
HMT310 can be connected to the hand-held MI70 indicator by using the
MI70 connection cable (DRW216050SP). The MI70 connection cable is
an optional accessory: for more information on ordering optional items,
see section Spare Parts and Accessories on page 93.
When HMT310 is connected to the MI70 indicator, you can view
HMT310 readings and perform HMT310 calibration and adjustment
using the MI70 indicator’s interface.
For instructions on performing HMT310 calibration and adjustment with
the MI70 indicator, see section Adjustment Using an MI70 Indicator on
page 81.
For instructions on using the MI70 handheld indicator to view and log
readings and further information on using MI70, see the MI70 user
documentation.
The instructions below describe how to connect to the HMT310 using the
PuTTY terminal application for Windows (available for download at
www.vaisala.com). Perform the necessary cabling and configuration of
the transmitter before following the instructions.
Opening a Serial/USB Connection with PuTTY
1. Power up the HMT310 and start the PuTTY application.
2. Select the Serial & USB settings category, and check that the
correct COM port is selected in the Serial or USB line to connect to
field. Change the port if necessary.
If you are using a Vaisala USB cable, you can check the port that it
uses by clicking the USB Finder... button. This opens the Vaisala
USB Instrument Finder program that has been installed along with
the USB drivers.
3. Check that the other serial/USB line settings are correct for your
connection, and change if necessary.
1311-113
Figure 15 PuTTY Serial Line Configuration
4.Click the Open button to open the connection window and start
using the serial connection command line interface.
If PuTTY is unable to open the serial port you selected, an error
message is displayed. If this happens, restart PuTTY and check the
settings.
Serial line settings (default: 4800 E 7 1)
baud: 300…19200
ADDR [0 … 99]
Set transmitter address for use in POLL
mode
OPEN [0 … 99]
Temporarily open transmitter from POLL
mode to receive commands
CLOSE
Set transmitter in POLL mode
Command
Description
FORM
Set serial output format
TIME
Set time
DATE
Set date
FTIME [ON/OFF]
Add time to SEND and R outputs
FDATE [ON/OFF]
Add date to SEND and R outputs
UNIT
Select metric or nonmetric output units
FST [ON/OFF]
Output state of optional probe heating and
commands)
List of Serial Commands
All commands can be issued either in uppercase or lowercase. In the
command examples, the keyboard input by the user is in bold type.
The notation <cr> refers to pressing the carriage return (Enter) key on
your computer keyboard. Enter a <cr> to clear the command buffer
before starting to enter commands.
In the tables below, the bold text in the brackets is the default setting.
In STOP mode: the command OPEN has no effect, CLOSE sets the
transmitter temporarily in POLL mode.
In POLL mode: the command OPEN sets the transmitter temporarily in
STOP mode, CLOSE returns the instrument to POLL mode.
Example:
Relative humidity calibration is performed at transmitter 2, which is in
POLL mode.
>open 2
Opens the line to transmitter 2.
>crh
Calibration started.
...
>close
Line closed.
Output Formatting
Set Serial Output Format
Syntax: FORM x<cr>
The FORM command can be used to change the format of the output
commands SEND and R.
The format string entered with the FORM command consists of
quantities and modifiers: use the quantity abbreviations and modifiers
presented in Table 13 and in Table 14 when selecting the output
quantities.
Use the command ? to check the current transmitter configuration. The
command ?? is similar but can also be used if the transmitter is in POLL
state.
Example (factory default settings):
>?
HMT310 / 1.07
PRB serial nr : A0000000
Calibration : 2004-05-07
Cal. info : Vaisala/HEL
Output units : metric
Pressure : 1013.25 hPa
RS232 settings
Address : 0
Output interval: 0 S
Baud P D S : 4800 E 7 1
Serial mode : STOP
Analog outputs
Ch1 output mode: 0 ...20mA
Ch2 output mode: 0 ...20mA
Ch1 error out : 0.000mA
Ch2 error out : 0.000mA
Ch1 RH lo : 0.00 %RH
Ch1 RH hi : 100.00 %RH
Ch2 T lo : -40.00 'C
Ch2 T hi : 60.00 'C
>
The XPRES command should be used if the value is changed frequently.
Its value is not retained at reset, and when set to 0, the value set with the
PRES command is used.
