Vaisala HMT310 Technical data

HMT310 Humidity and Temperature Transmitter
www.vaisala.com
The Vaisala HUMICAP
(from left to right): HMT313, HMT317, HMT314, HMT318, HMT315 and HMT311.
Features/Benefits
Next-generation Vaisala
HUMICAP® Sensor for excellent accuracy and stability
Full 0 ... 100 %RH
measurement, temperature range up to +180 °C (depending on model)
Small size, easy to integrate
Insensitive to dust and most
chemicals
NIST traceable
®
Humidity and Temperature Transmitter HMT310 models
Reliable Vaisala HUMICAP® technology
The HMT310 incorporates the latest­generation Vaisala HUMICAP® Sensor. The Vaisala HUMICAP® Sensor is a capacitive thin-lm polymer sensor. It features high accuracy, excellent long-term stability and negligible hysteresis. It is insensitive to dust, particulate dirt and most chemicals.
Several outputs, one connector
The HMT310 is powered up with 12 ... 35 VDC. It has two analog outputs and an RS-232 serial output. The output signal and the supply power travel in the same cable, the only cable connected to the unit.
Chemical purge
Chemical purge helps to maintain measurement accuracy between calibraton intervals and it involves heating the sensor to remove harmful chemicals. The function can be initiated manually or programmed to occur at set intervals.
Optional functions
The following optional functions are available: several probes for various applications, calculated humidity quantities, different mounting kits, sensor protection options and probe cable lengths, warmed probe and sensor heating for high humidity conditions (HMT317), and chemical purge for applications risking an interference with chemicals in the measuring environment.
Technical data
Δ
°C
°C
-0.7
0.7
0.6
0.5
0.4
0.3
0.2
0.1
-0.1
-0.2
-0.3
-0.4
-0.5
-0.6
0
0-60 -40 -20 20 40 60 80 100
120
140
160
180
-80
Measured values
RELATIVE HUMIDITY
Measurement range 0 ... 100 %RH Sensor
Vaisala HUMICAP Vaisala HUMICAP
Accuracy (incl. non-linearity, hysteresis and repeatability) at a temperature range of
+15 ... +25 °C ±1 %RH (0 ... 90 %RH)
-20 ... +40 °C ±(1.0 + 0.008 x reading) %RH
-40 ... +180 °C ±(1.5 + 0.015 x reading) %RH Factory calibration uncertainty (+20 °C)
* Defined as ±2 standard deviation limits. Small variations possible, see also calibration certificate.
Response time (90 %) at +20 °C 17 s with grid filter in 0.1 m/s air flow 50 s with grid and steel, netting filter
TEMPERATURE
HMT311 -40 ... +60 °C (-40 ... +140 °F) HMT313 -40 ... +80 °C (-40 ... +176 °F)
HMT314, HMT315, HMT317, HMT318 -70 ... +180 °C (-94 ... +356 °F)
Accuracy over temperature range (see graph below)
®
180R typical applications
®
180RC applications with chemical
purge and/or warmed probe
±1.7 %RH (90 ... 100 %RH)
±0.6 %RH (0 ... 40 %RH)*
±1.0 %RH (40 ... 97 %RH)*
60 s with sintered filter
or -40 ... +120 °C (-40 ... +248 °F)
Electrical connections
Two analog outputs, selectable and scalable Typical accuracy of analog output at +20 °C Typical temperature dependence of analog output Serial output RS--232C Connections M12 8-pole connector with RS-232C,
current outputs (two channels) and U
Operating voltage 12 ... 35 VDC, the maximum
operating voltage for a device with
Power consumption 30 mA with RS-232 External load R Startup time after powerup 3 s
0 ... 20 mA or 4 ... 20 mA
+0.05 % full scale
0.005 %/°C (0.003 %/°F) of full scale
sensor heating is 24 VDC
< 500 Ohm
l
General
Operating temperature range for electronics -40 ... +60 °C (-40 ... +140 °F)
Storage temperature range -55 .... +80 °C (-67 ... +176 °C)
Operating pressure
HMT314 0 ... 100 bar HMT318 0 ... 40 bar
HMT315, HMT317 vapor tight Transmitter housing material G-AlSi10Mg Transmitter base material ABS/PC Housing classification IP65 Cable feed through alternatives
Sensor protection PPS grid with stainless steel net,
Complies with EMC standard EN61326-1, Industrial environment
8-pole connector with 5 m cable,
Female 8-pin connector screw joint for
cable diameter 4 ... 8 mm
PPS grid, Sintered filter
Membrane stainless steel filter
in
Note: When using the current output, the RF field susceptibility level according to standard EN61000-4-3 with a frequency band of 110 ... 165 MHz, is only 3V/m (generic environment) with the specified accuracy.
Typical temperature
+0.05 °C/°C (+0.005 °F/°F) dependence of electronics Temperature sensor Pt100 IEC751/3 class B
For more information, visit www.vaisala.com or contact us at sales@vaisala.com
HUMICAP® is a registered trademark of Vaisala.
Ref. B210769EN-B ©Vaisala 2009
This material is subject to copyright protection, with all copyrights retained by Vaisala and its individual partners. All rights reserved. Any logos and/or product names are trademarks of Vaisala or its individual partners. The reproduction, transfer, distribution or storage of information contained in this brochure in any form without the prior written consent of Vaisala is strictly prohibited. All specifications — technical included — are subject to change without notice.
Loading...