Example:
If the gauge pressure is 1.4 bar, set the pressure value to 2.4 bar (= 2400
hPa)
x = ON, OFF or EXT
ON = Short
show
OFF = No filtering (default)
EXT = Extended filter of about 1 min (
will show the average value of the last 1 min measurement data)
This command resets the device, resulting in a short start-up delay. The
serial interface mode (STOP, RUN or POLL) selected with the SMODE
command is taken into use at restart.
Display Error Messages
Syntax: ERRS<cr>
Display transmitter error messages. If there are no errors present,
a PASS will be returned.
Examples:
>errs
PASS
>
>errs
FAIL
Error: Temperature measurement malfunction
Error: Humidity sensor open circuit
>
1 … 5 V output can be selected when ordering the device. The 1 ... 5 V
output is based on the 0 … 5 V output mode, but has been adjusted to
1 … 5 V by scaling the valid range to 1 … 5. In the scaled 1 … 5 V
output mode, 0 V is reserved for error output (AERR set to 0).
Select Parameter for Analog Outputs
Syntax: ASEL xxx yyy<cr>
After entering the command, the low and high values of each channel are
displayed one row at a time and a new value can be entered. Enter a new
value and press <cr> to move to the next row, or press <cr> without
entering a value to keep the current setting.
Use the abbreviations shown in Table 17 on page 55.
x = Output quantity of analog channel (use
abbreviations, see
on page
45)
yyy.yyy
=
Value
Test Analog Outputs
Syntax: ITEST aa.aaa bb.bbb<cr>
The operation of the analog outputs is tested by forcing the outputs to
given values. The values in the analog outputs can then be measured with
a current/voltage meter.
Example:
>itest 8.30 6.40
8.30000mA 2c30 6.40000mA 2278
>
The ITEST command output includes internal debugging values shown
in hexadecimal format (2c30 and 2278 in the example above).
The set values remain valid until you issue the ITEST command without
readings or reset the transmitter.
Test Analog Outputs for Desired
Readings
Syntax: AQTEST x yyy.yyy<cr>
Using the AQTEST command, you can test the analog output by forcing
the output to correspond to a chosen value.
The error output value is
faults such as a humidity sensor open circuit. When there is a
device malfunction, such as an
microprocessor ROM/RAM failure, the error output value is not
necessarily
The c
Set Error Outputs
Syntax: AERR<cr>
The factory default state for analog outputs during error condition is
0 mA. Be careful when selecting the new error value, the error state of
the transmitter should not cause problems in process monitoring.
Example:
>aerr
Ch1 error out : 0.000mA ?
Ch2 error out : 0.000mA ?
NOTE
NOTE
utput value needs to be within the valid range of the output type.
shown only when there are minor electrical
analog output electronics failure or a
shown.
Chemical Purge (Optional)
General
In some specific applications the sensor gain may decrease gradually due
to an interference caused by some particular chemical present in the
ambient (see Figure 16 on page 58). The sensor polymer absorbs the
interfering chemical; this reduces the ability of the polymer to absorb
water molecules and so decreases the sensor gain. In chemical purge, the
humidity sensor is heated to a temperature level of approximately
+180 °C for several minutes, which evaporates the interfering chemical.
severe
The purge function starts with a heating stage, continues with a settling
stage, and when the temperature of the sensor is decreased, the
transmitter returns to normal mode. The whole cycle takes about 8
minutes.
Using purge is not recommended when the transmitter is powered with
USB. For information on HMT310 operating voltages, see section Power
Supply Requirements on page 34.
0507-056
Figure 16 Decrease of Sensor Gain Due to Interfering Chemical
and Effect of Chemical Purge Process
Automatic/Manual Chemical Purge
When HMT310 leaves the factory the automatic chemical purge (if
selected) takes place repeatedly with the time intervals set in the factory.
You can change the interval in which the purge takes place by using
serial commands. This may be necessary if the measuring environment
contains high concentrations of interfering chemicals. The automatic
chemical purge can be also turned off if necessary.
NOTE
It is not recommended to alter any other purge settings than the interval.
This command enables or disables automatic chemical purge.
Syntax: PUR x<cr>
Example:
>pur off
Chemical Purge : OFF
>pur on
Chemical Purge : ON
>
Set Chemical Purge Interval
If the sensor is exposed to chemicals it is recommended to have the
chemical purge done at least once in 720 min (= 12 hours). In
applications where the chemical exposure is not likely, the interval may
be longer.
In normal use, it is not recommended to change other parameters than the
purge interval. If changing any other parameters is required, always take
note of the current settings before making changes so you are able to
restore the configuration.
Syntax: PUR<cr>
Issue the PUR command and type the new interval (in minutes). Press
Enter without entering any values to keep the other parameters in their
current configuration and finish the command. The maximum interval is
14400 minutes (= 10 days).
Example:
>pur
Chemical Purge : ON <cr>
Interval min : 1440 ? 720
Duration s : 120 ? <cr>
Settling s : 240 ? <cr>
Power : 160 ? <cr>
Temp diff : 0.50 ? <cr>
>
To activate the new interval setting immediately, reset the transmitter.
Chemical Purge in Powerup
This command enables or disables the function which turns on chemical
purge within 10 seconds from the powerup of the device.
Syntax: PURR x<cr>
NOTE
When you enable this function, wait about 8 min after powerup before
taking measurements. The powerup chemical purge locks the output
values for the first operation minutes.
Chemical Purge Activated Manually
Chemical purge needs to be performed always before calibration (see
Chapter Calibration and Adjustment on page 69) or whenever there is
reason to believe that the sensor has been exposed to an interfering
chemical. Make sure that the temperature of the sensor has come down to
normal before starting calibration.
Before starting chemical purge note the following:
- The sensor must be protected with a PPS grid with stainless steel
netting, a stainless steel sintered filter or with a membrane SST filter.
- The sensor temperature needs to be below 100 °C. At higher
temperatures the chemicals evaporate spontaneously from the sensor
and the chemical purge is not necessary.
During the sensor heating the outputs are locked to the values measured
before the heating cycle.
Using sensor heating is not recommended when the transmitter is
powered with USB. For information on HMT310 operating voltages, see
section
The prompt ">" appears when the heating period is over. However, the
transmitter outputs are locked to values measured before chemical purge
until the settling time is over.
Sensor Heating (Optional)
General
NOTE
This function is optionally available only in transmitters with
HUMICAP
probe.
The sensor heating is recommended for the high humidity environments
where even a small temperature differences can cause water to condense
on the sensor. The sensor heating speeds up the recovery of the humidity
sensor from condensation.
Sensor heating starts up when the relative humidity of the measuring
environment reaches the RH value set by the user (RH limit). The user
can define the RH sensor heating temperature as well as the duration of
the heating.
After the heating cycle the humidity conditions are checked and new
sensor heating is performed if the predefined conditions are reached
again.
®
180RC sensor. It should be used only with the warmed
NOTE
Power Supply Requirements on page 34.
HMT317 Fuel Cell Model Heating Settings
In the HMT317 fuel cell model, the probe heating temperature has been
set to the maximum in the factory configuration in order to minimize
condensation.
Sensor heating (XHEAT command) has also been enabled in the factory
configuration for the fuel cell option. See section Setting Heating
Parameters on page 62 for instructions on configuring the sensor heating
settings.
When HMT310 leaves the factory the sensor heating follows the factory
default values. You can enable or disable the function, change the RH
limit and define the heating temperature and duration of the heating
period.
Sensor Heating ON/OFF
Enables or disables the sensor heating.
Syntax: XHEAT x<cr>
* In the HMT317 fuel cell model, XHEAT is enabled in the factory
settings.
Example:
>xheat on
Extra heat : ON
>xheat off
Extra heat : OFF
>
Setting Heating Parameters
Whenever the RH value seen by the sensor exceeds the predefined RH
limit the humidity sensor is warmed up to the predefined temperature.
The duration of the heating can also be defined.
After the heating cycle the humidity conditions are checked and a new
sensor heating is activated if the predefined conditions are reached again.
Syntax: XHEAT<cr>
Type the values after the question mark. The ranges available:
Be careful when changing the filter, since it is easy to break the sensor
when the filter is removed.
If the transmitter is installed in a pressurized chamber, always make sure
that the pressure of the chamber is equalized with the ambient pressure
prior to removing the transmitter. When the transmitter is removed for
maintenance, cap the hole with a capp
be pressurized although the transmitter is not in place.
Plugs are available from Vaisala for ISO threaded connections.
CHAPTER 5
MAINTENANCE
This chapter provides information that is needed in basic maintenance of
the HMT310.
Periodic Maintenance
Calibration Interval
HMT310 is fully calibrated and adjusted as shipped from factory. Typical
calibration interval is one year. In certain applications it may be
advisable to do more frequent checks. Calibration needs to be done
always when there is reason to believe that the device is not within the
accuracy specifications.
You can carry out calibration and adjustment by yourself, or you can
send the transmitter to Vaisala Service Centers for recalibration. For
more information on calibration, see section Calibration and Adjustment
on page 69.
Replacing Consumables
Changing the Filter
Replace the filter if it is contaminated. New filters are available from
Vaisala; see section Spare Parts and Accessories on page 93.
When replacing the sensor, the new sensor must be of the same type as
the old sensor (for example, HUMICAP180R). The sensor type ca
be changed at a Vaisala Service Center. Information on Vaisala sensors
is available online at
When replacing the filter, wear clean gloves to avoid depositing dirt or
oil on the filter.
Replace the filter as follows:
1. Turn the filter counterclockwise until it is loose.
2. Pull the filter straight out carefully. Do not damage the sensor.
3. Take the new filter, and insert it to the filter thread.
4. Tighten the new filter to 5 Nm by turning it clockwise.
Changing the Sensor
NOTE
n only
www.vaisala.com/sensorinfo.
Changing HUMICAP180 and HUMICAP180R humidity sensors is
simple. If the transmitter has the chemical purge and/or warmed probe
option which utilizes the HUMICAP180C or HUMICAP180RC sensor,
the change is more difficult as soldering is required to attach the pins of
the integrated temperature sensor element. Vaisala recommends that you
leave the replacement of HUMICAP180C and HUMICAP180RC sensors
to Vaisala Service Centers, but spare parts are available if you want do it
yourself. Soldering instructions are not provided.
Changing the sensor should be considered corrective maintenance, and it
is not necessary in normal operation. If the accuracy of the transmitter
does not seem to be within specification, it is more likely that the
transmitter is in need of calibration and adjustment, and not sensor
replacement. Refer to chapter Maintenance on page 65.
1. Remove the filter from the probe. See the instructions in section
Changing the Filter on page 65.
2. Remove the damaged sensor and insert a new one. Handle the new
sensor by the plastic socket. DO NOT TOUCH THE SENSOR
PLATE.
3. Perform a calibration and adjustment as instructed in section
Relative Humidity Calibration after Sensor Change on page 71.
4. Attach a new filter on the probe. When using the stainless steel
filter, take care to tighten the filter properly (recommended force
5 Nm).
Chapter 6 ___________________________________________________ Calibration and Adjustment
If you think the HMT310 is not measuring correctly, calibration and
adjustment is not the first th
-
-
-
For
book from Vaisala at the following address:
www.vais
CHAPTER 6
CALIBRATION AND ADJUSTMENT
This chapter describes the relative humidity and temperature adjustment
procedures. Note that after adjustment, the original calibration certificate
shipped with the product is not valid anymore.
Vaisala recommends using the calibration services offered by Vaisala.
Vaisala Service Centers perform calibrations and adjustments as well as
repair and spare part services, and offer accredited calibrations and
maintenance contracts.
Calibration services can also be ordered online at http://store.vaisala.com.
CAUTION
Make sure nothing is interfering with the measurement: heat sources.
temperature differences, or condensation
Check that there is no moisture on the probe. If the sensor has become
wet, you must allow it to dry before you can measure.
Always wait for the measurement to stabilize.
NOTE
an introduction to calibration, order or download the free calibration
Chapter 6 ___________________________________________________ Calibration and Adjustment
Relative Humidity Calibration after
Sensor Change
Syntax: FCRH<cr>
The transmitter asks and measures relative humidity and calculates the
calibration coefficients. This two-point adjustment needs to be performed
after a sensor change. Follow the more detailed calibration instructions
on page 73, but instead of the CRH command, use the FCRH command.
Example:
>fcrh
RH : 1.82 1. ref ? 0
Press any key when ready <cr>
RH : 74.22 2. ref ? 75
OK
>
The OK indicates that the calibration has succeeded.
Set Calibration Information Text
Syntax: CTEXT<cr>
1. Remove the transmitter unit from the mounting plate (see
Mounting the Transmitter/Removing the Transmitter Unit on page
22) and press the adjustment button once (see Figure 18 on page
74).
2. Issue the LI command and set the offset value to 0 and gain value
to 1.
3. Issue command CTEXT and enter calibration information text.
1. Remove the transmitter unit from the mounting plate (see
Mounting the Transmitter/Removing the Transmitter Unit on page
22) and press the adjustment button once (see Figure 18 on page
74).
2. Issue the LI command and set the offset value to 0 and gain value
to 1.
3. Issue the CDATE command and set the calibration date.
Example:
>cdate 2001 12 11
Calibration : 2001-12-11
>
Analog Output Calibration
Syntax: ACAL<cr>
Connect the HMT310 to a multimeter in order to measure either current
or voltage depending on the selected output type. Issue the ACAL
command. Then type the multimeter readings and press Enter.
Chapter 6 ___________________________________________________ Calibration and Adjustment
where
quantity
=
Quantity to be adjusted. The only available quantity is
RH.
function
=
Desired function. The options are:
ON
OFF
LIST
INIT
CLEAR
table and
disable multipoint correction.
Multipoint Relative Humidity Adjustment
Syntax: MPC [quantity] [function]<cr>
– Enable multipoint correction.
– Disable multipoint correction.
– Show correction table.
– Start entering values in correction table.
– Remove all values from correction
Multipoint calibration (applicable from software version 2.0.1 onward)
can be used to optimize RH measurement in a narrow range of
measurement by using 3-8 adjustment points. For more information and
an example multipoint calibration, see section Multipoint Adjustment
Using Serial Line on page 78.
Relative Humidity Calibration and Adjustment
(in Two Points)
Use two reference humidities over the measurement range. The
references need to have a difference of at least 50 %RH.
Before calibration HMT310 needs to be set to adjustment mode by
pressing the adjustment button once, see Figure 18 on page 74. To return
to normal operation mode, use the RESET command: see Reset
Transmitter on page 53.
With a warmed probe (HMT317) or sensor heating, heating is interrupted
when the adjustment button is pressed. Allow enough time for the probe
to reach ambient temperature and then start the calibration procedure.
Low End Adjustment
1. Remove the transmitter unit from the mounting plate (see
Mounting the Transmitter/Removing the Transmitter Unit on page
22), and press the adjustment button once (see Figure 18 above).
2. Remove the filter from the probe (see instructions on page 65) and
insert the probe into the dry end reference chamber (for example,
LiCl: 11 % RH in the humidity calibrator HMK15, use adapter
fitting 13.5 mm for HMT314, HMT315, HMT317, and HMT318).
3. Wait at least 30 minutes for the sensor to stabilize.
4. Issue the CRH command and press Enter.
5. Type C and press Enter a few times to check if the reading is
stabilized.
6. When the reading is stabilized, type the reference humidity after
the question mark and press Enter.
>crh
RH : 11.25 Ref1 ? c
RH : 11.25 Ref1 ? c
RH : 11.25 Ref1 ? c
RH : 11.24 Ref1 ? c
RH : 11.24 Ref1 ? 11.3
Press any key when ready ...
Now the device is waiting for the high-end reference.
Chapter 6 ___________________________________________________ Calibration and Adjustment
High End Adjustment
7. After the low end adjustment, insert the probe into the
measurement hole of the high end reference chamber (for example,
NaCl: 75 % RH chamber in the humidity calibrator HMK15, use
the adapter fitting 13.5 mm for HMT314, HMT315, HMT317, and
HMT318). Note that the difference between the two humidity
references must be at least 50% RH.
8. Let the probe stabilize for at least 30 minutes. You can follow the
stabilization by pressing any key, typing C and pressing Enter.
9. When stabilized, type the high-end reference value after the
question mark and press Enter.
>crh
RH : 11.25 Ref1 ? c
RH : 11.24 Ref1 ? c
RH : 11.24 Ref1 ? 11.3
Press any key when ready ...
RH : 75.45 Ref2 ? c
RH : 75.57 Ref2 ? c
RH : 75.55 Ref2 ? c
RH : 75.59 Ref2 ? 75.5
OK
>
10. Take the probe out of the reference conditions and replace the
filter. When using the stainless steel filter, take care to tighten the
filter properly, recommended force is 130 Ncm.
11. If needed, record the calibration information (text and date) to the
memory of the transmitter: see Set Calibration Information Text on
page 71 and Set Calibration Date on page 72.
12. Reset the transmitter by issuing the RESET command. The
Temperature Calibration and Adjustment (in
One Point)
1. Remove the transmitter unit from the mounting plate (see
Mounting the Transmitter/Removing the Transmitter Unit on page
22) and press the adjustment button once (see Figure 18 on page
74).
2. Remove the probe filter (see instructions on page 65) and insert the
probe into the reference temperature.
3. Let the sensor stabilize.
4. Issue the CT command and press Enter.
5. Type C and press Enter a few times to check if the reading is
stabilized.
6. When the reading is stabilized, type the Ref1 reference temperature
after the question mark and press Enter three times.
>ct
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? 16.0
Press any key when ready ...
T : 16.06 Ref2 ?
OK
>
7. Take the probe out of the reference conditions and replace the
filter. When using the stainless steel filter, take care to tighten the
filter properly (recommended force 130 Ncm).
8. If needed, record the calibration information (text and date) to the
memory of the transmitter, see Set Calibration Information Text on
page 71 and Set Calibration Date on page 72.
9. Reset the transmitter by issuing the RESET command. The
transmitter returns to normal mode.
Chapter 6 ___________________________________________________ Calibration and Adjustment
Humidity Calibration and Adjustment (in One
Point)
1. Remove the transmitter unit from the mounting plate (see
Mounting the Transmitter/Removing the Transmitter Unit on page
22) and press the adjustment button once (see Figure 18 on page
74).
2. Remove the probe filter (see instructions on page 65) and insert the
probe into the reference humidity (for example, use a humidity
generator such as the HMK 15 Humidity Calibrator or salt
solutions).
3. Let the sensor stabilize.
4. Issue the CRH command and press Enter.
5. Type C and press Enter a few times to check if the reading is
stabilized.
6. When the reading is stabilized, type the reference humidity after
the question mark and press Enter three times (to set the first RH
reference, acknowledge the “Press any key when ready …”
message, and leave the second reference point unset).
>crh
RH : 26.00 Ref1 ? c
RH : 26.00 Ref1 ? c
RH : 26.00 Ref1 ? c
RH : 26.00 Ref1 ? c
RH : 26.00 Ref1 ? c
RH : 26.00 Ref1 ? 33
Press any key when ready ...
RH : 26.00 Ref2 ?
OK
>
OK indicates that the calibration has succeeded.
7. Take the probe out of the reference conditions and replace the
filter. When using the stainless steel filter, take care to tighten the
filter properly (recommended force 130 Ncm).
8. If needed, record the calibration information (text and date) to the
memory of the transmitter: see Set Calibration Information Text on
page 71 and and Set Calibration Date on page 72.
9. Reset the transmitter by issuing the RESET command. The
Quantity to be adjusted. The only available quantity is
RH.
function
=
Desired function. The options are:
ON
OFF
LIST
INIT
CLEAR – Remove all values from correction table and
disable multipoint correction.
Multipoint Adjustment Using Serial Line
You can use the MPC command (applicable from software version 2.0.1
onward) to perform a multipoint adjustment to the relative humidity
measurement. Due to the linear response of the HUMICAP® sensor, it is
recommended to use the standard two-point adjustments. Use multipoint
correction only if:
- You need to adjust the relative humidity measurement in 3 ... 8
adjustment points.
- You are interested in a narrow range of measurement and want to
optimize the RH measurement performance for that range.
Multipoint adjustment works by means of a correction table that holds
2 ... 6 adjustment points. The idea is that for every point, you enter the
RH reading that the transmitter shows, and the correct humidity reference
value. You can select any reference points from the range of 0 ... 100 %
RH, but Vaisala recommends that the difference between the lowest and
highest reference point is at least 50 %RH.
NOTE
Multipoint adjustment is applied “on top” of the standard user-made two
point adjustment (made using serial line commands or MI70). It is
recommended that before you start using the MPC command, you clear
the existing humidity adjustment. This is done as part of the example
multipoint adjustment procedure provided here.
Chapter 6 ___________________________________________________ Calibration and Adjustment
Point Number
Transmitter’s Reading
Reference
1
12.2% RH
11% RH
2
34% RH
33% RH
3
75.5% RH
75% RH
4
97.3% RH
97% RH
Example Multipoint Adjustment Procedure
1. Connect the HMT310 to a computer (see section Serial Line and
Analog Communication Options on page 36). Open a terminal
program.
2. Issue the PURGE command to carry out the chemical purge, and
wait for six minutes for it to complete. Transmitters without the
chemical purge feature (selected when ordering) do not respond to
the command.
>purge
Purge started, press any key to abort...
3. Press the adjustment button once (see Figure 18 on page 74).
4. View the currently applied standard user adjustment offset and gain
using the L command. If offset is zero and gain is one (example
shown below), no user correction is applied:
>l
RH offset : 0.00000000
RH gain : 1.00000000
T offset : 0.00000000
T gain : 1.00000000
5. If an RH offset or gain is in place, write down the adjustment and
then clear it by writing zero to RH offset and one to RH gain using
the LI command (see section Revert to Factory Calibration on page
70) :
>li
RH offset : 0.00130000 ? 0
RH gain : 1.04000000 ? 1
T offset : 0.00000000 ?
T gain : 1.00000000 ?
6. You must now prepare a list of 3 ... 8 correction points using a
humidity generator or salt solutions (for example, using the
HMK15 Humidity Calibrator). You must measure the transmitter’s
RH reading at each point, and make a note of the reference RH
reading. Here is an example list for a four-point correction:
To start, remove the filter from the probe and insert the probe into a
measurement hole of the driest reference chamber you intend to
use. Proceed from low to high humidity. Remember to allow at
least 30 minutes for the sensor to stabilize at each point.
7. Once you have completed the list, remove the probe from the last
measurement chamber and re-attach the filter.
8. Using the serial connection command line, issue the MPC RH INIT command and start entering the readings and reference points
from the list you prepared in step 6. When you have no more points
to enter, press enter to save the points.
10. If you are satisfied with the correction list, enable multipoint
correction by issuing the command MPC RH ON.
>mpc rh on
11. Multipoint adjustment is now complete. To exit adjustment mode
and return to the normal mode, issue the RESET command.
The multipoint adjustment can be disabled with the MPC RH OFF
command, and the correction list can be cleared with the MPC RH CLEAR command. Note that using MPC configuration commands
requires pressing the adjustment button. The current status can be viewed
with the MPC RH LIST command also outside of the adjustment mode.
Chapter 6 ___________________________________________________ Calibration and Adjustment
When adjusting both relative humidity and temperature, make the
temperature adjustment first.
Adjustment Using an MI70 Indicator
You can use the MI70 handheld indicator to adjust the following in
HMT310:
- Temperature (T)
- Relative humidity (RH)
- Pressure compensation value for humidity measurement (environment
settings)
Note that starting the calibration in MI70 requires pressing the
adjustment button on the HMT310 board. See Figure 18 on page 74 for
the location of the adjustment button. To return to normal operation
mode, use the RESET command (see section Reset Transmitter on page
53).
NOTE
Starting MI70 Adjustment
1. Connect the HMT310 series transmitter to the MI70 handheld
indicator using the connection cable (optional accessory, Vaisala
order code DRW216050SP).
2. Turn on the MI70.
3. Press the HMT310 adjustment button (see Figure 18 on page 74) to
start the adjustment mode. A notification about starting the
adjustment mode is displayed on the MI70 screen.
4. Press OK to enter the adjustment menu. The available adjustment
options are displayed (RH or T). You can also view the date of last
adjustment.
Use the arrow buttons to highlight RH or T and press SELECT to choose
the parameter for adjustment.
For RH adjustment instructions, see section Adjusting RH with MI70.
For T adjustment instructions, see section Adjusting T with MI70.
Selecting RH in the main adjustment menu brings up a notification about
checking environment settings. Select YES to access the environment
settings before performing the RH adjustment, or NO to continue to RH
adjustment without changing the current environment settings. Select
ADJUST to access the RH adjustment options.
The following RH adjustment options are available:
- 1-point adjustment
- 2-point adjustment
- LiCl – NaCl auto
1-point RH Adjustment
Usually it is recommended to make an adjustment in two reference
humidities. If the adjustment is done by using one reference humidity (1point adjustment), select the reference humidity so that it represents the
measuring environment. Make the adjustment in one point as instructed
in the 2-point RH Adjustment instructions (select 1-point adjustment
instead of 2-point adjustment and follow the display instructions).
2-point RH Adjustment
Select 2-point adjustment in the RH adjustment menu and follow the
steps below:
1. Set the probe to a lower reference relative humidity. You can
follow the stabilization from the GRAPH display. Press READY
when the reading is stabilized.
2. Give the lower reference humidity value by using the arrow
buttons. Press OK.
3. Set the probe to the higher reference humidity. Press READY
when the reading is stabilized. You can follow the stabilization
from the GRAPH display.
4. Give the higher reference relative humidity value by using the
arrow buttons. Press OK.
5. To confirm the adjustment, press YES. To cancel entering the
adjustment, press NO to return to adjustment mode display
without making any changes. Note that if the difference between
the two references is less than 50 %, adjustment cannot be done.
Once confirmed, the adjustment has been carried out. Press BACK to
exit the RH adjustment mode and return to the main adjustment options.
Chapter 6 ___________________________________________________ Calibration and Adjustment
Temperature adjustment in one or two points can be done if there is
reason to believe that the adjustment is changed. If
done only in one point, the reference condition is used to represent the
measuring environment.
Press EXIT to quit the adjustment mode and return to the basic MI70
view.
LiCl – NaCl automatic adjustment
LiCl-NaCl automatic adjustment is a two-point adjustment in reference
humidities of 11.3 % (LiCl) and 75 % (NaCl). You do not have to enter
the reference values, the MI70 displays the accurate value based on the
measured temperature and the Greenspan table stored into the memory of
the MI70.
To make the adjustment, select LiCl – NaCl auto in the RH adjustment
menu and follow the steps below:
1. Set the probe to the LiCl reference humidity. You can follow the
stabilization from the GRAPH display. Press READY when the
reading is stabilized.
2. Set the probe to the NaCl reference humidity. You can follow the
stabilization from the GRAPH display. Press READY when the
reading is stabilized.
NOTE
3. After the NaCl reference humidity measurement is done, MI70
checks if you want to use this 2-point adjustment. Press OK to
confirm the adjustment.
Once confirmed, the adjustment has been carried out. Press BACK to
exit the RH adjustment mode and return to the main adjustment options.
Press EXIT to quit the adjustment mode and return to the basic MI70
view.
Adjusting T with MI70
Selecting T in the main adjustment menu brings up the current T
measurement. Select ADJUST to choose the adjustment type. The
following T adjustment options are available:
Chapter 6 ___________________________________________________ Calibration and Adjustment
Environment Settings
When you select RH adjustment for the first time from the adjustment
main menu, MI70 asks if you want to check the environment settings
(pressure compensation value for humidity measurement). If you need to
access the environment settings after this, exit the adjustment mode by
selecting EXIT in the main adjustment menu and navigate to the
environment menu from the MI70 main menu. To return to the
adjustment mode, press the HMT310 adjustment button.
Last Adjustment Date
To see the last adjustment date, select Last adjustment date in the main
adjustment menu. You can also view this date from the Device
information menu in the MI70 main menu.
Accuracies of the calculated variables depend on the calibration accuracy
of the humidity and temperature sensors; here the accuracies are given
for ± 2 %RH and ± 0.2 °C.
Find the intersection of the dewpoint temperature curve and the dewpoint
difference reading (process temperature-dewpoint temperature) on the xaxis and read the accuracy in dewpoint measurement on the y-axis.
Appendix A ____________________ PROBE INSTALLATION KITS AND INSTALLATION EXAMPLES
The following numbers refer to Figure 30 above:
1 = Duct wall
2 = Flange
3 = Sealing ring
4 = Supporting bar (not included in the kit for HMT315)
5 = Probe attaching part (to be fixed with the supporting bar)
6 = Relative humidity probe
When the temperature difference between the duct and the air outside the
duct is significant, the supporting bar must be installed as deep in the
duct as possible. This helps to prevent errors caused by heat conduction
in the bar and cable.
APPENDIX A
PROBE INSTALLATION KITS AND
INSTALLATION EXAMPLES
This appendix describes the available probe installation kits and provides
some installation examples.
Duct Installation Kits (for HMT313/317/315)
Duct installation kit includes a flange, a sealing ring, a supporting bar
and probe attaching part for the probe and screws for attaching the flange
to the duct wall. Vaisala order codes: 210697 (for HMT313) and 210696
(for HMT315, no supporting bar).